Philips Series 50 (M1350) User manual

Series 50 XM (M1350B) Series 50 XMO (M1350C) Fetal/Maternal Monitors
INSTRUCTIONS FOR USE
M1350-9001S Printed in Germany March 2002 Edition 1
The information contained in this document is subject to change without notice. Philips assumes no responsibility for the use or reliability of its software on equipment that is
not furnished by Philips.
Responsibility of the Manufacturer
Philips only considers itself responsible for any effects on safety, reliability and performance of the equipment if:
• assembly operations, extensions, re- adjustments, modifications or repa irs are carried out by persons authorized by Philips, and
• the electrical installation of the relevant room complies with national standards, and
• the instrument is used in accordance with the Instructions for Use or User’s Guide.
Intended Use (M1350B)
The Series 50 XM Fetal/Maternal Monitor (M1350B) allows non-invasive or invasive monitoring of an ambulant patient during both antepartum testing and labor and delivery in that the monitoring of the fetal hear t rate (FHR) via ultrasound or direct electrocardiogram (DECG), and uterine activity via an external Toco transducer or an internal intrauterine pressure (IUP) transducer is possible, additionally it allows maternal heart rate recording via the MECG transducer. Alarms are generated from maternal heart rate.
The Series 50 XM allows the non-invasive measurement of the Noninvasive Blood Pressure and the Oxygen Saturation parameters, generate alerts, and the generation of alerts and recordings on matern al patients.
The device is intended to be used in Labor-Rooms and Delivery-Rooms and in Antepartum­Testing-Areas. It is not intended to be u sed for transport monitoring and home use.
ii
Intended Use (M1350C)
The Series 50 XMO Fetal/Maternal Monitor (M1350C) allows non-invasive or invasive monitoring of an ambulant patient during both antepartum testing and labor and delivery in that the monitoring of the fetal hear t rate (FHR) via ultrasound or direct electrocardiogram (DECG), and uterine activity via an external Toco transducer or an internal intrauterine pressure (IUP) transducer is possible, additionally it allows maternal heart rate recording via the MECG transducer. Alarms are generated from maternal heart rate.
The Series 50 XMO allows the non-invasiv e measurement of the Noninvasive Blood Pressure and the Oxygen Saturation parameters, generate alerts, and the generation of alerts and recordings on maternal patients. Additionally, the Series 50 XMO allows you to record fetal pulse oximetry (FSpO
).
2
The device is intended to be used in Labor-Rooms and Delivery-Rooms and in Antepartum­Testing-Areas. It is not intended to be u sed for transport monitoring and home use.
The Series 50 XMO (M1350C) is not available in the U.S.A.
Conventions Used in This Guide
Warning A warning alerts you to a potential serious outcome, adverse event or safety hazard.
Failure to observe a warning may result in death or serious injury to the user or patient.
Caution A caution alerts you where special care is necessary for the safe and effective use of the product. Failure to observe a caution may result in minor or moderate personal injury or damage to the product or other property, and possibly in a remote risk of more serious injury.
Note—A note
calls your attention to an important point in the text.
iii
On your monitor, this sign indicates that there is detailed information in this book which you must read before proceeding with your task
The monitor should only be used by, or under the direct supervision of, a licensed physician or other health care practitioner who is trained in the use of fetal and maternal heart rate monitors and in the interpretation of fetal and maternal heart rate traces. US law restricts this device to sale by, or on the order of, a physician.
2002 Philips Medizinsysteme GmbH
All rights are reserved. Reproduction in whole or in part is prohibited without the prior written consent of the copyright holder.
iv
Contents
1. Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
General Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Maximum Input/Output Voltages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Service Socket for Upgrade Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Protective Earth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Environment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Spillage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Electromagnetic Compatibility (EMC). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
EMC Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
System Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Avoiding Interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
ESU, MRI and Defibrillation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Leakage Current. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2. Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
About This Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
About the Monitors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Major Parts and Keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Cardio, Toco a nd FSpO
Maternal Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Recorder Keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Setting Keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Channels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2
3. General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Suspected Fetal Demise. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Fastening a Belt. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Clipping a Transducer to the Belt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Attaching a Patient Module to the Belt. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Connecting a Transducer or Patient Module to the Monitor . . . . . . . . . . . . . . . . . . . . 23
Signal Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Input Channels at a Glance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Example Trace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Marking an Event . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Contents v
4. Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Before Connecting Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Switching On the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Loading Paper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Switching on the Recorder. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Displaying the Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Setting the Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Choosing Paper Speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Setting the Paper Speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Switching Off After Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5. Monitoring FHR and FMP Using Ultrasound . . . . . . . . . . . . . . . . 39
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
What You Need . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Fetal Movement Profile. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
FMP Statistics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Switching FMP Off and On. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Using Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Using the Barcode Reader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6. Monitoring FHR Using DECG . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
What You Need . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Using the Traditional Open-wire Method to Monitor DECG . . . . . . . . . . . . . . . . . . . . 50
With DECG legplate M1357A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
With Patient Module M1364A or M1365A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Using the DECG Adapter Cable M1362B to Monitor DECG. . . . . . . . . . . . . . . . . . . . . 53
With DECG Legplate M1357A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
With Patient Module M1364A or M1365A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Monitoring DECG. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Why use Arrhythmia Logic? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
After Monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Heart Rate Out of Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
vi Contents
7. Monitoring Twin FHRs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Things to Remember During Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Monitoring Internally . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Monitoring Externally. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Separating Twin FHR Traces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Twins Offset: On. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Twins Offset: Off. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
8. Monitoring Uterine Activity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
External Toco Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Internal Toco Monitoring (IUP Monitoring) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
External Toco . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Internal Toco . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
9. Fetal Pulse Oximetry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
To Begin Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Understanding the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Alarm Limits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Triggering an Alarm. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Acknowledging an Alarm. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Turning the Alarm On and Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Changing Alarm Limits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Inop Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Loss of Pulse Signal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Inaccurate Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
10. Monitoring Using External Devices . . . . . . . . . . . . . . . . . . . . . . 83
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Recording From an External Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
What You Need . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Connecting the External Devices to the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Contents vii
Example Trace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Trace Display on External Monitoring Systems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
FSpO
2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90
External Devices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
11. Recording Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Recording A Note. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Deleting a Barcode Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Recording Several Barcodes as One Note. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Recording Several Barcodes as Separate Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Recording a Patient’s Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
12. Fetal Heart Rate Alerting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Alerts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Recognizing an Alert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Acknowledging an Alert. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Turning Alerting ON or OFF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Changing Alert Limits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
13. Non Stress Test Timer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Setting the NST Timer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
14. Maternal Monitoring - Overview . . . . . . . . . . . . . . . . . . . . . . . . 101
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Maternal Parameters - Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Softkeys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Reset Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Power on Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
To Begin Monitoring Immediately . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Maternal Main Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Returning to the Maternal Main Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Alarms - Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Acknowledging an Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Reviewing All Alarm Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Setting an Alarm. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
viii Contents
Warning Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Acknowledging a Warning Message. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Volume and Contrast Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
15. Non-invasive Blood Pressure Monitoring. . . . . . . . . . . . . . . . . 111
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
To Begin Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Setup Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Measurement Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Changing Measurement Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Pulse Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Turning Alarms On and Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Changing Alarm Limits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Recording. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Warning Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Measuring Problems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
16. Maternal ECG, Heart and Pulse Rate. . . . . . . . . . . . . . . . . . . . 121
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
To Begin Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Connecting the Transducer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Applying the Electrodes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Using MECG transducer M1359A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Using a Patient Module (M1364A or M1365A) . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
Cross-Channel Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
Displaying the MECG Waveform. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Changing Display Speed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Freezing and Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Maternal Heartrate Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Alarm (MECG and SpO
Turning Alarms On and Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Changing Alarm Limits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
2
17. Maternal Pulse Oximetry (SpO2) . . . . . . . . . . . . . . . . . . . . . . . 133
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
To Begin Monitoring Immediately . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Contents ix
Alarm (SpO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Turning Alarm On and Off. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Changing Alarm Limit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Alarm (Pulse rate). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Warning Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
18. Maintenance and Performance Assurance . . . . . . . . . . . . . . . . 139
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Visual Inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Routine Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Mechanical Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Calibration and Electrical Safety Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Care and Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Care of the Fetal Monitoring System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Transducers and Patient Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
Adapter Cables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Sterilizing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Monitor, Patient Modules, Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Adapter cables M1347A, M1362B. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Adapter cables M1362A, M1363A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
IUP (1290C/CPJ840J5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
IUP Transducer Adapter Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Belts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
Reusable SpO
Non-invasive Blood Pressure Cuff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
Storing Recorder Paper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
Self Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
Quick Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Parameter Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152
Testing Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154
Toco . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154
Ultrasound . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154
IUP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Replacing the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Replacing the Fuses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
Transducers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
2
x Contents
19. Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159
Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
20. Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Standard Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Fetal Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Maternal Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164
Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164
Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164
Paper. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166
Gels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166
Transducers and Patient Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
MECG Electrodes and Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
DECG Cables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
Disposable Scalp Electrodes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
Fetal Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169
IUP Transducers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169
IUP Catheters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169
Domes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
IUP Transducer Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
Maternal Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
NIBP Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
SpO
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 171
2
Adapter Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 171
Belts and Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
21. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173
Patient Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173
Operating and Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174
Fetal Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
Maternal Non-invasive Blood Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
Maternal Pulse Oximetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
Maternal ECG and Heart Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
Maternal Display Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
Numerical Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
Maternal Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
Mode Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
Contents xi
Ultrasound, External and Internal Toco. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179
Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179
FHR (Cardio) Scales . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180
Testing Facilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180
Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
22. Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183
Manufacturers Responsibility. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183
Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183
23. Training Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .187
M1350 XM, XMO Fetal Monitoring Lab Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . 187
Competency Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187
Performance Objectives. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187
Maternal Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
Fetal SpO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
Fetal Telemetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
Resources. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
Fetal Monitor Lab Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
Maternal Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192
Fetal SpO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
Fetal Telemetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
Answers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
Maternal Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 198
Fetal SpO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
Fetal Telemetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
xii Contents

General Safety Information

1

Safety

This symbol indicates that you should consult the Instructions For Use (this guide), and particularly any warning messages.
Equipotential Terminal
This symbol identifies terminals which are connected together, bringing various equipment or parts of a system to the same potential. This is not necessarily earth potential. The value of potentials of earth may be indicated adjacent to the symbol.
Protective Earth Terminal
This symbol identifies the terminal for connection to an external protective earth system.
2 x 1.5 V
The monitor is de signed to fulfil safety requirements according to IEC 60601-1/EN 60601-1 (Class I), UL 2601-1, CSA-C22.2 No 601.1­M90.
It complies with the essential requirements of the Medical Device Directive 93/42/EEC. The monitor is classed:
Chapter 1 - Safety 1
Battery
This symbol identifies the battery holder conta ining two 1.5 V batteries.
2 x 1.5V

Electrical Safety

The Series 50 XMO and the Series 50 XM are not “ECG-Monitors, are not defibrillator-protected, and are not designed for direct cardiac application.
Electrical Safety
Equipment Equipment Type Information
ORDINARY EQUIPMEN T - Enclosed equipment protected against ingress of water.
CONTINUOUS OPERATION - Can be operated continuously.
External Toco transduce r (M135 5A) and IUP quartz transducer (1290C) and IUP transducer (CPJ840J5)
Ultrasound transducer (M1356A)
DECG transducer (M1357A) and MECG transducer (M1359A)
US/MECG combi transducer (M1358A)
SpO
transducer (M1191A)
2
Adapter cable (M1940A) NIPB cuffs (M1574A and M1575 A)
Interco nn ect tubing (M1599A)
FSpO
/ECG patient module
2
(M1365A) ECG only patient module (M1364A)
B
B
CF
B (Ultrasound) CF (MECG)
CF
CF
CF
Patient Connector is electrically isolated. Connecting the transducer to the Toco channel results in CF condition.
Connecting the trans ducer to a Cardio channel results in a B condit i on .
Connecting the transducers to a Cardio channel results in a CF condition.
Connecting the combi transducer to the Cardio 1 channel results in a CF condition for MECG and a B condition for US application.
Connecting the SpO2 transducer results in a CF condition.
Connecting the NIBP tubing results in a CF condition.
Connecting a patient module results in a CF condition.
2 Chapter 1 - Safety

Maximum Input/Ou tp ut Volt a ges

The following diagram shows the sockets for peripheral devices.
Maximum Input/Output Volta ge s
Chapter 1 - Safety
3
Maximum Input/Output Voltages
1. Mains Socket.
2. Equipotential Grounding P oint. To use the monitor with other equipment in an operating room environment, connect the equipotential grounding point (2) to earth potential. Use the grounding cable supplied with the monitor.
3. +5V input socket for the HBSW8200 Barcode Reader.
4. Socket for the Philips M1310A Series 50 T Fetal Telemetry System. +5V input except for: Pins 1, 14, 15 and 16: ± 12V input Pin 2 -12V output Pin 3 +5V output Pin 4 ±12V output
5. RS232 Digital System Interface: Pin2 ±12V input Pin 3 ±12V output
6. Socket for one of the following:
80225A or 80235A/B Obstetrical Information Management System ( OBMS)
M1370A Obstetrical Display Information System (ODIS)
±12V except for Pins 17, 18 and 22 which are +5V input.
7. Socket (9-pin) for an external device: Pin 3 ±12V
8. Socket (25-pin) for an external device: Pin 2 ±12V Pins 9 and 10 +5 Volt
4 Chapter 1 - Safety

Service Socket for Upgrade Key

The Service En gineer can conne ct a compatible PC to this socket (1) to carry out extended configuration and service functions.
Maximum voltage of ±12V.
Service Socket for Upgrade Key

Protective Ea rth

To prote ct hospital personnel and the patien t, the monitor’s casing must be ground ed. Accordingly, the monitor has a 3-wire power cable that grounds it to the power line ground when plugged into an appropriate 3­wire receptacle. Do not use a 3-wire to 2-wire adapter with the monitor. Any interruption of the protective earth grounding will cause a potential shock hazard that could result in serious personal injury.
Whenever it is likely that the protection has been impaired, the monitor must be made inoperative and be secured against any unintended operation.
Chapter 1 - Safety
5

Environment

Warning Check each time before use that the monitor is in perfect working
order and properly grounded.
Position the patient cable so that it does not come into contact with any other electrical equipment. The cable connecting the patient to the monitor must be free of electrolyte.
Make sure that during operation, the monitor is free from condensation. This can form when equipment is moved from one building to another, and is exposed to moisture and differences in temperature.
Warning Possible explosion hazard if used in the presence of flammable
anaesthetics.
Environment
Use the monitor in a n environment that is r easonably fr ee from vibration, dust, corrosive or explosive gases, flammable agents, extremes of temperature, humidity and so forth. It operates within specifications at
ambient temperatures between 0 and 55 exceed these limits can affect the accuracy of the monitor and cause damage to the components and circuits. Only products that fulfil the necessary safety and electrical standards should be used in conjunction with the monitor (contact your local response center for details).
Allow at least 5cm (2in) clearance around the monitor for proper air circulation. If the monitor is mounted in a cabinet, allow sufficient space at the front for operation and at the rear for servicing with the cabinet door open.
6 Chapter 1 - Safety
o
C. Ambient temperatures that

Spillage

When the maternal display is in a tilted position, take additional care to prevent spillage of liquid. If liquid accidentally enters the monitor through the maternal display recess, you must cease using the monitor immediately. Contact an authorized engineer for a safety inspection.

Electromagnetic Compatibility (EMC)

The electromagnetic compatibility (EMC) validation includes testing performed according to the international standard for EMC with medical devices. See the Manufacturer’s Declaration for details.

EMC Testing

Caution If operating under conditions according to the EMC-st andard
60601-1- 2 (Radiated I mmunity 3 V /m), fi eld st rength s a bove 1 V/m may cause erroneous measurements at various frquencies. Therefore it is recommended to avoid the use of electr ical ly r adiating equipment in close proximity to these measurem en ts.
Spillage
Chapter 1 - Safety
During the test prog ram the monitor was sub j e cted to international standard and Philips proprietary EMC tests. During most of the testing no anomalies were observed. Some reduced performance was observed with the IEC 801-3 Radiated Immunity and IEC 801-4 Fast Transient/ Bursts Immunity tests.
IEC 801-3 specifies that the pro duct must be subjected to a field of 3 V/m over a frequency range of 26 to 1000 MHz with no degradation of performance. At many of the test frequencies over the specified range no anomalies were ob served. However increase d jitter of SpO
values and
2
DECG derived fetal heartrate values was observed at a number of test points. For these test points the radiated field was reduced to the level at which the trace returned to normal. These reduced levels are shown in the following table.
7
Electromagnetic Compatibility (EMC)
Parameter Sensor Frequency Range
SpO
2
IEC 801-4 specified that the pro duct is subjected to high speed p ulses up to 1000 V applied to the powe r cord and al l I/O cables. D uring and af ter most of the test pulses, no anomalies were observed. However in rare cases the ultrasound channel had been triggered.
Parameter Sensor
Ultrasound M1356A
SpO
with adapter
2
cable M1940A

System Characteristics

M1191A M1192A M1194A
M1358A M1191A
M1192A M1194A
52 MHz to 70 MHz
Immunity
Level
2.3 V/m
Burst Immunity
Level
600 V
200V
The phenomena discussed above are not unique to the monitor but are characteristic of pat ient monitors in use today. This performance is due to very sensitive high gain front end amplifiers used to process the physiological signals from the patien t. Among the many similarly performing monitors already in use by customers, interference from electromagnetic sources is rarely a problem.
8 Chapter 1 - Safety

Avoiding Interference

When electromagnetic interference (EMI) is encountered there are a number of things that can be done to mitigate the problem.
1. Eliminate the source. Possible sources of EMI can be turned off or moved away to reduce their strength.
2. Attenuate the coupling. If the coup ling path is through the patient cables the interference may be reduced by moving and/or rearranging th e cables to a diffe rent location of the monitor. If the coupling path is throu gh the power cord, plugg ing the monitor into a differen t mains circuit may help.
3. Connect the equipotential terminal of the moni tor to the corresponding terminal of your mains installation.
4. Add external attenuators. If EMI becomes an unusually difficult problem, external devices such as an isolation transformer or a transient suppressor may help. An Philips customer engineer can assist you in determining the need for external devices.

ESU, MRI and Defibrillation

ESU, MRI and Defibrillation
Warning Remove all transducers, patient modules, sensors and accessor ies
before performing electrical surgery, defibrillation, and MRI. High frequency current can flow through the equipment and burn the skin.
The equipment has not been tested with defibrillators.
Chapter 1 - Safety
9

Leakage Current

Leakage Current
Leakage current can be hazardous to the patient.
Warning If the monitor is connected directly to other equipment, such as an
additional patient monitor, or a second monitor is to be connected directly to the mother, you must carry out all relevant safety tests in accordance with safety standard IEC 60601-1-1.
10 Chapter 1 - Safety

About This Guide

This guide tells midwives, nurses and other healthcare professionals how to use the Series 50 XMO fetal/maternal monitor and the Series 50 XM fetal/maternal monitor . I t discusses and ill ustrates all possible featur es and parameters of both mo nitors. Your monitor may not have every one o f these features and may lo ok slightly different to the monito r shown in t he illustrations in this guide. It can be upgraded to incorporate them.

About the Monitors

Both monitors let you observe and record:
2

Overview

Fetal heart rate (inclu ding twins)
Uterine activity
Maternal heart rate (MHR) and ECG waveform
Maternal pulse oximetry (SpO
Fetal movement profile (FMP)
Maternal blood pressure, non-invasively.
The Series 50 XMO also allows you to record fetal pulse oximetry (FSpO
You can monitor fetal heart rate externally with u ltrasound from approximately 20 weeks and with the other parameters internally during labor and delivery.
Not all parame ters and features detai led in this manual are ava ilable on all monitors.
Chapter 2 - Overview 11
).
2
)
2

Major Parts and Keys

Major Parts and Keys
.
The display panel can be viewed flat in the monitor, or tilted at an angle.
1. Monitor on/off switch
2. Monitor on/off light
3. Recorder keys
4. Recorder
5. Maternal parameters
6. Cardio 2 channel
7. Toco channel
9. Function key
10. Telemetry indicator
11. FSpO
parameter
2
12. Opening recess
13. Socket for remote event marker
14. Setting keys
15. Service socket
8. Cardio 1/ combi channel
12 Chapter 2 - Overview

Cardio, Toco and FSpO2 Channels

Major Parts and Keys
Chapter 2 - Overview
1. Toco Display shows uterine activity.
2. Fetal SpO
display shows fetal pulse indicator, signal quality,
2
alarm status, and cross channel verifi cation plus indicator.
3. Cardio Display shows the FHR.
4. Signal Quality Indicator shows the quality of heart rate signal detected by the transducer:
Green (good). Yellow (fair to potentially poor). Red (unacceptable).
5. Function Key: selects menus for:
– FMP, twins offset, logic, FHR alert and FSpO
.
2
– returns to the normal display.
6. MECG Indicator shows when MECG is being measured through this channel. (Indicator location different for Series 50 XM.)
13
Major Parts and Keys
7. Fetal SpO2 display shows current value of FSpO2.
8. Speaker Lamp shows which heartbeat is heard from the loudspeaker.
9. Volume Keys set the volume and select the channel to which you are listening. Changes current setting of FMP, twins offset, logic, FHR alert and FSpO
alarms.
2
10. Recess for use when tilting the display.
11. Remote Event Marker Socket for connecting remote event marker (15249A).
12. Cardio 1/Combi Transducer Socket for connecting:
– FSpO
/ECG combined patient module (M1365A).
2
ECG patient module (M1364A).Ultrasound transducer (M1356A).DECG transducer (M1357A).US/MECG Combi transducer (M1358A).MECG transducer (M1359A).
13. Transducer Socket for connecting:
external Toco transducer (M1355A). IUP transducer (1290C, 13972A, or M1333A).
14. Toco Baseline Key zeroes the Toco display and trace to 20 units (when monitoring uterine activity externally) or 0 units (when monitoring uter ine activity internally).
15. Cardio 2 Transducer Socket for connecting:
ECG patient module (M1364A). US transducer (M1356A).DECG transducer (M1357A).MECG transducer (M1359A).
14 Chapter 2 - Overview

Maternal Parameters

1. NBP Transducer Socket for connecting:
2. SpO
Major Parts and Keys
– NBP cuff interconnect tubing (M1599A) and blood pressure
cuff.
Transducer Socket for connecting:
2
–SpO2 transducer (M1940A adapter cable connected to M1191A
transducer).
Chapter 2 - Overview
3. Softkeys for setting maternal parameters.
selects modes and alarm limits for NBP. selects modes and alarm limits for MHR. selects modes and alarm limits for SpO
.
2
4. Reset Key acknowledges al arms and re turns monitor from setting mode to maternal parameter display.
5. SpO
Value indicates the current reading for patients SpO2 level.
2
15
Major Parts and Keys
6. Maternal heart/pulse rate icon indicates measurement source of maternal heart/pulse rate.
indicates heart rate value comes from MECG.
indicates pulse rate value comes from SpO
.
2
indicates average pulse rate value comes from NIBP.
7. Maternal Heart Rate shows the current heart/pulse rate .
8. Systolic Value shows the value for the systolic parameter of th e non-invasive blood pressure measurement.
9. Diastolic Value shows the value for the diastolic parameter of the non-invasive blood pressure measurement.
16 Chapter 2 - Overview

Recorder Keys

Major Parts and Keys
1. Recorder On/Off Light lights when the recorder is working. Flashes when monitor detects five or fewer pages remaining in the pack, or if the paper runs out.
2. Recorder On/Off Key switches recorder on and off. Also starts NST timer (switch off recorder and press for two seconds).
3. Event Marker Key records event on paper. Acknowledges all alerts and alarms.
4. Paper Advance Key advances paper automatical ly to the next fold. Tear paper at fold. Never pull paper to advance it.
5. Paper-Eject Key unlocks drawer when you press it once. Press a second time and hold when removing paper.
Chapter 2 - Overview
17
Major Parts and Keys

Setting Keys

1. Time and Date Key for changing the time and date. Press to show the current time in the Cardio 1/Combi and Toco di splays, to cycle through the settings to be ch anged (hours, minutes, day, month and year ) and to return to the normal display.
2. Paper Speed Key for changing the paper speed. Press to show the current paper speed in the Cardio 1/Combi display, and to return to the normal display.
3. Test Key for starting monito rs self test.
18 Chapter 2 - Overview
Loading...
+ 187 hidden pages