This manual uses the following conventions:
“User” or “Operator” refers to anyone using a VB14N.
“Device” refers to the VB14N.
“You” refers to the System Administrator or Technical Support person using this manual to install,
mount, operate, maintain or troubleshoot a VB14N.
REFERENCE DOCUMENTATION
The documentation related to the VB14N management is listed below:
CBX100 Installation Manual
CBX500 Installation Manual
CBX Accessory Manuals
Genius™ Help On Line
v
VB14N
SAFETY AND COMPLIANCE NOTICES
LASER SAFETY
The following information is provided to comply with the rules imposed by international authorities and
refers to the correct use of the VB14N scanner.
Standard Regulations
This scanner utilizes a low-power laser diode. Although staring directly at the laser beam momentarily
causes no known biological damage, avoid staring at the beam as one would with any very strong light
source, such as the sun. Avoid that the laser beam hits the eye of an observer, even through reflective
surfaces such as mirrors, etc.
This product conforms to the applicable requirements of IEC 60825-1 and com plies with 21 CFR
1040.10 except for deviations pursuant to Laser Notice N° 50, date June 24, 2007. The scanner is
classified as a Class 2 laser product according to IEC 60825-1 regulations.
There is a safety device, which allows the laser to be switched on only if the motor is rotating above
the threshold for its correct scanning speed.
The laser beam can be switched off through a software command (see also the Genius™ Help On
Line).
Use of controls or adjustments or performance of procedures other than those
specified herein may result in exposure to hazardous visible la ser light.
WARNING
The laser light is visible to the human eye and is emitted from the window on the front of the scanner
Figure A, 7).
(
Warning labels indicating exposure to laser light and the device classification are applied onto the
body of the scanner (
Figure A, 1).
vi
VB14N
Disconnect the power supply when opening the
device during maintenance or installation to avoid
exposure to hazardous laser light.
The laser diode used in this device is classified as
a class 3B laser product according to EN 60825-1
regulations and as a Class IIIb laser product
according to CDRH regulations.
Any violation of the optic parts in particular can
cause radiation up to the maximum level of the
laser diode (35 mW at 630 to 680 nm).
AVOID EXPOSURE LASER LIGHT
IS EMITTED FROM THIS APERTURE
Warning and Device Class Labels
FCC COMPLIANCE
Modifications or changes to this equipment without the expressed written approval of Pepperl+Fuchs
could void the authority to use the equipment.
This device complies with PART 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must accept any
interference received, including interference which may cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against harmful interference when the equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case
the user will be required to correct the interference at his own expense.
POWER SUPPLY
This product is intended to be installed by Qualified Personnel only.
This accessory device is intended to be supplied by a UL Listed or CSA Certified Power Unit with
«Class 2» or LPS power source, which supplies power directly to the scanner via the 25-pin
connector.
CE COMPLIANCE
Warning:
This is a Class A product. In a domestic environment this product may cause radio interference in
which case the user may be required to take adequate measures.
vii
VB14N
HANDLING
The VB14N is designed to be used in an industrial environment and is built to withstand vibration and
shock when correctly installed, however it is also a precision product and therefore before and durin g
installation it must be handled correctly to avoid damage.
avoid that the scanners hit one another causing damage. They should be handled separately.
avoid that the scanners are dropped (exceeding shock limits).
do not fine tune the positioning by striking the scanner or bracket.
viii
VB14N
do not weld the scanner into position which can cause electrostatic, heat or output window
damage.
do not spray paint near the scanner which can cause output window damage.
ix
VB14N
GENERAL VIEW
VB14N
4
5
6
2
1
2
3
4
1
Warning and Device Class Labels
"POWER ON" LED
Mounting Holes
"READY" LED
3
11
Figure A
Laser Beam Output Window
7
"COM" LED
8
"STATUS" LED
9
Push Button
10
10
9
8
7
"GOOD" LED
5
"TRIGGER" LED
6
Accessory Mounting Holes
11
x
VB14N
1 RAPID CONFIGURATION
This chapter illustrates a Stand Alone application. For other types of installations, such
as ID-NET™, Fieldbus, Pass-Through, Multiplexer Layout, etc., refer to chapters 4, 5
and 6. For complete scanner configuration using the Genius™ configuration program,
NOTE
refer to the Context-Sen
STEP 1 – CONNECT THE SYSTEM
To connect the system in a Stand Alone configuration, you need the hardware indicated in Figure 1.
layout the data is trans
In this
In Local Echo communication mode, data is transmitted on the RS232 auxiliary interface
independently from the main interface selection.
When On-Line Operating mode is used, the scanner is activated by an External Trigger (photoelectric
sensor) when the object enters its reading zone.
sitive Help On-Line.
mitted to the Host on the main serial interface.
24V DC
MAIN
VB14N
P.S.*
Figure 1 – VB14N in Stand Alone Layout
CBX100/500
I/O, AUX
Host
* Presence Sensor
(for On-Line mode)
1
VB14N
CBX100/500 Pinout for VB14N
The table below gives the pinout of the CBX100/500 terminal block connectors. Use this pinout when
the VB14N reader is connected by means of the CBX100/500:
CBX100/500 Terminal Block Connectors
Input Power Outputs
Vdc Power Supply Input Voltage + +V Power Source - Outputs
GND Power Supply Input Voltage - -V Power Reference - Outputs
Earth Protection Earth Ground O1+ Output 1 +
O1- Output 1 -
Inputs
+V Power Source – External Trigger O2- Output 2 I1A External Trigger A (polarity insensitive)
I1B External Trigger B (polarity insensitive)TX Auxiliary Interface TX
-V Powe r R e f er e n ce – Ex t e rn a l Tr i g g er RX Auxiliary Interface RX
+V Power Source – Inputs SGND Auxiliary Interface Reference
I2A Input 2 A (polarity insensitive)
I2B Input 2 B (polarity insensitive) REF Network Reference
* Do not leave floating, see par. 4.2.2 for connection details.
O2+ Output 2 +
Auxiliary Interface
ID-NET™
ID- ID-NET™ network -
CAUTION
2
Do not connect GND, SGND and REF to different (external) ground references.
GND, SGND and REF are internally connected through filtering circuitry which can
be permanently damaged if subjected to voltage drops over 0.8 Vdc.
VB14N
25-pin Connector Pinout for VB14N
The table below gives the pinout of the 25-pin male D-sub connector for connection to the power
supply and input/output signals. Use this pinout when the VB14N reader is connected by means of the
25-pin connector:
1
Figure 2 - 25-pin Male D-sub Connector
25-pin D-sub male connector pinout
Pin Name Function
13, 9 Vdc Power supply input voltage +
25, 7 GND Power supply input voltage -
1 CHASSIS Cable shield connected to chassis
18 I1 A External Trigger A (polari ty i nsensitive)
19 I1B External Trigger B (polarity insensitive)
6 I2A Input 2 A (polarity i nsensiti ve)
10 I2B Input 2 B (polarity insensitive)
* Do not leave floating, see par. 5.2.2 for connection details.
13
2514
RS485
Full-Duplex
RS485
Half-Duplex
3
VB14N
STEP 2 – MOUNTING AND POSITIONING THE SYSTEM
1. To mount the VB14N, use the mounting bracket to obtain the most suitable position for the reader
as shown in the figures below.
Skew
Tilt
Pitch
Skew
Figure 3 - Positioning with Mounting Bracket
2. When mounting the VB14N take into consideration these th ree ideal label po sition an gles: Skew 10° to
30°, Tilt 0° and Pitch 0°.
S
T
Assure at least 10° Minimize
Figure 4 –Skew and Tilt Angles
Minimize
Figure 5 – Pitch Angle
3. Refer to the Reading Diagrams in par. 7.4 to decide the distance your scanner should be
positioned at.
4
VB14N
STEP 3 – MODE BUTTON CONFIGURATION
The Mode Button is the intuitive Human Machine Interface designed to improve ease of installation
and maintenance.
Status and diagnostic information are clearly presented by means of the five colored LEDs, whereas
the single push button gives immediate access to the following relevant functions:
AutoSetup to self-optimize and auto-configure reading
performance in demanding applications
AutoLearn to self-detect and auto-configure for
reading unknown barcodes (by type and length)
Test Mode with bar-graph visualization to check static
reading performance
The colors and meaning of the five LEDs are illustrated in the following table:
READY (green) This LED indicates the device is ready to operate.
GOOD (green) This LED confirms successful reading.
TRIGGER (yellow) This LED indicates the status of the reading phase.
COM (yellow) This LED indicates active communication on main serial port. *
STATUS (red) This LED indicates a NO READ result.
* When connected to a Fieldbus network through the CBX500, the COM LED is al ways active, even in the
absence of data transmission, because of polling activity on the Fieldbus network.
During the reader startup (reset or restart phase), all the LEDs blink for one second.
On the back of the reader near the cable, the “POWER ON” LED indicates the laser scanner is
correctly powered.
5
VB14N
Auto Learn
If you are configuring your scanner using the Mode Button, you must start with the Auto Learn
procedure.
1. Enter the Auto Learn function by holding the Mode push button pressed until the LEARN LED is
on.
2. Release the button to enter the Auto Learn function.
Once entered, the reader starts a procedure to automatically detect and recognize barcodes (by
type and length), which are presented to it (*). The laser turns on and the LEARN LED blinks to
indicate the ongoing process.
The procedure is as follows:
A) place the desired barcode on the
scanline.
B) wait until the LEARN LED stays
steady on (indicating the reader
has detected the barcode).
C) repeat, if needed, the above two
steps to program up to 10 different
barcodes (the LEARN LED returns
to the blinking state for the next
code). If more than one barcode is
detected in the scan line, the Multi
Label mode is enabled (refer to the
Figure 6 – Auto Learn Function
3. Exit the process by pressing the Mode push button once. The scanner will restart at the end of the
process, and then the detected barcodes are automatically configured in scanner memory.
If the barcode cannot be read because of low contrast or excessive ambient light,
NOTE
NOTE
* In case of Programming Barcodes (refer to the “ID-NET™: Programming Barcodes And Setup Procedure”
document).
you can perform the AutoSetup function to optimize the optical parameters. Then
you can perform AutoLearn to recognize the barcode symbology.
On exit from Autolearn, the following parameters are forced: Code Combination =
Single Label, Reading Mode = Linear. If necessary, these parameters can be
changed through Genius™.
“Software Configuration Parameter
Guide” Help file).
6
VB14N
Auto Setup (Optional)
At the end of the Auto Learn procedure, you have the possibility to follow the Auto Setup procedure to
set up the reading parameters.
1. Enter the Auto Setup function by holding the Mode push button pressed until the SETUP LED is
on.
2. Release the button to enter the Auto Setup function.
3. Once entered, if a barcode label is positioned in front of the scanline, the scanner automatically
performs the optimal setup of the reading parameters for that specific barcode.
The procedure is as follows:
A) place the desired barcode on the
scanline.
B) enter the AutoSetup function (the
laser turns on and the SETUP LED
blinks to indicate the ongoing
process)
C) wait until the SETUP LED stays
steady on (indicating the reader
has detected the barcode)
Figure 7 – Auto Setup Function
This procedure ends either when the barcode is successfully decoded or after a timeout of about 7
(seven) seconds.
The scanner will restart at the end of the process, and then the optimized reading parameters for that
barcode are automatically configured in scanner memory.
NOTE
If your application has been configured using Mode Button, go to STEP 5.
Reset Scanner to Factory Default (Optional)
If it ever becomes necessary to reset the scanner to the factory default values, you can perform this
procedure by holding the Mode push button pressed while powering up the scanner. At the end of the
procedure (about 5-6 seconds), the Configuration and Environmental parameters are reset, and all
LEDs blink simultaneously 3 times.
7
VB14N
STEP 4 – INSTALLING GENIUS™ CONFIGURATION PROGRAM
Genius
Wizard approach for new users;
Multi-language version;
Defined configuration directly stored in the reader;
Communication protocol independent from the physical interface allowing to consider the reader as
Install the Genius
This configuration procedure assumes scanner conne ction to a CBX100/500. Genius™, running on a
laptop computer, is connected to the scanner auxiliary port through the CBX100/500 9-pin connector.
To communicate with the scanner, Genius™ performs an auto baudrate detection starting from its
default parameters which are 115200, 8, N, 1. These parameters can also be set in the Genius™
Tools>Options>Communications window.
Wizard for Quick Reader Setup
After installing the Genius™ software program the following window appears asking the user to
choose the desired configuration level.
™
is a Barcode scanner configuration tool providing several important advantages:
a remote object to be configured and monitored.
™
Software! (Download from www.pepperl-fuchs.com)
Figure 8 - Genius™ Wizard Opening Window
The Wizard option is advised for rapid configuration or new users, since it show s a step-by-step
scanner configuration.
8
VB14N
1. Select the Create a new configuration button.
You will be guided through the configuration being asked to define the following parameters:
a. Barcode selection and definition
9
VB14N
b. Operating mode selection and definition
c. Digital Outputs configuration
10
VB14N
d. Hardware interface selection
e. Output data format configuration
The On Line operating Mode requires the reader to be connected to an External
Trigger/Presence Sensor using I1A and I1B inputs.
The Automatic operating mode does not require connection to an external Presence Sensor.
When working in this mode the reader is continuously scanning, while the reading phase is
activated each time a barcode enters the reader reading zone. The reader stops reading after
an N number of scans without a code. Barcode characters are transmitted on the serial
interface. In case of a failed reading phase no message is sent to the host computer.
11
VB14N
2. After defining the parameter values the following window appears allowing to complete the reader
configuration as follows:
Saving the configuration to disk;
Switching to Advanced mode;
Sending the configuration to the scanner.
3. After sending the configuration to the
scanner you have completed the
configuration process.
4. By clicking Finish, the System Information
window will be displayed with specific
information concerning the scanner.
12
VB14N
STEP 5 – TEST MODE
Use a code suitable to your application to test the system.
1. Enter the Test mode function by holding the Mode push button pressed until the TEST LED is on.
2. Release the button to enter the Test mode function.
Once entered, the Bar-Graph on the five LEDs is activated and if the scanner starts reading
barcodes the Bar-Graph shows the Good Read Rate. In case of no read condition, only the
STATUS LED is on and blinks.
Figure 9 – Test Mode Function
3. To exit the Test Mode, press the Mode push button once.
NOTE
By default, the Test Mode exits automatically after two minutes.
13
VB14N
ADVANCED SCANNER CONFIGURATION
The following are alternative or advanced scanner configuration methods:
Host Mode Programming
The scanner can also be configured from a host computer using the Host Mode programming
procedure, by commands via the serial interface.
Advanced Genius™ Configuration
The ADVANCED selection available when starting the Genius™ program is addressed to expert users
being able to complete a detailed scanner configuration. By choosing this option it is possible either to
start a new scanner configuration or to open and modify an old one. The desired parameters can be
defined in the following window, similar to the MS Explorer:
Figure 10 - Genius™ Parameter Explorer Window
Alternative Layouts
The ID-NET™ is a built-in high-speed interface dedicated for high-speed scanner interconne ction.
ID-NET™ is in addition to the Main and Auxiliary serial interfaces.
If you need to install an ID-NET™ network refer to the VB14N Reference Manual.
The scanner can also be configured by reading programming barcodes. See the ID-NET™ Setup
Procedure Using Programming Barcodes.
If you need to install a Pass-Through network refer to the VB14N Reference Manual.
If you need to install a Multiplexer network refer to the VB14N Reference Manual.
If you need to install an RS232 Master/Slave (for backward compatibility) refer to the VB14N
Reference Manual.
14
VB14N
2 INTRODUCTION
2.1 PRODUCT DESCRIPTION
The VB14N laser scanner satisfies the most advanced needs of a wide range of users. It has been
developed focusing on the realistic requirements of its target market. The outstanding result is an
extremely compact, cost-effective and easy to use industrial scanner.
Standard Application
Program
Some of the main features of VB14N are listed below:
ACB (Advanced Code Builder)
A standard application program is factory-loaded onto the VB14N. This
program controls barcode reading, serial port interfacing, data formatting
and many other operating and control parameters.
It is completely configurable from a host computer through the Genius™
utility program.
small dimensions and light weight
software programmable scanning speed on all models
linear and raster version
completely configurable via serial interface (Genius™)
3 serial communication interfaces (Main, Auxiliary, ID-NET™)
supply voltage from 10 to 30 Vdc
reads all popular codes
test mode to verify the reading features and exact positioning of the scanner without the need for
external tools
programmable in 4 different operating modes to suit the most various barcode reading system
requirements
code verifier
low power consumption
The VB14N uses a solid-state laser diode as a light source; the light emitted has a wavelength
between 630 and 680 nm. Refer to the section “Safety Precautions” at the beginning of this manual for
information on laser safety.
The protection class of the enclosure is IP65, the reader is therefore suitable for industrial
environments where high protection against harsh external conditions is required.
15
VB14N
2.1.1 Indicators
The five LEDs on the side of the scanner (
READY
(green)
This LED indicates the device is ready to operate.
Figure A) indicate the following:
GOOD
TRIGGER
COM
STATUS
(green)
(yellow)
(yellow)
(red)
This LED confirms successful reading.
This LED indicates the status of the reading phase. *
This LED indicates active communication on main serial port. **
This LED indicates a NO READ result.
* In On-Line mode the TRIGGER LED corresponds to the active reading phase signaled by the Presence Sensor.
In Automatic and Continuous modes the TRIGGER LED is always on indicating that the reader is ready to read a
code.
** When connected to a Fieldbus network through the CBX500, the COM LED is always active, even in the
absence of data transmission, because of polling activity on the Fieldbus network.
During the reader startup (reset or restart phase), all the LEDs blink for one second.
On the back of the reader near the cable, the “POWER ON” LED indicates the laser scanner is
correctly powered.
2.2 ID-NET™
The ID-NET™ is a built-in high-speed interface dedicated for high-speed
scanner interconnection. The ID-NET™ is in addition to the Main and
Auxiliary serial interfaces.
The following network configurations are available:
ID-NET™ M/S Synchronized: Single station – multiple scanners
CBX100
CBX100CBX100
ID-NET™ interface allows local connection of
multiple scanners reading different sides of the
same target. All scanners share a single
presence sensor and activate/deactivate
simultaneously.
At the end of each reading phase a single data
message is transmitted to the host.
Thanks to ID-NET™, data communication
among scanners is highly efficient so that an
immediate result will be available.
16
VB14N
ID-NET™ M/S Multidata: Multiple stations – single scanner
CBX100
CBX100
CBX100
ID-NET™ interface allows connection of scanners reading objects placed on independent conveyors.
All scanners are typically located far away from each other and they use a dedicated presence sensor.
At the end of each reading phase, each scanner transmits its own data message to the host.
Thanks to ID-NET™, data collection among readers is accomplished at a high speed without the need
of external multiplexing device. This leads to an overall cost reduction and to a simple system wiring.
17
VB14N
2.2.1 How To Setup/Configure the Scanner Network
A complete ID-NET™ scanner network can be rapidly setup, as follows:
Mounting & Connection
1. Mechanically mount/install all the readers (refer to par. 3.1 and 3.2).
2. Wire ID
3.
Con
4. Power up the entire system.
Configuration
1. Launch Genius™.
2. From the Genius™ Device Menu select “Local Device Network Settings” and program the Role of
the Master scanner (Synchronized or Multidata).
This procedure requires the Network Baud Rate be the same for all Slave s and M aster, (500 kbs
is the default value). It can be changed after network setup using Geniu s™ thro ugh the Master
scanner. See also the alternative procedure in the note below.
3. At the prompt to "Send updated Network configuration to the Local Device" (Master) choose
"Yes".
4. Then run the NET-AUTOSET procedure from the Icon in the Devices Area. Genius™ sets all
slave scanners according to the Master Role (Synchronized or Multidata), and assign s each a
random address. If necessary, this address can be changed through the Network Wizard.
5. Configure the System parameters via Genius™.
6. If using the CBX connection box equipped with a BM100 Backup module, perform System Backup
at the Master.
The scanner network is ready.
NOTE
-NET™ (refer to par. 4.3 or 5.3).
n
ect the planned Master scanner to a PC by means of the Genius™ configuration software.
If necessary, the ID-NET™ baudrate can be set individually on each Slave scanner
to match the Master. Connect each Slave to Genius™ and set the
Reading System Layout > Network Baudrate parameter. Then follow the procedure
above.
18
NOTE
An alternative method of programming scanner address and rol e assignment can be
accomplished by using the “Connectivity Programm ing Barcodes”.
VB14N
2.3 HUMAN MACHINE INTERFACE
The Mode Button is the intuitive Human Machine Interface
designed with the precise goal of improving ease of
installation and maintenance.
Status and diagnostic information are clearly presented by
means of five-colored LEDs, whereas the single multi-function
key gives immediate access to relevant functions:
Autosetup to self-optimize reading performance in
demanding applications
Autolearn to self-detect unknown barcodes
Test Mode with bar-graph visualization to check
static reading performance
The colors and meaning of the five LEDs when in the one of the operating modes (On-Line, Automatic
or Continuous) are illustrated in par 2.1.1.
The Mode Button functions do not work if the motor or laser are turned off, see chp.
9 for details.
NOTE
2.3.1 Diagnostic Indication
The “STATUS” and “READY” LEDs blink simultaneously to signal the presence of a failure.
Diagnostic message transmission on interfaces can be enabled to provide details about specific failure
conditions.
At the same time one or more LEDs light up according to the following scheme:
LED STATUS
READY BLINK
GOOD
TRIGGER ON to indicate a Motor Failure.
COM ON to indicate a Laser Failure.
STATUS BLINK
ON to indicate any Failure different than
Motor or Laser failures.
19
VB14N
2.3.2 Mode Button Functions
Quick access to the following functions is provided by an
easy procedure using the push button:
1 – Press the button (the STATUS LED will give a visual
feedback).
2 – Hold the button until the specific function LED is on
(TEST, LEARN or SETUP).
3 – Release the button to enter the specific function.
Once button is pressed, the cycle of LEDs activation is as follows:
Release button to
Exit
Release button to
enter Test Mode
Release button to
enter AutoLearn
Release button to
enter AutoSetup
Release button to
Exit
(cycle)
Test Mode Function
Once entered, the Bar-Graph on the five LEDs is activated and if the scanner starts reading barcodes
the Bar-Graph shows the Good Read Rate. In case of no read condition, only the STATUS LED is on
and blinks.
To exit the Test Mode, press the Mode push button once.
20
VB14N
AutoLearn Function
Once entered, the reader starts a procedure to automatically detect and recognize barcodes (by type
and length), which are presented to it
ongoing process.
The procedure is as follows:
- place the desired barcode on the scanline.
- wait until the LEARN LED stays steady on (indicating the reader has detected the barcode).
- repeat, if needed, the above two steps to program up to 10 different barcodes (the LEARN LED
returns to the blinking state for the next code). If more than one barcode is detected in the scan
line, the Multi Label mode is enabled (refer to the “Software Configuration Parameter Guide” Help
file).
- exit the process by pressing the Mode push button once.
The scanner will restart at the end of the process, and then the detected barcodes are automatically
configured in scanner memory.
AutoSetup Function
Once entered, if a barcode label is positioned in front of the scanline, the scanner automatically
performs the optimal setup of the reading parameters for that specific barcode.
The procedure is as follows:
1
. The laser turns on and the LEARN LED blinks to indicate the
- place the desired barcode on the scanline.
- enter the AutoSetup function (the laser turns on and the SETUP LED blinks to indicate the
ongoing process).
- wait until the SETUP LED stays steady on (indicating the reader has detected the barcode).
This procedure ends either when the barcode is successfully decoded or after a timeout of about 7
(seven) seconds.
The scanner will restart at the end of the process, and then the optimized reading parameters for that
barcode are automatically configured in scanner memory.
The AutoSetup function does not modify the programmed barcode symbologies. If
needed, the AutoLearn function can be performed after Autosetup.
NOTE
Reset Scanner to Factory Default
If it ever becomes necessary to reset the scanner to the factory default values, you can perform this
procedure by holding the Mode push button pressed while powering up the scanner. At the end of the
procedure (about 5-6 seconds), the Configuration and Environmental parameters are reset, all LEDs
blink simultaneously 3 times.
1
In case of Programming Barcodes (refer to the “ID-NET™: Programming Barcodes And Setup Procedure”
document)
21
VB14N
3 INSTALLATION
3.1 MECHANICAL INSTALLATION
VB14N can be installed to operate in different positions. The four screw holes (M4 x 5) on the body of
the reader are for mechanical fixture (
the scanner and mounting bracket and may be used for installation.
Refer to par. 3.1.1 and 3.2 for correct positioning.
23.3
14
Figure A, 3). The diagrams below give the overall dimensions of
84
4
A
14.76
M4
10.3
READY
mode
GOOD
TRIGGER
COM
STATUS
B
32.7
auto setup
46
68
40
auto learn
test mode
exit
10.3
40
Figure 11 –VB14N Overall Dimensions
9
4.2
7.8
90°
73
R
20°
4
0
13.8
30
4.
2
17.5
42
2.5
mm
2.5
Figure 12 – Mounting Bracket Overall Dimensions
23
4
.
2
1 x 45° n° 2
n
°
2
22
VB14N
3.1.1 Mounting VB14N
Using the VB14N mounting bracket you can obtain the most suitable position for the reader as shown
in the figure below:
Tilt
Skew
Skew
Pitch
Figure 13 – Positioning with Mounting Bracket
23
VB14N
3.1.2 Mounting Scanner Accessories
DM-VB14A is an accessory deflection mirror available on request for VB14N.
The DM-VB14A is a 90° deflection mirror
The installation of the deflection mirror is very easy (Figure 14).
Avoid any contact with the deflection mirror, mirrored rotor, the lenses or other
optical components; otherwise the performance of the reader will be reduced.
CAUTION
1. Turn off the device.
2. Remove the VB14N scanning window unscrewing the two cover screws.
3. Fix the mirror to the device by means of the two fixing screws.
4. Remount the scanning window so that the opening face is now at 90° with respect to the VB14N
body.
deflection mi rror
Cover screws
Scanning window
Fixing screws
Figure 14 - Installation of the Deflection Mirror
24
VB14N
3.2 POSITIONING
The VB14N scanner is able to decode moving barcode labels at a variety of angles, however
significant angular distortion may degrade reading performance.
When mounting the VB14N take int o consi deration th ese three id eal labe l positio n angl es: Skew 10° to 30°, Tilt 0° and Pitch 0°.
Follow the suggestions for the best orientation:
The Skew angle is represented by the value S in Figure 15. Position the reader to assure
for the Skew angle. This avoids the direct reflection of the laser light emitted by the VB14N.
For the raster version, this angle refers to the most inclined or external raster line, so that all other raster
lines assure more than 10° Skew.
at least 10°
S
Figure 15 - Skew Angle
The Tilt angle is represented by the value T in Figure 16. Position the reader in order to minimize
Tilt angle.
Figure 16 - Tilt Angle
By using the ACB (Advanced Code Builder) software parameter, the tilt angle is less critical and can
be decoded even if the scan line doesn’t cross the entire co de.
See par. 7.1 or the Help On Line for details.
T
the
25
VB14N
The Pitch angle is represented by the value P in Figure 17. Position the reader in order to minimize
the Pitch ang
le.
Figure 17 - Pitch Angle
P
26
VB14N
4 CBX ELECTRICAL CONNECTIONS
All VB14N models are equipped with a cable terminated by a 25-pin male D-sub connector for
connection to the power supply and input/output signals.
We recommend making system connections through one of the CBX connection boxes since they
offer the advantages of easy connection, easy device replacement and filtered reference signa ls.
If you require direct wiring to the scanner the details of the connector pins and
NOTE
The table below gives the pinout of the CBX100/500 terminal block connectors. Use this pinout when
the VB14N reader is connected by means of the CBX100/500:
Vdc Power Supply Input Voltage +
GND Power Supply Input Voltage Earth Protection Earth Ground
+V Power Source – External Trigger
I1A External Trigger A (polarity insensitive)
I1B External Trigger B (polarity insensitive)
-V Power Reference – External Trigger
+V Power Source – Inputs
I2A Input 2 A (polarity insensitive)
I2B Input 2 B (polarity insensitive)
* Do not leave floating, see par. 4.2.2 for connection details.
27
VB14N
To avoid electromagnetic interference when the scanner is connected to a CBX
NOTE
4.1 POWER SUPPLY
Power can be supplied to the scanner through the CBX100/500 spring clamp terminal pins as shown
in Figure 18:
The power must be between 10 and 30 Vdc only.
It is recommended to connect the device CHASSIS to earth ground (Earth) by setting the appropriate
jumper in the CBX connection box. See the CBX Installation Manual for details.
4.2 MAIN SERIAL INTERFACE
connection box, verify the jumper positions in the CBX as indicated in its Installation
Manual.
Power Supply
VGND
V+
Earth
Ground
in
Figure 18 - Power Supply Connections
Do not connect to the Main Interface spring clamp terminals if using Host Interface
Modules (Fieldbus) with the CBX500.
CAUTION
The signals relative to the following serial interface types are available on the CBX spring clamp
terminal blocks.
If the interface type is not compatible with the current communication handshaking, then the system
forces the handshake to none.
The main interface type and the relative parameters (baud rate, data bits, etc.) can be set using
the Genius™ utility program or the Genius™ based Host Mode Programming procedure.
Details regarding the connections and use of the interfaces are given in the next paragraphs.
28
VB14N
4.2.1 RS232 Interface
The serial interface is used in this case for point-to-point connections; it handle s co mmunication with
the host computer and allows both transmission of code data and the programming of the scanner.
This is the default setting.
The following pins are used for RS232 interface connection:
CBX100/500
TX Transmit Data
RX Receive Data
RTS Request To Send
CTS Clear To Send
SGND Signal Ground
It is always advisable to use shielded cables. The overall maximum cable length must be less than 15
m (49.2 ft).
Function
USER INTERFACE
SGND RXD TXD
CTS RTS
SCANNER
SGND TX RX
RTS CTS
Figure 19 – RS232 Main Interface Connections Using Hardware Handshaking
The RTS and CTS signals control data transmission and synchronize the connected devices.
+ V
RTS
- V
+ V
TX DATA
- V
START
OF
TRANSMISSION
TRANSMISSION
C1
DATA
C2
TRANSMISSION
STOPPED
DATA
TRANSMISSION
C4
C3
END
OF
TRANSMISSION
C5
+ V
CTS
- V
IDLE
ENABLED
DISABLED
ENABLED
IDLE
Figure 20 - RS232 Control Signals
If the RTS/CTS handshaking protocol is enabled, the VB14N activates the RTS output to indicate a
message is to be transmitted. The receiving unit activates the CTS input to enable the transmission.
29
VB14N
4.2.2 RS485 Full-Duplex Interface
The RS485 full-duplex (5 wires + shield) interface is used for non -polled communication protocols in
point-to-point connections over longer distances (max 1200 m / 3940 ft) than those acceptable for
RS232 communications or in electrically noisy environments.
The CBX pinout follows:
CBX100/500
TX+ RS485 Transmit Data +
RX+ RS485 Receive Data +
TX- RS485 Transmit Data -
RX- RS485 Receive Data -
SGND Signal Ground
Function
USER INTERFACE
RX485+ TX485+
SGND RX485- TX485-
SCANNER
Figure 21 - RS485 Full-duplex Connections
SGND TX+ RX+
TX- RX-
For applications that do not use RX485 signals, do not leave these lines floating but
NOTE
connect them to SGND as shown below.
SCANNER
Figure 22 - RS485 Full-duplex Connections using Only TX Signals
USER INTERFACE
RX485+
SGND RX485-
SGND TX+
TX-
30
VB14N
4.2.3 RS485 Half-Duplex Interface
This interface is provided for backward compatibility. We recommend using the more
NOTE
The RS485 half-duplex (3 wires + shield) interface is used for polled communication protocols.
efficient ID-NET™ network for Master/Slave or Multiplexer layouts.
CBX100/500
Function
RTX+ RS485 Receive/Transmit Data +
RTX- RS485 Receive/Transmit Data -
SGND Signal Ground
USER INTERFACE
RTX485+
SGND RTX485-
SCANNER
Figure 23 - RS485 Half-duplex Connections
SGND RTX+
RTX-
This interface is forced by software when the protocol selected is MUX32 protocol.
31
VB14N
4.3 ID-NET™ INTERFACE
CBX100/500 Function
Shield Network Cable Shield
ID+ ID-NET™ network +
ID- ID-NET™ network -
REF Network Reference
4.3.1 ID-NET™ Cables
The following instructions are referred to Figure 25, Figure 26 and Figure 27.
The general cable type specifications are: CAT5 twisted pair + additional CAT5 twisted pair,
shielded cable AWG 24 (or AWG 22) stranded flexible.
We recommend using
DeviceNet cables (drop or trunk type) to the following reference standards:
AN50325 – IEC 62026
UL STYLE 2502 80°C 30V
Cable Shield MUST be connected to earth ground ONLY at the Master.
NEVER use ID-NET™ cable shield as common reference.
The ID-NET™ max cable length depends on the baudrate used, (see the Baudrate Table below).
For Common Power Connections use only 2 wires (ID+ and ID-).
- DC Voltage Power cable (Vdc – GND) should be handled as a signal cable (i.e. do not put it
together with AC cable):
- Wire dimensioning must be checked in order to avoid voltage drops greater than 0.8 Volts.
- Cable should lie down as near as possible to the ID-NET™ cable (avoiding wide loops
between them).
Scanner's chassis may be connected to earth.
Network inside the same building.
Baudrate Table
Baud Rate 125 kbps 250 kbps
Cable Length 1200 m 900 m
500 kbps
700 m
32
NOTE
The default ID-NET™ baudrate is 500 kbps. Lower ID-NET™ baudrates allow longe r
cable lengths.
VB14N
4.3.2 ID-NET™ Response Time
The following figure shows the response time of the ID-NET™ network. This time is defined as the
period between the Trigger activation and the beginning of data transmission to the Ho st.
Max ID-NET™ Response Time
240
220
200
180
160
140
120
100
80
Response Time (ms)
60
40
20
0
0 1 2 3 4 56789 10 1112 1314 15 16
500 kbps250 kbps
Figure 24 – ID-NET™ Response Time
CONDITIONS:
ID-NET™ M/S Synchronized layout
message length = 50 bytes per node
Number of Nodes
125 kbps
33
VB14N
24V DC
24V DC
24V DC
Figure 25 – ID-NET™ Network Connections with isolated power blocks
34
VB14N
Figure 26 - ID-NET™ Network Connections with Common Power Branch Network
35
VB14N
36
Figure 27 – ID-NET™ Network Connections with Common Power Star Network
VB14N
4.3.3 ID-NET™ Network Termination
The network must be properly terminated in the first and last scanner of the network. This is done by
setting the ID-NET™ Termination Resistance Switch in the CBX100/500 to ON.
4.4 AUXILIARY RS232 INTERFACE
The auxiliary serial interface is used exclusively for RS232 point-to-point connections.
The parameters relative to the aux interface (baud rate, data bits, etc.) as well as particular
communication modes such as LOCAL ECHO can be defined using the Genius™ utility program or
Genius™ based Host Mode Programming.
The 9-pin female Auxiliary Interface connector inside the CBX is the preferred connector for device
configuration or communication monitoring.
5
Figure 28 - 9-pin female connector
If permanent system wiring is required, the following pins are used to connect the RS232 auxiliary
interface:
CBX100/500Function
RX Auxiliary Interface Receive Data
TX Auxiliary Interface Transmit Data
SGND Auxiliary Interface Reference
1
69
USER INTERFACE
RX TX
Reference
Figure 29 - RS232 Auxiliary Interface Connections
Do not connect the Aux Interface to the CBX spring clamp connectors and the 9-pin
NOTE
connector simultaneously.
37
VB14N
r
4.5 INPUTS
There are two optocoupled polarity insensitive inputs available on the scanner: Input 1 (External
Trigger) and Input 2, a generic input:
The electrical features of both inputs are:
Maximum voltage: 30 Vdc
Maximum current: 12 mA (scanner) + 12 mA (CBX)
An anti-disturbance filter is implemented in software on both inputs so that the minimum pulse duration
is 5 milliseconds. This value can be increased through the software parameter Debounce Filter, see
the "Software Configuration Parameter Guide” Help file".
CBX100/500
+V Power Source - External Trigger
I1A External Trigger A (polarity insensitive)
I1B External Trigger B (polarity insensitive)
-V Power Reference - External Trigger
The External Trigger input is used in the On-Line operating Mode and tells the scanner to scan for a
code. The active state of this input is selected in software. Refer to the Genius™ Help On Line.
The yellow Trigger LED (
to ON.
This input is optocoupled and can be driven by both an NPN and PNP type command. The
connections are indicated in the following diagrams:
EXTERNAL TRIGGER INPUT CONNECTIONS USING VB14N POWER
Function
Figure A, 3) is on when the active state of the External Trigger corresponds
Triggersensor (PNP)
(brown)
(b la c k) (blue)
Figure 30 – Trigger sensor (PNP) External Trigger Using VB14N Power
NPN senso
Power to In p ut
Photocell Signal
Photocell
Reference
Figure 31 - NPN External Trigger Using VB14N Power
EXTERNAL TRIGGER INPUT CONNECTIONS USING EXTERNAL POWER
38
VB14N
PNP Photocell
Input
Signal
Pulled down to External
Input Device Reference
Figure 32 - PNP External Trigger Using External Power
NPN Photocell
Pulled up to External
Input Device Power
Input
Signal
Figure 33 - NPN External Trigger Using External Power
39
VB14N
CBX100/500 Function
+V Power Source - Inputs
I2A Input 2 A (polarity insensitive)
I2B Input 2 B (polarity insensitive)
-V Power Reference - Inputs
INPUT 2 CONNECTIONS USING VB14N POWER
PNP Input 2 Using VB14N Power
Input Device
Power to
Input Device
Input Input Device
Signal Reference
Input Device
Power to Input
Input Device Signal
Input Device
Reference
NPN Input 2 Using VB14N Power
INPUT 2 CONNECTIONS USING EXTERNAL POWER
Input Device
Input
Signal
Pulled down to External
Input Device Reference
Figure 34 - PNP Input 2 Using External Power
Input Device
Pulled up to External
Input Device Power
Input
Signal
40
Figure 35 - NPN Input 2 Using External Power
VB14N
4.5.1 Code Verifier
If the VB14N is used as a Code Verifier, the verifier code can be configured in software through the
Genius™ configuration program. However it is also possible to use one of the inputs to trigger when
the scanner should store a code read as the verifier code.
The Code Verifier parameter must be enabled, and the configuration parameters to allow correct Co de
Type reading must be saved to the scanner in order to read the verifier code.
When the selected input is activated, the next read code will be stored as the verifier code in the
scanner's non-volatile (Flash) memory.
For more details see the Verifier Parameters in the "Software Configuration Parameter Guide” Help
file.
4.6 OUTPUTS
Two general purpose outputs are available.
CBX100/500
+V Power Source - Outputs
O1+ Output 1 +
O1- Output 1 -
O2+ Output 2 +
O2- Output 2 -
-V Power Reference Outputs
The meaning of the two outputs Output 1 and Output 2 can be defined by the user (No Read, Right,
Wrong, etc.). Refer to the Genius™ Help On Line.
By default, Output 1 is associated with the No Read event, which activates when the code signaled by
the external trigger is not decoded, and Output 2 is associated with the Complete Read event, which
activates when all the selected codes are correctly decoded.
The output signals are fully programmable being determined by the configured Activation/Deactivation
events, Deactivation Timeout or a combination of the two.
Function
41
VB14N
OUTPUT CONNECTIONS USING VB14N POWER
Output Device
Power to Output
Output device Signal
Output device
Reference
Figure 36 - Open Emitter Output Using VB14N Power
Output Device
Power to
Output device
Output device
Reference
Output
Signal
Figure 37 - Open Collector Output Using VB14N Power
OUTPUT CONNECTIONS USING EXTERNAL POWER
Output Device
Pulled up to External
Output Device Power
Output
Signal
Figure 38 - Open Emitter Output Using External Power
Output Device
Output
Signal
Pulled down to External
Output Device Reference
Figure 39 - Open Collector Output Using External Power
max = 30 Vdc
V
CE
I max = 40 mA continuous; 130 mA pulsed
42
VB14N
4.7 USER INTERFACE - HOST
The following table contains the pinout for standard RS232 PC Host interface. For other user interface
types please refer to their own manual.
All VB14N models are equipped with a cable terminated by a 25-pin male D-sub connector for
connection to the power supply and input/output signals. The details of the connector pins are
indicated in the following table.
1
Figure 40 - 25-pin Male D-sub Connector
25-pin D-sub male connector pinout
Pin Name Function
13, 9 Vdc Power supply input voltage +
25, 7 GND Power supply input voltage -
1 CHASSIS Cable shield connected to chassis
18 I1 A External Trigger A (polari ty i nsensitive)
19 I1B External Trigger B (polarity insensitive)
6 I2A Input 2 A (polarity insensitive)
10 I2B Input 2 B (polarity insensitive)
* Do not leave floating, see par. 5.2.2 for connection details.
44
VB14N
V
5.1 POWER SUPPLY
Power can be supplied to the scanner through the pins provided on the 25-pin connector used for
communication with the host (Figure 41):
B14N
13
25
1
Vdc
GND
CHASSIS
POWER SUPPLY
V+ (10 - 30 Vdc )
VGND
CHASSIS
Earth Ground
Figure 41 - Power Supply Connections
The power must be between 10 and 30 Vdc only.
It is recommended to connect pin 1 (CHASSIS) to a common earth ground.
5.2 MAIN SERIAL INTERFACE
The signals relative to the following serial interface types are available on the input/output connector of
VB14N.
If the interface type is not compatible with the current communication handshaking, then the system
forces the handshake to none.
The main interface type and the relative parameters (baud rate, data bits, etc.) can be set using
the Genius™ utility program or the Genius™ based Host Mode Programming procedure.
Details regarding the connections and use of the interfaces are given in the next paragraphs.
45
VB14N
5.2.1 RS232 Interface
The serial interface is used in this case for point-to-point connections; it handle s co mmunication with
the host computer and allows both transmission of code data and the programming of the scanner.
This is the default setting.
The following pins are used for RS232 interface connection:
NameFunction
25-pin
2 TX Transmit Data
3 RX Receive Data
4 RTS Request To Send
5 CTS Clear To Send
7 GND Ground
It is always advisable to use shielded cables. The overall maximum cable length must be less than 15
m (49.2 ft).
VB14N
TX
2
RX
3
RTS
4
5
CTS
7
GND
1Chassis
USER IN TERF ACE
RXD
TXD
CTS
RTS
GND
Figure 42 – RS232 Main Interface Connections Using Hardware Handshaking
The RTS and CTS signals control data transmission and synchronize the connected devices.
+ V
RTS
- V
+ V
TX DATA
- V
+ V
CTS
- V
START
OF
TRANSMISSION
ENABLED
IDLE
Figure 43 - RS232 Control Signals
DATA
TRANSMISSION
C2
C1
TRANSMISSION
DISABLED
STOPPED
DATA
TRANSMISSION
C4
C3
ENABLED
END
OF
TRANSMISSION
C5
IDLE
If the RTS/CTS handshaking protocol is enabled, the VB14N activates the RTS output to indicate a
message is to be transmitted. The receiving unit activates the CTS input to enable the transmission.
46
VB14N
V
V
5.2.2 RS485 Full-Duplex Interface
The RS485 full-duplex (5 wires + shield) interface is used for non -polled communication protocols in
point-to-point connections over longer distances (max 1200 m / 3940 ft) than those acceptable for
RS232 communications or in electrically noisy environments.
The connector pinout follows:
NameFunction
25-pin
2 TX+ RS485 Transmit Data +
3 RX+ RS485 Receive Data +
4 TX- RS485 Transmit Data 5 RX- RS485 Receive Data 7 GND Ground
Chassis
B14N
TX+ 2
4
TXRX+ 3
RX-
5
7
GND
1
USER INTERFACE
+
RX485
-
+
TX485
-
GND
Figure 44 - RS485 Full-duplex Connections
For applications that do not use RX signals, do not leave these lines floating but
NOTE
connect them to GND as shown below.
Chassis
B14N
TX+ 2
4
TXRX+ 3
RX-
5
7
GND
1
USER INTERFACE
+
RX485
-
GND
Figure 45 - RS485 Full-duplex Connections using Only TX Signals
47
VB14N
5.2.3 RS485 Half-Duplex Interface
This interface is provided for backward compatibility. We recommend using the more
NOTE
The RS485 half-duplex (3 wires + shield) interface is used for polled communication protocols.
The connector pinout follows:
25-pin
2 RTX+ RS485 Receive/Transmit Data +
4 RTX- RS485 Receive/Transmit Data 7 GND Ground
efficient ID-NET™ network for Master/Slave or Multiplexer layouts.
NameFunction
VB14N
MULTIPLEXER
2
RTX+
4
RTX-
7
GND
1Chassis
RTX485 +
RTX485 RS485REF
Figure 46 - RS485 Half-duplex Connections
This interface is forced by software when the protocol selected is MUX32 protocol.
48
VB14N
5.3 ID-NET™ INTERFACE
25-pin Name Function
23 ID+ ID-NET™ network +
24 ID- ID-NET™ network -
7 GND Ground
5.3.1 ID-NET™ Cables
The following instructions are referred to Figure 48, Figure 49 and Figure 50.
The general cable type specifications are: CAT5 twisted pair + additional CAT5 twisted pair,
shielded cable AWG 24 (or AWG 22) stranded flexible.
We recommend using
DeviceNet cables (drop or trunk type) to the following reference standards:
AN50325 – IEC 62026
UL STYLE 2502 80°C 30V
Cable Shield MUST be connected to earth ground ONLY at the Master.
NEVER use ID-NET™ cable shield as common reference.
The ID-NET™ max cable length depends on the baudrate used, (see the Baudrate Table below).
For Common Power Connections use only 2 wires (23 and 24).
- DC Voltage Power cable (Vdc – GND) should be handled as a signal cable (i.e. do not put it
together with AC cable):
- Wire dimensioning must be checked in order to avoid voltage drops greater than 0.8 Volts.
- Cable should lie down as near as possible to the ID-NET™ cable (avoiding wide loops
between them).
Scanner's chassis may be connected to earth.
Network inside the same building.
Baudrate Table
Baud Rate 125 kbps 250 kbps
Cable Length 1200 m 900 m
500 kbps
700 m
* Application dependent, contact your Datalogic Automation representative for details.
The default ID-NET™ baudrate is 500 kbps. Lower ID-NET™ baudrates allow longe r
NOTE
cable lengths.
49
VB14N
5.3.2 ID-NET™ Response Time
The following figure shows the response time of the ID-NET™ network. This time is defined as the
period between the Trigger activation and the beginning of data transmission to the Ho st.
Max ID-NET™ Response Time
240
220
200
180
160
140
120
100
80
Response Time (ms)
60
40
20
0
0 1 2 3 4 56789 10 1112 1314 15 16
500 kbps250 kbps
Figure 47 – ID-NET™ Response Time
CONDITIONS:
ID-NET™ M/S Synchronized layout
message length = 50 bytes per node
Number of Nodes
125 kbps
50
VB14N
24V DC
24V DC
24V DC
Figure 48 – ID-NET™ Network Connections with isolated power blocks
51
VB14N
52
Figure 49 - ID-NET™ Network Connections with Common Power Branch Network
VB14N
Figure 50 – ID-NET™ Network Connections with Common Power Star Network
53
VB14N
V
5.3.3 ID-NET™ Network Termination
The network must be properly terminated by a 120 Ohm resistor at the first and last scanner of the
network.
5.4 AUXILIARY RS232 INTERFACE
The auxiliary serial interface is used exclusively for RS232 point-to-point connections.
The parameters relative to the aux interface (baud rate, data bits, etc.) as well as particular
communication modes such as LOCAL ECHO can be defined using the Genius™ utility program or
Genius™ based Host Mode Programming.
The following pins of the 25-pin connector are used to connect the RS232 auxiliary interface:
Pin Name Function
20 RX Receive Data
21 TX Transmit Data
7 GND Ground
B14N
USER INTERFACE
20
Chassis
21
1
RX
TX
7
GND
TXD
RXD
GND
Figure 51 - RS232 Auxiliary Interface Connections
54
VB14N
V
-
5.5 INPUTS
There are two optocoupled polarity insensitive inputs available on the scanner: Input 1 (External
Trigger) and Input 2, a generic input:
The electrical features of both inputs are:
Maximum voltage: 30 Vdc
Maximum current: 12 mA
An anti-disturbance filter is implemented in software on both inputs so that the minimum pulse duration
is 5 milliseconds. This value can be increased through the software parameter Debounce Filter, see
the "Software Configuration Parameter Guide” Help file".
NameFunction
25-pin
9 Vdc Power Source - External Trigger
18 I1A External Trigger A (polarity insensitive)
19 I1B External Trigger B (polarity insensitive)
7 GND Power Reference - External Trigger
The External Trigger input is used in the On-Line operating Mode and tells the scanner to scan for a
code. The active state of this input is selected in software. Refer to the Genius™ Help On Line.
The yellow Trigger LED (
to ON.
This input is optocoupled and can be driven by both an NPN and PNP type command. The
connections are indicated in the following diagrams:
Figure A, 3) is on when the active state of the External Trigger corresponds
EXTERNAL TRIGGER INPUT PNP Photocell
B14N
9
Vdc
PNP Photocell
(brown) +10-30 Vdc
I1A
VCC
18
~
~
19
I1B
7
GND
(black) NO
(blue) 0 V
Figure 52 - Photocell (PNP) External Trigger Using VB14N Power
55
VB14N
V
8
9
9
7
-
8
9
7
-
8
V
8
-
EXTERNAL TRIGGER INPUT CONNECTIONS USING VB14N POWER
B14N
Vdc
EXTE RNAL TR IGGER
V
Signal
VCC ~
I1A
1
~
I1B
1
GND
Ground
Figure 53 - PNP External Trigger Using VB14N Power
Vdc
9
I1A
1
CC
+
~
I1B
1
GND
EXTE RNAL TR IGGERVB14N
V
Signal
Ground
Figure 54 - NPN External Trigger using VB14N Power
EXTERNAL TRIGGER INPUT CONNECTIONS USING EXTERNAL POWER
VB14N
Vext 30 Vdc max.
EXTERNAL TRIGG ER
V
Signal
VCC
1
I1A
~
+
~
19
I1B
Figure 55 - PNP External Trigger Using External Power
B14N
1
I1A
VCC
~
+
~
19
I1B
EXTERNAL TRIGG ER Vext 30 Vdc max.
V
Signal
Figure 56 - NPN External Trigger Using External Power
56
VB14N
V
6
9
7
-
7
-
V
-
25-pin Name Function
9 Vdc Power Source Inputs
6 I2A Input 2 A (polarity insensitive)
10 I2B Input 2 B (polarity insensitive)
7 GND Power Reference - Inputs
INPUT 2 CONNECTIONS USING VB14N POWER
B14N
Vdc
INPUT DEVICE
V
Signal
VCC ~
I2A
~
10
I2B
GND
Ground
Figure 57 - PNP Input 2 Using VB14N Power
Vdc
9
I2A
6
CC
+
~
10
I2B
GND
INPUT DEVICEVB14N
V
Signal
Ground
Figure 58 - NPN Input 2 Using VB14N Power
INPUT 2 CONNECTIONS USING EXTERNAL POWER
VB14N
Vext 30 Vdc max.
EXTERNAL TRIGGER
V
Signal
VCC
6
I2A
~
+
~
10
I2B
Figure 59 - PNP Input 2 Using External Power
B14N
6
I2A
VCC
~
+
~
10
I2B
EXTERNAL TRIGGER Vext 30 Vdc max.
V
Signal
Figure 60 - NPN Input 2 Using External Power
57
VB14N
V
V
V
V
5.5.1 Code Verifier
If the VB14N is used as a Code Verifier, the verifier code can be configured in software through the
Genius™ configuration program. However it is also possible to use one of the inputs to trigger when
the scanner should store a code read as the verifier code.
The Code Verifier parameter must be enabled, and the configuration parameters to allow correct Co de
Type reading must be saved to the scanner in order to read the verifier code.
When the selected input is activated, the next read code will be stored as the verifier code in the
scanner's non-volatile (Flash) memory.
For more details see the Verifier Parameters in the " Software Configuration Parameter Guide” Help
file.
5.6 OUTPUTS
Two general purpose outputs are available. The following pins are present on the 25-pin connector of
the scanner:
The meaning of the two outputs Output 1 and Output 2 can be defined by the user (No Read, Right,
Wrong, etc.). Refer to the Genius™ Help On Line.
By default, Output 1 is associated with the No Read event, which activates when the code signaled by
the external trigger is not decoded, and Output 2 is associated with the Complete Read event, which
activates when all the selected codes are correctly decoded.
The output signals are fully programmable being determined by the configured Activation/Deactivation
events, Deactivation Timeout or a combination of the two.
B14N
USER INTERFACE
ext 30 Vdc max.
8/11
C
22/12
E
O+
O-
Figure 61 - Open Emitter Output Connections
B14N
C
E
8/11
22/12
O+
USER INTERFACE
O-
ext 30 Vdc max.
Figure 62 - Open Collector Output Connections
VCE max = 30 Vdc
I max = 40 mA continuous; 130 mA pulsed
58
VB14N
V
X
V
y
5.7 USER INTERFACE - HOST
The following table contains the pinout for standard RS232 PC Host interface. For other user interface
types please refer to their own manual.
The following wiring diagram shows a simple test cable including power, external (push-button) trigger
and PC RS232 COM port connections.
B14N
25 - p in D-sub ma le
TX
21
20
RX
GND
7
13
Vdc
25
GND
13
dc
18
I1A
I1B19
9-pin D-sub female
2
3
5
Pow e r Sup pl
Vdc (10 – 30 Vdc)
Power GND
RX
T
GND
PC
Trigger
Test Cable for VB14N
59
VB14N
6 TYPICAL LAYOUTS
The following typical layouts refer to system hardware configurations
to optional hardware configurations within the particular layout.
These layouts also require the correct setup of the software configuration parameters. Complete
software configuration procedures can be found in the Guide To Rapid Configuration in the
Genius™ Help On Line.
6.1 POINT-TO-POINT
In this layout the data is transmitted to the Host on the main serial interface. A Genius™ based Host
Mode programming can be accomplished either through the main interface or the Auxiliary interface.
In Local Echo communication mode, data is transmitted on the RS232 auxiliary interface
independently from the main interface selection.
When On-Line Operating mode is used, the scanner is activated by an External Trigger (photoelectric
sensor) when the object enters its reading zone.
. Dotted lines in the figures refer
24V DC
PG6000
VB14N
Figure 63 – Serial Interface Point-to-Point Layout
3
CBX
1
2
Terminal
Main Serial Interface (RS232 or RS485 Full-Duplex)
Auxiliary Serial Interface (Local Echo) (RS232)
External Trigger (for On-Line Mod e)
Host
60
VB14N
In this layout a single scanner functions as a Slave node on a Fieldbus network. The data is
transmitted to the Host through an accessory Fieldbus interface board installed inside the CBX500
connection box.
Scanner configuration can be accomplished through the Auxiliary interface using the Geniu s ™
configuration program or Genius™ based Host Mode programming.
In Local Echo communication mode, data is transmitted on the RS232 auxiliary interface
independently from the Fieldbus interface selection.
When On-Line Operating mode is used, the scanner is activated by an External Trigger (photoelectric
sensor) when the object enters its reading zone.
Pass-through mode allows two or more devices to be connected to a single external serial interface.
Each VB14N transmits the messages received by the Auxiliary interface onto the Main interface. All
messages will be passed through this chain to the host.
When On-Line Operating mode is used, the scanner is activated by an External Trigger (photoelectric
sensor) when the object enters its reading zone.
Applications can be implemented to connect a device such as a hand-held reader to the Auxiliary port
of the last scanner in the chain for manual code reading capability.
The Main and Auxiliary ports are connected as shown in the figure below:
1 2
1
21
Device#1
Device#2Device#n
3 3
Power
Main Serial Interface (RS232 only)
Host
Figure 65 – Pass-Through Layout
Auxiliary Serial Interface (RS232)
External Trigger (for On-Line Mod e)
3
2
62
VB14N
An alternative Pass-Through layout allows the more efficient ID-NET™ network to be used. This
layout is really an ID-NET Master/Slave Multidata layout which also allows each scanner (Master and
Slaves) to accept input on the Auxiliary interface, for example to connect a device such as a hand-held
reader for manual code reading capability.
Each VB14N transmits its own messages plus any messages received by its Auxiliary interface onto
the ID-NET™ interface. The Master passes all messages to the Host.
When On-Line Operating mode is used, the scanner is activated by an External Trigger (photoelectric
sensor) when the object enters its reading zone.
14
Master
33
Power
Slave#2Slave#n
2
3
2
Host
Main Serial Interface (RS232 or RS485)
Auxiliary Serial Interface (RS232)
External Trigger (for On-Line Mod e)
ID-NET™
Figure 66 – Pass-Through On ID-NET™ Layout
63
VB14N
6.3 ID-NET™
The ID-NET™ connection is used to collect data from several scanners to build a multi-point or a
multi-sided reading system; there can be one master and up to 31 slaves connected together.
The slave scanners are connected together using the ID-NET™ interface. Every slave scan ner must
have a ID-NET™ address in the range 1-31.
The master scanner is also connected to the Host on the RS232/RS485 main serial interface.
For a Master/Slave Synchronized layout the External Trigger signal is unique to the system; there is a
single reading phase and a single message from the master scanner to the Host computer. It is not
necessary to bring the External Trigger signal to all the scanners.
The main, auxiliary, and ID-NET™ interfaces are connected as shown in the figure below.
13
2
Master
Power
Host
Figure 67 – ID-NET™ M/S Synchronized Layout
Slave#1Slave#n
Main Serial Interface (RS232 or RS485)
External Trigger (for On-Line Mod e)
ID-NET ™ (up to 16 devices - practical limit)
64
VB14N
For a Master/Slave Multidata layout each scanner has its own reading phase independent from the
others; each single message is sent from the master scanner to the Host computer.
Master Slave#1 Slave#n
1 4
2
Terminal
3
Power
Host
Main Serial Interface (RS232 or RS485)
Auxiliary Serial Interface (Local Echo) (RS232)
External Trigger (for On-Line Mod e)
ID-NET™ (up to 32 devices, max network extension of 1000 m)
Figure 68 – ID-NET™ M/S Multidata
The auxiliary serial interface of the slave scanners can be used in Local Echo
communication mode to control any single scanner (visualize collected data) o r to
configure it using the Genius™ utility or the Genius™ based Host Mode
programming procedure.
NOTE
The ID-NET™ termination resistor switches must be set to ON only in the first and last
CBX connection box.
65
VB14N
Alternatively, the Master scanner can communicate to the Host as a Slave node on a Fieldbus
network. This requires using an accessory Fieldbus interface board installed inside the CBX500
connection box.
System configuration can be accomplished through the Auxiliary interface of the Master scanner
(internal CBX500 9-pin connector) using the Genius™ configuration program or Geniu s™ ba sed Host
Mode programming.
Power
Master
1
3
Slave#1Slave#n
2
Host
Fieldbus Interface
External Trigger (for On-Line Mod e)
ID-NET™ (up to 16 devices - practical limit)
Figure 69 – ID-NET™ Fieldbus M/S Synchronized Layout
Power
4
Master Slave#1 Slave#n
2
3
1
Host
Fieldbus Interface
Auxiliary Serial Interface (Local Echo) (RS232)
Terminal
External Trigger (for On-Line Mod e)
ID-NET™ (up to 32 devices, max network extension of 1000 m)
66
Figure 70 – ID-NET™ Fieldbus M/S Multidata
VB14N
6.4 RS232 MASTER/SLAVE
This interface is provided for backward compatibility. We recommend using the more
NOTE
The RS232 master/slave connection is used to collect data from several scanners to build either a multipoint or a multi-sided reading system; there can be one master and up to 9 slaves connected together.
The Slave scanners use RS232 only on the main and auxiliary serial interfaces. Each slave VB14N
transmits the messages received by the auxiliary interface onto the main interface. All messages will be
passed through this chain to the Master.
The Master scanner is connected to the Host on the RS232/RS485 main se r i al i nt e rf a c e .
There is a single reading phase and a single message from the master scanner to the Host computer.
Either On-Line
When On-Line Operating mode is used, the external trigger signal is unique to the system,
however it is not necessary to bring the external trigger signal to the Slave scanners.
The main and auxiliary ports are connected as shown in the figure below.
efficient ID-NET™ network for Master/Slave or Multiplexer layouts.
or Serial On-Line Operating modes can be used in this layout.
12
1
21
Power
Master
Host
Slave#1Slave#n
3
Main Serial Interface (RS232 only)
Auxiliary Serial Interface (RS232)
External Trigger (for On-Line Mod e)
Figure 71 – RS232 Master/Slave Layout
67
VB14N
7 READING FEATURES
7.1 ADVANCED CODE BUILDER (ACB)
In addition to linear reading, the A
together two partial reads of it. ACB is not as powerful as Advanced Code Reconstruction due to limits
on tilt angle, speed and Multi-label function; but it is effective in the case of close-to-linear, small
height codes, damaged codes, or poor print quality codes.
ACB is used to read a code label when the scan line does not cross the label along its entire length
(excessive tilt angle).
dvanced Code Builder (ACB) allows code reading by “stitchi ng”
Linear Reading
Linear Reading
ACB Reading
ACB Reading
ACB reads two fragments of a label containing a start or a stop character and a number of digits, and
puts them together to build the complete label.
ACB also has an intrinsic ability to increase the reading percentage of damaged codes as in the
examples below:
ACB Readable
Not ACB Readable
ACB Readable
ACB is disabled by default but can be enabled for the following code types:
Do not use ACB for omni-directional reading stations.
ACB can be activated for each symbology independently from the others.
ACB requires that the code be in movement with respect to the scanner.
ACB requires fixed length barcode reading.
The codes read with ACB enabled must pass in front of the scanner one at a time.
Not valid for ACB
Code concatenation and ACB are not compatible and therefore cannot be enabled simultaneously
on the same code.
For correct operation, ACB requires at least 5 scans for each of the two fragments.
Valid for ACB
7.1.2 Tilt Angle Improvement with ACB
ACB allows barcode reading with an increased tilt angle. The tilt angle depends upon the code aspect
ratio defined as H/L according to the table below:
H
Aspect Ratio
H/L
0.33 18° 30°
0.25 14° 23°
0.125 7° 11°
Max theoretical linear tilt
L
barcode label
angle
Max practical
ACB angle
7.2 LINEAR CODE READING
The number of scans performed on the code by the VB14N and therefore the decoding capability is
influenced by the following parameters:
number of scans per second
code motion speed
label dimensions
scan direction with respect to code motion
At least 5 scans during the code passage should be allowed to ensure a successful rea d.
69
VB14N
7.2.1 Step-Ladder Mode
If scanning is perpendicular to the code motion direction (Figure 72), the number of effective scans
performed by the reade
SN = [(LH/LS) * SS] – 2 Where: SN = number of effective scans LH = label height (in mm)
LS = label movement speed in (mm/s)
SS = number of scans per second
r is given by the following formula:
Direction of code
movement at LS speed
VB14N
Figure 72 - "Step-Ladder" Scanning Mode
For example, the VB14N-300 (500 scans/sec.) for a 25 mm high code moving at 1000 mm/s performs:
[(25/1000) * 500] - 2 = 10 effective scans.
LH
Laser beam
70
VB14N
7.2.2 Picket-Fence Mode
If scanning is parallel to the code motion, (Figure 73), the number of effective scans is given by the
following formula:
((FW-LW)/LS) * SS] -2 Where: SN = number of effective scans
SN = [
FW = reading field width (in mm)
LW = label width (in mm)
LS = label movement speed (in mm/s)
SS = scans per second
Direction of code movement
at LS speed
VB14N
LW
FW
Laser beam
Figure 73 - "Picket-Fence" Scanning Mode
For example, for a 60 mm wide code moving in a point where the reading field is 160 mm wide at a
1500 mm/s speed, the VB14N-300 (500 scans per sec.), performs:
[((160-60)/1500) * 500] - 2 = 31 effective scans
71
VB14N
7.3 PERFORMANCE
The VB14N scanner is available in different versions according to the reading performance.
Version Max Code Resolution Speed
mm (mils) scans/s
VB14N-300 0.20 (8) 500 to 800
VB14N-600 0.35 (14) 600 to 1000
VB14N-300-R 0.20 (8) 500 to 800
VB14N-600-R 0.35 (14) 600 to 1000
Version Reading Distance
VB14N-300 /-R 40 mm (1.6 in) - 300 mm (11.8 in) on 0.50 mm (20 mils) codes
VB14N-600 /-R 190 mm (7.5 in) - 600 mm (23,6 in) on 0.50 mm (20 mils) codes
Refer to the diagrams given in par. 7.4 for further details on the reading features. They are taken on
various resol
notes under the diagrams.
7.3.1 Raster
Raster versions are available. The distance between the top and bottom scan lines is called capture
and is measured from the laser beam output window.
VB14N-300-R: The maximum capture is 18 mm (0.7 in) at 300 mm (11.8 in).
VB14N-600-R: The maximum capture is 35 mm (1.4 in) at 600 mm (23.6 in).
ution sample codes at a 25 C ambient temperature, depending on the conditions in the
72
VB14N
7.4 READING DIAGRAMS
VB14N-300 /-R
NOTE: (0,0) is the center of the laser beam output window.
CONDITIONS
Optic Version = Linear
Code = Interleaved 2/5 or Code 39
PCS = 0.90
"Pitch" angle =
"Skew" angle =
"Tilt" angle =
*Reading Conditions = Standard
*Scan Speed = 500 scans/sec
* Parameter selectable in Genius™
0
15
0
73
VB14N
VB14N-600 /-R
NOTE: (0,0) is the center of the laser beam output window.
CONDITIONS
Optic Version = Linear
Code = Interleaved 2/5 or Code 39
PCS = 0.90
"Pitch" angle =
"Skew" angle =
"Tilt" angle =
0
10
0
*Code Resolution = = High for 0.35mm (14 mils) codes
Standard for 0.50mm (20 mils) codes or greater
*Reading Conditions = Standard
*Scan Speed = 800 scans/sec
* Parameter selectable in Genius™
74
VB14N
0.50 mm
0.35 mm
0.30 mm
VB14N-300 /-R
Reading Distance vs Scanning Speed
Distance
0 13 2 10 6
0 20 40 60 80 100 120 140 160 180 200 220 240 260
457 89
11
280
12
300
(in)
(mm)
Resolution
0.20 mm
Code
500 scans/s
800 scans/s
75
VB14N
8 MAINTENANCE
8.1 CLEANING
Clean the laser beam output window periodically for continued correct operation of the reader.
Dust, dirt, etc. on the window may alter the reading performance.
Repeat the operation frequently in particularly dirty environments.
Use soft material and alcohol to clean the window and avoid any abrasive substances.
Clean the window of the VB14N when the scanner is turned off or, at least, when
the laser beam is deactivated.
WARNING
76
VB14N
9 TROUBLESHOOTING
9.1 GENERAL GUIDELINES
When wiring the device, pay careful attention to the signal name (acronym) on the CBX100/500 spring
clamp connectors (chp. 4). If you are connecting directly to the sca
to the pin number of the signals (chp 5).
If you need information a
files.
ProblemSuggestions
Power On:
the “Power On”/”Ready”
LED is not lit
On line Mode:
TRIGGER LED is not lit
(when external trigger
activates)
On line Mode:
TRIGGER LED is
correctly lit but nothing
happens (no reading
results)
Serial On line Mode:
the reader is not triggered
(no reading results)
On line Mode and Serial
On Line:
Reader doesn’t respond
correctly to the expected
external signals end
Mode:
Mode Button functions
don't work. LEDs light up
but do not allow access to
the functions.
bout a certain
Is power connected?
If using a power adapter, is it connected to a wall outlet?
If using rail power, does rail have power?
If using CBX100, does it have power (check switch and LED)?
Measure voltage either at pin 13 and pin 25 (for 25-pin connector) or at
spring clamp Vdc and GND (for CBX).
Is sensor connected to I1A, I1B spring clamps (for CBX) or to pins 18
and 19 (for 25-pin connector)?
Is power supplied to photo sensor?
Are the photo sensor LEDS (if any) working correctly?
Is the sensor/reflector system aligned?
Is the software configuration consistent with the application condition
(operating mode etc.)?
In the Genius™ program select the Operating Mode branch and check
for related parameters.
In the Genius™ program select the Operating Mode branch and check
if Serial On Line is selected in the On Line Options.
Are the Start – Stop characters correctly assigned?
Is the serial trigger source correctly connected and configured?
In the Genius™ program select the Operating Mode branch and check
the Reading Phase Timeout parameterization.
The Mode Button functions don't work if the scanner motor or laser are
turned off. Check if the motor or laser are turned off through the
following parameters:
Beam Shutter = enabled
Scan Speed = Motor Off
Energy Saving>Serial Motor Off has been sent
reader parameter you can refer to the Genius™ program help
TROUBLESHOOTING GUIDE
nner 25
-pin connector pay attention
77
VB14N
TROUBLESHOOTING GUIDE
ProblemSuggestions
Reading:
Not possible to read the
target barcode (always
returns No Read) or the
Auto Setup procedure
Fails.
Communication:
Device is not transmitting
anything to the host
Communication:
Data transferred to the
host are incorrect,
corrupted or incomplete
Communication:
Always returns the
Reader Failure Character
(<BEL> char as default)
How do I obtain my
units’ serial numbers?
Check synchronization of reading pulse with object to read:
Is the scan line correctly positioned?
Place barcode in the center of scan line and run Test mode (selectable
by Genius™ as an Operating Mode). If you still have trouble, check the
following:
Is the reading distance within that allowed (see reading
diagrams)?
Is the Tilt angle too large?
Is the Skew angle less than 10° (direct reflection)?
Choose the Code Definition branch and enable different Code
Symbologies (except Pharmacode). Length = Min and Max
(variable).
Is the Bar Code quality sufficient?
Is the serial cable connected?
Is the correct wiring respected?
Are serial host settings equivalent to the serial device setting?
If using CBX, be sure the RS485 termination switch is positioned to
OFF.
In the Genius™ program select the Data Communication
Settings/Data Format/Standard Parameters branch and check the
Header, Separators, and Terminator values
Also check the Code Field Length and Fill Character values.
Are the COM port parameters correctly assigned?
Note the exact model and Serial Number of the device.
Contact Pepperl+Fuchs.
The device’s serial number is printed on a label that is affixed to the
body of the reader.
Serial numbers consist of 9 characters: one letter, 2 numbers, and
another letter followed by 5 numbers.
78
VB14N
10 TECHNICAL FEATURES
ELECTRICAL FEATURES VB14N-300 /-R VB14N-600 /-R
Input Power
Supply Voltage 10 to 30 Vdc
Power consumption max. 0.3 to 0.1 A; 3 W 0.5 to 0.17 A; 5 W
Serial Interfaces
Main Serial Interface Sw programmable: RS232; RS485 FD and HD
Baudrate 1200 - 115200
Auxiliary RS232
Baudrate 1200 - 115200
ID-NET™ RS485 Half-duplex
Baudrate Up to 1 MBaud
Inputs
Input 1 (External Trigger), Input 2 Optocoupled, polarity insensitive
Voltage 10 to 30 Vdc
Current Consumption 12 mA max.
Minimum Pulse Duration 5 ms.
Outputs
Output 1, Output 2 Optocoupled
VCE 30 Vdc max.
Collector Current 40 mA continuous max.; 130 mA pulsed max.
V
CE saturation
Power Dissipation
OPTICAL FEATURES
Light Source Semiconductor laser diode
Wave Length In the range 630 to 680 nm
Safety Class Class 2 - EN 60825-1; Class II - CDRH
READING FEATURES (Note 1)
Scan Rate (software program.) (500 to 800 scans/sec) (600 to 1000 scans/sec)
Aperture Angle 50°
Maximum Reading Distance
Maximum Resolution
ENVIRONMENTAL FEATURES
Operating Temperature (Note 2)
Storage Temperature
Humidity max. 90% non condensing
Vibration Resistance 14 mm @ 2 to 10 Hz; 1.5 mm @ 13 to 55 Hz;
EN 60068-2-6 2 g @ 70 to 200 Hz; 2 hours on each axis
Bump Resistance 30g; 6 ms;
EN 60068-2-29 5000 shocks on each axis
Shock Resistance 30g; 11 ms;
EN 60068-2-27 3 shocks on each axis
Protection Class – EN 60529 IP65
PHYSICAL FEATURES
Mechanical Dimensions 68 x 84 x 34 mm (2.7 x 3.3 x 1.3 in)
Weight 330 g (11.6 oz.)
1V max. at 10 mA
Note 1: Further details given in par. 7.3.
Note 2: If the reader is used in high temperature environments (over 40 C), use of the Beam Shutter is advised (see the
Genius™ configuration program) and/or a thermally conductive support (such as the metal bracket provided).
* Code 39 (Standard and Full ASCII) * EAN 128
* Codabar ISBT 128
*ABC Codabar Pharmacode
Plessey
*ACB Readable.
Code Selection
Decoding Safety
Headers and Terminators
Operating Modes
Configuration Modes
Genius™ based Host Mode Programming
Special Functions
Parameter Storage
USER INTERFACE
LED Indicators Read y, Good, Trigger, Com, Status, Power On
Multi-function Key Mode Button
up to ten different symbologies during one reading phase
can enable multiple good reads of same code
Up to 128-byte header string
Up to 128-byte terminator string
On-Line, Serial On-Line, Verifier,
Automatic, Continuous, Test
Mode Button Functions
Genius™ utility program
Code Verifier
ACB (Advanced Code Builder)
Motor Off and SW_Speed Control
Programmable Diagnostic and Statistic Messages
Non-volatile internal Flash
80
VB14N
GLOSSARY
ACB (Advanced Code Builder)
Advanced Code Builder (ACB) allows code reading by “stitching” together two partial reads of it. ACB
is effective in reading codes positioned close-to-linear, small height codes, damaged codes, or poor
print quality codes. See par. 7.1.
ture
Aper
Term used on the required CDRH warning labels to describe the laser exit window.
Barcode
A pattern of variable-width bars and spaces which represents numeric or alphanumeric data in
machine-readable form. The general format of a barcode symbol consists of a leading margin, start
character, data or message character, check character (if any), stop character, and trailing margin.
Within this framework, each recognizable symbology uses its own unique format.
Barcode Label
A label that carries a barcode and can be affixed to an article.
Baud Rate
A unit used to measure communications speed or data transfer rate.
CDRH (Center for Devices and Radiological Health)
This organization (a service of the Food and Drug Administration) is responsible for the safety
regulations governing acceptable limitations on electronic radiation from laser devices. Datalogic
devices are in compliance with the CDRH regulations.
Code Positioning
Variation in code placement that affects the ability of a scanner to read a code. The terms Pitch, Skew,
and Tilt deal with the angular variations of code positioning in the X, Y and Z axes. See pars. 3.1.1
and 3.2. Variations in code placement affect the pul
Pulse width is defined as a change from the leading edge of a bar or space to the trailing edge of a bar
or space over time. Pulse width is also referred to as a transition. Tilt, pitch, and skew impact the pulse
width of the code.
Decode
The process of translating a barcode into data characters using a specific set of rules for each
symbology.
Decoder
As part of a barcode reading system, the electronic package which receives the signals from the
scanner, performs the algorithm to interpret the signals into meaningful data and provides the interface
to other devices. The decoder is normally integrated into the scanner.
EAN
European Article Number System. The international standard barcode for retail food packages.
FLASH
An on-board non-volatile memory chip.
Full Duplex
Simultaneous, two-way, independent transmission in both directions.
Half Duplex
Transmission in either direction, but not simultaneously.
se width and therefore the decoding of the code.
81
VB14N
Host
A computer that serves other terminals in a network, providing services such as network control,
database access, special programs, supervisory programs, or programming languages.
Interface
A shared boundary defined by common physical interconnection characteristics, signal characteristics
and meanings of interchanged signals.
LED (Light Emitting Diode)
A low power electronic device that can serve as a visible or near infrared light source when voltage is
applied continuously or in pulses. It is commonly used as an indicator light and uses less power than
an incandescent light bulb but more than a Liquid Crystal Display (LCD). LEDs have extremely long
lifetimes when properly operated.
Multidrop Line
A single communications circuit that interconnects many stations, each of which contains terminal
devices. See RS485.
Parameter
A value that you specify to a program. Typically parameters are set to configure a device to have
particular operating characteristics.
Picket-Fence Orientation
When the barcode’s bars are positioned vertically on the product, causing them to appear as a picket
fence. The first bar will enter the scan window first. See par. 7.2.2.
Pitch
Rotation of a
characters. See pars. 3.1.1 and 3.2.
Position
The po
Protocol
A formal set of conventions governing the formatting and relative timing of message exchange
between two communicating systems.
Raster
The process of projecting the laser beam at varied angles spaced evenly from each other. Typically,
the mirrored rotor surfaces are angled to create multiple scan lines instead of a si ngle beam.
Resolution
The narrowest element dimension, which can be distinguished by a particular reading devi ce o r
printed with a particular device or method.
RS232
Interface between data terminal equipment and data communication equipment employing serial
binary data interchange.
RS485
Interface that specifies the electrical characteristics of generators and receivers for use in balanced
digital multipoint systems such as on a Multidrop line.
Scanner
A device that examines a printed pattern (barcode) and either passes the uninterpreted data to a
decoder or decodes the data and passes it onto the Host system.
Serial Port
An I/O port used to connect a scanner to your computer, identifiable by a 9-pin or 25-pin connector.
code pattern about the X-axis. The normal distance between center line or adjacent
sition of a scanner or light source in relation to the target of a receiving element.
82
VB14N
Signal
An impulse or fluctuating electrical quantity (i.e.: a voltage or current) the variations of which represent
changes in information.
Skew
Rotation about the Y-axis. Rotational deviation from correct horizontal and vertical orientation; may
apply to single character, line or entire encoded item. See pars. 3.1.1 and 3.2.
Step-Ladd
When the barcode’s bars are positioned horizontally on the product, causing them to appear as a
ladder. The ends of all bars will enter the scan window first. See par. 7.2.1.
mbol
Sy
A combination of characters including start/stop and checksum characters, as required, that form a
complete scannable barcode.
Tilt
Rotation around the Z axis. Used to describe the position of the barcode with respect to the laser scan
line. See pars. 3.1.1 and 3.2.
Trigger Sign
A signal, typically provided by a photoelectric sensor or proximity switch, which informs the scanner of
the presence of an object within its reading zone.
UPC
Acronym for Universal Product Code. The standard barcode type for retail food packaging in the
United States.
Visible Laser Diode
A light source used in scanners to illuminate the barcode symbol. Generates visible red light at
wavelengths between 630 and 680 nm.