Pendulum CNT-90, CNT-91, CNT-90XL, CNT-91R Getting Started Manual

II
4031 600 90401
May 2017- Seventh Edition
© 2017 Pendulum Instruments / Altaria Services
III
Table of Contents
Introduction ................ 1-2
About this Manual............ 1-2
Warranty................. 1-2
DeclarationofConformity .......1-2
Safety Precautions ............. 1-3
CautionandWarningStatements....1-3
Symbols................. 1-3
If in Doubt about Safety .........1-4
2 Preparation for Use
Unpacking................. 2-2
CheckList................ 2-2
Identification .............. 2-2
ReadingtheElectronicManuals......2-2
Installation................. 2-2
Supply Voltage ............. 2-2
Setting ................ 2-2
Fuse.................. 2-2
Grounding ................ 2-3
RearPanel................ 2-3
LinePowerInlet ...........2-3
ReferenceOutput...........2-3
External Reference Input .......2-3
External Arming Input ........2-3
GPIB/IEEE-488 Connector ......2-3
USBConnector............ 2-4
Optional Main Input Connectors . . . 2-4
Fan.................. 2-4
TypePlate............... 2-4
OrientationandCooling......... 2-4
Fold-Down Support ...........2-4
Rackmount Adapter ........... 2-4
3 Operating the Counter
Introduction ................ 3-2
UserInterface............... 3-2
Getting Familiar with the Counter .....3-2
DescriptionofKeys............3-3
Power.................. 3-3
SelectFunction............. 3-3
Autoset/Preset.............. 3-3
MoveCursor .............. 3-3
Enter................... 3-3
Exit/OK ................. 3-3
Cancel.................. 3-3
PresentationModes...........3-4
VALUE................ 3-4
STAT/PLOT.............. 3-4
Remote................ 3-5
HardMenuKeys ............ 3-5
Input A ................ 3-5
Input B ................ 3-5
Settings................ 3-6
Math/Limit.............. 3-7
UserOptions............. 3-8
Hold/Run............... 3-9
Restart ................ 3-9
EnteringNumericValues........3-9
IV
4 Exercises
Preparation................. 4-2
BasicStartup................ 4-2
High-Resolution Frequency Measurement . 4-2
Built-InMathProcessing .........4-2
High-SpeedMeasurements ........ 4-3
UndertheINPUTMenus:........ 4-3
Under the SETTINGS Menu: ......4-3
TimeMeasurements............4-4
Jitter Measurements (Statistics) ......4-4
Other Single-Channel Measurements . . . 4-4
DutyCycle ............... 4-4
RiseTime................ 4-5
FallTime ................ 4-5
Remarks................. 4-5
VoltMax/MinMeasurements......4-5
ChannelSwapping............ 4-5
Two-ChannelMeasurements........4-6
TimeInterval.............. 4-6
Phase .................. 4-6
UsingAutotosetfixedtriggerlevels
(AutoOnce).............. 4-6
Procedure............... 4-6
MemorySettings ............ 4-6
Auxiliary Functions ............ 4-7
Time-Out ................ 4-7
Summary.................. 4-7
Sales&ServiceOffices...........V
Chapter 1
Introduction
Introduction
Congratulations on your choice of instrument. It will serve you well for many years to come.
Even though we know that you are eager to get going, we urge you to take a few minutes to read through the sections on safety in the first two chapters carefully before plugging the line connector into the wall outlet.
It is essential for your own safety to know the restrictions that are applicable to all equip
­ment that can be connected to line power. Therefore, read about Safety Precautions on page 1-3 and Installation on page 2-2.
That chapter is also the key to the compre­hensive information that can be found on the included CD, if you need closer information on a subject.
About this Manual
This manual contains directions for use that are common to all Timer/Counter/Analyzers in the CNT-9X series.
In order to simplify the references, the CNT-9X is further referred to throughout this manual as the '9X'.
Warranty
The Warranty Statement is part of the folder Important Information that is included with the shipment.
Declaration of Conformity
The complete text with formal statements con­cerning product identification, manufacturer and standards used for type testing is available on request.
Introduction
1-2 Introduction
Safety Precautions
This instrument has been designed and tested for Measurement Category I, Pollution Degree 2, in accordance with EN/IEC 61010-1:2001 and CAN/CSA-C22.2 No. 61010-1-04 (in
­cluding approval). It has been supplied in a safe condition. The user must have acquired adequate knowledge of it by thoroughly studying this manual.
To ensure the correct and safe operation of the instrument, it is essential that you follow gen
­erally accepted safety procedures in addition to the safety precautions specified in this man
-
ual.
The instrument is designed to be used by trained personnel only. Removing the cover for repair, maintenance, and adjustment of the instrument must be done by qualified person­nel who are aware of the hazards involved.
The warranty commitments are rendered void if unauthorized access to the interior
of the instrument has taken place during the given warranty period.
Caution and Warning Statements
CAUTION: Shows where incorrect
procedures can cause damage to, or destruction of equipment or other property.
WARNING: Shows a potential danger
that requires correct procedures or practices to prevent personal in
-
jury.
Symbols
Shows where the protective ground terminal is connected inside the instrument. Never remove or loosen this screw.
This symbol is used for identifying the functional ground of an I/O signal. It is always connected to the instrument chassis.
Tells the operator to consult the
manual.
Example:
One such symbol is printed on the instrument below the A and B inputs. It indicates that the damage level for the input voltage decreases from 350 V
p
to 12 V
RMS
whenyouswitchthe
input impedance from 1 MW to 50 W.
Safety Precautions 1-3
Introduction
Do not overlook the safety instruc
-
tions!
If in Doubt about Safety
Whenever you suspect that it is unsafe to use the instrument, you must make it inoperative by doing the following:
Disconnect the line cord.
Clearly mark the instrument to prevent its further operation.
Inform your Pendulum representative.
For example, the instrument is likely to be un
-
safe if it is visibly damaged.
Introduction
1-4 Safety Precautions
Chapter 2
Preparation for Use
Unpacking
Check that the shipment is complete and that no damage has occurred during transportation. If the contents are incomplete or damaged, file a claim with the carrier immediately. Also no
-
tify your local Pendulum sales or service orga
­nization in case repair or replacement may be required.
Check List
The shipment should contain the following:
The counter, Model CNT-9X
Line cord
N-to-BNC Adapter (only if an optional C-channel input with a Type N connector
was ordered)
Built-in options as ordered should be in­stalled. See Identification below.
Folder with Important Information
Certificate of Calibration
A CD-ROM including the following docu­mentation in PDF:
Getting Started Manual
User's Manual
Programmer's Handbook
Identification
The type plate on the rear panel shows the type number and the serial number. See illus
­tration on page 2-4. Installed options are listed under the menu User Options - About,where you can also find information on firmware version and calibration date. See page 3-8.
Installed options can also be identified by checking the full type number on the type plate.
Reading the Electronic Manuals
You need the Adobe® Reader® software to be able to read the manuals on the CD. It is in
­cluded on the CD or can be downloaded free of charge from www.adobe.com
.
Insert the CD into the CD-ROM unit of your PC or Mac and select the file you are looking for from the index.
Installation
Supply Voltage
n
Setting
The Counter can be connected to any AC sup­ply with a voltage rating of 90 to 265 V
RMS
, 45 to 440 Hz. The counter automatically ad­justs itself to the available line voltage.
n
Fuse
The secondary supply voltages are electroni
-
cally protected against overload or short cir
­cuit. The primary line voltage side is protected by a fuse located in the power supply unit. The fuse rating covers the full voltage range. Consequently there is no need for the user to replace the fuse under any operating condi
-
tions, nor is it accessible from the outside.
CAUTION: If this fuse is blown, it is
likely that the power supply is badly damaged. Do not replace the fuse. Send the counter to the local Service Center.
Removing the cover for repair, maintenance and adjustment must be done by qualified and
Preparation for Use
2-2 Unpacking
trained personnel only, who are fully aware of the hazards involved.
The warranty commitments are rendered void if unauthorized access to the interior of the instrument has taken place during the given warranty period.
Grounding
Grounding faults in the line voltage supply will make all instruments connected to it po
­tentially dangerous. Before connecting any unit to the power line, you must make sure that the protective ground functions correctly. Only then can a unit be connected to the power line and only by using a three-wire line cord. No other method of grounding is permit­ted. Extension cords must always have a pro­tective ground conductor.
CAUTION: If a unit is moved from a
cold to a warm environment, con­densation may cause a shock hazard. Ensure, therefore, that the grounding requirements are strictly met. Allow enough time for the in­strument to adapt to new ambient conditions before connecting it to line power.
WARNING: Never interrupt the
grounding cord. Any interruption of the protective ground connection inside or outside the instrument or disconnection of the protective ground terminal is likely to make the instrument dangerous.
Rear Panel
See the figure overleaf for the location of the connectors. The labels refer to the headings below, where the characteristics are summa
­rized. See the User's Manual for technical specifications.
n
Line Power Inlet
AC 90-265 V
RMS
, 45-440 Hz, no range
switching needed.
n
Reference Output
10 MHz derived from the internal or the exter
­nal reference, depending on which of them is the active Measurement Reference. The choice is made from the Settings Menu.
n
External Reference Input
If the Measurement Reference is set to Auto in the Settings Menu, this input will be automati
-
cally selected, provided a valid signal is pres
-
ent.
n
External Arming Input
Supports external arming (synchronization) of measurements. The main inputs A & B can also be selected for measurement arming via the Settings Menu.
n
GPIB/IEEE-488 Connector
The address is set from the User Options Menu.
n
USB Connector
A serial interface according to USB 2.0 12 Mb/s provides a fast communication link to your PC.
n
Optional Main Input Connectors
This factory-installed option replaces the ordi
-
nary front panel main input connectors.
n
Fan
Forced cooling is provided by means of a speed-controlled fan.
Installation 2-3
Preparation for Use
n
Type Plate
Here you can find the type number and the se- rial number as well as information on rated line voltage and power.
Orientation and Cooling
The counter can be operated in any position desired. Make sure the air flow through the ventilation slots at the side panels is not ob
­structed. Leave 5 centimeters (2 inches) of space around the counter.
Fold-Down Support
For bench top use, a fold-down support is available for use underneath the counter. This support can also be used as a handle to carry the instrument.
Rackmount Adapter
An optional rackmount kit is available. See the User's Manual for installation details.
Preparation for Use
2-4 Installation
!
!
!
191125
The CNT-90 rear panel layout.
Ext. Ref. In
Ref. Out Ext. Arm In GPIB Conn.
USB Conn.
Line Power Inlet
Type Plate
Fan
Opt. Main Input Conn.
Fold-down support for comfortable bench-top use.
Chapter 3
Opera ting the
Counter
Introduction
While this counter has a vast array of features and operating modes, the majority of the most useful functions can be learned in just a few minutes. The following descriptions and exer
-
cises will help you grasp the basics of opera
­tion. They will also serve as an introduction to some of the more advanced features. It will take you about half an hour to gather experi
­ence enough to continue exploring the world of counting on your own.
User Interface
The fundamental idea has been to facilitate even complex measurements by a consistent interactive user interface that fully exploits the large graphic display. The casual as well as the frequent user will benefit from a combina
­tion of hard and soft keys. In this context hard keys either act immediately or open fixed menus, whereas the function of soft keys de
-
pends on the display information.
Getting Familiar with the Counter
The front panel is laid out in a logical fashion. Take a few seconds to find the sections de­scribed below on the front panel of your coun­ter. See also the figure on this page. It will help you locate the different keys faster.
Operating the Counter
3-2 Introduction
TRIG
GATE
TRIG
pendulum
TIMER / COUNTER / ANALYZE R
100ps / 300MHz
CNT-90
INPUTINPUT
INPUT A
SETTINGSETTING
SETTINGS
MATH/LIM USER OPUSER OP
USER OPT
HOLD / RU
HOLD / RUN
RESTARTINPUTINPUT
INPUT B
A
300 MHz
B
MAX 12Vrms 50 30Vrms 1M
!
C
MAX +30dBm
50
0.2 8GHz
!
Move CursorSave&Exit Don't Save&ExitPower
Hard Menu Keys
Freeze
Select Function Autoset/Preset Enter
Presentation Modes
Soft Keys
Description of Keys
Power
The ON/OFF key is a toggling secondary power switch. Part of the instrument is always on as soon as line power is applied, and this so-called standby condition is indicated by a red LED above the key.
Select Function
This hard key is marked MEAS FUNC. When you depress it, the menu below will open.
The current selection is indicated by text in­version. Select the measurement function you want by depressing the corresponding soft key right below the display. A new menu will ap­pear where the contents depend on the func­tion. If you for instance have selected Fre
-
quency, you can then select between Fre
-
quency, Frequency Ratio and Fre
-
quency Burst. Finally you can also change
the preselected input channel.
Autoset/Preset
By depressing the AUTOSET key once after selecting the wanted measurement function and input channel, you will most probably get a measurement result. The AUTOSET system ensures that the trigger levels are set optimally for each combination of measurement func
­tion and input signal amplitude, provided that you apply relatively normal signals.
By depressing the AUTOSET key twice within two seconds, you will enter the Preset mode, which takes you another step further to
­wards fully automatic settings for your current measurement. Such auxiliary functions as
measuring time, mathematics, filter and arm
-
ing are then reset to their default values. The
intention is to prevent possible lockups and misinterpretations when changing measure
-
ment function or test setup, for instance.
Move Cursor
There are four cursor keys for moving the cur
­sor, normally marked by text inversion, around the menu trees.
Enter
When you want to confirm a choice without leaving your menu position, press this key.
Exit/OK
This hard key performs the Save & Exit opera­tion. You will confirm your selections by de­pressing the key, and at the same time you will exit the current level and enter the next higher level in the menu tree.
Cancel
This hard key performs the Don't Save & Exit operation. You will exit the current level by depressing the key and enter the next higher level in the menu tree without confirming any selections made.
Description of Keys 3-3
Operating the Counter
Presentation Modes
n
VALUE
The Value Mode gives the result of the main measurement function as a numerical value in large characters with full resolution. In addi
-
tion, the results of supplementary measure
­ments are displayed in smaller characters with limited resolution near the bottom of the screen.
In case the Limits Alarm function is enabled, Range is the selected Limit Mode,andAlarm is the selected Limit Behavior,thenasimple graph is also given in which the result of the current measurement is shown as an 'emoticon' at a position relative to the limits set by the operator. Values are represented by a smiling face when inside the limits and a frowning face when outside.
If one of the limits has been exceeded, the limit indicator at the top of the display will be flashing. Only data inside the limits is used for autoscaling, so results outside the visible graph area are replaced by an arrowhead at the left or the right edge of the display.
n
STAT/PLOT
If you want to analyze a number of successive measurements using statistical methods, this is the key to operate. There are three display modes available by toggling this key:
Numerical Mode
Histogram Mode
Trend Plot Mode
Numerical Mode
In this mode the statistical information is dis
-
played as numerical data containing the fol
-
lowing elements:
Mean: running mean value of the main measurement over N samples
Max: maximum value
Min: minimum value
P-P: peak-to-peak deviation
Adev: Allan deviation
Std: standard deviation
Histogram Mode
Successive measurement results are converted into a histogram.
The number of bins along the horizontal axis can be set by the user through the Settings ® Stat menu. The bins in the histogram are autoscaled based on the measured data. Lim
-
its, if enabled, and the running mean value
X
are shown as vertical dotted lines. The center of the graph is indicated by a filled triangle on the X-axis. The corresponding numerical value is displayed below it and so is the scale
Operating the Counter
3-4 Description of Keys
factor. Enabled limits affect the autoscaling so as to visualize the current measurements and the set limits simultaneously.
Trend Plot Mode
This mode is used for observing periodic fluc
-
tuations or possible trends.
A trend plot terminates (if HOLD is activated) or restarts (if RUN is activated) after the set number of samples has been completed. The trend plot is always autoscaled based on the measured data, starting with 0 at restart. Lim­its are shown as horizontal lines, if enabled.
n
Remote
When the instrument is controlled from the GPIB bus, and the remote line is asserted, or when the instrument is controlled from the USB bus, then the presentation mode changes to Remote, indicated by the label REM on the display. The main measurement result and the input settings are displayed in this mode.
Hard Menu Keys
These keys are mainly used for opening fixed menus, from which further selections can be made, for instance by means of the soft keys.
n
Input A
By depressing this key, the bottom part of the display will show the settings for Input A.
The active settings are in bold characters and can be changed by pressing the corresponding soft key below the display. You can also move the cursor, indicated by text inversion, with the RIGHT and the LEFT arrow keys to the
desired position and then make the selection by pressing the UP or the DOWN arrow keys. You can also use the ENTER key.
The selections that can be made in this menu are:
Trigger Slope: positive or negative, indi
-
cated by corresponding symbols
Coupling: AC or DC
Impedance: 50 W or1MW
Attenuation: 1x or 10x
Trigger: Manual or Auto (always Auto 10/90 % when measuring transition times).
Trigger Level: If Manual has been se­lected, the absolute level can be ad­justed with the up/down arrow keys or by entering a numerical value from the keyboard.
Filter: On or Off. Pressing ENTER or the soft key below the text opens up the Fil
-
ter Settings menu. See below.
You can select a fixed 100 kHz analog
filter or an adjustable digital filter. The equivalent cutoff frequency is set via the value input menu that opens if you se
­lect Digital LP Frequency from the menu above.
n
Input B
The available settings under Input B are equal to those available under Input A.
Description of Keys 3-5
Operating the Counter
n
Settings
This key accesses a range of more sophisti
­cated instrument settings that usually need not be changed for basic measurements. The fig
-
ure below shows the display after recalling the default settings via the USER OPT menu.
A detailed description of all the possibilities is far beyond the scope of this introductory man
­ual. See the User's Manual for full details on functions mentioned and unmentioned here.
Meas Time
This value input menu is only useful if you se­lect a frequency function. Longer measuring time means fewer measurements per second but gives higher resolution. You can change the Measuring Time by entering a numerical value, or you can use the UP/DOWN arrow keys to increase or decrease the current value.
Burst
This settings menu facilitates measurements on pulse-modulated signals and is only useful if Frequency Burst is the main measurement function.
Both the carrier frequency and the modulating frequency – also known as the pulse repetition frequency (PRF) – can be measured, often without the support of an external arming sig
-
nal (see below).
Arm
Arming is the general term used for the means to control the actual start or stop of a measure
­ment. When arming is used, the normal free-running mode is inhibited, and triggering takes place only when certain pretrigger con
-
ditions are fulfilled.
The signal(s) used for initiating the arming can be applied to three channels, and the start channel can be different from the stop chan
­nel. All conditions can be set via the menu below.
Trigger Hold-Off
In this menu you can set the delay during which the stop trigger conditions are ignored after the measurement start. A typical use is to clean up signals generated by bouncing relay contacts.
Statistics
In this menu you can find the following submenus:
No. of Samples: Set the number of sam
-
ples used for calculation of various sta
-
tistical measures.
No. of Bins: Set the number of bins used in the histogram mode.
Pacing: Set the delay between succes
­sive measurements, called Pacing Time, to ON or OFF.
Operating the Counter
3-6 Description of Keys
Pacing Time: Set the pacing time to a value between 2 ms – 1000 s.
Timebase Reference
Here you can select if the counter is to use the internal or an external timebase reference. If Auto is selected, an external timebase will be used only if it is interpreted as a valid signal, i.e. both amplitude and frequency must be within specified limits. This does not imply, however, that an external reference source has to be better in any sense than the internal timebase oscillator. The EXT REF indicator at the upper right corner of the display shows that the instrument is using an external timebase reference.
Miscellaneous
The options in this menu are:
Smart Time Interval: When selected, the counter decides by means of timestamping which measurement channel is leading.
Auto Trig LF: In a value input menu you can set the lower frequency limit for au
-
tomatic triggering and voltage measure
­ments in the range 1 Hz – 100 kHz. A higher limit means faster settling time and consequently faster measurements.
Timeout
Switch the Timeout function ON or OFF. In case the input signal gets interrupted,
the timeout system (if enabled) will hold the last measurement result on screen only during the selected period of time (see next paragraph). Then the screen will be blanked, and a pending bus query will read a zero result. In case timeout is switched off when the signal gets interrupted, the display will freeze, i.e. the result of the last complete mea
­surement will stay on screen indefinitely. A pending bus query will also wait indef
­initely for a response, unless the test system controller has enabled its own timeout.
Timeout Time
Set the maximum time the instrument will wait for a pending measurement to finish before outputting a zero result. The range is 10 ms to 1000 s.
n
Math/Limit
Here you will find the menus for mathemati­cal postprocessing of the measurement result and for setting up the limit testing function.
Math
You can modify the measurement result math
-
ematically by scaling or offsetting before pre
­sentation on the display. This feature can be used for getting revolutions/min instead of Hz or for recalculating the frequency in case a de
-
vice causing frequency conversion (e.g. a
Description of Keys 3-7
Operating the Counter
multiplier or a mixer) is part of the system un
-
der test.
Select one of four formulas and enter the con
­stants K, L and M to make the counter show directly what you want, without tedious recal
­culations. X stands for the current unmodified measurement result. See the User's Manual for a closer description.
Limits
This menu is used for setting numerical limits and selecting the way the instrument will re­port the measurement results in relation to them. See the section Presentation Modes on page 3-4 for a short description or the User's Manual for a more detailed description.
n
User Options
From this menu you can reach a number of functions that are not directly involved in the measurement process.
Save/Recall Menu
Twenty complete front panel setups can be stored in non-volatile memory; the first ten of them can be user-protected. The different set
-
ups can be individually labeled to make it eas
-
ier for the operator to remember the applica
-
tion.
The following can be done:
Save Current Setup: Select one of twenty positions.
Recall Setup: Here you will find a fac­tory-programmed default setup as well as any setups you may have stored be­fore.
Modify Labels: The seven soft keys right
below the display plus the numeric input keys 6, 7, 8, 9, 0 are used for entering lower-case letters and digits much in the same way as you write SMS messages on a cell phone. Each label can hold seven characters.
Setup Protect: ON or OFF. Only setup positions 1-10 can be protected against inadvertent overwriting.
Calibration Menu
This counter has an extensive system for closed-case calibration by software. Refer to the Service Manual for details.
Operating the Counter
3-8 Description of Keys
Interface Menu
Set the active interface to GPIB or USB and enter the GPIB address.
Bus Type: Choose GPIB or USB
GPIB Mode: There are two alternative command systems, Native and Compat
-
ible. See the User's Manual and the
Programmer's Handbook for details.
GPIB Address: Enter the bus address, an integer between 0 and 31. The fac
-
tory default value is 10.
Test Menu
Different parts of the hardware can be tested by means of built-in software support.
Test Mode
You can select among the following test modes:
All (The five individual tests below are performed in sequence)
Memory (RAM) test
Memory (ROM) test
Logic hardware test
LCD & display drivers test
Interface test
Start Test
Press this key to run the selected test.
About (Information)
Read status information about the instrument.
Model
Serial number
Firmware version
Factory-installed options
Calibration date
n
Hold/Run
This key serves the purpose of manual arm
­ing. A pending measurement will be finished and the result will remain on the display until a new measurement is triggered by pressing
the RESTART key. The HOLD sign in the upper right corner of the screen indicates that no new measurements are taking place.
Pressing the key again will resume the contin
-
uous measurement mode.
n
Restart
Often this key is operated in conjunction with the HOLD/RUN key (see above), but it can also be used in free-running mode, especially when long measuring times are being used, to initiate a new measurement after a change in the input signal. RESTART will not affect any front panel settings.
Entering Numeric Values
Sometimes you may want to enter constants and limits in a value input menu, for instance after you have pressed the MATH/LIMIT key.
You may also want to select a value that is not in the list of fixed values available by pressing the UP/DOWN arrow keys. One example is Meas T ime under SETTINGS.
Whenever it is possible to enter numeric val­ues, the keys marked with digits (0,1,...9), decimal point (.)andchange sign (±)takeon this alternative numeric meaning.
It is often convenient to enter numbers using the scientific format with mantissa and exponent. When this is supported in a particu
­lar menu, the rightmost soft key will be marked EE (stands for Enter Exponent), mak
­ing it easy to switch between the mantissa and the exponent.
Press EXIT/OK to store the new value or CANCEL to keep the old one.
Description of Keys 3-9
Operating the Counter
This page is intentionally left blank.
Operating the Counter
3-10 Description of Keys
Chapter 4
Ex ercises
Preparation
For these exercises you will need a '9X' coun
­ter, a function generator and two BNC cables of approximately equal length. Set up the gen
­erator according to the following table, and connect the main output of the generator to in
-
put A of the counter.
Basic Startup
Since the counter will remember its previous setting in nonvolatile memory, it is recom
-
mended that you recall the factory default set
-
ting before you begin.
To do so, first key in the following sequence:
USER OPT ® Save/Recall ®
® Recall Setup ® Default.Thenpress
the AUTOSET key twice within two seconds. The preselected measurement function will be Frequency, and the trigger levels will be au
­tomatically adjusted in relation to the applied signal.
If a signal with a frequency of 20 Hz or higher and an amplitude large enough to trigger the counter is connected to input A, its frequency will now be displayed.
High-Resolution Frequency Measurement
Note the high resolution of the primary readout. Twelve digits are displayed in a mea
-
suring time of 1 s. Don’t worry about the trig
­ger settings, the AUTO trigger will take care of any input signal above 20 Hz. This fre
­quency limit can, if so desired, be changed down to 1 Hz or up to 100 kHz with a tradeoff between frequency and measurement speed. If the generator has been set up properly, you will now read the frequency (1 MHz) on the display.
Built-In Math Processing
With the built-in math functions, you can make post-processing operations like scaling and offsetting an easy benchtop task without having to hook up a computer to the instru
-
ment. You can, for example, display any devi
­ation from the desired value instead of directly showing the result of the measurement itself. This is known as offsetting.
In the following, the direct, unscaled result of the counter's measurement process will be re
-
ferred to as 'X'.
Exercises
4-2 Preparation
Set up the generator: Function: Square
Wave
Frequency: 1 MHz
Amplitude: 5 V
pp
Modulation: Off
TRIG
GATE
TRIG
pendulum
TIMER / COUNTER/ ANALYZE R
100ps / 300MHz
CNT-90
INPUTINPUT
INPUTA
SETTINGSETTING
SETTINGS
MATH/ LIM USER OPUSER OP
USEROPT
HOLD/ RU
HOLD/ RUN
RESTARTINPUTINPUT
INPUTB
A
300MHz
B
MAX 12Vrms 50 30Vrms 1M
!
C
MAX+30dBm
50
0.2 8GHz
!
Autoset
5V
pp
1 MHz
TRIG
GATE
TRIG
pendulum
TIMER / COUNTER/ ANALYZE R
100ps / 300MHz
CNT-90
INPUTINPUT
INPUTA
SETTINGSETTING
SETTINGS
MATH/ LIM USER OPUSER OP
USEROPT
HOLD/ RU
HOLD/ RUN
RESTARTINPUTINPUT
INPUTB
A
300MHz
B
MAX 12Vrms 50 30Vrms 1M
!
C
MAX+30dBm
50
0.2 8GHz
!
Math/Limit
To set up the counter to display any deviation from 1 MHz, press the MATH/LIM key and select Math. The display will show that Math is still Off. Press the soft key below the Math Off indicator to enter the Formula Select menu. Use the cursor keys to mark the for
-
mula K*X/M + L and confirm by pressing the EXIT/OK key.
Now enter the numerical values for the con
-
stants K, L and M. The default values are: K=1, L=0 and M=1. In this case only L has to bealteredto-1*10
6
in order to get the job
done.
Open the value input menu for L by pressing the soft key below the menu heading marked L on the display. Press 1 followed by ±, EE (short for Enter Exponent) and 6. Confirm and exit by pressing EXIT/OK.PressEXIT/OK repeatedly until the display is showing the measurement result, now modified to reflect any deviation from 1 MHz.
Change the generator frequency upward and downward just a little, and watch the counter's display.
High-Speed Measurements
The benefits of high speed measurements for benchtop use become obvious when you use statistics. For instance, the '9X' can make 1000 7-digit measurements and present the standard deviation (jitter) in less than one sec
-
ond.
Recall the default setting and press AUTOSET twice within two seconds before tweaking the controls to reach the optimum measurement speed for your application. See the following two paragraphs.
Under the INPUT Menus:
Auto trigger level settings in this model is so
fast that you will normally not notice any dif
­ference in speed if auto is on or off. However, if you use statistics to make hundreds or thou
-
sands of measurements, the fractions of a sec
­ond it takes to calculate trigger levels before each measurement add up to a considerable time over the total sequence.
The measuring speed can be increased sub
­stantially if you set the trigger levels manually once, before a lengthy measurement sequence starts.
Press INPUT A and choose MAN for the MAN/AUTO setting. Now press Trig to open the trigger level value input menu. Enter a level of +0.5 V. Confirm the selection by pressing EXIT/OK twice.
Under the SETTINGS Menu:
By making the measuring time for each sample as short as possible you can also in­crease the overall measuring speed. Remem­ber, however, that there is always a tradeoff between measuring time and resolution.
Press SETTINGS and then MeasTime.Use the DOWN arrow key to set the measuring time to the minimum value. Confirm by press
-
ing EXIT/OK twice.
Now the counter is taking measurements at a very high speed.
Toggle the STAT/PLOT key to see the statis
­tical measures and the graphical presentation of the result.
The measuring speed can be increased even more, if you take advantage of the features that are available over the GPIB. These in
­clude commands for turning off the display update, etc. In this way you can reach speeds
High-Speed Measurements 4-3
Exercises
up to 2000 transferred measurements/s, each individually triggered. If you save the results to the internal memory of the instrument for later transfer, you can even attain an impres
-
sive speed of 250,000 measurements/s.
Time Measurements
So far we have only been using the frequency function. Now we will use some others.
To measure the period of the signal, simply press the key MEAS FUNC followed by PE
-
RIOD. Select SINGLE or AVERAGE de
-
pending on the desired resolution and measur
-
ing speed. Finish by selecting the input chan
-
nel.
To measure pulse width, use the keys MEAS
FUNC and PULSE followed by POS or NEG depending on whether you want to mea-
sure the positive or the negative pulse width. Finish by selecting the input channel.
Jitter Measurements (Statistics)
You can make statistical measurements, such as pulse width jitter, directly via the front panel of the counter.
Make sure the counter is measuring positive pulse width on input A. To turn on statistics, press the STAT/PLOT key. Change the dis
­play mode by toggling the same key. Return to the numerical mode and watch the results. The display gives you a survey of several sta
­tistical measures. See also page 3-4. The rms jitter is equivalent to the standard deviation of 100 measurements, where 100 is the default value for the sample size.
If you want to change the default values for sample size, bin size or pacing time, then you can press the SETTINGS menu key and after that the STAT soft key.
Other Single-Channel Measurements
The counter/timer can help you characterize signals even further. Recall the default settings and press AUTOSET twice.
Duty Cycle
Use the MEAS FUNC key to select Duty and then Positive A (for positive-pulse duty cy­cle measurement on channel A). Change the frequency of the generator to 10 kHz. Then vary the symmetry (duty cycle) of the square wave on your generator. The counter’s display will directly show the duty cycle as a number between 0 and 1. There is no need to manually calculate duty cycle by dividing pulse width by period.
The duty cycle (D) is defined as: D=W/T
D = 0.5 for a square wave.
Exercises
4-4 Time Measurements
W
1
W
2
T
LOW
DUTY CYCLE
HIGH
DUTY CYCLE
Rise Time
Use the MEAS FUNC key to select Time, Rise Time and A (for channel A). Press IN
-
PUT A and make sure the following settings
are active:
Coupling DC
Impedance 50 W.
Attenuation 1x
Notice the rise time for a square wave, which is measured totally automatically.
Now change the generator output to a triangular waveform. See next figure.
Vary the rise time and notice the difference on the display.
Fall Time
Fall time is measured in a similar way by se
-
lecting Fall Time instead of Rise Time above.
Remarks
Varying the symmetry of the waveform will also vary the rise and fall times.
As you have noticed, there is no need to check max and min voltages and calculate 10 % and 90 % levels. The counter does it all automati
-
cally.
Volt Max/Min Measurements
The counter can also measure the peak voltage values of your input signal. Use the MEAS
FUNC key to select Volt. Then select Vpp, Max or Min as the primary function. The re
­sults of the secondary functions are calculated simultaneously and displayed in a smaller font.
Set up the Generator:
Frequency: 200 Hz
Waveform: Square
Symmetry: 50%
Amplitude: 2-5 V
Note that the voltage reading on the counter may be lower than the setting indicated on the generator, as some generators indicate the open output voltage, while the counter is now set to 50 W input impedance.
Channel Swapping
There is no separate key or function for swap­ping input channels, as found on some other counters, yet by exploring the menu trees un­der MEAS FUNC you will find that all mea
­surement functions can be performed on each of the two input channels, thus eliminating the need for a swap function.
As both input channels have the same specifi
-
cation, the definition of primary and second
-
ary channel (as seen with some other coun
-
ters) is not relevant with the '9X' from a tech
­nical point of view.
Other Single-Channel Measurements 4-5
Exercises
'  
  
'  
  
Two-Channel Measurements
The counter can measure the timing relationship between two channels with the Time Interval
AtoBand Phase A rel B functions.
Time Interval
Often function generators have a TTL output in addition to the main output. We will make use of this now. Connect a BNC cable from the TTL output to Input A on the counter and another BNC cable from the main output on the function generator to Input B.
Set up the counter: Trigger Level:
AUTO
Function: TIME INT A-B
Coupling: DC (both A and B)
You can now read the time difference between the TTL and main outputs of the generator.
Phase
So far the counter has shown the timing rela
­tionship between the TTL and the main output of the generator as a time delay. You can also show this as phase shift between the signals.
n
Using Auto to set fixed trigger levels (Auto Once)
Some measurements like Phase A rel B ben
­efit from having fixed trigger levels. This is because a change in trigger level causes a change in the measured phase shift, and Auto might change the trigger level between mea
-
surements.
To avoid calculating the trigger levels yourself, you can let the counter measure the Auto levels and then store them as fixed values.
Press INPUT A and check that Auto is still se
-
lected. Read the Auto trigger level for Input A. Select Man. Note that the automatically calculated trigger level is now entered as a fixed manual level.
Press INPUT B and run through the same steps once more to store this level as a manu
-
ally set trigger level as well.
n
Procedure
Use the MEAS FUNC key to select Phase A rel B. Now the channel delay is expressed as
phase shift in degrees.
Increase the frequency of the generator to 2 kHz, 20 kHz, 200 kHz, 2 MHz, and 20 MHz. Watch the phase difference change.
Memory Settings
See also page 3-8.
The counter has 20 memory locations in which you can store frequently used instru
-
ment settings.
To save an instrument setting, press the USER OPT key.
Select the Save/Recall key.
Press Save Current Setup, and select one of the memory locations using the LEFT/RIGHT arrow keys. Note that the first ten positions may be user-protected.
Press EXIT/OK three times to return to the normal display mode.
Now change some settings on the counter, and repeat the first steps above, until you have pressed Save/Recall.
Exercises
4-6 Two-Channel Measurements
Press Recall Setup and select the mem
­ory location in which you stored your original setup.
Press EXIT/OK three times to return to the normal display mode.
Watch the counter restore your original settings.
The memory is nonvolatile, so it will not change if you switch off the counter, nor if you restore the default setup. All settings are stored including trigger levels, mathematical constants etc. You can even attach a label to each stored setup, which helps you identify the application easier.
Auxiliary Functions
There is a collection of useful, however sel­dom used auxiliary functions. Only one will be mentioned here. See the User's Manual for a more comprehensive description. Before we go on, first recall the default setup and then press AUTOSET twice.
Time-Out
Check that the counter is measuring. Discon
­nect the Input A signal, and the measurement result will freeze on the display. Connect the signal again and the counter resumes measuring immediately.
Is this a feature? Well, yes and no. Yes, since this function gives touch-hold characteristics. No, since interrupted measurements giving false readouts may be misinterpreted by auto
-
matic test equipment and can even cause con
­trol SW to halt. The solution to this problem is the time-out function.
Press the SETTINGS key and then Misc ® Timeout Time.
Calculate the time needed to decide if there is a signal present at the input.
Enter this time using the numerical soft keys, for instance 1 s.
Activate the time-out function by toggling the soft key Timeout to ON.
Disconnect the signal and check that the counter shows '---' after one second.
Keep in mind that the value you select for this time-out must always be longer than the selected measuring time.
Summary
This concludes the Getting Started Manual. Now you have been familiarized with the con­trol structure and the display modes of the model '9X'.
There are still a great many features to dis­cover, so have a look at the User's Manual and the Programmer's Handbook, which can be found on the CD-ROM. Continue to explore the vast realm of measurement possibilities that opens up to the user of this instrument.
Auxiliary Functions 4-7
Exercises
V
service@pendulum.se
www.pendulum.se
Altaria Services Sp. z.o.o.
Loading...