M3500-OPT01
10 Channel Scanner Card
USER’S GUIDE
August 2007 Version 1 Printed
in Taiwan
PICOTEST CORP.
8F-1, 286-9, Hsin-Ya Rd., 80673, Kaoshiung, Taiwan
TEL: +886-7-815-7183
FAX: +886-7-815-8312
URL http://www.picotest.com.tw
Product Introduction
Picotest thanks you to purchase the
“M3500-OPT01 10 channel Scanner Card”.
To reach the best performance from the
product, please read this guide carefully.
1. Overview
The M3500-OPT01 supports the following
measurements, such as for voltage 10
points, for current 2 points, for 2-wired
resistance 10 points and for 4-wired
resistance 5 points at most.
2. Inspection & Upkeep
Inspection:
When you open the package, inspect it
carefully to make sure whether defects
occur on the appearance or malfunctions
show in the operation. Please contact with
your local reseller or PICOTEST
representative for more help.
Upkeep:
To clean the product, wipe its cover
(excluding the circuit) gently with a soft
and moistened cloth. Prevent using
solvents, such as benzene, alcohol,
acetone, ether, ketones, thinners or
gasoline because of their destructive
capabilities.
3. Safety
This safety information with the
warning and danger marks on the
user’s guide reminds users to avoid
risks as they are using it.
Warning: The triangle symbol in
black indicates that incorrect operation
might cause an injury to users or
damage to the product.
Danger: The triangle symbol in
red indicates that incorrect operation
might cause an extreme hazard to
users’ life.
4. Prenotion
Danger
z To avoid electrical shock and
personal injury, please don’t
measure the source out of
specification.
z The maximum AC voltage is 125V
rms or 175V peak, 100kHz, 1A
switched 62.5VA (resistive load),
and DC voltage is 110V, 1A
switched, 30VA (resistive load).
Warning
z To avoid breaking the product,
please do not pull it away when
measurement is executed.
5. Specifications
Maximum AC Voltage
Maximum DC Voltage 110V, 1A switched, 30VA (resistive load)
Contact Life
Contact Resistance <1ohm at end of contact life
Actuation Time 5ms maximum on/off
Contact Potential
Connector Type Screw terminal, #22 AWG wire size
Isolation btw Any >10 Gohm, < 75pF
125V rms or 175V peak, 100kHz, 1A
switched, 62.5VA (resistive load)
>100000 operations at maximum signal
level; >100000000 operations cold switching.
<±500nV typical per contact, 1μV max
<±500nV typical per contact pair, 1μV max
Two te rmi na l s
Isolation btw Any
Terminal and Earth
Common Mode Voltage 350V peak btw any terminal and earth
Max. Voltage btw Any
Two Term in als
Max. Voltage btw Any
Terminal and M3500A
Input LO
Environmental
6. Efficiency
Rate of the Scanner Card Measurement
AutoZero OFF,AutoGain OFF,AutoRange OFF, Scan Timer=0,60Hz
single
function(VDC)
Mix function
(VDC+2WRES)
7. SCPI Commands
>10 Gohm, < 150pF
200V peak
200V peak
Meets all M3500A Environmental Spec.
NPLC
(Fast 4.5) 0.02 68 29.4
(Slow 4.5 & Fast 5.5) 0.1
(Slow 5.5 & Fast 6.5) 1
(Slow 6.5) 10 408 4.9
NPLC
(Fast 4.5) 0.02 306 6.5
(Slow 4.5 & Fast 5.5) 0.1
(Slow 5.5 & Fast 6.5) 1
(Slow 6.5) 10 1710 1.2
Take Time with
2000 Readings(sec)
74 27.0
105 19.0
Take Time with
2000 Readings(sec)
318 6.3
442 4.5
rate(ch/s)
rate(ch/s)
Command Description
ROUTe:CLOSe <channel> Close channels <1 ~ 10>.
ROUTe:CLOSe? Query the closed channels
ROUTe:OPEN Open all channels.
ROUTe:STATe?
ROUTe:SCAN:FUNCtion
<channel>,{<function>|“V
OLT:DC”|“VOLT:AC”|“FREQ
uency”|
“RESistance”|“FRESistan ce”
|“NONE”}
ROUTe:SCAN:FUNC?
<channel>
ROUTe:SCAN:TIMER? Read the time interval of scanning.
ROUTe:SCAN:TIMER
<value>
ROUTe:SCAN:COUNT? Read the number of times of scanning.
ROUTe:SCAN:COUNT
<value>
ROUTe:SCAN:STATe?
ROUTe:SCAN:SCAN Run SCAN mode
ROUTe:SCAN:STEP Run STEP mode
Ask the state. The state 1 means Card
inserted or 0 means Card not i nserted.
Set card states which might measue
the VAC, VDC, Frequency, 2-Wire
Resistance, 4-Wire Resistance or
disabling the channel.
Ask the channel’s state of the card.
Set the time interval of scanning <The
unit is second>.
Set the number of times of scanning.
Read the state of scanning. 1 means
“finished”. 0 means “not finished”.
8. Measurement
z Voltage Application
z 2-Wired Resistance Application
z 4-Wired Resistance Application
z Current Application
※ Note: CH1 & CH6 can be used on other
measurement when the shunts are released.
z Output Terminal Connection
※ Note: Output B is only for 4-Wired resistance
measurement.