Powertronix Isolation Station Series Instruction Manual

 
300VA – 3000VA
Instruction Manual
P2 POWERTRONIX Corporation Page1of10
Thank you for your purchase of a POWERTRONIX Power Isolation Unit and congratulations on your
choice. The Isolation Station series of power isolation units is ideal for applications where medical and
dental devices require improved electrical isolation and/or reduced leakage current in order to comply with existing safety standards, and where sensitive electronic systems require protection from electrical power disturbances.
SAFETY SYMBOLS
Warning, dangerous
voltage
Alternating Current/AC
General warning sign
Caution
Protective Earth
(Ground)
Direct Current/DC
Equipotentiality
Earth (Ground)
Refer to Instruction
Manual/booklet
AND
“ON” and OFF (Power)
Indicator lights
GREEN,WHITE
&AMBER
Indicator lights
“Ready for Use”
INSPECTION
Remove the power isolation unit from the shipping container and inspect it for shipping damage. Do not install or operate the product if it appears to be damaged in any way. If the unit is damaged, notify the carrier, and the company that sold you the unit, IMMEDIATELY.
ENVIRONMENTAL ISSUES
 
has been designed for indoor use only, in areas where it will not be exposed to excessive dust or moisture. Make certain there is adequate airflow around the unit. Do not place objects on top of or near the unit, which could obstruct the air vents.
This equipment may generate, use and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. If this equipment does cause harmful interference to other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures. However, there is no guarantee that interference will not occur in an installation.
This power isolation unit has environmental requirements similar to other electronic equipment. It
P2 POWERTRONIX Corporation Page2of10
Only authorized personnel are allowed to operate the power isolation station
- Reorient or relocate the equipment.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit different from that to which the other device (s) are connected.
- Consult the manufacturer or field service technician for help.
 
WARNINGS
WARNING:
WARNING:
WARNING:
WARNING:
WARNING: TO PREVENT ELECTRICAL SHOCK, DO NOT REMOVE THE POWER
ISOLATION UNIT COVER.
To reduce the risk of electrical shock, do not open enclosure. There are no serviceable parts inside. Refer service to POWERTRONIX Corporation or its
authorized agents only.
This product requires convection cooling. Adequate ventilation is required. When installing this product, a minimum clearance of 2.0" must be provided to all sides with air ventilation holes, except the mounting surface.
Risk of explosion exists if used in the presence of flammable anesthetics, or other flammable gases or liquids.
Grounding continuity should be checked periodically.
NOTE: USA and Canada, grounding reliability can only be achieved when this equipment is connected to a matching receptacle marked "Hospital Only", or "Hospital Grade".
The top and bottom surfaces of the power supply enclosure are not intended to be
WARNING:
touched while the power supply is in use. While the unit is in use, the top and bottom surfaces of the power supply enclosure may reach temperatures which cause discomfort when touched for more than 1 second. There is no user- accessible parts on the top and bottom surface of the power supply. Use only the front-panel and rear­panel switches, nobs and connectors – which are intended to be used while the power supply is in operation. [See the “Installation” and “Operation” sections of this
instruction manual for additional information.]
WARNING:
If the power isolation station provided with an equipotential or protective earth stud, where it connects to a potential equalization or protective earth conductor then the resulting medical electrical system should meet all applicable IEC60601-1 requirements.
P2 POWERTRONIX Corporation Page3of10