This work is protected under Title 17 of the United States copyright code and is the sole property of Respironics.
No par of this document may be copied or otherwise reproduced, or stored in any electronic information retrieval
system, except as specifically permitted under United States copyright law, without the prior written consent of
Respironics.
For Technical support, contact:
Respironics, Inc. Customer Service
Within the U.S.A. 1-800-345-6443
Outside the U.S.A. 724-387-4000
Facsimile 724-387-5012
service@respironics.com
United States of America
Respironics California, Inc.
2271 Cosmos Court
Carlsbad, CA 92011
USA
1-800-345-6443
or 724-387-4000
Authorized Representative
Respironics Deutschland, Inc.
Gewerbestrasse 17
D-82211 Herrsching Deutschland
+49-8-15-29-30-60
The Esprit Ventilator is a microprocessor-controlled, electrically powered
mechanical ventilator. It is intended for use by qualified medical personnel to
provide continuous or intermittent ventilatory support for adult and pediatric
patients as prescribed by a physician. The Esprit Ventilator is intended for use
in either invasive or non-invasive applications.
The Esprit Ventilator meets or exceeds all applicable safety requirements,
consensus guidelines, US regulatory statutes, and international regulatory
standards for life support/mechanical ventilation devices.
Read this manual thoroughly prior to performing service or maintenance on the
Esprit Ventilator. This manual contains advanced troubleshooting, calibration,
and maintenance instructions for the Esprit Ventilator. All maintenance and
repair work should be performed by qualified biomedical technicians who have
received appropriate training and authorization to provide maintenance, repair,
and service for the Esprit Ventilator.
Review the Esprit Ventilator Operator’s Manual and become familiar with Esprit
Ventilator operation before running tests, checking operational readiness or
initiating patient use. The operator’s manual includes important information
about ventilator safety and operation.
Schematic diagrams of the Esprit Ventilator are available upon request.
For additional information about accessories or related equipment, such as
humidifiers and remote alarm systems, refer to the appropriate instruction
manual prior to operating with the Esprit Ventilator.
WARNING:Patients on life-support equipment should be visually monitored by competent
medical personnel, since life-threatening circumstances may arise that may
not activate alarms. Heed all appropriate alarms and follow the instructions
and warnings in this service manual and the operator’s manual. Always check
life-support equipment for proper operation before use.
NOTE:The Esprit Ventilator Operator’s Manual lists all applicable warnings and
cautions. Review these notices thoroughly before operating the ventilator.
The following table lists the recommended tools, test equipment, and
materials required to service and maintain the Esprit Ventilator (Table 1-1).
Test equipment must meet the requirements in Table 1-2.
DescriptionManufacturer and Model
Test Equipment
Digital multimeter and frequency counter (DMM)
accurate to three decimal places
Pneumatic calibration analyzer capable of
measuring low pressure (cmH
(LPM), and volume (liters)
Electrical safety analyzerDale LT 544D or equivalent
Oxygen analyzer with accuracy of ± 2%TSI Certifier Plus with oxygen sensor kit or
Pressure analyzer with accuracy of measuring
high pressure (PSI)
10 mL calibrated syringe (Neonatal testing)Hans Rudolph 5220 or equivalent
Adapter, USB to serialRespironics P/N 1022895
Analog output port signal selectorRespironics P/N 1010891
Esprit Service kit (*included in the kit)Respironics P/N 1021670
Cable, nurse call test*Respironics P/N 1001375
Test adapter, O2 regulator*Respironics P/N 1001376
Cork, silicone*Respironics P/N 1001735 or equivalent
Adapter, parallel port*Respironics P/N 1004644
Test lung, 1 liter*Respironics P/N 1021671 or equivalent
Cable assy, null modem*Respironics P/N 1022815 or equivalent
O), flow rate
2
Local Supplier
TSI Certifier Plus (Respironics P/N 1040311)
or equivalent
equivalent
TSI Certifier Plus or equivalent
Ventilator Accessories
Tubing, silicone, 3/16 in. ID x 6.5 ft., PAPRespironics P/N C06686 or equivalent
Throughout this manual the following definitions apply:
WARNING:A condition that could cause injury to a patient or operator if the operating
instructions in this manual are not followed correctly.
CAUTION:A condition that could cause damage to, or shorten the service life of,
the Esprit Ventilator.
General Warnings
and Cautions
WARNING:Do not obstruct the emergency air intake near the oxygen water trap/inlet
filter assembly.
WARNING:Never troubleshoot while a patient is connected to the ventilator, since normal
operation is suspended.
WARNING:If the ventilator has been operating, the exhalation filter heater conductor may
be hot. Use caution when removing the filter.
WARNING:To prevent disease transmission, use protective equipment when handling
contaminated bacterial filters or other patient accessories.
WARNING:To avoid personal injury, always disconnect external AC and DC power
sources and high-pressure oxygen sources from the ventilator before
servicing.
WARNING:Explosion hazard. Do no operate the ventilator in the presence of flammable
anesthetic agents.
CAUTION:Troubleshooting and repair should be performed only by a qualified
service technician. Respironics Esprit Factory Service Training is highly
recommended prior to performing service procedures on the Esprit
Ventilator. Contact Customer Service at 1-800-345-6443 or 724-3874000 for more information.
CAUTION:Use only Respironics Esprit repair/service parts. Only Respironics parts
are designed for use in this ventilator. Use of non-Respironics repair parts
may alter ventilator reliability resulting in damage. Use of nonRespironics repair parts will affect your warranty. Contact Customer
Service at 1-800-345-6443 or 724-387-4000 for more information.
CAUTION:Do not modify oxygen diameter index safety systems (DISS) connector on
The Esprit mechanical ventilator is a microprocessor-controlled device that can
deliver air, oxygen, or a mixture of air and oxygen to the patient’s lungs in a
predetermined manner to augment or replace the work normally performed by
the patient’s respiratory system. It uses electromechanical control circuits,
flow and pressure monitors, and software programs to deliver breaths as a flow
or pressure controller.
The Esprit Ventilator includes a graphic user interface (GUI), internal blower,
and inspiratory module that mixes air and oxygen. The ventilator can operate
from a 40 to 90 psig (276 to 620 kPa) medical grade oxygen source for
enriched oxygen operation. It also includes multiple communications
interfaces and an internal power supply that can run from a 100 to 240 V AC
50/60 Hz or 24 V DC power sources.
Schematic diagrams of the Esprit Ventilator are available upon request.
Pneumatic SystemThe Esprit Ventilator pneumatic system consists of these subsystems (see
Figure 3-1):
•Internal blower (air source)
•Oxygen regulator (oxygen source)
•Inspiratory module
•Heated exhalation filter assembly
•Exhalation valve assembly
•Expiratory flow sensor
The internal blower generates the air pressure necessary for breath delivery,
eliminating the need for an external source of medical-grade compressed air.
An internal regulator regulates wall oxygen pressure. The ventilator mixes air
and oxygen in the inspiratory module before delivery to the patient.
Based on operator settings, the central processing unit (CPU) controls the air
valve, oxygen valve, and exhalation valve through stepper motor controller
printed circuit boards (PCBs). As flow is delivered to the patient, the air and
oxygen flow sensors and two pressure sensors provide feedback to the CPU.
The pressure relief and safety valves in the inspiratory module provide for
patient safety in the event of an over-pressure condition or any component or
system failure that could interfere with the patient’s ability to breathe when
connected to the ventilator.
The exhalation filter conditions the exhaled gas, reducing the risk of
contamination or component damage due to bacteria or moisture in expired
gases. The exhalation filter is housed in a heated sleeve, which reduces the
relative moisture condensation in the exhalation filter, exhalation valve, and
expiratory flow sensor. Exhaled gas is then vented to atmosphere.
Air
O
2
Exhalation
Valve
Inspiratory Module
• Gas Mixing
• Pressure Relief Valve
• Safety Valve
To Patient
From Patient
Heated
Exhalation
Filter
Assembly
Oxygen
Supply
Room
Air
Blower
Oxygen
Regulator
Expiratory
Flow
Sensor
Figure 3-1: Pneumatic System Block Diagram
Figure 3-2 shows the Esprit ventilator pneumatic system and its components.
Delivery SystemThe delivery system includes the components that condition and control the
flow delivered to the patient based upon operator-selected parameters. The
blower draws room air through the blower inlet filter (F1) and the muffler
(silencer) and outputs flow to the air valve assembly (AV). A pressure switch
(PS1) monitors oxygen input pressure. The oxygen water trap/inlet filter
assembly filters wall oxygen, and the oxygen regulator (REG1) regulates oxygen
down to 23 pounds per square inch, psi (1.5 kilopascals, kPa). Regulated
oxygen then enters the oxygen valve assembly (OV) and the crossover solenoid
(SOL1). The air and oxygen valves (AV and OV) are controlled by the
microprocessor, based on continuous feedback from the air and oxygen flow
sensors (FS1, T1 and FS2, T2).
Delivery System Components
Blower Inlet Filter
(F1)
Cooling Fan FilterThe cooling fan filter removes coarse particulate from ambient air
Muffler (Silencer)The muffler reduces the noise of air flow into the blower by channeling
BlowerThe blower draws room air though the air inlet filter and outputs the air
The blower inlet filter removes coarse particulate from ambient air as it
is entrained into the blower assembly. See section 4 for periodic
maintenance information.
entrained by the cooling fan. See section 4 for periodic maintenance
information
the air through a baffled system lined with sound absorbing material.
that is delivered to the patient and provides the pilot pressure that can
actuate the safety valve. The blower contains a DC motor and a series of
stator and impeller assemblies. It can provide at least 200 LPM of flow.
Blower speed is automatically adjusted to account for differences in gas
density due to altitude. The altitude can be adjusted from the hardware
screen in diagnostics mode. The High Pressure alarm limit setting also
affects blower speed.
Cooling CoilThe cooling coil is a copper tube connected to the outlet of the blower
Cooling Coil FanThe 24 V DC cooling coil fan removes the heat dissipated by the cooling
Air Valve Assembly
(AV)
Air Flow Sensor (FS1) The air flow sensor measures flow from the air valve. The ventilator uses
Oxygen Inlet
Connector (O
Oxygen Water Trap/
Inlet Filter Assembly
(F2)
)
2
that reduces the temperature of the gas from the blower before it
reaches the air valve.
coil and blower.
The air valve assembly contains a stepper motor that meters air flow
from the blower to achieve the target flow under CPU control, based on
operator selected parameters. It can deliver up to 200 LPM of flow.
this measurement to provide closed loop control of the air valve and to
compute the flow and volume delivered to the patient. A thermistor in
the flow sensor measures the temperature of the air and provides the
microprocessor with information to compensate the delivered flow.
The oxygen inlet connector provides a country-specific connection point
for an external oxygen gas supply of 40 to 90 psig (276 to 620 kPa).
The oxygen water trap/inlet filter assembly consists of a 5-micron (µ)
filter to remove particulate (both dry and liquid) from the oxygen gas
supply, a bowl with drain for accumulated water, and an oxygen inlet
connector.
Oxygen Supply
Pressure Switch (PS1)
The oxygen supply pressure switch is part of the oxygen regulator. PS1
is a normally open (NO) switch that closes when measured pressure is
greater than 40 psig (276 kPa), and provides a digital signal to the
sensor PCB indicating whether supply pressure is adequate at the
oxygen inlet.
PS1 opens if measured pressure is less than 35 psig (241.3 kPa). If the
oxygen supply pressure switch opens during normal ventilation (at O
21%), a Low O
Oxygen Valve (OV)The oxygen valve assembly contains a stepper motor that meters flow
Oxygen Flow Sensor
(FS2)
Crossover Solenoid
(SOL1)
The oxygen regulator reduces the oxygen supply pressure to the proper
inlet pressure for the oxygen valve (22-24 psig, or 152-165 kPa @ 180
LPM) and supplies the regulated pressure to the crossover solenoid,
which pilots the safety valve.
from the oxygen regulator to achieve the target flow under CPU control,
based on operator-selected parameters. It can deliver up to 200 LPM of
flow.
The oxygen flow sensor measures the flow from the oxygen valve. The
ventilator uses this measurement to provide closed loop control of the
oxygen valve and to compute the flow and volume delivered to the
patient. A thermistor contained in the flow sensor measures the
temperature of the oxygen and provides temperature compensation
information to the microprocessor for delivered flow.
The crossover solenoid is a three-way valve that supplies either air or
oxygen pressure to pilot (hold) the safety valve closed during normal
ventilation. In its normal state, SOL1 is normally de-energized to pilot
the safety valve with oxygen. If oxygen pressure is lost, SOL1 is
energized and air (rather than oxygen) controls the safety valve.
Check Valve (CV5)The cross contamination check valve prevents the oxygen supply from
entering the air delivery system pneumatics (blower) in the event of a
crossover solenoid leak.
Inspiratory SystemThe inspiratory system includes a manifold where air and oxygen are
blended and the inspiratory pressure transducer (PT3) is connected
through the inspiratory pressure transducer solenoid (SOL4). The
manifold also houses several components designed to ensure patient
safety, including the safety valve pilot solenoid (SOL2), safety valve
(SV), (CV2), inspiratory non-rebreathing check valve (CV3), pressure
relief valve (PRV), and oxygen sensor (OS).
The safety valve pilot solenoid directs the output of the crossover
solenoid to the safety valve or vents the pilot pressure line to
atmosphere.
During normal operation, SOL2 is energized and directs pressure from
the crossover solenoid to close the safety valve. During a high priority
alarm condition such as an occlusion or ventilator failure mode (VENT
INOP), SOL2 is deenergized to open the safety valve and allow the
patient to breathe room air.
Chapter 3
Theory of Operation
Delivery System Components
Safety Valve (SV)The safety valve contains a spring-loaded diaphragm that is controlled
Air System Check
Valve (CV2)
Inspiratory Nonrebreathing Check
Valve (CV3)
Pressure Relief Valve
(PRV)
by safety valve pilot solenoid (SOL 2). Under normal conditions SV is
closed, allowing delivered flow to reach the patient. In the event of a
safety valve open (SVO) condition, pilot pressure is vented to
atmosphere, which opens SV and allows the patient to breathe room air
through the safety port at the rear of the ventilator.
The air system check valve (CV2) prevents oxygen from entering the air
delivery system in the event of a blower failure.
The inspiratory non-rebreathing check valve prevents the patient from
exhaling through the inspiratory limb during a safety valve open
condition, which prevents the patient from rebreathing exhaled gas.
The pressure relief valve provides a backup to the operator adjustable
high-pressure alarm and prevents excessive pressures in the patient
circuit. The PRV is spring-loaded to limit the maximum circuit pressure
to 130 to 140 cmH2O.
Oxygen Sensor (OS)The oxygen sensor is an optional device that can be installed between
Inspiratory Pressure
Transducer (PT3)
the 22-mm inspiratory port and the inspiratory bacteria filter. The
oxygen sensor is a galvanic device that measures the oxygen
concentration of the blended gas as it leaves the inspiratory manifold.
The output signal from the sensor is used for the high and low oxygen
concentration alarms. When the sensor is installed and calibrated, the
ventilator alarms if the measured oxygen concentration is not within 6%
of the %O
The inspiratory pressure transducer on the sensor PCB monitors system
pressure from the inspiratory side of the patient circuit during
exhalation pressure transducer autozeroing, ensuring uninterrupted
pressure monitoring. It is also used with the exhalation pressure
transducer to detect patient circuit occlusions.
Exhalation SystemThe exhalation system maintains circuit pressure and conditions, filters,
Heated Exhalation
Filter (F3)
Exhalation Pressure
Transducer (PT2)
Exhalation Pressure
Transducer Solenoid
(SOL3)
The inspiratory pressure transducer solenoid periodically vents the
inspiratory pressure transducer to atmosphere and makes a
measurement at zero (atmospheric) pressure. Periodically autozeroing
the transducer allows it to correct the slight zero voltage drift that can
occur over time, and improves the overall accuracy of the pressure
measurement.
During normal operation, SOL4 is de-energized and applies patient
circuit pressure to the inspiratory pressure transducer. During an
autozero, SOL4 is energized, venting the transducer to atmosphere.
This occurs during power on self test (POST), at the beginning of a
breath one minute after POST, six minutes after POST, eleven minutes
after POST, and hourly thereafter.
and monitors exhaled gas. It contains the heated exhalation filter (F3),
exhalation pressure transducer (PT2), exhalation pressure transducer
solenoid (SOL3), exhalation valve (EV), exhalation non-rebreathing
check valve (CV4), and the exhalation flow sensor (FS3).
The heated exhalation filter includes a heated filter sleeve and a
bacteria filter. The heater protects the exhalation flow sensor and
exhalation system components from condensation by heating exhaled
gas (which has cooled in the exhalation limb) above the dew point.
The exhalation bacteria filter protects the exhalation flow sensor and
exhalation system component from contaminants and filters exhaled
gas before it is vented to atmosphere.
The exhalation pressure transducer on the sensor PCB measures patient
circuit pressure from the exhalation side of the patient circuit. During
normal operation PT2 is the primary transducer for measuring patient
pressures, including peak inspiratory pressure (PIP), mean airway
pressure (MAP), end inspiratory pressure, and auto-PEEP. The
exhalation pressure transducer provides monitoring data for closed loop
control.
The exhalation pressure transducer solenoid periodically vents the
exhalation pressure transducer to atmosphere and makes a
measurement at zero (atmospheric) pressure. Periodically autozeroing
the transducer allows it to correct the slight zero voltage drift that can
occur over time, and improves the overall accuracy of the pressure
measurement.
During normal operation, SOL 3 is de-energized and applies patient
circuit pressure to the exhalation pressure transducer. During an
autozero, SOL3 is energized, venting the transducer to atmosphere.
This occurs during POST, at the beginning of a breath one minute after
POST, six minutes after POST, eleven minutes after POST, and hourly
thereafter.
Exhalation Valve (EV)The exhalation valve assembly is a stepper motor-controlled valve. At
Exhalation Flow
Sensor (FS3)
The exhalation non-rebreathing check valve prevents the patient from
inspiring room air through the exhalation limb of the patient circuit.
During normal operation, it blocks the exhalation system from
atmosphere, allowing the patient to trigger a breath.
the beginning of an inspiration, the exhalation valve shuts to create a
closed circuit and allow the patient system to pressurize. The exhalation
valve opens at the beginning of exhalation, allowing system pressure to
vent to atmosphere.
The exhalation valve also regulates positive end expiratory pressure
(PEEP) and expiratory positive airway pressure (EPAP) levels during
exhalation.
The exhalation flow sensor measures the flow leaving the ventilator. This
flow includes gas exhaled by the patient, tubing compliance volume,
and bias flow if flow triggering or Auto-Trak triggering is selected. The
ventilator uses the exhaled flow measurement to compute flow and
volume coming from the patient and the circuit.
A thermistor in the flow sensor measures the temperature of the gas and
provides the microprocessor with information to compensate the
measured flow.
Figure 3-3 shows the electronic system. Schematics are available upon
request.
Figure 3-3: Electronic System Diagram
The ventilator can be powered by a 100 to 240 VAC 50/60 Hz or external 24 V
DC power sources (Backup Battery or External Battery). The power supply
conditions the input voltage and distributes +5 V, +12 V, -12 V, and +29 V to
the main PCB and blower motor controller to power digital electronics,
electropneumatic components, and displays. AC power to the humidifier port
can be used on 100-120 V units only.
The microprocessor on the CPU PCB and programs stored in memory control
the interaction of the pneumatic and electronic subsystems. Using inputs from
electropneumatic sensors and the operator, the CPU controls the flow,
pressure, and volume of air and oxygen to be delivered to the patient. The CPU
also monitors alarms and independently monitors software execution.
The CPU interfaces with the pneumatics and displays through the main PCB
and daughter boards that are vertically mounted on the main PCB. The
daughter boards include the CPU PCB, digital PCB, analog PCB, VGA
controller PCB, and three stepper motor controller PCBs.
The digital control signals from the CPU are sent to the analog PCB where they
are converted into analog signals that control blower speed and chart recorder
outputs (pressure, flow, and volume). Analog data from the flow, pressure, and
oxygen sensors is conditioned and converted by the sensor PCB. The sensor
PCB conditions and coverts the data, and sends it to the analog PCB, where it
is read by the CPU.
Ventilator data from the CPU is conditioned by the VGA and man-machine
interface (MMI) PCBs, then displayed on an LCD.
The following table summarizes the electronic signal path for Esprit
components.
ComponentSignal Path Sequence
100% O2 indicatorFront panel overlay, MMI PCB, main PCB, digital PCB, CPU
PCB
29 V EnableMain PCB, CPU PCB, power supply
Air (AV), oxygen (OV), exhalation
(EV) valves
Air (FS1), oxygen (FS2), exhalation
(FS3) flow sensors
Alarm High IndicatorFront panel overlay, MMI PCB, main PCB, CPU PCB
Alarm Med/Low IndicatorFront panel overlay, MMI PCB, main PCB, CPU PCB
Alarm Silence IndicatorFront panel overlay, MMI PCB, main PCB, digital PCB, CPU
Backlight (9.5-in. display)Backlight inverter PCB, MMI PCB, backlight control
Backlight (10.4-in. display)Backlight inverter PCB, DC/DC converter PCB, main PCB,
Backup AlarmMMI PCB, main PCB, digital PCB, CPU PCB
Backup Battery, External BatteryPower supply
Battery/Charging IndicatorFront panel overlay, MMI PCB, power supply, main PCB,
Battery/In Use IndicatorFront panel overlay, MMI PCB, main PCB, digital PCB, CPU
Battery/Low IndicatorFront panel overlay, MMI PCB, digital PCB, CPU PCB
Blower DACBlower controller PCB, sensor PCB, main PCB, analog PCB,
Blower on/offBlower controller PCB, sensor PCB, main PCB, CPU PCB
Blower temperature switchSensor PCB, main PCB, digital PCB, CPU PCB
Console: all keysFront panel overlay, MMI PCB, main PCB, CPU PCB
The CPU and other ventilator logic interact through the system data, address,
and control buses on the main PCB. The main PCB receives input signals from
various keys on the console or touch screen display and sends them to the
CPU. The main PCB also contains signal inputs for non-maskable interrupt,
running on AC, and running on external battery.
The main PCB receives control signals from the CPU and outputs them to
various pneumatic components and console indicators. The main PCB receives
signals from the digital PCB to turn on the indicators for alarm silence, 100%
oxygen, AC power, external battery power, and backup battery status. The main
PCB receives signal from the CPU PCB to turn on the backup alarm, enable
24V, and the Screen Lock, Battery/Charging, and Vent Inop indicators. The
CPU PCB reads the Accept key from the main PCB.
The main PCB includes a normal open and normal closed relay that can trigger
the remote nurse call alarm. Interface connectors on the main PCB include the
RS-232, parallel printer, analog output, and remote alarm connectors.
Other signals routed by the main PCB are the reset, MMI PCB reset, sensor
PCB reset, primary alarm, primary alarm failure detection logic, backup alarm,
remote alarm, printer, POST timer, clocked serial interface (CSI) signals, and
the battery backed +3.6 V.
CPU PCB
The CPU PCB contains the microprocessor, memory, I/O ports, and associated
control circuitry that controls the ventilator. Functional circuits contained on
the CPU PCB include:
•V851 microprocessor with a 25-Mhz clock.
•Static RAM that stores ventilator data.
•EEPROM that stores patient settings.
•Flash memory that contains ventilator operating software.
•One time programmable (OTP) memory that stores the POST routine.
•Internal RS-232 port that receives ventilator data from the touch
screen.
•Non-maskable interrupt that tells the CPU a power source has been
lost or interrupted.
•5-msec bus timer that monitors hardware operation.
•169-msec watchdog timer that monitors software operation.
•Data address and control bus to the main PCB.
•Current version includes 2 MB memory capacity (previous versions
included 1 and 1.5, MB).
The analog PCB performs a digital-to-analog conversion of signals from the
CPU to the blower controller PCB and analog output port. The analog PCB
connects directly to the system bus on the main PCB, and includes these
functional circuits:
•An eight-bit digital to analog converter (DAC) that converts digital
signals from the CPU to analog for the blower and external devices
such as chart recorders and bedside monitors.
•Clocked serial interface (CSI), a high-speed communication link
between the air, oxygen, and exhalation motor controllers and flow
sensor lookup tables contained on the CPU and the voltage monitor
register.
•A circuit that retrieves converted data from the sensor PCB.
Digital PCB
The digital PCB conditions serial port signals coming from and going to the
CPU PCB. It also contains control circuitry for the power fail alarm, primary
alarm, backup alarm, RS-232 port, and rotary encoder.
Digital inputs include analog-to-digital converter (ADC) out of range,
compressor temperature switch, and oxygen present. Digital outputs include
the alarm silence indicator, 100% oxygen indicator, running on AC indicator,
running on external battery indicator, backup battery status indicators, printer
ready signal, and printer direction.
VGA Controller PCB
The VGA controller PCB contains the date and real time clock and LCD VGA
display controller drivers.
Blower Controller PCB
The blower motor controller PCB controls the speed of the blower motor based
on analog input conditioned by the sensor PCB. It includes a lockup sensing
circuit, which monitors sensors in the blower motor to detect a locked rotor
condition. If the blower motor stops running, the lockup sensing circuit shuts
off power to the blower.
There are three motor controller PCBs for the air valve, oxygen valve, and
exhalation valve. The three boards are physically the same, and are
differentiated by the slot they occupy on the main PCB:
•Exhalation valve motor controller PCB: slot CN11
•Oxygen valve motor controller PCB: slot CN12
•Air valve motor controller PCB: slot CN13
Each motor controller PCB includes a microprocessor dedicated to controlling
the corresponding motor, and drives the step positions of the motor based on
input from the CPU.
Sensor PCB
The sensor PCB contains an analog to digital converter (ADC) that converts
analog signals from various pneumatic components and the power supply into
digital signals for the CPU. Signals include: air flow and temperature, oxygen
flow and temperature, exhalation flow and temperature, inspiratory and
exhalation pressure, battery voltage, FIO
enclosure oxygen concentration.
2, enclosure temperature, and
The sensor PCB conditions blower speed analog input and the on/off control to
the blower controller PCB, and routes signals for the oxygen pressure and
blower temperature switches.
The sensor PCB also includes voltage monitors. LEDs on the sensor PCB light
to indicate under- and over-voltage conditions, as summarized in Table 3-2.
The MMI PCB interfaces the front panel overlay, VGA display, rotary encoder,
and touch screen to the CPU via the main PCB. The MMI PCB contains control
circuitry for the primary and back-up alarms, and includes the hard keys and
LEDs on the front panel membrane keypad.
Power Supply
The power supply converts AC voltage to DC voltage to be used by the system
electronics. The switching power supply can accept voltage from 100 to 240 V
AC (50/60 Hz), and converts it to +5 V, + 12 V, and +29 V DC voltages. In the
absence of AC voltage, the power supply converts the +24V DC input voltage
from an external DC power source (Backup Battery or External Battery). The
power supply also includes power fail logic and charging circuitry for the
backup battery.
Backlight Inverter PCB
For 9.5-in. displays: the backlight inverter PCB converts 5 V to approximately
500 V to drive the backlight on the VGA display assembly.
For original 10.4-in. displays: the backlight inverter PCB converts 12 V to
approximately 500 V to drive the backlight on the VGA display assembly.
For 2nd generation 10.4-in. displays: the backlight inverter PCB converts 5V
to approximately 500V to drive the backlight on the VGA display assembly.
WARNING:The backlight inverter PCB generates high voltage. To avoid personal injury,
verify that the AC and external DC power sources (Backup Battery or External
Battery) are disconnected from the ventilator.
Real-Time Clock Battery
The real time clock battery is a 3.6-V lithium battery that supplies power to the
real time clock on the VGA controller PCB when ventilator power is off.
Backup Battery
The optional Backup Battery can power the ventilator for approximately 30
minutes under nominal settings if AC power is lost.
The optional External Battery supplements the Backup Battery, and can
provide an additional two hours of ventilator operation (depending on ventilator
settings). The ventilator runs on AC power when available, then External
Battery power if installed, and then switches to Backup Battery power when
External Battery power is depleted.
DC/DC Converter PCB (use with original 10.4-in. displays only)
The DC/DC converter PCB converts a 5-V input to a 12-V output for the
backlight inverter PCB on 10.4-in. displays.
Optical Rotary Encoder
The knob on the user interface is an optical rotary encoder. It converts a
mechanical position into a representative electrical signal using a patterned
disk or scale, a light source, and photosensitive elements.
Graphic User Interface (GUI)
Esprit ventilators include a 9.5-in. monochrome or 10.4-in. color-capable
liquid crystal display (LCD) screen. The 9.5-in. LCD is a monochrome 640 x
480 active matrix display. The 10.4-in LCD is a color 640 x 480 active matrix
display capable of operating in monochrome or color mode. The 10.4-in. GUI
has mounting screws at the bottom corners, while the 9.5-in. GUI does not.
The GUI includes an infrared (IR) touchframe that contains 24 vertical and 32
horizontal IR emitter detector pairs, each of which is sequenced at a high
frequency. When the screen is touched, breaking the IR beam, the x and y
coordinates that correspond to the position on the screen are communicated to
the microprocessor.