Philips Optimus RAD User manual

X--RAY GENERATION
LIST OF DOCUMENTATION IN THIS BINDER:

SUBSYSTEM MANUAL OPTIMUS RAD

UNIT MANUAL Surge Arrester WN
UNIT MANUAL Extension set for an additional tube assembly WG / GWB
UNIT MANUAL 26 V DC / 230 V AC Adapter
UNIT MANUAL Handswitch for OPTIMUS
UNIT MANUAL Extension of Photo Pick-Up OPTIMUS (SEV)
UNIT MANUAL Mains group EWD
Note: indicated document present
LIST OF ALL BINDERS FOR X--RAY GENERA
-- SUBSYSTEM MANUAL OPTIMUS RAD (this binder)
TION:
4512 984 28171 REV AB
OPTIMUS_RAD_28171AB_01
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
0.1
Philips Medical Systems
OPTIMUS RAD
W
SERVICE MANUAL
742
SUBSYSTEM
INTRODUCTION AND TECHNICAL DATA
INSTALLATION
1
2
9890 000 0218.
FAULT FINDING
REPLACEMENT
PROGRAMMINGS
ADJUSTMENTS
3
4
5
6
358J92
CAN-controlled X-ray generator of the converter type
DMC Hamburg
Printed in Hamburg, Germany
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
ACCEPTANCE
SERVICE INFORMATION
PARTS LIST
SCHEMATIC DRAWINGS
7
8
P
Z1-
4512 984 28171 REV AB 1
OPTIMUS_RAD_28171AB_r eg
WIRING DIAGRAMS
Z2-
OPTIMUS RAD
SERVICE MANUAL -- SUBSYSTEM
OPTIMUS RAD Author: T. Frenscheck
Type No: 9890 000 0218x
Techn. No: Basis 4512 104 70625
Release: 3.6
In case there are any questions concerning this manual, please send this LOPAD via fax to 49/(0)40/5078 2481
File: OPTIMUS_RAD_28171AB
List of pages and drawings (LOPAD) Manual Order No: 4512 984 28171
released: 10/2004
0.1
1
3.1
3.2
_________________________________________
_________________________________________
1--0.1 ... 0.2 (e/04.0)
1--1 ... 15 (e/04.0)
Z--1.1 (01.0) A4 4512 982 0092. Z--1.2 (01.0) A4 4512 982 0092. Z--1.3 (01.0) A4 4512 982 0099. Z--6.1 (01.0) A4 4512 982 0099. Z--7.1 (01.0) A3 4512 982 0092. Z--7.2 (01.0) A3 4512 982 0092. Z--7.3 (01.0) A3 4512 982 0092. Z--7.4 (04.0) A3 4512 982 0092. Z--7.10 (01.0) A3 4512 982 0010. _________________________________________
2--0.1 ... 0.2 (d/04.1)
2--1 ... 66 (d/04.1)
2Z--2.0 3x (01.0) A4 4512 983 05591 2Z--2.2 (a/01.0) A4 4512 983 0561 1 2Z--2.4 (a/01.0) A4 4512 983 05631 2Z--2.5 (a/01.0) A4 4512 983 05641 2Z--2.6 (01.0) A4 4512 983 05651 2Z--2.8 (01.0) A4 4512 983 05671 2Z--2.9 (01.0) A4 4512 983 05681 2Z--2.10 (01.0) A4 4512 983 05691 2Z--3 (97.0) A3 4512 983 05721 2Z--4 (97.0) A4 4512 983 05771 2Z--5 (97.0) A4 4512 983 05731 2Z--10 (a/02.0) A3 4512 983 05831 _________________________________________
3--0.1 (c/04.1)
3--1 ... 88 (c/04.1)
3Z--1 (a/03.0) A4 OPTIMUS R/F 3Z--21 (97.1) A4 OPTIMUS R/F _________________________________________
4--0.1 (c/04.1)
4--1 ... 27 (c/04.1) _________________________________________
5Z--1 (b/04.0) A3 OPTIMUS R/F 5Z--2 (c/04.0) A3 OPTIMUS R/F _________________________________________
6--0.1 (c/04.0)
6--1 ... 14 (c/04.0) _________________________________________
7--0.1 (b/04.0)
7--1 ... 4 (b/04.0) _________________________________________
8--1 (00.0) FCO-- Checklist 8--2 (00.0) FCO-- Checklist _________________________________________
4512 984 28171 REV AB
OPTIMUS_RAD_28171AB_l opad
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
3.1
Z0--1 (02.0)
Z1--1.1 (04.0) A3 OPTIMUS R/F Z1--1.2 (04.0) A3 4512 983 05751 Z1--2.1 (04.0) A3 4512 983 05761 Z1--2.1.1 (04.0) A3 4512 983 05951 Z1--2.2 (04.0) A3 4512 983 05941 Z1--2.2.1 (04.0) A3 4512 983 05961 Z1--2.3 (04.0) A3 OPTIMUS R/F Z1--3.3 (a/04.0) A3 OPTIMUS R/F Z1--4.1 (a/04.0) A3 OPTIMUS R/F Z1--4.2 (04.0) A3 Z1--5.1 (04.0) A3 4512 983 06551 Z1--6 (04.0) A3 4512 983 05531 Z1--11.1 (a/04.0) A3 4512 983 05521 Z1--11.2 (96.0) A4 OPTIMUS R/F Z1--12 (a/04.1) A3 Z1--13.2 (d/04.1) A3 OPTIMUS R/F Z1--14.1 (b/98.0) A3 4512 983 05541 Z1--14.2 (c/97.1) A3 4512 983 05551 Z1--15.1 (97.0) A3 4512 983 05571 _________________________________________ Z0--2 (02.0)
OPTIMUS RAD
Z2--1.0 (96.0) A4 OPTIMUS R/F Z2--1.1 (b/99.0) A3 Z2--1.2 (b/99.0) A3 Z2--1.3 (a/04.0) A3 Z2--2.1 (a/96.0) A3 Z2--2.2 (a/96.0) A3 Z2--2.3 (94.0) A3 Z2--5.1 (b/04.0) A3 OPTIMUS R/F Z2--5.2 (b/04.0) A3 4512 983 0551 1 Z2--5.3 (a/04.0) A3 OPTIMUS R/F Z2--5.4 (02.0) A3 OPTIMUS R/F Z2--12 (a/04.0) A3 Z2--13 (d/00.0) A3 OPTIMUS R/F Z2--14.1.1 (a/99.0) A3 Z2--14.1.2 (a/99.0) A3 Z2--14.2 (a/04.0) A3 OPTIMUS R/F Z2--14.3 (a/04.0) A3 OPTIMUS R/F Z2--15.1 (97.0) A3 4512 983 05741 Z2--16 (97.0) A3 OPTIMUS R/F Z2--17 (97.0) A3 OPTIMUS R/F _________________________________________
3.2
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
4512 984 28171 REV AB
OPTIMUS_RAD_28171AB_l opad
OPTIMUS RAD
1. Product information 1--1..................................................
1.1. Applications 1--1..........................................................
1.2. Options 1--1..............................................................
1.2.1. Hardware options 1--1......................................................
1.2.2. Software options 1--2......................................................
2. Compatibility 1--3........................................................
2.1. Generator components 1--3.................................................
2.2. Tubes 1--3................................................................
2.3. Five-field bucky chamber 1--3...............................................

INTRODUCTION AND TECHNICAL DATA

INTRODUCTION AND TECHNICAL DATA
Contents
TEXT
Contents 1--0.1............................................................
3. Mechanical data 1--4......................................................
4. Environmental data 1--5...................................................
4.1. Electrical environment 1--5..................................................
4.2. Climatic conditions 1--5.....................................................
4.3. Emission 1--5.............................................................
5. Electrical data 1--6........................................................
5.1. Power data and mains conditions 1--6........................................
5.2. Power supply for applications 1--7...........................................
5.3. Operating data 1--7........................................................
5.4. Power supply 1--8.........................................................
5.4.1. Type of power supply 1--8..................................................
5.4.2. Calculating the mains resistances 1--9........................................
5.4.3. Earth-leakage circuit-breaker 1--10............................................
5.4.4. Emergency--OFF device 1--10................................................
6. Tools 1--11................................................................
7. Traceable items 1--12......................................................
8. Preparation 1--13..........................................................
8.1. Installation material 1--13....................................................
8.2. Cables 1--13...............................................................
8.3. Manpower 1--14............................................................
9. Planned maintenance 1--15.................................................
OPTIMUS RAD (e/04.0)
OPTIMUS_RAD_1_e040_inh
2004 Philips Medical Systems
ALL RIGHTS RESERVED
1--0.1
OPTIMUS RADINTRODUCTION AND TECHNICAL DATA
DRAWINGS
Generator cabinet Z--1.1......................................................
Room layout Z--1.2...........................................................
Operating panel Z--1.3........................................................
Connection of generator Z--6.1................................................
Connection diagram 1Z--7.1...................................................
Connection diagram 1Z--7.2...................................................
Connection diagram 1Z--7.3...................................................
Earthing diagram 1Z--7.4......................................................
Legend for earthing and cabling diagram Z--20.1..................................
2004 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS RAD(e/04.0)1--0.2
OPTIMUS_RAD_1_e040_inh
OPTIMUS RAD
INTRODUCTION AND TECHNICAL DATA

1. Product information

The Optimus family of generators for radiography is based on computer-controlled converter technology. The converter operates in the non-audible frequency range.
Application options are essentially achieved by releasing software modules using customized PAL ICs. Control between the internal Function Units (FUs) and the external online equipment takes place by a CAN bus.
Safety-relevant signals are transferred directly on the so-called “Signal bus”. Units without any CAN interface are operated by the ”Adapter for 4 auxiliary units WA” option.
1.1. Applications
-- Radiography
-- Tomography
1.2. Options
Component overview according to the commercial catalogue.
Only the versions in the current commercial catalogue can be ordered.
If an existing generator is to be upgraded the commercial department must order:
MGR0011 (upgrade of an existing configuration) + MGRxxxx +S/N
1.2.1. Hardware options + MGR0011 + S/N
-- Low-speed rotor control 9890 000 0220x...............................
-- Dual-speed rotor control 9890 000 0268x MGR2082.............................. ............
-- Mains transformer 400 -- 480V; 50 / 60Hz,
also for 400V mains supply without neutral lead N
with taps for 400 / 440 / 460 / 480V 9890 000 0230x.....................
-- Mains transformer 190--390V; 50 / 60Hz with taps for 190 / 200 / 207 / 220 / 230 / 240 / 250 /
343 / 380 / 390V max. 50kW! 9803 720 8100x.......................
-- Adapter for 4 aux. units WA 9890 000 0231x MGR2131........................... ............
-- Option rack 9890 000 0232x..........................................
-- Extension set for one additional tube 9890 000 0234x...................
-- Tube extension WG 9890 000 0238x..................................
-- Operating panel 9890 000 0240x.....................................
-- Operating module Optimus 9890 000 0278x............................
-- Operating desk data cable 10m, 20m, 30m 9890 000 0241x / 2x / 3x..............
-- Stand for operating panel 9890 000 0244x MGR1482............................. ............
OPTIMUS RAD (e/04.0)
OPTIMUS_RAD_1_e040_BW
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
1--1
OPTIMUS RADINTRODUCTION AND TECHNICAL DATA
-- Wall mounting of operating panel 9890 000 0245x.......................
-- 26VDC / 230VAC adapter 9890 000 0246x MGR2281............................. ............
-- Surge arrester WN 9890 000 0247x...................................
-- Handswitch for Optimus 9890 000 0249x...............................
-- Patient Data Organizer (PDO) 9890 000 0255x MGR2091......................... ............
-- Decade cable set 14 x 4m top decade AMP decade 9803 704 2010x....
-- APR extension 9890 000 0257x.......................................
-- Extension photo pick-up Optimus 9890 000 0258x.......................
1.2.2. Software options + MGR0011 + S/N
Software options are provided by the function key (see also 5Z--1, EZ 139 Central Unit D38). Additional hardware components are not required.
-- Automatic Exposure Control (AEC) 9890 000 0281x
-- Anatomically Programmed Radiography / Fluorography (APR/F) 9890 000 0282x
1)
1)
-- Automatic Tomo Time Input (TTI) 9890 000 0222x MGR2121............................ .......
-- Tomo Density Control (TDC) 9890 000 0223x MGR2122................................. .......
-- VARIOFOCUS 9890 000 0227x MGR2101............................................. .......
-- Area dose calculator 9890 000 0256x MGR2141........................................ .......
1)
= Options only for base 9890 000 0218x
Options are always included in base 9890 000 0216x
MGR2171.......................... .....
MGR2181. .....
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS RAD(e/04.0)1--2
OPTIMUS_RAD_1_e040_BW
OPTIMUS RAD
INTRODUCTION AND TECHNICAL DATA

2. Compatibility

2.1. Generator components
-- Base Optimus 9890 000 0218x...............................
-- H.V. transformer 1 tube, 50kW 9890 000 0203x...... .........
-- H.V. transformer 2 tubes, 50kW 9890 000 0204x MGR2051 (Upgrade 1 ----> 2) tubes...... ........ ....
-- H.V. transformer 1 tube, 65 / 80kW 9890 000 0205x...... ......
-- H.V. transformer 2 tubes, 65 / 80kW 9890 000 0206x MGR2052 (Upgrade 1 ----> 2) tubes...... ..... ....
-- 50kW extension -- RAD 9890 000 0262x.......................
-- 50kW extension -- RAD 480V 9890 000 0208x..................
-- 65 / 80kW extension -- RAD 9890 000 0264x...................
-- 65 / 80kW extension -- RAD 480V 9890 000 0209x..............
-- Firmware Rel. 3.6 9890 000 0251x............................
2.2. Tubes
Recommended standard tubes:
-- RO 17 50
-- SRO 25 50
-- SRO 33 100
Further compatible tubes:
-- RO 30 -- SRO 09 51 -- SRO 20 55
-- RO 12 30 -- SRO 13 30 -- SRO 22 50
-- RO 16 48 -- SRO 20 50 -- SRO 32 100
-- RO 30 50
Compatible tube housings:
-- ROT 350
-- ROT 351
The latest information on further tubes which are connectable is available at the service center Hamburg.
NOTE
When the generator is retrofitted it is important to use the screened cable 3 x 1.31mm
2
(0722 215 02054) as
the stator cable.
If necessary, exchange the old stator cable.
___________________
2.3. Five-field bucky chamber
-- Five-field bucky chamber 9890 000 7000x
The five-field bucky chamber is compatible with rel. 3.6 or higher.
OPTIMUS RAD (e/04.0)
OPTIMUS_RAD_1_e040_BW
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
1--3

3. Mechanical data

C
For installation dimensions and weights see drawings Z--1.1.
Transport data:
Weights [kg] Dimensions [cm]
aseNo.Contents
-- Generator cabinet
1
-- Operating panel
-- Cables
net gross length width height
178 226 210 82 84
OPTIMUS RADINTRODUCTION AND TECHNICAL DATA
1--tube version:
2
Contents: H.V. generator
2--tube version:
73
88
100
77 67 80
115

4. Environmental data

The environmental data comply with PMS standard UXW 13600.
4.1. Electrical environment
Class S0 -- Dedicated mains supply, 3 phases and neutral. Thus single phase voltage is also available.
A low impedance, permanently installed connection, fed in by the step-down transformer of the hospital to supply large systems like in MR, CT and X--ray departments is required.
NOTE
Use always a mains cable with 4 wires and concentric PE--shield, type NYCY .
___________________
4.2. Climatic conditions
Ambient temperature 10_C--40_C.............
Relative humidity 15% -- 90%; no condensation.................
Relative atmospheric pressure 70kPa -- 110kPa.....
4.3. Emission
Heat dissipation max. 500W; average per hour.....................
Noise level 46dBA.........................
EMC IEC 950...............................
To avoid any possible annoying noise of the implemented fans it is advisable to install the generator cabinet outside the examination room.
OPTIMUS RAD(e/04.0)1--4
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS_RAD_1_e040_BW
OPTIMUS RAD
(
)

5. Electrical data

5.1. Power data and mains conditions
50kW 65kW 80kW
Mains voltage 3 x 400V ±10% (415V
3 x 400 / 440 / 460V ±10% *
6%
3 x 480V
3 x 190 ... 343V ±10% **
* = with internal mains transformer (option)
** = with external mains transformer; max. 50kW (option)
+
|
--10%
Frequency
INTRODUCTION AND TECHNICAL DATA
Voltage
+6%
/ 380V
-- 5 %
)
*
50kW 65kW 80kW
Mains frequency 49 ... 61Hz
Max. mains current
Voltage
50kW 65kW 80kW
Exposure: 400V 145A 190A 230A
440V 135A 180A 215A
460V 125A 170A 210A
480V 120A 160A 205A
190V 300A -- --
Short--time
power consumption
100kVA 132kVA 160kVA
[I x U x 3]
Fuse protection
slow-blow
35A 50A
100A at 240V --
Connected load [I
xUx3]
Fuse
25kVA 35kVA
Emergency power supply:
static
(inverter)
dynamic
(diesel generator
with flywheel mass)
OPTIMUS RAD (e/04.0)
OPTIMUS_RAD_1_e040_BW
Short--time power consumption
[I x U x 3]
Connected load
[I
xUx3]
Fuse
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
1--5
Mains resistance
OPTIMUS RADINTRODUCTION AND TECHNICAL DATA
Voltage
50kW 65kW 80kW
400V 300m 200m
440V 350m 240m
460V 350m 240m
480V 400m 300m
480V valid for DOD only 300m 240m 180m
NOTE
500m
is the absolute max. mains resistance.
5.2. Power supply for applications
Generator power
50kW 65kW 80kW
Max. output power 230V / 400V; max. 5A
5.3. Operating data
Generator power
Data
Tube current 1 ... 650mA 1 ... 900mA 1 ... 1100mA
Tube voltage 40 ... 150kV in kV-- or %--steps
mAs product 0.5 ... 850mAs
Exposure time 1ms ... 6s / 16s
Exposure frequency 12exp./s
Interfacing option for door contact, external radiation warning indicator
50kW 65kW 80kW
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS RAD(e/04.0)1--6
OPTIMUS_RAD_1_e040_BW
OPTIMUS RAD
5.4. Power supply
5.4.1. Type of power supply
3-phase WYE
INTRODUCTION AND TECHNICAL DATA
L1
L2
L3 (N)
3-phase DELTA, balanced earth or floating
L1
L2
L3
X-ray installation
X-ray installation
-- 400V
-- 440V / 460V / 480V with mains transformer 9890 000 0230x.
-- Surge arrester WN is required if the mains transformer 9890 000 0230x is ordered.
-- Neutral not required if the mains transformer 9890 000 0230x is ordered.
-- 190V ... 343V with external mains transformer 9803 720 8100x (max. 50kW).
-- Mains transformer 9890 000 0230x is required.
-- 400V / 440V / 460V / 480V
-- Surge arrester WN is required.
-- 190V ... 343V with external mains transformer 9803 720 8100x (max. 50kW). Works only together with the internal mains transformer.
3-phase DELTA, grounded
L1
-- Mains transformer 9890 000 0230x is required.
-- 400V / 440V / 460V / 480V
X-ray installation
L2
L3
CAUTION
Ensure the sequence of phases in the wall junction box corresponds to designations L1, L2, L3.
___________________
-- Surge arrester WN is required (requires modification at the EMC--filter of the kV power unit).
-- 190V ... 343V with external mains transformer 9803 720 8100x (max. 50kW). Works only together with the internal mains transformer.
OPTIMUS RAD (e/04.0)
OPTIMUS_RAD_1_e040_BW
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
1--7
5.4.2. Calculating the mains resistances
OPTIMUS RADINTRODUCTION AND TECHNICAL DATA
NOTE
The cross section of lead l
must not exceed 25mm2(see figure below).
3
___________________
If possible the sum of R smaller than the R
requires.
XG
0,R1,R2
and R3should be
R
+R1+R2+R3<R
0
With higher internal mains resistances the generator output is reduced correspondingly.
R
designates the mains resistance of the distributor transformer
0
R
depends on the length of lead l1between distributor transformer and mains distributor
1
L1 L2 L3
l
1
and on the selected cross section as well:
==> R
R
consists of upstream elements such as:
2
1=l1
× R
Cu
RCufrom table below
-- Emergency-OFF switch 4.0m.................
-- Earth-leakage circuit -breaker 5.5m............
l
3
XG
MEX 100
L1 L2 L3
357J94
-- Fuse 5.5m.................................
-- Surge arrester WN 23.0m...................
R
depends on the length of lead l3between mains distributor and wall junction box
3
and on the selected cross section as well:
==> R
3=l3
× R
Cu
RCufrom table below
The resistances consider the go and return lines so that the calculation can be based on simple cable lengths.
Copper cross section
[mm
2
]
Resistance R
[m/m]
Cu
16 2.19 25 1.4 35 1.0 50 0.7 70 0.5
95 0.38 120 0.30 150 0.24
NOTE
500m
is the absolute max. mains resistance.
___________________
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS RAD(e/04.0)1--8
OPTIMUS_RAD_1_e040_BW
OPTIMUS RAD
INTRODUCTION AND TECHNICAL DATA
5.4.3. Earth-leakage circuit-breaker
To be provided between mains fuse and X-ray installation depending on local regulations.
Siemens earth-leakage circuit--breaker N 5SZ type B:
-- Order No.: 5SZ3 466 OKG00
-- Rated fault current 30mA
-- Rated current 63A
-- Connection terminals for wire cross sections of up to 25mm
2
5.4.4. Emergency--OFF device
To be provided depending on local regulations.
There are 2 possibilities:
1. All the Emergency--OFF buttons are connected in series and looped into the switch--ON circuit (12VDC) of the generator.
2. The Emergency--OFF circuit acts on an external mains contactor which switches OFF the power before it is fed into the generator.
OPTIMUS RAD (e/04.0)
OPTIMUS_RAD_1_e040_BW
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
1--9
OPTIMUS RADINTRODUCTION AND TECHNICAL DATA

6. Tools

-- Service engineer standard tool kit
-- Service-PC:
Zeppelin standard, Win 2000 compatible.
-- Installation and service software AGent.
-- Security device, parallel port key or smart card, PMS security.
Necessary to carry out the installation and to run the service software (special programming, fault finding).
-- 0-modem cable:
The minimum length is the distance between generator cabinet and operating desk. Male 25-pole D-Sub connector at the generator side. A 5m data cable of the bucky controller can be used: 4512 130 5693x
-- Mains resistance measuring instrument
-- Dose measuring instrument
-- mAs--meter
-- Multimeter
-- Digital oscilloscope with 2-beam memory
-- Recommended PLCC extraction tool (AMP 822154--1): 2422 487 89772

7. Traceable items

The following items have serial numbers of the following format when delivered ex factory:
1. Generator cabinet 6-digit serial number........
2. H.V. tank 7-digit serial number................
3. Operating desk 8-digit serial number..........
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS RAD(e/04.0)1--10
OPTIMUS_RAD_1_e040_BW
OPTIMUS RAD
INTRODUCTION AND TECHNICAL DATA

8. Preparation

Connection of the generator: see drawing Z--6.1................
Operating panel: see drawing Z--1.3..........................
Connection diagram: see drawing Z--7.1/.2/.3/.4.......................
Earthing diagram: see drawing Z--7.5.........................
Legend for earthing and cabling: see drawing Z--7.10.............
8.1. Installation material
To be ordered from the service department of PMS Hamburg:
-- Wall junction box 4512 103 7538x.............................
2
including connection block (25mm
-- Relay for radiation warning indicator 4512 100 4523x............
One interface relay with a floating contact (230V/1A) is included in the scope of delivery of the generator.
) for mains supply and connection block (10mm2) for unit supply.
8.2. Cables
H.V . cables
with O3 / O3 plugs: 9806 402 6xx02................................
length: 6m -- 30m in steps of 2m...........................................
capacity: 155pF/m.........................................
diameter: 16.5mm........................................
The cable length is indicated by the 9th and 10th digit of the numeric code.
Thermal contact cable
-- 3-wire screened for 1 excess temperature switch 4512 100 66162.............
(3 x 0.5mm2, 5.3mm)
-- 10-wire screened for additional supervision like
temperature alarm switch, buzzer, selection indicator 0722 215 19005.........
Stator cable
3×1.31mm
2
, screened 0722 215 02054.....................................
NOTE
The above described cables are part of the pre--installed systems.
___________________
OPTIMUS RAD (e/04.0)
OPTIMUS_RAD_1_e040_BW
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
1--11
OPTIMUS RADINTRODUCTION AND TECHNICAL DATA
AMPLIMAT cable
with D-Sub and 3--Plus plug:
12m 9890 000 01721......................
16m 9890 000 01731......................
20m 9890 000 01741......................
24m 9890 000 01751......................
CAUTION
AMPLIMAT cables 9803 507 0xx02 (for hybrid measuring chambers 9803 509 xxxxx) with 3- Plus plugs
at both ends must be connected in the generator by the following adapter for each cable:
___________________
Adapter for AMPLIMAT cable: 4512 108 09042. The generator includes 1 adapter.
The hybrid measuring chambers 9803 509 xxxxx require connection (chassis) between contacts:
D--Sub end GND (13) <-- ----> RF 0V (8) (generator input)
or
3--Plus end GND (N) <-- ----> RF 0V (J) (generator input)
This connection is established by the adapter for the AMPLIMAT cable. See drawing Z1--6 ”Basic interface”.
In case a hybrid measuring chamber 9803 509 xxxxx is not operated with the required
AMPLIMAT cable 3--Plus / 3--Plus 9803 507 0xx02... .....
but with
AMPLIMAT cable D--Sub / 3--Plus 9890 000 017xx... .....
make sure to establish this connection (13 <-- -- --> 8) in the D--Sub connector!
For ALC measuring chambers 9890 000 016xx connection GND <------> RF 0V is not permitted. Therefore, ALC measuring chambers AMPLIMAT cables 9890 000 017xx should always be used.
Operating desk
NOTE
Use the shortest cables. Noise immunity increases.
___________________
Cable set 10m 9890 000 02411.......................
20m 9890 000 02421.......................
30m 9890 000 02431.......................
8.3. Manpower
At least two persons are necessary to insert the H.V . tank in the generator cabinet. The weight of the 2--tube version is about 88kg.
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS RAD(e/04.0)1--12
OPTIMUS_RAD_1_e040_BW
OPTIMUS RAD
INTRODUCTION AND TECHNICAL DATA

9. Planned maintenance

The technical documentation for carrying out maintenance work in compliance with the applicable regulations are available at the responsible authority of Philips Medical Systems.
The importance of having maintenance implemented is pointed out to the operator in the operating instructions.
It must be guaranteed that the person carrying out maintenance work knows about the respective national regulations and that this person observes these regulations throughout all steps of maintenance work.
OPTIMUS RAD (e/04.0)
OPTIMUS_RAD_1_e040_BW
E 2004 Philips Medical Systems
ALL RIGHTS RESERVED
1--13
OPTIMUS RAD
1. Installing the wall junction box 2--1........................................
1.1. Fixing of the wall junction plates to the walll junction plates 2--2..................
1.2. Use of the cable support and the fixing rail of the wall junction box 2--3...........
2. Preparatory work 2--4.....................................................
2.1. Mounting of the H.V. generator in the cabinet 2--4..............................
2.2. Electrical connection of the H.V. generator 2--5................................
3. Installing the operating panel 2--7.........................................
3.1. Desk version 2--7..........................................................
3.2. Stand version 2--9.........................................................
3.3. Wall mounted version 2--10..................................................
3.4. Supporting angle version 2--11...............................................
3.5. Additional release switch 2--11...............................................
4. Electrical connection 2--12.................................................
4.1. Earthing 2--12..............................................................
4.2. Mains connection 2--12......................................................
4.2.1. Mains connection of the generator 2--12.......................................
4.3. Stator connection 2--13......................................................
4.3.1. Shielding 2--13.............................................................
4.3.2. Connection 2--14...........................................................
4.4. Signal cables 2--17.........................................................
4.4.1. Room decade cable 2--17....................................................
4.4.2. Tube supervision 2--18......................................................
4.4.3. CAN interface 2--19.........................................................
4.4.4. Adapter for four auxiliary units 2--20...........................................
4.4.5. Dose inputs 2--21...........................................................
4.4.6. Patient data organizer PDO (option) 2--21......................................
4.5. H.V . cables generator side 2--22..............................................
4.6. Emergency--OFF circuit 2--22................................................
5. Hardware programming 2--23...............................................
5.1. Mains transformer (option) 2--23..............................................
5.2. PCB EZ150 basic interface 2--24.............................................
6. Switch-ON of the generator 2--25...........................................
7. Installation software AGenT 2--26...........................................
7.1. PC and generator settings to avoid problems during up/downloading
7.1.1. Preparation of the service PC to guarantee a safe loading process 2--26...........
7.1.2. Preparation of the generator 2--27............................................
7.2. Interface 2--28.............................................................
8. Setting-to-work overview 2--30.............................................
8.1. Configuration 2--31.........................................................
8.1.1. Date and time 2--31.........................................................
8.1.2. Mains data 2--31...........................................................
8.1.3. Tubes 2--32................................................................
8.1.3.1. Tube data set 2--33.........................................................
8.1.3.2. Tube speed selection 2--34..................................................

INSTALLATION

INSTALLATION
Contents
TEXT
Contents 2--0.1............................................................
of CU complete files 2--26...................................................
OPTIMUS RAD (d/04.1)
OPTIMUS_RAD_2_d041_inh
2004 Philips Medical Systems
ALL RIGHTS RESERVED
2--0.1
Loading...
+ 262 hidden pages