1. WARNINGS AND CAUTIONS
The following WARNINGS and CAUTIONS
must be read and understood before using
this ventilator.
WARNINGS
General Information
1. Personnel must make themselves
familiar with the contents of this
manual and the machine’s function
before using the ventilator.
Before Using the Ventilator
2. Before the ventilator is used clinically
for the first time, verify that the hospital
engineering department has carried out
an earth continuity test.
3. Excessive electronic noise caused by
other poorly regulated devices, such as
an electrocautery unit, may adversely
interfere with the proper functioning of
the ventilator.
To avoid this problem, do not connect
the ventilator’s power cord into the
same electrical wall outlet or adaptor
strip into which an electrocautery unit
is connected.
4. If used with a mains extension cord, the
unit may be subject to electro-magnetic
interference.
5. The driving gas supply must be clean
and dry to prevent ventilator
malfunction.
6. This ventilator is designed to be driven
by oxygen or medical air only. It is
calibrated during manufacture for use
with either gas.
Before the ventilator is used clinically
for the first time, the commissioning
engineer must confirm that the internal
Air/Oxygen switch is set correctly for
the gas that is to be used.
The use of any other gas will cause
inaccurate operation and may damage
the ventilator, resulting in potential
injury to the patient.
7. The driving gas is discharged through
the opening in the back of the ventilator
control unit.
The discharged gas may contaminate
the environment, and should therefore
be extracted using a gas scavenging
system.
8. The bellows can only support
approximately 1 kPa (10 cmH
2
O)
differential positive pressure, above
which it may be dislodged from the
mounting ring, resulting in dangerous
malfunction of the ventilator.
Do not connect a positive end
expiratory pressure (PEEP) valve or
other restrictive device to the exhaust
port on the bellows base.
This would increase the pressure inside
the bellows and the bellows could
detach from the base, causing serious
malfunction.
9. Breathing System
The breathing system which conveys
gases from the anaesthetic machine to
the patient, and disposes of expired
gases, must conform to the
requirements of ISO 8835-2.
Because breathing systems require
frequent cleaning and disinfection they
are not a permanent part of the
anaesthetic ventilator and therefore
cannot be directly under the control of
the anaesthetic ventilator manufacturer.
However, we strongly recommend that
only breathing systems which have
been approved and authorised by the
manufacturer for use with AV-S should
be employed.
Do not use conductive breathing
system hoses.
When mechanical ventilation is
employed the patient breathing system
must be connected directly to a
pressure relief valve to prevent the
possibility of barotrauma.
10. Do not connect a spirometer to the
exhaust port on the bellows base.
The device will not measure exhaled
volumes in that position.
2