PIE Practical Instrument Electronics 422, 422 Plus Data sheet

PIE 422
Preliminary
& Milliamp Calibrator
Best Thermocouple & Milliamp Calibrator on the market!
• The only 2-in-1 thermocouple & full function milliamp calibrator on the market
Only calibrator that combines ALL the functions of a diagnostic milliamp calibrator with the accuracy of a laboratory thermocouple
calibrator.
=
• Protect instruments & technicians from potentially dangerous catastrophic failures due to hidden loop problems
Quickly diagnose ground fault and current leakage often caused by water in conduits and junction boxes with patented loop diagnostic technology (US Patent# 7,248,058).
• Best accuracy & stability found in a handheld calibrator
The internal cold junction is accurate to ±0.05°C and is traceable to NIST. Isothermal brass blocks with screw terminals allow for connection of bare thermocouple wires plus a miniature thermocouple connector for fast connections. More accurate than any other handheld thermocouple calibrator.
• Verify Heat Treating Uniformity Survey Recorders & perform System Accuracy Tests
The PIE 422Plus is capable of meeting the requirements of an AMS 2750 Field Test Instrument when certified by an accredited laboratory.
Half the size of the closest competitor & easiest to use
Fits the palm of your hand like a cell phone and weighs less than a pound. Automatic indication of connections on the display for simple hookups. Carry it without worry - it comes protected with a rubber boot and rugged, low profile switches. Easy to operate even in dark areas of the plant with the backlit display.
Actual Size
Practical Instrument Electronics
82 E. Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • sales@piecal.com • www.piecal.com
More functions and better accuracy in one calibrator
• Two complete calibrators
Unlike any other combination thermocouple/milliamp calibrator on the market the PIE 422Plus includes a FULL function diagnostic milliamp calibrator that sources and reads 0.000 to 24.000 mA, simulates two wire transmitters, powers and measures transmitters with its internal 24V supply, has a built-in 250 ohm resistor for HART compatibility, and PIE’s patented Ground Leak detection. Competitive calibrators allow you to calibrate temperature transmitters but don’t have the ability to test other instruments in a 4 to 20 milliamp loop. The other half of the calibrator sources and reads thermocouple types J, K, T, E, R, S, B, N, G, C, D, L (J-DIN), U (T-DIN), P (Platinel II), and from -13.000 to
80.000 mV.
• Protect instruments & technicians from potentially dangerous catastrophic failures due to hidden loop problems
It’s important to find hidden loop problems BEFORE they become cause product problems or catastrophic failures. Only PIE Calibrators have Ground Leak Detection that can detect these problems. Some technicians routinely check their critical loops for ground faults and current leaks using PIE’s Ground Leak detection as part of their preventative maintenance programs. When a significant leak or fault is detected it is located and fixed before catastrophic failures occur. Sometimes the control room detects a problem that isn’t fixed with routine calibration of the instruments. The technician then reaches for a calibrator with Ground Leak Detection to power up the loop and confirm that the issue is hiding somewhere in the loop wiring.
• State of the art accuracy & stability
It’s important to have a calibrator that is at least 4 times more accurate than the instruments being calibrated. Both the milliamp and thermocouple sections were designed with state of the art components that are the most accurate and stable found in a handheld calibrator. The internal cold junction is accurate to ±0.05°C and is traceable to NIST. The sensor is thermally bonded to an isothermal mass including brass blocks with screw terminals for connection of bare thermocouple wires along with a miniature thermocouple connector for fast connections. An extremely stable voltage reference and low drift components make it more accurate than other handheld thermocouple calibrators.
• Verify Heat Treating Uniformity Survey Recorders & perform System Accuracy Tests
The AMS 2750 standard calls out particular specifications that need to be met for a calibrator to be used as a Field Test Instrument. The Field Test Instrument is then used to verify the thermocouple measuring thermometers and controllers that are used in the heat treating of parts for the aerospace industry. The PIE 422Plus is capable of meeting the requirements of an AMS 2750 Field Test Instrument when certified by an accredited laboratory.
• Half t
he size
of the closest competitor &
easiest to use
Don’t get tired by lugging around heavy oversized test equipment when you can carry the palm sized PIE 422Plus that weighs less than a pound. It comes with a deluxe carrying case for simple, hands free operation. The simple and intuitive double click menu and EZ-Check switch are faster and easier to use than calibrator with lots of confusing buttons or with confusing menus. And with the optional magnet strap you can attach it to a panel or loop the strap around a
conduit or pipe. Many calibrators don’t have automatic stepping and ramping while others limit you to selecting either slow or fast fixed 25% steps. The PIE 422Plus lets you choose between 2, 3, 5, 11 and 21 steps to automatically increment the output in 100%, 50%, 25%, 10% or 5% of span. Select the step time to match your system from 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 and 60 seconds.
Ground Leak Detection
Prevent Small Problems from Becoming Major Plant Failures
The PIE 422Plus offers powerful troubleshooting tools that offer visibility to the health of your loop that is not possible with any calibrator brand on the market. Ground Leak Detection allows technicians to find and fix undetected problems before they cause quality issues or become catastrophic failures - ending in poor product, injury, loss of life, or equipment damage. Common loop problems caused by moisture, corrosion and contamination left unchecked can lead to dangerous conditions that are easily avoided with the innovation technology in the PIE 422Plus.
Even the most common small errors caused by loop wiring issues can lead to inferior product, lost production time and risk to personnel. Intrinsically safe loops are protected by barriers only against extreme overvoltage and over current conditions but allowing small but significant ground faults to go undetected. PIE’s unique and powerful Ground Leak Detection technology quickly and easily finds a fault that would otherwise go undetected. These undetected faults could lead to potentially disastrous outcomes.
Have you ever replaced a “faulty” transmitter only to find the problem was somewhere else in the loop? And did you end up throwing the transmitter away after you fixed the other problem “just in case” the transmitter was faulty? If you find a loop where the transmitter is calibrated correctly but all the readings elsewhere in the loop have a fixed offset this is due to a Zero Shift. This Zero Shift is typically caused by some of the current in the loop bypassing the transmitter. If you have some loops that are erratic after it rains there may be moisture present in a junction box or where insulation has broken down.
Only a PIE Calibrator with patented Ground Leak Detection can find all these hidden problems. Turn on Ground Leak Detection and use the PIE 422Plus to power up the loop. Any current that isn’t controlled by the transmitter or other current control element will be clearly indicated as leakage on the PIE 422Plus display.
Ground Leak detection allows technicians to troubleshoot loop problems with ease and confidence - minimizing risk and maximizing process uptime.
Hidden Problem Found with Ground Leak Detection
Safe loop
Sealed junction box is dry and the control system is operating within parameters.
Dry
Junction
Transmitter
input signal at
50% of Span
Box
+ IN - REF +OUT-
Transmitter
regulating
current to
12.00 mA
12.00mA
+
Control
­System
Dangerous loop left undetected
The flooded junction box goes undetected and the control system potentially goes further out of control until...
Flooded
Junction
Transmitter
input signal at
50% of Span
Box
+ IN - REF +OUT-
Transmitter
regulating
current to
12.00 mA
12.51mA
+
Control
­System
Sealed junction box is wet* and the control system is operating with a fixed offset of 0.51 milliamps!
Transmitter
input signal at
50% of Span
* The value of the current leak increases when the water in the junction box is contaminated with rust, salt or process chemicals.
OR
Hidden problem is detected by the PIE 422Plus. The flooded sealed junction box is emptied and the control system is once again operating within parameters. The product, equipment, and personnel are safe!
Dangerous loop
Flooded
Junction
+ IN - REF +OUT-
Transmitter
regulating
current to
12.00 mA
Disaster averted
Box
12.51mA
+
Control
­System
Accessories
Deluxe Hands Free
Carrying Case (Included)
Thermocouple Wire Kit 1
J, K, T & E (Optional)
Magnetic Hanging Strap
(Optional)
Included:
Blue Rubber Boot, Four “AA” Alkaline batteries, Certificate of Calibration Deluxe Hands Free Carrying Case Part No. 020-0211 Evolution mA/V Wire Kit Part No. 020-0207
1 Red & 1 Black Lead with Banana Plugs & Alligator Clips
Thermocouple, Millivolt & Milliamp Connections
Simulating or reading thermocouples requires the use of thermocouple or extension grade thermocouple wire. Plug thermocouple wires into the miniature thermocouple jack or place bare thermocouple wires onto the brass block under the screws.
The PIE 422Plus has two banana jacks (1+ and 2-) mounted in the top end of the housing. These are not temperature compensated and are to be used only for millivolt signals. The two banana jacks (5+ and 6-) are for all milliamp signals.
Thermocouple Wire Kit 2
B, R/S, N (Optional)
Optional:
T/C Wire Kit 1 for Types J, K, T & E Part No. 020-0202 T/C Wire Kit 2 for Types B, R/S & N Part No. 020-0203
Three feet (1 meter) of T/C extension wire, stripped on one end
with a miniature T/C male connector on the other end. Magnetic Hanging Strap Part No. 020-0236 Ni-MH Charger with 4 Ni-MH AA Batteries Part No. 020-0103
PIE 422Plus Specifications
(Unless otherwise indicated all specifications are rated from a nominal 23°C, 70% RH for 1 year from calibration)
General
Operating Temperature Range
Relative Humidity Range 10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
Temperature Drift
Size
Weight 12.1 ounces, 0.34 kg (including boot & batteries)
Batteries Four “AA” Alkaline 1.5V (LR6)
Battery Life 25 Hours Thermocouple, 8 Hours milliamp
Isolation: Voltage 60V rms between all milliamp functions/Read V DC
Normal Mode Rejection 50/60 Hz, 50 dB
Common Mode Rejection 50/60 Hz, 120 dB
Optional NiMh
Rechargeable battery kit
Low Battery Low battery indication with nominal 1 hour of
Protection against misconnection
Display High contrast graphic liquid crystal display. LED
-25 to 60 °C (-10 to 140 °F)
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing
± 0.01% of span outside of 23°C ±10 °C (73°C ±18 °F)
5.63 x 3.00 x 1.60 in, 143 x 76 x 41 mm (L x W x H)
and Source V DC/Thermocouple
Charger, four
operation left
Over-voltage protection to 60 V dc (rated for 30 seconds)
backlighting for use in low lit areas.
NiMh batteries [Part # 020-0103]
Source Thermocouple
Accuracy ±(0.008% of Reading + 0.006 mV)
Cold Junction
Compensation
Millivolt Range -13.000 to 80.000 mV
Output Impedance < 0.3 Ohms
Source Current > 20 mA (drives 80 mV into 10 Ohms)
Noise ≤ 4 microvolts p-p for frequencies of 10 Hz or below
Read Thermocouple
Accuracy ±(0.008% of Reading + 0.006 mV)
Cold Junction
Compensation
Millivolt Range -13.000 to 80.000 mV
Input Impedance
Open Thermocouple
Threshold Pulse
Normal Mode Rejection 50/60 Hz, 50 dB
Common Mode Rejection 50/60 Hz, 120 dB
± 0.09°F (±0.05 °C) - Thermistor traceable to NIST for 11 years
± 0.09°F (±0.05 °C) - Thermistor traceable to NIST for 11 years
> 10 Megohms
10,000 Ohms nominal
< 10 microamp pulse for 400 milliseconds
Read mA
Ranges and Resolution
Accuracy ≤ ± (0.02 % of Reading + 0.003 mA)
Voltage burden ≤ 2V at 24 mA
Overload/Current limit protection
Source mA / Power & Measure Two Wire Transmitters & PWRM LEAK
Ranges and Resolution
Accuracy ≤ ± (0.02 % of Reading + 0.003 mA)
Loop compliance voltage
Loop drive capability 1200 Ω at 20 mA for 15 hours nominal;
mA 2-Wire Transmitter Simulation
Accuracy Same as Source/Power & Measure
Voltage burden ≤ 2V at 20 mA
Overload/Current limit protection
Loop voltage limits 2 to 60 VDC (fuse-less protected from reverse
0.000 to 24.000 mA or -25.00 to 125.00% of 4-20 mA
25 mA nominal
0.000 to 24.000 mA or -25.00 to 125.00% of 4-20 mA
≥ 24 DCV @ 20.00mA
950 Ω with Hart Resistor or leak detection running
24 mA nominal
polarity connections)
Warranty
Our equipment is warranted against defective material and workmanship (excluding batteries) for a period of three years from the date of shipment. Claims under warranty can be made by returning the equipment prepaid to our factory. The equipment will be repaired, replaced or adjusted at our option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our warranty. No responsibility is accepted for damage, loss or other expense incurred through sale or use of our equipment. Under no condition shall Practical Instrument Electronics, Inc. be liable for any special, incidental or consequential damage.
Additional Information
PIE Calibrators are designed, assembled, and calibrated in the USA. This product is calibrated on equipment traceable to NIST and includes a Certificate of Calibration. Test Data is available for an additional charge.
Practical Instrument Electronics recommends a calibration interval of one year. Contact your local representative for recalibration and repair services.
Thermocouple Ranges & Accuracies
Based on ≤ ± (0.008 % of Reading + 0.006 mV)
T/C Degrees C
Range
J -200.0 to -180.0 ±0.3° -328.0 to -292.0 ±0.5° +Iron
-180.0 to -50.0 ±0.2° -292.0 to -58.0 ±0.4°
-50.0 to 500.0 ±0.1° -58.0 to 932.0 ±0.2°
500.0 to 1200,0 ±0.2° 932.0 to 2192.0 ±0.4°
K -230.0 to -100.0 ±0.6° -382.0 to -148.0 ±1.1° + Chromel®
-100.0 to 1050.0 ±0.2° -148.0 to 1922.0 ±0.4°
1050.0 to 1371.1 ±0.3° 1922.0 to 2500.0 ±0.5°
T -260.0 to -200.0 ±1.0° -436.0 to -328.0 ±1.8° +Copper
-200.0 to -50.0 ±0.5° -328.0 to -58.0 ±0.9°
-50.0 to 0.0 ±0.2° -58.0 to 32.0 ±0.4°
0.0 to 400.0 ±0.1° 32.0 to 752.0 ±0.2°
E -240.0 to -200.0 ±0.4° -400.0 to -328.0 ±0.7° +Chromel
-200.0 to -100.0 ±0.2° -328.0 to -148.0 ±0.4°
-100.0 to 850.0 ±0.1° -148.0 to 1562.0 ±0.2°
850.0 to 1000.0 ±0.2° 1562.0 to 1832.0 ±0.4°
R -18.3 to 250.0 ±1.2° -1.0 to 482.0 ±2.2° +Pt/13Rh
250.0 to 750.0 ±0.6° 482.0 to 1382.0 ±1.1°
750.0 to 1600.0 ±0.5° 1382.0 to 2192.0 ±0.9°
1600.0 to 1767.8 ±0.6° 2192.0 to 3214.0 ±1.1°
S -18.3 to 100.0 ±1.2° -1.0 to 212.0 ±2.1° +Pt/10Rh
100.0 to 400.0 ±0.8° 212.0 to 752.0 ±1.4°
400.0 to 1700.0 ±0.6° 752.0 to 3092.0 ±1.1°
1700.0 to 1767.8 ±0.7° 3092.0 to 3214.0 ±1.3°
B 315.6 to 550.0 ±1.8° 600 to 1022.0 ±3.2° +Pt/30Rh
550.0 to 900.0 ±1.1° 1022.0 to 1652.0 ±2.0°
900.0 to 1150.0 ±0.7° 1652.0 to 2102.0 ±1.3°
1150.0 to 1820.0 ±0.6° 2102.0 to 3308.0 ±1.1°
°C Degrees F
Range
°F T/C
Material
-Connstantan
-Alumel®
-Constantan
-Constantan
-Platinum
-Platinum
-Pt/6Rh
Note: Doesn’t include cold junction error of ±0.05°C
T/C Degrees C
Range
N -230.0 to -180.0 ±1.0° -382.0 to -292.0 ±1.8° +Nicrosil
-180.0 to -50.0 ±0.5° -292.0 to -58.0 ±0.9°
-50.0 to 1100.0 ±0.2° -58.0 to 2012.0 ±0.4°
1100.0 to 1300.0 ±0.3° 2012.0 to 2372.0 ±0.5°
G
100.0 to 150.0 ±1.2° 212.0 to 302.0 ±2.2° +Tungsten
(W)
150.0 to 400.0 ±0.8° 302.0 to 752.0 ±1.4°
400.0 to 1700.0 ±0.4° 752.0 to 3092.0 ±0.7°
1700.0 to 2320.0 ±0.7° 3092.0 to 4208.0 ±1.3°
C
-1.1 to 1500 ±0.5° 30.0 to 2372.0 ±0.9° +W5/Re
(W5)
(W3)
Platinel®
J-DIN
T-DIN
1500 to 1900 ±0.6° 2372.0 to 3452.0 ±101°
1900.0 to 2100.0 ±0.7° 3452.0 to 3812.0 ±1.3°
2100.0 to 2320.0 ±0.9° 3812.0 to 4208.0 ±1.6°
D
-1.1 to 50.0 ±0.6° 30.0 to 122.0 ±1.1° +W3/Re
50.0 to 1400.0 ±0.4° 122.0 to 2552.0 ±0.7°
1400.0 to 1800.0 ±0.5° 2552.0 to 3272.0 ±0.9°
1800.0 to 2320.0 ±0.9° 3272.0 to 4208.0 ±1.6°
0.0 to 1000.0 ±0.2° 32.0 to 1832.0 ±0.4°
P
1000.0 to 1395.0 ±0.3° 1832.0 to 2543.0 ±0.5°
L
-200.0 to -50.0 ±0.2° -328.0 to -58.0 ±0.4° +Iron
-50.0 to 500.0 ±0.1° -58.0 to 932.0 ±0.2°
500.0 to 900.0 ±0.2° 932.0 to 1652.0 ±0.4°
U
-200.0 to -75.0 ±0.3° -328.0 to -103.0 ±0.5° +Copper
-75.0 to 100.0 ±0.2° -103.0 to 212.0 ±0.4°
100.0 to 600.0 ±0.1° 212.0 to 1112.0 ±0.2°
°C Degrees F
Range
DIN Colors
°F T/C
Material
-Nisil
-W26/Re
-W26/Re
-W25/Re
+Pd55/Pt31/Au14
-Au65/Pd35
-Connstantan
-Constantan
Practical Instrument Electronics
82 Est Main Street Suite 3.14 Webster, NY 14580
Tel: 585.872.9350 • Fax: 585.872.2638
sales@piecal.com • www.piecal.com
Copyright © 2017 All rights reserved 422Plus-9001 - Rev A 23 June 2017
Available From:
Loading...