Philips EC10, EC5 User manual

Page 1
IntelliBridge EC10 & EC5
Technical Data Sheet
Features
• Single Interface Port for connecting medical
devices.
• Together with the IntelliBridge EC5 Device ID
module, the IntelliBridge EC10 module and EC10
Interface Board support automatic device
• Pre-programmed RS-232 or LAN connection
available for certain external devices.
• Supports up to four waveforms.
• Supports up to 80 parameters.
• Supports external devices with multiple channels
(for example, IV-pump systems).
• Easily upgradeable to interface additional
supported devices as new drivers become available.
• Supports the transfer of data to Electronic Medical
Records (EMRs) and Patient Data Management
Systems through the IntelliVue Network and
IntelliVue Information Center, for example,
settings, status and numbers, from connected
medical devices.
• Depending on the patient monitor, up to four
modules can be used at the same time, to allow
interfacing of numerous devices. Combinations of
IntelliBridge EC10 and VueLink modules is
supported.
• Provides comprehensive alarm identification.
• Device-specific settings are stored inside the
module, so that it can be reconnected after
transport without repeating the setup process.
Page 2
• Can be used with the following patient monitor families1:
– IntelliVue MP40–MP90 and MX600–MX800
– IntelliVue MX400, MX450, MX500 and MX550
Description
The IntelliBridge EC10 module/EC10 IF-Board2 is designed to be used
in an OR or ICU environment by trained healthcare professionals.
Together with an EC5 ID module it provides external device interface
capability to external devices at the bedside which have a serial RS-232
and/or LAN output. The EC10 modules/EC10 IF-Boards are ‘universal’
that is, they can be configured for various external devices3 (for
example, Anesthesia, Ventilators, Monitors, Infusion pumps). The
EC5 ID modules3 are specific to an external device.
Displayed Information
The following key data from the external device can be displayed
alongside the measurement information sourced from the monitor:
• Device Type (Auxiliary, Ventilator, Gas Analyzer, Anesthesia Machine,
Infusion Pumps, etc.)
•Device Name.
• Waves received by the IntelliBridge EC10 module/EC10 IF-Board.
• Numerics associated with the current wave.
• Alarm Status (ignored, accepted, or not available).
• Current device mode, if available (for example, ventilation mode:
CMV).
• Selected alphanumeric information received by the IntelliBridge EC10
module/EC10 IF-Board.
• Wave scale information: Several choices of wave scale are available for
most waves.
Connecting your Device to IntelliBridge EC10
One external device can be connected to the IntelliBridge EC10
module/EC10 IF-Board at a time. External devices are connected via
the IntelliBridge EC5 ID module. The IntelliBridge EC5 ID module
provides unique identification information to enable automatic device
detection and driver selection.
Up to four IntelliBridge EC10 modules can be used simultaneously with
the IntelliVue patient monitors MP60–MP90 and MX800.
Up to two IntelliBridge EC10 modules can be used simultaneously with
the IntelliVue patient monitors MP40–MP50 and MX500, MX550,
MX600, MX700.
Up to two EC10 IF-Boards can be used simultaneously with the
IntelliVue patient monitors MX400, MX450, MX500 and MX550.
1
patient monitors with software release G.0 or higher for devices using a dedicated driver, and software release H.0 or higher for devices using the OpenInterface driver.
2
The MX400, MX450, MX500 and MX550 can be optionally fitted with an IntelliBridge EC10 Interface Board (Option # J32).
3
See Ordering Information for a list of supported devices.
The IntelliBridge EC10 module and IntelliBridge EC10 Interface Board
can be used simultaneously with the MX500 and MX550 patient
monitors, but only two at any one time. For example; 2 x EC10
module, 2 x EC10 IF-Board, or 1 EC10 module + 1 EC10 IF-Board.
Installation Policy
Authorized technical personnel are responsible for the set-up and
configuration of the IntelliBridge EC10 module/EC10 IF-Board and all
related equipment.
Configuration or set-up of external devices furnished by other
manufacturers must be carried out by the local site technical
personnel, or by representatives of the external device manufacturer.
Repair Policy
Authorized Philips technical personnel are responsible for repairing
faults in the IntelliBridge EC10 module/EC10 IF-Board, or any related
equipment. Philips is not responsible for repairing faults in external
devices.
Upgrade Policy
As additional drivers become available for connection to external
devices, upgrades can be purchased to enable their connection.
Specifications
IntelliBridge EC10 Interface Module
Size (W x H x D): 36 x 102 x 111 mm (1.4 x 4.0 x 4.4 in) Weight: 0.3 kg (0.7 lb)
IntelliBridge EC10 Interface Board
The IntelliBridge EC10 IF-Board is an option (# J32) for the MX400,
MX450, MX500 and MX550 patient monitors. Refer to the relevant
patient monitor Technical Data Sheet for details, especially the
environmental specifications.
IntelliBridge EC5 Device ID Module
Size (W x H x D): 35 x 17 x 57 mm (1.4 x 0.7 x 2.2 in) ±5% Weight: 35 g (1.2 oz.)
Standards Compliance
US Federal Law restricts this device to sale by (or on the order of) a
physician. The IntelliBridge EC10 module/EC10 IF-Board and
IntelliBridge EC5 ID Module are in conformity in conjunction with the
Philips IntelliVue patient monitors with the essential requirements of
the European Medical Devices Directive 93/42/EEC and the following
major standards:
2
Page 3
Safety
• EN 60601-1:1990+A1:1993+A2:1995/IEC 60601-
1:1988+A1:1991+A2:1995
• EN 60601-1-1:2001/IEC 60601-1-1:2000
• UL 60601-1:2003
• CAN/CSA C22.2#601.1-M90:1990
• AS/NZS 3200.1.0-1998
• EN/IEC 60601-2-49:2001
Protected against damage from defibrillation and electrosurgery.
Electromagnetic Compatibility
EN/IEC 60601-1-2:2001
Wave Channel Size
Waveform size and positioning depends on the configuration of the
system in which the module is used.
Ordering Information
Ordering Structure
EC10 IntelliBridge
a
Module
RS-232/LAN
865115 #A01
Driver Options See IntelliBridge Device Library
EC5 Device ID Module 865114
ID Module Option See IntelliBridge Device Library
Connection Cables
b
Connection Cable 1.5 m #L01
Connection Cable 3.0 m #L02
Connection Cable 10.0 m #L03
a Requires relevant EC5 Device ID Module. Up to ten Driver options can be ordered
with one EC10 Module.
b A minimum of one of the Connection Cable options must be ordered.
Environmental Specifications
a
Temperature Range
Operating -5–50°C (+23–122°F)
Storage -40–70°C (-40–158°F)
Humidity Range
Operating 15–95% RH max. @ 45°C (113°F)
Storage Up to 90% RH max. @ 65°C (149°F)
Altitude Range
Operating Up to 4,600 m (-1500–15,000 ft.)
Storage Up to 15,300 m (-1500–50,000 ft.)
a As the IntelliBridge EC10 Interface Board is built into the MX400, MX450, MX500
and MX550 Patient Monitors, refer to the relevant patient monitor for the environmental specifications.
Supported Devices (Dedicated device drivers)
For a list and details of the external device interfaces available in your
country, please consult your Sales Representative.
Supported Devices (OpenInterface)
Many devices provide interfaces for Philips IntelliVue Patient Monitors
based on the Philips OpenInterface protocol. For a list and details of
devices that support the OpenInterface (Option 101), please consult
your Sales Representative.
IntelliBridge Device Library
For availability check with your Philips sales representative.
Respiratory Care
EC10
Option
Puritan Bennett 840 201 201
Dräger Babylog VN500 213 213
Dräger Carina 218 218
Dräger Evita 4 203 203
Dräger Evita V500 206 206
Dräger Savina 300 207 207
Dräger Evita XL 202 202
GE Engström Carestation 217 217
Maquet SERVO-i 209 209
Maquet SERVO-s 210 210
Maquet SERVO-U 224 224
Maquet SERVO-n 225 225
Philips Respironics V60 101 105
Respironics (See, Philips Respironics)
EC5
Option
Anesthesia Delivery
EC10
Option
EC5
Option
Dräger Apollo 402 402
Dräger Fabius 404 404
Dräger Pallas 403 403
Dräger Perseus A500 412 412
Dräger Primus 401 401
Dräger Zeus IE 420 420
GE Aisys 407 407
3
Page 4
GE Avance 406 406
GE Avance CS2 421 421
GE Aisys CS2 422 422
GE Aespire View 419 419
GE Aespire 7100 414 414
GE Aespire 7900 418 418
GE Aestiva 7100 408 408
GE Aestiva 7900 417 417
Maquet FLOW-i 411 411
Philips IntelliSave AX700 101 104
Gas Analyzers
TcG10 301
a Can only be ordered together with an IntelliVue TcG10 (865298)
Open Interface supported
devices
EC10
Option
a
EC10
Option
EC5
Option
a
301
EC5
Option
Open Interface 101 101 – 105
Infusion Devices
EC10
Option
EC5
Option
B.Braun Space 501 501
Fresenius-Kabi Link+ Agilia 512 512
Standalone Patient Monitoring
Systems
EC10
Option
EC5
Option
Philips Respironics NM3 101 102
Covidien BIS VISTA 808 808
Edwards EV1000 711 711
EdwardsVigilance 701 701
Edwards Vigilance II 702 702
Edwards Vigileo 703 703
Hemodialysis
EC10
Option
EC5
Option
Gambro Prismaflex 803 803
Infant Thermal Care
EC10
Option
EC5
Option
Dräger Babytherm 801 801
Dräger Caleo 810 810
New drivers are continuously being developed. For a complete and
updated Device Library, please contact your local Sales Representative.
EC10 Patient Monitor Compatibility
IntelliVue MP40, MP50,
up to 2 EC10 modules
IntelliVue MX500, MX550,
MX600, MX700
IntelliVue MP60, MP90,
up to 4 EC10 modules
IntelliVue MX800
IntelliVue MX400, MX450,
up to 2 EC10 Interface Boards
MX500, MX550,
Disclaimer
Philips makes every effort to ensure that signal names are maintained
between their monitors and any interfaced external devices. However,
in some cases, the wave and numeric labels used on the monitor may
be different from those used on the interfaced external device.
Philips will not be held responsible for any erroneous or inaccurate
data that is interfaced from external devices furnished by other
manufacturers.
If manufacturers other than Philips make changes to any IntelliBridge
EC10-supported device, Philips will try to ensure that interfacing to
that device remains possible. However, Philips reserves the right to
discontinue interfacing to any device.
Philips IntelliVue Demo Devices
EC10
Option
Philips Anesthesia Demo 400 400
Philips Ventilator Demo 200 200
4
EC5
Option
Page 5
5
Page 6
Philips Healthcare is part of
0366
Royal Philips
How to reach us
www.healthcare.philips.com
healthcare@philips.com
fax: +31 40 27 64 887
Asia
+852 2821 5888
Europe, Middle East, Africa
+49 7031 463 2254
Latin America
+55 11 2125 0744
North America
+1 425 487 7000
800 285 5585 (toll free, US only)
865115 & 865114 comply with the
requirements of the Council Directive 93/
42/EEC of 14 June 1993 (Medical Device
Directive).
Please visit www.philips.com/
© 2015 Koninklijke Philips N.V. All rights are reserved.
Philips Healthcare reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication.
Published in The Netherlands. 4522 991 09691 * May 2015