With the PIECAL 322-1 you can check & calibrate all
your thermocouple instruments and measure
thermocouple Sensors. Automatic indication of
connections on the display for simple hookups.
• Take it without into the shop, plant or field
Carry it without worry - it comes protected with a
rubber boot and rugged, low profile switch. Easy to
operate even in the dark areas of the plant with the
backlit display.
• Calibrate directly in temperature (°C & °F)
Stop carrying around a millivolt source and
thermocouple tables. The PIECAL 322-1 works with
the T/Cs you use including types J, K, T & E. Easily
set any value quickly to within 0.1° with the adjustable
digital potentiometer “DIAL” plus store any three
temperatures for instant recall with the EZ-CHECK™
switch.
• Calibrate quickly with automatic output stepping
Choose between 2, 3, 5, 11 and 21 steps to
automatically increment the output in 100%, 50%,
25%, 10% or 5% of span. Select the step time to
match your system from 5, 6, 7, 8, 10, 15, 20, 25, 30
and 60 seconds.
• Compatible with all process instruments
Connect directly to the thermocouple inputs of smart
transmitters, PLCs, DCS and multichannel recorders
and verify their outputs or displays.
• Measure Thermocouple Sensors
Trouble shoot sensor connections and find broken wires or high resistance junctions. Connect
your sensor and the PIECAL 322 automatically measures the thermocouple in degrees C or F.
Select “SOURCE” to output in °C, °F, or millivolts.
Select “READ” to read a thermocouple sensor
or millivolts.
HI
322.1
TYPE K °C
THE RM OCO UP LE
CAL IB RATO R
J, T, E, K, mV
SOURCE
MAX
HI
READ
SET
OFF
MIN
LO
OVERLOAD
READ
Push & Turn
Fast Dialing
Push & Hold
Store/Ramp
Double Click
for Options
for
to
q EZ-CHECK™ SWITCH
SOURCE: Instantly output t wo preset thermocouple
temperatures by moving the EZ-CHECK™ switch to
the “LO” position or “HI” position.
point checks select the “DIAL” position. The
PIECAL 322-1 will remember the last “DIAL”
value, even with the power off.
These values can easily be changed to suit the
calibration requirements. The temperatures stored in
the HI and LO positions are also used for Auto
Stepping.
READ: Slide the switch to the DIAL position. The
PIECAL 322-1 will display the current temperature
from the thermocouple sensor. Slide the switch to HI
and the highest temperature measured since turn-on
or reset will be displayed; slide the switch to LO and
the lowest temperature measured since turn-on or
reset will be displayed.
For fast three
e EZ-DIAL™ KNOB
SOURCE: Turn the knob to adjust the output
level. Turn clockwise to increase the output,
counter clockwise to decrease the output in 0.1°
steps at a time. Push down and turn the
EZ-DIAL knob for faster dialing.
Press and hold the knob for two seconds to store
desired EZ-Check™ HI/LO points in SOURCE
mode. Continue to press and hold the knob for
two more seconds to start the automatic
ramping.
READ: Press and hold to transfer the current
temperature into the EZ-Check™ HI/LO points.
This resets the HI/LO temperature readings
which will update as the temperature changes.
Double click the knob to get into the PIECAL
322-1 Configuration Mode. Use configuration to
select °C or °F, T/C Type, Backlight On/Off, Step
Size, Step Time and Auto Off On/Off.
CHANGING BATTERIES
Low battery is indicated by “BAT” on the display.
Approximately one to four hours of typical
operation remain before the PIECAL 322-1 will
automatically turn off. To change the batteries;
remove the rubber boot, remove the battery door
from the back of the unit by sliding the door
downward. This allows access to the battery
compartment. Replace with four (4) “AA” 1.5V
batteries being careful to check the polarity.
Replace the battery door and replace the boot. All
stored configuration options (T/C Type,
EZ-CHECK Memories, etc., are reset to factory
settings when the batteries are removed.
Note: Alkaline batteries are supplied and
recommended for maximum battery life and
performance.
saves any changes. Menu will automatically exit
after 15 seconds of inactivity (countdown timer is
displayed).
TEMP UNITS - pressing the knob will toggle
between °C and °F.
T/C - pressing the knob will cycle through T/C
types J, K, T, E and mV.
Setup
Double click the
e DIAL KNOB at any time the
unit is on and the following displays will appear
for 15 seconds:
> EXIT 15
TEMP UNITS °C
T/C K
BACKLIGHT ON
> EXIT 15
STEPS 3
STEP TIME 8
AUTO OFF ON
Turn the e DIAL KNOB to move through the
menu. Press the e DIAL KNOB to toggle
between OFF and ON or to scroll through the
settings.
Basic Operation
Connecting the Calibrator
Thermocouple
Sourcing or reading thermocouples requires the
use of thermocouple or extension grade
thermocouple wire between the PIECAL 322 and
the device being calibrated.
The PIECAL 322 has a female miniature
thermocouple connector mounted in the top end
of the housing for fast, easy connection of wires.
Millivolt
The PIECAL 322 has two banana jacks mounted
in the top end of the housing. These jacks are not
temperature compensated and are to be used
only for generating and measuring millivolt signals.
BACKLIGHT - If BACKLIGHT is ON the
backlight will light all the time the unit is powered
up. For maximum battery life turn the backlight
off when using the calibrator in areas with
enough ambient light to read the display.
STEPS - pressing the knob will cycle through 2,
3, 5, 11 and 21 steps. The endpoints of the steps
are based on the values stored in the HI and LO
EZ-CHECK outputs.
STEP TIME - pressing the knob will cycle
through 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 and 60
seconds.
AUTO OFF - If AUTO OFF is ON, the unit will
turn off after 30 minutes of inactivity to save
battery life. If AUTO OFF is OFF the unit will stay
on until the POWER SWITCH is moved to the off
position.
Note: All settings are remembered even with the
power off. Removing the batteries resets the
values to factory defaults.
Choose this function to provide a simulated thermocouple signal into controllers, temperature
transmitters, indicators or any input devices that measure thermocouple sensors..
1) Disconnect the thermocouple sensor from the
device to be calibrated.
2) Select “SOURCE” with slide switch
w.
3) Connect a thermocouple wire (matching the type
of wire to sensor being simulated) with miniature
male T/C connector to the inputs of the device
being calibrated, making sure to check polarity.
Millivolt outputs (without cold junction compensation) may be connected with a copper (white)
miniature thermocouple connector or through
the banana jacks with cooper wire.
The output is adjusted in 0.1° (or 0.001 mV)
increments by turning the knob
e while the
EZ-CHECK™ switch q is in the “HI”, “LO” or “SET”
position. Press and turn the knob for faster dialing
HI
322.1
THE RM OCO UP LE
CAL IB RATO R
J, T, E, K, mV
MAX
HI
READ
SET
MIN
LO
OVERLOAD
TYPE K °C
SOURCE
OFF
Push & Turn
Fast Dialing
Push & Hold
Store/Ramp
Double Click
for Options
READ
for
to
with 10° (or 0.100) increments.
The OVERLOAD indicator will light if excessive
voltage or current is detected by the calibrator.
Reading Thermocouple Sensors
READ
Choose this function to measure temperatures with a thermocouple probe or sensor.
1) Disconnect the thermocouple sensor from any other device.
2) Select “READ” with slide switch
w.
3) Place the EZ-CHECK switch into the READ position.
4) Connect a thermocouple probe (matching the type of wire to sensor being measured) with miniature male T/C connector to the sensor, making sure to check polarity. Millivolt outputs (without
cold junction compensation) may be connected with a copper (white) miniature thermocouple
connector or through the banana jacks with cooper wire.
The PIECAL 322 measures the temperature signal and constantly updates the display with the
current temperature reading. Move the EZ-CHECK switch to MAX to see the highest temperature
reading and to MIN to see the lowest temperature reading. Press and hold the knob
e to clear the
MAX and MIN readings.
The OVERLOAD indicator will light if excessive voltage or current is detected by the calibrator.
Choose this function to provide a simulated thermocouple signal into controllers, temperature
transmitters, indicators or any other input device that measure thermocouple sensors.
1) Store your high (SPAN) output temperature by moving the EZ-CHECK switch to the HI position
and turn the
EZ-Dial knob until STORED appears to store the value. Release the EZ-Dial knob.
2) Store your low (ZERO) output temperature by moving the EZ-CHECK switch to the LO position
and turn the
EZ-Dial knob until STORED appears to store the value. Release the EZ-Dial knob.
3) Instantly output your SPAN and ZERO temperature outputs by moving the EZ-CHECK switch
between HI and LO. You may also select any third temperature output (such as mid-range) using
the SET position on the EZ-CHECK switch.
e EZ-Dial knob until the desired temperature is on the display. Press and hold the
e EZ-Dial knob until the desired temperature is on the display. Press and hold the
Automatic Stepping
To change the Automatic Stepping settings
Double click the e DIAL KNOB at any time the unit is on and the following display will appear for 30
seconds:
Turn the e DIAL KNOB to move through the menu. Press the e
DIAL KNOB to toggle between OFF and ON or to change the
STEPS and the STEP TIME settings. These settings are
remembered even with the power off.
EXIT MENU - exits this menu immediately and saves any
changes. Menu will automatically exit after 30 seconds of
inactivity.
STEPS - pressing the knob will cycle through 2, 3, 5, 11 and 21 steps then reverse dircetion. The
endpoints of the steps are based on the values stored in the HI and LO EZ-CHECK outputs.
2 steps will automatically switch between the values stored in the HI & LO EZ-CHECK (0 &
100%).
3 steps between the HI, Midpoint and LO EZ-CHECK (0, 50 & 100%).
5 steps between the HI and LO EZ-CHECK in 25% increments (0, 25, 50, 75 & 100%).
11 steps between the HI and LO EZ-CHECK in 10% increments (0, 10, 20...80, 90 &100%).
EXIT 15
> STEPS 3
STEP TIME 8
AUTO OFF ON
21 steps between the HI and LO EZ-CHECK in 5% increments (0, 5, 10... 90, 95 & 100%).
STEP TIME - pressing the knob will cycle through 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 and 60 seconds.
To start Automatic Stepping
Start automatic stepping or ramping by placing the EZ-CHECK Switch into the HI or LO position
then press and hold the
after 3 seconds - continue to press the DIAL KNOB) until the word STEPPING appears on the
display. The word STEPPING will appear on the display anytime the selected automatic function is
running. Stop the stepping by again pressing and holding the e DIAL KNOB for 3 seconds.
e DIAL KNOB for 6 seconds (the word STORE will appear on the display
1050.0 to 1371.1±0.3°1922.0 to 2500.0±0.5°JacketYellow
T-260.0 to -200.0±1.0°-436.0 to -328.0±1.8°+CopperBlue
-200.0 to -50.0±0.5°-328.0 to -58.0±0.9°-ConstantanRed
-50.0 to 0.0±0.2°-58.0 to 32.0±0.4°JacketBlue
0.0 to 400.0±0.1°32.0 to 752.0±0.2°
E-240.0 to -200.0±0.4°-400.0 to -328.0±0.7°+ChromelPurple
-200.0 to -100.0±0.2°-328.0 to -148.0±0.4°-ConstantanRed
-100.0 to 850.0±0.1°-148.0 to 1562.0±0.2°JacketPurple
850.0 to 1000.0±0.2°1562.0 to 1832.0±0.4°
AccuracyDegrees F
Range
AccuracyT/C
Material
ISA/ANSI
Color
Accessories
T/C Wire Kit 1 for Types J, K, T & E Part No. 020-0202
Three feet (1 meter) of T/C extension wire,
stripped on one end with a miniature T/C
male connector on the other end.
Standard Test Leads Part No. 020-0207
Three feet (1 meter) of wire with an alligator
clip on one end and a spring loaded protected
banana plug on the other end.
Ni-MH 1 Hour Charger w/4 Ni-MH AA Batteries Part No. 020-0103
(100-120 V AC input for North America Only)
Warranty
Our equipment is warranted against defective material and workmanship (excluding batteries) for a
period of three years from the date of shipment. Claims under warranty can be made by returning
the equipment prepaid to our factory. The equipment will be repaired, replaced or adjusted at our
option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our
warranty. No responsibility is accepted for damage, loss or other expense incurred through sale or
use of our equipment. Under no condition shall Practical Instrument Electronics, Inc. be liable for any
special, incidental or consequential damage.