Philips Trilogy200 Quick Start Manual

Whisper Swivel II (exhalation port)
Whisper Swivel II (exhalation port)
Mask with integrated leak
Mask without integrated leak
Tracheostomy tubing
Flexible tubing
Bacterial/
viral filter
(optional)
Exhalation valve line
Exhalation valve
Flexible tubing
Tracheostomy tubing
Mask without integrated leak
Bacterial/
viral filter
(optional)
Proximal pressure line
Proximal pressure port
Trilogy universal active PAP tube adapter
1
Overview
2
Unlock procedure
3
Circuit set-up
Before operating Trilogy200, you need to configure the circuit type.
Proximal pressure line
Exhalation valve line
Quick start guide
Overview
1
Unlock procedure
2
Circuit set-up
3
Ventilation set-up
4
7
1
Cancel action or exit menu
2
Navigate menus or change settings
3
Select/modify/confirm parameters
4
Audio pause button
5
Start/stop button
6
AC power inlet
7
SD card to download ventilation data
Symbols and icons
1
2 3
Prescriber environment, full access mode
SD card inserted in the ventilator
Audio pause active
Porting block
4
5
6
Nurse call connector
Serial connector – to connect to a computer or Alice polysomnograph
Air inlet and grey foam filter
Detachable battery
Ethernet connector
Status of detachable battery
Status of internal battery
Status of external battery
External battery connector
Removable air pathway
Oxygen valve port (flow source up to 15 LPM)
To access Full Menu and change settings
Press and hold and for a few seconds to access Set-up page
To keep full access to the settings
A
vigate to Options and press
Na
C
If shows on the main screen, full menu access is already allowed. Press to enter menu
B
Press to enter Menu Access
D
Passive exhalation port circuit
Active exhalation valve with proximal airway pressure circuit
Ramp active
Battery is charging
Change to Full using and press
to confirm
Press and to return to Initial Screen
In the Set-up menu,select "Passive" for circuit type.
In the Set-up menu,select "Active PAP" for circuit type.
Exit
Settings And Alarms
Options
Alarm log
Event log
Information
Menu 1/10
Select
Navigate
Cancel
Dual Prescription OFF
Circuit Type Active PAP
Mode S/T
Menu Settings and Alarms 2/23
OK
Edit
Exit
Dual Prescription OFF
Circuit Type Passive
Mode S/T
Menu Settings and Alarms 2/23
Modify
Navigate
Exit
Settings And Alarms
Options
Alarm log
M
Select
Navigate
Menu
Dual Prescription ON
Mode S/T
AVAPS ON
Tidal Volume 400 ml
IPAP Max Pressure 25.0 cmH20
Menu
Settings And Alarms 1/20
Navigate
Modify
Finish
Exit
Switch to Secondary Settings
Safely Remove SD Card
Primary Settings and Alarms
Secondary Settings and Alarms
Options
Menu 3/8
Select
Navigate
RR 20BPM
PIP
5.9
cm H2O
Vte
435ml
Leak
21l/min
Peak Flow
67.9l/min
MinVent
8.7l/min
Cancel
AVAPS ON
Tidal Volume 400 ml
IPAP Max Pressure 25.0 cmH20
OK
Edit
Cancel
AVAPS ON
Tidal Volume 400 ml
IPAP Max Pressure 25.0 cmH20
OK
Edit
Finish
AVAPS OFF
IPAP 30.0 cmH20
EPAP 8.0 cmH20
M
Modify
Navigate
Finish
Mode S/
AVAPS OFF
IPAP 30.0 cmH20
EPAP 8.0 cmH20
M
Modify
Navigate
Mode S/T
AVAPS ON
Tidal Volume 400 ml
IPAP Max Pressure 25.0 cmH20
Navigate
Modify
Finish
Flow line
Exhalation valve line
Tracheostomy tubing
Exhalation valve
Mask without integrated leak
Bacterial/
viral filter
(optional)
Flow sensor
White striped flow line
Flexible tubing
Flow line
Exhalation valve line
Tracheostomy tubing
Exhalation valve
Mask without integrated leak
Bacterial/
viral filter
(optional)
Flow sensor
White striped flow line
Flexible tubing
3
Circuit set-up (cont.)
Ventilation set-up
4
Philips Healthcare is part
of Royal Philips Electronics
How to reach us
www.philips.com/healthcare
healthcare@philips.com
fax: +31 40 27 64 887
Asia
+852 2821 5888
Europe, Middle East,Africa
+49 7031 463 2254
Latin America
+55 11 2125 0744
North America
+1 425 487 7000
800 285 5585
(toll free, US only)
Please visit www.philips.com
Philips Respironics
1010 Murry Ridge Lane
Murrysville, PA 15668
Customer Service
+1 724 387 4000
800 345 6443
(toll free, US only)
Respironics Asia Pacific
+65 6882 5282
Respironics Australia
+61 (2) 9666 4444
Respironics Deutschland
+49 8152 93 06 0
Respironics Europe,
Middle East,Africa
+33 1 47 52 30 00
Respironics France
+33 2 51 89 36 00
Respironics Italy
+39 039.203.1
Respironics Sweden
+46 8 120 45 900
Respironics Switzerland
+41 6 27 45 17 50
Respironics United Kingdom
+44 800 1300 845
www.philips.com/respironics
Active exhalation valve flow sensor circuit
Porting block
White striped flow line
Setting the circuit type
Make sureTrilogy200 is turned off before taking the following steps
A
Press and hold and for a few seconds to access Set-up page
C
Navigate to CircuitType and press
B
Select Settings And Alarms
D
Use to choose Active PAP, Active flow, or Passive and then press twice to confirm
A
From main screen, press Menu
E
Use to change the value or status of setting
Dual Prescription
B
Select Settings And Alarms
F
Press to validate new setting
C D
Scroll up and down using to navigate through the settings and alarms
G
Press and to return to the main screen
Press to change the parameter highlighted in blue
Respironics,Trilogy,Whisper Swivel, and Alice are trademarks of Respironics, Inc. and its affiliates.All rights reserved.
© 2010 Koninklijke Philips Electronics N.V. All rights are reserved.
Philips Healthcare reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication.
Broudy KB 1/4/10 MCI 4103124 PN 1070648
Flow line
In the Set-up menu,select "Active flow" for circuit type.
Exhalation valve line
If Dual Prescription is enabled,both prescriptions need to be adjusted as previously explained
Two prescriptions to set