WARNING Before making connections to the in p ut leads of this
output lead is grounded.
You may do this either by connecting the BNC shell
, or by
the auxiliary grounding terminal to a
ad the next paragraph for
2. General safety summary
Please review the following safety precautions to avoid injury and prevent damage
to this probe or any products that are connected to it.
Observe maximum working voltage
To avoid any injury, do not use the probe above 2500 V rms CAT I between each
input lead and earth or above 5000 V rms CAT I between the two input leads.
Must be grounded
This probe is grounded with the shell of the BNC connector and an auxiliary
grounding terminal.
probe, ensure that its
You must verify that the probe is securely grounded before connecting the probe
input leads. Some types of measuring instrument, such as a USB oscilloscope
connected to a laptop, are unlikely to be grounded even if the laptop is powered
from the mains. Bench-top oscilloscopes are us ually gro unded, b ut in so me cases
may have been isolated from ground. A USB oscillosco pe connected to a deskto p
computer is usually grounded. In any case, do no t assume tha t the mea surement
instrument is grounded. Always verify the ground connection for yourself before connecting the probe input leads.
Use fused test prods if necessary
If this probe is intended to use for measurements in circuits of INSTALLATION
CATEGORY III, it should incorporate with fused test prods.
to a grounded measurement instrument
connecting
reliable ground po int. Re
further advice.
Do not operate without covers
To avoid electric shock or fire hazard, do not operate this probe with the covers
removed.
Do not operate in wet or damp conditions
To avoid electric shock, do not operate this probe in wet or damp conditions.
Do not operate in explosive atmosphere
To avoid injury or fire hazard, do not operate this probe in an explosive
atmosphere.
Avoid exposed circuitry
To avoid injury, remove jewelry such as rings, watches and other metall ic objects.
Do not touch exposed connections and components when power is present.
WARNING To protect against electric shock, use only the
accessories designed for use with this differential
CAUTION This probe is designed for carrying out differential
the circuit
under test. It is not intended for electrically
insulating the circuit under test and the measuring
To ensure proper functioning of this probe, use four AA cells or a 6 VDC / 200 mA
mains adaptor or 9 VDC / 120 mA mains adaptor or power leads.
Do not operate the probe if it is damaged
If you suspect there is damage to thi s probe , have it inspecte d by qualified service
personnel.
3. Description
By enabling conventional oscilloscopes to display and measure in-circuit
waveforms that are referenced to high common-mode voltages, this differential
probe extends the measurement capability of oscilloscopes to electronic power
converters, inverters, motor speed controls, switch-mode power supplies and
many other applications.
4. Installation
a. Simply plug in the BNC output connector to the vertical input of a
general-purpose oscilloscope or other meas urement instrument, and connect
the auxiliary grounding terminal to a proper ground. The measurement
instrument must have a ground ref erence.
b. Install four AA cells or connect an appropriate power source to this probe.
c. Select the proper attenuation ratio. When measuring signals below 700 V,
switch the attenuation ratio to 1/100 in order to get higher resolution and less
noise. Otherwise , set the attenuation to 1/1000.
probe.
d. Using the appropriate probe accessories, connect the input to the circuits
The derating curve for absolute maximum input voltage is as follows:
9. Overrange indicator
The overrange indicator lights up red if the voltage of the inp ut signal exceed s the
linear operating range of the probe. When this happens, the signal on the probe
output may not accurately represent the signa l o n the probe input.
10. Test procedure
a. Connect the BNC output connector to the vertica l input of a general-purpose
oscilloscope.
b. Install four AA cells or connect an appropriate mains adaptor to this probe.
c. Set the oscilloscop e input coupling to DC and 1 V/div. Center the trace on
the display.
d. Connect the sprung hooks of the probe to power lines.
e. Set the range of the probe to 1/100.
f. A 50 Hz / 60 Hz sine-wave of proper amplitude will be displayed on the scree n
of the oscilloscope. This means the probe is working properly.
11. Cleaning
Use a soft cloth to clean the probe, taking care not to cause damage.
a. Do not immerse the probe.
b. Do not use abrasive cleaners.
c. Do not use chemicals contains benzene or similar solvents.