Pfeiffer Vacuum UNO 120 A, DUO 120 A Operating Instructions Manual

Rotary Vane Pumps
PK 800 151 BE/G (0311)
UNO/DUO 120 A
Betriebsanleitung • Operating Instructions
2
1. Safety Precautions .................................. 3
1.1. For Your Orientation........................................................... 3
2. Understanding The Pumps...................... 4
2.1. Main Features..................................................................... 4
2.2. Differences Between The Pump Types.......................... 4
3. Installation ............................................... 5
3.1. Preparations For Installation............................................ 5
3.2. Pump Assembly And Location ......................................... 5
3.3. Connecting The Vacuum Side.......................................... 5
3.4. Connecting The Exhaust Side .......................................... 5
3.5. Connecting To The Mains Power Supply....................... 5
3.6. Monitoring Connection ..................................................... 7
3.7. Special Models................................................................... 7
4. Operations ................................................ 9
4.1. Filling Up And Checking The Operating Fluid ................ 9
4.1.2.Operating Fluid F5 .............................................................. 9
4.2. Switching The Pump ON And OFF................................... 9
4.3.1.Standard Model................................................................ 10
4.3.2 Corrosive Gas Model ....................................................... 10
4.3.3.Magnetic controllable gas ballast valve ..................... 11
4.4. Pumping Off Dry Gases ................................................... 11
4.5. Pumping Off Vapours....................................................... 11
5. What To Do In Case Of Breakdown...... 11
6. Maintenance .......................................... 12
6.1. Replacing The Operating Fluid....................................... 12
6.2. Re-Setting The Silencer ................................................. 12
6.3. Cleaning The Gas Ballast Valve..................................... 14
6.4. Shutting Down For Longer Periods ............................... 14
7. Service .................................................... 15
8. Technical Data ....................................... 16
8.1. Dimensions.........................................................................17
9. Accessories............................................ 17
10. Spare Parts ..............................................17
Contamination Declaration...................................................... 18
Conformity Declaration ............................................. (last page)
Index
Page
3
Danger of damage to the pump or system.
Danger of burns from touching hot parts.
Danger of an electric shock.
Read and follow all the instructions in this manual.Inform yourself regarding:Hazards which can be caused by the pump;Hazards which can arise in your system;Hazards which can be caused by the medium being
pumped.
Avoid exposing any part of your body to vacuum.Observe all safety and accident prevention regulations.Check regularly that all safety requirements are being
complied with.
Do not carry out any unauthorised conversions or
modifications on the pump.
When returning the pump to us please note the shipping
instructions in Section 7.
1.1. For Your Orientation
Instructions in the text
Operating instructions: Here, you have to do something
Symbols Used
The following symbols are used throughout in the illustrations:
Vacuum-flange Gas ballast
Exhaust flange Power connection
Position-Numbers
Identical pump and accessory parts have the same position numbers in all illustrations.
Danger of injury from moving parts.
Danger of personal injury.
1. Safety Precautions
1.2. Pictogramm-Definition
Modifications reserved
Attention to particularly important information on the product, handling the product, or to a particular part of the documentation.
WARNING
WARNING
WARNING
WARNING
CAUTION
PLEASE NOTE
4
2.1. Main Features
2. Understanding The Rotary Vane Pumps
2.2. Differences Between The Pump Types
Type Stage Nominal Attainable
flow rate Final Pressure at 50 Hz (Total, without
Gasballast)
UNO 120 A one stage 128 m3/h < 3 10
-2
mbar
DUO 120 A two-stage 128 m3/h < 2 10
-3
mbar
The UNO 120 A is a single stage and the DUO 120 A is a two stage rotary vane pump. Special features: – Pressurised oil lubrication – Hydraulically controlled high vacuum safety valve – Optional: Oil filter systems, Condensate separator,
Operational monitoring systems.
Proper Use
– Rotary Vane Pumps UNO/DUO 120 A may only be used to
generate vacuum. – Do not use the pump as a compressor. – Do not pump corrosive or explosive gases (a special versi-
on for these operations is available on request). – Do not operate the pumps where explosive processes are
involved. This is only permitted with the "explosive" version
of the pump. – Using accessories not named in this manual is not permit-
ted without authorisation from PFEIFFER. Improper use will cause any rights regarding liability and gua­rantees to be forfeited.
Special Models
Helium tight model:
– Operating Fluid P3 – Leak rate < 1
10
-5
mbar l/s
– Additional radial shaft seal with sealing oil chamber
Korrosivgasausführung:
– Operating fluid F5 – Leak rate < 1
10
-5
mbar l/s – Additional radial shaft seal with sealing oil chamber – Vanes made from special materials – Gas ballast valve in special version
Rotary Vane Vacuum Pumps UNO120 A, DUO 120 A
1 Sight glass
2 Gas ballast valve, complete;
3 Motor,
4 Exhaust flange,
5 Vacuum flange, 15 Casing 18 Baffle Plate 28 Support Stand 42 Operating fluid filler screw 43 Operating medium drain screw 45 Connection socket
142 15 5 4 2 3
43 18 45
28
Improper Use
The following are regarded as improper: – Uses not covered above, and, in particular:
– Connection to pumps and units which is not permitted
according to their operating instructions.
– Connection to units which contain touchable and voltage
carrying parts.
Improper use will cause any rights regarding liability and gua­rantees to be forfeited
5
3.4. Connecting The Exhaust Side
Pressure can rise to dangerous levels in exhaust lines. Lay exhaust lines without built-in shutoff devices. If a shut-off device is installed for
internal operating reasons or if there is a risk of excess pressure developing in the lines, official accident pre­vention regulations must be complied with. The exhaust line diameter should not be less than 63 DN. The line from the pump must be laid with a downward gradient so that condensate is prevented from flowing back into the pump. Otherwise a trap must be installed.
Expelled gases and vapours can represent a
hazard to health and/or the environment.
For this reason the relevant regulations must
always be complied with when working with
toxic substances. Only officially approved filters may be used for the separation of such substances. This also applies when handling toxic or chemically unstable gases and to gases susceptible to polymerisation or to the formation of peroxide (see 9. Accessories).
3. Installation
3.1. Preparations For Installation
Do not carry out any unauthorised conversions or modifications on the pump.
Fill up with operating fluid before operating the first time (included in a separate pack in the delivery consignment). When transporting use only the eyebolts provided for the purpose. When filled with operating fluid, do not tilt the pump otherwise operating fluid could escape via the exhaust opening.
3.2. Pump Assembly And Location
Where stationary installation is involved, anchor the pump
on site.
Sight glass 1 on the front side must be visible and gas bal-
last valve readily accessible.
When installing the pump in a closed housing, ensure the-
re is sufficient air circulation.
3.3. Connecting The Vacuum Side
Keep the connection between the vacuum chamber and
pump as short as possible.
Remove all scaling and loose particles from welded lines
prior to assembly.
– To protect the pump, separators, filters and cooling traps
etc.(see Section 9. Accessories) can be fitted upstream of the pump. However, the conductance value of the accessory causes a reduction in volume flow rate.
Where rotary vane vacuum pumps are
anchored, no forces from the piping system
should be allowed to act on the pump.
Suspend, decouple or support all piping
concerned with the pump.
1
Maximum angle of inclination to each side 10Þ Maximum erection height 2000 m N.N. Ambient temperature maximum 40 °C and minimum 12 °C Maximum relative humidity 85%
WARNING
CAUTION
CAUTION
CAUTION
max. 10°
WARNING
6
3.5. Connecting To The Mains Power Supply
Power connections must comply with local
regulations. Voltage and frequency information given on the rating plate must correspond to the mains voltage and frequency values.
When the electrical connections have been completed, carry out the earthed conductor test. The pump must be disconnected from the power before opening the ter­minal box. When switching on for the first time, check the direction of rotation:
Checking The Direction Of Rotation
Check the direction of rotation of the pump
before filling up with operating medium.
Remove blank flange on the exhaust side.Switch on pump for a short time (max. 5 seconds).
Motor/motor fan must turn clockwise (see arrow on motor fan cover).
If direction of rotation is incorrect: Interchange two
phases on the connecting cable.
Switching parts list
S3 Control switch S1 Pushbutton "OFF" S2 Pushbutton "ON" K1 Contactor F1-3 Motor fuse F5 Control fuse F8 Triggering Unit TMA 4 or TMA 4R H41 Signal lamp in Triggering Unit F8 H42 Signal lamp M1 Motor B1 Temperature sensor
R Reset button in Triggering Unit F8 (only for type TMA 4R)
Contactor control (short contact making with Triggering Unit Type TMA 4R).Manual reset (R) at triggering unit after thermal tripping. Thermal tripping displayed by H42 and H41.
Three phase PTC motors
The pumps are three phase current motor driven and are equipped with three PTC resistors built into the stator windings. The connection of a tripping unit will protect the pump against overloading (see connecting example).
Power section with motor to be protected
Tripping unit:
PTC tripping unit with two switchover contacts at the output manufactured by Ziehl, type MSR 220 F (see Section 9, „Accessories“), AEG type TMA 4R or Siemens type 3 UN8 004.
CAUTION
CAUTION
7
The Three Phase Current Motor Circuit
The three phase motor winding consists of three identical coils.
Delta Connection
The three coils are connected in series with the connection point connected to the mains. The voltage of each coil is the same as the mains voltage whereas the mains current is the cube root of the coil current. Delta connections are denoted by the symbol ∆. The voltage between the mains supply lines is called mains voltage. The mains current is the current which flows in the supply lines.
Star Connection
The ends of the three coils are connected at the star center. The terminal voltage is the cube root of the coil voltage; the mains and the coil current are the same. Star connections are denoted by the symbol Y. A star connection three phase motor for 400 V mains voltage has a coil voltage of 230 V.
Motor coil and connecting plate of Delta Connection
Motor coil and connecting plate of Star Connection
Rotation Speed and Operating Hours Monitor RCI 002
3.6. Monitoring Connection
The Rotary Vane Pumps UNO/ DUO 120 A have been designed so that they can be connected to a Rotation Speed and Operating Hours Monitor RCI 002, which is connected at connecting socket 45. For the operation and application of this device please refer to the relevant operations instructions.
Important: The instructions in 4.1.2 must be observed when working with synthetic operating medium F5.
3.7. Special Models
Helium Tight and Corrosive Gas Models
Helium tight and corrosive gas models are becoming increasingly important. – Helium tight models are distinguished by their high level
tightness to the atmosphere. Corrosive gas models are
manufactured with special materials. – There is a different high vacuum safety valve control. – The rotor shaft of both versions is sealed with an
additional radial shaft seal. The cavity between the radial
shaft seals 118 and 133 is filled with operating medium and
is supplied by the oiler 136. – If the gas ballast valve 2 is not available, the opening
should be closed with the locking screw 169 or a special
gas ballast valve with dosing device and hose connection
for flushing gas is available (on request). – Helium tight models are operated, as standard, with opera-
ting medium P3 and the corrosive gas version with synthe-
tic operating medium F5, also as standard.
L1
W2
W1
V2
L3
V1
U1
U2
L2
W2
L1
U1
V2
U2
V1
L2 L3
W1
L1
U1
V2
U2
W2
L1
U1
V1
L2 L3
W1
W1
L3
U2
W2
V2
V1
L2
CAUTION
8
Helium tight and corrosive gas models UNO/DUO 120 A
118 Radial shaft seal 122 Armature 126 O-ring 128 O-ring 132 Support washer 133 Radial shaft seal 134 Double supporting ring 135 Spacer 136 Oiler 138 O-ring 168 O-ring 169 Locking screw 180 Hose nozzle for inert gas inlet
169 180
122
132
133 135
126
128
118
134
136
168
138
Do not exceed 200 mbar over pressure
at inlet for inert gas at hose nozzle 180.
CAUTION
9
Rotary Vane Vacuum Pumps UNO120 A, DUO 120 A
1 Sight glass
2 Gas ballast valve, complete;
3 Motor,
4 Exhaust flange,
5 Vacuum flange, 15 Casing 28 Support Stand
14244 15 5 4 2 3
43
4.1. Filling Up And Checking The Operating Fluid
Standard version pumps are tested with opera­ting fluid P3. Guarantees relating to attainment of final pressures and trouble free functioning apply only providing this operating fluid is used.
The delivery consignment contains sufficient operating fluid for one filling.
Pumps for special applications (e.g. for pumping corrosive gases) can be operated with special operating fluids such as F5 and other PFEIFFER operating fluids. The use of such operating fluids requires prior authorisation from the manufacturer. P3 must be disposed of in accordance with local waste dispo­sal regulations.
Remove operating fluid filler screw 42.Fill up with operating fluid included in the consignment
(only for P3); refer to rating plate for correct quantity.
Replace operating fluid filler screw 42; be careful with the
O-ring.
Check operating fluid level only when the pump is warm
and running and vacuum connection 5 and gas ballast val­ve are closed.
The level of operating fluid must always be between the
top and the middle of the sight glass 1.
Topping up operating fluid is possible when the pump is
running and in final vacuum operations.
In non-stop operations check operating fluid daily,
otherwise whenever the pump is switched on.
4.1.2. Operating Fluid F5
Only the synthetic operating fluid F5 may be used for corrosi­ve gas model pumps.
The use of other operating fluid must be cleared by the manu­facturer.
When working with synthetic operating media, the manufacturer’s directions must be observed. If ‘’F5’’ is heated above 300°C, toxis vapours which can harm the airways are given off.
Do not allow ‘’F5’’ to come into contact with tobacco products (danger of poisoning when ignited). The precautions necessary in the handling of chemicals must be observed.
Please note:
If there is a noticeable loss of operating fluid, the radial shaft seals should be checked. If operating fluid is leaking out underneath the casing 15 from the opening between the support stand 28 and motor 3, the radial shaft seal must be replaced. Please contact your loca Pfeiffer-Service (see Section 7.).
4.2. Switching The Pump ON And OFF
When operating the pump gets hot.
The pump can be switched ON and OFF in any pressure range. On switching off, the built-in high vacuum safety valve 6 to the vacuum plant closes automatically and vents the pump. Minimum start-up temperature, in compliance with German Industrial Standard DIN 28 426: + 12 °C. On switching OFF (intake pressure < 750 mbar) the built-in high vacuum safety valve to the vacuum chamber closes automatically and the pump is vented.
4. Operations
42 Operating fluid filler
screw
43 Operating medium drain
screw
28
WARNING
CAUTION
CAUTION
10
4.3. Gas Ballast Valve
4.3.1 Standard Model
Gas ballast valve 2 is fitted with a sliding sleeve which, when pushed towards the pump, opens the valve and when pushed away from the pump, shuts the valve (Fig. 7). Intermediate settings are not possible.
Gas ballast valve standard model
2 Gas ballast valve with sliding sleeve
28 Support stand
2
28
Gasballast valve helium tight and corrosive gas models
4.3.2 Corrosive Gas Model
Gas ballast valve 2 is fitted with a dosing device and a flushing gas connection. It has a sliding sleeve which opens the valve when pushed towards the pump and shuts the valve when pushed away from the pump (Fig. 8). The dosing device is closed when the groove in the dosing screw is in the horizontal position; in the vertical position the dosing device is fully open. A flow meter enables any intermediate gas volu­me setting between fully open and fully closed to be made.
Please note:
Flushing gas must be connected when operating corrosive gas version pumps. The gas ballast valve must be open. Flus­hing gas pressure may not exceed 1.5 bar. Our specialists would be pleased to give advice regarding the type and volu­me of flushing gas required.
11
4.4. Pumping Off Dry Gases
No special measures are necessary. The gas ballast valve can remain closed.
4.5. Pumping Off Vapours
To prevent their condensation in the pump,
vapours should only be pumped off with the
pump at operating temperature and the gas bal-
last valve 2 open.
Observe pump water vapour tolerance levels (see Section 8.0 Technical Data). When the process has been completed, allow the pump to continue running for approx. 30 minutes for operating medium cleaning purposes.
5. What To Do In Case Of Breakdown?
Problem Possible Causes Remedy
Pump doesn't attain final pressure Pump dirty Operate pump for a longer period with open
Operating fluid dirty gasballast valve or change operating fluid Leak in system Repair leak
Unusual operating noises Damaged pump stages Request repair by PFEIFFER Service
Damaged coupling piece Request repair by PFEIFFER Service Oil leaking from the foot of the pump Radial shaft seal defective Request repair by PFEIFFER Service Pump doesn't start Ambient temperature < 12 °C Warm up pump
Dirty pump stages Request cleaning by PFEIFFER Service
Damaged pump stages Request repair by PFEIFFER Service
Further repairs can only be carried out by trained specialists in accordance with Service Instructions.
4.3.3.Magnetic controllable gas ballast valve
21 Solenoid valve 22 Hose nozzle 23 Double nipple 24 Sealing ring 25 O ring 26 Silencer 27 Sleeve 29 Valve housing
Solenoid valve scheme stromlos closed
26
24
21
23
22
24 25 27
29
Do not exceed 500 mbar over pressure at inlet for inert gas at hose nozzle 22.
Assemble solenoid valve reversion kit for gas ballast valve
Push back sliding sleeve of gas ballast valve 2 in standard
version.
Unscrew valve housing 29 from the stand by means of a
pin (Ø 8).
Assemble sleeve 27 and valve housing 29 and screw in the
stand by means of a pin (Ø 8).
Screw in double nipple 23 and solenoid valve as shown in
the illustration.
hole for pin Ø 8
CAUTION
CAUTION
12
Screw in operating fluid drain screw 43.Allow pump to run for about 10 seconds with open vacuum
flange.
Drain off remaining operating fluid.In the case of serious contamination, the operating fluid
will have to be changed several times.
Screw in operating fluid drain screw 43 with O-ring.Fill in fresh operating fluid and check level as per
section 4.1.
Operating fluid can contain substances from the medium pumped. Operating fluid must be disposed of in accordance with the local respective regulations.
Please request safety instruction data sheets for operating fluids from PFEIFFER or download from the INTRANET/ INTERNET.
6.2. Re-Setting The Silencer
If the silencer 13 is not functioning, a knocking noise becomes audible at end vacuum. This does not cause dama­ge to the pump but it may be disturbing in quiet rooms. Setting can be carried out while the pump is running and the intake socket and gas ballast valve are closed, using a normal screwdriver:
The pump must have reached its operating temperature.Remove locking screw 44 (UNO 120A) or operating medium
filler screw 42 (DUO 120 A).
Using a screwdriver, open the valve spindle 47 approx. 3
turns (flushing) for some seconds and then close it again completely (clockwise).
Slowly open the valve spindle 47 (1/2 to a complete turn)
until the knocking noise is no longer audible.
Replace locking screw 44 or operating medium filler screw
42 paying attention to the O-ring.
Please note:
Silencer cleaning is described in the service instructions.
When carrying out servicing work, take the following precautions:
Always ensure the pump cannot be switched
on when carrying out any work on the pump. If necessary, remove the pump from the system to carry out inspection work.
Only dismantle the pump as far as is necessary in order to
repair defects.
Dispose of used operating fluid in compliance with local
regulations.
When working with synthetic operating media, toxic
substances and substances contaminated with corrosive gases, the relevant instructions governing their use must be observed.
Only use benzine or similar agents for cleaning pump
parts. Do not use soluble cleaning agents.
6.1. Changing The Operating Fluid
The rate of deterioration of the operating fluid depends on the pump applications. The level of deterioration of organic operating fluids (for
example P3) can be read off from the colour scale in accordance with DIN 51578 on the supplementary sheet PK 800 219 BN/B, on request. – The colour scale enables precise colour determination.
Fill a specimen in a test tube or some similar vessel and
test by holding against the light.
Where the discolouration is dark yellow to red brown
(equivalent to 4 ... 5 on the scale) change operating fluid.
Where organic operating fluids, for example P3,
are involved, the operating fluid should be changed at least once a year!
Replacing The Operating Fluid
Switch off the pump.Unscrew operating fluid drain screw 43 and drain
operating fluid.
The operating fluid temperature can be as high
as 80 °C. During maintenance and repair work, process related toxic gases and vapours can escape from the operating fluid which may become contaminated with harmful substances
(radioactive, chemical etc.). Disposal of used operating fluid is subject to the relevant regulations.
6. Maintenance
WARNING
CAUTION
WARNING
WARNUNG
13
47
48
50
49
129
51
52
56
54
55
51
52
53
179
58
57
174 31 28 34
61
59
68
60
174 28 31 66 62
72 34 71 70 69
63
64 65 67
68
59
Helium-tight and korrosive gas models
2 Gas ballast 13 Silencer 14 Non-return valve 28 Support stand 31 O-ring 34 Cover 47 Valve spindle 48 O-ring 49 O-ring 50 Valve housing 51 Plate spring
52 Screw 53 O-ring 54 Screw 55 Insert 56 Screw 57 O-ring 58 Centering sleeve 59 Hole 60 O-ring 61 Sleeve 62 Flushing gas connection 63 O-ring
64 Dosing valve unit 65 O-ring 66 Dosing insert 67 Safety ring 68 Valve housing 69 O-ring 70 Insert 71 O-ring 72 Intermediate piece 129 Centering ring 174 Screw connection 179 Housing non-return valve
14
6.3. Cleaning The Gas Ballast Valve
Gas ballast valve 2 can only get dirty if the ambient air sucked in is dusty.
Assembling and dismantling the Standard Model
Insert round steel bar into boring 59 and remove gas
ballast valve 2; take care with the O-ring.
Remove sleeve 61.Remove insert 70 (see "Corrosive Gas Model"). Four bore
holes have been provided on the front side for this operati­on.
Clean all parts.Check O-ring 60 for wear and replace if necessary.Lightly oil O-rings with operating medium.Re-assemble the parts in reverse order.
Assembling and dismantling the Helium Tight and Corrosive Gas Models
Insert round steel bar into boring 59 and remove valve hou-
sing 68 taking care with the O-ring 71. Connecting piece 72 is not removed.
Remove dosing valve unit 64 from 68.Remove safety ring 67 and dosing insert 66 from dosing
valve unit 64.
Remove insert 70. Four borings have been provided on the
front side for this operation.
Clean all parts.Check O-rings 65 and 69 for wear and replace if necessary.Lightly oil O-rings with operating medium.Re-assemble the parts in reverse order.
6.4. Shutting Down For Longer Periods
Should the pump have to be shut down for an longer period, the complete pumping system must be adequately protected against corrosion.
Switch off pump.Unscrew operating fluid drain screw 43 and drain
operating fluid as per Section 6.1.
Screw in operating fluid drain screw 43; keep an eye on
the position of the O-ring.
Allow pump to run for about 10 seconds with open vacuum
flange. .
Drain off remaining operating fluid.Fill up pump with fresh operating fluid:
Filling: UNO 120 A --> 21 l
DUO 120 A --> 18 l
If this amount of operating fluid is filled in, the level is extremely high but this is necessary to protect all parts of the pump against corrosion.
Before operating, the operating fluid must be reduced to its normal level.
CAUTION
15
7. Service
Decontaminate units before returning or possible disposal. Do not return any units which are microbiologically, explosively or radio­actively contaminated.
Returning Contaminated Units
If contaminated units have to be returned for maintenance/repair, the following instructions concerning shipping must be followed without fail:
Neutralise the pump by flushing with nitrogen or dry air.Seal all openings to the air.Seal pump or unit in suitable protective foil.Ship units only in appropriate transport containers.
Repair orders are carried out according to our
general conditions of sale and supply.
If repairs are necessary, please send the unit together
with a short damage description to your nearest PFEIFFER Service Center.
Contact Addresses And Telephone Hotline
Contact addresses and service hotlines can be found on the back cover of these operating instructions.
Do Make Use Of Our Service Facilities
In the event that repairs are necessary to your pump, a number of options are available to you to ensure any system down time is kept to a minimum: – Have the pump repaired on the spot by our PFEIFFER
Service Engineers;
– Return the individual components to the manufacturer for
repairs;
– Replace individual components with a new value
exchange units.
Local PFEIFFER representatives can provide full details.
Before Returning:
Dismantle all accessories.Drain lubricant.Attach a clearly visible notice: “Free of contamination” (to
the unit being returned, the delivery note and accompany-
ing paperwork). Harmful substances" are substances and preparations as defined in current legislation. PFEIFFER will carry out the decontamination and invoice this work to you if you have not attached this note. This also applies where the operator does not have the facilities to carry out the decontamination work. Units which are contaminated microbiologically, explosively or radioactively cannot be accepted as a matter of principle.
Fill Out The Contamination Declaration
In every case the "Contamination Declaration" must be
completed diligently and truthfully. A copy of the completed declaration must accompany the
unit; any additional copies must be sent to your local
PFEIFFER Service Center. Please get in touch with your local PFEIFFER representatives if there are any questions regarding contamination.
WARNING
PLEASE NOTE
16
Single stage Unit UNO 120 A Two stage DUO 120 A
Connection nominal diameter
Inlet ISO-F DN 63 DN 63 Outlet ISO-F DN 63 DN 63
Rated volume flow rate at
50 Hz m3/h 128 128 60 Hz m3/h 154 154
Volume flow rate
50 Hz m3/h 120 120 60 Hz m3/h 144 144
Final pressure
total without gas ballast mbar < 3 10
-2
< 3 10
-3
total with gas ballast mbar < 1 < 6 10
-3
Auspuffdruck min mbar > 250 > 250 Auspuffdruck max mbar < 1500 < 1500
Water vapour compatibility
50 Hz mbar 33 20
60 Hz mbar 33 20 Water vapour acceptable capacity 50 Hz g/h 2800 1800 60 Hz g/h 3360 2150
Noise development
without gas ballast dB(A) 60 60 Operating fluid, amount l 17 13
Max. admissible operating temperature
1)
°C 90 90
Nominal rotation speed of pump
at 50 Hz 1/min 965 965
at 60 Hz 1/min 1158 1158 Nominal motor power
three phase motor with motor protection switch at 50 Hz kW 4 4
at 60 Hz kW 4 4 three phase motor with temperature sensor at 50 Hz kW 4 4
at 60 Hz kW 4 4 Weight pump with
three-phase motor. kg 193 215
1)
The maximum admissible operating temperature of 90 °C is valid for an ambient temperature of 25 °C and operating fluid P3, without gas ballast. Operating temperatures in the range of 110 ... 115 °C could be caused by increased ambient temperature to max. 40 °C (for operating fluid P3).
8. Technische Daten
17
8.1. Dimensions
9. Accessories
10. Spare parts
Pos. Description Quantity Size Number Comment Order Quantity
Set of seals 1 PK 215 099-T
1 Sight glass 1 R 15’’ PK 213 140 2 Gas ballast valve PK 215 022-U
29 Sieve insert 2 DN 63 ISO PK 300 010 -X
45 PCB 1 PK 215 306 -X 172 O-ring 2 75,56 x 5,33 P 4071 054 PV 173 O-ring 1 38 x 3 P 4070 621 PV
187 Sealing ring 2 20/25,5 x 2,4 P 3191 476
A comprehensive range of accessories can be found in section B of the Pfeiffer catalogue (order no. PA 800 066 PE), to: – extend the application range of the pumps and – protect the pump from wear, when used under extreme
operating conditions.
The tripping device to protect against overloading the pump can be ordered under the order number P 4768 051 FQ.
When ordering accessories and spare parts please be sure to state the full part number. When ordering spare parts please state additionally the unit type and unit number (see rating plate). Please use this list as an order form (by taking a copy).
18
Tradename Chemical name Danger class Precautions associated Action if spillage or human Product name (or Symbol) with substance contact Manufacturer
1.
2.
3.
4.
5.
5. Legally Binding Declaration
I hereby declare that the information supplied on this form is complete and accurate. The despatch of equipment will be in accordance with the appropriate regulations covering Packaging, Transportation and Labelling of Dangerous Substances.
Name of Organisation: _______________________________________________________________________________
Address: _____________________________________ Post code: _____________________________________
Tel.: ______________________________________________________________________________________
Fax: _____________________________________ Telex: ________________________________________
Name: ______________________________________________________________________________________
Job title: ______________________________________________________________________________________
Date: _____________________________________ Company stamp: ________________________________
Legally binding signature: _____________________________________________________________________________
Declaration of Contamination of Vacuum Equipment and Components
The repair and/or service of vacuum components will only be carried out if a correctly completed declaration has been submitted. Non-completion will result in delay.
The manufacturer could refuse to accept any equipment without a declaration.
This declaration can only be completed and signed by authorised and qualified staff:
1. Description of component:
- Equipment type/model: _________________________
- Code No.: __________________________
- Serial No.: __________________________
- Invoice No.: __________________________
- Delivery Date: __________________________
3. Equipment condition
- Has the equipment been used? yes no ❐
- What type of pump oil was used?
___________________________________________
- Is the equipment free from potentially harmful
substances?
yes (go to section 5) no (go to section 4)
4. Process related contamination
of equipment
- toxic yes ❐ no ❐
- corrosive yes ❐ no ❐
- microbiological hazard*) yes ❐ no ❐
- explosive*) yes ❐ no ❐
- radioactive*) yes no
- other harmful substances yes ❐ no ❐
*) We will not accept delivery of any equipment that has been radioactively or microbiologically contaminated without written
evidence of decontamination!
2. Reason for return:
_____________________________________________
_____________________________________________
_____________________________________________
_____________________________________________
_____________________________________________
Please list all substances, gases and by-products which may have come into contact with the equipment:
im Sinne folgender EU-Richtlinien:
pursuant to the following EU directives:
- Maschinen/
Machinery
98/37/EG
- Elektromagnetische Verträglichkeit/
Electromagnetic Compatibility
89/336/EWG
- Niederspannung/
Low Voltage
73/23/EWG
Hiermit erklären wir, daß das unten aufgeführte Produkt den Bestimmungen der EU-Maschinenrichtlinie 98/37/EG - Anhang IIA, der EU-Richtlinie über elektromagnetische Verträglichkeit 89/336/EWG und der EU-Niederspannungsrichtlinie 73/23/EWG entspricht.
We hereby certify that the product specified below is in accordance with the provision of EU Machinery Directive 98/37/EEC - Annex II A, EU Electromagentic Compatibility Directive 89/336/EEC and EU Low Voltage Directive 73/23/EEC.
Produkt/
Product
:
UNO/DUO 120 A
Angewendete Richtlinien, harmonisierte Normen und angewendete, nationale Normen:
Guidelines, harmonised standards, national standards which have been applied:
EN 292-1 60335-1, -41 EN 292-2 EN 61 010 EN 294 EN 60 204 EN 50 081-1 EN 1012-2 EN 50 082-2
Unterschrift/
Signature:
Unterschriften:
(W. Dondorf) Geschäftsführer
Managing Director
Pfeiffer-Vacuum GmbH Emmeliusstrasse 33 35614 Asslar Germany
Konformitätserklärung
Declaration of Conformity
Konf.I/2000
(03/04)
Adressen • Addresses
Argentina
ARO S.A., Av. Belgrano 369, 1092 Buenos Aires, Phone: +54 / 11 4331 5766, Fax: +54 / 11 4331 3572, Email: info@aroline.com.ar
Australia
Scitek Australia Pty. Ltd., Suite 1B, 10-18 Cliff Street, Milsons Point, NSW 2061, Phone: +61 / 2 9954 1925, Fax: +61 / 2 9954 1939, Email: contact@scitek.com.au
Austria
Pfeiffer Vacuum Austria GmbH Diefenbachgasse 35, A-1150 Wien, Phone: +43 / 1 8941 704, Fax: +43 / 1 8941 707 Service Hotline: +43 / 1 8941704, Email: office@pfeiffer-vacuum.at
Branch Office, Czech Republic
Pfeiffer Vacuum Austria GmbH, Branch Prague Zvonarska 885, CZ-156 00 Praha 5 Phone: + 420/257 923 888, Fax: + 420/257 923 014 Email: office@pfeiffer-vacuum.cz
Belgium / Luxemburg
Pfeiffer Vacuum Belgium N.V./S.A. Luxemburgstraat 5, B-9140 Temse Phone: +32 / 3 710 5920, Fax: +32 / 3 710 5929 Service Hotline: +32 / 3 710 5922, Email: Sales@pfeiffer-vacuum.be
Brazil
Elmi Tec Assistencia Técnica e Representação S/C Ltda. Rua Bernadino de Compos, 551-Brooklin CEP 04620-002 São Paulo, SP - Brasil Phone: +55 / 11 5532 0740, Fax: +55 / 11 5535 3598, Email: elmi-tec@elmi-tec.com.br
Chile
BERMAT S.A., Coyancura 2283, Oficina 601 Providencia, P.O. Box 9781, Santiago Phone: +56 / 2 231 8877, Fax: +56 / 2 231 4294, Email: bermat@bermat.cl
Colombia
Arotec Colombiana S.A., Carrera 16 No. 36-95 Apartado 050 862, Santafe de Bogota / Colombia Phone: +57 / 1 288 7799, Fax: +57 / 1 285 3604, Email: arotec@arotec.net
Denmark
Pfeiffer Vacuum Scandinavia AB, Vesterengen 2, DK-2630 Taastrup, Phone: +45 / 43 52 38 00, Fax: +45 / 43 52 38 50, Email: sales@pfeiffer-vacuum.dk
France
Pfeiffer Vacuum France SAS 45, rue Senouque, BP 139 F-78531 BUC Cedex, Phone: +33 / (0)1 30 83 04 00, Fax: +33 / (0)1 30 83 04 04, Email: info@pfeiffer-vacuum.fr
Germany
Pfeiffer Vacuum GmbH, Berliner Strasse 43, D-35614 Asslar Phone: +49 / 6441 802 400, Fax: +49 / 6441 802 399 Service Hotline: +49 / 6441 802 333, Email: Info@pfeiffer-vacuum.de
Great Britain
Pfeiffer Vacuum Ltd. 2-4 Cromwell Business Centre Howard Way, Interchange Park Newport Pagnell, MK16 9QS, United Kingdom Phone: +44 / 1 908 500615, Fax: +44 / 1 908 500616, Email: sales@pfeiffer-vacuum.co.uk
Greece
Analytical Instruments S.A., 1 Mantzarou St., GR-15451 Athens, Phone: +30 / 210 674 8973-7, Fax: +30 / 210 674 8978, Email: contact@analytical.gr
Hong Kong
Pfeiffer Vacuum Asia Ltd. c/o Hakuto Enterprises Ltd. 8th Floor World Trade Center No. 280 Gloucester Road Causeway Bay Hong Kong Phone: +852 2578 4921, Fax: +852 2807 2498, Email: sales@pfeiffer-vacuum.com.hk
India
Pfeiffer Vacuum India Ltd. 25-5 Nicholson Road, Tarbund Secunderabad 500 009, Phone: +91 / 40 2775 0014, Fax: +91 / 40 2775 7774, Email: pfeiffer@vsnl.net
Israel
ODEM Scientific Applications Ltd. 9 Hamazmera St., P.O.B. 2001 Nes Ziona Phone: +972 8 938 0333, Fax: +972 8 938 0334, Email: odemltd@isdn.net.il
Italy
Pfeiffer Vacuum Italia S.p.a. Via San Martino, 44 I-20017 RHO (Milano) Phone: +39 / 2 93 99 051, Fax: +39 / 2 93 99 05 33, Email: contact@pfeiffer-vacuum.it
Japan
Hakuto Co. Ltd., C.P.O. Box 25, Vacuum Scientific Instruments Division C.P.O. Box 25, Tokyo 100-91, Phone: +81 / 3 32 258 910, Fax: +81 / 3 32 259 011, Email: pfeiffer@hakuto.co.jp
Korea
Pfeiffer Vacuum Korea Ltd., 703 Ho, 853-1, Hankuk Mulru, Dongchonri, Suzi-Eup, 449-843 Yong-in City, Kyungkido, Phone: +82 / 31 266 0741, Fax: +82 / 31 266 0747, Email: sales@pfeiffer-vacuum.co.kr
Netherlands
Pfeiffer Vacuum Nederland BV Veldzigt 30a, NL-3454 PW De Meern, Phone: +31 / 30 6666050, Fax: +31 / 30 6662794, Email: Sales@pfeiffer-vacuum.nl
Peru
Ing. E. Brammertz s.r.l., José Pardo 182, Apartado 173, PE-18 Miraflores, Phone: +51 / 1 445 8178, Fax: +51 / 1 445-1931, Email: braming@terra.com.pe
Poland
Softrade Sp.z.o.o, ul. Grunwaldzka 391, PL-60-173 Poznan, Phone: +48 / 61 8677 168, Fax: +48 / 61 8677 111, Email: softrade@softrade.com.pl
Portugal
Tecnovac Tecnologia de Vacio S.L., Sector Literatos, 38 - Local 1, ES-28760 Tres Cantos (Madrid) Phone: +34 / 91 804 11 34, Fax: +34 / 91 804 30 91, Email: tecnovac@tecnovac.es
Singapore
APP Systems Services Pte Ltd, 11, Toh Guan Road East, #03-01 APP Enterprise Building, Singapore 608603, Phone: +65 / 6425 6611, Fax +65 / 6560 6616, Email: sales@appsystems.com.sg
Spain
Tecnovac Tecnologia de Vacio S.L., Sector Literatos, 38 - Local 1, ES-28760 Tres Cantos (Madrid) Phone: +34 / 91 804 11 34, Fax: +34 / 91 804 30 91, Email: tecnovac@tecnovac.es
Sweden
Pfeiffer Vacuum Scandinavia AB Johanneslundsvägen 3 SE-194 61 Upplands Väsby, Phone: +46 / 8 590 748 10, Fax: +46 /8 590 748 88, Email: sales@pfeiffer-vacuum.se
Switzerland
Pfeiffer Vacuum Schweiz S.A. Förrlibuckstraße 30, CH-8005 Zürich Phone: +41 / 1 444 2255, Fax: +41 / 1 444 2266 Email: info@pfeiffer-vacuum.ch
South Africa
Labotec Pty Ltd., P.O. Box 6553, Halfway House, 1685 Midrand Phone: +27 / 11 315 5434, Fax: +27 / 11 315 5882, Email: sales@labotec.co.za
Taiwan
HAKUTO Taiwan Ltd. Hsinchu office No. 103, Hsien Chen 11th Street, Chupei, HsinChu County, Taiwan, R.O.C. (zip/postal code: 302) Phone: +886 / 3 554 1020, Fax: +886 / 3 554 0867, Email: leo-shih@hakuto.com.tw
Thailand
Hakuto (Thailand) Ltd. 18th Floor, Chokchail Intíl Bldg. 690 Sukhumvit Road, Klongton, Klongtoey, Bangkok 10110 Phone: +662 / 259 6244, Fax: +662 / 259 6243, Email: tmgroup@hakutothailand.com
U.S.A.
Pfeiffer Vacuum, Inc. 24 Trafalgar Square Nashua, NH 03063-1988, USA Phone: +1/ 603 578 6500, Fax: +1/ 603 578 6550, Email: contact@pfeiffer-vacuum.com
Venezuela
Secotec S.A., Apartado 3452, Caracas 1010-A, Phone: +58 / 212 573 8687, Fax: +58 / 212 573 1932, Email: secotec@cantv.net
Other countries
AVI - Applied Vacuum Industries GmbH Leginglenstrasse 17A; CH-7320 Sargans Switzerland Phone: +41 / 81 710 03 80 Fax: +41 / 81 710 03 81 Email: avi@bluewin.ch
Scope of represented countries
Armenia, Azerbaijan, Bangladesh, Belarus, Bulgaria, Cambo­dia, Estonia, Georgia, Kazakhstan, Kingdom of Nepal, Kirghi­zia, Latvia, Lithuania, Maldavia, Philippines, P.R. China, Rumania, Russia, Tajikistan, Turkmenistan, Ukraine, Uzbeki­stan, Vietnam
A.E.M.S.
Advanced Equipment Materials and Systems P.O. Box 25 Föhrenweg 18 FL-9496 Balzers Phone: +423/ 380 0550 Fax: +423/ 380 0551 Email: aems@adon.li
Scope of represented countries
Bahrain, Egypt, Iraq, Iran, Jordan, Kuwait, Lebanon, Lybia, Oman, Pakistan, Saudi-Arabia, Sudan, Syria, Turkey, United Arab Emirates, Yemen
Zentrale/Headquarters
Pfeiffer Vacuum GmbH Berliner Strasse 43 D-35614 Asslar
Telefon 06441/802-0 Telefax 06441/802-202 Hotline 06441/802-333 www.pfeiffer-vacuum.net
Loading...