Pickering Laboratories Pinnacle PCX Operator's Manual

Page 1
G
U
A
R
A
N
T
E
D
C
H
E
M
I
S
T
R
Y
APPROVED
OperatOr Manual
Post-Column
Derivatization
Instrument
1280 Space Park Way Mountain View CA 94043 USA
800-654-3330 • 650-694-6700
650-968-0749
fax
www.pickeringlabs.com support@pickeringlabs.com http://pickeringlabs.blogspot.com
Cat. No. 0101-0008
Version 10, April 2016
Copyright © 2005 by Pickering Laboratories, Inc. All rights reserved. No Portion of this manual or any artwork contained herein may be reproduced in any shape or form without the express written consent of Pickering Laboratories, Inc.
Page 2
Page 3
table Of COntents
1 Getting Started
How to Use this Manual 1
Read this First! 2
Symbols and Warnings 2
Specifications 2
Site Requirements 4
1.1 Section 1 Introduction
What is Post-column Derivatization? 1.1
Requirements for a Successful Post-column Method 1.1
Design of a HPLC system 1.2
Designing a Post-column System 1.4
Design of the Pinnacle PCX 1.6
I
2.1 Section 2 Overview
Introduction 2.2
Check Valve 2.3
Column Connections and Column Oven 2.3
Reagent Pump 2.4
Reagent Valves 2.5
Reservoir Tray / Bottles 2.5
Fluidics Panel 2.7
Quick Change Reactor 2.8
Detector Connections 2.8
Pump Compartment 2.9
Gas Manifold 2.9
Display Module 2.10
Electronics Compartment 2.11
Back Panel 2.11
Safety Features in the Pinnacle PCX 2.13
Standard Configurations of Pinnacle PCX 2.14
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 4
II
3.1 Section 3 Installation
Site Requirements 3.2
Instrument Unpacking and Preparation 3.4
Unpacking 3.4
Gas Connections to Back Panel of Pinnacle PCX 3.5
Computer Connections 3.6
Software Installation 3.6
Pump and Autosampler Connections 3.8
Eluant Priming 3.10
Column and Guard Installation 3.10
Detector Connections 3.11
Reservoir Connections 3.12
Reagent Pump Preparation 3.13
Running a Chromatogram 3.14
Shutdown 3.15
4.1 Section 4 Pinnacle PCX Operation
Software Overview 4.1
Status Window 4.2
Menu Options/Functions 4.4
File
Control
Method
Sequence
Instrument
Configuring your Pinnacle PCX 4.6
Log Files 4.7
Creating and Editing Methods 4.7
Creating and Editing Sequences 4.10
Tutorials 4.11
Running post-column analysis using Pinnacle PCX 4.14
STOP/RESET Pinnacle PCX 4.17
Pinnacle Shutdown 4.17
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 5
5.1 Section 5 Maintenance
Maintenance Suggestions 5.1
Operating Suggestions 5.2
Proper Shutdown Procedures 5.3
Basic Maintenance Procedures 5.6
Reagent Filter(s) Replacement 5.7
Ambient Reactor Replacement 5.8
Replacement of Heated Reactor Cartridge 5.8
Valve Maintenance 5.9
Pump Seal Replacement 5.12
Fuse Replacement 5.16
Tubing Kit Guides 5.17
III
6.1 Section 6 Applications
Amino Acids 6.2
Introduction 6.2
Background 6.2-1
Basic Sample Preparation 6.2-2
Reagent Preparation 6.2-3
Post-Column Conditions 6.2-4
Procedure 6.2-5
Sample Chromatograms and Gradient Programs 6.2-6
Precautions 6.2-18
Carbamates 6.3
Introduction 6.3
Background 6.3
Basic Sample Preparation 6.3-2
Reagent Preparation 6.3-3
Post-Column Conditions 6.3-5
Analytical Procedure 6.3-5
Sample Chromatograms and Gradient Programs 6.3-5
Precautions 6.3-12
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 6
IV
Glyphosate 6.4
Introduction 6.4
Background 6.4-1
Basic Sample Preparation 6.4-1
Reagent Preparation 6.4-2
Analytical and Post-Column Conditions 6.4-4
Analytical Procedure 6.4-4
Sample Chromatograms 6.4-5
Precautions 6.4-6
7.1 Section 7 Troubleshooting
Contact Pickering Laboratories For Support 7.1
Instrument Parameter Log 7.2
General Troubleshooting Advice 7.2
Common System Problems 7.3
Common Chromatography Problems 7.4
Common Column Problems 7.6
Application Specific Troubleshooting 7.7
• Amino Acids
• Carbamates
• Glyphosate
Amino Acid 7.7
Software Troubleshooting 7.8
Procedures 7.11
• To Remove Silica Deposits From Reactor
• To Remove Mineral Deposits In The Reactor From Hard Water
• To Remove Grease Deposits
• If Reagent Backflows Onto Column
• If TRIONE Backflows Onto Column
• If NaOH Is On Column
• To Remove Iron Contamination From Column
• To Pump RESTORE Through The Glyphosate Column
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 7
8.1 Appendices
Installation/Operational Qualitication of the Pinnacle PCX 8.2
Installation/Operational Qualitication of the Pinnacle PCX Data Sheet 8.7
Performance Qualifications of the Pinnacle PCX Data Sheet 8.8
Sample Instrument Parameter Log 8.15
Flow Diagram – Simplex 8.16
Flow Diagram – Duplex 8.17
Parts List for Pinnacle PCX 8.18
Recommended Consumables and Spare Parts 8.21
Limited Warranty 8.24
References 8.25
V
9.1 Index
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 8
VI
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 9
Ge tt in G St ar te d
GettinG started
How to Use this Manual
1 2 Read this First! 2 Symbols and Warnings 2 Specifications 4 Site Requirements
How to Use this Manual
The Pinnacle PCX manual is designed to contain all of the information necessary for the installation, operation, maintenance and troubleshooting of the Pinnacle PCX. It is designed to contain any and all information that you may require during the lifetime of your Pinnacle PCX.
1
In addition to the general operation information, this manual also contains sections devoted to a particular application. Since Pickering Laboratories provides the complete solution, we included the chromatograms, operating conditions, and some troubleshooting information for our most commonly supported methods.
As well as instructional information, this manual also contains a section for recording service information such as serial numbers, installation dates, service dates etc.
Each section of the manual is divided by a tab with the title of that section. For easy reference, simply select the tab that you require based on the information you are looking for. For example, if you are running Amino Acids, and have a question about the gradient program, simply go to the section titled “Amino Acids”.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 10
2
!
Ge tt in G St ar te d
Read this First!
Before attempting to install the Pinnacle PCX post-column derivatization instrument, it is vitally important that you read this manual first, and attend to site, HPLC, and accessories requirements: HPLC – Page 5 Gas Supply Requirements – Page 5 Reagent Reservoir bottles – Page 5 Computer – Page 5
Symbols and Warnings
Caution – this symbol indicates that caution must be used when dealing with this part.
Hot – this symbol is located on the Heated reactor, which can reach scalding temperatures.
Specifications
INSTRUMENT
Dimentions
21.50 H x 10.63 W x 18.25 D inches (54.0 x 26.7 x 46.4 cm), instrument only, doors closed
Weight
77lbs for Dual-pump systems 67lbs for Single-pump systems
Reagent Pumps
Max operating pressure 500 psi (35bar) Flow rate range 50μl –1500μl/minute Refill cycle 60 seconds
Heated Reactor
5°C above ambient to 130°C Thermal Safety switch limits temperature to 150°C Stability +/- 0.5°C Accuracy +/- 1°C
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 11
Ge tt in G St ar te d
Electrical
720 W 120 VAC +/- 10%, 240 VAC +/- 10% 5 A maximum at 108 VAC 47 – 63 Hz Installation over voltage category: Pinnacle PCX complies with Class B Emission Test Specifications.
Fuses
2 ea, 5mm x 20 mm, 6 A, time lag
LCD
Backlit, positive mode, high contrast, viewing area – 125mm (L) x 75mm (W)
3
COMMUNICATION
USB
Requires a USB port.
Relay
Any machine that drives this relay input shall provide a relay contact pair that is electrically isolated from all other electrical devices. The relay contacts must be capable of switching 1mA at 24 +/- 2 Vdc.
Ethernet (optional)Connector: RJ45
Connection Speed: 10/100-BASE-T Ethernet Protocol: TCP/IP IP Address: User set or DHCP (DHCP requires the Network Administrator to “Reserve” an IP address with the DHCP server)
Note: Recommend connection to a Network Switch or Router, or direct connection to a dedicated Network card in the PC.
ENVIRONMENTAL
Indoor use only Altitude up to 6500 ft (1981 m) Ambient Temperature – 40°C Relative Humidity up to 80% at 31°C
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 12
4
Ge tt in G St ar te d
Radio Frequency Pinnacle PCX complies with IEC 61000-4-3
This device complies with Part 15 CHECK Class B of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
WETTED MATERIALS
• PEEK, PTFE, Air Barrier Tubing
• 99.9% Ceramic
• Perlast
• Borosilicate glass
• EPR
®
SOFTWARE
PCX Control Software runs on Microsoft Windows XP, Vista, or Windows 7. 10Mb hard disk space
PCX History Log file: Maximum file Size on the computer: 1.2 Mb
Site Requirements
INSTRUMENT
Bench Space
33 H x 17 W x 21 D inches (84 x 56 x 59 cm), both doors fully opened, with bottles and electrical connections in place.
Minimum 3 inches clearance at back of instrument for venting.
Note: Space quoted above is for Pinnacle PCX.
The total space requirement depends on the brand and model of HPLC.
Electrical Outlet
One grounded outlet must be provided for the Pinnacle PCX.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 13
Ge tt in G St ar te d
Computer
IBM-compatible PCX Control Software runs on Microsoft Windows XP, Vista, or Windows 7. Ethernet or USB port Available Memory: Minimum 10Mb
Ethernet (optional)Connector: RJ45
Connection Speed: 10/100-BASE-T Ethernet Protocol: TCP/IP IP Address: User set or DHCP (DHCP requires the Network Administrator to “Reserve” an IP address with the DHCP server)
5
Note: Recommend connection to a Network Switch or Router, or direct connection to a dedicated Network
card in the PC.
Relay
For synchronization, the HPLC system must be capable of sending a relay signal to an external instrument. No relay connection is needed for Agilent 1100 or Agilent 1200. Pinnacle PCX software directly synchronizes with Chemstation version 9.0 or higher.
Gas Supply
High purity Nitrogen, 45-75 psi (min - max) Outlet of regulator must connect to 1/8” OD tubing
Reagent Reservoir Bottles
The Pinnacle PCX includes one pressurized reagent reservoir for the one reagent system and two for the two reagent system.
Note: For your safety, the bottles are coated with a tough plastic film and rated to a maximum of 15 psig (1 bar). Do not use uncoated bottles.
HPLC Pump
Binary gradient for glyphosate, carbamate applications. Quaternary for all others.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 14
6
Ge tt in G St ar te d
Autosampler
Minimum injection volume 10μl, preferably by full-loop injection For drinking water, minimum injection volume 200μl
Tefzel rotor seal required for all applications using eluants with pH>10 PEEK needle seat required for all applications using eluants with pH>10
Detector
Pressure rating of flow cell must be > 110 psi Inlet capillary must be > 0.17 mm ID
MISCELLANEOUS SUPPLIES
For Amino Acid analysis, a Dead-Head kit is required. This can be purchased from Pickering Laboratories.
Chemistry
The user must check the chemistry requirements for the specific application.
For Carbamate Analysis
HPLC Grade Methanol HPLC Grade Water Materials for calibration standards Carbamate hydrolysis reagent (Cat. No. CB910) Carbamate OPA diluent (Cat. No. CB130 or CB130.2) o-phthaladehyde (Cat. No. O120)
Thiofluor
For Glyphosate Analysis
(Cat. No. 3700-2000)
5% Sodium hypochlorite solution Materials for calibration standards Methanol for OPA reagent preparation Glyphosate Eluant, pH2.0 (Cat. No. K200) Glyphosate Regenerate (Cat. No. RG019) Glyphosate Hypochlorite diluent (Cat. No. GA116) Glyphosate OPA diluent (Cat. No. GA104) o-phthaladehyde (Cat. No. 0120)
Thiofluor
(Cat. No. 3700-2000)
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 15
For Amino Acid Analysis
A: Fluorescence detection
• Methanol for OPA reagent preparation
• 5% Sodium hypochlorite if using the 2-reagent method
• Brij 35 solution for OPA reagent preparation
• DI Water
• Pickering sodium or lithium elution buffers (see application section for cat. nos.) B: UV – Visible detection
• DI Water
®
• TRIONE
Ninhydrin reagent (cat. no. T1OOC or T200)
• Pickering sodium or lithium elution buffers (see application section for cat. nos.)
7
Ge tt in G St ar te d
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 16
8
Ge tt in G St ar te d
Notes
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 17
Section 1
intrOduCtiOn
1.1
Section 1
In tr od uc tI on
1.1
1.1
1.2
1.4
1.6
What is Post-column derivatization? Requirements for a Successful Post-column Method Design of a HPLC system Designing a Post-column system Design of the Pinnacle PCX
What is Post-column Derivatization?
This is a method which renders visible certain compounds that are normally invisible. Since this reaction occurs after, or post-separation, it is referred to as post-column deriviatization. The analytes of interest are separated on the column first, and then reacted with a chemical that will render them detectable at a desirable wavelength, voltage, or any number of various means of detection.
Post column derivatization enhances the sensitivity of HPLC by several means:
1) Most reagents are selective for a particular class of substances, so analytes of that class are more easily seen against a complex background.
2) Since the separation is performed first, the matrix of the sample is either washed off of the column before the analytes, or is retained by the column. This leaves a very pure sample of analyte to react. This eliminates the need for extensive sample clean-up, and provides a very reproducible reaction because there are no matrix interferences.
The Pinnacle PCX post-column derivatization instrument automatically mixes the stream of effluent flowing from the HPLC column with a stream of reagent solution. The mixture flows through a reactor to allow enough time for the chemical reactions to complete. In many cases, the reaction is very slow at room temperature. For this reason, the reactor can be heated. There are some methods that require two or more reagents added in sequence. This is done by the addition of a second reagent pump. In many cases, the second reaction occurs at a much faster rate, and can be efficiently accomplished at room temperature. After the reaction is complete, the derivatives flow into the detector, where the absorbance or the fluorescence (usually) is measured by the HPLC system. These two means of detection are the most common, but they are certainly not the only means of detection.
Requirements for a Successful Post-column Method
There are many things to take into consideration when developing a method and instrument for post­column derivatization. For example, many pumps have a periodic motion when drawing and dispensing that
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 18
1.2
Section 1
In tr od uc tI on
will manifest itself in the baseline of a chromatogram unless it is properly dampened. Below are the basic requirements for a successful automated post-column method:
1)
Reagent Stability. The minimum reagent stability sufficient for routine work is one day. This means that
the yield and signal-to-noise ratio for a given sample must remain constant for at least 8 hours.
2)
Reaction Speed. The analytical separation is complete when the reagent is mixed with the column effluent.
Therefore it is important that the analyte react as quickly as possible. The longer the reaction time, the larger the reactor volume required. With larger volumes, the peak shape will become distorted. To minimize band spreading, it is important to keep the overall time (and therefore volume) as low as possible between the column and detector. If the reaction is slow (in excess of one minute), an elevated temperature can be used to decrease the reaction time.
Reproducibility. Because the reaction is occurring “on the fly,” as the combined column and reagent
3) stream flows toward the detector, the reproducibility is linked to the flow rate precision of the pumps and to the temperature. Accordingly, even an incomplete reaction will be as repeatable as the retention time for any given species. Therefore, it is important that the pumps maintain a constant flow rate, and that the reactor maintain a constant temperature. It is also very important that the column be maintained at constant temperature to ensure that the analytes are properly separated and identified.
Minimal Detector Response of Reagents. The color or background fluorescence of the reagent (or
4) its by-products) represents a continuous noise source. Because the reagent is present in excess relative to the analyte, the analyte’s signal could be obliterated by the reagent’s strong background signal. The baseline noise is proportional to the background signal.
Solubility. All species must remain in solution, including the combined components of the eluants and the
5) reagent(s), as well as the newly formed derivative(s). Precipitates can block capillary tubes, burst reactors, and foul detector flow cells.
Uniformity of flow. The baseline noise is a function of the flow-noise in the eluant and reagent pumps.
6) Non-uniform flow causes non-uniform mixing leading to modulation of the background signal which appears as noise. Refractive index noise can be even more objectionable than absorbance noise. Common techniques for evening the flow of the pumps is the addition of a pulse dampener, or the use of a syringe pump.
Design of a HPLC system
This next section is a simplified view of a HPLC, followed by the ideas behind a post-column system. This section is intended to help novice HPLC operators.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 19
Section 1
In tr od uc tI on
In order to understand post-column HPLC, we need to understand the design of an HPLC. If we connect an HPLC pump directly to a detector (with nothing in between), the baseline from the detector shows a periodic noise (Figure 1-1); the time period is equivalent to the pump stroke.
Injector
Detector
1.3
LC Pump
Eluants
Baseline
Waste
FIGURE 1-1
Now add a commercial pulse dampener. The baseline is still not smooth; the periodic noise is still there although less pronounced (Figure 1- 2). The pulse dampener absorbs most of the pulses from the pump, but the flow requires more stabilization.
Injector
Commmercial Pulse Dampener (capacitor)
LC Pump
Baseline
Eluants
Detector
Waste
FIGURE 1-2
A restriction inline will cause the flow at the outlet of the restriction to be constant. In an HPLC system, this is accomplished with the analytical column. Actually, the column does more than separation; it creates a back­pressure. It is the combination of the pulse dampener and the column that creates a smooth baseline. (Figure 1-3)
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 20
1.4
Eluants
Section 1
In tr od uc tI on
Injector
Pulse Damper
LC Pump
Pre-column Check Valve
Detector
Thermostatted Column & Guard
Baseline
W aste
FIGURE 1-3
An analogy will help us understand the concept. Let us use a river as an example. If it rains; the river swells. If it stops raining; the level goes down. As the level fluctuates, it is equivalent to a periodic noise. To obtain a constant flow, we need to add a reservoir (pulse dampener) and a dam (column). The flow downstream from the dam is constant (smooth baseline).
Designing a Post-column System
The same principals that are used in HPLC can be applied to the post-column system. What happens if we simply add a post-column pump, a mixing tee, and a reactor? The periodic noise returns to the baseline (created by the post-column pump; Figure 1-4).
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 21
Section 1
In tr od uc tI on
1.5
Pulse Damper
LC Pump
Eluants
Injector
Filter
Baseline
Thermostatted Column & Guard
Heated Reactor
Post-column
1
Pump
Reagent
Bottle
Detector
Back-pressure Regulator
Waste
FIGURE 1-4
However we do not need to invent anything new; we just need a pulse dampener and a column. In the generation of Post-column systems prior to the Pinnacle PCX, the restriction performed by the column was achieved using a Restrictor, which is packed with very inert material. With this “flow conditioner” in place, the baseline is now acceptable.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 22
1.6
Section 1
In tr od uc tI on
Design of the Pinnacle PCX
The Pinnacle PCX has taken the post-column system to a new level. The introduction of a syringe pump has eliminated the need for a pulse-dampener and restrictor by:
• The Pickering syringe pumps complete a filling cycle prior to the injection of a run, and deliver reagent during the run at a constant rate.
• There is a valve between the pump and reactor which also helps to regulate the reagent flow, by opening key ports at the appropriate time.
• There is also a pressure transducer which we have added in line before the valve to determine if there are any blockages in the reactors.
Now you are ready for Section 2.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 23
Section 2
Overview
2.2 Introduction
2.3 Check Valve
2.3 Column Connections and Column Oven
2.4 Reagent Pump
2.5 Reagent Valves
2.5 Reservoir Tray / Bottles
2.7 Fluidics Panel
2.8 Quick Change Reactor
2.8 Detector Connections
2.9 Pump Compartment
2.9 Gas Manifold
2.10 Display Module
2.11 Electronics Compartment
2.11 Back Panel
2.13 Safety Features in the Pinnacle PCX
2.14 Standard Configurations of Pinnacle PCX
2.1
Section 2
Ov er vi ew
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 24
2.2
Section 2
Ov er vi ew
Introduction
This chapter is designed to familiarize you with the components, layout, and function of the Pinnacle PCX. Here you will find descriptions of each key component of the instrument and what it does. At the most basic level, the Pinnacle PCX performs three main tasks: 1 .Regulates the temperature of the analytical column
2. Delivers the reagent
3. Heats the reaction
In addition to accomplishing the above three tasks, the Pinnacle PCX has various features to make the analysis more reliable, convenient, and simple. It also contains features to protect the instrument from accidental damage.
R1
FLUSH
FROM
HPLC
COLUMN
FILTER
FILTER
WASTE
PUMP 2
HEATED
REACTOR
PUMP 1
TRANSDUCER 2
2
3
TRANSDUCER 1
2
3
1
VALVE 2
4
AMBIENT REACTOR
1
VALVE 1
4
R2
6
5
6
5
TO DETECTOR
FLUSH
R1
Flow Path – Dual Pump Figure 2-2
2
3
FILTER
VALVE 1
1
6
4
FROM
HPLC
COLUMN
HEATED
REACTOR
TRANSDUCER
PUMP
Flow Path – Single Pump Figure 2-1
Pinnacle Operators Manual
Pickering Laboratories Inc.
The flow path of the Pinnacle PCX is extremely inert, rendering it very versatile. The same instrument can be used for many different applications, and will tolerate a high percentage of reagents.
5
WASTE
TO DETECTOR
Page 25
Section 2
Ov er vi ew
2.3
Check Valve
The column check valve is located just outside of the column oven (figure 2-3). The check valve is designed to protect the column against backflow of the reagents in case of unexpected HPLC shutdown.
Figure 2-3
Column outlet tubing
Column oven door
Rubber foam seal
Column Connections and Column Oven
The column oven contains the analytical column and guard column (figure 2-3). The oven is a convection air oven. The column heater utilizes re-circulating air flow technology to provide quick, uniform column heating. Fast column cooling is assisted by the introduction of fresh air flow into the chamber by means of two fans that are controlled by the Pinnacle PCX control software. The temperature range holds within ± 1˚C resolution from 5˚C above ambient to 75˚C. A temperature gradient can be performed on the column to effect an increase in the speed of the separation. The temperatures can be programmed for a gradient with as many steps as required for fine-tuning an analysis.
The heater block is designed to receive the analytical column (5, 10, 15, and 25 cm long) and the guard column when they are connected via a column easy­connect fitting (included with the instrument, and sold separately by Pickering, Cat. #3102-3064). The outlet end of the column hangs in the slot at the top of the column oven. The inlet to the guard column is at the bottom of the column oven door. The column oven door contains a pocket to fit around the analytical column and guard. There is a groove in the door that is lined with Santoprene foam rubber cord
to Guard
Column
Blue Peek Tubing
PEEK Nut
(3101-0011)
Check Valve
(3106-1007)
that seals the door to the chassis and provides a stable insulated heating environment. The last part of the lead-in capillary is located inside the column oven to preheat the eluant for a more uniform temperature within the column.
Analytical column
Inlet end
Coupler/column connector
Guard column
Check valve
(1452-0226)
from HPLC
PEEK Nut
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 26
2.4
Section 2
Ov er vi ew
Make sure that the column door is securely closed before starting operation.
Warning. The column heating block may become hotter than 75°C. For your safety, wear insulating gloves when the column oven is warm. Attention. La résistance chauffante de la colonne peut dépasser une température de 75°C. Pour votre sécurité, prière de porter des gants isolants lorsque le four de la colonne est chaud. Warnung! Der Heizblock des Säulenofens könnte heißer als 75°C werden. Für Ihre Sicherheit sollten Sie isolierende Handschuhe tragen, wenn der Säulenofen warm ist. Atención. El bloque calefactor de columnas puede estar por encima de 75°C. Para su seguridad use guantes aislantes cuando el horno de columnas esté caliente. Avvertimento. Il blocco della colonna potrá diventare molto caldo e superare ai 75°C. Per la sua protezione usa guanti con insulazione per questa applicazione.
Reagent Pump
The reagent pump is a syringe pump. Refer to figure 2-4 and 2-8 (items 4 and 10). The syringe has a volume of 70 ml. The flow rate can be programmed from 0.05 ml/min to 1.5 ml/min. Once programmed, the syringe pump delivers at a constant speed. Since there is no need for constant refilling and dispensing (as with a reciprocating pump), there is a very high flow precision that does not require external pulse dampening features.
The syringe pump cylinder and head is made from a single piece of 99.9% Alumina for ruggedness and non­reactivity. The piston surface is made from PEEK with an inert o-ring seal.
The pump contains a piston wash for extended seal life. The piston wash is automatically performed with the movement of the piston.
For purposes of fast refill and flushing, the pump can refill or dispense (to waste) at 55 ml/min.
Syringe pump cylider
Pinnacle Operators Manual
Pickering Laboratories Inc.
Figure 2-4
Page 27
Reagent Valves
Reagent Valves are placed immediately after the pump. The valves have 5 possible operating positions:
REAGENT: From Reagent Bottle to Pump (3-4)
1)
REACTOR: From Pump to Reactor (1-2)
2)
WASTE: From Pump to Waste (1-6)
3)
FLUSH: From Flush Bottle to Pump (5-4)
4)
BLOCKED: The valve ports are blocked.
5)
The valves are connected to a Flush Bottle which can be used to flush the reactor in the event of a shutdown or to aid in flushing the system. This will reduce the likelihood that the reactor becomes blocked. The valves are also connected to the reagent bottles and are used for filling the syringe pump.
Pump
2
3
Section 2
Ov er vi ew
1
6
5
4
Figure 2-5
2.5
Valve
Reservoir Tray / Bottles
This is the tray located on top of the instrument. It holds the reagent bottles, wash bottle and flush bottle. It is removable for cleaning. It will hold 1L of liquid and as such is considered secondary containment. It does not seal any openings at the top of the instrument in case of heavy spillage.
There are 4 bottles located here:
1. Reagent Reservoir 1
2. Reagent Reservoir 2
3. Flush Bottle – contains water or 80-20 water/alcohol (IPA or Methanol) for flushing the instrument.
4. Piston Wash bottle – contains 90-10 water/alcohol (IPA or Methanol). This bottle is connected via 1/4 inch OD flexible tubing to the piston wash of each pump in the instrument. Change Wash and Flush solutions at least once a week to prevent contamination.
The pressurized reagent reservoir serves two purposes: It protects air-sensitive reagents from oxidation. It helps the syringe pump fill consistently and quickly by providing a source of pressure.
Flush
Reagent
Pinnacle Operators Manual
Pickering Laboratories Inc.
Piston Wash
Figure 2-6
Page 28
2.6
Section 2
Ov er vi ew
Reagent tubing is pre-connected to the pump at the factory. All Reagents lines are Air Barrier tubing. Reagent 1 is labeled R1, and Reagent 2 is labeled R2 throughout the instrument. Nuts and ferrules (1/4-28 ) are provided for connecting the tubing to the Reservoir caps.
The reservoir cap (Figure 2-7A) has two connecting ports on the top – one for connecting reagent line and one for connecting the gas line. The port for the gas line has a one-way valve to prevent backflow of the reagent into the gas manifold in case of a drop in gas pressure. Use nuts and ferrules (1/4-28) to make the connections. There are two stopcocks, one is for the vent port and the other is for the reagent line. Keep reagent line stopcock open for operation. Open the vent stopcock to sparge the reagent and close it to pressurize the reservoir. This is the normal operating position.
When changing reagent, first turn off the gas using the toggle valve on the fluidics panel. Then vent the reagent bottle by opening vent stopcock. Now you can safely remove the cap. It is convenient to have extra bottles so that you can simply transfer the cap without setting it down and risking contamination.
Figure 2-7A
If you have a reservoir cap shown on figure 2-7B, the large white knob is the valve; pull it up for CLOSED, and push it down for OPEN. If the gas is turned on, opening the vent valve will sparge the reagent. Closing the valve will pressurize the reservoir; this is the normal operating position. On the side of the cap, away from the on-off valve, there is a 1/4-28 fitting; you may optionally connect a tube here to carry vapors to an exhaust vent.
Warning. For your safety, the bottles are coated with a tough plastic film and are rated to a maximum of 15 psig (1 bar). Do not use uncoated bottles. Attention. Pour votre sécurité, les bouteilles sont recouvertes d’un film de plastique dur, et sont calibrées à un maximum de 10 psig (0.7 bar). Ne pas utiliser les bouteilles non recouvertes. Warnung! Für Ihre Sicherheit wurden die Reagenzienflaschen mit einem festen Schutzüberzug aus Kunststoff versehen. Die Flaschen sind bis max. 0.7 bar (10 psig) zugelassen Flaschen mit beschädigtem Schutzüberzug dürfen nicht mehr benutzt werden Verwenden Sie keine Flaschen ohne Schutzüberzug! Atención. Para su seguridad, las botellas están recubiertas con una resistente película plástica, y están constrastadas a 10 psig (0.7 bar). No utilice botellas sin recubrimiento. Avvertimento. Per la sua protezione, le bottiglie sono construite forti con un percentuale du plastica, e sono usabili per un massimo di 0.7 Bar (10 psi). Non usare bottiglie normali.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Valve
Figure 2-7B
Page 29
Section 2
Ov er vi ew
There are three types of Reservoir/Cap assemblies used on the Pinnacle PCX:
1) Reagent Reservoir is 1L, 45-430, safety-coated. These are shipped with blue caps with white valve assembly. These can be pressurized.
2) The wash Reservoir is 1L, 38-430, clear glass. These are shipped with caps with 1ea 1/4” opening and 2ea 1/8” openings. These bottles cannot be pressurized, and do not contain threads for tubing connections.
3) The flush Reservoir is 1L, 38-430, clear glass. These are also shipped with caps, but with 3 each 1/8” openings. These bottles cannot be pressurized, and do not contain threads for tubing connections.
Note: The two types of caps cannot be interchanged because the necks of the bottles are different sizes.
Fluidics Panel
The fluidics panel is the busiest area of the instrument. Everything that you will need is located on the front of the panel. The door can be removed if desired by the user for easy­access. Refer to figure 2-8 for parts identification throughout the next section.
5
3
2.7
Starting at the valves, the parts of the Fluidics Panel are numbered counter-clockwise.
1. Valve 2 (on dual systems only)
2. Pressure Transducer 2 (on dual systems only)
3. Gas is controlled by the toggle valve. Lever ON
pressurizes the manifold.
4. Pump 2 (on dual systems only)
5. Column Outlet
6. Mixing Manifold 2 (with integrated reagent
filter) (on dual systems only)
7. Heated Reactor
8. Mixing Manifold 1 (with integrated reagent
filter and over-pressure relief valve)
9. Over pressure Relief valve
10. Pump 1
11. Pressure Transducer 1
12. Valve 1
13. Ambient reactor (on dual systems only)
14. Reactor outlet and union to connect to
detector
2
4
1
6
7
8
9
10
13
11
12
14
Figure 2-8
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 30
2.8
Section 2
Ov er vi ew
Quick Change Reactor
The standard reactor is a PTFE capillary tube 0.011” I.D. wrapped on a heated mandrel. The narrow diameter reduces band-spreading, and the PTFE is corrosion resistant. The reactor can be easily switched if it becomes blocked or if a different volume is required. The electronics of the reactor are contained within the Pinnacle PCX, making replacement affordable and easy.
Figure 2-9
Volumes ranging from 0.15 ml to 3 ml are available. The most commonly used volumes are available off the shelf, but if you require a volume that is not in our price list, we can make it for you.
There is a 500 psi relief valve in case there is a blockage in the reactor or detector (Item 9 on Fluids Panel).
Detector Connections
from Detector Exit
There is a 100 psi (5 bar) back-pressure regulator (cat. # 3102-9025) on the exit line from the detector; it suppresses boiling inside the hot reactor and prevents bubbles from forming in the detector flow cell. This in turn provides for a smooth baseline and therefore high-sensitivity.
Flow direction
to Waste
Pinnacle Operators Manual
Pickering Laboratories Inc.
Figure 2-10
Page 31
Section 2
Ov er vi ew
2.9
Pump Compartment
This compartment is located on the right side of the Pinnacle PCX. It contains the reagent pump(s), gas manifold, and electronics for the heated reactor and valves(figure 2-11).
The only time you will need to access this panel is for pump piston seal replacement. All other times, qualified personnel should make any repairs.
The Pump, Valves, and Heated Reactor are described in other sections.
Here you will see the Piston Wash feature for the reagent pumps. The Piston wash is a 1/4” OD tubing connected to the back chamber of the syringe.
Valve Motor
Heated Reactor
Pump 2
Pump 1
Gas Manifold
Piston Wash
Figure 2-11
Gas Manifold
The “Gas Inlet” fitting is a 1/4-28 fitting located on the back panel of the Pinnacle PCX (reference fig 2-15). The inert gas source is connected via a 1/8” OD Air Barrier tubing, and the pressure is regulated to 5psi before pressurizing the reagent reservoirs.
The gas regulator requires an input pressure of 45–75 psi (3–5 bar) to function properly. The manifold has a safety relief valve that opens at about 10 psi to prevent dangerous over­pressurizing of the reagent reservoirs.
Figure 2-12
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 32
2.10
Section 2
Ov er vi ew
Display Module
The display module (figure 2-13) is the information panel of the Pinnacle PCX. It displays: the flow rate of each pump, the set point and current temperatures of the column and reactor the Pump pressure(s), which are also indicators of the pressure inside the post-column system and status information.
Figure 2-13
In addition to the electronic information on the LCD, the display module contains 3 buttons and one LED. STOP - this button is used for an instant stoppage of the instrument in the case of an emergency. If this button is pushed, the instrument will cease all activity, regardless of what is happening.
Stop LED
Stop
Mode
Select
The STOP LED will illuminate when the Pinnacle PCX has been stopped manually by the user. When STOP is pressed, the LED will glow red, indicating that there has been an emergency manual stoppage. To reset the instrument after emergency stop, turn instrument OFF and then turn it back ON.
MODE - Used during initial installation to determine the IP address. This function is for instruments that are connected to a company network. SELECT - Used during initial installation to determine the IP address. This function is for instruments that are connected to a company network.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 33
Electronics Compartment
This compartment contains the central nervous system of the Pinnacle PCX – the main PC board. From here, all communication is coordinated.
This compartment is separate from the pump compartment for two reasons:
1) To segregate the liquid end form the electronic end to avoid any damage to the PC board by reagents and liquids
2) To keep the board cool from the warm action chamber of the pumps
2.11
Section 2
Ov er vi ew
This compartment also contains: Power supply Cooling fans Column oven cooling mechanism
Figure 2-14
Since this compartment is very sensitive to liquids, and shocks, it is strongly recommended that no repairs are attempted by the user.
If there is a need for the user to open this panel, please do so only under the recommendation/ instruction of qualified Pickering support personnel.
Back Panel
This panel contains: Communication Ports Gas Inlet
Gas Inlet
USB Port
Ethernet Port
Relay Input
Power Fuses Vents and other openings
COMMUNICATION PORTS
USB Port
Figure 2-15
Power Switch and Main Fuse
The USB port on the Pinnacle PCX connects directly to one of the USB ports on the computer.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 34
2.12
Section 2
Ov er vi ew
Relay Ports
The relays will be explained in the operations chapter under Site Requirements.
Ethernet Port
This port is used for connecting the Pinnacle PCX to a network.
POWER SWITCH AND MAIN FUSE
The power connector is a standard IEC 320 type connector. Use the appropriate power cord for your local wall outlet and electrical code.
The 120V version comes with a standard North American cord set.
The 240V version comes with a cord set used in much of continental Europe.
Your local reseller may have provided the correct local cord set. If your local power outlets are different, you will need to obtain the appropriate grounded cord set. The main power switch is located in the power connector assembly.
The fuse holder is located in the power connector assembly. To change the fuse, first remove the power cord from the connector. Carefully pry out the fuse clip with a small screwdriver. Replace with the specified-type fuse: 2 ea, 5mm x 20 mm, 6 A, time lag.
Warning. Ensure that the power cord is disconnected before replacing a fuse. Use only the specified-type fuse. Attention. Assurez vous que le cable secteur n’est pas connecté avant de changer un fusible. Warnung. Sicherungen dürfen nur bei nicht angeschlossenem Netzkabel ersetzt oder gewechselt werden. Cuidado. Asegúrese que el cable de red está desconectado antes de instalar o cambiar un fusible. Attenzione. Assicuratevi che il cavo di alimentazione sia scollegato prima di installare o sostituire un fusible. Waarschuwing. Zorg dat de voedingskabel losgekoppeld is, voordat een zekering wordt geplaatst of vervangen. Avvertimento. Fare atenzione che la corda del voltaggio sia staccata prima di cambiare valvole. Usa solo
valvole di capacitá precisata dalla fattoria.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 35
Safety Features in the Pinnacle PCX
The Pinnacle PCX system has features designed into the instrument and operation that will prevent reagent back-flow onto the column and bursting of reactor tubing due to a blockage.
Post-column reagent can immediately damage the analytical column if the reagent flow is diverted in the wrong direction by a lack of HPLC flow. This is prevented by:
Inline Check valve
2.13
Section 2
Ov er vi ew
Over-pressure relief valve
1) In-line Check Valve
Figure 2-16
A one-way check valve placed before the column prevents the flow of liquid in the direction of the column.
2) Over-pressure Relief Valve In the rare event there is a blockage in the post-column system, an integrated 500 psi over-pressure relief valve will open and divert the liquid into the drip tray. This will prevent a bursting of tubing or fittings. By relieving the pressure, this will give you the opportunity to correct the blockage rather than having to replace the heated reactor.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 36
2.14
Section 2
Ov er vi ew
Standard Configurations of Pinnacle PCX
The Pinnacle PCX is available in standard configurations for one-reagent or two-reagents, standard or micro volume, 120V or 240V operation, and is shipped completely assembled, calibrated, and tested. Custom instruments are also available; contact Pickering Laboratories for a quote.
LCD
Valve
Ambient Reactor
Transducer
Check Valve
Dual Reagent System Figure 2-17
Pump
Pump
Single Reagent System Figure 2-18
LCD
Transducer
Valve
The one-reagent instrument consists of a single reagent pump, valve, heated reactor, column heater, backflow and over-pressure safety devices, filters, reagent reservoir, gas manifold, Air Barrier gas tubing, and other accessories.
The two-reagent Pinnacle PCX contains two reagent pumps, two valves, heated and ambient reactors, column heater, backflow and over-pressure safety devices, filters , reagent reservoirs, gas manifold, Air Barrier gas tubing, and other accessories.
The Pinnacle PCX includes one pressurized reagent reservoir for the one-reagent system and two for the two­reagent system.
The instrument also includes a 1L wash bottle and a 1L flush bottle for the piston wash and flush feature.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 37
Section 3
installatiOn
3.2 Site Requirements
3.4 Instrument Unpacking and Preparation
3.5 Gas Connections to Back Panel of Pinnacle PCX
3.6 Computer Connections
3.6 Software Installation
3.8 Pump and Autosampler Connections
3.10 Eluant Priming
3.10 Column and Guard Installation
3.11 Detector Connections
3.12 Reservoir Connections
3.13 Reagent Pump Preparation
3.14 Running a Chromatogram
3.15 Shutdown
3.1
Section 3
In st al la tI on
Installation Checklists (see Appendix)
The Pinnacle PCX instrument is shipped in one carton. Application Kits may be shipped in one or more cartons each. Report any carton damage to the carrier. Unpack all cartons and review the contents using the Packing List to ensure that your order is complete. If any items are missing, immediately contact Pickering Laboratories at (650) 694-6700 or by fax at (650) 968-0749.
Store any standards in the freezer or refrigerate immediately upon arrival. Pickering Amino Acid, Carbamate, and Glyphosate columns are shipped with test mixtures. Remove the vials from the box and freeze upon arrival.
While installing the Pinnacle PCX complete the Installation Checklist located in the Appendix. When the installation is completed, fax or mail the checklist back to Pickering Laboratories and place in the front of the Operation Manual under the tab labeled Installation. If the customer requires it, complete the IQ/OQ procedure (also located in Appendix) and add that to the tab labeled Installation.
Read all installation instructions and material safety data sheets (MSDS’s) before operating your post-column derivatization instrument and HPLC system.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 38
3.2
Section 3
In st al la tI on
Note: Before the Pinnacle PCX can be installed and qualified properly, the HPLC must be completely installed and in good working order (including pump, injector, detector and data collection system). The user prior to Pinnacle PCX installation must remove all organic compounds that are immiscible with the Pickering Laboratories’ eluants as well as any hazardous chemicals.
The Pinnacle PCX uses three main styles of fittings.
1. Lite-touch fittings These are 10-32 nut with 2-part ferrules for 1/16" OD tubing. These fittings are found at the column connections, and at the inlet to the reagent filters. (nut: 1452-0118, ferrule: 1452-0117)
2. Long PEEK fittings These are 1/4-28 nuts with 2-part ferrules for 1/8" OD tubing. These fittings are found at the inlet and outlet of the reagent pumps. Low-pressure gas and reagent fittings are 1/4-28 x 1/8 inch size. (nut: 1452-0116, ferrule: 1452-0115)
3. Short PEEK fittings These are 1/4-28 nuts with 2-part ferrules for 1/16" OD tubing or 1/8" OD tubing. These fittings are found at the connections to the electronic valves and transducers. (nut: 1452-0113, ferrule: 1452-0115)
Pickering Laboratories supplies all the matching nuts and ferrules needed for normal installation.
Site Requirements
PINNACLE PCX SITE REQUIREMENTS
The minimum bench top space required for the Pinnacle PCX system is approximately 32H x 16W x 20D inches (81 x 41 x 51 cm), both doors fully opened, with bottles and electrical connections in place. The Pinnacle PCX weighs approximately 67 lbs (30kg) for simplex systems, and approximately 77 lbs (35kg) for duplex systems. The minimum bench space does not include the HPLC system. The total space requirement depends on the brand and model of HPLC.
For most cases, it is best to place the LC pump and injector system on the left side of the Pinnacle PCX, and the detector on the right.
In addition to the power outlets required for the HPLC system, one grounded outlet will be needed.
Nitrogen is required to pressurize the reagent reservoir(s). The Pinnacle PCX requires gas pressure of 45-75 psi (3-5 bar) at the gas inlet. An adaptor from the gas regulator to 1/8 inch OD tubing is required.
®
To minimize oxidation of the TRIONE gas supply line (Air Barrier or metal).
ninhydrin or OPA reagent, use oxygen-impermeable tubing for the entire
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 39
Section 3
In st al la tI on
Note: If TRIONE® is to be used for Reagent 1, Nitrogen must be used to prevent out-gassing.
A waste container should be provided for the waste lines from the Pinnacle PCX and the HPLC detector.
HPLC SYSTEM REQUIREMENTS
Since every HPLC is different, the following procedure has been generalized. Before attempting to connect any tubing, examine the HPLC setup, and determine the best possible means of making the connections. Small ID tubing (0.011") should be used wherever the sample is in the flow path.
Important! If the system will be used for amino acids, glufosinate, glyphosate, polyamines, or diquat & paraquat analysis, be aware that the column regenerant is strongly alkaline. Any polymers or other materials in the HPLC pump, injector, needle seat, and detector must be compatible. For example, the standard rotor seal in Rheodyne injector valves is Vespel® polyimide, which is not recommended at pH >9; a Tefzel® or PEEK rotor seal must be installed.
3.3
For all applications, the pressure rating of the detector flow cell must be > 110 psi (7.5 bar)
FOR AMINO ACID ANALYSIS
Pump
Minimum ternary gradient elution Piston wash capability is preferable
Injector
Tefzel or PEEK rotor seal for injector valve Tefzel or PEEK needle seat if it is an autosampler
FOR GLYPHOSATE ANALYSIS
Pump
Minimum binary gradient elution Piston wash capability is preferable
Injector
Tefzel or PEEK rotor seal for injector valve Tefzel or PEEK needle seat if it is an autosampler For water samples, at least 200 μl injection
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 40
3.4
Section 3
In st al la tI on
FOR CARBAMATE ANALYSIS
Pump
Minimum binary gradient elution
Injector
For water samples, at least 200μl injection
For all other applications, review the method notes for chemistry requirements.
HPLC RELAY REQUIREMENTS
For HPLC systems other than Agilent 1100, or 1200, the Software and system must be capable of sending a relay signal to an external piece of equipment to achieve synchronization.
Chemstation version 9.0 or higher is needed for Agilent 1100 or Agilent 1200. Pinnacle PCX software will communicate with Chemstation directly – no relay connection is needed.
Any machine that drives this relay input shall provide a relay contact pair that is electrically isolated from all other electrical devices. The relay signal must have:
Relay detection voltage 24 +/- 2 V Relay detection current Approximately 1 mA
COMPUTER REQUIREMENTS
For the Installation, we strongly recommend installing the USB cable. Use of the network connection is optional at the user’s discretion. The computer must have: Microsoft Windows XP, Vista, and Windows 7. Minimum of one extra USB port for installation Available Memory: Minimum 10Mb
Instrument Unpacking and Preparation
UNPACKING
Unpack all cartons and review the contents using the Packing List to ensure that all of the items are present. If any items are missing, immediately contact Pickering Laboratories, Inc.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 41
Section 3
In st al la tI on
Toll Free: (800) 654-3330 International: (650) 694-6700 Email: support@pickeringlabs.com Internet: www.pickeringlabs.com
Unpack the instrument and place it on the bench. Place it so there is enough clearance between the Pinnacle PCX and the HPLC, detector, and the edge of the bench.
Ensure that there is enough room to open the column oven door and the fluidics door.
Ensure that there is at least 3 inches clearance between the outlet vent at the back of the Pinnacle PCX and any walls or other instruments. This is very important for proper cooling of the column oven.
3.5
PREPARATION AND INSPECTION
Place the Pinnacle PCX on the bench. Examine the external chassis for damage. Open the column oven door. Check that it is straight. Check that the latch is tight and undamaged. Open the Fluidics chamber door. Check that it is straight. Examine the connections on the face of the fluidics panel. Check that there are no broken, bent, or loose tubings. Remove the right side panel. Check that the pump/s is/are firmly attached to the chassis. Uncoil the 1/4" C-flex tubing. This tube is for the piston wash. Place the tubing on top of the instrument in the Reservoir tray.
Check that the piston wash tubing is firmly connected to the Y fitting, see figure 2-11.
Replace the right side panel.
Connect the power cord to the outlet in the back of the Pinnacle PCX, just above the power switch.
Gas Connections to Back Panel of Pinnacle PCX
Connect the Inlet Gas line to the back of the Pinnacle PCX. [Set the regulator on the inert gas supply to between 45-75 psi] Using either 1/8" OD Air Barrier tubing, or 1/8" copper tubing, connect the gas supply to the Gas Inlet fitting on the back panel of the Pinnacle PCX with the 1/4-28-1/8" nuts and ferrules (PN 1452-0181 and 1452-0180).
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 42
3.6
Section 3
In st al la tI on
Do not connect the 1/8" Air Barrier tubings to the bottles yet. Lay them carefully in the Reservoir tray. The gas connections will be completed later in the installation.
Carefully turn on the main gas supply. Switch the toggle valve to the ON position to start gas flow. Let the gas system purge for about one minute. Check for leaks at the Gas Inlet connection, and check that there is gas flowing out of the 1/8" tubes. Switch the toggle valve OFF.
Slide the Pinnacle PCX into place on the bench. Allow a minimum of 3 inches of space for venting at the back of the instrument.
Computer Connections
RELAY CONNECTIONS
Relay Input Connector can be used to trigger Pinnacle PCX operations.
Connect the relay cable to the Relay Input Connector on the back of the Pinnacle PCX. The opposite end of the cable will be connected to the HPLC system’s location for relay output. Refer to the HPLC operation manual for further information.
POWER-UP
Turn on the power switch, but do not operate the buttons.
Upon power-up, the valves will find their home positions. Ensure that this happens. Check that the LCD displays the status of the instrument.
Software Installation
Insert the CD into the computer and follow the on-screen instructions. The Installation wizard will guide you through the installation and configuration process.
MULTI INSTRUMENT SUPPORT
Select the number of Pinnacle PCXs that will be connected to the same PC. For each Pinnacle an icon will be created on the desktop. Configure each instrument separately.
COMMUNICATION METHOD
Select Network or USB cable. These settings can be modified later in the Configuration window of the Pinnacle PCX software.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 43
Section 3
In st al la tI on
SECURITY LEVEL
Select who can access the Pinnacle PCX program. “Everyone” will allow access to many users while “Just me” will only one user.
REGISTER
Complete the registration form and e-mail directly to support @pickeringlabs.com or print and fax to (650) 968 – 0749.
OPTIONAL
Connecting the Pinnacle PCX to the network
It is at the user’s discretion if they would like their Pinnacle PCX placed on their company network. This will involve working closely with the Company’s IT personnel, to determine proper address and their network security procedures.
3.7
A network cable is provided with the Pinnacle PCX
Connector RJ45 Connection Speed 10/100-Base-T Ethernet protocol TCP/IP IP address User set or DHCP
DHCP IP address is assigned by a DHCP server. The network administrator must configure the DHCP server so that the same address is reserved for and always given to Pinnacle PCX. User set or Static IP address is set in the Pinnacle PCX. The network administrator must provide an address specifically for Pinnacle PCX.
Note: We recommend connection to a Network Switch or Router, or direct connection to a dedicated Network card in the PC.
Setting up or Changing User set (Static) IP address Follow the chart below:
Note: Use the MODE button to go through options and the SELECT button to select the option. Messages appear at the bottom of the main LCD screen.
Turn Pinnacle PCX ON. While instrument is initializing press the STOP button and hold till the message reads: BOOT: 1-04 CPLD: 2.5 DHCP IP: 192.168.111.111
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 44
3.8
Section 3
In st al la tI on
Press MODE button. The message will read: RUN MAIN F/W DHCP IP: 192.168.111.111
Follow the chart below to make changes in IP address.
Main LCD Display
DHCP IP: 152.168.111.111
Select
RUN MAIN F/W
DHCP IP: 192.168.111.111
Mode
NETWORK
IP FROM DHCP
Mode
DHCP IP: 192.168.111.111
MAC: 00:50:C2:44:30:45
Mode
PRESSURE UNITS
PSI
Restart the instrument after completing the steps above.
Pump and Autosampler Connections
Note: Do not fit the analytical column and guard yet.
Mode
Select
Select
NETWORK
STATIC IP
Mode
STATIC IP:192.168.111.111
MAC:00:50:C2:44:30:45
Select
EDIT IP: 192.168.111.111 Select changes number
Mode moves next to position
PRESSURE UNITS
BAR
Replace any mobile phase that is more than 2 weeks old, especially DI water. Rinse the reservoir bottles thoroughly with soap and water. Wipe down the dip tubes on the caps with methanol and a clean, lint-free cellulose tissue. Avoid touching the tubing or the interior of the reservoir with your skin and do not leave caps and lines dangling without a reservoir because this can cause contamination.
Fill reservoirs with 80/20 Water/Methanol
Open the prime-purge valve on the HPLC. Purge each line in the system with a water/methanol mixture to flush the system. Set the flow rate to the maximum and purge at least 25 ml through each line.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 45
outlet check valve
from solvent switching valve
purge valve
3.9
Section 3
In st al la tI on
pulse dampener
Te e
Plug
inlet check valve
To autosampler
Figure 3-1
Drop the flow rate to 1 ml/min and close the prime-purge valve to flush the lines to the injector.
If this is an amino acid system, it is highly recommended that you dead-head the pulse dampener of the pump. This will provide reliable gradient formation, and will prevent corrosion of the low-grade steel used in most pulse dampeners.
Use one of the two methods below for making the connections. Amino Acid systems must be connected according to method 2.
1.Connect the outlet of the injector to the inlet of the pre-column check valve assembly. OR
2.Connect a tee to the inlet of the pulse dampener. Connect a line from the outlet of the pulse dampener to a tee and place a high-pressure plug in the outlet. Connect the inlet and outlet of the pump through the tee (figure 3-1).
Connect the outlet of the injector to the Check Valve assembly. Use 0.011" ID tubing and a PEEK nut and ferrule. (PNs 1452-0118, 1452-0117, 2104-0210, 1452-0226)
Place the open end of the check valve tubing into a beaker.
Flush line from injector for 5 minutes at 1 ml/min with 80/20 Water/Methanol.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 46
3.10
Section 3
In st al la tI on
Eluant Priming
Before proceeding, check for and repair any leaks between the pump and the pre-column check valve. Once you are certain there are no leaks, do not open the connections between the pump and injector.
IMPORTANT! If any application other than Carbamates is to be used, remove water/methanol and replace with water. Flush at least 10 ml through lines and flush injector line. This is to prevent any precipitation issues between Methanol and the buffers and to prevent any organic solvents from entering the column. Pickering Laboratories cation exchange columns for Amino Acids and Glyphosate analysis will be damaged by organic solvents.
Figure 3-2
Column outlet tubing
Column oven door
Rubber foam seal
Fill the reservoirs with the appropriate eluants/mobile phases and again flush at least 25 ml OF EACH ELUANT with the purge valve in the open position.
Close the purge valve, and pump each eluant at 2 ml/ min for 5 minutes.
If buffers will be used as the eluant, use pH paper to measure the pH of the solution coming through the tubing.
Column and Guard Installation
Set the HPLC to 100% of the storage solution in the column. Install the pre-column check valve and female nut inline before the column over door. Reference the Flow Diagram in section 8.5 for complete parts list of the pre-column check valve. Start the flow rate to 0.2 ml/min. Connect the line from the outlet of the pre-column check valve to the inlet of the guard column. When liquid exits the guard column, connect the “column connector/coupler” provided in the packing kit. When liquid exits the connector, connect the
to Guard
Column
Blue Peek Tubing
PEEK Nut
(3101-0011)
Check Valve
(3106-1007)
Analytical column
Inlet end
Coupler/column connector
Guard column
Check valve
from HPLC
PEEK Nut
(1452-0226)
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 47
Section 3
In st al la tI on
analytical column. Connect the Column Outlet tubing (pre-formed and cut to the right length at Pickering). Hang the column in the column oven, use care to ensure the tubing passes through the guide hooks at the top of the column oven. The guard column should be at the bottom, and the exit of the column at the top.
NOTE: If the application is one that uses sodium or lithium buffers, use deionized water to clean any stainless steel fittings that have come into contact with buffer to prevent corrosion.
Monitor the pressures and stop the HPLC pump when the pressure stabilizes.
Detector Connections
Connect a 1/16 inch x 0.020" ID tubing from the outlet of the Detector to the external 100psi back-pressure regulator (PN 3102-9025) using a 1/4-28 nut with a 1/16 inch reversed-ferrule. There is an arrow on the back pressure regulator indicating direction of flow. Insure that the arrow is pointing away from the detector and toward the waste line.
3.11
from Detector Exit
Connect the 0.020" ID PTFE tubing provided in the packing kit (PN 2101-
0225) to the outlet of the external 100 psi back-pressure regulator. Place the other end in an appropriately labeled waste container.
Connect the PEEK union to the outlet of the Ambient reactor (or Heated reactor if it is a single-pump system) using a Lite-touch Nut and Ferrule.
Connect a 0.011" ID tubing from the outlet of the union at the exit of the
Figure 3-3
fluidics panel on the right-hand side of the instrument to the inlet of the detector flow cell.
Set the time constant on the detector to 2–4 seconds.
The pressure rating of the detector flowcell must be >110 psi (7.5 bar). If your detector flowcell is rated lower, consult Pickering Laboratories.
Special Note to Hewlett-Packard 1046A end-users: Replace the 0.12 mm ID inlet tubing (red) and heat­exchanger from the left side of the detector to the flowcell (behind the front panel of the detector) with a 0.25 mm ID (blue) tubing (HP Cat. No. 79881-67302) to reduce the back-pressure.
Caution! The 100psi back-pressure regulator is directional. Do not reverse flow!
Flow direction
to Waste
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 48
3.12
Section 3
In st al la tI on
Reservoir Connections
Wash all of the Pickering Pinnacle PCX reservoirs with laboratory detergent and hot water.
Rinse with methanol then with deionized water.
Wipe down the dip tubes on the caps with methanol and a clean, lint-free cellulose tissue. Avoid touching the tubing or the interior of the reservoir with your skin and do not leave caps and lines dangling without a reservoir because this can cause contamination.
There are three types of Reservoir/Cap assemblies used on the Pinnacle PCX: Reagent Reservoirs are 1L, 45-430, safety-coated. These are shipped with blue caps with white valve assembly. These can be pressurized. Wash Reservoir are 1L, 38-430, clear glass. These are shipped with blue caps with 1ea 1/4" opening and 2ea 1/8" openings. These bottles cannot be pressurized, and do not contain threads for tubing connections. Flush Reservoir, is 1L, 38-430, clear glass, from Wheaton. These are also shipped with caps, but with 3ea 1/8" openings. These bottles cannot be pressurized, and do not contain threads for tubing connections.
The two types of caps cannot be interchanged because the necks of the bottles are different sizes.
REAGENT RESERVOIR CONNECTIONS
Connect 1/8" Air Barrier lines from “Gas Supply OUT” port on the back of the Pinnacle PCX to the CHECK VALVE on the top of the reagent cap. Use a 1/4-28 Nut and ferrule (PN 3101-0005 and 3101-0006). Don’t over tighten!
Connect 1/8" Air Barrier lines from the Reagent ports of the valves (No 3) ports at the front of the Pinnacle PCX to the empty port on the top of the reagent cap. Use a 1/4-28 Nut and ferrule (PN 3101-0005 and 3101-0006). Don’t over tighten!
Put the cap on the bottle.
Toggle the Gas switch to ON and check that gas is flowing through the reagent cap. Close the stopcock under the vent hole to pressurize the bottle. Open the stopcock under the reagent line before refilling the pump.
Caution! The reagent bottles are specially coated with a protective polymer to ensure operator safety if the reservoirs should become over-pressurized. Non-coated bottles must not be substituted in the Pinnacle PCX system. Replacement 1 L, 2 L, or 5 L reagent bottles may be ordered directly from Pickering Laboratories.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 49
Section 3
In st al la tI on
Reservoirs should be labeled with an appropriate label using the GLP of the laboratory. Fill the Flush reservoir with water or 80-20 water/alcohol (IPA or Methanol) Fill the Wash reservoir with 90-10 water/alcohol (IPA or Methanol). Change Wash and Flush solutions at least once a week to prevent contamination.
Rinse the Reagent reservoir(s) with a small amount of reagent or diluent, and then if necessary prepare the post-column reagents as described in the appropriate Application section of this manual.
FLUSH RESERVOIR CONNECTIONS
Take the 1/8" FEP tubes leading from the Flush (No. 5) position on the valves and slide the tubing through the 1/8" openings on the proper cap. This will be a snug fit to prevent slipping.
3.13
(PISTON) WASH RESERVOIR CONNECTIONS
Take the 1/4" C-flex tubing leading from the piston wash of the reagent pump(s) and slide it through the 1/4" opening in the proper cap. This will be a snug fit to prevent slipping.
Reagent Pump Preparation
All of the reservoirs should now be filled with the appropriate solutions. Ensure that the Flush and Piston Wash lines are immersed in liquid.
The Pinnacle PCX has a maximum of two waste lines that must be fed into a waste container. They can either be fed into a separate container, or into the waste stream of the HPLC. These are the 1/8" OD clear FEP tubes that are connected to the Waste port (No. 6) on the valves.
Dual pump systems will have two lines, and single pump systems will have one.
In the Pinnacle PCX software, select Empty pump(s) under the Control menu. Select Both pumps for 2-pump system. The valve(s) will move to waste (1-6) position and pump(s) will dispense all the liquid.
In the Pinnacle PCX software, select Flush pump(s) under the Control menu. Select Both pumps for 2-pump system. The system will flush pumps using Flush solution. The pump(s) will be empty at the end of Flush cycle.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 50
3.14
Section 3
In st al la tI on
Running a Chromatogram
Look in the column box for the gradient program and conditions.
Set the HPLC pump to run the starting eluant conditions, and while the system is equilibrating, set up the gradient method in the HPLC software. We recommend that equilibration is set up at the end of each run (as a post-run, if possible). Equilibration time should be at least 5 min long to give Pinnacle PCX enough time to refill between the runs. If using relays to synchronize with Pinnacle PCX set up relay signal at time 0.0 in the HPLC method. Contact your HPLC support representative if you have questions about setting up relay signal.
Create a sequence for the HPLC system.
Set up method and sequence for Pinnacle PCX (refer to section 4 OPERATION of this manual). Load the sequence. Make sure correct sequence name and method name are displayed in the status bar at the bottom of the Pinnacle software (If sequence and method fail to load: close Pinnacle PCX software, reopen Pinnacle PCX software, reload sequence and method).
In the Pinnacle PCX software, select Enable in the Control menu. Column and reactor heater will be turned ON. Wait for temperatures to reach the set point.
Select Refill pump(s) in the Control menu. Select Both pumps for 2-pump system. The valve(s) will move to reagent (3-4) position and the pump(s) will refill enough reagent for one run.
NOTE: Pinnacle software will calculate needed reagent volume based on your run time and reagent flow rate. Some extra volume will be added to that to ensure pumps will not run out of reagent during operation.
Select Pump(s) ON in the Control menu. Select Both pump(s) for 2-pump system. The valve(s) will move to reactor (1-2) position and pump(s) will start dispensing reagent at set flow rate. Wait until the pressures are stable. Select SequenceStart Sequence. Run time (0.0) will be displayed in the Status Bar at the bottom of the Pinnacle PCX software. The reagent pump(s) will start dispensing reagent. Pinnacle now waits for injection signal from HPLC in order to start running the method.
®
For TRIONE
Amino Acid Analysis:
Set the detector wavelength(s) to 570 and 440nm (if applicable).
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 51
Section 3
In st al la tI on
Do four runs of the Amino Acid Standard that was provided with the chemical kit. Inject 10μl. Discard the first injection. Compare the chromatogram with that of the QC test of the system and column. Verify that the system is functioning by using the IQ/OQ documents as reference (See Operation Manual, Appendix)
For Carbamate and Glyphosate Analyses: Set the detector excitation wavelength to 330 nm and the emission wavelength to 465 nm Do at least four runs of the appropriate test mixture. Inject 10ul. Discard the results of the first injection. Compare the chromatogram with that of the QC test of the system and column. Verify that the system is functioning by using the IQ/OQ documents as reference (See Operation Manual, Appendix)
3.15
Shutdown
Upon completion of the analyses, use one of the following two procedures to shut down the Pinnacle PCX system properly. These procedures can prevent potential column damage, reaction coil blockage, high background fluorescence, reagent precipitation, or other problems.
The Pinnacle PCX must be flushed out at the end of a series of injections, and the reactors must be cooled down. If the instrument is simply stopped, with reagent inside the reactor in a hot state, with no movement, then the heated reactor will become blocked. It is very important for a long useful life that the reactors be flushed out until the temperature is cool.
MANUAL SHUTDOWN
You may shutdown the Pinnacle PCX manually by pressing/selecting the Disable function in the Control menu of the PCX control software. If you choose this function, allow the HPLC to pump for at least 30 minutes to allow for the reactor to cool and to flush reagent from the reactor.
OR AUTOMATIC SHUTDOWN
Create a Shutdown Method for Pinnacle PCX according to section 4 of this manual. Set it up as the last method in the Pinnacle PCX sequence. Make sure pump flow rate is set to 0 mL/min and reactor temperature is set close to room temperature. You can leave the column at the operating temperature or set it at room temperature.
Create corresponding slowdown Method for HPLC. Set it up as the last method in the HPLC sequence. Make sure HPLC flushes column and reactor with column storage eluant for at least 30 min.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 52
3.16
Section 3
In st al la tI on
RECOMMENDED ACCOMPANYING HPLC SLOWDOWN METHOD
Set the HPLC to 100% Storage Eluant (see application section for proper eluant for your column), and set the HPLC pump at the normal analytical flow rate. Choose an eluant that elutes contaminants from the column; for example, methanol for a reversed-phase column and regenerant for an ion-exchange column.
Time (min) % Storage Eluant Flow (ml/min)
0 100 Analytical flow rate*
30 100 Analytical flow rate*
30.1 100 0.0 or 0.02mL/min
*Follow instructions that come with the column.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 53
Section 4
pinnaCle pCx OperatiOn
4.1
Section 4
Pi nn ac le PcX OPerati On
4.1
4.2
4.4
4.6
4.7
4.7
4.10
4.11
4.14
4.17
4.17
Software Overview Status Window Menu Options/Functions Configuring your Pinnacle PCX Log Files Creating and Editing Methods Creating and Editing Sequences Tutorials Running Post-column Analysis Using Pinnacle PCX Stop/Reset Instrument Pinnacle Shutdown
Software Overview
Pinnacle PCX can be controlled only with PC using Pinnacle PCX software. The system can be connected to PC through USB or Ethernet cable.
Pinnacle PCX software controls column and reactor heaters, electronic valves and pumps. It allows to create, store and run post-column methods and sequences. It monitors reactor and column temperatures, pump pressures and flow rates. It keeps logs that records system parameters and HPLC flags.
COMPUTER REQUIREMENTS
Microsoft Windows XP, Vista, or Windows 7 operating environment Minimum one free USB port or network connection Available memory: 10 Mb
HPLC SYNCHRONIZATION
Pinnacle PCX needs to receive injection signal from HPLC to synchronize operation.
For HPLC systems other then Agilent 1100 and Agilent 1200 the HPLC software and system must be capable of sending a relay signal to an external piece of equipment to achieve synchronization.
Agilent Chemstation version 9.0 or higher is needed for Agilent 1100 or Agilent 1200. Pinnacle PCX software will read injection signal from Chemstation directly – no relay connection is needed. If any other software program is used to control Agilent 1100 or Agilent 1200, relays board should be installed on the Agilent pump.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 54
4.2
Section 4
Pi nn ac le PcX OPerati On
Any machine that drives the relay output shall provide a relay contact pair that is electrically isolated from all other electrical devices. The relay signal must have:
Relay detection voltage 24 ±2V Relay detection current Approximately 1 mA
Status Window
The Status window is the main screen of the Pinnacle PCX software (Figure 4-1). The Status window can be moved, or minimized into an icon at the bottom of the PC screen. It has a menu bar on the top, status bar at the bottom and icons for the Pinnacle PCX components.
ICONS
Pump 1:
Displays the actual volume in ml of liquid in the syringe. The status bars will increase/decrease with volume. The actual volume is displayed at the bottom of the icon.
Pump 2: Displays the actual volume in ml
of liquid in the syringe. The status bars will increase/decrease with volume. The actual volume is displayed at the bottom of the icon.
Column oven temperature: Displays the
Set and actual temperature in ºC. The actual
Figure 4.1
temperature is displayed at the bottom of the icon. The status bars on the column oven icon will fill in and increase as the column heats. The reverse is true when the column cools. The icon shows the temperature of the column oven to the nearest degree. The recommended maximum temperature for the column heater is 75°C. A thermal safety switch limits the heater at ca. 80°C.
Reactor temperature: Displays the Set and actual temperature in ºC. The actual temperature is displayed
at the bottom of the icon. The reactor temperature is measured to the nearest degree. The status bars on the reactor oven icon will fill in and increase as the reactor heats. The reverse is true when the reactor cools. The recommended maximum temperature for the heated reactor is 130°C. Above this temperature the reaction coil begins to lose strength. A thermal safety switch limits the heater at ca. 150°C.
Pressure: Displays the actual pressure in bar or psi. Pump 1 pressure corresponds to the actual pressure on
Pump 1, and through post-column system beginning at the first mixing tee. Pump 2 pressure corresponds to the actual pressure on Pump 2, and through the post-column system beginning at the second mixing tee.
Flow: Displays the actual flow rate in ml/min of the reagent pumps. When the pump is dispensing at the
analytical flow rate, the black square in the icon will move slowly in a clockwise direction. When the pump is refilling, the icon changes to an alternating segments that rotate in a counter-clockwise direction.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 55
Section 4
Pi nn ac le PcX OPerati On
MESSAGE AREA
The message area gives details about the status of the instrument. This area describes s, and actions. Some sample messages include: Enabled Heaters are On Instrument Stopped Sequence Stopped Finishing Current Run Sequence Done No Relay Signal
STATUS BAR
The status bar at the bottom of the screen displays the running status (elapsed run time, refilling, ready), loaded method, loaded sequence. The status bar gives very basic information for a quick glance for method loaded and run time. If in ‘Run’ mode: Run Status, i.e. 3 of 7 Elapsed run time in minutes
4.3
BUTTONS
The buttons located on the right side of the Main Window emulate the buttons on the right of the LCD on the Pinnacle PCX instrument.
Stop: This will stop the instrument. It does not matter what task the instrument is performing when this is
pressed. The Pinnacle PCX will stop whatever it is doing and will not resume until the user has Reset the system. The button activates a red LED when the instrument is stopped. This is an emergency function designed to allow the user to fix any catastrophic problems before they do damage or make a mess.
Mode and Select buttons are used during installation and function only on the instrument.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 56
4.4
Section 4
Pi nn ac le PcX OPerati On
Menu Options/Functions
FILE
Import/Export
another instance of the Pinnacle PCX software.
Exit closes Pinnacle PCX software.
CONTROL
Enable
selecting Enable will cause the reactor and column heaters to begin heating. This will not start the pumps. The Enable button changes to Disable when pressed. Press Disable for the instrument to stop heating. Instrument must be enabled before turning pump(s) ON or starting the sequence.
Refill pump(s) this option will refill pumps to the volume calculated for the currently loaded method. If no
method is loaded the full syringe will be refilled. For two-pump instrument there is a choice of refilling pump1, pump2 or both pumps. The pump(s) must be refilled before the sequence is started. The pump(s) will refill automatically between the runs.
Flush pump(s) this option will flush the pumps using solution from the flush bottle. For two-pump
instrument there is a choice of flushing pump1, pump2 or both pumps.
Empty pump(s) this option will empty pump(s). For two-pump instrument there is a choice of empting
pump1, pump2 or both pumps.
Pump1 ON turns on Pump1. This option changes to Pump1 OFF when selected. Pump2 ON turns on Pump2. This option changes to Pump2 OFF when selected. Both pumps ON turns on both pumps. After this option is selected, Both Pumps OFF option becomes
available.
exports existing methods and sequences into a file. The file then can be imported to
CONTROL menu choices are not available when sequence is running
METHOD
Edit/Delete
this option opens Method editing screen where Methods can be created, edited, saved or
deleted.
Print this will print Method information to default printer.
SEQUENCE
Edit/Delete
this option opens Sequence editing screen where sequences can be created, edited, saved,
loaded or deleted.
Print this will print Sequence information to default printer. Load this loads the selected sequence.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 57
Section 4
Pi nn ac le PcX OPerati On
Start sequence starts the loaded sequence. Instrument must be enabled and pump(s) must be refilled
before starting the sequence. After Start Sequence is pressed the following choices become available:
Stop sequence after this button is pressed the Pinnacle PCX finishes current run then stops the pump(s)
and turns the heaters OFF.
Pause sequence this option allows the Pinnacle PCX to finish current run and then stops the pump(s). It
leaves the heaters ON. The sequence will continue after Resume Sequence is selected.
INSTRUMENT
Configuration Maintenance gives the selection of the following operations:
Flush Instrument flushes the pumps and the instrument using solution from the flush bottle. Create and load a method for Pinnacle PCX with suitable reagent flow rates, reactor and column temperature before executing this command. Make sure HPLC is running to avoid backflow into the column. Change seals prepares the pump(s) for seal change. Follow directions on the screen to complete this procedure Prepare for storage/shipping flushes the pumps and the instrument in preparation for long term storage or shipping. Follow directions on the screen to complete this procedure. Pump and valve seal pressure test & system pressure test. Builds up pressure to test the integrity of the seals after a PM.
opens the configuration screen.
4.5
INSTRUMENT menu choices are not available when sequence is running.
STOP/RESET This option performs the same action as Stop button on the front of the instrument. This
option is used for an instant stoppage of the instrument in the case of an emergency. The instrument will cease all the activities, regardless of what is happening. The pump(s) and heaters will be turned OFF and all the loaded methods and sequences will be cleared. After
STOP is selected the option will change to RESET. Select RESET function when you are
ready to start running the instrument again.
HELP
Index this will open searchable Help features for the Pinnacle PCX software.
Send Log to Support
saves logs to file that can be e-mailed to support@pickeringlabs.com . Logs carry information about instrument status from the last 3 days and are necessary to troubleshoot software and instrument problems.
About gives software version number.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 58
4.6
Section 4
Pi nn ac le PcX OPerati On
Configuring your Pinnacle PCX
Select Instrument then Configuration to open configuration window (figure 4-2) The tabs open corresponding windows that allow you to configure the Pinnacle PCX.
Instrument Tab:
Network:
PCX will be connected to the computer through a company’s network. Enter the network IP address in the spaces provided.
USB Port: Select this option if the
Pinnacle PCX will be connected directly to the computer via a USB cable.
Column Tab:
Temperature Range:
displays the current minimum and maximum column temperature range for the Pinnacle PCX.
Reactor Tab:
Volume, ml:
Pinnacle PCX. The temperature range depends on the volume of the reactor. Enter the volume of the reactor that you are using.
Pump Tab:
Max Volume: Flow Rate: 0.05 – 1.5 ml/min: This is the flow rate range for the installed Reagent pump(s). Pressure in: Bar or psi: Select bar if you would like to see the pressure displayed in bars, or psi if you
would like to see the pressure displayed in pounds per square inch.
More Tab:
HPLC type: Select the brand of HPLC that you will be using with Pinnacle PCX. This will determine the type of synchronization between Pinnacle PCX and HPLC system.
Select this option if the Pinnacle
This window
This window displays the current minimum and maximum reactor temperature range for the
70 ml: This is the volume of the syringe installed on the Reagent pumps(s).
Figure 4.2
For HPLC systems other then Agilent 1100 and Agilent 1200 the HPLC software and system must be capable of sending a relay signal to an external piece of equipment to achieve synchronization. Agilent Chemstation version 9.0 or higher is needed for Agilent 1100 or Agilent 1200. Pinnacle PCX software will read injection signal from Chemstation directly – no relay connection is needed. If any other software program is used to control Agilent 1100 or Agilent 1200, relays board should be installed on the Agilent pump.
If your HPLC does not have Relay capabilities, select “No connection to HPLC”.
If you need to connect two Pinnacle PCX to the same PC you need to select multiple instruments during the software installation. For each instrument a separate icon will created on the desktop that will open a main Pinnacle software window. Configure each instrument before starting operation.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 59
Section 4
Pi nn ac le PcX OPerati On
Log Files
Pinnacle PCX records detailed information about the instrument parameters in the Log files. Pump(s) flow rates and pressures, reagent volume, column and reactor temperature, s and flags from HPLC are all stored for 3 days before being overwritten. Log files are necessary to troubleshoot Pinnacle PCX system and software and should always be collected after a problem is detected. To collect Log files go to Help, select Send Log to Support and select file from the day the problem happened. Save the file with the date stamp and e-mail it to support@pickeringlabs.com. If you are not sure about the time collect all 3 log files.
Creating and Editing Methods
To run post-column analysis using Pinnacle PCX a post-column method must be created and executed in the Pinnacle PCX software. To create or edit a post-column method select Method then Edit/Delete. This will open Methods editing screen (figure 4-3). To execute methods (even as a single run) they should be part of the sequence.
4.7
Methods can be edited when they are currently running. Once edited the method should be saved and the sequence reloaded. New methods and methods not currently running can be added/edited/deleted at any time.
Status Bar This area is located at the top of the
Method Edit screen. It displays messages regarding the saving, loading and deleting of methods. For example, when a method called Test 1 is saved, this area will display Method “Test 1” saved.
Method Box This area displays all of the methods
currently in memory of the Pinnacle PCX software. Here you can Edit an existing method, Delete a method, or Add a new method. To perform an action on an existing method, click on the method name.
Add Click here to add a new method. The status
bar will display Method “name” added.
Delete To delete a method, click once on the
method name, then click Delete. The Status bar will display Method “name” deleted.
PUMPS TAB Timetable for both pumps (one table for both). Pump 2 flow rate column is grayed out for
1-pump systems. If timetables are empty the initial conditions are the conditions for the whole run.
Time, min: Enter the time in minutes to the nearest tenth, but not greater than the Runtime. (For example,
0.5 minutes)
P1, ml/min Enter the desired flow rate for Pump1, between 0.05 and 1.5 ml/min. (For example, 1.3 ml/min.
This means that Pump 1 will begin pumping 1.3 ml/min at 0.5 minutes)
Figure 4.3
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 60
4.8
Section 4
Pi nn ac le PcX OPerati On
P2, ml/min Enter the desired flow rate for Pump2, between 0.05 and 1.5 ml/min. (For example, 0.08 ml/min.
This means that Pump 2 will begin pumping 0.08 ml/min at 0.5 minutes). Enter 0.0 ml/min if you want to stop pumping reagent at a set time during the run.
Add This will add a new line below the currently selected line in the table. Insert This will insert a new line above the currently selected line in the table. Delete This will delete the currently selected line in the table.
COLUMN TAB Timetable for the column oven. If timetables are empty the initial conditions are the conditions
for the whole run.
Time, min Enter the time in minutes to the nearest tenth, but not greater than the Runtime. (For example,
15.5 minutes)
Temp, °C Enter the desired temperature, in whole numbers, in ºC. (For example, 53ºC. This means that the
column will reach 53ºC at 15.5 minutes)
Add This will add a new line below the currently selected line in the table. Insert This will insert a new line above the currently selected line in the table. Delete This will delete the currently selected line in the table.
INITIAL SETTINGS In this section, set up the initial conditions for the Pinnacle PCX. If no pump timetable or
column timetable is required, Pinnacle will use this information for analysis. The items marked with * indicate required parameters.
*Run Time, min In this window, enter the total run time for the Pinnacle PCX method. This must be a whole
number, greater than 1
*Equilibration Time, min Enter the time between runs. The Pinnacle PCX will use this information to
determine when to begin pumping again prior to the next injection. Minimum time is 5 minutes.
*Column t, °C Enter the initial column temperature in ºC. Enter a whole number between 30ºC and 75ºC. Column Type Enter the column type that is used for this application. Any characters are acceptable. There is
no limit on number of characters, but 20 are visible in the window at a time.
*Reactor t, °C Enter the reactor temperature for the analysis. Enter a whole number between 30ºC-130ºC for
reactor volumes 2.0ml, or between 30ºC-80ºC for reactor volumes >2.0 ml. (The reactor volume is set in Configuration).
Reactor Volume, ml Enter the volume of the reactor used in this application. *Pump 1 Flow rate, ml/min Enter the initial flow rate for Pump 1. Enter a number between 0.05-1.50 ml/
min, or 0
Reagent 1 Enter the name or description of Reagent 1. Any characters are acceptable. There is no limit on
number of characters, but 20 are visible in the window at a time.
*Pump 2 Flow rate, ml/min Enter the initial flow rate for Pump 2. Enter a number between 0.05-1.50 ml/
min, or 0. Put 0.0 for a single-pump system.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 61
Section 4
Pi nn ac le PcX OPerati On
Reagent 2 Enter the name or description of Reagent 2. Any characters are acceptable. There is no limit on
number of characters, but 20 are visible in the window at a time.
ADVANCED SETTINGS
This tab allows you to designate currently open method as a Flush Method (figure 4-4). This method can be set as a last method in the Pinnacle PCX sequence in order to flush the reagent pumps and the Pinnacle PCX reagent lines and reactor. Pinnacle PCX will use solution in the Flush bottle to do this. Always set Flush Method as the last method in the sequence and make sure corresponding HPLC method is created and set as the last method in the HPLC sequence. Flushing is not necessary after every sequence and only recommended before changing applications and long time storage or in the case of extremely aggressive reagents.
4.9
ACTION TABS
Save
This will save the current method
Save As This will save the current method
Figure 4.4
information using a different name.
Cancel This will close the window and cancel any changes. No changes will be made to the method. Print This will print the method to the default printer. The information printed will be: print date, operator,
method name, save date, initial conditions, pump time table, column time table.
Help This will open the searchable Help features for the Pinnacle PCX software. Close This will close the window.
IMPORTANT PRACTICAL CONSIDERATIONS
In order to start executing the method Pinnacle PCX needs to receive injection signal from HPLC. To avoid missing the signal it is very important to match Pinnacle PCX method run time and equilibration time to that of your HPLC method.
HPLC pump uses equilibration to return to original conditions before the next analysis. HPLC software programs have different ways of setting up equilibration. Most often used are the following:
- as a PostRun;
- at the end of the pump gradient table;
- as additional time before the injection (set as a PreRun or as negative time in the pump gradient table).
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 62
4.10
Section 4
Pi nn ac le PcX OPerati On
Even if isocratic method is used and no equilibration time is required for HPLC pump Pinnacle PCX still needs at least 5 min of Equilibration time to refill reagent pumps and stabilize the baseline signal before the next injection.
If equilibration for HPLC pump is set up as PostRun or PreRun match Equilibration time in the Pinnacle PCX method to respectively PostRun or PreRun time of your HPLC method. The run time of your HPLC method (the actual time of the analysis) should be the same as Pinnacle PCX Run time. Pinnacle PCX delivers reagents only during the Run time set up in the Pinnacle PCX method.
If returning to initial conditions is set as part of the HPLC pump gradient table (at the end or as a negative time step in the beginning) consider how long the actual analysis and equilibration steps are. Match analysis time to Run time of the Pinnacle PCX and match time of the equilibration step to Equilibration time of the Pinnacle PCX.
Creating and Editing Sequences
To create or edit a post-column sequence select Sequence then Edit/Delete. This will open Sequence editing screen (figure 4-5).
Status Bar This area is located at the top of the
Sequence Edit screen. It displays messages regarding the saving, creative and deleting of sequences. For example, when a sequence called Test 1 is saved, this area will display Sequence “Test 1” saved.
Select/Edit Sequence Box This area displays all of
the sequences currently in memory of the Pinnacle PCX software. Here you can Edit an existing sequence, Delete a sequence, or Add a new sequence. To perform an action on an existing sequence, click on the sequence name.
Create Click here to add a new sequence Delete To delete a sequence, click once on the sequence name, then click Delete.
SEQUENCE TABLE WINDOW
Method name Number of runs
Figure 4.5
Select the method name from the drop down menu
Pinnacle Operators Manual
Pickering Laboratories Inc.
1-...
Page 63
Section 4
Pi nn ac le PcX OPerati On
Add This will add a new line below the currently selected line in the sequence.
Insert This will insert a new line above the currently selected line in the sequence. Delete This will delete the currently selected line in the sequence.
The end of the sequence should be with a shutdown method or a flush method.
ACTION TABS
Save
This will save the current sequence.
Save As This will save the current sequence information using a different name. Load This will load the current sequence into the Pinnacle PCX software. Cancel This will close the window and cancel any changes. Print This will print the sequence to the default printer. The information printed will be: print date, operator,
sequence name, sequence table (with method names), save date.
Help This will open the searchable Help features for the Pinnacle PCX software. Close This will close the editing window.
4.11
EDITING THE RUNNING SEQUENCE:
- select the Sequence name from the Sequence Box.
- change number of runs of the currently running Method or add a new line to the Sequence anywhere below the Method in progress.
- select Save and then Load from the Actions tabs.
- check number of runs on the bottom of the main software window to confirm that changes has taken affect.
Tutorials
CREATE A METHOD
The following is a tutorial on how to create a method and start a run on the Pinnacle PCX. It is not intended to be used for a real run because the flow rates and temperatures are not representative. If you like, you may use this tutorial as a guide to create your own methods.
This example is a fictitious method and is not intended to analyze samples.
Double-click on the Pinnacle PCX icon on your desktop. This will open the main Status Window of the Pinnacle PCX software. Go to Method, then select Edit/Delete. This will open the Method Edit Screen. Here you will set the initial conditions, flow rates, temperatures for your method.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 64
4.12
Section 4
Pi nn ac le PcX OPerati On
Click Add Method.
Type a Name for your method. For this example, type “Tutorial”, and click OK.
The Pinnacle PCX software has now added the method “Tutorial” to the list of available methods. Highlight this method if it is not already done so.
Under Initial Settings Tab, type: Run Time, min: 60 Equilibration Time, min: 5 Column t, °C: 38 Column Type: 1154150 Na cation exchange Reactor t, °C: 36 Reactor Volume, ml: 0.5 Pump 1 Flow rate, ml/min: 0.3 Reagent 1: Hypochlorite Pump 2 Flow rate, ml/min: 0.3 Reagent 2: OPA
Click on the Pumps Tab, then click once in the box under Time. The cursor should now be flashing here. Type in your desired Times and flow rates below. Use the Add button to add lines to the table. Use the data from the table below as a guide. Time is the elapsed time in minutes. P1 and P2 are the flow rates of each pump in ml/min.
Time, min P1, ml/min P2, ml/min
1.0 0.3 0.5
1.6 0.8 0.0
3.8 1.0 0.05
12.5 0.0 0.05
30 0.1 0.0
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 65
Section 4
Pi nn ac le PcX OPerati On
Next, click on the Column Tab, then click once in the box under Time. The cursor should now be flashing here. Type in your desired Times and temperatures below. Use the Add button to add lines to the table. Use the data from the table below as a guide. Time is the elapsed time in minutes. Temperature is in °C and is the desired temperature of the column at the time specified.
Time, min Temp, °C
5.0 45
9.5 48
25 56
50 75
Now Click Save. This has saved the conditions you have created in the method “Tutorial”.
4.13
Click Close to close the Edit Method Window.
Next, we will go through the Create Sequence Tutorial.
CREATE A SEQUENCE
As with the above method, this example sequence is not designed to be used for analysis. It will guide you through the steps and you may use your own method for the Sequence Table.
From the main Status Window, go to Sequence, then select Edit/Delete.
This will open the Sequence Edit Screen. Here you will set the method names, and the number of injections for each method.
Click Create Sequence.
Type a Name for your Sequence. For this example, type “Tutorial 1” and click OK.
The Pinnacle PCX software has now added the Sequence “Tutorial 1” to the list of available sequences. Highlight this sequence if it is not already done so.
Under Method Name, there is a drop down menu. Choose the Method “Tutorial”.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 66
4.14
Section 4
Pi nn ac le PcX OPerati On
Next, under the Runs, type in the number of injections you plan to make using this method. In this example, it will be 3.
Your table should now look like:
Method Name Runs
Tutorial 3
The number of Runs should match the number of runs you plan to make on the HPLC using that method. In this example, your system will perform 3 injections using the Pinnacle PCX method “Tutorial”. The sequence table on your HPLC system should also have 3 injections programmed.
Now Click Save. This has saved the conditions you have created in the sequence “Tutorial 1”.
Click Close to close the Edit Sequence Window.
Running post-column analysis using Pinnacle PCX
To start running post-column analysis using Pinnacle PCX follow the steps below:
Create analytical method(s) for your HPLC system using gradient conditions and detector settings from Pickering Laboratories application notes or other sources. For any other HPLC software except Agilent Chemstation, set the relay start signal close to time 0. Depending on the software the relay stop signal can be either a few minutes after injection or at the end of the run. Contact your HPLC support representative if you have any questions about setting up relay signal in the HPLC method.
Create post-column method(s) for your application using post-column conditions from Pickering Laboratories application notes or other sources.
Set up HPLC sequence. Contact your HPLC support representative if you have any questions about creating and running HPLC method or sequence.
Start HPLC pump. Increase the flow rate gradually to avoid overpressuring the analytical column.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Figure 4.6
Page 67
Section 4
Pi nn ac le PcX OPerati On
Create sequence for Pinnacle PCX using your post-column method(s). Set up the same number of runs as in the HPLC sequence.
Select Save to save Pinnacle PCX sequence.
Select Load to load Pinnacle PCX sequence. Sequence name and the first method name will appear in the status bar at the bottom of the Pinnacle PCX software.
Select ControlEnable to turn column and reactor heaters ON. Set and actual temperatures will be displayed on the heaters icons. Wait until the actual temperature reaches the set point.
Select ControlFlush both pumps if this is the first run after the installation or if the instrument was not in use for a long time.
4.15
Select ControlRefill Pump(s) to fill pumps with reagent. Empty pumps first if there is any old reagent left or if instrument was turned OFF earlier. The pump icon should show partially filled syringe and actual volume of reagent inside (Figure 4-6).
NOTE: If the instrument was turned OFF with partially filled pumps the icons will show alternated black and white segments and no volume data will be displayed. This means that the instrument does not know the position of the piston inside the syringe. Select Control Empty Pump(s). This will allow Pinnacle PCX to find empty position.
Once reactor and column temperatures are stable and analytical column is equilibrated select ControlBoth pumps ON (or Pump ON for single-pump instrument) to start pumping reagents. Wait until the pressures are stable.
NOTE: Pinnacle PCX software will calculate needed reagent volume based on your run time and reagent flow rate. Some extra volume will be added to that to ensure pumps will not run out of reagent during operation. Do not run pumps for more then 5 min before starting the sequence to avoid running out during the analysis.
Select SequenceStart Sequence. Run time (0.0) will be displayed in the Status Bar at the bottom of the Pinnacle PCX software. Pinnacle now waits for injection signal from HPLC in order to start running the method.
Start HPLC sequence. Once an injection is made Pinnacle PCX software will receive an injection signal via relays or directly from Agilent Chemstation. After that Pinnacle will start the post-column method. Check the timer at the bottom of the Pinnacle PCX software to make sure it matches the elapsed time of the HPLC run.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 68
4.16
Section 4
Pi nn ac le PcX OPerati On
NOTE: after starting a new sequence always make sure that first injection signal was received normally by checking that Pinnacle PCX timer matches elapsed time of the HPLC run. Pinnacle PCX will stop the sequence if no injection signal is received from HPLC in 15 min from selecting Start Sequence.
After the post-column method Run time is done Pinnacle PCX turns the pump(s) OFF , checks temperature and flow rate settings for the next run and automatically refills the reagent. Pinnacle PCX will turn the pump(s) ON 5 min before the end of the Equilibration time.
Once next injection signal is received Pinnacle PCX starts the second method in the sequence. The number of the current run and total number of runs in the Sequence are shown on the bottom of the Pinnacle PCX software.
After Sequence is complete Pinnacle PCX turns OFF the pump(s). Message “Sequence done. (Runs n of N)” is displayed in the software window.
Important Practical Considerations
Always match number of runs in HPLC and Pinnacle PCX sequences. If more samples are added to HPLC sequence or samples are removed edit Pinnacle PCX sequence accordingly. Total number of samples displayed by HPLC software often does not include calibrators and controls. Make sure you count total number of runs in HPLC sequence not just number of samples before creating Pinnacle PCX Sequence.
To edit the running Sequence:
- select SequenceEdit/Delete
- select the Sequence name from the Sequence Box.
- change number of runs of the currently running Method or add a new line to the Sequence anywhere below the Method in progress.
- select Save and then Load from the Actions tabs.
- check number of runs on the bottom of the main software window to confirm that changes has taken affect.
We recommend restarting Pinnacle PCX software after Sequence is completed.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 69
Section 4
Pi nn ac le PcX OPerati On
STOP/RESET Pinnacle PCX
In a case of an emergency there are two ways of stopping Pinnacle PCX:
On the front of the instrument press STOP button. To reset turn the instrument OFF and ON again. In the Pinnacle PCX software select STOP. Once selected this option changes to RESET. Select RESET once the emergency is cleared.
These option are used for an instant complete stoppage of the instrument. Pinnacle PCX will cease all the activities, regardless of what is happening. The pump(s) and heaters will be turned OFF and all the loaded methods and sequences will be cleared. When ready (after the problem is fixed and you have rinsed the analytical column) load a sequence again and follow steps to start post-column analysis.
4.17
Pinnacle Shutdown
Pinnacle can be shut down at the sequence by either using a Shutdown method or a Flush method. The latter is recommended if the instrument is be stored for a long time or if reagents will be changed.
Upon completion of the analyses, use one of the following two procedures to shut down the Pinnacle PCX system properly. These procedures can prevent potential column damage, reaction coil blockage, high background fluorescence, reagent precipitation, or other problems. The Pinnacle PCX must be flushed out at the end of a series of injections, and the reactors must be cooled down. If the instrument is simply stopped, with reagent inside the reactor in a hot state, with no movement, then the heated reactor will become blocked. It is very important for a long useful life that the reactors be flushed out until the temperature is cool.
MANUAL SHUTDOWN
You may shutdown the Pinnacle PCX manually by pressing/selecting the Disable function in the Control menu of the PCX control software. If you choose this function, allow the HPLC to pump for at least 30 minutes to allow for the reactor to cool and to flush reagent from the reactor.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 70
4.18
Section 4
Pi nn ac le PcX OPerati On
AUTOMATIC SHUTDOWN
Create a Shutdown Method for Pinnacle PCX using the following settings as an example:
Run time: 30 min Equilibration time: 5 min Reactor temperature: 35 C Column temperature: 35 C Pump(s) flow rate: 0 mL/min
Set it up as the last method in the Pinnacle PCX sequence. Make sure pump flow rate is set to 0 mL/min and reactor temperature is set close to room temperature. You can leave your column at your operating temperature or set it at room temperature.
Create corresponding slowdown Method for HPLC. Set it up as
Figure 4.7
the last method in the HPLC sequence. Make sure HPLC flushes column and reactor with column storage eluant for at least 30 min.
RECOMMENDED ACCOMPANYING HPLC SLOWDOWN METHOD
Set the HPLC to 100% Storage Eluant (see application section for proper eluant for your column), and set the HPLC pump at the normal analytical flow rate. Choose an eluant that elutes contaminants from the column; for example, methanol for a reversed-phase column and regenerant for an ion-exchange column.
Time (min) % Storage Eluant Flow (mL/min)
0 100 Analytical flow rate*
30 100 Analytical flow rate*
30.1 100 0.0
*Follow instructions that come with the column.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 71
Section 5
MaintenanCe
5.1
Section 5
Ma in te na nc e
5.1
5.2
5.3
5.6
5.7
5.8
5.8
5.9
5.12
5.16
5.17
Your Pickering Pinnacle PCX will require some routine maintenance to stay in top condition. Ordinarily, little maintenance is needed beyond good operating procedures. In this section you will find some suggestions for maintaining your Pinnacle PCX, as well as Proper Shutdown procedures for typical, Long-term, and Storage of the Pinnacle PCX.
Maintenance Suggestions Operating Suggestions Proper Shutdown Procedures Basic Maintenance Procedures Reagent Filter Replacement Ambient Reactor Replacement Replacement of Heated Reactor Cartridge Valve Maintenance Pump Seal Replacement Fuse Replacement Tubing Kit Guides
Maintenance Suggestions
Make copies of the blank forms in the Appendix and complete the parameter log on the photocopy.
Record the pressures for the system equilibrated under initial conditions. Keep a daily log of Column, Pump 1 and Pump 2 (if applicable) pressures for diagnostic use. Include all the settings for the pump, injector, detector, and integrator.
Record any repairs, or problems in the Instrument Log.
Keep copies of the QC chromatograms and logs, and of the initial chromatograms generated at installation.
Create a maintenance schedule for your laboratory to ensure that the instrument is kept in good working order.
Always use PEEK nuts and ferrules in the Pinnacle PCX fluidics lines. Use of stainless steel nuts and ferrules in the PEEK ports will void the warranty. Stainless steel nuts and ferrules will cut the threads at the bottom of the ports, causing them to leak. If this happens, the entire item may need replacing.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 72
5.2
Section 5
Ma in te na nc e
Operating Suggestions
Below are several helpful suggestions and bits of information which will help keep your Pinnacle PCX, columns, and reagents in top condition. The following information is broken down into sections for convenience.
There are two items that should be part of any good laboratory practices, regardless of what equipment is being used:
1) Always wear safety glasses or goggles, laboratory coat, gloves, and other appropriate safety-clothing.
2) Read and understand the instructions in the MSDS’s shipped with the chemicals. If the MSDS’s are missing, please contact Pickering Laboratories and we can fax you a copy, or you can download them from our website at www.pickeringlabs.com.
PINNACLE PCX
Check for leaks daily at the column fittings; the eluants can be corrosive. Frequently observe and record the pressures and check for leaks. You may find a problem before it becomes serious. Always use the correct fuse. Use of the incorrect fuse can seriously damage the PC board and even the valves, reactor and pump. Use of an incorrect fuse will void the warranty. Disconnect the power cord before removing the case of the Pinnacle PCX. Always follow the proper shutdown procedures. See Below. Use the proper start-up and shutdown procedures consistently (see Section 4). Periodically Flush the instrument to ensure long life. Do not operate the heated reactor above the boiling point of the eluant unless the back-pressure regulator is connected to the waste line of the detector. Boiling inside the reactor can cause precipitates to block the reactor.
Operating above the boiling point without a back pressure regulator will void your warranty.
Thoroughly clean any leaks from fittings with water and dry with paper towels, especially if the solution is a buffer or hydroxide. Standing salt and hydroxide solution are corrosive.
Soak up spills with rags, paper towels, or sponge. Clean spill-area with a wet towel and thoroughly dry. Do not spray water directly into the instrument.
Rinse out the drip tray periodically.
MAINTENANCE SCHEDULE
Check for leaks around all fittings and record column and reagent pressures daily. Change the reagent filter every 3 months. Change valve and pump seals every 12 months. Change the heated reactor and all tubing every two years.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 73
Section 5
Ma in te na nc e
PICKERING COLUMNS
Do not operate with a column pressure above 2800 psi (193 bars) for an extended period of time. Isolate the source of the high pressure and replace those items. Never disconnect any fittings between the pre-column check valve and the column until the post-column system has been shut down and depressurized (loosen the detector inlet fitting first). If the column pressure drops below the post-column pressure, it is possible for reagent to back-flow onto the column.
When removing the column, remove the outlet fitting first.
Never pump organic solvent through an ion-exchange column. This will swell the resin and over-pressure the column. Never pump high pH buffers or hydroxide through a silica column. This can dissolve the silica and cause extensive damage to the post-column system by permanently blocking tubing. When switching a system between ion-exchange and reversed-phase applications, be sure to flush the HPLC and injector with water before connecting the column. Eluants for one analysis may damage the column for the other.
5.3
ELUANTS AND REAGENTS
Thiofluor is extremely hygroscopic. Always keep it in a tightly closed container. Sodium borate (any grades) contains excessive amounts of heavy metal contaminants and insoluble matter. These impurities will eventually precipitate in the reactor and flowcell. The one year warranty does not
cover damage caused by these contaminants.
If you must prepare your own borate buffer for the OPA reagent, do not use sodium tetraborate as suggested by the EPA methods. Instead, use molar equivalents of boric acid and sodium hydroxide, because they are available in higher purity (ACS-grade or better) and have very little insoluble matter. Use Pickering Laboratories reagents and eluants. The quality of the chemicals is guaranteed and the cost is low relative to the worth of your analytical results. The one year warranty does not cover damage caused by
poor-quality reagents and eluants not purchased from Pickering Laboratories.
Do not touch the interior of the mobile phase reservoirs and the dip tubes with your fingers. Amino acids in fingerprints will cause contamination. Gloves are suggested. Do not leave caps and lines dangling without a reservoir. To fill reservoir, transfer caps and lines into a spare bottle or an Erlenmeyer flask filled with deionized water.
Proper Shutdown Procedures
Upon completion of the analyses, use one of the following two procedures to shut down the Pinnacle PCX system properly. These procedures can prevent potential column damage, reaction coil blockage, high background fluorescence, reagent precipitation, or other problems.
The Pinnacle PCX must be flushed out at the end of a series of injections, and the reactors must be cooled down. If the instrument is simply stopped, with reagent inside the reactor in a hot state, with no movement, then the heated reactor will become blocked. It is very important for a long useful life that the reactors be flushed out until the temperature is cool.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 74
5.4
Section 5
Ma in te na nc e
MANUAL SHUTDOWN
You may shutdown the Pinnacle PCX manually by pressing/selecting the Disable function in the Control menu of the PCX control software. If you choose this function, allow the HPLC to pump for at least 30 minutes to allow for the reactor to cool and to flush reagent from the reactor.
OR AUTOMATIC SHUTDOWN
Create a Shutdown Method for Pinnacle PCX according to section 4 of this manual. Set it up as the last method in the Pinnacle PCX sequence. Make sure pump flow rate is set to 0 mL/min and reactor temperature is set close to room temperature. You can leave the column at the operating temperature or set it at room temperature.
Create corresponding slowdown Method for HPLC. Set it up as the last method in the HPLC sequence. Make sure HPLC flushes column and reactor with column storage eluant for at least 30 min.
RECOMMENDED ACCOMPANYING HPLC SLOWDOWN METHOD
Set the HPLC to 100% Storage Eluant (see application section for proper eluant for your column), and set the HPLC pump at the normal analytical flow rate. Choose an eluant that elutes contaminants from the column; for example, methanol for a reversed-phase column and regenerant for an ion-exchange column.
Time (min) % Storage Eluant Flow (mL/min)
0 100 Analytical flow rate*
30 100 Analytical flow rate*
30.1 100 0.0
*Follow instructions that come with the column.
LONG-TERM SHUTDOWN
The above methods are used if the Pinnacle PCX will be used again within 3 days of the last run. If you plan to not use the Pinnacle PCX for more than 3 days, we recommend that you use the Long-term Shutdown.
In this procedure, the reactor is flushed free from reagent, thus ensuring that it will be free from blockage upon your return.
There are two ways to perform this shutdown; automatic, or manual.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 75
Section 5
Ma in te na nc e
MANUAL LONG-TERM SHUTDOWN
Make sure your HPLC pump is running storage eluant through the analytical column (see Application section for proper eluant) and then remove the column for storage. Replace the column with a union and the eluants with Water and Methanol. Set the HPLC pump to 80% Water and 20% Methanol at 1.0 ml/min. This will prevent backflow into the column during Flush procedure.
Create a Flush method as described in section 4.9 under the Advanced Settings tab of this manual. Set up a Sequence with one run of a Flush method and load it into Pinnacle PCX software.
Select Enable function in the Control menu.
Select InstrumentMaintenance- Flush Instrument. Follow directions on the screen to complete this procedure.
5.5
After Flush Instrument is completed select Disable function in the Control menu of the Pinnacle PCX software to turn OFF heaters and pumps.
Allow the HPLC to pump at least 30 minutes to allow for the reactor to cool.
AUTOMATIC LONG-TERM SHUTDOWN
Create a Flush method as described in section 4.9 under the Advanced Settings tab of this manual. Set it up as the last method of your Pinnacle PCX sequence.
Create a corresponding method for your HPLC to flush analytical column with the storage eluant (see Application section for proper eluant) and set it up as a last method of your HPLC sequence.
After sequence is completed select Disable function in the Control menu of the Pinnacle PCX software to turn OFF heaters and pumps.
Allow the HPLC to pump at least 30 minutes to allow for the reactor to cool.
STORAGE
If the Instrument will not be used for a period longer than 2 weeks, it must be put into Storage state. This becomes necessary because as reagents age, the chances of precipitation in the heated reactor increases.
Perform Manual or Automatic Long-Term Shutdown as described above.
Remove the reagents and replace them with a solution of 80/20 Water/Methanol.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 76
5.6
Section 5
Ma in te na nc e
Select Refill Pump(s) from Control menu of the Pinnacle PCX software. This will flush the reagent lines from the bottles from the remaining reagent.
Select in the Pinnacle PCX software Instrument – Maintenance – Prepare for Storage/Shipping. Follow directions on the screen to complete the procedure.
Pump the storage eluant through the analytical column (see Application section for proper eluant), and remove the column and guard. Cap both column and guard tightly. Replace any buffers with water and flush the HPLC lines for 5 minutes. Replace the water with 80/20 Water/Methanol and flush the lines. Place a restrictor or union in the column oven in place of the column. Pump 80/20 Water/Methanol through the system.
3
Basic Maintenance Procedures
The following section will describe the procedure for cleaning/replacing/rebuilding key elements of the Pinnacle PCX instrument. For the following procedures, refer to Figure 5-1 below, which calls out the various parts of the Pinnacle PCX.
Starting at the valves, the parts of the Fluidics Panel are numbered counter-clockwise.
1. Valve 2 (on dual systems only)
2. Pressure Transducer 2 (on dual systems only)
3. Gas is controlled by the toggle valve. Lever ON
pressurizes the manifold.
4. Pump 2 (on dual systems only)
5. Column Outlet
6. Mixing Manifold 2 (with integrated reagent
filter) (on dual systems only)
7. Heated Reactor
8. Mixing Manifold 1 (with integrated reagent
filter and over-pressure relief valve)
5
2
4
1
6
7
8
9
10
13
11
12
14
Figure 5-1
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 77
9. Over pressure Relief valve
10. Pump 1
11. Pressure Transducer 1
12. Valve 1
13. Ambient reactor (on dual systems only)
14. Reactor outlet and union to connect to detector
Reagent Filter(s) Replacement
Refer to Figures 5-1 and Figure 5-2 for the location of the Reagent filter(s).
There are two filters in the Duplex system, and one filter on the Simplex system. They are located in the reagent manifolds; Items 6 and 8.
5.7
Section 5
Ma in te na nc e
Reagent Filter
1. Be sure that the Pinnacle PCX is shut down and depressurized before changing a filter.
2. Remove the 1/16" fitting from the inlet of the reagent filter.
3. Use a 7/16" wrench to remove the filter.
4. Replace the filter element. The correct filter element is 3102-9040, a 10 µm frit. Tighten with fingers first, then
Figure 5-2
give it a 1/4 turn with a 7/16" wrench Reconnect the tubings to the filter. Use care not to strip the threads, and do not overtighten. Start the LC pump and inspect for leaks.
Note: Always use a PEEK nut and ferrule to connect to the reagent filters. They are made of soft PEEK material, and stainless steel nuts and ferrules will cut them, causing them to leak.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 78
5.8
Section 5
Ma in te na nc e
Ambient Reactor Replacement
The Ambient Reactor is Item 13 (on dual systems only).
1. Shutdown the post-column system and let the reactor cool for at least 30 min.
2. Disconnect the two fingertight fittings at either end of the ambient reactor. These will be located on the manifold (Item 6) and the inlet of the detector.
3. Loosen the screw that holds the black spindle to the fluidics panel with a 3/32" Alan wrench.
4. Remove the spindle and remove the used reactor.
5. Slide the new reactor onto the spindle and fasten to the fluidics panel.
6. Re-connect the fingertight fittings to the manifold and the union (Items 6 and 14). Do not overtighten the fittings.
7. Start the LC pump and inspect for leaks.
Figure 5-3
Replacement of Heated Reactor Cartridge
The Heated Reactor is Item 7. Please refer to Figure 5-4 for removal of the Reactor Cartridge.
1. Shutdown the post-column system and let the reactor cool until it is a safe temperature for handling (below 30˚C).
2. Disconnect the fingertight fittings connected to the heated reactor. These will be located on the manifolds (Items 6 and 8) for duplex systems or on the manifold and union in simplex systems (Items 8 and 14).
3. Loosen the retaining screw at the top of the reactor cartridge using a flathead screwdriver (see Figure 5-4).
4. Gently slide the Reactor Cartridge out of the reactor housing.
5. Replace the heated reactor cartridge with the proper volume (see Appendix).
6. Reconnect the two fingertight fittings. Do not overtighten. The inlet tubing of the heated reactor is marked with a piece of black shrink tube and connects to mixing manifold I (Item 8). Start the LC pump and inspect for leaks.
Retaining Screw
Figure 5-4
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 79
Section 5
Ma in te na nc e
Valve Maintenance
Flushing the Valves is recommended before valves replacement or resealing to remove reagents from lines and pumps. To flush the valve follow the steps below:
Remove the reagents and replace them with a solution of 80/20 Water/Methanol.
Select Refill Pump(s) from Control menu of the Pinnacle PCX software. This will flush the reagent lines from the bottles from the remaining reagent. Select empty pump(s) to dispense liquid from pump.
Select Flush Pump(s) from Control menu of the Pinnacle PCX software.
5.9
Now Valves are ready for maintenance.
VALVE FACE REMOVAL PROCEDURE
Replace the post-column reagents with 80/20 water/methanol. Flush the pump connected to the valve that requires maintenance (in the software, go to Control, then Flush Pump). When the flush is complete, pull the Flush tubing out of the liquid to prevent siphoning of the liquid back into the flush line.
Power down the Pinnacle PCX and turn the power back on again. This will move the valve to the blocked position. This is important for alignment of the new or repaired valve face.
Place some paper towels under the Pinnacle PCX to absorb any excess liquid that is in the Flush line. Disconnect the Waste line (position 6) from the Valve face. Hold the end of the tubing upright for a moment to allow any liquid to drain into the waste
1
2
6
container. Disconnect the remaining fittings from the valve face.
3
5
4
Flat-head screw
Flat-head screw
Using an Alan wrench and a flat-head screwdriver, loosen the cap screw that holds the valve indicator plate in the center of the valve, and the 2 flat-head screws located between ports 1-2 and 4-5.
Cap screw
Figure 5-5
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 80
5.10
Section 5
Ma in te na nc e
Pull the entire face horizontally forward. There will be some resistance.
If there has been any crystal build-up behind the valve, insert the jack screws contained in the Valve Maintenance kit and tighten them until the face of the valve comes forward and you are able to remove it (figure 5-7).
Body
Warning: When removing the valve face with the jack screws, it is important to remember to remove the valve using equal turns on each side. It is important that it come off of the instrument as straight as possible to prevent deforming the soft PEEK material.
Remove the jackscrews after the valve is removed.
Inspect the face of the motor mount. Ensure that there is no excess crystal build up. Carefully clean any crystals with a paper towel and some DI Water.
Using the Pinnacle Valve Seal tool, turn insert clockwise and push to remove spool (figure 5-8).
O-RING REMOVAL
DO NOT use any sharp tools as this will scratch the soft plastic of the insert.
For the 5 smaller O-rings (Size 4) – use a length of 1/16 OD Peek tubing to remove. For the 2 large O-rings (Size 18) – pinch o-rings with fingers and roll to remove.
Spool
Figure 5-6
jack scews
Figure 5-7
Once O-rings are removed, sonicate PEEK ring and PTFE insert to remove any crystals from valve. O-rings will be replaced.
O-RING REPLACEMENT
Clean the new O-rings first by sonicating them in a soapy water bath for 20 minutes and then rinse thoroughly with DI Water. Wear gloves and use care not to get dust or dirt on them after cleaning. Push the 5 new Size 4 O-rings into the ports of the PEEK ring. Port 6 (Waste port) does not have an O-ring.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Figure 5-8
Figure 5-9
Page 81
Section 5
Ma in te na nc e
Carefully slide the 2 new Size 18 O-rings onto the white center spool.
Note: Never reuse the O-rings. They take a shape after they are installed and will not conform to a new shape.
Note: Never use substitute O-rings.
Only use O-rings provided and approved by Pickering Laboratories.
INSERTING THE SPOOL INTO THE PEEK RING
Using the special Pinnacle Valve Seal tool (the same tool used to remove the spool), carefully put the PTFE center back into place in the PEEK ring. Rotate the insert clockwise to prevent slicing of the O-rings. Look into the valve, and push insert toward you to ensure that no O-rings pop out.
1
2
6
5.11
Line up the holes in the spool so that when it is placed on the Pinnacle, the V is between ports 2-3. This is the equivalent to the blocked position of the insert.
3
5
4
REPLACING THE NEW/REBUILT VALVE ON THE PINNACLE
Ensure that at the front face of the new valve, the body and spool are completely flush. These two pieces must be level.
Slide the new valve onto the motor mount, so that the Number 1 is at the top.
Ensure that the dowels match the openings in the back of the valve. Ensure that the V is between ports 2-3 (that is, ensure that the spool is in the home blocked position) refer to figure 5-10.
Push the valve face firmly against the Pinnacle PCX. Ensure that the Valve Spool is firmly pushed against the motor mount.
Replace the cap screws and tighten until you feel slight resistance. Then make 1/8 turn past snug.
Perform a pump flush to exercise the new valve and to properly position the spool within the body (remember to put flush lines back into the solution).
Figure 5-10
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 82
5.12
Section 5
Ma in te na nc e
Pump Seal Replacement
The Pinnacle PCX uses Pickering’s own pulse-free syringe pump with piston wash. The pump is specially designed for extended seal life, however the piston seals will require periodic replacement. The length of service to be expected from the seal depends on a wide variety of factors, including whether proper shutdown procedures were followed, how often the system was turned on and off, and whether the piston-wash system was wetted.
It is critically important that the seal be replaced immediately upon failure, or better yet, before failure, because the reagent can leak into the mechanical housing of the pump and cause corrosion. When a leak occurs, you may notice fluid on the side of the pump.
The pumps are Items 4 and 10.
SOFTWARE SET UP
To replace the seals, open the PCX control software and select Maintenance under the Instrument menu. From the Maintenance sub-menu, select Change Seal, and follow the onscreen instructions.
Wait until the end of a run before performing this procedure to conserve reagent. The Pinnacle will perform the Flush pump sequence, and will flush reagent out of the syringe. Pull the Piston Wash and Flush lines out of the liquid, so they are just inside the caps. Leave enough space for the pump to dispense any remaining liquid into the bottles.
When you are ready, click OK on the software. The pump will retract the piston to dispense any piston wash back into the bottle. Next, the pump will move the piston all the way forward to empty any remaining liquid into the Flush bottle. The Valve will move to the blocked position to prevent any flow from the bottles. Disconnect the Reagent lines from the pump head and plug the openings to prevent any spillage. Open the panel on the right side of the Pinnacle PCX by removing the captive screw located at the back of the Pinnacle PCX, then push the cover back, and lift it up.
TO REMOVE THE PUMP FROM PINNACLE PCX
Turn off Pinnacle PCX. Disconnect the electronic pin connector from the inside wall of the Pinnacle PCX. Remove the two Alan bolts from the bottom of the pump first. See Figure 5-11.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Figure 5-11
Page 83
Section 5
Ma in te na nc e
Note: It is very important to remove the bottom bolts first. There are specially designed hooks in the chassis that will support the pump.
Loosen and remove the third top screw. Use caution as this will cause the pump to lean toward you. Brace the pump toward the wall of the Pinnacle PCX to prevent damage. Carefully remove the pump from the chassis. If it is Pump 1, tilt the back end up and remove the pump in a vertical position. If it is Pump 2, tilt the back end down, and remove the pump in a vertical position. Place the pump on clean paper towels on the bench.
Wear gloves for this procedure.
REMOVE SYRINGE FROM BODY OF THE PUMP
Once the pump is on the bench, carefully remove the syringe from the body of the pump: Using a 3/32” Alan wrench, loosen the set screw (DO NOT remove set screw) at the front of the syringe. It is important to perform this step first, as it will remove the tension from the bracket pins and cap screws (figure 5-12).
Set Screw
Once the set screw is loose, using the 3/16” Alan wrench, loosen the two cap screws that secure the bracket to the pump base.
Figure 5-12
5.13
Cap Screw Bracket
Cylinder Cap
Remove the bracket and cylinder cap by gently pulling the bracket ends out and away from the pins.
Carefully remove the ceramic syringe body by turning it in a clockwise direction as you pull toward you (figure 5-13). Warning: Never turn ceramic syringe body counter clockwise. This will cause the Piston Head to unscrew and fall off inside of the cylinder, making it very difficult to remove.
There will be some liquid inside the body, which will spill onto the paper towels. Set the syringe body aside on a clean dry surface. The piston should be fully extended from the pump.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Figure 5-13
Page 84
5.14
Section 5
Ma in te na nc e
PISTON HEAD UNIT DISSASEMBLY AND
Metal Piston Face
SEAL REPLACEMENT
Note: All items to be replaced in the Pump Seal Replacement Procedure can be found in the the Pinnacle Pump Seal Kit (PN 1452-0122). In the following steps, you will replace 5 items: Brass-ended set screw Piston Head O-Ring Secondary Cylinder O-Ring Piston Rod O-Ring Three, 4 - 40 x 1/4" capnut screws and washers
Using a 1/16” Alan wrench (do not use ball wrench because it is too weak) remove the set screw from the Piston Head (figure 5-15).
Note: Very Important! Never re-use the set screw. It has a soft brass end that is designed to be used only once. A new set screw is included with the seal kit. Re-use of the set screw can cause the piston head to come loose.
Piston Head O-Ring
Set Screw
Secondary Cylinder O-Ring
PEEK Piston Head
Piston O-Ring Guide
Piston Rod O-Ring
Figure 5-14
O-Ring Guide
Pump Base
Lead Screw
Piston Head Unit
Figure 5-15
Remove the piston head by turning it in a counter-clockwise direction. It is threaded onto the leadscrew. If needed, use a cresent wrench to lossen the piston head.
Place the piston head on the bench with the PEEK side down. Using a Phillips Head screwdriver, remove the metal plate by removing the 3, 4 - 40 x 1/4" capnuts and washers. These cannot be reused and should be replaced. The piston head will disassemble into 3 parts:
1. Metal Piston Face
2. Piston Head O-ring
3. PEEK Piston Head
Blow out debris from brass-ended set screw that may be inside the piston head. Remove and discard the old O-ring. Sonicate the metal plate, PEEK piston head, and new O-rings in a beaker of water.
Note: Do not over tighten cap screws. Over tightening will strip the threads on the PEEK piston head.
Rinse with DI water and reassemble the Piston Head using the new O-ring and cap screws. Set the reassembled Piston Head Unit aside until Pump Reassembly.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 85
Section 5
Ma in te na nc e
SECONDARY O-RING REPLACEMENT
To replace the secondary O-ring at the base of the pump, first remove the piston wash connection and then “back piston” by turning the lead screw 4 to 5 full turns manually. Reverse direction of turn by 4 to 5 full turns by turning the lead screw in the opposite direction. This will extend the O-ring guide at the base of the pump (figure 5-16). Note: O-ring guide will not come out with piston wash connection still attached.
Slide the O-ring guide out and replce the small piston rod O-ring that is located at the back of the guide (figure 5-17). Slide the guide back to the base of the pump and replace the large secondary O-ring with the new one (figures 5-14 and 5-16).
Figure 5-16
5.15
Piston Wash Tubing
O-ring Guide
Placement of Secondary O-ring
PUMP REASSEMBLY
Screw the newly assembled Piston Head Unit onto the head screw (hand tight).
Using a new brass-ended setscrew, tighten the setscrew with the 1/16" Alan wrench. It is important to make this tight.
Figure 5-17
Lubricate the ceramic body with some methanol or isopropanol. Turning it in a clockwise direction, push it back over the piston, using care to keep it horizontal. Ensure that the through holes are orientated top and bottom (figure 5-18).
Place the cylinder cap onto the head of the cylinder and the bracket on the pins, but do not tighten the cap screws. Reference figure 5-12.
The set screw at front of cylinder has “adjusment” depression to center bracket. Center the bracket by sliding it back and forth a little bit until you feel it in the depression. Once it is centered, tighten the set screw. Tighten the set screw so that the syringe is tight against the base of the pump.
Figure 5-18
Finally, tighten the cap screws using the 3/16" alan wrench.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 86
5.16
Section 5
Ma in te na nc e
REPLACE THE PUMP IN THE PINNACLE PCX
Replace the pumps in the Pinnacle PCX by following the reverse of the directions in section titled “To Remove the Pump From Pinnacle PCX” on page 5.12.
Be sure that the body of the pump rests on the support brackets, and that the front of the pump is protruding through the front of the chassis.
Note: Be very careful not to pinch any of the other wires inside this area. Be especially careful of the reactor wires and the valve wires when removing Pump 1.
Attach the reagent and piston wash lines.
FLUSH AND REFILL PUMPS
Turn ON Pinnacle PCX. Select Control – Flush pump(s). Make sure all flush lines are immersed in solution. After seals have been exercised for at least 60 minutes, perform Pump and Valve Seal Pressure Test followed by a System Pressure Test. Follow the onscreen instructions.
Fuse Replacement
The fuse is on the back panel under the power switch. Warning: The Pinnacle PCX must be shutdown to replace the fuse. Remove the cord from the power inlet. Use a small flat screwdriver to pry up the fuse holder then pull the fuse out. Only use the correct type of fuse: 2 ea, 5mm x 20 mm, 5 amp, time lag Reinstall the fuse holder and the power cord.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 87
1100-0500 Tubing Kit, Pinnacle, Dual Pump
TRANSDUCER 2
R23
COLUMN
PUMP 2
R25
5.17
Section 5
Ma in te na nc e
R2
R24
1
2
W2
6
FLUSH
R1
S2
FILTER
FILTER
HEATED
REACTOR
PUMP 1
R13
3
VALVE 2
R21
R22
AMBIENT REACTOR
TRANSDUCER 1
R15
2
VALVE 1
3
R 11
R12
5
4
F2
R14
1
4
W1
6
5
F1
FROM
HPLC
WASTE
TO DETECTOR
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 88
5.18
Section 5
Ma in te na nc e
1100-0501 Tubing Kit, Pinnacle, Single Pump
COLUMN
R1
S2
HEATED
REACTOR
TRANSDUCER
R13
FILTER
R15
2
FLUSH
R14
W1
1
6
FROM
HPLC
Pinnacle Operators Manual
Pickering Laboratories Inc.
PUMP
R 11
3
VALVE 1
4
5
F1R12
WASTE
TO DETECTOR
Page 89
Section 6
appliCatiOns
6.1
Section 6
Ap pl ic At io ns
6.2
6.3
6.4
AMINO ACIDS CARBAMATES GLYPHOSATE
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 90
6.2
Section 6
Ap pl ic At io ns
AMINO ACIDS
6.2 Introduction
6.2-1 Background
6.2-2 Basic Sample Preparation
6.2-3 Reagent Preparation
6.2-4 Analytical and Post-column Conditions
6.2-5 Procedure
6.2-6 Sample Chromatograms and Gradient Programs
6.2-18 Precautions
6.2-21 Notes
Introduction
High performance liquid chromatography (HPLC) with post-column derivatization is a technique for rendering analytes more detectable than they would otherwise be in their native forms. Post-column derivatization gives improved sensitivity or better selectivity (reduction of interference) leading to lower detection limits. The Pickering Laboratories Pinnacle PCX was developed to facilitate the determination of amino acids in using sodium ion-exchange or in native samples using lithium ion-exchange columns. There are two options
®
for post-column detection of amino acids. The first is the use of Pickering’s patented TRIONE reagent, which will react with both primary and secondary amino acids. The second is the use of o-phthalaldehyde (OPA), a fluorescent reagent that gives greater sensitivity but will detect only primary amino acids.
A complete post-column analysis system for amino acids consists of the following components:
• HPLC ternary or greater gradient pump
®
• Manual injector or autosampler equipped with high pH compatible Tefzel
• Pickering Laboratories ion-exchange columns
• Pickering Pinnacle PCX post-column derivatization instrument
• Eluants, reagents, and standards
• Visible or fluorescence detector
• Chart recorder, integrator, or data system
Ion-exchange chromatography followed by post-column derivatization has been the method of choice for amino acid analysis since S. Moore, D.H. Spackman and W.H. Stein published it in 1958—work which merited a Nobel prize.
or PEEK™ seals
ninhydrin
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 91
Section 6
Ap pl ic At io ns
Background
The separation is a multi-modal process wherein ion-exchange, ion-exclusion, and partition all take place. The primary process is cation-exchange where a pH gradient mobilizes amino acids in order of their isoelectric points; acidic amino acids such as glutamic acid elute early and basic amino acids such as lysine elute late. Partitioning is affected by ionic strength and organic modifiers; for example threonine and serine are resolved by partition effects. Ion-exclusion only occurs for highly acidic amino acids such as taurine.
Sodium ion-exchange is used for fast analysis of the 22 amino acids found in hydrolyzed protein or in simple formulated products. Lithium ion-exchange is a slower technique with higher resolution to separate as many as 46 amino acids and compounds found in the complex mixtures of biological fluids or tissue extracts.
6.2-1
The most popular reagent for post-column detection is ninhydrin. Ninhydrin reacts with primary amines and hydrindantin to form Ruhemann’s Purple (Figure 6.2-A) which is detectable at 570nm. Ninhydrin reacts with secondary amines to form a yellow complex detectable at 440nm. The ninhydrin reaction is carried out at 130°C with a reactor volume of 500 µl. The elevated temperature is required because at room temperature, the ninhydrin reaction is very slow and takes hours to go to completion.
Figure 6.2-A
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 92
6.2-2
Section 6
Ap pl ic At io ns
An alternative reagent system based on o-phthalaldehyde (OPA) can be used for high-sensitivity detection of
primary amino acids. OPA reacts rapidly with primary amines and Thiofluor
mercaptoethylamine) under mild basic conditions to produce a strongly fluorescent isoindole derivative
(Figure 6.2-B). OPA does not react with secondary amines or aryl amines, so fails to detect Proline and other
secondary amino acids.
OPA
CHO
CHO
10 Amino Acid
OO
H
-
NH
R
+
3
pH>9, ambient T
(N,N-dimethyl-2-
CH
N
S
R
N
COOH
H
CH
3
3
HS
N
CH
Thiofluor
CH
3
3
Fluorescent Isoindole
Figure 6.2-B
However, it is possible to detect secondary amino acids by using a two-step reaction in which they are first
oxidized and then reacted with OPA. This technique has some disadvantages, and is not often used. Contact
Pickering for details.
The Pickering Pinnacle PCX derivatization instrument for fluorescent detection of amino acids is similarly
designed to the ninhydrin instrument, except that it contains a 150 µl reactor and the reacton is carried out
at 45°C.
Basic Sample Preparation
The following is a general sample preparation for Physiologic Fluid and Protein Hydrolysate samples. There are
many more types of samples that can be used for amino acid analysis. For details, consult the AOAC methods
and laboratory procedures.
Native amino acids are those found “free” in samples such as serum, urine and other physiological fluids, plant
extracts, foods and beverages. Although preparation of these samples for amino acid analysis is much simpler
and less time-consuming than protein hydrolysis, control of pH and normality, and removal of soluble protein
are critical factors which can affect the chromatography.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 93
Section 6
Ap pl ic At io ns
The early-eluting amino acids — taurine, urea, aspartic acid, threonine, serine, etc. — are particularly sensitive to pH and normality. Accordingly the samples must be held to a narrow pH range between 2.1 and
2.5, and the proper lithium ion concentration to ensure reproducibility in the early part of the chromatogram. The later-eluting compounds are more tolerant of initial sample conditions, and their retention times are not
as likely to be affected. SERAPREP as acetonitrile, perchloric acid and picric acid, and eliminate the need for dialysis, ultrafiltration, and repeated centrifugation steps, followed by pH adjustment.
• Filter all samples through a 0.45µm membrane filter. Some samples may require even more stringent filtration, especially if colloids are present.
• Samples must always be properly buffered. The ideal pH for sample injection is pH 2.3 ± 0.2.
• For native samples, be sure that all proteins have been removed before analysis.
and URIPREP™ replace commonly-used protein precipitation reagents such
6.2-3
PHYSIOLOGIC FLUID SAMPLES:
PREPARATION USING SERAPREP™ OR URIPREP
Use SERAPREP™ for preparing serum and other samples with a high buffering capacity, e.g. sardine oil. Use
URIPREP
for preparing urine and other samples with low buffering capacity, such as fruit juices, musts and warts. The efficiency of protein precipitation and the need for post-centrifugation pH adjustment of the sample determine which reagent is best for your particular sample.
1. In a microcentrifuge tube thoroughly mix equal portions of sample and SERAPREP
or URIPREP™.
2. Let stand for 5 minutes. Centrifuge the mixture at 13,000 rpm for 5 minutes. Check the supernate pH to ensure that the range is pH 2.3±0.2. Adjust the initial mixing ratio as necessary.
3. Filter the supernate with a syringe filter (0.2 or 0.45 um).The filtrate is ready to be injected into an auto­sampler vial for amino acid analysis.
4. If further dilution is needed, use Li 220 to adjust the concentration of analyte.
Reagent Preparation
TRIONE PREPARATION
TRIONE® reagent requires little to no preparation, depending on what type you use.
T100: The one-part TRIONE
into the reagent reservoir and put the cap on the reservoir.
T200: To prepare two-part TRIONE
reservoir, and cap tightly under Nitrogen. Swirl until homogeneous.
®
(Cat. No. T100C) requires no preparation - simply pour the TRIONE® directly
®
(Cat. No. T200), pour Bottle 1 into the reservoir, add Bottle 2 to the
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 94
6.2-4
Section 6
Ap pl ic At io ns
Note: TRIONE® is air sensitive, and must be kept under Nitrogen. The useful lifetime of T100 is three months* unopened, and one month in the reservoir. The shelf-life of T200 is one year* unmixed, and one month in the reservoir. *From date of manufacture.
OPA PREPARATION
1. Pour 945ml of the OPA Diluent (Cat.No. OD104) into the reagent reservoir. Save approximately 5 ml for Step 5.
2. Put the cap on the bottle, open the vent valve, and turn on the gas supply. Thoroughly deaerate the contents by sparging with inert gas. Continue bubbling for at least 10 minutes.
3. Dissolve 300mg of OPA (Cat. No. O120) in 10mL of HPLC-grade methanol in a clean, dry container.
4. Turn off the gas supply and remove the cap from the bottle. Add the OPA solution to the deoxygenated diluent in the reservoir. Wash any residual mixture into the reservoir with an additional 1–2 ml of methanol.
5. Dissolve 2g of Thiofluor™ (Cat. No. 3700-2000) in the reserved 5 ml of OPA Diluent and add to the reservoir.
®
6. Add 3ml of 30% Brij-35
7. Replace the cap and close the vent valve. Gently swirl the reagent to complete the mixing. Turn on the inert gas.
(Sigma) solution.
Note: OPA reagent is sensitive to air oxidation and will degrade over time. The Pinnacle PCX system is designed to minimize this oxidation. When the OPA reagent reservoir is maintained under inert gas pressure, the OPA reagent can maintain its activity for up to one week without significant loss of activity.
Post Column Conditions
These are the recommended post-column conditions for the most common methods of amino acid analysis. For the HPLC conditions, refer to the section titled Sample Chromatograms and Gradient Programs.
USING TRIONE NINHYDRIN REAGENT:
Reagent 1: TRIONE® (Cat. No. T100 or T200) Pump 1 Flow Rate: 0.30 ml/min Reactor 1 Volume: 500 µl Reactor 1 Temp: 130°C
USING OPA REAGENT:
Reagent 1: o-Phthalaldehyde and Thiofluor™ in OD104 Diluent, plus 35% Brij-35 Pump 1 Flow Rate: 0.30 ml/min Reactor 1 Volume: 150 µl Reactor 1 Temp: 45°C
®
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 95
USING SODIUM HYPOCHLORITE, FOLLOWED BY OPA REAGENT:
Reagent 1: 250 µl of 5% Sodium Hypochlorite in GA116 Diluent
Reagent 2: o-Phthalaldehyde and Thiofluor
in OD104 Diluent, plus 30% Brij-35 Pump 1 Flow Rate: 0.30 ml/min Pump 2 Flow Rate: 0.30 ml/min Reactor 1 Volume: 500 µl Reactor 2 Volume: 100 µl Reactor 1 Temp: 55°C Reactor 2 Temp: Ambient
Procedure
6.2-5
Section 6
Ap pl ic At io ns
®
Pickering Laboratories recommends different gradient conditions depending on the column and type of sample. Use the program recommended on the column data sheet for the initial testing. Do not change this program until you are sure that the other aspects of the system are functioning properly.
The column oven temperature programming gives additional flexibility when optimizing methods. Using temperature gradient allows to improve separation, shorten analysis time and fine-tune the method for detecting compounds of interest. Please refer to page 4.8 for details on how to set up timetable for the column oven.
Set the maximum pressure limit on the HPLC to 220 bar to protect the column. Allow the column to equilibrate for about 30 minutes under initial conditions. Inject 10µl of the Amino Acid Standard, and collect three chromatograms. The first chromatogram will not be representative of the systems performance, so use the second two to evaluate the performance.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 96
6.2-6
Section 6
Ap pl ic At io ns
Sample Chromatograms and Gradient Programs
METHOD 1: 0354675T 70-MIN PHYSIOLOGICAL FLUID LITHIUM (4.6 x 75mm) GUARD COLUMN: GARD COLUMN TEMP: SEE GRADIENT HPLC FLOW RATE: 0.55ml/min REAGENT FLOW RATE: 0.5ml/min INJECTION VOLUME: 10ul
HPLC PROGRAM
Time 1700-1125 % Li365 % Li375 % RG003 %
0 100 0 0 0 10 100 0 0 0 19 40 60 0 0 32 0 100 0 0 43 0 100 0 0
43.1 0 0 100 0 57 0 0 100 0
57.1 0 0 70 30 72 0 0 70 30
72.1 100 0 0 0
COLUMN OVEN PROGRAM
Time Temp ºC
0 34 6 34 17 65 25 70 70 70 71 34
1
20
23
15
17
16
18
21
24
2
3
456
0 10 20 30 40 50 70
8
7
11
44
10
9
12
19
25
22
26
27
13
:
14
30
29
28
31
45
36
37
39
35
34
32
33
38
40
42
41
60
PEAK IDENTIFICATION
1 Phosphoserine 2 Taurine 3 Phosphoethanolamine 4 Urea 5 Aspartic acid 6 Hydroxyproline 7 Threonine 8 Serine 9 Asparagine 10 Glutamic acid 11 Glutamine 12 Sarcosine
*Internal Standard
NOTE: This method utilizes column temperature gradient. Use Pinnacle PCX column oven or HPLC column oven with temperature gradient capabilities.
13 -Aminoadipic acid 14 Proline 15 Glycine 16 Alanine 17 Citrulline 18 -Amino-n-butyric acid 19 Valine 20 Cystine 21 Methionine 22 Allo-Isoleucine 23 Cystathionine 24 Isoleucine
25 Leucine 26 Tyrosine 27 Phenylalanine 28 -Alanine 29 -Amino-i-butyric acid 30 Homocystine 31 -Aminobutyric acid 32 Tryptophan 33 Ethanolamine 34 Ammonia 35 Hydroxylysines 36 Ornitine
37 Lysine 38 1-Methylhistidine 39 Histidine 40 3-Methylhistidine 41 Anserine 42 Carnosine 43 Arginine 44 Glucosaminic Acid* 45 2-Aminoethyl-cysteine (AEC)*
Pinnacle Operators Manual
Pickering Laboratories Inc.
43
min
Page 97
Section 6
Ap pl ic At io ns
METHOD 2: 0354100T ISOTHERMIC PHYSIOLOGICAL FLUID LITHIUM (4.0 x 100mm) GUARD COLUMN: GARD COLUMN TEMP: 37°C HPLC FLOW RATE: 0.35ml/min REAGENT FLOW RATE: 0.3ml/min INJECTION VOLUME: 10ul
CONDITIONS
Time (Min) Li275 % Li750 % RG003 % Comment
0 100 0 0 Inject 8 100 0 0 Isocratic 46 65 35 0 Linear Gradient 86 0 100 0 Linear Gradient 90 0 100 0 Isocratic 115 0 94 6 Linear Gradient 122 0 94 6 Isocratic
122.1 100 0 0 Step Change 140 100 0 0 Re-equilibration
6.2-7
2
1
3
20
8
5
7
4
6
17
18
16
19
10
15
9
12
13
11
14
PEAK IDENTIFICATION
1. Phosphoserine
2. Taurine
3. Phosphoethanolamine
4. Urea
5. Aspartic acid
6. Hydroxyproline
7. Threonine
8. Serine
9. Aspargine
10. Glutamic acid
11. Glutamine
12. Sarcosine
22
30
21
23
24
25
29
28
26
27
13. α-Aminoadipic acid
14. Proline
15. Glycine
16. Alanine
17. Citrulline
18. α-Amino-n-butyric acid
19. Valine
20. Cystine
21. Methionine
22. Cystathionine
23. Isoleucine
24. Leucine
44
40
39
41
31
42
37
38
35
34
33
32
36
43
45
12020 40 60 80 1000
25. Norleucine
26. Tyrosine
27. Phenylalanine
28. β-Alanine
29. β-Amino-i –butyric acid
30. Homocystine
31. γ-Aminobutyric acid
32. Tryptophan
33. Ethanolamine
34. Hydroxylysines
37. Ornithine
38. Lysine
39. Histidine
40. 3-Methylhistidine
41. 1-Methylhistidine
42. Carnosine
43. Anserine
44. α-Amino-β guanidinopropionic acid
45. Arginie
35. Ammonia
36. Creatinine
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 98
6.2-8
Section 6
Ap pl ic At io ns
METHOD 3: 0354100T TEMP GRADIENT PHYIOLOGICAL FLUID LITHIUM (4.0 x 100mm) GUARD COLUMN: GARD COLUMN TEMP: SEE GRADIENT HPLC FLOW RATE: 0.4ml/min REAGENT FLOW RATE: 0.3ml/min INJECTION VOLUME: 10ul
CONDITIONS
Time (Min) 1700-1125 % Li365 % Li375 % RG003 % Comment
0 100 0 0 0 Inject 15 100 0 0 0 Isocratic 27 40 60 0 0 Linear Gradient 45 0 100 0 0 Linear Gradient 60 0 100 0 0 Isocratic
60.1 0 0 100 0 Step Gradient 78 0 0 100 0 Isocratic
78.1 0 0 70 30 Step Gradient 95 0 0 70 30 Isocratic
95.1 100 0 0 0 Step Gradient 115 100 0 0 0 Equilibration
COLUMN OVEN PROGRAM
Time (Min) Temp. ºC
0 34 13 34 30 65 67 66 80 70 97 70 98 34
1
22
20
2
8
5
7
3
43
4
0 20 40 60
10
9
11
6
16
15
17
18
19
21
23
24
25
26
13
14
12
29
31
30
28
27
35
36
37
34
32
33
38
39
42
41
40
80
min
PEAK IDENTIFICATION
1 Phosphoserine 2 Taurine 3 Phosphoethanolamine 4 Urea 5 Aspartic acid 6 Hydroxyproline 7 Threonine 8 Serine 9 Asparagine 10 Glutamic acid 11 Glutamine
12 Sarcosine 13 a-Aminoadipic acid 14 Proline 15 Glycine 16 Alanine 17 Citrulline 18 α-Amino-n-butyric acid 19 Valine 20 Cystine 21 Methionine 22 Cystathionine
23 Isoleucine 24 Leucine 25 Tyrosine 26 Phenylalanine 27 β-Alanine 28 β-Amino-i-butyric acid 29 Homocystine 30 γ-Aminobutyric acid 31 Tryptophan 32 Ethanolamine 33 Ammonia
34 Hydroxylysines 35 Ornitine 36 Lysine 37 1-Methylhistidine 38 Histidine 39 3-Methylhistidine 40 Anserine 41 Carnosine 42 Arginine 43 Glucosaminic Acid*
*Internal Standard
NOTE: This method utilizes column temperature gradient. Use Pinnacle PCX column oven or HPLC column oven with temperature gradient capabilities.
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 99
Section 6
Ap pl ic At io ns
METHOD 4: 0393250 ISOTHERMIC PHYSIOLOGICAL FLUID LITHIUM (3.0 x 250mm) GUARD COLUMN: GARD COLUMN TEMP: 40°C HPLC FLOW RATE: 0.30ml/min REAGENT FLOW RATE: 0.3ml/min INJECTION VOLUME: 10ul
Time(min) %Li275 %Li750 %RG003 Comment
0 100 0 0 inject 17 100 0 0 isocratic 65 65 35 0 linear gradient 128 0 100 0 linear gradient 145 0 100 0 isocratic 185 0 94 6 linear gradient
185.1 100 0 0 step change 210 100 0 0 re-equilibration
6.2-9
1
2
3
22
20
5
8
7
10
9
4
6
020406080100 min120140 160180
18
17
16
19
23
15
13
12
11
14
24
21
30
25
26
27
31
29
28
35
33
34
32
36
40
39
41
42
37
38
44
43
PEAK IDENTIFICATION
1. Phosphoserine
2. Taurine
3. Phosphoethanolamine
4. Urea
5. Aspartic acid
6. Hydroxyproline
7. Threonine
8. Serine
9. Aspargine
10. Glutamic acid
11. Glutamine
12. Sarcosine
13. α-Aminoadipic acid
14. Proline
15. Glycine
16. Alanine
17. Citrulline
18. α-Amino-n-butyric acid
19. Valine
20. Cystine
21. Methionine
22. Cystathionine
23. Isoleucine
24. Leucine
25. Norleucine
26. Tyrosine
27. Phenylalanine
28. β-Alanine
29. β-Amino-i –butyric acid
30. Homocystine
31. γ-Aminobutyric acid
32. Tryptophan
33. Ethanolamine
34. Hydroxylysines
35. Ammonia
36. Creatinine
45
37. Ornithine
38. Lysine
39. Histidine
40. 3-Methylhistidine
41. 1-Methylhistidine
42. Carnosine
43. Anserine
44. α-Amino-β­ guanidinopropionic acid
45. Arginie
Pinnacle Operators Manual
Pickering Laboratories Inc.
Page 100
6.2-10
Section 6
Ap pl ic At io ns
METHOD 5: 0354100T PKU AND MSUD SCREENING LITHIUM (4.0 x 100mm) GUARD COLUMN: GARD COLUMN TEMP: 38°C HPLC FLOW RATE: 0.35ml/min REAGENT FLOW RATE: 0.3ml/min INJECTION VOLUME: 10ul
Time (min) %Li275 %Li750 %RG003 Comment
0 86 14 0 Inject 25 73 27 0 Linear gradient
25.1 0 0 100 Step gradient 30 0 0 100 Isocratic
30.1 86 14 0 Step gradient 42 86 14 0 Re-equilibration
a-Amino-n-Butyric Acid
Cystathionine
Valine
Methionine
Isoleucine
Leucine
Norleucine
Tyrosine
Phenylalanine
5
10 2015
25 min
Pinnacle Operators Manual
Pickering Laboratories Inc.
Loading...