Practical Instrument Electronics
841 Holt Road #1 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • sales@piecal.com • www.piecal.com
Automated RTD Calibrator
• Easy to use
With the PIECAL 211 you can check & calibrate all your RTD
instruments and measure RTD Sensors. Automatic indication of
connections on the display for simple hookups.
• Take it into the shop, plant or field
Carry it without worry - it comes protected with a rubber boot and
rugged, low profile switch. Easy to operate even in the dark areas
of the plant with the backlit display.
• Calibrate directly in temperature (°C & °F)
Stop carrying around a decade box and RTD resistance tables. The
PIECAL 211 works with the RTDs you use including Platinum 100
(alpha = 3850, 3902, 3926) & 1000 (alpha = 3850, 3750) Ohm,
Copper 10 & 50 Ohm, Nickel 100 and 120 Ohm. Easily set any
value quickly to within 0.1° with the adjustable digital potentiometer
“DIAL” plus store any three temperatures for instant recall with the
EZ-CHECK™ switch. Or use like a decade box from 0.00 to 410.00
and from 410.0 to 2010.0 Ohms.
PIECAL 211
•
Fast calibration with automatic output stepping
Choose between 2, 3, 5, 11 and 21 steps to automatically increment
the output in 100%, 50%, 25%, 10% or 5% of span. Select the step
time to match your system from 5, 6, 7, 8, 10, 15, 20, 25, 30 and 60
seconds.
• Compatible with ALL process instruments
No competitor’s calibrator is compatible with as many process
instruments! Connect directly to the RTD inputs of smart transmitters,
PLCs, DCS and multichannel recorders and verify their outputs or
displays. Works with older instruments with fixed excitation currents and
newer multichannel instruments that switch the excitation current
between input channels.
Actual Size
• Measure RTD sensors
Trouble shoot sensor connections and find broken wires with patented technology. Connect your two, three or four
wire RTDs and the PIECAL 211 automatically detects the connections and measures the RTD in degrees C or F.
• Evolutionary design
PIECAL Calibrators are designed and built by members of the same team that designed and built the calibrators
manufactured by Fluke* under the Altek* label. The PIECAL 211 improves upon other brands by including a rubber
boot, tilt stand, backlit display with larger digits, rugged switches and a battery compartment for fast battery
changes.
* PIECAL Calibrators are not manufactured or distributed by Fluke
Corp or Altek Industries Inc, manufacturers of Altek Calibrators.

Practical Instrument Electronics
841 Holt Road #1 • Webster, NY 14580
Tel: 585.872.9350 • Fax: 585.872.2638
sales@piecal.com • www.piecal.com
Copyright © 2009 All rights reserved 211-9001 - Rev B 9 Feb 2010
PIECAL 211 Specifications
(Unless otherwise indicated all specifications are rated from a nominal 23°C, 70% RH for 1 year from calibration)
General
Accuracy ±(0.025% of Reading + 0.05 Ohms)
Temperature Drift
Operating Temperature
Range
Relative Humidity Range 10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
Size L=5.63 x W=3.00 x H=1.60 inches
Weight 12.1 ounces (including boot & batteries)
Batteries Four “AA” Alkaline 1.5V (LR6)
Battery Life 50 Hours
Optional NiMh
Rechargeable battery kit
Low Battery Low battery indication with nominal 1 hour of
Protection against
misconnection
Display High contrast graphic liquid crystal display. LED
± 0.01% of span outside of 23°C ±10 °C (73°C ±18 °F)
-25 to 60 °C (-10 to 140 °F)
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing
120 VAC for North America Only; charger, four
NiMh batteries, AC & DC cords [Part # 020-0103]
operation left
Over-voltage protection to 60 V dc (rated for 30
seconds)
backlighting for use in low lit areas.
Ranges & Accuracies
RTD
Type
Pt 100 Ohm
DIN/IEC/JIS 1989
Based on ITS-90
Pt 100 Ohm
(Burns)
Pt 100 Ohm
(Old JIS 1981)
Pt 100 Ohm
(US Lab)
Pt 1000 Ohm
DIN/IEC/JIS 1989
Pt 1000 Ohm
Hy-Cal HVAC
Copper 10 Ohm
(Minco)
Copper 50 Ohm
Ni 120 Ohm
(Pure)
Ni 110
(Bristol 7 NA)
Alpha Degrees C
Range
1.3850
(0.00385)
1.3902
(0.003902)
1.3916
(0.003916)
1.3926
(0.003926)
1.3850
(0.00385)
1.3750
(0.00375)
1.4274
(0.004274)
1.4280
(0.00428)
1.6720
(0.00672)
1.5801
(0.005801)
-200.0 to 200.0
200.0 to 600.0
600.0 to 850.0
-195.6 to 200.0
200.0 to 648.9
-200.0 to 200.0
200.0 to 648.9
-200.0 to 100.0
100.0 to 700.0
700.0 to 850.0
-200.0 to 200.0
200.0 to 266.0
-200.0 to 200.0
200.0 to 274.0
-200.0 to 260.0
-50.0 to 150.0
-80.0 to 260.0
-100.0 to 260.0
°C
±0.2°
±0.3°
±0.4°
±0.2°
±0.3°
±0.2°
±0.3°
±0.2°
±0.3°
±0.4°
±0.2°
±0.3°
±0.2°
±0.3°
±2.0°
±1.0°
±0.1°
±0.2°
Degrees F
Range
-328.0 to 392.0
392.0 to 1112.0
1112.0 to 1562.0
-320.0 to 392.0
392.0 to 1200
-328.0 to 392.0
392.0 to 1200
-328.0 to 212.0
212.0 to 1292.0
1292.0 to 1562.0
-328.0 to 392.0
392.0 to 511.0
-328.0 to 392.0
392.0 to 525.0
-328.0 to 500.0
-58.0 to 302.0
-112.0 to 500.0
-148.0 to 500.0
°F
±0.4°
±0.6°
±0.7°
±0.4°
±0.6°
±0.4°
±0.6°
±0.4°
±0.6°
±0.7°
±0.4°
±0.6°
±0.4°
±0.6°
±4.0°
±1.6°
±0.2°
±0.3°
Read
Excitation Current
Normal Mode Rejection 50/60 Hz, 50 dB
Common Mode Rejection 50/60 Hz, 120 dB
Source
Accuracy
From 1 to 10.2 mA
External Excitation Current
Below 1 mA of External
Excitation Current
Resistance Ranges 0.00 to 410.00, 410.1 to 2001.0 Ohms
Allowable Excitation
Current Range
Pulsed Excitation
Current Compatibility
0.5 mA nominal
±(0.025% of Reading + 0.05 Ohms)
±(0.025% of Rdg + + 0.05 Ohms)
<410 Ohms: 10.2 mA max; steady or pulsed/intermittent
410 to 2001 Ohms: 1 mA max; steady or pulsed/
intermittent
DC to 0.01 second pulse width
0.025 mV
mA Excitation Current
Accessories
Included:
Rubber Boot, Four “AA” Alkaline batteries, Certificate of Calibration
Small Carrying Case with PIE Logo Part No. 020-0205
Evolution RTD Wire Kit Part No. 020-0208
2 Red & 2 Black Leads with Retractable Shield Banana Plugs & Spade Lugs
Optional:
Ni-MH 1 Hour Charger with 4 Ni-MH AA Batteries Part No. 020-0103
(100-120 V AC input for North America Only)
Additional Information
PIE Calibrators are manufactured in the USA.
calibrated on equipment traceable to NIST and includes a
Certificate of Calibration. Test Data is available for an additional
charge.
Practical Instrument Electronics recommends a calibration
interval of one year. Contact your local representative for
recalibration and repair services.
This product is
Warranty
Our equipment is warranted against defective material and
workmanship (excluding batteries) for a period of three years
from the date of shipment. Claims under warranty can be made
by returning the equipment prepaid to our factory. The equipment
will be repaired, replaced or adjusted at our option. The liability
of Practical Instrument Electronics (PIE) is restricted to that
given under our warranty. No responsibility is accepted for
damage, loss or other expense incurred through sale or use of
our equipment. Under no condition shall Practical Instrument
Electronics, Inc. be liable for any special, incidental or
consequential damage.