PME miniDOT User Manual

Page 1
Precision Measurement Engineering, Inc. • www.pme.com
User’s Manual
2018
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
1
Page 2
1-YEAR LIMITED WARRANTY ON miniDOT HARDWARE
Precision Measurement Engineering, Inc. (PME) warrant that the miniDOT Logger shall be free of defects in workmanship and materials, under normal use, for a period of one year from the date of shipment. This warranty is made only to the original purchaser. In the event a miniDOT Logger covered by this warranty fails to operate according to our published specifications, then return it freight pre-paid to PME or an authorized Service Provider. PME will repair the unit at no charge to the customer and bear the cost of return shipment. Carefully pack all components, as the customer is responsible for any freight damage.
This warranty does not apply to services or consumable / expendable items (such as batteries, fuses and ropes) required for general maintenance. Equipment manufactured by other companies (such as meteorology sensors, solar panels, etc) are warranted only to the limit of the warranties provided by their original manufacturer.
PME makes no warranty, either expressed or implied, that the sensors will be operable after they are exposed to adverse environmental conditions, such as bio-fouling, oil fouling, freezing temperatures or others.
This warranty is void if, in our opinion, the miniDOT Logger has been damaged by accident, mishandled, altered, or repaired by the customer, where such treatment has affected its performance or reliability. In the event of such treatment by the customer, costs for repairs plus two-way freight costs (no COD shipments will be accepted) will be borne by the customer. In such cases, an estimate will be submitted for approval before repair work is started. Items found to be defective should be returned to PME carefully packed, as the customer will be responsible for freight damage.
Incidental or consequential damages or costs incurred as a result of the product malfunction are not the responsibility of PME.
For all warranty or non-warranty returns please obtain, complete, and submit a RMA to PME. This RMA form may be obtained at
http://www.pme.com/HTML%20Docs/RMAform.html.
After submission of this form, then PME will respond with a RMA number. Please place this number on all shipments and related communications.
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
2
Page 3
Safety Information
BURSTING HAZARD
Should water enter the miniDOT Logger and come into contact with the enclosed batteries, then the batteries may generate gas, causing the internal pressure to increase. This gas will likely exit via the same location where the water entered, but not necessarily. The miniDOT Logger is designed to release internal pressure, as the black end cap is unscrewed, prior to the disengagement of the black end cap threads. If internal pressure is suspected, then treat the miniDOT Logger with extreme caution.
Revision History
Date Revision Description 2014JUN09 Initial Document, revised from 6881 2014JUL01 Extensively Revised 2014JUL16 Added Storage When Not in Use to sec 1.2 2014SEP24 Added pictures, reworded. Removed wire mesh installation instructions 2014NOV14 Added update Java instructions. Removed statement saying alkaline batteries
supplied. 2014DEC17 Added 1 green flash for sample intervals of 1 minute or less 2015JAN09 Minor revision 2015JAN23 Added note about miniDOTPlot large sample sets 2015FEB03 Modified battery endurance, modified wording, added # files warning 2015APR15 Added file upload to Android phone instructions 2015SEP11 Added the Cu mesh / plate instructions 2015SEP11 Updated wording to be consistent throughout manual 2016APR13 Expanded description of Android use 2017FEB28 Revised recalibration recommendation from 1/2 million samples to yearly. 2018JAN08 Corrected "oxygen saturation" to be "air saturation" 2018JAN25 Changed wording 2018MAR29 Updated for 16GB SD cards, added maintenance section, changed battery
section
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
3
Page 4
CONTENTS
Chapter 1 Quick Start
1.1 The Quickest Start Possible
1.2 A Few Details
1.3 Overview General Maintenance
Chapter 2 Software
2.1 Overview and Software Installation
2.2 miniDOTControl
2.3 miniDOTPlot
2.4 miniDOTConcatenate
Chapter 3 miniDOT Logger
3.1 Overview
3.2 Opening and Closing the miniDOT Logger
3.3 Electrical Connections and Controls
3.4 Battery Replacement
3.5 Installing Copper Mesh or Plate
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
4
Page 5
Chapter 1: Quick Start
1.1 The Quickest Start Possible
Your miniDOT Logger has arrived completely ready to go. It is set to measure and record time, battery voltage, temperature, oxygen concentration, and measurement quality once every 10 minutes and write one file of measurements daily. Open the miniDOT Logger and move the Logger Control Switch to the “Record” position. In this condition, the miniDOT Logger will record measurements for a year before the internal batteries are expended. You must re-close the miniDOT Logger prior to deploying it.
At the end of the deployment period, open the miniDOT Logger and connect it to a HOST computer via the USB connection. The miniDOT Logger will appear as a ‘thumb drive’. Your temperature and oxygen concentration measurements, together with a time stamp indicating the time the measurements were made, are recorded in text files in the folder having the serial number of your miniDOT Logger. These files can be copied onto any Windows or Mac HOST computer.
This Manual and other software programs are also recorded on the miniDOT Logger. miniDOTControl program allows you to see the state of the miniDOT Logger as well as
set the recording interval.
miniDOTPlot program allows you to see the plots of the recorded measurements.  miniDOTConcatenate program gathers all the daily files into one CAT.txt file.
Your miniDOT Logger will return to recording measurements after you disconnect the USB connection. If you wish to stop recording, then move the Logger Control Switch to the “Halt” position.
You may move the Logger Control Switch at any time.
Follow these steps to start the deployment, logging DO & T once each 10 minutes:
1) Open the miniDOT Logger by unscrewing the white pressure housing from the black end cap. It opens like a flashlight. Remove the white pressure housing completely. Inside you will see the circuit pictured below:
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
5
Page 6
2) Move the Logger Control Switch to the “Record” position. The LED will flash green 5 times. The miniDOT Logger will now record a measurement of time, battery voltage, temperature, and dissolved oxygen every 10 minutes (or at some other interval you may have set using the miniDOTControl program).
3) Inspect the o-ring seal for debris.
4) Close the miniDOT Logger by screwing the white pressure housing back onto the black end cap.
5) Deploy the miniDOT Logger.
Follow these steps to end the deployment
1) Recover the miniDOT Logger
2) Clean and dry all accessible surfaces except the ‘sensing foil’.
3) Open the miniDOT Logger by unscrewing the white pressure housing from the black end cap. Remove the white pressure housing completely, taking care that water does not drip onto interior surfaces of circuits or other items inside the miniDOT Logger.
4) Connect to a Windows HOST computer via the USB connection. The miniDOT Logger will appear as a ‘thumb drive’.
5) Copy the folder having the same serial number as the miniDOT Logger (example 7450-
0001) to the HOST computer.
6) (Suggested, but optional) Delete the measurement folder, but NOT the miniDOTControl program or the other .jar programs.
7) (Optional) Run the miniDOTControl program to see the state of the miniDOT Logger such as battery voltage or to select a different recording interval.
8) (Optional) Run the miniDOTPLOT program to see a plot of measurements.
9) (Optional) Run the miniDOTConcatenate program to gather together all the daily files of measurements into one CAT.txt file.
10) If no more recording is desired, then move the Logger Control Switch to “Halt”, otherwise leave it set to “Record” to continue recording measurements.
11) Disconnect the miniDOT Logger from the USB connection.
12) Inspect the o-ring seal for debris.
13) Close the miniDOT Logger by screwing the white pressure housing back onto the black end cap.
Remove the batteries if storing the miniDOT Logger for extended periods.
1.2 A Few Details
The previous section gives instructions for sampling at 10-minute intervals. However, there are a few additional details that will enhance use of the miniDOT Logger.
Recording Interval – The miniDOT Logger measures and records time, battery voltage, temperature, dissolved oxygen concentration and measurement quality at equal time intervals. The default time interval is 10 minutes. However, it is also possible to instruct the
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
6
Page 7
miniDOT Logger to record at different intervals. This is accomplished by running the miniDOTControl.jar program supplied with the miniDOT Logger. Recording intervals must be 1 or more minutes and must be less than or equal to 60 minutes. Intervals outside this range will be rejected by the miniDOTControl program. (Contact PME for other recording intervals.)
Please refer to Chapter 2 for instructions on operating the miniDOTControl program.
Time – All miniDOT Logger times are UTC (formerly known as Greenwich mean time (GMT)). The miniDOT Logger internal clock will drift in the <10 ppm range (< about 30 seconds/month) so you should plan to connect it occasionally to a HOST computer having an Internet connection. The miniDOTControl program will automatically set time based on an Internet time server.
Please refer to Chapter 2 for instructions on operating the miniDOTControl program.
File Information – The miniDOT Logger software creates 1 file daily on the miniDOT Logger’s internal SD card. The number of measurements in each file will depend upon the sample interval. Files are named by the time of the first measurement within the file based on the miniDOT Logger’s internal clock and expressed in YYYY-MM-DD HHMMSSZ.txt format. For example a file having the first measurement on September 9, 2014 at 17:39:00 UTC will be named
2014-09-09 173900Z.txt.
Files can be uploaded from the miniDOT Logger by connecting it to a HOST computer. Use the HOST computer’s copy/paste functions to move the files from the miniDOT Logger to the HOST computer.
Each measurement within the files has a time stamp. The time stamp format is Unix Epoch 1970, the number of seconds that have passed since the first moment of 1970. This may be inconvenient in some cases. If so, then the miniDOTConcatenate program not only concatenates all the measurement files, but also adds more readable statements of the time stamp.
Please refer to Chapter 2 for instructions on operating the miniDOTConcatenate program.
The miniDOT Logger requires time and battery energy to work through the file directory on the SD card to allocate new file space. A few hundred files on the SD card is not a problem, but as the number of files grows large into the thousands then the miniDOT Logger may suffer decreased battery life or other performance problems. Please, at the earliest convenient time, copy the recorded files to a HOST computer and delete them from miniDOT Logger’s SD card. Also, do not use miniDOT Logger to store files unrelated to the miniDOT Logger’s operation.
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
7
Page 8
Control miniDOT and Upload Files to an Android device – It is possible to control miniDOT and to transfer your measurement files from the miniDOT Logger to your Android device. This operation has 3 prerequisites.
1. An Android device equipped with USB On-The-Go (OTG) feature and one that can act as a USB host. OTG is a feature for most modern Android devices that allow them to act as a USB host. OTG was introduced with Android version 3.1 and should be a useable feature for most devices. It is still possible that your device doesn't support OTG or that it does support OTG but will either not see miniDOT as a local drive or will not control miniDOT. Please test your device with a miniDOT Logger prior to attempting to upload while in the field.
2. A USB OTG cable. This is essentially an adapter with a micro USB connector on one end that connects to your Android device and a standard A type USB plug on the other. The cable supplied with the miniDOT Logger connects to this end.
3. A "File Explorer" app downloaded on your Android device. These types of apps assist with being able to view and move files from device to device. "ES File Explorer" is a free app available on the Google app store and is recommended for use with the miniDOT Logger. Samsung products have their own file explorer called "MyFiles".
If the 3 requirements are met, then connect the miniDOT Logger to your Android device. You should see the miniDOT Logger’s green connection light come on and your Android device will display a message about "preparing USB storage". When "USB storage ready" is shown, then use your "file explorer" program to move your data from the miniDOT Logger onto your Android device. You will later use the device to move the file to a new location via internet or possibly move the file onto plugged SD card storage.
miniDOT can be controlled by using a PME app. This can be downloaded from Google Play Store. Search for "pmecontrolapp". Download. The app is free. After the app is downloaded just plug miniDOT into the device. This will bring up the app. There may be some confusion at plug in time since the app comes up but also the system file explorer may appear. The app will appear requesting a sample interval be entered. If not entering an interval touch Done to make the entry pad disappear. The app will scold you for not entering a valid sample interval. Ignore this. The sample interval within miniDOT will be unchanged and appears in the app.
Again, it is highly recommended that you try these operations prior to depending upon this in the field. Not all devices that support OTG will communicate with miniDOT.
Currently the associated miniDOT Logger programs miniDOTPlot, and miniDOTConcatenate are not supported on the Android devices.
1.3 Overview General Maintenance
Cleaning the Sensing Foil –The sensing foil can be cleaned at regular intervals depending
on the fouling condition at the site. The cleaning procedure of the sensing foil should be done with caution so that the protective coating is not removed. If the fouling is calcareous it can normally be dissolved with household vinegar.
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
8
Page 9
If the marine growth remains, then use Q-tips to gently wipe the sensing foil after it has been softened by soaking in vinegar or perhaps diluted HCl. After cleaning the sensing foil, then it should be rinsed well in clean tap water before storing or reuse. Do not use other organic solvents such as acetone, chloroform, and toluene since these and others will damage the sensing foil.
The sensing foil can also be cleaned using a 3% H2O2 solution or rinsing it with ethanol.
The white pressure housing and black end cap can be gently scrubbed.
AA Alkaline Battery Life – Alkaline batteries will give somewhat less performance than lithium, especially at low temperatures. Alkaline batteries are superior to lithium in one way: you can determine how much battery life remains by measuring the battery terminal voltage. For short deployments of a month or two, then alkaline batteries will provide adequate performance. For longer deployments, or for deployments in cold environments, then substitute lithium batteries.
AA Lithium Battery Life – The miniDOT Logger consumes battery power mostly from the measurement of dissolved oxygen, but also slightly from simply keeping track of time, writing files, sleeping, and other activities. The following table shows the approximate endurance of the miniDOT Logger when powered by the Energizer L91 AA lithium / ferrous disulfide batteries:
SAMPLE INTERVAL
(minutes)
Main AA Battery Life
(months)
Number of Samples
1 12 500K
10 >12 >52,000
60 >12 >8,000
Keep a general record of the miniDOT Logger’s number of samples. It is not possible to accurately tell the charge state of a lithium battery by measuring its terminal voltage. If you have a general idea of the number of samples already obtained on a battery, then you can make a guess as to how many more samples remain.
The numbers in the table above are, at the time of this writing, based upon extrapolations of testing of 500K samples acquired at a 5-second interval. The 1-year performance at 1 minute is very likely. Performance at longer sample intervals will be much longer, but how long is difficult to predict. In any event, these AA batteries are easily available and relatively inexpensive compared to the cost of the miniDOT Logger. PME suggests you replace the batteries often, especially before any long (months) measurement deployments.
Monitor battery voltage in the miniDOTControl program. You cannot tell from terminal voltage of a lithium battery how long the battery will last, but you can tell if it will die soon.
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
9
Page 10
The Low Drain Performance plot below gives an estimate of terminal voltage for both lithium and alkaline batteries. You can operate batteries down to about 2.4 Volts (for two in series,
1.2 Volts on the graph below). Remove the batteries and measure each of them. If your
combined battery voltage is less than 2.4 Volts, replace the batteries.
You may also use alkaline AA batteries such as Duracell Coppertop. They will not last nearly as long, especially at low temperatures, but will likely be adequate for several weeks at 10­minute interval.
When replacing batteries use only fresh batteries. Don’t mix battery types. If one battery differs in type or charge level from the other and the miniDOT Logger runs them to full discharge, then one battery may leak. SEE SECTION 3.4 FOR CAUTION ON BATTERY
PLACEMENT.
Err on the side of caution when planning your deployment.
The recommended battery is the Energizer L91 lithium battery. For more information including the performance at low temperatures, then click the link:
http://data.energizer.com/PDFs/l91.pdf
The figure below gives a general idea of terminal voltage vs. lifetime. Service life in hours is incorrect since miniDOT Logger draws far less than 50 mA continuously, but the general shape of the voltage vs. time gives an estimate of life remaining. This plot is taken from the manufacturer’s specification. The plot is for a single battery. Double the voltages shown to give the terminal voltage as measured in the picture above. The miniDOT Logger halts operation at a total of 2.4 Volts.
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
10
Page 11
Coin Cell Battery Life – The miniDOT Logger uses a coin cell battery for backup of the clock when the power is switched off. This coin cell battery will supply many years of clock operation. Should the coin cell battery discharge, then it must be replaced by PME. Contact PME.
Recalibration – The miniDOT Logger will maintain its calibration without the necessity of adjustment by the user. The miniDOT Logger should be returned to PME for recalibration. We suggest that this be done yearly.
O-Ring and Seal – When the white pressure housing is screwed on to the black end cap, then it passes along the o-ring located in the black end cap several revolutions. Keep this o­ring lightly lubricated with silicone grease or an oil compatible with buna-N o-ring material.
When the miniDOT Logger is opened after deployment, then a small number of water drops are deposited on the inner surface of the o-ring. When the white pressure housing is screwed back on to the black end cap, then these drops can become trapped inside the miniDOT Logger. Be sure to carefully dry the o-ring and adjacent surfaces (especially underneath) prior to closing the miniDOT Logger. Re-lubricate the o-ring at this time.
LED Indications – The miniDOT Logger indicates its operation with its LED. The table below presents LED indications:
LED Reason 1 Green Flash 1 Green Flash 5 Green Flashes
Normal. Presented immediately after new batteries are installed. Indicates that the CPU has started its program. Occurs at the time of sampling for sample intervals of 1 minute or less. Normal. Indicates that miniDOT Logger is beginning to record measurements. This indication appears in response to
switching the Logger Control Switch to “Record.” 5 Red Flashes
Normal. Indicates the miniDOT Logger is ending recording of
measurements. This indication appears in response to
switching the Logger Control Switch to “Halt.” Continuously Green Continuously Flashing Red
Normal. Indicates the miniDOT Logger is connected to a HOST
computer via the USB connection.
SD card write error. Try removing/reinstalling batteries.
Contact PME.
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
11
Page 12
Verifying Calibration – You may from time to time want to verify the calibration of your miniDOT Logger. Do this by placing the miniDOT Logger in a black 5-gallon bucket containing 4 gallons of fresh water. (The picture below shows a white bucket so that the miniDOT Loggers are more easily seen.) The miniDOT Logger’s black end cap is heavy and the miniDOT Logger will tend to flip so that this end is down. Prevent this somehow. The miniDOT Logger must be placed in the bucket with the black end cap upwards. Otherwise bubbles will accumulate in the black end cap area and miniDOT Logger will not sense the DO in the water correctly. Use an aquarium pump and air stone in the water to provide a bubble stream. Cover the bucket with a black lid. The idea is to prevent light from enabling algal growth.
Record measurements for several hours or a day, but in any event long enough for the miniDOT Logger’s temperature to come to equilibrium with the water. During the experiment, find the local air pressure, either from measurements or from a local weather station. Watch out... weather stations often report barometric pressure referenced to sea level. You must determine the absolute barometric pressure at your elevation.
A more comprehensive experiment is to additionally place ice in the bucket and mix until the water temperature is close to zero degrees. Next, remove the ice. Place the bucket on a towel or piece of cardboard and cover the top of the bucket with a towel. Record for 24 hours as the bucket temperature gradually returns to room temperature.
After recording the bubbled water, you may also remove the air stone and gently mix a packet of baker’s yeast into the bucket together with a tablespoon of sugar. The water must be only slightly warm to the touch but not more than 30 deg C. These organisms will deplete all the dissolved oxygen in the water. Cut a disc of thin plastic film just large enough to lay on top of the water. Place this on top of the water. Do not stir or bubble after placing the film. Record measurements for at least an hour or more.
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
12
Page 13
Use the miniDOT Logger’s miniDOTPlot program to examine the measurements. Saturation values should be very close to 100%, depending upon the accuracy that you have determined barometric pressure. If you placed ice in the bucket, then saturation values will still be 100%. You will see the DO concentration and temperature change greatly as the bucket warms.
The recorded data, when using yeast should show 0% saturation and 0 mg/l dissolved oxygen concentration. In practice the miniDOT Logger often reports slightly positive values of about 0.1 mg/l, but within the accuracy of the miniDOT Logger.
Closing and Opening – Close and open the miniDOT Logger like you would a flashlight; open by unscrewing the white pressure housing from the black end cap. Close by screwing the white pressuring housing on to the black end cap. When closing, do not tighten the white pressure housing. Just screw it on until it makes contact with the black end cap. See Chapter 3 for more instructions.
Caution: DO NOT remove the stainless steel screws in the black end cap. There are no user serviceable parts here. If the screws are removed, then you will damage the miniDOT Logger and it will have to be returned for repair.
Storage When Not in Use – Remove the batteries. Keep the black end covered with the cap supplied by PME. If the cap is lost, then cover the black end cap with aluminum foil. There may be a calibration effect of ambient lighting so attempt to keep ambient light from reaching the sensing foil as much as possible.
Java – miniDOT programs depend on Java and require Java 1.7 or higher. Update Java at
https://java.com/en/download/index.jsp
Chapter 2: Software
2.1 Overview and Software Installation
The miniDOT Logger arrives with these files on the SD card:
miniDOTControl.jar program allows you to see the state of the miniDOT Logger as well
as set the recording interval.
miniDOTPlot.jar program allows you to see the plots of the recorded measurements.  miniDOTConcatenate.jar program gathers all the daily files into one CAT.txt file.  Manual.pdf is the manual.
These files are located on the root directory of the miniDOT Logger.
PME suggests you leave these programs where they are on the miniDOT Logger, but you may copy them to any folder on your HOST computer’s hard drive.
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
13
Page 14
miniDOTControl, miniDOTPlot, and miniDOTConcatenate programs are Java language programs that require the HOST computer to have the Java Runtime Engine V1.7 (JRE) or later versions installed. This engine is commonly required for Internet applications and will likely already be installed on the HOST computer. You can test this by running the miniDOTPlot program. If this program displays its graphical user interface, then the JRE is installed. If not, then the JRE can be downloaded via the Internet from
http://www.java.com/en/download/windows_xpi.jsp
At this time the miniDOT Logger is supported on the Windows operating systems, but may also operate on Macintosh and perhaps Linux.
2.2 miniDOTControl
Begin program operation by clicking “miniDOTControl.jar”. The program presents the screen shown below:
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
14
Page 15
The miniDOT Logger must be connected to the HOST computer via the USB connection at this time. When correctly connected, the miniDOT Logger’s LED will display a constant green light.
Click the “Connect” button. The program will contact the miniDOT Logger. If the connection is successful, then the button will turn green and display “Connected”. The Serial Number and other parameters will be filled in from the information taken from the miniDOT Logger.
If the HOST computer is connected to the Internet, then the current difference between an Internet time server’s time and the miniDOT Logger’s internal clock will be displayed. If more than a week has passed since time was last set, then the miniDOT Logger’s clock will be set and the checkmark icon will appear. If the HOST computer is not connected to the Internet, then no time services will occur.
The current miniDOT Logger’s sample interval will be displayed next to the “Set Sample Interval” button.
To set the interval, enter an interval not less than 1 minute and not greater than 60 minutes. Click the “Set Sample Interval” button. Shorter and faster intervals are available. Contact PME.
If this interval is acceptable, then the interval need not be set.
End the miniDOTControl program by closing the window. Disconnect the miniDOT Logger’s USB cable.
Upon disconnection of the USB cable, the miniDOT Logger will begin logging or remain halted as indicated by the position of the Logger Control Switch.
2.3 miniDOTPlot
Begin the program operation by clicking “miniDOTPlot.jar”. The program presents the screen shown below.
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
15
Page 16
The miniDOTPlot program plots the files recorded by the miniDOT Logger. The program reads all the miniDOT Logger files in a folder, except the CAT.txt file. The program will also compute air saturation from the dissolved oxygen measurements. To do this, the program must know the air pressure and salinity. It calculates air pressure based on elevation of the water surface above sea level or uses the barometric pressure you enter if Barometric Pressure is selected. If the Surface Elevation is entered, then no compensation for weather­induced barometric pressure variation is made. Enter elevation or barometric pressure. Enter water salinity.
Select the folder that contains the files recorded by the miniDOT Logger. If the miniDOTPlot program is run directly from the miniDOT Logger, then the program will suggest the folder that is located on the miniDOT Logger’s SD card. You may accept this by clicking “Plot”, or you may click “Select DATA Folder” to browse to your HOST computer’s hard drive. If the number of measurements recorded is small, for example a few thousand, then these can conveniently be plotted directly from the miniDOT Logger’s storage. However, it is best to copy large measurement sets to the HOST computer and select them there. The file access to the miniDOT Logger is slow.
The miniDOT Logger’s measurement folders must NOT contain any files besides those the miniDOT Logger recorded and the CAT.txt file.
Click “Plot” to begin plotting.
The program reads all the miniDOT Logger’s data files in the selected folder. It concatenates these and presents the plot shown below
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
16
Page 17
You may zoom this plot by drawing a square from upper left to lower right (click and hold left mouse button) that defines the zoom region. To zoom completely out, attempt to draw a square from lower right to upper left. Right click on the plot for options such as copy and print. The plot can be scrolled with the mouse while the Control key is held depressed. Copies of the plot can be obtained by right clicking on the plot and selecting Copy from the pop-up menu.
Different DATA Folders can be selected during one session of the program. In this case the software produces multiple plots. Unfortunately, the plots are presented exactly on top of one other and so when a new plot appears it is not obvious that the old plot is still there. It is. Just move the new plot to see previous plots.
The program can be re-run at any time. If an already processed DATA Folder is selected, then the progrm simply reads the miniDOT Logger’s measurement files again.
End the miniDOTPlot program by closing the window.
Special note: plotting of sample sets of more than 200K samples may consume all the memory available to the JRE. The miniDOTPlot program will present a partial plot and freeze in this case. A simple solution is to separate the files into multiple folders and plot each folder individually. A special miniDOTPlot that sub-samples can be provided by PME. Please contact PME in this case.
2.4 miniDOTConcatenate
Begin the program operation by clicking “miniDOTConcatenate.jar”. The program presents the screen shown below.
The miniDOTConcatenate program reads and concatenates the files recorded by the miniDOT Logger. This program produces a CAT.txt file in the same folder as selected for the data. The CAT.txt file contains all the original measurements and contains two additional statements of time and air saturation. To compute saturation, the program must know the air pressure and salinity. It calculates air pressure based on elevation of the water surface above sea level or uses the barometric pressure you entered if the Barometric Pressure was
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
17
Page 18
selected. If Surface Elevation is entered, then no compensation for weather-induced barometric pressure variation is made. Enter elevation or barometric pressure. Enter water salinity.
Select the folder that contains the files recorded by the miniDOT Logger. If the miniDOTPlot program is run directly from the miniDOT Logger, then the program will suggest the folder located on the miniDOT Logger. You may accept this by clicking “Concatenate”, or you may click “Select DATA Folder” to browse your HOST computer’s hard drive. If the number of measurements recorded is small, for example a few thousand, then these can conveniently be plotted directly from the miniDOT Logger’s storage. However it is best to copy large measurement sets to the HOST computer and select them there. The file access to the miniDOT Logger is slow.
The miniDOT Logger measurement folders must NOT contain any files besides those the miniDOT Logger recorded and the CAT.txt file.
Click “Concatenate” to begin concatenating files and create the CAT.txt file.
The CAT.txt file will resemble the following:
End the miniDOTConcatenate program by closing the window.
Chapter 3: miniDOT Logger
3.1 Overview
All miniDOT Logger measurements are saved in files on the SD card inside the miniDOT Logger. The files are transferred to a HOST computer via a USB connection where the miniDOT Logger appears as a “thumb drive”. Measurements may be plotted by the miniDOTPlot program and files concatenated by the miniDOTConcatenate program. The miniDOT Logger itself is controlled by the miniDOTControl program. Customers are required to open the logger each time measurements are transferred to the HOST computer. This chapter describes the miniDOT Logger’s internal features.
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
18
Page 19
3.2 Opening and Closing the miniDOT Logger
The miniDOT Logger’s circuitry is contained in a waterproof housing that must be opened. Unscrewing the white pressure housing from the black end cap opens the miniDOT Logger. This is similar to opening a flashlight. Turn the white pressure housing counter clockwise relative to the black end cap. Close the miniDOT Logger by reversing this procedure after being sure that the o-ring is free from debris. Lubricate the o-ring occasionally with silicone oil intended for buna-N o-ring material.
Please attempt to handle the miniDOT Logger by only touching the aluminum chassis. Try not to touch the circuit board.
When closing the miniDOT Logger, inspect the o-ring and the interior of the white pressure housing for debris. Lubricate the o-ring, and screw the white pressure housing onto the black end cap until the white pressure housing just touches the black end cap. Do not tighten! The miniDOT Logger tends to get a little tighter during deployment.
If you cannot open the miniDOT Logger by yourself, then find another person with strong hands. This person should grip the black end cap while the other person turns the white pressure housing.
Caution: Do NOT remove the stainless screws in the black end cap. If this is done, then the miniDOT Logger will be permanently damaged and must be returned for repair.
3.3 Electrical Connections and Controls
Removal of the cover reveals the miniDOT Logger’s connections and controls, shown below.
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
19
Page 20
The LED Light is a LED that can display either red or green light. This is used to indicate the different features described in Chapter 1 in this manual.
The Logger Control Switch controls the miniDOT Logger’s mode:
Record – When the switch is in this position the miniDOT Logger is recording
measurements.
Halt – When the switch is in this position the miniDOT Logger is not recording and is
sleeping at low power.
The USB Connection allows communication between the miniDOT Logger and an external HOST computer. When connected, the miniDOT Logger is in HALT mode regardless of the Logger Control Switch position. When disconnected, the miniDOT Logger’s mode is controlled by the Logger Control Switch position. The switch position may be changed while the USB is connected.
NOTE: miniDOT units equipped with a 16GB SD card will take longer to appear as a useable removable disk drive on your computer.
The Main Batteries (2 X AA on side opposite to pictured side above) provide main power to the miniDOT Logger. Note the positive (+) terminal. Batteries are described in Chapter 1 of this manual.
3.4 Battery Replacement
Be sure that the replacement batteries are compatible with the miniDOT Logger. PME recommends Energizer L91 AA size lithium batteries or Duracell AA size alkaline batteries.
http://data.energizer.com/PDFs/l91.pdf
http://ww2.duracell.com/media/en­US/pdf/gtcl/Product_Data_Sheet/NA_DATASHEETS/MN1500_US_CT.pdf
Caution: Improper replacement of the batteries will damage the miniDOT Logger.
Follow these steps:
1) Move the miniDOT Logger’s Control Switch to the “Halt” position.
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
20
Page 21
2) Remove the depleted batteries noting the position of the (+) terminal.
3) Use only new, fully charged batteries, both of the same type.
4) Install fresh batteries with the (+) position the same as the removed batteries. The
(+) position is also marked on the inside of the battery holder.
5) The miniDOT Logger’s LED Light should flash to indicate that the software is
beginning operation within a second or two after you complete the battery installation. At this time the logger will enter the mode selected by the Logger Control Switch (which should initially be “Haltfrom Step 1).
Please be aware that the warranty will become void if the batteries are installed backwards.
3.5 Installing Copper Mesh or Plate
miniDOT Anti-Fouling Copper Kit includes
1 Cu Wire Mesh Disc 1 Cu Plate 1 Nylon Ring 3 Phillips Pan Head Screws
How to install the Cu mesh on a miniDOT Logger:
1. Use a Phillips screwdriver to remove 3 of the 6 screws (every other one). Do NOT remove all screws. At least 3 must remain screwed in at all times.
Save the ss316 screws. They will be needed if the Cu mesh is removed.
:
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
21
Page 22
2. Position the nylon ring under the Cu mesh so that the notches in the nylon ring and Cu mesh align over the screw holes.
3. Install the three pan head screws included in the kit. Gently tighten.
CAUTION: Environments where debris can get trapped inside the sensing area should be avoided when using this product. PME recommends using the Cu plate in such environments.
How to install the Cu plate on a miniDOT Logger:
1. Use a Phillips screwdriver to remove 3 of the 6 screws (every other one). Do NOT remove all screws. At least 3 must remain screwed in at all times.
Save the ss316 screws. They will be needed if the Cu plate is removed.
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
22
Page 23
2. Position the Copper Plate faced down so that the notches in the Copper Plate perfectly align over sensing foil and over the screw holes.
4. Install the three pan head screws included in the kit. Gently tighten.
Enjoy your new miniDOT!
Precision Measurement Engineering, Inc. (760) 727-0300 www.pme.com
23