Model 510/511
RTD Simulators
Practical Instrument Electronics
Features
Simulate RTD Temperature Outputs
Calibrate directly in temperature for your RTD curve
Adjustable output for full temperature range
Several Manufacturers’ RTD Curves Available
Platinum, Copper & Nickel
Accurate to ±0.25°C (±0.5°F) with up to 0.1° Resolution
Available on the Model 511
Works with a wide variety of transmitters including
popular Rosemount and Honeywell Models
Compatible with devices using pulsed excitation currents
including PLCs, DCS, recorders, and all others
EZ-Dial Knob
Easily adjust output by 0.1° (Model 511) or 1°F (Model 510)
Pressing down and turning will select a faster dialing speed
EZ-Check Switch
User settable EZ-Check for 0% and 100% span adjustments
Store new EZ-Check values by pressing the EZ-Dial Knob
Uses a standard 9V Alkaline Battery
Superior battery life of 45 hours under typical continuous
usage
Easy access to battery compartment
Lightweight, Rugged and Reliable
Small, tough and protected to 60V
Datasheet
Models pictured above:
Model 510 – Single Curve, 1°, selectable °C or °F
Model 511 – Multi-type 7 Curves, 0.1° with selectable °C or
°F and Ω with 0.01Ω resolution
(Pt100: α=1.3850, 1.3902, 1.3916, 1.3926 and Cu10, Ni110,
Ni120, Ω)
APPROVED!
Description
The Practical Instrument Electronics Models 510 and 511 RTD Simulators simulate standard RTD curves over the entire industrial
temperature range. Choose between seven standard RTD curves. The Model 510 can also be supplied with a custom curve for your
specific application such as Pt 200, 500, 1000 with up to 0.1° resolution. Contact the factory for details.
The Model 510/511 will simulate RTD resistances into all types of instruments such as transmitters, recorders, controllers, alarms,
data acquisition, and computer systems. Rest easy knowing these calibrators are 100% compatible with pulsed systems and
transmitters like the Rosemount 3144 Transmitter. The Model 510/511 is a superior replacement for decade boxes, so there is no
need to lug a decade box around or be prone to error by reading RTD tables incorrectly. The Models 510 and 511 have better
accuracy, functions and compatibility than many higher priced RTD calibrators.
The EZ-Check function allows the user to store three output temperatures for real convenience. This will save time for repetitive
calibrations by instantly recalling the three stored temperature values. Three output settings can be stored, and all settings are
saved, even with the power off.
Practical Instrument Electronics, Inc. Copyright 2003. All rights reserved.
510-9001 2/28/05
1-2

Model 510/511 Specifications
Specifications
General Specifications:
(Unless otherwise indicated all specifications are rated from a nominal 23 °C, 70 % RH for 1 year from calibration)
Temperature Range -25 to 60 °C (-10 to 140 °F)
Relative Humidity Range 10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing
Size 4.9 X 3.15 X 1.82 inches (125.5 X 80 X 46.2 mm)
Weight 9.1 oz (258 grams)
Battery 9V Alkaline provides 45 hours of continuous use
Miscellaneous Low battery indication with nominal 1 hour of operation left
Protection to 60V DC or AC peak up to 30 seconds in duration
High contrast graphic liquid crystal display with 0.357” (9.07 mm) high digits
RTD Curve Simulation Specifications (ITS-90 Curves):
Accuracy
Typical accuracies for RTD curves are:
Pt100
Cu10
Ni110, Ni 120Ω
Allowable Excitation Current
for accuracies below 100µA add
Pulsed Excitation Current Compatibility DC to 0.01 second pulse widths
Output Dial Adjustment Resolution 0.1°F or 0.1°C Adjustment Resolution for Model 511
Temperature Coefficient
Available Options:
Carrying Case Part Number: 020-0201
±(0.015% of Setting in Ω + 0.05Ω)
±0.25°C (±0.5°F)
±1.5°C (±3°F)
±0.25°C (±0.5°F)
100 µA to 10.2 mA, steady or pulsed/intermittent/smart
±10µV/Excitation Current (units are in Ω)
1°C or 1°F Adjustment Resolution for Model 510
±0.05Ω/°C Ambient
Ordering Information
Model 510 - RTD Source (Single Type/1° resolution)
Model 511 - RTD Source (7 Types, Ω/0.1° resolution) Order Code: Model 511
Order Code:
Model 510-Pt100-1 (α=1.3850)
Model 510-Pt100-2 (α=1.3902)
Model 510-Pt100-3 (α=1.3916)
Model 510-Pt100-4 (α=1.3926)
Model 510-Cu10
Model 510-Ni110
Model 510-Ni120
Warranty
Our equipment is guaranteed against defective material and workmanship (excluding batteries) for a period of three years from the
date of shipment. Claims under guarantee can be made by returning the equipment prepaid to our factory. The equipment will be
repaired, replaced or adjusted at our option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our
guarantee. No responsibility is accepted for damage, loss or other expense incurred through sale or use of our equipment. Under no
condition shall Practical Instrument Electronics, Inc. be liable for any special, incidental or Consequential damage.
Practical Instrument Electronics, Inc. Copyright 2003. All rights reserved.
510-9001 2/28/05
2-2