PIE Practical Instrument Electronics 520 Operating Manual

Model 520/521
TC Simulator(s)
Practical Instrument Electronics

Basic Keypad Operations

EZ-Check™ Switch
Slide the switch to select from three user stored values for the desired calibration points. The User can select HI, DIAL and LO positions. These values can easily be changed to suit the calibration requirements.
Operating Instructions
ON/OFF Switch
Slide the ON/OFF Switch to turn the Model 520/521 on or off.
3 EZ-Dial™ Knob
The EZ-Dial™ Knob has two adjustment speeds. Simply turning the EZ-Dial™ Knob will select fine adjustments. While pressing down and turning the EZ-Dial™ Knob will make course adjustments.
Note: When the EZ-Check™ Switch is in the HI or LO position, pressing and holding the EZ- Dial™ Knob without turning or LO values.
will store new HI
.

Model 520/521 Configuration

Instructions for Enabling and Disabling the Configuration Options
1. Turn the Model 520/521 on with the ON/OFF Switch
2. Press the EZ-Dial Knob CONFIGURATION” message is displayed.
3. Select options by turning the EZ-Dial Knob until the arrow points to the desired option.
4. The option can be enabled or disabled by tapping the EZ-Dial Knob.
The Model 520/521 configuration menu will exit automatically after 5 seconds of inactivity and go to normal operation with the options selected. These options are recalled at turn on until they are changed again.
841 Holt Road #1 Webster, NY 14580 Tel: 585-872-9350 Fax: 585-872-2638 sales@piecal.com http://www.piecal.com/ Practical Instrument Electronics, Inc. Copyright 2003. All rights reserved.
while the “PRESS EZ-DIAL KNOB FOR
3
2
.
520/521-9002 2/27/03 1-3
Model 520/521
Operating Instructions

Model 520/521 Configuration

Model 520/521 Configuration Menu

Auto Off ON (default)/OFF

If Auto Off is ON, the unit will turn off after 30 minutes to save battery life, if there is no user activity. If Auto Off is OFF the unit will stay on until it is turned off from the keypad. This is typically useful for manual loading or continuous use.
Display Units °C (default)/°F
Pressing the EZ-Dial Knob to toggles between °C or °F
TC {tc type} Model 520: The T/C type is fixed as
ordered from the factory and cannot be changed.
Model 521: To change T/C type, press the EZ-Dial Knob. Turn the EZ-Dial Knob to scroll through the list of available types. Press again to save and return to the configuration menu.
{tc type} is one of:
B, E, J, K, N, R, S, T, or mV
{tc type} is any of:
B, E, J, K, N, R, S, T, or mV

EZ-Dial™ Knob

Turning the EZ-Dial Knob to adjust the output up or down. Fine adjustments can be made by turning the EZ-Dial Knob. Coarse adjustments can be made by pressing and turning the EZ-Dial™. New values can be stored into the HI and LO EZ-Check™ positions by pressing down on the EZ-Dial Knob until “STORED” is indicated on the display.

EZ-Check™ Switch

The EZ-Check™ Switch has three positions: HI, DIAL, and LO. Its position is shown at the left edge of the display with “HI” and “LO” indicators. The output is adjustable in all three positions. The EZ-Check™ Switch allows user-selected values to be stored in the HI and LO positions when used in combination with the EZ-Dial Knob.
To store new EZ-Check™ value(s):
1. Dial the display to match the desired stored outputs for the HI or LO positions.
2. Press down on the EZ-Dial Knob until the confirmation message “STORED” appears.
3. Recall the values by moving the switch between HI, DIAL, and LO.
Operational description: When returning to previously set EZ-Check™ positions, the DIAL position always recalls the last output value it was dialed to. The HI and LO positions will recall the last STORED value, NOT the last output value it was last dialed to.
Hint: For faster calibrations, the product has been designed so the position of the switch can be felt. This tactile feature allows continuous monitoring of the device being calibrated without looking back at the Model 520/521 display. This is also useful in poor lighting or under difficult operating conditions.
841 Holt Road #1 Webster, NY 14580 Tel: 585-872-9350 Fax: 585-872-2638 sales@piecal.com http://www.piecal.com/ Practical Instrument Electronics, Inc. Copyright 2003. All rights reserved.
520/521-9002 2/27/03 2-3
Model 520/521
Operating Instructions

Connection Diagram

Two Wire Connection to Transmitter

WARRANTY

Our equipment comes with a NIST traceable Certificate and is guaranteed against defective material and workmanship (excluding batteries) for a period of three years from the date of shipment. Claims under guarantee can be made by returning the equipment prepaid to our factory. The equipment will be repaired, replaced or adjusted at our option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our guarantee. No responsibility is accepted for damage, loss or other expense incurred through sale or use of our equipment. Under no condition shall Practical Instrument Electronics, Inc. be liable for any special, incidental or consequential damage.
841 Holt Road #1 Webster, NY 14580 Tel: 585-872-9350 Fax: 585-872-2638 sales@piecal.com http://www.piecal.com/ Practical Instrument Electronics, Inc. Copyright 2003. All rights reserved.
520/521-9002 2/27/03 3-3