LIFEPAK, FAST-PATCH, DERMA JEL, QUIK-LOOK, QUIK-COMBO, and LIFE•PATCH are registered trademarks of Physio-Control, Inc. ADAPTIV,
CODE-STAT, CODE SUMMARY, REDI-PAK, and Shock Advisory System are trademarks of Physio-Control, Inc. Masimo and LNOP are registered
trademarks of Masimo Corporation. EDGE System is a trademark of Ludlow Technical Products. Microsoft and Windows are registered trademarks
of Microsoft Corporation. Specifications are subject to change without notice.
About Automated External Defibrillationpage viii
About Defibrillation Therapyviii
About Noninvasive Pacingix
About SpO2 Monitoringx
About ECG Monitoringx
The following considerations and guidelines apply when using the LIFEPAK 20 defibrillator/monitor as
an automated external defibrillator (AED).
Operator Considerations
The LIFEPAK 20 defibrillator/monitor, when in AED mode, is a semiautomatic defibrillator that uses a
patented Shock Advisory System
(ECG) rhythm and indicates whether or not it detects a shockable rhythm. The LIFEPAK 20
defibrillator/monitor in AED mode requires operator interaction to defibrillate the patient.
The LIFEPAK 20 defibrillator/monitor in AED mode is intended for use by personnel who are
authorized by a physician/medical director and have, at a minimum, the following skills and training:
•CPR training.
•AED training equivalent to that recommended by the American Heart Association.
•Training in the use of the LIFEPAK 20 defibrillator/monitor in AED mode.
™
. This software algorithm analyzes the patient’s electrocardiographic
Indications
The AED mode is to be used only on patients in cardiopulmonary arrest. The patient must be
unconscious, pulseless, and not breathing normally before using the defibrillator to analyze the
patient’s ECG rhythm.
In AED mode, the LIFEPAK 20 defibrillator/monitor is not intended for use on pediatric patients less
than 8 years old.
Contraindications
None known.
ABOUT DEFIBRILLATION THERAPY
Operator Considerations
A direct current defibrillator applies a brief, intense pulse of electricity to the heart muscle. The
LIFEPAK 20 defibrillator/monitor delivers this energy through disposable electrodes, standard paddles
or internal paddles applied to the patient’s chest.
Defibrillation is only one aspect of the medical care required to resuscitate a patient with a shockable
ECG rhythm. Depending on the situation, other supportive measures may include:
•Cardiopulmonary resuscitation (CPR)
•Administration of supplemental oxygen
•Drug therapy
Successful resuscitation is related to the length of time between the onset of a heart rhythm that does
not circulate blood (ventricular fibrillation, pulseless ventricular tachycardia) and defibrillation. The
American Heart Association has identified the following as critical links in the chain of survival from
cardiac arrest:
The physiological state of the patient may affect the likelihood of successful defibrillation. Thus, failure
to resuscitate a patient is not a reliable indicator of defibrillator performance. Patients will often exhibit
a muscular response (such as jumping or twitching) during an energy transfer. The absence of such a
response is not a reliable indicator of actual energy delivery or device performance. For further
information, refer to the booklet, Defibrillation: What You Should Know.
Indications
Defibrillation is a recognized means of terminating certain potentially fatal arrhythmias, such as
ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of this energy in the
synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular
tachycardia, and, in relatively stable patients, ventricular tachycardia.
The biphasic defibrillation waveform used in this device has only been clinically tested on adults; it has
not been tested on pediatric patients.
Contraindications
Defibrillation is contraindicated in the treatment of Pulseless Electrical Activity (PEA) such as
idioventricular or ventricular escape rhythms, and in the treatment of asystole.
Preface
ABOUT NONINVASIVE PACING
A noninvasive pacemaker is a device that delivers an electrical stimulus to the heart, causing cardiac
depolarization and myocardial contraction. The energy is delivered through large adhesive electrodes
placed on the chest. In addition to noninvasive pacing, other supportive measures may be necessary.
Among other factors, it is recognized that successful pacing of a patient is related to the length of time
between the onset of a dysrhythmia and the initiation of pacing. Rapid pacing and prompt follow-up
care are essential. The physiologic state of the patient may affect the likelihood of successful pacing or
of skeletal muscle activity. The failure to successfully pace a patient is not a reliable indicator of
pacemaker performance. Similarly, the patient’s muscular response to pacing is not a reliable indicator
of energy delivered. Refer to the booklet, Noninvasive Pacing: What You Should Know for further
information.
Indications
Noninvasive pacing is indicated for symptomatic bradycardia in patients with a pulse.
Contraindications
Noninvasive pacing is contraindicated for the treatment of ventricular fibrillation and asystole.
A pulse oximeter is a noninvasive device that checks the saturation of oxygen in arterial blood (SpO2).
It uses an optical sensor that directs light through the patient’s finger and then measures the received
light with a detector. This received light is translated into a saturation percentage and is displayed as
an SpO2 reading.
Indications
The pulse oximeter is indicated for use in any patient who is at risk of developing hypoxemia.
Contraindications
None known.
ABOUT ECG MONITORING
The ECG (electrocardiogram) is a recording of the electrical activity of the heart. ECG monitoring
allows for identification and interpretation of cardiac rhythms or dysrhythmias and calculation of heart
rate. The ECG is obtained by placing either electrodes or paddles on the patient and allows the heart’s
electrical activity to be monitored and recorded.
This section provides important information to help you operate the LIFEPAK 20 defibrillator/monitor.
Familiarize yourself with all of these terms, warnings, and symbols.
Te r mspage 1-2
General Warnings and Cautions1-2
Symbols1-3
The following terms are used either in these Operating Instructions or on the LIFEPAK 20 defibrillator/
monitor:
Danger:Immediate hazards that will result in serious personal injury or death.
Warning:Hazards or unsafe practices that may result in serious personal injury or death.
Caution:Hazards or unsafe practices that may result in minor personal injury, product damage, or
property damage.
GENERAL WARNINGS AND CAUTIONS
The following are general warning and caution statements. Other specific warnings and cautions are
provided as needed in other sections of these operating instructions.
WARNINGS!
Shock hazard.
The defibrillator delivers up to 360 J of electrical energy. Unless properly used as described in these
Operating Instructions, this electrical energy may cause serious injury or death. Do not attempt to
operate this device unless thoroughly familiar with these operating instructions and the function of all
controls, indicators, connectors, and accessories.
Shock hazard.
Do not disassemble the defibrillator. It contains no operator serviceable components and dangerous
high voltages may be present. Contact authorized service personnel for repair.
Shock or fire hazard.
Do not immerse any portion of this defibrillator in water or other fluids. Avoid spilling any fluids on
defibrillator or accessories. Spilled liquids may cause the defibrillator and accessories to perform
inaccurately or fail. Do not clean with ketones or other flammable agents. Do not autoclave or sterilize
this defibrillator or accessories unless otherwise specified.
Possible fire or explosion.
Do not use this device in the presence of flammable gases or anesthetics. Use care when operating
this device close to oxygen sources (such as bag-valve-mask devices or ventilator tubing). Turn off
gas source or move source away from patient during defibrillation.
Possible electrical interference with device performance.
Equipment operating in close proximity may emit strong electromagnetic or radio frequency
interference (RFI) which could affect the performance of this device. RFI may result in improper device
operation, distorted ECG, failure to detect a shockable rhythm, or cessation of pacing. Avoid operating
the device near cauterizers, diathermy equipment, cellular phones, or other portable and mobile RF
communications equipment. Maintain equipment separation of at least 1.2 m (4 ft) and do not rapidly
key EMS radios on and off. Contact a technical support representative if assistance is required.
Possible electrical interference
Using cables, electrodes, or accessories not specified for use with this device may result in increased
emissions or decreased resistance to electromagnetic interference which could affect the
performance of this device or of equipment in close proximity. Use only parts and accessories
specified in these operating instructions.
This defibrillator may cause electromagnetic interference (EMI) especially during charge and energy
transfers. EMI may affect the performance of equipment operating in close proximity. Verify the
effects of defibrillator discharge on other equipment prior to using defibrillator in an emergency
situation, if possible.
Possible defibrillator shutdown.
When operating on battery power, the large current draw required for defibrillator changing may cause
the defibrillator to reach shutdown voltage levels with no low battery warning. If the defibrillator shuts
down without warning, or if a
LOW BATTERY: CONNECT TO AC POWER message appears on the
monitor screen, immediately connect the AC power cord to an outlet.
Possible improper device performance.
Changing factory default settings will change the behavior of the device. Changes to the default
settings must only be made by authorized personnel.
Possible improper device performance.
Using other manufacturers’ cables, electrodes, or batteries may cause the device to perform
improperly and invalidates the safety agency certification. Use only the accessories specified in these
Operating Instructions.
1 Safety Information
Possible failure to detect an out of range condition.
Reselecting QUICK SET will reset the alarm limits around the patient’s current vital sign values. This
may be outside the safe range for the patient.
Safety risk and possible equipment damage.
Monitors, defibrillators, and their accessories (including electrodes and cables) contain ferromagnetic
materials. As with all ferromagnetic equipment, these products must not be used in the presence of
the high magnetic field created by a Magnetic Resonance Imaging (MRI) device. The high magnetic
field created by an MRI device will attract the equipment with a force sufficient to cause death or
serious personal injury to persons between the equipment and the MRI device. This magnetic
attraction may also damage the equipment and affect the performance of the equipment. Skin burns
will also occur due to heating of electrically conductive materials such as patient leads and pulse
oximeter sensors. Consult the MRI manufacturer for more information.
SYMBOLS
The symbols below may be found in these operating instructions or on various configurations of
LIFEPAK 20 defibrillator/monitor and accessories:
Defibrillation-proof type CF terminal
Defibrillation protected, type BF patient connection
Recommended storage temperature: 5° to 45°C (41° to 113°F).
Storage at extreme temperatures of -20° or 60°C (-4° or 140°F) is
limited to seven days. If storage at these temperatures exceeds one
week, the electrode shelf-life is reduced.
Recycle this item
System connector/Data in
LIFEPAK 20 defibrillator/monitor to LIFEPAK 20 defibrillator/monitor
cable (Refer to Send Configuration Setup Menu, page 8-11)
Do not dispose of this product in the unsorted municipal waste stream.
Dispose of this product according to local regulations. See http://
recycling.medtronic.com for instructions on disposing of this product.
Federal law restricts this device to sale by or on the order of a
physician
The LIFEPAK 20 defibrillator/monitor is an acute cardiac care response system used by authorized
healthcare providers in hospital and clinic settings.
The LIFEPAK 20 defibrillator/monitor offers the following optional features:
•Semiautomatic defibrillator
•Noninvasive pacemaker
•Pulse oximeter
•Paddle accessories
Note: These operating instructions include information and procedures related to all features of
the LIFEPAK 20 defibrillator/monitor. Your LIFEPAK 20 defibrillator/monitor may not have all of
these features. For more information, contact your Physio-Control representative or call the
number listed on the back cover of these operating instructions.
The LIFEPAK 20 defibrillator/monitor is available only with the biphasic defibrillation waveform. For a
description of the defibrillation waveform, refer to Appendix A.
™
The LIFEPAK 20 defibrillator/monitor uses QUIK-COMBO
FAST-PATCH
®
disposable defibrillation/ECG electrodes for ECG monitoring and patient therapy. The
therapy cable connects the QUIK-COMBO or FAST-PATCH electrodes to the defibrillator. For more
information about QUIK-COMBO or FAST-PATCH electrodes, refer to Section 3 of these operating
instructions.
The standard paddle set is an accessory for the LIFEPAK 20 defibrillator/monitor and includes adult
and pediatric defibrillator (hard) paddles. The standard paddles can be used for QUIK-LOOK
monitoring, defibrillation, and synchronized cardioversion therapies. When using standard paddles, a
conductive interface designed for defibrillation, such as defibrillation gel or gel pads, must be used
between the paddle electrode surface and the skin.
The adult standard paddles can be used for any pediatric patient weighing approximately 10 kg (22 lb)
or more as long as the paddles fit completely on the chest and there is at least 2.5 cm (1 in.) of space
between the paddle electrodes. Pediatric paddles should be used for patients weighing 10 kg (22 lb)
or less or those whose chests are too small to accommodate the adult paddles.
Optional internal paddles are also available.
For more information about using paddle accessories, refer to Section 5 of these operating
instructions.
pacing/defibrillation/ECG electrodes or
®
ECG
UNPACKING AND INSPECTING
After you have removed the LIFEPAK 20 defibrillator/monitor from the shipping container, make sure
you have all the required supplies and accessories including cables and ECG paper. Examine the
defibrillator and all accessories for any sign of damage that may have occurred during shipping. If
possible, save the shipping container and foam inserts in case you have to ship the defibrillator at a
later date.
CONTROLS, INDICATORS, AND CONNECTORS
The following figures provide a brief description of the controls, indicators, and connectors for the
LIFEPAK 20 defibrillator/monitor. Figure 2-1 shows the front view of the LIFEPAK 20 defibrillator/
monitor and Figure 2-2 shows the front view divided into six areas. Figure 2-3 through Figure 2-12
show details of each area. Figure 2-13 shows the back view of the defibrillator. Additional information
about areas 3, 4, and 6 follow the applicable figures. The light emitting diode (LED) in a function button
is on when the corresponding function is active. For example, the
The door on the LIFEPAK 20 defibrillator/monitor hides the manual defibrillation and noninvasive
pacing buttons. When the door is closed, the appearance and operation of the device is simplified for
the automated external defibrillator (AED) user.
To enter manual mode, press the
MANUAL button located on the lower left corner of the door. This
opens the door and automatically takes the device out of AED mode and allows access to manual
mode defibrillation and pacing. After entering manual mode, closing the door does not affect
operation.
Sets the date and time.
For changes to take
effect, cycle power.
ARCHIVES
Accesses archived
patient records.
Refer to page 6-6.
USER TEST
Initiates automatic
self-test.
Refer to page 7-2.
ALARM VOLUME
Adjusts volume for
alarms, tones, and
voice prompts.
PRINT
Selects printer report,
format, and mode for
printing a current patient
report.
Generic is automatically selected when EVENT is pressed and no other selection is made. The selected
event and time stamp appear in the message/status area on the screen. Events are printed in the
™
CODE SUMMARY
Event Log. Refer to page 8-9 for information about configuring events.
Options
After pressing
OPTIONS, the screen displays the overlay shown in Figure 2-6. Use the Speed Dial to
scroll through and select menu choices.
Figure 2-6 Options
Alarms
Refer to page 2-14 for information about setting alarms.
Scrolls through and selects menu items.
LED illuminates when Speed Dial is active.
Refer to this page.
ECG CABLE CONNECTOR
Refer to page 3-5.
THERAPY CABLE
CONNECTOR
Refer to page 2-7.
SpO2 CABLE CONNECTOR
Refer to page 3-10.
SPEAKER
Area 4
IrDA PORT
Refer to page 6-11
4
Refer to warning, page 2-12.
Figure 2-7 Area 4
Area 4
The following paragraphs provide additional information about the Speed Dial and the therapy cable
connector shown in Area 4.
Speed Dial
The Speed Dial is active when the indicator LED is illuminated. When active, you can rotate the Speed
Dial to highlight and select certain areas of the screen and displayed menu items. Pressing the Speed
Dial activates the highlighted menu item. Default menu items are highlighted with a gray background;
after a menu item is selected, the background is black.
Therapy Cable Connector
WARNING!
Possible equipment damage and inability to deliver therapy.
To protect the therapy cable connector from damage or contamination, keep the therapy cable
connected to the defibrillator at all times.
Connecting the Therapy Cable
To connect a therapy cable to the therapy cable connector:
1 Orient the therapy cable so that the arrow is on top with the cable angled to the right (refer to
Figure 2-8).
2 Insert the therapy cable into the therapy cable connector on the defibrillator.
2 Basic Orientation
3 Rotate the locking ring on the therapy cable clockwise until you feel the connector “click.” Pull
gently on the locking ring to check that the cable is locked in place.
The following paragraphs provide additional information aboutArea 6.
Monitoring Area—Heart Rate
The LIFEPAK 20 defibrillator/monitor displays a heart rate between 20 and 300 beats per minute
(bpm). A heart rate symbol flashes with each beat. If the heart rate is below 20 bpm or pacing is
enabled, the screen displays dashes (– – –). Heart rates above 300 bpm do not yield valid systole
tones and the displayed heart rate will not be valid. The heart rate indicator is a tool to be used in
addition to patient assessment. Care should be taken to assess the patient at all times and not to rely
solely on the heart rate displayed.
Heart rate meters may continue to count the internal pacing pulses during occurrences of cardiac
arrest or some arrhythmias. Do not rely entirely on heart rate meter alarms. Keep pacemaker patients
under close surveillance.
QRS detection is essential for using the digital heart rate display, systole tone, synchronized
cardioversion, and noninvasive demand pacing. The QRS detector in the LIFEPAK 20 defibrillator/
monitor selectively detects QRS complexes. It discriminates against most noise, muscle artifact, Twaves, and other spurious signals.
The QRS detect algorithm automatically adjusts itself to the amplitude of the QRS complexes.
Changing the gain of the ECG has no effect on QRS detection. For optimum QRS detection
performance, use the lead with the greatest QRS amplitude.
Monitoring Area—Pulse Rate. If the ECG is not active, the SpO2 monitor can display pulse rate. The
pulse rate source is indicated by
PR (SPO2).
Monitoring Area—SpO2 (pulse oximeter). The oxygen saturation level is shown as a percentage
from 50 to 100. Saturation below 50% is shown as <50%. A fluctuating bar graph represents the pulse
signal strength.
Waveform Channel Area
Channel 1. This is the top channel. It displays the primary ECG waveform and is always visible when
ECG is displayed.
Channel 2. This is the bottom channel. It can display an additional waveform or a continuation of the
Channel 1 ECG.
Selecting Waveform Channels
The monitor power must be turned on.
1 At the home screen, rotate the Speed Dial
to highlight Channel 1 or 2.
2 Press the Speed Dial. An overlay appears
with the monitoring choices for the
selected channel.
3 Rotate and press the Speed Dial to select
monitoring choices for that channel.
Changing Printer Paper
CAUTION!
Possible printer malfunction.
Using other manufacturers’ printer paper may cause the printer to function improperly and/or damage
the print head. Use only the printer paper specified in these operating instructions.
The printer is equipped with an out-of-paper sensor to protect the printhead. The sensor automatically
turns off the printer if paper runs out or if the printer door is open.
To load the paper:
1 Press the black button to open the printer door.
2 Remove the empty paper roll.
3 Insert the new paper roll, grid facing upward.
4 Pull out a short length of paper.
5 Push the printer door in to close.
Figure 2-12 illustrates the steps for loading 50 mm paper.
The following paragraphs provide additional information aboutthe back view (refer to Figure 2-13).
Figure 2-13 Back View
WARNING!
Shock hazard.
If you are monitoring a patient and using the system connector, all equipment connected to the
system connector must be battery powered or electrically isolated from AC power according to
EN 60601-1. If in doubt, disconnect the patient from the monitor before using the system connector.
For more information, contact Physio-Control Technical Support.
System Connector
The system connector allows access to another LIFEPAK 20 defibrillator/monitor, so that setup
information can be transferred between devices.
ECG/SYNC Connector
The ECG/SYNC connector provides remote synchronization and real-time ECG output to a third party
monitor.
1 Select SEX.
2 Rotate the Speed Dial to highlight
FEMALE.
MALE or
3 Press the Speed Dial.
SETTING ALARMS
Alarms for the LIFEPAK 20 defibrillator/monitor can be configured to ON or OFF, and are enabled when
the monitor is turned on. When the alarms are configured
these limits, press the ALARMS button. The limits will appear to the right of the parameter value. To
change the limits, select
When the alarms are configured
QUICK SET.
OFF, the ALARMS button must be pressed and QUICK SET selected to
enable alarms.
When you press the
ALARMS button, the following Alarmsoverlay appears:
ON, predetermined limits are set. To view
1 Select QUICK SET to activate the alarms
for all active parameters. The quick set
limits are set automatically based on the
patient’s current vital sign values (refer to
Table 2-1). The alarm limits default to the
setting (
WIDE or NARROW) displayed on
the overlay.
2 Select LIMITS to change the alarm limits to
WIDE or NARROW (refer to Table 2-1).
3 Select SUSPEND to turn off the audible
alarm for up to 15 minutes. If an alarm
limit is exceeded while the alarm is
silenced, the violated parameter flashes,
an alarm message appears, but the alarm
tone remains silent.
monitoring for ventricular fibrillation and
ventricular tachycardia in manual mode.
A symbol appears above the primary
ECG when the alarm is on .
Reselect
VF/VT ALARM to turn off this
alarm.
Note: When the VF/VT alarm is
on, you are limited to paddles lead
or lead II. Refer to Selecting ECG
Lead and Size, page 3-2.
Note: The VF/VT alarm will be
suspended when the noninvasive
pacemaker is on and when standard
paddles are attached and
LEAD
is selected. The alarm is also
PADDLES
suspended when the device is
charging or is fully charged.
1
Narrow Limits
1
Limits Range
2
2 Basic Orientation
MANAGING ALARMS
The alarm bell symbol indicates when alarms are on or off . When alarms are on and an alarm
limit is exceeded, a tone sounds, the violated parameter flashes, and an alarm message appears.
To manage an alarm:
1 Press
2 Assess the cause of the alarm.
3 Assess the appropriateness of the limits setting (
4 If the patient is unstable, consider silencing the alarm for up to 15 minutes while attending to the
5 Once the patient is stable, reselect QUICK SET if necessary.
Possible failure to detect an out of range condition.
Reselecting QUICK SET will reset the alarm limits around the patient’s current vital sign values. This
may be outside the safe range for the patient.
When alarms are on, you can silence them preemptively for up to 15 minutes.
To silence alarms preemptively:
1 Press
ALARMS.
2 Select
SUSPEND.
3 Select a silence duration of 2, 5, 10, or 15 minutes.
4 The message ALARMS SUSPENDED appears on the bottom of the screen.
CONNECTING TO POWER
The LIFEPAK 20 defibrillator/monitor operates on AC (line) power or its internal battery. You can switch
from battery to AC power or AC power to battery while the device is on and in use by plugging in or
unplugging the AC power cord.
WARNING!
Possible defibrillator shutdown.
When operating on battery power, the large current draw required for defibrillator charging may cause
the defibrillator to reach shutdown voltage levels with no low battery warning, If the defibrillator shuts
down without warning, or if a LOW BATTERY: CONNECT TO AC POWERmessage appears on the
monitor screen, immediately connect the AC power cord to an outlet.
AC Operation
When the LIFEPAK 20 defibrillator/monitor operates on AC power, the AC Mains LED illuminates.
When the defibrillator is not in use, the battery charge is best maintained if the power cord is
connected to an AC outlet and the defibrillator is turned off.
Battery Operation
The internal nickel-metal hydride battery is rechargeable and intended to be used for standby
operation. The defibrillator automatically switches to battery power when the power cord is
disconnected from an AC outlet or from the defibrillator.
A new, fully charged battery provides approximately 90 360-joule discharges, 70 minutes of pacing, or
approximately 120 minutes of continuous monitoring before the defibrillator turns off. When the LOW BATTERY: CONNECT TO AC POWER message appears on the screen, immediately plug the power
cord into an AC outlet to continue use and begin recharging the battery. If low battery messages
frequently appear, the battery may need to be replaced. Contact Physio-Control Technical Service or
qualified service personnel for assistance.
Connect the defibrillator to AC power after each use to recharge the battery. Typically, new fully
depleted batteries recharge for 2.5 hours to regain full capacity. Partially depleted batteries recharge
for a time period equivalent to the time the defibrillator was in use. For example, if the defibrillator was
used one hour, the required recharge time will be approximately one hour.
New batteries or batteries that have been stored for an extended time should be recharged before
they are used. Connect the defibrillator to an AC power outlet for 2.5 hours to bring the battery to full
charge.
Battery Performance and Life
Several factors can contribute to the nickel-metal hydride battery’s performance and life cycle,
including:
1 The defibrillator’s use for assisting patients (“on time” and shocks).
2 The defibrillator’s use when the battery is at minimum capacity (low battery condition).
3 The battery’s normal self-discharge rate and the energy used by defibrillator self-tests.
To maximize battery performance and life, plug the defibrillator/monitor into an AC outlet to recharge
the battery whenever the defibrillator/monitor is not in use. As a reminder, you can set up the
LIFEPAK 20 defibrillator/monitor to sound a series of warning beeps, identified as AC LOSS ALERT,
whenever the defibrillator is turned off and not plugged into an AC outlet (refer to the General Setup
Menu, page 8-3).
Note: Even when properly maintained, the internal rechargeable nickel-metal hydride battery
should be replaced every two years.
The frequency response of the monitor screen is intended only for basic ECG rhythm identification; it
does not provide the resolution required for diagnostic and ST segment interpretation. For diagnostic or
ST segment interpretation, or to enhance internal pacemaker pulse visibility, attach the ECG cable.
Then print the ECG rhythm in diagnostic frequency response (DIAG).
Selecting ECG Lead and Size
There are two methods for selecting or changing the ECG lead. Both methods are available on your
LIFEPAK 20 defibrillator/monitor. The leads available depend on the ECG cable (3-wire or 5-wire)
connected to the device.
To change the ECG lead using the
Note: If one or more lead sets are preconfigured, the menu will display the lead sets. Refer to
page 8-8 for information about configuring lead sets.
LEAD button:
1 Press the LEAD button. If an ECG lead
appears, the lead automatically changes to
paddles. If paddles lead appears, the lead
automatically changes to lead II.
2 When the Lead menu appears, press the
LEAD button again or rotate the Speed Dial
to select another lead. The highlighted
selection shows the ECG lead.
Note: When the VF/VT alarm is on,
you are limited to paddles lead or
lead II in Channel 1. Refer to Setting
Anterior-lateral placement is the only placement that should be used for ECG monitoring with paddles
accessories.
1 Place either the ♥ or
midaxillary line, with the center of the electrode in the midaxillary line, if possible.
Figure 3-1.
+ therapy electrode or the apex paddle lateral to the patient’s left nipple in the
Refer to
Figure 3-1 Anterior-lateral Placement
2 Place the other therapy electrode or sternum paddle on the patient’s upper right torso, lateral to the
sternum and below the clavicle as shown in Figure 3-1.
Special Placement Situations
When placing therapy electrodes or standard paddles, be aware of the special requirements in the
following possible situations.
Obese Patients or Patients with Large Breasts
Apply therapy electrodes or standard paddles to a flat area on the chest, if possible. If skin folds or
breast tissue prevent good adhesion, it may be necessary to spread skin folds apart to create a flat
surface.
Thin Patients
Follow the contour of the ribs and spaces when pressing the therapy electrodes or standard paddles
onto the torso. This limits air spaces or gaps under the electrodes and promotes good skin contact.
Patients with Implanted Pacemakers
If possible, place therapy electrodes or standard paddles away from internal pacemaker generator.
Patients with Implanted Defibrillators
Apply therapy electrodes or standard paddles in the anterior-lateral position and treat this patient as
any other patient requiring emergency care.
Paddles Monitoring Procedure
To monitor using therapy electrodes or standard paddles:
1 Press
ON. Adjust contrast if necessary.
2 Prepare the patient’s skin:
• Remove excessive chest hair as much as possible. Avoid nicking or cutting the skin. If possible,
avoid placing therapy electrodes or standard paddles over broken skin.
• Clean and dry the skin.
• Do not use alcohol, tincture of benzoin, or antiperspirant to prep the skin.
3 Apply the therapy electrodes or standard paddles in the anterior-lateral position.
For therapy electrodes, confirm that the package is sealed and the Use By date has not passed.
For standard paddles, apply conductive gel over the entire electrode surface.
4 Connect the disposable therapy electrodes to the therapy cable.
5 Select paddles lead.
Monitoring with the Patient ECG Cable
There are two ECG cables available for ECG monitoring as shown in Figure 3-2: the 3-wire and 5-wire
cables.
Connecting the Patient ECG Cable
Connect the cable by inserting the main cable connector into the green electrically isolated ECG
connector on the monitor.
3 Monitoring
Figure 3-2 3-wire and 5-wire ECG Cables
ECG Monitoring Procedure
1 Press
ON. Adjust contrast if necessary.
2 Attach the ECG cable to the monitor.
3 Identify the appropriate electrode sites on the patient as shown in Figure 3-3.
Figure 3-3 Electrode Placement for ECG monitoring
4 Prepare the patient’s skin for electrode application:
• Shave excessive hair at electrode site. Avoid locating electrodes over tendons and major
muscle masses.
• Confirm package is sealed and Use By date has not passed.
• Attach an electrode to each of the lead wires.
• Grasp electrode tab and peel electrode from carrier.
• Inspect electrode gel and ensure the gel is intact (discard electrode if gel is not intact).
• Hold electrode taut with both hands. Apply the electrode flat to the skin. Smooth tape outwardly.
Avoid pressing the center of the electrode.
• Secure the trunk cable clasp to the patient’s clothing.
6 Select the lead on the monitor screen.
7 If necessary, adjust ECG size.
8 Press
PRINT to obtain an ECG printout.
ECG Electrode Requirements
Electrode quality is critical for obtaining an undistorted ECG signal. Always check the date code on
electrode packages for the Use By date before applying the electrodes to a patient. Do not use
electrodes with expired Use By date codes. Disposable electrodes are intended for a single use.
For best ECG monitoring results, use silver/silver chloride (Ag/AgCl) electrodes. The post-defibrillation
ECG will display in less time than expected with other types of electrodes.
Leads Off Messages
If an electrode or lead wire disconnects during ECG monitoring, the monitor emits an audible alarm
and displays a leads off message. The ECG trace becomes a dashed line
. The alarm and messages
continue until the electrode or lead wire is replaced.
Color Coding for ECG Leads
The lead wires and the electrode snaps for the patient ECG cable are color coded according to AHA or
IEC standards as listed in Table 3-1.
Table 3-1 ECG Leads Color Codes
LeadsAHA LabelAHA ColorIEC LabelIEC Color
Limb LeadsRAWhiteRRed
LABlackLYellow
RLGreenNBlack
LLRedFGreen
CBrownCBrown
Monitoring Patients with Internal Pacemakers
The LIFEPAK 20 defibrillator/monitor typically does not use internal pacemaker pulses to calculate the
heart rate. However, the monitor may detect internal pacemaker pulses as QRS complexes. This may
result in an inaccurate heart rate display.
Smaller amplitude internal pacemaker pulses may not be distinguished clearly. For improved detection
and visibility of internal pacemaker pulses, turn on the internal pacemaker detector, and/or connect the
ECG cable, select an ECG lead, and print the ECG in diagnostic frequency response.
Large amplitude pacemaker pulses may overload the QRS complex detector circuitry so that no paced
QRS complexes are counted. To help minimize ECG pickup of large unipolar pacemaker pulses when
monitoring patients with internal pacemakers, place ECG electrodes so the line between the positive
and negative electrodes is perpendicular to the line between the pacemaker generator and the heart.
The LIFEPAK 20 defibrillator/monitor annotates internal pacemaker pulses with a hollow arrow on
the display and the printed ECG if this feature is configured or selected
ON. False annotations of this
arrow may occur if ECG artifacts mimic internal pacer pulses. If false annotations occur, you may
deactivate the detection feature using the Options/Pacing/Internal Pacer menu (refer to Figure 2-6).
Also refer to the Pacing Setup Menu in Tab le 8 - 8. Patient history and other ECG waveform data, such
as wide QRS complexes, should be used to verify the presence of an internal pacemaker.
Troubleshooting Tips for ECG Monitoring
If problems occur while monitoring the ECG, check the list of observations in Table 3-2 for aid in
troubleshooting. For basic troubleshooting problems such as no power, refer to General
Troubleshooting Tips in Section 7.
Table 3-2 Troubleshooting Tips for ECG Monitoring
ObservationPossible CauseCorrective Action
1 Screen blank and ON
LED lighted.
2 Any of these messages
displayed:
CONNECT ELECTRODES
CONNECT ECG LEADS
ECG LEADS OFF
XX LEADS OFF
3 Poor ECG signal quality. Poor electrode-skin contact.• Reposition cable and/or lead wires to
Screen not functioning properly.• Print ECG on recorder as backup.
• Contact service personnel for repair.
Therapy electrodes are not
connected.
One or more ECG electrodes are
• Confirm therapy electrode
connections.
• Confirm ECG electrode connections.
disconnected.
ECG cable is not connected to
• Confirm ECG cable connections.
monitor.
Poor electrode-to-patient
adhesion.
• Reposition cable and/or lead wires to
prevent electrodes from pulling away
from patient.
• Prepare skin and replace electrode(s).
• Select another lead.
Broken ECG cable lead wire.• Select paddles lead and use standard
paddles or therapy electrodes for ECG
monitoring.
• Check ECG cable continuity.
prevent electrodes from pulling away
from patient. Secure trunk cable clasp
to patient’s clothing.
• Prepare skin and replace electrode(s).
Outdated, corroded, or dried-out
electrodes.
• Check date codes on electrode
packages.
• Use only silver/silver chloride
electrodes with Use By dates that have
not passed.
• Leave electrodes in sealed pouch until
time of use.
Loose connection.
Damaged cable or connector/lead
wire.
• Check/reconnect cable connections.
• Inspect ECG and therapy cables.
• Replace if damaged.
• Check cable with simulator and replace
if malfunction observed.
Noise because of radio frequency
interference (RFI).
• Check for equipment causing RFI
(such as a radio transmitter) and
relocate or turn off equipment power.
Table 3-2 Troubleshooting Tips for ECG Monitoring (Continued)
ObservationPossible CauseCorrective Action
4 Baseline wander
(low frequency/high
amplitude artifact).
Inadequate skin preparation.
Poor electrode-skin contact.
Diagnostic frequency response.
• Prepare skin and reapply electrodes.
• Check electrodes for proper adhesion.
• Print ECG in monitor frequency
response.
5 Fine baseline artifact
(high frequency/low
amplitude).
Inadequate skin preparation.
Isometric muscle tension in arms/
legs.
• Prepare skin and reapply electrodes.
• Confirm that limbs are resting on a
supportive surface.
• Check electrodes for proper adhesion.
6 Systole beeps not heard
or do not occur with
Volume too low.
QRS amplitude too small to detect.
• Adjust volume.
• Change ECG lead.
each QRS complex.
7 Monitor displays dashed
lines with no ECG leads
Paddles lead selected but patient
connected to ECG cable.
• Select one of the limb leads.
off messages.
8 Heart rate (HR) display
different than pulse rate.
Monitor is detecting the patient’s
internal pacemaker pulses.
• Change ECG lead.
• Change monitor lead to reduce internal
pacemaker pulse size.
9 Internal pacemaker
pulses difficult to see.
Pulses from pacemaker are very
small. Monitor the visibility of
frequency response limits.
• Turn on internal pacemaker detector
(refer to page 2-7).
• Connect ECG cable and select ECG
lead instead of paddles.
• Print ECG in diagnostic mode (refer to
page 2-7).
MONITORING SpO2
The following paragraphs describe:
•SpO2 Warnings and Cautions
•When to Use a Pulse Oximeter
•How a Pulse Oximeter Works
•SpO2 Monitoring Considerations
•SpO2 Monitoring Procedure
•Pulse Oximeter Sensors
•SpO2 Volume
•Troubleshooting Tips for SpO2
SpO2 Warnings and Cautions
WARNINGS!
Shock or burn hazard.
Before use, carefully read these operating instructions, the sensor and extension cable directions for
use, and precautionary information.
Shock or burn hazard.
Other manufacturers’ oxygen transducers may cause improper oximeter performance and invalidate
safety agency certification. Use only oxygen transducers approved for this product.
Do not use a damaged extension cable or one with exposed electrical contacts. Never use more than
one extension cable between the pulse oximeter and the sensor.
Inaccurate pulse oximeter readings.
Do not use a damaged sensor. Do not alter the sensor in any way. Alterations or modifications may
affect performance and/or accuracy.
Inaccurate pulse oximeter readings.
Sensors exposed to ambient light when incorrectly applied to a patient may exhibit inaccurate
saturation readings. Securely place the sensor on the patient and check the sensor’s application
frequently to help ensure accurate readings.
Inaccurate pulse oximeter readings.
Severe anemia, significant blood levels of carboxyhemoglobin or methemoglobin, intravascular dyes
that change usual blood pigmentation, excessive patient movement, venous pulsations,
electrosurgical interference, exposure to irradiation and placement of the sensor on an extremity that
has a blood pressure cuff, intravascular line or externally applied coloring (such as nail polish) may
interfere with oximeter performance. The operator should be thoroughly familiar with the operation of
the oximeter prior to use.
3 Monitoring
Inaccurate pulse oximeter readings.
The pulsations from intra-aortic balloon support can be additive to the pulse rate. Verify patient’s pulse
rate against the ECG heart rate.
Skin injury.
Prolonged, continuous use of a sensor may cause irritation, blistering, or pressure necrosis of the
skin. Check the sensor site regularly based on patient condition and type of sensor. Change the
sensor site if skin changes occur. Do not use tape to hold the sensor in place, as this may cause
inaccurate readings or damage to the sensor or skin.
Possible strangulation.
Carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.
CAUTION!
Possible equipment damage.
To avoid damaging the extension cable or the sensor, hold the connectors, rather than the cables,
when disconnecting.
When to Use a Pulse Oximeter
A pulse oximeter is a noninvasive tool that checks the saturation of oxygen in arterial blood (SpO2)
and is not to be used as an apnea monitor. It is used for monitoring patients who are at risk of
developing hypoxemia. If a pulse oximeter is not used, the only indications of hypoxemia are a
patient’s dusky skin, nail beds, and mucous membranes, accompanied by restlessness and confusion.
These indications are not conclusive, however, and do not appear until after the patient has developed
hypoxemia. Pulse oximetry is to be used in addition to patient assessment. Care should be taken to
assess the patient at all times and to not solely rely on the SpO2 reading. If a trend toward patient
deoxygenation is indicated, blood samples should be analyzed using laboratory instruments to
completely understand the patient’s condition.
A pulse oximeter sensor directs light through a fleshy body site (usually the finger or earlobe). The
sensor sends light from the emitting diodes to the receiving detector as shown in Figure 3-4. Oxygen
saturated blood absorbs light differently as compared to unsaturated blood. The pulse oximeter
translates the amount of light received into a saturation percentage and displays an SpO2 reading.
Normal values typically range from 95% to 100% at sea level.
Figure 3-4 How a Pulse Oximeter Works
The quality of the SpO2 reading depends on correct sensor size and placement, adequate blood flow
through the sensor site, patient motion, and exposure to ambient light. For example, with very low
perfusion at the monitored site, readings may read lower than core arterial oxygen saturation. Test
methods for accuracy are available by contacting your local Physio-Control representative.
SpO2 Monitoring Considerations
Each oximeter sensor is applied to a specific site on the patient. Use the following criteria to select the
appropriate sensor:
•Patient weight
•Patient perfusion to extremities
•Patient activity level
•Available application sites on the patient’s body
•Sterility requirements
•Anticipated duration of monitoring
To help ensure optimal and accurate performance:
•Use a dry and appropriately sized sensor.
•Keep the sensor site at the same level as the patient’s heart.
•Apply it according to the Directions for Use provided with the sensor.
•Observe all warnings and cautions noted in the sensor’s Directions for Use.
The sensors are sensitive to light. If excessive ambient light is present, cover the sensor site with an
opaque material to block the light. Failure to do so could result in inaccurate measurements.
If patient movements present a problem, consider the following possible solutions:
•Be sure the sensor is secure and properly aligned.
•Use a new sensor with fresh adhesive backing.
•If possible, move the sensor to a less active site.
Note: Wrapping the sensor too tightly or using supplemental tape to hold the sensor in place may
cause inaccurate measurements.
SpO2 Monitoring Procedure
The defibrillator controls power to the pulse oximeter. When the defibrillator is turned on, the oximeter
turns on and performs a self-test that requires up to 10 seconds. When the defibrillator is turned off,
the oximeter also turns off.
To conserve battery power, the pulse oximeter goes into “sleep mode” when not in use. Sleep mode is
activated within 10 seconds of disconnecting the sensor. During sleep mode, the screen does not
display SpO2 information. The oximeter returns to normal mode after detecting a sensor or a patient
signal. The oximeter performs the self-test when it returns from sleep mode to active mode.
The pulse oximeter measures SpO2 levels between 1% and 100%. When SpO2 levels are between
70% and 100%, oximeter measurements are accurate from ±3 digits. When the pulse oximeter
measures SpO2 levels less than 50%, the display shows < 50%.
3 Monitoring
To measure the patient’s SpO2 levels:
1 Connect the SpO2 cable to the monitor.
2 Attach the sensor to the SpO2 cable and the patient.
3 Press
ON.
4 Observe the pulse bar for fluctuation. Amplitude of the pulse bar indicates relative signal strength.
5 Adjust sensitivity, averaging time, and SpO2 volume as necessary.
SpO2 Waveform
The SpO2 waveform can be displayed on waveform Channel 2 by selecting waveform Channel 2 and
then selecting SpO2 from the Waveform menu. The SpO2 waveform automatically sizes itself to
provide optimum waveform viewing.
To adjust the pulse tone volume, highlight and select SPO2 on the home screen.
The following overlay appears:
1 Highlight and select SPO2 VOLUME.
2 Rotate the Speed Dial to the desired
volume.
3 Press the Speed Dial to set the volume.
Sensitivity
The sensitivity setting allows you to adjust the oximeter for differing perfusion states. To adjust the
sensitivity to either normal or high, highlight and select
SENSITIVITY.
The normal sensitivity setting is the recommended setting for most patients. The high sensitivity
setting allows for SpO2 monitoring under low perfusion states such as the severe hypotension of
shock. However, when the SpO2 sensitivity is set to high, the signal is more susceptible to artifact. It is
recommended that the patient be monitored closely when the high sensitivity setting is in use.
SPO2 on the home screen and then select
Averaging Time
The averaging time setting allows you to adjust the time period used to average the SpO2 value. Four
time periods are provided for averaging: 4, 8, 12, and 16 seconds. To adjust the averaging time,
highlight and select
The averaging time of 8 seconds is recommended for most patients. For patients with rapidly changing
SpO2 values, the 4-second time is recommended. The 12 and 16 second periods are used when
artifact is affecting the performance of the pulse oximeter.
SPO2 on the home screen and select AVERAGING TIME.
Pulse Oximeter Sensors
Refer to the LIFEPAK 20 Accessories Catalog for the sensors and extension cables to be used with the
LIFEPAK 20 defibrillator/monitor. Carefully read the directions for use provided with these sensors and
cables for complete descriptions, instructions, warnings, cautions, and specifications. To order sensors
and extension cables, contact your local Physio-Control representative
No Implied License
Possession or purchase of this oximeter does not convey any express or implied license to use the
oximeter with replacement parts which would, alone or in combination with the oximeter, fall within the
scope of one or more of the patents relating to this device.
Cleaning
To clean the sensors, first remove them from the patient and disconnect them from the connector
cable. Clean LNC and LNOP DCI sensors by wiping them with a 70% isopropyl alcohol pad. Allow the
sensors to dry before placing them on a patient. Do not attempt to sterilize.
To clean the connector and adapter cables, first remove them from the defibrillator and then wipe them
with a 70% isopropyl alcohol pad. Allow them to dry before reconnecting them to the defibrillator. Do
not attempt to sterilize.
Note: Do not soak or immerse the cables in any liquid solution.
Troubleshooting Tips for SpO2
Table 3-3 Troubleshooting Tips for SpO2
ObservationPossible CauseCorrective Action
1 The oximeter measures a
pulse, but there is no oxygen
saturation or pulse rate.
SpO2 or pulse rate changes
2
rapidly; pulse amplitude is
erratic.
SpO2 value is displayed.
3 No
SPO2: NO SENSOR DETECTED
4
message appears.
SPO2: CHECK SENSOR
5
message appears.
Excessive patient motion.• Keep patient still.
• Check that sensor is secure.
• Relocate sensor.
• Replace sensor.
Patient perfusion may be too low.• Check patient.
• Increase sensitivity.
Excessive patient motion.• Keep patient still.
• Check that sensor is secure.
• Relocate sensor.
• Replace sensor.
• Increase sensitivity.
An electrosurgical unit (ESU) may
be interfering with performance.
• Move the monitor as far as
possible from the ESU.
• Plug the ESU and monitor into
different circuits.
• Move the ESU ground pad as
close to the surgical site as
possible.
• Sensor may be damp, replace
it.
• Remove sensor extension
cable and connect the sensor
directly.
Oximeter may be performing a
• Wait for completion.
self-test (requires 10 seconds).
Sensor is not connected to patient
or cable disconnects from device.
Sensor is disconnected from
patient or cable.
• Check that sensor and cable
are connected properly.
• Attach the sensor.
• Check that sensor is secure.
Excessive ambient light.• Remove or block light source if
possible.
• Cover sensor with opaque
material, if necessary.
Patient has a weak pulse or low
blood pressure.
• Test sensor on someone else.
• Check if patient perfusion is
adequate for sensor location.
• Check if sensor is secure and
not too tight.
• Check that sensor is not on
extremity with blood pressure
cuff or intravascular line.
The defibrillator delivers up to 360 J of electrical energy. When discharging the defibrillator, do not
touch the paddle electrode surfaces or disposable therapy electrodes.
Shock hazard.
If a person is touching the patient, bed, or any conductive material in contact with the patient during
defibrillation, the delivered energy may be partially discharged through that person. Clear everyone
away from contact with the patient, bed, and other conductive material before discharging the
defibrillator.
Shock hazard.
Do not discharge the defibrillator into the open air. To remove an unwanted charge, change the
energy selection, select disarm, or turn off the defibrillator.
Possible fire, burns, and ineffective energy delivery.
Do not discharge standard paddles on top of therapy electrodes or ECG electrodes. Do not allow
standard paddles (or therapy electrodes) to touch each other, ECG electrodes, lead wires, dressings,
transdermal patches, etc. Such contact can cause electrical arcing and patient skin burns during
defibrillation and may divert defibrillating energy away from the heart muscle.
Possible defibrillator shutdown.
When operating on battery power, the large current draw required for defibrillator charging may cause
the defibrillator to reach a shutdown voltage level with no low battery warning. If the defibrillator shuts
down without warning, or if a LOW BATTERY: CONNECT TO AC POWER message appears on the monitor
screen, immediately connect the AC power cord to an outlet.
Possible skin burns.
During defibrillation or pacing, air pockets between the skin and therapy electrodes may cause patient
skin burns. Apply therapy electrodes so that entire electrode adheres to skin. Do not reposition the
electrodes once applied. If the position must be changed, remove and replace with new electrodes.
Possible skin burns and ineffective energy delivery.
Therapy electrodes that are dried out or damaged may cause electrical arcing and patient skin burns
during defibrillation. Do not use therapy electrodes that have been removed from foil package for more
than 24 hours. Do not use electrodes beyond expiration date. Check that electrode adhesive is intact
and undamaged. Replace therapy electrodes after 50 shocks.
Possible interference with implanted electrical device.
Defibrillation may cause implanted devices to malfunction. Place standard paddles or therapy
electrodes away from implanted devices if possible. Check implanted device function after
defibrillation.
CAUTION!
Possible equipment damage.
Prior to using this defibrillator, disconnect all equipment from the patient that is not defibrillatorprotected.
The following paragraphs describe therapy electrodes and standard paddles placement, including
special placement situations.
Anterior-lateral Placement
Anterior-lateral placement allows for ECG monitoring, defibrillation, synchronized cardioversion, and
noninvasive pacing.
1 Place either the ♥ or + therapy electrode, or apex paddle lateral to the patient's left nipple in the
midaxillary line, with the center of the electrode in the midaxillary line, if possible. Refer to
Figure 4-1.
4Therapy
Figure 4-1 Anterior-lateral Placement
2 Place the other therapy electrode or sternum paddle on the patient’s upper right torso, lateral to the
sternum, and below the clavicle as shown in Figure 4-1.
Anterior-posterior Placement
Anterior-posterior is an alternative position for noninvasive pacing, manual defibrillation, and
synchronized cardioversion, but not for ECG monitoring or automated defibrillation. The ECG signal
obtained through electrodes in this position is not a standard lead.
1 Place either the ♥ or + therapy electrode over the left precordium as shown in Figure 4-2. The
upper edge of the electrode should be below the nipple. Avoid placement over the nipple, the
diaphragm, or the bony prominence of the sternum if possible.
2 Place the other electrode behind the heart in the infrascapular area as shown in Figure 4-2. For
patient comfort, place the cable connection away from the spine. Do not place the electrode over
the bony prominences of the spine or scapula.
Figure 4-2 Anterior-posterior Placement for Noninvasive Pacing or Defibrillation
Special Placement Situations
When placing therapy electrodes or standard paddles, be aware of the special requirements in the
following possible situations.
Synchronized Cardioversion
Alternative anterior-posterior placements for cardioversion of supraventricular arrhythmias include:
•Place the ♥ or + therapy electrode over the left precordium and the other electrode on the patient’s
right posterior infrascapular area
– or –
•Place the ♥ or + therapy electrode to the right of the sternum and the other electrode on the
patient’s posterior left infrascapular area.
Obese Patients or Patients with Large Breasts
Apply therapy electrodes or standard paddles to a flat area on the chest, if possible. If skin folds or
breast tissue prevent good adhesion, it may be necessary to spread skin folds apart to create a flat
surface.
Thin Patients
Follow the contour of the ribs and spaces when pressing the therapy electrodes onto the torso. This
limits air spaces or gaps under the electrodes and promotes good skin contact.
Patients with Implanted Pacemakers
If possible, place therapy electrodes or standard paddles away from the internal pacemaker generator
to help prevent damage to the pacemaker. Treat the patient like any other patient requiring care. When
operating the defibrillator in AED mode, pacemaker pulses may prevent advisement of an appropriate
shock, regardless of the patient’s underlying rhythm.
Patients with Implanted Defibrillators
Apply therapy electrodes or standard paddles in the anterior-lateral position and treat this patient as
any other patient requiring emergency care. If defibrillation is unsuccessful, it may be necessary to try
alternate electrode placement (anterior-posterior) due to the insulative properties of implanted
defibrillator electrodes.
Do not analyze while patient is moving or being transported. Motion artifact may affect the ECG signal
resulting in an inappropriate shock or no shock advised message. Motion detection may delay
analysis. Stop motion and stand clear of patient during analysis.
Pediatric patient safety risk.
In AED mode, this defibrillator is not designed to administer energy at pediatric joule settings. The
American Heart Association recommends AEDs be used only on patients who are more than eight
years old.
AED Setup
You can setup the LIFEPAK 20 defibrillator/monitor to turn on in AED mode before placing the
defibrillator in use (refer to Section 8).
When illuminated, the AED mode LED indicates that the Continuous Patient Surveillance System
(CPSS) is active. CPSS automatically monitors the patient ECG for a potentially shockable rhythm.
When the
analysis system that advises the user if it detects a shockable or nonshockable rhythm.
The LIFEPAK 20 defibrillator/monitor can be setup to display the ECG waveform in AED mode or to
not display a waveform. The operation in AED mode remains the same whether or not the ECG
waveform is displayed.
ANALYZE button is pressed, the Shock Advisory System (SAS) is activated. SAS is an ECG
When the ECG waveform is set to ON in the
setup options (refer to Section 8), the ECG
appears with all of the AED messages and
prompts, as shown in the screen to the left.
When the ECG waveform is set to OFF in the
setup options, the messages and prompts fill
the screen as shown in the screen to the left.
AED Procedure
The following descriptions of voice prompts and messages are based on the factory default settings
for AED mode. The default settings are consistent with 2005 American Heart Association (AHA) and
European Resuscitation Council (ERC) guidelines. Changing the setup options may result in different
AED behavior. Refer to Section 8 for setup option choices.
1 Verify that the patient is in cardiopulmonary arrest (unresponsive, not breathing normally, and
showing no signs of circulation.
2 Press
3 Prepare the patient for electrode placement (refer to Paddles Monitoring Procedure, page 3-4).
ON.
The CONNECT ELECTRODES message and
voice prompt occur until the patient is
connected to the AED.
4 Connect the therapy electrodes to the therapy cable, and confirm cable connection to the
defibrillator.
5 Apply the electrodes to the patient's chest in the anterior-lateral position (refer to Anterior-lateral
Placement, page 4-3).
6 Press the
ANALYZE button to initiate analysis. Stop CPR.
The PUSH ANALYZE message and voice
prompt occur when the patient is properly
connected to the AED. The
message will stay on the screen and the
analyze LED flashes until
7 Follow screen messages and voice prompts provided by the AED.
You will see and hear ANALYZING NOW,
STAND CLEAR
patient or therapy cable during analysis.
ECG analysis requires approximately 6 to 9
seconds. The analyze LED illuminates
during analysis.
Therapy
. Do not touch or move the
The SAS analyzes the patient's ECG and advises either
Shock Advised
SHOCK ADVISED or NO SHOCK ADVISED.
If the AED detects a shockable ECG rhythm,
you will see and hear SHOCK ADVISED. The
AED begins charging to the setup joule
setting for shock #1. A rising tone indicates
that the AED is charging.
When charging is complete, the AED
displays the available energy.
You will see and hear STAND CLEAR, PUSH
TO SHOCK
() followed by a “shock ready”
tone. The shock LED flashes.
Clear everyone away from the patient,
bed, or any equipment connected to the
patient.
4Therapy
Press the button to discharge the AED.
Note: If you do not press the
button within 60 seconds, the AED
disarms the shock button, and the
DISARMING message appears.
When the button is pressed you will see
the message
When energy transfer is complete the shock
counter increases by 1. This will continue to
increase incrementally with every energy
transfer.
No Shock Advised
After a shock is delivered you will see and
hear START CPR. A countdown timer (min:sec
format) continues for the duration specified in
CPR TIME 1 setup option.
the
When the CPR countdown time ends, you
will see and hear
message stays on the screen and the voice
prompt will repeat every 20 seconds until you
press the
If the AED detects a nonshockable rhythm,
you will see and hear NO SHOCK ADVISED.
The AED will not charge, and a shock can
not be delivered.
ANALYZE button.
PUSH ANALYZE. This
After a NO SHOCK ADVISED prompt you will
see and hear
(min:sec format) continues for the duration
specified in the CPR TIME 2 setup option.
(Refer to Section 8).
When the CPR countdown time ends, you
will see and hear PUSH ANALYZE. This
message stays on the screen and the voice
prompt will repeat every 20 seconds until
you press the
If therapy electrodes are not connected to
the therapy cable or not placed on the
patient’s chest, the CONNECT ELECTRODES
message and voice prompt occur until the
patient is connected to the AED.
If the therapy cable is not connected to the
defibrillator, you will see the
CABLE
message until the cable is
CONNECT
connected.
If the test plug is connected to the therapy
cable when analysis is initiated, the REMOVE
TEST PLUG
message and voice prompt
occurs.
4Therapy
After you remove the test plug from the
therapy cable, the message and voice
prompt
If the motion is detected during the ECG
analysis, you will see and hear MOTION
DETECTED, STOP MOTION
followed by a
warning tone. Analysis is inhibited up to 10
seconds. After 10 seconds, even if motion is
still present, the analysis proceeds to
completion. Refer to Table 4-1 for possible
motion causes and suggested solutions.
When the AED is not analyzing the ECG or
is in CPR Time, it continuously monitors the
ECG for a potentially shockable rhythm
(CPSS).
If the AED detects a shockable rhythm, you
will see and hear
ANALYZE
is off.
PUSH ANALYZE, if AUTO
You should:
1 Confirm the patient is unconscious,
pulseless, and not breathing normally.
2 Confirm no motion is present. Stop CPR.
3 Press
ANALYZE.
The AED begins to analyze the patient’s
ECG.
For information about changing the
defibrillation mode, refer to Section 8.
Special AED Setup Options
The following descriptions of voice prompts and messages explain special cprMAX technology setup
options (refer to Appendix F).
Initial CPR - CPR First
When the
the AED is turned on.
INITIAL CPR option is set to CPR FIRST, you will be prompted to START CPR immediately after
After 3 seconds, a countdown timer
continues for the duration specified in the
initial CPR period and you will see and hear
IF YOU WITNESSED THE ARREST, PUSH
ANALYZE. This provides an opportunity to end
the initial CPR early and proceed to analysis.
•If you did witness the arrest, you should press ANALYZE and proceed directly to analysis. This will
end the CPR period and you will see and hear
Note: The decision to end CPR early is based on your hospital protocol.
•If you did not witness the arrest, you should perform CPR and not press
countdown timer continues for the duration specified in the
example, 90 seconds. When initial CPR time ends, you will see and hear
Initial CPR - Analyze First
When the
AED is turned on. CPR is prompted after the AED completes the analysis.
If the electrodes are not attached to the patient, you will see and hear
you are prompted to perform analysis.
INITIAL CPR option is set to ANALYZE FIRST, you will be prompted to perform analysis after the
ANALYZING NOW, STAND CLEAR.
ANALYZE. The Initial CPR
INITIAL CPR TIME setup option, for
PUSH ANALYZE.
CONNECT ELECTRODES before
4Therapy
No Shock Advised
If the AED detects a nonshockable rhythm,
you will be prompted to START CPR. A
countdown timer (min:sec format) continues
for the duration specified in the
setup option. (Refer to Section 8).
TIME
When initial CPR time ends, you will see and hear NO SHOCK ADVISED followed by PUSH ANALYZE.
Shock Advised
If the AED detects a shockable rhythm, you
will see and hear
YOU WITNESSED THE ARREST, PUSH
ANALYZE
end the initial CPR early and proceed
directly to delivering a shock.
Table 4-1 Troubleshooting Tips for AED Mode (Continued)
ObservationPossible CauseCorrective Action
3 MOTION DETECTED and STOP
MOTION
messages appear
during analysis.
Patient movement.• Stop CPR during analysis.
• When patient is being manually
ventilated, press
complete exhalation.
• Move patient to stable location
when possible.
Patient movement because of
agonal respirations.
• Press
ANALYZE immediately after
exhalation or wait until agonal
respirations are slower or
absent.
Electrical/radio frequency
interference.
• Move hand-held communication
devices or other suspected
devices away from the
defibrillator when possible.
DISARMING message appears. Electrode disconnected from
4
patient or AED.
Shock button not pressed within
60 seconds or door is open.
• Replace electrode and press
ANALYZE.
• Press
ANALYZE again.
• Press SHOCK immediately when
directed.
5 Voice prompts sound faint or
Low battery power.• Connect to AC power.
distorted.
LOW IMPEDANCE–
6
RECHARGING
message
Patient impedance <15 ohms
detected.
• No action required.
appears.
ANALYZE after
4Therapy
Switching from AED to Manual Mode
If the front console door is closed, you can enter manual mode by pressing the Manual button located
in the lower left corner of the door. This opens the door and automatically takes the defibrillator out of
AED mode, allowing you to access manual mode defibrillation and pacing.
Note: Closing the door again will not place the defibrillator in AED mode. Pressing Analyze while
the defibrillator is in manual mode will place the defibrillator in AED mode.
If the door is not installed, or to manually switch to manual mode, press one of the following keys:
ENERGY SELECT, CHARGE, PACER, LEAD.
Depending on the defibrillator’s configuration, continue to manual mode as follows:
•Direct. No restrictions to manual mode - access is immediate.
Refer to Section 8 for information about changing the defibrillation mode.
MANUAL DEFIBRILLATION
The following paragraphs describe:
•Manual Defibrillation Warnings
•Impedance
•Defibrillation Procedure
•Synchronized Cardioversion Procedure
•Remote Synchronization Procedure
Manual Defibrillation Warnings
WARNINGS!
Possible fire, burns, and ineffective energy delivery.
Precordial lead electrodes and lead wires may interfere with the placement of standard paddles or
therapy electrodes. Before defibrillation, remove any interfering precordial lead electrodes and lead
wires.
Shock hazard.
Conductive gel (wet or dry) on the paddle handles can allow the electrical energy to discharge through
the operator during defibrillation. Completely clean the paddle electrode surfaces, handles, and
storage area after defibrillation.
Possible patient skin burns.
During defibrillation, air pockets between the skin and standard paddles can cause patient skin burns.
Completely cover paddle electrode surfaces with fresh conductive gel and apply 11.3 kg (25lbs.) of
pressure per paddle during discharge.
Possible paddle damage and patient skin burns.
Discharging the defibrillator with the standard paddle surfaces shorted together can pit or damage the
paddle electrode surface. Pitted or damaged paddle surfaces may cause patient skin burns during
defibrillation. Discharge the defibrillator only as described in these Operating Instructions.
Possible burns and ineffective energy delivery.
A gel pathway on the skin between the standard paddles will cause defibrillating energy to arc
between paddles and divert energy away from the heart muscle. Do not allow conductive gel (wet or
dry) to become continuous between paddle sites.
Possible damage to defibrillator and defibrillator shutdown.
When used in conjunction with another defibrillator to deliver more than 360J, one or both
defibrillators may be damaged and shutdown may occur due to excessive currents. Avoid
simultaneous discharge from both defibrillators and maintain a backup device in case one or both
defibrillators shut down. If the defibrillator shuts down, take the device out of service and contact a
qualified service technician.
To use the LIFEPAK 20 defibrillator/monitor primarily as a manual defibrillator, configure the
defibrillator before placing the device in use. To configure the defibrillator, refer to Section 8.
Impedance
LIFEPAK biphasic defibrillators measure the patient's transthoracic impedance and automatically
adjust the defibrillation waveform voltage and current duration to meet the needs of the individual
patient.
Impedance is measured whenever the defibrillator is charged. To ensure correct patient impedance
readings, you should always charge the defibrillator when the standard paddles (hard paddles) or
QUIK-COMBO electrodes are in contact with the patient's chest.
If the standard paddles are in the paddle wells or touching face to face (shorted paddles) when the
defibrillator is charged, the defibrillator limits the available energy to 79 joules. This prevents damage
to the internal circuits, in the event the energy is discharged while the hard paddles are still in the
paddle wells or shorted together. If the defibrillator is charged to 80 joules or more and energy is
discharged when the hard paddles are seated in the paddle wells, the display indicates the selected
energy available and the automatic printout annotates time, date and Shock 79 J.
If the defibrillator is charged to 80 joules or more with paddles in the paddle wells, and then the
paddles are removed and placed on a patient, the defibrillator continues charging to the selected
energy and defibrillation may be completed as usual. When the defibrillator is charged with the
paddles on the patient’s chest, the defibrillator automatically adjusts the waveform voltage and current
duration based on the patient's impedance. It is important to understand that this automatic
adjustment does not occur when the defibrillator is charged with the paddles seated in the paddle
wells.
When the paddles or QUIK-COMBO electrodes are on the patient's chest when the defibrillator is
charged and the defibrillator measures an impedance of 15 ohms or less, the defibrillator disarms the
capacitor and automatically recharges to a lower energy setting. When this condition occurs, the
IMPEDANCE-RECHARGING
message appears on the display. When charging is complete, defibrillation
may be completed as usual.
LOW
4Therapy
Defibrillation Procedure
You can setup the LIFEPAK 20 defibrillator/monitor to automatically sequence energy levels. Refer to
Manual Mode Setup Menu, page 8-4.
1 Press
2 Identify the electrode or paddle sites on the patient. Use either the anterior-lateral or anterior-
3 Prepare the patient’s skin for electrode application:
4 Connect the therapy electrodes to the therapy cable, and confirm cable connection to the device.
• If possible, place the patient on a firm surface away from standing water or conductive material.
• Remove clothing from the patient’s upper torso.
• Remove excessive hair from the electrode sites; if shaving is necessary, avoid cutting the skin.
• Clean the skin and dry it briskly with a towel or gauze.
• Do not apply alcohol, tincture of benzoin, or antiperspirant to the skin.
Therapy
Note: When in paddles lead, if the REMOVE TEST PLUG message appears, disconnect test plug
and connect therapy electrodes to QUIK-COMBO therapy cable.
5 Apply therapy electrodes to the patient in anterior-lateral or anterior-posterior position. If using
standard paddles, apply conductive gel to the paddles and place paddles on the patient’s chest.
6 Press
7 Press
ENERGY SELECT.
CHARGE. While the defibrillator is charging, a charging bar appears and a ramping tone
sounds, indicating the charging energy level. When the defibrillator is fully charged, an overlay
appears (refer to Defibrillation Procedure, page 4-15).
8 Make certain all personnel, including the operator, stand clear of the patient, bed, and any
equipment connected to the patient.
9 Confirm ECG rhythms and available energy.
10 Press the
the charge. If the
SHOCK button(s) to discharge energy to the patient or press the Speed Dial to remove
SHOCK button(s) are not pressed within 60 seconds, stored energy is internally
removed.
Note: If you change the energy selection after charging has started, the energy is removed.
Press
CHARGE to restart charging.
11 Observe the patient and the ECG rhythm. If an additional shock is necessary, repeat the procedure
beginning at Step 6.
Note: If the
ABNORMAL ENERGY DELIVERY message appears and the shock is not effective,
increase energy, if necessary, and repeat shock. (Also refer to page 4-20.)
For more information about defibrillation, refer to the booklet, Defibrillation: What You Should Know.
Synchronized Cardioversion Procedure
Note: The LIFEPAK 20 defibrillator/monitor can be configured to remain in synchronous mode or
to return to asynchronous mode after discharge. It is important that you know how your defibrillator
is configured. Refer to Manual Mode Setup Menu, page 8-4.
1 Press
2 Attach patient ECG cable and ECG electrodes as described previously on page 3-5.
3 Select lead II or the lead with greatest QRS complex amplitude (positive or negative).
4 Press
5 Observe the ECG rhythm. Confirm that a triangle sense marker appears near the middle of each
ON.
Note: To monitor the ECG through therapy electrodes, place the electrodes in the anteriorlateral position and select paddles lead.
WARNING!
Possible lethal arrhythmia.
Ventricular fibrillation may be induced with improper synchronization. DO NOT use the ECG from
another monitor (slaving) to synchronize the LIFEPAK 20 defibrillator/monitor discharge. Always
monitor the patient’s ECG directly through the ECG cable, therapy cable or use the remote
synchronization procedure. Confirm proper placement of the sense markers on the ECG.
SYNC. Confirm the sync LED blinks with each detected QRS complex.
Note: Press
SYNC again to deactivate synchronous mode.
QRS complex. If the sense markers do not appear or are displayed in the wrong locations (for
example, on the T-wave), select another lead. (It is normal for the sense marker location to vary
slightly on each QRS complex.)
6 Prepare the patient’s skin for therapy electrode application as described in Step 3 of Defibrillation
Procedure, page 4-15.
7 Connect the therapy electrodes to the therapy cable, and confirm cable connection to the device.
8 Apply therapy electrodes to the patient in the anterior-lateral or anterior-posterior position. If using
standard paddles, apply conductive gel to the paddles and place paddles on the patient’s chest.
9 Press
10 Press
ENERGY SELECT.
CHARGE.
Note: If the
REMOVE TEST PLUG message appears, disconnect test plug and connect therapy
electrodes to QUIK-COMBO therapy cable and press
CHARGE.
11 Make certain all personnel, including operator, stand clear of the patient, bed, and any equipment
connected to the patient.
12 Confirm ECG rhythm. Confirm available energy.
13 Press and hold
release
SHOCK button(s). If SHOCK buttons are not pressed within 60 seconds, stored energy is
SHOCK button(s) until discharge occurs with next detected QRS complex and then
internally removed.
Note: If you change the energy selection after charging has started, the energy is removed
internally. Press
CHARGE to restart charging.
14 Observe patient and ECG rhythm. Repeat procedure from Step 4, if necessary.
4Therapy
Remote Synchronization Procedure
WARNINGS!
Possible lethal arrhythmia.
Ventricular fibrillation may be induced with improper synchronization. The hospital’s biomedical
engineering staff should perform synchronization delay measurements on the system, as a whole, to
ensure that the 60 ms limit for synchronization delay is not exceeded, per requirements as specified in
AAMI DF2 (1996). Always confirm proper placement of the sense markers on the ECG.
Possible monitor incompatibility
If the R-wave synchronization markers do not appear to be nearly simultaneous with the R-waves on
the remote monitor display or are not present, do not proceed with synchronized cardioversion.
The LIFEPAK 20 defibrillator/monitor can be configured to receive an ECG source from a remote
monitor (such as a bedside ECG monitoring system) by means of the ECG/SYNC connector on the
back of the defibrillator/monitor. Refer to Manual Mode Setup Menu, page 8-4. The remote monitor
must have a sync out connector and a cable must be provided to make this connection. Refer to the
LIFEPAK 20 Defibrillator/Monitor Service Manual for more details.
Note: The LIFEPAK 20 defibrillator/monitor can be configured to remain in synchronous mode or
to return to asynchronous mode after discharge. It is important that you know how your defibrillator
is configured. Refer to Manual Mode Setup Menu, page 8-4.
To perform the synchronized cardioversion using a remote monitoring ECG source:
1 Ensure defibrillator/monitor is connected to AC power.
2 Connect the sync cable to the defibrillator/monitor system connector and the remote monitor.
3 Press
4 Attach the ECG cable from the remote monitor to the patient.
Note: The screen on the defibrillator/monitor will display the message REMOTE SYNC in place of
any waveforms.
7 Observe the ECG rhythm on the remote monitor. Confirm that a sense marker appears above each
QRS complex.
8 Confirm that the sync LED on the defibrillator/monitor blinks with each detected QRS on the
remote monitor.
9 Follow steps 6 through 14 from Synchronized Cardioversion Procedure provided previously.
PEDIATRIC DEFIBRILLATION
Pediatric paddles are part of the standard paddle set (refer to page 5-6).
Pediatric Paddle Placement
Pediatric paddles should be used for patients weighing less than 10 kg (22 lb) or for patients whose
chest size cannot accommodate the adult therapy electrodes.
Adult paddles are recommended if the paddles will fit completely on the patient’s chest. Allow at least
2.5 cm (1 in.) of space between the paddles.
For neonates with very small chests, pediatric paddles may be too large to place in the anterior-lateral
position. In this situation, place paddles in the anterior-posterior position. Holding the paddles against
the chest and back will support the patient on his/her side.
Do not use the pediatric paddles on adults or older children. Delivery of recommended adult energies
through this relatively small electrode surface increases the possibility of skin burns.
Anterior-Lateral
The following is the standard pediatric paddle placement (refer to Figure 4-3):
•Sternum paddle to the patient’s right upper torso, lateral to the sternum and below the clavicle.
•Apex paddle lateral to the patient’s left nipple in the midaxillary line, with the center of the paddle in
the midaxillary line if possible.
Place the sternum paddle anteriorly over the left precordium and the apex paddle posteriorly behind
the heart in the infrascapular area (refer to Figure 4-4).
Figure 4-4 Anterior-posterior Paddle Position
Defibrillation Procedure
To defibrillate the patient:
1 Press
2 To access the pediatric paddles, slide the adult paddle forward until it releases.
3 Apply defibrillation gel to the pediatric paddle electrode surfaces.
4 Select the appropriate energy for the weight of the child according to American Heart Association
5 Place the paddles firmly on the patient's chest.
6 Press
7 Make certain all personnel, including the operator, are clear of the patient, the bed, and any
8 Confirm ECG rhythm and available energy.
9 Press the
Note: If the
increase energy, if necessary, and repeat shock. (Also refer to page 4-20.)
ON to turn on the defibrillator.
recommendations (or equivalent guidelines).
CHARGE.
equipment connected to the patient.
SHOCK button(s) to discharge energy to the patient or press the Speed Dial to remove
the charge. If
SHOCK buttons are not pressed within 60 seconds, stored energy is internally
removed.
Note: If you change the energy selection after charging has started, the energy is removed.
CHARGE to restart charging.
Press
ABNORMAL ENERGY DELIVERY message appears and the shock is not effective,
4Therapy
Troubleshooting Tips for Defibrillation and Synchronized Cardioversion
Tab l e 4 - 2 Troubleshooting Tips for Defibrillation and Synchronized Cardioversion
Tab l e 4 - 2 Troubleshooting Tips for Defibrillation and Synchronized Cardioversion (Continued)
ObservationPossible CauseCorrective Action
9 CONNECT ELECTRODES
message appears.
Inadequate connection to
defibrillator.
Electrodes do not adhere
properly to the patient.
• Check for electrode connection.
• Press electrodes firmly on
patient’s skin.
• Clean, shave, and dry the
patient’s skin as recommended.
Electrodes are dry, damaged, or
• Replace the electrodes.
out of date.
CONNECT TO AC POWER
10
message appears.
CONNECT SYNC CABLE TO
11
REMOTE MONITOR
message
appears.
LOW IMPEDANCE–RECHARGING
12
message appears.
SEARCHING FOR SIGNAL
13
message appears.
Remote sync is selected and the
device is not connected to AC
power.
Remote sync is selected and the
device is not connected to the
remote monitor.
Patient impedance of <15 ohms
detected.
Remote sync is selected and the
device is qualifying the input
• Connect to AC power.
• Press SYNC to turn off remote
sync.
• Connect to remote monitor.
• Press SYNC to turn off remote
synchronization.
• No action required.
• No action required.
signal.
4Therapy
NONINVASIVE PACING
The LIFEPAK 20 defibrillator/monitor provides noninvasive pacing using QUIK-COMBO electrodes.
The following paragraphs include:
•Noninvasive Pacing Warnings
•Demand and Nondemand Pacing
•Noninvasive Pacing Procedure
•Troubleshooting Tips for Noninvasive Pacing
For information about noninvasive pediatric pacing, refer to the Physio-Control Therapy Electrodes Operating Instructions MIN 3200346.
Noninvasive Pacing Warnings
WARNINGS!
Possible inducement of ventricular fibrillation.
Sinus bradycardia may be physiologic in severe hypothermia (i.e., appropriate to maintain sufficient
oxygen delivery when hypothermia is present), and cardiac pacing is usually not indicated.
Possible interruption of therapy.
Observe the patient continuously while the pacemaker is in use. Patient response to pacing therapy,
e.g., capture threshold, may change over time.
Possible inability to pace.
Using other manufacturers’ combination therapy electrodes with this device could cause a decrease in
pacing efficacy or the inability to pace because of unacceptably high impedance levels.
Prolonged noninvasive pacing may cause patient skin irritation and burns, especially with higher
pacing current levels. Discontinue noninvasive pacing if skin becomes burned and another method of
pacing is available.
Demand and Nondemand Pacing
The noninvasive pacemaker can be used for either demand (synchronous) or nondemand
(asynchronous) pacing modes.
The demand mode is used for most patients. In the demand mode, the LIFEPAK 20 defibrillator/
monitor/pacemaker inhibits pacing when it senses the patient’s own beats (intrinsic QRSs). In demand
mode, if the ECG amplitude is too low to detect the patient’s beats, or if an ECG lead becomes
detached so that the ECG rhythm is not present, the pacemaker generates pacing pulses
asynchronously. This means that the pacemaker generates pacing pulses at the selected rate,
regardless of the patient’s ECG rhythm.
Asynchronous or nondemand mode can be selected if noise or artifact interferes with proper sensing
of QRS complexes. Press the
The LIFEPAK 20 defibrillator/monitor has an integrated pulse oximeter that can be used in conjunction
with a noninvasive pacemaker to help confirm capture. To confirm capture, compare the pulse rate
measured by the oximeter to the set pacing rate of the pacemaker.
OPTIONS button to access the nondemand mode. (Refer to page 2-6.)
Noninvasive Pacing Procedure
ECG monitoring during pacing must be performed with the ECG electrodes and patient ECG cable.
Pacing therapy electrodes cannot be used to monitor ECG rhythm and deliver pacing current at the
same time. Be sure to place the therapy electrodes in the proper locations as described in the pacing
procedure. Improper electrode placement may make a difference in the capture threshold.
To pace, perform the following:
1 Press
2 Connect the patient ECG cable, apply ECG electrodes to the ECG cable and patient, and select
3 Identify the QUIK-COMBO electrode sites on the patient. For pacing, use either the anterior-lateral
4 Prepare patient's skin for electrode application as described in Step 3 of the Defibrillation
5 Apply QUIK-COMBO electrodes to the patient.
6 Connect the therapy electrodes to the therapy cable.
7 Press
8 Observe the ECG rhythm. Confirm that a triangle sense marker appears near the middle of each
9 Press
ON.
Lead I, II, or III. To receive the best monitoring signal, ensure there is adequate space between the
ECG electrodes and the therapy electrodes.
or anterior-posterior position (refer to page 4-3).
Procedure.
PACER. Confirm the LED illuminates, indicating that the power is on.
Note: If the
REMOVE TEST PLUG message appears, disconnect test plug and connect therapy
electrodes to QUIK-COMBO therapy cable.
QRS complex. If the sense markers do not appear or are displayed in the wrong location (for
example, on the T-wave)
, select another lead. (It is normal for the sense marker location to vary
slightly on each QRS complex.)
RATE or rotate the Speed Dial to select the desired pacing rate.
Note: The
RATE button changes the rate in 10 pulse per minute (ppm) increments; the Speed
10 Press CURRENT or rotate the Speed Dial to increase current until electrical capture occurs. For
each delivered pacing stimulus, the
PACER indicator flashes off and a positive pace marker displays
on the ECG waveform.
11 Palpate the patient's pulse or check blood pressure and compare the SpO2 pulse rate with the set
pacing rate to assess for mechanical capture. Consider use of sedation or analgesia if patient is
uncomfortable.
Note: The
CURRENT button changes the current in 10 mA increments; the Speed Dial changes
the current in 5 mA increments.
Note: To change rate or current during pacing, press
RATE or CURRENT, then rotate the Speed
Dial.
Note: To interrupt pacing and view the patient's intrinsic rhythm, press and hold the
button. This causes the pacer to pace at 25% of the set rate. Release the
PAUSE button to
PAUSE
resume pacing at the set rate.
12 To stop pacing, reduce current to zero or press
Note: To defibrillate and stop noninvasive pacing, press
PACER.
ENERGY SELECT or charge the
defibrillator. Pacing automatically stops. Proceed with defibrillation.
If the monitor detects ECG leads off during pacing, pacing continues at a fixed rate until the ECG lead
is reattached. During fixed-rate pacing, the pacemaker delivers pulses at the set pace rate regardless
of any intrinsic beats that the patient may have. The monitor continues to display the pacing rate (ppm)
and the current (mA). To reestablish demand pacing, reattach the ECG lead.
While pacing, visually monitor the patient at all times, do not rely on the
ECG LEADS OFF warning to
detect changes in pacing function. Routinely assess the ECG for proper sensing, pace pulse delivery,
electrical capture, and mechanical capture.
If pacing electrodes detach during pacing, the
CONNECT ELECTRODES and PACING STOPPED messages
appear and an alarm sounds. The pacing rate is maintained and the current resets to 0 mA.
Reattaching the pacing electrodes silences the alarm and removes the
CONNECT ELECTRODES
message. The current remains at 0 mA until you increase the current manually.
4Therapy
Troubleshooting Tips for Noninvasive Pacing
Tab l e 4 - 3 Troubleshooting Tips for Noninvasive Pacing
ObservationPossible CauseCorrective Action
1 Device does not function
when
PACER is pressed.
PACER LED on, but
2
CURRENT (MA) will not
increase.
PACER LED on,
3
CURRENT (MA) >0, but pace
markers absent (not pacing).
4 Pacing stops spontaneously.
Power off.• Check if power is
Low battery.• Connect to AC power.
Therapy electrodes off. • Check for message displayed.
• Inspect therapy cable and
electrode connections.
Pacing rate set below patient's
• Increase
PPM.
intrinsic rate.
Pacer oversensing (ECG artifact,
ECG size too high).
• Establish clean ECG; decrease
ECG size.
• Select nondemand pacing.
PACER button pressed off.• Press PACER and increase the
current.
Internal error detected. Service
message indicates an internal
failure.
There are two pre-gelled, self-adhesive therapy electrodes available: QUIK-COMBO pacing/
defibrillation/ECG electrodes and FAST-PATCH defibrillation/ECG electrodes (Figure 5-1).
QUIK-COMBO electrodes are used for defibrillation, synchronized cardioversion, ECG monitoring,
and pacing. FAST-PATCH electrodes can be used for defibrillation, synchronized cardioversion, and
ECG monitoring, but not for pacing. To use FAST-PATCH electrodes with the LIFEPAK 20 defibrillator/
monitor requires the addition of a FAST-PATCH defibrillation adapter cable (MIN 3011030).
Figure 5-1 QUIK-COMBO and FAST-PATCH Electrodes
A QUIK-COMBO or FAST-PATCH electrode set:
•Is a substitute for standard paddles.
•Provides a Lead II monitoring signal when placed in the anterior-lateral position.
•Quickly restores the ECG trace on the monitor following defibrillation.
To help prevent electrode damage:
•Do not fold the electrodes.
•Do not trim the electrodes.
•Do not crush, fold, or store the electrodes under heavy objects.
•Store electrodes in a cool, dry location (59° to 95°F or 15° to 35°C).
There are several types of QUIK-COMBO electrodes available as described in Table 5-1.
Table 5-1 QUIK-COMBO Electrodes
TypeDescription
QUIK-COMBOElectrodes, with .6 m (2 ft) of lead wire, designed for
patients weighing 15 kg (33 lb) or more.
QUIK-COMBO - RTSElectrodes, providing a radio-transparent electrode and
lead wire set, designed for patients weighing 15 kg (33 lb)
or more.
QUIK-COMBO with REDI-PAK
system
™
preconnect
Electrodes designed for patients weighing 15 kg (33 lb) or
more and allow preconnection of the electrode set to the
device while maintaining electrode shelf life and integrity.
Pediatric QUIK-COMBOElectrodes designed for patients weighing 15 kg (33 lb) or
less.
FAST-PATCH electrodes can be used on pediatric patients if the placement meets conditions noted in
the following paragraphs. Usually, these conditions can be met by patients weighing 15 kg (33 lb) or
more.
Electrode Placement
When using QUIK-COMBO or FAST-PATCH electrodes, ensure that the electrodes:
•Fit completely on the chest as described on page 3-4 or page 4-3.
•Have at least 2.5 cm (1 in.) of space between electrodes.
•Do not overlap bony prominences of sternum or spine.
To place the electrodes:
1 Prepare the patient for electrode placement:
• Remove all clothing from the patient’s chest.
• Remove excessive chest hair as much as possible. Avoid nicking or cutting the skin if using a
shaver or razor. If possible, avoid placing the electrodes over broken skin.
• Clean and dry the skin. If there is ointment on the patient’s chest, use soap and water to clean
the skin. Briskly wipe the skin dry with a towel or gauze. This mildly abrades the skin and
removes oils, dirt, and other debris for better electrode adhesion to the skin. Do not use alcohol,
tincture of benzoin, or antiperspirant to prepare the skin.
2 Slowly peel back the protective liner on the electrodes, beginning with the cable connection end
(refer to Figure 5-2).
Figure 5-2 Peeling the Liner from the Electrode
3 Place the electrodes in the anterior-lateral or anterior-posterior position, as described on page 3-4
or page 4-3, depending on the therapy to be provided and special placement considerations.
4 Starting from one edge, firmly press the electrode on the patient’s chest to eliminate air pockets
between the gel surface and the skin. Firmly press all adhesive edges to the skin.
Note: Once applied, therapy electrodes should not be repositioned.
To connect QUIK-COMBO electrodes to the QUIK-COMBO therapy cable:
1 Open the protective cover on the QUIK-COMBO therapy cable connector (refer to Figure 5-3).
2 Insert the QUIK-COMBO electrode connector into the therapy cable connector by aligning the
arrows and pressing the connectors firmly together for proper attachment.
Figure 5-3 Connecting QUIK-COMBO Electrodes to Therapy Cable
To properly connect FAST-PATCH electrodes to the FAST-PATCH defibrillation adapter cable and
ensure energy delivery:
1 Attach the defibrillation cable to the electrode post (before applying electrodes to the patient, when
possible).
2 Support the electrode post when attaching the defibrillation cable (refer to Figure 5-4). Firmly press
the cable onto the electrode until a click is heard or felt.
Figure 5-4 Connecting FAST-PATCH Electrodes to Defibrillation Cable
3 Pull up gently on the connector to confirm that the defibrillation cable is securely connected to the
electrode.
Note: If you are reattaching the defibrillation cable to an electrode that is already on the patient,
lift the adhesive edge under the electrode post slightly and place your finger under the post.
Connect the cable as described in the preceding steps.
ECG Monitoring and Therapy Procedures
WARNINGS!
Possible patient skin burns.
Do not use Pediatric QUIK-COMBO electrodes on adults or larger children. Delivery of defibrillation
energies equal to or greater than 100 J (typically used on adults) through these smaller electrodes
increases the possibility of skin burns.
Possible pediatric patient skin burns.
Noninvasive pacing may cause patient skin irritation and burns, especially with higher pacing current
levels. Inspect underlying skin of the ♥ electrode frequently after 30 minutes of continuous pacing.
Discontinue noninvasive pacing if skin burn develops and another method of pacing is available. On
cessation of pacing, immediately remove or replace electrodes with new ones.
For adult patients, follow the procedures for ECG monitoring, AED defibrillation, manual defibrillation,
synchronized cardioversion, and pacing described in Section 3 or Section 4.
For pediatric patients, follow the procedures for ECG monitoring, manual defibrillation, synchronized
cardioversion, and pacing except for the following:
•Select the appropriate defibrillation energy for the weight of the pediatric patient according to the
American Heart Association (AHA) recommendations or local protocol. Using energy levels of
100 J or greater is likely to cause burns.
•When pacing, frequently inspect the patient’s skin under the heart electrode for signs of burns.
Note: The amount of pacing current needed for capture is similar to the pacing current needed
for adults.
Replacing and Removing Electrodes
Replace QUIK-COMBO, QUIK-COMBO RTS, QUIK-COMBO REDI-PAK, or FAST-PATCH Plus
electrodes after 50 defibrillation shocks or 24 hours on the patient’s skin, or after 8 hours of
continuous pacing. Replace pediatric QUIK-COMBO electrodes after 25 defibrillation shocks or
24 hours on the patient’s skin, or after 8 hours of continuous pacing.
To remove QUIK-COMBO or FAST-PATCH electrodes from the patient:
1 Slowly peel back the electrode from the edge, supporting the skin as shown in Figure 5-5.
Figure 5-5 Removing Therapy Electrodes from Skin
2 Clean and dry the patient’s skin.
3 When replacing electrodes, adjust the electrode positions slightly to help prevent skin burns.
4 Close the protective cover on the QUIK-COMBO therapy cable connector when the cable is not in
use.
To disconnect the defibrillation cable from the FAST-PATCH electrodes:
1 Press down around the electrode post.
2 Pinch the snap connector with the fingers of the other hand and pull straight up (refer to
Figure 5-6).
Figure 5-6 Disconnecting Defibrillation Cable from FAST-PATCH Electrodes
WARNING!
Possible cable damage and ineffective energy delivery or loss of monitoring.
Improper disconnection of the defibrillation cable may damage the cable wires. This can result in
failure to deliver energy or loss of ECG signal during patient care. Position the cable so that it will not
be pulled, snagged, or tripped over. Do not disconnect the defibrillation cable snap connectors from
electrode posts or posts on testing devices by pulling on the cable. Disconnect the cable by pulling
each cable connector straight out (refer to Figure 5-7).
Figure 5-7 Disconnecting Defibrillation Cable from Test Post
Testing
As part of your defibrillator test routine, inspect and test the QUIK-COMBO therapy cable or FASTPATCH defibrillation adapter cable. Daily inspection and testing will help ensure that the defibrillator
and therapy cables are in good operating condition and are ready for use when needed.
Note: During the daily auto test and the user test, energy is discharged through the therapy
cable, testing the defibrillation component of the therapy cable. Performing the therapy cable test
on the Operator’s Checklist checks the sensing component of the therapy cable. Complete therapy
cable integrity is checked when both tests are performed.
If you detect any discrepancy during inspection and testing, remove the therapy cable from use and
immediately notify a qualified service technician.
Cleaning and Sterilizing
QUIK-COMBO and FAST-PATCH electrodes are not sterile or sterilizable. They are disposable and are
to be used for a single patient application. Do not autoclave, gas sterilize, immerse in fluids, or clean
electrodes with alcohol or solvents.
3 Slide the paddle handle forward until you hear a click. (Refer toFigure 5-11.)
Figure 5-11 Replacing a Pediatric Paddle
Each adult paddle attachment has a contact spring plate that transfers energy from the pediatric
paddle to the adult paddle. Routinely inspect the spring plates and pediatric paddle surfaces to make
sure that they are clean and intact.
Cleaning the Standard Paddle Set
Individually protect paddles before and after cleaning to prevent damage to paddle surfaces. After
each use:
1 Separate the adult and pediatric paddles.
2 Wipe or rinse paddle electrodes, cable connector, handles, and cables with mild soap and water or
disinfectant using a damp sponge, towel, or brush. Do not immerse or soak.
3 Dry all parts thoroughly.
4 Examine paddles (including electrode surfaces), cables, and connectors for damage or signs of
wear.
Note: Cables showing signs of wear, such as loose cable connections, exposed wires, or
cable connector corrosion, should be removed from use immediately.
Note: Paddles with rough or pitted electrodes should be removed from use immediately.
Internal handles with discharge control (Figure 5-12) are specifically designed for open chest cardiac
defibrillation and connect directly to the LIFEPAK 20 defibrillator/monitor.
Figure 5-12 Internal Handles with Discharge Control
Internal handles with discharge control are designed to be used only with internal paddles that have
the cam locking end as shown in Figure 5-13. No other paddles are compatible with these handles.
Figure 5-13 Internal Paddle
The internal paddles are available in the sizes listed in Ta bl e 5 - 2 :
1 Using a sterile technique, insert paddle fully into handle until a positive stop is reached.
2 Press and rotate the paddle as shown (clockwise) until a second stop is reached.
3 Release the paddle to lock in place. A correctly installed and locked paddle cannot be directly
withdrawn or rotated.
Removing the Paddles
To remove the paddles:
1 Push the paddle into the handle until a positive stop is reached.
2 Rotate the handle counterclockwise until a second stop is reached.
3 Slide the paddle out of the handle.
Internal Defibrillation Procedure
When internal handles are connected, energy selection is automatically limited to 50 J because of
possible cardiac damage from higher energies. To initiate internal defibrillation:
1 Press
2 Press
3 Press
ON. The Joules Selected symbol appears on the screen.
ENERGY SELECT if energy other than 10 J is desired.
CHARGE.
4 Place conductive surface of paddles against the right atrium and left ventricle.
5 Make certain that all personnel, including the operator, are clear of the patient, operating table or
bed, or any other equipment connected to the patient.
6 Press the discharge control located on the internal handle when the defibrillator has reached the
selected energy level. The defibrillator will not discharge until it completes charging to the selected
energy level. If discharge control is not pressed within 60 seconds, stored energy is removed
automatically.
7 Press the Speed Dial to manually remove an unwanted charge.
When internal handles are connected, the energy selection is automatically limited to 50 joules. To use
internal paddles for synchronized cardioversion:
1 Connect the internal paddles to the defibrillator.
2 Turn on the defibrillator, and then select Paddles lead.
3 Change the ECG size (gain) to the lowest setting, 0.25.
4 Select the desired energy setting.
5 Place the conductible surface of the paddles against the patient’s atrium and ventricle.
7 Confirm that a stable ECG signal is present and that triangle sense markers appear near the
middle of each QRS complex.
Note: The patient’s ECG acquired through internal paddles may be unreliable for
synchronized cardioversion due to excessive noise or artifact, causing inappropriate R-wave
detection. If the sense markers do not appear or are displayed in the wrong location (for
example, on the T-wave), acquire the patient’s ECG through standard ECG electrodes and
cable.
8 Press
CHARGE.
9 Make certain that all personnel, including the operator, are clear of the patient, operating table or
bed, or any other equipment that is connected to the patient.
10 When the defibrillator reaches the selected energy level, press and hold the discharge control
located on the internal handle. Discharge will occur with the next detected QRS complex.
11 Observe the patient’s ECG rhythm.
12 If necessary, repeat steps 4 through 11.
Handling Internal Paddles
Observe the following precautionary measures to avoid damage to the coating on internal paddles.
•Immediately following surgery and after removing the handle(s), cover each paddle to help protect
the paddles from impact to each other, other instruments, or hard surfaces.
•Use caution while handling the paddles during and after cleaning and before the sterilization
wrapping process.
•Inspect the paddles for chips and scratches after each use. If any damage is found, remove the
paddle(s) from use immediately.
•Ensure each paddle surface is protected from direct contact with the other while inside the
sterilization wrapping.
Cleaning and Sterilizing
Clean and sterilize this accessory according to the sterilization instructions provided with the
accessory.
Testing
Perform comprehensive electrical testing using a defibrillator analyzer no less than quarterly or after
ten sterilization cycles, whichever comes first.
The following paragraphs describe patient data storage and retrieval using the LIFEPAK 20
defibrillator/monitor.
Data Storage
When you turn on the LIFEPAK 20 defibrillator/monitor, you create a new Patient Record stamped with
the current date and time. All events and associated waveforms are digitally stored in the Patient
Record as patient reports, which you can print. When you turn off the device, the current Patient
Record data is saved in the patient archives.
To access the patient archives, press
records stored in the archived Patient Record. When you enter the archives mode, patient monitoring
ends and the current Patient Record is saved and closed. Turn off the device to exit the archives mode.
OPTIONS and select ARCHIVES. You can print or delete patient
Report Types
Patient reports within a Patient Record are stored as a CODE SUMMARY Critical Event Record, which
includes patient information, event and vital signs logs, and waveforms associated with events (for
example, defibrillation) as described on page 6-4.
Memory Capacity
The LIFEPAK 20 defibrillator/monitor retains data for two or more patients when you switch power off
or remove the batteries. The number of patient reports that the defibrillator can store depends on
various factors, including the number of displayed waveforms, the duration of each use, and the type of
therapy. Typically, memory capacity includes up to 100 single waveform reports. When the defibrillator
reaches the limits of its memory capacity, the defibrillator deletes an entire Patient Record using a “first
in, first out” priority to accommodate a new Patient Record. Deleted Patient Records cannot be
retrieved.
CODE SUMMARY REPORT
The LIFEPAK 20 defibrillator/monitor automatically stores a CODE SUMMARY report as part of the
Patient Record for each patient. The report consists of the following:
Figure 6-1 is an example of a CODE SUMMARY report. Press CODE SUMMARY to print the report.
6 Data Management
Figure 6-1 CODE SUMMARY Report
Preamble
The preamble contains patient information (name, event identification, patient identification, location,
age, and sex) and device information (date, time, and therapy information) as shown in Figure 6-1.
The event identification is a unique identifier that the defibrillator automatically enters in the ID field for
each Patient Report. This identifier is composed of the date and time that the defibrillator is turned on.
The location field allows you to enter up to 25 alpha-numeric characters to identify where the device
was used. You can link the data you enter to other patient information.
Event/Vital Signs Log
The LIFEPAK 20 defibrillator/monitor documents events and vital signs in chronological order. Events
are operator or device actions that are related to monitoring, pacing, AED therapy, data transmission,
and more. Table 6-1 shows a complete listing of events that can be found in the event log.
Vital signs (or active parameters) are entered into the log automatically every 5 minutes (or for each
event; refer to
You can configure the LIFEPAK 20 defibrillator/monitor to print a CODE SUMMARY report in one of
the formats described in Ta bl e 6 -3 . CODE SUMMARY reports are always stored in the medium
format.
Tab l e 6 - 3 CODE SUMMARY Formats
FormatAttributes
6 Data Management
Medium format• Preamble
• Event waveforms
• Event/vital signs log
Short format• Preamble• Event/vital signs log
The format determines only which reports are printed when the
CODE SUMMARY button is pressed. If
you interrupt printing of a CODE SUMMARY report, the entire CODE SUMMARY report will be
reprinted when printing is resumed.
CODE SUMMARY Complete is printed immediately following the last waveform event.
Refer to Figure 6-2 for examples of waveform data event printouts in the CODE SUMMARY report.
When you turn off the LIFEPAK 20 defibrillator/monitor, the current Patient Record is saved in the
archives. There are three options for managing archived Patient Records:
•Print archived patient reports
•Edit archived patient records
•Delete archived patient records
To perform any or all of these options, you must first enter the archives mode and then proceed with
the desired option.
ENTERING ARCHIVES MODE
To enter the archives mode:
1 Press OPTIONS.
2 Select ARCHIVES.
3 Select YES to enter the patient archives.
YES closes and saves the current
Patient Record and ends patient
monitoring.
–or–
Select
Note: If, after you select DELETE,
you decide you do not want to
remove the patient record,
immediately select
UNDO. If you
continue operations, you cannot
reverse the
6 Press HOME SCREEN and then turn off the
device.
DELETE selection.
OVERVIEW OF CONNECTIONS FOR TRANSMITTING REPORTS
Patient reports can be transmitted from the LIFEPAK 20 defibrillator/monitor to a PC-compatible
computer equipped with the Physio-Control CODE-STAT Suite, SYSTEM VIEW™, QUIK-VIEW™, or
CODE-STAT Data Review Software. Refer to the specific release of software for compatible Microsoft
Windows
An IrDA port, located on the front of the LIFEPAK 20 defibrillator/monitor (refer to page 2-7), supports
wireless, infrared communications for transmitting reports from the defibrillator to your computer. To
receive the transmission, your computer must have an operational IrDA port.
If your computer does not have an IrDA port, you can install an IrDA adapter to provide the needed
interface. Physio-Control recommends installing an IrDA adapter on all computers to ensure
successful communication connections and data transmissions.
IrDA adapters are available for serial or USB computer ports. Follow the installation and usage
instructions provided with the adapter, ensuring that the adapter mount (receiving end) is positioned on
a stable surface. Figure 6-3 provides guidelines to follow for positioning the defibrillator and the IrDA
adapter before initiating a transmission.
Note: The shaded cone in Figure 6-3 represents the approximate parameters for positioning the
defibrillator’s IrDA port opposite the IrDA adapter. As the distance between the two increases, so
does the possible range for aligning them.
You initiate and control transmission of device data at your computer using the Physio-Control CODESTAT Suite, SYSTEM VIEW™, QUIK-VIEW™, or CODE-STAT Data Review Software. This includes
entering patient information, selecting reports to be transmitted, and monitoring transmission
progress. More information about configuring the Data Review Software and instructions for
transmitting device data are provided in the User’s Guide and Reference Card that accompany your
Data Review Software.