The monitor is intended for use with
Medical Equipment to display alpha,
numerical and graphical data. The
subject equipment Philips monitor is
powered by an external recognized AC/
DC adaptor. (IEC/EN60601-1).
1.1 Power adapter safety
Information
Power adapter
This adapter (Manufacture: Philips,
Model: PMP60-13-1-HJ-S) is a forming
part of the monitor.
The connection of external equipments
External equipment intended for
connection to signal input/output or
other connectors, shall comply with
relevant UL / IEC standard (e.g. UL
60950 for IT equipment, UL 60601-1-1
and ANSI/AAMI ES60601-1 / IEC 60601
series for systems – shall comply with
the standard IEC 60601-1-1, Safety
requirements for medical electrical
systems.
Disconnecting Device
The mains plug or appliance coupler
is used as the disconnect device, the
disconnect device shall remain readily
operable. Always completely disconnect
the power cord set from your product
whenever you are working or cleaning
on it. Do not make connections while
the power is on, because a sudden
rush of power can damage sensitive
electronic components.
Classification
• Degree of protection against the
ingress of water: IPX0
• Equipment not suitable for use
in the presence of a flammable
anesthetic mixture with air or with
oxygen or nitrous oxide. (Non AP or
APG Category)
• Mode of operation: Continuous
• Type of protection against electric
shock: Class I ME Equipment
• No Applied part.
Shutdown Procedure
We strongly recommended that
you should shut down the system
before you start to clean any single
components.
Please follow the steps below.
• Close all application programs
• Close operating software
• Turn off power switch
• Disconnect power cord set
• Remove all devices
Safety Symbol Description
The following safety symbols are the
further explanations for your reference.
With respect to electric shock, fire and
mechanical hazards only in accordance
with ANSI/AAMI ES60601-1, and CAN/
CSA C22.2 NO. 60601-1
Attention, consult ACCOMPANYING
DOCUMENTS.
Type of Current- AC
Dircet Current
European Community Approval,
The monitor complies with the 93/42/
EEC and 2007/47/EC and conforms
to the applicable following standards:
EN60601-1, EN 60601-1-2, EN 610003-2 and EN 61000-3-3.
TUV Type Testing Approval,
The monitor complies with the
EN60601-1 and IEC60601-1 of
European Standards.
Power "ON"
Power "OFF"
Medical Equipment With respect to
electric shock,, fire and mechanical
hazards only in accordance with ANSI/
AAMI ES 60601-1:2005, and CAN/CSA
C22.2 NO.60601-1:2008
1
1. Important
Note
• Caution: Use suitable mounting
apparatus to avoid risk of injury.
• Use a power cord that matches the
voltage of the power outlet, which
has been approved and complies
with the safety standards of your
particular country.
• Make sure user does not contact
SIP/SOPs and the patient at the
same time.
2
1. Important
1.2 EMC information
Guidance and manufacturer’s declaration – electromagnetic emissions – for all
EQUIPMENT and SYTEMS
The monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the monitor should assure that it is used in such an environment.
The monitor uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The monitor is suitable for use in all establishments, including
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
3
1. Important
Guidance and manufacturer’s declaration – electromagnetic immunity – for all
EQUIPMENT and SYTEMS:
The monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the monitor should assure that it is used in such an environment.
Immunity testIEC 60601 test levelCompliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
6 kV contact
8 kV air
2 kV for power
supply lines
1 kV for input/output
lines
1 kV line(s) to line(s)
2 kV line(s) to earth
<5 % UT (>95 % dip
in UT) for 0.5 cycle
40 % UT (60 % dip
in UT) for 5 cycles
70 % UT (30 % dip in
UT) for 25 cycles
<5 % UT (>95 % dip
in UT) for 5 sec
3 A/m3 A/mPower frequency magnetic fields
6 kV contact
8 kV air
2 kV for power
supply lines
1 kV for input/output
lines
1 kV line(s) to line(s)
2 kV line(s) to earth
<5 % UT (>95 % dip
in UT) for 0.5 cycle
40 % UT (60 % dip
in UT) for 5 cycles
70 % UT (30 % dip in
UT) for 25 cycles
<5 % UT (>95 % dip
in UT) for 5 sec
Electromagnetic environment –
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the monitor requires continued
operation during power mains
interruptions, it is recommended
that the monitor be powered
from an uninterruptible power
supply or a battery.
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.
Note
UT is the a.c. mains voltage prior to application of the test level.
4
1. Important
Guidance and manufacturer’s declaration – electromagnetic immunity – for
EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING:
The monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the monitor should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the monitor, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance:
√P
d = 1.2
Conducted
RF
IEC 610004-6
Radiated RF
IEC 610004-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2.5
GHz
3 Vrms
3 V/m
d = 1.2 √P 80 MHz to 800 MHz
d = 1.2 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey:
a. Should be less than the compliance level in
each frequency range.
b. Interference may occur in the vicinity of
equipment marked with the following
symbol:
Note
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
• Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength
in the location in which the monitor is used exceeds the applicable RF compliance
level above, the monitor should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the monitor.
• Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3
V/m.
5
1. Important
Recommended separation distances between portable and mobile RF
communications equipment and the EQUIPMENT or SYSTEM – for EQUIPMENT and
SYSTEMS that are not LIFE-SUPPORTING:
The monitor is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the monitor
can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the
monitor as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output
power of transmitter (W)
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223
Note
Separation distance according to frequency of transmitter (Meter)
150 kHz to 80 MHz
d = 1.2 √P
80 MHz to 800 MHz
d = 1.2 √P
800 MHz to 2.5 GHz
d = 2.3 √P
• For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
• At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
6
1. Important
1.3 Safety precautions and
maintenance
Warnings
• We strongly recommended that
you should shut down the system
before you start to clean any single
components.
• No modification of this equipment is
allowed.
• Use of controls, adjustments or
procedures other than those
specified in this documentation
may result in exposure to shock,
electrical hazards and/or mechanical
hazards.
• Read and follow these instructions
when connecting and using your
computer monitor.
Operation
• Please keep the monitor out of
direct sunlight, very strong bright
lights and away from any other heat
source. Lengthy exposure to this
type of environment may result in
discoloration and damage to the
monitor.
• Remove any object that could fall
into ventilation holes or prevent
proper cooling of the monitor’s
electronics.
• Do not block the ventilation holes
on the cabinet.
• When positioning the monitor, make
sure the power plug and outlet are
easily accessible.
• If turning off the monitor by
detaching the power cable or DC
power cord, wait for 6 seconds
before attaching the power cable
or DC power cord for normal
operation.
• Please use approved power cord
provided by Philips all the time. If
your power cord is missing, please
contact with your local service
center. (Please refer to Customer
Care Consumer Information Center)
• Do not subject the monitor to
severe vibration or high impact
conditions during operation.
• Do not knock or drop the monitor
during operation or transportation.
Maintenance
• To protect your monitor from
possible damage, do not put
excessive pressure on the monitor
panel. When moving your monitor,
grasp the frame to lift; do not lift
the monitor by placing your hand or
fingers on the monitor panel.
• Unplug the monitor if you are not
going to use it for an extensive
period of time.
• Unplug the monitor if you need to
clean it with a slightly damp cloth.
The screen may be wiped with a
dry cloth when the power is off.
However, never use organic solvent,
such as, alcohol, or ammonia-based
liquids to clean your monitor.
• To avoid the risk of shock or
permanent damage to the set, do
not expose the monitor to dust,
rain, water, or excessive moisture
environment.
• If your monitor gets wet, wipe it
with dry cloth as soon as possible.
• If foreign substance or water gets in
your monitor, please turn the power
off immediately and disconnect
the power cord. Then, remove the
foreign substance or water, and
send it to the maintenance center.
7
1. Important
• Do not store or use the monitor in
locations exposed to heat, direct
sunlight or extreme cold.
• In order to maintain the best
performance of your monitor and
use it for a longer lifetime, please
use the monitor in a location
that falls within the following
temperature and humidity ranges.
• Temperature: 10°C to 40°C
• Humidity: 30% to 75%
• Atmospheric pressure: 700 to
1060 hPa
Important information for Burn-in/
Ghost image
• Always activate a moving screen
saver program when you leave
your monitor unattended. Always
activate a periodic screen refresh
application if your monitor will
display unchanging static content.
Uninterrupted display of still or
static images over an extended
period may cause “burn in”, also
known as “after-imaging” or “ghost
imaging”, on your screen.
• “Burn-in”, “after-imaging”, or
“ghost imaging” is a well-known
phenomenon in LCD panel
technology. In most cases, the
“burned in” or “after-imaging” or
“ghost imaging” will disappear
gradually over a period of time after
the power has been switched off.
Warning
Failure to activate a screen saver, or
a periodic screen refresh application
may result in severe “burn-in” or “afterimage” or “ghost image” symptoms
that will not disappear and cannot be
repaired. The damage mentioned above
is not covered under your warranty.
Service
• The casing cover should be opened
only by qualified service personnel.
• If there is any need for any
document for repair or integration,
please contact with your local
service center. (please refer to the
chapter of “Consumer Information
Center”)
• For transportation information,
please refer to “Technical
Specifications”.
• Do not leave your monitor in a car/
trunk under direct sun light.
Note
Consult a service technician if the
monitor does not operate normally, or
you are not sure what procedure to take
when the operating instructions given
in this manual have been followed.
8
1. Important
1.4 Notational Descriptions
The following subsections describe
notational conventions used in this
document.
Notes, Cautions and Warnings
Throughout this guide, blocks of text
may be accompanied by an icon and
printed in bold or italic type. These
blocks contain notes, cautions or
warnings. They are used as follows:
Note
This icon indicates important
information and tips that help you make
better use of your computer system.
Caution
This icon indicates information that
tells you how to avoid either potential
damage to hardware or loss of data.
Warning
This icon indicates the potential for
bodily harm and tells you how to avoid
the problem.
Some warnings may appear in alternate
formats and may not be accompanied
by an icon. In such cases, the specific
presentation of the warning is
mandated by the relevant regulatory
authority.
Do not modify this equipment without
authorization of the manufacturer.
The monitor shall not be used for
critical diagnosis purpose or for life
supporting system.
WARNING
TO AVOID RISK OF ELECTRIC SHOCK,
THIS EQUIPMENT MUST ONLY BE
CONNECTED TO A SUPPLY MAINS WITH
PROTECTIVE EARTH.
1.5 Disposal of product and
packing material
Waste Electrical and Electronic
Equipment-WEEE
This marking on the product or on
its packaging illustrates that, under
European Directive 2012/19/EU
governing used electrical and electronic
appliances, this product may not be
disposed of with normal household
waste. You are responsible for disposal
of this equipment through a designated
waste electrical and electronic
equipment collection. To determine the
locations for dropping off such waste
electrical and electronic, contact your
local government office, the waste
disposal organization that serves your
household or the store at which you
purchased the product.
Your new monitor contains materials
that can be recycled and reused.
Specialized companies can recycle
your product to increase the amount of
reusable materials and to minimize the
amount to be disposed of.
All redundant packing material has been
omitted. We have done our utmost to
make the packaging easily separable
into mono materials.
Please find out about the local
regulations on how to dispose of your
old monitor and packing from your
sales representative.
9
Loading...
+ 25 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.