LIFEPAK, LIFENET, QUIK-LOOK, and QUIK-COMBO are registered trademarks of Physio-Control, Inc. CODE SUMMARY, Shock Advisory
System, CODE-STAT, DT EXPRESS, LIFEPAK CR, REDI-PAK, LIFEPAK EXPRESS, SunVue, and cprMAX are trademarks of Physio-Control,
Inc. Bluetooth is a registered trademark of Bluetooth SIG, Inc. CADEX is a registered trademark of Cadex Electronics Inc. Microstream ,
CapnoLine, and FilterLine are registered trademarks of Oridion Systems Ltd. The Oridion medical capnography in this product is covered by
one or more of the following US patents: 6,428,483; 6,997,880; 5,300,859; 6,437,316 and their foreign equivalents. Additional patent
applications pending. PC Card is a trademark of the Personal Computer Memory Card International Association. Masimo, SET, Signal
Extraction Software, LNCS, and LNOP are registered trademarks of Masimo Corporation. Rainbow, SpCO, and SpMet are trademarks of
Masimo Corporation. EDGE System Technology is a trademark of Ludlow Technical Products. Formula 409 is a registered trademark of
The Clorox Company. Specifications are subject to change without notice.
The U.S. Food and Drug Administration requires defibrillator manufacturers and distributors to track the
location of their defibrillators. If the device is located somewhere other than the shipping address or the
device has been sold, donated, lost, stolen, exported, destroyed, permanently retired from use, or if the
device was not obtained directly from Physio-Control, please do one of the following: register the device at
http://www.physio-control.com, call the device tracking coordinator at 1.800.426.4448, or use one of the
postage-paid address change cards located in the back of this manual to update this vital tracking
information.
Text Conventions
Throughout these operating instructions, special text characters (for example, CAPITAL LETTERS such as CHECK
PATIENT and SPEED DIAL) are used to indicate labels, screen messages, and voice prompts.
The LIFEPAK 15 monitor/defibrillator is a complete acute cardiac care response system designed
for basic life support (BLS) and advanced life support (ALS) patient management protocols.
These operating instructions include information and procedures related to all features of the
LIFEPAK 15 monitor/defibrillator. Your LIFEPAK 15 monitor/defibrillator may not have all of these
features.
These operating instructions describe the operation of the LIFEPAK 15 monitor/defibrillator when
the factory default settings are used. The factory default settings for all setup options are identified
in Table A-5 on page A-14. Your device may be set up with different default settings, based on your
protocols. For information about changing default settings, see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device.
IMPORTANT! Some LIFEPAK 15 monitor/defibrillator accessories are not interchangeable with
accessories that are used with other LIFEPAK monitor/defibrillators. Specific accessory
incompatibilities are noted in the related sections.
1
Intended Use
The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in out-ofdoors and indoor emergency care settings within the environmental conditions specified on
page A-10. The LIFEPAK 15 monitor/defibrillator is designed to be used during ground
transportation except when specified otherwise.
Manual mode monitoring and therapy functions are intended for use on adult and pediatric
patients. Automated external defibrillation mode is intended for use on patients eight years of age
and older.
For additional intended use information, and information about the indications and
contraindications of the monitoring and therapy functions, see the individual sections identified
below.
• Setup mode—for changing default settings of the operating functions. For more information, see
the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device.
• Demo mode—for simulated waveforms and trend graphs for demonstration purposes. For more
information, see LIFEPAK 15 Monitor/Defibrillator Demo Mode at www.physio-control.com.
• Service mode—for authorized personnel to perform diagnostic tests and calibrations. For more
information, see the LIFEPAK 15 Monitor/Defibrillator Service Manual.
This chapter provides important information to help you operate the LIFEPAK 15 monitor/
defibrillator. Familiarize yourself with all of these terms and warnings.
The following terms are used either in these operating instructions or on the LIFEPAK 15 monitor/
defibrillator:
Danger: Immediate hazards that will result in serious personal injury or death.
2
Warning:
Caution: Hazards or unsafe practices that may result in minor personal injury, product damage, or
property damage.
Hazards or unsafe practices that may result in serious personal injury or death.
General Dangers and Warnings
The following are general danger and warning statements. Other specific warnings and cautions are
provided as needed in other sections of these operating instructions.
DANGER!
EXPLOSION HAZARD
Do not use this defibrillator in the presence of flammable gases or anesthetics.
WARNINGS
SHOCK OR FIRE HAZARDS
SHOCK HAZARD
The defibrillator delivers up to 360 joules of electrical energy. Unless properly used as
described in these operating instructions, this electrical energy may cause serious
injury or death. Do not attempt to operate this device unless thoroughly familiar with
these operating instructions and the function of all controls, indicators, connectors,
and accessories.
SHOCK HAZARD
Do not disassemble the defibrillator. It contains no operator serviceable components
and dangerous high voltages may be present. Contact authorized service personnel for
repair.
Do not immerse any portion of this defibrillator in water or other fluids. Avoid spilling
any fluids on defibrillator or accessories. Spilled liquids may cause the defibrillator
and accessories to perform inaccurately or fail. Do not clean with ketones or other
flammable agents. Do not autoclave or sterilize this defibrillator or accessories unless
otherwise specified.
POSSIBLE FIRE
Use care when operating this device close to oxygen sources (such as bag-valve-mask
devices or ventilator tubing). Turn off gas source or move source away from patient
during defibrillation.
ELECTRICAL INTERFERENCE HAZARDS
POSSIBLE ELECTRICAL INTERFERENCE WITH DEVICE PERFORMANCE
Equipment operating in close proximity may emit strong electromagnetic or radio
frequency interference (RFI), which could affect the performance of this device. RFI
may result in distorted ECG, incorrect ECG lead status, failure to detect a shockable
rhythm, cessation of pacing, or incorrect vital sign measurements. Avoid operating the
device near cauterizers, diathermy equipment, or other portable and mobile RF
communications equipment. Do not rapidly key EMS radios on and off. Refer to
Appendix D for recommended distances of equipment. Contact Physio-Control
Technical Support if assistance is required.
POSSIBLE ELECTRICAL INTERFERENCE
Using cables, electrodes, or accessories not specified for use with this defibrillator
may result in increased emissions or immunity from electromagnetic or radio
frequency interference (RFI) which could affect the performance of this defibrillator or
of equipment in close proximity. Use only parts and accessories specified in these
operating instructions.
POSSIBLE ELECTRICAL INTERFERENCE
This defibrillator may cause electromagnetic interference (EMI) especially during
charge and energy transfers. EMI may affect the performance of equipment operating
in close proximity. Verify the effects of defibrillator discharge on other equipment prior
to using the defibrillator in an emergency situation, if possible.
IMPROPER DEVICE PERFORMANCE HAZARDS
POSSIBLE IMPROPER DEVICE PERFORMANCE
Using other manufacturers’ cables, electrodes, or batteries may cause the device to
perform improperly and may invalidate the safety agency certifications. Use only the
accessories that are specified in these operating instructions.
Changing factory default settings will change the behavior of the device. Changes to
the default settings must only be made by authorized personnel.
POSSIBLE DEVICE SHUTDOWN
Always have immediate access to a spare, fully charged, properly maintained battery.
Replace the battery when the device displays a low battery warning.
SAFETY RISK AND POSSIBLE EQUIPMENT DAMAGE
Monitors, defibrillators, and their accessories (including electrodes and cables)
contain ferromagnetic materials. As with all ferromagnetic equipment, these products
must not be used in the presence of the high magnetic field created by a Magnetic
Resonance Imaging (MRI) device. The high magnetic field created by an MRI device
will attract the equipment with a force sufficient to cause death or serious personal
injury to persons between the equipment and the MRI device. This magnetic attraction
may also damage and affect the performance of the equipment. Skin burns will also
occur due to heating of electrically conductive materials such as patient leads and
pulse oximeter sensors. Consult the MRI manufacturer for more information.
2
Note: The LIFEPAK 15 monitor/defibrillator and its accessories that are intended for direct or
Disconnect: Rotate FilterLine connector counterclockwise and pull
connector out.
SpO2/
SpCO/
SpMet
Connect: Align cable connector with SpO
2
port and push in until
connector clicks into place.
Disconnect: Press the gray buttons on each side of the cable
connector simultaneously and pull connector out.
NIBP
Connect: Insert NIBP tubing connector into the NIBP port.
Disconnect: Press the latch on the left side of the port and pull
tubing connector out.
ECG
Connect: Align the green ECG connector with the ECG port;
position the white line on the cable facing left. Insert the cable
connector into the port until the connector is firmly seated.
Disconnect: Pull the ECG connector straight out.
P1/P2
Connect: Align the IP (invasive pressure) cable connector with the
P1 or P2 port; position the gap on the connector facing up. Insert
the cable connector into the port until the connector is firmly
seated.
Disconnect: Grip the connector and pull straight out.
1. Align the therapy cable connector with the
receptacle.
2. Slide the therapy cable until you feel the
connector lock in place. You will also hear a
“click.”
1. Press the release button on the therapy
cable connector.
2. Slide the therapy cable connector out.
Connecting and Disconnecting the Therapy Cable
WARNING
POSSIBLE EQUIPMENT DAMAGE AND INABILITY TO DELIVER THERAPY
To help protect the therapy cable connector from damage or contamination, keep
therapy cable connected to the defibrillator at all times. Inspect and test the therapy
cable daily according to the Operator’s Checklist in the back of this manual. PhysioControl recommends replacement of therapy cables every three years to reduce the
possibility of failure during patient use.
3
IMPORTANT!
(hard) paddles have the same type of connector and connect to the defibrillator at the same
location. These therapy cables are not compatible with other LIFEPAK defibrillator/monitors.
To connect a therapy cable to the defibrillator:
Figure 3-7 Connect Therapy Cable
To disconnect the therapy cable from the defibrillator:
The LIFEPAK 15 monitor/defibrillator QUIK-COMBO therapy cable and standard
Standard paddle wells,
retainers, and test
contacts
USB port coverProtects USB port from the environment.For future use
System connectorConnects device to a gateway or external
Each well holds one Lithium-ion battery. Two
pins in each well transfer the battery power.
Battery contacts transfer battery status
information.
monitoring EtCO
Paddle wells stow standard (hard) paddles.
Retainers provide secure retention and quick
removal of the paddles. Test contacts allow
complete paddles defibrillation checks
according to the Operator’s Checklist.
computer for transfer of patient reports. Also
provides real-time ECG output.
during use of anesthetics.
2
See page 9-12
See page 4-43
See page 6-5 and
Operator’s Checklist
the back of this manual
See page 7-3
WARNING
SHOCK HAZARD
If you are monitoring a patient and using the system connector, all equipment
connected to the system connector must be battery powered or electrically isolated
from AC power according to EN 60601-1. If in doubt, disconnect the patient from the
defibrillator before using the system connector. Only use Physio-Control recommended
data transmission cables. For more information, contact Physio-Control Technical
Support.
in
Note: To prevent inadvertent depletion of the defibrillator batteries, disconnect external devices
The LIFEPAK 15 monitor/defibrillator operates only on battery power using two Lithium-ion
batteries, which must be removed from the device and charged in the Station or Mobile Li-ion
Battery Charger.
IMPORTANT! The LIFEPAK 15 monitor/defibrillator Lithium-ion batteries are not interchangeable
with batteries that are used in other LIFEPAK defibrillators.
Routinely inspect batteries for damage or leakage. Recycle or discard damaged or leaking batteries.
Each battery has a fuel gauge that indicates the approximate charge level in the battery. Press the
gray button above the battery symbol to check the battery’s charge level prior to installing it in the
defibrillator. The four battery indicators shown here represent approximate charge—greater than
70%, greater than 50%, greater than 25%, and 25% or less, respectively.
Figure 3-10 Battery Charge Indicators
Battery warning indicators are shown below. A single flashing LED indicates that the battery is very
low and needs to be charged. Any two or more flashing LEDs indicate that the battery is faulty and
should be returned to your authorized service personnel.
Figure 3-11 Battery Warning Indicators
Note: Older or heavily used batteries lose charge capacity. If a battery fuel gauge indicates fewer
than four LEDs immediately after completing a charge cycle, the battery has reduced capacity. If
the battery fuel gauge shows two or fewer LEDs after the battery completes a charge cycle, the
battery should be replaced.
To install a battery:
1. Confirm that the battery is fully charged.
2. Inspect battery pins in the battery wells for signs of damage.
3. Align battery so battery clip is over the pins in the battery well.
4. Insert the end of the battery that is opposite the battery clip into the battery well.
5. Firmly press the clip end of the battery into the battery well until it clicks into place.
6. Repeat Step 1 through Step 5 to insert second battery.
To remove a battery, press the battery clip in and tilt the battery out of the battery well.
WARNING
POSSIBLE LOSS OF POWER DURING PATIENT CARE
Battery pins in the defibrillator may be damaged if batteries are dropped or forced into
battery wells. Inspect pins routinely for signs of damage. Keep batteries installed at all
times except when the device is removed from service for storage.
For information about battery maintenance, see "Battery Maintenance" on page 9-12.
The Home Screen is the main screen that displays ECG and other information. When a monitoring
cable is attached to the device, the corresponding monitoring area on the screen is activated and
the current patient values for that function are displayed. For example, when you connect an SpO
cable, the SpO
is connected. When the cable is disconnected, the SpO
area is activated on the screen. SpO2 values for the patient appear after the patient
2
patient values are replaced by dashes (--).
2
Separate controls do not activate the monitoring functions, except for NIBP.
Each vital sign monitoring area is colored to match its waveform. This color scheme aids in
associating the displayed waveform with its vital sign value. When a function does not have a
waveform displayed, the vital sign area is gray.
WARNINGS
FAILURE TO ACCURATELY COUNT HEART RATE
Patient heart rates above 300 bpm may not be counted accurately and may be
displayed as dashes (---), a value near 300, or a value that is approximately one-half
the actual patient heart rate. Increasing the ECG size to 2.0 or greater may improve
the accuracy of the displayed heart rate value. Do not rely solely on the displayed heart
rate for patient assessment. Use a printout of the ECG to calculate actual patient heart
rate.
Heart rate meters may count internal pacing pulses during cardiac arrest or some
arrhythmias. Do not rely entirely on heart rate meter alarms. Keep pacemaker patients
under close surveillance.
IMPORTANT! Set the high heart rate alarm for patients who have heart rates above 300 bpm.
• For patient heart rates of 20 to 317 bpm, the device consistently sounds the alarm when ECG
size is set to 1.0 or greater.
• For patient heart rates of 318 to 350 bpm, the device may intermittently silence the alarm for
up to five seconds.
Table 3-7 Home Screen
AREADESCRIPTIONFOR MORE INFORMATION
3
Alarm limitsLimits display along the right side of the
parameter.
Heart symbolFlashes with detected QRS signals.
Alarm indicatorIndicates whether alarms are on or silenced.
Absence of indicator means alarms are off.
Heart rateDevice accurately detects and displays heart rates
between 20 and 300 beats per minute (bpm). If
heart rate is below 20 bpm or pacing is active,
dashes (– – –) appear. If ECG is not active, the
or NIBP monitor can display pulse rate,
SpO
2
indicated by
patient’s heart rate is above 300 bpm, dashes (--) may appear or the displayed rate may be less
than the patient’s heart rate.
SpO2/SpCO/SpMetOxygen saturation level displays as a percentage
from 50 to 100. Saturation below 50% displays
as <50%. A fluctuating bar graph represents the
pulse signal strength. When available and
selected, the SpCO or SpMet value is displayed
as a percent for 10 seconds, and then the SpO
area reverts to the SpO
EtCO2End-tidal CO
kPa. Respiratory rate (RR) displays in breaths per
minute.
Use the SPEED DIAL to navigate around the Home Screen. As you rotate the SPEED DIAL, the individual
vital sign areas and waveform channels on the Home Screen are outlined. If you outline a vital sign
area or channel and then press the
SPEED DIAL, a menu appears.
3
For example, rotate the
following menu appears.
Whenever a menu is displayed, the ECG is always visible in Channel 1. To return to the Home
Screen from any menu, press the
Rotate and press the
SPEED DIAL to outline Channel 3, and then press the SPEED DIAL. The
The Home Screen displays battery indicators that show the following information about the
batteries installed in the defibrillator:
• Presence or absence of battery in battery well
• Battery in use
• Battery charge state
When two batteries are installed, the defibrillator uses the battery with the lowest level of charge
first. The battery in use is indicated by a white battery number in a black box. When a battery
reaches the replace battery state, the defibrillator automatically switches to the other battery.
Table 3-8 provides a description of the various battery status indicators.
Table 3-8 Battery Status Indicators
INDICATORMEANINGDESCRIPTION
The defibrillator is using the battery in well 1 for power. Battery status
Active battery
indicators display up to four green bars. Each green bar represents
approximately 25% remaining charge. For example, three green bars
indicate about 75% remaining charge.
Low battery
Very low
battery
Unrecognized
battery
No battery
installed or
fault
detected
Battery in well 1 is in use and is low. One yellow bar indicates 5% to
10% remaining charge.
Battery in well 1 is in use and is very low. One red flashing bar
indicates 0 to 5% remaining charge. The defibrillator automatically
switches to the other battery only if adequate charge is available. If
both batteries show red bars, the
occurs.
Battery in well 2 is not in use. Battery communication failed or a nonPhysio-Control battery is installed. The battery may power the
defibrillator but the level of charge is unknown and low battery
messages and prompts will not occur.
No battery is installed in battery well 1, or a fault was detected in the
battery in well 1 and the device will not use the battery.
REPLACE BATTERY voice prompt
Note: Older or heavily used batteries lose charge capacity. If a fully charged battery is installed in
the defibrillator and the battery status indicator shows less than four bars, the battery has reduced
capacity. If a battery status indicator shows only one or two bars after a fully charged battery is
installed, the battery has less than half the normal use time and should be recycled.
Select QUICK SET to activate the alarms for all
active monitoring functions.
The Quick Set
limits automatically set high and
low limits based on the patient’s current vital
sign values. For example, if the patient’s HR is
70, selecting
WIDE results in a high limit of
110 and a low limit of 45; selecting NARROW
results in a high limit of 100 and a low limit of
50. The default is
WIDE.
Alarms
Quick Set
Limits
Silence
VF/VT Alarm
Wide
2 Min
Off
Select LIMITS to change alarm limits to WIDE or
NARROW. See Table A-3 on page A-12.
Select SILENCE to turn off the audible alarm for
up to 15 minutes. If an alarm limit is exceeded
while the alarm is silenced, the violated vital
sign flashes and an alarm message appears,
but the alarm tone remains silent.
Alarms
Quick Set
Limits
Silence
VF/VT Alarm
Wide
Narrow
Alarms
LIFEPAK 15 monitor/defibrillator alarms can be set up to be ON or OFF when the defibrillator is
turned on. For more information, see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided
with your device.
When alarms are set up to be ON, default limits are set. The limits temporarily appear to the right of
the active vital signs. For all vital sign default alarm limits, see Table A-3 on page A-12.
3
If alarms are set up to be OFF, press
ON or are enabled by pressing
device. If power is lost for less than 30 seconds, for example due to a system reset or changing the
only active battery, alarm settings are restored automatically.
IMPORTANT! Set the high heart rate alarm for patients who have heart rates above 300 bpm.
• For patient heart rates of 20 to 317 bpm, the device consistently sounds the alarm when ECG
size is set to 1.0 or greater.
• For patient heart rates of 318 to 350 bpm, the device may intermittently silence the alarm for
up to five seconds.
ALARMS to enable the alarms. Whether alarms are set up to be
ALARMS, they can only be turned off by pressing ON to turn off the
Setting Alarms
When you press ALARMS, the following menu appears:
When an event is selected, the event and time
stamp appear in the message area on the Home
Screen.
Generic 12:20:30
Events
Use the Events menu to annotate patient events. A selected event appears in the Event log of the
CODE SUMMARY critical event record. Events can be customized in Setup mode. For more
information, see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device.
To select an event:
3
Notes:
• If you highlight an event but do not select it and the menu times out, a Generic event and time
stamp are annotated in the event log.
• If you highlight an event but do not select it and then press
time stamp are annotated in the event log.
The electrocardiogram (ECG) is a recording of the electrical activity of the heart. ECG monitoring
allows for identification and interpretation of cardiac rhythms or dysrhythmias and calculation of
heart rate. The ECG is obtained by placing either electrodes or paddles on the patient and allows
the heart’s electrical activity to be monitored and recorded.
ECG monitoring is a tool to be used in addition to patient assessment. Care should be taken to
assess the patient at all times; do not rely solely on the ECG monitor.
ECG Monitoring Warning
WARNINGS
POSSIBLE MISINTERPRETATION OF ECG DATA
The frequency response of the monitor screen is intended only for basic ECG rhythm
identification; it does not provide the resolution required for diagnostic and ST
segment interpretation. For diagnostic or ST segment interpretation, or to enhance
internal pacemaker pulse visibility, attach the multi-lead ECG cable. Then print the
ECG rhythm in diagnostic frequency response (DIAG) or obtain a 12-lead ECG.
4
FAILURE TO ACCURATELY COUNT HEART RATE
Patient heart rates above 300 bpm may not be counted accurately and may be
displayed as dashes (---), a value near 300, or a value that is approximately one-half
the actual patient heart rate. Increasing the ECG size to 2.0 or greater may improve
the accuracy of the displayed heart rate value. Do not rely solely on the displayed heart
rate for patient assessment. Use a printout of the ECG to calculate actual patient heart
rate.
IMPORTANT! Set the high heart rate alarm for patients who have heart rates above 300 bpm.
• For patient heart rates of 20 to 317 bpm, the device consistently sounds the alarm when ECG
size is set to 1.0 or greater.
• For patient heart rates of 318 to 350 bpm, the device may intermittently silence the alarm for
up to five seconds.
1. Press LEAD. If any ECG lead currently appears
on the Home Screen, the lead changes to
PADDLES. If PADDLES lead is currently
displayed, the lead changes to Lead
II.
2. While the
LEAD menu is displayed, press LEAD
again or rotate the
SPEED DIAL to the desired
lead.
Note: If lead sets are predefined for Channels 2
and 3, the lead sets show on the menu. The ECG
cable that is connected to the device, such as 3lead or 5-wire, determines the leads you can
select. For information about defining lead sets,
see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device.
Lead
III
AVR
AVL
AVF
II, CO2
I, II, III
AVR, AVL, AVF
Paddles
I
II
1. For the primary ECG, outline and select
CHANNEL 1 and then select LEAD.
2. Rotate the
SPEED DIAL to the desired ECG
lead.
3. Press the
SPEED DIAL to select the ECG lead.
4. Repeat this procedure to select or change
displayed ECG waveforms for Channels 2
and 3.
Channel 1
Lead
Size1.0
The LIFEPAK 15 monitor/defibrillator includes two methods for selecting or changing the ECG lead.
To select or change the displayed ECG lead using the
:
LEAD button:
To select or change the displayed ECG lead using the SPEED DIAL:
Note: The ECG shows dashed lines until the electrodes are connected to the patient.
Note: When the VF/VT ALARM is ON, you are limited to PADDLES lead or Lead II in Channel 1. See
To adjust the systole beep volume, use the SPEED DIAL to outline and select the HR area on the Home
Screen.
The following menu appears:
Note: The volume is reset to OFF each time the device is turned off.
Monitoring Using Paddle Accessories
To monitor ECG using paddles, you can use either QUIK-COMBO therapy electrodes or standard
(hard) paddles. For more information about paddle accessories, see Chapter 6, "Paddle Accessory
Options."
Anterior-Lateral Placement
Anterior-lateral placement is the only placement that should be used for ECG monitoring using
paddle accessories.
To place the therapy electrodes or paddles:
1. Place either the ♥ therapy electrode or
midaxillary line, with the center of the electrode in the midaxillary line, if possible (see
Figure 4-1).
Figure 4-1 Anterior-Lateral Placement
APEX paddle lateral to the patient’s left nipple in the
2. Place the other therapy electrode or STERNUM paddle on the patient’s upper right torso, lateral
to the sternum and below the clavicle as shown in Figure 4-1.
Special Situations for Electrode or Paddle Placement
When placing therapy electrodes or standard paddles, be aware of the special requirements in
the following possible situations:
Obese Patients or Patients with Large Breasts
Apply therapy electrodes or standard paddles to a flat area on the chest, if possible. If skin folds
or breast tissue prevent good adhesion, it may be necessary to spread skin folds apart to create
a flat surface.
Thin Patients
Follow the contour of the ribs and spaces when pressing the therapy electrodes or standard
paddles onto the torso. This limits air spaces or gaps under the electrodes and promotes good
skin contact.
Patients with Implanted Devices Such as Pacemakers or Defibrillators
4
If possible, place therapy electrodes or standard paddles away from implanted device.
Paddles ECG Monitoring Procedure
To monitor using standard paddles or therapy electrodes:
1. Press
2. Prepare the patient’s skin:
3. Apply the standard paddles or therapy electrodes in the anterior-lateral position. For therapy
4. Connect the therapy electrodes to the therapy cable.
ON.
• Remove all clothing from the patient’s chest.
• Remove excessive chest hair as much as possible. Avoid nicking or cutting the skin if
using a shaver or razor. If possible, avoid placing electrodes over broken skin.
• Clean and dry the skin, if necessary. Remove any medication patches and ointment on the
patient’s chest.
• Briskly wipe the skin dry with a towel or gauze. This mildly abrades the skin and removes
oils, dirt, and other debris for better electrode adhesion to the skin.
• Do not use alcohol, tincture of benzoin, or antiperspirant to prep the skin.
electrodes, confirm that the package is sealed and the Use By date is not passed. For
standard paddles, apply conductive gel over the entire electrode surface.
4. Prepare the patient’s skin for electrode application:
• Shave excessive hair at electrode site.
• For oily skin, clean skin with alcohol pad.
• Gently scrape skin to remove surface layer of dead cells and improve conduction of
electrical signals.
• Avoid locating electrodes over tendons and major muscle masses.
• Clean and dry the skin.
5. Apply ECG electrodes:
• Confirm that the package is sealed and the Use By date is not passed.
• Attach an electrode to each of the lead wires.
• Grasp electrode tab and peel electrode from carrier.
• Inspect electrode gel and make sure gel is intact (discard electrode if gel is not intact).
• Hold electrode taut with both hands. Apply the electrode flat to the skin. Smooth tape
outwardly. Avoid pressing the center of the electrode.
4
• Secure the trunk cable clasp to the patient’s clothing.
Note: Electrode quality is critical for obtaining an undistorted ECG signal. Always check the
date code on electrode packages for expiration date before using on a patient. Do not use
electrodes that have expired. Disposable electrodes are intended for a single use.
6. Select the desired ECG lead on the monitor screen.
7. If necessary, adjust ECG size for accurate heart rate counting.
8. Press
PRINT to obtain an ECG printout.
Precordial Lead ECG Monitoring
The precordial (chest) leads (see ECG Leads Color Codes on page 4-10) can be used for
monitoring when using the 12-lead cable or 5-wire cable.
To perform precordial lead ECG monitoring:
1. Insert the precordial lead attachment into the main cable as shown in Figure 4-2 on
page 4-8.
2. Place the precordial lead electrodes on the chest as described in the 12-lead ECG procedure
and shown in Figure 4-5 on page 4-15.
Note: When using a 5-wire cable, attach the limb leads as described in "ECG Monitoring
Procedure" on page 4-8, and place the C-lead electrode on the chest in the precordial position
desired. Note that the LIFEPAK 15 monitor labels the ECG for this lead as V1 on the screen and
printout, regardless of the location of the C-lead electrode.
If an electrode or lead wire disconnects during ECG monitoring, the monitor emits an audible
alarm and displays a
LEADS OFF message. The ECG trace becomes a dashed line. The alarm and
messages continue until the electrode or lead wire is replaced.
Color Coding for ECG Leads
The lead wires and the electrode snaps for the patient ECG cable are color coded according to
American Heart Association (AHA) or International Electrotechnical Commission (IEC) standards
as listed in Table 4-1.
Table 4-1 ECG Leads Color Codes
LEADSAHA LABELAHA COLORIEC LABELIEC COLOR
Limb LeadsRA
LA
RL
LL
C
Precordial LeadsV1
V2
V3
V4
V5
V6
White
Black
Green
Red
Brown
Red
Yellow
Green
Blue
Orange
Violet
R
L
N
F
C
C1
C2
C3
C4
C5
C6
Red
Yellow
Black
Green
Brown
Red
Yellow
Green
Brown
Black
Violet
Monitoring Patients Who Have Internal Pacemakers
The LIFEPAK 15 monitor/defibrillator internal pacemaker detection feature can be used to help
identify internal pacemaker pulses on the printed ECG. When enabled, this feature uses lead V4
to detect internal pacemaker pulses. If V4 is not available because it is not attached or is too
noisy, Lead II is used.
When the internal pacemaker detection feature is ON, the LIFEPAK 15 monitor/defibrillator
annotates a hollow arrow on the printed ECG if internal pacemaker pulses are detected.
Patient history and other ECG waveform data, such as wide QRS complexes, should be used to
verify the presence of an internal pacemaker. False annotations of this arrow may occur if ECG
artifacts mimic internal pacemaker pulses. If false annotations occur frequently, deactivate the
detection feature using the
OPTIONS / PACING / INTERNAL PACER menu (see "Options" on page 3-23).
4-10
The LIFEPAK 15 monitor/defibrillator typically does not use internal pacemaker pulses to
calculate the heart rate. However, when using therapy electrodes or standard paddles to monitor
in
PADDLES lead, the monitor may detect internal pacemaker pulses as QRS complexes, resulting
Large amplitude pacemaker pulses may overload the QRS complex detector circuitry so that no
paced QRS complexes are counted. To help minimize ECG pickup of large unipolar pacemaker
pulses, place ECG electrodes so the line between the positive and negative electrodes is
perpendicular to the line between the pacemaker generator and the heart.
4
Smaller amplitude internal pacemaker pulses may not be distinguished clearly in
For improved detection and visibility of internal pacemaker pulses, turn on the internal
pacemaker detector function using the
ECG cable, select an ECG lead, and print the ECG in diagnostic frequency response. For
information about configuring internal pacemaker detection, see the Pacing Setup menu in the
LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device.
OPTIONS / PACING / INTERNAL PACER menu or connect the
PADDLES lead.
Troubleshooting Tips
If problems occur while monitoring the ECG, check Table 4-2 for aid in troubleshooting. For basic
troubleshooting problems, such as no power, see General Troubleshooting Tips on page 9-18.
Table 4-2 Troubleshooting Tips for ECG Monitoring
OBSERVATIONPOSSIBLE CAUSECORRECTIVE ACTION
Any of these messages
displayed:
CONNECT ELECTRODES
CONNECT ECG LEADS
ECG LEADS OFF
XX LEADS OFF
• Remove the Test Load and
connect therapy electrodes to
cable.
• Connect ECG cable and select
another lead.
• Turn on internal pacemaker
detector (see "Monitoring Patients
Who Have Internal Pacemakers"
on page 4-10).
• Connect ECG cable and select a
lead other than
• Print ECG in Diagnostic mode
(see "How to Print a Current
Report" on page 7-10).
PADDLES.
4-13
Acquiring a 12-Lead ECG
Intended Use
The 12-lead ECG offers paramedics and emergency physicians significant advantages over the
single lead ECG trace typically available in EMS. The 12-lead ECG not only provides a diagnostic
quality ECG for use in the detection of ST elevation myocardial infarction (STEMI), but also allows
the knowledgeable paramedic to determine the area of myocardial injury, anticipate associated
potential complications, and implement treatment strategies accordingly. In addition, the 12-lead
ECG provides a baseline for serial ECG evaluations.
The 12-lead ECG transmission to the emergency department (ED) is recommended by the AHA and
ERC for patients with Acute Coronary Syndrome (ACS). When transmitted from the field, 12-lead
ECG has been shown to shorten time to in-hospital treatment by an estimated 10 to 60 minutes.
Patients may also benefit from triage and transport to the most appropriate facility. Documentation
of transient or intermittent arrhythmias and other electrophysiologic events that occur in the
prehospital setting can assist in diagnosis and treatment decisions in the ED.
Indications
The 12-lead electrocardiogram is used to identify, diagnose, and treat patients with cardiac
disorders and is useful in the early detection and prompt treatment of patients with acute STelevation myocardial infarction (STEMI).
Contraindications
None known.
12-Lead ECG Warning
WARNING
POSSIBLE INABILITY TO OBTAIN A DIAGNOSTIC 12-LEAD ECG
Using previously unpackaged electrodes or electrodes past the Use By date may impair
ECG signal quality. Remove electrodes from a sealed package immediately before use
and follow the procedure for applying the electrodes.
Identifying Electrode Sites
To obtain a 12-lead ECG, place the electrodes on the limbs and the chest (precordium) as
described in the following paragraph.
V1C1Fourth intercostal space to the right of the sternum
V2C2Fourth intercostal space to the left of the sternum
V3C3Directly between leads V2/C2 and V4/C4
V4C4Fifth intercostal space at midclavicular line
V5C5Level with V4/C4 at left anterior axillary line
V6C6Level with V5/C5 at left midaxillary line
Angle of
Louis
Limb Lead Electrode Sites
When acquiring a 12-lead ECG, limb lead electrodes are typically placed on the wrists and
ankles as shown in Figure 4-4. The limb lead electrodes can be placed anywhere along the
limbs. Do not place the limb lead electrodes on the torso when acquiring a 12-lead ECG.
Figure 4-4 Limb Lead Electrode Placement for 12-Lead ECG
4
Precordial Lead Electrode Sites
The six precordial (chest) leads are placed on specific locations as shown and summarized in
Figure 4-5. Proper placement is important for accurate diagnosis and should be identified as
follows: leads are V1 through V6 for AHA, or C1 through C6 for IEC. See ECG Leads Color Codes
on page 4-10 for color codes.
Figure 4-5 Precordial Lead Electrode Placement
Locating the V1/C1 position (fourth intercostal space) is critically important, because it is the
reference point for locating the placement of the remaining V/C leads.
1. Place your finger at the notch in the top of the sternum.
2. Move your finger slowly downward about 3.8 centimeters (1.5 inches) until you feel a slight
horizontal ridge or elevation. This is the Angle of Louis where the manubrium joins the body
of the sternum.
3. Locate the second intercostal space on the patient’s right side, lateral to and just below the
Angle of Louis.
4. Move your finger down two more intercostal spaces to the fourth intercostal space, which is
the V1/C1 position.
5. Continue locating other positions from V1/C1 (see Figure 4-5).
Other important considerations:
• When placing electrodes on female or obese patients, always place leads V3-V6 and C3-C6
under the breast rather than on the breast.
• Never use the nipples as reference points for locating the electrodes for men or women
patients, because nipple locations vary widely.
12-Lead ECG Procedure
To acquire a 12-lead ECG:
1. Press
ON.
2. Insert the limb lead and the precordial lead attachments into the main cable as shown in
Figure 4-6.
Figure 4-6 12-Lead ECG Cable
3. Insert the cable connector into the green ECG connector on the monitor.
4. Prepare patient’s skin for electrode application (see page 4-9).
6. Encourage the patient to remain as still as possible.
WARNING
POSSIBLE INACCURATE DIAGNOSIS
If age and sex are not entered when a 12-lead ECG is obtained, the interpretive
statements are based on a default of a 50-year-old male and may provide incorrect
analysis for that patient.
MONITORING
4
7. Press
8. The
The monitor acquires, analyzes, and automatically prints the 12-lead ECG. An ECG leads-off
condition for any lead is indicated on the report by a dashed line.
12-LEAD.The 12-LEAD / AGE menu appears, prompting you to enter the patient’s age.
Use the
or younger. If you do not enter an age, the default value of 50 years is used by the interpretive
analysis program and annotated on the 12-lead ECG report.
Use the
used by the interpretive analysis program and is annotated on the 12-lead ECG report.
SPEED DIAL to select the age. Always enter the patient’s age if the patient is 15 years old
12-LEAD / SEX menu appears, prompting you to enter the patient’s sex.
SPEED DIAL to select the patient’s sex. If you do not enter the sex, the default of male is
Note: If 15 years or less is entered for patient age, the 12-lead ECG prints at diagnostic frequency
response of 0.05–150 Hz, even when 0.05–40 Hz is set up as the print default.
Note: When 12-LEAD is pressed, internal pacemaker detection is automatically enabled, even if the
function is set up to be OFF.
ECG Override
If the monitor detects signal noise while acquiring data (such as patient motion or a disconnected
electrode), the screen displays the message:
remains and 12-lead ECG acquisition is interrupted until noise is eliminated. Take appropriate
action to eliminate the signal noise. This message remains as long as signal noise is detected.
When signal noise is eliminated, the monitor resumes acquiring data. To override the message and
acquire the 12-lead ECG in spite of the signal noise, press
acquired and printed with no interpretive statements. Any 12-lead ECG report acquired in this way
is annotated with the following statement:
NOISY DATA! PRESS 12-LEAD TO ACCEPT. The message
12-LEAD again. The 12-lead ECG will be
ECG OVERRIDE: DATA QUALITY PROHIBITS INTERPRETATION.
If the signal noise persists for longer than 30 seconds, 12-lead ECG acquisition stops. The screen
displays
acquisition.
EXCESSIVE NOISE–12-LEAD CANCELLED. You must then press 12-LEAD to restart 12-lead ECG
Note: If 12-LEAD is pressed immediately after ECG electrodes are applied, the message NOISY DATA
may occur. This message is due to the temporary instability between the electrode gel and the
patient’s skin that is not viewable on the ECG monitor screen, but is detected as noisy data. In
general, it is best to wait at least 30 seconds after applying the last electrode before pressing the
12-LEAD button, to allow for electrode/skin stabilization. Also, good skin preparation shortens the
stabilization time.
Computerized ECG Analysis
Computerized ECG analysis statements are automatically printed on 12-lead ECG reports. Printing
of the interpretive statements is a setup option and may be turned off in Setup mode. For
information on how to change this setup option, see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device.
The interpretative statements pertaining to myocardial injury, infarct, and ischemia are derived
from measurements made on a signal-averaged beat (median beat) formed for each of the 12 leads.
The computerized ECG analysis selects three representative beats from the ten seconds of data for
each lead and averages the three beats to derive the median beat for that lead. The ECG analysis is
always based on ECG data obtained at 0.05–150 Hz frequency response.
The analysis program is adjusted for patient age and sex. The 12-lead ECG interpretive algorithm
used by the LIFEPAK 15 monitor/defibrillator is the University of Glasgow 12-Lead ECG Analysis
Program. For more information, contact your Physio-Control representative for a copy of the Physio-Control Glasgow 12-Lead ECG Analysis Program Physician’s Guide.
WARNING
POSSIBLE INCORRECT TREATMENT WITH REPERFUSION THERAPY
Computerized ECG interpretive statements should not be used to withhold or prescribe
patient treatment without review of the ECG data by qualified medical personnel. All
12-lead ECG interpretation statements provided by the LIFEPAK 15 monitor/
defibrillator include the printed message **UNCONFIRMED**. Always confirm
interpretive statements by over-reading the ECG data.
Printed 12-Lead ECG Report Formats
Two 12-lead ECG report formats are available for printing: 3-channel or 4-channel. In addition,
each of those formats can be printed in standard and cabrera styles.
The 3-Channel Format
The 3-channel format prints 2.5 seconds of data for each lead. Figure 4-7 is an example of a
12-lead ECG report printed in the 3-channel format, standard style. Figure 4-8 is an example of
a 12-lead ECG report printed in the 3-channel format, cabrera style. The sequence in which the
limb leads are presented differs between the standard and cabrera styles, as shown. The default
format for printing 12-lead ECG reports is 3-channel standard. To change the printed format of
12-lead ECG reports, see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with
your device. Alternatively, press
FORMAT.
OPTIONS, select PRINT, select REPORT: 12-LEAD, and then select
4
Figure 4-7 Example of Printed 3-Channel, Standard 12-Lead ECG Report
Figure 4-8 Example of Printed 3-Channel, Cabrera 12-Lead ECG Report
The 4-Channel Format
Figure 4-9 and Figure 4-10 are examples of 12-lead ECG reports printed in the 4-channel
format. The 4-channel format consists of the median complex (or median beat) derived for each
of the 12 leads and 10 seconds of data for Lead II.
Note: The fiducial marks displayed in the 4-channel format identify the measurement intervals
Fiducial marks
Name:Lee, William
Record ID:
Patient ID:
Incident:
Age: 50
Sex: M
12-Lead 1
P-QRS-T Axes
24 Apr 08
PR .168s
QT/QTc
HR 78 bpm
QRS 0.104s
Abnormal ECG **Unconfirmed**
***Meets ST ELEVATION MI CRITERIA***
Sinus rhythm
Anteroseptal ST elevation, CONSIDER ACUTE INFARCT
aVR
aVL
aVF
V1
V2
V3
V4
V5
V6
Record ID:
Patient ID:
Incident:
Age: 50
Sex: M
12-Lead 1
24 Apr 08
PR .168s
QT/QTc
P-QRS-T Axes
Abnormal ECG **Unconfirmed**
***Meets ST ELEVATION MI CRITERIA***
Sinus rhythm
Anteroseptal ST elevation, CONSIDER ACUTE INFARCT
Name:
Lee, William
HR 89 bpm
QRS 0.104s
aVL
-aVR
aVF
V1
V2
V3
V4
V5
V6
used for the interpretive statements of the analysis program. These marks are part of the
analysis program and cannot be turned off.
Figure 4-9 Example of Printed 4-Channel, Standard 12-Lead ECG Report
Figure 4-10 Example of Printed 4-Channel, Cabrera 12-Lead ECG Report
Printed 12-Lead ECG Frequency Response
The 12-lead ECG can be printed in two diagnostic frequency responses (or bandwidths):
0.05–40 Hz and 0.05–150 Hz. The frequency response of 0.05–150 Hz is the Association for the
Advancement of Medical Instrumentation (AAMI) standard for diagnostic ECGs. The 0.05–40 Hz
setting preserves the low frequency limit that is needed for the diagnosis of myocardial ischemia
and infarction while reducing high frequency artifact (in particular from patient muscle tension) to
help make the diagnostic printout less noisy and more readable.
Note: The LIFEPAK 15 monitor/defibrillator acquires ECG data and performs the interpretive
analysis based on the full frequency of 0.05–150 Hz. The 0.05–40 Hz bandwidth affects only the
printed appearance of the ECG data.
The 12-lead ECG printed in the 0.05–40 Hz setting can be used to diagnose acute myocardial
ischemia and ST-segment elevation myocardial infarction (STEMI). This is because the low
frequency limit of 0.05 Hz is not changed from the standard diagnostic setting of 0.05–150 Hz.
The 0.05 Hz frequency provides accurate representation of low frequency signals, that is, the P, ST
segment, and T waves. The presence or absence of ST segment changes indicative of myocardial
ischemia or infarction will be accurately reproduced. In addition, the criteria for visual analysis and
interpretation of cardiac rhythm and PR, QRS, and QT intervals are preserved, as is true with
hospital cardiac monitors that have an upper frequency limit of 40 Hz.
However, in some adult patients, the amplitude (that is, voltage) of the QRS may be reduced when
12-lead ECGs are printed at the upper limit of 40 Hz rather than at 150 Hz. Therefore, certain
diagnoses, which depend on R wave amplitude (for example, ventricular hypertrophy), should not
be made using this setting. In the pediatric patient, this effect on R wave amplitude is particularly
noticeable because QRS durations in children are typically quite narrow. Because R wave amplitude
reduction is more likely with pediatric patients, the 12-lead ECG automatically prints at 0.05–150
Hz, overriding the 40 Hz limit, when a patient age of 15 years or younger is entered.
SpO2, SpCO™, and SpMet™ are optional features for the LIFEPAK 15 monitor/defibrillator. When
all three options (SpO
oxygen saturation (SpO
concentration (SpMet) in the blood.
, SpCO, and SpMet) are installed, the pulse oximeter measures functional
2
), carboxyhemoglobin concentration (SpCO), and methemoglobin
2
IMPORTANT! SpO
for use. Masimo
monitor. Masimo Rainbow™ sensors are necessary to monitor SpO
sensors are not compatible with other LIFEPAK defibrillator/monitors.
For a list of Masimo sensors and connector cables that are intended for use with the LIFEPAK 15
monitor/defibrillator, see the Physio-Control web site. Carefully read the Directions for Use that are
provided with the sensors and connector cables for a complete description, instructions, warnings,
cautions, and specifications. To order sensors and connector cables, contact your Physio-Control
representative or order online at store.physio-control.com.
-only sensors and combination SpO2, SpCO, and SpMet sensors are available
2
®
SpO2-only sensors that have a red connector are compatible with the LIFEPAK 15
, SpCO, and SpMet. These
2
Intended Use
A pulse oximeter is a noninvasive device that continuously measures functional oxygen saturations
(SpO
), carboxyhemoglobin concentration (SpCO), and methemoglobin concentration (SpMet) in
2
the blood. Continuously monitoring SpO
decreasing and can help the clinician act rapidly before the patient develops the later signs of
hypoxemia. Previously, the blood parameters SpCO and SpMet could only be obtained from invasive
blood gas samples. This new technology assists in identifying the often hidden conditions of
carboxyhemoglobinemia (carbon monoxide poisoning) and methemoglobinemia (a condition that
impedes delivery of oxygen to the tissues). Low levels of both SpCO and SpMet are normally found
in the blood; however, early detection of significantly high levels can lead to proper diagnosis and
treatment, and can help improve patient outcome.
can provide an early warning when oxygen saturation is
2
Pulse oximetry is a tool to be used in addition to patient assessment. Care should be taken to
assess the patient at all times; do not rely solely on the SpO
a trend toward patient deoxygenation is evident or carbon monoxide poisoning or
methemoglobinemia is suspected, blood samples should also be analyzed using laboratory
instruments to completely understand the patient’s condition.
Do not use the pulse oximeter to monitor patients for apnea.
Pulse oximetry is indicated for use in any patient who is at risk of developing hypoxemia,
carboxyhemoglobinemia, or methemoglobinemia. SpO
motion and motion conditions, and in patients who are well or poorly perfused. SpCO and SpMet
accuracies have not been validated under motion or low perfusion conditions.
Before use, carefully read these operating instructions, the sensor and cable directions
for use, and precautionary information.
SHOCK OR BURN HAZARD
Using other manufacturers’ sensors or cables may cause improper oximeter
performance and invalidate safety agency certifications. Use only sensors and cables
that are specified in these operating instructions.
INACCURATE READINGS HAZARDS
INACCURATE PULSE OXIMETER READINGS
Do not use a damaged sensor or cable. Do not alter the sensor or cable in any way.
Alterations or modification may affect performance and/or accuracy. Never use more
than one cable between the pulse oximeter and the sensor to extend the length.
INACCURATE PULSE OXIMETER READINGS
Sensors exposed to ambient light when incorrectly applied to a patient may exhibit
inaccurate saturation readings. Securely place the sensor on the patient and check the
sensor’s application frequently to help ensure accurate readings.
INACCURATE PULSE OXIMETER READINGS
Severe anemia, methemoglobin, intravascular dyes that change usual blood
pigmentation, excessive patient movement, venous pulsations, electrosurgical
interference, exposure to irradiation and placement of the sensor on an extremity that
has a blood pressure cuff, intravascular line, or externally applied coloring (such as
nail polish) may interfere with oximeter performance. The operator should be
thoroughly familiar with the operation of the oximeter prior to use.
INACCURATE PULSE OXIMETER READINGS
The pulsations from intra-aortic balloon support can be additive to the pulse rate on
the oximeter pulse rate display. Verify patient’s pulse rate against the ECG heart rate.
POSSIBLE SKIN INJURY
Prolonged, continuous use of a sensor may cause irritation, blistering, or pressure
necrosis of the skin. Check the sensor site regularly based on patient condition and
type of sensor. Change the sensor site if skin changes occur. Do not use tape to hold
the sensor in place as this may cause inaccurate readings or damage to the sensor or
skin.
Carefully route patient cabling to reduce the possibility of patient entanglement or
strangulation.
CAUTIONS
EQUIPMENT HAZARDS
POSSIBLE EQUIPMENT DAMAGE
To avoid damage to the cable, always hold by the connector rather than the cable,
when connecting or disconnecting either end.
POSSIBLE EQUIPMENT DAMAGE
Do not soak or immerse the sensors or cables in any liquid solution. Do not attempt to
sterilize.
4
No Implied License
Possession or purchase of the pulse oximeter does not convey any expressed or implied license to
use the pulse oximeter with unauthorized sensors or cables which would, alone or in combination
with this device, fall within the scope of one or more of the patents relating to this device.
How a Pulse Oximeter Works
A pulse oximeter sensor directs light through a patient’s fleshy body site (usually a finger or toe).
The sensor sends wavelengths of light from the emitter to the receiving detector as shown in
Figure 4-11.
The pulse oximeter translates the amount of light received by the detector to the various forms of
hemoglobin saturation levels and displays them as SpO
values for SpO
9% (the higher range of normal is often seen in smokers). Normal values for SpMet are typically
less than 2% and may be caused by exposure to some pharmaceuticals including local anesthetic
agents and chemical agents such as nitrites.
typically range from 95% to 100%. Normal values for SpCO are typically less than
2
, SpCO, and SpMet percentages. Normal
2
SpO2, SpCO, and SpMet Monitoring Considerations
The quality of the SpO2, SpCO, and SpMet readings depends on correct sensor size and placement,
adequate blood flow through the sensor site, and limiting patient motion and sensor exposure to
ambient light. For example, with very low perfusion at the sensor site, readings may be lower than
core arterial oxygen saturation. Test methods for accuracy are available by contacting your local
Physio-Control representative.
Use the following criteria to select the appropriate pulse oximeter sensor:
• Patient size (adult, pediatric, infant) and weight
• Patient perfusion to extremities
• Patient activity level
• Available application sites on the patient’s body
• Sterility requirements
• Anticipated duration of monitoring
To help ensure optimal performance:
• Use a dry and appropriately sized sensor.
• Choose a site that is well perfused. The ring finger is preferred.
• Choose a site that least restricts patient movement, such as finger of the non-dominant hand.
• Be sure the fleshy part of the digit completely covers the detector.
• Keep the sensor site at the same level as the patient’s heart.
• Apply the sensor according to the Directions for Use provided with the sensor.
• Observe all warnings and cautions noted in the sensor’s Directions for Use.
Sensor Application
The preferred site for sensor application is the ring finger of the non-dominant hand. To position
the sensor:
1. Orient the sensor so the cable is on the back of the patient’s hand.
2. Place the finger in the sensor until the tip of the finger touches the “raised digit stop.”
3. The hinged tabs of the sensor should open to evenly distribute the grip pressure of the sensor
along the length of the finger. Check the arrangement of the sensor to verify correct
positioning. Complete coverage of the detector window is needed to ensure accurate data.
The sensors are sensitive to light. If excessive ambient light is present, remove or reduce
lighting, cover the sensor site with an opaque material to block the light, and check
appropriateness of sensor site. Failure to do so could result in inaccurate measurements.
If excessive movement presents a problem during SpCO/SpMet monitoring, consider the
following possible solutions:
• Be sure the sensor is secure and properly aligned.
• Use a disposable adhesive sensor.
• If possible, move the sensor to a less active site.
Note: Wrapping the sensor too tightly or using supplemental tape to hold the sensor in place
may cause inaccurate oximeter readings.
Note: Circulation distal to the sensor site should be checked routinely.
IMPORTANT! Masimo Rainbow sensors are necessary to monitor SpCO and SpMet and are not
compatible with other LIFEPAK defibrillator/monitors.
4
Oximeter Monitoring Procedure
Power to the pulse oximeter is controlled by the LIFEPAK 15 monitor/defibrillator. When the
defibrillator is turned on, the oximeter turns on and performs a calibration and self-test that
requires approximately 20 seconds. During the calibration and self-test, the screen does not display
SpO
, SpCO, or SpMet information.
2
To conserve battery power, the pulse oximeter goes into “sleep mode” when not in use. Sleep mode
is activated within 10 seconds of disconnecting the sensor. During sleep mode, the screen does not
display SpO
oximeter performs a self-test and then returns to normal mode.
The pulse oximeter measures and displays SpO
than 50% are displayed as <50. When SpO
measurements are accurate ±3 digits. The pulse oximeter measures and displays SpCO in the range
of 0–40% with accuracy of ±3 digits. The pulse oximeter measures and displays SpMet in the range
of 0–15% with accuracy of ± 1 digit.
, SpCO, or SpMet information. When a sensor or patient signal is detected, the
1. Rotate the SPEED DIAL to outline the SpO2
area on the Home Screen.
2. Press the
SPEED DIAL.
3. Highlight and select
SPO2 VOLUME.
4. Rotate the
SPEED DIAL to the desired
volume.
5. Press the
SPEED DIAL to set the volume.
SpO2_SpCo_SpMet
Parameter
SpO2 Volume
Sensitivity
Averaging Time
SpO2
Normal
8 Seconds
To cancel the Advisory, press ALARMS. The SpO2 area reverts to the SpO2 reading. The Advisory
message remains on the screen until the elevated value returns to within normal limits or the device
is turned off.
WARNING
INACCURATE SPO2 READINGS
Carboxyhemoglobin and methemoglobin may erroneously increase SpO2 readings. The
amount that SpO
carboxyhemoglobin or methemoglobin that is present.
The Pleth Waveform
You can display the plethysmographic (pleth) waveform in Channel 2 or 3.
To display the pleth waveform:
increases is approximately equal to the amount of
2
4
1. Rotate the
2. Press the
3. Select
The waveform is automatically sized for optimum waveform viewing.
SPEED DIAL to outline waveform CHANNEL 2 or 3.
SPEED DIAL. The Channel menu appears.
WAVEFORM and then select SPO2. The SpO
waveform appears in the selected channel.
2
Volume
To adjust the pulse tone volume:
Sensitivity
The sensitivity setting allows you to adjust the oximeter to either NORMAL or HIGH for differing
perfusion states.
• If values do not display
within 30 seconds,
disconnect and reconnect
sensor. If values do not
display within another 30
seconds, replace sensor.
• None. If values do not
display within 30 seconds,
disconnect and reconnect
sensor. If values do not
display within another 30
seconds, replace sensor.
• Cover sensor with opaque
material, if necessary.
4-33
Table 4-4 Troubleshooting Tips for SpO
, SpCO, and SpMet (Continued)
2
OBSERVATIONPOSSIBLE CAUSECORRECTIVE ACTION
Damaged cable or sensor• Replace damaged cable or
sensor.
Different SpCO or SpMet
measurements on same patient
XXX appears in place of SpO
2
reading
SPO2: CHECK SENSOR message
appears
Every measurement, even on
the same patient, can be
different
• Confirm by taking three
measurements: ring finger,
middle finger, and then
index finger; average the
results.
SpO2 module failed.
Internal cable failed.
Sensor is disconnected from
patient or cable
• Contact qualified service
personnel.
• Attach the sensor.
• Check that sensor is secure.
Excessive ambient light• Remove or block light
source, if possible.
• Cover sensor with opaque
material, if necessary.
Faulty or defective sensor• Replace sensor.
Patient has a weak pulse or low
blood pressure, or the sensor is
not properly placed
• Change sensor location.
• Check if patient perfusion is
adequate for sensor
location.
• Check that sensor is secure
and not too tight.
• Check that sensor is not on
extremity with blood
pressure cuff or
intravascular line.
• Test sensor on someone
else.
4-34
SPO2: UNKNOWN SENSOR
message appears
SPO2: SEARCHING FOR PULSE
message appears
SPO2: LOW PERFUSION message
appears
A sensor that is not PhysioControl approved is connected
to the device.
A sensor is connected to the
• Check that the sensor is an
approved Physio-Control
sensor.
For general troubleshooting tips, see Table 9-2 on page 9-18.
, SpCO, and SpMet (Continued)
2
When the signal quality is low,
the accuracy of the
measurement may be
compromised
SpO
-only sensor used with
2
SpCO/SpMet capable device
, SpCO, and SpMet) are reported as SPO2: (MESSAGE).
2
• Check that sensor and cable
are connected properly.
• Move sensor to a better
perfused site.
• None necessary, or use
Rainbow sensor to measure
SpCO or SpMet.
Monitoring Noninvasive Blood Pressure
Intended Use
The LIFEPAK 15 noninvasive blood pressure (NIBP) monitor measures blood pressure (BP) using
the oscillometric measurement technique to determine systolic, diastolic, and mean arterial
pressures, and pulse rate. The measurement can be initiated manually or set to recur automatically
at predetermined intervals.
Blood pressure measurements determined using this device are equivalent to those obtained by a
trained observer using the cuff/stethoscope auscultation method, within the limits prescribed by the
American National Standard Electronic or automated sphygmomanometers (AAMI SP-10).
NIBP is a tool to be used in addition to patient assessment. Care should be taken to assess the
patient at all times; do not rely solely on the NIBP monitor.
Indications
Noninvasive blood pressure monitoring is intended for detection of hypertension or hypotension
and monitoring BP trends in patient conditions such as, but not limited to, shock, acute
dysrhythmia, or major fluid imbalance.
POSSIBLE LOSS OF INTRAVENOUS ACCESS AND INACCURATE INFUSION RATE
Do not apply the blood pressure cuff on an extremity that is used for an intravenous
infusion. Patency of the intravenous infusion may be affected by blood pressure
measurement due to the occlusion of blood flow.
INACCURATE READINGS HAZARDS
POSSIBLE INACCURATE BLOOD PRESSURE READINGS
Do not alter the NIBP monitor’s pneumatic tubing. Altering NIBP tubing may cause
improper performance and may void the warranty. Avoid compression or restriction of
pressure tubes.
POSSIBLE INACCURATE BLOOD PRESSURE READINGS
Using NIBP accessories not recommended by Physio-Control may cause the device to
perform improperly and invalidate the safety agency certifications. Use only the
accessories that are specified in these operating instructions.
POSSIBLE INACCURATE OXYGEN SATURATION READINGS
Do not perform NIBP measurement on an extremity used for oxygen saturation
monitoring. Oxygen saturation measurement is affected by blood pressure
measurement due to the occlusion of blood flow.
CAUTION
EQUIPMENT DAMAGE
Do not inflate a cuff unless it is placed on an extremity.
How NIBP Monitoring Works
The NIBP monitor uses the oscillometric measurement technique. The oscillometric technique does
not use Korotkoff sounds to determine blood pressure; rather, it monitors the changes in pressure
pulses that are caused by the flow of blood through the artery. The NIBP monitor inflates the cuff
around the patient’s arm to a value that occludes the artery, and then deflates the cuff in steps.
When blood starts to flow through the artery, the increasing blood flow causes the amplitude of the
pressure pulses in the cuff to increase. As the NIBP monitor steps the pressure down, the pulses
reach a peak amplitude and then start to decrease. The rising and falling amplitude values form a
curve that is analyzed to yield systolic pressure, diastolic pressure, and mean arterial pressure
(MAP).
The NIBP monitor measures the pulse rate by tracking the number of pulses over time. The NIBP
monitor uses artifact rejection techniques to provide accurate results under most operating
conditions. When a patient is experiencing arrhythmias during a measurement, the accuracy of the
pulse determination may be affected or the time needed to complete a measurement may be
extended. In shock conditions, the low amplitude of blood pressure waveforms makes it difficult for
the monitor to accurately determine the systolic and diastolic pressures.
NIBP Monitoring Considerations
As with any noninvasive oscillometric blood pressure monitor, clinical conditions can affect the
accuracy of the measurements obtained, including the following:
• The patient’s physiological condition. For example, shock may result in a blood pressure
waveform that has a low amplitude, making it difficult for the monitor to accurately determine
the systolic and diastolic pressures.
• The position of the patient.
• Motion may prolong the measurement process since motion artifacts have to be rejected in the
data stream. Motion that affects measurement can include patient movement, patient seizure,
bumping the cuff, and flexing the extremity under the cuff.
4
• The presence of other medical devices. The NIBP monitor does not operate effectively if the
patient is connected to a heart/lung machine.
• When a patient is experiencing arrhythmias, pulse rate accuracy may be affected or the time
needed to complete an NIBP measurement may be extended. The device automatically deflates
if a blood pressure measurement cannot be obtained in 120 seconds.
• Blood pressure and pulse can fluctuate greatly between measurements; the monitor cannot alert
the operator of changes in vital signs that occur between measurement cycles.
• There may be some difference between readings taken manually and readings from the NIBP
monitor due to the differing sensitivity of the two methods. The NIBP monitor meets the ANSI/
SP10 AAMI standard that requires a mean difference of ±5 mmHg, with a standard deviation no
greater than 8 mmHg, compared to auscultatory readings.
• When using the NIBP monitor during defibrillation, the NIBP monitor is not available when the
defibrillator is being charged. Upon shock, the monitor resets and dashes (– – –) appear in place
of pressure readings. After defibrillation, you can resume blood pressure measurement
according to "NIBP Monitoring Procedure" on page 4-38.
• If the blood pressure cuff fails to deflate for any reason or causes undue discomfort to the
patient, remove the cuff from the arm or disconnect the tubing from the defibrillator.
1. Rotate the SPEED DIAL to outline the NIBP
area.
2. Press the
SPEED DIAL. The NIBP menu
appears.
3. Select
INITIAL PRESSURE.
4. Rotate the
SPEED DIAL to the desired
pressure.
5. Press the
SPEED DIAL to set the initial
pressure.
NIBP
Start
Interval
Initial Pressure
Off
160 mmHg
The use of properly designed and sized cuffs is essential for the accurate measurement of blood
pressure. The cuff must fit snugly around the extremity to occlude the artery. For a list of BP cuffs
that are intended for use with the LIFEPAK 15 monitor/defibrillator, see the LIFEPAK 15 Monitor/
Defibrillator Accessories Catalog at store.physio-control.com.
NIBP Monitoring Procedure
The NIBP monitor inflates an occluding cuff and determines systolic and diastolic pressures, mean
arterial pressure (MAP), and pulse rate. Pressure measurements are reported in mmHg and pulse
rate in beats per minute (bpm).
Both single-measurement and specified-interval (timer-controlled) methods of blood pressure
reading are available.
The NIBP monitor draws power from the defibrillator. When the defibrillator is turned on, the NIBP
monitor conducts a self-test that takes approximately three seconds.
IMPORTANT! The LIFEPAK 15 monitor NIBP port and tubing are not compatible or
interchangeable with the NIBP tubing that is used with other LIFEPAK monitor/defibrillators.
Changing the Initial Inflation Pressure
4-38
The initial cuff pressure should be set approximately 30 mmHg higher than the patient’s
anticipated systolic pressure. The factory default initial inflation pressure for the first
measurement is 160 mmHg. For pediatric patients, the initial cuff pressure may need to be
lowered. Initial inflation settings are 80, 100, 120, 140, 160, or 180 mmHg.
Caution should be taken not to lower the initial pressure below the adult patient’s systolic
measurement. Doing so may cause the cuff to reinflate and cause patient discomfort. For
subsequent measurements, the monitor inflates approximately 30 mmHg higher than the
previously determined systolic pressure.
Countdown timer—displays time
until next measurement
Systolic pressure
Diastolic pressure
Mean arterial pressure (MAP)
Note: Measurement data is recorded in the LIFEPAK 15 monitor/defibrillator Vital Sign Log. For
more information about the Vital Sign Log and its use, see Chapter 7, "Data Management."
Manual Single-Measurement Procedure
The NIBP measurement typically takes 40 seconds to complete. If the measurement is not
completed within 120 seconds, the cuff automatically deflates.
To obtain a manual single measurement:
4
1. Press
2. Select the appropriately-sized cuff and apply it snugly to the extremity.
3. Connect the tubing to the cuff and to the NIBP port on the monitor.
4. Change the initial inflation pressure, if necessary.
5. Position the extremity in a relaxed and supported position at approximately the same level as
6. Press
To cancel a measurement, press
Note: NIBP pulse rate is displayed only when ECG or SpO
ON.
the patient’s heart. Inform the patient that the cuff will inflate and cause a “big squeeze”
around the arm and that the patient’s fingers may tingle.
NIBP to start the measurement, and check that the patient’s arm is not moving. When
the measurement is complete, systolic, diastolic, and mean arterial pressures are displayed.
NIBP again.
is not active.
2
Timer-Controlled Measurement Procedure
When the timer is set, the monitor performs recurring measurements at a fixed interval. When
using timer-controlled measurement, the interval is counted from the start of the measurement
to the start of the next measurement. Choices are
minutes.
To take a manual measurement between timer-controlled measurements, press
interval is counted from the beginning of the manual measurement.
OFF (factory default), 2, 3, 5, 10, 15, 30, and 60
2. Select the appropriately-sized cuff and apply it snugly to the extremity.
3. Connect the tubing to the cuff and to the NIBP port on the monitor.
4. Rotate the
5. Press the
6. Select
7. Position the extremity in a relaxed and supported position at approximately the same level as
8. Press
To cancel a measurement in progress, press
ON.
SPEED DIAL to outline the NIBP area.
SPEED DIAL. The NIBP menu appears.
INTERVAL and then select the desired time interval.
the patient’s heart. Inform the patient that the cuff will inflate and cause a “big squeeze”
around the arm and that the patient’s fingers may tingle.
NIBP to start the measurement, and check that the patient’s arm is not moving. When
the measurement is complete, systolic, diastolic, and mean arterial pressures are displayed.
The countdown timer shows the time to the next automatic NIBP measurement.
NIBP again.
Note: If at any time the cuff pressure exceeds 290 mmHg or there is a system failure of the
NIBP module, timer-controlled NIBP is terminated. To reactivate, follow the Timer-Controlled
Measurement Procedure.
Cleaning
To clean the cuff and pneumatic tubing:
1. Disconnect the tubing from the cuff and monitor. Use a clean, soft cloth dampened with a
germicidal solution to wipe clean.
2. Inspect the tubing for cracks or kinks. If any damage is noted, replace the tubing.
3. Inspect the cuff for damage or excessive wear. If any damage is noted, replace the cuff.
4. Allow both to dry before placing the cuff on a patient or reconnecting the tubing to the monitor.
For information about cleaning the device, see "Cleaning the Device" on page 9-15.
The end-tidal CO2 (EtCO2) monitor is a capnometric device that uses non-dispersive infrared
spectroscopy to continuously measure the amount of CO
present at the end of exhalation (EtCO
can be used with intubated or nonintubated patients. Respiration rate is also measured and
displayed in breaths per minute.
). The sample is obtained by the side stream method and
2
during each breath and report the amount
2
4
The EtCO
assess the patient at all times; do not rely solely on the EtCO
monitor is a tool to be used in addition to patient assessment. Care should be taken to
2
monitor.
2
Indications
EtCO2 monitoring is used to detect trends in the level of expired CO2. It is used for monitoring
breathing efficacy and treatment effectiveness in acute cardiopulmonary care, for example, to
determine if adequate compressions are being performed during CPR or to rapidly detect
whether an endotracheal tube has been placed successfully.
Contraindications
None known.
EtCO2 Monitoring Warnings
WARNINGS
FIRE HAZARDS
FIRE HAZARD
Before use, carefully read these operating instructions, the FilterLine
directions for use, and precautionary information.
®
tubing
FIRE HAZARD
The FilterLine tubing may ignite in the presence of O
electrosurgical devices, or high heat. Use with caution to prevent flammability of the
FilterLine tubing.
FIRE HAZARD
Flammable anesthetics become mixed with the patient’s air that is sampled by the
capnometer. When using the EtCO
as nitrous oxide or certain other anesthetics, connect the EtCO
scavenger system.
The EtCO
be used as a diagnostic apnea monitor. An apnea message appears if a valid breath
has not been detected for 30 seconds and indicates the time elapsed since the last
valid breath. It must be used in conjunction with clinical signs and symptoms.
POSSIBLE INACCURATE CO2 READINGS
Using other manufacturers’ CO
improperly and invalidate the safety agency certifications. Use only the accessories
that are specified in these operating instructions.
HEALTH HAZARDS
POSSIBLE STRANGULATION
Carefully route the patient tubing (FilterLine) to reduce the possibility of patient
entanglement or strangulation.
INFECTION HAZARD
Do not reuse, sterilize, or clean Microstream
single-patient one-time use.
monitor is intended only as an adjunct in patient assessment and is not to
2
accessories may cause the device to perform
2
®
CO2 accessories as they are designed for
How Capnography Works
An EtCO2 sensor continuously monitors carbon dioxide (CO2) that is inspired and exhaled by the
patient. The sensor employs Microstream non-dispersive infrared (IR) spectroscopy to measure the
concentration of CO
The CO
FilterLine system delivers a sample of the exhaled gases directly from the patient into the
2
LIFEPAK 15 monitor for CO
and secretion accumulation and prevents obstruction, which maintains the shape of the CO
waveform.
The CO
sensor captures a micro sample (15 microliters). This extremely small volume allows for
2
fast rise time and accurate CO
The Microbeam IR source illuminates the sample cell and the reference cell. This proprietary IR
light source generates only the specific wavelengths characteristic of the CO
Therefore, no compensations are required when concentrations of O
vapor are present in the exhaled breath.
molecules that absorb infrared light.
2
measurement. The low sampling flow rate (50 ml/min) reduces liquid
You can set up the LIFEPAK 15 monitor/defibrillator to use the capnography Body Temperature
Pressure Saturated (BTPS) conversion method. This option corrects for the difference in
temperature and moisture between the sampling site and alveoli. The correction formula is 0.97 ×
the measured EtCO
your device.
EtCO2 Monitoring Waveform Analysis
Valuable information concerning the patient’s expired CO2 can be acquired by examination and
interpretation of the waveform.
The Phases of the Waveform
Figure 4-13 is a graphic representation of a normal capnograph waveform. Four phases of the
waveform require analysis. The flat I–II baseline segment (Respiratory Baseline) represents
continued inhalation of CO
Upstroke), a sharp rise, represents exhalation of a mixture of dead space gases and alveolar
gases from acini with the shortest transit times. Phase III–IV (Expiratory Plateau) represents the
alveolar plateau, characterized by exhalation of mostly alveolar gas. Point IV is the end-tidal
(EtCO
Downstroke), a sharp fall, reflects the inhalation of gases that are CO
normal capnograph or EtCO
ventilation, or equipment function.
) value that is recorded and displayed by the monitor. Phase IV–V (Inspiratory
2
value. See the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with
2
-free gas. This value normally is zero. The II–III segment (Expiratory
2
-free. Alterations of the
2
values are the result of changes in metabolism, circulation,
2
4
Figure 4-13 Phases of the Respiratory Waveform
Respiratory Baseline Elevation of the waveform baseline (I–II segment) usually represents
rebreathing CO
Rebreathing CO
hypoventilation. Precipitous rises in both baseline and EtCO
contamination of the sensor.
Expiratory Upstroke In the normal waveform, the rising phase (II–III segment) is usually steep.
When this segment becomes less steep, CO
site. The causes of this delay can be physiologic or mechanical and include bronchospasm,
obstruction of the upper airway, or obstruction (or kinking) of an endotracheal tube (ETT).
Expiratory Plateau The plateau of the waveform, which represents the remainder of expiration
(III-IV segment), should be nearly horizontal. The end of the plateau represents the EtCO
Upward slanting of the expiratory plateau occurs when there is uneven emptying of the alveoli.
Similar to the diminished slope of the Expiratory Upstroke, this pattern can occur in asthma,
chronic obstructive pulmonary disease (COPD), partial upper-airway obstruction, or partial
mechanical obstruction such as a partially kinked ETT.
Inspiratory Downstroke The fall to baseline (IV-V segment) is a nearly vertical drop. This slope
can be prolonged and can blend with the expiratory plateau in cases of leakage in the exhale
portion of the breathing circuit. The peak EtCO
value (IV) is often not reached. Relying on the
2
numeric end-tidal value without observing the breathing waveform may obscure the presence of
a leak.
EtCO2 Monitoring Procedure
When activated, the EtCO2 monitor draws power from the defibrillator. The LIFEPAK 15 monitor/
defibrillator activates the EtCO
Initialization, self-test, and warm up of the EtCO
may take up to two-and-one-half minutes.
CAUTION
POSSIBLE EQUIPMENT DAMAGE
Failure to replace a broken or missing CO2 port door may allow water or particulate
contamination of the internal CO
malfunction.
monitor when it senses the attachment of the FilterLine set.
2
monitor is typically less than 30 seconds, but
2
sensor. This may cause the CO2 module to
2
To monitor EtCO
1. Press
ON.
2. Select the appropriate EtCO
3. Open the CO
4. Verify that the CO
:
2
accessory for the patient.
2
port door and insert the FilterLine connector; turn connector clockwise until tight.
2
area is displayed. The EtCO2 monitor performs the autozero routine as part
2
of the initialization self-test.
Note: If you use a ventilation system, do not connect the FilterLine set to the patient/ventilation
system until the EtCO
5. Display CO
6. Connect the CO
waveform in Channel 2 or 3.
2
2
7. Confirm that the EtCO
scale for the best visualization of the waveform. You can change the scale, if desired, as
described in the next section.
monitor has completed its self-test and warm-up.
2
FilterLine set to the patient.
value and waveform are displayed. The monitor automatically selects the
2
Note: It is possible for the FilterLine set to become loose at the device connection and still have an
EtCO
value and CO2 waveform, but they may be erroneously low. Make sure the FilterLine
Note: The capnography module performs self-maintenance within the first hour of monitoring and
once an hour during continuous monitoring. The self-maintenance includes “auto-zeroing.” Selfmaintenance is also initiated when the surrounding temperature changes 8°C (14.4°F) or more, or
the surrounding pressure changes greater than 20 mmHg. The CO
attempts to purge the tubing. To clear the
messages, remove the FilterLine tubing and reconnect it to the monitor.
CO2 FILTERLINE PURGING or CO2 FILTERLINE BLOCKAGE
CO2 Display
The following scales are available to display the CO2 waveform. The LIFEPAK 15 monitor/
defibrillator automatically selects the scale based on the measured EtCO
scale, outline and select the CO
the scale menu.
• Autoscale (default)
• 0–20 mmHg (0–4 Vol% or kPa).
• 0–50 mmHg (0–7 Vol% or kPa).
• 0–100 mmHg (0–14 Vol% or kPa).
area using the SPEED DIAL and then select the desired scale from
2
module detects this change and
2
value. To change the CO2
2
4
The CO
the 4-second screen. There is a slight delay between when the breath occurs and when it appears
on the screen. Printouts are at 25 mm/sec. Continuous print may be changed to 12.5 mm/sec, if
desired.
The monitor shows the maximum CO
increasing, the change can be seen with every breath. However, if the values are continually
decreasing, it will take up to 20 seconds for a lower numerical value to be displayed. Because of
this, the EtCO
waveform is compressed (displayed at 12.5 mm/sec sweep speed) to provide more data in
2
value over the last 20 seconds. If the EtCO2 values are
2
value may not always match the level of the CO2 waveform.
2
CO2 Alarms
The EtCO2 monitor provides:
•EtCO
•FiCO
• Apnea alarm (automatic and not adjustable)
Note: The apnea alarm occurs when a breath has not been detected for 30 seconds. The message
ALARM APNEA appears in the message area along with the time since the last detected breath.
high and low alarms controlled by activating ALARMS (see "Alarms" on page 3-21)
2
(inspired CO2) alarm (automatic and not adjustable)
A CO2 waveform appears when any CO2 is detected, but CO2 must be greater than 3.5 mmHg for a
numerical value to be displayed. However, the CO
is at least 8 mmHg (1.0% or kPa). Valid breaths must be detected in order for the apnea alarm to
function and to count the respiratory rate (RR). The RR represents an average over the last eight
breaths.
module will not recognize a breath until the CO2
2
When CO
dashes “---” or a flat solid line at or near zero—several factors must be quickly evaluated. Assess
for the following causes:
Equipment issues
• Disconnection of the FilterLine set from the endotracheal tube (ETT)
• System is purging due to fluid in the patient/sensor connection from ET administration of
medications
• System is auto-zeroing
• Shock was delivered and system is resetting
• Loose FilterLine set to device connection
Loss of airway function
• Improper placement of ETT
• ETT dislodgment
• ETT obstruction
Physiological factors
• Apnea• Loss of perfusion• Exsanguination
• Massive pulmonary embolism• Inadequate CPR
is not detected in the cardiac arrest situation—for example, the CO2 waveform is either
2
Cleaning
Accessories for CO2 monitoring are disposable and are intended for single-patient use. Do not clean
and reuse a FilterLine set. Dispose of the contaminated waste according to local protocols.
For information about cleaning the device, see "Cleaning the Device" on page 9-15.
Note: To decrease the likelihood of the FilterLine connection coming loose during use, hand-
straighten the tubing after removal from the package before connecting to patient or device.
For general troubleshooting tips, see Table 9-2 on page 9-18.
Monitoring Invasive Pressure
Intended Use
The LIFEPAK 15 invasive pressure (IP) monitor is intended for measuring arterial, venous,
intracranial, and other physiological pressures using an invasive catheter system with a compatible
transducer.
The IP monitor is a tool to be used in addition to patient assessment. Care should be taken to
assess the patient at all times; do not rely solely on the IP monitor.
4
Indications
Invasive pressure monitoring is indicated for use in patients who require continuous monitoring
of physiological pressures in order to rapidly assess changes in the patient’s condition or
response to therapy. It may also be used to aid in medical diagnosis.
Contraindications
None known.
IP Monitoring Warnings
WARNINGS
INACCURATE READINGS HAZARDS
POSSIBLE INACCURATE PRESSURE READINGS, AIR EMBOLISM, BLOOD LOSS, OR LOSS OF
STERILITY
Before use, carefully read these operating instructions, and the transducer and
infusion set instructions for use and precautionary information.
INACCURATE PRESSURE READINGS
Pressure readings should correlate with the patient’s clinical presentation. If readings
do not correlate, verify that the zeroing stopcock is positioned at the patient’s zero
reference, rezero the transducer, and/or check the transducer with a known or
calibrated pressure. Manually check cuff blood pressure.
Changing the patient’s position changes the zero reference level. Relevel the
transducer’s zeroing stopcock any time the patient’s position is changed.
HEALTH HAZARDS
POSSIBLE LETHAL ARRHYTHMIA
Ventricular fibrillation may be induced if the isoelectric barrier of the transducer is
disrupted. The isoelectric barrier within the transducer may be disrupted if the
transducer body is damaged. Do not use a transducer that is visibly damaged or
leaking fluid.
INCREASED INTRACRANIAL PRESSURE
Do not use a continuous flush device with transducers used for intracranial
monitoring.
IP Monitoring
Two channels are available for invasive pressure monitoring, with default labels P1 and P2 and the
user-selectable labels shown in Table 4-7.
Table 4-7 IP Labels and Descriptions
LABELDESCRIPTION
ARTArterial Pressure
PAPulmonary Artery Pressure
CVPCentral Venous Pressure
ICPIntracranial Pressure
LAPLeft Atrial Pressure
When the default labels P1 and P2 are used, the IP monitoring area displays systolic, diastolic, and
mean pressures. When ICP, LAP, or CVP labels are used, the IP monitoring area displays mean
pressure in large type. Systolic and diastolic pressures are not displayed.
Because pressures can change in a short time, data should be checked regularly during vital sign
monitoring.
How IP Monitoring Works
IP monitoring involves the conversion of fluid pressure into an electrical signal. The conversion is
accomplished with a pressure transducer. The transducer is connected to a patient’s indwelling
pressure catheter using a special assembly of tubing, stopcocks, adapters, flush valves, and fluids,
commonly known as a flush system. The transducer translates the pressure wave into an electrical
signal. A well-functioning flush system is essential for obtaining undistorted waveforms and
accurate information.
4
IP monitoring is available on either Channel 2 or 3. The IP connector (6-pin type 3102A-14S-6S) is
compatible with industry standard (60601-2-34 and AAMI-BP22) pressure transducers with 5μV/V/
mmHg sensitivity. For a list of IP transducers that are compatible with the LIFEPAK 15 monitor/
defibrillator, see the LIFEPAK 15 Monitor/Defibrillator Accessory Catalog at www.physiocontrol.com. If the use of other transducers is desired, the customer must be responsible for
determining if the transducers comply with standards and are compatible with the monitor.
The IP connector pinout has the following configuration, counterclockwise from 12 o’clock, viewed
from the front of the LIFEPAK 15 monitor/defibrillator.
A pin = - signalB pin = + excitationC pin = + signal
An invasive pressure adapter cable is used to connect the transducer to the monitor.
IP Monitoring Procedure
Prepare a flush system according to local protocols. Position the transducer at the patient’s
phlebostatic axis (zero-reference level).
To avoid offset errors, a zero reference must be established before any meaningful pressure
readings are obtained. This is done by opening the transducer stopcock to air so that atmospheric
pressure becomes the reference.
The P1 or P2 connector and Channel 2 or 3 can be used for IP monitoring. P1 and Channel 2 are
used in these instructions.
To monitor IP:
1. Prepare the transducer system according to the operating instructions provided with the
transducer and your local protocol.
2. Press
3. Connect the IP cable to the transducer and to the P1 port on the monitor.
4. Use the default label
5. Use the
6. Open the transducer’s stopcock to air to zero the transducer and remove stopcock cap. Select
7. Close the stopcock to air. The patient’s pressure waveform should be displayed. A scale is
ON.
P1 or select ART, PA, CVP, ICP, or LAP. To change the label, select the P1
area. From the menu, select
SPEED DIAL to outline and select CHANNEL 2 on the Home Screen. From the Channel 2
menu, select
the
P1 area. Select ZERO from the menu. The message P1 ZEROED appears when zeroing is
complete and the pressure values are displayed as zeros.
automatically selected to display the pressure. Confirm that pressure amplitude correlates with
the digital readout.
WAVEFORM and then select the label that is desiredfor the waveform.
P1. Select a label from the list.
Note: If you place a cap on an open port before you close the port to air, an error message may
appear. You will be required to zero the transducer again.
If pressure alarms are desired, set the alarms after you obtain a satisfactory waveform. Error or
alarm messages appear in the message area at the bottom of the screen. For more information, see
"Alarms" on page 3-21.
IP Scale Options
The IP monitor can display pressures from -30 to 300 mmHg. After zeroing the transducer
pressure, the monitor automatically selects one of the following scales based on the patient’s
measured pressure:
• -30 to 30 mmHg
•0 to 60mmHg
• 0 to 120 mmHg
• 0 to 150 mmHg
• 0 to 180 mmHg
• 0 to 300 mmHg
You can also manually select one of these scales or autoscale to readjust the waveform within the
channel.
SPEED DIAL to outline and select the P1 area. The P1 menu appears.
SCALE and then choose a scale from the list.
Cleaning
IP transducers are disposable and are intended for single-patient use. Do not clean and reuse
transducers. Dispose of the contaminated waste according to local protocols.
IP cables are reusable and may be cleaned. To clean the reusable IP cable:
1. Disconnect the cable from the monitor.
2. Use a clean, soft cloth dampened with a germicidal solution to wipe clean.
3. Allow to dry before reconnecting the cable to the monitor.
For information about cleaning the device, see "Cleaning the Device" on page 9-15.
Troubleshooting Tips
The error messages in Table 4-8 use the text PX to represent any of the labels for invasive pressure,
including P1, P2, and the user-selectable labels ART, PA, CVP, ICP, and LAP.
Table 4-8 Troubleshooting Tips for IP Monitoring
OBSERVATIONPOSSIBLE CAUSECORRECTIVE ACTION
Invasive pressure value is blankNo transducer is connected • Connect the transducer to the
cable, and the cable to the
monitor.
No scale appears next to the
waveform
PX NOT ZEROED message appearsThe zero reference has not
PX ZERO FAILED message appearsAn unsuccessful attempt
Dampened waveformLoose connection• Check the entire system for
The zero reference has not
been established
been established
has been made to set a zero
reference value
Tubing too long or too
compliant
Thrombus formation, air
bubbles, or blood left in
catheter after blood draw
• Zero the transducer.
• Zero the transducer.
• Make sure that the transducer
is open to air and repeat the
attempt to zero.
leaks. Tighten all connections.
Replace any defective
stopcocks.
• Use short, stiff tubing with a
large diameter.
• Use syringe to draw back air or
particles in catheter, and then
flush system.
Excessive catheter
movement. Motion of the
catheter tip within the
vessel accelerates fluid
movement in the catheter,
causing artifact to be
superimposed on the
pressure wave, increasing
readings by 10–20 mmHg.
Catheter tip partially clotted• Use syringe to aspirate, and
Catheter migrated distally
in pulmonary artery
• Close stopcock to air before
placing cap on port.
• Change catheter tip position.
• Use mean pressure values
(mean pressure is less
affected by extremes and
therefore reflects a more
accurate reading).
then flush.
• Observe PA waveform before
balloon inflation. Flattening of
the waveform could indicate
wedging with balloon deflated.
Turn patient side to side in
Trendelenburg position, or
stimulate cough in attempt to
dislodge catheter.
• Retract catheter with balloon
deflated until proper position
is obtained.
• Minimize chances of catheter
advancement by firmly
anchoring catheter at insertion
site.
Failure to obtain PWPMalposition of catheter tip• Reposition catheter.
Leak in balloon.
Ruptured balloon.
Progressive elevation of PWPOverinflation• Inflate balloon in small
Catheter migrated distally
in pulmonary artery
For general troubleshooting tips, see Table 9-2 on page 9-18.
increments while watching
scope for confirmation of
wedging. Use only enough air
to wedge. Do not use more
than the volume
recommended by the
manufacturer.
• Reposition catheter.
4-57
Vital Sign and ST Segment Trends
Intended Use
The trends feature of the LIFEPAK 15 monitor/defibrillator provides the ability to graphically
display and document the patient’s vital signs (VS) and ST segment measurements for up to eight
hours. VS trending is intended for use with any patient who requires continuous monitoring of vital
signs over an extended period of time to identify changes in patient condition and to document
patient response to therapy. ST trending is intended for use with patients suspected of having acute
ischemic events, such as unstable angina, and for patients during treatment of an acute ischemic
event. ST segment measurement is initiated using a 12-lead ECG and is derived using the
University of Glasgow 12-Lead ECG Analysis Program.
VS and ST Trends Warning
WARNING
INACCURATE INTERPRETATION OF PATIENT STATUS
Vital sign and ST graphs are tools to be used in addition to patient assessment.
Artifact and noise may produce spurious data. Ensure artifact-free monitoring as much
as possible and assess the patient frequently to confirm the appropriateness of
monitor data.
Each active vital sign can be displayed graphically for time ranges of 30 minutes, and 1, 2, 4, and
8 hours. The vital signs are HR, SpO
mean pressures. Data is sampled every 30 seconds. If valid data is not available, a blank space is
substituted on the graph. NIBP values are plotted only when an NIBP measurement is obtained. VS
measurements are not averaged or filtered. No messages or alarms occur based on changes in VS
measurements.
, SpCO, SpMet, CO2, and RR; and systolic, diastolic, and
ST measurements can be displayed graphically for time ranges of 30 minutes, and 1, 2, 4, and 8
hours. ST trending is initiated by obtaining the patient’s first 12-lead ECG. The ST J-point (STJ) is
the part of the ST segment that is measured (see Figure 4-17). The STJ measurement is plotted on
the ST trend graph (see Figure 4-18).
Figure 4-17 STJ Measurement
When all leads of the 12-lead ECG cable are attached to the patient, STJ measurements are
obtained automatically every 30 seconds. If a lead is off, or the ECG data is too noisy, ST
measurements are not obtained and the graph shows a blank for that time period. If an STJ
measurement in any lead deviates from the initial measurement by 1 mm (0.1 mV) or more and the
deviation persists for 2.5 minutes, the monitor automatically prints another 12-lead ECG.
Interpreting the ST Trend Graph
Using the first 12-lead ECG, the monitor identifies the presence of any STJ displacement, either
negative or positive, and the lead that has the most STJ displacement. When
lead that has the most STJ displacement is shown on the graph. The STJ is measured every 30
seconds thereafter.
Figure 4-18 shows an example of an ST trend graph. The elapsed time goes from right to left across
the screen. The most current STJ measurement is on the far right. Each time an STJ measurement
is obtained, it is compared to the first STJ or baseline measurement. The bars represent the change
in the STJ compared to the first measurement.