Piezotronics 378C10 Installation And Operating Manual

Model 378C10
ICP Microphone System
Installation and Operating Manual
This manual contains the 377C10, 426B03 installation and operating manuals that
comprise a Model 378C10 ICP Microphone System kit.
For assistance with the operation of this product,
contact PCB Piezotronics, Inc.
Toll-free: 800-828-8840
Fax: 716-684-0987
E-mail: info@pcb.com
Web: www.pcb.com
Warranty, Service, Repair, and
Return Policies and Instructions
The informati on contained in this document supersedes all similar i nformation th at may be found elsewhere in this manual.
Total Customer Satisfaction – PCB
Piezotronics guarantees Total Customer Satisfaction. If, at any time, for any reason, you are not completely satisfied with any PCB product, PCB will repair, replace, or exchange it at no charge. You may also choose to have your purchase price refunded in lieu of the repair, replacement, or exchange of the product.
Service – Due to the sophisticated nature of the sensors and associated instrumentation provided by PCB Piezotronics, user servicing or repair is not recommended and, if attempted, may void the factory warranty. Routine maintenance, such as the cleaning of electrical connectors, housings, and mounting surfaces with solutions and techniques that will not harm the physical material of construction, is acceptable. Caution should be observed to insure that liquids are not permitted to migrate into devices that are not hermetically sealed. Such devices should only be wiped with a dampened cloth and never submerged or have liquids poured upon them.
Repair – In the event that equipment becomes damaged or ceases to operate, arrangements should be made to return the equipment to PCB Piezotronics for repair. User servicing or repair is not recommended and, if attempted, may void the factory warranty.
Calibration – Routine calibration of sensors and associated instrumentation is
recommended as this helps build confidence in measurement accuracy and acquired data. Equipment calibration cycles are typically established by the users own quality regimen. When in doubt about a calibration cycle, a good “rule of thumb” is to recalibrate on an annual basis. It is also good practice to recalibrate after exposure to any severe temperature extreme, shock, load, or other environmental influence, or prior to any critical test.
PCB Piezotronics maintains an ISO­9001 certified metrology laboratory and offers calibration services, which are accredited by A2LA to ISO/IEC 17025, with full traceablility to N.I.S.T. In addition to the normally supplied calibration, special testing is also available, such as: sensitivity at elevated or cryogenic temperatures, phase response, extended high or low frequency response, extended range, leak testing, hydrostatic pressure testing, and others. For information on standard recalibration services or special testing, contact your local PCB Piezotronics distributor, sales representative, or factory customer service representative.
Returning Equipment – Following these procedures will insure that your returned materials are handled in the most expedient manner. Before returning
any equipment to PCB Piezotronics, contact your local distributor, sales representative, or factory customer service representative to obtain a Return
Materials Authorization (RMA) Number. This RMA number should be clearly marked on the outside of all package(s) and on the packing list(s) accompanying the shipment. A detailed account of the nature of the problem(s) being experienced with the equipment should also be included inside the package(s) containing any returned materials.
A Purchase Order, included with the returned materials, will expedite the turn-around of serviced equipment. It is recommended to include authorization on the Purchase Order for PCB to proceed with any repairs, as long as they do not exceed 50% of the replacement cost of the returned item(s). PCB will provide a price quotation or replacement recommendation for any item whose repair costs would exceed 50% of replacement cost, or any item that i s not economically feasible to repair. For routine calibration services, the Purchase Order should include authorization to proceed and return at current pricing, which can be obtained from a factory customer service representative.
Warranty – All equipment and repair services provided by PCB Piezotronics, Inc. are covered by a limited warranty against defective material and workmanship for a period of one year from date of original purchase. Contact
DOCUMENT NUMBER: 21354 DOCUMENT REVISION: B ECN: 17900
PCB for a complete statement of our warranty. Expendable items, such as batteries and mounting hardware, are not covered by warranty. Mechanical damage to equipment due to improper use is not covered by warranty. Electronic circuitry failure caused by the introduction of unregulated or improper excitation power or electrostatic discharge is not covered by warranty.
Contact Information – International customers should direct all inquiries to their local distributor or sales office. A complete list of distributors and offices can be found at www.pcb.com. Customers within the United States may contact their local sales representative or a factory customer service representative. A complete list of sales representatives can be found at www.pcb.com. Toll-free telephone numbers for a factory customer service representative, in the division responsible for this product, can be found on the title page at the front of this manual. Our ship to address and general contact numbers are:
PCB Piezotronics, I nc. 3425 Walden Ave. Depew, NY 14043 USA Toll-free: (800) 828-8840 24-hour SensorLine
SM
: (716) 684-0001 Website: www.pcb.com E-mail: info@pcb.com
Model Number
378C10
ICP MICROPHONE SYSTEM
Revision: A
ECN #: 42896
[9]
Performance
ENGLISH
SI
Nominal Microphone Diameter
1/4"
1/4"
Frequency Response Characteristic
Pressure
Pressure
Open Circuit Sensitivity
1.0 mV/Pa
1.0 mV/Pa
[3]
Open Circuit Sensitivity(+/-3 dB)
-60 dB re 1 V/Pa -60 dB re 1 V/Pa
[3]
Frequency Range(+/-2 dB)
5 to 70,000 Hz 5 to 70,000 Hz
Lower Limiting Frequency
0.75 to 4.0 Hz 0.75 to 4.0 Hz
Inherent Noise(Linear)
<60 dB re 20 µPa <60 dB re 20 µPa
[4]
Inherent Noise
<53 dB(A) re 20
µPa
<53 dB(A) re 20
µPa
[5]
Dynamic Range(3% Distortion Limit)
>174 dB re 20 µPa >174 dB re 20 µPa
Dynamic Range(Maximum without Clipping)
>177 dB re 20 µPa >177 dB re 20 µPa
[6]
TEDS Compliant
Yes
Yes
[7]
Environmental
Temperature Range(Operating)
-40 to +158 °F -40 to +70 °C
Temperature Coefficient of Sensitivity(+14 to +158°F (-10 to +70°C)) 0.004 dB/°F
0.007 dB/°C
[2][3]
Static Pressure Coefficient
-0.003 dB/kPa -0.003 dB/kPa
[2][3]
Humidity Coefficient of Sensitivity(0 to 100%, non-condensing)
± 0.0014
± 0.0014
[3]
Influence of Axial Vibration(0.1g (1 m/s²))
66 dB re 20 µPa 66 dB re 20 µPa
[2]
Electrical
Polarization Voltage
0 V
0 V
[1]
Excitation Voltage
20 to 32 VDC 20 to 32 VDC
Constant Current Excitation
2 to 20 mA
2 to 20 mA
Output Bias Voltage
10 to 14 VDC 10 to 14 VDC
Output Impedance
<50 Ohm
<50 Ohm
Maximum Output Voltage
+/-8 Vpk
+/-8 Vpk
Physical
Housing Material
Stainless Alloy Stainless Alloy
Venting
Rear
Rear
[8]
Electrical Connector
10-32 Jack
10-32 Jack
Mounting Thread(Grid)
0.25 - 60 UNS 0.25 - 60 UNS
Size (Diameter x Height)(with grid)
0.28 in x 2.07 in 7.1 mm x 52.6 mm
Size (Diameter x Height)(without grid)
0.25 in x 2.02 in 6.35 x 51.3 mm
Weight
0.27 oz
7.8 gm
[2]
All specifications are at room temperature unless otherwise specified. In the interest of constant product improvement, we reserve the right to change specifications without notice.
ICP® is a registered trademark of PCB Group, Inc.
OPTIONAL VERSIONS
Optional versions have identical specifications and accessories as listed for the standard model
except where noted below. More than one option may be used.
TLD
- TEDS Capable of Digital Memory and Communication Compliant with IEEE 1451.4 Optional Accessory : Model ACS-63 Calibration (with TEDS) of Precision Condenser Microphones and Preamplifiers together (mated pair). (1) replaces Model ACS-42
NOTES:
[1]
Prepolarized
[2]
Typical.
[3] re 250 Hz
[4]
Linear inherent noise is specified over the operating bandwidth (20 Hz
- 70 kHz).
[5]A-
Weighted inherent noise is specified from 20 Hz
- 20 kHz.
[6]
Peak.
[7]
TEDS Capable Digital Memory and Communication, compliant with IEEE P1451.4
[8]
Venting through Preamp.
[9]
See PCB Declaration of Conformance PS076 for details.
SUPPLIED ACCESSORIES:
Model ACS
-
42 Calibration of microphone with preamplifer (1)
Entered: AP Engineer: MJN Sales: MV
Approved: MT
Spec Number:
Date: 5/7/2014 Date: 5/7/2014 Date: 5/7/2014 Date: 5/7/2014
57855
3425 Walden Avenue, Depew, NY 14043
Phone: 716-684-0001 Fax: 716-684-0987 E-Mail: info@pcb.com
1
1
2
2
3
3
4
4
A A
B B
CODE
IDENT. NO.
52681
DWG. NO.
SCALE: SHEET
DRAWN CHECKED ENGINEER
TITLE
UNLESS OTHERWISE SPECIFIED TOLERANCES ARE:
DIMENSIONS IN MILLIMETERS
[ IN BRACKETS ]
ANGLES 2 DEGREES
3425 WALDEN AVE. DEPEW, NY 14043
(716) 684-0001 E-MAIL: sales@pcb.com
DIMENSIONS IN INCHES
ANGLES 2 DEGREES
FILLETS AND RADII
.003 - .005
FILLETS AND RADII
0.07 - 0.13
OUTLINE DRAWING
57747
1 OF 13X
1/4" ICP MICROPHONE SYSTEM
DECIMALS XX ±.03
XXX ±.010
DECIMALS X ± 0.8
XX ± 0.25
JDM
7/7/14
ECB
7/7/14
NJL
7/7/14
j
m
o
r
a
w
s
k
i
0
9
:
0
1
7
/
8
/
2
0
1
4
5
7
7
4
7
PCB Piezotronics Inc. claims proprietary rights in the information disclosed hereon. Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Piezotronics Inc.
2.07 [52.6]
1.86 [47.2]
.11 [2.9]
.41 [10.4]
10-32 COAXIAL CONNECTOR
.03 [0.8]2X
PREAMPLIFIER
.25 [6.4]
MICROPHONE CARTRIDGE
.28 [7.0]
.50 [12.7]
REVISIONS
REV DESCRIPTION DIN
A ADDED ADDITIONAL VENT HOLE 43127
Loading...