The IntelliVue MP2 portable patient monitor is
compact in size, ergonomic, and modular in design.
It provides an easy-to-use touchscreen user
interface, is highly customizable and shares a
technological platform with the Philips IntelliVue
MP5-MP90 patient monitors.
The IntelliVue series offers a complete monitoring
solution that is flexible and modular, designed to
suit a broad spectrum of monitoring needs.
Measurement Features
• Compact, rugged, lightweight monitor with built in
measurements
• ECG monitoring using any combination of three
to 10 electrodes.
• 12-lead ECG monitoring with five electrodes using
the EASI method or with 10 electrodes using the
conventional method.
• Multi-lead arrhythmia and ST segment analysis at
the bedside on all available leads.
• Mainstream or Sidestream CO
• FAST SpO2 for accurate performance even with
low perfusion.
• Invasive Pressure and Temperature measurement
• The monitor can operate using battery power for
up to 3 hours with basic monitoring configuration
to let you safely and easily monitor patients during
in-hospital transfer. AC power is provided by an
external power supply.
2
Usability Features
• Touchscreen and hardkeys as input device.
• Intuitive user interface.
PAD
Page 2
• Simple menu hierarchy gives fast access to all basic monitoring tasks.
• Patient data management with tabular and graphic trends.
• Settings “Profiles” for rapid case turnover.
• Patented automatic alarm limits help clinicians provide care more
efficiently.
• 3.5" TFT flat panel display with QVGA
(320 x 240) resolution, wide viewing angle, large numerics,
permanently visible alarm limits, and up to three real-time waves.
• Capable of functioning in a wireless infrastructure (IIT)
Intended Use
The monitor is intended to be used for monitoring and recording of,
and to generate alarms for, multiple physiological parameters of adults,
pediatrics, and neonates in a hospital environment and during patient
transport inside and outside of hospitals. The monitor is intended for
use by health care professionals.
The monitor is only for use on one patient at a time. It is not intended
for home use. Not a therapeutic device.
Rx only: U.S. Federal Law restricts this device to sale by or on the
order of a physician.
ST segment monitoring is intended for use with adult patients only and
is not clinically validated for use with neonatal and pediatric patients.
The ECG measurement is intended to be used for diagnostic recording
of rhythm and detailed morphology of complex cardiac complexes
(according to AAMI EC 11).
EMC Environment:
Main Components
Monitor
The monitor has a color LCD TFT display with a wide viewing angle,
providing high resolution waveform and data presentation.
The display, processing unit and measurements are integrated into one
device. An external power supply provides power for the monitor.
User Interface
The user interface is designed for fast and intuitive operation. The
color graphical user interface ensures that clinicians quickly feel at ease
using the monitor.
Configurable SmartKeys with intuitive icons allow monitoring tasks to
be performed quickly and easily, directly on the monitor screen.
Waves and numerics are color-coded.
The monitor displays up to three measurement waves simultaneously.
For 12-lead ECG monitoring it can display 12 real-time ECG waves,
with a rhythm strip and all ST values.
The MP2 monitor is supplied with a resistive touchscreen.
Simulated Keyboard
If alpha or numeric data entry is required, for example to enter patient
demographics, an on-screen keyboard will automatically appear on the
screen.
The monitor is suitable for use in all establishments incl. those directly
connected to the public low-voltage supply network that supplies
buildings for domestic purposes. When used with the M2741A
Sidestream CO2 sensor, the monitor can only be used in hospital
environments.
Upgradability
The MP2 monitor allows new capabilities to be added in the future as
your monitoring requirements evolve. This upgradability gives the
security of knowing that the monitors can be enhanced and updated as
practices and technologies advance, and it protects long-term
investments.
2
Mounting
The mounting options available enable flexible, space saving placement
of the monitors for an ergonomic work space. The monitor is shipped
with a low cost mounting plate if not specified otherwise.
Application Features
Critical and Cardiac Care Features
• The monitor performs multi-lead arrhythmia detection analysis
on the patient’s ECG waveform at the bedside. It analyzes for
ventricular arrhythmias, calculates heart rate, and generates alarms,
including asystole, bradycardia, and ventricular fibrillation.
• Up to 12 leads of ST segment analysis can be performed on adult
patients at the bedside, measuring ST segment elevation and
depression and generating alarms and events. The user can trend ST
changes, set high and low alarm limits, and set both ST and isoelectric
measurement points. Using ST Snippets, one-second wave segments
can be compared with a baseline segment for each measured ST lead.
• optional ST Map application shows ST changes over time in two
Page 3
multi-axis spider diagrams.
• QT/QTc interval monitoring provides the measured QT interval,
the calculated heart-rate corrected QTc value and a
'QTc value,
which tracks variation in the QT interval in relation to a baseline
value.
• optional 12-lead ECG data can be measured, using either the EASI
placement method with five standard electrodes or conventional
electrode placement with 10 electrodes.
1
• 12 realtime ECG waveforms can be displayed simultaneously.
• FAST-SpO2, using Fourier Artifact Suppression Technology,
performs accurately even in cases with low perfusion.
• Choice of sidestream or mainstream CO
monitoring for high
2
quality measurements with intubated and non-intubated patients.
- Vital Signs
- Graphic Trends
- Realtime Wave Reports
Report templates can be defined in advance, enabling print-outs
tailored to each hospital’s specific requirements to be started
quickly. Reports can be printed on centrally-connected printers or
via the IntelliVue PC Printing Solution, and they can be initiated
manually or automatically at user-defined intervals.
• The IntelliVue PC Printing Solution allows printing of reports,
waveform captures and trends from the MP2 to a standard off-the
shelf printer or to an electronic file.
Alarms
Ease of Use
• Screen layouts are easily adjustable, allowing flexible display of
measurement information.
• Temperature, height, and weight can be configured either in metric
or imperial units. Pressure measurements can be displayed in kPa or
mmHg. Gases can be displayed in kPa, mmHg.
Trends
• The trend database stores patient data from up to 30
measurement numerics. The measurement information can be
sampled every 12 seconds, one minute, or five minutes, and stored
for a period ranging from four to 48 hours.
• Horizon Trends show the deviation from a stored baseline.
Transport Features
• The monitor’s portable design means it can be used for in and out-ofhospital transport: a basic monitor weighs 1.5 kg.
• The monitor can operate using battery power for up to 3 hours, to
let you safely and easily monitor patients during procedures or inhospital transfer.
• Specially-designed mounting solutions let you quickly disconnect the
monitor for transport and reconnect to the mount after transport.
• The Universal Admit, Discharge and Transfer (ADT) feature means
that all ADT information is shared between the networked monitor
and the Information Center. Information need only be entered once.
Patient Data Documentation
• An extensive range of Patient Reports can be printed:
- 12-lead ECG Reports
- Alarm Limit Reports
The alarm system can be configured to present either the traditional
HP/Agilent/Philips alarm sounds or sounds compliant with the draft
ISO/IEC 9703-2 Standard.
Alarm limits are permanently visible on the main screen. The Alarm
Limits page provides a graphic depiction of alarm limits in relation to
the currently monitored measurement values and lets you adjust alarm
limits. It also lets you preview wide and narrow automatic alarm limits
before you apply them.
When an alarm limit is exceeded, it is signalled by the monitor in the
following ways:
• an alarm tone sounds, graded according to severity
• an alarm message is shown on the screen, color-coded according to
severity
• the numeric of the alarming measurement flashes on the screen
• alarm lamps flash for red and yellow alarms and are illuminated for
technical INOPs
If the monitor is connected via a network to a central monitoring
station, alarming is simultaneous at the monitor and at the Information
Center.
Alarms are graded and prioritized according to severity:
• Red Alarms*** identify a potentially life threatening situation for a
patient .
problems, equipment malfunction or equipment disconnect.
The Silence/Pause Alarms function (equivalent to Silence/Suspend
with previous monitor generations) allows the user to switch off
alarm tones with one touch.
1.EASI-derived 12-lead ECGs and their measurements are approximations to conventional
12-lead ECGs. As the 12-lead ECG derived with EASI is not exactly identical to the 12lead conventional ECG obtained from an electrocardiograph, it should not be used for
diagnostic purposes.
All alarms can be paused for a period of one, two, three, five, or 10
minutes or turned off indefinitely.
3
Page 4
Alarm strip recordings are available on a centrally-connected
recorder or via the IntelliVue PC Printing Solution.
Patented automatic alarm limits automatically adapt the alarm limits
to the patient’s currently measured vital signs within a safe margin
defined individually for each patient.
Visual and/or audible latching and non-latching alarm handling is
available.
Profiles
Profiles are predefined configuration settings for Screens,
measurement settings, and monitor settings. Each Profile can be
designed for a specific application area and patient category, for
example OR adult, or ICU neonatal. Profiles enable a quick reaction to
patient and care location changes: activating a Profile with a particular
patient category (Adult, Pediatric or Neonatal) automatically applies
suitable alarm and safety limits and saves time usually spent carrying
out a complete set-up procedure.
Profiles can be created directly on the monitor or remotely on a
personal computer and transferred to the monitor using the IntelliVue
Support Tool. A selection of Profiles for common monitoring
situations is provided with the monitor. These profiles can be changed,
added to, renamed, or deleted.
Network Interface
The network interface provides the system with networking capability
via a wired or wireless network connection.
Wireless Network (optional)
The monitor can function within a telemetry infrastructure compatible
with the Philips Cellular Telemetry System (CTS) in the WMTS and
ISM bands. Additional components are required to complete the
system. Please refer to the M3185A IntelliVue Clinical Network
Technical Data Sheet for further information.
Monitor Specifications
Safety Specifications
The monitor complies with the Medical Device Directive 93/42/EEC
) and with IEC 60601-1:1988 + A1:1991 + A2:1995; EN60601-
(CE
0366
1:1990 + A1:1993 + A2:1995; UL 60601-1:2003; CAN/CSA
C22.2#601.1-M90; JIS T 0601-1:1999; IEC 60601-1-1:2000; EN 606011-1:2001.
All applied parts are Type CF unless otherwise specified. They are
protected against damage from defibrillation and electrosurgery.
Optional Networking Capabilities
The monitor can operate as part of a wired or wireless hospital
network system, using the Philips IntelliVue Clinical Network interface.
Service Features
• The Support Tool helps technical personnel to
- carry out configuration, upgrades and troubleshooting via the
network, or on an individual monitor
- share configuration settings between monitors
- back up the monitor settings.
• A password-protected Service Mode ensures that only trained staff
can access service tests and tasks.
• The Configuration Mode is password-protected and allows trained
users to customize the monitor configuration.
Device Connections
The monitor can be connected to:
• an Information Center (for example M3150B)
•a PC
• MMS Extensions (M3012A, M3014A, M3015A, M3016A)
1
The possibility of hazards arising from software errors was minimized
in compliance with
and IEC 60601-1-4:1996 + A1:1999.
The monitor complies with the EMC standards
IEC 60601-1-2:2001; EN 60601-1-2:2001
This ISM device complies with Canadian ICES-001. Cet appareil ISM
est conforme a la norme NMB-001 du Canada.
The MP2 patient monitor can be used in a transport environment such
as road ambulance, airplane or helicopter, except when used with the
M2741A Sidestream CO2 sensor. For this purpose, the monitor fulfills
the following additional mechanical, EMC and environmental
requirements:
• Shock Tests according to IEC TR 60721-4-7, Class 7M3. Test
procedure according to IEC/EN 60068-2-27 (peak acceleration up to
100g).
• Random Vibration according to IEC TR 60721-4-7, Class 7M3. Test
procedure according to IEC/EN 60068-2-64 (RMS acceleration 5g).
• Sinusoidal Vibration according to IEC TR 60721-4-7, Class 7M3.
Test procedure according to IEC/EN 60068-2-6 (acceleration up to
amplitude 2g).
1.The MMS Extensions will only function when the monitor is connected to the
external power.
ISO 14971:2000, EN60601-1-4:1996 + A1:1999
4
Page 5
• Bump Test according to IEC/EN60068-2-29 (peak acceleration 15g,
1000 bumps).
• Free Fall Test according to EN1789 (covers also IEC TR 60721-4-7
and Class 7M3). Test procedure according to EN 60068-2-32 (height
0.75 m).
• Specification for degrees of protection provided by enclosures
according to IEC/EN 60529: IP 32
• EN 1789 +A1:2003 Medical vehicles and their equipment - Road
ambulances (chapter 6 - Medical Devices).
• Radiated susceptibility 20 V/m according to EN ISO 9919 (SpO2)
and EN ISO 21647 (CO2).
• Altitude Range from -500 to 3000 m operating and -500 to 4600 m
storage and transportation.
• Extended radiated susceptibility tests
The MP2 patient monitor with its out-of-hospital parameter set
provides a general immunity level of 20 V/m with only few restrictions.
Details are as listed below:
- GSM 900: Immunity at 900 MHz (uplink mobile phone), 20V/m,
duty cycle 1:8
- GSM 1800: Immunity at 1800 MHz (uplink mobile phone), 20V/m,
duty cycle 1:8.
- DECT: Immunity at 1800 MHz (digital cordless phone), 20V/m,
duty cycle 1:24
- AM: 1 kHz Immunity from 80 MHz to 2.5 GHz (any radio
communication unit, broadcasting and TV transmitter), 20V/m,
modulation factor 80%. (ECG: 20 V/m except 0.8-1.2 GHz where
it is 10V/m)
• Operating ambient temperature testing over the range from 0 to
40 °C (32 to 104 °F).
• Operating ambient humidity testing up to 95% RH at 40 °C
(104 °F), non condensing.
Physical Specifications
Environmental Specifications
ItemConditionRange
Temperature
Range
Temperature
Range when
charging the
battery
Humidity RangeOperating15% to 95% Relative
Altitude RangeOperating-500 m to 3000 m
Ingress
Protection
Operating
0 to 40oC
(32 to 104 oF)
Storage
(incl. Transport)
Operating
-20 to 60oC
(-4 to 140
0 to 35oC
(32 to 95
o
o
F)
F)
Humidity (RH) (non
condensing)
Storage and
Transport
5% to 95% Relative
Humidity (RH)
(10000 ft)
Storage and
Transport
-500 m to 4600 m
(15000 ft)
a
MonitorIP32 (protected against
the ingress of solid
foreign objects 2.5 mm
in diameter or larger,
and the ingress of water
when the water is
dripping vertically and
the monitor is tilted up
to 15°).
ProductMax Weight W x H x D
M8102A
IntelliVue MP2 (without
1.5 kg
(3.3 lb)
< 188 x 99 x 86 mm
(7.4 x 3.9 x 3.4 in)
handle and options)
External Power
Supply (M8023A)
IP31(protected against
the ingress of solid
foreign objects 2.5 mm
in diameter or larger,
and the ingress of water
when the water is
dripping vertically) when
rested on its rubber feet
on a flat, level surface.
IP32 when mounted
with the connectors
facing downwards
a. sufficient for flight altitudes up to 12,000 m with pressurized cabins
5
Page 6
Performance Specifications
Monitor Performance Specifications
Alarm SignalSystem delayless than 3 seconds
Monitor Performance Specifications
Power
Specifications
Battery
Specifications
IndicatorsAlarms Offred LED
Power
consumption
Line Voltage100 to 240 V ~
Current1.3 to 0.7A
Frequency50/60 Hz
Operating Time
(with new, fully
charged battery at
25 °C)
Charge TimeWhen MP2 is off: 2 h
Alarmsred/yellow/cyan LED
On/Standby/Errorgreen/red LED
< 40W average, <65W
peak
Basic monitor
configuration: 3 hours
When MP2 is in use and
connected to MP20/30/
40/50/60/70/80/90
without extensions: 12 h
approx.
When MP2 is in use and
connected to the external
power
Pause duration1,2,3 minutes or infinite,
depending on
configuration
Extended alarm
pause
Review
Alarms
Real Time
Clock
Buffered
Memory
Restart time: After power interruption, an ECG wave will be
shown on the display after 30 seconds maximum.
Information: all alarms / inops, main alarms on/off,
alarms acknowledged and time of occurrence
capacity500 items
Range: from: January 1, 1997, 00:00 to: December
31, 2080, 23:59
Accuracy: < 4 seconds per day (typically)
Hold Time: infinite if powered by host monitor or
external power supply; otherwise at least 48 hours
Contents: Active settings, trends, patient data,
realtime reports, review alarms
Hold Time: infinite if powered by external power
supply; otherwise at least 48 hours
5 or 10 minutes
M8023A External Power Supply Performance
Specifications
AC Powergreen LED
Batteryyellow (charging)/red
blinking (empty) LED
External Powergreen LED
SoundsAudible feedback for user input. Prompt tone.
QRS tones, or SpO
different alarm sounds.
Trends:
12, or 16 numerics @ 12 sec, 1 minute, 5 minute resolution.
Multiple choices of number of numerics, resolution and duration
depending on trend option and application area.
6
modulation tone. Four
2
M8023A External Power Supply Performance
Specifications
Power
Specifications
IndicatorsAC Powergreen LED
Power
Consumption
Line Voltage100 to 240 V ~
Current0.7 to 0.4 A
Frequency50/60 Hz ~
< 12 W average
< 30 W peak
Page 7
Interface Specifications
MP2 (M8102A) Interface Specifications
Measurement
Link (MSL)
ConnectorsFemale ODU (Proprietary)
Power30 V to 60 V input
Power SyncRS-422 compliant input
78.125 kHz (typical
LAN signalsIEEE 802.3 10-Base-T
complaint
Serial signalsRS-422 compliant
Local signalsProvided for connecting MMS
extensions
ECG Sync Pulse
Output
Cable
Detection
Yes
Marker InNo
Wave
No
Output
ConnectorBinder Series 709/719
Output
Levels
Output low <0.8V @ I = 4mA
Output high >2.4 V @ I=
4mA
IsolationNone
M8023A External Power Supply Interface Specifications
Measurement Link
(MSL)
ConnectorsMale ODU (Proprietary)
Power48 V output
Power Sync.RS-422 compliant output
78.125 kHz (typical)
LAN signalsIEEE 802.3 10-Base-T complaint
Serial
signals
RS-422 compliant output
78.125 kHz (typical)
Local signalsNot connected
Display Specifications
Integrated
QVGA
Display
Sweep
6.25, 12.5, 25 and 50 mm/s;
Speeds
Resolution320 x 240
Refresh
60 Hz
frequency
Useful
72x54mm (2.8x2.1in)
screen
Pixel size0.22 x 0.22 mm
Pulse Width100 +/- 10 ms (high)
Delay from
R-wave peak
20 ms maximum per AAMI
EC13
to start of
pulse
Minimum
0.5 V
required Rwave
amplitude
MP2 (M8102A) Compatible Devices
IntelliVue Instrument Telemetry Wireless Network
(USA only)
Internal WMTS
Adapter
Technologycompatible with Philips
Cellular Telemetry System
(CTS), cellular infrastructure
Frequency
Band
WMTS, 1395-1400 MHz and
1427-1432 MHz
IntelliVue Instrument Telemetry Wireless Network
(except USA)
7
Page 8
MP2 (M8102A) Compatible Devices
Measurement Specifications
Internal ISM
Adapter
M4607A Battery Specifications
Physical Specifications
W x D x H66 mm (2.36 in) x 80 mm (3.15 in) x
Weight 160 g ±5%
Performance Specifications
Nominal Voltage 10.8 Volt
Rated Capacity at
discharge C/5
Environmental Specifications
Temperature Range Discharge 0 to 60qC (32 to 122qF)
Humidity Range Operating: 15 % to 95 % Relative
Battery TypeLithium Ion Mangan, 10.8 V,
Safety complies with UL 1642 (UL
Technologycompatible with Philips
Cellular Telemetry System
(CTS), cellular infrastructure
Frequency
Band
2.4 GHz ISM
20 mm (0.79 in)
1000 mAh (typical)
Charge 0 to 60qC (32 to 122qF)
Storage and Transportation: -20 to
65qC (-4 to 140qF)
Humidity (RH)
Storage and Transportation: 5 % to
95 % Relative Humidity (RH)
1000 mAh,
Recognized)
ECG/Arrhythmia/ST/QT
Complies with IEC 60601-2-25:1993 + A1:1999 /EN60601-2-25:1995
+ A1:1999, IEC 60601-2-27:2005/EN60601-2-27:2006, IEC 60601-251:2003 /EN 60601-2-51:2003 and AAMI EC11/EC13:1991/2002.
ECG/Arrhythmia/ST Performance Specifications
CardiotachRangeAdult/pedi:
15 to 300 bpm
Neo range:
15 to 350 bpm
Accuracy±1% of range
Resolution1 bpm
Sensitivityt200 μV
PVC RateRange0 to 300 bpm
Resolution1 bpm
ST NumericRange-20 to +20 mm
Accuracy±0.5 mm or 15%, whichever is
greater
Resolution0.1 mm
QT NumericRange200 to 800 ms
Accuracy±30 ms
Resolution8ms
QTc NumericRange200 to 800 ms
Resolution1ms
'QTc NumericRange-600 to +600 ms
Resolution1ms
peak
Electromagnetic
Compatibility (EMC)
Communication Standardcomplies with the SMBus specification
8
complies with the requirements for
FCC Type B computing Device, and
EN 61000-4-2 and EN 61000-3-2
Heart Rate Averaging MethodThree different methods are
used:
Normally, heart rate is computed
by averaging the 12 most recent
RR intervals.
For runs of PVCs, up to 8 RR
intervals are averaged to
compute the HR.
If each of 3 consecutive RR
intervals is greater than 1200 ms
(that is, rate less than 50 bpm),
then the 4 most recent RR
intervals are averaged to
compute the HR.
Response Time of Heart Rate
Meter to Change in Heart
Rate
HR change from 80 to 120 bpm:
Range: [6.4 to 7.2 seconds]
Average: 6.8 seconds
HR change from 80 to 40 bpm:
Range: [5.6 to 6.4 sec] Average:
6.0 seconds
Accuracy of Input Signal
Reproduction
Methods A and D were used to
establish overall system error
and frequency response.
Respiration
Respiration Performance Specifications
Respiration
Rate
Bandwidth0.3 to 2.5Hz (–6dB)
NoiseLess than 25m:(rms) referred to
Respiration
Alarm Speci-
fications
HighAdult/pedi:
LowAdult/pedi: 0
RangeAdult/pedi: 0 to 120 rpm
Neo: 0 to 170 rpm
Accuracyat 0 to 120 rpm ±1 rpm
at 120 to 170 rpm ±2 rpm
Resolution1 rpm
the input
RangeAdjustmentDelay
under 20 rpm: 1
10 to 100
rpm
Neo: 30 to
rpm steps
over 20 rpm: 5
rpm steps
150 rpm
under 20 rpm: 1
to 95 rpm
Neo:
0 to 145 rpm
rpm steps
over 20 rpm: 5
rpm steps
max. 14
seconds
for limits
from 0 to
20 rpm:
max. 4
seconds
for limits
above 20
rpm: max.
14 seconds
10
Page 11
Respiration
Alarm Speci-
RangeAdjustmentDelay
fications
Apnea Alarm10 to 40
5 second steps
seconds
SpO
2
Complies with EN ISO 9919:2005 (except alarm system; alarm system
complies with IEC 60601-2-49:2001).
Measurement Validation: The SpO
accuracy has been validated in
2
human studies against arterial blood sample reference measured with a
CO-oximeter. Pulse oximeter measurements are statistically
distributed, only about two-thirds of the measurements can be
expected to fall within the specified accuracy compared to COoximeter measurements. Display Update Period: Typical: 2 seconds,
Maximum: 30 seconds. Max. with NBP INOP suppression on60
seconds.
Measurement Validation: In adult and pediatric mode, the blood
pressure measurements determined with this device comply with the
American National Standard for Electronic or Automated
Sphygmomanometers (ANSI/AAMI SP10 - 1992) in relation to mean
error and standard deviation, when compared to intra-arterial or
auscultatory measurements (depending on the configuration) in a
representative patient population. For the auscultatory reference the
5th Korotkoff sound was used to determine the diastolic pressure.
In neonatal mode, the blood pressure measurements determined with
this device comply with the American National Standard for Electronic
or Automated Sphygmomanometers (ANSI/AAMI SP10 - 1992 and
AAMI/ANSI SP10A -1996) in relation to mean error and standard
deviation, when compared to intra-arterial measurements in a
representative patient population.
1 bpm steps (25
to 40 bpm)
5 bpm steps (40
to 300 bpm)
Pedi/Neo:
1 bpm steps (25
to 50 bpm)
5 bpm steps (50
to 300 bpm)
max. 12
seconds
TachycardiaDifference to
high limit 0 to
5 bpm stepsmax. 14
seconds
50 bpm
Clamping at
5 bpm steps
150 to 300
bpm
14
Page 15
Invasive
Pressure
Alarm
Specifications
RangeAdjustmentDelay
CO
2
The CO2 measurement in the monitor, M3014A and M3015A
complies with EN ISO 21647:2004 + Cor.1:2005 (except alarm system;
alarm system complies with IEC 60601-2-49:2001).
BradycardiaDifference to
5 bpm stepsmax. 14
low limit 0 to
50 bpm
Clamping at 25
5 bpm steps
to 100 bpm
Temp
Complies with EN 12470-4:2000
Temp Performance Specifications
TempRange–1 to 45 qC
(30 to 113 qF)
Resolution0.1 qC (0.2 qF)
Accuracy±0.1 qC (±0.2 qF)
Average Time ConstantLess than 10 seconds
AlarmsRange–1 to 45 qC (30 to 113 qF)
Adjustment-1 to 35 qC (30 to 95 qF): 0.5 qC
(1.0 qF) steps
35 to 45 qC (95 to 113 qF): 0.1 qC
(0.2 qF) steps
seconds
M3015A Microstream CO2 Performance Specifications
CO
2
Range0 to 98 mmHg (0 to 13 kPa), or
13 % CO
, whichever is lower
2
AccuracyUp to 5 minutes during warmup:
±4 mmHg or 12 %, whichever is
greater
After 5 minutes warmup:
0 to 40 mmHg (0 to 5.3 kPa):
±2.2 mmHg (±0.3 kPa)
Above 40 mmHg (5.3 kPa):±(5.5 % +
(0.08 %/mmHg above 40 mmHg)) of
reading
These specifications are valid for
21 % O
and N2 balance, up to 35qC
2
ambient temperature, up to 60 rpm
in adult mode and 100 rpm in
neonatal mode.
Outside of these
conditions the accuracy reaches at a
minimum ±4 mmHg or ±12 % of the
reading, whichever is greater.
ResolutionNumeric: 1.0 mmHg (0.1 kPa)
Wave: 0.1 mmHg (0.01 kPa)
StabilityIncluded in Accuracy specifications
Temp Alarm
Specifications
Temp High/
Low Alarms
RangeAdjustment
–1 to 45 ºC (30
to 113 ºF)
-1 to 35 qC (30 to 95 qF), 0.5
qC (1.0 qF) steps
35 to 45 qC (95 to 113 qF),
0.1 qC (0.2 qF) steps
awRRRange0 to 150 rpm
Accuracy0 to 40 rpm: ±1 rpm
41 to 70 rpm: ±2 rpm
71 to 100 rpm: ±3 rpm
>100 rpm: ±5 % of reading
Warm-up Time5 minutes for full accuracy
specification
Rise Time 190 ms for neonatal mode
(measured with FilterLine H for
neonatal)
240 ms for adult mode
(measured with FilterLine H for
adult)
Sample Flow Rate50 + 15/-7.5 ml/minute
15
Page 16
M3015A Microstream CO2 Performance Specifications
Gas Sampling Delay TimeTypical:2.3 seconds
Maximum:3 seconds
Sound PressureAcoustic noise: <45 dBA
Total System Response
Time
The total system response time is
the sum of the delay time and the
rise time.
M3014A Mainstream CO2 Performance Specifications
CO
2
Range0 to 150 mmHg (0 to 20.0 kPa)
Accuracyafter 2 minutes warmup:
For values between 0 and 40 mmHg:
±2.0 mmHg (±0.29 kPa)
For values from 41 to 70 mmHg: ±5 %
of reading
For values from 71 to 100 mmHg:
±8 % of reading
The specifications are valid for
standard gas mixtures, balance air, fully
hydrated at 35°C, P
= 760 mmHg,
abs
flow rate = 2 l/min.
ResolutionNumeric: 1.0 mmHg (0.1 kPa)
Wave: 0.1 mmHg (0.01 kPa)
Stability:
Short term
drift
Long term
±0.8 mmHg over four hours
Accuracy specification will be
maintained over a 120 hour period
drift
awRRRange2 to 150 rpm
Accuracy±1 rpm
M3014A Sidestream CO2 Performance Specifications
CO
2
Range0 to 150 mmHg (0 to 20.0 kPa)
Accuracyafter 2 minutes warmup:
For values between 0 and 40 mmHg:
±2.0 mmHg (±0.29 kPa)
For values from 41 to 70 mmHg: ±5 %
of reading
For values from 71 to 100 mmHg:
±8 % of reading
For values from 101 to 150 mmHg:
±10 % of reading
At respiration rates above 80 rpm, all
ranges are ±12 % of actual. The
specifications are valid for gas mixtures
, balance N2, dry gas at
of CO
2
760 mmHg within specified operating
temperature range.
ResolutionNumeric: 1.0 mmHg (0.1 kPa)
Wave: 0.1 mmHg (0.01 kPa)
Stability:
Short term
drift
Long term
±0.8 mmHg over four hours
Accuracy specification will be
maintained over a 120 hour period
drift
awRRRange2 to 150 rpm
Accuracy±1 rpm
Warm-up Time2 minutes with CO2 sensor attached
for full accuracy specification
Sample Flow Rate50 ±10 ml/minute
Total System Response
3seconds
Time
Warm-up Time2 minutes with CO2 transducer
attached for full accuracy specification
Response TimeLess than 60 ms (with adult or infant
reusable or disposable adapter)
16
Operating Temperature0 to 40°C (32 to 104°F)
M3014A Mainstream and Sidestream CO2 Humidity Correction
Factor
Either BTPS or STPD can be selected as the humidity correction factor
for the CO
Where p = partial pressure, P
readings. The formula for the correction calculation is:
2
P
abs
---------------------------
P
STPDPBTPS
=
P
–
absPH2 O
= absolute pressure, and P
abs
H2O
= 42
mmHg @35°C and 100 % RH.
Page 17
M3016A Mainstream CO2 Performance Specifications
CO
2
Range–4 to 150 mmHg (-0.5 to 20.0 kPa)
Accuracyafter 20 minutes warmup and
calibration:
For values between 0 and 40 mmHg:
±2.2 mmHg (±0.29 kPa)
For values between 40 and 76 mmHg:
±5.5 % of reading
The specifications are valid for 45 %
and N2 or N2O balance. Outside
O
2
these conditions the accuracy reaches
at a minimum the requirements of
EN864/ISO9918.
ResolutionNumeric: 1.0 mmHg (0.1 kPa)
Wave: 0.1 mmHg (0.01 kPa)
Stability±1.0 mmHg over a 7 day period
awRRRange0 to 150 rpm
Accuracy±2 rpm
Warm-up Time20 minutes with CO2 transducer
attached for full accuracy specification
Response TimeLess than 125 ms (for step from 10 %
to 90 %)
Mainstream CO2 Humidity Correction Factor
Either BTPS or STPD can be selected as the humidity correction factor
for the Mainstream CO
readings. The formula for the correction
2
calculation is:
P
abs
P
STPDPBTPS
Where p = partial pressure, P
=
---------------------------
P
–
absPH2 O
= absolute pressure, and P
abs
H2O
=
47 mmHg @37qC and 100 % RH.
CO2 Alarm
Specifications
Range
etCO2 High20 to
95 mmHg (2
to 13 kPa)
etCO2 Low10 to
90 mmHg (1
to 12 kPa)
imCO2 High2 to
20 mmHg
(0.3 to
3.0 kPa)
awRR HighAdult/pedi:
10 to
100 rpm
Neo: 30 to
150 rpm
awRR LowAdult/pedi:
0 to 95 rpm
Neo: 0 to
145 rpm
Apnea delay10 to
40 seconds
Adjust-
ment
1mmHg
(0.1 kPa)
steps of
1mmHg
(0.1 kPa)
under
20 rpm: 1
rpm steps
over
20 rpm:5
rpm steps
5 second
steps
Delay
M3014A/
M3016A: less
than 14 seconds
M3015A: less
than18 seconds.
M3014A/
M3016A: less
than 14 seconds
M3015A: less
than18 seconds.
M3014A/
M3016A: less
than 14 seconds
M3015A: less
than18 seconds.
M3015A:
settings
<20 rpm: less
than 8 seconds
>20 rpm: less
than 18 seconds
M3014A/
M3016A
settings
<20 rpm: less
than 4 seconds
>20 rpm: less
than 14 seconds
set apnea delay
time + 4 seconds
(M3014A/
M3016A) or
8seconds
(M3015A)
17
Page 18
Ordering Information
Ordering information for the M81052A patient monitor is given here.
The Intellivue MP5 Roll Stand Mounting Kit (Order No.
989803002531) is compatible with the table top mount and the
standard mounting plate. For information on other mounting
hardware, contact your local Philips sales representative. For GCX
mounting hardware information, see www.gcx.com/philips.
Documentation
Trunk Cables
Part
No.
M1669AM1668AM1667AM1663AM1665A
3-Electrode Cable Set
5-Electrode Cable Set
6-Electrode Cable Set
10-Electrode Cable set
(5+5)
10-Electrode Cable set
(6+4)
All documentation is available in .pdf format on documentation CDROM. Additionally, a printed copy of the Instructions for Use and
Quick Guide ships with each monitor.
• Instructions for Use (printed)
• Quick Guide (printed)
• Installation and Service Guide
• Configuration Guide
• Documentation CD-ROM
• Training Guide (printed)
• Computer Based Training (optional)
ECG Accessories
This symbol indicates that the cables and accessories are
designed to have special protection against electric shocks
(particularly regarding allowable leakage currents), and
are defibrillator proof.
Length2.7m2.7m2.7m2.0m2.7m
3-Electrode Cable Sets
DescriptionLength
OR Grabber
shielded
ICU Grabber
shielded
ICU snap
shielded
ICU Clip nonshielded
ICU Clip nonshielded
1.0mM1675AM1678A
1.0mM1671AM1672A
1.0mM1673AM1674A
0.45mM1622A--
0.7mM1624AM1626A
AAMI Part
No.
IEC Part No.
5-Electrode Cable Sets
DescriptionLength
OR Grabber
shielded
ICU Grabber
shielded
ICU Snap shielded1.0m/1.6mM1644AM1645A
ICU Miniclip nonshielded
1.0m/1.6mM1973AM1974A
1.0m/1.6mM1968AM1971A
0.7m/1.3mM1647AM1648A
AAMI
Part No.
IEC Part
No.
19
Page 20
6-Electrode Cable Sets
One-piece Cables
DescriptionLength
OR Grabber 1.0m/1.6mM1684AM1685A
ICU Grabber1.0m/1.6mM1680A M1681A
ICU Snap1.0m/1.6mM1682A M1683A
AAMI Part
No.
IEC Part
No.
10-Electrode (5+5)Cable Sets
DescriptionLength
ICU Grabber,
chest, shielded
ICU Snap, chest,
shielded
OR Grabber,
chest, shielded
For Limb Leads see 5-electrode cable sets
1.0mM1976AM1978A
1.0mM1602AM1604A
1.0mM1979AM1984A
AAMI Part
No.
IEC Part
No.
DescriptionLength
3-lead Grabber,
ICU
5-lead Grabber,
ICU
1.0m989803143181989803143171
1.0m989803143201989803143191
AAMI Part
No.
IEC Part No.
Radio-translucent Cables
Pack of five single wires, radio-translucent, 0.9m, M1649A
Set Combiners and Organizers
Set combiners and organizersPart No.
Set combiner3-electrode M1501A
5-electrodeM1502A
Set organizer for
shielded leadsets grabber and snap
3-electrodeM1503A
4-electrodeM1664A
5-electrodeM1504A
10-Electrode (6+4)Cable Sets
DescriptionLength
ICU Grabber,
chest, shielded
ICU Snap, chest,
shielded
OR Grabber,
chest, shielded
For Limb Leads see 6-electrode cable sets
1.0mM1532AM1533A
1.0mM1537AM1538A
1.0mM1557AM1558A
AAMI Part
No.
IEC Part
No.
6-electrodeM1679A
Set organizer for nonshielded lead sets miniclip
Bedsheet clipM1509A
Replacement red cover for trunk cable (for 5electrode cable sets)
3-electrodeM1636A
5-electrodeM1638A
989808148861
20
Page 21
Philips FAST SpO2Accessories
Part NumberDescriptionConnector Type
Philips Reusable Sensors
Part NumberDescriptionConnector Type
M1191A/BAdult Sensor
(2m cable)
M1191AL/
BL
Adult Sensor
(3m cable)
M1191ANLAdult Sensor
(3m cable)
Nellcor OxiMax-
compatible
a
M1191TAdult Sensor
(requires M1943A
(1.1m) or M1943AL
(3m) adapter cable)
M1192ASmall Adult/Pediatric
sensor (1.5m cable)
M1192ANSmall Adult/Pediatric
sensor (1.5m cable)
Nellcor OxiMax-
compatible
a
M1192TSmall Adult Pediatric
sensor (requires
M1943A (1.1m) or
M1943AL (3m)
adapter cable)
Philips 8-pin
Generic D-Sub
Philips 8-pin
Generic D-Sub
M1194AAdult/Pediatric Clip
Philips 8-pin
Sensor (ear) (1.5m
cable)
M1194ANAdult/Pediatric Clip
Sensor (ear) (1.5m
cable)
Nellcor OxiMax-
compatible
M1195AInfant Sensor (1.5m
a
Philips 8-pin
cable)
M1195ANInfant Sensor (1.5m
cable)
Nellcor OxiMax-
compatible
M1196AAdult Clip Sensor
a
Philips 8-pin
(3m cable)
M1196TAdult Clip Sensor
Generic D-Sub
(requires M1943A
(1.1m) or M1943AL
(3m) adapter cable)
a. only in combination with Philips FAST-SpO2 and Philips OxiMax-compatible patient
monitors.
Philips Disposable Sensors
M1193ANeonatal Hand/Foot
Sensor (1.5m cable)
M1193ANNeonatal Hand/Foot
Sensor (1.5m cable)
Nellcor OxiMax-
compatible
a
M1193TNeonatal Sensor
(requires M1943A
(1.1m) or M1943AL
(3m) adapter cable)
Purchase Nellcor OxiCliq sensors and adapter cables directly from
Tyco Healthcare.
Philips
Product NumberDescription
OxiMax MAX-A
OxiMax MAX-AL
a
a
Adult Sensor
Adult Sensor
(long cable)
OxiMax MAX-P
OxiMax MAX-I
OxiMax MAX-N
a
a
a
Pediatric Sensor
Infant Sensor
Neonatal Sensor
Part
Number
M1904B
n/a
M1903B
M1902B
M1901B
Product
Number
LNOP DC-IAdult Sensor989803140321
LNOP DC-IPPediatric Sensor989803140331
LNOP-YIReusable Multi-Site
LNOP TC-ITip Clip reusable
MASIMO LNCS®
Product
b
b
b
b
Number
LNCS DC-IAdult Sensor989803148281
LNCS DC-IPPediatric Sensor989803148291
LNCS-TC-IReusable Ear
LNCS TF-IReusable Forehead
Description
Sensor
Sensor
1
Reusable Sensors:
Description
Sensor
Sensor
Philips
Part Number
n/a
989803140341
Philips
Part Number
989803148301
989803148311
Oxisensor II D-25
Oxisensor II D-20
Oxisensor II I-20
Oxisensor II N-25
OxiCliq A
Oxicliq P
OxiCliq I
OxiCliq N
a. Requires M1943 A(L) adapter cable
b.not available from Philips in the U.S.A.
c. Requires M1943 A(L) and OC3 adapter cables
1.Nellcor, OxiMax and OxiCliq are trademarks of Nellcor Puritan Bennett Inc., a part
of Tyco Healthcare.
c
c
c
c
a
a
a
a
Adult Sensorn/a
Pediatric Sensorn/a
Infant Sensorn/a
Neonatal Sensorn/a
Adult Sensorn/a
Pediatric Sensorn/a
Infant Sensorn/a
Neonatal Sensorn/a
2.LNOP and LNCS are federally registered trademarks of Masimo Corporation
22
Page 23
MASIMO LNOP® Disposable Adhesive Sensors:
Extension/Adapter Cables:
Product
Number
Description
Philips
Part Number
LNOP AdtAdult Sensor989803140231
LNOP AdtxAdult Sensorn/a
LNOP PdtPediatric Adhesive
989803140261
Sensor
LNOP PdtxPediatric Sensorn/a
LNOP INF-LNeo/Infant
989803140311
Adhesive Sensor
LNOP NEO-LNeo Adhesive
989803140291
Sensor
LNOP NEOPT-LNeo Pre-Term
989803140301
Sensitive Skin
Adhesive Sensors
MASIMO LNCS® Disposable Adhesive Sensors:
Product
Number
Description
LNCS AdtxAdult Finger
Sensor
Philips
Part Number
989803148231
Part NumberDescription
M1941AExtension Cable (2m) (8-pin to 8-pin)
M1943AAdapter Cable (1.1m) for Philips and
Nellcor disposable sensors
(8-pin to 9-pin D-Sub)
M1943ALAdapter Cable (3m) for Philips and
Nellcor disposable sensors
(8-pin to 9-pin D-Sub)
OC3Adapter cable for OxiCliq Sensors
(available from Nellcor only)
LNOP MP12
(451261000761)
LNOP MP Series Patient Cable (3.6 m)
Adapter Cable for Masimo LNOP
Sensors
LNC MP10
(989803148221)
LNCS MP Series Patient CAble (3.0 m)
Adapter Cable for Masimo LNCS
Sensors
Non Invasive Blood Pressure Accessories
These cuffs and tubings are designed to have special
protection against electric shocks (particularly regarding
allowable leakage currents), and are defibrillator proof.
LNCS PdtxPediatric Finger
989803148241
Sensor
LNCS INF-LInfant Toe Sensor989803148251
LNCS NEO-LNeo Foot Sensor
989803148271
or Adult Finger
Sensor
LNCS NEOPT-LNeo Pre-Term
989803148261
Sensitive Skin
Adhesive Sensors
The Philips M8102A uses Masimo certified pulse
oximetry for reduced noise and low perfusion
performance with Masimo Sensors under the
Masimo NR&LP protocol available from Masimo.
Multi-Patient Comfort Cuffs and Disposable Cuffs
Patient CategoryDisposable cuff Reusable cuff
Adult (Thigh)M1879AM1576A
Large AdultM1878AM1575A
AdultM1877AM1574A
Small AdultM1876AM1573A
PediatricM1875AM1572A
InfantM1874AM1571A
Tubing: Use M1598B or M1599B
Reusable Cuff KitsPart No.
Infant, pediatric, small adult, adultM1577A
23
Page 24
Reusable Cuff KitsPart No.
Small adult, adult, large adult, thighM1578A
Infant, pediatric, small adult, adult, large
M1579A
adult, thigh
Adult/Pediatric Soft Single Patient Single-Hose Disposable
Cuffs
Patient
Category
Limb
Circumference
Bladder
Width
Disposable
cuff
Part No.
Adult/Pediatric Antimicrobial Coated
Reusable cuffs
Cuff Size
(color)
Infant
(orange)
Pediatric
(green)
Small Adult
(royal blue)
Adult
(navy blue)
Adult
X-long
Circumference
(cm)
Bladder
Width
9.0 - 14.85.4 cm
2.1 inches
13.8 - 21.58.0 cm
3.1 inches
20.5 - 28.510.6 cm
4.2 inches
27.5 - 36.513.5 cm
5.3 inches
27.5 - 36.513.5 cm
5.3 inches
(navy blue)
Large Adult
(burgundy)
Large Adult
X-long
35.5 - 46.017.0 cm
6.7 inches
35.5 - 46.017.0 cm
6.7 inches
(burgundy)
Thigh
(grey)
45 - 56.521.0 cm
8.3 inches
Single-
Hose Part
No.
M4552A
M4553A
M4554A
M4555A
M4556A
M4557A
M4558A
M4559A
Large Adult
35.5-46.0 cm16.4 cmM4578A
X-long
Large Adult35.5-46.0 cm16.4 cmM4577A
Adult
27.5 to 36.5 cm16.4 cmM4576A
X-long
Adult27.5-36.5 cm13.1 cmM4575A
Small Adult20.5-28.5 cm10.4 cmM4574A
Pediatric15.0-21.5 cm8.0 cmM4573A
Infant9.0-15.0 cm5.6 cmM4572A
Tubing: Use M1598B or M1599B
Neonatal/Infant Cuffs (Disposable, non-sterile)
Cuffs
Limb
Circumference
Bladder
Width
Part No.
Size 13.1 to 5.7 cm2.2 cmM1866A
Size 24.3 to 8.0 cm2.8 cmM1868A
Size 35.8 to 10.9 cm3.9 cmM1870A
Size 47.1 to 13.1 cm4.7 cmM1872A
Tubing: Use M1596B or M1597B
Tubing: Use M1598B or M1599B
Adult/Pediatric Soft Single Patient Single-Hose Disposable
Cuffs
Patient
Category
Limb
Circumference
Bladder
Width
Disposable
cuff
Part No.
Adult (Thigh)45.0-56.5 cm20.4 cmM4579A
24
Cuff Tubing
Adult1.5 m /4.9’M1598B
3.0 m/9.8’M1599B
Neonatal1.5 m /4.9’M1596B
3.0 m/9.8’M1597B
Page 25
Temperature Accessories
Pressure Transducers and AccessoriesPart No.
Temperature ProbesPart No.
Reusable
General purpose probe21075A
Small flexible vinyl probe
(Infant/Pediatric)
Attachable surface probe21078A
Disposable
General purpose probeM1837A
Skin probe21091A
Esophageal/Stethoscope Probe
(12 French)
Esophageal/Stethoscope Probe
(18 French)
Esophageal/Stethoscope Probe
(24 French)
Foley Catheter Probe
(12 French)
21076A
21093A
21094A
21095A
M2255A
Transducer holder for CPJ840J6
(pack of 4)
IV pole mount for CPJ840J6CPJ84447
Disposable (EU/EFTA only. Not available in USA)
Single channel disposable sensor kit (20)M1567A
Dual channel disposable sensor kit (20)M1568A
Transducer holder for M1567/8AM2271A
IV pole mount for M1567/8AM2272C
Adapter cable for disposable sensor kit, 3.0m,
for M1567/8A
CPJ84046
M1634A
Foley Catheter Probe
(16 French)
Foley Catheter Probe
(18 French)
Adapter cable 1.5m/4.9’21082B
Adapter cable 3.0m/9.8’21082A
21096A
21097A
PRESS Accessories
These transducers and accessories are designed to have
special protection against electric shocks (particularly
regarding allowable leakage currents), and are defibrillator proof.
Philips Medical Systems Nederland B.V. reserves the right to make changes in specifications and/or to discontinue any product at any time without notice
or obligation and will not be liable for any consequences resulting from the use of this publication.
Printed in The Netherlands.
4522 962 26701/862 * DEC 2007
0366
M8102A complies with the requirements of the
Council Directive 93/42/EEC of 14 June 1993
(Medical Device Directive).
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.