Series 50 XM (M1350B)
Series 50 XMO (M1350C)
Fetal/Maternal Monitors
INSTRUCTIONS FOR USE
M1350-9001S
Printed in Germany March 2002
Edition 1
Philips makes no warranty of any kind with regard to this material, including, but not limited
to, the implied warranties of merchantability and fitness for a particular purpose. Philips shall
not be liable for errors contained herein or for incidental or consequential damages in
connection with the furnishing, performance or use of this material.
The information contained in this document is subject to change without notice.
Philips assumes no responsibility for the use or reliability of its software on equipment that is
not furnished by Philips.
Responsibility of the Manufacturer
Philips only considers itself responsible for any effects on safety, reliability and performance of
the equipment if:
• assembly operations, extensions, re- adjustments, modifications or repa irs are carried out by
persons authorized by Philips, and
• the electrical installation of the relevant room complies with national standards, and
• the instrument is used in accordance with the Instructions for Use or User’s Guide.
Intended Use (M1350B)
The Series 50 XM Fetal/Maternal Monitor (M1350B) allows non-invasive or invasive
monitoring of an ambulant patient during both antepartum testing and labor and delivery in
that the monitoring of the fetal hear t rate (FHR) via ultrasound or direct electrocardiogram
(DECG), and uterine activity via an external Toco transducer or an internal intrauterine
pressure (IUP) transducer is possible, additionally it allows maternal heart rate recording via
the MECG transducer. Alarms are generated from maternal heart rate.
The Series 50 XM allows the non-invasive measurement of the Noninvasive Blood Pressure
and the Oxygen Saturation parameters, generate alerts, and the generation of alerts and
recordings on matern al patients.
The device is intended to be used in Labor-Rooms and Delivery-Rooms and in AntepartumTesting-Areas. It is not intended to be u sed for transport monitoring and home use.
ii
Intended Use (M1350C)
The Series 50 XMO Fetal/Maternal Monitor (M1350C) allows non-invasive or invasive
monitoring of an ambulant patient during both antepartum testing and labor and delivery in
that the monitoring of the fetal hear t rate (FHR) via ultrasound or direct electrocardiogram
(DECG), and uterine activity via an external Toco transducer or an internal intrauterine
pressure (IUP) transducer is possible, additionally it allows maternal heart rate recording via
the MECG transducer. Alarms are generated from maternal heart rate.
The Series 50 XMO allows the non-invasiv e measurement of the Noninvasive Blood Pressure
and the Oxygen Saturation parameters, generate alerts, and the generation of alerts and
recordings on maternal patients. Additionally, the Series 50 XMO allows you to record fetal
pulse oximetry (FSpO
).
2
The device is intended to be used in Labor-Rooms and Delivery-Rooms and in AntepartumTesting-Areas. It is not intended to be u sed for transport monitoring and home use.
The Series 50 XMO (M1350C) is not available in the U.S.A.
Conventions Used in This Guide
Warning
A warning alerts you to a potential serious outcome, adverse event or safety hazard.
Failure to observe a warning may result in death or serious injury to the user or patient.
Caution
A caution alerts you where special care is necessary for the safe and effective use of the
product. Failure to observe a caution may result in minor or moderate personal injury
or damage to the product or other property, and possibly in a remote risk of more
serious injury.
Note—A note
calls your attention to an important point in the text.
iii
On your monitor, this sign indicates that there is
detailed information in this book which you must read
before proceeding with your task
The monitor should only be used by, or under the direct supervision of, a licensed physician
or other health care practitioner who is trained in the use of fetal and maternal heart rate
monitors and in the interpretation of fetal and maternal heart rate traces. US law restricts this
device to sale by, or on the order of, a physician.
2002 Philips Medizinsysteme GmbH
All rights are reserved. Reproduction in whole or in part is prohibited without the prior
written consent of the copyright holder.
This symbol indicates that you should consult the
Instructions For Use (this guide), and particularly any
warning messages.
Equipotential Terminal
This symbol identifies terminals which are
connected together, bringing various equipment or
parts of a system to the same potential. This is not
necessarily earth potential. The value of potentials
of earth may be indicated adjacent to the symbol.
Protective Earth Terminal
This symbol identifies the terminal for connection
to an external protective earth system.
2 x 1.5 V
The monitor is de signed to fulfil safety requirements according to IEC
60601-1/EN 60601-1 (Class I), UL 2601-1, CSA-C22.2 No 601.1M90.
It complies with the essential requirements of the Medical Device
Directive 93/42/EEC. The monitor is classed:
Chapter 1 - Safety1
Battery
This symbol identifies the battery holder conta ining
two 1.5 V batteries.
2 x 1.5V
Electrical Safety
The Series 50 XMO and the Series 50 XM are not “ECG-Monitors”, are
not defibrillator-protected, and are not designed for direct cardiac
application.
Electrical Safety
EquipmentEquipment TypeInformation
ORDINARY EQUIPMEN T - Enclosed equipment protected
against ingress of water.
CONTINUOUS OPERATION - Can be operated continuously.
External Toco transduce r (M135 5A)
and
IUP quartz transducer (1290C) and
IUP transducer (CPJ840J5)
Ultrasound transducer (M1356A)
DECG transducer (M1357A) and
MECG transducer (M1359A)
US/MECG combi transducer
(M1358A)
SpO
transducer (M1191A)
2
Adapter cable (M1940A)
NIPB cuffs (M1574A and M1575 A)
Interco nn ect tubing (M1599A)
FSpO
/ECG patient module
2
(M1365A)
ECG only patient module
(M1364A)
B
B
CF
B (Ultrasound)
CF (MECG)
CF
CF
CF
Patient Connector is electrically isolated.
Connecting the transducer to the Toco
channel results in CF condition.
Connecting the trans ducer to a Cardio
channel results in a B condit i on .
Connecting the transducers to a Cardio
channel results in a CF condition.
Connecting the combi transducer to the
Cardio 1 channel results in a CF condition
for MECG and a B condition for US
application.
Connecting the SpO2 transducer results in
a CF condition.
Connecting the NIBP tubing results in a
CF condition.
Connecting a patient module results in a
CF condition.
2Chapter 1 - Safety
Maximum Input/Ou tp ut Volt a ges
The following diagram shows the sockets for peripheral devices.
Maximum Input/Output Volta ge s
Chapter 1 - Safety
3
Maximum Input/Output Voltages
1. Mains Socket.
2. Equipotential Grounding P oint.
To use the monitor with other equipment in an operating room
environment, connect the equipotential grounding point (2) to
earth potential. Use the grounding cable supplied with the
monitor.
3. +5V input socket for the HBSW8200 Barcode Reader.
4. Socket for the Philips M1310A Series 50 T Fetal Telemetry
System. +5V input except for:
Pins 1, 14, 15 and 16: ± 12V input
Pin 2 -12V output
Pin 3 +5V output
Pin 4 ±12V output
5. RS232 Digital System Interface:
Pin2 ±12V input
Pin 3 ±12V output
6. Socket for one of the following:
•80225A or 80235A/B Obstetrical Information Management
System ( OBMS)
•M1370A Obstetrical Display Information System (ODIS)
•±12V except for Pins 17, 18 and 22 which are +5V input.
7. Socket (9-pin) for an external device:
Pin 3 ±12V
8. Socket (25-pin) for an external device:
Pin 2 ±12V
Pins 9 and 10 +5 Volt
4Chapter 1 - Safety
Service Socket for Upgrade Key
The Service En gineer can conne ct a compatible PC to this socket (1) to
carry out extended configuration and service functions.
Maximum voltage of ±12V.
Service Socket for Upgrade Key
Protective Ea rth
To prote ct hospital personnel and the patien t, the monitor’s casing must
be ground ed. Accordingly, the monitor has a 3-wire power cable that
grounds it to the power line ground when plugged into an appropriate 3wire receptacle. Do not use a 3-wire to 2-wire adapter with the monitor.
Any interruption of the protective earth grounding will cause a potential
shock hazard that could result in serious personal injury.
Whenever it is likely that the protection has been impaired, the monitor
must be made inoperative and be secured against any unintended
operation.
Chapter 1 - Safety
5
Environment
Warning
Check each time before use that the monitor is in perfect working
order and properly grounded.
Position the patient cable so that it does not come into contact with any
other electrical equipment. The cable connecting the patient to the
monitor must be free of electrolyte.
Make sure that during operation, the monitor is free from condensation.
This can form when equipment is moved from one building to another,
and is exposed to moisture and differences in temperature.
Warning
Possible explosion hazard if used in the presence of flammable
anaesthetics.
Environment
Use the monitor in a n environment that is r easonably fr ee from vibration,
dust, corrosive or explosive gases, flammable agents, extremes of
temperature, humidity and so forth. It operates within specifications at
ambient temperatures between 0 and 55
exceed these limits can affect the accuracy of the monitor and cause
damage to the components and circuits. Only products that fulfil the
necessary safety and electrical standards should be used in conjunction
with the monitor (contact your local response center for details).
Allow at least 5cm (2in) clearance around the monitor for proper air
circulation. If the monitor is mounted in a cabinet, allow sufficient space
at the front for operation and at the rear for servicing with the cabinet
door open.
6Chapter 1 - Safety
o
C. Ambient temperatures that
Spillage
When the maternal display is in a tilted position, take additional care to
prevent spillage of liquid. If liquid accidentally enters the monitor
through the maternal display recess, you must cease using the monitor
immediately. Contact an authorized engineer for a safety inspection.
Electromagnetic Compatibility (EMC)
The electromagnetic compatibility (EMC) validation includes testing
performed according to the international standard for EMC with medical
devices. See the Manufacturer’s Declaration for details.
EMC Testing
Caution
If operating under conditions according to the EMC-st andard
60601-1- 2 (Radiated I mmunity 3 V /m), fi eld st rength s a bove 1 V/m
may cause erroneous measurements at various frquencies. Therefore
it is recommended to avoid the use of electr ical ly r adiating
equipment in close proximity to these measurem en ts.
Spillage
Chapter 1 - Safety
During the test prog ram the monitor was sub j e cted to international
standard and Philips proprietary EMC tests. During most of the testing
no anomalies were observed. Some reduced performance was observed
with the IEC 801-3 Radiated Immunity and IEC 801-4 Fast Transient/
Bursts Immunity tests.
IEC 801-3 specifies that the pro duct must be subjected to a field of 3 V/m
over a frequency range of 26 to 1000 MHz with no degradation of
performance. At many of the test frequencies over the specified range no
anomalies were ob served. However increase d jitter of SpO
values and
2
DECG derived fetal heartrate values was observed at a number of test
points. For these test points the radiated field was reduced to the level at
which the trace returned to normal. These reduced levels are shown in
the following table.
7
Electromagnetic Compatibility (EMC)
ParameterSensorFrequency Range
SpO
2
IEC 801-4 specified that the pro duct is subjected to high speed p ulses up
to 1000 V applied to the powe r cord and al l I/O cables. D uring and af ter
most of the test pulses, no anomalies were observed. However in rare
cases the ultrasound channel had been triggered.
ParameterSensor
UltrasoundM1356A
SpO
with adapter
2
cable M1940A
System Characteristics
M1191A
M1192A
M1194A
M1358A
M1191A
M1192A
M1194A
52 MHz to 70
MHz
Immunity
Level
2.3 V/m
Burst Immunity
Level
600 V
≤ 200V
The phenomena discussed above are not unique to the monitor but are
characteristic of pat ient monitors in use today. This performance is due
to very sensitive high gain front end amplifiers used to process the
physiological signals from the patien t. Among the many similarly
performing monitors already in use by customers, interference from
electromagnetic sources is rarely a problem.
8Chapter 1 - Safety
Avoiding Interference
When electromagnetic interference (EMI) is encountered there are a
number of things that can be done to mitigate the problem.
1. Eliminate the source. Possible sources of EMI can be turned off or
moved away to reduce their strength.
2. Attenuate the coupling. If the coup ling path is through the patient
cables the interference may be reduced by moving and/or
rearranging th e cables to a diffe rent location of the monitor. If the
coupling path is throu gh the power cord, plugg ing the monitor
into a differen t mains circuit may help.
3. Connect the equipotential terminal of the moni tor to the
corresponding terminal of your mains installation.
4. Add external attenuators. If EMI becomes an unusually difficult
problem, external devices such as an isolation transformer or a
transient suppressor may help. An Philips customer engineer can
assist you in determining the need for external devices.
ESU, MRI and Defibrillation
ESU, MRI and Defibrillation
Warning
Remove all transducers, patient modules, sensors and accessor ies
before performing electrical surgery, defibrillation, and MRI. High
frequency current can flow through the equipment and burn the
skin.
The equipment has not been tested with defibrillators.
Chapter 1 - Safety
9
Leakage Current
Leakage Current
Leakage current can be hazardous to the patient.
Warning
If the monitor is connected directly to other equipment, such as an
additional patient monitor, or a second monitor is to be connected
directly to the mother, you must carry out all relevant safety tests in
accordance with safety standard IEC 60601-1-1.
10Chapter 1 - Safety
About This Guide
This guide tells midwives, nurses and other healthcare professionals how
to use the Series 50 XMO fetal/maternal monitor and the Series 50 XM
fetal/maternal monitor . I t discusses and ill ustrates all possible featur es and
parameters of both mo nitors. Your monitor may not have every one o f
these features and may lo ok slightly different to the monito r shown in t he
illustrations in this guide. It can be upgraded to incorporate them.
About the Monitors
Both monitors let you observe and record:
2
Overview
•Fetal heart rate (inclu ding twins)
•Uterine activity
•Maternal heart rate (MHR) and ECG waveform
•Maternal pulse oximetry (SpO
•Fetal movement profile (FMP)
•Maternal blood pressure, non-invasively.
The Series 50 XMO also allows you to record fetal pulse oximetry
(FSpO
You can monitor fetal heart rate externally with u ltrasound from
approximately 20 weeks and with the other parameters internally during
labor and delivery.
Not all parame ters and features detai led in this manual are ava ilable on all
monitors.
Chapter 2 - Overview11
).
2
)
2
Major Parts and Keys
Major Parts and Keys
.
The display panel can be viewed flat in the monitor, or tilted at an angle.
1. Monitor on/off switch
2. Monitor on/off light
3. Recorder keys
4. Recorder
5. Maternal parameters
6. Cardio 2 channel
7. Toco channel
9. Function key
10. Telemetry indicator
11. FSpO
parameter
2
12. Opening recess
13. Socket for remote event marker
14. Setting keys
15. Service socket
8. Cardio 1/ combi channel
12Chapter 2 - Overview
Cardio, Toco and FSpO2 Channels
Major Parts and Keys
Chapter 2 - Overview
1. Toco Display shows uterine activity.
2. Fetal SpO
display shows fetal pulse indicator, signal quality,
2
alarm status, and cross channel verifi cation plus indicator.
3. Cardio Display shows the FHR.
4. Signal Quality Indicator shows the quality of heart rate signal
detected by the transducer:
– Green (good).
– Yellow (fair to potentially poor).
– Red (unacceptable).
5. Function Key: selects menus for:
– FMP, twins offset, logic, FHR alert and FSpO
.
2
– returns to the normal display.
6. MECG Indicator shows when MECG is being measured through
this channel. (Indicator location different for Series 50 XM.)
13
Major Parts and Keys
7. Fetal SpO2 display shows current value of FSpO2.
8. Speaker Lamp shows which heartbeat is heard from the
loudspeaker.
9. Volume Keys set the volume and select the channel to which you
are listening. Changes current setting of FMP, twins offset, logic,
FHR alert and FSpO
14. Toco Baseline Key zeroes the Toco display and trace to 20 units
(when monitoring uterine activity externally) or 0 units (when
monitoring uter ine activity internally).
indicates average pulse rate value comes from NIBP.
7. Maternal Heart Rate shows the current heart/pulse rate .
8. Systolic Value shows the value for the systolic parameter of th e
non-invasive blood pressure measurement.
9. Diastolic Value shows the value for the diastolic parameter of the
non-invasive blood pressure measurement.
16Chapter 2 - Overview
Recorder Keys
Major Parts and Keys
1. Recorder On/Off Light lights when the recorder is working.
Flashes when monitor detects five or fewer pages remaining in the
pack, or if the paper runs out.
2. Recorder On/Off Key switches recorder on and off. Also starts
NST timer (switch off recorder and press for two seconds).
3. Event Marker Key records event on paper. Acknowledges all alerts
and alarms.
4. Paper Advance Key advances paper automatical ly to the next fold.
Tear paper at fold. Never pull paper to advance it.
5. Paper-Eject Key unlocks drawer when you press it once. Press a
second time and hold when removing paper.
Chapter 2 - Overview
17
Major Parts and Keys
Setting Keys
1. Time and Date Key for changing the time and date. Press to show
the current time in the Cardio 1/Combi and Toco di splays, to
cycle through the settings to be ch anged (hours, minutes, day,
month and year ) and to return to the normal display.
2. Paper Speed Key for changing the paper speed. Press to show the
current paper speed in the Cardio 1/Combi display, and to return
to the normal display.
3. Test Key for starting monito r’s self test.
18Chapter 2 - Overview
Loading...
+ 187 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.