Legend for earthing and cabling diagramZ--7.10.................................
OPTIMUS C(a/02.1)
OPTIMUS_C_1_a021_inh
E 2003 Philips Medical Systems
ALL RIGHTS RESERVED
1--0.1
OPTIMUS C
INTRODUCTION AND TECHNICAL DATA
1.Product information
The Optimus family of generators for radiography is based on computer-controlled converter technology.
The converter operates in the non-audible frequency range.
Applicable options can be enabled by releasing software modules using customized PAL ICs depending on the
respective order.
Control between the internal Function Units (FUs) and the external online equipment takes place by a CAN bus.
Safety-relevant signals are transferred directly on the so-called “Signal bus”.
1.1.Applications
-- Radiography
-- Tomography
-- Fluoroscopy
1.2.Options
1.2.1. Hardware options
-- Mains transformer PDU: 400 -- 480V; 50 / 60Hz,
also for 400V mains supply without neutral lead N
with taps for 400 / 440 / 460 / 480V9890 000 0260x......................................
-- Stand for operating panel9890 000 0244x.............................
-- Wall mounting of operating panel9890 000 0245x.......................
-- User interface support9890 000 0283x................................
-- Cable set for COCKPIT9890 000 0279x...............................
3.Mechanical data
For installation dimensions and weights see drawings Z--1.1.
Transport data:
Weights [kg]Dimensions [cm]
ase no.Contents
-- Generator cabinet
1
-- Operating panel
-- Cables
netgrosslengthwidthheight
1782262108284
2-- H.V. generator; 1--tube version73100776780
(a/02.1)1--2
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS C
OPTIMUS_C_1_a021
OPTIMUS C
INTRODUCTION AND TECHNICAL DATA
4.Environmental data
The environmental data comply with to PMS standard UXW 13600.
4.1.Electrical environment
Class S0 -- Dedicated mains supply, 3 phases and neutral. Thus single phase voltage is also available.
A low impedance, permanently installed connection, fed in by the step down transformer of the hospital to supply
large systems like in MR, CT and X--ray departments is required.
Note
Use always a mains cable with 4 wires and concentric PE--shield, type NYCY.
4.2.Climatic conditions
Ambient temperature10_C--40_C...................
Relative humidity15% -- 90%; no condensation.......................
-- External mains transformer PDU
9890 000 0260x is required
(requiresmodificationatthe
EMC-- filter of the kV power unit).
X-ray
installation
L2
L3
Caution!
Ensure the sequence of phases in the wall junction box corresponds to designations L1, L2, L3.
OPTIMUS C(a/02.1)
OPTIMUS_C_1_a021
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
-- 400V / 440V / 460V / 480V
1--7
5.4.2. Calculating the mains resistances
Note
The cross section of lead l
must not exceed 25mm2. (See figure below).
3
OPTIMUS CINTRODUCTION AND TECHNICAL DATA
If possible the sum of R
smaller than the R
requires.
XG
0,R1,R2
and R3should be
R
+R1+R2+R3<R
0
With higher internal mains resistances the generator
output is reduced correspondingly.
R
designates the mains resistance of the distributor transformer
0
R
depends on the length of lead l1between distributor transformer and mains distributor
1
L1
L2
L3
l
1
and on the selected cross section as well:
==>R
R
consists of upstream elements such as:
2
1=l1
× R
Cu
RCufrom table below
-- Emergency-OFF switch4.0mΩ.................
-- Earth-leakage circuit breaker5.5mΩ............
l
3
XG
MEX 100
L1
L2
L3
357J94
-- Fuse5.5mΩ.................................
-- Surge arrester WN23.0mΩ...................
R
depends on the length of lead l3between mains distributor and wall junction box
3
and on the selected cross section as well:
==>R
3=l3
× R
Cu
RCufrom table below
The resistances consider the go and return lines so that the calculation can be based on simple cable lengths.
Copper cross section
[mm
2
]
Resistance R
[mΩ/m]
Cu
162.19
251.4
351.0
500.7
700.5
950.38
1200.30
1500.24
Note
500m
Ω
is the absolute max. mains resistance.
(a/02.1)1--8
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS C
OPTIMUS_C_1_a021
OPTIMUS C
INTRODUCTION AND TECHNICAL DATA
5.4.3. Earth-leakage circuit breaker
To be provided between mains fuse and X-ray installation depending on local regulations.
Siemens earth-leakage circuit breaker:
-- Order No.: 5SZ3 466 OKG00
-- Rated fault current 30mA
-- Rated current 63A
-- Connection terminals for wire cross sections of up to 25mm
2
5.4.4. Emergency--OFF device
To be provided depending on local regulations.
There are 2 possibilities:
1.All the emergency--OFF buttons are connected in series and looped into the switch--ON circuit (12VDC) of the
generator.
2.The emergency--OFF circuit acts on an external mains contactor which switches OFF the power before it is
fed into the generator.
6.Tools
-- Service engineer standard tool kit
-- Service-PC:
IBM-compatible, 640kB RAM, 3.5” floppy disk drive, ≥ 1 serial port
-- Installation and service software OMC: 4512 116 024xx.
Supplied on a floppy disk within this generator service manual.
-- PC-hardkey (DIAGGEN):
Necessary to carry out the installation and to run the service software (special programmings, fault finding).
-- 0-modem cable:
Minimum length is distance between generator cabinet and operating desk.
Male 25-pole D-Sub connector at the generator side.
A 5m data cable of bucky controller can be used: 4512 130 5693x
The above described cables are part of the pre--assembled systems.
(a/02.1)1--10
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS C
OPTIMUS_C_1_a021
OPTIMUS C
INTRODUCTION AND TECHNICAL DATA
8.3.Manpower
At least two persons are necessary to insert the H.V. tank in the generator cabinet.
The weight is about 73kg.
9.Planned maintenance
The technical documentation for carrying out maintenance work in compliance with the applicable regulations are
available at the responsible authority of Philips Medical Systems.
The importance of having maintenance implemented is pointed out to the operator in the operating instructions.
It must be guaranteed that the person carrying out maintenance work knows about the respective national
regulations and that this person observes these regulations throughout all steps of maintenance work.
OPTIMUS C(a/02.1)
OPTIMUS_C_1_a021
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
1--11
OPTIMUS C
1.Installing the wall junction box2--1........................................
2.Mounting the H.V. generator in the cabinet2--2.............................
2.1.Mounting of the H.V. generator in the cabinet2--2..............................
2.2.Electrical connection of the H.V. generator2--3................................
3.Installing the operating panel2--4.........................................
S Mount the wall junction box at the place where the generator is intended to be installed.
(See drawing “Connection of generator” in section 1 and manual UNIT 4512 103 75380 for wall junction boxes).
S If necessary, install the optional surge arrester WN inside the wall junction box.
(See surge arrester documentation.)
S If applicable, mount the filler panels of the generator to the wall junction box.
S Have the mains cable present at the clinic connected to mains terminal MEX by a person who is authorized for
this job.
S Check the phase sequence of L1, L2 and L3.
Warning!
Switch OFF the mains supply present at the clinic and make sure that it cannot be switched
ON again accidentally
.
OPTIMUS C(a/02.0)
OPTIMUS_C_2_a021
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
2--1
OPTIMUS CINSTALLATION
2.Mounting the H.V. generator in the cabinet
2.1.Mounting of the H.V. generator in the cabinet
Caution!
Do not tilt the H.V. generator while transporting it.
In case of a tilting angle larger than 45_, the setting-to-work of the generator can be started not before a
waiting time of about 8 hours has passed. Otherwise the H.V. generator may be destroyed by electrical
sparkover.
S Unpack generator cabinet E.
S In case the packing material is strongly soiled with
oil check the oil level. Repair it if necessary.
deaerating hole
Tolerance:2mm
Oil:Shell Diala G in 2.5l container
4512 148 43172
24mm at 20 - 30_C
22mm at 30 - 36_C
329H97
S Remove the deaerating screw completely from the cover of the H.V. generator. Only this way the precision of
the high voltage measuring divider corresponds to the specification.
In case of return shipment of the H.V. generator this screw must be fixed again. Therefore, keep the screw laying
on top of the cover.
Caution!
Make sure that no foreign matter falls into the oil. Otherwise the transformer must be exchanged.
S Take the two transport bars from the rear side of the cabinet.
S Lift the H.V. generator into the generator cabinet with the transport bars.
The 4 connecting bolts GX1001 to 1004 must point at the front of the generator cabinet.
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS C(a/02.0)2--2
OPTIMUS_C_2_a021
Loading...
+ 198 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.