5.1Selecting general exposure settings ..................................................................................... 13
5.2Patient positioning in lateral projection ................................................................................. 15
5.3Patient positioning in posteroanterior or anteroposterior projection ..................................... 19
5.4Patient positioning in submental-vertex projection ............................................................... 23
5.5Patient positioning in Waters’ view ....................................................................................... 24
5.6Patient positioning in AP Towne projection .......................................................................... 25
5.7Patient positioning in oblique projection ............................................................................... 26
5.8Positioning in hand exposure ...............................................................................................28
5.9Adjusting exposure values for current exposure .................................................................. 29
5.10Selecting Dynamic Exposure Control (DEC) ........................................................................ 31
5.11Taking an exposure .............................................................................................................. 32
User’s ManualPlanmeca ProMax Cephalostat1
The manufacturer, assembler and importer are responsible for the safety, reliability and
performance of the unit only if:
- installation, calibration, modification and repairs are carried out by qualified
authorised personnel
- electrical installations are carried out according to the appropriate requirements
such as IEC 60364
- equipment is used according to the operating instructions.
Planmeca pursues a policy of continual product development. Although every effort is
made to produce up-to-date product documentation this publication should not be
regarded as an infallible guide to current specifications. We reserve the right to make
changes without prior notice.
COPYRIGHT PLANMECA
Publication Number 10033034 Revision 9
Release Date 15 December 2016
2Planmeca ProMax CephalostatUser’s Manual
1 INTRODUCTION
NOTE
NOTE
NOTE
1 INTRODUCTION
This manual describes how to operate the Planmeca
ProMax Cephalostat.
The Planmeca ProMax Cephalostat is used to position the
patient’s head accurately for taking X-ray images of the
skull. The device can also be used for taking hand images
for bone age studies.
The cephalostat is attached to a 2D or 3D Planmeca
ProMax X-ray unit. The X-ray unit has a separate manual.
Refer to the User’s Manual of the X-ray unit for any
general information about the X-ray unit.
This manual is valid for software version 3.8.0.0.r or later.
This software version is compatible with Planmeca
Romexis software version 4.5.0.r or later. To check the
software version of your X-ray unit, select Settings > About
> 4100 Component Information > ProMax SW version.
Make sure that you are fully acquainted with the
appropriate radiation protection measures and these
instructions before you use the X-ray unit.
The X-ray unit may be used by health care professionals
only.
2 SWITCHING X-RAY UNIT ON
The on / off switch is located on the underside of the
stationary column top.
To prolong the lifetime of your X-ray unit, always switch the
X-ray unit off when it is not in active use.
User’s ManualPlanmeca ProMax Cephalostat1
3 MAIN PARTS
Ceph&Dimax2.eps
5
1
2
3
4
7
8
6
3 MAIN PARTS
3.1 General view of cephalostat
1Height adjusting button
2Head support plate
3Rotation scale
4Nasal positioner
5Dimax sensor
6Ear posts
(see section 3.2 “Patient supports” on page 3)
7Release lever on ear post holder (patient’s left ear)
8Collimator
2Planmeca ProMax CephalostatUser’s Manual
3 MAIN PARTS
1
2
3
L
4
5
3.2 Patient supports
Hand_imaging_support.eps
1Ear post for patient’s left ear (marked with letter L)
2Ear post for patient’s right ear (marked with letter R)
3Positioning cones
4Nasal positioner
5Hand imaging support
User’s ManualPlanmeca ProMax Cephalostat3
3 MAIN PARTS
NOTE
NOTE
NOTE
Ceph&Dimax2.eps
3.3 Patient positioning controls
The patient positioning controls are located above the
touch screen.
The up and down buttons are used to adjust the
cephalostat to suit the height of the patient.
The cephalostat moves slowly at first, then faster.
If for some reason either of the buttons gets stuck during
operation, you can stop the up / down movement by
pressing any of the other control buttons or the positioning
joystick. This is a safety measure that guarantees that the
up / down movement can be stopped in an emergency.
There is an additional height adjusting button on the
underside of the cephalostat. The cephalostat moves up
or down depending on which side the button is pressed.
Be careful that the X-ray unit does not hit the ceiling when
you press the up button. The maximum height can be
adjusted to suit offices with a low ceiling, contact your
service technician for help.
Make sure that there is no object under the X-ray unit
when you press the down button. If something is in danger
of becoming trapped, release the button immediately to
stop the movement. Clear any obstruction before pressing
the button again.
4Planmeca ProMax CephalostatUser’s Manual
3 MAIN PARTS
NOTE
NOTE
emergency_stop.eps
The column movement stops automatically if the
emergency stop plate at the bottom is pressed upwards.
Clear any obstruction before moving the column again.
When positioning seated patients (e.g. in a wheelchair)
always first move the cephalostat down before you
position the patient.
3.4 Emergency stop button
The emergency stop button is located on the top of the
stationary column. Press the button to stop the X-ray unit
operating in an emergency. When the emergency stop
button is pressed down, all movements of the X-ray unit
are blocked and the unit will not generate radiation. The
up / down movement will stop within a distance of 10 mm
(0.4 in.).
A help message will appear on the touch screen. Guide
the patient away from the X-ray unit. Then release the
emergency stop button. The X-ray unit will automatically
restart.
User’s ManualPlanmeca ProMax Cephalostat5
4 PREPARATIONS FOR EXPOSURE
NOTE
NOTE
L
R
.
O
L
L
4 PREPARATIONS FOR EXPOSURE
4.1 Preparations on all X-ray units
4.1.1 Attaching cephalometric patient supports
The ear posts are marked with the letter L for the patient’s
left ear and letter R for the patient’s right ear.
The positioning cones contain small metal objects that are
visible in the image. They help you to identify the patient’s
left and right side.
•The left ear post contains a ball.
•The right ear post contains a ring.
Ensure that you position the ear posts the right way round.
The nasal positioner has to point inward as shown.
1.Lift the locking clip up and slide the ear post / nasal
positioner as far up as it will go.
2.Press the locking clip down to lock the ear post /
nasal positioner in place.
6Planmeca ProMax CephalostatUser’s Manual
4 PREPARATIONS FOR EXPOSURE
enter_to_ceph_scr2.eps
Head_support_attach.eps
ProMax_5_2.eps
4.1.2 Removing 2D or 3D patient supports
If the temple supports are in place, turn them down and
remove all patient supports from the patient support table.
If the support bars are in place, remove all patient
supports from the patient support table, including support
bars, chin cup and adapter.
•To remove the support bars, first loosen the locking
knobs and then pull the support bars out.
User’s ManualPlanmeca ProMax Cephalostat7
4 PREPARATIONS FOR EXPOSURE
NOTE
Attach Dimax sensor
to cephalostat
(refer to section 4.2.2)
Remove Dimax sensor
from C-arm
(refer to section 4.2.1)
X-ray unit with
Dimax sensor
Sensor1.eps
4.2 Additional preparations on X-ray units with Dimax sensor
If a Dimax sensor is attached to the C-arm, the sensor
must be moved to the cephalostat before you can take
cephalometric exposures.
4.2.1 Removing sensor from C-arm
8Planmeca ProMax CephalostatUser’s Manual
Do not remove the sensor during imaging process.
1.Push in the C-arm electrical connector. This will
disconnect the electrical connection between the
sensor and C-arm.
4 PREPARATIONS FOR EXPOSURE
CAUTION
Sensor2.eps
Sensor3 .eps
1
2.Turn the locking knob 180 degrees. This will release
the locking mechanism.
3.Carefully pull the sensor out.
Do not drop the sensor. Planmeca limited warranty
does not cover damage which is due to misuse, e.g.
dropping the sensor, neglect, or any cause other than
ordinary use.
Do not use the sensor if the shock indicator (1) is red
- contact your service technician for help.
If you have any reason to believe that the sensor
might be faulty, take a test exposure before taking a
patient exposure.
shockwarningD3.eps
User’s ManualPlanmeca ProMax Cephalostat9
4 PREPARATIONS FOR EXPOSURE
Sensor4.eps
Sensor5.eps
Sensor6.eps
4.2.2 Attaching Dimax sensor to cephalostat
1.Push the sensor onto the adapter on the cephalostat.
2.Turn the locking knob 180 degrees in order to
complete the mechanical connection.
3.Push in the cephalostat electrical connector button on
the other side. This will make the electrical
connection between the sensor and cephalostat.
10Planmeca ProMax CephalostatUser’s Manual
4 PREPARATIONS FOR EXPOSURE
Tube_head_non_rotating.eps
No joint
Joint
(tube head rotates
(tube head does
not rotate)
Non-rotating tube head
-> remove 3D sensor from C-arm
Rotating tube head
-> no need to remove 3D sensor
to ceph position)
(refer to section 4.2.1)
3D unit
C-arm
4.3 Additional preparations on 3D X-ray units
If your X-ray unit has a non-rotating tube head, you must
remove the 3D sensor from the C-arm before you can
take cephalometric exposures.
Tube_head_rotating.eps
4.4 Preparing Planmeca Romexis
First select the patient.
Then select the cephalometric exposure.
Refer to the Planmeca Romexis User’s Manual for details
on Romexis functions.
User’s ManualPlanmeca ProMax Cephalostat11
4 PREPARATIONS FOR EXPOSURE
NOTE
4.5 Preparing patient
Ask the patient to remove any spectacles, hearing aids,
dentures, hairpins, and personal jewellery such as
earrings, necklaces and piercings as these can produce
shadows or reflections in the image. The patient should
also remove any loose items of clothing (e.g. scarf, tie)
that might get caught in the arm structures of the X-ray
unit.
Place a protective lead apron over the patient’s back if
required.
We recommend that you image patients with poor health
in a sitting position.
12Planmeca ProMax CephalostatUser’s Manual
5 CEPHALOMETRIC EXPOSURE
5 CEPHALOMETRIC EXPOSURE
5.1 Selecting general exposure settings
Refer to section “Touch screen” in the User’s Manual of
the X-ray unit for general information on how to make or
cancel selections on the touch screen.
5.1.1 Selecting program
Select the cephalometric program (2D Dental >
Cephalostat).
5.1.2 Selecting program type
Select the program type from the drop-down menu at the
top of the screen.
User’s ManualPlanmeca ProMax Cephalostat13
5 CEPHALOMETRIC EXPOSURE
NOTE
NOTE
5.1.3 Selecting patient size
Use this button to select the patient size:
•XS = Child
•S = Small adult
•M = Medium-sized adult
•L = Large adult
•XL = Extra large adult
The exposure values will automatically change according
to the selected patient size and program type.
The patient size can also be selected on the next screen.
14Planmeca ProMax CephalostatUser’s Manual
5 CEPHALOMETRIC EXPOSURE
X2_Di_Ceph2.eps
2 a) 0°
2 b) 180°
5.2 Patient positioning in lateral projection
1.Select the lateral program from the drop-down menu
at the top of the screen.
2.Rotate the head support plate
a.to the 0° position if you want the patient to face
away from the wall
OR
b.to the 180° position if you want the patient to
face towards the wall.
X2_Di_Ceph2.eps
3.Select the patient size as described in section 5.1.3
“Selecting patient size” on page 14.
User’s ManualPlanmeca ProMax Cephalostat15
5 CEPHALOMETRIC EXPOSURE
NOTE
NOTE
4.Select the area that you wish to expose:
•Gray prohibition sign = area to be exposed
•Red prohibition sign = area NOT to be exposed
The size of the exposure area is indicated on the screen.
The exposure area is automatically reduced for child
patients (XS).
The illustration on the screen is only an example.
5.To go to the next screen, select
•this symbol
OR
•the forward button.
16Planmeca ProMax CephalostatUser’s Manual
5 CEPHALOMETRIC EXPOSURE
X2_Di_Ceph2.eps
Ceph&Dimax2.eps
6.Press the release lever on the left ear post holder and
slide the ear posts and nasal positioner out as far as
they will go.
7.Adjust the height of the cephalostat by pressing a
height adjusting button until the positioning cones at
the ends of the ear posts are level with the patient’s
ears.
User’s ManualPlanmeca ProMax Cephalostat17
5 CEPHALOMETRIC EXPOSURE
8 a)
8 b)
X2_Di_Ceph.eps
8.Position the patient between the ear posts so that
a.they face away from the wall
OR
b.they face towards the wall.
9.Press the release lever on the left ear post holder and
very carefully slide the positioning cones into the
patient’s ears.
10. Slide the nasal positioner until it touches the patient’s
nasion.
11. Adjust the angle of the patient’s head by sliding the
nasal positioner up or down until the Frankfort plane
is horizontal.
18Planmeca ProMax CephalostatUser’s Manual
5 CEPHALOMETRIC EXPOSURE
Ceph_side.eps
90°
5.3 Patient positioning in posteroanterior or anteroposterior projection
1.Select the posteroanterior program from the dropdown menu at the top of the screen.
2.Rotate the head support plate to either of the 90°
positions.
a.To take an anteroposterior exposure, place the
nasal positioner next to the collimator as shown
below.
b.To take a posteroanterior exposure, place the
nasal positioner next to the sensor.
3.Select the patient size as described in section 5.1.3
“Selecting patient size” on page 14.
User’s ManualPlanmeca ProMax Cephalostat19
5 CEPHALOMETRIC EXPOSURE
NOTE
NOTE
4.Select the area that you wish to expose:
•Gray prohibition sign = area to be exposed
•Red prohibition sign = area NOT to be exposed
The size of the exposure area is indicated on the screen.
The exposure area is automatically reduced for child
patients (XS).
The illustration on the screen is only an example.
5.To go to the next screen, select
•this symbol
OR
•the forward button.
20Planmeca ProMax CephalostatUser’s Manual
5 CEPHALOMETRIC EXPOSURE
Ceph_side.eps
Ceph&Dimax2.eps
6.Press the release lever on the left ear post holder and
slide the ear posts and nasal positioner out as far as
they will go.
7.Adjust the height of the cephalostat by pressing a
height adjusting button until the positioning cones at
the ends of the ear posts are level with the patient’s
ears.
User’s ManualPlanmeca ProMax Cephalostat21
5 CEPHALOMETRIC EXPOSURE
8 a)
8 b)
Ceph_app.eps
8.Position the patient between the ear posts.
a.To take a posteroanterior exposure, position the
patient so that they look at the sensor.
b.To take an anteroposterior exposure, position
the patient so that they look at the collimator.
9.Press the release lever on the left ear post holder and
very carefully slide the positioning cones into the
patient’s ears.
10. Slide the nasal positioner until it touches the patient’s
nasion.
11. Adjust the tilt of the patient’s head so that the
Frankfort plane is horizontal.
22Planmeca ProMax CephalostatUser’s Manual
5 CEPHALOMETRIC EXPOSURE
Ceph_pos_up.eps
5.4 Patient positioning in submental-vertex projection
1.Perform steps one to seven as described in section
5.3 “Patient positioning in posteroanterior or
anteroposterior projection” on page 19. Rotate the
head support plate to the 90° position so that the
nasal positioner is placed next to the collimator.
2.Turn the nasal positioner up.
3.Position the patient between the ear posts and adjust
the tilt of the patient’s head until the alar-tragal line is
perpendicular to the floor.
4.Press the release lever on the left ear post holder and
very carefully slide the positioning cones into the
patient’s ears.
User’s ManualPlanmeca ProMax Cephalostat23
5 CEPHALOMETRIC EXPOSURE
Ceph_atline.eps
35-40°
5.5 Patient positioning in Waters’ view
1.Perform steps one to seven as described in section
5.3 “Patient positioning in posteroanterior or
anteroposterior projection” on page 19. Rotate the
head support plate to the 90° position so that the
nasal positioner is placed next to the sensor.
2.Turn the nasal positioner up.
3.Position the patient between the ear posts so that
they face the sensor.
4.Position the patient’s head so that the alar-tragal line
is tilted up about 35 to 40 degrees.
5.Press the release lever on the left ear post holder and
very carefully slide the positioning cones into the
patient’s ears.
24Planmeca ProMax CephalostatUser’s Manual
5 CEPHALOMETRIC EXPOSURE
Ceph_pos_ap.eps
30°
5.6 Patient positioning in AP Towne projection
1.Perform steps one to seven as described in section
5.3 “Patient positioning in posteroanterior or
anteroposterior projection” on page 19. Rotate the
head support plate to the 90° position so that the
nasal positioner is placed next to the collimator.
2.Turn the nasal positioner up.
3.Position the patient between the ear posts so that
they face the collimator.
4.Position the patient’s head so that the Frankfort plane
is tilted down about 30 degrees.
5.Ask the patient to open their mouth as wide as
possible.
6.Press the release lever on the left ear post holder and
very carefully slide the positioning cones into the
patient’s ears.
User’s ManualPlanmeca ProMax Cephalostat25
5 CEPHALOMETRIC EXPOSURE
e.g. 30°
5.7 Patient positioning in oblique projection
1.Select the oblique program from the drop-down menu
at the top of the screen.
2.Rotate the head support plate to an oblique position
(other than 0, 90 or 180 degrees).
X2_Di_Ceph2.eps
3.Select the patient size as described in section 5.1.3
“Selecting patient size” on page 14.
4.To go to the next screen, select
•this symbol
OR
•the forward button.
26Planmeca ProMax CephalostatUser’s Manual
5 CEPHALOMETRIC EXPOSURE
Ceph&Dimax2.eps
5.Press the release lever on the left ear post holder and
slide the ear posts and nasal positioner out as far as
they will go.
6.Adjust the height of the cephalostat by pressing a
height adjusting button until the positioning cones at
the ends of the ear posts are level with the patient’s
ears.
7.Position the patient between the ear posts so that the
patient faces the nasal positioner.
8.Press the release lever on the left ear post holder and
very carefully slide the positioning cones into the
patient’s ears.
9.Slide the nasal positioner until it touches the patient’s
nasion.
10. Adjust the angle of the patient’s head by sliding the
nasal positioner up or down until the Frankfort plane
is horizontal.
User’s ManualPlanmeca ProMax Cephalostat27
5 CEPHALOMETRIC EXPOSURE
5.8 Positioning in hand exposure
1.Select the hand program from the drop-down menu
at the top of the screen.
2.Rotate the head support plate to either of the 90°
positions as described in section 5.3 “Patient
positioning in posteroanterior or anteroposterior
projection” on page 19.
3.Remove the ear posts from the ear post holders.
4.Attach the hand imaging support to the ear post
holders. Ensure that you slide the hand imaging
support as far up as it will go.
5.Turn the nasal positioner up.
6.To go to the next screen, select
•this symbol
OR
•the forward button.
7.Ask the patient to place their hand against the hand
imaging support.
28Planmeca ProMax CephalostatUser’s Manual
5 CEPHALOMETRIC EXPOSURE
NOTE
5.9 Adjusting exposure values for current exposure
The exposure values have been preset at the factory for
each patient size and program type. The preset exposure
values are average values and they are only meant to
guide the user.
Always try to minimize the radiation dose to the patient.
The preset exposure values are shown in the following
tables.
Factory presets for program types Lateral and Oblique
PATIENT SIZEkV VALUEmA VALUE
Child (XS)6210
Small adult (S)6410
Medium-sized adult (M)6610
Large adult (L)6810
Extra large adult (XL)7010
Factory presets for program type Posteroanterior
PATIENT SIZEkV VALUEmA VALUE
Child (XS)7611
Small adult (S)7811
Medium-sized adult (M)8011
Large adult (L)8211
Extra large adult (XL)8411
Factory presets for program type Hand
PATIENT SIZEkV VALUEmA VALUE
All patient sizes (XS, S, M, L & XL)604
User’s ManualPlanmeca ProMax Cephalostat29
5 CEPHALOMETRIC EXPOSURE
NOTE
1
2
3
If you need to adjust the preset exposure values for the
current exposure:
1.Select the kV / mA field.
2.Use the minus or plus buttons to set the exposure
values you wish to use. To improve the image
contrast, reduce the kV value. To reduce the
radiation dose, reduce the mA value.
3.Select the green check mark button.
To adjust the preset exposure values permanently, select
Settings > Program > 2100 Programs > 2140
2D Cephalostat. Refer to section “Program settings” in the
User’s Manual of the X-ray unit for details.
30Planmeca ProMax CephalostatUser’s Manual
5 CEPHALOMETRIC EXPOSURE
NOTE
5.10 Selecting Dynamic Exposure Control (DEC)
Dynamic Exposure Control (DEC) is an optional feature.
Select the DEC button.
Dynamic Exposure Control (DEC) automatically provides
optimal exposure values for each patient during exposure.
The function adjusts the exposure values individually for
each patient based on their anatomic structure and bone
density. Switching DEC on improves the image quality as
the function produces images of more consistent
brightness and contrast.
When DEC is switched on the exposure values are
automatically adjusted during exposure. The kV value is
adjusted by max ±4 kV and the mA value is adjusted by
max +4 / -3 mA within the available scale.
5.10.1 Adjusting DEC density
If the images appear to be too bright or too dark, DEC
density can be adjusted. Select Settings > Program >
2200 Program Features > 2230 Cephalostat > Ceph DEC
Density Lat or Ceph DEC Density PA / AP. Then use the
minus or plus sign to adjust the setting.
The setting can be adjusted between 20% (lower
exposure values -> brighter image) and 200% (higher
exposure values -> darker image). The recommended
setting is 100% (default setting).
User’s ManualPlanmeca ProMax Cephalostat31
5 CEPHALOMETRIC EXPOSURE
NOTE
NOTE
NOTE
1
5.11 Taking an exposure
Make sure that you have selected the correct patient and
exposure mode in the Planmeca Romexis program.
1.Select the radiation symbol or the forward button.
Green lights flash on the touch screen and exposure
button when the X-ray system is getting ready for an
exposure. The green lights stop flashing and stay on
continuously when the X-ray system is ready for an
exposure.
2.Ask the patient to stay as still as possible.
3.Move to a protected area.
4.Press and hold down the exposure button for the
duration of the exposure.
During exposure yellow radiation warning lights
illuminate on the exposure switch and on the touch
screen, and you hear a radiation warning tone.
Additionally, a radiation warning symbol is shown on
the touch screen.
If the optional DEC (Dynamic Exposure Control)
function is switched on, the exposure is taken in two
stages. The exposure values are adjusted during the
first (short) exposure and the second exposure
produces the actual image. Press and hold down the
exposure button for the duration of both exposures.
Maintain audio and visual contact with the patient and
X-ray unit during exposure. If the sensor and / or
collimator stop moving during exposure, release the
exposure button immediately.
5.The image is shown on the computer screen.
•Note that you must accept the image in the
Planmeca Romexis program.
32Planmeca ProMax CephalostatUser’s Manual
6.Press the release lever on the left ear post holder and
very carefully slide the positioning cones out of the
patient’s ears.
7.Guide the patient away from the cephalostat.
Planmeca Oy | Asentajankatu 6 | 00880 Helsinki | Finland