Phynix SURFIX PRO X Instruction Manual

Page 1
SURFIX® PRO X
Instruction Manual
WWW.PHYNIX.COM
Page 2
Page 3
USB 2.0 interface
Indicator for Bluetooth connection
Cursor-up-key
Confirm key
Probe socket Cursor-down-key
Escape key
Power key
Page 4
2
Instruction manual
Coating thickness gauge Surfix Pro X
About this manual 4 Meaning of keys 4
1. Measuring gauge, probes and their applications 5
2. Preparing for measurement 6
3. Menu structure 7
4. Calibration (1st menu item) 7
4.1. Activating the works calibration 8
4.2. Zero setting (1-point calibration) 9
4.3. One-foil calibration (2-point calibration) 9
4.4. Two-foil calibration, e.g. for rough surfaces 10
4.5. Cal-Through-Coat calibration (calibration on a coated test object) 11
4.6. Deletion of calibration 12
4.7. Measurement 13
5. Statistics (2nd menu item) 13
5.1. Transfer of statistic values only 14
5.2. Transfer of readings and statistics 15
5.3. Display of statistics on the screen 16
5.4. Display of stored readings on the screen 17
5.5. Deletion of readings and statistics 17
5.6. Quick deletion of readings and statistics 18
5.7. Single-value/block-value statistics 18
5.7.1. Activation of single-value statistics 18
5.7.2. Activation of block-value statistics 19
5.8. Trend display 20
5.9. Histogram display 20
6. Measuring mode (3rd menu item) 21
6.1. Setting of measurement methods: Automatic FN identification, Ferrous mode and Non-Ferrous mode 21
6.2. Setting of limit values 22
6.2.1. Setting of limit values 22
6.2.2. Deletion of limit values 23
6.3. Continuous mode (Scan mode) 23
6.3.1. Activation of the continuous mode with all values 24
6.3.2. Activation of the continuous mode with stable values 24
6.3.3. Deactivation of the continuous mode 25
6.4. Zero offset mode 25
6.4.1. Setting the zero offset value 26
6.4.2. Deleting the zero offset value 26
7. Data memory (4th menu item) 27
7.1. New file 27
7.2. Opening a file 29
Page 5
3
7.3. Closing a file 29
7.4. Renaming the file 30
7.5. Deleting files 31
7.5.1. Deleting a selected file 31
7.5.2. Deleting all files 31
8. File transfer to PC (5th menu item) 32
8.1. Sending an active file to PC 32
8.2. Selecting and sending a stored file to PC 33
8.3. Sending all stored files to PC 34
8.4. Setting the transfer contents (statistics and/or readings) 34
9. Options (6th menu item) 35
9.1. Selection of switch-off modes 35
9.2. Beeper on/off 35
9.3. Measuring unit µm/mils 36
9.4. Welcome screen on/off 36
9.5. Power supply 37
9.6. Setting of date and time 37
9.7. Selecting the language 38
10. Total-Reset to factory settings (Total Reset) 39
11. Display of firmware version number and serial numbers of gauge and probe 39
12. Status information, error messages und malfunctions 40
13. Maintenance / Service 41
14. Technical Data 41
15. Probe overview 42
16. Scope of delivery 43
17. Display 43
18. Changing the probe 44
19. Default settings 44
20. Protect the environment 45
Page 6
4
Introduction
The following operating manual is intended for the Surfix® Pro X instrument family with separate exchangeable probe.
The gauge conforms to the following standards for coating thickness measurement:
Magnetic method: DIN EN ISO 2808
DIN EN ISO 2178 ASTM B499 ISO 19840
Eddy current method: DIN EN ISO 2808 DIN EN ISO 2360
ASTM D7091
About this manual
The well-proven menu technique is applied to the use of this coating thickness gauge. This manual informs you step by step about the correct operation of the gauge Surfix® Pro X. If you follow these step-by-step instructions, it will only take a few minutes before you are able to carry out the first measurements.
Meaning of keys
The operation of Surfix® Pro X is carried out by means of the five buttons below the display.
switching the gauge on and off
The arrow keys have several functions. Depending on the menu item you are currently at, by using these keys you will:
browse through the main menu and submenus set calibration values, limit values and zero offset value browse through stored single readings select a character during the input of a filename select a file
Use this key to: confirm the choice of a menu item, e.g. Statistics 2/6 and
then reach the corresponding submenu, e.g. Transfer of statistics1/6
close a setting action, e.g. Measuring unit µm, or confirm
the calibration or limit value settings
Page 7
5
confirm a charater during the input of a file name confirm a file name (if pressed longer than 2 seconds.
This key enables you to: go back to the next higher menu level up to the start
screen – – – –
delete the last reading immediately after the
measurement
delete readings and statistics quickly (see chapter 5.6) abort a setting action, e.g. when setting calibration and
limit values
delete a character during the input of a filename abort the input of a file name (if pressed longer than 2
seconds).
1. Measuring gauge, probes and their applications
Depending of the used probe the Surfix® Pro X is a coating thickness gauge for non­destructive coating thickness measurement according to the magnetic induction principle and the eddy current principle.
The magnetic induction principle is used for all non-magnetic coatings, such as
varnish, paint, enamel, chromium, copper, zinc... coatings on steel and iron (F probes)
The eddy current principle is used for all electrically insulating coatings, such as
varnish, paint and anodising coatings on non-ferrous metals, even on austenitic stainless steels (N probes)
FN probes are dual probes for measurements on steel and non-ferrous metal (F = Ferrous und N = Non-Ferrous).
The following probes can be connected to the gauge Surfix® Pro X:
Probe
Range
Application
Accuracy
(referred to PHYNIX standards)
1
works calibration
2
1-point calibration
3
2-point calibration
FN1.5
0 … 1500 µm
dual probe for measurements on steel/iron and on non-ferrous metals
1
3 µm or 3 % of reading;
whichever is greater
2
(1 µm + 2 % of reading)
3
(1 µm + 1 % of reading)
F1.5
0 … 1500 µm
for measurements on steel/iron
as FN1.5
N1.5
0 … 1500 µm
for measurements on non-ferrous metals
as FN1.5
FN1.5/90°
0 … 1500 µm
angled dual probe for measure­ments on steel/iron and on non­ferrous metals in tubes or grooves
as FN1.5
Page 8
6
FN1.5R
0 … 1500 µm
angled dual probe for measure­ments on steel/iron and on non­ferrous metals
as FN1.5
F1.5R
0 … 1500 µm
angled probe for measure­ments on steel/iron
as FN1.5
FN0.2
0 … 200 µm
high-precision combination probe for thin layers on steel/iron and on non-ferrous metals
1
2,5 µm or 2,5 % of reading;
whichever is greater
2
(0,7 µm + 1,5 % of reading)
3
(0,7 µm + 1 % of reading)
FN3.5
F: 0 … 3.5 mm N: 0 … 3.0mm
angled dual probe for measure­ments on steel/iron and on non­ferrous metals with extended range
1
5 µm or 3 % of reading;
whichever is greater
2
(2 µm + 2 % of reading)
3
(2 µm + 1 % of reading)
F3.5
0 … 3.5 mm
angled probe for measurements on steel/iron with extended range
as FN3.5
F10
0 … 10 mm
angled probe for measurements on steel/iron with large range
1
10 µm or 3 % of reading;
whichever is greater
2
(5 µm + 2 % of reading)
3
(3 µm + 1 % of reading)
F20
0 … 20 mm
probe for measurements on steel/iron with very large range
1
n.a.
2
(10 µm + 4 % of reading)
3
(10 µm + 2 % of reading)
F10-1 F10-2 F10-3
0 … 6.5 mm
angled probes for measure­ments on steel mesh
1
n.a.
2
(0,1 mm + 5 % of reading)
3
(0,1 mm + 3 % of reading)
F10-cp
0 … 10 mm
angled probes for measure­ments of wall thickness (with counter-pole)
1
n.a.
2
(0,05 mm + 5 % of reading)
3
(0,05 mm + 3 % of reading)
F30T
0 … 30 mm
probe with large range for measurements on steel mesh
1
n.a.
2
(0,2 mm + 5 % of reading)
3
(0,2 mm + 3 % of reading)
F30C
0 … 20 mm
probe with large range for measurements on conveyor belts
1
n.a.
2
(0,2 mm + 5 % of reading)
3
(0,2 mm + 3 % of reading)
You can identify the present probe and its working principle by the colour of the plug’s
reinforcing sleeve:
green FN dual probes (magnetic induction and eddy current principle) blue F probes (magnetic induction principle) yellow N probe (eddy current principle)
Please avoid bending or too tight reeling of the probe cable. Cable breaks due to this are excluded from warranty.
2. Preparing for measurement
Power supply: Surfix Pro® X can be operated using 2 x round cell (AA) 1.5V alkaline batteries or also 2 rechargeable batteries (AA) 1,2V.
Page 9
7
Empty batteries are indicated by a red or empty battery icon. Please remove used batteries at once and dispose of them in an environmentally acceptable manner. When inserting new batteries, please pay attention to the correct polarity. Wrongly inserted batteries may lead to a destruction of the gauge.
In order to avoid any possible loss of data, please pay attention when exchanging batteries: a) Always switch off the gauge by means of the button before exchanging batteries. b) Exchange the batteries within 20 seconds.
After inserting the batteries and switching-on for the first time by means of the button, the display shows a) English as proposed language: confirm with , or use the arrow keys to select another language and confirm with , after which the display shows b) the proposed date and time: confirm with , or use the arrow keys to change and confirm with (pay attention to the instructions on the display).
Normally the gauge works in the automatic switch-off mode. It switches off automatically 90 seconds after the last measurement or operation. For permanent opertion mode see chapter 9.1.
3. Menu structure
There are 6 main menu items: Calibration, Statistics, Measuring mode, Data memory, Transfer to PC and Options. All other menu items or functions are subordinate to these 6
menu items. The operation is carried out by means of the two arrow keys and , the confirmation
key and the escape key . The steps to be carried out are displayed on the screen,
e.g. “Place probe on foil standard“. After pressing the confirmation is displayed, e.g. „Calibration has been set“.
You can find a diagram of the menu structure at the end of this manual. A single look at this structure is sufficient to recognize the applicable menu items to select a desired function.
4. Calibration (1st menu item)
This section informs you about the different calibration modes and the application cases that they are especially suitable for. To meet the demands of coating thickness measurement tasks the gauge has to be checked by means of the standards included in the supply package before the measurement.
Surfix® Pro X has two operating modes: in the Direct mode you can directly measure without any further user interaction. Readings, calibration values and settings are stored consecutively in the direct memory, in case of a FN combination probe either in the F memory or in the N memory. The differentiation between F memory and N memory means also that e.g. setting of limit values has to be done in both measuring modes. The direct mode enables very quick measurements; but all readings are within one single memory (FN probe: two single memories).
Page 10
8
In the data memory mode the readings, calibration values and settings are stored in freely nameable data files. This enables the creation and resumption of many different, also very large test series. Due to the storage also of the calibration values a recalibration is normally not necessary. During a running test session a switchover between F measuring mode and N measuring mode is not possible.
If you decide to use the direct mode you can simply start calibration and measuring without selecting or defining a file name. If the readings should be stored in the data memory you first have to open an existing file or create a new one. After that you can do calibrations, measurements and settings. Calibrations have to be done in the file where the readings should be stored. After closing the file the gauge will be in direct mode again.
Surfix® Pro X and all probes are works calibrated. If the accuracy of the works calibration is not sufficient or if the measurement task makes it necessary you can do some additional calibrations as described below.
A zero calibration on an uncoated specimem reduces measurement errors induced by surface geometry e.g. by curved surface or near edges. Calibration data is preserved during power off. Surfix
®
Pro X uses five different calibration methods (see Technical Data
chapter 14).
The different calibration methods can be done in direct mode or in data memory mode. The usage of different data files helps storing calibration data for different applications.
4.1. Activating the works calibration
This calibration mode is suitable for measurements on flat surfaces larger than 35mm x
35mm / 1.4” x 1.4”. A calibration that has been carried out previously has to be deleted.
You need neither calibration standards nor any uncoated test objects.
Procedure:
1.
press this key; Calibration appears on the display
2.
press this key; Zero setting appears on the display
3.
press this key; Deletion of calibration appears on the display
4.
press this key; Works calibration activated appears on the display followed by the start screen with the four dashes – – – – .
Page 11
9
You can now measure using the works calibration.
4.2. Zero setting (1-point calibration)
For this purpose, you have to use an uncoated test object having similar dimensions and material properties as the coated test object, e.g. when you plan to measure on small cylindrical parts or near edges. The zero plates delivered with your gauge are not intended to be used for zero calibration; instead use these plates together with the calibration foils to check the accuracy of the gauge.
Procedure:
1.
press this key; Calibration appears on the display
2.
press this key; Zero setting appears on the display
3.
press this key; Place probe on “0” appears on the display, and the ZERO symbol in the display begins to flash
4.
place the probe repeatedly perpendicular to the uncoated surface of the test object
5.
press this key. Zero has been set appears briefly on the display followed by the start screen with the four dashes – – – – . The ZERO symbol is visable permanently.
You can now measure using the 1-point calibration.
4.3. One-foil calibration (2-point calibration)
The one-foil-calibration has to be done on an uncoated test object (as with the zero setting). For the one-foil-calibration choose a calibration foil that comes closest to the expected coating thickness.
Procedure:
1.
Carry out ‚zeroing’ (1-point calibration) as described above (please see chapter 4.2).
2.
press this key; Calibration appears on the display
Page 12
10
3.
press this key; Zero setting appears on the screen
4.
press this key one more time; ONE-foil calibration appears on the screen
5.
press this key; Place probe on foil standard appears on the display and the CAL symbol in the display begins to flash
6.
Put the measurement foil on the uncoated test object and place the probe repeatedly on the foil
7. press this key; set standard value appears on the display
8.
use the arrow keys to enter the calibration foil thickness as indicated on the standard
9.
press this key; Calibration has been set appears briefly on the display followed by the start screen with the four dashes – – – – . The CAL symbol is visable permanently.
You can now measure using the ONE-foil calibration.
4.4. Two-foil calibration, e.g. for rough surfaces
This calibration mode is used especially for the measurement on rough surfaces, e.g. on sand-blasted base metals. It is also suitable for low-intensity magnetic test objects when carrying out the measurement either according to the magnetic-inductive method or according to the eddy current method. In this calibration mode, two calibration foils varying in thickness are used. The thickness of the foils should be selected so that the thickness of the thinner foil is 0.5 to 0.9 times that of the expected coating thickness and the thickness of the thicker foil is 1.1 to 1.5 times that of the expected coating thickness.
Procedure:
1.
press this key; Calibration appears on the display
2.
press this key; Zero setting appears on the screen
3.
press this key twice until TWO-foil calibration appears on the display
Page 13
11
4.
press this key; Place probe on 1st foil appears on the display.
5.
Put one of the two calibration foils on the uncoated test object and place the probe repeatedly on the foil
6.
press this key; Set standard value appears on the display
7.
use the arrow keys to enter the calibration foil thickness as indicated on the standard
8.
press this key; Place probe on 2nd foil appears on the display
9.
Put the second calibration foil on the uncoated test object at the same point and place the probe repeatedly on the foil
10. press this key; Set standard value appears on the display.
11.
use the arrow keys to enter the calibration foil thickness as indicated on the standard
12.
press this key; Calibration has been set appears briefly on the display followed by the start screen with the four dashes – – – – .
You can now measure using the Two-foil calibration. In measurements on rough base metals, the average coating thickness is measured across the peaks. The measuring uncertainty increases with increasing roughness.
4.5. Cal-Through-Coat calibration (calibration on a coated test
object)
This calibration method is used if no uncoated test object is available for „zeroing“. The method can only be applied with the probes F1.5 or FN1.5 (F mode). The method may only be applied to cases in which the coating at the calibration point is smooth and reproducible values are measured. Don’t use on structural coats of varnish. The errors of measurement produced may be too high.
Procedure:
1.
press this key; Calibration appears on the display
Page 14
12
2.
press this key; Zero setting appears on the screen
3.
press this key three times until Cal-Through Coat calibration appears on the display
4.
press this key; Place probe on coating appears on the display
5.
Now place the probe repeatedly on the coated test object at a defined calibration point. A mean value is displayed
6. press this key; Place probe on foil standard appears on the display
7.
Select a calibration foil no thicker than the expected coating thickness and put it on the same calibration point. Place the probe then repeatedly on the calibration foil covering the calibration point
8. press this key; Set standard value appears on the display
9.
use the arrow keys to enter the calibration foil thickness as indicated on the standard
10.
press this key; Calibration has been set appears briefly on the display followed by the start screen with the four dashes – – – – .
You can now measure using the Cal-Through-Coat calibration.
4.6. Deletion of calibration
This menu item allows you to delete the active calibration. However, if you would like to overwrite a calibration, a deletion is not necessary beforehand.
Procedure:
1.
press this key; Calibration appears on the screen display
Page 15
13
2.
press this key; Zero setting appears on the screen
3.
press this key once until Deletion of calibration appears on the screen display
4.
press this key; Works calibration activated appears briefly on the display followed by the start screen with the four dashes – – – – .
You can now measure using the works calibration.
4.7. Measurement
This section describes the suitable application cases of the different probes. You can use F probes for measurements on all non-magnetic coatings, e.g. varnish, paint, plastics, chrome, zinc, etc. coatings on iron/steel. The display shows Ferr. You can use N probes for measurements on all electrically insulating coatings, e.g. varnish, paint, plastics, anodized coatings, etc. on non-ferrous metals. The display shows Non-Ferr. The FN combination probes allow both measurements and is factory-set to the measuring mode Auto FN identification. The display shows AUTO FN. This means that the gauge automatically identifies the base material (iron/steel or non-ferrous metal) after postioning of the probe. Subsequently the gauge switches to the corresponding measuring method, magnetic (F) or eddy current method (N). After each measurement the display shows Ferr or Non-Ferr to indicate the base material.
Note: In some cases, especially with varnish on zinc on steel, it is nevertheless recommended to use a fixed measuring method setting, e.g. Ferrous mode (F) for measurements on iron/steel or to Non-Ferrous mode (N) for measurements on non­ferrous metals.
5. Statistics (2nd menu item)
After each measurement Surfix® Pro X stores the readings in a data memory and calculates a variety of statistic parameters. Surfix® Pro X has two kinds of memory:
the data memory in direct mode stores up to 2 x 1000 readings (1000 in F mode,
1000 in N mode)
the file based data memory stores up to 100.000 readings.
In both modes the statistical values are updated after each measurement and stored together with the thickness values. The display shows the number of measurements N, the mean value x¯ , the standard deviation s, the maximum value ↑ of the series and the minimum value ↓ .
Page 16
14
With the option block value statistics Surfix® Pro X can collect a selectable number of readings to a single measurement value. This feature facilitates the data acquisition if mean values are required instead of single values. With this option the statistical values in display become a different meaning. They are now related to a series of mean values instead of a series of single values.
Parameter
Single value statistics
Block value statistics
N
number of measurements
number of mean values
mean value
mean value of the mean values
σ
standard deviation
standard deviation of the mean values
Max
maximum value of the series
maximum value of the mean values
Min
minimum value of the series
minimum value of the mean values
Kvar
Coefficient of variation
Coefficient of variation of the mean values
Cp
Process capability index
Process capability index of the mean values
Cpk
Process capability index
Process capability index of the mean values
All readings recorded and statistics calculated by Surfix® Pro X and belonging to the currently active file can be completely transferred or displayed at the menu item Statistics.
The process capability indices Cp and Cpk are calculated, displayed and transferred only if the limits are set (menu item Measuring mode/Limits).
5.1. Transfer of statistic values only
The gauge calculates a maximum of 8 statistical values from a measurement series (see table above). These can be quickly and simply transferred to a PC in the 2nd menu item Statistics. Transfer can be made immediately if you are in the Direct Mode or in an active file which is still open. If you wish to print or transfer data from a file already closed then you can do this under menu item Transfer to PC.
Data transfer to a PC can be done by means of the USB interface or a Bluetooth connection. The supplied data transfer program PHYNIX.connect receives the data and shows it on the screen.
Procedure:
Page 17
15
1.
Connect Surfix® Pro X with a PC by means of a USB interface cable or a Bluetooth connection. The connection has to be initilized from the PC.
2. press this key twice until Statistics appears on the display
3. press this key; Transfer of statistics appears on the display
4.
press this key. Data is transferred and In progress... appears briefly on the display, followed by the start screen with the four dashes – – – – .
5.2. Transfer of readings and statistics
The statistical values of the currently active file (both direct mode or file mode)can also be transferred together with the single readings
Page 18
16
Procedure:
1.
Connect Surfix® Pro X with a PC by means of a USB interface cable or a Bluetooth connection. The connection has to be initilized from the PC.
2. press this key twice until Statistics appears on the display
drücken; es erscheint Kalibrierung
3. press this key; Transfer of statistics appears on the display
drücken; es erscheint Null-Einstellung
4. press this key; Transfer statistics and readings appears on the display
sooft drücken bis schen der Kalibrierung erscheint
5.
press this key. Data is transferred and In progress... appears briefly on the display, followed by the start screen with the four dashes – – – – .
If you want to transfer data from an already closed file, you can do this at the menu item Transfer to PC.
5.3. Display of statistics on the screen
This section explains how to display the different statistical values of single-value or of block-value statistics.
Procedure:
1. press this key twice until Statistics appears on the display
2. press this key; Transfer of statistics appears on the display
3. press this key twice until Display of statistics appears on the display
4.
press this key. Number of readings appears on the display (with block­value statistics: Number of mean values)
5.
press this key repeatedly; the sequence of statistics appears on the display as follows: Mean value, Standard deviation, Maximum, Minimum, Variation coefficient and the Process capability indices Cp
Page 19
17
and Cpk (if the limits has been set) With block-value statistics the sequence will be: Average of mean values, the Standard deviation of mean values, the Maximum of mean values, the Minimum of mean values, the Variation coefficient of the mean values and the Process capability indices.
6.
press this key to continue measuring. The start screen with the four dashes – – – – is displayed.
5.4. Display of stored readings on the screen
This section explains how to display the stored single values on the screen at any time. Procedure:
1. press this key twice until Statistics appears on the display
2.
press this key; Transfer of statistics appears on the display
3.
press this key three times until Display of readings appears on the display
4. press this key; the first single value appears on the display
5.
press this key repeatedly. The following single values with a
consecutive number are displayed. Symbols ( ↓ , ↑ ) indicate a reading
out of the limits. With the block-value statistics the standard deviation for each block is shown, too.
6.
press this key to continue measuring. The start screen with the four dashes – – – – is displayed.
5.5. Deletion of readings and statistics
This section explains how to delete the readings and statistics either in the Direct Mode or in a currently active file in the data memory. Procedure:
1. press this key twice until Statistics appears on the display
2.
press this key; Transfer of statistics appears on the display
Page 20
18
3.
press this key four times until Deletion of statistics and readings appears on the display
4.
press this key. The statistics are deleted together with the single values. Readings and statistics deleted appears briefly on the display followed by the start screen with the four dashes – – – – .
5.6. Quick deletion of readings and statistics
This procedure offers a simple and fast way to delete all readings and statistical values by circumventing the menu structure.
Procedure:
1.
press this key repeatedly until the start screen with the four dashes – – – – appears on the display
2.
press this key once more; Delete readings and statistics? appears on the display
3.
press this key; Readings and statistics deleted appears briefly on the display followed by the start screen with the four dashes – – – – .
5.7. Single-value/block-value statistics
This section describes the activation of single-value or block-value statistics. With single­value statistics all single readings are statistically evaluated: N, x¯ , σ, maximum, minimum,
Kvar. With block-value statistics a mean value is calculated from every block of readings, e.g. with 5 readings per block. All statistical values refer only to the mean value of the blocks. The block size can be pre-selected individually from 3 to 20 readings for each data file.
5.7.1. Activation of single-value statistics
The single-value statistics is always active at the delivery state and after a total reset. If this should be changed and reactivated, please proceed as follows.
Procedure:
1. press this key twice until Statistics appears on the display
2. press this key. Transfer of statistics appears on the display
Page 21
19
3.
press this key three times. Single-value-/Block-value-statistics appears on the display
4. press this key
5. press this key until Single val.stat. switch on appears on the display
6.
press this key. Single val. stat. switched on appears on the display followed by the start screen with the four dashes – – – – .
The following measurements are now evaluated in the typical statistics mode.
5.7.2. Activation of block-value statistics
Procedure:
1. press this key twice until Statistics appears on the display
2. press this key. Transfer of statistics appears on the display
3.
press this key three times. Single-value-/Block-value-statistics appears on the display
4. press this key
5. press this key until Block val.stat. switch on appears on the display
6. press this key
7. the actual block size and Set block size appear on the display
8.
use these keys to adjust the block size (between 3 and 20)
9.
press this key. Block statistics switched on appears on the display followed by the start screen with the four dashes – – – – .
The following measurements are now evaluated in the block-value statistics mode.
Page 22
20
5.8. Trend display
The trend display shows a chart of the measurement values. To navigate through longer measurement series use the arrow keys and ; the key will bring you back to the start screen. Limits are shown as red lines. Green dots mark values within the limits, red dots show values outside the limits.
Procedure:
1. press this key twice until Statistics appears on the display
2. press this key. Transfer of statistics appears on the display
3. press this key twice. Trend appears on the display
4. press this key; the trend display is shown. You can now
5. press this key to display older values;
6. press this key to display newer values;
7. press this key to get back to the statistics menu;
8. press this key to go to the start screen with the four dashes – – – – .
5.9. Histogram display
The histogram display shows the distribution of the measurement values around the mean value x¯ . Limits are shown as red lines. Green dots mark values within the limits, red dots mark values outside the limits. The key will finish the histogram display.
Procedure:
1. press this key twice until Statistics appears on the display
2. press this key. Transfer of statistics appears on the display
Page 23
21
3. press this key. Histogram appears on the display
4. press this key; the histogram display is shown. You can now
5. press this key to get back to the statistics menu;
6. press this key to go to the start screen with the four dashes – – – – .
6. Measuring mode (3rd menu item)
This section describes the activation of the
measurement method (only possible with FN probes), upper and lower limit value, continuous measuring mode, zero offset function.
6.1. Setting of measurement methods: Automatic FN identification,
Ferrous mode and Non-Ferrous mode
When using FN combination probes the display shows Auto FN. This indicates the automatic base material recognition after positioning of the probe and the activiation of the appropriate measuring method.
In some cases, e.g. when measuring paint on zinc on steel, it may be useful to select the measuring method manually, e.g. selecting Ferrous when measuring the total coating thickness, and Non-Ferrous when measuring the paint thickness on zinc.
Procedure:
1.
press this key three times until Measuring mode appears on the display
2. press this key; Autom. Detection FN, F, N appears on the display
3. press this key; Automatic identification (FN) appears on the display
4.
use these keys to select the desired method: Automatic identification (FN), Ferrous (F) or Non-Ferrous (N)
5.
press this key to confirm your choice. The chosen mode is displayed, e.g. Measuring mode FN selected, followed by the start
Page 24
22
screen with the four dashes – – – – .
6.2. Setting of limit values
This section describes the setting of limit values. To monitor your readings, you can set an upper and a lower limit value. This setting is useful:
during the measurement for the evaluation of readings later on when displaying the single values during the analysis of the readings with the PC.
If the reading is outside of the limit values, you will hear a characteristical beep, the reading is displayed in red and an indicator LIMIT ↑ or LIMIT ↓ at the top of the display shows whether the reading fell below or exceeded the limit values. Readings within the limit values are dispalyed green.
6.2.1. Setting of limit values
Procedure:
1.
press this key three times until Measuring mode appears on the display
2. press this key; Autom. Detection FN, F, N appears on the display
3. press this key; Limits appears on the display
4. press this key; Limits setting appears on the display
5.
press this key; Set lower limit appears on the display and a number as suggestion for the lower limit value
6.
use these keys to adjust the lower limit value; with continued pressing of these keys the set value increases or decreases faster
7.
press this key; Set upper limit appears on the display and a number as suggestion for the upper limit value; the upper limit value must be at least 10 µm higher than the lower limit value
8.
use these keys to adjust the upper limit value; with continued pressing of these keys the set value increases or decreases faster
9.
press this key; Limits have been set appears briefly on the display followed by the start screen with the four dashes – – – – .
Page 25
23
The limit values are set for the measuring method (Ferr, Non-Ferr) which was activated last. When the base material changes the limit values have to be set again, if desired.
When transferring the readings to a PC a “<”-sign in front of the thickness value indicates a reading below the lower limit value; a “>”-sign indicates a reading above the upper
limit value.
6.2.2. Deletion of limit values
Procedure:
1.
press this key three times until Measuring mode appears on the display
2. press this key; Autom. Detection FN, F, N appears on the display
3. press this key; Limits appears on the display
4. press this key; Limits setting appears on the display
5. press this key; Deletion of Limits appears on the display
6.
press this key. Limits deleted appears briefly on the display followed by the start screen with the four dashes – – – – .
6.3. Continuous mode (Scan mode)
This section describes the usefulness and the activation/deactivation of the continuous mode. After activation of the Scan mode, the readings are displayed virtually in continuous mode after the probe has been placed on the test object’s surface. The display rate is maximum 180 readings/minute. The Scan mode is useful: when searching the thinnest or thickest point of a coating. By means of the
minimum-maximum-display (↑ and ↓) in the display you can easily find such points by simply moving the probe over the surface without lifting it.
when using the tube probe in small tubes. The space may be not sufficient to lift the
probe after each measurement.
When measuring large surfaces. You can check rapidly the coating thickness of the
complete surface at different positions without lifting the probe after each measurement. Discrepancies in thickness values can be detected very fast.
Normally Surfix® Pro X only shows a thickness value if the underlying measurements are stable. This function prevents fluctuating or faulty readings and increases the accuracy of
Page 26
24
the gauge. With the continuous mode this function can lead to readings at irregular intervals or even no readings due to the movement of the probe. Therefore you can choose whether you want to measure with all values or only with stable values.
6.3.1. Activation of the continuous mode with all values
Procedure:
1. press this key three times until Measuring mode appears on the display
2. press this key; Autom. Detection FN, F, N appears on the display
3. press this key twice; Continuous mode appears on the display
4. press this key;
5. press this key until Conti-mode with all values appears on the display
6.
press this key. Conti-mode with all values activated appears briefly on the display followed by the start screen with the four dashes – – – – .
The continuous mode is switched on only for the measuring method (Ferr, Non-Ferr) which was activated last. When the base material changes the continuous mode has to be activated again, if desired.
6.3.2. Activation of the continuous mode with stable values
Procedure:
1. press this key three times until Measuring mode appears on the display
2. press this key; Autom. Detection FN, F, N appears on the display
3. press this key twice; Continuous mode appears on the display
4. press this key;
Page 27
25
5.
press this key until Conti-mode with stable values appears on the display
6.
press this key. Conti-mode with stable values activated appears briefly on the display followed by the start screen with the four dashes – – – – .
The continuous mode is switched on only for the measuring method (Ferr, Non-Ferr) which was activated last. When the base material changes the continuous mode has to be activated again, if desired.
6.3.3. Deactivation of the continuous mode
Procedure:
1. press this key three times until Measuring mode appears on the display
2. press this key; Autom. Detection FN, F, N appears on the display
3. press this key twice; Continuous mode appears on the display
4. press this key;
5. press this key until Conti-mode deactivate appears on the display
6.
press this key. Conti-mode deactivated appears briefly on the display followed by the start screen with the four dashes – – – – .
6.4. Zero offset mode
With this mode you can set a constant positive or negative value (zero offset value). This value will be added automatically to the reading after each measurement. The sum of original reading and zero offset value is then shown on the display. Moreover, the zero offset value is always displayed at the top of the display as a reminder (OFFSET).
This measuring mode is useful when it comes to determining the upper layer in a coating system consisting of several layers. If the lower layer(s) is (are) known and show(s) an adequately uniform thickness, the coating thickness of this (these) layer(s) can be set as a negative offset value. After measuring of the total coating thickness the thickness of the lower layer(s) is subtracted (addition of a negative offset value) and the thickness of the top layer is displayed.
Page 28
26
This mode can also be selected for measurements on coated rough metal surfaces. The influence of the roughness is determined by measuring the uncoated rough metal surface and set as a negative zero offset value. The gauge will then indicate the coating thickness across the peaks of the rough surface.
6.4.1. Setting the zero offset value
Procedure:
1.
press this key three times until Measuring mode appears on the display
2. press this key; Autom. Detection FN, F, N appears on the display
3. press this key; Zero offset appears on the display
4. press this key; Zero offset setting appears on the display
5. press this key; Set zero offset value appears on the display
6.
use these keys to adjust the zero offset value; with continued pressing of these keys the set value increases or decreases faster
7.
press this key; Zero offset set appears briefly on the display followed by the start screen with the four dashes – – – – . As indicator the zero offset value can be seen at the top of the display (OFFSET).
After this all readings will be corrected by the zero offset value. Example: with a real coating thickness of 100 µm and a zero offset value of –10 µm the display shows 90 µm. With a real coating thickness of 100 µm and a zero offset value of +10 µm the display will show 110 µm.
The zero offset value is only set for the measuring method (Ferr, Non-Ferr) which was activated last. When the base material changes the zero offset value has to be set again, if desired.
6.4.2. Deleting the zero offset value
Procedure:
1. press this key three times until Measuring mode appears on the display
Page 29
27
2. press this key; Autom. Detection FN, F, N appears on the display
3. press this key; Zero offset appears on the display
4. press this key; Zero offset setting appears on the display
5. press this key; Deletion of zero offset appears on the display
6.
press this key; Zero offset deleted appears briefly on the display followed by the start screen with the four dashes – – – – . The zero offset value indicator (OFFSET) disappears.
7. Data memory (4th menu item)
The Surfix® Pro X has two types of data memory:
1. The direct mode with a maximum of 1.000 storable readings each for the F or N
mode. The direct mode enables instantaneous measurements without any further actions, e.g. the input of a file name. All settings as limit values, the zero offset value etc. are available in the direct mode. These settings remain valid until they are changed or a total reset has been performed.
2. The data files with alphanumerical names. A maximum of 100.000 readings can be
stored in this memory. The file name is input using an character-numeral-structure via the gauge’s keypad. All characters (A-Z, a -z) and numerals (0-9) as well as the space character ( ) can be used for file names. The individual settings, such as calibration values, limit values, zero offset value and scan mode, are likewise stored together with the readings.
7.1. New file
This section describes the creation of a new file and the input of a new, automatically predefined or individual file name.
You can transfer the “currently active settings” that you have defined last in another memory, e.g. calibration values, limit values etc. to the new file. This makes the complete creation process of a new file easier if only the file name is meant to be new but all other settings are meant to be kept. If required, you can of course overwrite or edit the transferred settings at the corresponding menu items.
Or you can give the factory settings to the new file, that means works calibration, no limits, no zero offset value etc. These settings can be changed or adjusted as well.
Procedure:
Page 30
28
1. press this key repeatedly until Data memory appears on the display
2. press this key; New file appears on the display
3.
press this key; Input file name appears on the display together with an automatically predefined file name, e.g. D000, followed by a black cursor █ .
Now you can
accept the predefined file name (continue with step 7.) or modify the predefined file name by means of the character-numeral-structure. For
this purpose
4.
press this key shortly to delete one character (pressing this key longer will abort the file name input); and/or
5.
use these keys to select one character out of the character­numeral-structure; all characters and numerals (A-Z, a-z, 0-9) as well as the space character (the white cursor ) are available. Continued pressing of these keys let the characters run through the structure.
6.
press this key shortly; the selected character is input and the black cursor appears again; repeat steps 5. and 6. until the file name is complete e.g. ABC123. Then
7.
press this key, when the black cursor █ is visible. The new file name appears on the display, e.g. D001 or ABC123, and Create file?
8.
press this key; Using actual calibration appears on the display. Now you can choose whether you
a. press this key to create the new file with the actual settings from direct mode or another file. File opened appears briefly on the display followed by the start screen with the four dashes – – – – . Or whether you
b. press this key; Using works calibration appears on the display. With this option you choose for the works calibration; all other settings such as limit values etc. are deleted.
c. press this key to create the new file with the works calibration. File opened appears briefly on the display followed by the start screen with the four dashes – – – – .
Page 31
29
Now you can measure in the newly created file. The file name is shown at the bottom line of the display to indicate the data file mode.
You should close the file after your measurements to get back to direct mode. The direct mode enables quick measurements on different base materials without any effort. The file will also be closed automatically after creating or opening of another file.
7.2. Opening a file
This section describes the process of opening existing files.
Procedure:
1. press this key repeatedly until Data memory appears on the display
2. press this key; New file appears on the display
3. press this key; Open file appears on the display
4.
press this key; Select file appears on the display and the name of the file used at last e.g. D001.
5. use these keys to select a file
6.
press this key; File opened appears briefly on the display followed by the start screen with the four dashes – – – – .
The file name is shown at the bottom line of the display to indicate the data file mode.
7.3. Closing a file
This section describes the process of closing an open file. The Direct mode will become active enabling quick measurements on different base materials without any effort.
Procedure:
1. press this key repeatedly until Data memory appears on the display
Page 32
30
2. press this key; New file appears on the display
3. press this key twice; Close file appears on the display
4.
press this key; File closed appears briefly on the display followed by the start screen with the four dashes – – – – .
After closing of the file Direct mode appears on the display.
7.4. Renaming the file
This section describes the process of renaming an open, active file. To rename a closed file you have to open it first (see chapter 7.2).
Procedure:
1. press this key repeatedly until Data memory appears on the display
2. press this key; New file appears on the display
3. press this key three times; Rename file appears on the display
4.
press this key; Rename file appears on the display together with the actual file name, e.g. D000, followed by a black cursor █ .
5.
press this key shortly to delete one character (pressing this key longer will abort the file name input); and/or
6.
use these keys to select one character out of the character­numeral-structure; all characters and numerals (A-Z, a-z, 0-9) as well as the space character (the white cursor ) are available. Continued pressing of these keys let the characters run through the structure.
7.
press this key shortly; the selected character is input and the black cursor appears again; repeat steps 6. and 7. until the file name is complete e.g. ABC123. Then
8.
press this key, when the black cursor █ is visible. The new file name appears on the display, e.g. D001 or ABC123, and Rename file? If Name already exists appears on the display choose another name for the file.
Page 33
31
9.
press this key; File renamed appears briefly on the display followed by the start screen with the four dashes – – – – .
The new file name is shown at the bottom line of the display to indicate the data file mode.
7.5. Deleting files
The two sections below describe the processes of deleting one selected file or all stored files.
7.5.1. Deleting a selected file
This section describes the process of deleting a specially selected file. The file name, the readings and the settings of the selected file are deleted.
Procedure:
1. press this key repeatedly until Data memory appears on the display
2. press this key; New file appears on the display
3. press this key four times; Delete file appears on the display
4. press this key; Delete one file appears on the display
5.
press this key; Select file appears on the display and the name of the file used at last e.g. D001
6. use these keys to select a file to be deleted e.g. D002
7.
press this key; Delete file? and the selected file name appears on the display
8.
press this key; File deleted appears briefly on the display followed by the start screen with the four dashes – – – – .
7.5.2. Deleting all files
This section describes the process of deleting all files. All file names, readings and settings are deleted. The contents of the Direct mode are not deleted by this action.
Page 34
32
Procedure:
1. press this key repeatedly until Data memory appears on the display
2. press this key; New file appears on the display
3. press this key four times; Delete file appears on the display
4. press this key; Delete one file appears on the display
5. press this key; Delete all files appears on the display
6. press this key; Delete all files? appears on the display
7.
press this key; All files deleted appears briefly on the display followed by the start screen with the four dashes – – – – .
After deleting, Direct appears on the display and you can now measure in the Direct mode.
8. File transfer to PC (5th menu item)
This section describes
1. the process of transferring one or all files to a smartphone, tablet computer or PC,
2. the process of selecting a file to be transferred,
3. the predefinition of transfer contents whether only the readings or only the statistics
or both should be transferred in the future transfer processes.
File transfer requires the data transfer software PHYNIX.connect which is delivered with your gauge.
8.1. Sending an active file to PC
This section describes the transfer process of the currently open file including readings and statistics via Bluetooth or USB interface of the Surfix® Pro X.
Alternatively you can transfer data from the actual file by means of the main menu item Statistics (see chapter 5.1 Transfer of statistic values or 5.2 Transfer of readings and statistics).
Page 35
33
Procedure:
1.
Connect Surfix® Pro X by means of an USB interface cable or a Bluetooth connection with the PC. The Bluetooth connection has to be initialized from the PC.
2. press this key three times; File transfer appears on the display
3. press this key; Send active file appears on the display
4.
press this key; In progress... appears briefly on the display followed by the start screen with the four dashes – – – – .
You can now continue measuring either in the currently active file or, after closing, in another file.
8.2. Selecting and sending a stored file to PC
Procedure:
1.
Connect Surfix® Pro X by means of an USB interface cable or a Bluetooth connection with the PC. The Bluetooth connection has to be initialized from the PC.
2. press this key three times; File transfer appears on the display
3. press this key; Send active file appears on the display
4. press this key; Select file appears on the display
5. press this key; Select file appears on the display
6. use these keys to select a file to be transferred e.g. D002
7.
press this key; In progress... appears briefly on the display followed by the start screen with the four dashes – – – – .
Page 36
34
8.3. Sending all stored files to PC
This section describes the process of transferring all files to PC. The data stored in the Direct mode is not transferred in this process.
Procedure:
1.
Connect Surfix® Pro X by means of an USB interface cable or a Bluetooth connection with the PC. The Bluetooth connection has to be initialized from the PC.
2. press this key three times; File transfer appears on the display
3. press this key; Send active file appears on the display
4. press this key twice; Send all files appears on the display
5.
press this key; In progress... appears briefly on the display followed by the start screen with the four dashes – – – – .
You can now continue measuring either in the currently active file or, after closing, in another file.
8.4. Setting the transfer contents (statistics and/or readings)
This section describes the presetting of the transfer contents, i.e. whether only the readings, only the statistics or both readings and statistics should be transferred.
Procedure:
1. press this key three times; File transfer appears on the display
2. press this key; Send active file appears on the display
3. press this key; Transfer content appears on the display
4. press this key;
5.
use these keys to select the desired transfer content: Statistics and readings, Statistics or Readings
Page 37
35
6.
press this key; Your choice, e.g. Transfer content: stat. + readings, appears briefly on the display followed by the start screen with the four dashes – – – – .
9. Options (6th menu item)
With this menu item you can define the settings described below. They are kept after the gauge switch-off. At the delivery state or after a total reset, always the first menu item of the corresponding menu is set (e.g. µm/mm).
9.1. Selection of switch-off modes
With this menu item you can choose between permanent operation of the gauge or auto-shut-off after 90 seconds without measurement or handling.
Procedure:
1. press this key twice; Options appears on the display
2. press this key; Switch-off mode appears on the display
3. press this key;
4.
use these keys to select the desired mode: Auto switch-off after 90 sec. or Permanent operation
5.
press this key for confirmation; your choice, e.g. Permanent operation selected, appears briefly on the display followed by the start screen with the four dashes – – – – .
9.2. Beeper on/off
With this menu item you can choose between silent operation of the gauge or a short beep after each reading and operation.
Procedure:
1. press this key twice; Options appears on the display
2. press this key; Switch-off mode appears on the display
Page 38
36
3. press this key; Beeper appears on the display
4. press this key;
5.
use these keys to select the desired mode: Beeper ON or Beeper OFF
6.
press this key for confirmation; your choice, e.g. Beeper switched ON, appears briefly on the display followed by the start screen with the four dashes – – – – .
9.3. Measuring unit µm/mils
With this menu item you can choose the measuring unit. The readings can be displayed using the metric system (µm/mm) or alternatively the imperial system (mils).
Procedure:
1. press this key twice; Options appears on the display
2. press this key; Switch-off mode appears on the display
3. press this key twice; Measuring unit appears on the display
4. press this key;
5. use these keys to select the desired mode: µm/mm unit or mils unit
6.
press this key for confirmation; your choice, e.g. Unit µm/mm selected, appears briefly on the display followed by the start screen with the four dashes – – – – .
9.4. Welcome screen on/off
With this menu item you can deactivate the welcome screen which appears after power on.
Procedure:
Page 39
37
1. press this key twice; Options appears on the display
2. press this key; Switch-off mode appears on the display
3. press this key three times; Welcome screen appears on the display
4. press this key;
5.
use these keys to select the desired mode: Welcome screen ON or Welcome screen OFF
6.
press this key for confirmation; your choice, e.g. Welcome screen switched ON, appears briefly on the display followed by the start screen with the four dashes – – – – .
9.5. Power supply
With this menu item you can set the currently used power supply. This setting is necessary for a correct display of the battery status.
Procedure:
1. press this key twice; Options appears on the display
2. press this key; Switch-off mode appears on the display
3. press this key three times; Power supply appears on the display
4. press this key; the current setting is displayed, e.g. Batteries
5.
use these keys to select the desired power supply: Batteries or Rechargeable batteries
6.
press this key for confirmation; your choice, e.g. Power supply with batteries, appears briefly on the display followed by the start screen with the four dashes – – – – .
9.6. Setting of date and time
With this menu item you can set the real time clock of the gauge.
Page 40
38
Procedure:
1. press this key twice; Options appears on the display
2. press this key; Switch-off mode appears on the display
3. press this key twice; Date + time appears on the display
4.
press this key; the actual date and time appears on the display and Adjust?
5. press this key; Set year appears on the display
6. use these keys to select the desired year
7. press this key
8.
repeat step 6. and 7. four times to adjust month, day, hour and minute. At the end Date + time have been set appears briefly on the display followed by the start screen with the four dashes – – – – .
After battery replacement and switching on for the first time an initial menu item apperas on the display with a) English as suggested language: confirm with or select another language with the
arrow keys ; confirm with ; then appears
b) a suggestion for Date and Time: confirm with or adjust date and time with the
arrow keys ; confirm with ; (see above).
9.7. Selecting the language
With this menu item you can choose the menu language. Normally you can choose from English, German, French, Italian, Spanish, Turkish, Czech and Chinese; more languages are available on request.
Procedure:
1. press this key twice; Options appears on the display
Page 41
39
2. press this key; Switch-off mode appears on the display
3. press this key; Language appears on the display
4.
press this key; the actual language appears on the display, e.g. English
5.
use these keys to select the desired menu language, e.g. Deutsch
6.
press this key for confirmation; your choice, e.g. Sprache Deutsch ausgewählt, appears briefly on the display followed by the start screen with the four dashes – – – – .
10. Total-Reset to factory settings (Total Reset)
With this function you can reset the gauge to factory settings.
That means: a) Readings, statistics, calibration settings, limit values and zero offset values in the direct
mode will be deleted (both F- and N-mode) b) works cailbration is activated (both F- and N-mode) c) continious mode is deactivated (both F- and N-mode) d) the gauge is set to automatic FN identification (if available) e) reset of all options to the first menu item, e.g. μm/mm or Beeper ON
Procedure:
1. switch off the gauge; a probe has to be connected
2. press this key and keep it pressed
3.
switch on the gauge; you will hear a longer lasting beep; the welcome screen comes up followed by the start screen with the four dashes – – – – .
11. Display of firmware version number and serial numbers of
gauge and probe
Procedure:
Page 42
40
1. switch off the gauge; a probe has to be connected
2. press this key and keep it pressed
3.
switch on the gauge; after releasing the keys you will see the version number of the firmware and its release date, the serial number of the gauge and probe and at last the start screen with the four dashes – – – – .
12. Status information, error messages und malfunctions
Different status information and error messages are shown on the display.
PC Remote Control
This message appears when the instrument is switched on with no probe connected or when the probe is damaged or not correctly connected. During this operational state the automatic switch-off function is deactivated. Switch off the gauge with the
power button .
No measurement possible
This message comes up if the probe has changed while a file was open. Close this file and proceed.
Batteries empty
Preferably replace empty batteries immediately to avoid battery leakage. Mind the polarity. Replace the batteries within 20 seconds to preserve your data in direct mode. In data file mode all data will be kept also without batteries.
Probe too close to metal
This message comes up if the probe was too close to metal during power on. During this phase some internal balancing measurements are done, and the probe may not be in close distance to metal. Keep the probe away from metal during power on. If this message comes up anyway this may be an indication for a damaged probe.
Magnetic interfering field
There are strong magnetic fields near by the gauge (e.g. transformer, power supplies etc.) Use the gauge in an environment without interfering fields. If this message comes up anyway this may be an indication for a damaged probe or gauge.
Probe defective
This message indicates a damaged probe. The probe together with the gauge should be sent for repair to your dealer or to the manufacturer PHYNIX.
Page 43
41
Other Errors
the gauge does not measure anymore the readings are implausible the keys are locked the gauge can not be switched on or off
In these cases normally a total reset provides remedy (see chapter 10).
Do this as follows:
remove the batteries for a short time and insert them again perform a total reset (switch off the gauge , press and keep it pressed, switch on
the gauge ; you will hear a longer lasting beep) maybe the batteries have to be replaced .
13. Maintenance / Service
The Surfix® Pro X gauge requires no special maintenance. However, it should be handled with care just as any other gauge. Please avoid bending or too tight reeling of the probe cable. Remove the batteries before longer lasting nonuse. Remove used batteries at once and dispose them in an environmentally acceptable manner.
Please send a damaged gauge or a gauge showing faulty operation to your dealer or to the manufacturer PHYNIX.
14. Technical Data
Resolution
0,1µm or <0,2% of reading (probes with ranges up to 1.5 mm) 1µm or <0,2% of reading (probes with ranges larger than 1.5 mm)
Display
high resolution color graphic display, backlight
Operator guidance
menu-driven, language selectable from English, German, French, Chinese; more languages on request
Memory
Direct mode: max. 1.000 readings each in F-mode and N-mode Data file mode: max. 100.000 readings
Calibration
Works calibration Zeroing (One-point calibration) Foil calibration (Two-point calibration) Two-foil calibration Cal-Through-Coat calibration
Zero offset
Addition of a constant value to the reading
Statistics
N, x¯ , σ, Max, Min, Cp, Cpk, Kvar
Online- Statistics
N, x¯ , σ, Max, Min
Limit values
adjustable with acoustic signal and indicator in the display
Interface
USB 2.0, Bluetooth 4.0
Page 44
42
Ambient temperature
0...50 °C
Power supply
2 x Batteries AA 1,5V or 2 x Rechargeable Batteries AA 1,5V 1,2V
Sizes / Weight
137 mm x 66 mm x 23 mm (H x B x T) 150g (incl. Batt.)
Type of protection
IP 52 (Protection against dust and dripping water)
15. Probe overview
Probe
Range
Application
Measuring uncertainty
(referred to PHYNIX standards)
1
works calibration
2
zero setting
3
foil calibration
FN1.5
0 … 1500 µm
combination probe for measurements on steel/iron and non-ferrous metals
1
3 µm or 3 % of reading;
whichever is greater
2
(1 µm + 2 % of reading)
3
(1 µm + 1 % of reading)
F1.5
0 … 1500 µm
for measurements on steel/iron
as FN1.5
N1.5
0 … 1500 µm
for measurements on non-ferrous metals
as FN1.5
FN1.5/90°
0 … 1500 µm
angled combination probe for measurements on steel/iron and non-ferrous metals in bores and grooves
as FN1.5
FN1.5R
0 … 1500 µm
angled combination probe for measurements on steel/iron and non-ferrous metals
as FN1.5
F1.5R
0 … 1500 µm
angled probe for measurements on steel/iron
as FN1.5
FN0.2
0 … 200 µm
high precision combination probe for measurements of thin layers on steel/iron and non-ferrous metals
1
2,5 µm or 2,5 % of reading;
whichever is greater
2
(0,7 µm + 1,5 % of reading)
3
(0,7 µm + 1 % of reading)
FN3.5
F: 0 … 3.5 mm N: 0 … 3.0mm
angled combination probe with extended range for measurements on steel/iron and non-ferrous metals
1
5 µm or 3 % of reading;
whichever is greater
2
(2 µm + 2 % of reading)
3
(2 µm + 1 % of reading)
F3.5
0 … 3.5 mm
angled probe with extended range for measurements on steel/iron
as FN3.5
F10
0 … 10 mm
angled probe with large range for measurements on steel/iron
1
10 µm or 3 % of reading;
whichever is greater
2
(5 µm + 2 % of reading)
3
(3 µm + 1 % of reading)
Page 45
43
F20
0 … 20 mm
angled probe with very large range for measurements on steel/iron
1
n.a.
2
(10 µm + 4 % of reading)
3
(10 µm + 2 % of reading)
F10-1 F10-2 F10-3
0 … 6.5 mm
angled probe for measurements on steel mesh
1
n.a.
2
(0,1 mm + 5 % of reading)
3
(0,1 mm + 3 % of reading)
F10-cp
0 … 10 mm
for side wall measurement with counter piece
1
n.a.
2
(0,05 mm + 5 % of reading)
3
(0,05 mm + 3 % of reading)
F30T
0 … 30 mm
angled probe with very large range for measurements on steel mesh
1
n.a.
2
(0,2 mm + 5 % of reading)
3
(0,2 mm + 3 % of reading)
F30C
0 … 20 mm
angled probe with very large range for measurements on conveyor belts
1
n.a.
2
(0,2 mm + 5 % of reading)
3
(0,2 mm + 3 % of reading)
16. Scope of delivery
Surfix® Pro X gauge. Probes are not part of the standard delivery und have to be
ordered separately. Usually they can be stored in the carrying case of the gauge.
Protective rubber cover
2 batteries AA
USB cable
Instruction manual
Manufacturer’s certificate
Data transfer software PHYNIX.connect (works with Microsoft
®
Excel®)
Carrying case for transport and storage
Optional accessories:
Probes from PHYNIX probe program
Manufacturer’s certificate type M DIN 55350 for gauge and probe(s)
Calibration standards with various thickness values
Probe guide for precise measurements on small parts
Compendium on coating thickness measurement
17. Display
Beside the reading and the menu structure the display shows other useful information icons and notes. The following characters and icons appear on the display: All segments and characters of the display can be displayed for checking purposes.
Bluetooth connection established
a red or empty battery symbol indicates low battery voltage
ZERO zero calibration has been carried out ZERO flashing zero calibration has begun but not yet finished CAL1F one-foil-calibration has been carried out CAL1F flashing one-foil-calibration has begun but not yet finished CAL2F two-foil-calibration has been carried out CAL2F blinkend two-foil-calibration has begun but not yet finished
Page 46
44
LIMIT limit values are set LIMIT flashing setting the limit values has begun but not yet finished
LIMIT ↑, LIMIT ↓ reading exceeds or fall below the limit values
AUTO FN Automatic FN identification is active Ferr Ferrous mode is active (for iron/steel) Non-Ferr Non-Ferrous mode is active (for non-ferrous metals) μm measuring unit is µm mm measuring unit is mm mils measuring unit is mils
18. Changing the probe
PHYNIX offers a variety of different probes. Surfix® Pro X automatically recognizes the connected probe and sets the specific probe parameters, e.g. measurement range, resolution, measurement method.
Please observe the following points:
Switch off the gauge before changing the probe.
Important: If you have carried out measurements in the Direct Mode before changing
the probe then you should transfer the data beforehand to a PC. The measurement values stored in the Direct Mode will be deleted after changing the probe and the works calibration will be activated.
When working in the File Mode you should close the file before changing the probe. If
the file stay open the message “No measurements possible” will be displayed after
power on. Close the file and proceed with the measurements. You can open a new file, open the same file again or work in the Direct Mode.
19. Default settings
This section describes the gauge settings
at initial start-up
after total reset
if the batteries have been removed for a longer period of time.
Gauge settings
Condition after first-time switch-on
Condition after total reset
Calibration
Works calibration
Works calibration
Single­value/block­value statistics
Single-value statistics
Single-value statistics
Online statistics display
N, x¯ , σ, Max, Min
N, x¯ , σ, Max, Min
Autom. FN, Ferr, Non-Ferr
Autom. FN (if available)
Autom. FN (if available)
Limit values
off
off
Continuous mode
off
off
Page 47
45
Zero offset
off
off
File
none
none
Transfer contents
Statistics
Statistics
Switch-off mode
automatic
automatic
Beeper
on
on
Units of measurement
µm
µm
Welcome screen
on
on
Date + Time
has to be set
unchanged
Language
has to be set
unchanged
After a battery interruption lasting more than 20 seconds the gauge behaves as after the total reset. In addition, the user is prompted to select the language and set the date and time.
20. Protect the environment
Your gauge contains a number of different reusable materials. Please do not throw your instrument away with the normal household waste but bring it to a special waste disposal depot for electrical equipment (material disposal depot).
Page 48
EIN-Folien
EIN-Folien
Kalibrierung 2/6
Surfix menu structure
®
Pro X
Two-foil
calibration 3/6
Set 1st standard value
Place probe on
2nd foil standard
Zero setting
1/6
One-foil
calibration 2/6
Set standard value
Zero has been set
Calibration has been set
Set standard value
Calibration has been set
Deletion of
calibration 6/6
Place probe on
1st foil standard
Place probe on "0"
Place probe on
foil standard
Place probe on
coating
Works calibration
activated
Transfer of
statistics 1/8
Display of
statistics 3/8
Single-value/block-value
statistics 6/8
Single val.stat.
switch on 1/2
Single val.stat.
switched on
Set block size
Block val.stat.
switch on 2/2
Set lower limit
Deletion of
limits 2/2
Set upper limit
Limits deleted
Limits
2/4
Limits have been set
Limits setting
1/2
Autom. detection
FN, F. N 1/4
Automatic
identification (FN) 1/3
Ferrous (F)
2/3
Non-Ferrous (N)
3/3
Measuring mode
FN selected
Measuring mode
F selected
Measuring mode
N selected
Conti-mode
deactivate 1/3
Conti-mode
deactivated
Close file
3/5
New file
1/5
Open file
2/5
Rename file
4/5
Rename file?
Delete file
5/5
File closed
Input file name
Select file
Input file name
Delete one
file 1/2
Delete all
files 2/2
Select file
Delete all files?
File: 1234ABCD
delete?
File deleted
Transfer of sta
and readings 2/8
tistics
Deletion of stat
and readings 5/8
istics
Display of
readings 4/8
Zero offset
4/4
Zero offset
setting 1/2
Set zero offset value
Zero offset set
Deletion of
zero offset 2/2
Zero offset deleted
Conti-mode with
stable values 2/3
Conti-mode with stable
values activated
Using actual
calibration 1/2
Using works
calibration 2/2
Place probe on
foil standard
Cal-Through-Coat
calibration 4/6
Continious
mode 3/4
Conti-mode with
all values 3/3
Conti-mode with all
values activated
Set 2nd standard value
Create file?
File opened
File opened
File opened
File renamed
All files deleted
Display active
calibration 5/6
Calibration
1/6
Statistics
2/6
Measuring mode
3/6
Data memory
4/6
Block val.stat.
switched on
Calibration has been set
Trend
7/8
Histogram
8/8
Page 49
Switch-off mode
1/7
Auto switch-off
after 90 sec. 1/2
Permanent
operation 2/2
Auto switch-off after
90 sec. selected
Permanent
operation selected
Beeper
2/7
Beeper ON
1/2
Beeper OFF
2/2
Beeper switched ON
Beeper switched OFF
Measuring unit
3/7
Measuring unit
µm/mm 1/2
Measuring unit
mils 2/2
Measuring unit
µm/mm selected
Measuring unit
mils selected
Statistics and
readings 1/3
Statistics
2/3
Transfer content:
stat. + readings
Transfer content:
statistics
Readings
3/3
Transfer content:
readings
Quick deletion of statistics
and readings
1. until "- - - -"
2.
3.
Power on/off
a) Scrolling through the menus
b) Setting of calibration, limit, offset values
c) Display of statistics and readings
a) Confirming of selected menu
b) Finishing an adjustment
a) Back to higher menu level
b) Deletion of the last reading immediate
after the measurement
d) Aborts an adjustment
c) Quick deletion of statistics and readings
d) Character selection during file name input
e) File selection
c) Confirming a character during file name input
d) Confirming a file name
f) Aborts file name input, if pressed longer
than 2 sec.
e) Deletes a character during file name input
Send active file
1/4
In progress...
Select file
2/4
Select
File
Send all files
3/4
In progress...
File transfer
5/6
+
Total Reset
Transfer content
4/4
Welcome screen
4/7
Welcome screen
ON 1/2
Welcome screen
OFF 2/2
Welcome screen
switched ON
Language
7/7
English
1/3
Deutsch
2/3
Français
3/3
Date + time
6/7
01-Jan-2010 00:00:00
Adjust?
Set date + time
Options
6/6
Power supply
5/7
Batteries
1/2
Rechargeable
batteries 2/2
Power supply with
rechargeable batteries
Welcome screen
switched OFF
Power supply with
batteries
In progress...
Language English
selected
Sprache Deutsch
ausgewählt
Option langue
Français
Page 50
Remarks:
Page 51
Page 52
PHYNIX GmbH & Co. KG Siemensstr. 14 D-41469 Neuss Germany Tel +49-2137-10978-0 Fax +49-2137-10978-28 www.phynix.com info@phynix.de
Loading...