Ponsel ACTEON 5000 User Manual

1
Version 2.1 Update: April 2016
ACTEON 5000
Digital multi-parameter transmitter
User manual
2
Version 2.1 Update: April 2016
CONTENTS
1. GENERAL ..............................................................................................................4
1.1 Safety instructions .................................................................................................................................... 4
1.2 Labeling .................................................................................................................................................... 4
2. DESCRIPTION OF THE EQUIPMENT ...................................................................5
2.1 ACTEON 5000 transmitter. ........................................................................................................................... 5
2.1.1 General description. ................................................................................................................................... 5
2.1.2 Technical characteristics. ........................................................................................................................... 5
2.2 Digital sensors. ................................................................................................................................................. 6
2.2.1 OPTOD sensor: dissolved oxygen (optical technology). ........................................................................... 6
2.2.2 PHEHT sensor: pH and temperature. ......................................................................................................... 7
2.2.3 EHAN sensor: Redox and temperature. ..................................................................................................... 8
2.2.4 NTU sensor: Turbidity in NTU-mg/l. ........................................................................................................ 9
2.2.5 C4E sensor: 4-electrode conductivity. ...................................................................................................... 10
2.2.6 CTZN sensor: inductive conductivity. ..................................................................................................... 11
2.2.7 VB5 sensor: Sludge Blanket detection. .................................................................................................... 12
2.2.8 MES5 sensor: Sludge Blanket detection, Suspended Solid, Turbidity. .................................................... 13
3. INSTALLATION ................................................................................................... 14
3.1 Description and mounting of the controller. ............................................................................................... 14
3.1.1 Description of the front face. .................................................................................................................... 14
3.1.2 Equipment required. ................................................................................................................................. 14
3.1.3 Outline drawings of the ACTEON 5000. ................................................................................................. 15
3.2 Electrical connections. .................................................................................................................................. 16
3.2.1 Safety instructions – Installation. ............................................................................................................. 16
3.2.2 Description of the wiring.......................................................................................................................... 17
4. USER INTERFACE .............................................................................................. 20
4.1 Home screen. .................................................................................................................................................. 20
4.2 Navigation icons. ........................................................................................................................................... 21
5. SETTING UP ........................................................................................................ 22
5.1 Initial start-up. .............................................................................................................................................. 22
5.2 Installation of digital sensors. ....................................................................................................................... 23
5.3 Selection of the parameters. ......................................................................................................................... 23
6. PROGRAMMING ................................................................................................. 25
6.1 Main menu. .................................................................................................................................................... 25
6.2 Description of the menus. ............................................................................................................................. 25
6.2.1 Configuring the digital inputs................................................................................................................... 25
6.2.2 Configuring the On/Off inputs. ................................................................................................................ 32
6.2.3 Configuring the Analog inputs. ................................................................................................................ 33
6.2.4 Configuring the ACTEON 5000. ............................................................................................................. 37
6.2.5 Configuring the Analog outputs. .............................................................................................................. 38
6.2.6 Configuring the relay outputs. .................................................................................................................. 41
6.2.7 Configuring the Recording output. ........................................................................................................... 44
7. MAINTENANCE ................................................................................................... 45
7.1 Maintaining the transmitter: ........................................................................................................................ 45
7.2 Maintenance of digital sensors: .................................................................................................................... 46
3
Version 2.1 Update: April 2016
7.2.1 OPTOD sensor. ........................................................................................................................................ 46
7.2.2 NTU sensor: Turbidity. ............................................................................................................................ 48
7.2.3 PHEHT sensor: pH/Temperature. ............................................................................................................ 50
7.2.4 ORP sensor: REDOX/Temperature.......................................................................................................... 51
7.2.5 C4E sensor: 4-electrode conductivity. ...................................................................................................... 52
7.2.6 CTZN sensor: Inductive conductivity. ..................................................................................................... 53
7.2.7 VB5 – MES 5 sensor: Sludge blanket detection – Sludge blanket detection/Suspended Solid/Turbidity.54
8. TROUBLESHOOTING ......................................................................................... 55
4
Version 2.1 Update: April 2016
1. General
1.1 Safety instructions
In order to maintain and ensure the good working order of the device, users must comply with the safety precautions and warnings featured in this manual.
Assembly and activation:
- Assembly, electrical connection, activation, operation and maintenance of the measuring system must only be carried out by specialist personnel authorized by the user of the facilities.
- Trained personnel must be familiar with and comply with the instructions in this manual.
- Make sure the power supply complies with the specifications on the nameplate before connecting the device.
- A clearly-labeled power switch must be installed near the device.
- Check all connections before turning the power on.
- Do not attempt to use damaged equipment: it may represent a hazard and should be labeled as faulty.
- Repairs must only be carried out by the manufacturer or by Ponsel's after-sales service department.
1.2 Labeling
Prior to any installation or start-up operation, check all the labels and symbols affixed to the measurement device.
This symbol indicates there is a risk of electric shock or electrocution associated with the use of the device.
This symbol indicates that the measuring device cannot be disposed of as conventional waste.
Label on outside of device:
The ACTEON 5000 label on the right-hand face indicates the required power supply and the device's serial number.
5
Version 2.1 Update: April 2016
2. Description of the equipment
2.1 ACTEON 5000 transmitter.
2.1.1 General description.
The ACTEON 5000 digital transmitter can be connected to two digital sensors in the PONSEL MESURE range to monitor the following parameters: pH, redox, temperature, dissolved oxygen (using optical technology), conductivity, salinity, turbidity (NTU, mg /L), Suspended Solid, Sludge blanket detection….
The values measured are displayed and transmitted using analog or digital technology. The preconfigured regulation functions can be used to optimize the control of processes.
The ACTEON 5000 is used in combination with a wide range of interference-resistant digital sensors, offering pre-amplification features built into the sensor and digital signal processing. All the data regarding the calibration, history, users and measurements is processed directly within the sensor, thus delivering very high levels of traceability and enhancing the reliability of the measurements.
2.1.2 Technical characteristics.
Software and functionalities
Digital sensor input
2 RS-485 digital sensor inputs
2 analog outputs
Choice of 2 programmable parameters depending on the sensor connected Programming in mode PI, PID
2 relay/digital outputs
Can be set to NO/NC Setpoint: the measurement range (hysteresis/direction) and activation time can both be selected,
Control of the external cleaning system Equipment sensor fault alarm output
Atmospheric pressure sensor
For oxygen pressure compensation
Technical characteristics of the transmitter
Display
Backlit LCD graphic touch screen – Size: 95x54 mm
Analog outputs
0/4.00 – 20.00 mA with galvanic isolation Max. load 250
Relay outputs
6 A /250 V
Operating conditions
Range of operating temperatures: -15 °C to 50 °C Storage/shipping temperature -15 °C to 50 °C
Power supply/Electrical protection
100-240 V ac/dc 50-60 Hz
- Option: 9-28 V dc/dc
- Electrical protection: complies with EN 61010-1: 2010
Casing
Dimensions (WxHxD)
213 x 185 x 84 mm
Weight
950 g
Material
Grey ABS
Ingress protection rating
IP 65
Front face
Non-reflective polyester
6
Version 2.1 Update: April 2016
2.2 Digital sensors.
The digital sensors in the PONSEL range are equipped with galvanic isolation and can perform digital signal processing to optimize the reliability of the measurements and data sent to the ACTEON 5000 terminal. All the data regarding the calibration, calibration history, users and measurements is processed directly within the sensor and transmitted via a Modbus RS-485 link.
The range of digital sensors can be used to measure a variety of parameters: temperature, dissolved oxygen, pH, redox potential, conductivity (4-electrode or inductive measurement principle), turbidity and suspended solids.
2.2.1 OPTOD sensor: dissolved oxygen (optical technology). The OPTOD dissolved oxygen sensor applies the luminescence-based optical measurement
technology and measures reliably and accurately without requiring calibration. With no consumables or maintenance required, the OPTOD sensor gives an immediate return on the investment. The only intervention required is to replace the DO disk every two years. Since it does not consume oxygen, the OPTOD sensor can be used in all media; even when there is a very weak flow of water. The body is made of passivated 316 L stainless steel or Titanium for applications in corrosive media.
Measurements
Measurement principle
Luminescence-based optical measurement
Measurement ranges
0.00 to 20.00 mg/L
0.00 to 20.00 ppm 0-200%
Resolution
0.01
Accuracy
+/- 0.1 mg/L +/- 0.1 ppm +/- 1 %
Response time
90% of the value in less than 60 seconds
Recommended measurement frequency
> 5 s
Water movement
No circulation required
Temperature compensation
Via an NTC thermistor
Storage temperature
- 10 °C to + 60 °C
Temperature measurement range
0 °C to 50 °C
Accuracy
+ /- 0.5 °C
Signal interface
Modbus RS-485 (or SDI-12)
Power supply for sensor
5 to 12 volts
Power consumption
Standby: 25 µA When sending via RS-485 (1 measurement/second): 4.4 mA When sending via SDI-12 (1 measurement/second): 7.3 mA Current pulse: 100 mA Warm-up time: 100 mS
Sensor
Dimensions
Diameter: 25 mm; Length not including cable: 146 mm
Weight
Stainless steel version: 450 g (sensor + 3 m of cable) Titanium version: 300 g (sensor + 3 m of cable)
Material in contact with the medium
Passivated 316L stainless steel. New: Titanium body
Maximum pressure
5 bar
Cable
9-wire shielded conductor, uncoated-strand polyurethane sleeve
Ingress protection rating
IP68
7
Version 2.1 Update: April 2016
2.2.2 PHEHT sensor: pH and temperature. This PONSEL sensor is fitted with an Ag/AgCl reference electrode, used to measure pH and redox, in
a "PLASTOGEL"® KCl-saturated plasticized electrolyte. The Plastogel® electrolyte is in direct contact with the external environment without interposition of capillary or porous material. There is therefore no risk of fouling or deactivating the reference electrode. The electrode used to measure the pH is a pH-sensitive glass bulb (made from special glass) welded to the end of a crystal tube, and the electrode for Redox measurements is a platinum disk.
Temperature: measured by an NTC thermistor inserted in a stainless steel sheath.
pH measurement
Measurement principle (pH)
pH/reference combined electrode: special glass, Ag/AgCI reference. Gel (KCI) electrolyte
Measurement range
0 – 14 pH
Resolution
0.01 pH
Accuracy
+/- 0.1 pH
Measurement of the Redox
Measurement principle (Redox)
Redox/reference combined electrode: Platinum disk, Ag/AgCI reference. Gel (KCI) electrolyte
Measurement range
- 1000.0 to + 1000.0 mV
Resolution
0.1 mV
Accuracy
2 mV
Response time
< 5 s
Temperature measurement
Measurement principle (T°C)
NTC thermistor
Operating temperature
0.00 °C to + 50.00 °C
Resolution
0.01 °C
Accuracy
0.5 °C
Response time
< 5 s
Storage temperature
0 °C to + 60 °C
Ingress protection rating
IP 68
Signal interface
Modbus RS-485 as standard and SDI-12 as an option
Measurement refresh rate
< 1 second maximum
Power supply for sensor
5 to 12 volts
Power consumption
Standby: 25 µA When sending via RS-485 (1 measurement/second):
3.9 mA When sending via SDI-12 (1 measurement/second):
6.8 mA Current pulse: 500 mA
Sensor
Dimensions of fitted sensor
Lower part: 21 mm in diameter; 92 mm long, Upper part: 27 mm in diameter; 103 mm long, Length of fitted sensor: without cable gland 210 mm; Length with cable gland: 260 mm.
Weight
350 g (sensor + cable)
Material in contact with the medium
PVC, POM-C, special pH glass, platinum, polyurethane
Maximum pressure
5 bar
Cable/connection hardware
9-wire shielded conductor, uncoated-strand polyurethane sleeve
8
Version 2.1 Update: April 2016
2.2.3 EHAN sensor: Redox and temperature. The PONSEL sensor incorporates a reference electrode, used for ORP measurements, such as Ag /
AgCl in saturated KCl electrolyte plasticized "PLASTOGEL" ®. The electrolyte "PLASTOGEL" ® communicates directly with the external environment without
interposition of capillary or porous. So there is no risk of clogging or defusing the reference. The measuring electrode is in platinum (3,5mm2) presented in sealed ring on a glass rod and is for in-
situ measurements continuously
Temperature: measures via CTN.
ORP Measure
Principle of ORP measure
Combination Electrode (ORP/reference) platinium ring, Reference Ag/AgCI. Gelled electrolyte (KCI)
Range of measures
- 1000.0 to + 1000.0 mV
Resolution
+/- 0.1 mV
Precision
+/- 10 mV
Answer time
< 90 s
Temperature measurement
Principle of measure T°C
CTN
Temperature
0,00 °C to + 50.00°C
Resolution
0.01 °C
Precision
+/- 0.5 °C
T90
< 300 s
Temperature of storage
0°C to + 60°C
Protection scale
IP 68
Interface signal
Modbus RS-485 standard and SDI-12 in option
Refresh rate measurement
Maximum < 1 second
Sensor power
5 to 12 volts
Consumption
Standby : 25 µA Average RS485 (1 measure/second) : 20 mA Pulse current : 500 mA Heating time : 100 mS
Sensor
Dimensions of sensor mounted
Mounted sensor length: gland (262 mm) not included ; Length with gland : 324 mm.
Weight
350 g (sensor + cable)
Material in contact with the environment
PVC, POM-C,platinum, Polyurethane
Maximum Pression
5 bars
Cable/ connection
9-wire shielded conductor, uncoated-strand polyurethane sleeve
9
Version 2.1 Update: April 2016
2.2.4 NTU sensor: Turbidity in NTU-mg/l. The measuring principle is based on nephelometry: a diode emits infrared light (850 nm) and an IR
receiving diode, set to one side at an angle of 90°, detects the amount of scattered light (standardized measurement). The sensor can be calibrated using a Formazine standard. This very economical optical technology requires very little maintenance and no consumables.
Measurements
Measurement principle
Scattering of IR at 90°
Measurement ranges
0 to 4,000 NTU in 5 ranges:
0 - 50 NTU 0 – 200 NTU 0 – 1,000 NTU 0 – 4,000 NTU AUTO range
0 to 4,500 mg/L range Calibration:
0-500 mg/L range, as per standard NF EN 872 range >500 mg/l as per standard NF T 90 105 2
Resolution
from 0.1 to 1, set automatically as a function of the range
Accuracy
< 5% of the NTU value recorded
Response time
< 5 s
Operating temperature
0 °C to + 50 °C
Temperature measurement
Via an NTC thermistor
Storage temperature
-10 °C to + 60 °C
Signal interface
Modbus RS-485 as standard and SDI-12 as an option
Maximum refresh rate
< 1 second
Power supply to sensor
5 to 12 volts
Power consumption
Standby: 40 µA / Warm-up time: 100 mS/ Current pulse: 500 mA When sending via RS-485 (1 measurement/second): 820 µA When sending via SDI-12 (1 measurement/second): 4.2 mA
Sensor
Dimensions
Diameter: 27 mm; Length not including cable: 170 mm
Weight
300 g (with 3 meters of cable)
Materials
PVC, POM-C, PMMA, Polyamide
Maximum pressure
5 bar
Cable/connection hardware
9-wire shielded conductor, uncoated-strand polyurethane sleeve
Ingress protection rating
IP68
10
Version 2.1 Update: April 2016
2.2.5 C4E sensor: 4-electrode conductivity. The operation of the sensor is based on 4-electrode conductivity technology: an alternating current at
constant voltage is set up between a pair of graphite primary electrodes. The secondary electrodes, made from platinum, adjust the drive potential at the primary electrodes to compensate for any fouling. The voltage measured between the primary electrodes varies depending on the resistance of the medium, and thus the conductivity.
Measurements
Measurement principle
4-electrode type conductivity sensor (2 graphite + 2 platinum).
Conductivity measurement range
0 - 200.0 µS/cm 0 - 2,000 µS/cm
0.00 - 20.00 mS/cm
0.0 - 200.0 mS/cm
Resolution
from 0.01 to 1, depending on the range
Accuracy
+/ 1 % of full scale
Salinity measurement range
5-60 g/kg
TDS-KCl range
0 – 133,000 ppm
Response time
< 5 s
Operating temperature
0 °C to 50 °C
Temperature compensation
Via an NTC thermistor
Storage temperature
- 10 °C to + 60 °C
Signal interface
Modbus RS-485 as standard and SDI-12 as an option
Measurement refresh rate
< 1 second maximum
Power supply for sensor
5 to 12 volts
Power consumption
Standby: 25 µA When sending via RS-485 (1 measurement/second):
6.3 mA When sending via SDI-12 (1 measurement/second):
9.2 mA Current pulse: 500 mA
Sensor
Dimensions
Diameter: 27 mm; Length not including cable: 177 mm (not including temperature sensor)
Weight
350 g (sensor + 3 m of cable)
Materials in contact with the medium
PVC, POM-C, stainless steel
Maximum pressure
5 bar
Cable/connection hardware
9-wire shielded conductor, uncoated-strand polyurethane sleeve
Ingress protection rating
IP68
11
Version 2.1 Update: April 2016
2.2.6 CTZN sensor: inductive conductivity. The operation of the CTZN sensor is based on a conductive induction measurement principle. A ring-type coil is excited at a fixed frequency and the response is retrieved on a second coil, linked to
the excited coil. The coupling between the coils varies depending on the conductivity of the conducting solution present.
Measurements
Measurement principle
Inductive conductivity sensor with temperature compensation
Conductivity measurement range
0.0 to 100.0 mS/cm
Resolution
0.1
Salinity measurement range
5-60 g/kg
Operating temperature
0 to 50 °C
Temperature compensation
Via an NTC thermistor or an external measurement
Measurement accuracy of T°C
0.1 °C over a range of 0-40 °C
Response time
T90<30 s
Storage temperature
-10 °C to 60 °C
Signal interface
Modbus RS-485 and SDI-12
Measurement refresh rate
< 1 second maximum
Power supply for sensor
5 to 28 volts, max. voltage: 30 V
Power consumption
Automatic standby of < 50 µA, warm-up time: 100 ms When sending via Modbus RS-485/ Range 0-100 mS/cm
Vin 5V
Vin 12 V
Vin 24 V
1 measurement/s
31 mA
15.5 mA
11.5 mA
Max. current peak of 700 mA for 2 mS, 350 mA for 150 mS
Sensor
Dimensions
Max. diameter: 62.4 mm, Length: 196 mm
Weight
700 g
Materials in contact with the medium
EPDM, PVC, stainless steel
Maximum pressure when immersed
5 bar
Cable/connection hardware
9-wire shielded conductor, uncoated-strand polyurethane sleeve
Ingress protection rating
IP68
12
Version 2.1 Update: April 2016
2.2.7 VB5 sensor: Sludge Blanket detection. The principle of measure is based on the mitigation of the Infra-Red signal in 870 nm through an
optical path of 5mm. The sensor delivers measures in Sludge Blanket detection in % of transmission IR. For a better precision, the optics of the sensor are regulated in temperature.
Measurements
Measurement principle
Optical IR (870 nm) based on IR absorption
Range of measure
Sludge blanket : 0-100 %
Resolution
Sludge blanket : 0.01 à 0.1 %
Accuracy
Sludge blanket : +/- 2%
Response time
< 35 secondes
Temperature measure
Principle of measure
NTC
Working temperature
-5.00 °C to + 60,00°C
Resolution
0,01 °C
Accuracy
+/- 0.5 °C
Storage Temperature
-10°C to + 60°C
Degree of protection
IP 68
Singal Interface
Modbus RS-485 or SDI-12
Refreshment of the measure
Maximum < 1 seconde
Power supply
5 to 28 volts
Consumption
Standby : 25 µA (5 V) Average RS485 (1 measure/seconde) : 4.5 mA (5V) Average SDI12 (1 measure/seconde) : 4.5 mA (5V) Curent Pulse : 100 mA during 30 mS Heating times : 100 mS
Sensor
Weight
750 g (sensor)
Material
DELRIN, Nickel-plated brass, EPDM
Pressure max.
5 bars
Cable/ connexions
9-wire shielded conductor, uncoated-strand polyurethane sleeve
Ingress protection rating
IP68
13
Version 2.1 Update: April 2016
2.2.8 MES5 sensor: Sludge Blanket detection, Suspended Solid, Turbidity. The principle of measure is based on the mitigation of the Infra-Red signal in 870 nm through an
optical path of 5mm. The sensor delivers measures in Suspended Solid (g/l), Turbidity (FAU) and Sludge Blanket detection in % of transmission IR. For a better precision, the optics of the sensor are regulated in temperature.
For a measure of Suspended Solid, the sensor is directly calibrated on the material to be measured (sample of sludge).
In Turbidimeter version the sensor delivers measures on a range 0-4000 FAU (Formazine Attenuation Unit) and is calibrated with solutions of Formazine.
Temperature: measures and regulation of optics via CTN.
Suspended Solid measure
Principle of measure
Optical IR (870 nm) based on IR absorption
Range of measure
SS : 0-50 g/L Turbidity : 0-4000 FAU Sludge blanket : 0-100 %
Resolution
SS : 0.01 g/L Turbidity : 0.01 à 1 FAU Sludge blanket : 0.01 à 0.1 %
Accuracy
SS< 10 % Turbidity : +/- 5% (range 200-4000 FAU) Sludge blanket : +/- 2%
Response time
< 35 secondes
Temperature measure
Principle of measure
NTC
Working temperature
-5.00 °C to + 60,00°C
Resolution
0,01 °C
Accuracy
+/- 0.5 °C
Storage Temperature
-10°C to + 60°C
Degree of protection
IP 68
Singal Interface
Modbus RS-485 or SDI-12
Refreshment of the measure
Maximum < 1 seconde
Power supply
5 to 28 volts
Consumption
Standby : 25 µA (5 V) Average RS485 (1 measure/seconde) : 4.5 mA (5V) Average SDI12 (1 measure/seconde) : 4.5 mA (5V) Curent Pulse : 100 mA during 30 mS Heating times : 100 mS
Sensor
Weight
750 g (sensor)
Material
DELRIN, Nickel-plated brass, EPDM
Pressure max.
5 bars
Cable/ connexions
9-wire shielded conductor, uncoated-strand polyurethane sleeve
Ingress protection rating
IP68
14
Version 2.1 Update: April 2016
3. Installation
3.1 Description and mounting of the controller.
3.1.1 Description of the front face.
1
Touch screen
2
Screws securing lower cover (2 screws)
3
Upper cover – do not open.
4
Lower cover - remove to make electrical connections
5
Four cable glands
3.1.2 Equipment required. The ACTEON 5000 unit is shipped with a bag containing 2 mounting brackets (with 2 pan head
screws) and a set of 12 connectors (two 5-contact connectors, three 4-contact connectors, three 3­contact connectors, three 2-contact connectors and one 1-contact connector to connect up the power supply).
Equipment required for installing the unit and for making the electrical connections:
- PH1x75 mm cross head screwdriver for attaching the unit's mounting brackets and for the screws securing the cover which protects the electrical connections,
- 2.0 x 75 mm flat-blade screwdriver for working on the various electrical connection terminals. To mount the ACTEON 5000 on a wall, use M5 screws (with a head size of < 10.8).
15
Version 2.1 Update: April 2016
3.1.3 Outline drawings of the ACTEON 5000.
Overall dimensions of the ACTEON 5000.
Diagram 1: Outline drawing for the ACTEON 5000 unit
Dimensions when mounted on a wall.
Diagram 2: Outline drawing for the ACTEON 5000 unit – wall mounting
A
225 B 198 C 165 D 178.5
E
198 A 213 mm
B
185 mm
C
84 mm
a1
179.3 mm
b1
125 mm
b2
111.3 mm
c1
71.5 mm
c2
53.5 mm
16
Version 2.1 Update: April 2016
3.2 Electrical connections.
3.2.1 Safety instructions – Installation. Isolate the power supply to the measuring device before performing any electrical connection work.
The electrical wiring and cabling work must be performed exclusively by authorized personnel.
Due to the risk of electrocution, systematically isolate the power supply to the controller before performing electrical connection work.
Warning for the model supplied with 10 – 30V
Risk of electrocution: do not connect a device operating in mains power mode to a model supplied with 10-30 V.
Risk of electrocution: the connection of a protective earth (PE) is compulsory when wiring and cabling both 100-240 Va.c. and 10-30 Vd.c. models.
Warning for wiring the relays
Fire risk. Since the relay contacts have a nominal value of 6 A, the external loads connected to the relays must be fitted with devices which limit the current to < 6 A.
Open the unit's lower cover to access the controller's wiring terminals. Unscrew the two screws which hold the lower protective cover. Before removing the cover, loosen all the cable glands so that the cables can slide in and out. Figure 3 shows the transmitter's wire connection zone when the lower protective cover is removed.
17
Version 2.1 Update: April 2016
3.2.2 Description of the wiring.
Diagram 3: Photograph showing the terminals on the circuit card
Diagram 4: Diagram of the terminals affixed (as a self-adhesive label) to the protective cover
Identification
Description
Terminal identification on circuit card
Terminal identification on protective cover (self-adhesive label)
1 – Power Supply
Option
110-230 V a.c. power supply
Yellow and green - Earth Blue - Neutral Brown - Phase
10-30 V d.c. power supply
Black Red
T (-) N (+) P
T (-) N (+) P
E N P
­+ NC
2 – ON/OFF INPUT
Two dry contract inputs
WAKE – Inactive GND – Inactive
ET1 GND ET2 GND
WAKE – Inactive GND – Inactive
ET1 GND ET2 GND
3 – Digital sensors
RS485 - INPUT
Two digital sensor inputs
V- sensor power supply: Black Shielding – Yellow and Green RS485 - Green
GND GND B
GND SHIELD B
  
A
B
18
Version 2.1 Update: April 2016
RS485 - White V+ sensor power supply: Red
A 12V A 12V
4 – ANALOG OUTPUT
Two analog outputs
0-20 mA or 20-0 mA or 4-20 mA or 20-4 mA
Output 1: - Output 1: +
Output 2: - Output 2: +
GND SA1
GND SA2
GND SA1
GND SA2
5 – RELAY OUTPUT
2 relay outputs Output 1: Release (default is NC) Output 1: Switch Output 1: Operate (default is NO)
Output 2: NC Output 2: Switch Output 2: NO
R1 C1 T1
R2 C2 T2
R1 C1 T1
R2 C2 T2
6 – USB
USB port For downloading data
USB
7 – Ethernet
Ethernet option
ETH
8 – 485 OUTPUT
One RS485 output
RS485 (-) RS485 (+)
G B A
GND B A
9 - ANALOG INPUT
Two analog inputs
Analog input 2 Input 2: - Input 2: +
Analog input 1 Input 1: - Input 1: +
Jumpers: Selection jumpers
Voltage or Current
GND EA2
GND EA1
U I
GND EA2
GND EA1
U I
A - RESET
RESET button
RESET
/
B- WAKE
A button which should not be activated
WAKE
/
Table 1: Description of the electrical connections
19
Version 2.1 Update: April 2016
Electrical connections for digital sensors:
Diagram 5 below shows the electrical connections required for digital sensors when the cable lengths are 15 meters or less, and when the cable lengths are more than 15 meters. When a CTZN sensor is connected, the wiring diagram to be used is that corresponding to a cable length of more than 15 meters, regardless of the length of the cable.
Diagram 5: Sensor connection for a cable length of 15 meters or less, and of more than 15 meters.
1 - red
V+ power
2 - blue
SDI-12 – Not connected
3 - black
GND
4 - green
B " RS-485 "
5 - white
A " RS-485 "
6 -
green/yellow
Cable shield
Connect cable 3 and 6
RED YELLOW ORANGE
VIOLET
PINK
V+ power
2- blue
SDI-12 – Not connected
3 - black
V- power
4 - green
B " RS-485 "
5 - white
A " RS-485 "
6 – green/
yellow
Cable shield
Cable length more than 15 meters Connect cable 3 and 6
20
Version 2.1 Update: April 2016
4. User Interface
4.1 Home screen.
The figure below shows the data displayed on the home screen when two sensors are connected to the ACTEON 5000. In this specific case, a pH/Temperature combined sensor and an oxygen sensor are connected to the ACTEON 5000.
1
Primary parameter measured by sensor 1
10
State of the two analog outputs: in this example, analog output 1 is assigned to sensor No.1. The red mark indicates 4 mA.
2
Stability indicator for the primary parameter measured by sensor 1
11
State of Relay 1. In this example, relay 1 is assigned to sensor 1 and is in an open position.
3
Real-time value of the primary parameter measured by sensor 1
12
Unit for the secondary parameter measured by sensor 2
4
Primary parameter measured by sensor 2
13
Real-time value of the secondary parameter measured by sensor 2
5
Stability indicator for the primary parameter measured by sensor 2
14
State of the two analog outputs: in this example, analog output 1 is assigned to sensor No.2. The red mark indicates 4 mA.
6
Real-time value of the primary parameter measured by sensor 2
15
State of Relay 2. In this example, relay 2 is assigned to sensor 2 and is in an open position.
7
Secondary parameter measured by sensor 1
16
Navigation icon used to access the main menu
8
Unit for the secondary parameter measured by sensor 1
17
Logos representing the analog outputs
9
Real-time value of the secondary parameter measured by sensor 1
18
Secondary parameter measured by sensor 2
Table 2: description of the home screen.
If only one sensor is connected to the ACTEON 5000 then lines are displayed in the lower part of the display where the parameter information and units would normally appear.
1 2 5 4 3 6 7 8 9
10
11
12
13
14
15
16
17
18
21
Version 2.1 Update: April 2016
4.2 Navigation icons.
The device has a touch screen so that the user can navigate through the various menus using the icons listed in the table below.
Icon
Functionality
Moves downwards through the options.
Moves upwards through the options.
Takes you back to the previous screen.
Opens the Main Menu containing all the main programming functions for the device. This icon appears on the home screen.
Takes you back to the home screen.
Cancels an action and takes you back to the previous screen.
Confirms a selection and opens a new screen.
Confirms a programming task or action.
Opens the menu used to modify the Modbus address for the digital sensors.
Indicates an unusual situation, and provides information which can be consulted. When it appears on the home screen, it indicates that the sensor is currently in a special state which may be consulted by pressing on the value displayed.
Transfers data to a USB stick.
Indicates that the screen is locked.
Table 3: Functionalities of the navigation icons.
22
Version 2.1 Update: April 2016
5. Setting up
5.1 Initial start-up.
When the transmitter is switched on, the home screen (i.e. the main measurement screen) appears with no indication of the sensor(s) installed if no sensors have yet been configured. If the sensors connected have already been configured, measured values may be displayed (primary and secondary parameters).
As an initial step, the operator can set the display language, the date format and the date and time.
Setting the language:
To access the language menu, follow the sequence shown below from the home screen:
At the home screen, select the Main menu icon, then the Device settings icon, and finally the Language icon.
Setting the date and time:
From the Device settings menu, select the Date/hour settings icon.
Option
Description
Format
Used to set the date format: the options are DD/MM/YY, YY/MM/DD and MM/DD/YY.
Place the pointer on the "Format" line using the up and down navigation icons, then select the Confirm icon.
Date
Select the "Date" line, press the confirm icon and then set the date by changing the values on the screen which, in order, correspond to the day, month and year. Use the up and down navigation icons to change the numbers and the Confirm icon to move from one variable to another.
Hour
Used to set the time: select the "Hour" line using the up and down navigation icons, press the Confirm icon, then set the hour and minutes.
Summer/winter time
To activate/deactivate automatic change-over to summer or winter time, select the "Summer/Winter time" line, confirm with the Confirm icon and select "Yes" or "No".
23
Version 2.1 Update: April 2016
5.2 Installation of digital sensors.
When sensors are connected to the transmitter for the first time they must be installed by running a SCAN (to scan the addresses from 1 to 243).
As soon as the transmitter detects a sensor, it displays its address, its serial number (which is also engraved on the body of the sensor) and a description of the sensor.
In the example above, pH and oxygen sensors have been detected: the pH sensor is at address 24, the sensor's serial number is SN-PPHRA-0762 and the description of the sensor is pH/redox/Temperature Ponsel Mesure. The OPTOD sensor is at address 10, the serial number is SN­PODOA-4032 and its description is OPTOD/Temperature Ponsel Mesure.
The operator can stop the SCAN operation at any time by pressing on the icon. If no sensor is detected, or 2 sensors have the same address (an address conflict) or a communication
error is detected, then the device displays a warning message accompanied by the icon. Pressing on the icon accesses the addressing menu (refer to section 6.2.1 for more details
about the addressing function) where the operator can change the sensor's address in the event of an address conflict (i.e. if 2 sensors have the same address).
5.3 Selection of the parameters.
Once ACTEON has detected the sensors, the parameters measured by the sensors must be set up. For each sensor, a primary and a secondary parameter can be selected. To access the measurement settings menu, follow the sequence shown below from the home screen:
To set up the primary and secondary parameters for sensor 1, select the corresponding icon.
To select a different line, press on the and icons.
Open the set-up line of interest using the Confirm icon, .
24
Version 2.1 Update: April 2016
Set-up line
Functionality
Average
Used to set the number of measurements (from 1 to 50) from which the moving average is calculated.
Primary parameter
This line lets the operator select the primary parameter for sensor 1, which shall then be displayed in the upper part of the home screen (refer to item 3 in Table 2). The next window then lets the operator select from a list of the parameters measured by sensor No. 1 (up to 4 parameters).
Secondary parameter
This parameter shall be displayed in the upper right-hand corner of the upper part of the home screen (refer to item 7 in Table 2). The next window then lets the operator select from a list of the parameters measured by sensor No. 1.
Measurement range
The measurement range can only be set for the conductivity and turbidity parameters, as indicated in the table below.
Temperature units
This line is used to select the units in which the temperature parameter will be displayed (°C or °F).
Refresh rate
This line lets the operator select an interval between each measurement of between 1 and 60 seconds.
The sensors can measure up to 4 parameters, as described below:
Sensor
Parameters measured
Measurement range options
OPTOD
Temperature Oxygen as a % of saturation Oxygen in mg/L Oxygen in ppm
PHEHT
Temperature pH Redox in mV
EHAN
Temperature Redox in mV
C4E
Temperature Conductivity in µS/cm or mS/cm Salinity in g/kg TDS in ppm
Conductivity: Auto (Automatic range) 0-200 µS/cm 0-2,000 µS/cm 0-20 mS/cm 0-200 mS/cm
NTU
Temperature Turbidity in NTU Turbidity in FNU Turbidity in mg/L
Turbidity: Auto (Automatic range) 0-50 NTU 0-200 NTU 0-1,000 NTU 0-4,000 NTU
CTZN
Temperature Conductivity in mS/cm (default parameter) Salinity in g/kg Conductivity (not compensated for temperature) in mS/cm
VB5
Temperature Sludge Blanket detection
MES
Temperature Sludge Blanket detection Suspended Solid Turbidity
25
Version 2.1 Update: April 2016
6. Programming
6.1 Main menu.
From the home screen, access the main menu using the icon.
Icon
Functionality
This menu is dedicated to PONSEL digital sensors and is used to set up the 2 digital inputs, including: the sensor detection function (SCAN), the setting-up of the parameters measured by the sensors, the calibration of the sensors and the setting of each sensor's Modbus address.
This menu is used to set up the two On/Off inputs for the washing, Event, Alert, etc. functions.
This menu is used to select voltage or current for the analog inputs.
This menu is dedicated to configuring the ACTEON 5000 and is used to: set-up the screen; consult the hardware/software version and the events log; set the date/time and language; return to factory settings and set a password.
This menu is used to set up the optional Ethernet digital output and the MODBUS functionality.
This menu is used to set up the two relay outputs in alarm/fault mode or based on thresholds.
This menu is used to set up the two analog outputs for the 0/4-2 mA or PID linking functions.
This menu is used to set up the recording/exporting to a USB stick function.
6.2 Description of the menus.
6.2.1 Configuring the digital inputs.
26
Version 2.1 Update: April 2016
This menu is used to set up two digital inputs when installing PONSEL sensors and is also used to: select the parameters measured by the sensors; scan the network of sensors connected to the ACTEON 5000; calibrate the sensors; and modify the Modbus address of the sensors.
To access the "Numeric input settings" window from the Main menu, select the icon.
Icon
Functionality
Used to access the setting up of digital sensors (parameter selection, etc.).
This menu lets you scan and detect the PONSEL digital sensors connected to the ACTEON 5000.
This icon lets the user calibrate the digital sensors and set up the compensation parameters.
This menu lets the operator modify the Modbus address of a sensor (notably used when two identical sensors are installed on the device).
Setting up digital sensors
The menu used to set up the digital sensors is used to: select the parameters which will be displayed on the home screen; select the number of values from which the average is calculated (moving average); select the units (if necessary); and set up the interval between each measurement.
Section 5.3 provides a detailed description of this functionality.
Scan of the sensors connected to the transmitter.
The SCAN functionality detects and identifies the sensors connected to the ACTEON 5000. This task is only performed when the device is switched on and when a sensor in the measurement chain is changed.
Refer to section 5.2 for more details about the SCAN function.
Calibration of digital sensors.
27
Version 2.1 Update: April 2016
From the "Numeric input settings" window, select the icon then select the sensor to be calibrated.
This window is used to calibrate the parameters measured by the selected sensor, or to set the external compensation data.
Icon
Functionality
Used to access the sensor's calibration menu
This menu is used to set up the external data used for compensation purposes
The external compensation data is detailed in the table below:
Set-up line
Functionality
Atmo. pressure
Atmospheric pressure compensation is applied to the OPTOD sensor (measurement of oxygen levels using optical technology). The atmospheric pressure sensor is built into the ACTEON 5000. The first line is used to activate the compensation and to access a second line which is used to adjust this parameter (possible values: 0 to 2,000 hPa).
Salinity
Salinity compensation can be applied to the Oxygen in mg/L parameter measured by the OPTOD oxygen sensor. The first line is used to activate the compensation and to access a second line which is used to adjust this parameter (possible values: 0 to 85.00 g/kg).
Temperature
Each sensor is equipped with its own temperature sensor, and the temperature recorded is considered if temperature compensation is required (for the pH, oxygen in mg/L and conductivity parameters). However, it is also possible to enter a fixed external compensation value. The first line is used to activate the compensation and to access a second line which is used to adjust this parameter (possible values: 0.00 to 40.00 °C).
Alpha
This coefficient can be set for the Non-compensated conductivity parameter measured by the CTZN sensor (possible values: 0.0 to 6.0 %/°C) in order to activate linear compensation.
To select the line above or below, or to increase or reduce a value, use the and icons.
The icon confirms the value set for the external compensation data. Any changes made become effective when the "Communication completed successfully" message appears.
To access the "Calibration settings" menu from the "Calibration: sensor 1 (or 2)" window, select the "Calibration" icon.
28
Version 2.1 Update: April 2016
Icon
Functionality
Used to access the sensor’s calibration menu.
This menu is used to apply the default calibration coefficients set in the factory.
Detailed calibration log: this menu contains the data for the last ten calibrations (offset/gain) … The coefficients for a specific calibration can be reused.
The Calibration settings window lets the operator: select the parameter to be calibrated; enter the operator's name; select a calibration mode and set the values of the standards used.
Set-up line
Functionality
Parameter
The parameter to be calibrated can be selected from a drop-down menu which offers the parameters measured by the sensor.
Operator
This menu is used to enter the Operator's name (maximum of 13 characters) which will be saved with the results of a calibration process.
Type
This line appears if the parameter selected for calibration is oxygen as a %Sat. To complete this line, the operator can select either a 2-point calibration (OFFSET and GAIN) or a single-point calibration (GAIN only).
Mode
The options are: "Single" or "Multiple" mode: when several sensors of the same type are connected, a parameter common to all the sensors can be calibrated. The "Single" option is used when calibrating one parameter for one sensor, and "Multiple" is used when calibrating one parameter for 2 sensors (e.g. the Temperature parameter measured by all the sensors, or the connection of 2 pH sensors).
Standard 1
In the "Standard 1" window, which only appears if the value can be modified, the value of the Standard can be set using a number pad and will be used during the first step of the calibration for the calculation of the OFFSET. The value which can be set cannot exceed the minimum and maximum values.
29
Version 2.1 Update: April 2016
Standard 2
In the "Standard 2" window, which only appears if the value can be modified, the value of the Standard can be set using a number pad and will be used during the second step of the calibration for the calculation of the GAIN. The value which can be set cannot exceed the minimum and maximum values.
Site
If a NTU sensor or a MES sensor is connected to the device and when selecting a calibration in mg/L or in g/L, the calibration process will be saved in a file saved under the name of the site. Up to 10 sites can be configured, each identified using 8 characters.
Pressing on the icon validates the setting of the calibration data and starts the first calibration step ("Standard 1" window).
In the window corresponding to the first step in the calibration, the first line indicates the value of the N1 standard and the second line presents the measurement delivered by the sensor in contact with the standard solution. A message indicates if the value measured by the sensor is stable or not. The confirmation of the first calibration step (gearwheel icon) initiates the second step ("Standard 2" window). Once the second calibration step has been confirmed, a summary window appears which presents the information about the values of the standards used and the coefficients calculated.
The appearance of the icon on a coefficient calculation line indicates that the coefficient has been confirmed, whereas the icon indicates that the coefficient obtained is out-of-range.
The table below presents the calibration steps for each parameter:
PARAMETER
STANDARD 1
STANDARD 2
Temperature °C
Water at a temperature close to 0 °C (bath of crushed ice, with T°C measured using a certified thermometer)
Bath of crushed ice, with T°C measured using a certified thermometer
O2 % Sat
0.00 % (water + sulfite (2% sulfite solution))
100.0 % (Humid, oxygen-saturated air)
pH
7.01 (buffer solution at 25 °C) Enter the corresponding pH value at the temperature of the standard solution.
4.01 (or 9.01; 10.01 buffer solution at 25 °C) Enter the corresponding pH value at the temperature of the standard solution.
Redox
0 mV (sensor exposed to air for an electronic 0)
240 mV (or 470 mV buffer solution)
Conductivity - C4E sensor Range 0.0-200.0 µS/cm Range 0-2,000 µS/cm Range 0.00-20.00 mS/cm Range 0.0-200.0 mS/cm
0 µS/cm Sensor exposed to air
Enter the value of the solution at 25 °C.
84 µS/cm (buffer solution at 25 °C) 1,413 µS/cm (buffer solution at 25 °C) 12,880 µS/cm (buffer solution at 25 °C)
111.8 mS/cm (buffer solution at 25 °C)
Conductivity – CTZ sensor Range 0.00 - 100.0 mS/cm
Deionised water at 0.00 mS/cm
Select the second point based on the range selected:
1.413 mS/cm (buffer solution at 25 °C)
2 mS/cm (buffer solution at 25 °C)
12.88 mS/cm (buffer solution at 25 °C)
56.84 mS/cm (buffer solution at 25 °C)
Turbidity Range 0.00-50.00 NTU Range 0.0-200.0 NTU Range 0-1,000 NTU Range 0 - 4,000 NTU Range 0-4000 FAU
0 NTU Distilled water
25.00 NTU (from a 4,000 NTU stock solution)
100.0 NTU (from a 4,000 NTU stock solution)
500.0 NTU (from a 4,000 NTU stock solution) 2,000 NTU (from a 4,000 NTU stock solution) 2,000 NTU (from a 4,000 NTU stock solution)
30
Version 2.1 Update: April 2016
Sludge blanket detection
100 % in tap water
0% occultation du faisceau optique à l’aide d’un réglet opaque ?
31
Version 2.1 Update: April 2016
Addressing of the digital sensors.
The first 3 lines of the "Addressing" window are used to define the serial number of the sensor affected by the change of address.
After configuring the Serial No. and address, confirm with the icon. The sensor's serial number is engraved on the body of the sensor and comprises the following
information:
Set-up line
Functionality
Type SN-XXXX
PODO for an OPTOD sensor PNEP for a Turbidity sensor PC4E for a C4E sensor PPHR for a PHEHT sensor PORP for an ORP sensor PTUR for a VB5 – MES 5 sensor
Revision
Represented by a letter (A in the example above)
Number
A 4-digit number (0000 in the example above)
Address
Between 1 and 243
After configuring the Serial No. and address, confirm with the icon. The change made to the Modbus address becomes effective when the "Communication completed successfully" message appears.
To access the "Addressing" menu, select the
icon in the Sensors scan window or select the
icon in the Numeric input settings screen.
32
Version 2.1 Update: April 2016
6.2.2 Configuring the On/Off inputs.
Set-up line
Functionality
I1: Function
Programming of On/Off Input No.1:
Inactive (no programmed action) Washing: used with an external washing system, for which one On/Off
output would be connected to the ACTEON 5000. The washing activation signal activates Automatic Maintenance Mode for the relay and analog outputs (refer to sections 6.2.5 and 6.2.6). A note regarding the activation and ending of the washing shall be recorded in the Events log (refer to section 6.2.4).
Event: the functionality is similar to that of the washing function, however, in the events log this information shall be recorded with a note regarding the start and end of the event.
Alert: a functionality used to monitor an alarm function fitted to an external item of equipment. The activation of this function triggers Automatic Maintenance Mode for the relay and analog outputs (refer to sections 6.2.5 and 6.2.6). A note shall be included in the Events log of the start and end of the alert.
I1: Release state
NO (for Normally Open) NC (for Normally Closed)
I2: Function
Programming of On/Off Input No.2 (see above for the configuration of On/Off Input No.1) Inactive Washing Event Alert
I2: Release state
NO (for Normally Open) NC (for Normally Closed)
33
Version 2.1 Update: April 2016
6.2.3 Configuring the Analog inputs.
Set-up line
Functionality
Input 1
Programming of Analog Input No.1:
Inactive (no programmed action) Current: if a sensor is connected which generates a signal of 0/4-20 mA
Voltage: if a sensor is connected which operates within a 0-10 V range
I1: Signal min
For a Current input: can be set to a value between 0 and 20 mA For a Voltage input: can be set to a value between 0 and 10V
I1: Signal max
For a Current input: can be set to a value between the minimum value set up in the "Signal min" line and 20 mA.
For a Voltage input: can be set to a value between the minimum value set up in the "Signal min" line and 10 V.
I1: Average
Used to set the number of measurements, from 1 (instantaneous value) to 50, from which the moving average is calculated.
I1: Refresh rate
This line lets the operator select the interval between each measurement. The possible values are:1 to 60 seconds.
I1: Error reporting
Inactive (no error reporting) Out-of-range alert: if the measurement falls outside of the range
programmed by the Signal min and Signal max lines, information is sent to the relay output if the programming is set to "Alarm mode" (refer to section
6.2.6).
I1: Source signal 1 units
----------- : units not selected
mH2O: water height measurement sensor mbar: pressure sensor bar: pressure sensor
I1: Source signal 1 min
This line lets the operator set the minimum value to between -9999.99 and
9999.9
I1: Source signal 1 max
This line lets the operator set the maximum value to between the minimum value set on the "Source signal 1 min" line and 9999.9.
I1: Source signal 1 display
This line lets the operator arrange the information displayed on the home screen (refer to Table 2 for parameter display details)
Not displayed Sens 1. Prim. Meas.: the measurement shall be displayed in the space
allocated to the primary parameter for sensor 1 (item 3 on the home screen) Sens 1. Sec. Meas.: item 9 on the home screen.
34
Version 2.1 Update: April 2016
Sens 2. Prim. Meas.: item 6 on the home screen. Sens 2. Sec. Meas.: item 13 on the home screen.
I1: Source signal 1 desig.
------------ : no designation
Water height Defined by User: in this case, an additional line will appear as soon as this
option is confirmed
I1: User desig.
A free text field for entering the user's designation (maximum of 13 characters).
I1: Source signal 1 stability delta
Can be set from 0 to 50 % (the default is 1 %) and is used to define when the arrow on the home screen will indicate that the measurement is stable if the parameter measured has been set as the primary parameter for sensor 1 or 2 (refer to items 2 and 5 in Table 2)
I1: Source signal 2
Inactive Active: source signal 2 is calculated from source signal 1 by a conversion
program written in a file stored on the USB stick plugged into the USB port. The name of the file must be of the form "CONFANA2.csv" and must
contain the following information: Line 1 of the csv file: a description (not more than 16 characters) of the calculated source signal.
Line 2: a blank line containing no characters. Lines 3 to 32: a number of conversion points in the form of portions of a
straight line Source signal 2 = f(Source signal 1). To activate the conversion, plug the stick into the ACTEON 5000's USB port
(refer to item 6 of diagram 3) and wait for the small green LED located above the USB port to stop flashing. Then navigate to menu I2: Source signal 2 and select the "Active" option.
If Source signal 2 is Activated
I1: Source signal 2 display
Not displayed Sens 1. Prim. Meas.: the measurement shall be displayed in the space
allocated to the primary parameter for sensor 1 (item 3 on the home screen)
Sens 1. Sec. Meas: item 9 on the home screen. Sens 2. Prim. Meas: item 6 on the home screen Sens 2. Sec. Meas: item 13 on the home screen.
I1: Source signal 2 stability delta
Can be set from 0 to 50 % (the default is 1 %) and is used to define when the arrow on the home screen will indicate that the measurement is stable if the parameter measured has been set as the primary parameter for sensor 1 or 2 (refer to items 2 and 5 in Table 2)
Setting up Input 2
Input 2
Programming of Analog Input No.2:
Inactive (no programmed action) Current: if a sensor is connected which generates a signal of 0/4-20 mA
Voltage: if a sensor is connected which operates within a 0-10 V range
I2: Signal min
For a Current input: can be set to a value between 0 and 20 mA For a Voltage input: can be set to a value between 0 and 10V
I2: Signal max
For a Current input: can be set to a value between the minimum value set up in the "Signal min" line and 20 mA.
35
Version 2.1 Update: April 2016
For a Voltage input: can be set to a value between the minimum value set up in the "Signal min" line and 10 V.
I2: Average
Used to set the number of measurements, from 1 (instantaneous value) to 50, from which the moving average is calculated.
I2: Refresh rate
This line lets the operator select the interval between each measurement. The possible values are:1 to 60 seconds.
I2: Error reporting
Inactive (no error reporting) Out-of-range alert: if the measurement falls outside of the range
programmed by the Signal min and Signal max lines, information is sent to the relay output if the programming is set to "Alarm mode" (refer to section
6.2.6).
I2: Source signal 1 units
----------- : units not selected
mH2O: water height measurement sensor mbar: pressure sensor bar : pressure sensor
I2: Source signal 1 min
This line lets the operator set the minimum value to between -9999.99 and
9999.9
I2: Source signal 1 max
This line lets the operator set the maximum value to between the minimum value set on the "Source signal 1 min" line and 9999.9.
I2: Source signal 1 display
This line lets the operator arrange the information displayed on the home screen (refer to Table 2 for parameter display details)
Not displayed Sens 1. Prim. Meas: the measurement shall be displayed in the space
allocated to the primary parameter for sensor 1 (item 3 on the home screen)
Sens 1. Sec. Meas.: item 9 on the home screen Sens 2. Prim. Meas: item 6 on the home screen Sens 2. Sec. Meas.: item 13 on the home screen.
I2: Source signal 1 desig.
------------ : no designation
Water height Defined by User: in this case, an additional line will appear as soon as this
option is confirmed.
I2: User desig.
A free text field for entering the user's designation (maximum of 13 characters).
I2: Source signal 1 stability delta
Can be set from 0 to 50 % (the default is 1 %) and is used to define when the arrow on the home screen will indicate that the measurement is stable if the parameter measured has been set as the primary parameter for sensor 1 or 2 (refer to items 2 and 5 in Table 2)
I2: Source signal 2
Inactive Active: source signal 2 is calculated from source signal 1 by a conversion
program written in a file stored on the USB stick plugged into the USB port. The name of the file must be of the form "CONFANA2.csv" and must
contain the following information: Line 1 of the csv file: a description (not more than 16 characters) of the calculated source signal. Line 2: a blank line containing no characters
36
Version 2.1 Update: April 2016
Lines 3 to 32: a number of conversion points in the form of portions of a straight line Source signal 2 = f(Source signal 1).
To activate the conversion, plug the stick into the ACTEON 5000's USB port (refer to item 6 of diagram 3) and wait for the small green LED located above the USB port to stop flashing. Then navigate to menu I2: Source signal 2 and select the "Active" option.
If Source signal 2 is Activated
I2: Source signal 2 display
Not displayed Sens 1. Prim. Meas: the measurement shall be displayed in the space allocated to the primary parameter for sensor 1 (item 3 on the home screen) Sens 1. Sec. Meas: item 9 on the home screen Sens 2. Prim. Meas: item 6 on the home screen Sens 2. Sec. Meas: item 13 on the home screen.
I2: Source signal 2 stability delta
Can be set from 0 to 50 % (the default is 1 %) and is used to define when the arrow on the home screen will indicate that the measurement is stable if the parameter measured has been set as the primary parameter for sensor 1 or 2 (refer to items 2 and 5 in Table 2)
37
Version 2.1 Update: April 2016
6.2.4 Configuring the ACTEON 5000.
From the main menu, select the icon to open the ACTEON 5000 settings window.
Icon
Functionality
Screen settings: A menu used: to set how long the backlight stays on for (from 1 to 60 minutes); to set the brightness (from 10 to 100 %); and to recalibrate the screen. A screen recalibration is only necessary when ACTEON 5000 is switched on for the first time, or after a software upgrade.
Hardware information: an information-only menu which displays the serial number and software version of the device, as well as the serial number, hardware version and model of the sensors connected to the device.
Events log: an information-only menu which lists all the events together with their time-
stamping and with the option to transfer this data to a USB stick via the icon
Date/time settings: a menu used to set the date and time, to choose the date format, and to enable/disable automatic summer/winter time change-over.
Language: used to select the display language.
Return to factory settings: used to reset all the ACTEON 5000's settings to the
defaults set during manufacture.
Password: Used to activate and define a password which prevents information from being displayed on the screen (a 4 digit code, between 0 and 9999). The universal emergency code is 3615.
38
Version 2.1 Update: April 2016
6.2.5 Configuring the Analog outputs.
To access the "Analog output settings" window, select the icon in the Main menu. The functionality offered for the setting up of the two analog outputs is described in the table below:
Set-up line
Functionality
MM: Err/Maintenance mode
Used to set the Maintenance Mode to one of the following modes: Automatic (the default option, and which is activated when the calibration menu is used, or if there is a loss of communication, etc.) Inactive Manual (this mode is activated manually when maintenance or other work is performed on the device).
State of outputs in MM
In Maintenance Mode, the analog outputs may return a value corresponding to:
- the last value measured,
- a fixed (unchanged) value which shall be defined in the "OUT. 1: Value mA in MM" and "OUT. 2: Value mA in MM" lines.
OUT. 1: Value mA in MM
This line is used to set the value which will be returned by analog output 1 when the device is in Maintenance Mode (between 0 and 21 mA).
OUT. 2: Value mA in MM
This line is used to set the value which will be returned by analog output 2 when the device is in Maintenance Mode (between 0 and 21 mA).
MM Final tempo (mn)
Used to set a time delay which shall begin when the Maintenance is completed, to allow a period of time for the measurement to stabilize. The possible values are: 0 to 60 minutes (1 min is the default setting).
Configuration of the Analog Output 1
OUT1 :Active function
Configuration of the output 1 according to 4 choices: No function Recorder Test – Fixed Value PID Controler
If Active Function : Recorder
OUT 1: Active Function
This function activates analog output 1. Used to set output 1 to Data logger mode (for when the output is connected to a controller/data logger).
OUT. 1: DL type
Used to set output 1 to operate within a range of 0-20 mA or 4-20 mA.
OUT. 1: Parameter linked to DL
Used to select the parameter whose value shall be outputted via analog output 1: Sensor 1: Primary parameter Sensor 1: Secondary parameter Sensor 2: Primary parameter
39
Version 2.1 Update: April 2016
Sensor 2: Secondary parameter
OUT. 1: DL Min. value
Used to set the minimum value of output 1
OUT. 1: DL Max. value
Used to set the maximum value of output 1
OUT. 1: DL gain mA/s
Used to set the "responsiveness" of the 4-20 mA output. Possible value: 0 to 40 mA/s
If Active Function : Test fixed value
OUT.1 : TEST Value mA
Possibility of simulation a value between 4.0 and 20.0 mA by step of 0.1 mA
If Active Function : PID Controler
OUT.1 : PID Type
Used to set output 1 to operate within a range of 0-20 mA or 4-20 mA
OUT.1 : PID linked param.
Used to select the parameter whose value shall be outputted for the PID function : Sens.1.pri.meas Sens.1.sec.meas Sens.2.pri.meas Sens.2.sec.meas
OUT.1 : PID % Band
Configurable between 0.00 and 1000.00 Proportional to the gap between the measure and the instruction
OUT.1 : PID Integral
Choice between Active (by fault) and Inactive
if OUT.1 : PID Integral: Active
OUT.1 : PID Integral
Customizable between 0.00 and 1000.00 s
OUT.1: PID Derivative
Choice between Inactive (by fault) and Active
If OUT.1 : PID Derivative : Active
OUT.1: PID Derivative
Customizable between 0.00 and 1000.00 / s
OUT.1: PID gain
Used to set the "responsiveness" of the 4-20 mA output. Possible value: 0 to 40 mA/s
OUT.1: PID Direction
Configurable in Direct or Reversed Mode
Configuration of the Analog Output 2
OUT2 :Active function
Configuration of the output 2 according to 4 choices: No function Recorder Test – Fixed Value PID Controler
If Active Function : Recorder
OUT. 2: Active Function
This function activates analog output 2. Used to set output 2 to Data logger mode (for when the output is connected to a controller/data logger).
OUT. 2: DL type
Used to set output 2 to operate within a range of 0-20 mA or 4-20 mA.
OUT. 2: Parameter linked to DL
Used to select the parameter whose value shall be outputted via analog output 2:
40
Version 2.1 Update: April 2016
Sensor 1: Primary parameter Sensor 1: Secondary parameter Sensor 2: Primary parameter Sensor 2: Secondary parameter
OUT. 2: DL Min. value
Used to set the minimum value of output 2
OUT. 2: DL Max. value
Used to set the maximum value of output 2
OUT. 2: DL gain mA/s
Used to set the "responsiveness" of the 4-20 mA output. Possible value: 0 to 40 mA/s
If Active Function : Test fixed value
OUT.2 : TEST Value mA
Possibility of simulation a value between 4.0 and 20.0 mA by step of 0.1 mA
If Active Function : PID Controler
OUT.2 : PID Type
Used to set output 2 to operate within a range of 0-20 mA or 4-20 mA
OUT.2 : PID linked param.
Used to select the parameter whose value shall be outputted for the PID function : Sens.1.pri.meas Sens.1.sec.meas Sens.2.pri.meas Sens.2.sec.meas
OUT.2 : PID % Band
Configurable between 0.00 and 1000.00
OUT.2 : PID Integral
Choice between Inactive and Active
if OUT.2 : PID Integral: Active
OUT.2 : PID Integral
Customizable between 0.00 and 1000.00 s
OUT.2: PID Derivative
Choice between Inactive and Active
If OUT.2 : PID Derivative : Active
OUT.2: PID Derivative
Customizable between 0.00 and 1000.00 / s
OUT.2: PID gain
Used to set the "responsiveness" of the 4-20 mA output. Possible value: 0 to 40 mA/s
OUT.2: PID Direction
Configurable in Direct or Reversed Mode
Maintenance Mode is activated under the following conditions:
- When a calibration is being performed,
- If communication with the sensor(s) is lost,
- If the measurement delivered by the sensor is incorrect,
- If an On/Off input in Washing Mode is activated.
41
Version 2.1 Update: April 2016
6.2.6 Configuring the relay outputs.
To access the "Relay output settings" window, select the icon from the Main menu.
Set-up line
Functionality
MM: Err/Maintenance mode
Used to set the Maintenance Mode to one of the following modes: Automatic (the default option, and which is activated when the calibration menu is used, or if there is a loss of communication, etc.) Inactive Manual (this mode is activated manually when maintenance or other work is performed on the device).
State of outputs in MM
In Maintenance Mode, the relay outputs may remain set as:
- the last value measured,
- a fixed (unchanged) value which shall be defined in the "R1: State in MM" and "R2: State in MM" lines.
R1: State in MM
In Maintenance Mode, Relay 1 can be set to Open or Closed mode.
R2: State in MM
In Maintenance Mode, Relay 2 can be set to Open or Closed mode.
MM Final tempo (mn)
Used to set a time delay which shall begin when the Maintenance is completed. The possible values are: 0 to 60 minutes.
R1 output settings
R1: Active Function
Used to set output 1 to one of the following: Not assigned Fault alarm Limit indicator
If Active function: Fault alarm
R1: Alarm state
Relay can be set to Open or Closed mode.
R1: Alarm tempo (s)
A time delay which must elapse prior to the triggering of the relay. Possible values: 0 to 3600 s.
If Active function: Limit indicator
R1: Action
Relay can be set to Open or Closed mode.
R1: Linked parameter
Used to select the parameter linked to output 1: Sensor 1: Primary parameter Sensor 1: Secondary parameter Sensor 2: Primary parameter Sensor 2: Secondary parameter
Set-up line
Functionality
R1: Type of limit
Used to define the limit triggering mode:
42
Version 2.1 Update: April 2016
Upper/overrange: when a limit is exceeded by being overshot Lower/underrange: when a limit is exceeded by being undershot Interval: the range is set by a combination of upper and lower limits.
If Active function: Limit Indicator and Type of Limit (Upper/overrange or Lower/underrange)
R1: Limit Value
Used to set the value of the Upper or Lower limit
R1: Hysteresis limit %
Possible values: 0 to 20 %
R1: Tempo
Possible values: 0 to 3600 s
If Active function: Limit Indicator and Type of Interval limit
R1: UPPER limit value
Used to set the value of the UPPER limit
R1: Hysteresis limit %
Possible values: 0 to 20 %
R1: Tempo
Possible values: 0 to 3600 s
R1: Forcing Tempo
Minimum time during which the contact (CLOSED or OPEN) shall be maintained in that state (between 0 and 120 minutes)
R1: LOWER limit value
Used to set the value of the LOWER limit
R1: Hysteresis limit %
Possible values: 0 to 20 %
R1: Tempo
Possible values: 0 to 3600 s
R1: Forcing Tempo
Minimum time during which the contact (CLOSED or OPEN) shall be maintained in that state (between 0 and 120 minutes)
R2 output settings
R2: Active Function
Used to set output 1 to one of the following: Not assigned Fault alarm Limit indicator
If Active function: Fault alarm
R2: Alarm state
Relay can be set to Open or Closed mode
R2: Alarm tempo (s)
A time delay which must elapse prior to the triggering of the relay. Possible values: 0 to 3600 s
If Active function: Limit indicator
R2: Action
Relay can be set to Open or Closed mode
R2: Linked parameter
Used to select the parameter linked to output 2: Sensor 1: Primary parameter Sensor 1: Secondary parameter Sensor 2: Primary parameter Sensor 2: Secondary parameter
Set-up line
Functionality
R2: Type of limit
Used to define the limit triggering mode: Upper/overrange: when a limit is exceeded by being overshot Lower/underrange: when a limit is exceeded by being undershot Interval: the range is set by a combination of upper and lower limits.
43
Version 2.1 Update: April 2016
If Active function: Limit Indicator and Type of Limit (Upper/overrange or Lower/underrange)
R2: Limit Value
Used to set the value of the Upper or Lower limit
R2: Hysteresis limit %
Possible values: 0 to 20 %
R2: Tempo
Possible values: 0 to 3600 s
If Active function: Limit Indicator and Type of Interval limit
R2: UPPER limit value
Used to set the value of the UPPER limit
R2: Hysteresis limit %
Possible values: 0 to 20 %
R2: Tempo
Possible values: 0 to 3600 s
R2: Forcing Tempo
Minimum time during which the contact (CLOSED or OPEN) shall be maintained in that state (between 0 and 120 minutes)
R2: LOWER limit value
Used to set the value of the LOWER limit
R2: Hysteresis limit %
Possible values: 0 to 20 %
R2: Tempo
Possible values: 0 to 3600 s
R2: Forcing Tempo
Minimum time during which the contact (CLOSED or OPEN) shall be maintained in that state (between 0 and 120 minutes)
Maintenance Mode is activated under the following conditions:
- When a calibration is being performed,
- If communication with the sensor(s) is lost,
- If the measurement delivered by the sensor is incorrect,
- If an On/Off input in Washing Mode is activated.
The conditions under which an On/Off input is activated in "Fault alarm" mode are:
- No response from sensor x parameter x,
- Poor response from sensor x parameter x,
- Out-of-range measurement for sensor x parameter x,
- Measurement not possible for sensor x parameter x,
- Low sensor power supply voltage,
- Low internal battery voltage,
- High internal temperature,
- Poor sensor communication response,
- On/Off input 1 alert
- Analysis of Actéon5000 program
- Low limit exceeded for analog input 1 or Upper limit exceeded for analog input 1
- On/Off input 2 alert
- Low limit exceeded for analog input 2 or Upper limit exceeded for analog input 2
44
Version 2.1 Update: April 2016
6.2.7 Configuring the Recording output.
Set-up line
Functionality
Recording
Inactive: the recording function is not activated Active: this option activates the recording function
Measurement variation
Possible values: 0.5 to 50 %. The measurement will not be recorded if its variation (as a %) from the last recorded value does not meet this requirement.
Interval
The interval between consecutive recordings can be set from 1 to 120 minutes
USB export
Inactive Active: after connecting a USB stick, the data recorded on the ACTEON
5000 is copied to the latter. To download the recorded data, plug the stick into the ACTEON 5000's USB
port (refer to item 6 of diagram 3) and wait for the small green LED located above the USB port to stop flashing. Navigate then to the "USB export" menu and select the "Active" option.
45
Version 2.1 Update: April 2016
7. Maintenance
7.1 Maintaining the transmitter:
Isolate the power to the measuring device when performing maintenance work.
Maintenance work must exclusively be carried out by authorized personnel.
Due to the risk of electrocution, disconnect the controller from the power supply systematically when performing maintenance work on the transmitter.
Do not use a corrosive or inflammable solvent to clean the transmitter. The use of this type of solvent could damage the device (its screen) and may invalidate the warranty.
After checking that the ACTEON 5000's protective covers are properly closed, wipe the outside of the transmitter with a soft cloth dampened with a mixture of water and non-corrosive detergent.
Replacing the lithium battery:
The lithium battery cannot be replaced by the operator; contact the after-sales service.
46
Version 2.1 Update: April 2016
7.2 Maintenance of digital sensors:
7.2.1 OPTOD sensor.
General description
Oxygen: luminescent membrane sensitive to the oxygen level in the medium
being analyzed. Gaseous exchange between the membrane’s material and
the medium. Temperature: NTC thermistor.
Materials
316L stainless steel version: polyamide, silicone and quartz; polyurethane sleeve around cable. Titanium version: polyamide, silicone and quartz; polyurethane sleeve around cable.
Precautions
The membrane is sensitive to:
- chemicals (organic solvents, acids, hydrogen peroxide),
- mechanical stresses (impacts, abrasion, tearing).
Measurement/Interference
When measuring, check that bubbles do not become trapped under the membrane.
If chlorine is present, the quality of the measurement will be impaired (overestimate of the level of dissolved oxygen). When immersing the sensor in the medium to be measured, wait until the sensor's temperature has stabilized before taking the reading. To optimize the service life of the sensor, we recommend a measurement interval (refresh rate) of more than 5 seconds.
Operating temperature
0 °C to 50 °C Temperature compensation effective over the range 0-40 °C
Servicing
After each use, rinse the sensor and the membrane carefully in clean water. If deposits such as biofilm or sludge persist, wipe the membrane carefully with a soft cloth or absorbent paper.
Caution: for the titanium version, clean the body of the sensor with acetone (do not use methylated spirits, ethanol or methanol).
Caution: only unscrew the sieve cap containing the DODISK when replacing the disk. If the sieve cap has to be replaced, screw the new sieve cap on slowly to allow the air to escape slowly.
Storage
Keep the membrane hydrated using the protective bag and a dampened absorbent wad of material (e.g. cotton wool. After being stored in dry conditions, rehydrate the membrane for 12 hours by immersing the sensor in water.
Storage temperature
- 10 °C to + 60 °C
Oxygen calibration
Using a clean sensor, occasionally check the 0% Sat value by immersing the sensor in a sulfite solution in water (sulfite concentration <2%). If there is a zero error, perform a complete calibration of the sensor.
Caution: do not allow the sensor to remain in contact with the sulfite solution for more than 1 hour.
The 2-point calibration is performed using a sulfite solution (for the offset)
then, after rinsing and drying, the sensor’s gain is determined by exposing the
sensor to moisture-saturated air (or to clean water saturated with air).
Temperature calibration
The calibration of the temperature sensor is a 2-step process:
- step 1 (offset): the sensor is placed in a container containing a water bath and ice,
47
Version 2.1 Update: April 2016
- step 2 (gain): the sensor is placed in a medium (with an even temperature distribution in the medium) at a known temperature. This temperature may be measured using a certified thermometer.
48
Version 2.1 Update: April 2016
7.2.2 NTU sensor: Turbidity.
General description
Turbidity: nephelometric measurement by IR diffusion (wavelength: 880 nm) at 90°. Temperature: NTC thermistor.
Materials
PVC, PMMA, Polyamide, POM-C, polyurethane sleeve on cable
Precautions
The optical windows are sensitive to:
- chemicals (organic solvents, strong acids and bases, hydrogen peroxide, hydrocarbons),
- mechanical stresses (impacts, abrasion).
Measurement/Interference
When in use, the sensor must not come into contact with the sides or bottom of a container. Maintain a minimum distance of 2 or 3 cm between the sides and the sensor (depending on the concentration of the medium).
Bubbles on the optical components can interfere with the measurements. When immersing the sensor in the medium to be measured, wait until the sensor's temperature has stabilized before taking the reading. If measurements are being made in a range between 0 and 20 NTU, or if the sensor saturates (a value of 9999), it is advisable to use the protective sieve to avoid interference from edge effects, sunlight, etc.
Operating temperature
0 °C to 50 °C
Servicing
After each use, rinse the sensor carefully in clean water. If deposits such as biofilm or sludge persist, clean the sensor carefully with soapy water and wipe the head with a soft cloth or absorbent paper.
Storage
Place the protective bag over the head of the sensor to prevent the optical part from being scratched.
Storage temperature
- 10 °C to + 60 °C
Turbidity calibration (in NTU)
The NTU sensor is an optical sensor which requires very little calibration. Using a clean sensor, occasionally check the 0 NTU value by immersing the sensor in clean, bubble-free water. If there is a zero error, perform a complete calibration of the sensor (over 1 or 4 measurement ranges).
This procedure requires a formazine solution whose concentration lies at the mid-point of the measurement range being calibrated. This solution is prepared from a 4,000-NTU stock solution.
Use a 200 ml volumetric flask when preparing the solutions. Add the required volume of formazine (refer to the table below) and make up to 200 ml with distilled water.
Formazine solutions with concentrations of less than 1,000 NTU degrade fairly quickly; do not keep the solution for more than a few days. The 2,000 NTU solution can be kept for 2 to 3 weeks in a fridge in an opaque bottle.
Measurement range
Concentration of formazine standard solution
Volume of formazine (mL)
0.0-50.0 NTU
25 NTU
1.25 mL
0.0-200.0 NTU
100 NTU
5 mL
0-1,000 NTU
500 NTU
25 mL
0-4,000 NTU
2,000 NTU
100 mL
49
Version 2.1 Update: April 2016
Turbidity calibration (in mg/L)
When a turbidity sensor is used to measure in a range whose units are mg/L, then the sensor must be calibrated using a field sample.
A 2-point calibration is performed:
- 1 offset using distilled water (0 mg/L),
- 1 gain using a sludge sample: immerse the sensor in the sample, with stirring, and log the theoretical value measured by the sensor. Using the same sample, analyze the dry weight in a laboratory in accordance with standard EN 872 within a range from 0-500 mg/L, and in accordance with standard NF T 90 105 2 if the concentration is > 500 mg/L.
Temperature calibration
Refer to section 7.2.1
50
Version 2.1 Update: April 2016
7.2.3 PHEHT sensor: pH/Temperature.
General description
pH/Redox: Potentiometric measurement; pH: a pair of electrodes; a reference (Ag/AgCl gel) and a glass bulb sensitive to H3O
+
ions Redox: a pair of electrodes; a reference (Ag/AgCl gel) and a platinum disk Temperature: NTC thermistor.
Materials
Glass, platinum, PVC, POM-C, Polyamide, Stainless steel 316L (sleeve protecting the temperature sensor); polyurethane sleeve on cable.
Precautions
The glass electrode is sensitive to:
- chemicals (organic solvents, strong acids and bases, hydrogen peroxide, hydrocarbons),
- mechanical stresses (impacts).
The electrode that measures the Redox potential is sensitive to sulfides, which are adsorbed onto the platinum.
Measurement/Interference
When immersing the sensor in the medium to be measured, wait until the sensor's temperature has stabilized before taking the reading.
Operating temperature
0 °C to 50 °C
Servicing
After each use, rinse the sensor carefully in clean water. pH: If deposits such as biofilm or sludge persist, immerse the sensor in a cleaning solution (PF-CSO-C-00010) for a few hours and rinse with plenty of water before use. Do not use a soft cloth or absorbent paper since the glass bulb is extremely sensitive to scratching. Redox: clean the platinum disk using fine, wet abrasive paper (such as P1200 or P220).
Storage
Keep the glass membrane hydrated by adding a few drops of storage solution (PF­CSO-C-00005) to the protective bag, or using a solution at pH 4. Rinse the glass bulb with plenty of water before use. If the sensor is stored in dry conditions, soak the sensor in a pH4 buffer solution for 12 hours before use.
The protective bag absorbs direct impacts on the head of the sensor. The platinum electrode is kept under dry conditions.
Storage temperature
0 °C to + 60 °C
pH calibration
Using a clean sensor, perform a 2-point calibration of the sensor (offset and gain, e.g. at pH7 and pH4).
Redox verification
Using a clean sensor, check the electronic 0 by exposing the sensor to air, and check a second point using a buffer solution at 240 mV (or 470 mV).
Temperature calibration
Refer to section 7.2.1
Changing the cartridge
In order not to damage the electronic part of the sensor, hold the cartridge in one hand and unscrew the connecting ring using the other hand. Remove the used
cartridge and insert the new cartridge before re-tightening the connecting ring.
51
Version 2.1 Update: April 2016
7.2.4 ORP sensor: REDOX/Temperature.
General Description
Redox : a pair of electrodes with a reference (Ag/AgCl gel) /platinum ring Temperature : NTC.
Materials
Glass, platinum, PVC, Polyamide, DELRIN; polyurethane jacketed cable.
Safeway
The redox potential electrode is sensitive to sulphide adsorption on platinum.
Measurement/ Interference
During the introduction of sensor in the measurement environment, wait sensor’s temperature stabilization before proceeding with measurement..
Operating temperature
0°C to 50°C
Maintenance
After each use, rinse meticulously the sensor with clear water. Clean the platinum ring with an abrasive moist paper (type P1200 or P220).
Storage
Maintain the head of the sensor hydrated in the protection case with a few drops of preservation agent solution (PF-CSO-C-00005) or, if it is not
available, with the solution of KCl. Rinse profusely the sensor before use. After storage in dry environment, put the sensor in a KCl solution for 12
hours. The case protects against direct impact on the head of the sensor.
Storage temperature
0°C to + 60°C
Redox verification
Using a clean sensor, check the electronic 0 by putting the sensor in free air and a second point with standard solution at 240 mV (or 470 mV).
Temperature calibration
Refer to section 7.2.1
Changing the cartridge
To avoid deteriorating the electronic part of the sensor, take the cartridge in one hand and unscrew the clamping ring with the other hand. Remove the
used cartridge and put the new cartridge before to screw back the clamping ring.
52
Version 2.1 Update: April 2016
7.2.5 C4E sensor: 4-electrode conductivity.
General description
Conductivity: Amperometric measurement with a 4-electrode system; Temperature: NTC thermistor.
Materials
Graphite, platinum, PVC, POM-C, Polyamide, Stainless steel 316L (sleeve protecting the temperature sensor); polyurethane sleeve on cable.
Precautions
The 4 electrodes are sensitive to deposits (greases, hydrocarbons, biofilm, sludges).
Measurement/Interference
When immersing the sensor in the medium to be measured, wait until the sensor's temperature has stabilized before taking the reading.
Operating temperature
0 °C to 50 °C
Servicing
After each use, rinse the sensor carefully in clean water. If deposits such as biofilm or sludge persists in the measurement slit or on the electrodes, use wet abrasive paper to remove a thin layer off the surface of the electrodes (type P1200 or P220).
Storage
The protective bag absorbs direct impacts on the head of the sensor. If storing for a short period between measurements, place a wad of cotton wool in the bottom of the storage bag, dampened with a few drops of buffer solution at 1,413 µS/cm.
Storage temperature
- 10 °C to + 60 °C
Conductivity calibration
Using a clean sensor, perform a 2-point calibration of the sensor (offset and gain using a standard solution whose conductivity is suited to the measurement range) for 1 or all 4 ranges:
Measurement range
Concentration of the standard conductivity solution
0.0-200.0 µS/cm
84 µS/cm
0-2,000 µS/cm
1,413 µS/cm
0.00-20.00 mS/cm
12.88 mS/cm
0.0-200.0 mS/cm
111.8 mS/cm
Temperature calibration
Refer to section 7.2.1
53
Version 2.1 Update: April 2016
7.2.6 CTZN sensor: Inductive conductivity.
General description
Conductivity: inductive measurement technology; Temperature: NTC thermistor.
Materials
EPDM, PVC, 316L stainless steel (sleeve protecting the temperature sensor); polyurethane sleeve on cable.
Precautions
This measuring technology is not sensitive to fouling. However, ensure that the loop is not obstructed.
Measurement/Interference
When immersing the sensor in the medium to be measured, wait until the sensor's temperature has stabilized before taking the reading. The sensor is not well suited to measuring in the weakest ranges (0-200 µS/cm).
Operating temperature
0° C to 50 °C
Servicing
After each use, rinse the sensor carefully in clean water.
Storage
The sensor should be dried before being stored.
Storage temperature
- 10 °C to + 60 °C
Conductivity calibration
Using a clean sensor, perform a 2-point calibration of the sensor (offset and gain using a standard solution whose conductivity is suited to the average value expected):
Measurement range
Concentration of the standard conductivity solution
0-2,000 µS/cm
1,413 µS/cm
0.00-20.00 mS/cm
12.88 mS/cm
0.-100.0 mS/cm
20.00 mS/cm
Temperature calibration
Refer to section 7.2.1
54
Version 2.1 Update: April 2016
7.2.7 VB5 – MES 5 sensor: Sludge blanket detection – Sludge blanket detection/Suspended Solid/Turbidity.
Description
Optical IR (870 nm) based on IR absorption Temperature: NTC.
Materials
DELRIN, Nickel-plated brass, EPDM ; Polyurethane jacketed cable
Safeway
The optical windows are vulnerable to: chemicals (organic solvents, acids and strong bases, peroxide, hydrocarbons), mechanical treatments (impact, abrasion).
Measure/ Interfering
Bubbles on optical parts can interfere with the measurement. On environment change, wait sensor’s temperature stabilization before proceeding with measurement.
Operating temperature
0°C to 50°C
Maintenance
After each use, rinse meticulously the sensor with clear water. If deposits like biofilm or mud persist, clean the sensor with soapy water and wipe the head with a soft cloth or an absorbent paper.
Storage
Place the protection case on the head of the sensor with a bottom of water to maintain the hydrated optical windows.
Temperature of storage
- 10°C to + 60°C
Turbidity calibration in FAU
On a clean sensor, check once in a while the 0 NTU value by dipping sensor in bubble free clear water. If the 0 point is shifted, proceed with the complete sensor calibration (on 1 or 4 ranges).
For this procedure, a Formazin solution, with concentration matching the middle of the measurement range, will be necessary. This solution will be prepared from a 4000 NTU main solution.
For the preparation of solutions, take a flask of 200 mL. Introduce the necessary volume of Formazin (cf. table below) and fill up to 200 mL with distilled water. The formazin solutions of concentrations lower at 1000 NTU deteriorate quickly, so do not preserve a solution during several days. The solution at 2000 NTU can be preserve in the refrigerator for 2 or 3 weeks in a opaque flask.
Measurement range
Concentration Formazin solution
Volume of Formazin (mL)
0.0-50.0 FAU
25 NTU
1,25 mL
0.0-200.0 FAU
100 NTU
5 mL
0-1000 FAU
500 NTU
25 mL
0-4000 FAU
2000 NTU
100 mL
Suspended Solid calibration in g/L
Turbidity in g / L, it is necessary to calibrate the sensor on a real sample. The calibration is achieved in 2 steps :
- Step 1 (offset) : immerse the sensor in distilled water (0 mg / L),
- Step 2 (slope) : immerse the sensor into a sample of sludge, maintained under agitation, and validate the theoretical value measured by the sensor. Analysis the sample dry weight in the laboratory according to the NF standard T 90 105 2.
Sludge blanket detection in %
For a use of the sensor in mode Sludge Blanket detection the sensor is tested on 1 point:
- 1 offset with some distilled water (100 %)
Temperature calibration
Refer to section 7.2.1
55
Version 2.1 Update: April 2016
8. Troubleshooting
Fault
Resolution
Sensor not recognized, Short lines displayed instead of the measurement
- Make sure that the sensor has been connected up correctly.
- Run a SCAN of the network of sensors (refer to section 5.2).
- Disconnect the sensor and connect it to the other digital input dedicated to the sensor.
- Contact the technical services department.
Unstable measurements
- Check that the sensor has been installed under the correct conditions.
- Clean the sensor and check the measuring elements (active disk for the OPTOD sensor, cartridge/glass bulb for the PHEHT sensor, etc.)
- Run a calibration of the sensor.
- Contact the technical services department.
Measurement displayed as 9999
Out-of-range measurement:
- Check that the sensor has been installed under the correct conditions.
- Clean the sensor.
- Run a calibration of the sensor.
- Contact the technical services department.
Current output not working, or value incorrect
- Check the settings for the current output.
- Check that the wiring is correct.
- Check that the device is not in Maintenance Mode.
- Disconnect the faulty analog output and connect it to the other analog output.
- Contact the technical services department.
No relay activation
- Check the settings for the relay output.
- Check that the wiring is correct.
- Check that the device is not in Maintenance Mode.
- Disconnect the faulty relay output and connect it to the other relay output.
- Contact the technical services department.
Frozen screen
Press the RESET button on the connection hardware part of the circuit card (refer to item A in diagram 3).
56
Version 2.1 Update: April 2016
PONSEL Mesure After-Sales Service
AQUALABO CONTROLE – PONSEL MESURE 35 Rue Michel MARION 56850 CAUDAN FRANCE
Tél : +33 (0)1 72 87 97 93 Fax: +33 (0)2 97 76 55 72 Email : sav.ponsel@aqualabo.fr
Loading...