Penn Scale SJ-1000H Installation Manual

Page 1
SJ-1000H SJ-2000H SJ-5000H SJ-12KH
WM+PD4001280A
Page 2
SAFETY PRECAUTIONS
All safety messages are identified by the following, “WARNING” or “CAUTION”, of ANSI Z535.4 (American National Standard Institute:
WARNING
CAUTION
This manual is subject to change without notice at any time to improve the
product.
Product specifications are subject to change without any obligation on the
part of the manufacture.
When using the SJ-H, the following safety precautions should always be
followed.
WARNING Use only specified AC adapter for the scale. Other
CAUTION Avoid installing the scale in direct sunlight, which may cause
A potentially hazardous situation which, if not avoided, could result in death or serious injury. A potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
adapter may cause damage (AC adapter is optional). Internal service or adjustment to this product should be performed by a qualified person.
discoloration or malfunctions.
Avoid over and/or shock load that could cause damage to the
scale.
Avoid using the weighing pan to move the scale, as that
could cause damage to the scale.
Do not mix the type of batteries. Replace all other batteries at
the same time.
If the scale is not to be used for a long period of time, remove
all batteries from the battery room to avoid leakage.
The scale is not watertight and avoid exposure to water .
1
Page 3
PARTS DESCRIPTION
r
p
Weighing Platform
Spirit level
Display
Four leveling feet
ON/OFF key
Turns the scale power on/off.
UNITS key
Changes weighing units, (k)g, lb, oz or lb-oz.
RE-ZERO key
Sets display to zero or subtracts weight of container.
Battery cover
AC adapter jack
Battery indicator
Indicates battery remaining amount. It turns on
STABLE indicato
Indicates when the reading is stable.
Polarity
about for 10 seconds when
owered on.
Weighing unit, (k)g, lb or oz
(depends on model)
ZERO indicator
Indicates when the scale is at zero point.
NET indicator
Indicates when NET weight is displayed (Tare function is used).
`
Weighing unit, lb-oz
(depends on model )
2
Page 4
PREPARATION
A
Installing Batteries
CAUTION
Take great care of the polarity of batteries. The polarity marks are shown in the battery box. Replace used batteries with four new ones when lb is displayed. Do not mix used and new batteries. Do not mix the battery type. Remove batteries when the scale is not to be used for a long time.
Battery box
2
1
New four batteries
3
(R20P / LR20 / D size)
Battery cover
1 Slide the battery cover downward to open. 2 Push the battery box to unhook and pull it out. 3 Insert new four batteries (R20P / LR20 / D size) into the battery box, taking
extreme care of the polarities.
Batteries are not included with shipment.
Connecting an AC adapter (Optional)
The AC input requirement could be 100, 120 or 230 volts (50/60Hz) depending on the area where us e d , s o p l e a s e v e r i f y t h a t the adapter is correct. Open the rubber cap and plug the AC adapter.
Open rubber cap.
C adapter plug
Setting up the Scale
Place the scale on the firm surface and adjust the feet so that the sprit level shows the scale is level for an accurate weighing.
OK NG
3
Page 5
WEIGHING
Before Weighing
Automatic power-off function.
If the scale left ON and the stable indicator is displayed, an automatic power-off function turns power off after approximately 5 minutes. To disable/enable this function, turn the scale on while pressing the [RE-ZERO] key. Then
poff0 Function disabled. poff1 Function enabled.
Select the setting by pressing the [RE-ZERO] key which toggles the display above, and press the [UNITS] key. The scale shows end and returns to the weighing mode.
Error message
e Overload: remove excess load on the weighing platform.
----- The scale zero is out of zero range or unstable when powered
on. Make sure no load is on the weighing platform or nothing touches the platform.
lb Low battery: replace the batteries or use the AC adapter.
poff0 or poff1 will be displayed.
Calibration may be required before weighing
Read “CALIBRATION” first and if necessary, calibrate your scale for accurate weighing.
Weighing Procedures
1. Turn on the scale.
Press the [ON/OFF] key to turn on the scale. Then the scale displays all segments for a few seconds and then
2. Start weighing
If you do not use a container for weighing,
Verify the reading is
0 . Place objects on the weighing platform to weigh. When the reading
becomes stable, the stable indicator is displayed.
If you use a container for weighing,
Place an empty container on the platform. Wait for the stable indicator to be displayed and press the [RE-ZERO] key. Place the objects to be weighed in the container. When the reading becomes stable, the stable indicator is displayed.
0 . If not, press the [RE-ZERO] key to display
0 will be displayed.
Select the weighing unit
Press the [UNITS] key to select weighing unit either “(k)g”, “lb”, “oz” or “lb-oz”. The scale will power up with the weighing unit that had been used when turned off last time.
4
Page 6
CALIBRATION
When to Calibrate
Calibration may be required when it initially installed or it is moved a substantial distance. Also periodical calibration will be recommended.
Calibration using a Calibration Weight
Calibration switch
1. Enter calibration mode.
Press the [ON/OFF] key to turn the power off. While
pressing both the [RE-ZERO] and [UNITS] keys, press the [ON/OFF] key.
Or remove the calibration switch cover on the bottom
and press the calibration switch while in the weighing mode. Then
Cal will be displayed.
Cal will be displayed.
Calibration switch cover
2. Zero calibration
Press the [RE-ZERO] key to calibrate zero. Cal0 will be displayed. Wait for the stable indicator to be displayed and press the [RE-ZERO] key.
Calf will be displayed in a few seconds. If the zero calibration only is to
be done, press the [UNITS] key. The scale will automatically return to the weighing mode.
3. Span calibration
When Calf is displayed, place the w ei g h t a t t he c e n t e r o f t h e platform.
Wait for the st able indicator to be displayed, then press the [RE-ZERO] key. The display will show weighing mode. See “SPECIFICATIONS” about the calibration weight.
end and the scale will automatically return to the
Calibration by Gravity Compensation
The scale is calibrated at 9.798 m/s2 at the factory. If the acceleration of gravity at your location is not same and if you do not have calibration weights, you can calibrate the scale by compensating the acceleration of gravity. (Refer “The Value of Gravity at Various Locations”.)
1. Set a new acceleration value.
When Cal is displayed, press the [UNITS] key and 9.798 will be
displayed. Press the [RE-ZERO] key to increment the blinking digit and press the [UNITS] key to move the blinking digit.
2. Store the value in memory.
While pressing the [UNITS] key, press and hold the [RE-ZERO] key and
release the [UNITS] key. Then the display will show
Cal . Turn the scale off to finish the procedure.
5
end and return to
Page 7
SPECIFICATIONS
MODEL SJ-1000H SJ-2000H SJ-5000H SJ-12KH
Capacity
x
Resolution
Non-linearity ±1 g ±2 g ±5 g ±10 g Repeatability 1 g 2 g 5 g 10 g Sp an drif t ±0.015% / °C TYP (5°C~35°C / 41°F~95°F) Operating temp. -10°C~40°C / 14°F~104°F, Less than 85%RH (No cond ensatio n) Display 25 mm / 0.98 inches, 7 segment liquid crystal display Display update Approximately 10 times per second Power 4 x R20P / LR20 / “D” size batteries or AC adapter Battery life Approximately 800 hours with alkaline cells at 20°C / 68°F Platform size 230 (W) x 190 (D) mm / 9.05 (W) x 7.48 (D) inches
Dimensions Weight Approximately 1.9 kg / 4.2 lb
Calibration weight Accessories This manual Options AC adapter
(k)g 1000 g x 0.5 g 2000 g x 1 g 5000 g x 2 g 12 kg x 5g
lb 2.2 lb x 0.001 l b 4.4 lb x 0.002 lb 11 lb x 0.005 lb 26 lb x 0.01 lb
oz 35 oz x 0.02 oz 70 oz x 0.05 oz 176 oz x 0.1 oz 423 oz x 0.2 oz
lb-oz 2.2 lb x 0.1 oz 4.4 lb x 0.1 oz 11 lb x 0.1 oz 26 lb x 1 oz
265 (W) x 250 (D) x 103 (H) mm
10.4 (W) x 9.8 (D) x 4.1 (H) inches
1000g ± 0.1g 2000g ± 0.2g 5000g ± 0.5g 10 kg ± 1g
The Value of Gravity at Various Locations
Amsterdam 9.813 m/s Athens 9.807 m/s Auckland NZ 9.799 m/s Bangkok 9.783 m/s Birmingham 9.813 m/s Brussels 9.811 m/s Buenos Aires 9.797 m/s Calcutta 9.788 m/s Cape Town 9.796 m/s Chicago 9.803 m/s Copenhagen 9.815 m/s Cyprus 9.797 m/s Djakarta 9.781 m/s Frankfurt 9.810 m/s Glasgow 9.816 m/s
2
Havana 9.788 m/s
2
Helsinki 9.819 m/s
2
Kuwait 9.793 m/s
2
Lisbon 9.801 m/s
2
London (
2
Los Angeles 9.796 m/s
2
Madrid 9.800 m/s
2
Manila 9.784 m/s
2
Melbourne 9.800 m/s
2
Mexico City 9.779 m/s
2
Milan 9.806 m/s
2
New York 9.802 m/s
2
Oslo 9.819 m/s
2
Ottawa 9.806 m/s
2
Paris 9.809 m/s
Greenwich
) 9.812 m/s
6
2
Rio de Janeiro 9.788 m/s
2
Rome 9.803 m/s
2
San Francisco 9.800 m/s
2
Singapore 9.781 m/s
2
Stockholm 9.818 m/s
2
Sydney 9.797 m/s
2
Taichung 9.789 m/s
2
Taiwan 9.788 m/s
2
Taipei 9.790 m/s
2
Tokyo 9.798 m/s
2
Vancouv er, BC 9.809 m/s
2
Washington DC 9.801 m/s
2
Wellington NZ 9.803 m/s
2
Zurich 9.807 m/s
2
2 2 2 2 2 2 2 2 2 2 2 2 2 2
Page 8
3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013 JAPAN Telephone: [81] (3) 5391-6132 Fax: [81] (3) 5391-6148
http://www.aandd.co.jp/
A&D ENGINEERING, INC.
1555, McCandless Drive, Milpitas, CA. 95035 U.S.A. Telephone: [1] (408) 263-5333 Fax: [1] (408)263-0119
A&D INSTRUMENTS LTD.
Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxon OX14 1DY United Kingdom Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485
<German Scales Office>
Großbe Straße 13 b 22926 Ahrensburg GERMANY Telephone: [49] (0) 4102 459230 Fax:[49] (0) 4102 459231
A&D MERCURY PTY. LTD.
32 Dew Street, Thebarton, South Australia 5031 AUSTRALIA Telephone: [61] (8) 8301-8100 Fax: [61] (8) 8352-7409
A&D KOREA Limited
8th Floor, Manhattan Bldg. 36-2 Yoido-dong, Youngdeungpo-ku, Seoul, KOREA Telephone: [82] (2) 780-4101 Fax: [82] (2) 782-4280
7
Loading...