Product Description
Easy to use
With the PIECAL 211 you can check & calibrate all your RTD instruments and measure RTD Sensors. Automatic indication of connections on the display for simple hookups.
Take it into the shop, plant or field
Carry it without worry - it comes protected with a rubber boot and rugged, low profile switch. Easy to operate even in the dark areas of the plant with the backlit display.
• Calibrate directly in temperature (°C & °F)
Stop carrying around a decade box and RTD resistance tables. The PIECAL 211 works with the RTDs you use including Platinum 100 (alpha = 3850, 3902, 3926) & 1000 (alpha = 3850, 3750) Ohm, Copper 10 & 50 Ohm, Nickel 100 and 120 Ohm. Easily set any value quickly to within 0.1° with the adjustable digital potentiometer “DIAL” plus store any three temperatures for instant recall with the EZ-CHECK™ switch.
• Calibrate quickly with automatic output stepping
Choose between 2, 3, 5, 11 and 21 steps to automatically increment the output in 100%, 50%, 25%, 10% or 5% of span. Select the step time to match your system from 5, 6, 7, 8, 10, 15, 20, 25, 30 and 60 seconds.
• Compatible with all process instruments
Connect directly to the RTD inputs of smart transmitters, PLCs, DCS and multichannel recorders and verify their outputs or displays. Works with older instruments with fixed excitation currents and newer multichannel instruments that switch the excitation current between input channels.
• Measure RTD Sensors
Trouble shoot sensor connections and find broken wires with patented technology. Connect your two, three or four wire RTDs and the PIECAL 211 automatically detects the connections and measures the RTD in degrees C or F.
PIECAL 211
Automated Universal RTD Calibrator
Operating Instructions
PIECAL 211
Automated Universal RTD Calibrator
Operating Instructions
Copyright © 2011 All rights reserved • 211-9002 - Rev D 13 July 2011
www.GlobalTestSupply.com
Practical Instrument Electronics
82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • sales@piecal.com • www.piecal.com
Page 2
q EZ-CHECK™ SWITCH
SOURCE: Instantly output two preset RTD
temperatures by moving the EZ-CHECK™ switch to the “LO” position or “HI” position.
For fast three point checks select the “DIAL” position. The PIECAL 211 will remember the last “DIAL” value, even with the power off. These values can easily be changed to suit the calibration requirements. The temperatures stored in the HI and LO positions are also used for Auto Stepping.
READ: Slide the switch to the DIAL position. The PIECAL 211 will display the current temperature from the RTD sensor. Slide the switch to HI and the highest temperature measured since turn-on or reset will be displayed; slide the switch to LO and the lowest temperature measured since turn-on or reset will be displayed.
w SOURCE/OFF/READ Switch
Select “SOURCE” to output in °C, °F or ohms. Select “READ” to read an RTD sensor or ohms.
e EZ-DIAL™ KNOB
SOURCE: Turn the knob to adjust the output
level. Turn clockwise to increase the output, counter clockwise to decrease the output in 0.1° steps at a time. Push down and turn the EZ-DIAL knob for faster dialing. Press and hold the knob for two seconds to store desired EZ-Check™ HI/LO points in SIMULATE mode. Continue to press and hold the knob for
two more seconds to start the automatic ramping.
READ: Press and hold to transfer the current temperature into the EZ-Check™ HI/LO points. This clears the HI/LO temperature readings which will update as the temperature changes.
Double click the knob to get into the PIECAL 211 Configuration Mode. Use configuration to select °C or °F, T/C Type, Backlight On/Off, Step Size, Step Time and Auto Off On/Off.
CHANGING BATTERIES
Low battery is indicated by “BAT” on the display. Approximately one to four hours of typical operation remain before the PIECAL 211 will automatically turn off. To change the batteries; remove the rubber boot, remove the battery door from the back of the unit by sliding the door downward. This allows access to the battery compartment. Replace with four (4) “AA” 1.5V batteries being careful to check the polarity. Replace the battery door and replace the boot. All stored configuration options (RTD Type, EZ-CHECK Memories, etc., are reset to factory settings when the batteries are removed.
Note: Alkaline batteries are supplied and recommended for maximum battery life and performance.
Basic Operation
322-1
THERMOCOUPLE
CALIBRATOR
J, T, E, K, mV
HI
Pt100 α=385
AUTOMATED
RTD
CALIBRATOR
READ
SOURCE
HI
SET
LO
MAX
READ
MIN
OVERLOAD
Push & Hold
to
Store/Step
Push & Turn
for
Fast Dialing
Double Click
for Options
OFF
°C
3W 2W 1W 4W
100.0
www.GlobalTestSupply.com
Practical Instrument Electronics
82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • sales@piecal.com • www.piecal.com
Page 3
Configuration
Configure the Calibrator
Move w POWER SWITCH to “SOURCE” or “READ”.
Setup
Double click the
e DIAL KNOB at any time the
unit is on and the following displays will appear for 15 seconds:
Turn the
e DIAL KNOB to move through the
menu. Press the e DIAL KNOB to toggle between OFF and ON or to scroll through the settings.
MODEL 211 V#.##
DOUBLE CLICK
EZ-DIAL KNOB
FOR CONFIGURATION
> EXIT 15 TEMP UNITS °C RTD Pt 100 a=3850 BACKLIGHT ON
> EXIT 15 STEPS 3 STEP TIME 8 AUTO OFF ON
EXIT MENU - exits this menu immediately and
saves any changes. Menu will automatically exit after 15 seconds of inactivity (countdown timer is displayed).
TEMP UNITS - pressing the knob will toggle between °C and °F.
RTD - pressing the knob will cycle through the various RTD (Pt, Cu & Ni) at different base resistances and alpha values.
BACKLIGHT - If BACKLIGHT is ON the backlight will light all the time the unit is powered up. For maximum battery life turn the backlight off when using the calibrator in areas with enough ambient light to read the display.
STEPS - pressing the knob will cycle through 2, 3, 5, 11 and 21 steps. The endpoints of the steps are based on the values stored in the HI and LO EZ-CHECK outputs.
STEP TIME - pressing the knob will cycle through 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 and 60 seconds.
AUTO OFF - If AUTO OFF is ON, the unit will turn off after 30 minutes of inactivity to save battery life. If AUTO OFF is OFF the unit will stay on until the POWER SWITCH is moved to the off position.
Note: All settings are remembered even with the power off. Removing the batteries resets the values to factory defaults.