Philips Optimus C User manual

X-RAY GENERATION
LIST OF DOCUMENTATION IN THIS BINDER:
SUBSYSTEM MANUAL OPTIMUS
UNIT MANUAL Cockpit for duo DIAGNOST
UNIT MANUAL Power Distribution Unit (PDU)
Note: indicates document present
LIST OF ALL BINDERS FOR X-RAY GENERA
TION:
-- SUBSYSTEM MANUAL OPTIMUS C (this binder)
4512 984 08497 REV AB
OPTIMUS_C_div 01_08497_AB
E 2003 Philips Medical Systems
ALL RIGHTS RESERVED
0.1
SERVICE MANUAL

OPTIMUS C

742
SUBSYSTEM
INTRODUCTION AND TECHNICAL DA TA
INSTALLATION
1
2
9890 000 02191
FAULT FINDING
REPLACEMENT
PROGRAMMINGS
ADJUSTMENTS
3
4
5
6
358J92
CAN-controlled X-ray generator of the converter type
DMC Hamburg
Printed in Hamburg, Germany
E 2003 Philips Medical Systems
ALL RIGHTS RESERVED
ACCEPTANCE
SERVICE INFORMATION
PARTS LIST
SCHEMATIC DRAWINGS
7
8
P
Z1-
4512 984 08497 REV AB 1
OPTIMUS_C_reg_08497_AB
WIRING DIAGRAMS
Z2-
SERVICE MANUAL -- SUBSYSTEM
OPTIMUS C
OPTIMUS C
Author: P. Thron
Type No: 9890 000 02191
Techn. No: Basis 4512 104 70625
Release: 1.2
In case there are any questions concerning this manual, please send this LOPAD via fax to 49/(0)40/5078 2481
File: OPTIMUS_C_08497_AB
List of pages and drawings (LOPAD) Manual Order No: 4512 984 08497
released: 04/2003
0.1
1
3.1
3.2
_________________________________________
1--0.1 (a/02.1)
1--1 ... 11 (a/02.1) Z--1.1 (01.0) A4 4512 982 0092.
Z--1.2 (02.0) A4 4512 982 0099. Z--1.3 (01.0) A4 4512 982 0099. Z--1.4 (01.0) A4 4512 982 0099. Z--6.1 (01.0) A4 4512 982 0099. Z--6.2 (01.0) A4 4512 982 0099. Z--7.1 (02.0) A3 4512 982 0099.
_________________________________________
6--0.1 (a/02.1)
6--1 ... 16 (a/02.1) _________________________________________
7--0.1 (a/02.0)
7--1 ... 4 (a/02.0) _________________________________________
8--1 (00.0) FCO--Checklist 8--2 (00.0) FCO--Checklist _________________________________________
Z--7.2 (02.0) A3 4512 982 0099. Z--7.3 (02.0) A3 4512 982 0099. Z--7.10 (01.0) A3 4512 982 0010. _________________________________________
2--0.1 ... 0.2 (a/02.1)
2--1 ... 36 (a/02.1)
2Z--4 (02.0) A4 4512 983 06161 2Z--10 (a/02.0) A3 4512 983 06171 _________________________________________
3--0.1 (a/02.1)
3--1 ... 85 (a/02.1)
3Z--1 (a/01.0) A4 OPTIMUS R/F _________________________________________
4--0.1 (a/02.1)
4--1 ... 22 (a/02.1) _________________________________________
5Z--1 (a/01.0) A3 OPTIMUS R/F 5Z--2 (a/01.0) A3 OPTIMUS R/F _________________________________________
E 2003 Philips Medical Systems
ALL RIGHTS RESERVED
4512 984 08497 REV AB3.1
OPTIMUS_C_lopad_08497_AB
OPTIMUS C
_________________________________________ Z0--1 (02.0)
Z1--1.1 (02.0) A3 Z1--2.1 (a/02.0) A3 4512 983 06181 Z1--2.2 (a/02.0) A3 4512 983 06191 Z1--2.3 (99.0) A3 OPTIMUS R/F Z1--3.3 (a/97.2) A3 OPTIMUS R/F Z1--4.1 (a/99.0) A3 OPTIMUS R/F Z1--4.2 (a/99.0) A3 OPTIMUS R/F Z1--5.1 (97.1) A3 4512 983 06551 Z1--6 (02.0) A3 4512 983 05531 Z1--13.2 (c/00.0) A3 OPTIMUS R/F _________________________________________
Z0--2 (02.0)
Z2--1.0 (a/00.0) A4 Z2--1.1 (d/99.0) A3 4512 983 06461 Z2--1.2 (e/99.0) A3 4512 983 06471 Z2--1.3 (c/98.0) A3 4512 983 06481 Z2--2.1.1 (97.0) A3 4512 983 06491 Z2--2.1.2 (97.0) A3 4512 983 06501 Z2--2.2.1 (96.0) A3 4512 983 06511 Z2--2.2.2 (a/97.0) A3 4512 983 06521 Z2--5.1 (a/02.0) A3 OPTIMUS R/F Z2--5.2 (b/02.0) A3 4512 983 05511 Z2--5.3 (a/02.0) A3 OPTIMUS R/F Z2--5.4 (02.0) A3 OPTIMUS R/F Z2--13 (d/00.0) A3 OPTIMUS R/F Z2--14.1 (b/99.0) A3 OPTIMUS R/F Z2--14.4 (97.0) A3 OPTIMUS R/F Z2--17 (98.0) A3 _________________________________________
(nur für Fabrik--Auslieferung)
Service software No:
-- OMC : 4512 116 0240x _________________________________________
4512 984 08497 REV AB
OPTIMUS_C_lopad_08497_AB
E 2003 Philips Medical Systems
ALL RIGHTS RESERVED
3.2
OPTIMUS C
1. Product information 1--1..................................................
1.1. Applications 1--1..........................................................
1.2. Options 1--1..............................................................
1.2.1. Hardware options 1--1......................................................
1.2.2. Software options 1--1......................................................
2. Compatibility 1--1........................................................
2.1. Generator components 1--1.................................................
2.2. Tubes 1--2................................................................
2.3. System components 1--2...................................................
3. Mechanical data 1--2......................................................
4. Environmental data 1--3...................................................
4.1. Electrical environment 1--3..................................................
4.2. Climatic conditions 1--3.....................................................
4.3. Emission 1--3.............................................................
5. Electrical data 1--4........................................................
5.1. Power data and mains conditions 1--4........................................
5.2. Power supply for application 1--6............................................
5.3. Operating data 1--6........................................................
5.4. Power supply 1--7.........................................................
5.4.1. Type of power supply 1--7..................................................
5.4.2. Calculating the mains resistances 1--8........................................
5.4.3. Earth-leakage circuit breaker 1--9............................................
5.4.4. Emergency--OFF device 1--9................................................
6. Tools 1--9................................................................
7. Traceable items 1--9......................................................
8. Preparation 1--10..........................................................
8.1. Installation material 1--10....................................................
8.2. Cables 1--10...............................................................
8.3. Manpower 1--1 1............................................................
9. Planned maintenance 1--11.................................................

INTRODUCTION AND TECHNICAL DATA

INTRODUCTION AND TECHNICAL DATA
TEXT
Contents 1--0.1............................................................
DRAWINGS
Generator cabinet: Mechanical dimensions Z--1.1................................
Overlayer for room layout Z--1.2..............................................
Operating panel: Mechanical dimensions Z--1.3.................................
Power Distribution Unit: Dimensions and weight Z--1.4...........................
Connection of generator Z--6.1...............................................
Connection of generator with PDU Z--6.2......................................
Connection diagram Z--7.1...................................................
Connection diagram Z--7.2...................................................
Earthing diagram Z--7.3.....................................................
Legend for earthing and cabling diagram Z--7.10.................................
OPTIMUS C (a/02.1)
OPTIMUS_C_1_a021_inh
E 2003 Philips Medical Systems
ALL RIGHTS RESERVED
1--0.1
OPTIMUS C
INTRODUCTION AND TECHNICAL DATA

1. Product information

The Optimus family of generators for radiography is based on computer-controlled converter technology. The converter operates in the non-audible frequency range.
Applicable options can be enabled by releasing software modules using customized PAL ICs depending on the respective order. Control between the internal Function Units (FUs) and the external online equipment takes place by a CAN bus.
Safety-relevant signals are transferred directly on the so-called “Signal bus”.

1.1. Applications

-- Radiography
-- Tomography
-- Fluoroscopy

1.2. Options

1.2.1. Hardware options
-- Mains transformer PDU: 400 -- 480V; 50 / 60Hz, also for 400V mains supply without neutral lead N
with taps for 400 / 440 / 460 / 480V 9890 000 0260x......................................
-- Mains transformer: 190--390V; 50 / 60Hz with taps for 190 / 200 / 207 / 220 / 230 / 240 / 250 / 343 / 380 / 390V
max. 50kW! 9803 720 8100x..........................................................
1.2.2. Software options
Software options are provided via the function key (see also 5Z--1, EZ 139 Central Unit D38). Additional hardware components are not required.
All system options are available. All Optimus C have the same function key configuration, there is only a difference in the power class 50/65/80kW.

2. Compatibility

2.1. Generator components

-- Base OPTIMUS C 9890 000 0219x....................................
-- H.V. transformer R/F 1 tube, 50kW 9890 000 0270x.. .................
-- H.V. transformer R/F 1 tube, 65/80kW 9890 000 0272x.. ...............
-- 50kW extension -- R/F 9890 000 0274x................................
-- 65kW extension -- R/F 9890 000 0275x................................
-- 80kW extension -- R/F 9890 000 0276x................................
-- Firmware Rel. 1.2 9890 000 0254x....................................
OPTIMUS C (a/02.1)
OPTIMUS_C_1_a021
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
1--1
OPTIMUS CINTRODUCTION AND TECHNICAL DATA
C

2.2. Tubes

Recommended standard tubes:
-- RO 17 50
-- SRM 06 12
-- SRO 25 50
-- SRO 33 100
Further compatible tubes:
-- RO 12 30 -- SRO 09 51 -- SRO 20 55
-- RO 16 48 -- SRO 13 30 -- SRO 22 50
-- RO 30 50 RE -- SRO 20 50 -- SRO 32 100
Compatible tube housings:
-- ROT 350
-- ROT 351
-- ROT 504
The latest information on further tubes which are connectable is available at the service center Hamburg.

2.3. System components

-- DuoDiagnost GEOMETRY 9890 000 0290x............................
-- Stand for operating panel 9890 000 0244x.............................
-- Wall mounting of operating panel 9890 000 0245x.......................
-- User interface support 9890 000 0283x................................
-- Cable set for COCKPIT 9890 000 0279x...............................

3. Mechanical data

For installation dimensions and weights see drawings Z--1.1.
Transport data:
Weights [kg] Dimensions [cm]
ase no.Contents
-- Generator cabinet
1
-- Operating panel
-- Cables
net gross length width height
178 226 210 82 84
2 -- H.V. generator; 1--tube version 73 100 77 67 80
(a/02.1)1--2
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS C
OPTIMUS_C_1_a021
OPTIMUS C
INTRODUCTION AND TECHNICAL DATA

4. Environmental data

The environmental data comply with to PMS standard UXW 13600.

4.1. Electrical environment

Class S0 -- Dedicated mains supply, 3 phases and neutral. Thus single phase voltage is also available.
A low impedance, permanently installed connection, fed in by the step down transformer of the hospital to supply large systems like in MR, CT and X--ray departments is required.
Note
Use always a mains cable with 4 wires and concentric PE--shield, type NYCY.

4.2. Climatic conditions

Ambient temperature 10_C--40_C...................
Relative humidity 15% -- 90%; no condensation.......................
Relative atmospheric pressure 70kPa -- 110kPa...........

4.3. Emission

Heat dissipation max. 1200W; average per hour........................
Noise level, normal operation 46dBA............
Noise level, maximum power operation 55dBA....
EMC IEC 950..................................
To avoid any possible annoying noise of the implemented fans it is advisable to install the generator cabinet outside the examination room.
OPTIMUS C (a/02.1)
OPTIMUS_C_1_a021
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
1--3

5. Electrical data

withexternaltransformerPDU(option
)
I
3+1x1
0
2
(L1,L
2,L
3,PE)

5.1. Power data and mains conditions

Voltag e
50kW 65kW 80kW
Mains voltage 3 x 400V ±10% (415V
3 x 400 / 440 / 460V ±10% *
6%
3 x 480V
=*=
The following connection cables are recommended:
+
|
--10%
+6%
/ 380V
*
-- 5 %
OPTIMUS CINTRODUCTION AND TECHNICAL DATA
)
nput:
Output: 4 x 4mm
4+1x4mm 3 x 1,5mm
mm
2
Generator supply 3 x 400V
2
Device supply 3 x 220V (option)
2
Switch control and temperature supervision
3 x 190 ... 343V ±10% **
** = with external transformer; max. 50kW (option)
Frequency
50kW 65kW 80kW
Mains frequency 49 ... 61Hz
(a/02.1)1--4
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS C
OPTIMUS_C_1_a021
OPTIMUS C
(
)
A
INTRODUCTION AND TECHNICAL DATA
Max. mains current
Voltag e
50kW 65kW 80kW
Exposure: 400V 145A 190A 230A
440V 135A 180A 215A
460V 125A 170A 210A
480V 120A 160A 205A
190V 300A -- --
Short--time power consumption
100kVA 132kVA 160kVA
[I x U x 3]
Fluoro: 400V -- 8A
480V -- 7A
Fuse protection
slow blow
Connected load [I
xUx3]
Fuse
Emergency static power supply: (Inverter)
dynamic
(Diesel generator
with flywheel mass)
35A
100A at 240V
25kVA 35kVA
Short--time power consumption
[I x U x 3]
Connected load
[I
xUx3]
Fuse
50
Mains resistance
Voltag e
50kW 65kW 80kW
400V 300m 200m
440V 350m 240m
460V 350m 240m
480V 400m 300m
480V valid for DOD only 300m 240m 180m
Note
500m
is the absolute max. mains resistance.
OPTIMUS C (a/02.1)
OPTIMUS_C_1_a021
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
1--5

5.2. Power supply for application

Generator power
Supply 50kW 65kW 80kW
OPTIMUS CINTRODUCTION AND TECHNICAL DATA
Output 1 230V / 400V; generator switched and protected ----> I
Output 2 230V / 400V; generator switched and protected ----> I
Output 3 230V / 400V; generator not switched and protected ----> I
Sum value I
max(1+2+3)
10A
Only with external transformer PDU
Output 4 127V / 220V generator switched and protected ----> I
Output 5 double socket 127V / 15A to phase L1;
generator not switched and protected ----> I
Sum value I
max (4 +5)
16A

5.3. Operating data

Generator power
1
2
3
4
5
Data
50kW 65kW 80kW
Exposure: Tube current 1 ... 650mA 1 ... 900mA 1 ... 1100mA
Tube voltage 40 ... 150kV in kV-- or %--steps
mAs product 0,5 ... 850mAs
Exposure time 1ms ... 6s / 16s
Exposure frequency 12exp./s
Interfacing option for door contact, external radiation warning indicator
Fluoro: Tube current 0,25 ... 6mA
Tube voltage 40 ... 110kV
kV/mA curves 3
Setting time 1s
(a/02.1)1--6
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS C
OPTIMUS_C_1_a021
OPTIMUS C

5.4. Power supply

5.4.1. Type of power supply
3 phase WYE
INTRODUCTION AND TECHNICAL DATA
L1
L2
L3 (N)
3 phase DELTA, balanced earth or floating
L1
L2
L3
X-ray installation
X-ray installation
-- 400V
-- 440V / 460V / 480V with external mains transformer PDU 9890 000 0260x.
-- Neutral not required if the external mains transformer PDU 9890 000 0260x is ordered.
-- 190V ... 343V with external mains transformer 9803 720 8100x (max. 50kW).
-- External mains transformer PDU 9890 000 0260x is required.
-- 400V / 440V / 460V / 480V
-- 190V ... 343V with external mains transformer 9803 720 8100x (max. 50kW).
3 phase DELTA, grounded
L1
-- External mains transformer PDU 9890 000 0260x is required (requires modification at the EMC-- filter of the kV power unit).
X-ray installation
L2
L3
Caution! Ensure the sequence of phases in the wall junction box corresponds to designations L1, L2, L3.
OPTIMUS C (a/02.1)
OPTIMUS_C_1_a021
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
-- 400V / 440V / 460V / 480V
1--7
5.4.2. Calculating the mains resistances
Note
The cross section of lead l
must not exceed 25mm2. (See figure below).
3
OPTIMUS CINTRODUCTION AND TECHNICAL DATA
If possible the sum of R smaller than the R
requires.
XG
0,R1,R2
and R3should be
R
+R1+R2+R3<R
0
With higher internal mains resistances the generator output is reduced correspondingly.
R
designates the mains resistance of the distributor transformer
0
R
depends on the length of lead l1between distributor transformer and mains distributor
1
L1 L2 L3
l
1
and on the selected cross section as well:
==> R
R
consists of upstream elements such as:
2
1=l1
× R
Cu
RCufrom table below
-- Emergency-OFF switch 4.0m.................
-- Earth-leakage circuit breaker 5.5m............
l
3
XG
MEX 100
L1 L2 L3
357J94
-- Fuse 5.5m.................................
-- Surge arrester WN 23.0m...................
R
depends on the length of lead l3between mains distributor and wall junction box
3
and on the selected cross section as well:
==> R
3=l3
× R
Cu
RCufrom table below
The resistances consider the go and return lines so that the calculation can be based on simple cable lengths.
Copper cross section
[mm
2
]
Resistance R
[m/m]
Cu
16 2.19 25 1.4 35 1.0 50 0.7 70 0.5
95 0.38 120 0.30 150 0.24
Note
500m
is the absolute max. mains resistance.
(a/02.1)1--8
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS C
OPTIMUS_C_1_a021
OPTIMUS C
INTRODUCTION AND TECHNICAL DATA
5.4.3. Earth-leakage circuit breaker
To be provided between mains fuse and X-ray installation depending on local regulations.
Siemens earth-leakage circuit breaker:
-- Order No.: 5SZ3 466 OKG00
-- Rated fault current 30mA
-- Rated current 63A
-- Connection terminals for wire cross sections of up to 25mm
2
5.4.4. Emergency--OFF device
To be provided depending on local regulations.
There are 2 possibilities:
1. All the emergency--OFF buttons are connected in series and looped into the switch--ON circuit (12VDC) of the generator.
2. The emergency--OFF circuit acts on an external mains contactor which switches OFF the power before it is fed into the generator.

6. Tools

-- Service engineer standard tool kit
-- Service-PC:
IBM-compatible, 640kB RAM, 3.5” floppy disk drive, 1 serial port
-- Installation and service software OMC: 4512 116 024xx.
Supplied on a floppy disk within this generator service manual.
-- PC-hardkey (DIAGGEN):
Necessary to carry out the installation and to run the service software (special programmings, fault finding).
-- 0-modem cable:
Minimum length is distance between generator cabinet and operating desk. Male 25-pole D-Sub connector at the generator side. A 5m data cable of bucky controller can be used: 4512 130 5693x
-- Mains resistance measuring instrument
-- Dose measuring instrument
-- mAs--meter
-- Multimeter
-- Digital oscilloscope with 2-beam memory
-- Recommended PLCC extraction tool (AMP 822154--1): 2422 487 89772

7. Traceable items

The following items have serial numbers of the following format when delivered ex factory:
1. Generator cabinet 6 digit serial number........
2. H.V. tank 7 digit serial number................
OPTIMUS C (a/02.1)
OPTIMUS_C_1_a021
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
1--9

8. Preparation

Connection of the generator: see drawing Z--6.1................
Connection of the generator with PDU: see drawing Z--6.2.......
Operating panel: see drawing Z--1.3..........................
Connection diagram: see drawing Z--7.1/.2.......................
Earthing diagram: see drawing Z--7.3.........................
Legend for earthing and cabeling: see drawing Z--7.10...........

8.1. Installation material

To be ordered from the service department of PMS Hamburg:
-- Wall junction box 4512 103 7538x.............................
2
inclusive connection block (25mm
-- Relay for radiation warning indicator 4512 100 4523x............
1 interface relay with a floating contact (230V/1A) is included in the scope of delivery of the generator.

8.2. Cables

) for mains supply and connection block (10mm2) for unit supply.
OPTIMUS CINTRODUCTION AND TECHNICAL DATA
H.V . cables
with O3 / O3 plugs: 9806 402 6xx02................................
length: 6m -- 30m in steps of 2m...........................................
capacity: 155pF/m.........................................
diameter: 16.5mm........................................
The cable length is indicated by the 9th and 10th digit of the numeric code.
Thermal contact cable
-- 2-wire screened for 1 excess temperature switch 4512 100 66151..................
-- 10-wire screened for additional supervision like
temperature alarm switch, buzzer, selection indicator 0722 215 19005..............
Stator cable
-- Tube SRM--0612 : 3x 1,31 mm
2
, screened with uMNL--connector 3322 405 14191....
-- all other tubes :
2
3×1.31mm
, screened 0722 215 02054........................................
AMPLIMAT cable
with D-Sub and 3--Plus plug:
12m 9890 000 01721......................
16m 9890 000 01731......................
20m 9890 000 01741......................
24m 9890 000 01751......................
Note
The above described cables are part of the pre--assembled systems.
(a/02.1)1--10
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS C
OPTIMUS_C_1_a021
OPTIMUS C
INTRODUCTION AND TECHNICAL DATA

8.3. Manpower

At least two persons are necessary to insert the H.V. tank in the generator cabinet. The weight is about 73kg.

9. Planned maintenance

The technical documentation for carrying out maintenance work in compliance with the applicable regulations are available at the responsible authority of Philips Medical Systems.
The importance of having maintenance implemented is pointed out to the operator in the operating instructions.
It must be guaranteed that the person carrying out maintenance work knows about the respective national regulations and that this person observes these regulations throughout all steps of maintenance work.
OPTIMUS C (a/02.1)
OPTIMUS_C_1_a021
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
1--11
OPTIMUS C
1. Installing the wall junction box 2--1........................................
2. Mounting the H.V. generator in the cabinet 2--2.............................
2.1. Mounting of the H.V. generator in the cabinet 2--2..............................
2.2. Electrical connection of the H.V. generator 2--3................................
3. Installing the operating panel 2--4.........................................
4. Electrical connection 2--5.................................................
4.1. Earthing 2--5..............................................................
4.2. Mains connection 2--5......................................................
4.2.1. Mains connection of the generator 2--5.......................................
4.2.2. Mains connection of the PDU 2--6...........................................
4.3. Stator connection 2--7......................................................
4.3.1. Shielding 2--7.............................................................
4.3.2. Connection 2--8...........................................................
4.4. Signal cables 2--9.........................................................
4.4.1. Room decade cable 2--9....................................................
4.4.2. Tube supervision 2--10......................................................
4.4.3. CAN interface 2--10.........................................................
4.4.4. Dose input 2--10............................................................
4.4.5. Dose rate input 2--10........................................................
4.5. H.V . cables generator side 2--11..............................................
4.6. Emergency--OFF circuit 2--1 1................................................
5. Hardware programming 2--12...............................................
6. Switch--ON of the generator 2--13...........................................
7. Installation software XRGSCOPE 2--13......................................
7.1. PC and generator settings to avoid problems during up/downloading of
7.2. Installation procedure 2--14..................................................
8. Setting-to-work overview 2--17.............................................
8.1. Configuration 2--17.........................................................
8.1.1. Date and time 2--17.........................................................
8.1.2. Mains data 2--17...........................................................
8.1.3. Tube data set 2--18.........................................................
8.1.4. Tube speed selection 2--18..................................................
8.1.5. Tube limits 2--18............................................................
8.1.6. Capacitance of tube connection 2--19.........................................
8.1.7. Tube operating modes 2--20.................................................
8.1.8. Disable tube 2--20..........................................................
8.2. Tube adjustment 2--21......................................................
8.2.1. Tube conditioning 2--21......................................................
8.2.2. Tube adaptation 2--25.......................................................
8.3. Dose rate control 2--28......................................................
8.3.1. AMPLIMAT sensitivity 2--28..................................................
8.3.2. Screen/film combinations 2--28...............................................
8.3.2.1. Automatic DRC processing 2--29.............................................
8.3.2.2. Manual DRC processing 2--30................................................
8.3.2.3. Density correction for AEC technique (option) 2--31.............................

INSTALLATION

INSTALLATION
TEXT
Contents 2--0.1............................................................
CU complete files 2--13......................................................
OPTIMUS C (a/02.1)
OPTIMUS_C_2_a021_inh
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
2--0.1
OPTIMUS CINSTALLATION
8.3.3. Image intensifier (II) 2--32....................................................
8.3.4. Fault exposure detection 2--32...............................................
8.3.5. Continuous fluoroscopy 2--32................................................
8.4. Application limits 2--34......................................................
8.4.1. X--mode limits 2--34.........................................................
8.4.2. Thoravison limits 2--34......................................................
8.4.3. Overload-dependent limits 2--34..............................................
8.5. Area exposure product calculatio(option) 2--35..................................
8.6. Acceptance test 2--35.......................................................
8.7. Backup of all configuration data 2--35.........................................
9. Labels 2--36...............................................................
10. Final installation work 2--36................................................
DRAWINGS
Data sets of chambers 2Z--4.........................................................
Labelling 2Z--10.....................................................................
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS C(a/02.1)2--0.2
OPTIMUS_C_2_a021_inh
OPTIMUS C
INSTALLATION

1. Installing the wall junction box

S Mount the wall junction box at the place where the generator is intended to be installed.
(See drawing “Connection of generator” in section 1 and manual UNIT 4512 103 75380 for wall junction boxes).
S If necessary, install the optional surge arrester WN inside the wall junction box.
(See surge arrester documentation.)
S If applicable, mount the filler panels of the generator to the wall junction box.
S Have the mains cable present at the clinic connected to mains terminal MEX by a person who is authorized for
this job.
S Check the phase sequence of L1, L2 and L3.
Warning!
Switch OFF the mains supply present at the clinic and make sure that it cannot be switched
ON again accidentally
.
OPTIMUS C (a/02.0)
OPTIMUS_C_2_a021
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
2--1
OPTIMUS CINSTALLATION

2. Mounting the H.V. generator in the cabinet

2.1. Mounting of the H.V. generator in the cabinet

Caution! Do not tilt the H.V. generator while transporting it.
In case of a tilting angle larger than 45_, the setting-to-work of the generator can be started not before a waiting time of about 8 hours has passed. Otherwise the H.V. generator may be destroyed by electrical sparkover.
S Unpack generator cabinet E.
S In case the packing material is strongly soiled with
oil check the oil level. Repair it if necessary.
deaerating hole
Tolerance: 2mm
Oil: Shell Diala G in 2.5l container
4512 148 43172
24mm at 20 - 30_C 22mm at 30 - 36_C
329H97
S Remove the deaerating screw completely from the cover of the H.V. generator. Only this way the precision of
the high voltage measuring divider corresponds to the specification. In case of return shipment of the H.V. generator this screw must be fixed again. Therefore, keep the screw laying on top of the cover.
Caution! Make sure that no foreign matter falls into the oil. Otherwise the transformer must be exchanged.
S Take the two transport bars from the rear side of the cabinet.
S Lift the H.V. generator into the generator cabinet with the transport bars.
The 4 connecting bolts GX1001 to 1004 must point at the front of the generator cabinet.
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS C(a/02.0)2--2
OPTIMUS_C_2_a021
Loading...
+ 198 hidden pages