This manual contains installation, operation and maintenance instructions for the
Valor™ 4000W Series. Please read the manual completely before using the scale.
1.1 Safety Precautions
Please follow these safety precautions:
Verify that the AC Adapter input voltage matches the local AC power supply.
Do not drop loads on the platform.
Do not place the scale upside down on the platform.
Disconnect the scale from power supply when cleaning.
Operate the scale only under ambient conditions specified in these
instructions.
Service should be performed only by authorized personnel.
Only use weights within the scale’s capacity as specified in these instructions.
Do not operate the scale in hostile environments.
Do not carry the scale by the pan or sub-platform. Use the handholds on the
side of the scale housing.
2. INSTALLATION
2.1 Package Contents
2.2 Installing Components
Place the stainless steel pan onto the weighing platform before turning the scale on.
Figure 2-1. Install the Stainless Steel Pan onto the weighing platform
EN-2 Valor™ 4000W Series
Level Bubble
Adjustable Feet
2.3 Selecting the Location
Use the scale on a clean, firm and flat surface. Avoid locations with excessive air
current, vibrations, heat sources, or rapid temperature changes.
2.4 Leveling the Equipment
Adjust the leveling feet so the bubble is centered in the circle. Be sure the
equipment is level each time its location is changed.
Figure 2-2. Level Indicator
2.5 Power
The AC Adapter is used to power the scale when battery power is not needed.
First connect the AC Adapter plug to the scale input jack located at the bottom of
the scale according to the description below.
Then connect the AC adapter to the proper AC supply.
Figure 2-3. Slide to plug in
Figure 2-4. Connecting the AC adapter to AC supply
Caution: The scale can only be used in a dry environment when
powered by the AC adapter.
Valor™ 4000W Series EN-3
Symbol
Charge Level
Battery in use:
Symbol displayed
2.5.1 Battery Power:
The scale can be used on AC power immediately. Allow the battery to charge
for 12 hours before using the scale on battery power. The scale will
automatically switch to battery operation if there is a power failure or the power
cord is removed. With AC power, the scale is constantly charging, so the
battery charge indicator (see item 11 in table 3-2) will remain lit. The scale can
be operated during charging, and the battery is protected against overcharging.
For maximum operating time, the battery should be charged at room
temperature.
During battery operation, the battery symbol indicates the battery charging
status. When charging, the symbol will blink slowly and when fully charged the
symbol will stop blinking.
TABLE 2-1
Notes:
When battery symbol blinks fast, approximately 30 minutes working time is left.
When [Lo.bAt] is displayed, the scale will shut off.
Charging the scale must be performed in a dry environment.
CAUTION: Battery is to be replaced only by an authorized
OHAUS service dealer. Risk of explosion can occur if the
rechargeable battery is replaced with the wrong type or if it is
not properly connected. Dispose of the lead acid battery
according to local laws and regulations.
EN-4 Valor™ 4000W Series
Button
Functions
Short Press1 (when on): Sets display to zero
Short Press (when off): Turns scale on
Long Press2 (when on): Turns the scale off
Short Press (in Menu): Selects / accepts displayed setting
Short Press: Accumulates the weight or displays the
accumulated information at 0 load.
Long Press: Toggles through active Units
Short Press (in Menu): Toggles through available settings
Short Press: Initiates an application mode specific response
Long Press: Selects active Mode
Short Press (in Menu): returns to previous settings
Short Press: Enter / clear a Tare value
Long Press: Enters User Menu
Short Press (in Menu): Quickly exit User Menu
IR Sensor3 can be programmed to act as “touchless” button.
See the User Menu section 4.3 for the available settings.
3. OPERATION
3.1 Controls
Figure 3-1. Front Control Panel Valor 4000W with display
TABLE 3-1
Notes:
1
Short Press: Press less than 2.5 seconds.
2
Long Press: Press and hold for more than 2.5 seconds.
3
The IR sensor can be activated by a hand or other object that is placed at a specified height (see
section 4.3) from the sensor. The sensor activation distance will vary based on the reflective nature
of the object. If unwanted activations occur due to unique situations the sensor can be turned off.
Valor™ 4000W Series EN-5
Item
Description
Item
Description
1
NET symbol
9
Calibration / Menu Mode symbol
2
Center of Zero symbol
10
Dynamic symbol*
3
Negative symbol
11
Battery charge symbol
4
Stable weight symbol
12
Pound, Ounce, Pound:Ounce symbol
5
Preset Tare symbol*
13
Gram, Kilogram symbols
6
Pointer symbols
14
Pieces symbol*
7
IR Sensor symbol
15
Percent symbol
8
Accumulation symbol
(Red) Loads > Upper limit
(Green) Loads ≥ Lower limit and ≤ Upper limit
(Yellow) Loads < Lower limit
1
1
IR Sensor
Figure 3-2. Valor 4000W Display
1
2
1
3
1
4
1
5
1
6
1
7
1
8
1
9
1
10
1
11
1
12
1
13
1
14
1
15
LED Indicators
Display
TABLE 3-2 Display Symbols
Note: * Not Used
Figure 3-3. Valor 4000W Rear display.
The colored LED indicators on the right side of the control panels are used in
Checkweigh mode (section 3.6) and will light up according to the following rules:
EN-6 Valor™ 4000W Series
Required Span Calibration Mass (sold separately)
Max
Mass
1
Max
Mass
1
1500g
1.5kg / 3lb
6000g
6kg / 15lb
3000g
3kg / 6lb
15000g
15kg / 30lb
3.2 Turning Scale On/Off
To turn the scale on, press the On/Zero Off button. The scale performs a display
test, momentarily displays the software version, and then enters the active weighing
mode.
To turn the scale off, press and hold the On/Zero Off button until OFF is displayed.
3.3 Initial Calibration
When the scale is first installed it should be calibrated to ensure accurate results.
Before performing the calibration, be sure to have the appropriate calibration
weights as listed in table 3-3.
Press and hold Menu until [mMeNU] (Menu) is displayed. When the button is released,
the display will display [C.A.L]. Press Yes to accept, [SpaN] will then be shown. Press
Yes again to begin the span calibration. [--C--] blinks while zero reading is stored.
Next, the display shows the calibration weight value. Place the specified calibration
mass on the pan and press Yes. [--C--] blinks while the reading is stored. The scale
returns to the previous application mode and is ready for use.
The message [Err 3.0 CAl] will be displayed if these calibration steps were not
followed or if the wrong weight was used.
The calibration process can be aborted by turning the scale off.
TABLE 3-3
1
Note:
Pound masses are used when calibrating in the lb unit.
3.4 Weigh Mode
1. Press and hold Mode until [wJeIGH] (Weigh) is displayed.
2. If required, place an empty container on the pan and press Tare.
3. Add sample to the pan or container. The display shows the weight of the
sample.
Valor™ 4000W Series EN-7
3.5 Percent Mode
This mode measures the weight of a sample as a percentage of a reference weight.
1. Press and hold Mode until [PerCNt] is displayed. [Clr.ref] (clear reference) will
then display.
Notes: Press Function to view the current reference weight.
If there is no reference data stored, [Set.ref] is displayed. Press Yes to
set reference data.
2. Press No to use the stored reference weight and proceed to step 6.
3. If required, place an empty container on the pan and press Tare. Press Yes to
establish a new reference. Scale will now display [PUt.ref].
4. Add the desired reference material to the container. Press Yes to store the
reference weight. The display shows 100%.
5. Remove the reference material. If required, place an empty container on the
pan and press Tare.
6. Add the sample material. The display shows the percentage of the sample
compared to reference weight.
7. To clear the stored reference data press and hold Mode until [PerCNt] is
displayed. Press Yes when [Clr.ref] is displayed.
3.6 Checkweigh Mode
This mode sets low and high weight limits for portion control processes.
1. Press and hold Mode until [CHeCk] (Check) is displayed. [CLr.ref] (clear
references) will then display.
Notes: Press Function to view the low and high reference weight limits.
If there is no reference data stored, [Set.ref] is displayed. Press Yes to
set reference data.
2. Press No to use the stored reference weight limits and proceed to step 5.
3. Press Yes to establish new reference values. The scale will then display [Set.
LO]. Press Yes to view the “Low” limit value. Press Yes to accept or No to edit
the “Low” limit value. The stored value then displays with the first digit
highlighted [000.000 kg]. Repeatedly press No until the desired number
appears. Press Yes to accept and highlight the next digit. Repeat until all the
digits are correct. Press Yesto accept the “low” limit value, [Set. HI] will be
displayed.
4. Repeat the same procedure to accept or edit the “high” value.
5. If required, place an empty container on the pan and press Tare. Place
sample material on the pan or in the container. If the sample weight is under
the target weight range, the yellow LED will light. If the sample is within the
target weight range, the green LED will light. If the sample is over the target
weight range, the red LED will light.
6. To clear the stored reference values press and hold Mode until [CHeCk] is
displayed. Press Yes when [Clr.ref] is displayed.
EN-8 Valor™ 4000W Series
In weighing mode; sub menu ACCUmM setting, select mMaNU:
If required, place an empty container on the pan and press
Tare.
Step i. Put the weight (0.04 kg) on the pan and press M+
key. The Σ icon indicator will keep flashing until
the weight is removed.
Step ii. Remove the weight from the pan. Put another
weight (0.03 kg) and press M+ key:
Step iii. Remove the weight from the pan:
Step iv. Press M+ key to display the statistical information:
Step v. If required, follow step 3 above to clear the
accumulation data.
N 2
tOtal
0.070 kg
mMIN
0.030 kg
mMax
0.040 kg
mMaNU
*
0.040 kg
*
0.030 kg
*
0.000 kg
3.7 Accumulation
Accumulation works together with each application mode. This function allows the
user to store the total of a series of weight measurements.
1. Press the M+ key to add the weight to the accumulation data. The Σ icon will
keep flashing until the weight is removed.
Note: When Accumulate is set to AUTO in the setup menu, it is not
necessary to press the M+ key.
2. When the pan is cleared, press the M+ key to display the statistical information
of the accumulation data.
3. To clear the accumulation data press the Tare key when the display is
showing the statistical information of the accumulation data. The display
shows [Clr.aCC]. Press the Yes key to clear the stored data and return to
current mode.
Notes:
Only stable weights are stored to accumulation total.
To prevent the same load from being added to the total multiple times, the load on
the pan must be returned to 0 before the next load can be added.
When Legal for Trade is OFF, the display must return to 0 gross, 0 net or a negative
net value; when Legal for Trade is ON, the display must return to 0 gross.
Otherwise, the Σ icon will continue flashing.
Gross loads and net loads cannot be added to the same total. If the first load is a
gross weight, future loads must also be gross weights. If the first load is a net
weight, future loads must also be net weights.
The max of accumulation times is 9999.
Changing the mode will clear the accumulation data.
Accumulation Example:
Menu:
C.A.L
S.E.t.U.P
r.E.A.d
M.O.d.E
U.n.i.t
L.O.C.k
E.n.d
Menu
Items:
Span
Lin
GEO
End
Reset
Pwr.Un
A.Tare
Ir.Func
Ir.Adj
Accum
End
Reset
Stable
Filter
AZT
Light
Sleep
A.Off
End
Reset
Percnt
Check
End
kg
g
oz
lb
lb:oz
End
Reset
L.Cal
L.Setup
L.Read
L.Mode
L.Unit
End
Valor™ 4000W Series EN-9
4. MENU SETTINGS
The User Menu allows the customizing of scale settings.
4.1 Menu Navigation
User Menu:
Notes:
Some Units/Modes may not be available in all models.
When LEGAL FOR TRADE is set to ON (see section 4.9), the menu settings are
affected.
To Enter the Menu Mode
Press and hold Menu until [mMeNU] (Menu) is displayed. When released the first submenu [C.a.l] (Cal) will be shown.
Press Yes to enter the displayed sub-menu or press No to advance to the next.
Selecting a sub-menu will display the first menu item. Press Yes to view the menu
item setting or press No to move to the next menu item. When viewing the setting,
press Yes to accept the setting, or press No to change the setting. When [End] is
displayed, press Yes to return to the sub-menu selections or No to return to the first
item in the current menu. Bold indicates factory default setting.
Note:The Calibration / Menu Mode indicator is displayed when in the Menu Mode.
4.2 Cal Menu
Enter this menu to perform calibrations.
Span[SpaN] (yes, no)
Initiates a span calibration procedure (zero and span).
Lin[LIN] (yes, no)
Initiates a linearity calibration procedure (zero, mid-point and span).
GEO[GEO]
Geographical Adjustment Factor (GEO) is used to adjust the calibration
based on the current location. Settings from 0 to 31 are available with 12
being the default. Refer to table 4-1 to determine the GEO factor that
corresponds to your location.
End Cal [End]
Advance to the next menu or return to the top of the current menu.
EN-10 Valor™ 4000W Series
4.3 Setup Menu
Enter this menu to set scale parameters.
Reset[reset] (no, yes)
Reset the Setup menu to factory defaults.
Power on unit[pwWr.UN] (auto, kg, g, lb, oz, lb:oz)
Set the unit of measure displayed at startup
Auto Tare[A.tArE] (off, on, on-acc)
Set the automatic tare functionality. If “on” is selected the first stable gross weight is tared. If “on-acc” is selected, stable gross loads within the accept
limits are tared (in Checkweigh mode)
IR Function [Ir.FUNC] (off, tare)
Set the IR sensor functionality.
IR Adj [Ir.adj] (hi, low)
Set the detecting level for the IR sensor.
(For reference: Hi: ~ 100 mm / 4 inches; Low: ~ 50mm / 2 inches)
Accumulation[aCCUmM] (off, auto, manu)
Set the accumulation functionality.
End Setup [End]
Advance to the next menu or return to the top of the current menu.
4.4 Read Menu
Enter this menu to set user preferences.
Reset [reset] (no, yes)
Reset the Read menu to factory defaults.
Stable Range[Stable] (0.5, 1, 2, 5)
Set the amount the reading can vary while the stability symbol remains on.
Filter[FILtEr] (low, medium, high)
Set the amount of signal filtering.
Auto-Zero Tracking[AZt] (off, 0.5, 1, 3)
Set the automatic zero tracking functionality.
Light [LIGHt] (hi, med, low)
Sets backlight functionality.
Sleep[SLEEP] (off, on)
Set the display shut off time.
Auto off[A.OFF] (off, 1, 5, 10)
Set the automatic shut off functionality.
End Readout[End]
Advance to the next menu or return to the top of the current menu.
Valor™ 4000W Series EN-11
4.5 Mode Menu
Enter this menu to activate modes so they will be available for use with the Mode
button. Weigh mode is always active by default.
Reset[reset] (no, yes)
Reset the Mode menu to factory defaults.
Percent[PerCNt] (off, on)
Set the status.
Check[CHECk] (off, on)
Set the sub-mode.
End Mode[End]
Advance to the next menu or return to the top of the current menu.
4.6 Unit Menu
Enter this menu to activate units so they will be accessible with the Units button.
The units in the menu must be turned “on” to be active.
Note: Available units vary by model and local regulations.
4.7 Lock Menu
The Lock Menu is a software controlled option which can lock Menu settings to
prevent tampering.
Reset[reset] (no, yes)
Reset the Lock menu to factory defaults.
Lock Cal [l.CAL] (off, on)
Set the status.
Lock Setup[l.SEtUP] (off, on)
Set the status.
Lock Read[l.rEAD] (off, on)
Set the status.
Lock Mode[l.mMOdE] (off, on)
Set the status.
Lock Unit[l.UNIt] (off, on)
Set the status.
End Lock[End]
Advance to the next menu or return to the top of the current menu.
4.8 End Menu
Press ‘Yes’ to advance to the Calibration menu. Press ‘No’ to exit the menu and
When the scale is used in trade or a legally controlled application it must be set up,
verified and sealed in accordance with local weights and measures regulations. It is
the responsibility of the purchaser to ensure that all pertinent legal requirements are
met.
4.9.1 Capacity Label
A label showing the capacity and readability of the scale must be installed near
each display. If the Capacity Labels were installed prior to delivery, no further action
is needed. If the Capacity Labels were not installed, they have been placed in the
packaging material. Affix the labels above the displays as shown in Figure 4-1.
Note: The Capacity Labels will be destroyed upon removal, so only attempt to
install them once.
Figure 4-1. Front and rear display capacity label locations
4.9.2 Settings
Before verification and sealing, perform the following steps:
1. Set Legal For Trade to OFF.
2. Verify that the menu settings meet the local weights and measures regulations.
3. Perform a calibration as explained in Section 3.3.
4. Set Legal For Trade to ON.
To change the scale’s Legal For Trade (LFT) setting, follow this procedure:
Caution: When accessing the bottom of the scale, avoid placing the scale
upside down on the pan or sub-platform. Place the scale on its side.
1. Turn the scale OFF.
2. Remove the Security Cover under the scale to access
the two pins located in a slot (see figure 4-2). Short these
pins (a slotted screwdriver may be used), while powering
the scale ON. The scale will perform the start-up
procedure, and then the display will show the current
status [lft OFF] (LFT OFF, unlocked) or [lft ON] (LFT
ON, locked); press the ‘No’ key to change this setting, or
press the ‘Yes’ key to confirm it.
Note: When Legal For Trade is set to On, the menu settings are affected as follows:
Calibration (C.A.L) menu is not accessible
IR Function, Filter and Unit menus are locked at the current settings.
Stable Range setting is locked at 1d
Auto-Zero Tracking setting is locked at 0.5d
3. Replace the Security Cover.
EN-14 Valor™ 4000W Series
Wire Seal
Paper Seal
Security Cover
4.9.3 Verification and Sealing
The local weights and measures official or authorized service agent must perform
the verification procedure.
4.9.3.1 Physical Seals
For jurisdictions that use the physical sealing method, the local weights and
measures official or authorized service agent must apply a security seal to prevent
tampering with the settings. Refer to the illustrations below for sealing methods.
Unlocked Locked
Figure 4-3. Sealing
4.9.3.2 Audit Trail Seal (USA and Canada only)
For jurisdictions that use the audit trail sealing method, the local weights and
measures official or authorized service agent must record the current configuration
and calibration event counter values at the time of sealing. These values will be
compared to values found during a future inspection.
Note: A change to an event counter value is equivalent to breaking a physical seal.
The audit trail uses two event counters to record changes to configuration and
calibration settings.
The configuration event counter (CFG) will index by 1 under the following
conditions:
- When the LFT setting is changed from ON to OFF.
- When exiting the menu if one or more of the following menu settings
are changed: Stable Range, Auto Zero Tracking (AZT), Filter, IR
Function and Units (kg, g, oz, lb or lb:oz).
The calibration event counter (CAL) will index by 1 when exiting the menu
if a Span Calibration or GEO setting change is made.
Note: The counter only indexes once, even if several settings are changed.
The event counters can be viewed by pressing and holding the MENU button.
While the button is held, the display will show MENU followed by Audit.
Release the button when Audit is displayed to view the audit trail information.
Valor™ 4000W Series EN-15
6
3
1
2
4
5
The audit trail information is displayed in the format CFGxxx and CALxxx.
The scale then returns to normal operation.
5. MAINTENANCE
5.1 Cleaning
The housing may be cleaned with a cloth dampened with a mild detergent if
necessary. Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean
the housing or control panels.
5.2 Cleaning the Plastic Pan
5.2.1 Uninstalling and cleaning the plastic pan
Please follow the steps below to clean the plastic pan:
1. Remove the stainless steel pan
2. Remove the four thumb screws holding the battery cover.
3. Unplug the two battery clips and remove the battery.
4. Using a Phillips screwdriver, remove the two screws located at the bottom of the
battery compartment.
5. Remove the plastic pan.
6. Clean the plastic pan.
Figure 5-1. Pan cleaning
5.2.2 Reinstalling the pan after cleaning
Please follow the steps below to reinstall the plastic pan (see also the steps in figure
5-1, in reversed order):
1. Using a Phillips screwdriver, attach the plastic pan to the housing with the
two screws.
2. Place the battery in the battery compartment and attach the two battery clips.
Attach the red wire to the positive (red) battery terminal and the black wire to
the negative (black) battery terminal.
3. Attach the battery cover with the four thumb screws.
4. Place the scale upright and install the stainless steel pan.
Caution: Risk of explosion can occur if the battery is not properly
connected.
EN-16 Valor™ 4000W Series
Symptom
Possible Cause
Remedy
Cannot turn on
No power to scale
Battery discharged
Verify connections, power source
and battery charge status.
Poor accuracy
Improper calibration
Unstable environment
Perform calibration
Move scale to suitable location
Cannot calibrate
Unstable environment
Incorrect calibration weight
Move the scale to suitable location
Use correct calibration weight
Cannot access mode
Mode not enabled
Enter menu and enable mode
Cannot access unit
Unit not enabled
Enter menu and enable unit
LO ref
Reference weight is too low
Increase reference weight.
Err 3.0 Cal
Incorrect calibration weight
See table 3-3 for correct weights
Err 8.1 ˉLOadˉ
Power on zero range exceeded
Clear pan
Err 8.2 _LOad_
Power on zero under range
Install pan
Err 8.3 ˉLOadˉ
Overload
Load exceeds scale maximum
capacity
Err 8.4 _LOad_
Under load
Reading below min. range - Reinstall pan.
Err 8.5 ˉtareˉ
Tare out of range
Tare value exceeds maximum.
Err 9 Data
Internal data error.
Contact an authorized service agent
Err13 mMEmM
Fail to write EEPROM.
Contact an authorized service agent
Err 53 C.SUmM
Invalid checksum data
Contact an authorized service agent
Lo.Bat
Battery is discharged
Connect the power and charge the
battery
NO.ACC
Gross and net weights cannot be
accumulated together
Accumulate only gross or net
weights.
Battery fails to charge
fully
Battery is defective
Have battery replaced by OHAUS
authorized service dealer.
5.3 Troubleshooting
The following table lists common problems and possible causes and remedies.
If the problem persists, contact OHAUS or your authorized dealer.
TABLE 5-1
5.4 Service Information
If the troubleshooting section does not resolve or describe your problem, contact
your authorized OHAUS service agent. Please visit our web site, www.ohaus.com
to locate the OHAUS office nearest you. An OHAUS product service specialist will
be available to provide assistance.
6. TECHNICAL DATA
The technical data is valid under the following ambient conditions:
Indoor use only
Operating temperature:-10ºC (14ºF) to 40ºC (104ºF)
Relative humidity: 10% to 90% relative humidity, non-condensing
Altitude: Up to 2000 m
Power: AC Adapter (supplied) - 12 VDC 0.84 A output, internal rechargeable sealed
lead-acid battery
Mains supply voltage fluctuations: up to ±10% of the nominal voltage
Installation Category: II
Pollution Degree: 2
Valor™ 4000W Series EN-17
MODEL
V41PWE1501T
V41XWE1501T
V41PWE3T
V41XWE3T
V41PWE6T
V41XWE6T
V41PWE15T
V41XWE15T
Capacity x Readability
(Max x d non-approved)
1.5 kg x 0.0002 kg
1500 g x 0.2 g
3 lb x 0.0005 lb
48 oz x 0.01 oz
3 kg x 0.0005 kg
3000 g x 0.5 g
6 lb x 0.001 lb
96 oz x 0.02 oz
6 kg x 0.001 kg
6000 g x 1 g
15 lb x 0.002 lb
240 oz x 0.05 oz
15 kg x 0.002 kg
15000 g x 2 g
30 lb x 0.005 lb
480 oz x 0.1 oz
Maximum Displayed Resolution
7500
6000
6000
7500
Capacity X Readability
(Max x e approved)
1.5 kg x 0.0005 kg
1500 g x 0.5 g
3 lb x 0.001 lb
48 oz x 0.02 oz
3 kg x 0.001 kg
3000 g x 1 g
6 lb x 0.002 lb
96 oz x 0.05 oz
6 kg x 0.002 kg
6000 g x 2 g
15 lb x 0.005 lb
240 oz x 0.1 oz
15 kg x 0.005 kg
15000 g x 5 g
30 lb x 0.01 lb
480 oz x 0.2 oz
Approved Resolution
3000
3000
3000
3000
Repeatability
0.0005 kg
0.001 kg
0.002 kg
0.005 kg
Linearity
± 0.0005kg
± 0.001kg
± 0.002kg
± 0.005kg
Weighing Units*
Non-Approved models: g, kg, lb, oz, lb:oz
EC and OIML Approved models: g, kg
Measurement Canada and NTEP Approved models: g, kg, lb, oz
Tare Range
To capacity by subtraction
Stabilization Time
≤ 0.5 seconds
Safe Overload Protection
150 % of scale capacity
Weight Display
2 x Red LED (front and rear)
6-digit 7-segment, 20.5 mm / 0.8 in characters
Keyboard
Four buttons
Application Modes
Weighing, Percent, Check Weighing (each with Accumulate function)
Battery Operating Time (at 20º C)
Typically 50 hours with 12-hour full charge
Construction
V41PW: ABS housing with 304 Stainless Steel Platform
V41XW: ABS bottom housing with 304 stainless steel top housing and platform
Ingress Protection
IPX8
Pan Dimensions
190 x 242 mm / 7.5 x 9.5 in
Approval Class
III
Net Weight
V41PW: 3.0 kg / 6.6 lb; V41XW: 3.9 kg / 8.6 lb
Shipping Weight
V41PW: 4.0 kg / 8.8 lb; V41XW: 4.9 kg / 10.8 lb
Shipping Dimensions
410 x 370 x 220 mm / 16.1 x 14.6 x 8.7 in
V41PW
A
256 mm / 10.1 in.
B
280 mm / 11.0 in.
C
121 mm / 4.8 in.
V41XW
A
256 mm / 10.1 in.
B
288 mm / 11.3 in.
C
124 mm / 4.9 in.
A
C
B
TABLE 6-2
Figure 6-1. Dimensions
6.1 Specifications
TABLE 6-1
Note: Turn off lb:oz unit with LFT ON
6.2 Drawings and Dimensions
EN-18 Valor™ 4000W Series
Mark
Standard
This product conforms to the EMC Directive 2004/108/EC, the Low
Voltage Directive 2006/95/EC and the Non-Automatic Weighing
Instrument Directive 2009/23/EC. The Declaration of Conformity is
available online at
europe.ohaus.com/europe/en/home/support/compliance.aspx.
AS/NZS CISPR 11
CAN/CSA-C22.2 No. 61010-1-12
UL Std. No. 61010-1 (3rd edition)
NSF/ANSI 169–2009
NSF/ANSI/3-A 14159-1-2010
Weighing Instruments verified at the place of manufacture bear one
of the preceding marks on the packing label and the green ‘M’
(metrology) sticker on the descriptive data plate. They may be put into
service immediately.
Weighing Instruments to be verified in two stages have no green ‘M’
(metrology) on the descriptive data plate and bear one of the
preceding identification marks on the packing label. The second
stage of the initial verification must be carried out by an authorized
and certified service organization established within the European
Community or by the National Notified Body.
The first stage of the initial verification has been carried out at the manufacturers
work. It comprises all tests according to the adopted European standard EN
45501:1992, paragraph 8.2.2.
If national regulations limit the validity period of the verification, the user of the
weighing instrument must strictly observe the re-verification period and inform the
respective weights and measures authorities.
6.3 Compliance
Compliance to the following standards is indicated by the corresponding mark on
the product.
Important notice for verified weighing instruments
Valor™ 4000W Series EN-19
Disposal
In conformance with the European Directive 2002/96/EC on Waste
Electrical and Electronic Equipment (WEEE) this device may not be
disposed of in domestic waste. This also applies to countries outside
the EU, per their specific requirements.
The Batteries Directive 2006/66/EC introduces new requirements
from September 2008 on removability of batteries from waste
equipment in EU Member States. To comply with this Directive, this
device has been designed for safe removal of the batteries at end-oflife by a waste treatment facility.
Please dispose of this product in accordance with local regulations at the collecting
point specified for electrical and electronic equipment. If you have any questions,
please contact the responsible authority or the distributor from which you
purchased this device.
Should this device be passed on to other parties (for private or professional use),
the content of this regulation must also be related.
Disposal instructions in Europe are available online at
europe.ohaus.com/europe/en/home/support/weee.aspx.
Thank you for your contribution to environmental protection.
FCC Note
This equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at his
own expense.
Industry Canada Note
This Class B digital apparatus complies with Canadian ICES-003.
ISO 9001 Registration
In 1994, OHAUS Corporation, USA, was awarded a certificate of registration to ISO
9001 by Bureau Veritus Quality International (BVQI), confirming that the OHAUS
quality management system is compliant with the ISO 9001 standard’s
requirements. On June 21, 2012, OHAUS Corporation, USA, was re-registered to
the ISO 9001:2008 standard.
Product Registration
Protect your investment. Register your product with your local OHAUS dealer or
online at www.ohaus.com.
EN-20 Valor™ 4000W Series
Limited Warranty
OHAUS products are warranted against defects in materials and workmanship from
the date of delivery through the duration of the warranty period. During the warranty
period OHAUS will repair, or, at its option, replace any component(s) that proves to
be defective at no charge, provided that the product is returned, freight prepaid, to
OHAUS. This warranty does not apply if the product has been damaged by accident
or misuse, exposed to radioactive or corrosive materials, has foreign material
penetrating to the inside of the product, or as a result of service or modification by
other than OHAUS. In lieu of a properly returned warranty registration card, the
warranty period shall begin on the date of shipment to the authorized dealer. No
other express or implied warranty is given by OHAUS Corporation. OHAUS
Corporation shall not be liable for any consequential damages.
As warranty legislation differs from state to state and country to country, please
contact OHAUS or your local OHAUS dealer for further details.
Serie Valor™ 4000W ES-1
Báscula
Plato de acero inoxidable
Tarjeta de garantía
Adaptador de corriente y conector
Manual de instrucciones
Etiqueta de capacidad
1. INTRODUCCIÓN
Este manual contiene instrucciones de instalación, funcionamiento y mantenimiento
de las series Valor™ 4000W. Por favor lea completamente el manual antes de usar
la báscula.
1.1 Precauciones de seguridad
Por favor siga estas precauciones de seguridad:
Verifique que el voltaje de entrada del adaptador de corriente alterna (CA)
coincida con el voltaje del suministro eléctrico de corriente alterna local.
No deje caer objetos en la plataforma.
No coloque la báscula al revés sobre la plataforma.
Desconecte la báscula de la fuente de corriente cuando la esté limpiando.
Utilice la báscula sólo en las condiciones ambientales especificadas en estas
instrucciones.
El mantenimiento debe realizarse solamente por el personal autorizado.
Utilice sólo pesos dentro de la capacidad de la báscula como se especifica en
estas instrucciones.
No use la báscula en ambientes hostiles.
No transporte la báscula por el plato o la subplataforma. Utilice los asideros
de sujeción que se encuentran en el lateral de la caja de la báscula.
2. INSTALACIÓN
2.1 Contenido del paquete
2.2 Instalación de componentes
Coloque el plato de acero inoxidable en la plataforma de pesaje antes de encender
la báscula.
Figura 2-1. Instalación del plato de acero inoxidable en la plataforma de pesaje
ES-2 Serie Valor™ 4000W
Burbuja de nivel
Pata niveladora
Sí No
2.3 Selección de la ubicación
Use la báscula en una superficie limpia, firme y segura. Evite lugares con corrientes
de aire excesivas, vibraciones, fuentes de calor o cambios bruscos de temperatura.
2.4 Nivelación del equipo
Ajuste las patas niveladoras para que la burbuja esté centrada en el círculo.
Asegúrese de que el equipo esté nivelado cada vez que cambia su ubicación.
Figura 2-2 Indicador de nivel
2.5 Alimentación
El adaptador de corriente alterna se utiliza para dar corriente a la báscula cuando
la alimentación por baterías no es necesaria. Primero conecte el adaptador de
corriente alterna al enchufe de entrada de la báscula ubicado en la parte inferior
de la báscula como se describe a continuación.
Después conecte el adaptador de corriente alterna a la alimentación de
corriente alterna adecuada.
Figura 2-4. Conexión del adaptador de corriente alterna a la alimentación de
Figura 2-3. Deslizar para conectar
corriente alterna
Precaución: Sólo se puede utilizar la báscula en un ambiente seco
cuando está alimentada por el adaptador de corriente alterna.
Serie Valor™ 4000W ES-3
Símbolo
Nivel de carga
Batería en uso:
Símbolo mostrado
2.5.1 Alimentación por baterías:
La báscula se puede utilizar de forma inmediata si se conecta a la corriente
alterna. Antes de utilizar la báscula alimentada por baterías, deje cargar la
batería durante 12 horas. La báscula cambiará automáticamente a
funcionamiento por batería si hay un fallo de corriente o si se quita el cable de
corriente. Con corriente alterna, la báscula está cargando constantemente, por
lo que el indicador de carga de la batería (consulte el elemento 11 en la tabla
3-2) permanecerá encendido. La báscula puede usarse durante la carga, y la
batería estará protegida contra sobrecarga.
Para un tiempo de funcionamiento máximo, la batería debe cargarse a
temperatura ambiente.
Durante el funcionamiento de la batería, el símbolo de batería indica el estado
de carga de la batería. Cuando esté cargando, el símbolo parpadeará
lentamente y cuando esté totalmente cargado, el símbolo dejará de parpadear.
TABLA 2-1
Notas:
Cuando el símbolo de batería parpadea rápidamente, queda
aproximadamente 30 minutos de tiempo de trabajo.
Cuando se muestre [Lo.bAt] , la báscula se apagará.
La carga de la báscula debe realizarse en un ambiente seco.
PRECAUCIÓN: La batería sólo debe ser reemplazada por un
distribuidor de OHAUS autorizado. Puede existir riesgo de
explosión si la batería recargable se reemplaza por el tipo
incorrecto o si no está correctamente conectada. El desecho
de la batería ácida de plomo debe hacerse de acuerdo a las
leyes y normativas locales.
ES-4 Serie Valor™ 4000W
Botón
Funciones
Pulsación corta1 (cuando está encendida): Puesta a cero de la
pantalla
Pulsación corta (cuando está apagada): Enciende la báscula
Pulsación larga2 (cuando está encendida): Apaga la báscula
Pulsación corta (en Menú): Selecciona/acepta la configuración que
se muestra en pantalla
Pulsación corta: Acumula un peso o muestra la información
acumulada a carga 0.
Pulsación larga: Alterna entre unidades activas
Pulsación corta (en Menú): Alterna entre las configuraciones
disponibles
Pulsación corta: Inicia una respuesta específica del modo de
aplicación
Pulsación larga: Selecciona el modo activo
Pulsación corta (en Menú): Vuelve a la configuración anterior
Pulsación corta: Introduce/borra un valor de tara
Pulsación larga: Accede al Menú usuario
Pulsación corta (en Menú): Sale rápidamente del Menú usuario
El sensor IR3 se puede programar para actuar como botones "sin
contacto". Consulte la sección 4.3 del menú de usuario para
consultar las configuraciones disponibles.
3. FUNCIONAMIENTO
3.1 Controles
Figura 3-1. Panel de control de la Valor 4000W con pantalla
TABLA 3-1
Notas:
1
Pulsación corta: Pulsar menos de 2,5 segundos.
2
Pulsación larga: Mantener pulsado más de 2,5 segundos.
3
El sensor IR se pueden activar con una mano u otro objeto que esté colocado a una altura
especificada (consulte la sección 4.3) desde el sensor.. La distancia de activación del sensor variará
la naturaleza reflectora del objeto. Si se produce una activación no deseada debido a situaciones
únicas el sensor se puede apagar.
Serie Valor™ 4000W ES-5
Elemento
Descripción
Elemento
Descripción
1
Símbolo de NET
9
Símbolo de Modo
calibración/menú
2
Símbolo de Centro de
cero
10
Símbolo de Dinámico*
3
Símbolo de Negativo
11
Símbolo de Cargador de batería
4
Símbolo de Peso estable
12
Símbolo de Libra, Onza,
Libra:Onza
5
Símbolo de Tara
preseleccionada*
13
Símbolos de gramo, kilogramo
6
Símbolos de Puntero
14
Símbolo de Piezas*
7
Símbolo de sensor IR
15
Símbolo de Porcentaje
8
Símbolo de Acumulación
(Rojo) Cargas > límite superior
(Verde) Cargas > límite inferior y ≤ límite superior
(Amarillo) Cargas < Límite inferior
1
1
Sensores
Figura 3-2. Pantalla de la Valor 4000W
1
2
1
3
1
4
1
5
1
6
1
7
1
8
1
9
1
10
1
11
1
12
1
13
1
14
1
15
Indicadores LED
Pantalla
TABLA 3-2 Símbolos de la pantalla
Nota: * no se utiliza
Figura 3-3: Pantalla trasera de la Valor 4000W.
Los indicadores LED coloreados de la parte derecha de los paneles de control se
utilizan en modo control de peso (sección 3.6) y se encienden según las siguientes
normas:
ES-6 Serie Valor™ 4000W
Masa de calibración de extensión necesario
(se vende por separado)
Máx.
Masa
1
Máx.
Masa
1
1.500 g
1,5 kg / 3 lb
6.000 g
6 kg / 15 lb
3.000 g
3 kg / 6 lb
15.000 g
15 kg / 30 lb
3.2 Apagado y encendido de la báscula
Para encender la báscula, presione el botón On/Zero Off. La báscula realiza una
prueba de pantalla, muestra momentáneamente la versión del software y después
entra al modo de pesaje activo.
Para apagar la báscula, mantenga pulsado el botón On/Zero Off hasta que se
muestre OFF.
3.3 Calibración inicial
Cuando la báscula se instala por primera vez, se debe calibrar para asegurar que
se van a lograr resultados precisos. Antes de realizar la calibración, asegúrese de
tener los pesos de calibración adecuados como se indican en la tabla 3-3.
Mantenga pulsado Menú hasta que se muestre en pantalla [mMeNU] . Cuando suelte
el botón, la pantalla mostrará [C.A.L]. Pulse Yes para aceptar y se mostrará [SpaN].
Pulse Yes de nuevo para comenzar la calibración de extensión. Mientras se
almacena la lectura cero parpadeará [--C--] . Después, la pantalla mostrará el valor
del peso de calibración. Coloque la masa de calibración especificada en el plato y
pulse Yes. Mientras se almacena la lectura parpadeará [--C--] . La báscula vuelve
al modo de aplicación anterior y estará lista para usar.
Se mostrará el mensaje [Err 3.0 CAl] si no se siguen estos pasos de calibración o
si se utilizó el peso incorrecto.
Se puede interrumpir el proceso de calibración apagando la báscula.
TABLA 3-3
Nota: 1 Las masas de libra se utilizan cuando se calibra en la unidad
de libra.
3.4 Modo de pesaje
1. Mantenga pulsado Mode hasta que se muestre en pantalla [wJeIGH].
2. Si fuera necesario, coloque un recipiente vacío en el plato y pulse Tare.
3. Añada una muestra al plato o contenedor. La pantalla muestra el peso de la
muestra.
Serie Valor™ 4000W ES-7
3.5 Modo de porcentaje
Este modo mide el peso de una muestra como porcentaje de un peso de
referencia.
1. Mantenga pulsado Mode hasta que se muestre en pantalla [PerCNt]. Se
mostrará en la pantalla [Clr.ref].
Notas: Pulse Function para visualizar el peso de referencia actual.
Si no hubiera almacenado ningún dato de referencia, se mostrará [Set.ref].
Pulse Yes para establecer datos de referencia.
2. Pulse No para almacenar el peso de referencia y continúe con el paso 6.
3. Si fuera necesario, coloque un recipiente vacío en el plato y pulse Tare. Pulse
Yes para establecer una nueva referencia. La báscula mostrará ahora en la
pantalla [PUt.ref].
4. Añada el material de referencia deseado al contenedor. Pulse Yes para
almacenar el peso de referencia. La pantalla muestra 100%.
5. Retire el material de referencia. Si fuera necesario, coloque un recipiente vacío
en el plato y pulse Tare.
6. Añada el material de muestra. La pantalla mostrará el porcentaje de la muestra
comparado con el peso de referencia.
7. Para borrar la referencia almacenada mantenga pulsado Mode hasta que se
muestre [PerCNt] en la pantalla. Pulse Yes cuando [Clr.ref] se muestre en la
pantalla.
3.6 Modo Control de peso
Este modo fija los limites alto y bajo de peso para los procesos de control de la
porción.
1. Mantenga pulsado Mode hasta que se muestre en pantalla [CHeCk]. Se mostrará
en la pantalla [CLr.ref].
Notas: Pulse Function para visualizar los límites de referencia bajo y alto.Si no hubiera almacenado ningún dato de referencia, se mostrará
[Set.ref]. Pulse Yes para establecer datos de referencia.
2. Pulse No para almacenar los límites del peso de referencia y continúe con el
paso 5.
3. Pulse Yes para establecer unos nuevos valores de referencia. La báscula
mostrará en la pantalla [Set. LO]. Pulse Yes para visualizar el valor límite "Bajo".
Pulse Yes para aceptar o No para editar el valor límite "Bajo". El valor
almacenado se mostrará entonces en la pantalla con el primer dígito resaltado
[000.000 kg]. Pulse repetidamente No hasta que aparezca el número deseado.
Pulse Yes para aceptar y resaltar el dígito siguiente. Repita hasta que todos los
dígitos sean correctos. Pulse Yes para aceptar el valor límite "bajo", se
mostrará en la pantalla [Set. HI].
4. Repita el mismo procedimiento para aceptar o editar el valor "alto".
5. Si fuera necesario, coloque un recipiente vacío en el plato y pulse Tare.
Coloque el material de muestra en el plato o en el contenedor. Si el peso de
muestra se encuentra por debajo del rango de peso objetivo, se encenderá el
LED amarillo. Si la muestra se encuentra dentro del rango de peso objetivo, se
encenderá el LED verde. Si la muestra se encuentra por encima del rango de
peso objetivo, se encenderá el LED rojo.
ES-8 Serie Valor™ 4000W
En modo pesaje, submenú ajuste ACCUmM, seleccione mMaNU:
Si fuera necesario, Coloque un recipiente vacío en el plato y
pulse Tare.
Paso i. Coloque el peso (0,04 kg) en el plato y pulse la tecla
M+. El icono Σ seguirá parpadeando hasta que
se retire el peso.
Paso ii. Retire el peso del plato. Coloque otro
peso (0,03 kg) y pulse la tecla M+:
Paso iii. Retire el peso del plato:
Paso iv. Pulse la tecla M+ para mostrar en la pantalla la
información estadística:
Paso v.
Si fuera necesario, siga el paso 3 indicado anteriormente para
borrar los datos acumulados.
N 2
tOtal
0.070 kg
mMIN
0,030 kg
mMax
0.040 kg
mMaNU
*
0,040 kg
*
0,030 kg
*
0,000 kg
6. Para borrar los valores de referencia almacenada mantenga pulsado Mode
hasta que se muestre [CHeCk]en la pantalla. Pulse Yes cuando [Clr.ref] se
muestre en la pantalla.
3.7 Acumulación
La acumulación funciona junto con cada modo de aplicación. Esta función permite
al usuario almacenar el total de una serie de mediciones de peso.
1. Pulse la tecla M+ para añadir el peso a los datos acumulados. El icono Σ
seguirá parpadeando hasta que se retire el peso.
Nota: Si se pone a AUTO la opción Acumular en el menú de configuración,
no es necesario pulsar la tecla M+.
2. Cuando el plato esté libre, pulse la tecla M+ para mostrar en la pantalla la
información estadística de los datos acumulados.
3. Para borrar los datos acumulados, pulse la tecla Tare cuando la pantalla
muestre la información estadística de los datos acumulados. La pantalla
muestra [Clr.aCC]. Pulse la tecla Yes para limpiar los datos almacenados y
volver al modo actual.
Notas:
Sólo se almacenan con los datos de acumulación total los pesos estables.
Para evitar añadir la misma carga varias veces al total, debe devolver 0 a la carga
del plato antes de que se pueda añadir la siguiente carga.
Cuando Legal for Trade (Comercio legal) esté OFF (APAGADO), la pantalla debe
mostrar el peso bruto 0, el peso neto 0 o un valor neto negativo, cuando Legal for
Trade (Comercio legal) esté ON (ENCENDIDO), la pantalla debe mostrar el valor
bruto 0. De lo contrario, la imagen Σ continuará parpadeando.
Las cargas brutas y las cargas netas no se pueden añadir al mismo total. Si la
primera carga es un peso bruto, las cargas futuras también deben ser pesos brutos.
Si la primera carga es un peso neto, las cargas futuras también deben ser pesos
netos.
El número máx. de veces que se puede acumular es 9999.
El cambio de modo borrará los datos acumulados.
Ejemplo de acumulación:
Loading...
+ 82 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.