Photon Systems Instruments PlantPen PRI Series, PlantPen NDVI Series, PlantPen PRI 210, PlantPen NDVI 310 Manual And User Manual

PlantPen
NDVI & PRI
Manual and User Guide
Please read this manual before operating this product
PSI, spol. s r. o., Drásov 470, 664 24 Drásov, Czech Republic FAX: +420 511 440 901, TEL: +420 511 440 011, www.psi.cz
© PSI (Photon Systems Instruments), spol. s r. o.
2
© PSI (Photon Systems Instruments), spol. s r. o. (hereinafter PSI), 2018
This document and its parts can be copied or provided to a third party only with the express permission of PSI.
The contents of this manual have been verified to correspond to the specifications of the device. However, deviations cannot be ruled out. Therefore, a complete correspondence between the manual and the real device cannot be guaranteed. The information in this manual is regularly checked, and corrections may be made in subsequent versions.
The visualizations shown in this manual are only illustrative.
This manual is an integral part of the purchase and delivery of equipment and its accessories and both Parties must abide by it.
© PSI (Photon Systems Instruments), spol. s r. o.
3
© PSI (Photon Systems Instruments), spol. s r. o.
4
TABLE OF CONTENTS
1 Information Before Using PlantPen Device ..................................................................................................................... 6
2 Technical Specification .................................................................................................................................................... 7
3 General Information ....................................................................................................................................................... 8
3.1 Device Description ................................................................................................................................................. 9
4 List of Equipment and Customer Information .............................................................................................................. 10
5 Care and Maintenance .................................................................................................................................................. 11
6 Principle of Measurement ............................................................................................................................................ 12
7 How to Get Started ....................................................................................................................................................... 13
7.1 PRI Measurement ................................................................................................................................................ 13
7.1.1 Calibration ....................................................................................................................................................... 13
7.1.2 Measurement .................................................................................................................................................. 13
7.2 NDVI Measurement ............................................................................................................................................. 14
7.2.1 Calibration ....................................................................................................................................................... 14
7.2.2 Measurement .................................................................................................................................................. 14
8 Control Menu Tree ........................................................................................................................................................ 15
9 USB Connection ............................................................................................................................................................ 20
10 Bluetooth Connection .............................................................................................................................................. 21
10.1 Bluetooth Pairing ................................................................................................................................................. 21
11 FluorPen Software ................................................................................................................................................... 25
11.1 Software Installation ........................................................................................................................................... 25
11.2 Menu and Icon Explanation ................................................................................................................................. 26
11.2.1 Main Menu ................................................................................................................................................. 26
11.2.2 Menu Settings ............................................................................................................................................. 27
11.2.3 Menu Online Control .................................................................................................................................. 28
11.3 Data Transfer and Visualization ........................................................................................................................... 29
11.4 Firmware Update ................................................................................................................................................. 31
12 GPS Module ............................................................................................................................................................. 33
12.1 GPS / PlantPen Operation ................................................................................................................................... 33
12.2 Data Download .................................................................................................................................................... 34
© PSI (Photon Systems Instruments), spol. s r. o.
5
13 Warranty Terms and Conditions .............................................................................................................................. 35
14 Troubleshooting and Customer Support ................................................................................................................. 36
© PSI (Photon Systems Instruments), spol. s r. o.
6
1 INFORMATION BEFORE USING PLANTPEN DEVICE
Read this manual carefully before operating the device. If you are not sure about anything in the manual, contact the manufacturer for clarification.
By accepting the device, the customer agrees to follow the instructions in this guide.
Always follow corresponding manuals while working with the PlantPen device or doing the maintenance. It is forbidden to interfere with the hardware or software of the PlantPen device in any way without previous agreement
with the manufacturer.
The following table presents basic highlight symbols used in this manual:
Symbol
Description
Important information, read carefully.
Complementary and additional information.
Tab. 1 Used symbols.
© PSI (Photon Systems Instruments), spol. s r. o.
7
2 TECHNICAL SPECIFICATION
Measured Parameters:
PRI (Photochemical Reflectance Index) = (R531 – R570) / (R531 + R570) Reference: Sellers et al. (1985) NDVI (Normalized Difference Vegetative Index) = (R740 – R660) / (R740 + R660) Reference: Rouse et al. (1974)
Measuring Light:
Internal dual wavelength light source R531 = 531 nm, R570 = 570 nm for PRI 210 Internal dual wavelength light source VIS = 635nm, NIR = 760nm for NDVI 310
Detector Wavelength Range:
PIN photodiode with 500 to 600 nm bandpass filters for PRI 210 PIN photodiode with 620 to 750 nm bandpass filters for NDVI 310
PC Communication modules:
Bluetooth and USB
FluorPen Software:
Windows 7 or higher compatible*
Memory Capacity:
Up to 16 Mb
Internal Data Logging:
Up to 100,000 data points
Display:
Graphical display
Keypad:
Sealed, 2-key tactile response
Keypad Escape Time:
Turns off after 5 minutes of no use
Power Supply:
Li-Ion rechargeable battery
Battery Capacity:
2000 mAh
Max. Charging Current:
0.5 A
Charging via USB port:
PC, power bank, USB charger, etc.
Battery Life:
48 hours typical with full operation
Low Battery Detection:
Low battery indication displayed
Size:
135 mm x 65 mm x 33 mm
Weight:
188 g
Sample Holder:
Mechanical leaf clip
Operating Conditions:
Temperature: 0 to +55 ºC, 32 to 130 ºF; Relative humidity: 0 to 95 % (non-condensing)
Storage Conditions:
Temperature: -10 to +60 ºC, 14 to 140 ºF; Relative humidity: 0 to 95 % (non-condensing)
Warranty:
1 year parts and labor (see the last page of this Operation Manual for precise warranty conditions)
* Windows is a registered trademark of Microsoft Corporation.
© PSI (Photon Systems Instruments), spol. s r. o.
8
3 GENERAL INFORMATION
The PlantPen model PRI 210 measures Photochemical Reflectance Index (PRI) in two narrow wavelength bands centered close to 531 nm and 570 nm. The photochemical reflectance index (PRI = (R531–R570)/(R531+R570)) correlates with the epoxidation state of xanthophyll cycle pigments and photosystem efficiency in many studies under diverse conditions, such as different 1. PRI is sensitive to changes in carotenoid pigments that are indicative of photosynthetic light use efficiency, the rate of carbon dioxide uptake, or as a reliable water-stress index. As such, it is used in studies of vegetation productivity and stress.
The PlantPen model NDVI 310 measures Normalized Difference Vegetation Index, which is an important indicator of
chlorophyll content in plants. The pigment in plant leaves, chlorophyll, strongly absorbs visible light (from 0.4 to 0.7 µm)
for use in photosynthesis. The cell structure of the leaves, on the other hand, strongly reflects near-infrared light (from 0.7 to 1.1 µm). The differences in plant reflectance in the visible and near-infrared wavelengths is used to calculate NDVI index as follows: NDVI=((NIR-VIS)/((NIR+VIS)), where VIS and NIR stand for the spectral reflectance measurements acquired in the visible (red) and near-infrared regions, respectively. NDVI is directly related to the photosynthetic capacity and hence energy absorption of plant canopies 2. The NDVI 310 device compares reflected light at two distinct wavelengths, 635 and 760 nm.
PlantPen can be operated as a stand-alone instrument and thus is ideal for field studies. Measured data are sequentially stored in the internal PlantPen memory. Data transfer is via USB or Bluetooth communication. Comprehensive FluorPen
1.1 software provides data transfer routines and many additional features for data presentation in tables and graphs.
1
John Gamon, Josep Penuelas, and Christopher Field (1992). A narrow-waveband spectral index that tracks diurnal changes in
photosynthetic efficiency. Remote Sensing of environment, 41, 35-44.
2
Sellers, P. J. (1985) 'Canopy reflectance, photosynthesis, and transpiration', International Journal of Remote Sensing, 6, 1335-1372.
© PSI (Photon Systems Instruments), spol. s r. o.
9
3.1 DEVICE DESCRIPTION
Fig. 1 Device description.
© PSI (Photon Systems Instruments), spol. s r. o.
10
4 LIST OF EQUIPMENT AND CUSTOMER INFORMATION
Standard version of the PlantPen device package consists:
PlantPen PRI 210 or PlantPen NDVI 310
Carrying Case
FluorPen software and driver (on a USB flash disc)
Operation Manual (PDF on a USB flash disc)
USB Cable
Self-Adhesive Rubber Pads for Optics Protection
Other Accessories or Optional Features (according to your specific order)
For USB connection you need to have the USB driver installed in your PC. You find the driver on the installation disk (USB driver folder). For more information go to
If any item is missing, please, contact the manufacturer. Also check the carton for any visible external damage. If you find any damage, notify the carrier and the manufacturer immediately. The carton and all packing materials should be retained for inspection by the carrier or insurer.
For customer support, please write to: support@psi.cz
© PSI (Photon Systems Instruments), spol. s r. o.
11
5 CARE AND MAINTENANCE
Never submerge the device in water!
The device should not come in contact with any organic solvents, strong acids or bases.
Keep the optical part clean and dry. If cleaning is needed, use soft, non-abrasive tissue.
Battery life is approximately 48 hours when the FluorPen is operated continuously.
© PSI (Photon Systems Instruments), spol. s r. o.
12
6 PRINCIPLE OF MEASUREMENT
PlantPen is a reflectance-based device that provides a convenient method of measuring the relative chlorophyll content
of a leaf sample.
The PlantPen model PRI 210 (Fig. 2)measures Photochemical Reflectance Index (PRI) by comparing leaf reflectance in two narrow wavelength bands centered close to 531 nm and 570 nm. PRI is sensitive to changes in carotenoid pigments that are indicative of changes in photosynthetic light use efficiency, the rate of CO2 uptake and as a reliable water-stress index. As such, it is used in studies of vegetation productivity and stress.
Fig. 2 PlantPen PRI 210.
The PlantPen model NDVI 310 (Fig. 3) measures Normalized Difference Vegetation Index (NDVI), which is an important indicator of chlorophyll content in plants. The pigment in plant leaves, chlorophyll, strongly absorbs visible light (from 0.4
to 0.7 µm) for use in photosynthesis. The cell structure of the leaves, on the other hand, strongly reflects near-infrared light (from 0.7 to 1.1 µm). The differences in plant reflectance in the visible and near-infrared wavelengths are used to
calculate NDVI index. NDVI is directly related to the photosynthetic capacity and hence energy absorption of plant canopies.
Fig. 3 PlantPen NDVI 310.
© PSI (Photon Systems Instruments), spol. s r. o.
13
7 HOW TO GET STARTED
For more detailed information on particular steps of PlantPen operation please refer to chapter 8. The device works on Li-Ion battery or is charged via USB cable from PC or grid.
The PlantPen is controlled using two buttons:
Use the MENU key to scroll through sequential menu options on the digital display.
Use the SET key to select a menu option based on cursor (>) position.
7.1 PRI MEASUREMENT
7.1.1 CALIBRATION Calibration assures you that your measurements are accurate within the specification limits that led you to select the
instrument in the first place. It is better to calibrate your device before every set of samples. For Calibration of PlantPen use white reflectance standard. The calibration of the white reflectance standard is an indispensable part of a correct calculation of PRI.
Prior PRI measurement perform the calibration as follows:
For calibration use white reflectance standard (It is part of leaf clip)
Turn ON device – hold SET button for 1 sec.
Close the leaf clip (white standard is on bottom part of the leaf clip)
Go to Calibration (“Insert white sample appears on display)
Press SET button to confirm the calibration and wait until the calibration step is completed.
Calibration is automatically stored into the device memory.
The device is now ready to measure other samples.
7.1.2 MEASUREMENT
The PlantPen model PRI 210 measures The PlantPen model PRI 210 measures Photochemical Reflectance Index (PRI) in two narrow wavelength bands centred close to 531 nm and 570 nm. The photochemical reflectance index (PRI = (R531– R570)/(R531+R570)) correlates with the epoxidation state of xanthophyll cycle pigments and photosystem efficiency in many studies under diverse conditions, such as different. PRI is sensitive to changes in carotenoid pigments that are indicative of photosynthetic light use efficiency, the rate of carbon dioxide uptake, or as a reliable water-stress index. As such, it is used in studies of vegetation productivity and stress.
Please follow these instructions to perform a measurement:
Insert a leaf sample into the leaf clip. The leaf clip is important for the sample stabilization.
Go to Measurement
It is necessary to calibrate PlantPen after every switch ON.
© PSI (Photon Systems Instruments), spol. s r. o.
14
Press SET to confirm.
PlantPen measure PRI
Value of PRI appears on display after measuring.
During the measurement all data are stored into the device memory.
7.2 NDVI MEASUREMENT
7.2.1 CALIBRATION Calibration assures you that your measurements are accurate within the specification limits that led you to select the
instrument in the first place. It is better to calibrate your device before every set of samples. For Calibration of PlantPen use white reflectance standard. The calibration of the white reflectance standard is an indispensable part of a correct calculation of NDVI.
Prior NDVI measurement perform the calibration as follows:
For calibration use white reflectance standard (It is part of leaf clip)
Turn ON device – hold SET button for 1 sec.
Close the leaf clip (white standard is on bottom part of the leaf clip)
Go to Calibration (“Insert white sample appears on display)
Press SET button to confirm the calibration and wait until the calibration step is completed.
Calibration is automatically stored into the device memory.
The device is now ready to measure other samples.
7.2.2 MEASUREMENT
The PlantPen model NDVI 310 measures Normalized Difference Vegetation Index, which is an important indicator of
chlorophyll content in plants. The pigment in plant leaves, chlorophyll, strongly absorbs visible light (from 0.4 to 0.7 µm)
for use in photosynthesis. The cell structure of the leaves, on the other hand, strongly reflects near-infrared light (from 0.7 to 1.1 µm). The NDVI 310 device compares reflected light at two distinct wavelengths, 635 and 760 nm.
Please follow these instructions to perform a measurement:
Insert a leaf sample into the leaf clip. The leaf clip is important for the sample stabilization.
Go to Measurement
Press SET to confirm.
PlantPen measure NDVI
Value of NDVI appears on display after measuring.
During the measurement all data are stored into the device memory.
It is necessary to calibrate PlantPen after every switch ON.
© PSI (Photon Systems Instruments), spol. s r. o.
15
8 CONTROL MENU TREE
The next pages show the structure of the operation scheme, which includes the Main Menu, first-level Sub-Menus and second-level Sub-Menus.
The blue color represents the Main Menu and its Options.
The yellow color represents the first-level Sub-Menus and their Options.
The green color represents the second-level Sub-Menus and their Options.
Full-line arrows are used for the SET key.
Dashed-line arrows are used for the MENU key.
© PSI (Photon Systems Instruments), spol. s r. o.
16
© PSI (Photon Systems Instruments), spol. s r. o.
17
© PSI (Photon Systems Instruments), spol. s r. o.
18
© PSI (Photon Systems Instruments), spol. s r. o.
19
© PSI (Photon Systems Instruments), spol. s r. o.
20
9 USB CONNECTION
Connect the USB cable with the PlantPen device. Please note that lock in system is used.
Pay attention when connecting the USB cable not to damage the outlet connector on the Pen device. Make sure that you orient the cable correctly prior connecting the inlet with the outlet and the cable is upright towards the device.
To connect PlantPen with your computer please follow steps below in Fig. 4:
Fig. 4 How to connect PlantPen with PC.
A) Outlet connector on PlantPen device. B) Inlet part on the USB cable. C – E) Position the cable horizontally, plug in the inlet and screw the securing screw. F) Correct connection of the USB cable and Pen device.
Connect the USB cable to a computer. The PlantPen switches ON automatically after connecting the cable to the PC. For USB connection you need to have the USB driver installed in your PC. You find the driver on the installation disk (USB driver folder). If you check the Device Manager in Windows you should see the USB serial port in the device tree. In case of missing driver, you may download it from PSI websites. When the driver is installed correctly you should be able to connect to the device in the FluorPen software menu Setup > Device ID.
For more information about FluorPen software see chapter 11.
© PSI (Photon Systems Instruments), spol. s r. o.
21
10 BLUETOOTH CONNECTION
Before you set up the Bluetooth connection between the PlantPen and your PC, make sure you have these components:
1. Bluetooth enabled PC
The PC with which you connect must have Bluetooth wireless technology, either built-in or through a Bluetooth card. Make sure that the PC's Bluetooth setting is "discoverable" (meaning that it shows up when other devices search for nearby Bluetooth connections). Consult the user guide for your PC or Bluetooth card to learn how to do this.
2. Bluetooth configuration software properly set up on PC
Before you can download files to your PC, you will need to set up the Bluetooth software that came with your PC, or your PC's Bluetooth card. This software varies by manufacturer. Please consult your PC's Bluetooth documentation for more information.
3. Bluetooth must be switched on visible on both devices
To pair the PlantPen with another Bluetooth device, such as a computer, you will need to ensure that Bluetooth is switched on visible on both devices.
10.1 BLUETOOTH PAIRING
1. Enabling Bluetooth in the PlantPen
Switch ON the PlantPen (press and hold the SET key).
Scroll to the Accessories menu (press the MENU key, then press the SET key).
Select Bluetooth On (press the MENU key, then press the SET key) to enable Bluetooth.
Keep in mind that the PlantPen turns off automatically after about 8 minutes of no action.
Turning off the PlantPen always turns off Bluetooth.
2. Starting Bluetooth Application on Your PC
Be aware that this description is working on Windows 7; some of the steps may be different on your PC.
Select: Start > Devices and Printers (Fig. 5).
You may also start your Bluetooth application via the Control Panel: Start > Control Panel > Hardware and Sound > Devices and Printers.
© PSI (Photon Systems Instruments), spol. s r. o.
22
Fig. 5 Start Bluetooth Application.
3. Opening the Add Bluetooth Device Application
Select: “Add a device” to start searching new Bluetooth device. Be sure that the PlantPen is in discoverable
mode (see step 1).
Fig. 6 Add a device.
4. Selecting the PlantPen
Select: PSI PlantPen icon.
Select: Next (Fig. 7).
Fig. 7 Select the PlantPen.
© PSI (Photon Systems Instruments), spol. s r. o.
23
5. Starting the Pairing Process
This step is different for old and new version of Bluetooth module.
Old version:
Your Bluetooth Pairing Code is: 0000
Select: “Enter the device’s pairing code”.
Enter: 0000 (four digits).
Select: Next (Fig. 8).
Fig. 8 Pairing process.
New version:
Select: Yes (Fig. 9). Please note that the PlantPen device does not display the verification number. The
verification code is not important for the BT connection.
Select: Next.
Fig. 9 Verifying of the BT pairing.
6. Completing the PlantPen Pairing
Select: Close (Fig. 10).
© PSI (Photon Systems Instruments), spol. s r. o.
24
Fig. 10 Finishing.
On computer run the program FluorPen 1.1. For more information about FluorPen software see chapter 11.
© PSI (Photon Systems Instruments), spol. s r. o.
25
11 FLUORPEN SOFTWARE
11.1 SOFTWARE INSTALLATION
1. Save the FluorPen software provided on the USB flash disk to your computer and launch the FluorPen program.
2. To connect and recognize your PlantPen device in the FluorPen software proceeds first with the registration of
your FluorPen software (Fig. 11).
Select: Help > Register
Enter: your serial registration number.
Select: OK
Fig. 11 Software registration.
Please note that you will find your serial (registration) number in the file SN.txt, which is included on the enclosed USB flash disk.
Please remember: it is not possible to download data from the PlantPen device without software registration.
3. Switch on the PlantPen and enable Bluetooth or connect USB cable to the PC.
4. Make sure that your PC and the PlantPen are properly paired (see chapter 9 and Error! Reference source not
found. for complete information on USB and Bluetooth pairing).
5. Select: Setup > Device ID (Ctrl+I). If properly connected, the message “Device: PlantPen” appears in the bottom
part of the screen (Fig. 12).
© PSI (Photon Systems Instruments), spol. s r. o.
26
Fig. 12 Connecting PlantPen with Software.
11.2 MENU AND ICON EXPLANATION
11.2.1 MAIN MENU
MENU: File
Load Loads previously saved data files.
Save Saves data to hard disc.
Export Exports data in .txt format.
Export to JSON Exports data in JavaScript Object Notation.
Close Closes the current experiment.
Close All Closes all running experiments.
Exit Exits the program.
MENU: Device
Download Downloads data from the PlantPen to your PC.
Erase Memory Erases data from the PlantPen memory.
Online Control Online control of AP device.
Attach GPS File Used for download data from GPS module (active
only in PlantPen version AP 100).
© PSI (Photon Systems Instruments), spol. s r. o.
27
MENU: Setup
Device ID Detects the connected device.
Update Firmware Used for firmware updates.
Settings Used for modification of the program
settings.
MENU: Help
About Offers basic information about the program.
Register Used for the FluorPen software registration.
Icon Explanation:
11.2.2 MENU SETTINGS
MENU > Setup > Settings
After Download – Memory Erase
If the box is checked the PlantPen memory is erased after each data download.
Data – Inverted
If the box is checked the polarity of data is inverted, e.g., multiplied by -1. This feature can be helpful for a certain type of experiment when the measured data are undesirably interpreted as negative values.
Data – Add to opened
If the box is checked the downloaded data are added to that of the current opened experiment.
Data – GPS Visible
This option is active only in older PlantPen version PP 100. In new version PP 110 the GPS data are automatically downloaded and paired with protocol measurements.
Download
Downloads data from the PlantPen to PC.
Load
Loads (opens) previously saved data files.
Save
Saves data to hard disc.
Export
Exports data in .txt format.
© PSI (Photon Systems Instruments), spol. s r. o.
28
Graf – Single
This function is not available for PlantPen PRI or NDVI.
11.2.3 MENU ONLINE CONTROL This function can be used for Online Control your PlantPen device after connection with your PC.
Select: Menu > Device > Online Control
Online Control – Switches
Sound On/Off - choose On/Off for device beeping. Measuring color – choose 455nm or 630nm for measuring fluorescence protocols. This function is active only in the
PlantPen-C 110-C model. Multi use averaging (YES/NO) – serves to confirm Repeat (number of repetitions) and Interval (time between
measurements) Options for each measurement within Multi Option – select YES or NO.
Online Control – Time
Set the PlantPen time a date. You can also synchronize time of PlantPen device with computer time.
Online Control – Protocols
Choose the protocol for measuring. PlantPen device start to measure. After measuring you can download data.
Online Control – Values
This function is not available for PlantPen PRI or NDVI.
© PSI (Photon Systems Instruments), spol. s r. o.
29
11.3 DATA TRANSFER AND VISUALIZATION
1. Perform a measurement with your PlantPen. In Fig. 13 we did an NDVI measurement.
2. Click the Download icon or select Device > Download.
3. The Data table appears.
Fig. 13 Example of Data Transfer and Visualization.
4. For export press File > Export or Export icon, select what you want to export (NDVI or PRI) -Fig. 14.
Selected only – exports only one measurement that is selected by mouse, otherwise it will export everything. Source data – exports raw data. Description – exports the data description if any. Computed values – export calculated data
© PSI (Photon Systems Instruments), spol. s r. o.
30
Fig. 14 Export data.
© PSI (Photon Systems Instruments), spol. s r. o.
31
11.4 FIRMWARE UPDATE
All data in the PlantPen memory are erased during the firmware update! Before starting any firmware update, export all your data from the PlantPen memory into your
computer!
1. Starting Update
Select: Setup > Update Firmware From File (Fig. 15).
Fig. 15 Update Firmware.
2. Warning
Select: OK to start update (Fig. 16).
Fig. 16 Warning.
3. Selecting .bxn file
Find: Binary file (with the extension .bxn) (Fig. 17).
Select: Open.
© PSI (Photon Systems Instruments), spol. s r. o.
32
Fig. 17 Select .bxn file.
4. Finishing Upload
Select: OK to start uploading of the update (Fig. 18).
Fig. 18 Data loss warning.
The bottom bar indicates the upload progress (Fig. 19).
Fig. 19 Upload progress.
Press: OK to finish upload (Fig. 20).
Fig. 20 Finish upload.
© PSI (Photon Systems Instruments), spol. s r. o.
33
12 GPS MODULE
PlantPen device has integrated GPS module which can be turned on during the measurement and the GPS coordinated will be added to the downloaded data.
For proper GPS reading, the time in your PlantPen and in your computer must be synchronized. Preset time and time zone must correspond to GPS time (time zone) in your location.
12.1 GPS / PLANTPEN OPERATION
1. Check the time setting in PlantPen device: Settings > Date & Time
2. Switch the GPS module on:
Select: Accessories > GPS
Press SET to confirm.
Wait until the GPS position is found – “Starting GPS”.
The GPS module is ready when the icon in upper panel change – see on Fig. 21.
Fig. 21 GPS icons.
3. Go to Measurement and choose required protocol.
The device may need a clear view of the sky to acquire satellite signal.
Keep in mind that the PlantPen turns off automatically after about 8 minutes of no action.
Turning off the PlantPen always turns off GPS module.
For prompt determination of the coordinates use the option Accessories > GPS > Location.
© PSI (Photon Systems Instruments), spol. s r. o.
34
12.2 DATA DOWNLOAD
1. Enabling Communication:
Switch on the computer and the PlantPen. Set your computer to PlantPen communication: enable Bluetooth
or connect to USB port.
2. Downloading Data from the PlantPen
Start FluorPen program.
Connect PlantPen device: Setup > Device ID (Ctrl+I)
Download measured data from the PlantPen to your PC. Data measured with activated GPS module are
downloaded with GPS coordinates (Fig. 22).
Fig. 22 GPS coordinates.
© PSI (Photon Systems Instruments), spol. s r. o.
35
13 WARRANTY TERMS AND CONDITIONS
This Limited Warranty applies only to the PlantPen device. It is valid for one year from the date of shipment.
If at any time within this warranty period the instrument does not function as warranted, return it and the
manufacturer will repair or replace it at no charge. The customer is responsible for shipping and insurance charges (for the full product value) to PSI. The manufacturer is responsible for shipping and insurance on return of the instrument to the customer.
No warranty will apply to any instrument that has been (i) modified, altered, or repaired by persons
unauthorized by the manufacturer; (ii) subjected to misuse, negligence, or accident; (iii) connected, installed, adjusted, or used otherwise than in accordance with the instructions supplied by the manufacturer.
The warranty is return-to-base only and does not include on-site repair charges such as labor, travel, or other
expenses associated with the repair or installation of replacement parts at the customer's site.
The manufacturer repairs or replaces faulty instruments as quickly as possible; the maximum time is one month.
The manufacturer will keep spare parts or their adequate substitutes for a period of at least five years.
Returned instruments must be packaged sufficiently so as not to assume any transit damage. If damage is
caused due to insufficient packaging, the instrument will be treated as an out-of-warranty repair and charged as such.
PSI also offers out-of-warranty repairs. These are usually returned to the customer on a cash-on-delivery basis.
Wear & Tear Items (such as sealing, tubing, padding, etc.) are excluded from this warranty. The term Wear &
Tear denotes the damage that naturally and inevitably occurs as a result of normal use or aging even when an item is used competently and with care and proper maintenance.
© PSI (Photon Systems Instruments), spol. s r. o.
36
14 TROUBLESHOOTING AND CUSTOMER SUPPORT
In case of troubles and for customer support, please, visit FAQ on our websites, write to support@psi.cz or contact your local distributor.