Philips makes no warranty of any kind with regard to this material, including, but not
limited to, the implied warranties of merchantability and fitness for a particular purpose.
Philips Medical Systems shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance or use of this
material.
This document contains proprietary information that is protected by copyright. All
rights are reserved. No part of this document may be photocopied, reproduced or
translated to another language without prior written consent of Philips Medical Systems.
The information contained in this document is subject to change without notice.
Philips assumes no responsibility for the use or reliability of its software on equipment
that is not furnished by Philips.
Purchase of this instrument confers no express or implied license under any Nellcor
patent or copyright to use this instrument with any fetal oximetry sensor that is not
manufactured or licensed by Nellcor.
Dinamap is a trademark of General Electric.
Press-Mate is a trademark of the COLIN Corporation.
Federal Law (US) restricts this device to sale by or on the order of a physician.
Caution
Failure on the part of the responsible individual hospital or institution employing
the use of this equipment to implement a satisfactory maintenance schedule may
cause undue equipment failure and possible health hazards.
This guide tells you how to install, service and repair a Series 50 XM Fetal/Maternal monitor or a
Series 50 XMO Fetal/Maternal monitor. It describes all the system hardware and software, and
tells you how to test the system and diagnose operating and service problems. It also tells you what
upgrade paths are available and how to follow them.
It gives instructions for both a Series 50 XM Fetal/Maternal monitor and a Series 50 XMO Fetal/
Maternal monitor. The features available on the monitor you are installing or servicing depend on
which options have been purchased. All instructions apply to a fully equipped monitor. If your
monitor does not have a described parameter, then you can ignore that part of the instruction and
skip to the next point.
The illustrations in this book show the Series 50 XMO Fetal/Maternal monitor, with the fetal
pulse oximetry
This guide is for technical personnel installing and servicing the monitor. They must have a good
understanding of medical equipment installation procedures. Conventional current technical
terms are used throughout this guide. Familiarity with such terms is assumed.
Familiarize yourself with the contents of this guide and the Instructions for Use before attempting to
install or service the monitor.
1
(FSpO2) parameter and maternal parameters.
Conventions Used in This Guide
This guide uses the following conventions for notes, cautions, and warnings.
Warning
A warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to
observe a warning may result in death or serious injury to the user or patient.
1. Currently not available in the U.S.A.
General Information 1
Page 20
Symbols
Caution
A caution alerts you to situations where special care is necessary for the safe and effective
use of the product. Failure to observe a caution may result in minor or moderate personal
injury or damage to the product or other property, and possibly in a remote risk of more
serious injury.
Note— A note calls your attention to an important point in the text.
Symbols used in this guide are:
SymbolMeaning
Equipotential Terminal
This symbol is used to identify terminals which are connected together,
bringing various parts of an equipment or system to the same potential, not
necessarily being earth potential (the value of potentials of earth may be
indicated adjacent to the symbol).
International Caution Symbol
This symbol indicates that the operator should refer to the product instruction
manual before beginning a procedure.
Protective Earth Terminal
This symbol identifies the terminal for connection to an external protective
earth.
Positive Confirmation Symbol
!
This symbol against an item in a table means "applies to the item".
Not Applicable Symbol
-
This symbol against an item in a table means "not applicable".
2 General Information
Page 21
Unpacking and Checking the Shipment
Initial
Inspection
The monitor and any supporting options ordered are supplied packed in protective shipping
cartons. Before unpacking, visually check the packaging and ensure that there are no signs of
mishandling or damage.
Open the package carefully and remove the instrument and accessories. Remove the accessories
packed in the base before you dispose of the packing.
Check that the contents are complete and that the correct options and accessories have been
delivered (refer to Table 1-1).
Claims for
Damage
If the shipping cartons are damaged, contact the carrier.
If any of the equipment is damaged, contact both the carrier and your local Philips service
organization for repair or replacement arrangements.
RepackingRetain the original packing carton and material, in case you need to return equipment to Philips
for service. If you no longer have the original packing materials, Philips can advise you on
alternatives.
Table 1-1Contents Checklist
2
Fetal and
maternal
and
FSpO
2
M1350C
1 pack of 3
pieces
Description
Fetal Monitor M1350BM1350B
Fetal Accessories
SpO
Combined F
DECG adapter cable M1362B
External Toco Transducer (M1355A)1111
Ultrasound Transducer (M1356A)2222
ECG-only Patient Module (M1364A) with DECG cable M1362B11--
MECG adapter cable M1363A for use with Patient Module
(M1364A)
Reusable Transducer Belts (includes belt fastening buttons)
(M1562A/B)
Transducer Knob Adapters (M1356-43203)1 pack of 3
Fetal Scalp Electrodes
15133D (Europe) or
15133E (USA)
Fetal Scalp Sensor FS14 (M1366-60001)--11
/DECG/MECG Patient Module M1365A with
2
Fetal and
maternal
--11
Only one transducer is supplied if option C01 was ordered
Instructions for Use
(language as appropriate for your country)
Quick Reference Guide
(language as appropriate for your country)
Service Guide
(CD-ROM, English only)
Error Reference Card (English only)1111
Sensor Placement Guide (for FSpO2)0011
Pocket Guide to Fetal Monitoring (only supplied with English
shipments)
1111
1111
1111
1111
4 General Information
Page 23
Optional
Accessories
In addition to the items mentioned in the previous table, the following are delivered when the
appropriate option has been ordered:
Table 1-2Optional Accessories
Description QuantityRe-Ordering Number
IUP Pressure Transducer, including
Transducer Holder CPJ84022.
IUP Sensor-Tip Pressure Catheters
(disposable)
Adapter Cable for disposable IUP
Sensor-Tip Pressure Catheters
Barcode Reader plus Barcode
Booklet
Dual Serial Interface Board1M1350-66533
Cable (Serial)1M1350-61609 (for external devices)
OBMS/ODIS System Interface Board
(inc. RS422)
Instrument Identification
1CPJ840J5
1 box of 10M1333A
1M1334A
1 M1350-68730
1M1350-66532
Philips uses a two-section serial number (“XXXXGOOOOO'') for instrument identification. This
is located on labels attached to the monitor. One label is on the right side of the monitor, the other
label is identical and attached to the base of the monitor. The first four digits are the Serial Prefix
Number, and identify the modification standard of the instrument. The last five digits are the
Serial Number of the instrument. The letter that separates the serial prefix and suffix designates the
country in which the instrument was manufactured, “G'' = Germany.
3650GO2272
Serial Number
Serial Prefix
Country of Manufacture
General Information 5
Page 24
Service Philosophy
Any errors produced are marked on the trace paper with the symbol and a specific error
message is shown on the monitor display. In addition, the errors are reported on the system error
log (see page 24 for instructions on how to print the error log). Read the error messages and system
error log to determine whether the fault lies in the hardware or firmware. These problems can be
repaired by board replacement or unit exchange. Do not attempt component-level board repair.
Overview of
the Service
Tests
The system contains resident tests to enable monitoring of system status and errors that occur.
"Power-On Self Test (see page 112)
This is invoked whenever the system is switched on.
"Performance Assurance Tests
•Parameter Test (see page 112)
This checks the processing of the signal from the transducers.
•Quick Test (see page 114)
This checks the display, recorder and complete instrument hardware and software.
"Service Tests
These are invoked by connecting the monitor to a PC and using a software service tool to
assess, test and configure the system.
•Cyclic Test (see page 116)
•Permanent Test (see page 116)
"Safety Tests
These test the electrical safety of the monitoring system and must be carried out after
upgrades and repairs. You must use a safety testing tool.
•Instrument Safety Test (see page 120)
•System Test (see page 120)
•Safety Test Blocks (see page 119)
6 General Information
Page 25
About the Monitor
The monitorhas five color- and mechanically-coded input sockets:
"Cardio 1/Combi for fetal heart rate (FHR) and fetal oxygen saturation (FSpO
"Cardio 2 for fetal heart rate
"Toco for uterine activity
"NPB for external blood pressure (NIBP)
"SpO
for maternal pulse oximetry
2
)
2
When a transducer is connected, the monitor automatically selects the correct operating mode.
The unit can:
"monitor FHR using both direct ECG (DECG) and ultrasound (US)
"monitor maternal heart rate using ECG
"monitor twins using either DECG and ultrasound or dual ultrasound
"monitor uterine activity (Toco) externally or internally
"detect fetal movements (FMP) using ultrasound (optional)
"display maternal ECG waveform and record “snapshot” ECG waveform (optional)
"monitor FSpO
(optional)
2
A series of error messages appear in the instrument's display if certain measurements cannot be
carried out. Built-in check facilities give you a greater monitoring confidence because most of the
instrument functions can be checked quickly and easily.
The Performance Plus Package signal processing technique used by the instrument ensures that the
information provided by the monitor is very accurate. Ultrasound FHR traces are comparable with
direct ECG FHR traces and variability recorded in the ultrasound mode is almost identical to that
recorded in the DECG mode.
The recorder uses a high resolution thermal array print head which gives a very reliable and good
quality trace. Automatic annotation capabilities include time, date, paper speed and monitoring
methods. Nursing notes can also be documented using an optional barcode reader.
Either of the monitors can be upgraded to include all parameters, depending on availability within
the country of purchase.
General Information 7
Page 26
Overview of the Monitor
Figure 1-1Major Parts
1.Monitor On/Off Switch9.Function Key
2.Monitor On/Off Light10. Telemetry Indicator
3.Recorder Keys11. FSpO
4.Recorder12. Opening Recess
5.Maternal Parameters (optional)13. Socket for Remote Event Marker
6.Cardio 2 Channel14. Setting Keys
7.Toco Channel15. Service Socket
8.Cardio 1/Combi Channel
parameter (optional)
2
8 General Information
Page 27
Cardio and
Toco Channels
ta4.tif
Figure 1-2Cardio and Toco Channels
1.Toco Display shows uterine activity.
2.Fetal SpO
display shows fetal pulse indicator, signal quality, alarm status, and cross channel
2
verification plus indicator.
3.Cardio Display shows the FHR.
4.Signal Quality Indicator shows the quality of heart rate signal detected by the transducer:
•Green (optimum).
•Yellow (fair to potentially poor).
•Red (unacceptable).
5.Function Key selects menus for:
•FMP, twins offset, logic, FHR alert and FSpO
2
•returns to normal display.
6.MECG Indicator shows when MECG is being measured through this channel. (Indicator
location different for Series 50 XM).
7.Fetal SpO
display shows current value of FSpO2.
2
8.Speaker Lamp shows which heartbeat is heard from the loudspeaker.
9.Volume Keys set the volume and select the channel to which you are listening. Changes
current setting of FMP, twins offset, logic, FHR alert and FSpO
12. Cardio 1/Combi Transducer Socket. You can connect:
•FSpO
/ECG combined patient module (M1365A)
2
•ECG only patient module (M1364A)
•An ultrasound transducer (M1356A).
•A DECG transducer (M1357A).
•A US/MECG Combi transducer (M1358A).
•An MECG transducer (M1359A).
13. Toco Transducer Socket. You can connect:
•An external Toco transducer (M1355A).
•An IUP transducer (1290C or M1333A).
14. Toco Baseline Key zeroes the Toco display and trace to 20 units (when monitoring uterine
activity externally) or 0 units (when monitoring uterine activity internally).
15. Cardio 2 Transducer Socket for connecting:
•ECG only patient module (M1364A)
•US transducer (M1356A)
•DECG transducer (M1357A)
•MECG transducer (M1359A)
10 General Information
Page 29
Maternal
Parameters
Figure 1-3Maternal Parameters
1.NBP Cuff Socket. You can connect:
•An NBP cuff interconnect tubing (M1599A) and cuff (M1574A).
2.SpO
Transducer Socket. You can connect:
2
•An SpO
transducer (M1940A adapter cable connected to M1191A transducer).
2
3.Softkeys for operating and setting maternal parameters. They are:
•selects modes and alarm limits for NBP.
•selects modes and alarm limits for MHR.
• selects modes and alarm limits forSpO
.
2
4.Reset Key (yellow)
One short press
•Acknowledges warning message
•Acknowledges alarm
One press, held for two seconds (from maternal main screen only)
•Accesses volume and contrast setup
Two presses within one second (from maternal main screen only)
•Displays current maternal alarm limits
5.SpO
Value indicates the current reading for patient's oxygen saturation level.
2
General Information11
Page 30
6.MHR Icon indicates source of MHR
indicates heart rate value taken from MECG measurement.
Recorder Keys
indicates pulse rate value taken from SpO
indicates average pulse rate taken from NIBP measurement.
7.Maternal Heart Rate shows current heart rate or pulse rate.
8.Systolic Value shows the value for the systolic parameter of the most recent external blood
pressure measurement.
9.Diastolic Value shows the value for the diastolic parameter of the most recent external blood
pressure measurement.
measurement.
2
Figure 1-4Recorder Keys
1.Recorder On/Off Light. On when the recorder is switched on. Flashes when monitor detects
five or fewer pages remaining in the pack or if the paper runs out.
2.Recorder On/Off Key.
Single press turns recorder on/off
2-second press starts NST timer. Recorder must be off.
3.Event Marker Key. Press to record an event on the paper. Acknowledges all alerts and
alarms.
4.Paper Advance Key. Press once to advance the paper automatically to the next fold. Press a
second time to stop paper before fold. Tear paper at fold. Never pull paper to advance it.
5.Paper-Eject Key. Press once to unlock the drawer, and then press a second time and hold to
remove the paper.
12 General Information
Page 31
Setting Keys
Figure 1-5Setting Keys
1.Time and Date Key. Used when changing the time and date. Press to show the current time
in the Cardio 1/Combi and Toco displays, to cycle through the settings to be changed
(hours, minutes, day, month and year) and to return to the normal display.
2.Paper Speed Key. Used when changing the paper speed. Press to show the current paper
speed in the Cardio 1/Combi display, and to return to the normal display.
3.Test Key. Used to start monitor self test.
General Information 13
Page 32
Rear Panel
(Slot
Positions)
.
3
1A1
Figure 1-6Rear Panel (Slot Positions)
Slot
Number
1Power Supply with Power Cord Connection, including Equipotential Grounding Point (1A)
2ROM Board (M1350-66506)
3CPU Board with Barcode Reader Interface (M1350-66513)
4Telemetry Interface Connector (M1350-66536) and System Interface Connector (RS232) (lower)
If the monitor has NIBP:
5 & 6External Blood Pressure Board (M1350-66535). Optional. Connected to Slot 5, but physically
7Digital Signal Processor (DSPII) (M1350-66507).
If the monitor does not have NIBP:
5Slot empty (there is a single slot cover in this case).
6Digital Signal Processor-CoP Board (M1350-66505). Optional. (Note: only in conjunction with
7Digital Signal Processor-CPU Board (M1350-66504). Optional. (Only with M1350-66505)
Other monitor options:
8Maternal SpO
9OBMS/ODIS Analog Interface (optional) (M1350-66532). Optional.
occupies both Slot 5 and Slot 6 (there is a double slot cover in this case).
M1350-66504.
OR Digital Signal Processor (DSPII) (M1350-66507). Optional. Here, Slots 5 and 6 are empty.
Board (M1350-66534). Optional.
2
Interface (M1350-66534). All optional.
Board (M1350-66540). Optional.
2
2
4578
6
9
10
11
Note— The DIF board (M1350-66515) is located in Slot 0 inside the monitor (not visible
from the rear).
14 General Information
Page 33
Power Source Requirements
The monitor can be operated from an ac source of 100 to 120 or 220 to 240 volts (±10%) at 50Hz
or 60Hz (±5%). The system will be set to the correct voltage at the factory.
Prior to installing the system, check that the systems voltage has been set correctly. The voltage and
fuse values are shown on the rear panel.
2
Pre-Installation Checks
Figure 2-1System Voltage and Fuse Values on the Rear Panel
Caution
If the voltage has been set incorrectly you must reset it before you connect the system to the
local line power supply. See “Setting the Required Voltage” on page 16. Ensure that you use
the correct fuse for the voltage setting.
Pre-Installation Checks 15
Page 34
Setting the Required Voltage
1.Using a flat blade screwdriver, open the panel on the rear of the monitor.
pin1sca.tif
Figure 2-2Opening the Rear Panel
16 Pre-Installation Checks
Page 35
2.Remove the voltage selector drum, turn it to the voltage required, and replace it.
pin2sca.tif
Installation
Figure 2-3Voltage Selector Removal and Replacement
3.Close the rear panel and check that the correct voltage setting is shown (see Figure 2-1).
For installation instructions, please refer to the Instructions for Use.
Pre-Installation Checks 17
Page 36
18 Pre-Installation Checks
Page 37
Introduction
3
Configuring the Monitor
This chapter describes the tasks you must carry out to configure the monitor. This involves the
following service tasks:
"Configure the monitor
"Configure the recorder
"Print and clear the error log
"Set the time and date format to the local standard
Some configuration tasks can be completed by connecting a PC to the monitor, some by using
pushbuttons on the monitor itself. Most can be performed using whichever method is more
convenient for you: see Table 3-1 for a list of tasks and methods. To find out how to set the time,
date, and paper speed, see Chapter 4.
Configuring the Monitor 19
Page 38
Configuration Tasks
The following table shows you which configuration actions can be carried out using pushbuttons
on the monitor and which can be carried out using a PC connected to the service socket.
Table 3-1Configuring the Monitor
Configure the time formatYesYes
Configure the date formatYesYes
Configure the IUP formatYesYes
Configure the paper formatYesYes
Configure the recorder print offsetYesYes
Configure the recorder heat adjustYesYes
Alert acknowledgment at marker button YesNo
TaskButtonsPC
Run the cyclic test
No
1
No
Read the options installedNoYes
Read the Error LogNoYes
Print the Error Log
Clear the Error Log
Ye s
Ye s
2
2
Yes
Yes
3
Write the serial number and set the optionsNoYes
Configure the language optionYesNo
Enable alert acknowledgementYesNo
Note Output to SystemYesNo
NIBP paper save modeYesNo
MECG trigger click volumeYesNo
response timeYesNo
FSpO
2
FSpO
Inop alarmYesNo
2
FSpO
alarm volumeYesNo
2
1. However you can run the Permanent Test, which is similar. See “Printing and Clearing the Error Log” on
page 24.
2. See “Printing and Clearing the Error Log” on page 24.
3. To print the error log from a PC, first save the error log from the pegserv tool as a text file.
20 Configuring the Monitor
Page 39
Configuring the Monitor Using Pushbuttons
You can set the time format, date format, IUP scale, paper format, recorder scale offset, recorder
heat and language option using pushbuttons. To change a setting:
1.Disconnect all transducers from the Monitor and disconnect or switch off Telemetry.
2.While pressing, press . The display shows C01 in the US1/US display and 0
or 1 in the Toco display.
3.Toggle through the menu by pressing or until you arrive at the menu item you
want to change. The menu items and their settings are shown in Table 3-2.
4.Press to change the setting (0 or 1).
5.Press to store the new settings. If yo do not press any keys for about 15 seconds,
the settings will be automatically stored.
1=RS232 System Interface on
Telemetry/System board active
C14Analog fetal movements0=off
1=on
C16NIBP paper save mode0=off
1=on
C17MECG trigger click volume0=off
1=quiet
2=medium
3=loud
C18
(XMO only)
C19
(XMO only)
C20
(XMO only)
response time
FSpO
2
FSpO
Inop alarm0=off
2
FSpO
alarm volume0=off
2
4
0=slow
1=fast
1=on
1=quiet
2=medium
3=loud
0
1
0
0
2
1
0
2
1. To find the correct setting, connect a Toco transducer to the monitor and then change the setting until the
trace is recording 20 units on the paper. Because of the 8-second time-out feature, and the delay between
changing the setting and seeing the change on the paper, you may have to repeat this procedure to set the
offset.
2. If the alert acknowledgement function is on, the fetal monitor adds the alert acknowledgement function on
the recorder marker button. This alert acknowledgement is given only to the central station after a request
from the central station (communication is via the digital interface.)
3. With this setting you can switch the control between the RS232 port on the new Telemetry/System board
M1350-66536 and the RS422 port implemented on the System Interface board M1350-66532. If only one
of them is built into the fetal monitor this one will be active regardless of the setting.
4. Response mode is an output-average filter. You can select whether the monitor averages samples over a short
time (approximately 10 seconds) or longer period (approximately 50 seconds). The times vary according to
the heartrate and signal quality and those quoted here are valid for good signal quality and a heartrate in the
120 bmp range. The fast response mode, because it requires fewer samples, allows you to see sudden changes
in output more easily. Fast response mode is indicated on the trace by .
1. The service setting C10=0 ensures the compatibility of older monitors with external
devices and will not normally be needed with the M1350B and M1350C
Table 3-4Service Setting C10 = 0 (DIP Switch Settings on the DSIF Board)
Connector 1
9 pin female
Dinamap 1846
(600Bd)
Philips CMS and
Philips 26/24 Series
(9600Bd)
Accutorr
(9600Bd)
Colin BP-8800
(4800Bd)
78352/4C
(9600Bd)
Connector 2
25 pin female
Nellcor N-200
(2400Bd)!!!!! XXX
Nellcor N-200
(2400 Bd) " ! ! !
Nellcor N-200
(2400Bd)
Nellcor N-200
(2400Bd)
Nellcor N-200
(2400Bd)
12345678
! " ! !
" " ! ! ! X X X
! ! " !
Switch Settings
! X X X
! X X X
! X X X
Configuring the Monitor 23
Page 42
Table 3-5Service Setting C12 Options
Printing and
Clearing the
Error Log
NST Timer
No No No 0
YesNoNo 1
Yes Yes No 2
NoNoYes 3
YesNoYes 4 (default)
YesYesYes 5 (Japan)
To print or clear the error log:
1.Disconnect all transducers from the monitor and disconnect, or switch off, Telemetry.
2.Make sure the recorder is on.
3.While pressing press
•C01 is shown in the US1/US display.
•0 or 1 is shown in the Toco display.
Auto Recording
Off
Acoustic Paper
Out Alert
Option \
Number
4.Press again to select the Function Menu:
•A01 (Print the Error Log) is shown in the US1/US display.
To print the Error Log, press . The display shows nnn nnn nnn.
•Press again to select A02 (Clear the Error Log).
To clear the Error Log, press. The display shows nnn nnn nnn.
The error log is divided into seven sections that contain the following information:
Section 1
1.The error code
For example 506
2.The number of times the error has occurred.
For example 0005 (note that the error counter uses hexadecimal notation)
3.The date and time the error first occurred.
For example 98/09/14 11:27
4.The date and time the error last occurred.
For example 98/09/14 16:23
The remaining lines in section 1 (if any) beneath the factory information code (FIC) are not
errors. Take no action.
Section 2
All built in options in the fetal monitor are written. For example Fet.Alert indicates fetal alerting.
Section 3
This shows information about the monitor itself:
DATE SW_REV INSTR_ID
Current date
1
Software revisionProduct number
and monitor serial number
1. This is taken from the monitor's internal clock.
Section 4
The final three digits of all boards connected into the fetal monitor. For example in Pos.05 is board
M1350-66536.
Section 5
The complete configuration made using the monitor's push buttons is listed for easy validation
and documentation.
Section 6
The revision number of the FSpO
software.
2
Example:
NPBFMHP5.0.1
Nellcor Puritan Bennett
Fetal Monitor
Manufacturer
Software revision number
Section 7
The error code, occurrences, and times and dates thereof are shown specifically for the FSpO
2
board in exactly the same way as other boards are displayed in Section 1 of the error log.
26 Configuring the Monitor
Page 45
Changing the
Time Format
To change the time format from AM/PM format to 24-hour format, do the following:
1.Disconnect all transducers from the Monitor and disconnect or switch off Telemetry.
2.While pressing press .
3.Press to change the setting to 1.
4.Press Test to store the new settings.
Full details about changing the time and date formats are given in Chapter 4.
Configuring the Monitor Using a PC
The service software kit (M1360C Option 875, part number M1360-68875) allows you to carry
out various extended configuration and service functions on the monitor. Contact your Philips
Sales Office for details.
The service software disk can be run on an industry standard PC. There is also an adapter cable
(M1360-61675) to connect the monitor to the PC.
Figure 3-1Cable for an Industry Standard compatible PC
Configuring the Monitor 27
Page 46
Installing the
Service
Program
Before you install the program, ensure that the serial port COM1 is set up as follows:
Transmission Rate9600 BPS
ParityNone
Word length (bits)8
Stop-bits1
This is done by giving the mode command (mode com1:96,N,8,1) at the DOS prompt.
Refer to your MS-DOS Manual for details.
To i ns t al l t he p ro g ra m:
1.Switch on the PC.
2.Create a directory for the program files. For example, to create a directory called service,
at the DOS prompt type in:
md service
and press.
3.Insert the program disk into drive A.
4.At the DOS prompt type A: and press .
5.At the A: prompt, type:
copy pegserv.exe c:\service
Serial Interface
Connection
and press . Where service is the directory you created to contain the program files.
The program is copied to c:\service
Make sure that the serial interface configuration on the PC is set up as follows:
Datacom-Configuration
ParameterSerial
Transmission Rate (BPS)9600
Word Length (bits)8
Stop-Bits1
ParityNone
%ON / %OFF PacingOff
CTS LineRegard
DSR LineRegard
DCD Line Ignore
Power to InterfaceOff
28 Configuring the Monitor
Page 47
Connecting
the PC to the
Monitor
Connect the cable from the PC to the service socket (1). If you are using an HP PC, use only the
COM1 port.
xmofull1.gif
Figure 3-2Service Socket
Loading the
Service
Program
1.Switch on both the monitor and the PC.
2.Now load the program.
If you are using the PC in DOS mode, do one of the following.
•If the program files are contained in a directory, change to that directory and enter:
pegserv
•If you want to load the program from drive A (or another drive):
i.Insert the program disk into drive A (or other).
ii.Select DOS-Commands and press .
iii. At the DOS prompt type A: and press .
iv. Type pegserv and press .
If you are using Windows 95/98/NT:
•Start the service program by double-clicking on the pegserv icon or on the filename
pegserv.exe
3.The program is now loaded and ready for use.
Configuring the Monitor 29
Page 48
Using the Service Program
When the program has been loaded onto your PC, the main menu is displayed. (If the menu is not
displayed, an error message is displayed along the bottom of the screen.)
Main Menu
Series 50 Fetal Monitor Configuration and Service Software Rev A.nn.nn
MESSAGE:
Select with >cursor keys< or >E, S, D, X< then press <ENTER> to execute!!
Main Menu
Configuration Settings
Recorder Adjustments
Service Menu
Exit
To select an item from the menu, move the cursor to the item you require and press :
"Use the arrow keys , , , keys on the keypad, or the E, S, D or X keys to
move the cursor.
Messages are displayed along the bottom of the screen while the Service Program is working, for
example:
Reading Recorder Adjustments from the Monitor.
Reading Serial Number from the Monitor.
Clearing the Error Log.
30 Configuring the Monitor
Page 49
Configuring
the Monitor
Series 50 Fetal Monitor Configurat ion and Service Softw are Rev A.nn.nn
Configuratio n Setting:
Function:Actual Settin g isSelect
Paper Scale :USUS EU
Time Format :AM/PMAM/PM 24h
Date Format :USUS EU
IUP Scale:mmHgmmHG kPA
MESSAGE:
Select with >cur sor keys< or >E, S, D, X< the n press <ENTER> to exec ute!!
To change a setting:
1.Using the appropriate keys, move the cursor to the selection you require.
2.Press .
To return to the Main Menu:
1.Move the cursor to Return to Main Menu.
2.Press .
Example
On the screen above, the actual setting of the paper scale is US. To change this to EU, move the
cursor to EU in the Select column and press . The program will reconfigure the monitor
for the European paper scale.
The settings on your monitor may be different from those shown.
Configuring the Monitor 31
Page 50
Adjusting the
Recorder
The Recorder Heat-Adjust is to compensate for different tolerances in the paper.
Series 50 Fetal Mo nitor Configuration and Service So ftware Rev A.nn.nn
Recorder Adju stments:
Recorder Heat Adjust:Actual Value is:|
Recorder Zero Adjust:Actual Value is:|
Return to Main Menu
MESSAGE:
Select with >cursor keys< or >E, S, D, X< then press <ENTER> to execute!!
Selected Value is|
∨
|……………|
Min Max
Selected Value is|
∨
|……………|
Left Right
To change a setting:
1.Using the appropriate keys, move the cursor to the new value you require.
2.Press .
To return to the Main Menu:
1.Move the cursor to Return to Main Menu.
2.Press .
Example
On the screen above, the actual value of the recorder zero-adjust is Left. To change this to
Right, move the cursor to Right in the Selected Value is: column and press.
The program will adjust the recorder accordingly. The values shown on your monitor may be
different from those shown.
32 Configuring the Monitor
Page 51
Service Menu To select an item from the menu, move the cursor to the item you require and press .
NoteBefore you select Cyclic Test, you must connect ultrasound, Toco and DECG transducers to
the monitor.
The serial number and software revision of your monitor may be different from those shown.
Series 50 Fetal Mo nitor Configurat ion and Service Software Rev A.nn.nn
Service Menu
Serial Number: 2948G00010
Software Revision: A.02.03
Cyclic Test
Read options
Read ErrorLog
Clear ErrorLog
Write Ser.nr./ Set Options
Fetal Heart Rate Alerting
MESSAGE:
Select with >cursor keys< or >E, S, D, X< then press <ENTER> to execute!!
Return to Main Menu
Running the Cyclic
The Cyclic Test is a continuous Self Test. Any errors are written to the Error Log and can be read
Test
using Read ErrorLog. Press any key to stop the test and return to the Service Menu. During
the test the keys on the front of the monitor are disabled.
Series 50 Fetal Mo nitor Configuration and Service So ftware Rev A.nn.nn
CYCLIC TEST is st arted
It starts a contin uous self test of the mo nitor
Press any key to sto p this test
Read ErrorLog to see the errors that occurred while sel f testing
MESSAGE:
Configuring the Monitor 33
Page 52
Reading the Options This lists the options installed on the monitor. Press any key to return to the Service Menu. The
options on your monitor may be different from those shown.
Series 50 Fetal Monitor Configuration and Service Sof tware Rev A.nn.nn
Implemented Options :
- Fetal Movement Profile
- External Twin Monitoring
MESSAGE:
Press any KEY to continue!
Reading the Error
Log
This displays the error log. (See Chapter 10 for a list of error log messages.)
Use >W< to Write ErrorLog Data to File >ErrLog.DAT<
use CursorKey >Down< or >X< to see next Page, <ENTER> to leave
The error log lists:
"The monitor's serial number.
"The error code and the number of times the error has occurred.
"The date and time the error first occurred.
"The date and time the error last occurred.
Up to 16 error messages can be displayed on a screen. To see the next or previous screen, use the s
or t keys on the keypad, or the E or X keys. At the end of the Error Log, No more Data
34Configuring the Monitor
Page 53
Clearing the Error
Log
available is displayed.
Press to print a page on LPT1.
Press to return to the Service Menu.
To write the error log data to a file, press the W key. The data is then stored as ERRLOG.DAT in
the directory containing PEGSERV.EXE. The data is stored as ASCII characters, and therefore
the file can be printed using the DOS Command print ERRLOG.DAT. The file is overwritten
each time you press the W key.
Series 50 Fetal Mo nitor Configurat ion and Service Soft ware Rev A.nn.nn
Service Menu
Serial Number: 2948G00010
Software Revision: A.02.03
Cyclic Test
Read options
Read ErrorLog
Clear ErrorLog Are you sure?
Write Ser.nr./ Set Options
Fetal Heart Rate Alerting
Return to Main Menu
MESSAGE:
Select with >cursor keys< or >E, S, D, X< then press <ENTER> to execute!!
The confirmation message Are You Sure? asks if you do want to clear the error log:
"Press Y to clear the error log, OR
"Press N.
You then return to the Service Menu.
Configuring the Monitor 35
Page 54
Writing the
Serial Number
You can write the serial number to the monitor only if you have replaced the CPU Board with an
exchange board. The serial number can be written only once to the exchanged board. If you
accidently enter the wrong serial number, you can reset it using an access code obtainable from
your Philips Response Center or from the Philips Technical Marketing Intranet page, “Softserver.”
Series 50 Fetal Mo nitor Configurat ion and Service Soft ware Rev A.nn.nn
NOTE: resettin g the Serial Number is protected by Acce ss Code
Serial Number: 3151G00026
Type in 4 digit code from factory TME and press <ENTER>
MESSAGE:
>----<
To write the serial number:
1.Using the keyboard, type in the serial number of the monitor. This is printed on the side and
on the bottom of the monitor
2.Press .
A confirmation message asks if the serial number is correct:
•Press Y if it is, OR
•Press N and retype the serial number.
You then go to the Set Options Screen to reset the options for the monitor.
36Configuring the Monitor
Page 55
Series 50 Fetal Mo nitor Configurat ion and Service Soft ware Rev A.nn.nn
Select Options according to the labels on the monitor
Option:Actual Setting is :Selection:
#C.01 TWIN:OFFOFF ON
#C.02 FMP:OFFOFF ON
Write Options
MESSAGE:
Select with >cursor keys< or >E, S, D, X< then press <ENTER> to execute!!
To select an option:
1.Using the appropriate keys, move the cursor to the selection you require.
2.Press .
To write the options to the monitor and return to the Main Menu:
1.Move the cursor to Write Options.
2.Press .
A confirmation message asks if the selected options are correct.
Configuring the Monitor 37
Page 56
Enabling/
Disabling FHR
Alerting
This screen enables you to enable or disable the fetal heart rate alerting. To change this
configuration you must enter an access code, which is available from your Philips Response Center
or from the Technical Marketing “Softserver” page on the Philips Intranet. There are two different
access codes, one enables the alerting, and the second disables alerting.
Series 50 Fetal Mo nitor Configurat ion and Service Soft ware Rev A.nn.nn
NOTE: Changin g the Alert Option is pr otected by Access Co de !
Alert Option is EN ABLED. Do you want to DISABLE the Alert op tion?
Press <Y> for YES or <N> for NO!
MESSAGE:
To enable or disable the alerting, enter the specific code to enable or disable the alerting.
38 Configuring the Monitor
Page 57
Getting Started (Basic Considerations
Setting the Time and Date
You can view and change the time, the date and the paper speed using the following keys:
Figure 4-1Setting the Time and Date
4
and Operation)
"Clock key displays the time (1).
"Paper speed displays the paper speed (2).
"Volume keys change the time, date and paper speed (3).
To set the time and date:
1.Press and releaseto display the current time. The Cardio 1/Combi display flashes to
show that the hour can be changed.
2.Press or to set the hour. Press and hold the keys to change the setting more
quickly.
3.Press and release and the Toco display flashes to show that the minutes can be
changed.
4.Press or to set the minutes.
5.Repeat the procedure:
•To set the month (in North America) or the day (in other countries).
•To set the day (in North America) or the month (in other countries).
•To set the year.
6.Press and release to return to the normal display.
You can also change the format of the time and date using the optional Barcode Reader and the
feature setting barcodes supplied with it.
Getting Started (Basic Considerations and Operation) 39
Page 58
Changing the
Time Format
To change from 12-hour to 24-hour format, enter
TIME FORMAT 24H (MILITARY) from the feature setting sheet.
To change from 24-hour to 12-hour format, enter
TIME FORMAT AM/PM from the feature setting sheet.
To change from Day.Month.Year to Month/Day/Year format, enter
DATE FORMAT NORTH AMERICAN from the feature setting sheet.
To change from Month/Day/Year to Day.Month.Year format, enter
DATE FORMAT EUROPEAN from the feature setting sheet.
If the batteries are not changed when they become low or empty of charge, the settings will return
to their default values: the date is set to 04.04.44 and the time is set to either 00.00 (European
Format) or 12:00A (US Format).
Setting Paper Speed and Scale
Setting the
Paper Speed
The paper speed can be 1, 2 or 3 cm/min. The default for North America is 3 cm/min; the default
for other countries is 2 cm/min. A change in paper speed will result in a change in the appearance
of an FHR trace, care must be taken to ensure ALL monitors in your institution are consistently
set at the same speed.
The ACOG technical bulletin on FHR monitoring states that “accurate pattern recognition is
difficult if not impossible at 1 cm/min and that 1 cm/min is only recommended for more
economic screening. When FHR abnormalities arise, the faster paper speeds will enhance FHR
pattern recognition.”
Use the Paper Speed Key to display the current paper speed and to return to the normal display.
You also return to the normal display automatically if you don't press any key for a few seconds.
When you return to the normal display, the new paper speed is set, and the time, date, speed and
monitoring modes are printed on the paper.
To set the paper speed:
1.Press and release to display the current speed.
2.Press or to set the speed.
3.Press and release to return to the normal display.
40 Getting Started (Basic Considerations and Operation)
Page 59
Setting the
FHR Paper
Scale
All chart paper available for the monitor is 2-channel chem/thermal paper, containing 150
numbered pages per pack. Each page is 100mm long and 151mm wide. Several configurations are
available, each with a 0 - 100 labor scale.
Table 4-1Types of Trace Paper Available
Product
Number
FHR
Scale
Color of
Grid
FMR
Filing
System
Packs
per Case
European
KPa scale
Highlighted
Three Lines
M1910A30 - 240OrangeNo40NoYes
M1911A50 - 210GreenNo40YesNo
M1913A50 - 210GreenNo40NoYes
Caution
Using recorder paper that is not approved by Philips can damage the monitor. This type of
damage will not be covered by warranty.
Some monitor configurations come with a fetal monitoring record (FMR) management system.
Included in the FMR system are labels that facilitate a total management system for organized
storage and retrieval of patient records.
If you wish to use paper with a different FHR scale (for example, 50 - 210 bpm instead of 30 - 240
bpm) you can change the paper scale settings:
1.Using the service software (see “Using the Service Program” on page 30).
2.Using pushbuttons (see “Configuring the Monitor Using Pushbuttons” on page 21).
Getting Started (Basic Considerations and Operation) 41
Page 60
Switching Logic Off and On
The default setting for arrhythmia logic is on. To change the setting, do the following:
"Connect a DECG transducer to the Cardio 1/Combi or Cardio 2 socket.
"Press repeatedly until is displayed.
"The Signal Quality Indicator shows:
•RED if logic is OFF.
•GREEN if logic is ON.
"Press or to change the setting.
"Press repeatedly to return to the normal display.
You also return to the normal display automatically if you do not press a key for a few seconds.
Separating FHR Traces (“Twins Offset”)
Twin monitoring is an option.
To help with the interpretation of traces with similar baselines, you can separate them.
Using Keys
Using the
Barcode Reader
"Connect two fetal heart rate transducers to the Cardio 1/Combi and Cardio 2 sockets.
"Press repeatedly to display .
"The Signal Quality Indicator shows:
•RED if the traces are NOT SEPARATED.
•GREEN if the traces are SEPARATED.
"Press or to change the setting.
"Press repeatedly to return to the normal display.
You also return to the normal display automatically if you do not press a key for a few seconds.
Enter “Twins Offset'' from the barcode sheet.
42 Getting Started (Basic Considerations and Operation)
Page 61
Switching FMP Off and On
FMP is an option.
The default setting for FMP is on.
Using Keys To change the setting, do the following:
"Connect a transducer to the Cardio 1/Combi socket.
"Press repeatedly until is displayed.
"The Signal Quality Indicator shows:
•RED if FMP is OFF.
•GREEN if FMP is ON.
"Press or to change the setting.
"Press repeatedly to return to the normal display.
You also return to the normal display automatically if you do not press a key for a few seconds.
Using the
Barcode Reader
Enter FMP Off or FMP On from the barcode sheet.
Getting Started (Basic Considerations and Operation) 43
Page 62
44 Getting Started (Basic Considerations and Operation)
Page 63
Monitor Rear Panel
5
System Interfaces
Figure 5-1Sockets for Peripheral Devices
1.Interface for optional barcode reader.
2.Interface for Philips Avalon CTS Cordless Fetal Transducer System (M2720A) and Series 50
T Fetal Telemetry System (M1310A). See “Connecting External Devices” on page 56 and
“Interface for Philips M2720A/M1310A” on page 46 for details.
3.System interface for connection to 80225A/80235A OBMS System and/or M1340A Fetal
Trace Transmitter or M1370A ODIS System (optional). See “System Interface Option J12”
on page 53 for details.
4.Serial RS232/RS422 system interface for connection of maternal monitors such as Philips
CMS. See “RS232/RS422 Dual Serial Interface Option J13” on page 55 for details.
5.RS232 digital system interface for connection to a Philips OB TraceVue system or an IBM
compatible PC. See the table “RS232 System Interface” on page 52 for details.
6.Serial RS232/RS422 system interface for connection of Nellcor N-200 maternal SpO
monitor (and N-400 fetal SpO
See “Rear Panel (Slot Positions)” on page 14 for a description of the board positions.
System Interfaces45
monitor for an M1350 XM without fetal SpO2).
2
2
Page 64
Interface for Philips M2720A/M1310A
The Telemetry interface allows connection to the Philips Fetal Telemetry Systems. Full
functionality is provided by the Philips Avalon CTS Cordless Fetal Transducer System (M2720A)
and Series 50 T Fetal Telemetry System (M1310A) with monitor software revision higher than
A.04.00.
The pin allocations for the various signals are shown below.
printelem.hpgl
Figure 5-2Telemetry Input Signal Allocations
For details of which parameters are supported with each monitor/interface combination, refer to
"Monitoring Using External Devices" in the Instructions for Use.
46 System Interfaces
Page 65
Table 5-1Telemetry Interface: Pin Connections
Pin Signal Function 1 (Telemetry)
Function 2 (External
Parameter)
1HR2HR-LF 2 MECG Analog: Input
2-12V Max. 100mA: Output
3+5V Max. 100mA: Output
4+12VMax. 100mA: Output
5S1
6S2
S3
Telemetry on (L)
DECG Inop (L)
1
:Input
1
Ext Par On (L)
1
INOP (L)1: Input
: Input
8S4DECG ModeExt Par Code MSB (scaling): Input
9S5Reserved ModeExt Par Code LSB (scaling): Input
10S6
11S7
12S8
13S9
US Mode (L)
IUP Mode (L)
Toco Ext Mode (L)
1
: InputCardio Channel (L1): Input
Toco Channel (L
1
1
Input
1
): Input
14HR1 (HR-LF)FHR low frequency 1
15TocoToco dc: Input
16Ext ParExt Par Analog: Input
17Gnd AnalogSignal GndSignal Gnd
18Gnd
19Mark
Telemetry Marker on (L)
1
20FMP onTelemetry FMP on
21FMTelemetry Fetal Movement event on
22
reserved (L)
1
23N.C. (if ID)New Interface= Do not
connect
Old Interface= connect to Gnd
24Gnd
25S10
HR2 Mode = MECG Mode (L)
1
1. If pin 5 is high (telemetry is off) the other inputs have different functions. Tele On (pin 5=L) and Ext.para.on
(pn 6=L) are not allowed at the same time.
HCMOS threshold:
L = Low = 0 to 1.5V
H = High = 3.5 to 5V
Pin 13 (S9). Toco Ext Mode (L). Toco mode programming. Activated Toco mode will be
displayed on the front panel.
48System Interfaces
Page 67
Pin 14 HR 1 Fetal Heart Rate INPUT. Input signal depends upon the selected HR mode. The
table below gives details of the input signals for each of the HR modes.
Fetal Heart Rate ModeInput signal Pin 14
US (S6 Low level)Doppler LF signal
Ri > 10 kΩ
Umax = ±3.5 Vpp
min = 35 mVpp
Pin 15 TOCO - DC INPUT
Ri > 10 kΩ
Input range = -3 V to +2 V
Sensitivity = -1 V Full scale
Pin 17. Signal ground for pins 14 and 15.
Pin 18 - 24. Ground for supply and status bits S1 to S10 activation. System outputs will be
controlled in the telemetry mode as in the front mode.
Pin 25. MECG - Mode. Input > (L). MECG mode programming. Activated MECG mode will
be displayed on the front panel. The input signal from pin 1 (MECG Analog Input) will be
processed by the instrument.
External
Parameter
Pin 6 (S2) External Parameter ON (L). The external parameter is printed on the recorder in
addition to the front mode information.
"Pin 5 (S1) must be High, if not appears in the display.
"Status bits S3 to S7 have function 2.
Pin 7 (S3) INOP (L). Printer disabled for the external parameter.
The external parameter is recorded but does not appear at the display and system interface output.
The digital serial interface can provide the external parameter signal.
Interface for Triplets Monitoring
Connect the M1360-61671 triplets cable to the 25 pin connector on the telemetry interface or
telemetry system interface of the M1350A or M1350B/C, and to the system interface of either the
8040A or the 8041A. Refer to the Instructions for Use for details.
RS232 System Interface
The 9 pin RS232 Telemetry interface system connector provides a digital interface protocol to
permit connection to a Philips Series 50 OB TraceVue system or an IBM compatible PC.
rs232sys.hpg
Figure 5-7RS232 System Interface Connector
52 System Interfaces
Page 71
NoteIf an OBMS Board (M1350-66532) is plugged in at the same time as the Combined Telemetry/
Digital System Interface Board (M1350-66536), the digital interface on the Telemetry Board is
switched off as per default. See Chapter 5 for information on how to activate RS232.
If the Software Revision installed is earlier than A.04.01 the RS232 system interface will remain
inactive.
System Interface Option J12
The system interface is provided for connection to the Obstetrical Display Information System
(ODIS), and the Obstetrical Information Management System (OBMS) Central Stations.
The pin allocations for the various signals are shown below.
Figure 5-8System Interface Connections
The following table lists the pin numbers, the signals at each pin and the signal details. If a
Telemetry/System Interface M1350-66536 is connected at the same time, you should check the
correct setting of C13. See Chapter 5 for the correct setting.
System Interfaces 53
Page 72
Table 5-2System Interface: Pin Connections
PinSignalSignal Details
1HR2 Output1 V/100 bpm ±40mV (0V if not valid) R_out = 100Ω
2Not connected
3RS422 Output +Digital Serial Output
4Toco Status OutputToco external/No Toco (L)
5HR1_valid OutputWhen heart rate is valid, output is high.
22External Test InputWhen input is low, external test is on.
23Instrument ID2.5V Regulated
24Chassis GroundChassis ground
1. In the event of a “paper end” or an “FHR coincidence” condition, an analog signal is applied to Pin 21 FHR
1 output. This lasts approximately 600msec and is repeated every minute while the condition remains. The
exact output voltage values are shown in the following table.
Internal Pullup 5kΩ
R_i = 100Ω
54 System Interfaces
Page 73
Table 5-3Pin 21: Output Voltage Values
FHR
Coincidence
Bit
0003.500 - 3.6503.575
0013.650 - 3.8003.725
0103.800 - 3.9503.875
0113.950 - 4.1004.025
1004.100 - 4.2504.175
1014.250 - 4.4004.325
1104.400 - 4.5504.475
1114.550 - 4.7004.625
Paper
End
Bit
Reserve
Bit
RS232/RS422 Dual Serial Interface Option J13
The Dual Serial Interface board provides data communication between the Philips monitor and
external devices. A model 8801 adapter is required from Critikon before the Dinamap 8100 can
be connected to the interface socket on the Philips monitor. External parameters are transmitted/
printed only if the monitor has no internal parameters installed.
1. An MHR measurement is provided in conjunction with maternal NIBP or SpO
2
Ye s
Ye s
NoNoNoNoYes
Yes
2
Yes
Temp
2
2
2
Ye s
Ye s
2
2
2
monitoring.
MHR
2
Ye s
2
Ye s
1
FSpO
2
No
No
System Interfaces55
Page 74
2. Only if this parameter is installed on the external device.
Measurement Priority Measurements made by internal maternal parameters (SpO
MHR) take precedence above the same measurements made by an externally connected device.
The MHR measurement obtained from the SpO
obtained from the NIBP measurement.
Connecting External Devices
The supported external devices are connected to the monitor via the Dual Serial interface at the
rear of the monitor. External devices connected to socket 1 that can monitor SpO
over the
Warning
Before connecting an external device to the monitor, connect the equipotential grounding
point (3) to earth potential. Use the grounding cable supplied with the monitor. The power
cord of the external device must be plugged into a wall-mounted power outlet - not into an
extension block.
The Dual Serial Interface has two sockets: see Figure 5-1.
To connect either a Philips CMS, a Philips 78352C or 78354C, a Dinamap 1846/8100, or a
Datascope Accutorr monitor follow the instructions below. The screws for the Philips CMS and
Philips 78352C or 78354C are metric and should be tightened by hand only, and not with a
screwdriver.
"Use interface cable M1350-61609.
"Connect the 9-pin end of the cable to the Philips Series 50 XM or XMO monitor.
"Connect the 25-pin end of the cable to the external device.
"Secure the interface cable by the screws at the cable connectors.
Figure 5-9Interface Cable M1350-61609
COLIN BP8800
To connect a COLIN BP-8800 Monitor to a Philips Series 50 XM or XMO monitor:
"Use the interface cable supplied by the COLIN Corporation with the NIBP monitor.
"Connect the 9-pin end of the cable to the Philips Series 50 XM or XMO monitor.
"Connect the 15-pin end of the cable to the COLIN.
"Secure the interface cable by the screws at the cable connectors.
Figure 5-10 COLIN Interface Cable
System Interfaces 57
Page 76
Nellcor N-200
or N-400
Monitor
To connect a Nellcor N-200 or N-400 monitor to the Philips Series 50 XM or XMO monitor:
"Use the interface cable M1350-61609.
"Connect the 25-pin end of the cable to the Philips Series 50 XM or XMO monitor.
"Connect the 9-pin end of the cable to the Nellcor N-200 or N-400.
"Secure the interface cable by the screws at the cable connectors.
"Set the dip switches on the rear panel of the Nellcor N-400 fetal SpO
monitor to:
2
•Up: 4, 8
•Down: 3, 6, 7,
This selects a baud rate of 2400 and an output format of “Conversation”. The positions
of the other dip switches do not matter.
The following tables list the pin numbers, the signal at each pin and the signal details.
Table 5-49-Pin Connector
PinSignalSignal Details
1- Connected to pin 4
2RxD (RS232)Received data input
3TxD (RS232)Transmitted data output
4Connected to pin 1
5GNDSignal ground
Table 5-525- Pin Connector
PinSignal Signal Details
2TxD (RS232)Transmitted data output
3RxD (RS232) Received data input
RxD-1 (RS422)Received differential data input +
7GNDSignal ground
9 TxD-1 (RS422)Transmitted differential data output +
10TxD-2 (RS422)Transmitted differential data output -
18RxD-2 (RS422) Received differential data input -
58 System Interfaces
Page 77
Figure 5-11 Dual Serial Interface: 9-Pin and 25-Pin
Philips V24/
V26
To connect a V24/V26 to the Philips Series 50 XM or XMO monitor:
"Use the interface cable M1353-61614.
"Secure the interface cable by the screws at the cable connectors.
System Interfaces 59
Page 78
External Device Baudrate Settings and Configurations
Philips CMS
Baudrate:9600
Startbit:1Parity:None
Databits:8TX/RX:High/Low
Stopbits:1Computer:On
Cable:M1350-61609 only
Port:Any port may be used, but ensure the above settings are stored
to the Philips CMS in configuration mode.
Philips 26/24
Baudrate:9600
Startbit:1Parity:None
Databits:8TX/RX:High/Low
Stopbits:1ComputerOn
Cable:M1353-61614
Philips 78352C
and Philips
78354C
COLIN BP8800
Baudrate:9600Switches:SW5 Open
Startbit:1SW6 Closed
Databits:8Parity:None
Stopbits:1
Cable:M1350-61609
Port:Only port B (sometimes referred to as port 2) can be used if
the Philips 78352C or 78354C is configured to 9600 Baud.
Baudrate:4800
Startbit:1
Databits:8Parity:Even
Stopbits:1
Cable:Supplied by the COLIN Corporation
60 System Interfaces
Page 79
Dinamap 1846/
8100
Datascope
Accutorr
Nellcor N-200
and N-400
Baudrate: 600
Startbit:1
Databits:8Parity:None
Stopbits:1
Cable:M1350-61609
Baudrate: 9600
Startbit:1
Databits:8Parity:None
Stopbits:1
Cable:M1350-61609
Baudrate: 2400
Startbit:1
Databits:8Parity:None
Stopbits:1
Cable:M1350-61609
System Interfaces 61
Page 80
External Maternal Measurement on the FHR Trace
Annotations on the trace of measurements made by external devices are always prefixed with an
asterisk, *. Maternal measurements made by the 50 XM/XMO monitor itself have higher priority
than maternal measurements made by a device externally connected to it.
Table 5-6External Maternal Measurement Frequency
Parameter
Measurement Printout
Interval
InternalExternal
SpO
2
NIBP
MECG WaveformOn demandNot available
TemperatureNot available5 minutes
1. If you are using automatic mode to measure maternal blood pressure, with a short interval between
repetitions, not all measurements will be recorded on the paper.
5 minutes
Each measurement
1
If you monitor maternal NIBP only, an MHR measurement is printed at the same time as the
NIBP measurement. If you monitor SpO
the maternal SpO
measurement, every 5 minutes.
2
, an MHR measurement is printed at the same time as
2
If you set the NIBP monitor to automatic mode, you must leave a minimum time interval between
each measurement. This time interval depends upon the paper speed setting.
Paper SpeedMinimum Time Interval
1 cm/min3 minutes.
2 cm/min2 minutes.
3cm/min1 minute.
62 System Interfaces
Page 81
Troubleshooting
You can use the following tables for solving general application problems and product-specific
problems that may occur. For solving general technical problems, refer to the troubleshooting
flowcharts.
ProblemPossible CausesSolutions
General
Not all maternal NIBP measurements are printed
on the trace and the NIBP Monitor is in
automatic mode.
Incorrect SpO
or Temperature measurements.The monitor has been switched on before the
2
ProblemPossible CausesSolutions
No maternal SpO2 and no MHR measurements
are printed on the trace.
ProblemPossible CausesSolutions
No FSpO2 annotation after Rec On in Toco grid.
mat HR and mat SpO
of trace.
appear as printout instead
2
The time interval between each measurement is
too short.
appropriate sensor was applied.
Nellcor N-200 Maternal SpO2 Monitor
The Nellcor monitor is powered by the internal
battery. (Battery power symbol is lit.)
Wrong DIP switch configuration of N-200
monitor.
Nellcor N-400 Fetal SpO2 Monitor
Wrong DIP switch configuration of N-400
monitor.
Wrong setting of C10.
No FSpO
Wrong C10 setting.
transducer plugged in.
2
Set the minimum time interval according to
paper speed. (See previous pages in this Chapter).
Apply sensor 5 minutes before the monitor is
switched on.
Switch on the AC power at the rear of the
monitor. (Battery power symbol is not lit.) See
also “FSpO2 Monitor” on page 164 for FSpO
troubleshooting instructions.
Check DIP switch configuration of N-200
monitor (see page 58).
Check DIP switches (see page 58).
Change C10 setting.
Connect transducer.
Change C10 > 10.
2
System Interfaces 63
Page 82
Philips CMS and Philips 26/24 Series
Problem Possible CausesSolutions
No maternal measurements are printed on the
trace.
No MHR measurement is printed on the trace.Incorrect parameter source.If the SpO
No maternal temperature is printed on the trace.TEMP1 is not labeled T1.
Mat.HR 0 and Mat.SpO
on the trace.
0% values are printed
2
The parameters are switched off.
Incorrect interface cable.
Incorrect Philips CMS software revision.
Other RS232 external devices are blocking the
communication to the Fetal Monitor.
Incorrect configuration of the RS232 port.
Temperature sensor is not connected to module
TEMP1.
The Nellcor and the Philips CMS are connected in
parallel and the Nellcor front power supply switch
is in the off position and the rear power supply
switch is in the on position.
Switch on the parameters in the Parameters On/
Off menu. (See the Operating Guide supplied with
the Component Monitoring System.)
Ensure the interface cable M1350-61609 is used.
(This number is printed on both ends of the cable.)
Ensure the Philips CMS has Revision C (or higher)
software.
Disconnect all other RS232 external devices,
except the fetal monitor from the Philips CMS and
turn the power supply switch for the Philips CMS
off. Wait at least 10 seconds, then turn the Philips
CMS power supply switch on again.
Check Philips CMS settings and refer to Philips
CMS Service Documentation.
HR/PULSE source to PLETH.
Change TEMP1 label to T1.
Check the cable connection of the temperature
sensor.
Ensure the Nellcor power supply switches (front
and rear) are both on or both off. Alternatively,
disconnect the Nellcor monitor completely (see
also page 164).
/PLETH module is plugged in, set the
2
64 System Interfaces
Page 83
Carrying Out the Checks
After you have installed the monitor, perform the following checks listed below:
1.Ensure you have loaded some paper and connected the power cord.
2.Turn the monitor on.
3.Check that the paper speed, and time and date are configured.
4.With no transducers connected, press .
You should see the following:
DISPLAY: All parts of the display are lit followed by all mode symbols. These will flash
alternately for about 10 seconds.
RECORDER: During the test the recorder speed will automatically set to 3cm/min and a
test pattern will be printed on the recorder paper to verify the condition of the thermal print
head and if the printer is correctly configured. See Figure 9-1. Recorder Test Pattern, for a
sample recorder test pattern.
6
Quick Installation Checks
5.Perform the Parameter Test as described in “Parameter Test” on page 112.
6.Check the Barcode Reader is correctly connected as outlined in Chapter 3.
The previous checks should verify the operation of the monitor. More complete tests, such as
transducer tests, are given later in this book.
Quick Installation Checks 65
Page 84
66 Quick Installation Checks
Page 85
Introduction
This chapter tells you how to upgrade the monitors, which options require which steps, and how
to restore the monitor’s original configuration. Many of the upgrade options (especially those
whose option number begins with “C”) require very similar steps to upgrade them, as you will see
from the tables.
Overview of Upgrade Options
There are many different upgrades possible. The following table shows you the upgrade option
numbers (upgrade option and part numbers are prefixed with M1360) and gives a brief
description of their function.
7
Upgrades
Existing Configuration
Series 50 XMSeries 50
add
Second US/Cardio channelC01C01C01C01
Fetal Movement ProfileC02C02C02C02
Maternal parameters onlyC04C03not applicableC30
FSpO
only C31C34C32not applicable
2
FSpO2 and maternal parametersC33C35not applicablenot applicable
latest software releaseS01S01S01S01
RS232 Interface (for Philips OB TraceVue)
or Telemetry Interface
Combined analog/digital system interface for
OBMS and ODIS
Maternal parameter interface to connect
external patient monitor
Series 50
IX
J10not applicablenot applicablenot applicable
J12J12J12J12
J13J13J13J13
without
maternal
parameters
with
maternal
parameters
XMO
without
maternal
parameters
Upgrades 67
Page 86
Work
Instructions at
a Glance
Each upgrade is described in a series of work instructions. Not all upgrades require all the work
instructions. The following table shows you exactly which instructions you must follow for the
upgrade you are performing.
Work Instruction
Upgrade
option
C01
C02
C03
C04
C30
C31
C32
C33
C34
C35
S01
J10
J12
1 2 3 4 567 8 910
---------
---------
!!!
-
!!!!
!!!
-
!!!!
!!!
-
!!!!
!!!
!!!
!!
!!
!!
------
------
--------
-
-
!!
-
--- -
!!
-
--- -
!
!!!
-
--- -
!
!!!
-
!!
!
!!
!
!
!!! ! !
!
!
!
!
-
-
-
-
-
-
-
-
-
-
J13
68 Upgrades
!!
------
!
-
Page 87
Options at a
Glance
The following table gives you an overview of what is involved in each upgrade option. All “Cxx”
upgrades include a mandatory CPU firmware upgrade for the ROM board.
OptionAdds.....Factory ships.....You must.....
C01second US/Cardio
• upgrade key• upgrade and test the monitor
channel
C02Fetal Movement Profile
C03maternal parameters to
an XM without maternal
capability
C04maternal parameters to IX• installed new base assembly
• new display
• NIBP board
• SpO
board
2
• new CPU firmware (4
EPROMs)
• new CPU firmware (4
EPROMs)
• use original base assembly
• fit new display
• connect NIBP and SpO2 and fit both boards
• upgrade ROM board with new CPU firmware
• configure and test monitor
• transfer original cover, recorder, power supply and
ROM board to new base assembly
• upgrade ROM board with new CPU firmware
• configure and test monitor
C30maternal parameters to
an XMO without
maternal capability
C31FSpO
only (no
2
maternal parameters) to
an IX
C32FSpO2 to an XM that
already has maternal
capability
• new display
• NIBP board
• SpO
board
2
• new CPU firmware (4
EPROMs)
• installed new base assembly
• new CPU firmware
• new display
• FSpO
board
2
• new CPU firmware (4
EPROMs)
• use original base assembly
• fit new display
• connect NIBP and SpO2 and fit both boards
• upgrade ROM board with new CPU firmware
• configure and test monitor
• transfer original cover, recorder, power supply and
ROM board to new base assembly
• upgrade ROM board with new CPU firmware
• configure and test monitor
• use original base assembly
• fit display assembly
• Fit FSpO
board
2
• upgrade ROM board with new CPU firmware
• configure and test monitor
C33maternal parameters and
FSpO
to an IX
2
• installed new base assembly
• new CPU firmware (4
EPROMs)
• transfer original cover, recorder, power supply and
ROM board to new base assembly
• upgrade ROM board with new CPU firmware
• configure and test monitor
C34FSpO
without maternal
capability
C35FSpO
parameters to an XM
without maternal
capability
only to an XM
2
and maternal
2
• FSpO2-only display
• FSpO
board
2
• new CPU firmware (4
EPROMs)
• new display
• FSpO
board
2
• NIBP board
• SpO
board
2
• new CPU firmware (4
EPROMs)
S01latest software• new CPU firmware (4
EPROMs)
• use original base assembly
• fit FSpO
board
2
• fit new display
• upgrade ROM board with new CPU firmware
• configure and test monitor
• use original base assembly
• connect NIBP and SpO
• fit FSpO
board
2
and fit boards
2
• fit new display
• upgrade ROM board with new CPU firmware
• configure and test monitor
• upgrade ROM board with new CPU firmware
• upgrade NIBP board with new firmware
• new NIBP firmware
J10RS232 interface for OB
TraceVue, or Telemetry
interface
J12combined analog/digital
• RS232/Telemetry board
• new CPU firmware (4
• upgrade ROM board with new CPU firmware
• Fit new board. Series 50 IX only.
EPROMs)
• Analog/digital interface board• Fit new board
system interface for
OBMS and ODIS
Upgrades 69
Page 88
OptionAdds.....Factory ships.....You must.....
J13maternal parameter
interface
• Interface Board
• CPU firmware (4 EPROMs)
• Cable to link monitor to external device
• Fit new interface board
• upgrade ROM board with new CPU firmware
Initial Inspection
The upgrades are supplied packed in protective shipping cartons. Before unpacking, visually check
the packaging and ensure there are no signs of mishandling or damage. Using the table above,
ensure that you have received the correct components for the upgrade option number, check that
the contents are complete and that you have the correct upgrade.
Claims for
Damage
Follow the instructions in the section “Unpacking and Checking the Shipment” on page 3 if the
shipping cartons show signs of damage.
What You Need
Upgrading a monitor requires simple tools:
"pozidrive screwdriver size 1
"safety test equipment
"PC for configuration
"Configuration software “pegserv.exe”
"Cable to link PC to fetal monitor
Before You
Start
Warning
Disconnect the electrical power to the monitor before you remove any component. Follow
the necessary electrostatic discharge (ESD) procedures throughout the upgrade process.
70 Upgrades
Page 89
Instruction 1 Verifying Monitor for Upgrade
Before you start the upgrade process, ensure that you are upgrading the correct monitor and that it
is working properly.
Options Requiring this Procedure
C01C02C03C04C30C31C32C33C34C35S01J10J12J13
--
!!! ! !!!!!!!!
1.Check the monitor’s serial number as described in the section on “Instrument Identification”
on page 5.
When you have upgraded the monitor, remove the old serial number labels from the
monitor and replace them with the new labels supplied in the upgrade pack.
2.Ensure that the monitor you are about to upgrade works correctly. Perform a Quick Test
and a Parameter Test on the monitor. These are detailed in Chapter 9.
If the monitor does not successfully complete its service tests, do not upgrade it. Make
appropriate arrangements for its repair. It is more expensive in time and money to
troubleshoot after an upgrade.
Instruction 2Checking Current Configuration
The first step in the upgrade process for all upgrade options is checking the current configuration.
You do this by printing the error log to provide you with a record of the current configuration.
After you have finished upgrading the monitor, you will need this record to help you to restore the
initial configuration. See page 24 for instructions on printing the error log.
Options Requiring this Procedure
C01C02C03C04C30C31C32C33C34C35S01J10J12J13
--
Upgrades 71
!!! ! !!!!!!!!
Page 90
Instruction 3Removing the Top Cover
p
See “Removing the Top Cover” on page 167 for instructions.
Options Requiring this Procedure
C01C02C03C04C30C31C32C33C34C35S01J10J12J13
--
!!!!!!!!
Instruction 4Using the New Base Assembly
When you upgrade a Philips Series 50 IX monitor, you receive a new base unit that is fully fitted
with the options that were ordered. All you have to do is transfer some components from the
original monitor into the new base unit and perform a firmware upgrade. From the original
monitor, you must reuse the cover, the power supply, the recorder and the ROM board.
Options Requiring this Procedure
C01C02C03C04C30C31C32C33C34C35S01J10J12J13
---
Reusing the
Power Supply
!
To remove the power supply from the monitor:
-
!
!
-
------
----
sm20sca.tif
72Upgrades
Page 91
1.Undo the four screws at the rear of the power supply.
2.Pull the power supply out of the rear of the monitor.
3.Put the power supply into the new card cache of the upgrade kit. This is basically a reversal
of the above procedure. Hold the ON/OFF switch (on the front of the monitor) as this can
be dislodged while inserting the power supply. If you have trouble locating the power supply
into the rear of the monitor, remove and check that the pins connecting the supply to the
backplane are not bent. Reinsert the power supply (see page 14 for the correct location).
Reusing
Boards
Depending on the configuration of the original monitor, you may have to swap up to three other
boards from the original monitor into the new one. The technique for removal and replacement is
the same for all of the boards. These boards are:
"Telemetry interface
"Dual serial interface
"OBMS interface
1.Undo screws (1) and (2) at the top and bottom of the board.
2.Move lever (2) upwards and pull the board out of the rear of the monitor.
2
1
psm23sca.tif
3.Put the reused board(s) into the appropriate slot(s). See page 14 for details.
4.Replace and tighten the screws.
Upgrades73
Page 92
Using the
p
Existing
Recorder
To swap the existing recorder to a new unit, follow the instructions in the section on “Recorder
Assembly” on page 174.
sm26sca.tif
Then replace the cover on the monitor, using the four screws you reserved earlier.
Labeling 1.Stick the option indicator label (which indicates whether the monitor has FMP and/or twin
capability) beneath the remote event marker socket.
2.Place the upgrade serial number label on the righthand side of the cover so that it can be
easily seen.
74Upgrades
Page 93
Instruction 5Removing the Display Assembly
To remove the display panel you need first to remove the front end board, and then the digital
interface board. This applies to upgrades to XM and XMO monitors.
Options Requiring this Procedure
C01C02C03C04C30C31C32C33C34C35S01J10J12J13
--
ARemoving
the Front E n d
Board
BRemoving
the Digital
Interface
Board
!
See “Frontend Board” on page 176 for instructions on how to remove the frontend board.
!
-
-
!
-
!!
----
difboard.hpg
1.Disconnect all the cables:
•loudspeaker cable (1)
•power LED connector (2)
•recorder switch board connector (3)
•recorder sensing board connector (5). Unlock the flex layer connector by pulling it
slightly forwards.
Upgrades75
Page 94
•recorder stepper motor (6)
•frontend board connector (7)
•backplane flat cable (13)
2.Disconnect the recorder print head (4) from the recorder assembly.
3.Remove the two screws (8) and (11) that hold the board in place.
4.Slide the digital interface (DIF) board towards the rear of the monitor before lifting the
board up, to give the marker connector a “chance”.
5.Remove the label from the front end connector area. Score carefully around the connectors
in the area with a flat bladed knife. Do not allow any adhesive to remain on the front panel.
CRemoving
See “Display Assembly” on page 180 for instructions on how to remove the maternal display panel.
the Maternal
Display Panel
Next StepIf your upgrade is Option C03, C30 or C35 go to “Adding SpO2 and NIBP Capability” on page
77.
If your upgrade is Option C32 or C34, now go to “Adding an FSpO2 Board” on page 79.
76Upgrades
Page 95
Instruction 6Adding SpO2 and NIBP Capability
m
This adds maternal pulse oximetry and blood pressure measurement capability.
Options Requiring this Procedure
C01C02C03C04C30C31C32C33C34C35S01J10J12J13
--
!
!
-
- ---
!
----
atcable.hpg
1.Remove the two plastic parts that pad the two holes already prepared for the NIBP and
SpO
connectors (screws 1 and 2).
2
2.Remove the three cover-blanks that cover slots 4 and 5 (NIPB) and 9 (SpO
of the monitor.
3.Fix the NIBP tubing and SpO2 cabling, with their connectors, to the front panel. Do not
force the screws drilled into the plastic on the front panel.
Upgrades77
), from the rear
2
Page 96
4.Replace the DIF board into the monitor, checking that the marker connector is raised
s
through the front end. Reconnect all the cables.
5.Secure the earth connection mounting (2) to hold the SpO
cable in place and to ensure the
2
correct shielding.
6.Place the SpO
cable and the NIBP tubing in wide soft curves from the front to the rear card
2
cache and feed them through the appropriate slots.
7.Insert the SpO2 and NIBP boards halfway into the appropriate slots at the rear of the
monitor. Connect the tubing and the cable and then fully insert the boards. Screw the boards
into position with the provided screws.
8.Attach the self-adhesive holders fixing cable and tubing (1) to the top of the card cache so
that it allows any subsequent exchange of the boards without removing the top cover. Make
sure that the cable and tubing lie in a soft curve without any crossover on top of the card
cache.
po2nib.tif
9.Put the frontend board back into the monitor, sliding it towards the front to click the
connectors back into their clips. Ensure that all connectors are fully inserted (check this from
the front) before you screw the board back in place.
10. Replace the top cover, ensuring that the cable and tubing are not squeezed.
Next StepIf your upgrade is Option C32, C34 or C35, go to “Adding an FSpO2 Board” on page 79.
If your upgrade is C03, C30, or C34, go to “Upgrading ROM Board EPROMs” on page 80.
78Upgrades
Page 97
Instruction 7Adding an FSpO2 Board
This adds fetal pulse oximetry capability.
Options Requiring this Procedure
C01C02C03C04C30C31C32C33C34C35S01J10J12J13
-----
1.Replace the DIF board into the monitor, checking that the marker connector is raised
through the front end. Reconnect all the cables.
2.Put the frontend board back into the monitor, sliding it towards the front to click the
connectors back into their clips. Ensure that all connectors are fully inserted (check this from
the front) before you screw the board back in place.
3.Remove the cover blank from slot 9 at the rear of the monitor.
4.Insert the FSpO
5.Screw the board into position with the screws provided.
6.Carefully remove the existing product label from the front of the monitor. The existing label
says either “Series 50 IX” or “Series 50 XM”. Replace this with the label that says “Series 50
XMO”(1).
board into the monitor. There is no tubing or cabling to connect.
2
!!!!!
----
xmolabel.gif
7.Go to “Upgrading ROM Board EPROMs” on page 80.
Upgrades79
Page 98
Instruction 8Upgrading ROM Board EPROMs
Most upgrades require new EPROMS. You must fit them now and then reconfigure the monitor.
For upgrade S01 this is the only procedure you need.
Options Requiring this Procedure
BoardC01C02C03C04C30C31C32C33C34C35S01J10J12J13
ROM
Board
NIBP
Board
--
- - --- - ----
1.Undo screws (1) at the top and bottom of the ROM Board.
2.Move lever (2) upwards and pull the board out of the rear of the monitor.
3.Using a flat-blade screwdriver, replace the EPROM set, according the diagram below.
!!! ! !!!!!!
!
---
!
-
pegsm34.hpg
4.Push the board back into the slot inside the rear of the monitor.
5.Replace the screws and tighten securely.
6.Stick the label (3) provided with the new EPROMS onto the board as shown.
7.Perform the appropriate safety tests, and initiate the monitor’s self test before allowing the
monitor to be used on a patient.
80Upgrades
Page 99
8.Carefully attach the new label with the additional holes for the NIBP and SpO
connectors
2
to the front panel.
9.If your S01 upgrade includes a new ROM for the NIBP board, follow these instructions:
a.Remove the NIBP board from the monitor. There is no need to remove the cover.
b.Undo the screws at the top and bottom of the NIBP board, which is in Slots 4 and 5 at
the rear of the monitor.
c.Move the lever upwards and pull the board out of the rear of the monitor.
d.Disconnect the part of the NIBP tubing that leads to the display panel at the front of
the monitor (1)
e.Remove screws (2) and (3) from the NIBP tubing board and gently ease the board
forwards so you can access the NIBP board beneath.
2
3
NIBP tubing board
NIBP
NIBP Board
tubing
f.Lever the NIBP EPROM gently off the board using a flat bladed screwdriver and
replace it with the upgrade EPROM.
NIBP EPROM
NIBP Board
1
g.Replace the NIBP tubing board and tighten the screws.
h.Reconnect the NIBP tubing that leads to the display panel.
i.Slide the NIBP board into the rear of the monitor and refasten the screws.
Upgrades81
Page 100
Instruction 9Reconfiguring the Monitor
When you exchange EPROMs on the ROM board you must rewrite the new serial number and
restore the original options configuration (Twins, FMP, Fetal Alerting).
Options Requiring this Procedure
C01C02C03C04C30C31C32C33C34C35S01J10J12J13
--
Restoring
Service
Settings Using
Pushbuttons
!!! ! !!!!!
1.If you have not yet done so, print the error log (see “Printing and Clearing the Error Log” on
page 24). This will serve as a useful record of the current monitor configuration.
2.Enter the new serial number from the serial number labels (included in this upgrade kit).
This must be done using the service software: see “Writing the Serial Number” on page 36
for instructions. A serial number can be written only once to the exchanged board. If you
accidentally confirm an incorrect serial number you can reset it using an access code
obtainable from your Philips Response Center.
3.You will be prompted for the option configuration. You can set all the option settings using
the pushbuttons, as described in “Writing the Serial Number” on page 36. Using the error
log you printed previously, ensure that you restore only those options that the customer had
before you commenced the upgrade.
4.Double check the settings carefully. If you accidentally confirm an incorrect serial number
you can reset it using an access code obtainable from your Philips Response Center.
5.Step back to the main menu of the service program and select “Configuration Tasks” to set
the paper speed, time format, and so forth, or perform the self tests, or read the error log.
You will find detailed instructions for all these tasks in Chapter 3, “Configuring the
Monitor.”
Using the data from the error log you must now restore the customer’s original service settings. Of
course, if you have added new functionality, you should ensure that default settings for the new
parameter(s) are satisfactory for the customer and if not, change them. See “Configuring the
Monitor Using Pushbuttons” on page 21 for instructions on changing the settings.
---
82 Upgrades
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.