PIE Practical Instrument Electronics 830, 830PM Operating Manual

PIE 830
Process Calibrator
Loop DiagnosticsTransmitter Supply
mA V • pH •TC • Ω • RTD • Freq • Pressure
Operating Instructions
PIE 830
Process Calibrator
Loop DiagnosticsTransmitter Supply
mA V • pH •TC • Ω • RTD • Freq • Pressure
Operating Instructions
Practical Instrument Electronics
82 East Main Street Suite 3.14 • Webster, NY 14580 USA
Contents
General Operations
Accessories ............................................................................. 3
Carrying Case, Boot & Changing Batteries ............................4
Storing EZ-CHECK Outputs & Connections ........................... 5
Basic Operation
Switches & Knobs ...............................................................6, 7
MAIN Menus - Functions, Units & Ranges .........................8, 9
FEATURE Menu - Stepping & Ramping / Auto Off ............... 10
FEATURE Menu - Stepping & Ramping / Backlight ............. 11
Functions and Hookup Diagrams
2-Wire Transmitter Calibration
Simulate sensor while powering 2-Wire Transmitters ...... 12, 13
Universal Isolated Transmitter
Use the 830 to replace a 2-Wire Transmitter ................... 14-17
Milliamp
Simulate 2 Wire Transmitters ........................................... 18, 19
Source mA, Read mA .....................................................20, 21
Power/Measure Transmitters ................................................. 22
Using Ground Leak Detection .........................................23, 24
Voltage & Millivolt
Source V & mV; Read V & mV ........................................25, 26
pH
Source pH ........................................................................27, 28
Thermocouple
Source T/C & Read T/C Sensors .................................... 29, 30
Resistance
Source Resistance, Read Resistance & Continuity ........ 31, 32
RTD
Source RTD & Read RTD Sensors................................. 33, 34
Frequency
Source KHz, Hz & CPM; Read KHz, Hz & CPM ............35, 36
Pressure
Read Pressure ...................................................................... 37
Pressure Accessories ............................................................ 38
Pressure Module Ranges & Specifications ........................... 39
Specifications
General ............................................................................40-44
Thermocouple Ranges & Accuracies...............................45-47
RTD Ranges & Accuracies ................................................... 48
Warranty
Warranty & Additional Information ........................................ 49
General Information
Get more tools in a smaller calibrator
Carry eight single function calibrators
plus
a milliamp calibrator with loop supply
plus
a loop troubleshooter in the palm of your hand!
Milliamp • Voltage • Frequency • pH • Ohms
Thermocouples • RTDs • Pressure
Loop Diagnostics • Transmitter Supply
Detect ‘hidden’ loop problems
Quickly diagnose troublesome ground faults & current leak­age with patented Loop Diagnostic technology. These prob-
1
lems are undetectable with other instruments!
Check all loop parameters at once with LoopScope™
Simultaneously displays current, voltage and resistance to let you know the condition of a live loop. Finds prob-
2
lems with power supplies & loops with too many loads.
Patented by PIE!
Automatically detect 2, 3, or 4 wire RTDs
Trouble shoot sensor connections and find broken wires with patented technology. LCD indicates which of the four wires
3
are connected to an RTD sensor.
Swap out faulty transmitters to diagnose control issues
Easily setup the 830 as a universal isolated transmitter for Thermocouple, RTD, frequency, millivolt, resistance or pres-
4
sure. If the loop is back under control you know the transmitter was faulty. Turn on the LoopScope to simultaneously see the loop current, loop voltage and loop loads for troubleshooting.
Troubleshoot wiring problems without a multimeter
Built in continuity checker with ‘beeper’ quickly finds broken
wires or shorts in instrumentation wiring. Also handy for
5
checking operation of relays and controller outputs.
Page 1
The PIE Model 830 is more than a multifunction calibrator. It is also a loop detective that is able to diagnose common problems that other test equipment just can’t find. Have a flooded junction box or unknown ground faults? Our Loop Diagnostic technology will detect it. Or use the LoopScope to see at a glance all the parameters - milliamps, voltage and resistance - in the loop.
Stop throwing away perfectly good transmitters only to find the problem is somewhere else in the loop. Setup the PIE Model 830 as an isolated universal transmitter and turn on the Loop Diagnostics. The display will tell you EXACTLY the sensor input, the current output and if there is any uncontrolled current in the loop due to a ground fault, corrosion bridge or moisture. If the control system sees no problem when the 830 is acting as a transmitter THEN you can replace the faulty transmitter. With RTDs the 830 automatically detects which of the 2, 3 or 4 wires are connected and unbroken quickly alerting you to a sensor problem. Diagnostic features are covered by
US Patent
#7,248,058.
Become a troubleshooting technician with Patented Diagnostic Technology - Available only
with PIE Calibrators!
Page 2
Accessories
INCLUDED:
Four “AA” Alkaline batteries, Certificate of Calibration
Evolution Hands Free Carrying Case Part No. 020-0211
Blue Rubber Boot Part No. 020-0213
Test Leads - one pair with Part No. 020-0207 banana plug & alligator clips
Evolution RTD Wire Kit Part No. 020-0208 2 Red & 2 Black Leads with Banana Plugs & Spade Lugs
OPTIONAL:
Ni-MH 1 Hour Charger with 4 Ni-MH AA Part No. 020-0103 Batteries (100-120 V AC input for North America Only)
T/C Wire Kit 1* for Types J, K, T & E Part No. 020-0202
T/C Wire Kit 2* for Types B, R/S & N Part No. 020-0203
*Three feet (1 meter) of T/C extension wire, stripped on one end with a miniature T/C male connector on the other end.
Page 3
Operating Instructions
FIELD & BENCH USE
PIE 830 comes with a carrying case designed for hands-free operation and a rubber boot with a built-in tilt stand. The PIE 830 is held in the case by elastic straps for use with the carrying case open. The tilt stand is easily raised by pulling the stand until it locks into place.
CHANGING BATTERIES
Low battery is indicated by a battery symbol on the display. Approximately one to four hours of typical operation remain before the PIE 830 will automatically turn off. To change the batteries remove the rubber boot and remove the battery door from the back of the unit by sliding the door downward. This allows access to the battery compartment. Replace with four (4) “AA” 1.5V batteries being careful to check the polarity. Replace the battery door and replace the boot. All stored configuration options (T/C Type, EZ-CHECK Memories, etc.) are reset to factory settings when the batteries are removed. Note: Alkaline batteries are supplied and recommended for typical battery life and performance. Optional rechargeable batteries (charged externally) are available.
Page 4
Operating Instructions
STORING HI and LO EZ-CHECK Source Outputs
Speed up your calibration by storing Span & Zero output setting for instant recall with the EZ-CHECK switch.
1) Store your high (SPAN) output temperature by moving the
EZ-CHECK switch to the HI position and turning the EZ-Dial knob until the desired output value is on the display. Press and hold the EZ-Dial knob until STORED appears to store the value. Release the EZ-Dial knob.
2) Store your low (ZERO) output value by moving the EZ-CHECK
switch to the LO position and turning the EZ-Dial knob until the desired output value is on the display. Press and hold the EZ-Dial knob until STORED appears to store the value. Release the EZ-Dial knob.
3) Instantly output your SPAN and ZERO output settings by
moving the EZ-CHECK switch between HI and LO. You may also select any third output setting (such as mid-range) using the SET position on the EZ-CHECK switch.
Connections
PIE 830 has banana jacks compatible with unshielded or retractable banana plugs. Included with your calibrator are a pair of test leads with alligator clips for mA, V, pH & Hz connections. Four test leads with spade lugs are also included for 2, 3 and 4 Wire RTD connections. Thermocouple connections are made through a miniature thermocouple socket.
23
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
Source V∙ pH
Page 5
4
1
RTD
+
-
T/C
Operating Instructions
Basic Operation
PWRM
+
-
5 6
18.300 mA
830.0 °C
TYPE K
SOURCE
MAX
HI
READ
SET
MIN
LO
READ
q EZ-CHECK™ SWITCH
SOURCE: Instantly output two preset settings by moving the
EZ-CHECK™ switch to the “LO” position or “HI” position. three point checks select the “SET” position. The PIE 830 will remember the last “SET” value, even with the power off. These values can easily be changed to suit the calibration requirements. The values stored in the HI and LO positions are also used for Auto Stepping.
For fast
READ: Slide the switch to the SET position. The PIE 830 will display the current reading from the sensor or device being measured. Slide the switch to MAX and the highest value measured since turn-on or reset will be displayed; slide the switch to MIN and the lowest value measured since turn-on or reset will be displayed.
Page 6
Operating Instructions
Basic Operation
w SOURCE/OFF/READ Switch
Select “SOURCE” to output mA, V, pH, T/C, Ω, RTD or frequency. Select “READ” to read mA, V, T/C, Ω, RTD, pressure or frequency. Select “OFF” to turn off the 830.
e EZ-DIAL™ KNOB
SOURCE: Turn the knob to adjust the output level. Turn clockwise to increase the output, counter clockwise to decrease the output in one least significant digit step at a time. Push down and turn the EZ-DIAL knob for faster dialing. Press and hold the knob for two seconds to store desired EZ-Check™ HI/LO points in SOURCE mode. Continue to press and hold the knob for two more seconds to start the automatic ramping.
READ: Press and hold to transfer the current temperature into the EZ-Check™ MIN/MAX points. This clears the MIN/MAX readings which will update as the input value changes.
SELECTING FUNCTIONS
The EZ-DIAL knob is used to setup the PIE 830 to match the instrument to be calibrated or signal to be measured. Each time you turn on the PIE 830 the LCD displays the following screen for about 1 second followed by operating in the function used the last time it was operated.
PIE 830
DOUBLE CLICK
EZ-DIAL KNOB
FOR CONFIGURATION
Double Click the EZ-DIAL knob to change the function of the calibrator and to select ranges, units and other user settings. Each function (mA, V, T/C, Ohms, RTD, Frequency) has up to four pages of menus. The first menu page has settings for the function and the last menu page has settings for STEPPING, AUTO OFF and BACKLIGHT. Settings are remembered even with the power off but are reset when the batteries are changed.
Page 7
Operating Instructions
Double Click Menus - MAIN Page
Double click the EZ-DIAL knob to access the Double Click Menus. Shown are the MAIN menus for each function. Turn the knob to scroll thru the menus and press the knob to select. Available choices are shown in grey.
Source mA & Simulate 2 Wire Transmitters
MAIN
> EXIT FUNCTION mA MODE SOURCE 2W SIM UNITS mA % HART 250Ω ON OFF
Read mA, Power/Measure Transmitters
& Leak Detect
> EXIT FUNCTION mA MODE READ PWR MEAS UNITS mA % HART 250Ω ON OFF
Source V & mV
>EXIT FUNCTION V RANGE 10V 100mV 1V
Read V & mV
>EXIT FUNCTION V RANGE 10V 1V 60V 100mV
Source pH
> EXIT FUNCTION pH
Source & Read Frequency
> EXIT FUNCTION FREQ RANGE 20KHZ 10000HZ 1000HZ 2000CPM
Page 8
Operating Instructions
Double Click Menus - MAIN Page
Source & Read Thermocouples
> EXIT FUNCTION T/C UNITS °C °F T/C TYPE J K E T R S B N L U G C D P COLD JUNC ON OFF
Source RTD
> EXIT FUNCTION RTD UNITS °C °F RTD Pt 100 a=3850 [*RTD Types - See Read RTD]
Read RTD
> EXIT FUNCTION RTD UNITS °C °F
RTD Pt 100 a=3850,
Pt 1000 a=3850; Cu 10 a=4274, Cu 50 a=4280 Ni 120 a=6720, Ni 110 a=5801
a=3902, a=3916, a=3926
Source Ohms
>EXIT FUNCTION OHMS RANGE 40 0Ω 4000Ω
>EXIT FUNCTION OHMS RANGE 400Ω 4000Ω CONT
Read Ohms
Read Pressure
>EXIT FUNCTION PRESSURE UNITS psi inH2O* ftH2O* mmH2O* cmH2O* mH2O* inHg mHg
cmHg mmHg torr kg/cm2 kg/m2 hPa kPa MPa bar mbar atm oz/in2
2
*Engineering unit available at 4°C, 20°C and 60°F.
lb/ft
Page 9
Operating Instructions
Double Click Menu - FEATURES
To change the Automatic Stepping settings
Double click the e DIAL KNOB at any time the unit is on and the following typical display (will be different for each FUNCTION) will appear for 15 seconds:
MAIN > EXIT (1/2) FUNCTION mA MODE SOURCE UNITS mA HART 250Ω ON
Turn the menu page so the word FEATURES appears at the top of the menu.
Turn the e DIAL KNOB to move through the menu. Press the e DIAL KNOB to toggle between OFF and ON or to change the STEPS/RAMP and the STEP/RAMP TIME settings. These settings are remembered even with the power off.
EXIT MENU - exits this menu immediately and saves any changes. Menu will automatically exit after 15 seconds of inactivity.
AUTO OFF - If AUTO OFF is ON, the unit will turn off after 30 minutes of inactivity to save battery life. If AUTO OFF is OFF the unit will stay on until the POWER SWITCH is moved to the off position.
e DIAL KNOB to move to the second, third or fourth
FEATURES
> EXIT (2/2) AUTO OFF ON BACKLIGHT ON STEPS/RAMP 3 STEP/RAMP TIME 5
Page 10
Operating Instructions
Double Click Menu - FEATURES
STEPS/RAMP - pressing the knob will cycle through 2, 3, 5, 11
and RAMP. The endpoints of the steps or ramp are based on the values stored in the HI and LO EZ-CHECK outputs.
2 steps will automatically switch bet ween the values stored in the HI & LO EZ-CHECK (0 & 100%).
3 steps between the HI, Midpoint and LO EZ-CHECK (0, 50 & 100%).
5 steps between the HI and LO EZ-CHECK in 25% increments (0, 25, 50, 75 & 100%).
11 steps between the HI and LO EZ-CHECK in 10% increments (0, 10, 20...80, 90 &100%).
RAMP continuously ramps up and down between the HI and LO EZ-CHECK outputs.
STEP/RAMP TIME - pressing the knob will cycle through 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 and 60 seconds.
To start the Automatic Stepping
Start automatic stepping or ramping by placing the EZ-CHECK Switch into the HI or LO position then press and hold the e DIAL KNOB for 6 seconds (the word STORE will appear on the display after 3 seconds and continue to press the EZ-DIAL KNOB) until the word STEPPING appears on the display. The word STEPPING will appear on the display anytime the selected automatic function is running. Stop the stepping by again pressing and holding the
e DIAL KNOB for 3 seconds.
BACKLIGHT - If BACKLIGHT is ON the backlight will light all the time the unit is powered up. For maximum battery life turn the backlight off when using the calibrator in areas with enough ambient light to read the display.
Page 11
Operating Instructions
Calibrate a 2-Wire Transmitter by sourcing the input while monitoring the output.
Works with SOURCE pH, T/C, DC V, OHMS, RTD, FREQ and READ PRESSURE.
Move the power switch DIAL KNOB and the MAIN menu for the function in use will appear for 15 seconds:
MAIN > EXIT FUNCTION T/C UNITS °C T/C TYPE K COLD JUNC ON
Turn the page so the word mA DISPLAY appears at the top of the menu.
Turn the e DIAL KNOB to move through the menu. Press the e DIAL KNOB to toggle between OFF and ON or to change the MODE setting.
HART 250Ω - turn on the 250Ω resistor if you are powering up a HART transmitter. This provides the loop load required for HART communication.
EXIT MENU - exits this menu immediately and saves any changes. Menu will automatically exit after 15 seconds of inactivity.
e DIAL KNOB to move to the second (or third) menu
mA DISPLAY
> EXIT MODE OFF READ PWRM READ% PWRM% HART 250Ω OFF ON LEAK DETECT OFF ON
w to READ and Double click the e
Page 12
Operating Instructions
MODE - pressing the knob will cycle through READ, PWRM,
READ% ,PWRM% and OFF.
READ turns on the mA display and indicates current passing through the loop proportional to the input of the transmitter which is controlled by the output of the 830. Choose READ% if your would like the mA display in percent of 4-20 milliamps.
PWRM is POWER MEASURE which uses the internal loop supply of the 830 to power up the transmitter while indicating the current passing through the loop proportional to the input of transmitter which is controlled by the output of the 830. Choose PWRM% if your would like the mA display in percent of 4-20 milliamps.
If PWRM or PWRM% is selected an additional menu selection of LEAK DETECT will appear. When LEAK DETECT is turned on the 830 will display LEAK on the display if there is more than 0.015 mA of uncontrolled current in the loop. This may be due to a faulty transmitter, corrosion causing a bridge to ground or moisture present at some connection point. When LEAK DETECT tests a loop with leakage the loop mA signal will be affected.
+ IN - REF +OUT-
Typical
2-Wire
Transmitter
Thermocouple Wire
PWRM
+
-
5 6
18.300 mA
830.0 °C
MAX
HI
READ
SET
MIN
LO
TYPE K
PWRM LEAK
18.300 mA
830.0 °c
TYPE K
(Enlarged Display)
SOURCE
READ
Page 13
Universal Isolated Transmitter Swap out a transmitter to diagnose control issues.
The 830 acts as an isolated universal T/C, mV, OHMS, RTD, FREQ and PRESSURE transmitter. Choose this function to temporarily replace a transmitter when you suspect the transmitter is faulty or to diagnose the parameters of the loop.
Move the power switch DIAL KNOB and the MAIN menu for the function in use will appear. Turn the and press the e KNOB until the desired FUNCTION appears. Turn the e DIAL KNOB to move through the menu. Press the e DIAL KNOB to toggle the units, range, T/C Type or RTD curve.
w to READ and Double click the e
e DIAL KNOB until the pointer is at FUNCTION
> EXIT FUNCTION T/C UNITS °C T/C TYPE K COLD JUNC ON
MAIN >EXIT FUNCTION V RANGE 100mV
MAIN > EXIT FUNCTION FREQ RANGE 20KHZ
Turn the e DIAL KNOB to move to the second (or third) menu page so the word XXX XMTR appears at the top of the menu.
Turn the e DIAL KNOB to move through the menu. Press the e DIAL KNOB to toggle between DOWN and UP, OFF and ON, or to change the MODE setting.
MAIN
MAIN > EXIT FUNCTION RTD UNITS °C
RTD Pt 100 a=3850
MAIN >EXIT FUNCTION OHMS RANGE 400Ω
MAIN >EXIT FUNCTION PRESSURE UNITS psi
Page 14
XXX XMTR > EXIT MODE OFF LINEAR NONLINEAR SQ ROOT BURNOUT DOWN UP LOOPSCOPE OFF ON
MODE - pressing the knob will cycle through OFF, LINEAR,
NONLINEAR. When setup to read pressure SQ ROOT replaces NONLINEAR.
LINEAR turns on the mA display and regulates the loop current linear with the input signal. For thermocouples & RTDs this is linear relative to the temperature between zero and span of the sensor input. This mimics the operation of a digital temperature transmitter or other linearizing transmitters.
NONLINEAR (applies only to thermocouples and RTDs) turns on the mA display and regulates the loop current linear with base units (millivolts or ohms) of zero and span of the sensor input. This mimics the operation of an analog temperature transmitter without linearizing.
SQ ROOT turns on the mA display and regulates the loop current with square root extraction of zero and span of the pressure module input. This mimics the operation of a differential pressure transmitter setup for square root extraction.
BURNOUT selects the failure mode of the calibrator when the input signal is lost or beyond the input scale. Down sets the output to 1.0 mA and UP to 21.5 mA when sensor failure is detected.
LOOPSCOPE turns on the Loop Diagnostic display which simultaneously indicates the current, voltage and resistance of the loop. When enabled the LOOPSCOPE also indicates errors by flashing the LED and writing diagnostic messages on the display. Messages include LOW LOOP VOLT, HIGH LOOP VOLT and HIGH LOOP IMPED. With LOOPSCOPE on the 830 occasionally performs a test on the loop. During this short test the loop mA signal will be affected.
Page 15
Setting up the 830 as a transmitter
Configure your 830 to the same Span (URV) and Zero (LRV) by storing Span & Zero setting with the q EZ-CHECK switch.
1) Store your SPAN input by moving the
to the MAX position and turning the e EZ-Dial knob until the desired output value is on the display. Press and hold the EZ-Dial knob until STORED appears to store the value. Release the EZ-Dial knob.
2) Store your ZERO input by moving the EZ-CHECK switch to the MIN position and turning the EZ-Dial knob until the desired output value is on the display. Press and hold the EZ-Dial knob until STORED appears to store the value. Release the EZ-Dial knob.
3) Move the EZ-CHECK switch to the READ position. The 830 constantly measures the input from the process sensor and regulates the current in the 4-20 mA loop.
Once the PIE 830 is setup and connected to the process sensor and the 4-20 mA loop it acts just like a calibrated isolated transmitter and will display the input signal from the sensor and the regulated 4-20 mA output signal. XMTR is shown on the display along with L, N or S for Linear, Nonlinear or Square Root respectively.
XMTRL
12.000 mA
q EZ-CHECK switch
12.000 mA
SUPPLY 23.4 V LOAD 987 Ω
500.0 °c
TYPE K
830 as Transmitter 830 with LOOPSCOPE
For full loop diagnostics turn on LOOPSCOPE which indicates the current, SUPPLY (voltage) and LOAD (resistance) of the loop. It also indicates errors by flashing the LED and writing diagnostic messages including LOW LOOP VOLT, HIGH LOOP
VO LT and HIGH LOOP IMPED.
500.0 °c
TYPE K
Page 16
Connecting the 830 in place of a transmitter
Connect the PIE 830 in place of the transmitter. The 4-20 mA loop connects to the mA jacks of the 830 and the sensor (or pressure module) connects to the other jacks of the 830.
Receiver
Powers External
2-Wire Transmitter
Power Supply
(2 to 60 VDC)
All milliamp connections
12.000mA
+
-
5 6
27.4 V 830 Ω
500.0
°C
23
RTD
All mA∙ Read V
-
+
65
To
Sensor
-
Source & Read Hz∙ mV
Source V∙ pH
+
4-20 mA Loop
+ IN - REF +OUT-
Typical
2-Wire
Transmitter
(Disconnected)
4
1
RTD
-
T/C
To
Note: With LOOPSCOPE on the 830 occasionally performs a test on the loop. During this short test the loop mA signal will be affected.
Page 17
2 Wire SIM mA, 2 Wire SIM % (Percent of 4 to 20 mA)
Choose this function to simulate a 2 Wire Transmitter output from
0.000 to 24.000 milliamps. Operates in loops with power supply
voltages from 2 to 60 VDC.
Move the power switch EZ-DIAL knob to get into the Menu. Turn the knob to scroll through the settings and press the knob to make your selection. Select mA for the FUNCTION and 2W SIM for the MODE. Choose either mA or % and whether you need the 250 HART resistor active in the loop.
If you would also like to see the condition of the loop turn the DIAL KNOB to move to the second (or third) menu page so the word mA DIAGNOSTIC appears at the top of the menu.
> EXIT LOOPSCOPE OFF ON
LOOPSCOPE turns on the Loop Diagnostic display which
simultaneously indicates the current, voltage and resistance of the loop. When enabled the LOOPSCOPE also indicates errors by flashing the LED and writing diagnostic messages on the display. Messages include LOW LOOP VOLT, HIGH LOOP VOLT and HIGH LOOP IMPED. With LOOPSCOPE on the 830 occasionally performs a test on the loop. During this short test the loop mA signal will be affected.
w to SOURCE then Double Click the
e
mA DIAGNOSTIC
2-WIRE
12.000 mA
Transmitter
2-WIRE
SUPPLY 25.4 V LOAD 485 Ω
12.000 mA
830 with LOOPSCOPE830 Simulating a
Page 18
2 Wire SIM mA, 2 Wire SIM % (Percent of 4 to 20 mA)
Connect the output leads of the PIE 830 to the inputs of the device being calibrated, making sure to check polarity. Red lead to the plus (+) input and black lead to the minus (-) input.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK switch between HI and LO (defaults to 20 & 4 mA). You may also select any third output setting (such as mid-range) using the SET position on the EZ-CHECK switch. The output is adjusted in 0.001 mA (0.01%) increments by turning the knob
e. Press and turn the knob for faster dialing
with 0.100 mA (1.00%) increments.
Receiver
Powers External
2-Wire Transmitter
Power Supply
(2 to 60 VDC)
All milliamp connections
2-WIRE
+
-
5 6
4.000mA
LO
All mA∙ Read V
RTD
+
23
-
65
To
Sensor
-
Source & Read Hz∙ mV
Source V∙ pH
+ IN - REF +OUT-
Typical 2-Wire
Transmitter
(Disconnected)
4
1
RTD
+
-
T/C
Note: With LOOPSCOPE on the 830 occasionally performs a test on the loop. During this short test the loop mA signal will be affected.
Page 19
mA SOURCE/ % SOURCE (Percent of 4 to 20 mA)
Choose this function to provide an output from 0.000 to 24.000 milliamps. The compliance voltage is a nominal 24 VDC to provide the driving power to your milliamp receivers.
Move the power switch
w to SOURCE then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll through the settings and press the knob to make your selection. Select mA for the FUNCTION and SOURCE for the MODE. Choose either mA or % and whether you need the 250 HART resistor active in the loop.
Connect the output leads of the PIE 830 to the inputs of the device being calibrated, making sure to check polarity. Red lead to the plus (+) input and black lead to the minus (-) input.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK switch between HI and LO (defaults to 20 & 4 mA). You may also select any third output setting (such as mid-range) using the SET position on the EZ-CHECK switch. The output is adjusted in 0.001 mA (0.01%) increments by turning the knob
e. Press and turn the knob for faster dialing
with 0.100 mA (1.00%) increments.
Milliamp Receiver Input
SOURCE
+
-
5 6
20.000mA
HI
Controller
Transmitter
Computer
Logger
I/P
DCS
Page 20
23
RTD
All mA∙ Read V
-
+
65
4
1
RTD
+
-
All milliamp connections
-
Source & Read Hz∙ mV
T/C
Source V∙ pH
READ mA, READ % (Percent of 4 to 20 mA)
Choose this function to measure from 0.000 to 24.000 milliamps or -25.00 to 125.00%.
Move the power switch
w to READ then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob e to scroll through the settings and press the knob to make your selection. Select mA for the FUNCTION and READ for the MODE. Choose either mA or % and whether you need the 250 HART resistor active in the loop.
Connect the red input lead (+) of the PIE 830 to the more positive point of the break and the black input to the more negative point.
Signals below 0 mA or open circuits are indicated by 0.000 mA (-25.00%) on the display. Signals above 24 mA are current limited by protection circuitry with “OVERRANGE” flashed on the display and the red OVERLOAD LED lit.
The PIE 830 measures the input signal and constantly updates the display with the current reading. Move the EZ-CHECK switch
q to MAX to see the highest reading and to MIN to see the
lowest reading. Press and hold the knob e to clear the MAX and MIN readings.
READ
+
-
5 6
19.999mA
MAX
All milliamp connections
Milliamp Output Signal
Controller
Transmitter
P/I
DCS
23
RTD
All mA∙ Read V
-
+
65
Page 21
-
Source & Read Hz∙ mV
Source V∙ pH
4
1
RTD
+
-
T/C
Power/Measure mA, Power/Measure % (Percent of 4 to 20 mA)
Choose this function to simultaneously supply power to a 2 Wire Transmitter while displaying the 4.000 to 20.000 mA output of the transmitter.
Move the power switch
w to READ then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob e to scroll through the settings and press the knob to make your selection. Select mA for the FUNCTION and PWR MEAS for the MODE. Choose either mA or % and whether you need the 250 HART resistor active in the loop.
Disconnect one or both input wires from the device to be calibrated. Connect the red source lead of the PIE 830 to the plus (+) input of the device and the black source lead to the minus (-).
The PIE 830 supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter. The current passed by the transmitter will be accurately displayed by the PIE 830. Calibrate the transmitter in the usual manner and disconnect the PIE 830. Signals above 24 mA are current limited by protection circuitry with “OVERRANGE” flashed on the display and the red OVERLOAD LED lit.
PWRM
+
-
5 6
11.999mA
Transmitter Input
Sensor
Process Signal
Simulated Input
+ IN - REF +OUT-
Typical
2-Wire
Transmitter
SOURCE
MAX
HI
READ
SET
MIN
LO
READ
Page 22
Using Ground Leak Detection
mA OUT, % OUT (Percent of 4 to 20 mA)
Find current leaks in loops caused by ground faults, moisture or corrosion. The 830 simultaneously supplies power to a 2 Wire Transmitter (or loop with a transmitter) while displaying the 4 to 20 mA output and the amount of current leaking in the loop.
1) Move the power switch
w to READ then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob e to scroll through the settings and press the knob to make your selec­tion. Select mA for the FUNCTION and PWR MEAS for the MODE. Choose either mA or %.
2) Turn the knob
e until the following menu appears.
mA DIAGNOSTIC > EXIT LEAK DETECT OFF ON
3) Turn the knob
e to scroll through the settings and press the
knob to make your selection. Turn on the LEAK DETECT.
4) Connect the red source lead from the mA (+) jack of the 830 to the plus (+) input of the device and the black source lead from the mA (-) to the minus (-).
The PIE 830 supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter or loop. The current passed by the transmitter will be accurately displayed by the 830 along with an indication of leakage current at the top of the display. If there is an uncontrolled loop, a transmitter with upscale burnout and bad or missing sensor or a short the display shows "OVER RANGE”
All milliamp connections
23
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
+
Source V∙ pH
4
1
RTD
-
T/C
Page 23
Using Ground Leak Detection
12.280
PROCESS
INDICATOR
Transmitter Input
Sensor Process Signal Simulated Input
+ IN - REF +OUT-
Typical
2-Wire
Transmitter
Typical Error Conditions
PWRM LEAK
+
-
5 6
00.28 mA
12.280mA
PWRM LEAK
00.28 mA
PWRM LEAK
mA
OVER RANGE
12.280 mA
The PIE 830 is supplying the loop voltage. A calibrated transmitter is limiting the loop current to
12.00 mA. An additional
0.28 mA is not controlled by the transmitter and is
mA
The PIE 830 is supplying the loop voltage. There is a control loop error. This may be a transmitter (set for upscale burnout) with a bad or missing sensor, or a
short in the loop. leaking somewhere in the loop.
Note: Many installed transmitters will normally indicate 0.01 to
0.02 mA leakage without significant control problem. Unstable readings may indicate loose connections or the presence of moisture.
Page 24
mV/V SOURCE
Choose this function to provide an output from -20.000 to 99.999 mV, -500.00 TO 999.99 mV or from 0.000 to 10.250 V. The source current is a nominal 20 mA to provide the driving power to your voltage receivers.
Move the power switch
w to SOURCE then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll through the settings and press the knob to make your selection. Select V for the FUNCTION and 1 V, 10V or 100 mV for the RANGE.
Connect the output leads of the PIE 830 to the inputs of the device being calibrated, making sure to check polarity. Red lead to the plus (+) input and black lead to the minus (-) input.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK switch between HI and LO. You may also select any third output setting (such as mid-range) using the SET position on the EZ-CHECK switch. The output is adjusted in
0.001 mV, 0.01 mV or 0.001 V increments by turning the knob
e.
Press and turn the knob for faster dialing with 0.100 mV, 1.00 mV or 0.100 V increments.
Voltage Receiver Input
-
2 1
5.000V
HI
+
Controller
Transmitter
Computer
Logger
DCS
V & mV Source
Page 25
23
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
Source V∙ pH
4
1
RTD
+
-
T/C
Read mV/V
Choose this function to measure from -99.999 to 99.999 millivolts, -999.99 to +999.99 mV, 0.000 to 10.250 V dc or 0.00 to 60.00 V dc.
Move the power switch
w to READ then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll through the settings and press the knob to make your selection. Select V for the FUNCTION and 1V, 10V, 60V or 100 mV for the RANGE.
Connect the red input lead (+) of the PIE 830 to the more positive point of the break and the black input to the more negative point.
Signals above the maximum scale are limited by protection circuitry with “OVER RANGE” flashed on the display and the red OVERLOAD LED lit.
The PIE 830 measures the input signal and constantly updates the display with the current reading. Move the EZ-CHECK switch
q to MAX to see the highest reading and to MIN to see the
lowest reading. Press and hold the knob e to clear the MAX and MIN readings.
Voltage Output Signal
+
-
2 1
8.300mV
MAX
Controller
Transmitter
Power Supply
All mA∙ Read V
Page 26
23
RTD
-
-
+
Source & Read Hz∙ mV
65
Connections for Read 10 V & 60V ranges
Connections for Read mV ranges
4
1
RTD
+
-
T/C
Source V∙ pH
pH SOURCE
Choose this function to provide an output from 0.000 to 14.000 pH @ 25°C (77°F) which corresponds to 414.12 to -414.12 mV. The source current is a nominal 20 mA to provide the driving power to your pH receivers.
Move the power switch
w to SOURCE then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll through the settings and press the knob to make your selection. Select pH for the FUNCTION.
Connect the output leads of the PIE 830 to the inputs of the device being calibrated, making sure to check polarity. Red lead from the mV (+) jack of the 830 to the plus (+) input and black lead from the mV (-) jack to the minus (-) input.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK switch between HI and LO. You may also select any third output setting (such as mid-range) using the DIAL position on the EZ-CHECK switch. The output is adjusted in
0.001 pH increments by turning the knob
e. Press and turn the
knob for faster dialing with 0.100 pH increments.
pH Receiver Input
+
-
2 1
4.010pH
ACID
Controller
Transmitter
Analyzer
pH Source
Page 27
23
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
Source V∙ pH
4
1
RTD
+
-
T/C
Simulate pH probes into transmitters & analyzers
Use the pH simulator to verify proper operation of pH devices before you place a probe into a calibrated buffer. Adjusting the pH transmitter or analyzer without a probe allows you to make sure the device is calibrated and operating correctly. The 830 simulates 0.000 to 14.000 pH @ 25°C corresponding to 414.12 to -414.12 mV.
Once the pH instrument has been adjusted against the PIE 830 reconnect the pH probe and check it against the proper buffer (typically 7 pH). If the instrument zero point requires more than the manufacturer’s recommendations (typically within 0.5 pH) it is time to clean or replace the probe.
Page 28
Thermocouple Source
Choose this function to provide a simulated thermocouple signal into controllers, temperature transmitters, indicators or any input devices that measure thermocouple sensors.
Move the power switch
w to SOURCE then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll through the settings and press the knob to make your selection. Select T/C for the FUNCTION, °F or °C for the UNITS, T/C Type (J, K, E, T, R, S, B, N, L (J-DIN), U (T-DIN), G, C, D or P (Platinel II)) and internal COLD JUNC ON or OFF (ON is the default).
Connect the PIE 830 to the inputs of the device being calibrated using the proper type of thermocouple wire via the miniature thermocouple socket.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK switch between HI and LO. You may also select any third output setting (such as mid-range) using the SET position on the EZ-CHECK switch. The output is adjusted in 0.1° increments by turning the knob
e. Press and turn the knob for
faster dialing with 10.0° increments.
Thermocouple Wire
Instrument with T/C Input
Controller
Temperature Transmitter
Temperature Indicator
Temperature Trip or Alarm
1083.0°C
HI
TYPE K
+
-
T/C
23
RTD
All mA∙ Read V
All thermocouple connections
-
-
+
65
Page 29
1
RTD
+
-
Source & Read Hz∙ mV
T/C
Source V∙ pH
4
Read Thermocouple Sensors
Choose this function to measure temperatures with a thermocouple probe, sensor or any device that output a thermocouple signal.
Move the power switch
w to READ then Double Click the
EZ-DIAL knob to get into the Double Click Menu. Turn the knob to scroll through the settings and press the knob to make your selection. Select T/C for the FUNCTION, °F or °C for the UNITS, T/C Type (J, K, E, T, R, S, B, N, L (J-DIN), U (T-DIN), G, C, D or P (Platinel II)) and COLD JUNC ON or OFF (ON is the default).
Connect the PIE 830 to the inputs of the device being calibrated using the proper type of thermocouple wire via the miniature thermocouple socket. If no sensor is connected, a wire is broken or the sensor is burned out, OPEN TC will appear on the display. Signals above the maximum scale are limited by protection circuitry with “OVER RANGE” on the display and the OVERLOAD LED blinks red.
The PIE 830 measures the input signal and constantly updates the display with the current reading. Move the EZ-CHECK switch
q to MAX to see the highest reading and to MIN to see the
lowest reading. Press and hold the knob e to clear the MAX and MIN readings.
830.0 °C
MIN
Thermocouple Wire
+
-
T/C
TYPE K
Page 30
Resistance Source
Choose this function to provide a simulated resistance into any device that measures resistance.
Move the power switch
w to SOURCE then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll through the settings and press the knob to make your selection. Select OHMS for the FUNCTION, 400Ω or 4000Ω for the RANGE.
Disconnect all sensor wires from the devices to be calibrated and connect the PIE 830 to the inputs of the device using 2, 3 or 4 wires.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK switch between HI and LO. You may also select any third output setting (such as mid-range) using the SET position on the EZ-CHECK switch. The output is adjusted in
0.01Ω/0.1Ω increments by turning the knob
e. Press and turn
the knob for faster dialing with 1.00Ω/10.0Ω increments.
Instrument with
Resistance Input
+-+
-
2 1
3 4
250.00
LO
Controller
Transmitter
PLC
2, 3 & 4 Wire Resistance Connections
Page 31
23
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
Source V∙ pH
4
1
RTD
+
-
T/C
Read Resistance & Check Continuity
Choose this function to measure resistance or check continuity.
Move the power switch
w to READ then Double Click the
EZ-DIAL knob to get into the Double Click Menu. Turn the knob to scroll through the settings and press the knob to make your selection. Select OHMS for the FUNCTION, 400Ω, 4000Ω or Continuity for the RANGE.
Connect the PIE 830 to the resistor or sensor using 2, 3 or 4 wires. The PIE 830 automatically detects how many wires are connected using a patented circuit and indicates each wire that is connected. Any wires that are not connected or broken are indicated by the 830. This is useful for troubleshooting the sensor.
If continuity is selected, resistance is measured up to 400.0Ω. The beeper will sound and
)))) appears on the display when
resistance below 100.0Ω is measured.
Signals above the maximum scale are limited by protection circuitry with “OVER RANGE” flashed on the display and the OVERLOAD LED blinks red.
The PIE 830 measures the input signal and constantly updates the display with the current reading. Move the EZ-CHECK switch
q to MAX to see the highest reading and to MIN to see the
lowest reading. Press and hold the knob e to clear the MAX and MIN readings.
+
-
2 1
830.00
Page 32
RTD Source
Choose this function to provide a simulated RTD signal into controllers, temperature transmitters, indicators or any input devices that measure RTD sensors.
Move the power switch
w to SOURCE then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll through the settings and press the knob to make your selection. Select RTD for the FUNCTION, °F or °C for the UNITS and RTD (Choose from one of Platinum 100Ω, or 1000Ω, Copper 10Ω or 50Ω, Nickel 120Ω or 110Ω curves). Note: Pt 100Ω 3850 is the most common RTD type.
Disconnect all sensor wires from the devices to be calibrated and connect the PIE 830 to the inputs of the device using 2, 3 or 4 wires.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK switch between HI and LO. You may also select any third output setting (such as mid-range) using the SET position on the EZ-CHECK switch. The output is adjusted in 0.1° increments by turning the knob
e. Press and turn the knob for
faster dialing with 10.0° increments.
Instrument with RTD Input
Controller
+-+
-
2 1
3 RTD 4
830.0 °C
HI
Pt 100 α=3850
Temperature Transmitter
Temperature Indicator
Temperature Trip or Alarm
2, 3 & 4 Wire RTD Connections
Page 33
23
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
Source V∙ pH
4
1
RTD
+
-
T/C
Read RTD Sensors
Choose this function to measure temperatures with an RTD probe, sensor or any device that output an RTD signal.
Move the power switch
w to READ then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll through the settings and press the knob to make your selection. Select RTD for the FUNCTION, °F or °C for the UNITS and RTD (Choose from one of Platinum 100Ω, or 1000Ω, Copper 10Ω or 50Ω, Nickel 120Ω or 110Ω curves). Note: Pt 100Ω 3850 is the
most common RTD type.
Connect the PIE 830 to the RTD sensor using 2, 3 or 4 wires. The PIE 830 automatically detects how many wires are connected using a patented circuit and indicates each wire that is connected. Any wires that are not connected or broken are indicated by the
830. This information is useful for troubleshooting the sensor.
Signals above the maximum scale are limited by protection circuitry with “OVER RANGE” flashed on the display and the red OVERLOAD LED lit.
The PIE 830 measures the input signal and constantly updates the display with the current reading. Move the EZ-CHECK switch
q to MAX to see the highest reading and to MIN to see the
lowest reading. Press and hold the knob e to clear the MAX and MIN readings.
+
-
-
2 1
3 RTD
83.0 °C
Pt 100 α=3850
Page 34
Frequency Source
Choose this function to provide a frequency signal into any input devices that measure frequency.
Move the power switch
w to SOURCE then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll through the settings and press the knob to make your selection. Select FREQ for the FUNCTION and 20KHZ, 10000HZ, 1000HZ or 2000CPM for the RANGE.
Disconnect all input wires from the devices to be calibrated and connect the PIE 830 to the input of the device matching polarity.
The green HZ SYNC LED pulses in synch with the output pulses and may be used to calibrate optical pickups. The output signal is a zero crossing square wave with a fixed amplitude of 6 V peak-to-peak from -1 and + 5 V.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK switch between HI and LO. You may also select any third output setting (such as mid-range) using the SET position on the EZ-CHECK switch. The output is adjusted in 1 count increments by turning the knob
e. Press and turn the
knob for faster dialing with 100 count increments.
Frequency Receiver Input
Flowmeter
+
-
2 1
8300.0Hz
HI
Transmitter
Computer
Logger
Controller
DCS
All Frequency Connections
Page 35
23
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
Source V∙ pH
4
1
RTD
+
-
T/C
Read Frequency
Choose this function to count frequency.
Move the power switch
w to READ then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll through the settings and press the knob to make your selection. Select FREQ for the FUNCTION and 20KHZ, 10000HZ, 1000HZ or 2000CPM for the RANGE.
Disconnect all input wires from the devices to be calibrated and connect the PIE 830 to the output of the device matching polarity.
The green HZ SYNC LED pulses in synch with the input frequency.
Signals above the maximum scale are limited by protection circuitry with “OVER RANGE” flashed on the display and the red OVERLOAD LED lit.
The PIE 830 measures the input signal and constantly updates the display with the current reading. Move the EZ-CHECK switch
q to MAX to see the highest reading and to MIN to see the
lowest reading. Press and hold the knob e to clear the MAX and MIN readings.
Frequency Output Signal
+
-
2 1
830.00 Hz
Flowmeter
Flow Sensor
Variable Speed Drive
Controller
Transmitter
All Frequency Connections
Page 36
23
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
Source V∙ pH
4
1
RTD
+
-
T/C
Read Pressure
Choose this function to measure pressure in one of 32 different engineering units using a PIE Pressure Module.
1) Move the power switch w to READ then Double Click the EZ-DIAL knob to get into the Menu. Turn the knob to scroll through the settings and press the knob to make your selection. Select PRESSURE for the FUNCTION and make your choice of UNITS to match the pressure instrument to be checked.
2) Remove the covers from the ends of the connector on the pressure module cable and the pressure connector of the 830. Align the white arrows and plug the cable into the 830.
3) Connect pressure hoses, fittings & pumps (if required) to the pressure instrument to be checked.
4) Press and hold the
e E-Z DIAL KNOB for 2 seconds (after
MAX/MIN RESET appears) to 'Zero' or 'Tare' the pressure. The display will briefly display 'TARE ON' then a '0' appears on the display indicating that all measurements are relative to the pressure measured when the calibrator was zeroed. Press and hold the
e E-Z DIAL KNOB for 2 seconds (after MAX/MIN
RESET appears) when you want to turn the 'Tare' off.
The PIE 830 measures the pressure and constantly updates the display with the current reading. Move the EZ-CHECK switch
q
to MAX to see the highest reading and to MIN to see the lowest reading. Press and hold the knob e for 1 second to clear the MAX and MIN readings.
To disconnect the pressure module gently pull up on the groved area just above the white arrow.
Port for Pressure Module
23
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
+
Source V∙ pH
4
1
RTD
-
T/C
Page 37
PIE 830 with Pressure Module,
Pressure/Vacuum Pump & Hose
Hands free carrying case with pockets for the PIE 830 and
the pressure module. Back of the case has a zippered pocket
for the manual, test leads, hoses and pressure fittings.
Page 38
Optional Pressure Modules
Sensor Code Application
DNxxxx Differential, Non-isolated
0 to 0010*, 0028, 0200, 0415, 2000” H2O
DIxxxx Differential, Isolated
0 to 0001, 0005, 0015, 0030, 0100, 0300, 0500 PSID
GIxxxx Gauge, Isolated
0 to 0015, 0030, 0050, 0100, 0300, 0500, 1000, 3000 PSIG
CIxxxx Compound, Isolated
-14.7 to +0015, 0030, 0050, 0100, 0300, 0500, 1000, 3000 PSIG
AIxxxx Absolute, Isolated
0 to 0017, 0038, 0100, 1000 PSIA
Media Compatibility
Non-isolated DN sensors: clean, dry, non-corrosive, non-condensing gases only Isolated DI sensors: any media compatible with 316L SS & Viton®
Isolated GI, CI & AI sensors: any media compatible with 316L SS
Pressure Module Accuracy
±0.025% of full scale including all effects of linearity, repeatability and hysteresis from -20° to +50°C (-4° to +122°F) * The DN0010 sensor accuracy is ±0.050% of full scale
32 Engineering Units: PSI • inches, feet, mm, cm and meter of H2O @ 4°C, 20°C & 60°F inches, meter, cm and mm of Hg @ 0°C torr • kg/cm2 • kg/m2 • Pa • hPa • kPa • MPa • Bar • mBar ATM • oz/in2 • lb/ft2
Page 39
Specifications
General
Operating Temp Range
Storage Temp Range -30 to 60 °C (-22 to 140 °F)
Temperature effect ≤ ± 0.01 %/°C of Full Scale; Cold
Relative Humidity
Range
Isolation: Voltage
Common Mode
Normal Mode Rejection
Noise
Size
Weight
Batteries Four “AA” Alkaline 1.5V (LR6)
Battery life
Low Battery Low battery indication with nominal 1
Protection against
misconnection
Display
-20 to 60 °C (-5 to 140 °F)
Junction Sensor ≤ ±25 ppm/°C
10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing
60V rms between all mA functions/Read V DC and Source V DC/thermocouple/ RTD/Ohms//Frequency/Pressure
50/60 Hz, 100 dB
50/60 Hz, 50 dB
≤ ± ½ Least Significant Digit from 0.1 to 10 Hz
5.63x3.00x1.60” 143x76x41mm (LxWxH)
12.1 ounces, 0.34 kg with boot & batteries
Read Functions: ≥ 20 hrs; Pressure ≥ 7 hrs Source mA: ≥ 14 hours @ 12 mA into 250Ω Pwr/Meas mA: ≥ 12 hours at 20 mA Source V, Ω, T/C, RTD & Hz:
hour of operation left
Over-voltage protection to 60 vrms (rated for 30 seconds)
Red LED indicates OVERLOAD or out of range conditions
High contrast graphic liquid crystal display,
0.315” (8.0 mm) high digits, backlighting
≥ 20 hours
Page 40
Specifications
Read mA
Ranges and Resolution
Accuracy ≤ ± (0.02 % of Reading + 0.003 mA)
Voltage burden ≤ 2V at 24 mA
Overload/Current limit protection
Source mA / Power & Measure Two Wire Transmitters
Ranges and Resolution
Accuracy ≤ ± (0.02 % of Reading + 0.003 mA)
Loop compliance voltage
Loop drive capability 1200 Ω at 20 mA for 15 hours nominal;
mA 2-Wire Transmitter Simulation
Accuracy Same as Source/Power & Measure
Voltage burden ≤ 2V at 20 mA
Overload/Current limit 24 mA nominal
Loop voltage limits 2 to 60 VDC (fuse-less protected)
DC Voltage Read
Range and Resolution
Accuracy
Input resistance ≥ 1 MΩ
0.000 to 24.000 mA or -25.00 to 125.00% of 4-20 mA
25 mA nominal
Same as Read mA
≥ 24 DCV @ 20.00mA
950 Ω with Hart Resistor enabled
±99.999 mV, ±999.99mV, 0 to 10.250 V, 0.00 to 60.00 V DC
≤ ± (0.02 % of Reading + 0.01% Full Scale)
Page 41
Specifications
Source V dc
Ranges and Resolution
Accuracy
Source Current
Sink Current > 16 mA
Output Impedance < 1 Ohm
Short Circuit Duration Infinite
pH Source
Range and Resolution
Accuracy in mV
Accuracy in pH ≤ ± 0.003 pH @ 25°C
Thermocouple Source
Accuracy ≤ ± (0.02 % of Reading + 0.01 mV)
Cold Junction Compensation
Output Impedance < 1 Ohm
Source Current > 20 mA (drives 80 mV into 10 Ohms)
Thermocouple Read
Accuracy & Cold
Junction Compensation
Input Impedance
Open TC Threshold; Pulse
-20.000 to 99.999 mV, -500.00 to 999.99 mV,
0.000 to 10.250V
≤ ± (0.02 % of Reading + 0.01% Full Scale)
≥ 20 mA
-414.00 to +414.00 pH
≤ ± (0.02 % of Reading in mV + 0.1 mV)
± 0.05°C; Thermistor traceable to NIST for 11 years
Same as Thermocouple Source
> 1 Megohms
10K Ohms; <5 µamp pulse for 300 milliseconds (nominal)
Page 42
Specifications
RTD, OHMS and Continuity Read
Resistance Ranges 0.00 to 401.00, 0.0 to 4010.0 Ohms
Accuracy ±(0.025% of Reading + 0.075 Ohms)
Excitation Current
Continuity 0.0 to 401.0 Ohms; Beeps from 0.0 to
RTD and OHMS Source
3 Wire & 4 Wire Accuracy
F ro m 1 t o 1 0. 2 m A External Excitation Current
B e l o w 1 m A o f E x t e r n a l Excitation Current
2 Wire Accuracy
Resistance Ranges 0.00 to 401.00, 0.0 to 4010.0 Ohms
Allowable Excitation
Current Range
Pulsed Excitation Current Compatibility
1.0 mA to 401 Ohms, 0.5 mA to 4010 Ohms (nominal)
100.0 Ohms
±(0.025% of Full Scale + 0.075 Ohms)
±(0.025% of Full Scale+0.075 Ohms +
0.025 mV
mA Excitation Current
Add 0.1 Ohm to specifications listed above
<401 Ohm:10.2 mA max; steady or pulsed/
intermittent
401 to 4000 Ohms: 1 mA max; steady or pulsed/intermittent
DC to 0.01 second pulse width
)
Page 43
Specifications
Frequency Source
Ranges 1 to 2000 CPM, 0.01 to 999.99 Hz, 0.1
Accuracy
Output Waveform
Risetime (10 to 90% of amplitude)
Output Impedance < 1 Ohm
Source Current > 1 mA rms at 20 kHz
Short Circuit Duration Infinite
Optical Coupling Green LED (HZ SYNC) flashes at
Frequency Read
Ranges & Accuracy Same as Frequency Source
Accuracy
Trigger Level 1 V rms, dc coupled
Input Impedance > 1 Meg Ohm + 60 pF
to 9999.9 Hz, 0.001 to 20.000 kHz
± (0.02 % of Reading + 0.01% Full Scale)
Square Wave, Zero Crossing -1.0 to +5 V peak-to-peak ±10%
< 10 microseconds
output frequency
± (0.02 % of Reading + 0.01% Full Scale)
Page 44
Thermocouple Ranges & Accuracies
Based on ≤ ± (0.02 % of Reading + 0.01 mV)
T/C Degrees C
Range
J -200.0 to -50.0 ±0.5° -328.0 to -58.0 ±1.0°
-50.0 to 300.0 ±0.2° -58.0 to 572.0 ±0.4°
300.0 to 900.0 ±0.3° 572.0 to 1652.0 ±0.6°
900.0 to 1200.0 ±0.4° 1652.0 to 2192.0 ±0.8°
K -230.0 to -50.0 ±1.2° -382.0 to -58.0 ±2.2°
-50.0 to 550.0 ±0.3° -58.0 to 1022.0 ±0.6°
550.0 to 1000.0 ±0.5° 1022.0 to 1832.0 ±0.8°
1000.0 to 1371.1 ±0.6° 1832.0 to 2500.0 ±1.1°
T -260.0 to -230.0 ±2.9° -436.0 to -382.0 ±5.2°
-230.0 to -210.0 ±1.0° -382.0 to -346.0 ±1.9°
-210.0 to -50.0 ±0.8° -346.0 to -58.0 ±1.4°
-58.0 to 50.0 ±0.3° -58.0 to 122.0 ±0.6°
50.0 to 400.0 ±0.2° 122.0 to 752.0 ±0.4°
E -240.0 to -200.0 ±0.9° -400.0 to -328.0 ±1.7°
-200.0 to 0.0 ±0.5° -328.0 to 32.0 ±0.8°
0.0 to 350.0 ±0.2° 32.0 to 662.0 ±0.3°
350.0 to 1000.0 ±0.3° 662.0 to 1832.0 ±0.6°
R -18.3 to 100.0 ±2.1° -1.0 to 212.0 ±3.8°
100.0 to 500.0 ±1.3° 212.0 to 932.0 ±2.4°
500.0 to 1400.0 ±1.0° 932.0 to 2552.0 ±1.8°
1400.0 to 1767.8 ±1.2° 2552.0 to 3214.0 ±2.0°
Note: Doesn’t include cold junction error of ±0.05°C
°C Degrees F
Range
°F
Page 45
Thermocouple Ranges & Accuracies
Based on ≤ ± (0.02 % of Reading + 0.01 mV)
T/C Degrees C
Range
S -18.3 to 100.0 ±2.0° -1.0 to 212.0 ±3.7°
100.0 to 350.0 ±1.4° 212.0 to 662.0 ±2.5°
350.0 to 1600.0 ±1.1° 662.0 to 2912.0 ±2.0°
1600.0 to 1767.8 ±1.3° 2912.0 to 3214.0 ±2.4°
B 315.6 to 600.0 ±3.2° 600.0 to 1122.0 ±5.7°
600.0 to 850.0 ±1.7° 1122.0 to 1562.0 ±3.1°
850.0 to 1100.0 ±1.3° 1562.0 to 2012.0 ±2.4°
1100.0 to 1820.0 ±1.1° 2012.0 to 3308.0 ±2.0°
N -230.0 to -150.0 ±1.9° -382.0 to -238.0 ±3.4°
-150.0 to -50.0 ±0.7° -238.0 to -58.0 ±1.2°
-50.0 to 950.0 ±0.4° -58.0 to 1742.0 ±0.8°
950.0 to 1300.0 ±0.5° 1742.0 to 2372.0 ±1.0°
G
100.0 to 350.0 ±1.7° 212.0 to 662.0 ±3.0°
(W)
350.0 to 1700.0 ±0.8° 662.0 to 3092.0 ±1.5°
1700.0 to 2000.0 ±1.0° 3092.0 to 3632.0 ±1.8°
2000.0 to 2320.0 ±1.1° 3632.0 to 4208.0 ±2.1°
C
-1.1 to 100.0 ±0.8° 30.1 to 212.0 ±1.4°
(W5)
100.0 to 1000.0 ±0.7° 212.0 to 1832.0 ±1.3°
1000.0 to 1750.0 ±1.2° 1832.0 to 3182.0 ±2.1°
1750.0 to 2320.0 ±2.0° 3182.0 to 4208.0 ±3.5°
Note: Doesn’t include cold junction error of ±0.05°C
°C Degrees F
Range
°F
Page 46
Thermocouple Ranges & Accuracies
Based on ≤ ± (0.02 % of Reading + 0.01 mV)
T/C Degrees C
Range
D -1.1 to 150.0 ±1.0° 30.1 to 302.0 ±1.8°
150.0 to 1100.0 ±0.7° 302.0 to 2012.0 ±1.3°
1100.0 to 1750.0 ±1.0° 2012.0 to 3182.0 ±1.8°
1750.0 to 2320.0 ±2.0° 3182.0 to 4208.0 ±3.6°
P 0.0 to 600.0 ±0.3° 32.0 to 1112.0 ±0.6°
600.0 to 900.0 ±0.4° 1112.0 to 1652.0 ±0.8°
900.0 to 1200.0 ±0.6° 1652.0 to 2192.0 ±1.1°
1200.0 to 1395.0 ±0.7° 2192.0 to 2543.0 ±1.2°
L
-200.0 to -50.0 ±0.4° -328.0 to -58.0 ±0.7°
J-DIN
T-DIN
Note: Doesn’t include cold junction error of ±0.05°C
-50.0 to 300.0 ±0.2° -58.0 to 572.0 ±0.4°
300.0 to 900.0 ±0.3° 572.0 to 1652.0 ±0.5°
U
-200.0 to -50.0 ±0.6° -328.0 to -58.0 ±1.1°
-50.0 to 50.0 ±0.3° -58.0 to 122.0 ±0.5°
50.0 to 550.0 ±0.2° 122.0 to 1022.0 ±0.4°
550.0 to 600.0 ±0.3° 1022.0 to 1112.0 ±0.5°
°C Degrees F
Range
°F
Page 47
RTD Ranges & Accuracies
RTD Accuracy Based on
RTD
Type
Pt 100 Ohm
DIN/IEC/JIS 1989
1.3850 (ITS-90)
Pt 100 Ohm
(Burns)
1.3902
Pt 100 Ohm
(Old JIS 1981)
1.3916
Pt 100 Ohm
(US Lab)
1.3926
Pt 1000 Ohm
DIN/IEC/JIS 1989
1.3850
Copper 10 Ohm
1.4274 (Minco)
Copper 50
1.4280
Ni 120 Ohm
1.6720 (Pure)
Ni 110 Ohm
Bristol 7 NA 1.5801
Degrees C
340.0 to 640.0
640.0 to 850.0
-200.0 to 10.0
10.0 to 350.0
350.0 to 650.0
650.0 to 850.0
-200.0 to 20.0
20.0 to 360.0
360.0 to 650.0
650.0 to 850.0
-200.0 to 20.0
20.0 to 360.0
360.0 to 660.0
660.0 to 850.0
340.0 to 640.0
640.0 to 850.0
-200.0 to 260.0
-50.0 to 150.0
-80.0 to 260.0
-100.0 to 260.0
±(0.025% of Reading in Ohms + 0.075 Ohms)
Range
-200.0 to 0.0
0.0 to 340.0
-200.0 to 0.0
0.0 to 340.0
°C
±0.2°
±0.3°
±0.4°
±0.5°
±0.2°
±0.3°
±0.4°
±0.5°
±0.2°
±0.3°
±0.4°
±0.5°
±0.2°
±0.3°
±0.4°
±0.5°
±0.2°
±0.3°
±0.4°
±0.5°
±2.0°
±0.4°
±0.1°
±0.2°
Degrees F
Range
-328.0 to 32.0
248.0 to 644.0
644.0 to 1184.0
1184.0 to 1562.0
-328.0 to 50.0
50.0 to 662.0
662.0 to 1202.0
1202.0 to 1562.0
-328.0 to 68.0
68.0 to 680.0
680.0 to 1202.0
1202.0 to 1562.0
-328.0 to 68.0
68.0 to 680.0
680.0 to 1220.0
1220.0 to 1562.0
-328.0 to 32.0
248.0 to 644.0
644.0 to 1184.0
1184.0 to 1562.0
-328.0 to 500.0
-58.0 to 302.0
-112.0 to 500.0
-148.0 to 500.0
°F
±0.4° ±0.6° ±0.8° ±1.0°
±0.4° ±0.6° ±0.8° ±0.9°
±0.4° ±0.6° ±0.8° ±0.9°
±0.4° ±0.6° ±0.8° ±0.9°
±0.4° ±0.6° ±0.8° ±1.0°
±3.6°
±0.8°
±0.3°
±0.4°
Page 48
Standard Warranty
Our equipment is warranted against defective material and workmanship (excluding batteries) for a period of three years from the date of shipment. Claims under warranty can be made by returning the equipment prepaid to our factory. The equipment will be repaired, replaced or adjusted at our option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our warranty. No responsibility is accepted for damage, loss or other expense incurred through sale or use of our equipment. Under no condition shall Practical Instrument Electronics, Inc. be liable for any special, incidental or consequential damage.
Pressure sensors that have been damaged by over pressurization or contaminated by process chemicals are not covered by our warranty. Pneumatic pumps that are contaminated with process chemicals are also not covered by our warranty.
Optional Repair/Replacement Warranty
Under our Repair/Replacement Warranty (RP-WAR-B), our equipment is warranted against ANY damage or malfunction that may cause the unit to fail for a period of three (3) years from the date of shipment.
This warranty is limited to one complete replacement against any damage or malfunction during the warranty period. If replaced, the new calibrator will carry our Standard Warranty for the remainder of the three (3) years or a minimum of one (1) year from the date of shipment.
Additional Information
PIE Calibrators are manufactured in the USA. calibrated on equipment traceable to NIST and includes a Certificate of Calibration. Test Data is available for an additional charge.
Practical Instrument Electronics recommends a calibration interval of one year. Contact your local representative for recalibration and repair services.
Page 49
This product is
Practical Instrument Electronics
82 East Main Street Suite 3.14
Webster, NY 14580 USA
Tel: 585.872.9350 • Fax: 585.872.2638
sales@piecal.com • www.piecal.com
Copyright © 2017 All rights reserved 830-9002 - Rev D 23 February 2017
Loading...