Philips M8102A User Manual

IntelliVue MP2 Patient Monitor
Philips M8102A Technical Data Sheet
The IntelliVue MP2 portable patient monitor is compact in size, ergonomic, and modular in design. It provides an easy-to-use touchscreen user interface, is highly customizable and shares a technological platform with the Philips IntelliVue MP5-MP90 patient monitors.
The IntelliVue series offers a complete monitoring solution that is flexible and modular, designed to suit a broad spectrum of monitoring needs.
Measurement Features
• Compact, rugged, lightweight monitor with built in measurements
• ECG monitoring using any combination of three to 10 electrodes.
• 12-lead ECG monitoring with five electrodes using
the EASI method or with 10 electrodes using the conventional method.
• Multi-lead arrhythmia and ST segment analysis at the bedside on all available leads.
• Mainstream or Sidestream CO
• FAST SpO2 for accurate performance even with low perfusion.
• Invasive Pressure and Temperature measurement
• The monitor can operate using battery power for up to 3 hours with basic monitoring configuration to let you safely and easily monitor patients during in-hospital transfer. AC power is provided by an external power supply.
2
Usability Features
• Touchscreen and hardkeys as input device.
• Intuitive user interface.
PAD
• Simple menu hierarchy gives fast access to all basic monitoring tasks.
• Patient data management with tabular and graphic trends.
• Settings “Profiles” for rapid case turnover.
• Patented automatic alarm limits help clinicians provide care more efficiently.
• 3.5" TFT flat panel display with QVGA (320 x 240) resolution, wide viewing angle, large numerics, permanently visible alarm limits, and up to three real-time waves.
• Capable of functioning in a wireless infrastructure (IIT)
Intended Use
The monitor is intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment and during patient transport inside and outside of hospitals. The monitor is intended for use by health care professionals.
The monitor is only for use on one patient at a time. It is not intended for home use. Not a therapeutic device.
Rx only: U.S. Federal Law restricts this device to sale by or on the order of a physician.
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11).
EMC Environment:
Main Components
Monitor
The monitor has a color LCD TFT display with a wide viewing angle, providing high resolution waveform and data presentation.
The display, processing unit and measurements are integrated into one device. An external power supply provides power for the monitor.
User Interface
The user interface is designed for fast and intuitive operation. The color graphical user interface ensures that clinicians quickly feel at ease using the monitor.
Configurable SmartKeys with intuitive icons allow monitoring tasks to be performed quickly and easily, directly on the monitor screen.
Waves and numerics are color-coded.
The monitor displays up to three measurement waves simultaneously. For 12-lead ECG monitoring it can display 12 real-time ECG waves, with a rhythm strip and all ST values.
The MP2 monitor is supplied with a resistive touchscreen.
Simulated Keyboard
If alpha or numeric data entry is required, for example to enter patient demographics, an on-screen keyboard will automatically appear on the screen.
The monitor is suitable for use in all establishments incl. those directly connected to the public low-voltage supply network that supplies buildings for domestic purposes. When used with the M2741A Sidestream CO2 sensor, the monitor can only be used in hospital environments.
Upgradability
The MP2 monitor allows new capabilities to be added in the future as your monitoring requirements evolve. This upgradability gives the security of knowing that the monitors can be enhanced and updated as practices and technologies advance, and it protects long-term investments.
2
Mounting
The mounting options available enable flexible, space saving placement of the monitors for an ergonomic work space. The monitor is shipped with a low cost mounting plate if not specified otherwise.
Application Features
Critical and Cardiac Care Features
• The monitor performs multi-lead arrhythmia detection analysis
on the patient’s ECG waveform at the bedside. It analyzes for ventricular arrhythmias, calculates heart rate, and generates alarms, including asystole, bradycardia, and ventricular fibrillation.
• Up to 12 leads of ST segment analysis can be performed on adult
patients at the bedside, measuring ST segment elevation and depression and generating alarms and events. The user can trend ST changes, set high and low alarm limits, and set both ST and isoelectric measurement points. Using ST Snippets, one-second wave segments can be compared with a baseline segment for each measured ST lead.
• optional ST Map application shows ST changes over time in two
multi-axis spider diagrams.
QT/QTc interval monitoring provides the measured QT interval, the calculated heart-rate corrected QTc value and a
'QTc value,
which tracks variation in the QT interval in relation to a baseline value.
• optional 12-lead ECG data can be measured, using either the EASI placement method with five standard electrodes or conventional electrode placement with 10 electrodes.
1
• 12 realtime ECG waveforms can be displayed simultaneously.
FAST-SpO2, using Fourier Artifact Suppression Technology, performs accurately even in cases with low perfusion.
• Choice of sidestream or mainstream CO
monitoring for high
2
quality measurements with intubated and non-intubated patients.
- Vital Signs
- Graphic Trends
- Realtime Wave Reports
Report templates can be defined in advance, enabling print-outs
tailored to each hospital’s specific requirements to be started
quickly. Reports can be printed on centrally-connected printers or
via the IntelliVue PC Printing Solution, and they can be initiated
manually or automatically at user-defined intervals.
• The IntelliVue PC Printing Solution allows printing of reports, waveform captures and trends from the MP2 to a standard off-the shelf printer or to an electronic file.
Alarms
Ease of Use
Screen layouts are easily adjustable, allowing flexible display of
measurement information.
• Temperature, height, and weight can be configured either in metric
or imperial units. Pressure measurements can be displayed in kPa or mmHg. Gases can be displayed in kPa, mmHg.
Trends
• The trend database stores patient data from up to 30
measurement numerics. The measurement information can be sampled every 12 seconds, one minute, or five minutes, and stored for a period ranging from four to 48 hours.
Horizon Trends show the deviation from a stored baseline.
Transport Features
• The monitor’s portable design means it can be used for in and out-of­hospital transport: a basic monitor weighs 1.5 kg.
• The monitor can operate using battery power for up to 3 hours, to let you safely and easily monitor patients during procedures or in­hospital transfer.
• Specially-designed mounting solutions let you quickly disconnect the monitor for transport and reconnect to the mount after transport.
• The Universal Admit, Discharge and Transfer (ADT) feature means that all ADT information is shared between the networked monitor and the Information Center. Information need only be entered once.
Patient Data Documentation
• An extensive range of Patient Reports can be printed:
- 12-lead ECG Reports
- Alarm Limit Reports
The alarm system can be configured to present either the traditional HP/Agilent/Philips alarm sounds or sounds compliant with the draft ISO/IEC 9703-2 Standard.
Alarm limits are permanently visible on the main screen. The Alarm Limits page provides a graphic depiction of alarm limits in relation to the currently monitored measurement values and lets you adjust alarm limits. It also lets you preview wide and narrow automatic alarm limits before you apply them.
When an alarm limit is exceeded, it is signalled by the monitor in the following ways:
• an alarm tone sounds, graded according to severity
• an alarm message is shown on the screen, color-coded according to severity
• the numeric of the alarming measurement flashes on the screen
• alarm lamps flash for red and yellow alarms and are illuminated for technical INOPs
If the monitor is connected via a network to a central monitoring station, alarming is simultaneous at the monitor and at the Information Center.
Alarms are graded and prioritized according to severity:
Red Alarms*** identify a potentially life threatening situation for a patient .
Yellow Alarms** indicate conditions violating preset vital signs limits.
Technical Alarms (INOPS)
are triggered by signal quality
problems, equipment malfunction or equipment disconnect.
The Silence/Pause Alarms function (equivalent to Silence/Suspend
with previous monitor generations) allows the user to switch off
alarm tones with one touch.
1.EASI-derived 12-lead ECGs and their measurements are approximations to conventional 12-lead ECGs. As the 12-lead ECG derived with EASI is not exactly identical to the 12­lead conventional ECG obtained from an electrocardiograph, it should not be used for diagnostic purposes.
All alarms can be paused for a period of one, two, three, five, or 10
minutes or turned off indefinitely.
3
Alarm strip recordings are available on a centrally-connected
recorder or via the IntelliVue PC Printing Solution.
Patented automatic alarm limits automatically adapt the alarm limits
to the patient’s currently measured vital signs within a safe margin
defined individually for each patient.
Visual and/or audible latching and non-latching alarm handling is
available.
Profiles
Profiles are predefined configuration settings for Screens, measurement settings, and monitor settings. Each Profile can be designed for a specific application area and patient category, for example OR adult, or ICU neonatal. Profiles enable a quick reaction to patient and care location changes: activating a Profile with a particular patient category (Adult, Pediatric or Neonatal) automatically applies suitable alarm and safety limits and saves time usually spent carrying out a complete set-up procedure.
Profiles can be created directly on the monitor or remotely on a personal computer and transferred to the monitor using the IntelliVue Support Tool. A selection of Profiles for common monitoring situations is provided with the monitor. These profiles can be changed, added to, renamed, or deleted.
Network Interface
The network interface provides the system with networking capability via a wired or wireless network connection.
Wireless Network (optional)
The monitor can function within a telemetry infrastructure compatible with the Philips Cellular Telemetry System (CTS) in the WMTS and ISM bands. Additional components are required to complete the system. Please refer to the M3185A IntelliVue Clinical Network Technical Data Sheet for further information.
Monitor Specifications
Safety Specifications
The monitor complies with the Medical Device Directive 93/42/EEC
) and with IEC 60601-1:1988 + A1:1991 + A2:1995; EN60601-
(CE
0366
1:1990 + A1:1993 + A2:1995; UL 60601-1:2003; CAN/CSA C22.2#601.1-M90; JIS T 0601-1:1999; IEC 60601-1-1:2000; EN 60601­1-1:2001.
All applied parts are Type CF unless otherwise specified. They are protected against damage from defibrillation and electrosurgery.
Optional Networking Capabilities
The monitor can operate as part of a wired or wireless hospital network system, using the Philips IntelliVue Clinical Network interface.
Service Features
• The Support Tool helps technical personnel to
- carry out configuration, upgrades and troubleshooting via the
network, or on an individual monitor
- share configuration settings between monitors
- back up the monitor settings.
• A password-protected Service Mode ensures that only trained staff can access service tests and tasks.
• The Configuration Mode is password-protected and allows trained users to customize the monitor configuration.
Device Connections
The monitor can be connected to:
• an Information Center (for example M3150B)
•a PC
• MMS Extensions (M3012A, M3014A, M3015A, M3016A)
1
The possibility of hazards arising from software errors was minimized in compliance with and IEC 60601-1-4:1996 + A1:1999.
The monitor complies with the EMC standards IEC 60601-1-2:2001; EN 60601-1-2:2001
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada.
The MP2 patient monitor can be used in a transport environment such as road ambulance, airplane or helicopter, except when used with the M2741A Sidestream CO2 sensor. For this purpose, the monitor fulfills the following additional mechanical, EMC and environmental requirements:
Shock Tests according to IEC TR 60721-4-7, Class 7M3. Test procedure according to IEC/EN 60068-2-27 (peak acceleration up to 100g).
Random Vibration according to IEC TR 60721-4-7, Class 7M3. Test procedure according to IEC/EN 60068-2-64 (RMS acceleration 5g).
Sinusoidal Vibration according to IEC TR 60721-4-7, Class 7M3. Test procedure according to IEC/EN 60068-2-6 (acceleration up to amplitude 2g).
1.The MMS Extensions will only function when the monitor is connected to the external power.
ISO 14971:2000, EN60601-1-4:1996 + A1:1999
4
Bump Test according to IEC/EN60068-2-29 (peak acceleration 15g, 1000 bumps).
Free Fall Test according to EN1789 (covers also IEC TR 60721-4-7 and Class 7M3). Test procedure according to EN 60068-2-32 (height
0.75 m).
• Specification for degrees of protection provided by enclosures according to IEC/EN 60529: IP 32
EN 1789 +A1:2003 Medical vehicles and their equipment - Road ambulances (chapter 6 - Medical Devices).
Radiated susceptibility 20 V/m according to EN ISO 9919 (SpO2) and EN ISO 21647 (CO2).
Altitude Range from -500 to 3000 m operating and -500 to 4600 m storage and transportation.
• Extended radiated susceptibility tests
The MP2 patient monitor with its out-of-hospital parameter set provides a general immunity level of 20 V/m with only few restrictions. Details are as listed below:
- GSM 900: Immunity at 900 MHz (uplink mobile phone), 20V/m,
duty cycle 1:8
- GSM 1800: Immunity at 1800 MHz (uplink mobile phone), 20V/m,
duty cycle 1:8.
- DECT: Immunity at 1800 MHz (digital cordless phone), 20V/m,
duty cycle 1:24
- AM: 1 kHz Immunity from 80 MHz to 2.5 GHz (any radio
communication unit, broadcasting and TV transmitter), 20V/m,
modulation factor 80%. (ECG: 20 V/m except 0.8-1.2 GHz where
it is 10V/m)
Operating ambient temperature testing over the range from 0 to 40 °C (32 to 104 °F).
Operating ambient humidity testing up to 95% RH at 40 °C (104 °F), non condensing.
Physical Specifications
Environmental Specifications
Item Condition Range
Temperature Range
Temperature Range when charging the battery
Humidity Range Operating 15% to 95% Relative
Altitude Range Operating -500 m to 3000 m
Ingress Protection
Operating
0 to 40oC
(32 to 104 oF)
Storage (incl. Transport)
Operating
-20 to 60oC
(-4 to 140
0 to 35oC
(32 to 95
o
o
F)
F)
Humidity (RH) (non condensing)
Storage and Transport
5% to 95% Relative Humidity (RH)
(10000 ft)
Storage and Transport
-500 m to 4600 m
(15000 ft)
a
Monitor IP32 (protected against
the ingress of solid foreign objects 2.5 mm in diameter or larger, and the ingress of water when the water is dripping vertically and the monitor is tilted up to 15°).
Product Max Weight W x H x D
M8102A IntelliVue MP2 (without
1.5 kg
(3.3 lb)
< 188 x 99 x 86 mm (7.4 x 3.9 x 3.4 in)
handle and options)
External Power Supply (M8023A)
IP31(protected against the ingress of solid foreign objects 2.5 mm in diameter or larger, and the ingress of water when the water is dripping vertically) when rested on its rubber feet on a flat, level surface.
IP32 when mounted with the connectors facing downwards
a. sufficient for flight altitudes up to 12,000 m with pressurized cabins
5
Performance Specifications
Monitor Performance Specifications
Alarm Signal System delay less than 3 seconds
Monitor Performance Specifications
Power Specifications
Battery Specifications
Indicators Alarms Off red LED
Power consumption
Line Voltage 100 to 240 V ~
Current 1.3 to 0.7A
Frequency 50/60 Hz
Operating Time (with new, fully charged battery at 25 °C)
Charge Time When MP2 is off: 2 h
Alarms red/yellow/cyan LED
On/Standby/Error green/red LED
< 40W average, <65W peak
Basic monitor configuration: 3 hours
When MP2 is in use and connected to MP20/30/ 40/50/60/70/80/90 without extensions: 12 h approx.
When MP2 is in use and connected to the external power
Pause duration 1,2,3 minutes or infinite,
depending on configuration
Extended alarm pause
Review Alarms
Real Time Clock
Buffered Memory
Restart time: After power interruption, an ECG wave will be shown on the display after 30 seconds maximum.
Information: all alarms / inops, main alarms on/off, alarms acknowledged and time of occurrence
capacity 500 items
Range: from: January 1, 1997, 00:00 to: December 31, 2080, 23:59
Accuracy: < 4 seconds per day (typically)
Hold Time: infinite if powered by host monitor or external power supply; otherwise at least 48 hours
Contents: Active settings, trends, patient data, realtime reports, review alarms
Hold Time: infinite if powered by external power supply; otherwise at least 48 hours
5 or 10 minutes
M8023A External Power Supply Performance Specifications
AC Power green LED
Battery yellow (charging)/red
blinking (empty) LED
External Power green LED
Sounds Audible feedback for user input. Prompt tone.
QRS tones, or SpO
different alarm sounds.
Trends:
12, or 16 numerics @ 12 sec, 1 minute, 5 minute resolution. Multiple choices of number of numerics, resolution and duration depending on trend option and application area.
6
modulation tone. Four
2
M8023A External Power Supply Performance
Specifications
Power Specifications
Indicators AC Power green LED
Power Consumption
Line Voltage 100 to 240 V ~
Current 0.7 to 0.4 A
Frequency 50/60 Hz ~
< 12 W average < 30 W peak
Interface Specifications
MP2 (M8102A) Interface Specifications
Measurement Link (MSL)
Connectors Female ODU (Proprietary)
Power 30 V to 60 V input
Power Sync RS-422 compliant input
78.125 kHz (typical
LAN signals IEEE 802.3 10-Base-T
complaint
Serial signals RS-422 compliant
Local signals Provided for connecting MMS
extensions
ECG Sync Pulse Output
Cable Detection
Yes
Marker In No
Wave
No
Output
Connector Binder Series 709/719
Output Levels
Output low <0.8V @ I = ­4mA
Output high >2.4 V @ I= 4mA
Isolation None
M8023A External Power Supply Interface Specifications
Measure­ment Link (MSL)
Connectors Male ODU (Proprietary)
Power 48 V output
Power Sync. RS-422 compliant output
78.125 kHz (typical)
LAN signals IEEE 802.3 10-Base-T complaint
Serial signals
RS-422 compliant output
78.125 kHz (typical)
Local signals Not connected
Display Specifications
Integrated QVGA Display
Sweep
6.25, 12.5, 25 and 50 mm/s;
Speeds
Resolution 320 x 240
Refresh
60 Hz
frequency
Useful
72x54mm (2.8x2.1in)
screen
Pixel size 0.22 x 0.22 mm
Pulse Width 100 +/- 10 ms (high)
Delay from R-wave peak
20 ms maximum per AAMI
EC13 to start of pulse
Minimum
0.5 V required R­wave amplitude
MP2 (M8102A) Compatible Devices
IntelliVue Instrument Telemetry Wireless Network
(USA only)
Internal WMTS Adapter
Technology compatible with Philips
Cellular Telemetry System (CTS), cellular infrastructure
Frequency Band
WMTS, 1395-1400 MHz and 1427-1432 MHz
IntelliVue Instrument Telemetry Wireless Network
(except USA)
7
MP2 (M8102A) Compatible Devices
Measurement Specifications
Internal ISM Adapter
M4607A Battery Specifications
Physical Specifications
W x D x H 66 mm (2.36 in) x 80 mm (3.15 in) x
Weight 160 g ±5%
Performance Specifications
Nominal Voltage 10.8 Volt
Rated Capacity at discharge C/5
Environmental Specifications
Temperature Range Discharge 0 to 60qC (32 to 122qF)
Humidity Range Operating: 15 % to 95 % Relative
Battery Type Lithium Ion Mangan, 10.8 V,
Safety complies with UL 1642 (UL
Technology compatible with Philips
Cellular Telemetry System (CTS), cellular infrastructure
Frequency Band
2.4 GHz ISM
20 mm (0.79 in)
1000 mAh (typical)
Charge 0 to 60qC (32 to 122qF) Storage and Transportation: -20 to 65qC (-4 to 140qF)
Humidity (RH) Storage and Transportation: 5 % to 95 % Relative Humidity (RH)
1000 mAh,
Recognized)
ECG/Arrhythmia/ST/QT
Complies with IEC 60601-2-25:1993 + A1:1999 /EN60601-2-25:1995 + A1:1999, IEC 60601-2-27:2005/EN60601-2-27:2006, IEC 60601-2­51:2003 /EN 60601-2-51:2003 and AAMI EC11/EC13:1991/2002.
ECG/Arrhythmia/ST Performance Specifications
Cardiotach Range Adult/pedi:
15 to 300 bpm Neo range: 15 to 350 bpm
Accuracy ±1% of range
Resolution 1 bpm
Sensitivity t200 μV
PVC Rate Range 0 to 300 bpm
Resolution 1 bpm
ST Numeric Range -20 to +20 mm
Accuracy ±0.5 mm or 15%, whichever is
greater
Resolution 0.1 mm
QT Numeric Range 200 to 800 ms
Accuracy ±30 ms
Resolution 8ms
QTc Numeric Range 200 to 800 ms
Resolution 1ms
'QTc Numeric Range -600 to +600 ms
Resolution 1ms
peak
Electromagnetic Compatibility (EMC)
Communication Standard complies with the SMBus specification
8
complies with the requirements for FCC Type B computing Device, and EN 61000-4-2 and EN 61000-3-2
v1.1
QT-HR Numeric
Range ­adult
Range ­pediatric and neonatal
15 to 300 bpm
15 to 350 bpm
ECG/Arrhythmia/ST Performance Specifications
Sinus and SV Rhythm Ranges
Brady Adult: 15 to 60 bpm
Pedi: 15 to 80 bpm Neo: 15 to 90 bpm
Normal Adult:60 to 100 bpm
Pedi: 80 to 160 bpm Neo: 90 to 180 bpm
Tachy Adult: > 100 bpm
Pedi: >160 bpm Neo: >180 bpm
Bandwidth Diagnostic
Adult/neo/pedi: 0.05 to 150Hz
Mode
Extended
Neo/pedi: 0.5 to 150Hz Monitoring Mode
Monitoring Mode
Adult: 0.5 to 40Hz
Neo/pedi: 0.5 to 55Hz
Filter Mode Adult/neo/pedi: 0.5 to 20Hz
Differential Input Impedance >2M: RA-LL leads (Resp)
>5M: at all other leads (at 10Hz
including patient cable)
Common Mode Rejection Ratio
Diagnostic mode: >86 dB (with a
51 k:/47 nF imbalance).
Filter mode: >106 dB (with a 51
k:/47 nF imbalance).
Electrode Offset Potential
±500mV
Tolerance
Auxiliary Current (Leads off Detection)
Active electrode: <100 nA
Reference electrode: <900 nA
Input Signal Range ±5 mV
ECG/
Arrhythmia/
ST Alarm
Range Adjustment
Specifications
HR 15 to 300 bpm
maximum delay: 10 seconds according to AAMI EC 13-1992 standard
Adult:1 bpm steps (15 to 40 bpm) 5 bpm steps (40 to 300 bpm) Pedi/Neo:1 bpm steps (15 to 50 bpm) 5 bpm steps (50 to 300 bpm)
Extreme Tachy Difference to high
5 bpm steps
limit 0 to 50 bpm
Clamping at 150 to
5 bpm steps
300 bpm
Extreme Brady Difference to low
5 bpm steps
limit 0 to 50 bpm
Clamping at 15 to
5 bpm steps
100 bpm
Run PVCs 2 PVCs Not adjustable by user
PVCs Rate 1 to 99 PVCs/minute 1 PVC
Vent Tach HR 20 to 300 bpm 5 bpm
Vent Tach Run 3 to 99 PVCs/minute 1 PVC
Vent Rhythm
2 to 99 PVCs/minute 1 PVC
Run
SVT HR 120 to 300 bpm 5 bpm
SVT Run 3 to 99 SV beats 1 SV beat
ST High -19.8 to +20 mm 0.2 mm
ST Low -20 to +19.8 mm 0.2 mm
QTc High 200 ms to 800 ms 10 ms steps
'QTc High 30 ms to 200 ms 10 ms steps
9
Loading...
+ 19 hidden pages