Philips 2610, 2620, Oralix 65S, 2611, 2630 Installation, Operation And Maintanance Manual

...
: )·.J1l,; :~:... .~.
.- .":"~1.-
-;~!~~'.. :~",,:.
.
.
~.
-
r
..
.
...
(:
""-l:?.•~.'
.
.,
.
.
:':'i.~.._
·.~r~••••--
-<
:::.
Dental
X.ray
System'-
2600 Series;~~··f:~··
-"".
\,:~.
~.
.
,
Installation, operation
andmaintenance
manual
~Phlllps Dent.I Syst.ml, 1984
,..
TABLE OF CONTENTS
Section
Title
Page
Obligation of the Installer and the User. . . . . . . . . . . .. 3
GENERAL DESCRIPTION
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 4
Models and Components. . . . . . . . . . . . . . . . . . . . . . .. 6
Description of Major Components. . . . . . . . . . . . . . . .. 7
II INFORMATION FORTHE ASSEMBLER
Equipment Compatibility 10 Pre-Installation Information 12
Structural and Electrical Requirements 12
Mechanical Data for Ceiling Model 14
ASSEMBLY AND INSTALLATION Wall Model #2610, #2611, #2612 15
Ceiling Model #2620 18 Mobile Model #2630 22 POlarity Test 24 Tubehead Mounting 24
" Operational Test 25
ADJUSTMENTS Folding Arm 26
Tubehead .. " 26
III OPERATING INSTRUCTIONS
Collimator/Position Indicator Device 27
Dens-o-rnat, Electronic Density Control 27
IV MAINTENANCE
Maintenance Schedule 28
TubeheadlP.I.D " 28
Dens-e-mat
Control 29
Return
POlicy
31.
V SPECIFICATIONS AND TECHNICAL DATA
Tube Housing Assembly .•....................... 32
Control and Generator ::".-.33
Measurement Basis Definitions 34
1
"-;.'
Figure
1 2 3 4 5 6 7 8 9
10 11 12 13 14 15 16 17 18
19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44
2
LIST OF ILLUSTRATIONS
Title
Page
Wall Model, #2610, #2611, #2612 5 Ceiling Model, #2620 5
Mobile Model, #2630. . . . . . . . . . . . . . . . . . . . . . . . . . .. 5
ORALIX 65 Tubehead with Collimator/P.I.D.. . . . . . . . .. .7
Dens-o-mat Electronic Density Control. . . . . . . . . . . .. 8
System Assembly 11
Dens-o-mat Dimensions 13
Wall Mount Dimensions 13
Ceiling Plate, Location and Dimensions 13 Ceiling Model, Dimensions and Positioning 14
Wall Mount 15
Extension Arm 15 Folding Arm 15 FOlding Arm Cable 16 FOlding Arm Cable to Wall Mount. 16 Removing Dens-o-mat Knobs 17 Removing Dens-o-mat Cover. 17 Dens-o-mat Mounting 17 Interconnect Wiring, Wall MountlDens-o-mat 17
Ceiling Plate Mounting ; 18 Ceiling Plate Cover 18 Telescopic Column Installed 19
Extension Arm in Telescopic Column 20
Ceiling Cable in Extension Arm 20
Folding Arm -.20
Folding Arm Cable Through Extension Arm 20
Dens-o-mat Knobs 21
Removing Dens-o-mat Cover 21
Interconnect Wiring, Ceiling MountiDens-o-mat 21
Mobile Model in Transport Position 22 Assembling Upright Column to Base 22
R ting Folding Arm Cable 22
Fling Arm Locking Screw 22
De -o-mat Knobs 23
Re oving Dens-o-mat Cover. 23
D mat Mounting 23
Interconnect Wiring, MobilelDens-o-mat 23
Installing Tubehead 24
Folding Arm Tension Adjustment 26 Tubehead Friction Adjustment ~ 26
Cooling Curve, ORALIX 65 35
Cooling Curve for Tube Housing, ORALIX 65 35 System SchematiC, ORALIX 65 36
Sample Form FD2579 37
Obligation of the Installer
1.Toinsure that the line voltage specified by the manufacturer of the equipment is available, and line voltage regulation is within specification.
2. To install and test the equipment and retain the recorded results
according to the installation instructions from the manufacturer and
with due regard to his training as a qualified serviceman.
3. Toprovide this Operator's Manual to the user.
4. To give the user, State and F.D.A.a copy of the FD.A. form "Report of Assembly of a Diagnostic X-ray System" #FD2579, properly filled out. This form should include the purchaser's name and address,
the installer or dealer's name and address, the identification, model and serial number of all Certified items installed, and the signature of the installer.
Obligation of the User
It is the responsibility of the user to maintain the equipment in compliance by following the manufacturer's recommended main- tenance schedule.
Failure of the user to properiy maintain the equipment may relieve the manufacturer, or his agent, from responsibility for any non-compliance personal injury or damage which may resuH.
3
SECTION I
General Description
Introduction
This manual provides instructions for installing, operating and maintaining the' Philips Oralix 65 Dental X-ray System (Series #2600) models listed below.
Tab/e 1. Oralix 65 Denta/ X-ray System - Series #2600
Model No. Description
2610,2611,2612 Wall Model 2620 Ceiling Model 2630 Mobile Model
All models of the Oralix 65 Dental X-ray System include the
Oralix 65 Tubehead with 8" Collimator/P.I.D., Folding Arm and
Dens-e-mat
Electronic Density Control
A Wall Mount, Extension Arm, Ceiling Mount or Mobile Pedes-
tal is also supplied, depending on the model selected.
The Oralix 65 Dental X-ray System is designed for intraoral
dental radiography and is ideally suited for use with A. S.A. type
"0"
and "E" film.
WARNING
X-ray units produce ionizing radiation that may be harmful if not properly controlled. It is therefore recommended that this equip- ment be operated in accordance with the guidelines of the National Council on Radiation Protection.
4
Fig. 2 Ceiling Model, 2620
Fig. 1 Wall Model. 2610. 2611. 2612
Fig. 3 Mobile Model, 2630
5
Table 2.
Oralix
65
Models and Components
Model Number
. and Description
Component and Catalog Number
Oralix 65 Tubehead · 9801 100 31304
8" Collimator/PJO. · 4519 100 28792
2610
Dens-e-mat
· 9801 710 80304
Wall Model Folding Arm · 9801 501 50104
2611,2612
Wall Mount ·9801 600 81004·
-
(see below)
* *
Extension Arm, 28"
· 9801 601 30104·
*Installation Kit
·9801 09390307·
Oralix 65 Tubehead
·9801 100 31304
8" Collimator/P.I.O.
· 4519 100 28792
2620 Oens-o-rnat
- 9801 71080304
Ceiling Model
FOlding Arm
- 9801 501 50104
Ceiling Mount · 9801 601 90004
Extension Arm, 20"
· 9801 601 20204
*Installation Kit
- 9801 09390307
Oralix 65 Tubehead - 9801 100 31304 8" Collimator/P.I.D.
·4519 100 28792
2630
Dens-o-mat
- 9801 71080304
Mobile Model FOlding Arm - 9801 501 50104
Mobile Pedestal
- 9801 50201204
*Installation Kit
·9801 09390307
Instruction Manual
* Installation Kit Sample and Blank Form F02579
9801 093 90307
Wire Nuts (4) Line Plug
**
2611
Same as Model 2610 except Extension Arm
Wall Model
Extension Arm, 20" - 9801 601 20104
2612 Same as Model 2610 except Extension Arm
Wall Model
Extension Arm, 12" - 9801 601 10104
6
Description of Major Components
Oralix 65 X·ray Tubehead '
The Oralix 65 X"ray Tubehead consists of a high·voltage/fila· ment transformer, and 3-element, grid-control X-ray tube, in a lead shielded, oil filled metal case, covered by a plastic housing. An
8"
collimator/position indicator device with quick-connect bayonet fit· ting is supplied with the tubehead.
Fig. 4
Oralix 65 Tubehead with Collimator/P.I.D.
7
Dens-e-mat,
Electronic Density Control
The Dens-o-mat Electronic Density Control is an object-pro- grammed exposure control which automatically provides optimal exposure based on patient type, film type, and tooth structure being radiographed. It incorporates a Line On-Off switch and indicator light, selector knobs for Patient Type, Film Type, and Tooth Struc- ture, and is provided with a "dead-man" type exposure switch with coil-cord for operation of the x-ray system. When the exposure but- ton is depressed, the x-ray ON indicator lights and an audible signal sounds. The exposure is terminated when the exposure button is released or when the programmed exposure is completed. This is indicated by the audible and visible x-ray ON indicators going off.
The Dens-o-mat automatically compensates for line voltage
fluctuation, consistently providing films with optimal density.
Note:
To determine actual mAs, multiply mAs setting by
correction factors for patient type and film type.
Example:
mAs Patient Film Actual setting x Type x Type
=
mAs
2.3 .66 1.5 2.3
Tooth structure
~.~,f
Fuse
-JI
Fig. 5
Dens-e-met Electronic Density Control
8
FOlding Ann
The FOlding Arm is a 2-section spring counter-balanced dental x-ray arm. The end of the Folding Arm has a 3-conductor cable which interconneCts with the Dens-a-mat Control via the Wall Mount, Ceiling Mount or Mobile Pedestal, when the system is
assembled.
Extension Anns
A 28" Extension Arm* is provided with the Wall Model and a
20" Extension Arm is provided with the Ceiling Model. Electrical
wiring and mechanical interconnection between the FOlding Arm and either the Wall Mount or Ceiling Mount is made via the Exten- sion Arm. *For Model 2610. Models 2611, 2612 supplied with 20",
12" Extension Arm respectively.
Wall Mount
The Wall Mount incorporates a mechanical fitting into which
the support tube of the Extension Arm is installed. A terminal strip is provided for wiring interconnection between the Dens-o-mat Elec- tronic Density Control and the Folding Arm and Tubehead.
Ceiling Mount
The Ceiling Mount Assembly consists of a ceiling column with
mounting plate, telescopic column and cover. The ceiling column is designed to be secured to a suitably reinforced ceiling with the mounting plate provided. A terminal strip on the mounting plate is provided for wiring interconnection between the Dens-o-mat Con- trol and the Folding Arm and Tubehead.
Mobile Pedestal
The Mobile Pedestal incorporates a vertical upright column which mounts on a weighted base with four casters. The upright column terminates in a mechanical fitting into which the F91ding Arm mounting is installed. Provision is made on the vertical upright column for mounting the Dens-o-mat Control.
9
SECTION II
Information for the Assembler
Equipment Compatibility
The Oralix 65 Dental X-ray System incorporates the following
certified compatible components:
Oralix 65 Tubehead - 9801 100 31304 Collimator/P.I.O. - 4519 100 28792 Dens-o-rnat Control - 9801 71080304
The certified components above are useable with the mounting components indicated below and with previous Philips Oralix mountings:
Folding Arm Extension Arm Extension Arm, 20"
Wall Mount
Ceiling Column Mobile Pedestal
- 9801 501 50104
- 9801 601 30104, 20104, 10104
- 9801 601 20204
- 9801 600 81004
- 9801 601 90004
- 9801 502 01204
Table3.Equipment Weight
Component
Catalog Number
Weight (Approx.)
Oralix 65 Tubehead
Folding Arm Extension Arm, 28" Extension Arm, 20"
WaJl Mount
Mobile Pedestal
Ceiling Mount
Oens-o-mat
10
9801 100 31304 9801 501 50104 9801 601 30104 9801 601 20204 9801 600 81004 .
980150201204 9801 601 90004 9801 71080304
121bs.
21
15 10 12
115
45
5
I I
o
0
0
I
o
I
'---+--
l!:=::!J
DENS-O-MAT,
9801 710 80304
20- INCH EXTENSION ARM
l
980160120204
(
I
I
IIJl
!J ,
t
28-INCH EXTENSION AR(
: 9801 60130104
WALL MOUNT \
J
9801 ·600 81004 '-..- .-/ .
I
I
'-----
CEILING MOUNT
9801 60190004
,
,
I
t
t
,
~'------rJ
FOLDING ARM
9801501 50104
ORALIX 65 TUBE HEAD
9801 100 31304
COLLlMA~
8-INCH P.I.D.
4519 100 28792
MOBilE STAND
9801 502 01204
Pre-tnstallatlon Information
All models of the Oralix 65 Dental X-ray System employ the
Dens-a-mat Electronic Density Control, which may be wall-mounted either in the same room as the x-ray system or remotely outside. On the Mobile Model, provision is made for mounting the Dens-o-mat on the vertical upright column. All equipment can be installed with standard hand tools.
The Dens-a-mat must be installed so that the patient and indi-
cation of technique factors are visible from the operator's position.
The Dens-a-mat and the Wall Mount are provided with electri- cal knockouts in the base plate for in-the-wall wiring and in both sides of plastic cover for surface wiring.
Structural Requirements
iMPORTANT:
The following data is provided as a guide for safe in-
stallation of the ORALIX 65 Dental X-ray System:
Wall Model, dead weight=65 Ibs.The mounting wall and hardware must be adequately reinforced to sup- port pull and thrust loads up to 600 Ibs.*
*A Wall Plate isavailable (optional). Ceiling Model, dead weight=100Ibs.The ceiling
and mounting hardware must be adequately reinforced to support pull and thrust loads as follows:
for 8' ceiling, up to 600 Ibs. for 10' ceiling, up to 1000 Ibs.
Electrical Requirements
WARNING:
This equipment operates at a nominal 120 volts A.C.
Appropriate safety precautions must be observed. Power: 115-130 volts, 60Hz., 15 amp. service
Line Voltage Regulation: 3.5%
@
115 V., 3.8%
@
120 \/,4.5%
@
130V
Wiring: 3-#14 AWG, Dens-o-mat to 120volt supply (30' max.)
3-#14 AWG, Dens-e-mat to Oralix 65 (30' max.)
All electrical installation described shall be in accordance with the National Electrical Code or Local Code, whichever is applicable to the area. Any electrical wall boxes required should be pre-instal-
led using the drawings supplied as a guide.
12
~----;:~
-1~r
5~z-----ll
I~
I
r--
I
,
IYt6
-$:
~
11
JL;:.
TYP
i '_'
U,
TVP-
$
1
7
9){6
~ ~,~I
KNOCKOUT
I
21Jf6
I
1
1
-~.
I
\
/
- -I
I
I
,
~
-$-.
'---
Fig. 7
Dens-a-mat
WALL MOUNT
MAClin
/
//
.
TOP
:%,-NTYP.
.-$.
r
I
!
,
74
.
'3/,;
~~.
r .
I;J
2~U
1
-%
3){
~.
-+
f~
i'-..i
~·I-
.
.-
3'/18
.
8
RXJT END
ILJ.LJI
~
ORAUX
~5
~r/£ING
,",O(lN7£D MODEL
2620
LOCATION OF CEILING PLATE
Fig.
8
Wall Mount
Fig. 9
Ceiling Plate Location
'-
I••
I~
13
Mechanical Data for Ceiling Model
Ceiling Height:
Max. Horiz. Reach: Ceiling Column Vertical Range: Vertical Range of Tubehead:
8 - 10 ft. 55"
33" min., 56" max.
43" - (30" to 73" above floor
on 8' ceiling)
3600around vertical axis 2600around extension arm
2700around ceiling column
Tubehead Rotation: Folding Arm Rotation:
Extension Arm Rotation:
1/
73 MAX.
14
Fig. 10
Ceiling Model
Dimensions&Positioning
Assembly and Installation Wall Model #2610, #2611, #2612
Wall Mount
1. Using the drawings supplied, mark location of four (4)Wall Mount mounting holes and electrical knockout
2. Remove electrical knockout frame Wall Mount
3. Using the four
(4)
mounting holes, secure wall Mount to wall
ensuring it is level. (Fig. 11)
Extension Arm
1. Remove round snap-out covers at both ends of Extension Arm.
2. Remove four
(4)
screws and blank panel on underside of
Extension Arm (see Fig. 13Afor 28" Extension Arm
9801-601-30104).
3. Push the chrome ring and the plastic ring over the support tube at end of Extension Arm. (Fig. 12)
4. Lightly grease support tube at base of Extension Arm and insert
into opening at top of Wall Mount.
5. Loosen locking screw on Extension Arm.
Folding Arm
WARNING
The springs in the Folding Arm are powerful and may cause injury to the installer (as well as possible damage to the arm) if not handled properly. Do not remove binding on Folding Arm until indi-
cated in instructions below.
1. Lightly grease mounting pin at base of Folding Arm.
2. Guide FOlding Arm cable down through hole in Extension Arm, insert mounting pin into Extension Arm an,g,Ji,ghtenthe locking
screw. (Fig.
13)
TO r'\l.L
CMlZ !tUo.O-l
IXt'INSlat
NIM.
m ac DC " ~
oeo
a.eu:
HID P\.U. "IIO.XJI.
9101 601 )0104 2r
uterW1.cn ••.•
Fig.13A
Extension Arm
,. 1Ot ••• 4
Fig. 11
Wall Mount
J
Fig. 12
Extension Arm
..II.
,
! •• '
\
Fig.
13
Folding Arm
15
3. Guide Folding Arm cable back through Extension Arm and out support tube at the other end, to Wall Mount terminal strip. (Figs. 14, 15)
Note:
Do not replace covers until Polarity Test (pg.
24)
is completed.
4. Hold Folding Arm sections together and carefully remove bind-
ing, allowing free section of arm to open slowly, away from installer.
5.
Connect3wires from Folding Arm connecting cable and3wires
from Dens-o-mat to Wall Mount terminal strip, as follows:
Folding Arm Cable Wall Mount Terminal Dens-o-mat Cable
#1 #101-1 Neutral -#101 #2 #102-2 Hot -#102
Yellow/Green #103-~ Ground -#103
~
l
NOTE: Locking Screw and Friction Screw must be tightened when
','_~'=
i~cOm~leled
v:.....
.--
-.--~.-~-
Fig. 14
Folding Arm Cable
Fig. 15
Folding Arm Cable to Wall Mount
16
4. Remove2 screws at bottom of Dens-o-matcover and 1 screw in center of front panel.
5. Using offset tool supplied, carefully remove snap-on cover
from Dens-o-mat (Fig. 17)
6. Carefully remove electrical knockout at back of Dens-a-mat and secure Dens-o-mat to wall, using 3 mounting holes (Fi9.18~"_
7.
Usingappropriate knockout,interconnect3-#14
AWG
wiresfrom"
WallMount terminal stripto Dens-o-mat,asfollows: -
Wall Mount Polarity Dens-o-mat
\
)
Dens-o-mat
1. Position the Dens-o-mattemplate on the wall so that the cable knockouts are over the wall box provided. Mark location of 3 mounting holes.
2. Set Patient Type and Film Type selector knobs to "1", Tooth Selector knob to 2.3 mAs.
3. Using socket wrench or needle-nose pliers, loosen nut on top of each knob and slide knobs off. (Fig. 16)
#101-1 Neutral #101 #102-2 Hot #102
~~* G~~
~ro
8. Turn off 120volt supply line.
9.
Interconnect 3-#12AWG wires between 120volt supply and
Dens-o-matterminal strip, as follows:
Dens-e-met
120 Volt Supply
#104 #105
#1~,,"
Neutral (White) Hot (Black)
Ground (Green)
'O~OING •••
CAlLI
Ag.19
Interconnect
Wiring Wall Mount
Dens-e-met
WAll MOUNT
DENS-O-MAT
10. Install BAF 10 fuse In Dens-o-mat. Note: Do not replace covers until Polarity Test (page 24)is com-
pleted. Proceed to page 24 for Polarity Test and Tubehead
Mounting.
J
C
Fig. 16
Dens-a-mat
Knobs
Fig.
17
Removing
Dens-a-mat
Cover
Fig. 18
Dens-o-mat
Mounting
17
Ceiling Model #2620
Ceiling Mount
. 1. Using the drawings supplied, mark location of four (4)ceiling
plate mounting holes and center hole for electrical cable from Dens-o-mat.
2. If not already done, run 3-#14 AWG wires between location of Dens-e-mat and Ceiling Plate.
3. Drill center hole and guide Dens-o-mat cable down through
ceiling.
4. The terminal strip on ceiling plate has two (2)3-conductor
cables connected. Disconnect and remove one of the cables.
Note:
When system is completely assembled, the Extension Arm
has 2700rotation around the ceiling column. For standard installation, the ceiling plate and column should be mounted so that the two (2) sides of the column without holes face the head end and side of chair. The orientation of the 270
0
rotation may be changed by rotating the ceiling plate mounting in 900steps.
5. Guide Dens-o-mat cable from ceiling through hole in ceiling
plate and secure ceiling plate to ceiling using appropriate hard- ware. (Fig. 20)
I ,
r::
";<'!
""
.:
~
Fig. 20
Ceiling Plate
Mounting
18
6. Using a level or plumb line, verify that the ceiling column is hanging vertically. Adjust, if necessary, with the four
(4)
adjust-
ing nuts in cross frame of ceiling plate.
7.
Connect
Dens-e-mat
cable to terminal strip and ground on ceil-
ing plate, as follows:
Dens-e-met Cable
Polarity Ceiling Mount
#101 #102 #103
Neutral Hot Ground
Black #1
Black #2
YellowlGreen, Ground
8. Carefully slide ceiling cover and square locking ring up the col- umn to the ceiling and secure with the Allen set screws.
(Fig. 21)
9. Fit the square trim bumper to the bottom of ceiling column.
10. Insert the telescopic column into the ceiling column, ensuring that the fastening holes in both columns are aligned. (Fig. 22)
11. Using the special bolts supplied, fasten the telescopic exten- sion column in the ceiling column at the required height.
12. Cover the holes in the telescopic column with the aluminum strips supplied.
,
\
1
Fig. 21
Ceiling Plate
Cover
Fig. 22
Telescopic
Column Installed
19
. Extension
Ann
~1~
1. Remove round snap-out covers at both ends of Extension Arm.
~.~.d'
1!!I!!!!!!!!!I!!!!!!::!!"t.::~=_~.~~~.
2.
Remove4screws and blank panel on underside of Extension
- -; Arm.
Fig. 23
Extension Arm
in Telescopic
Column
NOTE: Locking Screw and Friction Screw must be tightened when . installation is completed
./
Fig. 24
Ceiling Cable
in Extension Arm
Fig. 25
Folding Arm
20
3.
Lightly grease support tube of Extension Arm.
4.
Remove split collar and two
(2)
retaining shoes at bottom of
telescopic column. (Fig.
23)
5. Guide cable from telescopic column down through Extension
Arm support tube and insert and fasten Extension Arm in col- umn, using the retaining shoes and split collar removed in #4.
6. Guide cable from column back through Extension Arm and out
rectangular opening on underside. (Fig.
24)
Folding
Ann
WARNING
The springs in the Folding Arm are powerful and may cause injury
to the installer (as well as possible damage to the arm) if not hand-
led properly. Do not remove binding on Folding Arm until indicated in instruction below.
f.
Lightly grease mounting pin at base of Folding Arm.
2.
Guide Folding Arm cable down through hole in Extension Arm,
insert mounting pin into Extension Arm and tighten locking screw. (Fig.
25)
3. Guide Folding Arm cable back through Extension Arm and out
rectangular opening on underside. (Fig.
26)
4. Using wire nuts supplied, interconnect wires from Folding Arm and Ceiling Column, as follows:
Folding Arm Cable Polarity
#1 Neutral
#2
Hot
Yellow/Green Ground
Ceiling Column Cable
#1
#2
YellowlGreen
Note:
Do not replace covers until Polarity Test (pg.
24)
is
completed.
5. Hold Folding Arm sections together and carefully remove bind- ing, allowing free section of arm to open slowly, away from installer.
Fig. 26
Folding Arm Cable
through Extension Arm
Dens·o-mat
1. Position the Dens-o-mat template on the wall so that the cable knockouts are over the junction box. Mark location of3mount- ing holes.
2. Set Patient Type and Film Type selector knobs to "1", Tooth
Selector knob to 2.3 mAs.
3.
Using socket wrench or needle-nose pliers, loosen nut on top
of each knob and slide knobs off. (Fig.
27)
4. Remove 2 screws at bottom of Oens-o-mat cover and 1 screw
in center of front panel.
5.
Using offset tool supplied, carefully remove snap-on cover from
Dens-o-mat. (Fig.
28)
6. Carefully remove electrical knockout at back of Dens-o-mat
and secure Dens-o-mat to wall using3mounting holes.
7. Using appropriate knockout, connect 3-# 14 AWG wires from
Ceiling Mount terminal strip to Dens-a-mat, as follows:
Ceiling Mount
Polarity
Dens-o-mat
#1 #2
Ground
Neutral Hot Ground
#101 #102 #103 .
8. Turn off 120 volt supply line.
9. Interconnect 3-#14 AWG wires between 120 volt supply and
Dens-a-mat terminal strip, as follows:
Dens-e-met 120 Volt Supply
#104
Neutral ('Nhite)
#105
Hot (Black)
#103
Ground (Green)
10.
Install BAF
10
fuse in Dens-o-mat.
Note: Do not replace covers until Polarity Test (page 24) is com-
pleted. Proceed to page 24 for Polarity Test and Tubehead Mounting.
j
c
Fig. 27
Dens-e-met
Knobs
Fig. 28
Removing Dens-e-met
Cover
Q)
0 0 0 0
1010102 103 10••0105
<30 0 0 0
Fig. 29
Interconnect Wiring
Ceiling MountlDens-o-mat
CEILING MOUNT
N H
=-
N H
DENS-O-MAT
21
Fig. 30
Mobile Model
in Transport
Position
Fig. 31
Assembling Upright
Column to Base
22
Mobile Model #2630
WARNING: Prior to moving this unit, the FOlding Arm must be
secured in the transport position. (Fig. 30)
Mobile Pedestal
1. Using 2 bolts supplied, assemble upright column to base so
that cable and cable storage hooks are at rear of base, over short legs. (Fig. 31)
2.
Back out locking screw at top of column ..
Folding
Ann
WARNING
The springs in the FOlding Arm are powerful and may cause injury to the installer (as well as possible damage to the arm) if not handled properly. Do not remove binding on Folding Arm until indi-
cated in instructions below.
1. Lightly grease mounting pin at base of FOlding Arm.
2. Guide FOlding Arm cable down through hole at top of upright column and out through opening on side. (Fig. 32)
Note: The FOlding Arm and Tubehead must be installed so that
they can only be extended between the long leg sections of the base.
3. Insert mounting pin down into hole at top of column and tighten locking screw. (fig. 33)
4. Hold Folding Arm sections together and carefully remove bind- ing, allowing free section of arm to open slowly, away from installer.
Fig. 32
Routing Folding
Arm Cable
Fig. 33
Folding Arm
Locking Screw
Dens·o-mat
1. Set Patient Type and Film Type selector knobs to "1", Tooth Selector knob to 2.3 mAs.
2. Using socket wrench or needle-nose pliers, loosen nut on top of each knob and slide knobs off. (Fig. 34)
3. Remove 2 screws at bottom of
Dens-e-mat
cover and 1 screw
in center of front panel.
4. Using offset tool supplied, carefully remove snap-on cover from Dens-o-mat. (Fig. 35)
5. Carefully remove electrical knockout at back of Dens-o-mat
and secure Dens-o-mat to mounting plate on column, using the 3 nuts and washers supplied. (Fig. 36)
6. Insert the two (2) 3-conductor cables (1 from Folding Arm and 1 from bottom of column) through knockout at back of Deris-o-
mat. Cut cables to required length and connect to Dens-o-mat terminal strip, as follows:
Dens-e-met Polarity" Wires in Upright Column
#101 Neutral #1-to
P
} .
#102 Hot #2-top ORALIX Folding Arm #103 Ground Ground, Yellow/Green
#104 Neutral #1-bottom} 120 Volt Supply #105 Hot #2·bottom
7. Install BAF 10 fuse in Dens-o-mat.
Note: Do not replace cover until Polarity Test (page 24) is
completed.
8. Connect 3-prong polarized line plug to 3-conductor cable at
bottom of upright column, observing correct polarity, as follows: .
Plug Contact
Polarity
Line Cable Wire
Silver Brass
Neutral Hot Ground
#1 #2 YellowlGreen"U"
TO
FOLDING -
C:;:~....
AU" -----{
TO
--, 120 VOLT
1
SU'rL1
o
0 0 0 0
1010102 103 1040105
e
0 0
e
0
I
~4
Ag.37
Interconnect Wiring
MobilelDens-o-mat
N H
=-
N H
DENS-O-MAT
J
C
Fig. 34
Dens-e-met
Knobs
Fig. 35
Removing Dens-o-met
Cover
:IJ~I
.
~~~
,'
..
1-;
1 '
Fig. 36
tiens-o-me:
Mounting
23
5.
••
6.
COVER PLATE
FOLDING
J
ARM
7.
Note: Do not mount Tubehead until Polarity Test has been
satisfactorily completed.
Polarity Test
1. Connect either an A.C.voltmeter or a test light as follows: a. Wall Model: One lead to Wall Mount terminal strip #103
(Ground). Connect other lead to #102-2(Hot).
b. Ceiling Model: Removewire nuts from the #2 and yellow/
green wires in Extension Arm. Connect one lead of meter
or test light to yellow/green wires (Ground). Connect other
lead to #2 wires (Hot).
c. Mobile Model: Connect one lead to Dens-o-mat terminal
#103 (Ground).Connect other lead to #102(Hot).
2. Turn on 120Volt Supply Line.
3. Turn on Dens-e-matline switch. Green Line ON indicator should light.
4. Take exposure switch from Dens-o-matand move to an area
from which the voltmeter or test light can be observed.
Make exposure. Voltmeter or test light should indicate full line
voltage. If a "no voltage" indication is obtained, the system is
incorrectlywired andmustbe rewired(recordvoltagemeasured).
Disconnect the test light or meter lead from the #102-2(Hot) wire and connect it to the #101-1(Neutral) wire.
Make exposure. Voltmeter or test light should indicate "no Voltage". If line voltage indication is obtained, the system is
incorrectlywired and mustberewired(recordvoltagemeasured).
RETAINING
~
~ CLIP
Tubehead Mounting, All Models
~, . 1. Remove retaining screw and cover plate on inner side of tube-
. end of Folding Arm.
2. Push Folding Arm handgrip up and remove "U" shaped retain-
ing clip. (Fig. 38)
3. Remove clear plastiCtube from Tubehead contact assembly.
4. Depressing the ground contact spring, Insert Tubehead contact
pin up into FOldingArm and secure with "U" shaped retaining clip, ensuring it is flush in slots.
5. Pull handgrip down and reinstall cover plate and retaining
screw.
TUBEHEAO
y ~
(-~~
--- ---'
Fig. 38
Installing
Tubehead
24
Operational Test
Upon completion of the ORALIX installation, perform the fol-
lowing test procedure to verify proper operation. WARNING: X-rays will be generated during this test. Observe
appropriate precautions.
1. Position Tubehead with Collimator toward radiation-safe area, away from personnel and block distal end with 2mm lead.
2. Turn on Dens-o-mat line switch. Green Line ON indicator should light.
3. Take exposure switch and move to a radiation-safe area.
4. Depress exposure button and verify that amber X-ray ON indi-
cator lights and buzzer signal is audible during exposure.
5. Verify that audible and visible X-ray ON indicators go off when
exposure terminates.
6. Reset Tooth Selector knob to 4.5 mAs.
7. Depress and immediately release exposure button, verifying
that audible and visible X-ray ON indicators go off immediately after exposure button is released.
8. Move Tubehead, Folding Arm and Extension Arm assembly
through all positions. Periodically depress exposure button and verify that audible and visible X-ray ON indicators operate properly. CAUTION: Do not make more than 8 seconds of exposure in any
10 minute period.
9. Turn power off.
10. Check for proper tension at joints of Folding Arm. If adjustment is necessary, refer to adjustment instructions on page 26.
11. Check Tubehead for ability to maintain its vertical angulation
with 8" Collimator/Position Indicator Device in place. If adjust- ment is necessary, refer to adjustment instructions on page 26.
12. Replace all covers and coverplates on Wall Mount, Extension
Arm and Dens-o-mat.
13. Replace Dens-o-mat knobs and caps ensuring that index marks
are correctly aligned.
NOTE: For complete schedule of tests which must be perform-
ed at installation, see Maintenance Schedule, page 28.
IMPORTANT: In case of product defect or failure to comply wHh the
Perfonnance Standards, refer to page 31.
25
Adjustments
Folding Ann, Tubeslde Section
To adjust spring tension of Folding Arm, proceed as follows:
1. Position tubeside section of Folding Arm, in highest horizontal position, parallel to floor; other section of Arm should be verti-
cal. (Fig. 39)
2. Removecaps at (A) at front of arm cover. Note:One side of pin (8) is keyed.
3. Tapout the pin, (8)at front of arm cover and raise cover (C)for
access to the threaded bushing (0) used to adjust spring
tension.
4. Turn bushing clockwise to increase spring tension, counter- clockwise to decrease tension.
5. Replace cover, pin and caps.
Note: Other section of Folding Arm may be adjusted by removing
its cover and adjusting spring tension as above.
Tubehead Adjustment
To adjust tubehead friction at pivot axis, proceed as follows:
1. Removesnap-out disc with PHILIPS emblem.
2. Adjust three (3)screws for desired Tubehead friction. (Fig. 40)
3. Replace snap-out cover disc.
4. Check tubehead in various positions for proper balance.
c
r.;"
I :,'
I
II I
1/ I
// I
1/ /
/ I
/
I
I
I
I
I
r
\ t
.
~:.•.,.,..
-Y"'('--
'--""'--,
::.:::-----
A _
- 0 0
o
Fig. 39
Folding Arm
Tension Adjustment
Fig. 40
Tubehead
Friction Adjustment
26
SECTION III
Operating Instructions
PHILIPS ORALIX 65 DENTAL X·RAY SYSTEM
FOR INTRAORAL DENTAL RADIOGRAPHY
Collimator/Position Indicator Device
The ORALIX 65 Tubehead is supplied with an 8" (20cm) open- ended COllimator/position indicator device, suitable for use with the bisecting angle or parallel film technique. It is designed for use in contact with the patient's cheek and provides a maximum beam diameter of 60mm for cone 4519 100 28792, at the distal end. The P.l.D.fastens to the tubehead window by pushing in and twisting the' quick-connect bayonet fitting a quarter turn to the right.
Dens-e-mat Electronic Density Control
1. Depress Main Line Switch to "ON" posltlon, Green "Mains ON"
indicator should light.
Note: "Patient Type" and "Film Type" selectors are mAs rnultlpllers"
2. Set "Patient Type" selector to one of the following: "0.66" - light adult or child "1" - average adult
"1.34" - heavy adult
3. Set "Film Type" selector to "1.5", for ASA Type
"0"
Film. Set to "0.7"
for ASA Type "E" Film.
4. Set rotary selector knob to tooth structure being radiographed.
5. Position tubehead with Collimator/P.I.O. in contact with cheek.
6. Remove exposure switch from mounting hook and move to a radiation-safe area.
7. Depress exposure button and hold depressed until exposure is terminated. Termination of exposure is indicated by "X-ray ON" indicator light and audible signal going oft. Note: Exposure button employs "dead-man switch" which terminates exposure as soon as button is released. To obtain full exposure selected, exposure button must be held depressed until exposure light goes out and buzzer stops.
8. When radiography is completed, replace exposure button on
mounting hook and depress Main Line Switch to "OFF" position. Green indicator light should go off.
*Note: To determine mAs, multiply mAs setting by correction
factors ("mAs multiplication factors"):
Examples: mAs Patient Film Actual
setting x Type x Type=mAs
2.3 1 1.5 3.5
2.3 .66 1.5 2.3
27
SECTION IV
Maintenance
Maintenance Schedule
The following schedule describes the maintenance required to keep the ORALIX 65 Dental X-ray System in compliance with the X-ray Equipment Performance Standard.
It is the responsibility of the User to maintain the equipment in
compliance by following this maintenance schedule. Failure of the
User to properly maintain the equipment may relieve the manufacturer,
or his agent, from responsibility for any non-compliance which may
result
\
The following tasks must be performed by a qualified individual upon installation and replacement of certified components and at 12 month intervals. Any defects or malfunctions should be corrected immediatelybya qualified individual with adequate training. Any product defect or noncompliance must be promptly reported by the
assembler to Philips*. Also see page 31.
TubeheadlP.I.D.
1. Inspect for damage and/or wear to the X-ray Tubehead and 8" Collimator/Positioner. Any defective item affecting safe opera- tion must be repaired or replaced.
2. Inspect Tubehead for oil leaks. Repair or replace if necessary.
3. Check Tubehead and Folding Arm for positioning stability. If adjustment is required, see page 26. .
*National Service Manager
Philips Dental Systems 102Commerce Rd.
Stamford, CT.06902
28
Dens-e-mat Control
1. Verify and record that the green indicator lamp on Dens-o-mat lights when "Mains ON" switch is depressed.
2. Verify and record that amber indicator lamp on Dens-o-mat lights during exposure and goes out when exposure is terminated.
3. Verify and record that audible signal from Dens-o-mat buzzes during exposure and stops when exposure is terminated.
4. Verify and record that exposure terminates when exposure push- button is released. Termination of exposure is indicated by amber
exposure light going out and audible signal stopping.
5. Inspect exposure switch and cable for wear. Check for continuo ity and for intermittent open or short circuit. Replace if
defective.
6. Measure tine voltage under load and no-load condition and cal-
culate line voltage regulation, confirming that both are within tolerance.
% line voltage regulation=100(Vn-VL)/VL where: Vn
=
No-load line potential and record results
VL=Load line potential
7. Verify and record that the exposure terminates at the preset interval selected. .
Check and record Dens-o-mat accuracy at several settings, using either a "spinning top", a radiation impulse counter, or an electrical line impulse counter.
Notes: a. There is a multiplicative relationship between the
variable factors for mAs, Patient Type, and Film Type.
b. Exposure times and/or impulses in the tables be-
low are obtained at nominal settings: 120 volts,
Patient Type "1", Film Type "1".
c. The Dens-o-mat allows a filament warm-up time
of 0.125 sec. or 7.5 pulses. The pulses obtained during this warm-up period have a low amplitude and are not to be counted as part of the exposure time for accuracy measurement purposes.
29
Radiation Impulse Counter Method
Since no significant radiation Isproduced during the warm-up
time, the first 0.125secJ7.5impulses are excluded from the mea-
surement with this method.·
Exposure Time
sec. impulses
mAs
accuracy
1.5
3.0
6.4
0.2 12
0.4 24
0.85 51
±
1 impulse
±
2 impulse
±
3 impulse
Electrical Line Impulse Counter Method
Electrical impulses can be measured across the Dens-o-mat
output terminals #101and #102.
The first 7.5 impulses (warm-up time) will be included in the Total Elapsed Time measuredwith this method. However,since no significant radiation is produced during the warm-up time, 7.5 im- pulses must be subtracted from the Total ElapsedTime measured, to determine ExposureTime.
Total
Elapsed Time
impulsesmAs
Pre-heat Time
impulses
Exposure Time Accuracy
sec. impulses impulses
1.5
3.0
6.4
19.5
31.5
58.5
7.5
7.5
7.5
0.2 12
±
1
0.4 24
±
2
0.85 51
±
3
For those instruments so designed, set threshold level at 40% of full amplitude.
30
Labels, Warning and Indicators
1. Verify and record that Certification, Identification, Place and Date of
Manufacture labels on Dens-e-mat Oralix Tubehead and 8" Collirnator/Pf.D, are permanently affixed, legible and readily
accessible to view.
2. Verify and record that Warning Label on the Dens-o-mat is legible and accessible to view.
3. Verify and record that the technique factors, 6SkV/7.SmA are
indicated on the ORAUX Tubehead and the Dens-o-mat Control.
IMPORTANT
All recorded data must be kept by the installer for possible review
and inspection by the F.DA
Return Polley
Authorization for returns must be received from Dental Sys-
tems Division, Stamford, Conn.
All X-ray materials returned must be shipped to:
Philips Dental Systems Division 102 Commerce Road
Stamford, Conn. 06902
31
SECTION V
Specifications and Technical Data
Tube Housing Assembly
Maximum rated peak tube potential - 65kVp Leakage Technique Factors - 0.1mA
@
65kVp
Minimum Filtration permanently in the beam -
2.0mm AI equiv. @ 65kVp
Tube Rating Chart - only 1 fixed working point -
65kVp, 7.5mA, 3 sec. Cooling curves - Tube Housing and Insert Tube (page 35) X-ray Tube - Three electrode, grid action X-ray Tubehead - oil-immersed, shielded Beam Quality - half value layer better than 1.7mm AI. (H.V.L.) Focal Spot - 0.7mm Collimator/Position Indicator Device - consists of a shielded
beam-limiting device approximately 25mm 1.0. x 65mm in
length, permanently attached to a plastic cylinder which
serves as a minimum 5.5.0. spacer and alignment aid and
for attachment to the tubehead. Distance from Focal Spot to base of P.I.O. - 42mm Distance from Focal Spot to distal end of P.LO. - 8" (20 cm) Beam Diameter at distal end of P.LO. -
Maximum 6.Ocm for P.1.0. #4519-100-28771
Mean 5.8cm for P.LO. #4519-100-28791
Nominal Radiation Output - 5OOmRisec. +25-40% at
65kVI7.5mA at distal end of 8" Collimator/P.1.0.
Measurements are made using ionization chamber in a special
fixture at 20.32cm with a Victoreen 666 dosimeter and
ionization chamber.
32
Control and Generator
Rated Line Voltage - 115-130 volts Line Voltage Regulation - 3.5%
@
115 volts
3.8% @ 120 volts
4.5% @ 130 volts
Maximum Line Current - 8 amps @ 115 volts, 7.0mA - 62 kVp.
9 amps @ 120 volts, 7.5 mA - 65kVp.
12 amps @ 130 volts, 8.6 mA - 68kVp.
Generator Rating and Duty Cycle - 120 volts, 7 amps, 60 Hz:
- 50 sec. on, 10 min. off.
NOTE: This Duty Cycle rating is furnished for compliance with
F.DA standards only; it refersto the high-voltage trans- former only. IT IS NOT TO BE USED FOR X-RAY EX-
POSURES. TO DO SO MAY CAUSE IRREPARABLE DAMAGE TO THE X-RAY TUBE.
Maximum Deviation From Indicated Technique Factors
Peak Tube Potential (kVp)
Line Voltage 115 volts 120 volts 130 volts
Tube Potential (kVp) 62.0
±
8% 65.0
±
8% 68.5
±
8%
Tube Current (mA)
Line Voltage 115 volts 120 volts 130 volts
Tube Current (mA) 7.0
±
14% * 7.5
±
11% * 8.6
±
16% *
Exposure Time
Line Voltage Timer
115 volts 120 volts 130 volts
20%* 20%* 20%*
sta tistically distributed
33
Measurement Basis Definitions
kVp -
stationary high voltage which settles under load after
the pre-heating time.
Direct measurements in the laboratory using a Philips Elec-
tronic kV Meter, F. S. 200kV, with a high voltage divider network, accuracy 1%.
Indirect measurements during production as the high voltage terminals are not accessible. The factors considered are (1) rated line voltage, (2)turns ratio of the high voltage transformer, (3)
peak
value of the tube current, (4) rated mains resistance, (5) primary re- sistance of high voltage transformer, and (6)secondary resistance of high voltage transformer. Maximum deviation based on worst case calculations.
mA -
The average value of the stationary current which sett/es
after the pre-heating time.
Direct measurements made at the midpoint of high voltage
secondary using a precision mA meter, F. S. 10mA, class 1.
Time -
The first 7.5 pulses (warm-up pulses) are pulses with less than a minimum threshold voltage pulse amplitude and are ex- cluded from the measurement.
Total time consists of "preheating time" and the "exposure time." The first 7.5 pulses (125ms) during which no significant X-rays are generated, are considered as "preheating time."
The "exposure time" is the time that can be selected and
during which X-rays are generated.
The exposure time is checked with a special crystal-controlled
timing device with an accuracy of 1o-eseconds. The Dens-o-mat is tested with a dummy load, at nominal conditions of 120 volts A.C., at all mAs settings and at each setting of the Patient and Film selectors. The Dens-o-mat compensates for line voltage fluctua- tions by increasing the exposure duration when the line voltage falls below 120 volts. It reduces the exposure duration when the line volt-
age exceeds 120 volts, thus providing consistent film density.
34
Loading...