Pfaudler TW Series, T Series, P Series, FT Series, FS Series Operating Instructions Manual

Temperature probes
type T/TW
Contents
Content
1 General
1 Preface
1.1 Introduction
2 General instructions
1.2 Scope
2.1 Operating range
2.2 Warnings contained in the operating
2 Safety instructions
instructions (OI)
2.1 General safety notes
2.3 Notes on deliveries
2.2 Product-specific safety notes
2.4 Transport and storage
2.3 Safety instructions for accessing reactors
2.5 Warranty notes
2.4 Avoiding damages to the glass
2.6 Notes on return deliveries
3 Storage
3 Safety
4 Transport
3.1 Proper use
5 Product description
3.2 Qualified personnel
4 Glass-lined temperature probes
4.1 Probe types
4.2 Joint features and properties
4.3 General information
4.4 Avoiding damages to peripheral devices
Contents
5 Temperature probe type T
5.1 Measuring principle
5.2 Construction of the T probe
1 General
5.3 Transmitter TTH300 Ex
1.1 Introduction
5.4 Technical data of T probe
1.2 Scope
5.5 Installation of the T probe
1.3 Continuous glass monitoring
5.6 Connection of the T probe
5.7 Start-up and maintenance of the T probe
2 Safety instructions
5.8 Calibration of the T probe
2.1 General safety notes
5.9 Explosion protection
2.2 Product-specific safety notes
6 Temperature probe type TW
2.3 Safety instructions for accessing
6.1 Measuring principle
reactors
6.2 Construction of the TW probe
6.3 Transmitter TTH300 Ex
2.4 Avoiding damages to the glass
6.4 Technical data of TW probe
3 Storage
6.5 Installation of the TW probe
4 Transport
6.6 Connection of the TW probe
5 Product description
6.7 Start-up and maintenance of the TW probe
6.8 Calibrating the TW probe
6.9 Explosion protection
Operating Instructions
302-10 e
Annex 1
PTB 03 ATEX 2132 X
Annex 2
Declaration of conformity
Annex 3
A 3.1 Potentially hazardous
atmospheres A 3.2 Atmospheric conditions A 3.3 Equipotential bonding A 3.4 Lightning protection A 3.5 Quatro-Pipe
Temperature probes type T/TW
Temperature probes type T/TW

1 Preface

The present operating instructions are designed to familiarize users with the construction of the glass-lined tempera­ture probes and their use.
The operating instructions should be accessible to the operating and mainte­nance personnel in order to ensure that all necessary information is available for any assembly and maintenance work. By knowing these operating instructions (OI), you can avoid damage to the mea­suring equipment and ensure trouble-free operation. The information contained in these operating instructions corre­sponds to the state of the art at the time it is printed and is provided to the best of our knowledge. We reserve the right to include any improvements, amendments and new developments in the operating instructions without prior notice. The actual design of products may differ from the information provided in the catalog if warranted by technical modifications resulting from product improvements. The proposal submitted by Pfaudler for a concrete application will be binding in this case. The latest edition will always supersede all previous ones.
The present operating instructions are made available to our customers and interested parties free of charge. Reprints and copies as well as transformation into electronic forms, in whole or in part, shall require our written approval.

2.2 Warnings contained in the operating instructions (OI)

In the operating instructions, the danger symbol m is used to draw your attention to especially impor­tant safety instructions. Compliance with these instructions is manda­tory, because adherence to the instructions can avoid severe dam­age to people and/or equipment.

2.3 Notes on deliveries

The respective scope of delivery is speci­fied on the shipping documents attached to the shipment and corresponds to the valid purchase agreement.
Check that the items delivered are com­plete and intact.
If possible, keep the packing material for re-use for possible return shipments.

2.4 Transport and storage

The probes should only be transported and stored in their closed original pack­aging. Where it is no longer available, the probes must be protected against shock and impacts. In order to guarantee an as-new condition of the probes, maintain the following storage conditions:
n dry and dust-free n steady temperature and ventilation
The probes do not need any preserva­tives, they are resistant to normal envi­ronmental influences.

2.6 Notes on return deliveries

Before sending used probes to Pfaudler Werke GmbH or third parties for repair or other purposes, all parts must be cleaned and decontaminated.
To protect our staff and for insurance reasons, your return shipment must be accompanied by a clearance certificate (refer to publication 332) on which you confirm that the probe was properly cleaned and decontaminated. You may obtain a form sheet for this purpose from us on request.

3 Safety

For detailed safety instructions and infor­mation concerning explosion protection, please refer to the end of the operating instructions.

3.1 Proper use

Any use of the probe for purposes other than described in the present operat­ing instructions will adversely affect the safety and functioning of the measuring device and is therefore not allowed.
It is important to note the safety instruc­tions applicable to the electrical systems and equipment and all explosion protec­tion provisions, if any.
m Do not practice any working
me th od s wh ic h ma y en d a n ge r sa fe ty.
All rights reserved.

2 General instructions

2.1 Operating range

Glass-lined temperature probes are used to measure the product temperature in reactors and storage vessels as well as in pipelines.
For the resistance of the probes, please refer to our publication no. 614.
Never operate this measuring device outside its permissible operating condi­tions.
2

2.5 Warranty notes

Any warranty claims shall not be extended or limited by the information contained in the present operating instructions.
For the exact warranty conditions, please refer to the Terms of Sale of Pfaudler Werke GmbH as amended.
3.2 Qualified personnel
The probe may only be installed, started up and serviced by authorized personnel with special skills in measuring technol­og y and in strict compliance with the pre­sent Operating Instructions as well as the valid provisions.
The failure to observe these instructions – no matter whether intentionally or neg­ligently – releases Pfaudler Werke GmbH from all liability and warranty claims.
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 2
4 Glass-lined temperature




 

 


 
probes
t [%]
Temperature probes type T/TW
Temperature probes type T/TW

4.1 Probe types

Pfaudler offers two different types of
100
probes for temperature measurement which operate according to different physical measuring principles.
These probe types are:
n Temperature probe type T
E T/TW Ta
With this probe, a thermocouple per­forms the sensing function. Thermo­couples are based on the physical
50
effect of a charge shift in an electrical conductor (wire) along a temperature gradient. The free electrons move from the hot end of the wire to the cold end of the wire.
n Temperature probe type TW
This probe is equipped with a resist­ance thermometer as the sensor. Resistance thermometers are based on the physical effect that the elec­trical resistance of conductors and semiconductors changes subject to the temperature.

4.2 Joint features and properties

Both probe types have a sensor that is fused into the glass. Thus, the sensor is excellently protected against aggres­sive media, and its placed directly in the product. This arrangement offers users the following advantages:
n The probes have very short response
times. The response behavior of the measuring probes type T and TW is shown in Fig. 1 in comparison to
0 10 20 30 40 50 [s]
Measured in inside the reactor (100 l) with water at 120 rpm T/TW measuring probe type T/TW on
thermometer well, D = 40 mm
Ta thermocouple with tantalum sensor in
thermometer well, D = 40 mm
E Reference measurement with unprotected
pallaplat element (cannot be used in glass-lined reactor)
Fig. 1 Time response
MT_05_00001
probes with plugged-in sensors.
n The sensors of the probes have not
been simply screwed-in, producing error-prone sealing surfaces in the product.
n Due to the quick and precise tem-
perature measurement, the produc­tion process can be controlled safely and economically. In many cases, the product quality is also improved.
n The sensors are not subject to ageing
because they are made of extremely pure and highly resistant noble met­als and have been fused into the glass layer in a chemically inert process. Glass is absolutely diffusion-tight, therefore, reaction with hydrogen is also excluded.
© Pfaudler GmbH ·
OI 302-8 e
© Pfaudler GmbH 3
3
Temperature probes type T/TW
Temperature probes type T/TW

4.3 General information

During a spark test, inflammable sparks may occur at the pores in the form of an electric arc. Therefore, spark testing may only be carried out outside of potentially explosive atmospheres.

4.4 Avoiding damages to peripheral devices

When performing a spark test, please note the following information; other­wise, the components and/or the elec­tronic transmitter may get damaged.
n The measuring transducers for tem-
perature, glass monitoring, capacitive sensors and other electronic/electric components that have been attached to the valve or the baffle must be dis­connected prior to the spark test.
n Suitable equipment must be employed
for the test (impulse voltage). We rec­ommend using the GlaSparker®, our high-voltage tester
n For glass lined measuring probes, the
max. test voltage must not exceed 7 k V .
n The contact window around the meas-
uring probe site (e. g. P) must not be tested.
In general, however, we recommend call­ing our technical service for performing the test.

5 Temperature probe type T

5.1 Measuring principle

The temperature probe Ty p e T will also be referred to as the T probe in the present Operating Instructions.
As already mentioned, a thermocouple serves as the sensor of the T probe. A thermocouple consists of two electrical conductors made of different materials (wires) that are connected (in contact) to each other at one end. The contact point is the measuring location of the thermo­couple. The two free wire ends constitute the reference location of the thermocou­ple. If the thermocouple is heated in the measuring location, a thermal voltage is present in the reference location. This thermal voltage depends on the temper­ature difference between the mea suring location and the reference location (tem­perature gradient along the thermocou­ple), and on the material combination of the two metal wires. For temperature measurements, the temperature of the reference location must be kept con­stant at 0 °C, or it must be precisely mea­sured in order to perform an appropriate adjustment in mV. To evaluate the ther­mal voltage, the free wire ends of the thermocouple (the reference location) are connected to a suitable transmitter.
m The T probe is not identical to
thermocouple type T.
The standardized thermocouple types are not suitable for the T probe for rea­sons of glassing technology. The positive branch of the T probe is made of a plati­num-rhodium alloy (PtRh), the negative branch is made of a gold-palladium-plati­num alloy (AuPdPt). This material com­bination is known as “pallaplat”. Table 1 shows the values of the basic character­istic of pallaplat. However, these values may only serve as reference values. The exact characteristic depends on the manufacturing batch of the pallaplat. For programming the transmitter (refer to Section 5.3), the characteristic that was determined for the specific pallaplat batch after production will be used. The values entered during programming will be indicated in the test report that is attached to each probe supplied.
Only the Pfaudler-approved pallaplat compensating line (part no. 029034) may be used as an extension for the thermocouple.
Standard EN 60584 defines the various material combinations for the produc­tion of thermocouples. These standard­ized material combinations are referred to as “thermocouple types” and marked by different capital letters. One of them is the letter “T“.
4
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 4
Temperature probes type T/TW
Temperature probes type T/TW

5.2 Construction of the T probe

5.2.1 Probe carrier
The T probe can be installed in various probe carriers:
n Baffles, thermometer well, Quatro-
Pipe and C baffle (the term tubular probe will be used below for this type
of probe carriers)
n Ring
probe
n Valve shaft of outlet valves
5.2.2 Tubular probe
The thermocouple of tubular probes con­sists of very thin, narrow bands made of the pallaplat material combination. The pallaplat bands are fused into the glass layer.
The measuring location, where the two pallaplat bands are in contact is normally at the end of the tube in order to mea­sure the product temperature even at low liquid levels inside the reactor. When using tubular probes, several measuring locations may be fitted at the same or at different levels. The maximum possible number of measuring locations depends on the size of the tubular probe. For more details, please contact the Pfaudler Instrumentation department.
In order to provide electrical connection to the thermocouple, the pallaplat bands fused into the glass are routed close to the terminal box and connected to the compensating line. In turn, the compen­sating line is introduced into the terminal box where it is connected to the trans­mitter.
Tubular probes equipped with T probes can optionally be combined with the following measuring probes made by Pfaudler:
n Typ P Measuring probe for glass
monitoring
n Typ FT Measuring probe for capaci-
tive detection of filling limits or interfaces between liquids
n Typ FS Measuring probe continuous,
capacitive detection of filling levels
n Typ TW Measuring probe for tem-
perature measurement us­ing a resistance thermometer
For more details about possible com­binations, please contact the Pfaudler Instrumentation department.
5.2.3 Ring probe
The thermocouple of ring probes con­sists of very thin, narrow bands made of the pallaplat material combination. The pallaplat bands are fused into the glass layer.
The measuring location is in the middle of the cylindrical inside area. Ring probes can be produced with a maximum of 4 measuring locations.
In order to provide electrical connec­tion to the thermocouples, the pallaplat bands fused into the glass are routed close to the terminal box and connected to the compensating line. In turn, the compensating line is introduced into the terminal box where it is connected to the transmitter.
Ring probes equipped with T probes can optionally be combined with the following measuring probes made by Pfaudler:
n Typ P Measuring probe for glass
monitoring
n Typ FT Measuring probe for capaci-
tive detection of filling limits or interfaces between liquids
For more details about possible com­binations, please contact the Pfaudler Instrumentation department.
5.2.4 Valve shaft
The measuring location of a probe mounted on a valve shaft is in the mid­dle of the valve shaft on the product side. Power supply to the thermocouple is by means of pallaplat wires that are insu­lated and routed through the inside bore of the valve shaft. The wire ends are con­nected to the compensating line which in turn is introduced into the terminal box where it is connected to the transmitter type TTH300 (or terminals).
With nominal sizes DN 80/50 or more, the T probe can be installed in a valve shaf t. A maximum of 2 T measuring loca­tions are possible. Optionally, a T probe can be combined with the a measuring probe Type P (for glass monitoring). For more details about this possible com­bination, please contact the Pfaudler Instrumentation department.
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 5
5
Temperature probes type T/TW
Temperature probes type T/TW

5.3 Transmitter

A transmitter of the type TTH300 made by ABB is used as a standard for evaluat­ing the signal measured by the T probe (the thermal voltage). The transmitter is a freely programmable unit that converts the measured signals of thermocouples and of resistance thermo meters into a standard potential-free 4-20 mA signal. The transmitter is integrated into the ter­minal box on the probe carrier already in the factory where it is also connected to the sensor lines.
Based on product-dependent tolerances in the manufacture of pallaplat, the val­ues of the characteristic may slightly deviate between batches. Therefore, the batch-specific characteristic is measured after the production of each batch. Dur­ing the internal acceptance test in the factory, the TTH300 is parameterized with this batch-specific characteristic.
The characteristic has a total of 32 inter­mediate points. The values of the char­acteristic are indicated in the test report that is attached to each probe supplied. The transmitter is programmed using the “Smart Vision” firmware. The TTH300 is also equipped with a HART interface which allows for programming of the transmitter on location.
As a rule, it is possible to use other types or makes instead of the TTH300. How­ever, the following conditions must be observed in this case:
n The transmitter must be freely pro-
grammable and must be programmed with the batch-specific pallaplat char­acteristic. Third-party transmitters cannot be programmed by Pfaudler because we do not have the hardware and software necessary for that pur­pose. However, Pfaudler will support you with the programming of third­party transmitters.
n The transmitter must have a reference
location.
n Pfaudler cannot provide any binding
information concerning the accuracy of the measuring system as a whole if third-party transmitters are used.
If the ambient temperature in the sur­roundings of the transmitter is exces­sively high on location (refer to Sect. 5.4), the transmitter must be installed next to the probe carrier in an area in which the temperature is lower. A pallaplat com­pensation line must be used to connect the terminal box on the probe carrier to the transmitter. Suitable compensat­ing lines are available from Pfaudler (part no. 029034).
For more details concerning the trans­mitter, please refer to the documentation that is attached to each probe supplied.
6
m The T probe must in all cases
be operated in conjunction with a freely programmable transmitter, otherwise, the temperature mea­surement will not be correct.
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 6

5.4 Technical data of T probe

Sensor material: Pallaplat
pos. branch made of platinum-rhodium neg. branch made of gold-palladium-platinum
Resistance of pallaplat: 30 Ω /m
Resistance of the compensating line: 0,33 Ω /m
min/max temperature in the terminal box: – 40/+80 °C
min/max operating temperature: – 25/+200 °C
Measuring variance: max. ± 1,5 °C
Temperature probes type T/TW
Temperature probes type T/TW
Electrical data if used in potentially explosive atmospheres:
Please note the details in the type examination certificate PTB 03 ATEX 2132 X in annex 1.
© Pfaudler GmbH ·
OI 302-8 e
© Pfaudler GmbH 7
7
Temperature probes type T/TW
Temperature probes type T/TW

5.5 Installation of the probe carriers

5.5.1 Installation of tubular probe
The term tubular probe is used for the following probe carriers in this context:
n Baffles n Thermometer well n Quatro-Pipe n C baffles
Before installing a tubular probe in a reactor or a pipeline, you should verify whether there is sufficient distance to the agitator and the reactor wall. If nec­essary, suitable spacers or reducing flanges must be used.
Assembly process:
t Place flange gasket on nozzle. t Protect the nozzle and the probe
against damage by inserting a piece of cloth or a PTFE sleeve into the noz­zle. Slowly introduce the tubular probe into the reactor through the nozzle. Avoid pendulum motion.
t Tighten flange screws evenly cross-
wise with the prescribed tightening torque; (cf. Tab. 3).
m When using your own reduc-
ing flanges, please make sure that the element covering the contact area below the terminal box does not sit on the reducing flange (cf. Fig. 2). In the event of non-compli­ance, the glass may be damaged at the bottom side of the flange, and the fused-in metal strips may be interrupted.
5.5.2 Installation of ring probe
The ring probe may generally be installed in any position inside the pipeline. How­ever, it must be noted that the T probe does not work unless the measuring loca­tion is sufficiently covered with product. In radial direction, the measuring location is in the same place as the terminal box.
The ring probe is be installed in the pipe­line between 2 flanges using gaskets.
Tighten flange screws evenly crosswise with the prescribed tightening torque; (cf. Tab. 3).
5.5.3 Installation of outlet valve
For details concerning the installation, please refer to our Operating Instruc­tions 322.
These Operating Instructions are attached to each valve supplied. They are available from Pfaudler on request.
collision area
MT0010_1E
Fig. 2 Installation example with reducing flange
Table 3 Tightening torques of glass-lined flange connections
Flange Screws Max. tightening torques in Nm with admis-
DN50 4 x M16 30 30
DN80 8 x M16 35 35
DN100 8 x M16 35 35
DN150 8 x M20 40 40
DN200 8 x M20 55
sible operating pressures of:
–1 to +10 bar –1 to +16 bar
8
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 8
Temperature probes type T/TW
Temperature probes type T/TW
+ red
– violett
T probe
Fi
g. 3 Connection diagram of the T probe
3
1
power supply unit with electrical isolation Ex ia/ib max. 30 V DC 100 mA
optional floating change-over contacts
}
+5
6
current output 4 -2 0 m A
power Supply 24 V DC/AC (230 V AC)
MT_05_003_1e

5.6 Connection of the T probe

The following information only applies to T probes in c
onnection with the TH02-Ex
transmitter.
A power supply unit with the following specifications is necessary for power supply:
n Supply current: I n Supply voltage: U n in potentially explosive
= 0 - 20 mA
S
= 8,5 ... 30 V DC
S
atmospheres: Ui = 8,5 ... 29,4 V DC
For applications in potentially explosive atmospheres, power supply units in intrinsically safe design must be used.
The power supply unit must be installed outside the potentially explosive area. The installation instructions of the manu­facturer are mandatory in this respect.
The connection between the power sup­ply unit and the transmitter can be made using standard signal cables.
The probe carriers must be grounded using copper or stainless steel wires with a minimum conductor area of 10 mm2.
The electrical connection is made in compliance with the connection diagram shown in Fig. 3. Connection is identical for all probe carriers.

5.7 Start-up and maintenance of the T probe

No start-up procedure is necessary for T probes supplied together with the stand­ard transmitter. When the transmitter has been connected to the power supply unit, the T probe is ready for use. How­ever, we recommend verifying whether the transmitter’s measuring range as set by Pfaudler complies with your require­ments. The measuring range setting is indicated on a rating plate each on the transmitter and in the terminal box as well as in the test report.
If necessary, the measuring range can be changed. Upon final acceptance in the factory, a functional test and a sin­gle-point calibration will be performed. A test report will be prepared which is attached to each probe.
If the T probe is ordered and delivered without a transmitter, it is the operator’s responsibility to ensure that a suitable transmitter is used and configured (refer to Sect. 5.3). The values comprised in the characteristic of the thermocouple are provided in the attached test report.
Due to its special construction (thermo­couple fused into glass layer) and non­critical operating conditions, the probes are not subject to ageing. Therefore, inspections (calibration) of the thermo­couple are not necessary for measuring reasons. As a result, the T probe is main­tenance-free.
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 9
9
Temperature probes type T/TW
Temperature probes type T/TW
If it is used for processes, however, which are associated with the risk of product deposits, it must be verified whether or not product has accumulated on the probe carrier. Deposits must be removed using suitable means and methods.

5.8 Calibration of the T probe

5.8.1 General instructions
As mentioned before, the thermocouple has been fused into the glass layer, thus forming an integral part of the probe car­rier. Calibrating the thermocouple sepa­rately is not possible for this reason. It is most practical to calibrate the T probe with the probe carrier installed because stable, reproducible temperature con­ditions at the T probe are most conve­niently achieved. Calibration outside the reactor or pipeline is very problematic because there is considerable heat dissi­pation, which results in instable temper­ature conditions at the thermocouple, due to the shape and size of the probe carriers. For calibration, the reactor or the pipeline must be filled either with product or with water. A steady tempera­ture at the transmitter must be ensured during calibration. For this reason, the terminal box should be closed.
For calibrating tubular probes, the reac­tor must be filled until the immersion depth of the measuring location is at least five times the tube diameter. When calibrating inside a reactor, the agitator must turn slowly in order to ensure an even temperature distribution inside the liquid.
For ring probes, the pipeline must be fully filled. Ring probes may also be cali­brated outside the pipeline due to their compact dimensions. For this purpose, the probe must then be immersed into a suitable vessel. Caution! Do not immerse the terminal box.
When calibrating outlet valves, the valve cone must be sufficiently covered with product so as to ensure a stable tem­perature condition in the measuring location. Furthermore, the agitator must be slowly turned during calibration.
5.8.2 Calibrating the measuring chain
In order to calibrate the entire measuring chain including the sensor, transmitter and display unit, immerse a calibrated reference thermometer into the liquid and compare its measured value (refer­ence value) with the value measured by the T probe (actual value). If it is neces­sary to calibrate the T probe at different temperatures, we recommend calibrat­ing it in icy water at 0 °C and in boiling water at 100 °C.
The result will be reliable information concerning the accuracy of the entire measuring chain and the actual tempera­ture conditions in the subsequent pro­duction process.
If an excessive, inadmissible measuring deviation is detected during calibration, the individual elements of the measuring chain must be verified or calibrated.
5.8.3 Calibrating the sensor
Actually, it is not necessary to calibrate the sensor (thermocouple) for technical reasons because the sensor is not sub­ject to ageing (refer to Sect. 5.7).
However, in some cases the sensor has to be calibrated to comply with internal or process-related requirements. The Pfaudler T-Calibrator (part no. 256572-) and the Smart Vision firmware must be available in order to calibrate the ther­mocouple. For more information and a detailed description of the calibration process, please refer to the Operating Instructions of the T-Calibrator.
5.8.4 Calibrating the transmitter
In order to calibrate the transmitter, a process calibrator is connected to the transmitter input that serves to simulate the thermocouple‘s characteristic. Then, the voltage values of the characteristic are entered one by one at the process calibrator. In parallel, the actual values of the transmitter (input and output sig­nal) are compared to the programmed values (values of the characteristic). The values of the characteristic are indicated in the test report that is attached to each probe supplied. It is important to deac­tivate the internal reference point of the transmitter during calibration. Please refer to the Operating Instructions of the TH02-Ex unit for a description of how to deactivate the reference point.
5.8.5 Calibrating the display unit
In order to calibrate the display unit, an additional, calibrated display unit is looped into the transmitter’s output circuit, and the two values displayed as actual and programmed values must be compared.

5.9 Explosion protection

The tubular and ring probes with an integrated T probe have been approved for use in potentially hazardous atmos­pheres of zone 0 in accordance with the EC type examination certificate No. PTB 03 ATEX 2132 X. For outlet valves with an integrated T probe, an EC type exami­nation certificate has been applied for.
10
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 10
Temperature probes type T/TW
Temperature probes type T/TW

6 Temperature probe type TW The resistance values of resistance ther-

mometers made of platinum have been

6.1 Measuring principle

The temperature probe Type TW w also be referred as the TW probe in the present Operating Instructions.
standardized and are referred to as Pt 100, Pt 500 and PT 1000. The number speci-
ill
fies the nominal resistance in Ohm at a temperature of 0 °C. Therefore, a Pt 100
has a nominal resistance of R0 = 100 Ω . As already mentioned, a resistance ther­mometer serves as the sensor of the TW probe. The measuring principle of resist­ance thermometers is based on the fact that the electrical resistance of all con­ductors and semiconductors changes
For these standardized measuring resist-
ances, basic value series (characteris-
tics) have been specified in EN 60 751.
The measured signal is evaluated by the
transmitter based on this characteristic
in order to determine the temperature. with the temperature. The amount of change in referred to as varies depending on the material used. The relative change of the electrical resistance as a function of the temperature, R/∆t, is referred to as the temperature coefficient. This value is not constant over the whole tempera­ture range. Rather, it is a function of the temperature. The relationship between resistance and temperature is subject to higher-order mathematical polynomi­als for calculating the characteristic of a
For the TW probe, no Pt 100 to the EN
standard was used as resistance ther-
mometer. For technical reasons, the
TW probe cannot be produced with an
exact resistance value of 100 Ω . The re-
sistance of the TW probe is 100 ± 3 Ω .
Therefore, it is absolutely necessary to
use a programmable transmitter (refer
to Sect. 6.3) that compensates the de-
viation from the 100 Ω characteristic in
order to evaluate the measured signal. resistance thermometer.
The special properties of platinum, such as high measuring accuracy, high tem­perature resistance, chemical resistance, an approximately linear characteristic as well as an excellent reproducibility of the thermo-electrical properties, were reason enough to produce most resistance ther­mometers from this metal today. Since platinum can also be easily glassed, it is also used for the TW probe.

6.2 Construction of the TW probe

The TW probe can be installed in the following probe carriers:
n Paddle-type baffles n Thermometer well n Quatro-Pipe
It is not possible to install the TW probe in a ring probe, a C baffle or a valve shaft.
The resistance thermometer consists of a very thin platinum band that is spirally wound around the end of the probe car­rier tube. The platinum band is fused into the glass layer. The maximum possible number of measuring locations on a tubular probe depends on the size of the probe carrier. For more details, please contact the Pfaudler Instrumentation department. The measuring locations are always at the same level. Measuring loca­tions at different levels can only be pro­vided using the T probe (refer to Sect. 5)
The supply wires of the resistance ther­mometers are also made of platinum. The supply wires fused into the glass are routed close to the terminal box where they are connected to terminal wires. In turn, the terminal wires are introduced into the terminal box where they are con­nected to the transmitter type TH02-Ex (or terminals).
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 11
11
Temperature probes type T/TW
Temperature probes type T/TW
The TW probe is designed in four-wire technology (2 parallel supply bands each). Therefore, TW probes can be con­nected in a four-wire circuit. With this measuring method, a constant measur­ing current flows through 2 wires, where­as the temperature-dependent voltage drop at the measuring resistor is mea­sured via the remaining 2 wires. In a four­wire circuit, the influence of the supply conductor resistance on the measured result is totally neutralized.
Probe carriers equipped with TW probes can optionally be combined with the following measuring probes made by Pfaudler:
n Typ P Measuring probe for glass
monitoring
n Typ FT Measuring probe for capaci-
tive detection of filling limits or interfaces between liquids
n Typ FS Measuring probe continuous,
capacitive detection of filling levels
n Typ T Measuring probe for tem-
perature measurement using a thermocouple
For more details about possible com­binations, please contact the Pfaudler Instrumentation department.

6.3 Transmitter TTH300

A transmitter of the type TTH300 made
by ABB is used as a standard for evaluat-
ing the signal measured by the TW probe.
The transmitter is a freely programmable
unit that converts the measured signals
of resistance thermometers and thermo-
couples into a standard potential-free
4-20 mA signal. The transmitter is inte-
grated into the terminal box already in
the factory where it is also connected to
the sensor lines.
As already mentioned in Sect. 6.1, the
resistance thermometer for the TW probe
cannot be produced to have a resistance
of exactly 100 Ω at 0 °C. For this reason,
the exact resistance in Ω of the resist-
ance thermometer is measured upon
completion of the TW probe. During the
internal acceptance test in the factory,
the transmitter is parameterized with the
characteristic of this Ω value. The char-
acteristic has a total of 32 intermediate
points. The values of the characteristic
are indicated in the test report that is
attached to each probe supplied.
The transmitter is programmed using the
Smart Vision” firmware. The TTH300
is also equipped with a HART interface
which allows for programming of the
transmitter on location.
For more details concerning the trans-
mitter, please refer to the documentation
that is attached to each probe supplied.
As a rule, it is possible to use other types or makes instead of the TTH300. How­ever, the following conditions must be observed in this case:
n The transmitter must be freely pro-
grammable and must be programmed with the probe-specific characteristic. Third-party transmitters cannot be programmed by Pfaudler because we do not have the hardware and software necessary for that purpose. However, Pfaudler will support you with the programming of third-party transmitters.
n Pfaudler cannot provide any binding
information concerning the accuracy of the measuring system as a whole in this case.
If the ambient temperature in the sur­roundings of the terminal box is exces­sively high on location (refer to Sect. 6.4), the transmitter must be installed next to the probe carrier in an area in which the ambient temperature is lower. The 4-wire signal lead may be used to connect the terminal box on the probe carrier to the transmitter.
m The TW probe must in all
cases be operated in conjunction with a programmable transmitter, otherwise, the temperature measure­ment will not be correct.
Table 3 Tightening torques of glass-lined flange connections
Reactor type AE AE/BE/CE E BE/CE
Nominal reactor size 63
baffle with stuffing box 1 2 2 2 2
baffle/thermowell with flange
Quatro Pipe 2 2 2 2
to
400
2 2 2 2
630
to
1000
1600
to
6300
1200
to
20000
8000
40000
12
to
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 12

6.4 Technical data of TW probe

Sensor material: Platin
Resistance of the sensor material: ca. 40 Ω/m
Resistance of the feed conductors: ca. 10 Ω/m
min/max temperature in the terminal box: – 40/+80 °C
min/max operating temperature: – 25/+200 °C
Measuring variance: max. ± 1 °C
Electrical data if used in potentially explosive atmospheres::
Please note the details in the type examination certificate PTB 03 ATEX 2132 X in the appendix
Temperature probes type T/TW
Temperature probes type T/TW
© Pfaudler GmbH ·
OI 302-8 e
© Pfaudler GmbH 13
13
6
+5
MT_05_0004_1d TW-Sonde.fh11
Temperature probes type T/TW
Temperature probes type T/TW
Pt
100
TW-Sonde
Fig. 4 Connection diagram of the TW probe
white red
green black
1 4
2 3
power supply unit with electrical isolation Ex ia/ib max. 30 V DC 100 mA
-
optional floating change-over contacts
}
current output 4 - 20 m A
power supply 24 V DC/AC (230 V AC)
MT_05_0004_1e

6.5 Installation of the TW probe

As already mentioned, the TW probe can be installed in the following probe car­riers: Paddle-type baffles, thermometer wells, Quatro Pipe These probe carriers are generally also referred to as tubular probes.
Before installing a tubular probe in a reactor or a pipeline, you should verify whether there is sufficient distance to the agitator and the reactor or pipeline wall. If necessary, suitable spacers or reducing flanges must be used.
Assembly process:
t Place flange gasket on nozzle. t Protect the nozzle and the probe
against damage by inserting a piece of cloth or a PTFE sleeve into the noz­zle. Slowly introduce the tubular probe into the reactor through the nozzle. Avoid pendulum motion.
t Tighten flange screws evenly cross-
wise with the prescribed tightening torque; (cf. Tab. 3).
m When using your own reduc-
ing flanges, please make sure that
the element covering the contact
area below the terminal box does
not sit on the reducing flange (cf.
Fig. 2). In the event of non-compli-
ance, the glass may be damaged at
the bottom side of the flange, and
the fused-in metal strips may be
interrupted.

6.6 Connection of the TW probe

The following information only applies
to TW probes in connection with the
transmitter.
A supply unit with the following specifica-
tions is necessary for power supply:
n Supply current: I
n Supply voltage: U
n in potentially explosive
atmospheres: Ui = 8,5 .. . 29,4 V DC
For applications in potentially explosive
atmospheres, supply units in intrinsically
safe design must be used.
= 0 - 20 mA
S
= 8,5 ... 30 V DC
S
The power supply unit must be installed outside the potentially explosive area. The installation instructions of the manu­facturer are mandatory in this respect.
The connection between the power sup­ply unit and the transmitter can be made using standard signal cables.
The probe carriers must be grounded using copper or stainless steel wires with a minimum conductor area of 10 mm2.
The electrical connection is made in compliance with the connection diagram shown in Fig. 4.
14
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 14
Temperature probes type T/TW
Temperature probes type T/TW
6.7 Start-up and maintenance of
the TW probe
No start-up procedure is necessary for TW probes supplied together with the standard transmitter. When the electri­cal connection of the transmitter to the power supply unit has been made, the TW probe is ready for use. However, we recommend verifying whether the transmitter’s measuring range as set by Pfaudler complies with your require­ments. The measuring range setting is indicated on a rating place each on the transmitter and in the terminal box as well as in the test report. If necessary, the measuring range can be changed.
Upon final acceptance in the factory, a functional test and a single-point calibra­ti
on will be performed. A test report will be prepared which is attached to each probe.
If the TW probe is ordered and delivered without a transmitter, it is the operator’s responsibility to ensure that a suitable transmitter is used and configured (refer to Sect. 5.3). The values comprised in the characteristic of the resistance ther­mometer are provided in the attached test report.
Due to its special construction (resist­ance thermometer fused into glass layer) and non-critical operating conditions, the probes are not subject to ageing. There­fore, inspections (calibration) of the resistance are not necessary for measur­ing reasons. As a result, the TW probe is maintenance-free.
If it is used for processes, however, which are associated with the risk of product deposits, it must be verified whether or not product has accumulated on the probe carrier. Deposits must be removed using suitable means and methods.

6.8 Calibrating the TW probe

6.8.1 General instructions
As me ntioned before, the resistance ther­mometer has been fused into the glass layer, thus forming an integral part of the probe carrier. Calibrating the resistance thermometer separately is not possible for this reason. It is most practical to calibrate the TW probe with the probe carrier installed because stable, repro­ducible temperature conditions at the T probe are most conveniently achieved. Calibration outside the reactor or pipe­line is very problematic because there is considerable heat dissipation, which results in instable temperature condi­tions at the resistance thermometer, due to the shape and size of the probe car­riers. For calibration, the reactor or the pipeline must be filled either with prod­uct or with water. A steady temperature at the transmitter must be ensured dur­ing calibration. For this reason, the termi­nal box should be closed.
For calibrating tubular probes, the reac­tor must be filled until the immersion depth of the measuring location is at least five times the tube diameter. When calibrating inside a reactor, the agitator must turn slowly in order to ensure an even temperature distribution inside the liquid.
6.8.2 Calibrating the measuring chain
In order to calibrate the entire measuring chain including the sensor, transmitter and display unit, immerse a calibrated reference thermometer into the liquid and compare its measured value (ref­erence value) with the value measured by the TW probe (actual value). If it is necessary to calibrate the TW probe at different temperatures, we recommend calibrating it in icy water at 0 °C and in boiling water at 100 °C.
The result will be reliable information concerning the accuracy of the entire measuring chain and the actual tempera­ture conditions in the subsequent pro­duction process.
If an excessive, inadmissible measuring deviation is detected during calibration, the individual elements of the measuring chain must be verified or calibrated.
6.8.3 Calibrating the sensor
Actually, it is not necessary to calibrate the sensor (resistance thermometer) for technical reasons because the sensor is not subject to ageing (refer to Sect. 5.7).
However, in some cases the sensor has to be calibrated to comply with internal or process-related requirements. For calibrating the probe, a process calibra­tor capable of measuring the resistance of the resistance thermometer and the product temperature in a four-wire cir­cuit must be used. The measured values (actual values) must then be compared to the programmed values (values taken from the characteristic).
6.8.4 Calibrating the transmitter
To calibrate the transmitter, an adjusta­ble precision resistor must be connected to the transmitter input. Then, the resist­ance values from the characteristic are input one by one, and the related temper­ature (programmed value) is compared to the output value of the transmitter (actual value). The values of the charac­teristic are indicated in the test report that is attached to each probe supplied.
6.8.5 Calibrating the display unit
In order to calibrate the display unit, an additional, calibrated display unit is looped into the transmitter’s output circuit, and the two values displayed as actual and programmed values must be compared.

6.9 Explosion protection

The tubular with an integrated TW probe have been approved for use in potentially hazardous atmospheres of zone 0 in accordance with the EC type examination certificate No. PTB 03 ATEX 2132 X.
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 15
15
Annex 1 PTB 03 ATEX 2132 X
Annex 1
PTB 03 ATEX 2132 X
16
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 16
Annex 1
Annex 1
PTB 03 ATEX 2132 X
PTB 03 ATEX 2132 X
© Pfaudler GmbH ·
OI 302-8 e
© Pfaudler GmbH 17
17
Annex 1 PTB 03 ATEX 2132 X
Annex 1
PTB 03 ATEX 2132 X
18
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 18
Annex 1
Annex 1
PTB 03 ATEX 2132 X
PTB 03 ATEX 2132 X
© Pfaudler GmbH ·
OI 302-8 e
© Pfaudler GmbH 19
19
Annex 2
Annex 2
Declaration of conformity
Declaration of conformity
20
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 20
nex 2
An
Declaration of conformity
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 21
21
Annex 3 Safety instructions and explosion protection
A 3.1 Potentially hazardous atmospheres
Please note the details in the type examination certificate PTB 03 ATEX 2132 X, Issue: 1 in the annex.
A 3.2 Atmospheric conditions
Please note the details in the type examination certificate PTB 03 ATEX 2132 X, Issue: 1 in the annex.
A 3.3 Equipotential bonding
Please note the details in the type examination certificate PTB 03 ATEX 2132 X, Issue: 1 in the annex.
A 3.4 Lightning protection
If the temperature probe is installed in systems that must be protected against ignition hazards due to lightning, the redox probe must be included in the lightning protection. The lightning protection must comply with the requirements of VDE 0165.
A 3.5 Quatro Pipe
When using a Quatro Pipe with T / TW probe, please observe the information in the type examination certificate
PTB 03 ATEX 2132 X,
Issue: 1.
Annex 3
22
© Pfaudler GmbH · OI 302-8 e
© Pfaudler GmbH 22
Notes
Notes
© Pfaudler GmbH
23
The information provided in this documentation corresponds
to the state of the art at the time of printing. It is published
in good faith. However, we will accept no warranty claims
based on the information provided in this documentation. We
reserve the right to include improvements, amendments and
new findings in this documentation without prior notice. The
actual design of products may deviate from the information
contained in the calatoge if technical alterations and product
improvements so require. The proposal made by Pfaudler for
a concrete application will be binding in such cases.
The present documentation is made available free of charge
to our customers and other interested parties. The right to
print or copy this documentation, or any parts there of, or to
convert the same into electronic form shall be subject to our
written permission.
All rights reserved by us.
Pfaudler GmbH
P.O. Box 1780 D-68721 Schwetzingen
Pfaudlerstraße D-68723 Schwetzingen
Phone +49 6202 85-233
Telefax +49 6202 85-273
E-mail info@pfaudler-instrumentation.com
www.pfaudler-instrumentation.com
Loading...