PolyScience WB05A11B, WB05A12B, WB28A11B, WB02A11B, WB10A11B Operator's Manual

...
110-031 PSC/EN
Operator’s Manual
General Purpose Water Baths
110-031 PSC/EN 25 February 2014
1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
110-031 PSC/EN
Table of Contents
Introduction ........................................................................................... 2
General Safety Information .............................................................. 2
Safety Recommendations ................................................................ 3
Regulatory Compliance & Testing .................................................... 4
Unpacking Your Water Bath ............................................................. 4
Contents ........................................................................................... 4
LCD Display ..................................................................................... 5
Keypad ............................................................................................. 5
Installation and Startup ......................................................................... 6
General Site Requirements ............................................................... 6
Adding Liquid to the Reservoir .......................................................... 6
Electrical Power ................................................................................ 8
Controller Setup ................................................................................ 8
Power On .......................................................................................... 8
Selecting the Temperature Unit ........................................................ 8
Setting the High Limit Value .............................................................. 9
Setting the Temperature Presets ...................................................... 9
Normal Operation .................................................................................10
Keypad Controls ............................................................................... 10
Turning Your Water Bath ON ............................................................ 10
Adjusting the Set Point Temperature ................................................10
Setting and Using the Timer ............................................................. 11
Using the Temperature Presets ........................................................11
Entering a Calibration Offset ............................................................12
Resetting the Factory Default Values ............................................... 13
Loss of Power Restart ...................................................................... 13
Displaying the Firmware Version ......................................................13
Display Messages and Alarms ............................................................14
Routine Maintenance & Troubleshooting ...........................................15
Draining the Bath Reservoir ............................................................. 15
Resetting the Over Temperature / Low Liquid Level Safety .............15
Cleaning the Bath ............................................................................16
Troubleshooting Chart ......................................................................16
Technical Information ..........................................................................17
Performance Specications .............................................................17
Reservoir Fluids ...............................................................................18
Application Notes .............................................................................19
Equipment Disposal (WEEE Directive) ............................................... 20
Service and Technical Support ...........................................................21
Warranty ................................................................................................21
1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
110-031 PSC/EN
Introduction
Thank you for choosing this Water Bath. It is designed to handle a wide range of laboratory procedures including: incubation, inactivation and ag­glutination as well as most pharmaceutical, serological, biomedical and industrial procedures.
To ensure optimium temperature uniformity, your Water Bath features
an energy efcient, large-area heater and thermostatic control. Its PID
microprocessor control (Proportional Integral Derivative) system provides proportional heat control by anticipating the approach to your set point temperature and preventing overshoot. A redundant safety thermostat is standard on all models.
General Safety Information
When installed, operated and maintained according to the directions in this manual and common safety procedures, your Water Bath should provide safe and reliable temperature control. Please ensure that all individuals involved in the installation, operation or maintenance of this Water Bath read this manual thoroughly prior to working with the unit.
This symbol marks chapters and sections of this instruction manual that are particularly relevant to safety. When attached
!
to the unit, this symbol draws attention to the relevant section of the instruction manual.
!
This symbol indicates that hazardous voltages may be present.
This symbol indicates that hot surfaces or liquids may be present.
This symbol marks information that is particularly important.
This symbol indicates alternating current.
These symbols on the Power Switch indicate that they place the main power supply ON/OFF.
2
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
110-031 PSC/EN
This symbol on the Power Button indicates that it places the unit in a standby mode. IT DOES NOT fully disconnect the unit from the power supply.
This symbol indicates a protective conductor terminal.
Read all instructions pertaining to safety, set-
up and operation. Proper operation is the user’s
responsibility.
Safety Recommendations
To prevent injury to personnel and/or damage to property, always follow your workplace’s safety procedures when operating this equipment. You should also comply with the following safety recommendations:
WARNING:
Always connect the power cord on this Water Bath to a grounded (3-prong) power outlet. Make certain that the outlet is the same voltage and frequency as your unit.
Never operate the Water Bath with a damaged power cord.
Always turn the Water Bath Off and disconnect mains power before performing any maintenance or service.
WARNING:
• Never operate the Water Bath without bath uid in the
reservoir. Periodically check the reservoir to ensure that
the liquid depth is within acceptable levels. Always rell the reservoir using the same bath uid type that is already
in the reservoir. Bath oil must not contain any water contaminants and should be preheated to the actual bath
!
temperature before adding as there is an explosion hazard at high temperatures.
• Use compatible bath uids only.
• Always drain all uid from the reservoir before moving
or lifting your Water Bath. Be sure to follow your organization’s procedures and practices regarding the safe lifting and relocation of heavy objects.
WARNING: Always allow the bath uid to cool to ambient
temperature before draining.
WARNING: It is the user’s responsibility to properly decontaminate the unit in the event hazardous materials are spilled on exterior or interior surfaces. Consult manufacturer
!
!
Find Quality Products Online at: sales@GlobalTestSupply.com
if there is any doubt regarding the compatibility of decontamination or cleaning agents.
CAUTION: If the equipment is used in a manner not
specied by the manufacturer, the protection provided by the
equipment may be impaired.
3
www.GlobalTestSupply.com
110-031 PSC/EN
Regulatory Compliance & Testing
This product has been tested to the requirements of CAN/CSA-C22.2 No. 61010-1, second edition, including Amendement 1, or a later version of the same standard incorporating the same level of testing requirements.
ETL (60 Hz units)
UL 61010-1 / CSA C22.2 No. 61010-1, 2nd Edition 07/12/04 (R 10/28/2008) IEC 61010-2-010: 2003, CSA C22.2 No. 61010-2-010:2004/07/01 Ed:2 (R2009)
CE (50 Hz units)
Low Voltage Directive 2006/95/EC EN 61010-1, 2nd Edition Issued March, 2001 EN 61010-2-010 2nd Edition Issued October, 2003
EMC Directive 2004/108/EC EN 61326-1: 2006
Machinery Directive 2006/42/EC Annex I
All Units - RoHS2 2011/65/EU
Unpacking Your Water Bath
Your water bath is shipped in a special carton. Retain the carton and all packing materials until the unit is completely assembled and working properly. Set up and run the unit immediately to conrm proper operation. Beyond one week, your unit may be warranty repaired, but not replaced. If the unit is damaged or does not operate properly, contact the
transporation company, le a damage claim, then contact the company
where your unit was purchased.
WARNING: Keep unit upright when moving. Be sure to follow
!
your company’s procedures and practices regarding the safe lifting and relocation of heavy objects.
Contents
The items included with your Water Bath are:
Water Bath
Gabled Reservoir Cover Assembly
Sample Tray
Operator’s Manual
Quick-Start Guide
Selection Guide
4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
110-031 PSC/EN
LCD Display
High Limit Warning
set point higher than the High Limit value.
Lit steady when High
Limit value is being
warning condition is
Set Point Indicator
changed; lit steady
Timer Symbol
Lit when timer is being set or
Indicator
Flashes if you
attempt to enter a
set
Heating Symbol
Lit when heater is
Fault / Warning
Lit when fault or
set point is being
active
Symbol
detected
Flashes when
during normal
operation
is in use
Over-Temperature / Low Liquid Symbol
Lit when over-
temperature or low
liquid level fault is
detected
Loss of Power Indicator
Lit when electrical power is restored after power disruption
Calibration Mode Indicators
Lit when unit calibration offset is being set
Version Indicator
Lit when rmware
version ID is being displayed on the lower readout
Temperature Unit Symbol
Shows temperature unit in use when lit
Upper Readout
Displays actual bath temperature
Lower Readout
Displays set point temperature during normal operation; displays operational information when Menu button is pressed
Temperature Resets Indicator
Lit when preset is being set or is in use
Keypad
Power Button
Turns power to
the Controller On
and Off
Timer Button
Used to start
and stop the
timer
Menu Button
Used to select and set various
operational parameters
Selection Buttons
Used to increase or decrease temperature settings and operational parameters
Temperature Preset Buttons
Used to select frequently used
user-dened
set point temperatures
5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
110-031 PSC/EN
Installation and Startup
Your Water Bath is designed to be simple to setup and install. All that is
required is a container for adding water or another suitable uid to the
bath reservoir.
General Site Requirements
Locate your Water Bath on a level surface in an area that is free from drafts and wide ambient temperature variations, such as near heater or air conditioning vents. Do not place it where there are corrosive fumes, excessive moisture, or in excessively dusty areas. Avoid voltage drops by using properly grounded power outlets wired with 14 gauge or larger diameter wire and, if possible, be close to the power distribution panel. the use of extension cords is not recommended; this will reduce the potential for problems caused by low line voltage.
Adding Liquid to the Bath Reservoir
WARNING: Read the safety data sheet for the bath uid
being used carefully before lling reservoir.
WARNING: See Technical Information, Reservoir Fluids
!
(page 18) for a list of compatible liquids.
WARNING: If the proper uid level is not maintained, the unit could possible by damaged (uid level too low) or the bath may overow (uid level too high).
CAUTION: Fumes from acidic solutions may cause corrosion
!
Fill bath so that the liquid level is approximately one inch (2.54 cm) from top when samples are placed in the bath.
To ensure accurate reading of temperature, the uid level should not be
less than 2 inches (5.08 cm) from the bottom of the unit. Operation of the
bath without uid will not damage the heater, but will cause permanent
discoloration of the tank and will not provide accurate temperature information. Distilled water is preferred for temperatures from 10° to 90°C (50° to
194°F). A variety of uids can be used with the bath depending on the application. The uid must be compatible with 300 series stainless
steel. See Technical Information, Reservoir Fluids (page 18) for a list of
compatible uids.
If using water, a few drops of polyclean Algaecide (part number 004-
300040) should be used to help prevent algae formation.
of the stainless steel reservoir. Care should be taken to maintain a neutral pH at all times.
6
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...
+ 15 hidden pages