Practical Instrument Electronics 522 User Manual

Practical Instrument Electronics
Model 522
Thermocouple Calibrator
82 E. Main Street Suite 3.14 Webster, NY 14580
Tel: 585-872-9350 Fax: 585 -872-2638
sales@piecal.com • http://www.piecal.com/
Practical Instrumen t Electro nics, Inc. Copyright 2011. All rights reserved.
Instruc tions for Enabling and Disabling the Configuration Options

Basic Keypad Operations

j EZ-Check™ Switch
SOURCE mode: Slide the switch to select from three user-stored values for the d e sired ca lib ra tion points. The user can select HI, DIAL and LO positio ns. These val u e s ca n e a sil y be changed t o su it the calibration requirements.
READ mode: Slide the switch t o re call minimum and max imum readings. P re ss t he EZ-Dial™ Knob to clear the stored value s.
Operati n g Instruct ions
k SOURCE/OFF/READ Switch
Slide the SOURCE/OFF/READ Switch to SOURCE to output a voltage corresp ond in g to the temperature on the display for the selected thermocouple type. Use the READ position to directly convert the rmocouple input to temp e ra ture .
l EZ-Dial™ Knob
Turn the knob to change temperature in 0.1º increments. Push and turn for faster dialing. Push without turning to store new EZ-Ch eck™ HI/LO points in SIMULA TE mod e , or to cle ar E Z -Check™ HI/LO points in READ mode.

Model 522 Configuration

1. Turn the Model 522 on with the SOURCE/OFF/READ Switch.
2. Press the EZ-Dial Knob while the “PRESS EZ-DIAL KNOB FOR CONFIGURAT ION ” message is d isplayed.
3. Select options by turning the EZ-Dial Knob until the arrow points to the d e sire d option.
4. The option can be enabled or disabled by tapping the EZ-Dial Knob.
The Mode l 52 2 conf ig u rat ion menu will exit automatically aft e r 5 secon d s of ina ctiv ity and go to normal operation with the options selected. These options are saved even when the unit is turned off.
522-9002 Rev D 7/13/11 1-4
Model 522 O pe rating Instructions
82 E. Main Street Suite 3.14 Webster, NY 14580
Tel: 585-872-9350 Fax: 585 -872-2638
sales@piecal.com • http://www.piecal.com/
Practical Instrumen t Electro nics, Inc. Copyr i ght 2011. All rights reserved.
OVERRANGE or UNDERRANGE
The millivoltage input exceeds the selected thermocouple type’s range.

Model 522 Configuration

Double-click the EZ-Dial™ Knob while in source or READ mode to enter the configuration menu. Hold the EZ-Dial™ Knob while turning the unit on to bypass the “PRESS EZ-DIAL KNOB FOR CONFIGURATION” message altogether.
Model 522 C o nfiguration Menu Auto Off ON/OFF
If Auto O f f is ON , th e uni t will tu rn of f a f te r 30 minut e s to save battery life, if there is no user activity. If Au to Off is OFF the unit will stay on until it is turned of f from the keypad . This is ty p ica lly u sef u l f or ma nua l loa d in g or contin uous use.
Display Units °C/°F
Pressing the EZ-Dial™ Knob to toggles between ° C or °F

T/C B, E, J, K, N, R, S, T, or mV

To change T/C type, press the EZ-Di a l™ Kn o b. T urn the EZ-Dia l™ Kn o b to sc roll through the list of av ai lab l e ty p e s. P ress again to save and return to the configuration menu.

READ Mode

Slide the SOURCE/OFF/READ Switch to READ for direct thermocouple in p u t. The Model 522 displa y s te mp e ra ture corresponding to input for the selected thermocouple type.
Connect the T/C sensor. “OPEN T/C” will be displayed until properly connected. Slide the EZ-Check™ Switch to HI and LO to recall maximum and minimum saved readings. Press and hold the EZ-Dial™ Knob to
clear saved read in g s. The d isp la y f la she s “CLEAR ED” as a confirmation. Double-click the E Z -Dial ™ Knob to return to the configuration menu. Turning the EZ-Dial™ Knob has no effect in READ mode. Other Display Indications:
OPEN T/C No thermocouple is connected, or the connected thermocouple exhibits > 10 kΩ.

SOURCE Mode

Slide the SOURCE/OFF/READ Switch to SOURCE for direct thermocoup le o utp u t. The Mo d e l 522 ou tp u ts voltage corre sp onding to temperature for the selected thermocouple type.
Slide the EZ-Ch e ck S wit ch to H I or LO to re call stored setting s. W h ile in th e H I or LO posit ion, dial a new setting and p re ss the EZ­Dial™ Knob to store. The DIAL position always h old s the last setting d ia le d th e re.
Turn the EZ-Dial™ Knob to change temperature, push and turn for faster dialing. Double-click the E Z -Dial ™ Knob to return to the configuration menu.
522-9002 Rev D 2-4
82 E. Main Street Suite 3.14 Webster, NY 14580
Tel: 585-872-9350 Fax: 585 -872-2638
sales@piecal.com • http://www.piecal.com/
Practical Instrumen t Electro nics, Inc. Copyr i ght 2011. All rights reserved.
Overall Size
4.9 X 3.15 X 1.82 inches (125.5 X 80 X 46.2 mm)
Miscellaneous
Low battery indication with nominal 1 hour of op e ra tion le f t
Temperature Coeff icie n t of mV S o urce
50 ppm/°C of output range
General Temperature Accuracy
±(0.008 % of mV setting + 0.006 mV) ± 0.1 °C (0.2 °F)
Resolution
0.1 °C or 0.1 °F
Output Noise
±5 µV pp from 0.1 Hz to 10 Hz
Output Impedance
0.2 (200 nV/uA)
Input Noise
< ±1 LSD from 0.1 Hz to 10 Hz
Two Wire Connection to Transmitter
Model 522 O pe rating Instructions

Connection Diagram

Specifications

General Specifications:

Unless otherwise ind icate d all specifications are rated from a nomin al 23 °C, 70 % RH f or 1 year from calibration. Temperature Range -25 to 60 °C (-10 to 140 °F)
Relative Humidi ty Range 10 % RH 90 % (0 to 35 °C), Non-condensing
10 % RH70 % (35 to 60 °C), Non-condensing
Overall Weigh t (inclu d ing 9V battery) 7.2 oz (204 grams) Battery 9V Alkaline p rov ides 45 hours of continuous use
Overload protecte d t o 60 v ol ts f or 30 se cond s or less High-contrast gra p h ic liq uid cry sta l d isp la y wit h 0.357” (9.07 mm) high digits
Millivolt Accuracy ±(0.008 % of mV Setting + 0.006 mV)
Cold Junction Calibration A ccuracy ±0.1 °C (0.2 °F) Cold Junction Sensor Tempe ra ture Coefficient ±0.025 °/° in ambient temperature (°C or °F)
SOURCE Thermocouple Specifications:
Output Range -13.000 to +80.000 mV
Source Current < 8 mA
READ Therm oc ouple Specifications:
Input Impedance > 1 M
522-9002 Rev D 3-4
Model 522 O pe rating Instructions
82 E. Main Street Suite 3.14 Webster, NY 14580
Tel: 585-872-9350 Fax: 585 -872-2638
sales@piecal.com • http://www.piecal.com/
Practical Instrumen t Electro nics, Inc. Copyr i ght 2011. All rights reserved.
Open T/C Te s t P u lse
< 1 uA for 300 ms
Open T/C Response Time
< 3 seconds
Carrying Case
Part Number: 020-0205
T/C Wire Kits
Part Number: 020-02 02 wire k it 1 includ e s J, T , E, K , lea d s with mini co nn e ctor s
TYPE B
594 to 1820 °C
0
0.2
0.4
0.6
0.8
1
1.2
500
700
900
1100
1300
1500
1700
1900
Temperature °C
Error °C
TYPE E
-260 to 1000 °C
0
0.2
0.4
0.6
0.8
1
1.2
1.4
-400
-200
0
200
400
600
800
1000
1200
Temperature °C
Error °C
TYPE J
-210 to 1200 °C
0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
0.4
-400
-200
0
200
400
600
800
1000
1200
1400
Temperature °C
Error °C
TYPE K
-245 to 1372 °C
0
0.2
0.4
0.6
0.8
1
1.2
-400
-200
0
200
400
600
800
1000
1200
1400
1600
Temperature °C
Error °C
TYPE N
-229 to 1300 °C
0
0.2
0.4
0.6
0.8
1
1.2
-400
-200
0
200
400
600
800
1000
1200
1400
Temperature °C
Error °C
TYPE R
24 to 1768 °C
0
0.2
0.4
0.6
0.8
1
1.2
0
200
400
600
800
1000
1200
1400
1600
1800
2000
Temperature °C
Error °C
TYPE S
21 to 1768 °C
0
0.2
0.4
0.6
0.8
1
1.2
0
200
400
600
800
1000
1200
1400
1600
1800
2000
Temperature °C
Error °C
TYPE T
-251 to 400 °C
0
0.2
0.4
0.6
0.8
1
1.2
-300
-200
-100
0
100
200
300
400
500
Temperature °C
Error °C
Open T/C Th reshold 10 k nominal
Available Options:
Part Number: 020-0203 wire kit 2 includes B, R /S , N le a d s with mini connectors

Temperature Accuracy

The following chart s g iv e worst-case temperature accuracy b ase d on state d millivolt a cc ura cy of ±(0.008 % + 0.006). Temperature is uncompensated on the horizontal axis, referenced to 0 °C. Cold Junction calibration accuracy of 0.1 °C is not included in the temperature error.

Warranty

Our equipment is g ua ran te e d a g ai nst d e f e ctive material and workmanship (e xcluding batteries) f or a p e riod o f thre e years from the date of shipment. Claims under guarantee can be made by returning the equipment prepaid to our factory. The equipment will be repaired, rep lace d or a d juste d a t our op t ion. The lia b ilit y of P ractical Instrument Ele ctronics (PIE) is restri cte d to th at given under our guarantee. No re sp onsib il ity is a ccep te d for damage, loss or oth e r expense incurred throug h sale o r use of our equipment. Under no condition shall Practical Instrument Ele ctronics, Inc. be liab le f or an y sp e cia l, incid e n tal or consequential damage.
522-9002 Rev D 4-4