Philips A1 A3 service manual

0 (0)

Philips A1 and A3 Patient Monitors

A1 Models: M3921A

M3922A

M3923A

M3924A

A3 Models: M3926A

M3927A

M3928A

M3929A

Service Guide

Part Number M3921-9000C

Printed in the U.S.A. February 2002

First Edition

*M3921-9000C

About this Manual

Proprietary Information

This document contains proprietary information, which is protected by copyright. All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws.

Philips Medical Systems 3000 Minuteman Road Andover, MA 01810-1085 (978) 687-1501

Publication number

M3921-9000C

Printed in USA

Warranty

The information contained in this document is subject to change without notice.

Philips Medical Systems makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties or merchantability and fitness for Philips Medical Systems shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

Copyright

Copyright © 2002 Philips Electronics North America Corporation

Printing History

New editions of this document incorporate all material updated since the previous edition. Update packages may be issued between editions and contain replacement and additional pages to be merged by a revision date at the bottom of the page. Pages that are rearranged due to changes on a previous page are not considered revised.

The documentation printing date and part number indicate its current edition. The printing date changes when a new edition is printed. (Minor corrections and updates that are incorporated at reprint do not cause the date to change.) The document part number changes when extensive technical changes are incorporated.

This edition, M3921-9000C combines both A1 and A3 Service Guides into one book. There is no M3921-9000B book.

First Edition...............................................................

February 2002

ii

About this Manual

Philips Software License Terms

ATTENTION: USE OF THE SOFTWARE IS SUBJECT TO THE PHILIPS SOFTWARE LICENSE TERMS SET FORTH BELOW. USING THE SOFTWARE INDICATES YOUR ACCEPTANCE OF THESE LICENSE TERMS. IF YOU DO NOT ACCEPT THESE LICENSE TERMS, YOU MAY RETURN THE SOFTWARE FOR A FULL REFUND. IF THE SOFTWARE IS BUNDLED WITH ANOTHER PRODUCT, YOU MAY RETURN THE ENTIRE UNUSED PRODUCT FOR A FULL REFUND.

PHILIPS SOFTWARE LICENSE TERMS

The following License Terms govern your use of the accompanying Software unless you have a separate signed agreement with Philips Medical Systems.

License Grant. Philips Medical Systems grants you a license to Use one copy of the Software. "Use" means storing, loading, installing, executing or displaying the Software. You may not modify the Software or disable any licensing or control features of the Software. If the Software is licensed for "concurrent use", you may not allow more than the maximum number of authorized users to Use the Software concurrently.

Ownership. The Software is owned and copyrighted by Philips or its third party suppliers. Your license confers no title to, or ownership in, the Software and is not a sale of any rights in the Software. Philips’ third party suppliers may protect their rights in the event of any violation of these License Terms.

Copies and Adaptations. You may only make copies or adaptations of the Software for archival purposes or when copying or adaptation is an essential step in the authorized Use of the Software. You must reproduce all copyright notices in the original Software on all copies or adaptations. You may not copy the Software onto any public network.

No Disassembly or Decryption. You may not disassemble or decompile the Software unless Philips prior written consent is obtained. In some jurisdictions, Philips consent may not be required for limited disassembly or decompilation. Upon request, you will provide Philips with reasonably detailed information regarding any disassembly or decompilation. You may not decrypt the Software unless decryption is a necessary part of the operation of the Software.

Transfer. Your license will automatically terminate upon any transfer of the Software. Upon transfer, you must deliver the Software, including any copies and related documentation, to the transferee. The transferee must accept these License Terms as a condition to the transfer.

Termination. Philips Medical Systems may terminate your license upon notice for failure to comply with any of these License Terms. Upon termination, you must immediately destroy the Software, together with all copies, adaptations and merged portions in any form.

Export Requirements. You may not export or re-export the Software or any copy or adaptation in violation of any applicable laws or regulations.

U.S. Government Restricted Rights. The Software and any accompanying documentation have been developed entirely at private expense. They are delivered and licensed as

iii

About this Manual

"commercial computer software" as defined in DFARS 252.227-7013 (Oct. 1988), DFARS 252.211-7015 (May 1991) or DFARS 252.227-7014 (Jun. 1995), as a "commercial item" as defined in FAR 2.101(a), or as "Restricted computer software" as defined in FAR 52.227-19 (Jun. 1987)(or any equivalent agency regulation or contract clause), whichever is applicable.

You have only those rights provided for such Software and any accompanying documentation by the applicable FAR or DFARS clause or the Philips standard software agreement for the product involved.

iv

Philips A1 A3 service manual

About this Manual

Text Conventions

The following conventions for Notes, Cautions, and Warnings are used in this manual.

NOTE

A Note calls attention to an important point in the text.

Caution

A Caution calls attention to a condition or possible situation that could damage or destroy the product or the user’s work.

Warning

A Warning calls attention to a condition or possible situation that could cause injury to the user and/or patient.

Explanation of Symbols

Symbols on products and packaging mean the following:

Battery Gauge icon

Defibrillator-proof type CF equipment

On/Standby button.

Alarm Limits icon.

Adjust Screen Contrast to display (monochrome) or invert the video to (color).

Adjust Heart Rate Tone volume.

Initiate NiBP measurement

v

About this Manual

NiBP

ECG

T Temperature

SpO2 SpO2

Located on recorder. Produces a 20 Second Snapshot

Located on recorder. Produces a Continuous Print

Signal (ECG) Input

Signal (ECG) Output

Data input/output

Alternating Current

Direct Current

Temperature

Humidity

Altitude or atmospheric pressure

vi

About this Manual

Contains parts that may not be put into normal waste disposal but must be recycled or disposed as chemical waste

Fragile, handle with care

Keep dry

Consult instructions for use

Date of manufacture

Refer to User’s Guide

DC Lines

DC Input

Defib Sync.

RS-232 I/O

Class II Equipment

Power supply if for indoor use only

Silence Alarm

Equipotential Ground

vii

About this Manual

Recorder’s front-panel charging LED

Recorder’s front-panel communication LED

Ground Wire

viii

Contents

1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Manual Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Patient Monitor Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Introduction to External Recorder (M3925A) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Related Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

2. Routine Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Periodic Safety And Functional Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Environmental Protection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

3. Performance Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Test And Inspection Matrix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Equipment Needed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Performance Tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Battery Charge. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Battery Performance Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Power-On Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Hardware and Software Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

SpO2 Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Operation with an ECG Simulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

Operation with a Respiration Simulator (A3 only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Verification of Pneumatic System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Operation with a Temperature Simulator (A1 only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Operation with a Temperature Simulator (A3 only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 General Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Serial Interface Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Printer Verification (For A1 Optional Printer M3925A) . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Ground Integrity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Electrical Leakage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

4. Power-up Defaults Menu And Diagnostic Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 37

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

Power-up Defaults Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

Diagnostic Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

System Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

System A/D Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

NIBP Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

Restoring Factory Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

Contents

ix

5. Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 How To Use This Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Who Should Perform Repairs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Replacement Level Supported . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Obtaining Replacement Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Generating an Error Log Printout (A1 only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Serviceable Hardware Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 Other Error Codes (All Monitors) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 Buttons/Wheels (All Monitors) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Display/Audible Tones (All Monitors) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59 Operational Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

6. Disassembly Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

How To Use This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

Disassembly Procedures (A1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65

Closed Case Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65

Separation of Front and Rear Case Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

Front Case Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70

Rear Case Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

Main PCB Disassembly Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77

Disassembly Procedures (A3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79

Closed Case Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79

Separation of Front and Rear Case Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80

Front Case Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82

Rear Case Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85

7. Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91

Small Parts Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91

Top Level Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92

Front Case Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97

Rear Case Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103

Main PCB Assembly (A1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107

Power Supply, Cables, and Miscellaneous Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109

Exchange Unit Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112

8. Packing For Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113

General Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113

Repacking In Original Carton . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113

Repacking In A Different Carton . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113

x

Contents

9. RS-232 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115

Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .116

Nurse Call . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116

10. Training Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117 Levels of Involvement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .117 Preventative Maintenance Only . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117 Phone Support or Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117 Training Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118 Essential Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118 Optional Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119 Support Strategies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 Unit exchange . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 Bench repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 Theory of Operation and System Architecture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121 System Overview - A1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121 Block Diagram - A1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121 Block Diagram - A1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122 System Overview - A3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123 Block Diagram - A3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123 Block Diagram - A3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129 Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130 Monitor Applications and Algorithms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132 Reference Documents and Other Sources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132 Preventative Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133

Index

Contents

xi

xii

Contents

1 Introduction

Manual Overview

This manual contains information for servicing the A1 and A3 patient monitors, subsequently referred to as the monitor throughout this manual. Only where there are differences in service procedures for the two monitor types is each monitor referred to specifically as A1 and A3. Only qualified service personnel should service this product. Before servicing the monitor, read the User Guide carefully for a thorough understanding of operation.

Introduction

1

Patient Monitor Description

Patient Monitor Description

The purpose and function of the patient monitor is to monitor: ECG; heart rate; noninvasive blood pressure (systolic, diastolic, and mean arterial pressures); functional arterial oxygen saturation; respiration rate (A3 only); and temperature for adult and pediatric patients in all hospital areas and hospital-type facilities. It can be used during hospital transport and in mobile, land-based environments, such as ambulances.

The physical and operational characteristics of the monitor are described in the User Guide.

The measurement parameters and features for each model are indicated below.

Model

 

Measurement Parameters and Features

 

 

 

 

 

 

 

 

 

 

NIBP

SpO2

Temp

ECG

Resp

Integral

Color/

 

 

 

 

 

 

Recorder

Mono

 

 

 

A1 Monitor

 

 

 

 

 

 

 

 

 

 

 

M3921A

Yes

Yes

No

No

No

No

Mono

 

 

 

 

 

 

 

 

M3922A

Yes

Yes

Yes

No

No

No

Mono

 

 

 

 

 

 

 

 

M3923A

Yes

Yes

No

Yes

No

No

Mono

 

 

 

 

 

 

 

 

M3924A

Yes

Yes

Yes

Yes

No

No

Mono

 

 

 

 

 

 

 

 

M3925A

Optional Standalone Recorder for A1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A3 Monitor

 

 

 

 

 

 

 

 

 

 

 

M3926A

Yes

Yes

Yes

Yes

Yes

No

Mono

 

 

 

 

 

 

 

 

M3927A

Yes

Yes

Yes

Yes

Yes

Yes

Mono

 

 

 

 

 

 

 

 

M3928A

Yes

Yes

Yes

Yes

Yes

No

Color

 

 

 

 

 

 

 

 

M3929A

Yes

Yes

Yes

Yes

Yes

Yes

Color

 

 

 

 

 

 

 

 

2

Introduction

Introduction to External Recorder (M3925A)

Introduction to External Recorder (M3925A)

The recorder is an optional, standalone printer designed for use with the A1 patient monitor. The recorder communicates with the monitor using a nullmodem cable connected between each device’s RS-232 connector.

The recorder contains an internal battery, which, when fully charged, will operate the recorder for 3 hours (typical, at 25°C, producing fifteen 20-second printouts per hour). The recorder can be connected to AC power using an external power supply. The recorder uses the same type of power supply as the A1 monitor, the PS-120V or PS-240V.

The recorder does not have an On/Off switch. The recorder can sense when it has an established communication link with the monitor. At that time, the green LINKED indicator on the front panel lights, indicating that the recorder is ready for operation. See the recorder’s User Guide for more information regarding its operation.

Related Documents

To perform test and troubleshooting procedures and to understand the principles of operation and circuit analysis sections of this manual, you must know how to operate the monitor. Refer to the User Guide to understand the various sensors, ECG lead, blood pressure cuffs, and temperature probes that work with the monitor. Refer also to the user guide and individual directions for use that accompany these accessories.

Introduction

3

Related Documents

4

Introduction

 

2

 

 

Routine Maintenance

 

 

 

 

 

 

Cleaning

 

 

 

 

 

 

 

 

Warning

 

Do not immerse the monitor or its accessories in liquid or

 

 

clean with caustic or abrasive cleaners. Do not spray or

 

 

pour any liquid on the monitor or its accessories.

 

 

 

 

 

 

 

 

To clean the monitor, dampen a cloth with a commercial, nonabrasive cleaner

 

 

and wipe the exterior surfaces lightly. Do not allow any liquids to come in

 

 

contact with the power connector or switches. Do not allow any liquids to

 

 

penetrate connectors or openings in the instrument. For cables, sensors and

 

 

cuffs, follow the cleaning instructions in the directions for use that accompany

 

 

these accessories.

Periodic Safety And Functional Checks

The monitor requires cleaning, battery maintenance and NiBP performance and verification check every two years. The following performance verification tests may be used following repair or during routine maintenance (if required by your local institution).

1.Inspect the exterior of the monitor for damage.

2.Inspect labels for legibility. If the labels are not legible, contact Philips’ Response Center or your local Philips representative.

3.Verify that the NiBP performs properly as described in “Verification of Pneumatic System” on page 21.

Battery

If the monitor has not been used for a long period of time, the battery will need charging. To charge the battery, connect the monitor to an AC outlet, or external DC supply in the case of the A3, as described in “Battery Charge” on page 13 or the “Setup and Use” chapter of the User Guide.

Note

Storing the monitor for a long period without charging the battery

 

can degrade the battery capacity. A complete battery recharge

 

requires 8 hours.

Chapter 2 - Routine Maintenance

5

Environmental Protection

The battery can be recharged while the monitor is in use, in which case, the battery requires 14 hours to be recharged.

The battery can require a full discharge/charge cycle to restore normal capacity.

If the monitor operates1 for less than one hour on battery power before the low battery alarm occurs, the battery should be conditioned.

If the same symptom persists even after the battery is conditioned and indicating a full charge, the battery should be replaced.

Refer to Chapter 6, “Disassembly Guide”.

Environmental Protection

Follow local governing ordinances and recycling plans regarding disposal or recycling batteries and other device components.

1.See “Battery Performance Test” on page 14 for typical battery operating times and conditions.

6

Chapter 2 - Routine Maintenance

3 Performance Verification

Introduction

This section discusses the tests used to verify performance following repairs or during routine maintenance. All tests can be performed without removing the covers of the monitor.

If the monitor fails to perform as specified in any test, repairs must correct the problem before the monitor is returned to the user.

Performance Verification

7

Test And Inspection Matrix

Test And Inspection Matrix

The following test map shows which tests are required in which situations.

Table 1

Test Map

 

Service Event (When

Test Blocks Required (…Complete

Performing…)

these Tests)

 

 

 

Installation

 

Visual and Power On Tests

 

 

 

Repairs

 

 

 

 

A. Unit Exchange

Visual and Power On Tests

 

 

B. Unit Opened

Power On Test

 

 

Basic Pneumatic Leakage Test (BPL)

 

 

Ground Integrity Test

 

 

Leakage Current Test

 

 

C. NBP Pump replaced

Power On Test

 

 

NBP Tests:

 

 

Pneumatic Leakage and Inflation Rate

 

 

D. SpO2 Module replaced

Power On Test

 

 

BPL Test

 

 

SpO2 Tests:

 

 

Dynamic Operating Range and

 

 

LED Excitation

 

 

E. Front End Connectors replaced

Power On Test

 

 

BPL Test

 

 

Performance test for the parameter that had

 

 

the connector replaced.

 

 

F. Power Supply replaced (A3 only)

Power On, BPL, and Safety tests

 

 

G. Component level repair on any

Power On Test

PC board

 

All Performance Tests

 

 

 

 

All Safety Tests

 

 

H. Main PC board replaced

Power On Test

 

 

All Performance tests except Battery

 

 

Charge Test and Battery Performance Test

 

 

All software upgrades

Power On

 

 

Preventative maintenance

Power On and NBP Performance Tests

 

 

 

8

Performance Verification

Test And Inspection Matrix

Table 2 Test and Inspection Reporting

Test1

Expected Results

What to Record2

Visual: Inspect exterior

No Visual Damage

V:P or V:F

of monitor for damage

 

 

 

 

 

Power-On Self-Test

Displays Normal Monitoring

PO:P or PO:F

 

Screen Configuration and emits

 

 

tones

 

 

 

 

Basic Pneumatic

After 1 minute at approximately

BPL: Px6

Leakage Test (NBP

250 mmHg, pressure drops no

Where x6 =Pressure

test)

more than 6 mmHg.

Drop in mmHg

 

 

 

 

 

NiBP Accuracy Test

250 mmHg; ±5 mmHg

PN:P/X1…X4 (or X53)

 

 

PN:F/X1…X4 (or X53)

 

 

Record Pressure in

 

 

mmHg = X1

 

 

 

NiBP Leakage Test

After 1 Minute at 250 mmHg,

Record Pressure

 

pressure drops no more than

Drop in mmHg = X2

 

6 mmHg

 

 

 

 

NiBP Inflation Rate

Monitor Reports NiBP = 280

Record Time to

Test

mmHg in <6 seconds

280 mmHg in

 

 

Seconds = X3

 

 

 

NiBP Overpressure

Overpressure Deflation is

Record value in

Test

activated at a value between

mmHg at which

 

280 mmHg to 330 mmHg

Overpressure

 

 

Deflation occurs =X4

 

 

 

NiBP Deflation Rate

Monitor reports NiBP >10

Record value in

Test (A3 only)

mmHg and <190 mmHg

mmHg on monitor at

 

 

1 min = X5

 

 

 

Safety Performance

All Safety Performance Tests

S:P or S:F

Tests

are in range of Table 5 to

 

 

Table 12

 

 

 

 

1.Details of the tests are included later in this chapter.

2.When authorized Philips Medical Systems personnel service the instrument. The results are reported back to Philips. The collected data form a database to be used in product development. It is not necessary for hospital personnel to report these results.

3.In the case of the NiBP Deflation Rate Test, applicable to the A3 only.

Performance Verification

9

Equipment Needed

Equipment Needed

The following table lists the equipment required for performance verification.

Table 3

Required Test Equipment

 

 

 

 

 

Required For

 

 

 

 

Equipment

Description

A1

A3

 

 

 

 

Digital multimeter

Fluke Model 87 or equivalent

 

 

(DMM)

 

 

 

 

 

 

 

 

Sensor extension cable

M4787A

 

 

 

 

 

 

Finger clip sensor

M4789A

 

 

 

 

 

 

Oxisensor® II adhesive

D-25

 

 

sensor

 

 

 

 

 

 

 

 

 

ECG cable

 

M3913A

M3923A and

 

 

 

 

M3924A only

 

 

 

 

 

ECG electrodes

Standard

M3923A and

 

 

 

 

M3924A only

 

 

 

 

 

 

ECG leads

 

M3914A (IEC) or M3915A

M3923A and

 

 

 

(AAMI)

M3924A only

 

 

 

 

 

 

NiBP tubing

 

M3918A

 

 

 

 

 

 

 

NiBP cuff

 

40401C

 

 

 

 

 

 

Pulse oximeter tester

Nellcor Puritan Bennett SRC-2

 

 

 

 

 

 

ECG simulator

Dynatech Nevada medSim

 

 

 

 

300 or equivalent

 

 

 

 

 

 

NiBP simulator

Bio-Tek BP Pump or

 

 

 

 

equivalent

 

 

 

 

 

 

Temperature simulator

medSim 300 or equivalent

M3922A and

 

 

 

 

M3924A only

 

 

 

 

 

Respiration simulator

medSim 300 or equivalent

 

 

 

 

 

 

Safety analyzer

Bio-Tek 601 Pro or equivalent

 

 

 

 

 

 

 

Stopwatch

 

Manual or electronic

 

 

 

 

 

 

 

10

Performance Verification

Performance Tests

Performance Tests

The battery charge and battery performance test should be performed before monitor repairs whenever the battery is suspected as being a source of the problems. All other tests can be used following repairs or during routine maintenance (if required by your local institution). Before performing the battery performance test, ensure that the battery is fully charged. (See “Battery Charge” below).

This section is written using factory-set power-up defaults. If your institution has pre configured custom defaults, those values display.

Battery Charge

To fully charge the battery:

1.Connect the monitor to an AC power source using the proper power cord. For the A1, use the PS-120V or PS-240V external power supply and power cord.

2.• For the A1, verify that the EXTERNAL POWER indicator is lit.

For the A3, verify that the BATTERY CHARGING/AC SOURCE indicator is lit.

3.Charge the battery for at least 8 hours.

The battery can require a complete discharge/charge cycle to restore its normal capacity, depending on its previous usage.

4.To check for a full charge, perform the procedure described in “Battery Performance Test” on page 12.

Performance Verification

11

Battery Performance Test

Battery Performance Test

The the A1 and A3 mono monitors are specified to typically operate on battery power for a minimum of 4 hours, at 25°C, with no printing, and one NiBP measurement every 15 minutes. The A3 color monitor is specified to typically operate on battery power for a minimum of 3 hours, at 25°C, with no printing, and one NiBP measurement every 15 minutes. Before performing this test, ensure that the battery is fully charged (see “Battery Charge”).

1.Connect the Nellcor Puritan Bennett SRC-2 pulse oximeter tester to the monitor via the M4787A sensor cable.

2.Connect the NiBP simulator to the monitor via the M3918A tubing.

3.Set the SRC-2 switches as follows:

SWITCH

POSITION

 

 

RATE

38

 

 

LIGHT

LOW

 

 

MODULATION

LOW

 

 

RCAL/MODE

RCAL 63/LOCAL

 

 

4.Set the NiBP simulator to simulate a pressure setting of 120/80 mmHg and heart rate of 80 bpm.

5.Ensure that the monitor is not connected to AC power.

6.With the monitor turned off, press the ON/STANDBY button and verify that the battery icon appears at the bottom of the display after the power-on selftest is completed. The boxes in the battery icon should all be filled, indicating that the battery is charged.

7.Verify that the monitor is responding to the SpO2 simulator signal and that the audible alarm is sounding. Use the wheel to select the SpO2 Menu and permanently silence the SpO2 audible alarm.

8.Use the wheel to select the NiBP Menu and set the Automatic Measurement Interval to 15 minutes. Exit the menu and press the front panel NiBP button

to manually start the first NiBP measurement. Subsequent NiBP measurements are taken automatically every 15 minutes.

12

Performance Verification

Power-On Self-Test

9.The A1 and A3 mono monitors must operate for at least 4 hours before they automatically power down due to low battery condition. The A3 color monitor must operate for at least 3 hours before the monitor automatically powers down due to low battery condition.

10.Verify that the low battery alarm occurs 15-30 minutes before the battery fully discharges.

11.Allow the monitor to operate until it automatically powers down due to low battery condition. Verify that the audible alarm sounds when the monitor automatically shuts down. Press the Alarm Silence button to terminate this audible alarm.

12.If monitor passes this test, immediately recharge the battery. (See “Battery Charge” on page 11).

Power-On Self-Test

1.Connect the monitor to an AC power source. For the A1, use the PS-120V or PS-240V power supply and power cord. For the A3, use the proper power cord supplied.

2.• For the A1, verify that the EXTERNAL POWER indicator is lit.

• For the A3, verify that the BATTERY CHARGING/AC SOURCE indicator is lit.

3.Do not connect any input cables to the monitor.

4.Observe the monitor front panel. With the monitor off, press the ON/ STANDBY button . The monitor must perform the following sequence:

a.The A1 emits a beep.

The A3 emits three consecutively higher pitched beeps.

b.A few seconds later, the display backlight illuminates, but the display is blank.

c.The version numbers of the boot and operational software display in the lower left corner of the display.

d.A beep signals the end of the power-on self-test. Power-on self-test takes approximately 10 seconds to complete.

e.Upon successful completion of the power-on self-test, the display is in normal monitoring screen configuration. No vital-sign numeric values or waveforms are displayed.

Performance Verification

13

Hardware and Software Tests

Hardware and Software Tests

Hardware and software testing includes the following tests applicable to the indicated models in the series:

Test

 

Applies to Model(s)

 

 

SpO2 Testing

A1, A3 - all models

Operation with an ECG Simulator

A1

- M3923A, M3924A

 

A3

- all models

 

 

 

Operation with a Respiration Simulator

A3

- all models

 

 

Verification of Pneumatic System

A1, A3 - all models

 

 

 

Operation with a Temperature Simulator

A1

- M3922A, M3924A

 

A3

- all models

 

 

General Operation

A1, A3 - all models

 

 

 

SpO2 Testing

SpO2 testing includes the following tests:

Alarms and Alarm Silence

Heart Rate Tone Volume Control

Dynamic Operating Range

LED Excitation Test

Alarms and Alarm 1.

Connect the SRC-2 pulse oximeter tester to the M4787A sensor cable and

Silence

connect the cable to the monitor.

 

 

2.

Set SRC-2 as follows:

 

 

 

 

 

 

 

 

 

 

SWITCH

 

POSITION

 

 

 

 

 

 

 

 

 

RATE

 

38

 

 

 

 

 

 

 

 

 

LIGHT

 

LOW

 

 

 

 

 

 

 

 

 

MODULATION

 

OFF

 

 

 

 

 

 

 

 

 

RCAL/MODE

 

RCAL 63/LOCAL

 

 

 

 

 

 

 

3.Press the ON/STANDBY button to turn the monitor on.

4.After the normal power-up sequence, verify that the SpO2% display initially

14

Performance Verification

Hardware and Software Tests

indicates zero or is blank. The pulse bar can occasionally indicate a step change as the monitor is in the pulse search mode.

5.Move the modulation switch on the SRC-2 to LOW.

6.Verify the following monitor reaction:

a.The pulse bar begins to track the artificial pulse signal from the SRC-2.

b.Initially, zero is displayed in the SpO2 frame, or it is blank.

c.After about 10 to 20 seconds, the monitor displays saturation and heart rate as specified by the tester. Verify that the values are within the following tolerances:

Oxygen Saturation Range 79% to 83%

Heart Rate Range 35 to 41 bpm

d.The audible alarm sounds and both the SpO2% and HEART RATE displays flash, indicating both parameters have violated the default alarm limits.

e.The heart rate tone is heard. For this test, the user should set the heart rate tone source to SpO2 from the Heart Rate Menu.

7.Press the ALARM SILENCE button on the front panel of the monitor. The audible alarm is temporarily silenced.

8.Verify the following:

a.The audible alarm remains silenced.

b.The slashed bell icon appears in each numeric frame on the display.

c.The SpO2% and HEART RATE displays continue flashing.

d.The heart rate tone remains audible.

e.The audible alarm returns in approximately 60 seconds.

Heart Rate Tone 1. Connect the SRC-2 pulse oximeter tester to the M4787A sensor cable and

Volume Control

connect the cable to the monitor.

 

2.

Set SRC-2 as follows:

 

 

 

 

 

 

 

SWITCH

POSITION

 

 

 

 

A1

A3

 

 

 

 

 

 

 

 

RATE

38

38

 

 

 

 

 

LIGHT

LOW

LOW

 

 

 

 

 

MODULATION

LOW

OFF

 

 

 

 

 

RCAL/MODE

RCAL 63/LOCAL

RCAL 63/LOCAL

 

 

 

 

 

Performance Verification

15

Hardware and Software Tests

3.Power on the monitor and verify that the SpO2 and heart rate values are correctly displayed.

4.Press the ALARM SILENCE button on the front panel of the monitor to temporarily silence the audible alarm.

5.Verify that the heart rate tone source, found in the Heart Rate Menu, is set to

SpO2.

Dynamic Operating

Range

6.Press the Heart Rate Tone VOLUME button on the front panel of the monitor. Within 3 seconds of having pressed the button, rotate the wheel clockwise and verify that the beeping heart rate tone sound level increases.

7.Rotate the wheel counterclockwise and verify that the beeping heart rate tone decreases until it is no longer audible. Rotate the wheel clockwise to return the beep volume to a comfortable level.

3 seconds after the last button-press or rotation of the wheel, function of the wheel reverts to moving the highlight on the display screen.

The following test sequence verifies proper monitor operation over a range of input signals.

1.Connect the pulse oximeter tester to the monitor and turn the monitor on.

2.Place the SRC-2 pulse oximeter tester in the RCAL 63/LOCAL mode.

3.Set the SRC-2 as indicated in below. Verify that the monitor readings are within the indicated tolerances. Allow the monitor several seconds to stabilize the readings.

Table 4 SRC 2 Settings and Monitor Indications

 

SRC-2 Settings

Monitor Indications

 

 

 

 

 

RATE

LIGHT

MODULATION

SpO2

Pulse Rate

 

 

 

 

 

38

HIGH2

LOW

79-83*

35-41*

 

 

 

 

 

112

HIGH1

HIGH

79-83*

109-115

 

 

 

 

 

201

LOW

LOW

79-83*

195-207*

 

 

 

 

 

201

LOW

HIGH

79-83*

195-207*

 

 

 

 

 

An * indicates values that produce an alarm. Press the ALARM SILENCE button to temporarily silence the audible alarm.

For the pulse rate setting of 201 bpm, the pulse rate tolerance of 195 to 207 bpm is greater than the ±3 bpm accuracy specification of the monitor, due to the performance characteristics of the SRC-2 tester.

4.Turn the monitor off.

LED Excitation Test This procedure uses normal system components to test circuit operation. A

16

Performance Verification

Hardware and Software Tests

Nellcor Puritan Bennett Oxisensor II adhesive sensor, model D-25, is used to examine LED intensity control. The red LED is used to verify intensity modulation caused by the LED intensity control circuit.

1.Connect a M4787A sensor extension cable to the monitor.

2.Connect a D-25 sensor to the sensor extension cable.

3.Press the ON/STANDBY button to turn the monitor on.

4.Leave the sensor open with the LED and photo detector visible.

5.After the monitor completes its normal power-up sequence, verify that the sensor LED is brightly lit.

6.Slowly move the sensor LED in proximity to the photo detector element of the sensor. Verify, as the LED approaches the optical sensor, that the LED intensity decreases.

7.Open the sensor and notice that the LED intensity increases.

8.Repeat step 6 and the intensity again decreases. This variation is an indication that the micro-processor is in proper control of LED intensity.

9.Turn the monitor off.

Operation with an ECG Simulator

1. With the monitor off, connect the ECG leads to the appropriate jacks on the ECG tester.

2. Connect the leads to the ECG cable.

3. Connect the cable to the ECG input port on the monitor. 4. Set the ECG simulator as follows:

Heart rate:

30 bpm

Amplitude:

1 millivolt

Lead select:

II

Wave Type:

Normal sinus rhythm

Patient Type:

Adult mode

Note

The accuracy of the monitor’s ECG measurements is ±5 bpm. In

 

the procedure below, add the tolerance of the simulator to the

 

acceptable range of readings.

 

 

 

 

5. Press On/Standby button

to turn monitor on.

Performance Verification

17

Hardware and Software Tests

6.After normal power-up sequence, verify the following monitor reactions:

a.After at least five heartbeats, the monitor displays a heart rate of 30 ±5 bpm.

b.The audible alarm sounds and the HEART RATE display flashes, indicating heart rate is below the default lower alarm limit.

7.Press the ALARM SILENCE button. Verify that the audible alarm is silenced.

8.Increase the heart rate setting on the ECG simulator to 240 bpm.

9.After at least five heartbeats, verify that the monitor displays a heart rate of 240 ±5 bpm.

10. Verify that the audible alarm sounds and the HEART RATE display flashes, indicating that the heart rate is above the default upper alarm limit.

11.Press ALARM SILENCE button to silence alarm.

12.Decrease the heart rate setting on ECG simulator to 120 bpm.

13.After at least five heartbeats, verify that the monitor displays a heart rate of 120 ±5 bpm.

14.Disconnect the LL lead from ECG simulator.

15.

Verify that the Leads Off alarm message appears, three dashes are

 

displayed in HEART RATE display, and the low priority audible alarm

 

sounds.

16.

Reconnect the LL lead to ECG simulator. Verify that the Leads Off alarm

 

message no longer appears and audible alarm is silenced.

17.Repeat steps 14 through 16 for LA and RA leads.

18.Turn the monitor off.

Operation with a Respiration Simulator (A3 only)

1.With the monitor off, connect the ECG leads to the appropriate jacks on the respiration simulator.

2.Connect the ECG leads to the ECG cable.

3.Connect the cable to the ECG input port on the monitor.

Note

The accuracy of A3 measurements is ±3 breaths per minute. In

 

the procedure below, add the tolerance of the simulator to the

 

acceptable range of readings.

 

 

 

 

 

4.

Set the simulator for a respiration rate of 120 breaths per minute.

 

5.

Press the ON/STANDBY button

to turn the monitor on.

18

Performance Verification

Loading...
+ 120 hidden pages