PIE Practical Instrument Electronics 820-Elite Specifications

Practical Instrument Electronics
82 E. Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • sales@piecal.com • www.piecal.com
Multifunction Diagnostic Process Calibrator
Carry eight single function calibrators in the palm of your hand!
• Lighten up your toolbox
Compact calibrator replaces toolbox of single function devices
Milliamp • Voltage • Frequency • pH • Resistance Thermocouples • RTDs • Check Continuity • Pressure
• Technician friendly operation
Intuitive EZ-DIAL Double Click Menu makes it easier to setup than other multifunction calibrators. Similar to the single function PIE Calibrators.
• Use it as a milliamp and voltage calibrator
Source 0 to 24.000 mA, 0 to 10.250 V dc, -20.000 to 99.999 mV and
-500.00 to 999.99 mV Read to 24.000 mA, 60.00 V dc ±99.999 mV and ±999.99 mV Simulate 2-Wire Transmitters Power up transmitters & loops with the built-in 24 V power supply. Simplify HART hookups with built-in 250 Ohm resistor
PIE
820-ELITE
• Troubleshoot loop problems
Quickly diagnose ground fault and current leakage with patented loop diagnostic technology (US Patent 7,248,058).
• Calibrate directly in temperature to 0.1°C & 0.1°F
Compatible with the instruments you use including all brands of smart transmitters and PLCs with
• Checkout flow and vibration systems
Source & read frequency to 2000 CPM (Counts-Per-Minute), 999.99 Hz, 9999.9 Hz & to 20.000 kHz.
• Troubleshoot loop & wiring problems
‘Beep’ out connections with the built-in continuity checker.
• Simulate pH probes into transmitters & analyzers
Source from 0.000 to 14.000 pH @ 25°C (77°F) corresponding to -414.12 to +414.12 mV
• Measure pressure with optional pressure modules
32 modules for gauge, differential, compound and absolute pressure.
• Easy to read - Turn on the backlight & easily see the display in dark areas of the plant.
• Quickly set any three outputs plus automatic stepping & ramping
Easily set any value with the adjustable “DIAL” plus store any three output settings for instant recall with the EZ-CHECK™ switch. 2, 3, 5 & 11 steps automatically increment output in 100%, 50%, 25% or 10% of span. Select RAMP to smoothly increase and decrease the output. Set step/ramp time to 5, 6, 7, 8, 10, 15, 20, 25, 30 & 60 seconds.
14 thermocouple & 9 RTD types.
Actual Size
• Measure temperature sensors, frequency pickups, loop currents, voltage levels & pressures
Check the values of your process sensors. Instantly recall MAX and MIN values to see process variability.
• Evolutionary design
PIE Calibrators are designed and built by members of the same team that designed and built the calibrators manufactured by Fluke* under the Altek* label. The 820-ELITE improves upon other brands by including a rubber boot, a backlit display with larger digits, higher accuracy and more ranges for greater flexibility.
* PIECAL Calibrators are not manufactured or distributed by Fluke Corp or Altek Industries Inc, manufacturers of Altek Calibrators.
Milliamp Calibrator
• Easy to use
With the 820-ELITE you can check, calibrate and measure all your current signal instruments in a 4 to 20 milliamp DC loop. It can be used at any access point in your loop.
Source & Read 0.000 to 24.000 mA, Simulate a 2 Wire Transmitter or use the 820-ELITE to simultaneously power your 2 Wire Transmitter and measure its output.
• Source milliamps
Calibrate recorders, digital indicators, stroke valves or any instruments that get their input from a 4 to 20 mA loop. Easily set any value quickly to within 0.001 mA with the adjustable digital potentiometer “EZ-DIAL” or use preset 4.000 mA (0.00%) and 20.000 mA (100.00%) EZ-CHECK™ settings.
Voltage Calibrator
• Source three ranges of mV & V dc
With the 820-ELITE you can check, calibrate and measure all your voltage, millivolt and pH signal instruments in your plant. Source
0.000 to 10.250 V dc, -500.00 to
999.99 mV and -20.000 to 99.999 m V.
• Calibrate using loop power
Check loop wiring and receivers by using the 820-ELITE in place of a 2 Wire transmitter. Uses any loop power from 2 to 60 V DC.
• Read loop current
Check controller outputs or measure the milliamp signal anywhere in the loop. The 820-ELITE measures 0.000 to 24.000 mA (-25.00 to 125.00%) signals with greater accuracy than a typical multimeter.
• Power & measure 2 wire transmitters
The 820-ELITE can simultaneously output 24V DC to power any and all devices in a process loop using the internal batteries and internal switching power supply, while measuring the output of a 2 Wire Transmitter and any other loop devices. Powers HART™ transmitters with built-in 250 ohm resistor simplifying hookups with HART communicators.
• Read DC volts
The 820-ELITE can measure from 0.000 to 10.250 V,
-999.99 to 999.99 mV, -99.999 to 99.999 mV and 0.00 to 60.0 VDC. Use it to check loop power supplies, I/V converters, 1 to 5 Volt signals, and other voltages.
Frequency Calibrator
• Calibrate flow meters and frequency instruments
Generate zero crossing square waves to check, calibrate and measure all the frequency signal instruments in your plant. Source and read frequencies from 1 to 2000 CPM (Counts-Per-Minute), 0.01 to
999.00 Hz, 0.1 to 9999.9 Hz and
0.001 to 20.000 kHz.
Thermocouple Calibrator
• Calibrate directly in temperature to 0.1°C & 0.1°F
Stop carrying around a millivolt source and thermocouple tables. The 820-ELITE works with the thermocouples you use including types J, K, T, E, R, S, B, N, G, C, D, L (J-DIN), U (T-DIN) and P (Platinel II). Easily set any value quickly to within 0.1° with the adjustable digital potentiometer “EZ-DIAL” plus recall any three temperatures for instant recall with the EZ-CHECK™ switch.
• Checkout optical pickups
The 820-ELITE has a green LED that flashes in sync with the output frequency. Select a frequency and hold the calibrator up to the optical sensor.
• Measure frequency signals
Check the values of your process frequency outputs. Instantly recall MAX and MIN values to see process variability.
• Measure thermocouple sensors
Trouble shoot sensor connections and find broken wires or corroded connections. Connect your thermocouple with a miniature thermocouple connector and the 820-ELITE measures the probe to 0.1 degree C or F.
RTD, Resistance Calibrator
• Easy to use
With the 820-ELITE you can check & calibrate all your RTD instruments and measure RTD Sensors.
• Calibrate directly in temperature (°C & °F)
Stop carrying around a decade box and RTD resistance tables. The 820-ELITE works with the RTDs you use including Platinum 100 (alpha = 3850, 3902, 3916,
3926) & 1000 (alpha = 3850) Ohm, Copper 10 & 50 Ohm, Nickel 100 and 120 Ohm. Easily set any value quickly to within 0.1° with the adjustable digital potentiometer “EZ-DIAL” plus store any three temperatures for instant recall with the EZ-CHECK™ switch. Or use like a decade box from 0.00 to 401.00 and from 0.0 to 4010.0 Ohms.
pH Simulator
• Simulate pH probes into transmitters & analyzers
Use the pH simulator to verify proper operation of pH devices before you place a probe into a calibrated buffer. Adjusting the pH transmitter or analyzer without a probe allows you to make sure the device is calibrated and doesn’t require too much offset with the probe. If the probe requires more than the manufacturer’s recommendations (typically 5%) it is time to replace the probe. The 820-ELITE simulates 0.000 to 14.000 pH @ 25°C corresponding to -414.12 to +414.12 mV.
Continuity Checker
• Compatible with ALL process instruments
No competitor’s calibrator is compatible with as many process instruments! Connect directly to the RTD inputs of smart transmitters, PLCs, DCS and multichannel recorders and verify their outputs or displays. Works with older instruments with fixed excitation currents and newer multichannel instruments that switch the excitation current between input channels.
• Measure RTD sensors
Connect your two, three or four wire RTDs and the 820-ELITE measures the RTD to 0.1 degree C or F.
• Troubleshoot wiring and connection problems
Use the built-in continuity checker to look at wiring and connections during installation or to locate shorts. Beeps from 0 to 100 Ohms.
Connections
• Simple connections labeled on LCD
PIE 820-ELITE has banana jacks compatible with unshielded or retractable banana plugs. Included with your calibrator are a pair of test leads with alligator clips for mA, V, pH & Hz connections. Four test leads with spade lugs are also included for 2, 3 and 4 Wire RTD connections. Thermocouple connections are made through a miniature thermocouple socket and pressure modules plug into the pressure connector.
Thermocouple wire kits and pressure modules are optional accessories.
23
RTD
All mA∙ Read V
-
+
65
-
Source & Read Hz∙ mV
Source V∙ pH
+
Measure Pressure
• Easily measure pressure with a plug in pressure module
Purchase any of our 32 modules to quickly measure gauge, differential, compound & absolute pressure. 32 selectable engineering units. PSI; inches, feet, mm, cm and meter of H2O @ 4°C, 20°C & 60°F; inches, meter, cm and mm of Hg @ 0°C; torr • kg/cm2 • kg/m2 • Pa • hPa • kPa • MPa • Bar • mBar • ATM • oz/in2 • lb/ft2
4
1
RTD
-
T/C
Troubleshoot Loop Problems
• Find current leaks in loops before swapping instruments
Automatic indication Loop Current and Leakage Current (US Patent #7,248,058). Measure ground current leakage from faulty wiring, flooded
820-ELITE detects uncontrolled current in the loop due to a flooded junction box.
conduit and corrosion bridges to help you decide if there is a wiring problem in the loop (diagrams below).
Typical problem found with Leak Detection
Have you ever replaced a “faulty” transmitter only to find the problem was somewhere else in the loop? And did you end up
throwing the transmitter away after you fixed the other problem “just in case” the transmitter was faulty?
If you find a loop where the transmitter is calibrated correctly but all the readings elsewhere in the loop have a fixed offset this is due to a Zero Shift. This zero shift is typically caused by some current in the loop bypassing the transmitter. This might be caused by ground faults, moisture or corrosion.
If you have some loops that are erratic after it rains there may be moisture present in a junction box or where insulation has broken down. Turn on Ground Leak Detection and use the PIE 820-ELITE to power up the loop. Any current that isn’t controlled by the transmitter or other current control element will be indicated as leakage on the PIE 820-ELITE display.
The PIE 820-ELITE powers up the 2-Wire transmitter or loop and indicates the total current and the uncontrolled current. This provides information useful in troubleshooting loop errors.
Transmitter
input signal at
50% of Span
Transmitter
regulating
Junction
Box
12.51mA
Control System
current to
12.00 mA
Here is a loop where a technician has just recalibrated the transmitter but the control room still sees a problem. The problem started just after a rainstorm.
Why buy a PIE calibrator with loop diagnostics
Undiagnosed loop problems often cause calibration errors which can lead to dangerous operating conditions or catastrophic results. The PIE 820-ELITE is the only single display multifunction calibrator that can detect and indicate these problems due to the patented troubleshooting features.
T
ransmitter
input signal at
50% of Span
Flooded junction
box or
corroded
terminals
Transmitter
regulating
current to
LEAK ERROR = 00.48
12.492 mA
LEAKAGE: 00.51mA
PWR MEAS
LEAK ERROR = 00.48
12.492 mA
PWR MEAS
AUTOMATED DIAGNOSTIC CALIBRATOR
mA ∙ V ∙ pH ∙ TC ∙ Ω ∙ RTD ∙ Freq ∙ Pressure
SOURCE
MAX
HI
READ
SET
OFF
MIN
LO
READ
Double Click
Hz SYNC
Menu
OVERLOAD
Push & Turn
for Fast Dialing Push & Hold
to
Store/Step
12.00 mA
Using the PIE 820-ELITE to power up the loop the technician detects a leakage of 0.48 mA - approximately the offset seen in the control room. He walks the loop and opens a junction box releasing a stream of water. The loop is again in control.
Designed to work where you work
• Easy to use on the bench or in the field
The rubber boot provides more than protection...flip out the tilt stand when you need to set it on your work bench. Free up BOTH hands when you work on the plant floor or in the field. Especially handy when you have to hang onto a ladder or railing while calibrating.
Flip out stand for bench use
Hang from your neck for
hands free calibrating
Ordering Information
Description Part No
PIE Model 820-ELITE Process Calibrator .................................................. PIE Model 820-ELITE
Included:
Four “AA” Alkaline batteries, Certificate of Calibration
Blue Rubber Boot ............................................................................................... 020-0213
Evolution Hands Free Carrying Case ................................................................... 020-0211
Evolution mA/V Test Leads ................................................................................. 020-0207
1 Red & 1 Black Lead
Evolution RTD Wire Kit ....................................................................................... 020-0208
2 Red & 2 Black Leads with Banana Plugs & Spade Lugs
with Banana Plugs & Alligator Clips
Accessories
Optional
Three Year Repair/Replacement Warranty ........................................................... RP-WAR-B
Ni-MH 1 Hour Charger with 4 Ni-MH AA Batteries ............................................. 020-0103
(100-120 V AC input for North America Only)
Evolution Hands Free Carrying Case with Pressure Module Pocket .................... 020-0233*
*This case will be substituted for the Hands Free Case without the Module Pocket (020-0211) when the 820-ELITE is ordered with a pressure module
Optional Hand Pumps, Tubing & Fitting Kits
AP0V 0-300 PSI/20.7 bar Pneumatic Scissor Hand Pump ................................. 020-0224
HP0V 0-3000 PSI/206.9 bar Hydraulic Scissor Hand Pump ............................... 020-0225
DPPV 0-125 PSI/8.6 bar Pressure, 23”/584 mm Hg Vacuum hand pump ......... 020-0226
1/8” male NPT x Male Quick-Test™ Fitting with Cap ........................................... 020-0227
Adapter kit (1/8” MNPT&FNPT; 1/4” MNPT, FNPT & Tube Adapter) ................... 020-0228
Quick-Test™ 6900 psi/475 bar hose, 3ft (1 m) ................................................... 020-0229
PKIT1 (020-0224 Pneu Scissor Pump, 020-0229 Hose & 020-0227 fitting) ...... 020-0230
PKIT2 (020-0225 Hydr Scissor Pump, 020-0229 Hose & 020-0227 fitting) ...... 020-0231
PKIT3 (020-0226 Press/Vac Pump, 020-0229 Hose & 020-0227 fitting) ........... 020-0232
Optional Pressure Modules
0-10“/24.9 mbar H20 Differential, Non Isolated .................................................. DN0010
0-28“/69.7 mbar H20 Differential, Non Isolated .................................................. DN0028
0-200“/498 mbar H20 Differential, Non Isolated ................................................. DN0200
0-415“/1 bar H20 Differential, Non Isolated ........................................................ DN0415
0-2000“/5 bar H20 Differential, Non Isolated ...................................................... DN2000
0-1 PSI/68.9 mbar Differential, Isolated .............................................................. DI0001
0-5 PSI/344.7 mbar Differential, Isolated ............................................................ DI0005
0-15 PSI/1 bar Differential, Isolated .................................................................... DI0015
0-30 PSI/2 bar Differential, Isolated .................................................................... DI0030
0-100 PSI/6.9 bar Differential, Isolated ............................................................... DI0100
0-300 PSI/20.7 bar Differential, Isolated ............................................................. DI0300
0-500 PSI/34.5 bar Differential, Isolated ............................................................. DI0500
0 to 15 PSI/1 bar Gauge, Isolated ....................................................................... GI0015
0 to 30 PSI/2 bar Gauge, Isolated ....................................................................... GI0030
0 to 50 PSI/3.4 bar Gauge, Isolated .................................................................... GI0050
0 to 100 PSI/6.9 bar Gauge, Isolated .................................................................. GI0100
0 to 300 PSI/20.7 bar Gauge, Isolated ................................................................ GI0300
0 to 500 PSI/34.5 bar Gauge, Isolated ................................................................ GI0500
0 to 1,000 PSI/69 bar Gauge, Isolated ................................................................ GI1000
0 to 3,000 PSI/206.8 bar Gauge, Isolated ........................................................... GI3000
-14.7 PSIG/1 bar to +15 PSIG/1 bar Compound, Isolated .................................. CI0015
-14.7 PSIG/1 bar to +30 PSIG/2 bar Compound, Isolated .................................. CI0030
-14.7 PSIG/1 bar to +50 PSIG/3.4 bar Compound, Isolated ............................... CI0050
-14.7 PSIG/1 bar to +100 PSIG/6.9 bar Compound, Isolated ............................. CI0100
-14.7 PSIG/1 bar to +300 PSIG/20.7 bar Compound, Isolated ........................... CI0300
-14.7 PSIG/1 bar to +500 PSIG/34.5 bar Compound, Isolated ........................... CI0500
-14.7 PSIG/1 bar to +1,000 PSIG/69 bar Compound, Isolated ........................... CI1000
-14.7 PSIG/1 bar to +3,000 PSIG/ 206.8 bar Compound, Isolated ..................... CI3000
Pressure Module Media Compatibility
Non-isolated DN sensors: clean, dry,
non-corrosive, non-condensing gases only
Isolated DI sensors: any media
compatible with 316L SS & Viton®
Isolated GI, CI & AI sensors: any media
compatible with 316L SS
0-17 PSIA/1.2 bar Absolute, Isolated .................................................................. AI0017
0-38 PSIA/2.6 bar Absolute, Isolated .................................................................. AI0038
0-100 PSIA/6.9 bar Absolute, Isolated ................................................................ AI0100
0-1,000 PSIA/69 bar Absolute, Isolated .............................................................. AI1000
Measure Pressure
• Easily measure pressure with a plug in pressure module
Purchase any of the pressure modules from the table below along with one of the three hand pumps and tubing kits for a complete pressure calibration system.
Sensor Code Application Ranges Available DNxxxx Differential, Non-isolated 0 to 0010*, 0028, 0200, 0415, 2000” H2O DIxxxx Differential, Isolated 0 to 0001, 0005, 0015, 0030, 0100, 0300, 0500 PSID GIxxxx Gauge, Isolated 0 to 0015, 0030, 0050, 0100, 0300, 0500, 1000, 3000 PSIG CIxxxx Compound, Isolated -14.7 to +0015, 0030, 0050, 0100, 0300, 0500, 1000, 3000 PSIG AIxxxx Absolute, Isolated 0 to 0017, 0038, 0100, 1000 PSIA
Media Compatibility
Non-isolated DN sensors: clean, dry, non-corrosive, non-condensing gases only Isolated DI sensors: any media compatible with 316L SS & Viton® Isolated GI, CI & AI sensors: any media compatible with 316L SS
Accuracy
±0.025% of full scale including all effects of linearity, repeatability and hysteresis from -20° to +50°C (-4° to +122°F) * The DN0010 sensor accuracy is ±0.050% of full scale
32 Engineering Units: PSI • inches, feet, mm, cm and meter of H2O @ 4°C, 20°C & 60°F • inches, meter, cm and mm of Hg @ 0°C; torr • kg/cm2 • kg/m2 • Pa • hPa • kPa • MPa • Bar • mBar • ATM • oz/in2 • lb/ft2
PIE 820-ELITE with Pressure Module, Pressure/Vacuum Pump & Hose
Hand Pumps, Tubing & Fitting Kits
• Generate pressure with a full set of hand pumps
Choose from a selection hand pumps, tubing & fittings made in the USA by Ralston Instruments. All pumps have two pressure ports - one port & hose go the PIE pressure module and the other to the pressure input of your instrument.
Pneumatic Pressure/Vacuum Pump
0 to 125 psi & 0 to 23 inches vacuum
Pneumatic Scissor Hand Pump
0 to 300 psi
Hydraulic Scissor Hand Pump
0 to 3,000 psi
Quick-test™ Hoses
Microbore hoses provide a very quick, low volume, high
pressure way of connecting any pressure
instrumentation to the hand pump and pressure module.
Pressure Fitting Kit
Adapts from Quick-test™ hose to 1/4” male & female
NPT, 1/8” male & female NPT and 1/4” tube fitting
Hands free carrying case with pockets for the PIE 820-ELITE and the Pressure Module.
Back of case has zipped pocket for the manual, test leads, hoses & pressure fittings.
Hands free carrying case with pockets for the PIE 820-ELITE
and the Pressure Module.
Designed to be worn around
your neck so that you can safely
use both hands to calibrate.
Thermocouple Ranges & Accuracies @ 23°C
T/C Degrees C
Range
J -200.0 to -50.0 ±0.5° -328.0 to -58.0 ±1.0° +Iron
-50.0 to 300.0 ±0.2° -58.0 to 572.0 ±0.4°
300.0 to 900.0 ±0.3° 572.0 to 1652.0 ±0.6°
900.0 to 1200.0 ±0.4° 1652.0 to 2192.0 ±0.8°
K -230.0 to -50.0 ±1.2° -382.0 to -58.0 ±2.2° + Chromel®
-50.0 to 550.0 ±0.3° -58.0 to 1022.0 ±0.6°
550.0 to 1000.0 ±0.5° 1022.0 to 1832.0 ±0.8°
1000.0 to 1371.1 ±0.6° 1832.0 to 2500.0 ±1.1°
T -260.0 to -230.0 ±2.9° -436.0 to -382.0 ±5.2° +Copper
-230.0 to -210.0 ±1.0° -382.0 to -346.0 ±1.9°
-210.0 to -50.0 ±0.8° -346.0 to -58.0 ±1.4°
-58.0 to 50.0 ±0.3° -58.0 to 122.0 ±0.6°
50.0 to 400.0 ±0.2° 122.0 to 752.0 ±0.4°
E -240.0 to -200.0 ±0.9° -400.0 to -328.0 ±1.7° +Chromel
-200.0 to 0.0 ±0.5° -328.0 to 32.0 ±0.8°
0.0 to 350.0 ±0.2° 32.0 to 662.0 ±0.3°
350.0 to 1000.0 ±0.3° 662.0 to 1832.0 ±0.6°
R -18.3 to 100.0 ±2.1° -1.0 to 212.0 ±3.8° +Pt/13Rh
100.0 to 500.0 ±1.3° 212.0 to 932.0 ±2.4°
500.0 to 1400.0 ±1.0° 932.0 to 2552.0 ±1.8°
1400.0 to 1767.8 ±1.2° 2552.0 to 3214.0 ±2.0°
S -18.3 to 100.0 ±2.0° -1.0 to 212.0 ±3.7° +Pt/10Rh
100.0 to 350.0 ±1.4° 212.0 to 662.0 ±2.5°
350.0 to 1600.0 ±1.1° 662.0 to 2912.0 ±2.0°
1600.0 to 1767.8 ±1.3° 2912.0 to 3214.0 ±2.4°
B 315.6 to 600.0 ±3.2° 600.0 to 1122.0 ±5.7° +Pt/30Rh
600.0 to 850.0 ±1.7° 1122.0 to 1562.0 ±3.1°
850.0 to 1100.0 ±1.3° 1562.0 to 2012.0 ±2.4°
1100.0 to 1820.0 ±1.1° 2012.0 to 3308.0 ±2.0°
°C Degrees F
Range
°F T/C
Material
-Connstantan
-Alumel®
-Constantan
-Constantan
-Platinum
-Platinum
-Pt/6Rh
T/C Degrees C
Range
N -230.0 to -150.0 ±1.9° -382.0 to -238.0 ±3.4° +Nicrosil
-150.0 to -50.0 ±0.7° -238.0 to -58.0 ±1.2°
-50.0 to 950.0 ±0.4° -58.0 to 1742.0 ±0.8°
950.0 to 1300.0 ±0.5° 1742.0 to 2372.0 ±1.0°
G
100.0 to 350.0 ±1.7° 212.0 to 662.0 ±3.0° +Tungsten
(W)
350.0 to 1700.0 ±0.8° 662.0 to 3092.0 ±1.5°
1700.0 to 2000.0 ±1.0° 3092.0 to 3632.0 ±1.8°
2000.0 to 2320.0 ±1.1° 3632.0 to 4208.0 ±2.1°
C
-1.1 to 100.0 ±0.8° 30.1 to 212.0 ±1.4° +W5/Re
(W5)
100.0 to 1000.0 ±0.7° 212.0 to 1832.0 ±1.3°
1000.0 to 1750.0 ±1.2° 1832.0 to 3182.0 ±2.1°
1750.0 to 2320.0 ±2.0° 3182.0 to 4208.0 ±3.5°
D
-1.1 to 150.0 ±1.0° 30.1 to 302.0 ±1.8° +W3/Re
(W3)
150.0 to 1100.0 ±0.7° 302.0 to 2012.0 ±1.3°
1100.0 to 1750.0 ±1.0° 2012.0 to 3182.0 ±1.8°
1750.0 to 2320.0 ±2.0° 3182.0 to 4208.0 ±3.6°
P 0.0 to 600.0 ±0.3° 32.0 to 1112.0 ±0.6°
600.0 to 900.0 ±0.4° 1112.0 to 1652.0 ±0.8°
900.0 to 1200.0 ±0.6° 1652.0 to 2192.0 ±1.1°
1200.0 to 1395.0 ±0.7° 2192.0 to 2543.0 ±1.2°
L
-200.0 to -50.0 ±0.4° -328.0 to -58.0 ±0.7° +Iron
J-DIN
T-DIN
Table based on Thermocouple Accuracy: ≤ ± (0.02 % of Reading +0.01 mV)
Note: Doesn’t include cold junction error of ±0.05°C
-50.0 to 300.0 ±0.2° -58.0 to 572.0 ±0.4°
300.0 to 900.0 ±0.3° 572.0 to 1652.0 ±0.5°
U
-200.0 to -50.0 ±0.6° -328.0 to -58.0 ±1.1° +Copper
-50.0 to 50.0 ±0.3° -58.0 to 122.0 ±0.5°
50.0 to 550.0 ±0.2° 122.0 to 1022.0 ±0.4°
550.0 to 600.0 ±0.3° 1022.0 to 1112.0 ±0.5°
°C Degrees F
Range
°F T/C
Material
-W26/Re
-W26/Re
-W25/Re
+Pd55/Pt31/
-Au65/Pd35
-Connstantan
-Constantan
-Nisil
Au14
RTD
Type
Pt 100 Ohm
DIN/IEC/JIS 1989
Based on ITS-90
Pt 100 Ohm
(Burns)
Pt 100 Ohm
(Old JIS 1981)
Pt 100 Ohm
(US Lab)
Alpha Degrees C
Range
1.3850
(0.00385)
1.3902
(0.003902)
1.3916
(0.003916)
1.3926
(0.003926)
-200.0 to 0.0
0.0 to 340.0
340.0 to 640.0
640.0 to 850.0
-200.0 to 10.0
10.0 to 350.0
350.0 to 650.0
650.0 to 850.0
-200.0 to 20.0
20.0 to 360.0
360.0 to 650.0
650.0 to 850.0
-200.0 to 20.0
20.0 to 360.0
360.0 to 660.0
660.0 to 850.0
°C
±0.2° ±0.3° ±0.4° ±0.5°
±0.2° ±0.3° ±0.4° ±0.5°
±0.2° ±0.3° ±0.4° ±0.5°
±0.2° ±0.3° ±0.4° ±0.5°
RTD Ranges & Accuracies
Degrees F
Range
-328.0 to 32.0
248.0 to 644.0
644.0 to 1184.0
1184.0 to 1562.0
-328.0 to 50.0
50.0 to 662.0
662.0 to 1202.0
1202.0 to 1562.0
-328.0 to 68.0
68.0 to 680.0
680.0 to 1202.0
1202.0 to 1562.0
-328.0 to 68.0
68.0 to 680.0
680.0 to 1220.0
1220.0 to 1562.0
°F
±0.4° ±0.6° ±0.8° ±1.0°
±0.4° ±0.6° ±0.8° ±0.9°
±0.4° ±0.6° ±0.8° ±0.9°
±0.4° ±0.6° ±0.8° ±0.9°
DIN/IEC/JIS 1989
Copper 10 Ohm
Copper 50 Ohm
(Bristol 7 NA)
Table based on 3 & 4 Wire RTD Accuracy: ≤ ± (0.02 % of Reading +0.075 Ohms)
RTD
Type
Pt 1000 Ohm
(Minco)
Ni 120 Ohm
(Pure)
Ni 110
Alpha Degrees C
Range
1.3850
(0.00385)
1.4274
(0.004274)
1.4280
(0.00428)
1.6720
(0.00672)
1.5801
(0.005801)
-200.0 to 0.0
0.0 to 340.0
340.0 to 640.0
640.0 to 850.0
-200.0 to 260.0
-50.0 to 150.0
-80.0 to 260.0
-100.0 to 260.0
°C
±0.2° ±0.3° ±0.4° ±0.5°
±2.0°
±0.4°
±0.1°
±0.2°
Degrees F
Range
-328.0 to 32.0
248.0 to 644.0
644.0 to 1184.0
1184.0 to 1562.0
-328.0 to 500.0
-58.0 to 302.0
-112.0 to 500.0
-148.0 to 500.0
°F
±0.4° ±0.6° ±0.8° ±1.0°
±3.6°
±0.8°
±0.3°
±0.4°
PIE 820-ELITE Specifications
(Unless otherwise indicated all specifications are rated from a nominal 23°C, 70% RH for 1 year from calibration)
General
Operating Temperature Range
Storage Temperature Range -30 to 60 °C (-22 to 140 °F)
Temperature effect ≤ ± 0.005 %/°C of Full Scale;
Relative Humidity Range 10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
Normal Mode Rejection 50/60 Hz, 50 dB
Common Mode Rejection 50/60 Hz, 120 dB
Noise ≤ ± ½ Least Significant Digit from 0.1 to 10 Hz
Size 5.63 x 3.00 x 1.60 in, 143 x 76 x 41mm (L x W x H)
Weight 12.1 ounces, 0.34 kg with boot & batteries
Batteries Four “AA” Alkaline 1.5V (LR6)
Optional NiMh battery kit
Battery Life Read Functions: ≥ 20 hours
Low Battery Low battery indication with nominal 1 hour of
Protection against misconnection
Display High contrast graphic liquid crystal display with
Read mA
Ranges and Resolution
Accuracy ≤ ± (0.02 % of Reading + 0.003 mA)
Voltage burden ≤ 2V at 24 mA
Overload/Current limit protection
Source mA / Power & Measure Two Wire Transmitters
Ranges and Resolution
Accuracy ≤ ± (0.02 % of Reading + 0.003 mA)
Loop compliance voltage
Loop drive capability 1200 Ω at 20 mA for 15 hours nominal;
mA 2-Wire Transmitter Simulation
Accuracy Same as Source/Power & Measure
Voltage burden ≤ 2V at 20 mA
Overload/Current limit protection
Loop voltage limits 2 to 60 VDC (fuse-less protected from reverse
Voltage Read
Range and Resolution
Accuracy ≤ ± (0.02 % of Reading + 0.01% Full Scale)
Input resistance ≥ 1 MΩ
Rechargeable
Specifications subject to change without notice.
-20 to 60 °C (-5 to 140 °F)
Cold Junction Sensor ≤ ±25 ppm/°C
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing
120 VAC for North America Only; charger, four NiMh batteries, AC & DC cords [Part # 020-0103]
Source mA: ≥ 14 hours @ 12 mA into 250Ω Pwr/Meas mA: ≥ 12 hours at 20 mA Source V, Ω, T/C, RTD & Hz: ≥ 20 hours
operation left
Over-voltage protection to 60 vrms (rated for 30 seconds)
Red LED indicates OVERLOAD or out of range conditions
0.315” (8.0 mm) high digits. LED backlighting for use in low lit areas.
0.000 to 24.000 mA or -25.00 to 125.00% of 4-20 mA
25 mA nominal
0.000 to 24.000 mA or -25.00 to 125.00% of 4-20 mA
≥ 24 DCV @ 20.00mA
950 Ω with Hart Resistor enabled
24 mA nominal
polarity connections)
±99.999 mV, ±999.99mV, 0 to 10.250 V, 0.00 to 60.00 V DC
Source V dc
Ranges and Resolution
Accuracy ≤ ± (0.02 % of Reading + 0.01% Full Scale)
Source Current
Sink Current > 16 mA
Output Impedance < 1 Ohm
Short Circuit Duration Infinite
pH Source
Range and Resolution 0.000 to 14.000 pH
Accuracy in mV ≤ ± (0.02 % of Reading in mV + 0.1 mV)
Accuracy in pH ≤ ± 0.003 pH @ 25°C
Thermocouple Source
Accuracy ≤ ± (0.02 % of Reading + 0.01 mV)
Cold Junction Compensation
Output Impedance < 1 Ohm
Source Current > 20 mA (drives 80 mV into 10 Ohms)
Thermocouple Read
Accuracy & Cold Junction Compensation
Input Impedance
Open TC Threshold; Pulse
RTD, OHMS and Continuity Read
Resistance Ranges 0.00 to 401.00, 0.0 to 4010.0 Ohms
Accuracy ±(0.02% of Reading + 0.075 Ohms)
Excitation Current
Continuity 0.0 to 401.0 Ohms; Beeps from 0.0 to 100.0 Ohms
RTD and OHMS Source
3 Wire & 4 Wire Accuracy
F r o m 1 t o 1 0. 2 m A
External Excitation Current
B e l o w 1 m A o f E x t e r n a l Excitation Current
2 Wire Accuracy
Resistance Ranges 0.00 to 401.00, 0.0 to 4010.0 Ohms
Allowable Excitation Current Range
Pulsed Excitation Current Compatibility
Frequency Source
Ranges 1 to 2000 CPM, 0.01 to 999.99 Hz, 0.1 to 9999.9 Hz,
Accuracy ±(0.02% of Reading + 0.01% of Full Scale)
Output Waveform
Risetime (10 to 90% of amplitude)
Output Impedance < 1 Ohm
Source Current > 1 mA rms at 20 kHz
Short Circuit Duration Infinite
Optical Coupling Green LED (HZ SYNC) flashes at output frequency
Frequency Read
Ranges & Accuracy Same as Frequency Source
Accuracy ±(0.02% of Reading + 0.01% of Full Scale)
Trigger Level 1 V rms, dc coupled
Input Impedance > 1 Meg Ohm + 60 pF
-20.000 to 99.999 mV, -500.00 to 999.99 mV, 0.000 to 10.250V
≥ 24 mA
± 0.05°C - Thermistor traceable to NIST for 11 years
Same as Thermocouple Source
> 1 Megohms
10K Ohms; <5 µamp pulse for 300 milliseconds (nominal)
1.0 mA to 401 Ohms, 0.5 mA to 4010 Ohms (nominal)
±(0.02% of Reading + 0.075 Ohms)
±(0.02% of Reading+0.075 Ohms + )
Add 0.1 Ohms to 3 Wire & 4 Wire Accuracy
<400 Ohm Range:10.2 mA max; steady or pulsed/intermittent 4000 Ohms Range: 1 mA max; steady or pulsed/intermittent
DC to 0.01 second pulse width
0.001 to 20.000 kHz
Square Wave, Zero Crossing -1.0 to +5 V peak-to-peak ±10%
< 10 microseconds
0.025 mV
mA Excitation Current
Standard Warranty
Practical Instrument Electronics
82 East Main Street Suite 3.14 • Webster, NY 14580 USA
Tel: 585.872.9350 • Fax: 585.872.2638
sales@piecal.com • www.piecal.com
Copyright © 2017 All rights reserved 820-ELITE-9001 Rev D 23 Feb 2017
Our equipment is warranted against defective material and workmanship (excluding batteries) for a period of three years from the date of shipment. Claims under warranty can be made by returning the equipment prepaid to our factory. The equipment will be repaired, replaced or adjusted at our option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our warranty. No responsibility is accepted for damage, loss or other expense incurred through sale or use of our equipment. Under no condition shall Practical Instrument Electronics, Inc. be liable for any special, incidental or consequential damage.
Optional Repair/Replacement Warranty
Under our Repair/Replacement Warranty (RP-WAR-B), our equipment is warranted against ANY damage or malfunction that may cause the unit to fail for a period of three (3) years from the date of shipment.
This warranty is limited to one complete replacement against any damage or malfunction during the warranty period. If replaced, the new calibrator will carry our Standard Warranty for the remainder of the three (3) years or a minimum of one (1) year from the date of shipment.
Additional Information
PIE Calibrators are manufactured in the USA. calibrated on equipment traceable to NIST and includes a Certificate of Calibration. Test Data is available for an additional charge.
Practical Instrument Electronics recommends a calibration interval of one year. Contact your local representative for recalibration and repair services.