11
PRINTOUT DESCRIPTION
1. Start time refers to the clock time when the RUN mode was initiated.
2. End time refers to the clock time when the last RUN was halted by
either turning the unit OFF or PRINTING.
3. Total run time is the difference between the start and end times. If
there were multiple runs an asterisk will appear after the total run
time indicating that there is a break in the run time data.
4. The current alarm level setting is printed.
5. The current sample rate indicates the amount of time between each set
of data printed.
6. The highest ear temperature logged is printed along with its time of
occurrence.
7. The lowest ear temperature logged is printed along with its time of
occurrence.
8. The calibration value is that of 'ORAL minus EAR' at the time the
calibration sequence was completed. If the unit is calibrated more
than once, each additional value and time will be printed.
9. Ear temperature data (including the calibration offset if the unit was
calibrated) is printed in both Celsius and Fahrenheit.
10. Earmold temperature data is printed in both Celsius and Fahrenheit.
This refers to the second sensor located in the blue earmold and is an
'indicator' of what was occurring in the environment. The values are
not exact ambient temperatures because they may be affected by the
close proximity to the head.
11. If the oral thermometer probe is in the CAL jack and its temperature
is in the range of 32°C to 42°C then the data will be recorded.
Typically, this would provide calibration information of the oral
temperature.
12. Asterisks appear in the alarm column next to the data if alarm had
gone off during that particular time period. Multiple asterisks
appear for each tenth of a degree Celsius above the set alarm level.
(For example: when the alarm is set to 38.0°C, 1 asterisk indicates
38.0°C, 2 asterisks indicates 38.1°C, etc.)
LOW BATTERY
The battery condition should be checked each time before using the
QUESTEMP°II. Pressing the BAT/TEST key will display the actual battery
voltage. When the level falls below 7.5 volts, "Lo" appears in the display
indicating that the battery should be replaced. The unit will continue to
function until the battery falls below 7.2 volts. At that time, the alarm
will sound continuously and the unit will no longer read or store data. If
this occurs, turn the unit OFF and replace the battery before its next use.
Under normal use, the QUESTEMP°II will function for an addition 8 hours
from the time when the battery voltage first hits its "Lo" condition at 7.5
volts.
An on-board 3 volt lithium battery will retain the memory of the
QUESTEMP°II even when the 9 volt battery is removed. This will not,
however, power the unit for operation.
Replace dead batteries with 9 volt ALKALINE, type NEDA 1604A or equivalent.
12
AUDIBLE ALARM
The speaker located in the ear assembly will sound under three conditions.
1. When the 10-second-average temperature in the ear canal exceeds
the set alarm level, the speaker will give two pulses every
second.
2. When the 10-second-average temperature in the ear canal exceeds
0.5°C above the set alarm level, the speaker will give four
pulses every second.
3. When the battery voltage falls below 7.2 volts, the speaker will
sound continuously warning that the unit has stopped reading
temperature data and the battery should be immediately replaced.
III. CALIBRATION
The purpose of calibration is to compensate for the difference between an
individual's ear and body temperatures. The QUESTEMPNII accomplishes this
by taking an oral temperature with the QUESTEMP° CALPROBE I thermometer and
comparing it to the ear temperature. [When taking an oral temperature, the
person should be rested and should not have had anything to eat or drink in
the past 15 minutes. The probe should be kept under the tongue with the
mouth closed.] Although typical oral to ear temperature differentials can
be found over a several degree range, each individual's differential will
remain stable provided that the ambient temperature does not change
drastically from the time of calibration.
At the time of calibration, an offset is added to the ear temperature to
equate it to the oral temperature. This offset will remain in effect until
modified by the next calibration or until the unit is reset. The time and
amount of offset will be listed on the printout.
The calibration procedure should be followed before each use. After the
initial calibration, if the worker's environment changes by more than 10°C,
then the QUESTEMP°II should be recalibrated in the new environment.
Calibration can be done at any time and will not disturb the data logging
in the RUN mode.
!Use the following sequence to calibrate the QUESTEMP°II using the
QUESTEMP° CALPROBE I thermometer.
1. Insert the ear sensor and allow 3 - 5 minutes for it to stabilize
to the correct temperature.
2. Turn the unit ON; RESET it to clear old data if desired; and
press RUN to record calibration data.
3. Place the CALPROBE under the tongue and plug it into the CAL jack
on the unit. Allow 3 - 5 minutes for it to stabilize. (A total
of 6 - 10 minutes for the ear sensor.)
4. Pressing CAL once will display (and update) the oral temperature.