Alarm Specifications and Warranty ............................................................................................. 33
2
www.posey.com 1.800.447.6739
Sitter On Cue
BEFORE YOU BEGIN
Before You Begin
The Posey Sitter On Cue is an easy to use, restraint-free addition to your falls management
program. The Sitter On Cue provides an early warning when a patient attempts an unassisted exit
from a sensor. This system does not prevent falls or injury from falls and is not a substitute for
patient care, rounding and a comprehensive falls management protocol in your facility.
The Sitter On Cue activates as the patient attempts to rise and removes pressure from the sensor.
The Sitter On Cue may be used with an optional cable to connect to the nurse call system.
Indications for Use:
Persons who may benefit from the use of the Posey Sitter On Cue include:
•
Patients with diminished cognitive or mobility skills.
•
Patients receiving medications that may cause disorientation, drowsiness, dizziness, or
f requent urination
•
Patients (new or existing) with a history of falls, or who are assessed to be at risk of falling based
on your selected fall-risk assessment
•
Patients who are restless or prone to get up in the middle of the night, e.g., due to incontinence,
or with nighttime voiding habits
•
Patients who require mandatory bed rest
Contraindications:
NOTE: The Sitter On Cue may not be suitable for all high fall-risk patients. See Posey
catalog for other options for such patients.
The Sitter On Cue should NEVER be used as the only means of surveillance for:
•
Agitated, combative or suicidal patients
•
Patients at extreme risk of a life-threatening fall
Response Policy:
Make sure your facility has a clearly defined falls management policy
•
Check alarm function every time before leaving patient unattended.
If the alarm and/or sensor do not function properly, remove the alarm and sensor from service and
replace them with a properly functioning alarm and/or sensor. DO NOT use the alarm or sensor if
it does not activate each time weight is removed from the sensor or the chair belt sensor is unfastened.
Sitter On Cue
www.posey.com 1.800.447.6739
3
FE ATU R E S
Features of the Sitter On Cue
FRONT VIEW
REAR VIEW
(OPEN BATTERY DOOR)
RIGHT SIDE VIEWLEFT SIDE VIEW
4
www.posey.com 1.800.447.6739
Sitter On Cue
FE ATU R E S
Features of the Sitter On Cue (Cont.)
•
Seven alarm tones. There are seven different alarm tone options which can be used in
different patient rooms, or to discern between different caregivers or levels of fall risk for
easy patient identification.
•
Four alarm modes. Four alarm modes allow for facility choice regarding caregiver notification
in the event the alarm is triggered. The alarm modes are “Voice and Tone”, “Tone Only”, “Voice
Only” and “Mute.”
•
Three alarm volumes. Three alarm volume settings allow for facility choice regarding
caregiver notification in the event the alarm is triggered. The volume options are low, medium
and high.
•
Custom Recording. Allows a family member, friend or caregiver to record a personal message
that will sound if pressure is removed from the sensor pad or chair belt sensor is unfastened. The
recording will play only if the alarm mode is set to “Voice” or “Voice and Tone”. This feature can
provide patient communication in their language, or by a caregiver or family member for familiarity.
•
HOLD Button. Allows you to remove a patient from a sensor pad without the need to turn off the
alarm (e.g. for therapy, toileting or socializing). Alarm will resume monitoring when patient returns
to the sensor pad.
•
Optional alarm delay of 0, 1, or 2 seconds. Allows you to adjust alarm timing for each patient.
The delay allows you to determine, by alarm, the number of seconds between the time pressure is
removed from the sensor, chair belt sensor is unfastened, and the time the alarm activates.
•
Dual Sensor monitoring. This alarm connects to any Posey sensor pad, (chair, toilet, commode,
stretcher, over-mattress and chair belt sensors). Alarm activates when weight is removed from
sensor pad, or chair belt sensor is unfastened. Two sensors can be connected to the alarm at the
same time.
•
Nurse Call Interface. Provides alarm notice at patient’s room and nursing station. If connected
to nurse call, the Sitter On Cue has the ability to be muted, silencing the alarm at the patient’s
bedside. It will still alarm at nurse call system notification points.
•
Failsafe sensor alarm. The Posey Sitter On Cue activates if the alarm is on and a sensor cord is
removed from the alarm.
•
Battery operated. The alarm utilizes four (4) “AA” alkaline batteries.
•
Visual and Audible low battery warning. Low battery LED will blink red when batteries are
low and an audible cue will say “low battery” when batteries are critically low, alerting caregivers
of the need to change batteries.
•
DC Jack. For AC connection.
•
AC Adapter. The Sitter On Cue 8645 adapter connection enables users to operate on AC
electrical power.
•
Impact resistant cover. Helps minimize damage if dropped.
another manufacturer’s sensors may damage the Posey alarm, cause the fall monitoring
system not to function as intended, and will void the factory warranty.
NEVER connect other manufacturers’ sensors to a Posey alarm. Use of
Sitter On Cue
www.posey.com 1.800.447.6739
5
CONTENTS AND BRACKET OPTIONS
Each Sitter On Cue is
shipped to you with:
•
Alarm (1)
•
“AA” Alkaline batteries (4)
•
8276 Bracket and Clip
•
Instructions for Use
NOTE: If you require the ability to run
on AC power, order 8645AC.
See the Posey catalog for additional
accessories sold separately, including a
variety of brackets and nurse call cable sets.
6
www.posey.com 1.800.447.6739
Sitter On Cue
PREPARING
Preparing the Sitter On Cue for Use
Battery Installation:
The battery-operated Sitter On Cue is portable and long lasting.
Fresh alkaline batteries have an estimated life of 30 days of daily use. Actual life depends on alarm mode, tone and volume
you select.
1. Slide the power switch to the OFF position before inserting/
changing batteries (Fig. 1).
2. Press down on the arrow and slide the battery compartment
door completely off (Fig. 2). Set battery door aside.
3. Insert four (4) new “AA” alkaline batteries as pictured inside
the battery compartment (Fig. 3). Take care not to damage
battery contacts.
4. Reattach the battery compartment door. Slide it shut, locking
it into place.
Alarm Low Battery LED will blink red when batteries are low
and an audible cue will say “low battery” every 15 seconds
when batteries are critically low. Change batteries at once.
•
Take care when installing new batteries. The alarm will
not work if batteries are installed improperly.
•
ALWAYS install a completely new set of batteries
when the Low Battery LED is blinking red. DO NOT
replace a single cell, but all cells in the alarm. DO NOT
mix old and new batteries, or battery brands within
a battery pack (4 batteries). Use of mixed batteries,
battery leakage, or batteries installed incorrectly may cause battery damage, and may
damage the alarm. Remove any alarm from use and send to the appropriate facility
authority if batteries are damaged or corroded or the battery compartment has signs
of previous battery corrosion such as white powder residue.
•
Batteries can explode or leak and cause damage to alarm if installed incorrectly, fully
discharged, or exposed to liquid, fire or high temperatures. If battery damage has
occurred, or you see any corrosion, remove the alarm from use IMMEDIATELY. DO NOT
use the alarm if battery damage has been detected.
•
After changing batteries, test the alarm and sensor for proper operation prior to putting
in service with a patient, and each time before leaving the patient unattended. If the
alarm and/or sensor do not function properly, remove the alarm and sensor from service
and replace them with a properly functioning alarm and/or sensor.
Fig. 1
Fig. 2
Fig. 3
Sitter On Cue
www.posey.com 1.800.447.6739
7
STATUS L I GHT K E Y
The Status LED will flash different colors providing a visual indication of the current state of the
alarm.
FLASHALARM STATEDESCRIPTION
Green
Active Monitoring
Sensor is connected and pressure is being
applied to the pad
Yellow
Rapid Yellow
Red
Rapid Red
Ready
Sensor Activation
Hold
ALARM
The sensor will activate and begin monitoring
as soon as pressure is applied to a connected
sensor
When pressure is applied to a sensor, the alarm
will rapidly flash yellow for 3 seconds before the
sensor is fully activated.
For 30 seconds after the hold button is pressed,
monitoring is suspended. The alarm is not actively monitoring and the alarm will not activate
if pressure is removed or applied to from the
sensor.
The patient has removed pressure from the pad
or another event has occurred that requires a
nurse’s attention
8
www.posey.com 1.800.447.6739
Sitter On Cue
SETTING MODE
Setting Alarm Mode
The Sitter On Cue has four (4) alarm modes. These allow you to select an alarm notification best
suited to patient and facility needs.
Mode Features
VOICE &
TONE
TONE
VOICE
ONLY
MUTE
Depending on whether a custom voice recording is made, either the factory default
voice or a custom voice recorded message plays once, followed by the selected
alarm tone. The tone continues to play until alarm is placed on hold or patient is
repositioned with the sensor.
This is the factory default mode. Selected alarm tone plays until alarm is placed on
hold or patient is repositioned with the sensor.
This mode requires the nurse call cable to be plugged into your alarm and the
appropriate wall jack from your nurse call system. Custom voice or factory default
voice message plays repeatedly until alarm is placed on hold or patient is repositioned with the sensor. This mode also activates the facility nurse call system.
Automatically switches to VOICE & TONE if a nurse call cable is removed from the
alarm.
This mode requires the nurse call cable to be plugged into your alarm and the
appropriate wall jack from your nurse call system. THERE IS NO AUDIBLE VOICE
MESSAGE OR ALARM TONE AT THE PATIENT’S BEDSIDE. This mode activates the
facility nurse call system. Automatically switches to VOICE & TONE if the nurse call
cable is removed from the alarm.
to both the alarm and the wall jack before leaving the patient unattended. Verify that an
alert is received at the nursing station if the cable is unplugged from the wall jack.
NOTE: There will be no alert at the nursing station or at the bedside if the nurse call cable is
unplugged from the alarm.
When using a nurse call cable, ensure the nurse call cable is plugged in
To Change/Select Mode
NOTE: “Voice Only” and “Mute” can be set only when
the nurse call cable is plugged into your alarm. With this
feature, all mode options are available. Be sure to plug the
nurse call cable into wall and test. YOU MUST PLUG THE
NURSE CALL CABLE INTO THE ALARM AND THE WALL
JACK IN ORDER TO ACTIVATE NURSING STATION ALERTS.
Follow these steps to change or select alarm mode:
1. Lift the programming control protective cover (Fig. 4).
2. Ensure alarm is on
3. Press MODE button on left side of alarm to scroll
through the mode selections (Fig. 5). The last mode
heard is the mode utilized when the alarm is activated.
Fig. 4Fig. 5
Sitter On Cue
www.posey.com 1.800.447.6739
9
SETTING MODE
Setting Alarm Mode
4. When “Voice Only” or “Mute” are selected, you must have the nurse call cable plugged into
your alarm and the appropriate wall jack for your nurse call system. Ensure that the nurse call
system indicators (nursing station and wall light) are activated when the cable is unplugged
from the wall jack.
5. Replace the programming control protective cover, securing it into place.
To test the modes requiring nurse call interface, remove pressure from the sensor or unfasten
the chair belt sensor. Check at the nurse’s station to ensure that the indicator for the appropriate
patient alarms on the nurse call system and the wall light, indicating there is no pressure on the
sensor or chair belt sensor is unfastened.
(Continued)
10
www.posey.com 1.800.447.6739
Sitter On Cue
SET TING TONE
Setting Alarm Tone
The Sitter On Cue has seven (7) available tones. This allows you to differentiate between patients
and other equipment alarms.
To Select Tone:
1. Lift the programming control protective
cover (Fig. 7).
2. Ensure alarm is on
3. Alarm mode must be set to either
“Voice and Tone” or “Tone” (see pages
9-10) in order to hear the selection.
4. Press TONE button on left side of alarm
to scroll through the tone selections
(Fig. 8).
5. Each time you press the button, tone
will change and a sample plays. The
last sample heard is the tone utilized
when the alarm is activated.
6. Replace the programming control
protective cover, securing it into place.
Fig. 8
Sitter On Cue
www.posey.com 1.800.447.6739
11
ADJUSTING VOLUME
Adjusting Alarm Volume
There are three (3) alarm volume settings: low, medium and high. These allow you to select the
right volume for facility and/or patient needs.
All volume settings are within OSHA standards. For maximum staff alert in noisy areas, use
HIGHEST (loudest) volume setting.
distance before leaving patient unattended.
cause hearing loss or other injury. For more information, see: OSHA OCCUPATIONAL
NOISE EXPOSURE STANDARDS 1910.95.
To Change Volume:
1. Lift the programming control protective cover (Fig. 10).
2. Ensure alarm is on.
3. Alarm mode must be set to either “Voice and Tone” or “Tone” (see pages 9-10) in order to hear
the selection.
4. Press VOLUME button on left side of alarm to scroll through the volume selections (Fig. 11).
5. Each time you press the button, the volume changes and a two (2) second sample plays.
6. Continue to press the VOLUME button until you hear the desired volume. The last sample heard
is the volume in use.
7. Replace the programming control protective cover, securing it into place.
ALWAYS check to ensure staff can hear alarm at the furthest possible
NEVER place alarm closer than two feet from patient’s ear. Doing so may
12
Fig. 11
www.posey.com 1.800.447.6739
Sitter On Cue
TIME DELAY
Time Delay
The delay is the time that will elapse AFTER weight is removed from sensor pad or chair belt sensor
is unfastened, and BEFORE alarm activates. Set a delay, if any, to meet the needs of each patient.
The Sitter On Cue gives you the option of a 0, 1, or 2 second alarm delay. If set to 1 or 2 seconds,
the alarm will provide an audible cue at both power on and sensor activation
NOTE: With a time delay set, the alarm will not activate if weight is removed from the sensor pad,
chair belt sensor is unfastened, before the time delay that you set expires.
For example, if you set a 2 second time delay, the alarm will not activate until after 2 seconds
have expired.
The delay should be long enough to allow some patient movement without setting off alarm, but still
give sufficient warning when patient attempts to get off of a sensor pad. This is especially useful
when connected to the nurse call system to minimize false alarms.
the delay at zero (0) with patients at EXTREME risk of injury from a fall associated with
an unassisted bed, chair or toilet exit.
To Set Time Delay
1. Lift the programming control protective cover (Fig. 12).
2. Slide the toggle switch to the desired delay setting, 0-1-2 (Fig. 13). An audible cue will confirm
the current delay setting.
3. Replace the programming control protective cover, securing it into place.
Assess patient frequently to ensure that a time delay is appropriate. Set
Sitter On Cue
Fig. 13
www.posey.com 1.800.447.6739
13
RECORDING A MESSAGE
To Record a Message
The Posey Sitter On Cue has a feature that allows a caregiver or family member to communicate
a verbal “warning” message about unassisted sensor pad exits to a patient without having to be
physically in the room with the patient.
1. Lift the programming control protective cover (Fig. 14).
2. Ensure alarm is on.
3. Ensure alarm mode is set to “Voice and Tone” or “Voice” (nurse
call option only) to hear recording.
4. Press and hold the RECORD button on left side of alarm (Fig. 15).
5. Wait for the “begin record” audible cue” before speaking.
Continue to hold RECORD button down until you are done
speaking.
6. In a normal voice, speak into the microphone aperture labeled
MIC on front, top left corner of alarm
(Fig. 16).
NOTE: Changing alarm volume will not change message recording
volume. Your distance from the MIC and how loud you speak
controls message volume. To increase volume, re-record message
in a louder voice, with your mouth closer to MIC.
7. Recording will stop when you release the RECORD button or time
(15 seconds) expires. If you exceed the time allowed, recording
will stop and the alarm will give an audible cue “end record”.
8. Replace the programming control protective cover, securing
it into place.
9. After recording message, remove weight from sensor, or
unfasten chair belt sensor to play message back. Ensure
alarm mode is set to “Voice and Tone” or “Voice” (nurse
call option only) to hear recording.
10. Check that the message is clear and volume is right for
your patient. Re-record if necessary.
11. If the RECORD button is pushed twice rapidly a 3 second
count down will beging to reset to the factory default
message. If you do not wish to reset, press RECORD again
during the countdown. Factory default message will play.
Microphone MIC
Fig. 15
14
www.posey.com 1.800.447.6739
Fig. 16
Sitter On Cue
THE HOLD BUTTON
The HOLD Button (Sensor Use ONLY)
NOTE: HOLD feature will work only when the alarm is actively monitoring or alarming.
To Place the Alarm on HOLD:
1. Press HOLD button on the top of alarm (Fig. 17) The
STATUS LED will begin flashing RED indicating the alarm is
in HOLD mode (Fig. 17a).
2. You have 30 seconds to assist patient into or out of bed
or chair before alarm returns to monitoring mode. When
assisting patient out of bed, move patient towards the edge
of the bed with his or her legs over the side of the bed before
pressing the HOLD button. This will allow you more time to
reposition the patient on the sensor without activating the
alarm after the 30 second HOLD period expires. Note: While the alarm is in a hold state and the STATUS
LED is flashing red, the alarm will not activate if pressure is
removed from the sensor pad.
3. If during the 30 seconds when the alarm is on hold, you wish to resume active monitoring, press
the hold button again to cancel the hold. Verify the STATUS LED is flashing green and the alarm
and sensors are actively monitoring before leaving the patient unattended.
4. After 30 seconds:
•
If weight is present on the sensor, or chair belt sensor is connected you will hear a single
“beep” and monitoring will begin. ALWAYS verify the STATUS LED is flashing green, and the
alarm and sensors are monitoring before leaving patient unattended.
•
If there is no weight on sensor, or chair belt sensor is not connected, alarm will be in Ready
state and status LED will flash yellow.
•
The Ready state allows patient to be away from bed or chair for extended periods without
alarm activating (e.g., for meals, therapy, toileting etc.). When patient returns and weight
is applied to sensor or chair belt sensor is connected, alarm will “beep” once to indicate
monitoring has resumed. The STATUS LED will flash green..
Test the alarm and sensor for proper operation prior to putting in service
with a patient, and each time before leaving the patient unattended. If the alarm and/
or sensor do not function properly, remove the alarm and sensor from service and
replace them with a properly functioning alarm and/or sensor. Ensure the STATUS LED is
blinking green, indicating the alarm is monitoring.
Fig. 17
Sitter On Cue
www.posey.com 1.800.447.6739
15
THE HOLD BUTTON
To Place the Alarm on EXTENDED HOLD:
1. Press HOLD button on the top of alarm (Fig. 18) for 3 seconds. The STATUS LED will begin
flashing RED indicating the alarm is in HOLD mode and an audio cue will say “EXTENDED
HOLD” (Fig. 18a).
2. You have 5 minutes to assist patient into or out of bed or chair before alarm returns to
monitoring mode. You can cancel the HOLD as described in the previous section.
3. After 5 minutes the alarm will either resume active monitoring or go to the Ready state, as
describe above for the 30 second hold.
with a patient, and each time before leaving the patient unattended. If the alarm and/
or sensor do not function properly, remove the alarm and sensor from service and
replace them with a properly functioning alarm and/or sensor. Ensure the STATUS LED is
blinking green, indicating the alarm is monitoring.
Test the alarm and sensor for proper operation prior to putting in service
16
www.posey.com 1.800.447.6739
Sitter On Cue
CONNECTING TO AC POWER
Connecting to AC Power (Optional)
The Posey Sitter On Cue is equipped with a DC jack to allow you to use AC power to operate the
alarm when the 8645AC is ordered (includes AC power adapter). To ensure proper alarm function
and to prevent damage to the alarm, only use the AC power adapter supplied by Posey.
To Connect to AC Power:
1. Connect the AC power adapter to the Sitter On Cue (Fig. 19).
2. Plug the other end into the wall.
3. Connect a sensor to the alarm.
adapter other than the Posey adapter may damage the alarm and void the warranty.
If power is interrupted, the alarm will not function. There is no notification at bedside or
nurse call station.
Only use the Posey AC power adapter to power your alarm. Using an
AC power
adapter
input
Fig. 19
Note: If you are using the Posey AC Power Adapter to power your alarm and the adapter is
accidentally unplugged from the wall socket, the alarm will not operate - even if there are also
batteries in the alarm. If the adapter is unplugged from AC power adapter input and there are
working batteries in the alarm, the alarm will then operate on battery power.
Sitter On Cue
www.posey.com 1.800.447.6739
17
NURSE CALL INTERFACE
Nurse Call Interface
You can connect the Sitter On Cue to your facility nurse call system when you connect the optional
Nurse Call Cable. This allows you to add the options and notification associated with your specific
nurse call system such as nursing station notification and patient room lights and sounds. With this
feature, the alarm can be:
•
Silenced in patient’s room to reduce roommate disturbance (“Mute”)
•
Set for “Voice Only”, instead of alarm tone
NOTE: Alarm automatically switches to “Voice and Tone” if nurse
call cable is removed from the alarm. This ensures in-room voice
and tone notice to staff if patient attempts to rise when nurse call
cable is no longer connected to the alarm.
NOTE: The alarm supports both normally open and normally closed
nurse call systems. The manufacturing default is normally open. If
you would like to change to normally closed, use a small pointed
object, for example a pen, to slide the switch down to the closed
position.
To Connect the Nurse Call Cable:
1. Insert one end of cable into “Nurse Call” jack on right side
of alarm (Fig. 20). Use care not to plug into sensor outlet as
damage to sensor outlet could occur.
2. Insert the other end of cable into wall jack of nurse call panel.
Depending on your system, you may need a “Y” nurse call
adapter to use the Posey Sitter On Cue and the patient nurse
call device at the same time.
Posey offers a variety of nurse call adapter cable sets. Please
request document M6253 for a list of all Posey nurse call
adapter systems.
When using a nurse call cable, ensure the nurse call cable is plugged in
to both the alarm and the wall jack before leaving the patient unattended. Verify that an
alert is received at the nursing station if the cable is unplugged from the wall jack.
NOTE: There will be no alert at the nursing station or at the bedside if the nurse call
cable is unplugged from the alarm.
If the nurse call cable is unplugged from the alarm when it is in “Voice Only” or “Mute”
mode, the alarm will default to “Voice and Tone” as a failsafe.
Fig. 20
18
www.posey.com 1.800.447.6739
Sitter On Cue
NURSE CALL INTERFACE
Nurse Call Interface (Continued)
•
DO NOT stretch or strain cable to avoid possible damage and possible malfunction.
•
DO NOT attach cable to moving parts of the bed or chair that will cause strain or
damage if the bed or chair is repositioned.
•
ALWAYS position the cable so that moving parts (side rails, wheels, etc.) will not cause
strain or damage the cable.
•
DO NOT run over the cable with carts or equipment.
•
DO NOT wrap the cable tightly during storage.
•
ALWAYS remove the cable by pulling on the plug. DO NOT pull on the cable.
•
ALWAYS secure the cable out of the way so it will not be a tripping hazard.
•
ALWAYS test alarm and nurse call function prior to leaving the patient unattended.
Activate the alarm (remove pressure from sensor or unfasten chair belt sensor) and
make sure the nurse call light for the proper bed and room activate in the hall and at
the nurse’s station.
•
DO NOT use the alarm or sensor if it does not activate each time weight is removed
from the sensor or the chair belt sensor is unfastened. Replace alarm and sensor with
working units and retest before leaving the patient unattended.
FOR SAFE USE WITH NURSE CALL CABLE:
Sitter On Cue
www.posey.com 1.800.447.6739
19
MONITORING
Monitoring with a Sensor
The following instructions will help you set up and safely use the Sitter On Cue with a sensor.
TIPS TO PROTECT SENSORS FROM DAMAGE
To avoid inconvenience to staff and patients, and to protect
sensors from damage, you should follow these steps:
•
Only use Posey sensors with the Posey alarm.
•
When routing sensor cord to alarm, check that there is no
stress on cord. Cord must be clear of all moving parts of bed,
stretcher or chair to prevent sensor failure.
•
NEVER jerk or pull on the cord to remove RJ11 plug. Doing so will damage cord wires or plug,
or alarm sensor outlet.
•
ALWAYS use the plastic tab to release plug (Fig. 21).
FAILSAFE FEATURE
The Posey Sitter On Cue contains a “failsafe” feature that activates the Posey alarm if
the sensor is removed from the alarm when actively monitoring. To silence the alarm,
reconnect the sensor to the alarm.
Fig. 21 - Plastic tab
20
www.posey.com 1.800.447.6739
Sitter On Cue
MONITORING
Monitoring with a Sensor (Continued)
FOR SAFE USE OF ALL SENSOR TYPES:
To reduce the risk of serious injury or death, ALWAYS follow these steps after putting the
sensor in place and before leaving patient unattended (see instructions below). DO NOT
use any alarm or sensor that does not alarm each time it is tested.
1. Make sure alarm is ON
2. Check that the RJ11 plug on the sensor cable is not damaged (plug broken, or wires
disconnected) and is securely connected to the alarm. An audio cue will play indicating the
sensor was connected and the STATUS LED will be yellow.
3. Apply pressure to the sensor. An audible cue will play indicating the sensor is activated.
4. Disconnecting the sensor from the alarm while it is actively monitoring will cause the alarm to
activate. This is called a “failsafe” mode. While applying pressue to the sensor pad, disconnect the
sensor to make sure the failsafe mode works. DO NOT use the alarm if the alarm does not sound
when the sensor is disconnected.
5. When connecting the alarm to the nurse call system, check that the nurse call cable is securely
connected to the alarm and the nurse call panel. ALWAYS test alarm and nurse call function if nurse
call cable is plugged into the alarm and wall jack. Activate the alarm (remove pressure from sensor,
unfasten chair belt sensor, and make sure the nurse call light for the proper bed and room activate
in the appropriate nurse’s station location). Remove the cable from the wall jack and make sure the
visual or audible alert at the nurse’s station immediately activates.
6. Inspect sensor cord and nurse call cable (if in use) to ensure they are out of the footpath and DO
NOT pose a tripping hazard.
Sitter On Cue
www.posey.com 1.800.447.6739
21
MONITORING, STEPS TO APPLY
Monitoring with Two Sensors
Two sensors can be connected to the alarm allowing the patient to be transferred from
one location to another without moving sensor pads or switching out cords.
Transferring a Patient between sensors
1. Connect the second sensor to the alarm.
2. Pess HOLD button to pause monitoring for 30 seconds. You have 30 seconds to remove the
patient from the current sensor pad. During this time the STATUS LED will flash red.
3. Place the patient on the second sensor. At the end of the 30 seconds, the second sensor will
activate when pressure is applied. If the patient is seated on the second pad sooner, the HOLD
button can be pressed again to resume monitoring.
4. If more time is desired, activate the extended hold feature by pressing and the hold button for 3
seconds. An audio cue will indicate the extended hold has been activated. You have 5 minutes
to transfer the patient from the first sensor the second sensor before monitoring begins.
NOTE: While two sensors can be connected to the alarm, only one can be used for monitoring at a
time. If a first sensor is actively monitoring and pressure is applied to the second sensor, an audible
cue will indicate a second sensor is about to be activated and gives 10 seconds for the person to
remove pressure from the sensor. If pressure is not removed, the alarm will sound. Press the HOLD
button. You have 30 seconds to remove pressure from at least one of the sensor pads.
22
www.posey.com 1.800.447.6739
Sitter On Cue
USE OF PHYSICAL RESTRAINTS
The Sitter On Cue and Use of Physical Restraints
If the patient’s care plan calls for the use of a restraint, staff should read and follow all instructions
and warnings for the device you choose.
The position of the sensor pad is vital when using a restraint. Make sure the restraint is applied correctly per instructions for that device. Straps must NOT cross over sensor pad.
For bed use, sensor pad should be placed at shoulder blade level so alarm will activate if patient
sits up, tries to climb over side rails, or scoots to bottom of bed.
prevent alarm from activating. If patient falls out of bed or chair and is suspended in the
restraint, serious injury or death may occur from chest compression or suffocation (Fig. 36).
•
Bed safety: Refer to the Food and Drug Administration (FDA) for the most recent Hospital
Bed Safety Guidelines as well as the Bed Manufacture for their Instructions for Use.
•
Full compliant bed side rails must be UP when restraints are used on a patient.
To reduce the risk of entrapment, use side rail covers, especially with split side rails.
A failure to do so may result in serious injury or death if patient’s body goes under,
around, through or between the bed side rails.
Use extreme caution with chair cushions. If a cushion dislodges, straps may loosen and
allow patient to slide off seat and become suspended.
If straps cross over sensor pad and patient moves, pressure from straps may
*http://www.fda.gov search keyword “HBSW”
Sitter On Cue
www.posey.com 1.800.447.6739
23
WARNINGS AND CAUTIONS
Warnings and Cautions
•
NEVER connect a Posey alarm to other manufacturers’ sensors.
•
NEVER connect a Posey sensor to other manufacturers’ alarm.
•
Make sure it is safe to drill and there are no pipes or electrical wires that could be damaged
when using screws to attach the wall mount bracket.
•
ALWAYS check to ensure staff can hear alarm at the furthest possible distance before leaving
patient unattended.
•
Check that there is no stress on the nurse call cable. Make sure cable is clear of all moving parts
of bed or chair, and does not pose a tripping hazard.
•
Check that both ends of cable are securely plugged in and the nurse call system has an alert
warning if the cable is disconnected from the wall jack.
•
Test alarm and nurse call functions by activating alarm and removing pressure from the sensor
pad, or unfastening chair belt sensor EACH TIME before leaving patient unattended.
•
Assess patient frequently to ensure that a time delay is appropriate. Set the delay at zero (0) with
patients at EXTREME risk of injury from a fall associated with an unassisted bed, chair or toilet exit.
•
Before each use, check that:
– Alarm is securely mounted out of the patient’s reach and functions properly by activating alarm.
– Indicator lights are in clear view of staff.
•
DO NOT mix old and new batteries or battery brands. This may cause rupture or leakage and
damage alarm.
•
DO NOT allow batteries to deplete while in the alarm. Change batteries immediately when seeing
the low battery LED blink red. Depleted batteries may leak and corrode, causing damage. When
storing the alarm for a short period with power “on” to maintain custom voice messages and
settings, check the alarm every week to make sure the batteries are still operable and the alarm is
still on. If the alarm low battery LED is blinking red or the alarm does not power up, the batteries
are depleted and must be removed. DO NOT leave depleted (“dead”) batteries in the alarm to
avoid corrosion.
•
Remove batteries when storing the alarm for an extended period to prevent depleting the
batteries and potential corrosion.
•
The Posey Sitter On Cue is an electronic device. It may fail to work if subjected to severe shock,
such as being dropped, or immersed in liquid. To reduce the risk of serious injury or death,
the alarm and sensor for proper operation prior to putting in service with a patient, and
each time before leaving the patient unattended. If the alarm and/or sensor do not function
properly, remove the alarm and sensor from service and replace them with a properly
functioning alarm and/or sensor
. DO NOT use the alarm or sensor if it does not activate each
time weight is removed from the sensor or the chair belt sensor is unfastened.
test
24
FOR SAFE USE WITH ALL SENSORS:
www.posey.com 1.800.447.6739
Sitter On Cue
WARNINGS AND CAUTIONS
Warnings and Cautions (Continued)
To reduce the risk of serious injury or death, ALWAYS follow these steps after putting the sensor in
place and before leaving patient unattended. DO NOT use any alarm or sensor that does not alarm
each time it is tested.
1. Make sure the alarm is ON
2. Check that the RJ11 plug on the sensor cable is not damaged (plug broken, or wires
disconnected) and is securely connected to the alarm.
3. Disconnecting the sensor from the alarm will cause the alarm to activate. This is called a
“failsafe” mode. Disconnect the sensor to make sure the failsafe mode works. DO NOT use the
alarm if the alarm does not sound when the sensor is disconnected.
4. When connecting the alarm to the nurse call system, check that the nurse call cable is securely
connected to the alarm and the nurse call panel. ALWAYS test alarm and nurse call function
if nurse call cable is plugged into the alarm and wall jack. Activate the alarm (remove pressure
from sensor, unfasten chair belt sensor) and make sure the nurse call light for the proper bed
and room activate in the appropriate nurse’s station location. Remove the cable from the wall
jack and make sure the visual or audible alert at the nurse’s station immediately activates.
5. Inspect sensor cord and nurse call cable (if in use) to ensure they are out of the footpath and DO
NOT pose a tripping hazard.
Sensor Not Functioning
If the alarm and/or sensor do not function properly, remove the alarm and sensor from service and
replace them with a properly functioning alarm and/or sensor. DO NOT use the alarm or sensor if it
does not activate each time weight is removed from the sensor, the chair belt sensor is unfastened.
Adapter Cable Replacement
Contact Posey Customer Service for nurse call cable adapters available for various nurse call systems. Posey offers 21 varieties of nurse call adapter cable sets. Please request document M6253 for
a list of all nurse call adapter systems.
Sitter On Cue
www.posey.com 1.800.447.6739
25
OFF
ON
PULL LEVER
SLIDE OUT
DELAY
MODE
TONE
VOLUME
RECORD
VOICE
&
TONE
TONE
MIC
VOICE
MUTE
HOLD
&
SUSPEND
HOLD/SUSPEND
NURSE CALL
PULL LEVER
SLIDE OUT
PRESS
ON
OFF
MOUNTING
Mounting the Sitter On Cue
There are multiple mounting options to choose from for the Sitter On Cue: The Sitter On Cue ships
with the 8276 bracket listed below.
If you plan to use the Sitter On Cue for stretcher sensor monitoring or commode sensor monitoring,
please refer to the Instruction Sheets for these products, 8316 and 8333.
To purchase additional brackets contact your Posey sales representative, or call Posey Customer
Service at 1.800.447.6739.
•
Alarm is securely mounted out of the patient’s reach and functions properly by activating alarm.
•
Alarm indicator lights are in clear view of staff.
Bed Mounting with the 8276 Wire Bracket
The 8276 Wire Bracket fits head and footboards that are ½" – 2" (1 cm - 5 cm) thick (Fig. 38).
PRIOR TO USING ANY BRACKET, YOU SHOULD CHECK THAT:
Fig. 38 - Wire Bracket
26
RELEASE
LEVER
SLIDE ALARM
DOWN
Fig. 39
www.posey.com 1.800.447.6739
8276 Wire Bracket
SLIDE ALARM
OUT
Fig. 40
Sitter On Cue
PRESS
RELEASE
LEVER
MOUNTING
PRESS LEVER
SLIDE OUT
PRESS LEVER
SLIDE OUT
OFF
ON
PULL LEVER
SLIDE OUT
PRESS LEVER
SLIDE OUT
DELAY
MODE
TONE
VOLUME
RECORD
VOICE
&
TONE
TONE
MIC
VOICE
MUTE
HOLD
&
SUSPEND
HOLD/SUSPEND
NURSE CALL
TO MOUNT THE 8276 WIRE BRACKET:
1. Slide alarm onto bracket from top down until it is firmly in place (Fig. 39).
2. Choose location on head or footboard where patient cannot reach or tamper with the alarm or
connections.
3. Pull bracket wire away from alarm to create an opening wide enough to fit the headboard or
footboard. Slide bracket onto bed and push down to ensure a snug fit. Make sure alarm indicator
lights are in clear view of staff.
4. To remove alarm, gently push release lever IN while sliding alarm up and out (Fig. 40).
5. Make sure sensor, nurse call cables, and/or AC adapter cables can be secured out of the way
and do not present a tripping hazard.
Sitter On Cue
SLIDE
ALARM
DOWN
INDICATOR
LIGHTS
www.posey.com 1.800.447.6739
SLIDE ALARM
OUT
PRESS
RELEASE
LEVER
SCREW
HOLES
RELEASE
LEVER
27
MOUNTING
Chair/Wheel Chair Mounting
WITH THE 8276 WIRE BRACKET
1. Choose a location on rear of chair, out of the patient’s reach.
2. Slide alarm onto bracket from top down until it is firmly in place (Fig. 39 page 26).
3. Choose location on back of chair back where patient cannot reach or tamper with the alarm or
connections.
4. Pull bracket wire away from alarm to create an opening wide enough to fit the chair back
5. Slide bracket onto chair and push down to ensure a snug fit. Make sure indicator lights are in
clear view of staff.
6. To remove alarm, gently push release lever IN while sliding alarm up and out (Fig. 40 page 26).
7. Make sure sensor and/or nurse call cables can be secured out of the way and do not present a
tripping hazard.
28
www.posey.com 1.800.447.6739
Sitter On Cue
CLEANING, STORAGE, MAINTENANCE
Alarm Cleaning, Storage and Battery Maintenance
Cleaning: Sensor, Cables and Alarm Housing (exterior ONLY)
Dampen (but DO NOT soak) a clean cloth with disinfectant. Wipe alarm clean, using care not to get
disinfectant inside speaker grill and connector ports.
To reduce the risk of damage, NEVER:
•
use any cleaning substance that contains Phenol
•
immerse in liquid
•
sterilize with heat
Always use a clean, DRY cloth to dry all parts.
Storage
•
This device is designed for use in normal indoor environments.
•
This device may be stored in ambient warehouse temperatures at normal humidity levels (10 to 50%).
Avoid excess moisture or high humidity that may damage product materials (greater than 90%).
•
Store pad sensors flat or hang in a dry secure environment. DO NOT fold or roll sensors, as it
may damage internal electronic parts and cause a malfunction.
Disposal
Dispose of per facility policy. Be sure to follow all laws that apply.
Battery Compartment
the alarm from use and notify the appropriate facility authority. The alarm should be
disposed of according to your facility disposal requirements. DO NOT use the alarm and
DO NOT attempt to clean it if there are any signs of battery leakage such as corrosion,
rust or white powder residue.
immediately when seeing the low battery LED blink red. Depleted batteries may leak and
corrode, causing damage to the electronics and reliability. If the alarm low battery LED
is blinking red or the alarm does not power up, the batteries are depleted and must be
removed. DO NOT leave depleted (“dead”) batteries in the alarm to avoid corrosion.
Remove batteries when storing the alarm for an extended period to prevent depleting
the batteries and potential corrosion.
Battery Leakage. If there is ANY evidence of battery leakage, remove
DO NOT allow batteries to deplete while in the alarm. Change batteries
Sitter On Cue
www.posey.com 1.800.447.6739
29
TROUBLESHOOTING GUIDE
Troubleshooting Guide
PROBLEM:Continuous alarm with patient in bed or chair.
SOLUTION:Chair Sensor Pad
•
Check that sensor pad cord and RJ11 plug are clean and undamaged. Check plug connection
to alarm.
•
Check sensor pad for creases or damage to vinyl cover.
•
Check “neck” of chair sensor pad for signs that urine or other liquids have leaked into pad.
•
Check that sensor pad is directly under patient’s weight.
– Pad should be towards front of chair seat if patient normally sits toward front.
– Pad should be towards back of chair seat if posture support is in use or if patient is at risk of
forward sliding.
•
Check seating/positioning aids such as wheelchair cushions or wedge cushions. Weight from
these may activate alarm, or prevent sensor from activating.
•
Check sensor pad expiration date. A continuous alarm may indicate sensor pad is “worn-out” and
should be replaced.
SOLUTION: Two Sensors Connected
•
Check that pressure is not being applied to two sensor pads simultaneously. Only one sensor
pad can be actively monitoring at a time.
•
SOLUTION:Over-Mattress Sensor Pad
•
Check that sensor cord and RJ11 plug are clean and undamaged. Check plug connection to alarm.
•
Check sensor pad for creases or damage to vinyl cover.
•
Check “neck” of over mattress sensor pad for signs that urine or other liquids have leaked into pad.
•
Patient may not be heavy enough to activate sensor.
– Shoulder Placement: Adjust sensor pad so it is centered at shoulder blade area and patient
makes contact with pad.
– Try a different sensor pad location. Most patient weight is normally under buttocks.
– Buttocks Placement: Check that sensor pad is directly under patient’s weight. Shoulder
placement may be needed for a very small individual or restless sleeper.
•
A foam pad on top of mattress may diffuse patient’s weight so sensor pad does not activate.
– Reposition over-mattress sensor pad above foam pad.
•
Mattress may not bend easily when head or knee sections are raised or lowered. Some
mattresses are very stiff and may form an air pocket between mattress and frame when bed is
adjusted. This may prevent weight from touching sensor pad. Try a different sensor pad location.
•
Check sensor pad expiration date. A continuous alarm may indicate sensor pad is “worn-out” and
should be replaced.
SOLUTION:Chair Belt Sensor
•
Check that sensor cord and RJ11 plug are clean and undamaged. Check plug connection to alarm.
•
Check that buckle is securely fastened and there are no loose wires.
SOLUTION:Exit Alarm Mat
•
Check that there is no weight on sensor.
•
Check that sensor cord and RJ11 plug are clean and undamaged. Check plug connection to alarm.
30
www.posey.com 1.800.447.6739
Sitter On Cue
TROUBLESHOOTING GUIDE
Troubleshooting Guide (Continued)
PROBLEM:Intermittent Alarm while the patient is in a bed or chair.
SOLUTION:
•
Check that sensor cord and RJ11 plug are clean and undamaged. Check plug connection to alarm.
•
Check sensor pad for creases or damage to vinyl cover.
•
Check that sensor pad is directly under patient’s weight.
•
Is the sensor getting caught in “hammocking” wheelchair seat? If so, place a foundation cushion
on seat, under sensor.
•
Make sure sensor cord is not folded back under pad.
•
Make sure sensor pad air intake (“neck” of over mattress or chair sensor pad) is clear and not
blocked. Air must flow freely in and out of sensor.
•
Try a new sensor if intermittent alarm can not be fixed.
•
Make sure mattress continues to make contact with the sensor and will activate the alarm when
pressure is removed, even if the head or foot of the bed is articulated.
•
Apply pressure to sensor in several areas to check that alarm activates.
•
Ensure batteries are not corroded.
PROBLEM:No Alarm when patient exits bed or chair.
SOLUTION:
•
Make sure alarm is ON (STATUS LED flashing green or flashing red.)
•
Check batteries. If needed, insert four (4) new “AA” alkaline batteries. DO NOT mix old and new
batteries, or different brands of batteries.
SOLUTION: Voice and Tone or Voice Only with custom recording
•
Press record twice to reset to factory default or press hold and record to record a new message.
SOLUTION: Two Sensors Connected
•
Check that there is not a second sensor activated that is currently actively monitoring. Only one
sensor pad can be actively monitoring at a time.
SOLUTION: Chair Sensor Pad
•
Make sure sensor pad cord is not folded back under pad.
•
Make sure sensor pad air intake (“neck” of chair sensor pad) is clear and not blocked. Air must
flow freely in and out of the sensor pad.
•
Check that there is no weight in the chair such as a box, bag or book.
•
Check seating/positioning aides. A heavy wheelchair cushion may prevent alarm from alarming.
Try a different position for the sensor pad, such as on top of the cushion.
•
Is the sensor pad getting caught in “hammocking” wheelchair seat? If so, place a foundation
cushion on seat, under sensor pad.
•
Try a new sensor pad if alarm does not sound.
SOLUTION:Over-Mattress Sensor Pads
•
Check that all connections are tight and properly plugged into the alarm.
•
Check the “DELAY” setting.
•
Check that there is no weight on the mattress such as a box, bag or book.
•
Check that the correct side of the bed sensor pad is “UP” under the mattress.
•
When the patient lies down they may not be making contact with the sensor to activate monitoring.
Try a different position for the sensor pad. Most patient weight is normally under buttocks.
Sitter On Cue
www.posey.com 1.800.447.6739
31
TROUBLESHOOTING GUIDE
Troubleshooting Guide (Continued)
SOLUTION:Chair Belt Sensor
•
Check that all connections are tight and properly plugged into the alarm.
•
Check the “DELAY” setting.
PROBLEM:Alarm volume is too low or too loud.
SOLUTION:
•
Press the VOLUME button on the side of alarm to change volume setting. Continue tapping the
button to scroll through the selections. The last sample heard is the volume in use.
PROBLEM:Cannot access the “Voice Only” or “Mute” mode settings.
SOLUTION:
•
The “Voice Only” and “Mute” modes are available ONLY while nurse call interface is in use.
Check that nurse call cable is properly connected to alarm and nurse call panel jacks.
PROBLEM:In “Voice and Tone” or “Voice Only” modes, custom voice message
does not play.
SOLUTION:
•
Press and hold the record button to record a new message. Hold the microphone close to your
mouth and speak loudly and clearly. To reset to the factory default recording, rapidly press the
record button twice.
PROBLEM:Sensors slide around when the head of the bed is raised or lowered
on beds with foam overlays.
SOLUTION:
•
Anchor the sensor(s) on top of the mattress, under the foam overlay with the Posey Grip and
clips provided.
PROBLEM:In-room alarm activates, but nurse call station does not activate.
SOLUTION:
•
Check that all connections are tight and the nurse call cable is connected to the alarm and
properly plugged into the facility’s nurse call system. The connections should snap tightly together.
•
Check for worn or damaged wires.
•
Verify use of proper adapter for the system.
PROBLEM:
Factory self-test failed audible cue plays when alarm is powered on
SOLUTION:
•
Power the alarm off. Wait 60 seconds and then power the alarm back on.
32
www.posey.com 1.800.447.6739
Sitter On Cue
Alarm Specifications:
SPECIFICATIONS, WARRANTY, REPAIR
Size
Weight
Power Supply
Battery Life
Expectancy
Current Drain
Alarm Maximum
Volume
Voltage Range
Low Battery
Warning
3.6”W x 6”L x 2.2”D (9 cm x 15 cm x 5.6 cm)
10.2 oz (13.7 oz with batteries), 0.29 kgs. (0.39 kgs with batteries)
Four (4) “AA” alkaline batteries. Optional AC power adapter (8645AC) 9 VDC.
Approximately 30 days of daily use; may vary
Non-alarmed monitoring mode 0.45 mA, maximum alarm volume 132 mA
100 ± 5dB
4.6-6.3 VDC
Low battery LED blinks red and audible cue will say “Low battery” when
batteries need changing
Two Year Warranty
The Posey Company is committed to manufacturing the best quality products. \Posey warrants to
the original purchaser for a period of two (2) years after date of purchase (the “Warranty Period”) that
the Posey Sitter On Cue is defect-free in materials and workmanship. If the product is found to be
defective in workmanship or materials, the Posey Company will replace it without charge within the
Warranty Period. This warranty does not cover accidental damage, water immersion, improper care,
alteration or misuse, and excludes claims for loss or theft. Service under this warranty is available
by contacting Posey Customer Service (1.800.447.6739) for a return authorization, and by sending
the product in clean condition, freight pre-paid, with dated proof of purchase to the Posey Company.
This warranty gives you specific legal rights, and you may also have other rights, which vary from
state to state.
For additional information or questions, call the Posey Company at 1.800.447.6739.
Sitter On Cue
www.posey.com 1.800.447.6739
33
Sitter On Cue
Instruction Manual
®
Posey Products, LLC (Posey) • 5635 Peck Road, Arcadia, CA 91006-0020 USA
Phone: 1.800.447.6739 • Fax: 1.800.767.3933 • www.posey.com