Philips Respironics Esprit Service Manual

Respironics V200/Esprit Ventilator
Service Manual
580-1000-02 H
This work is protected under Title 17 of the United States copyright code and is the sole property of Respironics. No part of this document may be copied or otherwise reproduced, or stored in any electronic information retrieval system, except as specifically permitted under United States copyright law, without the prior written consent of Respironics.
For Technical support, contact: Respironics, Inc. Customer Service
Within the U.S.A. 1-800-345-6443 Outside the U.S.A. 724-387-4000 Facsimile 724-387-5012
service@respironics.com
United States of America Respironics California, Inc. 2271 Cosmos Court Carlsbad, CA 92011 USA 1-800-345-6443 or 724-387-4000
Authorized Representative Respironics Deutschland, Inc. Gewerbestrasse 17 D-82211 Herrsching Deutschland +49-8-15-29-30-60
ii Respirionics V200/Esprit Ventilator Service Manual © Respironics, Inc. REF 580-1000-02 H
Table of Contents
1 Introduction and Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1 Recommended Tools and Test Equipment . . . . . . . . . . . . . . . . . . . . . . . 1-2
2 Warnings, Cautions, and Notes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.1 General Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.2 General Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.3 General Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
3 Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.1 Pneumatic System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.1.1 Blower Inlet Filter (F1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.1.2 Muffler (Silencer). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.1.3 Blower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.1.4 Cooling Fan Filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.1.5 Cooling Coil. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.1.6 Cooling Coil Fan. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.1.7 Air Valve Assembly (AV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3.1.8 Air Flow Sensor (FS1). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3.1.9 Oxygen Water Trap/Inlet Filter Assembly (F2) . . . . . . . . . . . . . . . 3-4
3.1.10 Oxygen Inlet Connector (O
3.1.11 Oxygen Supply Pressure Switch (PS1) . . . . . . . . . . . . . . . . . . . . 3-4
3.1.12 Oxygen Regulator (REG 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3.1.13 Oxygen Valve (OV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.1.14 Oxygen Flow Sensor (FS2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.1.15 Crossover Solenoid (SOL1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.1.16 Cross-Contamination Check Valve (CV5) . . . . . . . . . . . . . . . . . . . 3-5
3.1.17 Safety Valve Pilot Solenoid (SOL2) . . . . . . . . . . . . . . . . . . . . . . 3-5
3.1.18 Safety Valve (SV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
3.1.19 Air System Check Valve (CV2) . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
3.1.20 Inspiratory Non-Rebreathing Check Valve (CV3) . . . . . . . . . . . . . 3-6
3.1.21 Pressure Relief Valve (PRV). . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
3.1.22 Oxygen Sensor (OS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
3.1.23 Inspiratory Pressure Transducer (PT3) . . . . . . . . . . . . . . . . . . . . 3-6
3.1.24 Inspiratory Pressure Transducer Solenoid (SOL4) . . . . . . . . . . . . 3-7
3.1.25 Heated Exhalation Filter (F3) . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
3.1.26 Exhalation Pressure Transducer (PT2) . . . . . . . . . . . . . . . . . . . . 3-7
) . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
2
REF 580-1000-02 H Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. iii
Table of Contents
3.1.27 Exhalation Pressure Transducer Solenoid (SOL3) . . . . . . . . . . . . 3-7
3.1.28 Exhalation Non-Rebreathing Check Valve (CV4) . . . . . . . . . . . . . 3-8
3.1.29 Exhalation Valve (EV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3.1.30 Exhalation Flow Sensor (FS3) . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3.2 Ventilator System Electronics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
3.2.1 Main PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
3.2.2 CPU PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
3.2.3 Analog PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
3.2.4 Digital PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
3.2.5 VGA Controller PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3.2.6 Blower Controller PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3.2.7 Motor Controller PCBs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3.2.8 Sensor PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3.2.9 Man-Machine Interface (MMI) PCB . . . . . . . . . . . . . . . . . . . . . 3-15
3.2.10 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
3.2.11 Backlight Inverter PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
3.2.12 Real-Time Clock Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
3.2.13 Backup Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
3.2.14 External Battery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
3.2.15 DC/DC Converter PCB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
3.2.16 Optical Rotary Encoder. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
3.2.17 Graphic User Interface (GUI) . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
3.2.18 Remote Alarm (Nurse Call) . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
4 Periodic Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.1 Cleaning the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.2 Annual PM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4.3 12,500-Hour PM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
5 Diagnostic Mode and
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.1 Entering Diagnostic Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.2 User Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.2.1 Setting the Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
5.2.2 Setting Altitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
5.2.3 Enabling Circuit Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5.2.4 Selecting 12 or 12 Hour Clock . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5.2.5 Confirm Backup Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5.3 Short Self Test (SST) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
5.4 Extended Self Test (EST) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
iv Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. REF 580-1000-02 H
Table of Contents
5.5 Software Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
5.6 Diagnostic Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
5.7 Hardware Diagnostics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
5.8 Pneumatic Component Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . 5-13
5.8.1 Oxygen Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
5.8.2 Oxygen Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
5.8.3 Oxygen Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
5.8.4 Air Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
5.8.5 Air Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
5.8.6 Inhalation and Exhalation Solenoids . . . . . . . . . . . . . . . . . . . . 5-18
5.8.7 Safety Valve and Safety Solenoid . . . . . . . . . . . . . . . . . . . . . . . 5-18
5.8.8 Pressure Relief Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19
5.8.9 Crossover Solenoid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-20
5.8.10 Exhalation Flow Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
5.8.11 Check Valve 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22
5.8.12 Check Valve 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22
5.8.13 Check Valve 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23
5.8.14 Filter Heater . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23
5.8.15 Blower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24
5.8.16 Inhalation/ Exhalation Pressure Transducers and Exhalation
Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-25
5.8.17 Oxygen Pressure Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-26
5.8.18 External Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-26
5.8.19 High Internal Oxygen Alarm Test . . . . . . . . . . . . . . . . . . . . . . . 5-26
5.8.20 Sensor PCB Voltage Indicators . . . . . . . . . . . . . . . . . . . . . . . . 5-27
5.9 Bacteria Filter Back Pressure Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-28
6 Diagnostic Codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6.1 Diagnostic Code 1012 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . 6-24
6.2 Diagnostic Code 5000 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . 6-26
7 Ventilator Communications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.1 Downloading Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.2 Programming the Ventilator Serial Number . . . . . . . . . . . . . . . . . . . . . . . 7-4
7.3 Enabling Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
7.4 Setting Up the Serial Interface for DRPT . . . . . . . . . . . . . . . . . . . . . . . . 7-6
7.5 Generating a Diagnostic Report (DRPT) . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
7.6 Analog Output Port (Chart Recorder) Pinout . . . . . . . . . . . . . . . . . . . . . 7-11
REF 580-1000-02 H Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. v
Table of Contents
8 Performance Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
8.1 Required Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
8.2 Preliminary Cleaning, Inspection and Setup . . . . . . . . . . . . . . . . . . . . . . 8-5
8.3 Preliminary Pneumatic Calibration Analyzer Setup . . . . . . . . . . . . . . . . . 8-6
8.4 Certifier FA Plus Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
8.4.1 Measurement Selection Screen . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
8.4.2 Averaging Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
8.4.3 Trigger Options Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
8.4.4 Configuration Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
8.4.5 Saving Configurations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-11
8.5 Performance Verification Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . 8-13
8.5.1 Test 1: Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-14
8.5.2 Test 2: Extended Self Test (EST). . . . . . . . . . . . . . . . . . . . . . . 8-15
8.5.3 Test 3: Air Flow Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-16
8.5.4 Test 4: Oxygen Flow Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . 8-19
8.5.5 Test 5: Pressure Accuracy. . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-21
8.5.6 Test 6: PEEP System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-23
8.5.7 Test 7: Breath Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-26
8.5.8 Test 8: Alarm/Analog Output Signals, Alarm Volume,
and Remote Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-27
8.5.9 Test 9: Gas Volume Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . 8-30
8.5.10 Test 10: Oxygen Accuracy. . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-32
8.5.11 Test 11: Heated Exhalation Bacteria Filter, Power Fail Alarm,
and Display Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-34
8.5.12 Test 12: Neonatal Option Testing . . . . . . . . . . . . . . . . . . . . . . 8-36
8.5.13 Test 13: Backup Battery and External Battery (If Installed) . . . . 8-38
8.6 Returning Ventilator to Normal Operation . . . . . . . . . . . . . . . . . . . . . . . 8-40
8.7 Performance Verification Troubleshooting/Repair . . . . . . . . . . . . . . . . . 8-42
8.7.1 Test 1: Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-42
8.7.2 Test 2: Extended Self Test . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-42
8.7.3 Test 3: Air Flow Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-43
8.7.4 Test 4: Oxygen Flow Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . 8-43
8.7.5 Test 5: Pressure Accuracy. . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-44
8.7.6 Test 6: PEEP System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-44
8.7.7 Test 7: Breath Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-44
8.7.8 Test 8: Alarm Output Signal, Volume, and Remote Alarm . . . . . 8-45
8.7.9 Test 9: Gas Volume Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . 8-45
8.7.10 Test 10: Oxygen Accuracy. . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-45
8.7.11 Test 11: Heated Exhalation Bacteria Filter, Power Failure Alarm,
and Display Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-46
vi Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. REF 580-1000-02 H
Table of Contents
8.7.12 Test 12: Neonatal Option Test. . . . . . . . . . . . . . . . . . . . . . . . . 8-47
8.7.13 Test 13: Backup Battery and External Battery Test . . . . . . . . . . 8-48
8.8 Neonatal Option Data Form. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-49
8.9 Electrical Safety/Extended Self Test (EST) Data Form . . . . . . . . . . . . . . 8-51
8.10 Performance Verification Data Form. . . . . . . . . . . . . . . . . . . . . . . . . . . 8-52
9 Esprit Ventilator Component
Removal/Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
9.1 Filter Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
9.2 Top Enclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
9.3 Sensor PCB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
9.4 Power Supply Fan/Shroud . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8
9.5 Backlight Inverter PCB (9.5-in. GUI) . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9
9.6 Backlight Inverter PCB (Original 10.4-in. GUI only) . . . . . . . . . . . . . . . 9-10
9.7 DC/DC Converter PCB (Original 10.4-in. GUI only) . . . . . . . . . . . . . . . . 9-11
9.8 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-12
9.9 Power Supply Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-15
9.10 MMI PCB (9.5-in. GUI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-17
9.11 MMI PCB (10.4-in. GUI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-20
9.12 GUI Assembly (9.5-in. GUI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-22
9.13 GUI Assembly (Original 10.4-in. GUI) . . . . . . . . . . . . . . . . . . . . . . . . . 9-24
9.14 GUI Assembly (2nd Generation 10.4-in. GUI). . . . . . . . . . . . . . . . . . . . 9-25
9.15 Intensity and Volume Potentiometers (9.5-in. GUI) . . . . . . . . . . . . . . . . 9-28
9.16 Intensity and Volume Potentiometers (10.4-in. GUI’s). . . . . . . . . . . . . . 9-29
9.17 Rotary Encoder (9.5-in. GUI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-30
9.18 Rotary Encoder (10.4-in. GUI’s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-31
9.19 GUI Front Panel Overlay (10.4-in. GUI’s) . . . . . . . . . . . . . . . . . . . . . . . 9-32
9.20 VGA Display Assembly (9.5-in. GUI) . . . . . . . . . . . . . . . . . . . . . . . . . . 9-33
9.21 Touch Screen/LED Indicator Assembly and Front Panel Overlay
(9.5-in. GUI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-34
9.22 Backlight Inverter PCB, VGA Display, and Touch Frame
(Original 10.4-in. GUI). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-35
9.23 Transition PCB, LCD, IR Touch Frame, and Backlight Inverter PCB
(2nd Generation 10.4-in. GUI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-36
9.24 GUI Cleaning and Dust Gasket Installation
(2nd Generation 10.4-in. GUI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-38
9.25 Increased Minimum Alarm Volume Harness . . . . . . . . . . . . . . . . . . . . . 9-41
9.26 Backup Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-42
9.27 AC Distribution Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-43
9.28 AC Cord Bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-45
REF 580-1000-02 H Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. vii
Table of Contents
9.29 Humidifier Receptacle Rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-46
9.30 Printed Circuit Boards (Daughter PCBs) (Except Main PCB) . . . . . . . . . . 9-48
9.31 CPU PCB with 4.10 or Greater . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-49
9.32 Main PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-50
9.33 Remote Alarm Jack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-53
9.34 Power Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-54
9.35 FIO
9.36 Exhalation Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-56
9.37 Exhalation Valve Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-58
9.38 Primary Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-61
9.39 Filter Heater Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-62
9.40 Oxygen Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-63
9.41 Inspiratory Manifold Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-65
9.42 Installing the Cable Extension Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-69
9.43 Three-Station Solenoid Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-71
9.44 Air Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-72
9.45 Air Valve Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-73
9.46 Oxygen Valve Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-74
9.47 Oxygen Regulator Assembly (with Oxygen Pressure Switch) . . . . . . . . . . 9-76
9.48 Elapsed Time Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-78
9.49 Oxygen Water Trap Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-79
9.50 Blower Motor Controller PCB (Original) . . . . . . . . . . . . . . . . . . . . . . . . 9-80
9.51 Blower Motor Controller PCB (Updated) . . . . . . . . . . . . . . . . . . . . . . . . 9-81
9.52 Blower Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-82
9.53 Blower Muffler Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-85
9.54 Cooling Fan/Cooling Coil Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-86
9.55 GUI Upgrades . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-88
9.56 Replacing the GUI (9.5-in. to original 10.4-in.) . . . . . . . . . . . . . . . . . . 9-90
9.57 Replacing the GUI (9.5-in. to 2nd Generation 10.4-in.) . . . . . . . . . . . . 9-95
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-55
2
9.56.1 Installing the CPU and DC/DC Converter PCBs . . . . . . . . . . . . . 9-93
9.56.2 Downloading Ventilator Software . . . . . . . . . . . . . . . . . . . . . . . 9-93
9.56.3 Calibrating the Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-94
9.56.4 Enabling Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-94
9.56.5 Final Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-94
9.57.1 Installing the CPU PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-96
9.57.2 Downloading Ventilator Software . . . . . . . . . . . . . . . . . . . . . . . 9-97
9.57.3 Calibrating the Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-97
9.57.4 Enabling Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-97
9.57.5 Final Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-97
viii Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. REF 580-1000-02 H
Table of Contents
9.58 Replacing the GUI (Original 10.4-in. to 2nd Generation 10.4-in) . . . . . . 9-98
9.58.1 Installing the CPU PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-102
9.58.2 Downloading Ventilator Software . . . . . . . . . . . . . . . . . . . . . . 9-102
9.58.3 Calibrating the Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-103
9.58.4 Enabling Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-103
9.58.5 Final Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-103
10 Respironics V200 Ventilator Component Removal/Installation . . . . . . . 10-1
10.1 Inlet Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
10.1.1 Air Inlet Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
10.1.2 Blower Inlet Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
10.1.3 Oxygen Inlet Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
10.2 GUI Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-6
10.2.1 GUI Assembly Removal/Installation . . . . . . . . . . . . . . . . . . . . . 10-6
10.2.2 Backlight Inverter PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8
10.2.3 Transition PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-9
10.2.4 LCD Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-9
10.2.5 Touch Frame and Alarm LED PCB . . . . . . . . . . . . . . . . . . . . . . 10-9
10.2.6 GUI Front Panel Overlays . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-10
10.2.7 GUI Cleaning and Dust Gasket Installation . . . . . . . . . . . . . . . 10-11
10.3 Separating the Top and Bottom Enclosures . . . . . . . . . . . . . . . . . . . . 10-14
10.4 Enclosure Brace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-16
10.5 Sensor PCB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-17
10.6 Intensity and Volume Potentiometers . . . . . . . . . . . . . . . . . . . . . . . . . 10-19
10.7 Rotary Encoder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-20
10.8 Power Supply Shroud . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-21
10.9 Power Supply Fan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-21
10.10 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-22
10.10.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-22
10.10.2Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-24
10.11 Power Supply Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-25
10.12 MMI PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-26
10.13 Backup Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-28
10.14 AC Distribution Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-29
10.15 AC Power Cord Bracket and Power Cord . . . . . . . . . . . . . . . . . . . . . . . 10-31
10.16 Elapsed Time Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-32
10.17 Printed Circuit Boards (Daughter PCBs) (Except Main PCB) . . . . . . . . . 10-33
10.18 Main PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-34
10.19 Remote Alarm Jack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-37
10.20 Power Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-38
REF 580-1000-02 H Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. ix
Table of Contents
10.21 FIO2 Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-39
10.22 Exhalation Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-40
10.22.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-40
10.22.2Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-40
10.23 Exhalation Valve Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-42
10.23.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-42
10.23.2Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-43
10.24 Primary Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-43
10.25 Filter Heater Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-44
10.26 Oxygen Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-45
10.26.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-45
10.26.2Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-46
10.27 Air Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-47
10.28 Inspiratory Manifold Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-48
10.28.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-48
10.28.2Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-50
10.29 Three-Station Solenoid Valve Assembly . . . . . . . . . . . . . . . . . . . . . . . 10-51
10.30 Air Valve Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-52
10.31 Oxygen Valve Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-53
10.32 Oxygen Water Trap/Inlet Filter Assembly . . . . . . . . . . . . . . . . . . . . . . 10-54
10.32.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-54
10.32.2Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-55
10.33 Oxygen Regulator Assembly (with Oxygen Pressure Switch) . . . . . . . . . 10-56
10.33.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-56
10.33.2Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-56
10.34 Blower Motor Controller PCB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-58
10.35 Blower Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-59
10.35.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-59
10.35.2Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-60
10.36 Blower Muffler Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-62
10.36.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-62
10.36.2Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-62
10.37 Cooling Fan/Cooling Coil Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . 10-63
10.37.1Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-63
10.37.2Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-64
10.38 Installing the CPU PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-66
10.38.1Downloading Ventilator Software . . . . . . . . . . . . . . . . . . . . . . 10-66
10.38.2Calibrating the Display Screen . . . . . . . . . . . . . . . . . . . . . . . 10-66
10.38.3Enabling Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-66
10.38.4Final Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-66
x Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. REF 580-1000-02 H
Table of Contents
11 Where to Go for Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1
12 Respironics V200/Esprit Ventilator Replacement Parts List . . . . . . . . . 12-1
12.1 Major Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-1
12.1.1 Electronics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-2
12.1.2 GUI Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-3
12.1.3 Oxygen Pneumatics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-7
12.1.4 Air Pneumatics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-8
12.1.5 Exhalation Pneumatics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-10
12.1.6 Inspiration Pneumatics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-11
12.1.7 Back Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-12
12.1.8 Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-13
12.2 Service Part Inventory List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-15
12.3 Complete Repair Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-17
A Pneumatic Schematics for EST. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
A.1 Block Patient Wye (Test 1, Step 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
A.2 Block Patient Wye (Test 1, Step 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
A.3 Block Patient Wye (Test 1, Step 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
A.4 Block Patient Wye (Test 1, Step 5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4
A.5 Safety Valve (Test 2, Step 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-5
A.6 Safety Valve (Test 2, Step 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-6
A.7 Blower (Test 3, Step 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-7
A.8 Blower (Test 3, Step 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-8
A.9 Oxygen Supply (Test 4, Step 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-9
A.10 Oxygen Supply (Test 4, Step 2). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-10
A.11 Crossover Circuit (Test 5, Step 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-11
A.12 Crossover Circuit (Test 5, Step 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-12
A.13 Crossover Circuit (Test 5, Step 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-13
A.14 Oxygen Delivery (Test 6, Step 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-14
A.15 Oxygen Sensor (Test 7, Step 1). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-15
A.16 Air Delivery (Test 8, Step 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-16
A.17 Pressure Relief Valve (Test 9, Step 1) . . . . . . . . . . . . . . . . . . . . . . . . . A-17
A.18 Exhalation Valve (Test 10, Step 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . A-18
A.19 Exhalation Valve (Test 10, Step 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . A-19
A.20 Patient Circuit (Test 11, Step 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-20
A.21 Patient Circuit (Test 11, Step 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-21
A.22 Heated Filter (Test 12, Step 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-22
A.23 Heated Filter (Test 12, Step 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-23
REF 580-1000-02 H Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. xi
Table of Contents
B Field Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
C Respi-Link . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Index-1
xii Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. REF 580-1000-02 H

Chapter 1. Introduction and Intended Use

The Esprit and Respironics V200 ventilators are microprocessor-controlled, electrically powered mechanical ventilators. They are intended for use by qualified medical personnel to provide continuous or intermittent ventilatory support for adult, pediatric, and neonatal patients as prescribed by a physician. They are intended for use in either invasive or non-invasive applications in institutional environments.
The Esprit and Respironics V200 ventilators meet or exceed all applicable safety requirements, consensus guidelines, U.S. regulatory statutes, and international regulatory standards for life support/mechanical ventilation devices.
Read this manual thoroughly prior to performing service or maintenance on the ventilators. This manual contains advanced troubleshooting, calibration, and maintenance instructions. All maintenance and repair work should be performed by qualified biomedical technicians who have received appropriate training and authorization to provide maintenance, repair, and service for the Esprit and Respironics V200 ventilators.
Review the Esprit or Respironics V200 Operator’ s Manual and become familiar
with the operation before running tests, checking operational readiness, or initiating patient use. The operator’s manual includes important information about ventilator safety and operation.
Schematic diagrams of the Esprit and Respironics V200 ventilators are available upon request.
For additional information about accessories or related equipment, such as humidifiers and remote alarm systems, refer to the appropriate instruction manual prior to operating the ventilator.
WARNING: Patients on life-support equipment should be visually monitored by competent
medical personnel, since life-threatening circumstances may arise that may not activate alarms. Heed all appropriate alarms and follow the instructions and warnings in this service manual and the operator’s manual. Always check life-support equipment for proper operation before use.
NOTE: The Esprit or Respironics V200 Operator’s Manual lists all applicable
warnings and cautions. Review these notices thoroughly before operating the ventilator.
REF 580-1000-02 H Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. 1-1
Chapter 1
Introduction and Intended Use
1.1 Recommended Tools and Test
Equipment
The following table lists the recommended tools, test equipment, and materials required for ventilator service and maintenance (Table 1-1). Test equipment must meet the requirements in Table 1-2.
Table 1-1: Recommended Test Equipment, Tools, and Materials
Description Manufacturer and Model
Test Equipment
Adapter, USB to serial Respironics P/N 1022895
Analog output port signal selector Respironics P/N 1010891
Digital multimeter (DMM) and frequency counter accurate to three decimal places
Electrical safety analyzer Dale LT 544D or equivalent
Pneumatic calibration analyzer capable of measuring low pressure (cmH (PSI), flow rate (LPM), volume (liters), and oxygen concentration (accuracy ± 2%)
Syringe, 10-mL calibrated (neonatal testing) Hans Rudolph 5220 or equivalent
Service kit (*included in the kit) Respironics P/N 1021670
Adapter, oxygen regulator test * Respironics P/N 1001376
Adapter, parallel port * Respironics P/N 1004644
Cable, remote alarm test * Respironics P/N 1027818
Cable adapter, remote alarm test Respironics P/N 1027817
Cable assembly, null modem * Respironics P/N 1022815 or equivalent
Cork, silicone * Respironics P/N 1001735 or equivalent
Test lung, 1 L * Respironics P/N 1021671 or equivalent
O), high pressure
2
Local supplier
Certifier FA Plus Ventilator Gas Analyzer with oxygen sensor kit (Respironics P/N 1040311 or 1040312) or equivalent
Ventilator Accessories
Adapter, oxygen sensor Respironics P/N 1001736 or equivalent
Connector, plastic, 22mm OD Respironics P/N C06335 or equivalent
Coupling, silicone rubber Respironics P/N C06348 or equivalent
Patient circuit tubes, adult, 42-in. smooth bore (2 each)
Tee, plastic with silicone rubber coupling Respironics P/N C06260 or equivalent
Tubing, silicone, 3/16 in. ID x 6.5 ft., PAP Respironics P/N C06686 or equivalent
Wye, reusable, 22mm OD/15mm ID Respironics P/N 1003070 or equivalent
Respironics P/N 1003643 or equivalent
Hand Tools and Materials
ESD-safe field service vacuum cleaner 3M model 497-AJM or equivalent
Hex key set, metric (rounded ends), 1.5 to 4 mm Local supplier
Hex key set, standard (rounded ends), 0.050 to 5/32 in.
Local supplier
1-2 Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. REF 580-1000-02 H
Introduction and Intended Use
Chapter 1
Table 1-1: Recommended Test Equipment, Tools, and Materials (Continued)
Description Manufacturer and Model
Loctite 222 Respironics P/N 200-1000-00
Lubricant, Dupont Krytox GPL226 Res pironics P/N 100-1012-00
Mild detergent or antiseptic wipes Local supplier
PC or laptop (for downloading software) with Windows 95 or later, serial port and CD ROM drive
Pliers Local supplier
Pliers, needle nose Local supplier
Ratchet, box, 8 mm Local supplier
Screwdriver, flat head, #2 Local supplier
Screwdriver, flat head, #3 Local supplier
Screwdriver, Phillips head, #2 Local supplier
Screwdriver, pen size flat head Local supplier
Screwdriver, pen size Phillips head Local supplier
Static dissipative field service kit Local supplier
Thread tape Local supplier
Tie wraps, 3 in. length Respironics P/N 500-1000-66 or equivalent
Tie wraps, 8 in. length Respironics P/N 500-1000-62 or equivalent
Tie wrap gun Local supplier
Torque driver capable of 5 to 25 in.-lbs. Local supplier
Tweezers, angled Local supplier
Wire cutters Local supplier
Wrench, open end, 7/32 in. Local supplier
Wrench, open end, 1/4 in. Local supplier
Wrench, open end, 5/16 in. Local supplier
Wrench, open end, 7/16 in. Local supplier
Wrench, open end, 1/2 in. Local supplier
Wrench, open end, 11/16 in. Local supplier
Wrench, open end, 5.5 mm Local supplier
Wrench, open end, 7 mm Local supplier
Wrench, open end or box, #10 Local supplier
Wrench, open end or socket, 3/16 in. Local supplier
Wrench, socket, 9/32 in. with removable 6-in. extension bar
Wrench, socket, #10 Local supplier
Local supplier
Local supplier
REF 580-1000-02 H Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. 1-3
Chapter 1
Introduction and Intended Use
Table 1-2: Test Equipment Specifications
Unit of
Measurement
Pressure -25 to 150 cmH2O 0 to
Flow (standard) 0 to 300 SLPM ± 2% of reading ± 0.20 SLPM (whichever is
Torque 5 to 25 in-lb. ± 1 in-lb. of reading
Volume (STP) 0 to 10 L STP ± 2% of reading ± 0.20 L STP
Voltage DC: ± 5 to 50 V
Range Accuracy
100 psi
AC: 2 to 300 V
± 1% of reading or ± 0.20 cmH2O (whichever is greater). ± 2% of reading @ -1 to 38ºC (30 to 100ºF)
greater)
± 2% of reading
1-4 Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. REF 580-1000-02 H

Chapter 2. Warnings, Cautions, and Notes

Throughout this manual the following definitions apply:
WARNING: A condition that could cause injury to a patient or operator if the operating
instructions in this manual are not followed correctly.
CAUTION: A condition that could cause damage to, or shorten the service life of,
the Esprit/Respironics V200 Ventilator.

2.1 General Warnings WARNING: Do not obstruct the emergency air intake near the oxygen water trap/inlet

filter assembly.
WARNING: Never troubleshoot while a patient is connected to the ventilator, since normal
operation is suspended.
WARNING: If the ventilator has been operating, the exhalation filter heater conductor may
be hot. Use caution when removing the filter.
WARNING: To prevent disease transmission, use protective equipment when handling
contaminated bacterial filters or other patient accessories.
WARNING: To avoid personal injury, always disconnect external AC and DC power
sources and high-pressure oxygen sources from the ventilator before servicing.
WARNING: Explosion hazard. Do no operate the ventilator in the presence of flammable
anesthetic agents.

2.2 General Cautions CAUTION: Troubleshooting and repair should be performed only by a qualified

service technician. Respironics Factory Service Training is highly recommended prior to performing service procedures on the Respironics Ventilator. Contact Customer Service at 1-800-345-6443 or 724-387­4000 for more information.
CAUTION: Use only Respironics ventilator repair/service parts. Only Respironics
parts are designed for use in this ventilator. Use of non-Respironics repair parts may alter ventilator reliability resulting in damage. Use of non­Respironics repair parts will affect your warranty. Contact Customer Service at 1-800-345-6443 or 724-387-4000 for more information.
CAUTION: Do not modify oxygen diameter index safety systems (DISS) connector on
rear panel. Use only medical grade oxygen.
REF 580-1000-02 H Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. 2-1
Chapter 2
Warnings, Cautions, and Notes
CAUTION: Always ensure that you are following proper electrostatic discharge (ESD)
grounding procedures before handling static-sensitive devices.
CAUTION: Be careful not to pull or crimp any cables, tubes, or wires.

2.3 General Notes NOTE: This manual covers multiple versions of the Esprit and Respironics V200

Ventilators. The information given may not exactly match your own ventilator.
2-2 Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. REF 580-1000-02 H

Chapter 3. Theory of Operation

The Esprit and Respironics V200 ventilators are microprocessor-controlled devices that can deliver air, oxygen, or a mixture of air and oxygen to the patient’s lungs in a predetermined manner to augment or replace the work normally performed by the patient’s respiratory system. It uses electromechanical control circuits, flow and pressure monitors, and software programs to deliver breaths as a flow or pressure controller.
The ventilators include a graphic user interface (GUI), internal blower, and inspiratory module that mixes air and oxygen. The ventilators can operate from a 40 to 90 psig (276 to 620 kPa) medical grade oxygen source for enriched oxygen operation. It also includes multiple communications interfaces and an internal power supply that can run from a 100 to 240 V AC 50/60 Hz or 24 V DC power sources.
Schematic diagrams of the Esprit and Respironics V200 ventilators are available upon request.

3.1 Pneumatic System

The pneumatic system consists of these subsystems (see Figure 3-1):
Internal blower (air source)
Oxygen regulator (oxygen source)
Inspiratory module
Heated exhalation filter assembly
Exhalation valve assembly
Expiratory flow sensor
The internal blower generates the air pressure necessary for breath delivery, eliminating the need for an external source of medical-grade compressed air. An internal oxygen regulator controls wall oxygen pressure. The ventilator mixes air and oxygen in the inspiratory module before delivery to the patient.
Based on operator settings, the central processing unit (CPU) controls the air valve, oxygen valve, and exhalation valve through stepper motor controller printed circuit boards (PCBs). As flow is delivered to the patient, the air and oxygen flow sensors and two pressure sensors provide feedback to the CPU.
The pressure relief and safety valves in the inspiratory module provide for patient safety in the event of an over-pressure condition or any component or system failure that could interfere with the patient’s ability to breathe when connected to the ventilator.
REF 580-1000-02 H Esprit® / V200 Ventilator Service Manual © Respironics, Inc. 3-1
Chapter 3
Blower
Oxygen
Regulator
Inspiratory Module
• Gas Mixing
• Pressure Relief Valve
• Safety Valve To P at ien t
From Patient
Heated
Exhalation
Filter
Assembly
Exhalation
Valve
Expiratory
Flow
Sensor
Room
Air
Oxygen Supply
Air
O
2
Theory of Operation
The exhalation filter reduces the risk of contamination or component damage due to bacteria or moisture in expired gases. The exhalation filter is housed in a heated sleeve, which reduces the relative moisture condensation in the exhalation filter, exhalation valve, and expiratory flow sensor. Exhaled gas is then vented to atmosphere.
Figure 3-1: Pneumatic System Block Diagram
Figure 3-2 shows the pneumatic system and its components.
3-2 Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. REF 580-1000-02 H
Figure 3-2: Pneumatic Schematic
Chapter 3
Theory of Operation

3.1.1 Blower Inlet Filter (F1)

The blower draws room air through the blower inlet filter (F1) and the muffler (silencer) and outputs flow to the air valve assembly (AV). The blower inlet filter removes coarse particulate from ambient air as it is entrained into the blower assembly. See Chapter 4 for periodic maintenance information.

3.1.2 Muffler (Silencer)

The muffler reduces the noise of air flow into the blower by channeling the air through a baffled system lined with sound absorbing material.

3.1.3 Blower

The blower draws room air though the air inlet filter and outputs the air that is delivered to the patient, and provides the pilot pressure that can actuate the safety valve. The blower contains a DC motor and a series of stator and impeller assemblies. It can provide at least 200 LPM of flow.
Blower speed is automatically adjusted to account for differences in gas density due to altitude. The altitude can be adjusted from the hardware screen in Diagnostics mode. The high pressure alarm limit setting also affects blower speed.

3.1.4 Cooling Fan Filter

The cooling fan filter removes coarse particulate from ambient air entrained by the cooling fan. See Chapter 4 for periodic maintenance information.

3.1.5 Cooling Coil

The cooling coil is a copper tube connected to the outlet of the blower dissipate heat from gas leaving the blower before it reaches the air valve.

3.1.6 Cooling Coil Fan

The 24 V DC cooling coil fan removes the heat dissipated by the cooling coil and blower.
REF 580-1000-02 H Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. 3-3
Chapter 3
Oxygen inlet connector
Theory of Operation

3.1.7 Air Valve Assembly (AV)

The air valve assembly includes a stepper motor that meters air flow from the blower to achieve the target flow under CPU control, based on operator selected parameters. It can deliver up to 200 LPM of flow.

3.1.8 Air Flow Sensor (FS1)

The air flow sensor measures flow from the air valve. The ventilator uses this measurement to provide closed loop control of the air valve and to compute the flow and volume delivered to the patient. A thermistor in the flow sensor measures the temperature of the air and provides the microprocessor with information to compensate the delivered flow.

3.1.9 Oxygen Water Trap/Inlet Filter Assembly (F2)

The oxygen water trap/inlet filter assembly consists of a 5-micron (µ) filter to remove particulate (both dry and liquid) from the oxygen gas supply, a bowl with drain for accumulated water, and an oxygen inlet connector.

3.1.10 Oxygen Inlet Connector (O2)

The oxygen inlet connector provides a country-specific connection point for an external oxygen gas supply of 40 to 90 psig (276 to 620 kPa).

3.1.11 Oxygen Supply Pressure Switch (PS1)

The oxygen supply pressure switch is part of the oxygen regulator. PS1 is a normally open (NO) switch that monitors oxygen supply pressure by closing when measured pressure is over 40 psig (276 kPa), and sends a signal to the sensor PCB to indicate if supply pressure is adequate at the oxygen inlet.
PS1 opens if measured pressure is less than 35 psig (241.3 kPa). If the oxygen supply pressure switch opens during normal ventilation (at O
> 21%),
2
a low oxygen supply alarm results.

3.1.12 Oxygen Regulator (REG 1)

The oxygen regulator reduces the oxygen supply pressure to the proper inlet pressure for the oxygen valve (22-24 psig, or 152-165 kPa @ 180 LPM) and supplies the regulated pressure to the crossover solenoid, which pilots the safety valve.
3-4 Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. REF 580-1000-02 H
Chapter 3
SOL2
Theory of Operation

3.1.13 Oxygen Valve (OV)

The oxygen valve assembly contains a stepper motor that meters flow from the oxygen regulator to achieve the target flow under CPU control, based on operator-selected parameters. It can deliver up to 200 LPM of flow.

3.1.14 Oxygen Flow Sensor (FS2)

The oxygen flow sensor measures the flow from the oxygen valve. The ventilator uses this measurement to provide closed loop control of the oxygen valve and to compute the flow and volume delivered to the patient. A thermistor contained in the flow sensor measures the temperature of the oxygen and provides temperature compensation information to the microprocessor for delivered flow.

3.1.15 Crossover Solenoid (SOL1)

The crossover solenoid is a three-way valve that supplies either air or oxygen pressure to pilot (hold) the safety valve closed during normal ventilation. In its normal state, SOL1 is normally de-energized to pilot the safety valve with oxygen. If oxygen pressure is lost, SOL1 is energized and air (rather than oxygen) controls the safety valve.

3.1.16 Cross-Contamination Check Valve (CV5)

The cross-contamination check valve prevents the oxygen supply from entering the air delivery system (blower) in the event of a crossover solenoid leak.

3.1.17 Safety Valve Pilot Solenoid (SOL2)

SOL2 directs the output of the crossover solenoid to the safety valve or vents the pilot pressure line to atmosphere.
During normal operation, SOL2 is energized and directs pressure from the crossover solenoid to close the safety valve. During a high priority alarm condition such as an occlusion or ventilator failure (VENT INOP) mode, SOL2 is deenergized, the safety valve opens, and the patient can breathe room air.
REF 580-1000-02 H Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. 3-5
Chapter 3
O ring
Diaphragm
Spring
CV2
CV3
PT3
Sensor PCB
Theory of Operation

3.1.18 Safety Valve (SV)

The safety valve contains a spring-loaded diaphragm that is controlled by safety valve pilot solenoid (SOL2). Under normal conditions the safety valve is closed, allowing delivered flow to reach the patient. In the event of a safety valve open (SVO) condition, pilot pressure is vented to atmosphere, which opens the safety valve and allows the patient to breathe room air through the safety port at the back of the ventilator.

3.1.19 Air System Check Valve (CV2)

CV2 prevents oxygen from entering the air delivery system in the event of a blower failure.

3.1.20 Inspiratory Non-Rebreathing Check Valve (CV3)

CV3 prevents the patient from exhaling through the inspiratory limb during a safety valve open condition, which prevents the patient from rebreathing exhaled gas.

3.1.21 Pressure Relief Valve (PRV)

The pressure relief valve provides a backup to the operator adjustable high­pressure alarm and prevents excessive pressures in the patient circuit. The pressure relief valve is spring-loaded to limit the maximum circuit pressure to 130 to 140 cmH
O.
2

3.1.22 Oxygen Sensor (OS)

The optional oxygen sensor is installed between the 22-mm inspiratory port and the inspiratory bacteria filter. The galvanic oxygen sensor measures the oxygen concentration of the blended gas as it leaves the inspiratory manifold. The output signal from the sensor is used to determine whether measured oxygen concentration is within 6% of the %O
setting. If not, a high or low
2
oxygen concentration alarm occurs.

3.1.23 Inspiratory Pressure Transducer (PT3)

PT3 on the sensor PCB monitors system pressure from the inspiratory side of the patient circuit during exhalation pressure transducer autozeroing, ensuring uninterrupted pressure monitoring. PT3 is also used with the exhalation pressure transducer (PT2) to detect patient circuit occlusions.
3-6 Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. REF 580-1000-02 H
Chapter 3
SOL4
PT2
Sensor PCB
SOL3
Theory of Operation

3.1.24 Inspiratory Pressure Transducer Solenoid (SOL4)

SOL4 periodically vents the inspiratory pressure transducer to atmosphere and makes a measurement at zero (atmospheric) pressure. Periodically autozeroing the transducer allows it to correct the slight zero voltage drift that can occur over time, and improves the overall accuracy of the pressure measurement.
During normal operation, SOL4 is de-energized and applies patient circuit pressure to the inspiratory pressure transducer. During an autozero, SOL4 is energized, venting the transducer to atmosphere. This occurs during power on self test (POST), at the beginning of a breath one minute after POST, six minutes after POST, eleven minutes after POST, and hourly thereafter.

3.1.25 Heated Exhalation Filter (F3)

The heated exhalation filter includes a heated filter sleeve and a bacteria filter. The heater protects the exhalation flow sensor and exhalation system components from condensation by heating exhaled gas (which has cooled in the exhalation limb) above its dew point.
The exhalation bacteria filter protects the exhalation flow sensor and exhalation system component from contaminants and filters exhaled gas before it is vented to atmosphere.

3.1.26 Exhalation Pressure Transducer (PT2)

The exhalation pressure transducer on the sensor PCB measures patient circuit pressure from the exhalation side of the patient circuit. During normal operation PT2 is the primary transducer for measuring patient pressures, including peak inspiratory pressure (PIP), mean airway pressure (MAP), end inspiratory pressure, and auto-PEEP. The exhalation pressure transducer provides monitoring data for closed loop control.

3.1.27 Exhalation Pressure Transducer Solenoid (SOL3)

The exhalation pressure transducer solenoid periodically vents the exhalation pressure transducer to atmosphere and measures zero (atmospheric) pressure. Periodically autozeroing the transducer allows it to correct the slight zero voltage drift that can occur over time, and improves overall pressure measurement accuracy.
During normal operation, SOL3 is de-energized and applies patient circuit pressure to the exhalation pressure transducer. During an autozero, SOL3 is energized, venting the transducer to atmosphere. This occurs during POST, at the beginning of a breath one minute after POST, six minutes after POST, eleven minutes after POST, and hourly thereafter.
REF 580-1000-02 H Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. 3-7
Chapter 3
CV4
Theory of Operation

3.1.28 Exhalation Non-Rebreathing Check Valve (CV4)

CV4 prevents the patient from inspiring room air through the exhalation limb of the patient circuit. During normal operation, it blocks the exhalation system from atmosphere, allowing the patient to trigger a breath.

3.1.29 Exhalation Valve (EV)

The exhalation valve is controlled by a stepper motor. At the beginning of an inspiration, the exhalation valve shuts to create a closed circuit and allow the patient system to pressurize. The exhalation valve opens at the beginning of exhalation, allowing system pressure to vent to atmosphere.
The exhalation valve also regulates positive end expiratory pressure (PEEP) and expiratory positive airway pressure (EPAP) levels during exhalation.

3.1.30 Exhalation Flow Sensor (FS3)

The exhalation flow sensor measures the flow leaving the ventilator, which includes gas exhaled by the patient, tubing compliance volume, and bias flow (if flow triggering or Auto-Trak triggering is selected). The ventilator uses the exhaled flow measurement to compute flow and volume coming from the patient and the circuit.
A thermistor in the flow sensor measures the temperature of the gas and provides the microprocessor with information to compensate the measured flow.
3-8 Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. REF 580-1000-02 H
Chapter 3
Theory of Operation

3.2 Ventilator System Electronics

Figure 3-3 shows the electronic system. Schematics are available upon request.
Figure 3-3: Electronic System Diagram
REF 580-1000-02 H Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. 3-9
Chapter 3
Theory of Operation
The ventilator can be powered by a 100 to 240 VAC 50/60 Hz or external 24 V DC power source (backup battery or external battery). The power supply conditions the input voltage and distributes + 5 V, +12 V, -12 V, a n d + 2 9 V t o the main PCB and blower motor controller to power digital electronics, electropneumatic components, and displays. AC power to the humidifier port can be used on 100-120 V units only.
The microprocessor on the CPU PCB and programs stored in memory control the interaction of the pneumatic and electronic subsystems. Using inputs from electropneumatic sensors and the operator, the CPU controls the flow, pressure, and volume of air and oxygen to be delivered to the patient. The CPU also monitors alarms and independently monitors software execution.
The the CPU PCB, digital PCB, analog PCB, VGA controller PCB, and three stepper motor controller PCBs are vertically mounted on the main PCB.
Digital control signals from the CPU are sent to the analog PCB, where they are converted into analog signals that control blower speed and chart recorder outputs (pressure, flow, and volume). The sensor PCB conditions and converts analog data from the flow, pressure, and oxygen sensors, then sends the data to the analog PCB, where it is read by the CPU.
Ventilator data from the CPU is conditioned by the VGA and man-machine interface (MMI) PCBs, then displayed on an LCD.
Table 3-1 summarizes the electronic signal path sequences for the ventilator components.
Table 3-1: Signal Path Sequences
Component Signal Path Sequence
100% O
29 V enable Main PCB, CPU PCB, power supply
Air (AV), oxygen (OV), exhalation (EV) valves
Air (FS1), oxygen (FS2), exhalation (FS3) flow sensors
Alarm High indicator Front panel overlay, MMI PCB, main PCB, CPU PCB Alarm Med/Low indicator Front panel overlay, MMI PCB, main PCB, CPU PCB
Alarm Silence indicator Front panel overlay, MMI PCB, main PCB, digital PCB, CPU
Backlight (9.5 in display, standard on Esprit ventilator)
Backlight (10.4-in display, optional on Esprit ventilator, standard on Respironics V200 ventilator)
indicator Front panel overlay, MMI PCB, main PCB, digital PCB, CPU
2
PCB
Main PCB, motor controller PCB, CPU PCB
Sensor PCB, main PCB, analog PCB, CPU PCB
PCB
Backlight inverter PCB, DC/DC converter PCB, main PCB, backlight control potentiometer
Backlight inverter PCB, DC/DC converter PCB, main PCB, backlight control potentiometer
3-10 Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. REF 580-1000-02 H
Chapter 3
Theory of Operation
Table 3-1: Signal Path Sequences (Continued)
Component Signal Path Sequence
Backup alarm MMI PCB, main PCB, digital PCB, CPU PCB Battery/Charging indicator Front panel overlay, MMI PCB, power supply, main PCB,
CPU PCB
Battery/In Use indicator Front panel overlay, MMI PCB, main PCB, digital PCB, CPU
PCB
Battery/Low indicator Front panel overlay, MMI PCB, digital PCB, CPU PCB
Blower DAC Blower controller PCB, sensor PCB, main PCB, analog PCB,
Blower on/off Blower controller PCB, sensor PCB, main PCB, CPU PCB
Blower temperature switch Sensor PCB, main PCB, digital PCB, CPU PCB
Console: all keys Front panel overlay, MMI PCB, main PCB, CPU PCB
Crossover solenoid (SOL1) Main PCB, CPU PCB
Enclosure temperature sensor Sensor PCB, main PCB, analog PCB, CPU PCB
Exhalation pressure transducer (PT2)
Exhalation pressure transducer solenoid (SOL3)
External backup battery Power supply External Battery indicator Front panel overlay, MMI PCB, power supply
Heater Main PCB, CPU PCB
Inspiratory pressure transducer (PT3)
Inspiratory pressure transducer solenoid (SOL4)
Mains indicator Front panel overlay, MMI PCB, power supply
Nonmaskable interrupt (NMI) signal
Normal indicator Front panel overlay, MMI PCB, main PCB, CPU PCB
Oxygen Sensor (OS) Sensor PCB, main PCB, analog PCB, CPU PCB
Oxygen supply pressure switch (PS1)
Primary alarm MMI PCB, main PCB, digital PCB, CPU PCB
Primary alarm potentiometer MMI PCB
Printer Main PCB, CPU PCB
Remote alarm (nurse call) Main PCB, CPU PCB
Rotary encoder MMI PCB, main PCB, digital PCB, CPU PCB Safety Valve indicator Front panel overlay, MMI PCB, main PCB, CPU PCB
Safety valve pilot solenoid (SOL2) Main PCB, CPU PCB Screen Lock indicator Front panel overlay, MMI PCB, main PCB, CPU PCB
Touch screen MMI PCB, main PCB, digital PCB, CPU PCB Vent Inop indicator Front panel overlay, MMI PCB, main PCB, CPU PCB
CPU PCB
Sensor PCB, main PCB, analog PCB, CPU PCB
Main PCB, CPU PCB
Sensor PCB, main PCB, analog PCB, CPU PCB
Main PCB, CPU PCB
Sensor PCB, main PCB, CPU PCB
Sensor PCB, main PCB, digital PCB, CPU PCB
REF 580-1000-02 H Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. 3-11
Chapter 3
Theory of Operation
Table 3-1: Signal Path Sequences (Continued)
Component Signal Path Sequence
VGA backlight intensity potentiometer (9.5 in display, standard on Esprit ventilator)
VGA backlight intensity potentiometer (10.4-in display, optional on Esprit ventilator, standard on Respironics V200 ventilator)
VGA display (9.5 in display, standard on Esprit ventilator)
VGA display (10.4-in display, optional on Esprit ventilator, standard on Respironics V200 ventilator)
MMI PCB, backlight inverter PCB
Backlight inverter PCB
MMI PCB, main PCB, VGA controller PCB, CPU PCB
Main PCB, VGA controller PCB, CPU PCB

3.2.1 Main PCB

The CPU and other ventilator logic interact through the system data, address, and control buses on the main PCB. The main PCB receives input signals from various keys on the console or touch screen display and sends them to the CPU. The main PCB also contains signal inputs for non-maskable interrupt, running on AC, and running on external battery.
The main PCB receives control signals from the CPU and outputs them to various pneumatic components and console indicators. The main PCB receives signals from the digital PCB to turn on the indicators for alarm silence, 100% oxygen, AC power, external battery power, and backup battery status. The main PCB receives signal from the CPU PCB to turn on the backup alarm, enable
24V, and the Screen Lock, Battery/Chargi ng, and Vent Inop indicators. The CPU PCB reads the Accept key from the main PCB.
The main PCB includes a normal open and normal closed relay that can trigger the remote nurse call alarm. The main PCB also includes the RS-232, parallel printer, analog output, and remote alarm connectors.
The main PCB also routes these signals: reset, MMI PCB reset, sensor PCB reset, primary alarm, primary alarm failure detection logic, backup alarm, remote alarm, printer, POST timer, clocked serial interface (CSI) signals, and the battery backed +3.6 V.
3-12 Respironics V200/Esprit Ventilator Service Manual © Respironics, Inc. REF 580-1000-02 H
Loading...
+ 378 hidden pages