The manufacturer, assembler and importer are responsible for the safety, reliability and
performance of the unit only if:
- installation, calibration, modification and repairs are carried out by qualified
authorised personnel
- electrical installations are carried out according to the appropriate requirements
such as IEC 60364
- equipment is used according to the operating instructions.
Planmeca pursues a policy of continual product development. Although every effort is
made to produce up-to-date product documentation this publication should not be
regarded as an infallible guide to current specifications. We reserve the right to make
changes without prior notice.
COPYRIGHT PLANMECA
Publication Number 10033256 Revision 12
Release Date 15 December 2016
2Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
1 INTRODUCTION
NOTE
NOTE
NOTE
1 INTRODUCTION
This manual describes how to take 2D panoramic
exposures. The manual applies to the following X-ray
units:
•Planmeca ProMax 2D S2
•Planmeca ProMax 2D S3
•Planmeca ProMax 3D s
•Planmeca ProMax 3D Classic
This manual is valid for software version 3.8.0.0.r or later.
This software version is compatible with Planmeca
Romexis software version 4.5.0.r or later. To check the
software version of your X-ray unit, select Settings > About
> 4100 Component Information > ProMax SW version.
The X-ray unit uses panoramic techniques to produce
two-dimensional (2D) X-ray images for the examination of
dentomaxillofacial anatomy.
You need a PC with the Planmeca Romexis program in
order to save, view and modify the images.
Make sure that you are fully acquainted with the
appropriate radiation protection measures and these
instructions before you use the X-ray unit. Note that your
X-ray unit may not feature all the options described in
these instructions.
These instructions include options that may not be
available in all countries.
The X-ray unit may be used by health care professionals
only.
You can register your X-ray unit online as described in
section 12.3.3 “Product Registration (4300)” on page 81.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic1
2 ASSOCIATED DOCUMENTATION
NOTE
2 ASSOCIATED DOCUMENTATION
The X-ray unit is supplied with the following manuals:
•User’s Manual(s) for
- 2D Imaging, Original English publication: 10033256
(- 2D Tomography, Original English publication:
10036593)
(- Cephalostat, Original English publication: 10033034 or
10033035)
(- 3D Imaging, Original English publication: 10033255)
•Installation Quick Guide,
Original English publication: 10039996
•Installation Manual,
Original English publication: 10038029
•Technical Manual,
Original English publication: 10033257
•Planmeca Device Tool Manual,
Original English publication: 10031558
These manuals are intended to be used in conjunction
with the documentation for the Planmeca Romexis
program. The Romexis package contains the following
manuals:
•User’s Manual,
Original English publication: 10014593
•Technical Manual,
Original English publication: 10037884
The User’s Manuals are available on Planmeca’s website.
•For X-ray units, select Material bank > Manuals >
Imaging.
•For software products, select Material bank >
Manuals > Software.
2Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
3 SYMBOLS ON PRODUCT LABELS
0598
3 SYMBOLS ON PRODUCT LABELS
3.1 Symbols on permanently installed X-ray units
CE marking according to European standard
(Directive 93/42/EEC)
SGS marking according to US and Canadian standards
(ANSI/UL 60601-1 and CAN/CSA C22.2 No. 601.1-M90)
Date of manufacture (Standard ISO 15223-1)
Type B applied part (Standard IEC 60601-1)
Separate collection for electrical and electronic equipment
(Directive 2002/96/EC WEEE)
Refer to instruction manual/booklet
(Standard ISO 7010-M002)
4Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
4 SAFETY PRECAUTIONS
WARNING
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
4 SAFETY PRECAUTIONS
The following safety precautions must be observed in
order to avoid the risk of personal injury or damage to
the X-ray unit.
FOR US USERS:
Federal law restricts this device to sale by or on the
order of a health care professional.
This X-ray unit may be dangerous to both patient and
operator unless safe exposure values are used and
correct operating procedures are observed.
It is very important that the place where the X-ray unit
is to be used and the position from which the user is
to operate the X-ray unit are correctly shielded.
Since radiation safety requirements vary from country
to country and state to state it is the responsibility of
the user to ensure that all local safety requirements
are met.
The patient positioning lights are laser lights. Do not
stare into the laser beam.
Do not drop the sensor. Planmeca limited warranty
does not cover damage which is due to misuse, e.g.
dropping the sensor, neglect, or any cause other than
ordinary use.
If you have any reason to believe that the sensor
might be faulty, take a test exposure before taking a
patient exposure.
If an exposure is interrupted (e.g. exposure button is
released or emergency stop button activated), the
patient must be guided away from the X-ray unit
before the C-arm is moved.
Do not connect items which are not specified as part
of the system.
Do not touch an electrical connector and the patient
at the same time.
If the X-ray unit shows any signs of oil leakage,
switch the X-ray unit off and contact your service
technician for help.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic5
4 SAFETY PRECAUTIONS
CAUTION
CAUTION
CAUTION
NOTE
NOTE
NOTE
NOTE
NOTE
NOTE
NOTE
NOTE
Do not use the X-ray unit in an oxygen rich
environment or in the presence of flammable
anesthetics.
Never use a defective or damaged X-ray system.
Contact your service technician for help.
Do not modify the X-ray unit. The X-ray unit must be
serviced by qualified personnel only.
Cone beam imaging should not be used for routine (or
screening) examinations. The imaging examinations must
be justified for each patient to demonstrate that the
benefits outweigh the risks.
When it is likely that evaluation of soft tissues will be
required as part of the patient’s radiological assessment,
conventional CT or MR medical imaging should be used
rather than CBCT.
Before taking an exposure, ask any female patient of
childbearing age whether she might be pregnant. The Xray unit is not intended for use on pregnant women.
FOR CANADIAN USERS:
All patients must be provided with a shielded apron for
gonad protection and a thyroid shield. The use of a thyroid
shield is especially important in children. The shielded
apron and thyroid shield should have a lead equivalence
of at least 0.25 mm on both sides (front and back of the
patient).
If the X-ray unit has been stored at temperatures under
+10°C for more than a few hours, time must be allowed for
the unit to reach room temperature before turning it on.
Ensure efficient air conditioning in the X-ray room. It is
recommended to keep the room temperature between
+20°C and +25°C at all times.
If exposures are taken in rapid succession, the X-ray tube
may overheat and a cooling time will flash on the touch
screen. The cooling time indicates the delay before the
next exposure can be taken.
If the X-ray system is not connected to an Uninterruptible
Power Supply (UPS), switch the X-ray unit off and
disconnect the PCs from the mains during lightning
storms.
6Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
4 SAFETY PRECAUTIONS
NOTE
NOTE
NOTE
NOTE
NOTE
NOTE
NOTE
NOTE
NOTE
CAUTION
LASER RADIATION -
DO NOT STARE
INTO BEAM
1mW
635nm
CLASS II
LASER
PRODUCT
LBL-X-099
FOR US & CANADIAN USERS:
The laser lights are class II laser products (21 CFR §
1040.10).
CLASS 1 LASER PRODUCT
APPAREIL À LASER DE CLASSE 1
IEC 60825-1:2007
FOR EUROPEAN USERS:
The laser lights are class 1 laser products (Standard IEC /
EN 60825-1: 2007).
EMC requirements have to be considered, and the
equipment must be installed and put into service according
to the specific EMC information provided in the
accompanying documents.
Portable and mobile RF communications equipment can
affect the X-ray unit.
External equipment intended for connection to signal
input, signal output or other connectors, shall comply with
relevant IEC standard (e.g. IEC 60950 for IT equipment
and the IEC 60601 series for medical electrical
equipment). In addition, all such combinations - systems shall comply with the standard IEC 60601-1-1, Safety
requirements for medical electrical systems. Equipment
not complying to IEC 60601 shall be kept outside the
patient area (more than 2m (79 in.) from the X-ray unit).
Any person who connects external equipment to signal
input, signal output or other connectors has formed a
system and is therefore responsible for the system to
comply with the requirements of IEC 60601-1-1. If in
doubt, contact your service technician or local
representative for help.
Contact your service technician if you notice a decrease in
image quality.
Contact your service technician if you have taken an
exposure but the image does not appear in the Planmeca
Romexis program. The last ten images can be manually
imported into Romexis.
Do not handle liquids near or on the X-ray unit.
Never place or hang any objects on any part of the X-ray
unit.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic7
4 SAFETY PRECAUTIONS
NOTE
NOTE
NOTE
NOTE
NOTE
Prof_6C.eps
Make sure that neither you nor your patient can get caught
or hooked up on any part of the X-ray unit. Keep loose
items of clothing, hair and jewellery tucked away safely.
If you have any reason to believe that the C-arm might hit
the patient during exposure (e.g. patients with wide
shoulders), take a test exposure without radiation first. To
turn radiation off, select Settings > User > 1300
Operational settings > 1310 Use Mode > 1311 Set Demo
Mode.
Do not touch the arm structures when the X-ray unit is
moving.
Patients are not allowed to hang on the patient handles.
FOR PROFACE SENSOR:
Do not touch the glass windows. Fingerprints or other
stains on the glass surface destroy image quality.
8Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
5 SWITCHING X-RAY UNIT ON
NOTE
5 SWITCHING X-RAY UNIT ON
The on / off switch is located on the underside of the
stationary column top.
To prolong the lifetime of your X-ray unit, always switch the
X-ray unit off when it is not in active use.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic9
6 MAIN PARTS
Ethernet
1
2 & 3
6 MAIN PARTS
6.1 General view of 2D X-ray system
1X-ray unit
2Planmeca Romexis program
3ProTouch desktop application
(optional, see section 6.10 “ProTouch desktop application” on page 19)
10Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
6 MAIN PARTS
Ethernet
1
3 & 4
2
6.2 General view of 3D X-ray system
1X-ray unit
23D reconstruction PC
3Planmeca Romexis program
4ProTouch desktop application
(optional, see section 6.10 “ProTouch desktop application” on page 19)
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic11
6 MAIN PARTS
1
2
3
4
5
6
7
8
9
10
6.3 General view of X-ray unit
1Sensor
(see section 6.4 “Sensor” on page 13)
2C-arm
3Patient supports
(see section 6.5 “Patient supports” on page 14)
4Patient support table
5Patient handles
6Patient positioning controls
(see section 6.11 “Patient positioning controls” on page 20)
7Touch screen
(see section 6.9 “Touch screen” on page 16)
8Telescopic column
(see section 6.6 “Telescopic column” on page 14)
9Stationary column
10 Emergency stop button
(see section 6.8 “Emergency stop button” on page 15)
12Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
6 MAIN PARTS
Prof_1B.eps
3D_sensor.eps
1
2
3
4
6.4 Sensor
Dimax_sensor.eps
3Ds_sensor.eps
1Dimax sensor
23D sensor for Planmeca ProMax 3D s
33D sensor for Planmeca ProMax 3D Classic
4ProFace sensor for Planmeca ProMax 3D s and Planmeca ProMax 3D Classic
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic13
6 MAIN PARTS
Bite1.eps
temple_support.ai
Chin2.eps
Bite2.eps
Chin1.eps
Lip1.eps
1
2
3
4
5
6
Digi_accessory_cabinet.eps
1
2
6.5 Patient supports
6.6 Telescopic column
1Bite piece
2Chin rest
3Adapter
4Chin cup
5Chin support
6Temple supports
1Telescopic column without accessory cabinet
2Telescopic column with accessory cabinet
(three accessory boxes)
14Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
6 MAIN PARTS
1) Flashing green =
Getting ready
2) Continuous green =
Ready
3) Yellow = Radiation
Exposure switch
Exposure button
6.7 Exposure switch
The exposure switch can be mounted on the wall, or it can
be hung from the hook provided on the stationary column
top if a protected area is within reach.
Green lights flash on the exposure button and on the
touch screen when the X-ray system is getting ready for
an exposure. The green lights stop flashing and stay on
continuously when the X-ray system is ready for an
exposure.
During exposure yellow radiation warning lights illuminate
on the exposure switch and on the touch screen. They
indicate that the X-ray unit is generating radiation.
6.8 Emergency stop button
The emergency stop button is located on the top of the
stationary column. Press the button to stop the X-ray unit
operating in an emergency. When the emergency stop
button is pressed down, all movements of the X-ray unit
are blocked and the unit will not generate radiation. The
up / down movement will stop within a distance of 10 mm
(0.4 in.).
A help message will appear on the touch screen. Guide
the patient away from the X-ray unit. Then release the
emergency stop button. The X-ray unit will automatically
restart.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic15
6 MAIN PARTS
NOTE
NOTE
NOTE
NOTE
6.9 Touch screen
The options shown on the touch screen depend on the unit
configuration. The X-ray unit can be upgraded with new
programs and features, contact your dealer for further
information. The views and values shown in this manual
are only examples.
The illustrations shown on the touch screen are based on
approximate patient anatomy. The actual exposure area
depends on the individual anatomy of the patient.
Never allow patients to touch the screen when they are
positioned in the X-ray unit. Touching the screen during
exposure will stop the imaging process.
The main screen shows the name and the imaging
programs of the X-ray unit. You can use the buttons at the
bottom of the main screen to change the appearance of
the main view.
•Home button:
To view up to five most recently used programs,
select the home button. The most recently used
program is shown first. This is the default view of the
main screen.
If you wish to use fast forward buttons on this view,
select Settings > Program > 2200 Program Features
> 2240 General > Fast Forward ON. Using a fast
forward button takes you directly to the screen where
you can take an exposure.
•Program group button(s):
To view all the programs that are available for a
program group, select the program group button.
16Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
6 MAIN PARTS
NOTE
•Making selections:
To make a selection on the touch screen, simply
touch a button or a field with your finger or a soft
stylus. The selected option is highlighted. To deselect
an option, touch the button or field again (or select an
other option if available).
You hear an audible signal when you make a
selection. If you wish to adjust the volume of the
signal, select Settings > User > 1300 Operational
Settings > 1320 Audio Settings > Touch Volume.
Do not use sharp objects to operate the touch screen.
•Accept button:
To accept a selection, select the green check mark
button.
•Cancel button:
To cancel a selection and close the screen, select the
red cross button.
•Pause button:
To pause a function (instead of cancelling it), select
the pause button.
•Moving forward:
To go to the next screen, select
the forward button
or
the next symbol at the bottom of the screen.
•Moving backward:
To go back to a previous screen, select a previous
symbol at the bottom of the screen.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic17
6 MAIN PARTS
•Scrolling lists:
To scroll a list down or up, slide your finger on the
touch screen.
•Checking patient’s name and ID number:
The patient’s name and ID number are shown in the
top right corner of the touch screen. The patient and
exposure mode have to be selected in Planmeca
Romexis.
•Checking DAP value:
The estimated value for DAP (Dose Area Product) is
shown with black text on the touch screen before you
take an exposure. The actual value is shown with
green text after the exposure.
•Changing settings:
To change a setting, select the settings symbol on
the main screen. This takes you to the settings menu
where you can adjust the settings of the X-ray unit.
Refer to section 12 “SETTINGS” on page 67 for
details.
•About stand-by mode:
The touch screen will automatically switch to standby mode if you do not touch the screen or the
exposure button for more than thirty minutes. In
stand-by mode the green light on the on / off switch
indicates that the X-ray unit is switched on even
though the screen is dark. The screen will switch on
as soon as you touch it again.
•Selecting demo mode:
You can switch demo mode on if you wish to practice
or demonstrate the functions of the X-ray unit without
radiation (Settings > User > 1300 Operational
settings > 1310 Use Mode > 1311 Set Demo Mode).
18Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
6 MAIN PARTS
NOTE
NOTE
6.10 ProTouch desktop application
If the ProTouch desktop application is installed on your
computer, you can use a virtual control panel on your
computer screen.
The application is identical to the touch screen that is
integrated to the X-ray unit. The two control panels are
synchronized, and you can use either or both of them.
You need a licence if you wish to use the application for
moving to the ready status. Contact your dealer for help.
To use the virtual control panel, double-click the
ProTouch icon on your computer screen. Then click your
mouse on the function that you wish to use.
A help message appears if you used the settings menu
before you opened the application. The application will
automatically close when you accept the help message.
You can reopen the application by double-clicking the icon
again.
The small icon in the bottom left corner of the screen
indicates when the application is connected to the X-ray
unit.
•A green icon indicates that the connection is working.
•A red icon indicates that the connection is not
working.
Restart the X-ray unit if the icon is red.
To move the control panel screen, drag it with your
mouse.
To close the application, click on the small red cross in the
top right corner of the screen.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic19
6 MAIN PARTS
NOTE
NOTE
NOTE
NOTE
NOTE
Open / close temple supports
Move X-ray unit up
Positioning joystick
Move X-ray unit down
(2D imaging)
Down
Up
6.11 Patient positioning controls
Never allow patients to press the positioning controls when
they are positioned in the X-ray unit.
Pressing any of the positioning controls (button or joystick)
will switch the patient positioning lights on. The lights will
automatically switch off after two minutes. To switch them
off earlier, press the positioning joystick.
X-ray unit up / down
The X-ray unit up and down buttons are used to adjust the
X-ray unit to suit the height of the patient.
The X-ray unit moves slowly at first, then faster.
If for some reason either of the buttons gets stuck during
operation, you can stop the up / down movement by
pressing any of the other control buttons or the positioning
joystick. This is a safety measure that guarantees that the
up / down movement can be stopped in an emergency.
Be careful that the X-ray unit does not hit the ceiling when
you press the up button. The maximum height can be
adjusted to suit offices with low ceiling, contact your
service technician for help.
Make sure that there is no object under the X-ray unit
when you press the down button. If something is in danger
of becoming trapped, release the button immediately to
stop the movement. Clear any obstruction before pressing
the button again.
20Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
6 MAIN PARTS
NOTE
NOTE
emergency_stop.eps
The column movement stops automatically if the
emergency stop plate at the bottom is pressed upwards.
Clear any obstruction before moving the column again.
When positioning seated patients (e.g. in a wheelchair)
always first move the X-ray unit down before you position
the patient in the unit.
Positioning joystick
The positioning joystick is used for adjusting the
positioning lights. It is used when the patient is positioned
in the X-ray unit.
Open / close temple supports
Press the temple support button to open the temple
supports in 2D imaging. The temple supports can be
closed by pressing the temple support button again.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic21
7 PROGRAMS
NOTE
NOTE
7 PROGRAMS
The X-ray unit uses SCARA (Selectively Compliant
Articulated Robot Arm) technology for arm movements.
Planmeca ProMax 2D S3 and Planmeca ProMax 3D
X-ray units have three joints (S3 = SCARA 3) and allow
limitless imaging possibilities. Planmeca ProMax 2D S2
X-ray units have two joints (S2 = SCARA 2) and offer a
more limited selection of imaging programs.
The image width and height depend on the X-ray unit
model and sensor. The images shown here are only
examples.
7.1 Panoramic programs
Standard
Bitewing
The standard panoramic program has a traditional path
and angles of the beam.
This program produces bitewing images from premolar
and molar areas including parts of maxilla, mandible and
rami. The bottom of the maxillary sinus as well as the
mandibular canal and the mental foramen are also visible.
This program is optimized for interproximal imaging and a
shadow of the opposite side teeth may therefore be visible
in the image.
22Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
7 PROGRAMS
NOTE
NOTE
1/2
1/2
2/2
2/2
Interproximal
The basic imaging geometry is the same as in the
standard panoramic program but the X-ray beam is more
parallel to the interproximal contacts of the teeth.
This program is optimized for interproximal imaging and a
shadow of the opposite side teeth may therefore be visible
in the image.
Orthogonal
The basic imaging geometry is the same as in the
standard panoramic program but the X-ray beam is
perpendicular to the jaw.
This program is optimized for orthogonal imaging and a
shadow of the opposite side teeth may therefore be visible
in the image.
7.2 Temporomandibular joint (TMJ) programs
Double lateral
Lateral exposures of closed (1/2) and open (2/2)
temporomandibular joints.
•Planmeca ProMax 2D S3 & Planmeca ProMax 3D:
The imaging position and angle can be adjusted. The
default imaging angle is 0°.
•Planmeca ProMax 2D S2:
The imaging position can be adjusted.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic23
7 PROGRAMS
1/2
1/2
2/22/2
1/2
1/2
2/22/2
1/3
2/33/3
Double posteroanterior
Posteroanterior exposures of closed (1/2) and open (2/2)
temporomandibular joints.
•Planmeca ProMax 2D S3 & Planmeca ProMax 3D:
The imaging position and angle can be adjusted. The
default imaging angle is 0°.
Double lateral-posteroanterior
Three angles lateral
Lateral (1/2) and posteroanterior (2/2) exposures of
closed or open temporomandibular joints. The imaging
angles (lateral and PA) are adjustable (default angle: 0°).
.
Three lateral multi-angle TMJ exposures (left or right).
The imaging angle for image no. 2 is adjustable (three
imaging angles: 0° ± 7° by default). The selected imaging
angle is in image no. 2.
24Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
7 PROGRAMS
1/3
2/33/3
Three angles posteroanterior
Three posteroanterior multi-angle TMJ exposures (left or
right).
The imaging angle for image no. 2 is adjustable (three
imaging angles: 0° ± 7° by default). The selected imaging
angle is in image no. 2.
7.3 Sinus programs
Posteroanterior
Posteroanterior linear
Posteroanterior sinus exposure.
Posteroanterior linear sinus exposure.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic25
7 PROGRAMS
Lateral
Lateral exposure of the left or right sinus area.
26Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
8 PREPARATIONS FOR EXPOSURE
NOTE
NOTE
NOTE
NOTE
CAUTION
Prof_6C.eps
shockwarningD3.eps
1
8 PREPARATIONS FOR EXPOSURE
8.1 Attaching and removing sensor
The available sensors are shown in section 6.4 “Sensor”
on page 13.
FOR PLANMECA PROMAX 2D X-RAY UNITS:
If the sensor is attached to the cephalostat (optional), the
sensor must be moved to the C-arm as described below.
FOR PLANMECA PROMAX 3D X-RAY UNITS:
If you wish to use a Dimax sensor, select Settings > User
> 1300 Operational Settings > 1340 Exposure Settings >
Panoramic System Dimax and change the sensor as
described below.
FOR PROFACE SENSOR:
Do not touch the glass windows. Fingerprints or other
stains on the glass surface destroy image quality.
Do not drop the sensor. Planmeca limited warranty
does not cover damage which is due to misuse, e.g.
dropping the sensor, neglect, or any cause other than
ordinary use.
Do not use the sensor if the shock indicator (1) is red
- contact your service technician for help.
If you have any reason to believe that the sensor
might be faulty, take a test exposure before taking a
patient exposure.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic27
8 PREPARATIONS FOR EXPOSURE
Sensor3 .eps
Sensor2.eps
Sensor7.eps
8.1.1 Attaching sensor to C-arm
1.Push the sensor onto the connector on the C-arm.
2.Turn the locking knob over the fastening mechanism.
This will secure the sensor in position.
3.Push in the C-arm electrical connector button on the
other side. This will make the electrical connection
between the sensor and C-arm.
28Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
8 PREPARATIONS FOR EXPOSURE
NOTE
NOTE
Sensor1.eps
Sensor2.eps
Sensor3 .eps
8.1.2 Detaching sensor from C-arm
Do not remove the sensor during imaging process.
1.Push in the C-arm electrical connector. This will
disconnect the electrical connection between the
sensor and C-arm.
2.Turn the locking knob 180 degrees. This will release
the locking mechanism.
3.Carefully pull the sensor out.
For safety reasons, wait for at least ten seconds before
you attach a sensor again. The blue indicator light under
the locking knob has to switch off first.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic29
8 PREPARATIONS FOR EXPOSURE
NOTE
NOTE
8.2 Preparing Planmeca Romexis
First select the patient.
Then select the panoramic exposure.
8.3 Preparing patient
Refer to the Planmeca Romexis User’s Manual for details
on Romexis functions.
Ask the patient to remove any spectacles, hearing aids,
dentures, hairpins, and personal jewellery such as
earrings, necklaces and piercings as these can produce
shadows or reflections in the image. The patient should
also remove any loose items of clothing (e.g. scarf, tie)
that might get caught in the arm structures of the X-ray
unit.
High contrast objects, such as gold teeth or amalgam, may
cause artefacts in the image.
Place a protective lead apron over the patient’s back if
required.
We recommend that you image patients with poor health
in a sitting position.
30Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
9 PANORAMIC EXPOSURE
NOTE
Insert_temple_supports.eps
ProMax_2_3.eps
Chin2.eps
9 PANORAMIC EXPOSURE
9.1 Before exposure
Insert the temple supports into the holes in the patient
support table.
Ensure that you insert the temple supports the right way
round (i.e. with the convex side facing outward as shown).
•Use these patient supports.
Bite2.eps
Bite1.eps
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic31
9 PANORAMIC EXPOSURE
NOTE
NOTE
Lip1.eps
Bite1.eps
Chin1.eps
Lip1.eps
Bite1.eps
FOR BITEWING EXPOSURES:
You can use these patient supports if the patient has a
very large jaw.
You can use these patient supports for edentulous
patients or for patients who are unable to bite.
32Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
9 PANORAMIC EXPOSURE
NOTE
9.2 Selecting exposure settings
Refer to section 6.9 “Touch screen” on page 16 for
general information on how to make or cancel selections
on the touch screen.
9.2.1 Selecting program
Select the panoramic program (2D Dental > Panoramic).
9.2.2 Selecting program type
Select the program type from the
•drop-down menu at the top
OR
•left of the screen.
Refer to section 7.1 “Panoramic programs” on page 22 for
details.
The available options depend on the X-ray unit model.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic33
9 PANORAMIC EXPOSURE
NOTE
NOTE
NOTE
9.2.3 Selecting patient size
Use this button to select the patient size:
•XS = Child
•S = Small adult
•M = Medium-sized adult
•L = Large adult
•XL = Extra large adult
Selecting child patient (XS) will automatically reduce the
exposure area and patient dose.
The exposure values will automatically change according
to the selected patient size.
The patient size can also be selected on the next screen.
34Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
9 PANORAMIC EXPOSURE
NOTE
NOTE
9.2.4 Selecting segmentation
The function allows you to take exposures of different jaw
segments. This will reduce the radiation dose of the
patient.
The illustration on the screen is divided into three
horizontal segments and five vertical segments. Select
the segments that should NOT be exposed. The
prohibition signs of these segments turn red.
It is not possible to expose two separate horizontal
segments.
The illustration on the screen is only an example. The
actual size of the exposed area depends on the patient’s
individual anatomy.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic35
9 PANORAMIC EXPOSURE
NOTE
NOTE
1
3
9.3 Patient positioning
9.3.1 Selecting patient entry position
Use the buttons on the right of the screen to select the
patient entry position.
•Select this button to move the C-arm to the back,
away from the patient positioning area. This full view
position allows you to monitor and adjust the patient’s
position freely from all directions.
•Select this button to position the C-arm around the
patient support table. This is the traditional closed
patient entry position.
If needed, the full view position (top button) can be
disabled (Settings > User > 1300 Operational settings >
1330 Patient positioning > Side entry OFF). This might be
necessary if there is no space for the C-arm to move back.
9.3.2 Positioning patient’s head
Planmeca ProMax 2D S2 X-ray units offer only one patient
entry position. This entry position is the closed patient
entry position where the C-arm is positioned around the
temple supports.
1.Press the temple support button to open the temple
supports if they are not already open.
2.Guide the patient to the X-ray unit.
3.Adjust the X-ray unit to suit the height of the patient.
To do this, press either of the height adjusting
buttons until the chin rest is approximately level with
the patient’s lower jaw.
36Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
9 PANORAMIC EXPOSURE
NOTE
NOTE
Dimax_mids_2A.eps
Frankfort plane light
Midsagittal
plane light
Layer light
4.Ask the patient to step forward, grasp the patient
handles, stretch and straighten their back and neck,
and bite the bite piece. The upper and lower incisors
must be in the groove in the bite piece.
If you use the chin support, position the patient so that the
chin touches the top bar as shown.
If you use the chin support or the chin cup, use for example
a cotton roll to ensure that the patient’s upper and lower
incisors do not overlap.
The positioning lights show on the patient’s face.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic37
9 PANORAMIC EXPOSURE
NOTE
patposlights3.eps
a.)
b.)
The positioning lights will automatically switch off after two
minutes. To switch them off earlier, press the positioning
joystick.
To switch them back on, do one of the following:
a.Press the thumb wheel on the underside of the
patient support table.
b.Press any of the positioning controls (button or
joystick).
9.3.3 Adjusting patient’s head position
The positioning lights and the illustrations on the touch
screen help you to position the patient’s head correctly.
The illustrations are for guidelines only.
1.To go to the next screen, select
•this symbol
OR
•the forward button.
38Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
2.Position the patient’s midsagittal plane so that it
coincides with the midsagittal plane light.
9 PANORAMIC EXPOSURE
patposlights3.eps
3.Position the patient’s Frankfort plane so that it
coincides with the Frankfort plane light.
To do this, adjust the tilt of the patient’s head by
raising or lowering the X-ray unit with the height
adjusting buttons. The patient’s back and neck must
be straight.
•The Frankfort plane light is located inside the column.
The light’s position can be adjusted if needed. This is
done by rotating the thumb wheel below the light slot.
4.Position the apices of the patient’s upper central
incisors within the image layer of the X-ray unit.
To do this, rotate the thumb wheel on the underside
of the patient support table to move the layer light
until it falls between the second incisor and the
canine.
5.Check that the midsagittal plane light and the
Frankfort plane light are still correctly positioned.
Reposition them if necessary.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic39
9 PANORAMIC EXPOSURE
NOTE
NOTE
12
3
A
C
B
1
2
3
9.4 Selecting jaw size and shape
Use this button to select the
•jaw size of the patient (columns):
1.Small
2.Medium
3.Large
and
•jaw shape of the patient (rows):
A.V-shaped
B.Standard
C.U-shaped
The jaw size will automatically change according to the
selected patient size (XS = 1, S & M & L = 2, XL = 3).
If needed, you can change the patient size on this screen.
40Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
9 PANORAMIC EXPOSURE
NOTE
NOTE
9.5 Selecting MultiView (Planmeca ProMax 3D X-ray units)
MultiView is an optional feature for Planmeca ProMax 3D
X-ray units.
Use this button to select the MultiView imaging mode.
This function allows you to adjust the viewing angle in the
Planmeca Romexis program.
•Panoramic layers with MultiView (button selected)
•Panoramic layers without MultiView (button not
selected)
The exposure values will automatically change according
to the selected patient size and imaging mode.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic41
9 PANORAMIC EXPOSURE
NOTE
NOTE
9.6 Adjusting exposure values for current exposure
The exposure values have been preset at the factory for
each patient size. The preset exposure values are
average values and they are only meant to guide the user.
FOR X-RAY UNITS WITH DIMAX SENSOR:
The preset exposure values are optimized for taking
exposures at enhanced resolution (Romexis setting). You
can use lower exposure values when taking exposures at
normal resolution.
Always try to minimize the radiation dose to the patient.
The preset exposure values are shown in the following
table.
Factory presets for panoramic exposures
PATIENT SIZEkV VALUEmA VALUE
X-ray units with Dimax sensor
Child (XS)625
Small adult (S)646.3
Medium-sized adult (M)668
Large adult (L)6810
Extra large adult (XL)7012.5
X-ray units with 3D sensor
SmartPan
Child (XS)625.6
Small adult (S)647.1
Medium-sized adult (M)669
Large adult (L)6811
Extra large adult (XL)7014
MultiView SmartPan
Child (XS)6210
Small adult (S)6411
Medium-sized adult (M)6612.5
Large adult (L)6814
Extra large adult (XL)7016
42Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
9 PANORAMIC EXPOSURE
NOTE
1
2
3
If you need to adjust the preset exposure values for this
exposure:
1.Select the kV / mA field.
2.Use the minus or plus buttons to set the exposure
values you wish to use. To improve the image
contrast, reduce the kV value. To reduce the
radiation dose, reduce the mA value.
3.Select the green check mark button.
You can adjust the preset exposure values permanently as
described in section 12.2.1 “Programs (2100)” on page 77.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic43
Dynamic Exposure Control (DEC) is an optional feature for
Planmeca ProMax 2D X-ray units.
DEC and vertical segmenting can be used simultaneously.
However, the right-most vertical segment cannot be
deselected.
DEC and horizontal segmenting cannot be used
simultaneously.
Select the DEC button.
Dynamic Exposure Control (DEC) automatically provides
optimal exposure values for each patient during exposure.
The function adjusts the exposure values individually for
each patient based on their anatomic structure and bone
density. Switching DEC on improves the image quality as
the function produces images of more consistent
brightness and contrast.
When DEC is switched on the exposure values are
automatically adjusted during exposure. The kV value is
adjusted by max ±4 kV and the mA value is adjusted by
max +4 / -3 mA within the available scale.
9.7.1 Adjusting DEC density
If the images appear to be too bright or too dark, DEC
density can be adjusted. Select Settings > Program >
2200 Program Features > 2210 2D Panoramic > Pan
DEC Density and use the minus or plus sign to adjust the
setting.
The setting can be adjusted between 20% (lower
exposure values -> brighter image) and 200% (higher
exposure values -> darker image). The recommended
setting is 100% (default setting).
44Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
9 PANORAMIC EXPOSURE
NOTE
NOTE
1
2
9.8 Taking an exposure
Make sure that you have selected the correct patient and
exposure mode in the Planmeca Romexis program.
1.Make sure that the Exposure tab is selected.
2.Select the radiation symbol or the forward button.
Green lights flash on the touch screen and exposure
button when the X-ray system is getting ready for an
exposure. The green lights stop flashing and stay on
continuously when the X-ray system is ready for an
exposure.
3.Ask the patient to swallow, place their tongue flat
against the roof of the mouth and stay as still as
possible.
4.Move to a protected area.
5.Press and hold down the exposure button for the
duration of the exposure.
The C-arm moves around the patient’s head. During
the exposure yellow radiation warning lights
illuminate on the exposure switch and on the touch
screen, and you hear a radiation warning tone.
Additionally, a radiation warning symbol is shown on
the touch screen.
Maintain audio and visual contact with the patient and
X-ray unit during exposure. If the C-arm st ops moving
during exposure, or moves in an erratic way, release
the exposure button immediately.
6.The image is shown on the computer screen.
•Note that you must accept the image in the
Planmeca Romexis program.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic45
7.Guide the patient away from the X-ray unit.
9 PANORAMIC EXPOSURE
NOTE
NOTE
NOTE
NOTE
9.9 Taking an exposure with Autofocus
(Planmeca ProMax 2D S3 X-ray units)
Make sure that you have selected the correct patient and
exposure mode in the Planmeca Romexis program.
Autofocus is an optional feature for Planmeca ProMax 2D
S3 X-ray units. It is available for standard, interproximal
and orthogonal panoramic programs.
To turn Autofocus on, select Settings > Program > 2200
Program Features > 2210 2D Panoramic > Autofocus ON.
If Autofocus is selected, DEC is automatically switched off
for the first exposure (scout image).
1.Select the Autofocus tab. This switches the layer light
off.
Autofocus adjusts the layer position automatically.
The function positions the image layer individually for
each patient based on the position and angle of the
apices of the upper central incisors.
The exposure is taken in two stages and the C-arm
moves twice.
2.Take the first exposure as described in section 9.8
“Taking an exposure” on page 45. The first exposure
is a short, low-dose exposure during which the
optimal position for the image layer is calculated.
3.The image appears on the touch screen and on the
computer screen. The calculated layer position is
shown with a white line on the image.
46Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
9 PANORAMIC EXPOSURE
NOTE
NOTE
•If needed, you can adjust the layer position
manually by using the plus or minus sign on the
touch screen. The new position is shown with a
red line on the image.
Make sure that the patient does not move between
exposures.
4.Press and hold down the exposure button again to
take the second exposure. The second exposure will
produce the actual image and the C-arm will now
move through one complete exposure cycle.
During exposure yellow radiation warning lights
illuminate on the exposure switch and on the touch
screen, and you hear a radiation warning tone.
Additionally, a radiation warning symbol is shown on
the touch screen.
Maintain audio and visual contact with the patient and
X-ray unit during exposure. If the C-arm st ops moving
during exposure, or moves in an erratic way, release
the exposure button immediately.
5.The image is shown on the computer screen.
•Note that you must accept the image in the
Planmeca Romexis program.
6.Guide the patient away from the X-ray unit.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic47
10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE
NOTE
Insert_temple_supports.eps
ProMax_2_3.eps
Bite1.eps
Lip1.eps
10 TEMPOROMANDIBULAR JOINT (TMJ)
EXPOSURE
Double TMJ programs produce closed and open views of
the left and right temporomandibular joints.
Three angle TMJ programs produce three exposures with
different angles from the left or right temporomandibular
joint.
10.1 Before exposure
Insert the temple supports into the holes in the patient
support table.
Ensure that you insert the temple supports the right way
round (i.e. with the convex side facing outward as shown).
Use these patient supports.
48Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE
NOTE
10.2 Selecting exposure settings
Refer to section 6.9 “Touch screen” on page 16 for
general information on how to make or cancel selections
on the touch screen.
10.2.1 Selecting program
Select the TMJ program (2D Dental > TMJ).
10.2.2 Selecting program type
Select the program type from the
•drop-down menu at the top
OR
•left of the screen.
Refer to section 7.2 “Temporomandibular joint (TMJ)
programs” on page 23 for details.
The available options depend on the X-ray unit model.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic49
10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE
NOTE
-
+
10.2.3 Selecting patient size
Select the patient size as described in section 9.2.3
“Selecting patient size” on page 34.
10.2.4 Selecting jaw side(s)
Use this button to select the right, left or both jaw sides.
Use the arrow buttons to select the imaging angle for this
exposure.
To change the default angle (0 degrees), select Settings >
Program > 2200 Program Features > 2220 2D TMJ >
Lateral TMJ Default Angle or PA TMJ Default Angle and
adjust the setting with the plus or minus sign.
50Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE
1
3
patposlights3.eps
a.)
b.)
10.3 Patient positioning
10.3.1 Selecting patient entry position
Select the patient entry position as described in section
9.3.1 “Selecting patient entry position” on page 36.
10.3.2 Positioning patient’s head
1.Press the temple support button to open the temple
supports if they are not already open.
2.Guide the patient to the X-ray unit.
3.Adjust the X-ray unit to suit the height of the patient.
To do this, press either of the height adjusting
buttons until the opening in the chin support is
approximately level with the patient’s mouth.
4.Ask the patient to step forward, grasp the patient
handles, stretch and straighten their back and neck,
and press their lips against the chin support. The
patient's nose must rest on top of the support and
their mouth must be closed, and their teeth must be
together.
The positioning lights for the midsagittal plane,
Frankfort plane (and TMJ position, depending on unit
configuration) show on the patient’s face. The lights
will automatically switch off after two minutes. To
switch them off earlier, press the positioning joystick.
To switch them back on, do one of the following:
a.Press the thumb wheel on the underside of the
patient support table.
b.Press any of the positioning controls (button or
joystick).
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic51
10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE
NOTE
10.3.3 Adjusting patient’s head position (Planmeca ProMax 2D S2 X-ray units)
The positioning lights and the illustrations on the touch
screen help you to position the patient’s head correctly.
The illustrations are for guidelines only.
1.To go to the next screen, select
•this symbol
OR
•the forward button.
2.Position the patient’s midsagittal plane so that it
coincides with the midsagittal plane light.
3.Position the patient’s Frankfort plane so that it
coincides with the Frankfort plane light. To do this,
adjust the tilt of the patient’s head by raising or
lowering the X-ray unit with the height adjusting
buttons. The patient’s back and neck must be
straight.
52Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE
NOTE
Adjust imaging position
•The Frankfort plane light is located inside the column.
The light’s position can be adjusted if needed. This is
done by rotating the thumb wheel below the light slot.
•FOR LATERAL EXPOSURES:
4.Use a ruler to measure the distance between the
layer light and the patient’s temporomandibular joint
so that you can determine the imaging position.
Adjust the imaging position with the minus or plus
sign according to your measurement.
FOR DOUBLE TMJ EXPOSURES:
You can set the X-ray unit so that the imaging position
is automatically moved forward for the open jaw
exposure (2/2). To do this, select Settings > Program
> 2200 Program Features > 2220 2D TMJ > TMJ 2/2
Shift and adjust the setting with the minus or plus sign
(e.g. 10 mm). The result is shown on the ruler (e.g.
1/2 = 77 mm and 2/2 = 67 mm).
5.Check that the midsagittal plane light and the
Frankfort plane light are still correctly positioned.
Reposition them if necessary.
•FOR POSTEROANTERIOR EXPOSURES:
It is not possible to adjust the imaging position.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic53
10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE
NOTE
10.3.4 Adjusting patient’s head position
(Planmeca ProMax 2D S3 & Planmeca ProMax 3D X-ray units)
The positioning lights and the illustrations on the touch
screen help you to position the patient’s head correctly.
The illustrations are for guidelines only.
1.To go to the next screen, select
•this symbol
OR
•the forward button.
2.Position the patient’s midsagittal plane so that it
coincides with the midsagittal plane light.
3.Position the patient’s Frankfort plane so that it
coincides with the Frankfort plane light. To do this,
adjust the tilt of the patient’s head by raising or
lowering the X-ray unit with the height adjusting
buttons. The patient’s back and neck must be
straight.
54Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE
NOTE
2
1
1
2
•The Frankfort plane light is located inside the column.
The light’s position can be adjusted if needed. This is
done by rotating the thumb wheel below the light slot.
4.Rotate the C-arm 90 degrees clockwise by selecting
the 90° button. This will give you a better view for
checking the TMJ light position.
Select the button again if you wish to rotate the
C-arm back to the original position.
5.Use the positioning joystick to adjust the TMJ light.
The TMJ light has to coincide with the patient’s
temporomandibular joint.
FOR DOUBLE TMJ EXPOSURES:
You can set the X-ray unit so that the imaging position is
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic55
automatically moved forward for the open jaw exposure (2/
2). To do this, select Settings > Program > 2200 Program
Features > 2220 2D TMJ > TMJ 2/2 Shift and adjust the
setting with the minus or plus sign (e.g. 10 mm). The result
is shown on the touch screen (e.g. 1/2 = 77 mm and 2/2 =
67 mm).
6.Check that the midsagittal plane light and the
Frankfort plane light are still correctly positioned.
Reposition them if necessary.
10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE
NOTE
10.4 Adjusting exposure values for current exposure
The exposure values have been preset at the factory for
each patient size and program type. The preset exposure
values are average values and they are only meant to
guide the user.
Always try to minimize the radiation dose to the patient.
The preset exposure values are shown in the following
tables.
Factory presets for lateral and lateral-PA TMJ exposures
PATIENT SIZEkV VALUEmA VALUE
Child (XS)624
Small adult (S)645
Medium-sized adult (M)666.3
Large adult (L)688
Extra large adult (XL)7010
Factory presets for posteroanterior TMJ exposures
PATIENT SIZEkV VALUEmA VALUE
Child (XS)644
Small adult (S)665
Medium-sized adult (M)686.3
Large adult (L)708
Extra large adult (XL)7210
You can adjust the preset exposure values (kV and mA)
as described in section 9.6 “Adjusting exposure values for
current exposure” on page 42.
56Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE
NOTE
NOTE
NOTE
1
10.5 Taking an exposure in double TMJ programs
Make sure that you have selected the correct patient and
exposure mode in the Planmeca Romexis program.
10.5.1 First exposure - jaw closed (1/2)
1.Select the radiation symbol or the forward button.
Green lights flash on the touch screen and exposure
button when the X-ray system is getting ready for an
exposure. The green lights stop flashing and stay on
continuously when the X-ray system is ready for an
exposure.
2.Ask the patient to stay as still as possible. Explain to
the patient that this is a double exposure and that the
C-arm will move twice around the patient’s head.
3.Move to a protected area.
4.Press and hold down the exposure button for the
duration of the first exposure.
The C-arm moves around the patient’s head. During
exposure yellow radiation warning lights illuminate on
the exposure switch and on the touch screen, and
you hear a radiation warning tone. Additionally, a
radiation warning symbol is shown on the touch
screen.
Maintain audio and visual contact with the patient and
X-ray unit during exposure. If the C-arm st ops moving
during exposure, or moves in an erratic way, release
the exposure button immediately.
If needed, you can stop the imaging process by
touching the red cross button after you have taken the
first exposure.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic57
10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE
1/2
1/2
2/2
2/2
10.5.2 Second exposure - jaw open (2/2)
5.Ask the patient to open their mouth as wide as
possible. Make sure that the patient’s upper lip is still
touching the chin support.
6.Select the forward button. Wait until the X-ray system
is ready, i.e. until the green lights stop flashing.
7.Press and hold down the exposure button for the
duration of the second exposure. The C-arm moves
around the patient’s head and the exposure is taken
in the same way as the first exposure.
8.The image is shown on the computer screen.
•Note that you must accept the image in the
Planmeca Romexis program.
9.Guide the patient away from the X-ray unit.
58Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE
NOTE
NOTE
1
1/32/3
3/3
10.6 Taking an exposure in three angle TMJ programs
Make sure that you have selected the correct patient and
exposure mode in the Planmeca Romexis program.
1.Select the radiation symbol or the forward button.
Green lights flash on the touch screen and exposure
button when the X-ray system is getting ready for an
exposure. The green lights stop flashing and stay on
continuously when the X-ray system is ready for an
exposure.
2.Ask the patient to stay as still as possible. Explain to
the patient that this is a multiple exposure and that
the C-arm will move through three exposure cycles.
3.Move to a protected area.
4.Press and hold down the exposure button for the
duration of the exposure.
During exposure yellow radiation warning lights
illuminate on the exposure switch and on the touch
screen, and you hear a radiation warning tone.
Additionally, a radiation warning symbol is shown on
the touch screen.
Maintain audio and visual contact with the patient and
X-ray unit during exposure. If the C-arm st ops moving
during exposure, or moves in an erratic way, release
the exposure button immediately.
5.The image is shown on the computer screen.
•Note that you must accept the image in the
Planmeca Romexis program.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic59
6.Guide the patient away from the X-ray unit.
11 SINUS EXPOSURE
NOTE
ProMax_2_3.eps
Bite1.eps
Lip1.eps
11 SINUS EXPOSURE
This procedure will produce an exposure of the maxillary
sinuses.
11.1 Before exposure
Insert the temple supports into the holes in the patient
support table.
Insert_temple_supports.eps
Ensure that you insert the temple supports the right way
round (i.e. with the convex side facing outward as shown).
Use these patient supports.
60Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
11 SINUS EXPOSURE
NOTE
11.2 Selecting exposure settings
Refer to section 6.9 “Touch screen” on page 16 for
general information on how to make or cancel selections
on the touch screen.
11.2.1 Selecting program
Select the sinus program (2D Dental > Sinus).
11.2.2 Selecting program type
Select the program type from the
•drop-down menu at the top
OR
•left of the screen.
Refer to section 7.3 “Sinus programs” on page 25 for
details.
The available options depend on the X-ray unit model.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic61
11 SINUS EXPOSURE
1
3
11.2.3 Selecting patient size
Select the patient size as described in section 9.2.3
“Selecting patient size” on page 34.
11.2.4 Selecting jaw side for lateral exposures
(Planmeca ProMax 2D S3 & Planmeca ProMax 3D X-ray units)
Use this button to select the right or left jaw side.
11.3 Patient positioning
11.3.1 Selecting patient entry position
11.3.2 Positioning patient’s head
Select the patient entry position as described in section
9.3.1 “Selecting patient entry position” on page 36.
1.Press the temple support button to open the temple
supports if they are not already open.
2.Guide the patient to the X-ray unit.
3.Adjust the X-ray unit to suit the height of the patient.
To do this, press either of the height adjusting
buttons until the opening in the chin support is
approximately level with the patient’s mouth.
62Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
11 SINUS EXPOSURE
NOTE
a.)
b.)
4.Ask the patient to step forward, grasp the patient
handles, stretch and straighten their back and neck,
and press their lips against the chin support. The
patient's nose must rest on top of the support and
their mouth must be closed.
The positioning lights for the midsagittal and
Frankfort plane show on the patient’s face. The lights
will automatically switch off after two minutes. To
switch them off earlier, press the positioning joystick.
patposlights3.eps
11.3.3 Adjusting patient’s head position
The positioning lights and the illustrations on the touch
screen help you to position the patient’s head correctly.
The illustrations are for guidelines only.
1.To go to the next screen, select
To switch them back on, do one of the following:
a.Press the thumb wheel on the underside of the
patient support table.
b.Press any of the positioning controls (button or
joystick).
•this symbol
OR
•the forward button.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic63
11 SINUS EXPOSURE
30°
2.Position the patient’s midsagittal plane so that it
coincides with the midsagittal plane light.
3.Position the patient’s Frankfort plane as follows.
•FOR POSTEROANTERIOR AND LATERAL
PROGRAMS:
Position the patient’s Frankfort plane so that it
coincides with the Frankfort plane light. To do this,
adjust the tilt of the patient’s head by raising or
lowering the X-ray unit with the height adjusting
buttons. The patient’s back and neck must be
straight.
•FOR POSTEROANTERIOR LINEAR PROGRAM:
Position the patient’s head so that the Frankfort plane
is tilted up about 30 degrees. To do this, support the
back of the patient’s head with your hand and, using
64Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
the Frankfort plane light as a reference line, adjust
the tilt of the patient’s head by raising or lowering the
X-ray unit with the height adjusting buttons. The
patient’s back and neck must be straight.
11 SINUS EXPOSURE
NOTE
•The Frankfort plane light is located inside the column.
The light’s position can be adjusted if needed. This is
done by rotating the thumb wheel below the light slot.
11.4 Adjusting exposure values for current exposure
The exposure values have been preset at the factory for
each patient size and program type. The preset exposure
values are average values and they are only meant to
guide the user.
Always try to minimize the radiation dose to the patient.
The preset exposure values are shown in the following
tables.
Factory presets for posteroanterior sinus exposures
PATIENT SIZEkV VALUEmA VALUE
Child (XS)724
Small adult (S)745
Medium-sized adult (M)766.3
Large adult (L)788
Extra large adult (XL)8010
Factory presets for lateral sinus exposures
PATIENT SIZEkV VALUEmA VALUE
Child (XS)604
Small adult (S)624
Medium-sized adult (M)644.5
Large adult (L)665
Extra large adult (XL)685
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic65
You can adjust the preset exposure values (kV and mA)
as described in section 9.6 “Adjusting exposure values for
current exposure” on page 42.
11 SINUS EXPOSURE
NOTE
NOTE
1
11.5 Taking an exposure
Make sure that you have selected the correct patient and
exposure mode in the Planmeca Romexis program.
1.Select the radiation symbol or the forward button.
Green lights flash on the touch screen and exposure
button when the X-ray system is getting ready for an
exposure. The green lights stop flashing and stay on
continuously when the X-ray system is ready for an
exposure.
2.Ask the patient to stay as still as possible.
3.Move to a protected area.
4.Press and hold down the exposure button for the
duration of the exposure.
The C-arm moves around the patient’s head. During
exposure yellow radiation warning lights illuminate on
the exposure switch and on the touch screen, and
you hear a radiation warning tone. Additionally, a
radiation warning symbol is shown on the touch
screen.
Maintain audio and visual contact with the patient and
X-ray unit during exposure. If the C-arm stops moving
during exposure, or moves in an erratic way, release
the exposure button immediately.
5.The image is shown on the computer screen.
•Note that you must accept the image in the
Planmeca Romexis program.
66Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
6.Guide the patient away from the X-ray unit.
12 SETTINGS
NOTE
NOTE
12 SETTINGS
Some of the settings can be used to alter the operation of
the X-ray unit. Never use functions that you are not familiar
with.
The contents of the displays depend on the unit
configuration. The displays shown here are only
examples.
Select the settings symbol on the main screen to enter the
settings menu.
Settings that can be entered by the user:
•User
•Program
•About
Settings that can be entered by service personnel only
(password required):
•Technical
To return to the main screen, select the settings symbol at
the top left corner.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic67
12 SETTINGS
12.1 User settings
12.1.1 Language (1100)
•To change language:
1.Select User > 1100 Language.
2.Select the language you wish to use.
3.Select the green check mark button.
12.1.2 Time and Date (1200)
•To set time display format:
1.Select User > 1200 Time and Date > 1210 Set
System Time and Time / Date Display Format > Time
Display Format.
2.Select the display format you wish to use.
3.Select the green check mark button.
68Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
12 SETTINGS
NOTE
•To set date display format:
1.Select User > 1200 Time and Date > 1210 Set
System Time and Time / Date Display Format > Date
Display Format.
2.Select the display format you wish to use.
3.Select the green check mark button.
•To set time:
1.Select User > 1200 Time and Date > 1210 Set
System Time and Time / Date Display Format >
Change System Time.
2.Use the plus and minus buttons to change the time.
3.Select the green check mark button.
The time is set to the local time at the factory. Change the
time setting to show the correct time before you start using
the X-ray unit.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic69
12 SETTINGS
•To set date:
1.Select User > 1200 Time and Date > 1220 Change
System Date.
2.Select the day or use the arrow buttons to change the
month (single arrow) or year (double arrow).
3.Select the green check mark button.
12.1.3 Operational Settings (1300)
•To select mode:
1.Select User > 1300 Operational Settings > 1310 Use
Mode.
2.Select the mode you wish to use.
In demo mode you can practice or demonstrate the
functions of the X-ray unit without radiation and PC
connection.
3.Select the green check mark button.
70Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
12 SETTINGS
•To control audio settings:
1.Select User > 1300 Operational Settings > 1320
Audio Settings.
2.Select
•Alarm Volume
This setting adjusts the volume of the radiation
warning tone. Use the minus or plus button to
reduce or increase the volume.
•Touch Volume
This setting adjusts the volume of the audible
signal that you hear when you make a selection
on the touch screen. Use the minus or plus
button to reduce or increase the volume. Set the
volume level to 0% if you do not wish to use this
function.
3.Select the green check mark button.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic71
12 SETTINGS
NOTE
•To manage settings for patient positioning:
1.Select User > 1300 Operational Settings > 1330
Patient Positioning.
2.Turn an option(s) ON or OFF:
•Side Entry
Turn this option OFF if you do not wish to use the
open (full view) patient entry position. This might
be necessary if there is no space for the C-arm
to move back.
Planmeca ProMax 2D S2 X-ray units offer only one
patient entry position. This entry position is the closed
patient entry position where the C-arm is positioned
around the temple supports.
•Midsagittal and Frankfort Lights in Tomo
Turn this option ON if you wish to use the
midsagittal and Frankfort plane lights when you
take 2D tomographic or 3D exposures.
•Temple Supports
Turn this option OFF if you do not wish to use the
temple supports.
•Return C-arm Automatically
Turn this option ON if you wish to set the X-ray
unit so that the C-arm will automatically return to
the starting position at the end of an exposure
cycle. Note, however, that the automatic function
works only if the exposure button is pressed and
held down for the entire duration of the
exposure.
3.Select the green check mark button.
72Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
12 SETTINGS
•To control exposure settings:
1.Select User > 1300 Operational Settings > 1340
Exposure Settings.
2.Select
•Panoramic System
On 3D X-ray units you can use either the Dimax
sensor or the 3D sensor for taking 2D
exposures. Select the Dimax or SmartPan (3D
sensor) system and change the sensor as
described in section 8.1 “Attaching and removing
sensor” on page 27.
•Continuous Exposure Indicator
Turn this option ON if you wish to set the X-ray
unit so that the yellow radiation warning light
stays on continuously (instead of flashing) when
you take an exposure.
Note that the setting affects all indicator lights
that are connected to your X-ray system
(indicator light on the hand-held exposure
switch, wall exposure switch(es) and remote
exposure lamp).
3.Select the green check mark button.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic73
12 SETTINGS
NOTE
12.1.4 Network Settings (1400)
•To view network settings:
1.Select User > 1400 Network Settings.
2.Select the network settings you wish to view.
3.Select the green check mark button.
Only a service technician or local administrator may
change the network settings.
74Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
12 SETTINGS
12.1.5 Testing Routines (1500)
•To take a test exposure:
1.Select User > 1500 Testing Routines > 1510 Test
Exposure.
2.Use the minus or plus buttons to set the exposure
values you wish to use.
3.Select the SET button.
4.Move to a protected area.
5.Press and hold down the exposure button for the
duration of the exposure. The C-arm will not move
when you take a test exposure.
6.Select the green check mark button.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic75
12 SETTINGS
NOTE
NOTE
•To perform tube head seasoning:
1.Select User > 1500 Testing Routines > 1520 Tube
Head Seasoning.
This option allows you to warm up the X-ray tube, i.e.
run a tube head seasoning process. This is
necessary if the X-ray unit has not been used for a
week or more and / or if you receive error message
E332 (Severe arcing across X-ray tube).
2.Select the Start button.
3.Move to a protected area.
4.Press the exposure button when the word Ready
appears.
You can press and hold down the exposure button for
the whole duration of the process or remove your
thumb from the exposure button when the word Wait
appears. Note that the seasoning process will take
several minutes. After a successful process the
message OK is shown.
5.Select the green check mark button.
Contact your service technician for help if error message
E332 (Severe arcing across X-ray tube) reappears after a
successful seasoning process.
12.1.6 Clinic Management (1600)
•To view network settings for Clinic Management:
Select User > 1600 Clinic Management to view the
network settings for the Planmeca Romexis Clinic
Management module.
Only a service technician or local administrator may
change the settings.
76Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
12 SETTINGS
NOTE
NOTE
12.2 Program settings
12.2.1 Programs (2100)
•To turn a program(s) ON or OFF:
1.Select Program > 2100 Programs.
2.Select a program group (e.g. 2D Panoramic).
3.Turn a program type(s) (e.g. Interproximal) ON or
OFF.
4.Select the green check mark button.
The function is available for programs that have been
activated in menu 2300 Licences.
•To permanently adjust preset exposure values:
1.Select Program > 2100 Programs.
2.Select a program group (e.g. 2110 2D Panoramic).
3.Select a program type (e.g. Interproximal).
4.Select the exposure values you wish to adjust (e.g.
66 kV / 9 mA for patient size M).
•For 2D tomographic programs:
In combined programs (cross-sectional +
longitudinal) only the cross-sectional exposure
values can be adjusted.
•For 3D X-ray units:
If you use both sensors (Dimax and 3D) for
taking 2D exposures, make sure that the
panoramic system in menu 1340 is set to the
option (Dimax or SmartPan) you wish to adjust.
•For 3D X-ray units:
In 2D panoramic programs with a 3D sensor (2D
Panoramic / SmartPan /) select also the
MultiView button if you wish to adjust the presets
for the MultiView imaging mode.
•For 3D X-ray units:
In 3D programs the exposure values are given
separately for each image resolution. The image
resolutions that are not available are shown with
faded buttons. Select also the ULD (Ultra Low
Dose) button if you wish to adjust the presets for
the ULD function.
5.Use the minus or plus buttons to set the exposure
values you wish to use.
6.Select the green check mark button.
7.Repeat for an other program type, patient size or
image resolution (3D) if needed.
8.Select the green check mark button.
Always try to minimize the radiation dose to the patient.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic77
12 SETTINGS
NOTE
NOTE
You can restore the exposure values that have been
preset at the factory (i.e. overrule your own settings) by
selecting Program > 2500 Reset to Factory Defaults.
You can adjust the preset exposure values temporarily as
described in section 9.6 “Adjusting exposure values for
current exposure” on page 42.
12.2.2 Program Features (2200)
•To manage program settings:
Select Program > 2200 Program Features to manage
program settings. For details on a specific setting, refer to
the manual section that contains the corresponding
function.
78Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
12 SETTINGS
12.2.3 Licences (2300)
•To activate a program licence:
1.Select Program > 2300 Licences.
2.Select the licence you wish to activate.
3.Enter the licence code that you have received for this
licence on this X-ray unit.
4.Select the green check mark button.
5.Repeat for an other program licence if needed.
6.Select the green check mark button.
12.2.4 Reset to Factory Defaults (2500)
•To reset to factory defaults:
1.Select Program > 2500 Reset to Factory Defaults.
2.Select the green check mark button.
The function will restore the exposure values that
have been preset at the factory (i.e. overrule your
own settings in menu Programs (2100)).
The preset exposure values are shown in sections
“Adjusting exposure values for current exposure”.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic79
12 SETTINGS
12.3 About tab
12.3.1 Component Information (4100)
•To view component information:
Select About > 4100 Component Information to view the
set-up or current software versions of the X-ray unit.
12.3.2 Archive (4200)
•To view error history:
Select About > 4200 Archive > Error History to view a list
of the error messages that have been generated by the Xray unit. The errors are shown in chronological order with
the latest error message on top.
•To view exposure statistics:
Select About > 4200 Archive > Exposure Statistics to view
statistical data about the X-ray unit.
80Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
12 SETTINGS
12.3.3 Product Registration (4300)
•To register the X-ray unit on Planmeca’s website:
1.Select About > 4300 Product Registration.
2.Do one of the following:
a.A QR (Quick Response) code is shown on the
screen. If you have a QR code reader installed
on your mobile device (e.g. smartphone), hold
the device steady over the QR code. You will be
directed to Planmeca’s product registration
page.
b.Go to Planmeca’s product registration page at
www.planmeca.com/register.
3.Follow the instructions on the registration page.
Note that when you enter the X-ray unit serial
number, you have to include any letters shown at the
beginning of the number.
4.Select the green check mark button.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic81
13 HELP MESSAGES
13 HELP MESSAGES
The X-ray unit incorporates a self-checking feature that
monitors the operation of the unit. If the system detects an
operating error, a help message (e.g. H101) appears on
the touch screen.
The X-ray unit will not accept any commands from the
user until the help message is cleared from the touch
screen. Clear the message by touching the green check
mark.
The following list shows, in numerical order, all the help
messages that can appear.
CodeExplanationComments
H101Exposure switchThe exposure button was released
before end of exposure.
H102The exposure button is stuck or the
cable is short circuited.
H103ProTouch desktop
application
H105Emergency stop
button
H115DECDEC is receiving too much radiation.Change the exposure values.
H116DEC is receiving too little radiation.Change the exposure values.
The ProTouch desktop application
needs to be closed.
The emergency stop button has been
activated.
Guide the patient away from the
X-ray unit before moving the
C-arm.
Press and hold down the
exposure button for the entire
duration of the exposure.
Release the exposure button.
If necessary, contact your
service technician to replace the
exposure switch.
Restart the application.
All movements of the X-ray unit
are blocked, no radiation is
generated.
Guide the patient away from the
X-ray unit. Then release the
emergency stop button to
resume normal operation.
H130Patient safety areaPatient safety area violation detected.Check the scan and layer
settings.
H131TomographyPatient safety area violation in tomo
slice 1.
H132Patient safety area violation in tomo
slice 2.
H133Patient safety area violation in tomo
slice 3.
H134Patient safety area violation in tomo
slice 4.
82Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
Check the slice count and / or
step between slices.
Check the slice count and / or
step between slices.
Check the slice count and / or
step between slices.
Check the slice count and / or
step between slices.
13 HELP MESSAGES
CodeExplanationComments
H142Height movementHeight movement is not possible
because the stop plate at the bottom of
the column was activated.
H144Height movement is not possible
because one (or more) of the
positioning control buttons or the
positioning joystick is stuck.
H151Line voltageThe line voltage was too low during
exposure.
H152The line voltage is too low.Exposure is not possible.
H161TemperatureThe temperature of the tube head is too
high.
H162The temperature of the lift motor is too
high.
H165The temperature of the tube head is too
high for the selected exposure values.
H166The maximum tube head energy level
was exceeded.
H170User related
H171The sensor is not attached properly to
H172The sensor is not attached properly to
messages
Wrong licence code.Check the licence code.
the C-arm.
the cephalostat.
Clear any obstruction before
moving the column again.
Release the button / joystick.
Exposure was interrupted.
Contact your service technician
for help.
Contact your service technician
for help.
Wait for a few minutes for the
tube head to cool down.
Wait for a few minutes for the lift
motor to cool down.
Wait for a few minutes for the
tube head to cool down.
Wait for a few minutes for the
tube head to cool down or use
lower exposure values.
Attach and / or lock the sensor
in position.
Attach and / or lock the sensor
in position.
H175PC program selection is in conflict with
the selected X-ray unit program.
H180DEC is not available.
H181The imaging process was cancelled in
Planmeca Romexis.
H182Timeout in image data transmission.Exposure was interrupted.
H183The attached sensor is not suitable for
the selected program.
H184Remove the 3D sensor.
H185The 3D sensor is not attached properly. Attach and / or lock the sensor
H186No IP address defined for 3D sensor.
H187Problem during image data
transmission.
H189The screen was touched during
exposure.
Select another exposure mode
in Planmeca Romexis.
Contact your service technician
for help.
Change the sensor.
in position.
Exposure was interrupted.
Contact your service technician
for help.
Exposure was interrupted.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic83
14 ERROR MESSAGES
NOTE
CodeExplanationComments
H193Invalid scan settings.Retry or contact your service
technician for help.
14 ERROR MESSAGES
Contact your service technician for help if you receive an
error message.
The X-ray unit incorporates a self-checking feature that
monitors the operation of the unit. If the system detects a
technical fault, an error message (e.g. E201) appears on
the touch screen.
An error message indicates that the X-ray unit has a
problem that needs to be solved before further exposures
can be taken. The X-ray unit will not accept any
commands from the user until the error message is
cleared from the touch screen. Guide the patient away
from the X-ray unit. Then clear the message by touching
the green check mark.
84Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
15 CLEANING AND DISINFECTION
NOTE
NOTE
NOTE
NOTE
temple_support.ai
Chin2.eps
Chin1.eps
Lip1.eps
Bite1.eps
15 CLEANING AND DISINFECTION
Switch the X-ray unit off before cleaning or disinfection.
Do not apply sprays or liquids directly on the surfaces.
Apply sparingly to a clean soft cloth and wipe the surface
with the cloth.
Contact your service technician for help if sprays or liquids
enter the system.
Follow the instructions provided by the manufacturer of the
cleaning solution, disinfectant and autoclave.
FOR PLANMECA PROMAX 3D X-RAY UNITS:
Refer to the User’s Manual for 3D imaging.
15.1 Patient supports and patient handles
Wipe these parts after each patient using a mild surface
disinfectant and a clean soft cloth.
Use a mild cleaning solution and a clean soft cloth for
cleaning stains and dirt if needed.
•Panoramic patient supports
Bite2.eps
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic85
15 CLEANING AND DISINFECTION
•Cephalometric patient supports
•Impression plate holder
•Patient handles
86Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
15 CLEANING AND DISINFECTION
NOTE
temple_support.ai
Bite2.eps
Chin1.eps
Lip1.eps
These parts can be autoclaved at 134°C (273°F) for 3
minutes. They can be autoclaved up to 100 times.
Chin2.eps
Bite1.eps
15.2 Touch screen
Wipe the touch screen regularly using a clean soft cloth
that has been slightly dampened with water.
Do not use any cleaning solutions.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic87
15 CLEANING AND DISINFECTION
NOTE
NOTE
Dimax_sensor.eps
Exp_sw_wall6_2.eps
3 x
15.3 Sensor
Wipe the sensor surfaces regularly using a clean soft
cloth that has been slightly dampened with water.
Do not use any cleaning solutions.
15.4 Other surfaces
Wipe the other surfaces regularly using a mild cleaning
solution and a clean soft cloth.
FOR TOUCH SCREEN AND SENSOR:
Refer to sections 15.2 “Touch screen” on page 87 and
15.3 “Sensor” on page 88.
Hand_imaging_support.eps
Accessory_box1.eps
88Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
16 SERVICE
NOTE
16 SERVICE
The X-ray unit must be checked by a qualified Planmeca
service technician once a year or after every 10 000
exposures (if this is sooner). This will guarantee patient
and user safety and ensure consistent image quality.
The annual maintenance service includes inspection of
the following:
•exposure switch
•exposure indicator lights and warning signals
•emergency stop button
•X-ray unit adjustments and quality control checks
•column motor nut
•labels
FOR SERVICE TECHNICIANS:
Refer to the Planmeca Technical Bulletin PX-081 for
details.
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic89
17 DISPOSAL
17 DISPOSAL
In order to reduce the environmental load over the
product’s entire lifecycle, Planmeca’s products are
designed to be as safe as possible to manufacture, use
and dispose of.
Parts which can be recycled should always be taken to
the appropriate processing centers, after hazardous
waste has been removed. Disposal of obsolete units is
the responsibility of the waste possessor.
All parts and components containing hazardous materials,
as well as batteries, must be disposed of in accordance
with waste legislation and instructions issued by the
environmental authorities. Batteries must be disposed of
in compliance with the requirements of Directive 2006 /
66 / EEC.
The risks involved and the necessary precautions must be
taken into account when handling waste products.
Recyclable
Part
Frame, covers &
patient supports
- metal
- plastic
Motors(X)
Component
boards
Cables,
transformers
Main materials
for disposal
Aluminium,
galvanized steel,
lead
PUR,
other plastics
Copper,
steel,
transformer oil
X
X
X
(X)
X
X
material
(X) =
if available
Waste
disposal site
X
Hazardous
waste
(separate
collection)
X
X
X-ray tubeX
PackingWood,
cardboard,
paper,
polystyrene
SensorReturn sensor to Planmeca.
Other partsX
90Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
X
X
X
X
18 TECHNICAL SPECIFICATIONS
18 TECHNICAL SPECIFICATIONS
18.1 Technical data for Planmeca ProMax product family
Classification
Medical Device Directive93/42/EEC (Class IIb)
RoHS2011/65/EU
IEC 60601-1Class I, type B
CISPR 11Class B
IP ClassificationIP20
Applied parts(according to IEC 60601-1: 2012)
Patient supportsAs shown in section Patient supports
in User Manuals
Patient handles
Generator(according to IEC 60601-2-7: 1998)
Resonant-mode, DSP-controlled, 80 - 160 kHz
X-ray tube
2D / 3D s / 3D ClassicToshiba D-054SB
3D Plus / 3D MidToshiba D-054SB, Toshiba D-059SBR or
SXR 130-10-0.5 SC
3D MaxToshiba D-067SB, Toshiba D-059SBR or
SXR 130-10-0.5 SC
Focal spot size(according to IEC 60336: 2005)
2D / 3D s / 3D Classic / 3D Plus /
3D Mid or 3D Max with Toshiba
D-059SBR or SXR 130-10-0.5 SC
3D Max with Toshiba D-067SB0.6 x 0.6 mm
Filtration
Pan / cephTotal 2.5 mm Al
0.5 x 0.5 mm
3DTotal 2.5 mm Al + 0.5 mm Cu
SmartPan3D s / 3D Classic / 3D Plus / 3D Mid:
total 2.5 mm Al
3D Max: total 2.5 mm Al + 0.5 mm Cu
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic91
18 TECHNICAL SPECIFICATIONS
Tube housing front cover
quality equivalent filtration
(not included in the specified total
filtration)
Anode voltage
Pan / SmartPan60 - 84 kV ±5%
Ceph60 - 84 kV ±5%
3D s / 3D Classic /
3D Plus or 3D Mid with Toshiba
D-054SB
3D Max with Toshiba D-067SB60 - 96 kV ±5%
3D Plus / 3D Mid / 3D Max
with Toshiba D-059SBR or
SXR 130-10-0.5 SC
Anode current
Pan1 - 16 mA ±10%
Ceph1 - 16 mA ±10%
0.3 mm Al @ 70 kV / HVL 2.6 mm Al
60 - 90 kV ±5%
60 - 120 kV ±5%
3D s / 3D Classic / 3D Plus or
3D Mid with Toshiba D-054SB or
3D Max with Toshiba D-067SB
3D Plus / 3D Mid / 3D Max
with Toshiba D-059SBR
3D Plus / 3D Mid / 3D Max
with Superior SXR 130-10-0.5 SC
3D1 - 14 mA ±10%
Pan / SmartPan1 - 16 mA ±10%
Scanning ceph1 - 16 mA ±10%
Planmeca ProCeph16 mA ±10%
3D1 - 14 mA ±10%
Pan / SmartPan1 - 14 mA ±10%
Scanning ceph1 - 14 mA ±10%
Planmeca ProCeph14 mA ±10%
3D1 - 14 mA ±10%
Pan / SmartPan1 - 16 mA ±10%
Scanning ceph1 - 16 mA ±10%
Planmeca ProCeph16 mA ±10%
mAs range
92Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
18 TECHNICAL SPECIFICATIONS
min. / max. as indicated ±(10% + 0.2 mAs)
mGy range
min. / max. as indicated ±40%
Linearity of radiation output
< 0.1
DEC accuracy
±10%
Cooling period
Automatically controlled
Exposure time
Pan2.7 - 16 s as indicated ±10%
SmartPan2.5 - 15.6 s as indicated ±10%
Scanning ceph6.7 - 10.5 s as indicated ±10%
Planmeca ProCeph0.1 - 0.8 s as indicated ±10%
3DPulsed, effective 3 - 36 s as indicated ±10%
SID
Pan2D / 3D s / 3D Classic: 501 mm (19.7 in.)
3D Plus / 3D Mid: 574 mm (22.6 in.)
Ceph1700 mm (66.9 in.)
3D / SmartPan3D s / 3D Classic:
528 mm (20.7 in.)
3D Plus or 3D Mid with Toshiba D-054SB /
3D Max with Toshiba D-067SB:
600 mm (23.6 in.)
3D Plus / 3D Mid / 3D Max with
Toshiba D-059SBR or SXR 130-10-0.5 SC:
632 mm (23.8 in.)
Magnification
Pan2D / 3D s / 3D Classic: 1.2 - 1.5
2D tomography: 1.5
3D Plus / 3D Mid: 1.4
SmartPan3D s / 3D Classic: 1.27
3D Plus / 3D Mid / 3D Max: 1.4
Ceph1.13
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic93
18 TECHNICAL SPECIFICATIONS
3D3D s / 3D Classic: 1.58
3D Plus / 3D Mid / 3D Max:
1.38, 1.40, 1.42, 1.43 or 1.8
Duty cycle for height adjustment
25 s ON / 400 s OFF
Line voltage
100 - 220 V~ / 50 - 60 Hz
230 - 240 V~ / 50 Hz
Line current
8 - 17 A
Line harmonics
Cos better than 0.9
Max. permissible apparent impedance of supply mains
Max. continuous heat dissipation
Internal fuse(s)
User replaceable
- 1 fuse on permanently
installed X-ray units
- 2 fuses on X-ray units with
detachable power supply cord
Type
External fuse(s)
Battery
Max. weight
0.5 (100 VAC)
250 W
100 - 220 V~ / 16A FF H 500 V
230 - 240 V~ / 8A FF H 500 V
195100 ELU
100 - 220 V ~ / 16A T 250 V
230 - 240 V ~ / 10A T 250 V
Lithium battery: CR2032, Panasonic
2D / 3D s / 3D Classic119 kg (262 lb)
3D Plus / 3D Mid141 kg (311 lb)
3D Max 139 kg (306 lb)
Scanning ceph26 kg (57 lb)
Planmeca ProCeph20 kg (44 lb)
Environmental requirements
94Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
18 TECHNICAL SPECIFICATIONS
Transport:
Temperature-20°C - +60°C (-4°F - +140°F)
Relative humidity10 - 90% RH (non-condensing)
Air pressure700 - 1060 hPa
Storage:
Temperature-10°C - +50°C (+14°F - +122°F)
Relative humidity10 - 90% RH (non-condensing)
Air pressure700 - 1060 hPa
Operating:
TemperaturePan / scanning ceph:
+10°C - +40°C (+50°F - +104°F)
3D / ProCeph: +10°C - +30°C (+50°F - +86°F)
Relative humidity10 - 90% RH (non-condensing)
Air pressure800 - 1060 hPa
Max. altitude2000 m (1.25 miles)
Image properties
Pan / ceph CCD:
Pixel size48 m
Pan active surface6 x 146 mm (0.24 x 5.74 in.)
Ceph active surface6 x 292 mm (0.24 x 11.15 in.)
Planmeca ProCeph:
Flat panel pixel size139 m
Flat panel active surface302 x 249 mm (11.89 x 9.80 in.)
3D:
Flat panel pixel size127 m
Flat panel active surface3D s: 80 x 130 mm (3.15 x 5.12 in.)
3D Classic / 3D Plus:
130 x 130 mm (5.12 x 5.12 in.)
3D Mid: 146 x 146 mm (5.74 x 5.74 in.)
3D Max: 193 x 242 mm (7.6 x 9.5 in.)
SmartPan:
Flat panel pixel size127 m
User’s Manual (2D Imaging)Planmeca ProMax 2D & 3D s & 3D Classic95
18 TECHNICAL SPECIFICATIONS
Flat panel active surface3D s / 3D Classic / 3D Plus:
8 - 25 x 130 mm (0.31 - 0.98 x 5.12 in.)
3D Mid: 8 - 25 x 146 mm (0.31 - 0.98 x 5.74 in.)
3D Max: 13 - 25 x 162 mm (0.51 - 0.98 x 6.38 in.)
Operating conditions for X-ray units with ProFace sensor
Optimum colour temperatureApprox. 6500 Kelvin
Frequency for fluorescent lamps100 Hz
Even and uniform lighting
No natural light(no windows in the room)
No green objects next to X-ray
18.2 Original manufacturer
PLANMECA Oy, Asentajankatu 6, FIN-00880 Helsinki, FINLAND
Phone: +358 20 7795 500, Fax: +358 20 7795 555, www.planmeca.com
96Planmeca ProMax 2D & 3D s & 3D ClassicUser’s Manual (2D Imaging)
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.