Pedigo ivNow Series, ivNow-2, ivNow-1, ivNow-3 Operation & Care Manual

manual
operation & care
individual warming pods
ivNow-1
ivNow-2
ivNow-3
120V / 220v / 230v
MM-1072 (MN-38645) • Rev 1 • 04/18Printed in the U.S.A.
Table of Contents
2
Transportation and Damage Claims
Environmental Conditions
Transport and storage environmental conditions (not to exceed 15 days)
Ambient temperature range of -40°C to +70°C (-40°F to +159°F) Relative humidity range of 10% to 95%, non-condensing Atmospheric pressure range of 50kPa to 106kPa
Operational environmental conditions
The appliance must acclimate to the room temperature in the environment it will be placed—24 hours is recommended The recommended environmental temperature range is 15°C to 32°C (60°F to 90°F) The recommended relative humidity is above 20%, non-condensing.
Receipt of Appliance
The appliance has been thoroughly tested and inspected to ensure only the highest quality appliance is provided. Upon receipt, inspect for any possible shipping damage and report it at once to the delivering carrier. See Transportation Damage and Claims section.
This appliance, complete with unattached items and accessories, may be delivered in one or more packages. Confirm that all standard items and options have been received with each appliance as ordered Save all the information packed with the appliance.
Indicates that the package contents should not be used if the package has been damaged or opened.
3
Unpacking
1. Carefully remove the appliance from the carton.
NOTE: Do not discard the carton and other packaging material until you have
inspected the unit for hidden damage and tested it for proper operation.
Should damage occur in shipment, do not put the warmer into service until the
damage has been inspected by an authorized service provider
Promptly contact Pedigo customer service to report damage.
2. Read all instructions in this manual carefully before initiating the installation of this appliance.
Do not discard this manual. This manual is considered to be part of the
appliance and is to be provided to the owner or manager of the business or to the person responsible for training operators Additional manuals are available from the Pedigo service department or online at www.pedigo-usa.com.
3. Remove all protective plastic film, packaging materials, and accessories from the appliance before connecting electrical power
4
Safety Procedures
Knowledge of proper procedues is essential to the safe operation of electrically energized appliances The following hazard signal words and symbols may be used throughout this manual.
Fluid warmers are only intended for warming medical solutions for irrigation and injection prior to use Refer to the labeling of the manufacturer of the products to be warmed regarding the recommended temperature and the duration of warming No other use for this appliance is authorized or recommended.
• If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
• Appliance must be accessible Do not place the appliance in a location where it is difficult to unplug.
• This warmer is intended for use in commercial establishments where all operators are familiar with the purpose, limitations, and associated hazards of this appliance The warmer can be used wherever there is appropriate space and electrical source including patient support areas, ER, ICU, PACU, surgical suites, patient rooms, and nursing stations Do not use the warmer in the presence of flammable anesthetic mixtures (with air, oxygen, or nitrous oxide).
• Operating instructions and warnings must be read and understood by all operators and users.
• Any troubleshooting guides, component views, and parts lists included in this manual are for general reference only and are intended for use by qualified and trained technicians.
• This manual should be considered a permanent part of this appliance This manual and all supplied instructions, diagrams, schematics, parts lists, notices, and labels must remain with the appliance if the item is sold or moved to another location.
5
Installation
How to Install/Mount the ivNow
ivNow modular units can be installed in a variety of ways.
Up to 3 units can be linked together with a jumper cord and connected via a mounting plate. Configurations can be placed on counter top, mounted to wall or pole.
The rear bracket provided with the ivNow can be used to mount the pod(s) to the wall. See mounting instructions on page 5.
Note: Rear bracket provided with the ivNow is not re-
quired for countertop single pod configuration. For other mounting options (i.e. Horizontal Rails or roll stands) refer to the instructions included with the mounting kit.
6
Installation
Mounting Instructions
7
Electrical Connection
ivNow Power Requirements
Electrical Information
The power specifications are located on the unit identifi cation nameplate This nameplate is permanently attached to the unit and must be located to verify power requirements.
120/220/230 V A C — 50/60 Hz, 1 ph 100W per pod (maximum of 400W with 4 connected pods) Safety Class I Equipment No Applied Parts Mode of Operation: Continuous
Applicable only on units with an equipotential-bonding terminal:
To prevent an electrical shock hazard between the appliance and other appliances or metal parts in close vicinity, an equipotential­bonding terminal is provided. An equalization bonding lead must be connected to this stud and the other appliances / metal parts to provide sufficient protection against potential difference. The terminal is marked with the following symbol.
Grounding reliability can only be achieved when
equipment is connected to an equivalent receptacle
marked “Hospital Grade.”
NEMA 5-15P 15A - 125V Plug Hospital Grade
8
Operation
General Information
The ivNow fluid warmer quickly warms and maintains the temperature of injection/intravenous solutions prior to their use. The specially contoured warming module cradles solution bags in 0.5-, 1-, 2- & 3-liter sizes. The control can easily be set to display temperatures in Celsius or Fahrenheit. A sensor in the heating plate detects the presence of a bag and engages the heating mechanism to quickly begin warming the fluid. Two (2) temperature sensors work in unison to precisely and continuously read the temperature of the bag and another sensor monitors the plate temperature. The heater will reengage as necessary to maintain the temperature within +0/-2°C (+0/-3ºF) of the setpoint. The electronic control monitors the length of time the bag has been held at temperature, beginning when the bag reaches setpoint temperature. The display shows the time the fluid has been held at temperature.
NOTE: In the event that fluid should spill inside the
cavity, unplug the unit to prevent an electrical shock hazard. Wipe excess fluid from module immediately. Refer to qualified service personnel. Qualified service personnel should remove the module control and remove any remaining liquid. Perform necessary hospital electrical safety checks before returning the unit to operation.
Safety Feature
• The control of the ivNow is designed to display an error
message (E-31) and stop heating if the temperature of the fluid bag (as monitored by the dual sensor switch) is ever above 40°C (104°F).
• The control also monitors the temperature of the
aluminum and limits the temperature to a maximum of 60°C (140°F).
• There are two protective devices wired in series
with the heating element; an automatic thermostat and a thermal fuse that disconnect the power to the heating element if the plate exceeds 65°C (149°F) and 84ºC (183ºF) respectively, in the event of a run away condition.
ivNow Capacity Information
The ivNow warming module accommodates fluids packaged in bags. Each warming module cradles solution bags in 0.5-, 1-, 2- and 3-liter sizes.
9
Operation
power
display
status indicator
light
status button
Operating Instructions
1. Before operating the module(s), clean the exterior of the unit with a damp cloth and general hospital cleaner (isopropyl alcohol).
2. Plug the appliance into an appropriate hospital-grade receptacle. See the appliance data tag for voltage requirements.
3. Turn ON by pressing the power button located on the front of the warmer. When on, the control display will be powered. The plate will not heat until a bag is placed in module.
4. Activate control by placing bag of fluid in module. See Figure 1. For optimum performance, the bag should be centered over the sensor.
NOTE: The warm-up stabilization time will vary slightly
depending on the bag temperature and ambient room temperature.
5. The display will continue to show the current bag temperature as the fluid is heated.
6. The green status indicator will illuminate when the fluid is within +0/-2°C (+0/-3°F) of the setpoint temperature.
10
Operation
power
display
status indicator
ivNow Status Indicators
Fluid bag is not present
The DISPLAY will show 4 dashes.
Fluid bag detected, measureing initial bag temperature
The DISPLAY will show a chasing dash.
Warming the fluid bag
The DISPLAY will show the current bag temperature.
Fluid bag temperature is within tolerance
The STATUS INDICATOR LIGHT will be illuminated and DISPLAY will show the current bag temperature.
Fluid bag temperature is over set point
The DISPLAY will flash current bag temperature. The display will continue flashing until the temperature falls below the set point. Do not use fluid bag until the display stops flashing.
Fluid bag heated residence time is in excess of limit
The DISPLAY will flash the word “DATE”. ivNow will monitor the time spent continuously heating a fluid bag and notify the user that the time limit has been exceeded. Remove the fluid bag from the warmer and check the expiration date. Discard fluid if expired.
Internal error is detected
The DISPLAY will show an error code starting with “E-” followed by a number. Refer to the Troubleshooting Guide for the error code description.
Display the set temperature
Press the STATUS button to show the set temperature.
Display the time spent at the set temperature
Press and hold the STATUS button for 2 seconds to see how long a bag has been at the set temperature. The first hour is indicated in minutes and seconds “MM:SS” and subsequent time is indicated in hours and minutes “HH:MM” for the first 24 hours. After 24 hours the time is displayed in days and fractions of a day “DD.DD”. The timer stops after 30.00 days. The time a bag is held at the set temperature will remain in memory until a new bag is placed in the ivNow.
status button
light
ivNow Adjustment
Display the firmware version
Press and hold the POWER button for 10 seconds. Press the POWER button to return to operation.
Temperature unit selection
Press and hold the POWER button for 20 seconds or until the current unit (default of °C) is displayed. When displayed, press the STATUS button to toggle the temperature unit between °C and °F. Press POWER to accept the change.
Heated residence time limit
Press and hold the POWER button for 30 seconds or until the current limit (default of 15) is displayed. When displayed, press the STATUS button to select a date range between 7 and 60 days in 1 day increments. Press the POWER button to accept the change.
Temperature set point
Press and hold the POWER button for 40 seconds or until the current set point (default of 40 °C or 104°F) is displayed. When displayed, press the STATUS button to select desired temperature set point between 35°C and 40°C (95°F and 104°F) in 1 degree increments. Press the POWER button to accept the change.
11
Cleaning and Preventative Maintenance
The cleanliness and appearance of this equipment will contribute considerably to its operating efficiency. Make certain the module is kept free of any debris that may accumulate. Good equipment that is kept clean works better and lasts longer.
Clean the Unit Regularly:
1. Turn the unit off.
2. Disconnect the module from the power source.
3. Wipe the metal, plastic, control surfaces, and sensor switch of the module with a cloth dampened with isopropyl alcohol or 10% bleach solution to clean and disinfect the unit Avoid the use of abrasive or corrosive cleaning compounds Avoid contact with the electrical connections and electrical components.
4. Wipe surfaces with a cloth dampened with clean, warm water.
5. Wipe dry with a clean cloth Always follow appropriate state or local health (hygiene) regulations regarding all applicable cleaning and sanitation requirements.
NOTE: In the event that fluid should spill inside the module, unplug the unit to prevent an electrical shock hazard. Wipe excess fluid from module immediately. Refer to qualified service personnel. Qualified service personnel should remove the module control and remove any remaining liquid Perform necessary hospital electrical safety checks before returning the unit to operation.
Annual Preventative Maintenance
1. Ensure that the correct Operation and Care Manual is available to all users.
2. Ensure that all users have been properly trained in unit’s operation.
3. Inspect condition of plug and cord. Replace if damaged.
4. Clean dust from the unit.
5. Check condition of wall mounting hardware. Ensure mounting screws and assembly are secure.
6. Check control panel overlay condition. Are there any tears or excessive wear on the graphic? Does the control work properly when buttons are pushed?
7. Check that all display LEDs light up. Test by turning the unit off and then on All LEDs illuminate for one second at start-up.
8. Contact the Pedigo Service department for immediate repair if any problems exist.
12
Troubleshooting
UTION
If your unit is not operating properly, check the following before calling your authorized service agent. Check the power applied to the unit. Is the plug in the outlet? Check to make sure the power is on by pressing the power button on the front of the unit. If it is a multi-pod unit, are the jumper cords between each pod fully inserted?
Do not attempt to repair or service beyond this point. Repairs made without prior authorization by manufacturer will void the warranty on the unit.
CA
Troubleshooting Guide
This chart is provided for the assistance of qualified technicians only and is not intended for use by untrained or unauthorized service personnel.
13