Upgrades Series 50 IX Fetal Monitor (M1350A) to
Series 50 XM (M1350B)
SERVICE SUPPLEMENT
Important! Return Your M1350A Components
Please call your local Philips Response Center for details of the return
procedure and address.
Please clean and disinfect the components of the M1350A monitor that you did
not reuse according to the instructions in this document, and return them, reusing
the kit carton, to the address given to you by your local Philips Response Center.
This entitles you to the exchange part pricing for the Support Upgrade Kit.
Otherwise, you will be charged the new part price.
Disposal of the returned parts will be handled by Philips.
Printed in Germany
January 2004
*M1360-9080B*
M1360-9080B
4512 610 01721
Page 2
What the M1360-69201 Support Upgrade Kit is For
What the M1360-69201 Support Upgrade Kit is For
If the front panel of your M1350A monitor is damaged or broken, you need the M1360-69201 Support
Upgrade Kit for repair. This service supplement tells you how to do it. This upgrade/repair should be
carried out by qualified service personnel, either by the hospital’s biomedical department, or by Philips
Support.
For further information on the fetal monitors please refer to the Instructions for Use and Service Guide for
your monitor.
Kit Contents
The M1360-69201 Support Upgrade Kit consists of:
• Chassis assembly fitted with:
– Front panel
–Frontend board
– Digital interface board
– Loudspeaker
– Display board
• New serial number label
•EPROMS
• Fetal Monitoring Documentation CD-ROM
• This Service Supplement
What You Need to Do
To complete the repair/upgrade, you need to remove the following components from your existing
monitor and fit them into the new chassis supplied in the Support Upgrade Kit:
•Top cover
•Power supply
• All original boards except those already fitted to the new chassis assembly (see “Kit Contents” above).
•Recorder
You need to enter the new serial number and configure the monitor with the appropriate settings using
the “pegserv.exe” service support software.
NOTECall your local Philips Response Center to obtain:
• Access codes for serial number, option settings and FHR Alerting
• Instructions for returning parts from the M1350A, and a return address
2
Page 3
Initial Inspection
Initial Inspection
The upgrades are supplied packed in protective shipping cartons. Before unpacking, visually check the
packaging and ensure there are no signs of mishandling or damage. With reference to the kit contents,
ensure that you have received the correct components and that the contents are complete.
Claims for Damage
If the shipping cartons show signs of damage, contact the carrier and arrange for an inspection to be made.
If any of the equipment supplied is damaged, you should contact both the carrier and your local Philips
service organization. Arrangements will then be made for repair or replacement, as appropriate.
What You Need
Upgrading a monitor requires simple tools:
• Pozidrive screwdriver size 1
• Safety test equipment
• PC for configuration
• Service software “pegserv.exe”
• Cable (part number M1360-61675) to link PC to fetal monitor
• Access code for serial number, and FHR Alerting, if applicable (obtain from your local Philips Response
Center). Record the serial number, its access code, and the FHR Alerting enabling code in the boxes
provided:
Serial Number:
Serial Number
Access Code:
FHR Alerting Enabling Code:
3
Page 4
What You Need
Instruction 1: Checking Current Configuration
The first step in the upgrade process is checking the current configuration. You do this by reading out the
configuration settings of the monitor with the “pegserve.exe” service software. See the Service Guide on the
enclosed CD-ROM for instructions on how to read the configuration settings.
Configuration Settings
The following table shows configuration settings for the monitor. Not all configuration settings are
available for every monitor, but can vary according to software revision. The “Menu Setting” is included
for reference purposes only.
NOTEOnly configuration settings C01 through C06 are handled by “pegserv.exe”.
Configuration Table
Menu
Setting
C01Time setting0 = AM/PM
C02Date form at0 = US
C03IUP format0 = mmHg
C04Paper format0 = 30-240 bpm
C05Recorder offset0 .. 11
C06Recorder heat0 .. 11 (SET THIS TO 11)
C07Language option1 = English (US)
2 = French
3 = German
4 = Dutch
5 = Spanish
6 = Italian
10 = Japanese
13 = Chinese (simplified)
17 = Russian
0 = off
1 = on
1 = Roman-8
1 = 50% gain
0 .. 5
Record Your
Setting
4
Page 5
What You Need
Configurat ion T a ble
Menu
Setting
DescriptionChoices
C13Serial port selection0 = serial port on the
System Interface board
(RS422) set to active
1 = serial port on the
Telemetry board (RS232)
set to a ctive
C14Analog FMP0 = analog fetal
movement print-out OFF
1 = analog fetal
movemen t ON
C15Not used
C16NiBP paper save mode0 = off
1 = on
C17MECG trigger click
volume
0 = off
1 = quiet
2 = medium
3 = loud
C18FSpO
response time0 = slow
2
1 = fast
C19FSpO
inop alarm0 = off
2
1 = on
C20FSpO
alarm volume0 = off
2
1 = quiet
2 = medium
3 = loud
Record Your
Setting
5
Page 6
What You Need
Instruction 2: Removing the Top Cover
WARNINGElectrical power is dangerous. Disconnect the electrical power to the monitor before you remove any
component. Follow the necessary electrostatic discharge (ESD) procedures throughout the upgrade
process.
Access to most of the items within the monitor is only possible with the top cover of the monitor
removed.
To remove the top cover:
1 Turn the system off and disconnect the power cable.
2 Undo the four screws situated on the sides of the monitor, and keep them safe for reuse.
3 Slide the cover towards the rear of the monitor and lift it off.
You will later refit the cover to the new chassis assembly supplied with the Upgrade Kit. Replacement of
the cover is a reversal of the above procedure.
6
Page 7
What You Need
Instruction 3: Using the New Chassis Assembly
When you upgrade a Philips Series 50 IX monitor, you receive a new chassis unit that is fitted with the
components listed under “Kit Contents” on page 2. All you have to do is transfer some components from
the original monitor into the new chassis unit, perform a firmware upgrade, set the new serial number and
configuration settings. From the original monitor, you must reuse the cover, the power supply, the
recorder, the CPU board and the ROM board.
Reusing the Power Supply
To remove the power supply from the monitor:
1 Remove the ON/OFF button by pulling it forwards. You hear a click as the end of the buttons’s
extension arm comes forward one notch from the switch unit located in the power supply. Release the
end of the extension arm with the aid of a flat-bladed screwdriver. Turn the button/arm 90 degrees
counter-clockwise, then pull it straight out.
2 Undo the four screws at the rear of the power supply.
3 Pull the power supply out of the rear of the monitor.
4 Put the power supply into the new card cage of the upgrade kit. This is basically a reversal of the above
procedure. DO NOT force the Power Supply into the rear of the monitor. If it will not locate, remove
and check that the pins connecting the supply to the Backplane are not bent. Reinsert the supply.
5 Replacement of the ON/OFF button is a reversal of the removal procedure. You should hear two clicks
as the end of the button’s extension arm locates on the notches on the switch unit in the power supply.
7
Page 8
Reusing Boards
Remove and reuse all the original boards, (except those already fitted to the new chassis assembly; see “Kit
Contents” on page 2), from your existing monitor and fit them into the new chassis supplied in the
Support Upgrade Kit.
The technique for removal and replacement is the same for all of the boards.
1 Undo screws (1) and (2) at the top and bottom of the board.
2 Move lever (2) upwards and pull the board out of the rear of the monitor.
What You Need
2
1
8
Page 9
What You Need
3
Put the new board(s) into the appropriate slot(s):.
3
1A1
Figure 0-1Rear Panel (Slot Positions)
Slot
Number
1Power Supply with Power Cord Connection, including Equipotential Grounding Point (1A)
2ROM Board (M1350-66506)
3CPU Board with Barcode Reader Interface (M1350-66513)
4Telemetry Interface Connector (M1350-66536) and System Interface Connector (RS232) (lower)
If the monitor has NIBP (never applies to M1350A):
5 & 6External Blood Pressure Board (M1350-66535). Optional. Connected to Slot 5, but physically occupies
7Digital Signal Processor (DSPII) (M1350-66507).
If the monitor does not have NIBP:
5Slot empty (there is a single slot cover in this case).
6Digital Signal Processor-CoP Board (M1350-66505). Optional. (Note: only in conjunction with M1350-
7Digital Signal Processor-CPU Board (M1350-66504). Optional. (Only with M1350-66505)
Other monitor options:
8Maternal SpO
9OBMS/ODIS Analog Interface (optional) (M1350-66532). Optional.
both Slot 5 and Slot 6 (there is a double slot cover in this case).
66504.)
OR Digital Signal Processor (DSPII) (M1350-66507). Optional. Here, Slots 5 and 6 are empty.
Board (M1350-66534). Optional.
2
Interface (M1350-66534). All optional.
Board (M1350-66540). Optional.
2
2
4578
6
9
10
11
NOTE
The DIF board (M1350-66515) is located in Slot 0 inside the monitor (not visible from the rear).
Reuse the rear blank covers from the M1350A monitor as appropriate.
4 Replace and tighten the screws.
9
Page 10
Using the Existing Recorder
To swap the existing recorder to a new unit, follow these instructions:
What You Need
1 Turn the system off and disconnect the power cable.
2 Remove the recorder paper.
3 Disconnect the two earth straps 1 from the front assembly.
NOTETake care not to lose the small O-rings when you have removed screws 2 and 3.
4 Loosen screws 2 and 3. Screws 2 are accessible via holes in the paper tray.
5 Disconnect the cables 4 and 5 between the Digital Interface board and the recorder. Remove screw 3.
6 Unlock the flex layer by pulling the plastic part A slightly forward, and then pull the flex layer to release
the flex.
A
7 Remove the recorder assembly by first sliding it away from the front panel, and then lifting it out.
10
8 Place it in the new chassis assembly, reversing the above procedure. Take care that the paper eject lever
fits back into the paper eject knob.
Then replace the cover on the new chassis, using the four screws you reserved earlier.
Page 11
What You Need
Labeling
Place the new serial number label on the right-hand side of the cover so that it can be easily seen.
Instruction 4: Upgrading ROM Board EPROMs
Upgrades require new EPROMS. You must fit them now and then reconfigure the monitor.
1 Undo screws (1) at the top and bottom of the ROM Board.
2 Move lever (2) upwards and pull the board out of the rear of the monitor.
3 Using a flat-blade screwdriver, replace the EPROM set, according the diagram below.
pegsm34.hp g
4 Push the board back into the slot inside the rear of the monitor.
5 Replace the screws and tighten securely.
6 Stick the label (3) provided with the new EPROMS onto the board as shown.
Instruction 5: Reconfiguring the Monitor
Reconfigure the monitor options after upgrade using the “pegserv.exe” service software. You can find the
“pegserv.exe” service software on the Fetal Monitoring Documentation CD-ROM, included with the
Support Upgrade Kit.
When you exchange EPROMs on the ROM board you must rewrite the new serial number and restore
the original options configuration (Twins, FPM, FHR Alerting). FHR Alerting is not approved for use in
the USA.
1 Switch on the monitor, use the “pegserv.exe” service software for option configuration, using the access
codes you obtained from your local Philips Response Center.
11
Page 12
Important! Return Your M1350A Components
2
Enter the new serial number from the serial number labels (included in this upgrade kit). This must be
done using the service software: see “Writing the Serial Number” in Chapter 4 of the Service Guide, on
the CD-ROM included with this kit, for instructions. If a wrong serial number has been programmed a
different access code is required corresponding to this serial number. If the correct serial number is
programmed then the FMP and US twi
code.
3
You will be prompted for the option configuration. Ensure that you restore only those options that the
customer had before you started the upgrade.
4 Double check the settings carefully.
5 Step back to the main menu of the service program and select “Configuration Tasks” to set the paper
speed, time format, and so forth, or perform the self tests, or read the error log. You will find detailed
instructions for all these tasks in Chapter 4 of the Service Guide on the CD-ROM included in this kit.
n options can be changed again and again with the identical access
Restoring Service Settings Using Push Buttons
Using the data from the configuration table starting on page 4, you must now restore the original service
settings. Of course, if you have added new functionality, you should ensure that default settings for the
new parameter(s) are satisfactory for the customer and if not, change them. Set C17 to ‘2’. Record any
new settings in the table starting on page 4. See “Configuring the Monitor Using Push buttons” in
Chapter 4 of the Service Guide, on the CD-ROM included with this kit, for instructions on changing the
settings.
Important! Return Your M1350A Components
Please call your local Philips Response Center for details of the
return procedure and address.
Please clean and disinfect the components of the M1350A monitor that you
did not reuse according to the instructions in this document, and return
them, reusing the kit carton, to the address given to you by your local
Philips Response Center. This entitles you to the exchange part pricing for
the Support Upgrade Kit. Otherwise, you will be charged the new part
price.
Disposal of the returned parts will be handled by Philips.
12
Page 13
Safety Tests
Safety Tests
This section defines the test and inspection procedures applicable to the upgrade of the fetal monitor.
Refer to the Test and Inspection Matrix on page 14 to determine what test and inspection results must be
reported after an installation or an exchange of system components has been carried out, and what safety
tests to carry out.
WARNINGSafety test requirements are set according to international standards, such as IEC/EN 60601-1 and IEC
60601-1-1, their national deviations, such as UL2601-1, CAN/CSA-C22.2 No. 601.1-M90 and No
601.1-S1-94, and specific local requirements.
The safety tests defined in this Service Supplement are derived from international standards but may not be
sufficient to meet local requirements.
CAUTIONThe correct and accurate functioning of the equipment is ensured by the successful completion of the
safety tests, performance test, and the system test (if applicable).
Safety Test Procedures
The test procedures outlined in this section are to be used only for verifying the safe installation or service
of the product in its place of use. The safety tests described here refer specifically to installation and setup
activities, and not to the aspects of safety that have already been tested during final acceptance at the
factory.
Use safety testers complying with IEC 60601-1 internationally, or any local regulations applicable to the
country of the installation. For safety test procedures see the operation instructions of the safety tester
used, and follow any local regulations.
If you use the Metron safety tester, the Metron Report should print results as detailed in this chapter,
along with other data.
For information and ordering guides for Metron products contact:
Metron AS, Vegamot 8, N-7048 Trondheim, Norway
www: http://www.metron-biomed.com
Performing Safety Tests
You must perform safety tests after completing the repair or upgrade as a standalone device, and each time
you combine equipment to form a system, or exchange system components. In the case of a system, you
must additionally perform the system test (see “Systems” on page 17).
13
Page 14
How to Carry Out the Test Blocks
Test and Inspection Matrix tells you which test blocks to carry out.
Key to Test and Inspection Matrix: P = Pass; F = Fail; X = test result value to be recorded.
Test and Inspection Matrix
Safety Tests
Test Block
VisualInspect the unit, transducers
Power OnPower on the unit.
SystemDo you have a system? For information about what constitutes a system, see “Systems” on page 17
Safety (1)Perform Safety Test: Protective
Safety (2)Perform Safety Test: Enclosure
Safety (3)Perform Safety Test: Enclosure
Safety (4)Perform Safety Test: Enclosure
PerformancePerform the parameter test
Test or Inspection to be
Performed
and cables for any damage.
Are they free of damage?
Does the self-test complete
successfully? For details of the
self-test, refer to the
Instructions for Use (included on
the CD_ROM supplied with
this kit)
If you do not have a system:
Perform Safety Tests (1) to (4)
on the standalone device
If you have a system:
Perform Safety Tests (1) to (4)
on all system components,
according to IEC/EN 60601-1-1
Earth (See page 15).
Leakage Current - Normal
Condition (see page 15).
Leakage Current - Single Fault
Condition - Open Supply
(see page 16).
Leakage Current - Single Fault
Condition - Open Earth
(see page 16.)
with all parameters as
described in the Instructions for Use for you monitor (included
on the CD_ROM supplied with
this kit).
Do these tests complete
without errors?
Expected Test Results
If Yes, Visual test is passed.V:P or V:F
If Yes, Power On test is passed.PO:P or PO:F
See Safety Tests (1) to (4)
With mains cable:
Maximum impedance = X1
(<= 100 mOhms)
With mains cables:
Maximum leakage current = X2
(<= 100µA)
With mains cables:
Maximum leakage current = X3
(<= 500µA)
(Note: maximum leakage current in the
US: 300µA)
With mains cables:
Maximum leakage current = X4
(<= 500µA)
(Note: maximum leakage current in the
US: 300µA)
If Yes, Performance Test is passed.P:P or P:F
What to Record
on Service
Record
See Safety Tests (1) to
(4)
S(1):P/X1 or
S(1):F/X1
S(2):P/X2or
S(2):F/X2
S(3):P/X3or
S(3):F/X3
S(4):P/X4or
S(4):F/X4
Enter
Result
See Safety Tests
(1) to (4)
14
Page 15
Safety Tests
Description of Applicable Safety Tests
Abbreviations:
AP: Applied PartsIUT: Instrument Under Test
GND: GroundPE: Protective Earth
S(1): Protective Earth Test
Test to perform:
The protective earth test measures impedance of Protective Earth (PE) terminal to all exposed metal parts
of Instrument under Test (IUT), which are connected to the Protective Earth (PE) for safety reasons.
Normally it includes the wiring in the mains cable (max. 100 mOhm).
A test current of 25 Amps is applied for five to ten seconds. It is recommended to flex the main cable
during the test to identify potential bad contact or damage to the earth wire.
Safety Test according to IEC 60601-1 (Clause 18).
Report the highest value.
Instrument under test
L (N)
Ri
50 Hz 25 A or 1.5 Ir
S(2): Enclosure Leakage Current Test
- Normal Condition (NC)
Test to perform:
The enclosure leakage current: normal condition is applicable to Class 1 and 2 equipment, type B, BF,
and CF Applied Parts. The test measures leakage current of exposed metal parts of the Instrument Under
Test; it tests normal and reversed polarity.
For Type BF and CF Applied Parts the test measures AP/GND.
Safety Test according to IEC 60601-1 (Clause 19.4g).
Report the highest value.
L (N)
S1
N (L)
S4
(*)
Ohm
MD
N (L)
PE
6V
S2
PE
(*)
Insulating pad
(*) If equipotential connection present : measure also
with yellow/green E.P. conductor connected.
Instrument under test
Applied part
Medical electrical system
Signal parts
Applied part
in- and/or
output
(**)
Signal parts
in- and/or
output
Other
Instrument
(*) Not present in Class 2.
(**) Can be multiple different connections to different equipment at same time.
15
Page 16
S(3): Enclosure Leakage Current Test
- Single Fault Condition (SFC) Open Supply
Test to perform:
The enclosure leakage current: single fault condition open supply is applicable to Class 1 and 2
equipment, type B, BF, and CF Applied Parts. The test measures leakage current of exposed metal parts of
Instrument Under Test with one supply lead interrupted; it tests normal and reversed polarity.
For type BF and CF Applied Parts measures AP/GND.
Safety Test according IEC 60601-1 (Clause 19.4g).
Report the highest value.
Safety Tests
S(4): Enclosure Leakage Current
- SFC Open Earth (Ground)
Test to perform:
The enclosure leakage current: single fault condition open earth (ground) is applicable to Class 1
equipment, type B, BF and CF Applied Parts. The test measures leakage current of exposed metal parts of
Instrument Under Test with Protective Earth open-circuit; it tests normal and reversed polarity.
For type BF and CF Applied Parts the test measures AP/GND.
Safety Test according IEC 60601-1 (Clause 19.4g).
Report the highest value.
L (N)
S1
N (L)
S4
MD
S2
PE
Me dical electrical system
Instrument under test
(**) Can be multiple different connections to different equipment at same time.
Signal parts
in- and/or
output
Applied part
(**)
Signal parts
in- and/or
output
Other
In s tr u m ent
16
Page 17
Safety Tests
Systems
Whenever you connect equipment together, a system is created, and you must perform a System Test after
installation for any combination of supported components that makes a system. For example, a system is
formed whenever a fetal monitor is connected to:
– another fetal monitor
– a fetal telemetry system or cordless fetal transducer system
– any other external device (for example, Nellcor Puritan Bennett OxiFirst
monitor (N-400)).
– an obstetrical surveillance system, such as OB TraceVue.
or any combination of the above equipment.
WARNINGDo not connect any devices that are not supported as part of a system.
WARNINGDo not use additional AC mains extension cords or multiple portable socket-outlets.
Fetal Oxygen Saturation
17
Page 18
Philips Medical Systems is part of
Royal Philips Electronics
INTERESTED?
Would you like to know more about our
imaginative products? Please do not hesitate to
contact us. We would be happy to provide specific
information about our products and services, or
put you on our mailing list for news about new
product developments, upcoming events or for our
clinical journal, MedicaMundi. We would be glad
to hear from you.
United States:
Philips Medical Systems
Cardiac and Monitoring Systems
3000 Minuteman Road
Andover, MA 01810
(800) 934-7372
Canada:
Philips Medical Systems Canada
281 Hillmount Road
Markham, ON
L6C 2S3
(800) 291-6743
On the web
Contact us through our web site:
www.medical.philips.com
Via e-mail
Our e-mail for all remarks and requests is:
medical@philips.com
By fax
We can be reached at the following fax number:
+31 40 27 64 887
By postal service
Please write to us at the following address:
Philips Medical Systems
Global Information Center
I.B.R.S. / C.C.R.I. Numéro 11088
5600 VC Eindhoven
Pays-Bas / The Netherlands
(no stamp required)
Europe, Middle East and Africa:
Philips Medizin Systeme Böblingen GmbH
Cardiac and Monitoring Systems
Hewlett-Packard Str. 2
71034 Böblingen
Germany
Fax: (+49) 7031 463 1552
Latin America Headquarters:
Philips Medical Systems
1550 Sawgrass Corporate Parkway #300
Sunrise, Fl 33323
Tel: (954) 835-2600
Fax: (954) 835-2626
Asia Pacific Headquarters:
Philips Medical Systems
30/F Hopewell Centre
17 Kennedy Road
Wan chai
Hong Kong
Tel: (852) 2821 5888
Fax: (852) 2527 6727