Warning statements identify conditions or practices
or practices
to this product or other
TA046
800 MHz ±30 V Differential Probe
User’s Manual
This probe complies with IEC-1010.1, IEC-1010.2-031 CAT III, Pollution Degree 2.
1. Safety terms and symbols
1 28.8.2008.
James House
Colmworth Business Park
St. Neots
PE19 8YP
United Kingdom
www.picotech.com
Terms appearing in this manual:
WARNING
CAUTION Caution statements identify conditions
that could result in injury or death.
that could result in damage property.
Symbols appearing on the product:
Danger
High Voltage
Protective
(Earth) Terminal
Refer to Manual
Attention
8 © Copyright Pico Technology 2008 DO142-1
DO142-1 © Copyright Pico Technology 2008 1
2. General safety summary
Please review the following safety precautions to avoid injury and prevent damage to this probe or any products that are connected to it.
Observe maximum working voltage
To avoid any injury, do not use the probe above 40 V between each input lead and earth or between the two inputs..
Do not operate without covers
To avoid electric shock or fire hazard, do not operate this probe with the covers removed.
Do not operate in wet or damp conditions
To avoid electric shock, do not operate this probe in wet or damp conditions.
Do not operate in explosive atmosphere
To avoid injury or fire hazard, do not operate this probe in an explosive atmosphere.
Avoid exposed circuitry
To avoid injury, remove jewelry such as rings, watches and other metallic objects. Do not touch exposed connections and components when power is present.
Use proper power source
To ensure proper functioning of this probe, use the 15 V DC / 100 mA regulated mains adaptor. Do not operate this probe from a power source that applies more than the voltage specified.
Do not operate the probe if it is damaged
If you suspect there is damage to this probe, have it inspected by qualified service personnel.
8. Derating curve
The derating curve for absolute maximum input voltage is as follows:
100
Voltage (V RMS)
40
10
1
1.E+08 1.E+09 1.E+10
40
20
4
Frequency (Hz)
4
9. Test procedure
a. Connect the BNC output connector to the vertical input of a general-purpose
oscilloscope.
b. Connect the AC mains adapter. c. Set the oscilloscope input coupling to DC and 0.5 V/div. Center the trace on
the display.
d. Connect the input pins of the probe to a function generator and select a
square-wave output with 10 V p-p amplitude and 100 kHz frequency.
e. A 100 kHz square-wave of 1 V p-p amplitude will be displayed on the screen of
the oscilloscope. This means the probe is working properly.
10. Cleaning
Use a soft cloth to clean the probe, taking care not to cause damage. a. Do not immerse the probe.
b. Do not use abrasive cleaners. c. Do not use chemicals contains benzene or similar solvents.
2 © Copyright Pico Technology 2008 DO142-1
DO142-1 © Copyright Pico Technology 2008 7
use only the
accessories designed for use with this differential
the circuit
It is not intended for electrically
insulating the circuit under test and the measuring
7. Specifications
Bandwidth DC to 800 MHz (Twin Pin configuration) Accuracy ±2% Input impedance 100 kΩ || 2 pF Attenuation 1/10 Input voltage
- Max. differential ±15 V
- Max. common mode ±30 V
- Absolute max. voltage ±40 V Output voltage
- Swing (into 50 Ω load) ±1.5 V
- Offset (typical) < ±5 mV Output impedance(typical) 50 Ω CMRR (typical) 60 dB @ 60 Hz; 15 dB @ 500 MHz Output BNC cable length 1.2 m
Power requirements
Ambient operating temperature -10 to +40 °C Ambient storage temperature -30 to +70 °C Ambient operating humidity 25 to 85% RH Ambient storage humidity 25 to 85% RH Weight 130 g Dimensions (LxWxH) 111 mm x 22 mm x 14 mm
15 V DC / 100 mA regulated mains adaptor
3. Description
With high bandwidth and high CMRR, this differential probe is ideal for timing analysis, troubleshooting ground bounce problems in high-speed logic, and design verification of disk drives, as well as for wireless and data communications design.
4. Installation
a. Simply plug in the BNC output connector to the vertical input of a
general-purpose oscilloscope.
b. Connect the AC mains adaptor to the correct line voltage, and then connect its
output to the power terminal of the differential probe.
c. Adjust the vertical offset (or position) of the oscilloscope.
WARNING To protect against electric shock,
CAUTION This probe is designed for carrying out differential
probe.
measurements between two points on under test.
instrument.
6 © Copyright Pico Technology 2008 DO142-1
DO142-1 © Copyright Pico Technology 2008 3
5. Appearance
The differential probe looks like this:
1. Input Pins. The input pins of the differential probe can be connected directly to the circuit under test or connected to optional accessories that come with the probe.
2. Output Lead. The BNC output connector.
3. Power Plug. The terminal connects to the 15 V DC / 100 mA regulated mains adaptor.
6. SMD micro test accessories
Descriptions Quantity
MicroFlex Pincer, Black 1 MicroFlex Pincer, Red 1 Micro Test Clip, Black 1 Micro Test Clip, Red 1 MicroLead, 0.8 mm J-P, 5 cm, Black 1 MicroLead, 0.8 mm J-P, 5 cm, Red 1 MicroLead, 0.8 mm J-P, 10 cm Black 1 MicroLead, 0.8 mm J-P, 10 cm Red 1 Twin Pin, 16.8 mm 2 Twin Pin, 12.8 mm 2 Test Tip, 0.8 mm 6
4 © Copyright Pico Technology 2008 DO142-1
DO142-1 © Copyright Pico Technology 2008 5