This product complies with the requirements of the following European Community
Directives: 2014/130/EU (Electromagnetic Compatibility) and 2013/35/EU (Low Voltage)
as amended by 2014/32/EU (CE-Marking).
Overvoltage category II; pollution degree 2.
To ensure safe operation of the equipment and eliminate the danger of serious injury due
to short-circuits (arcing), the following safety precautions must be observed.
Damages resulting from failure to observe these safety precautions are exempt from any
legal claims whatever.
* The instrument must be set up so that the power plug can be removed from the
socket easily.
* Prior to connection of the equipment to the mains outlet, check that the available
mains voltage corresponds to the voltage setting of the equipment.
* Connect the mains plug of the equipment only to a mains outlet with earth connection.
* Do not place the equipment on damp or wet surfaces.
* Do not cover the ventilation slots of the cabinet to ensure that the air is able to
circulate freely inside.
* Do not insert metal objects into the equipment by way of the ventilation slots.
* Do not place water filled containers on the equipment (danger of short-circuit in case
of knock over of the container).
* Do not exceed the maximum permissible input ratings (danger of serious injury and/or
destruction of the equipment).
* To avoid electric shock, disconnect power to the unit under test and discharge all
capacitors before taking any resistance measurements.
* Check test leads and probes for faulty insulation or bare wires before connection to
the equipment.
* To avoid electric shock, do not operate this product in wet or damp conditions.
Conduct measuring works only in dry clothing and rubber shoes, i. e. on isolating
mats.
* Comply with the warning labels and other info on the equipment.
* The measurement instrument is not to be to operated unattended.
* Do not subject the equipment to direct sunlight or extreme temperatures, humidity or
dampness.
2
* Do not subject the equipment to shocks or strong vibrations.
* Do not operate the equipment near strong magnetic fields (motors, transformers etc.)
* Keep hot soldering irons or guns away from the equipment.
* Allow the equipment to stabilize at room temperature before taking up measurement
(important for exact measurements).
* Do not input values over the maximum range of each measurement to avoid
damages of the meter.
* Use caution when working with voltages above 35V DC or 25V AC. These Voltages
pose shock hazard.
* Periodically wipe the cabinet with a damp cloth and mid detergent. Do not use
abrasives or solvents.
* The meter is suitable for indoor use only
* Do not operate the meter before the cabinet has been closed and screwed safely as
terminal can carry voltage.
* Do not store the meter in a place of explosive, inflammable substances.
* Do not modify the equipment in any way
* Do not place the equipment face-down on any table or work bench to prevent
damaging the controls at the front.
* Opening the equipment and service – and repair work must only be performed by
qualified service personnel
* -Measuring instruments don’t belong to children hands-
Cleaning the cabinet
Prior to cleaning the cabinet, withdraw the mains plug from the power outlet.
Clean only with a damp, soft cloth and a commercially available mild household cleanser.
Ensure that no water gets inside the equipment to prevent possible shorts and damage
to the equipment.
3
2. Quick Start
If it’s the first time for y ou to use the generator or you have no time to read the
guide carefully, you can get the basic operation as soon as you finish browsing
the chapter1. If more complicated functions are needed or meet difficulties in
operation, please read Features and Functions in chapter 3.
2.1 Prepare the waveform generator for use
2.1.1 Check the list of supplied items
Verify that you have received the complete unit according to the packing list. If
you find package damaged badly, leave it until the instrument passes
performance test. If anything is missing, please contact sales office
2.1.2 Connect the power
Turn on the instrument only the following conditions are met,
Power: AC 100 ~ 240 V
Frequency: 45 ~ 65 Hz
Power consumption: < 30VA
Temperature: 0 ~ 40°C Humidity: <80%
Plug the power cord into an AC100 ~240V socket with ground wire and press On
/Off switch below socket on rear panel. Then blinking power button on front
panel indicating the generator well connected with power but still in off state.
Only press power button, the generator initializes itself and obtains the default
parameters, outputting continuous Sine under normal working state, with
signal’s parameters displayed.
4
Warning: In order to ensure the security of the operator, use triple- core power
socket with ground wire.
While, 【Utility】key is used to set common parameter and 【Output】key is used to enable
or disenable output port.
12 keys in numeric keypad:
【0】【1】【2】【3】【4】【5】【6】【7】【8】【9】are used to enter numbers.
【.】is used to enter decimal point and 【-】is only available to enter allowed minus.
Four arrow keys:
【<】【>】were used to move the cursor left or right.
【∧ 】【∨ 】were used to increase and decrease the displayed number when setting
frequency and amplitude.
The balance 6 keys under display are menu operation softkeys, embraced with 〖〗 and
used to select menu or unit.
4.1.2 Display description
The display screen is divided into three sections:
Top section: output waveform
Middle section: parameters display of Frequency or Amplitude or Offset etc.
Bottom section: display of menu or unit.
8
1. Voltage Scale 2.Waveform 3. Amplitude 4. Time Scale
5. Output Information 6. Working Parameters 7. Operation Menu
4.2 Number Entry
4.2.1 Use the keypad to enter numbers and softkey to select unit.
Please use 【<】key to cancel mistake enter before select unit key. Don’t forget to select
unit key after finishing entering number. Only this, the enter data can come into effect.
Press 〖Cancel〗softkey in unit menu to cancel the enter data that has come into effect.
4.2.2 Use the knob and arrow keys to modify the displayed number.
Use the arrow keys 【<】【>】to move the cursor left or right. While, rotate the knob to
change digits. (clockwise to increase and the inverse to decrease). It’s not necessary for
user to select unit if entering number by this way.
Users can enter numbers by one of three mentioned methods based on different
applications.
4.3 Basic Operation
4.3.1 To select the output channel
Press 【CHA/CHB】key to open desired channel configuration screen. Note that fonts of
channel name, mode and waveform were indicated by different color (yellow for CHA
and blue for CHB). Use softkey together with knob or numeric keypad to set the
waveform and parameters of desired channel. Enable or disable the output signal of
desired channel by pressing 【Output】key.
4.3.2 To select a waveform
Press 【Waveform】key to see the first page of the option list Sine, Square, Ramp, Pulse,
Arbitrary and Noise. If select Arbitrary option, there are some operations for arbitrary
waveform. Select one desired waveform to see the waveform diagram under Continuous
mode.
9
Press〖Waveform〗key again to back to current menu.
4.3.3 To set duty cycle
For example, specify a duty cycle of Square to 20%.
Press 〖Duty Cycle〗softkey and then set the duty cycle to 20% using the numeric keypad
or the knob and arrow keys 【<】【 >】. If you use the keypad, press 〖%〗softkey to finish
entering the value.
4.3.4 To set Frequency
For example, specify a frequency to 2.5kHz.
Press 〖Freq/period〗softkey and then set the frequency to 2.5kHz using the numeric
keypad or the knob and arrow keys 【<】【 >】. If you use the keypad, press 〖kHz〗 softkey
to finish entering the value.
4.3.5 To set Amplitude
For example, specify amplitude to 1.6Vrms.
Press 〖Ampl/High〗softkey and then set the amplitude to 1.6Vrms using the numeric
keypad or the knob and arrow keys 【<】【>】. If you use the keypad, press 〖Vrms〗
softkey to finish entering the value.
10
4.3.6 To set offset
For example, specify an offset to -25mVdc.
Press 〖Offset〗softkey and then set the offset to -25mVdc using the numeric keypad or
the knob and arrow keys 【<】【>】. If you use the keypad, press 〖mVdc〗 softkey to
finish entering the value.
4.3.7 To output an AM waveform
A modulated waveform consists of a carrier and a modulation waveform. In AM, for
example, you want to output an AM waveform with 80% modulation depth, the carrier will
be 10kHz and the modulation waveform will be a 10Hz Ramp wave.
1. Select AM
Press 【Modulate】key and then select 〖AM〗softkey by pressing the〖Mod Type〗softkey.
2. To set Carrier frequency
Press〖Carrier〗to enter into the parameter setting. Then press〖Freq〗softkey, enter 10
kHz with the numeric keypad or the knob and arrow keys. Press 〖kHz〗softkey to finish
entering the number.
3. To set the modulation depth
11
Press 〖Return〗key to back to parameter setting and press the 〖Depth〗softkey and then
set the value to 80% using the numeric keypad or the knob and arrow keys. Press 〖%〗
softkey finish entering the number if you are using the numeric keypad.
4. To set modulating waveform frequency
Press 〖AM Freq〗softkey and then set the value to 10 using the numeric keypad and
finally press〖Hz〗softkey.
5. To select the modulation waveform shape
Press the 〖Shape〗softkey then press 【Ramp】key to set the shape of modulating
waveform as Ramp.
6. To set AM parameters by knob or arrow keys
4.3.8 To output an Sum waveform
If you want to output a Sum waveform with amplitude 10% and modulation waveform is
Ramp.
1. Select Sum
Press 【Modulate】key and then select 〖Sum〗softkey by pressing the 〖Mod Type〗
softkey.
2. To set the Sum amplitude
Press the 〖Sum Ampl〗softkey and then set the value to 10 using the numeric keypad
and finally press〖%〗softkey, or set by the knob and arrow keys.
3. To select the modulation waveform shape
Press the 〖Shape〗 softkey then press 【Ramp】 key to set the shape of modulating
waveform as Ramp.
4. Generator output a sum waveform as preset modulating parameters and you also
can adjust amplitude of summed ramp by 【<】or 【>】keys.
12
4.3.9 To output a FSK waveform
If you want to output a FSK waveform with hop frequency to 300Hz, with an FSK rate of
50Hz,
1. Select FSK
Press 【Modulate】key and then select 〖FSK〗softkey by pressing the 〖Mod Type〗
softkey.
2. Set the hop frequency
Press the 〖Hop Freq〗softkey then set the value to 300Hz using the numeric keypad or
the knob and arrow keys.
3. Set the FSK rate
Press the 〖Hop Rate〗softkey then set the value to 50Hz using the numeric keypad or
the knob and arrow keys.
4. Set the hop frequency and FSK rate by knob or arrow keys.
4.3.10 To output a Frequency Sweep
If you want to output an swept waveform with sweep time of 5 second, logarithmic
sweep,
1. Select Frequency sweep
13
Press 【Sweep】key and then verify that the Sweep is selected in default.
2. Select sweep time
Press the 〖Sweep Time〗softkey then set the value to 5s using the numeric keypad or
the knob and arrow keys.
3. Select sweep mode
Press the 〖Mode Line/Log〗softkey and then verify that the logarithmic sweep mode is
currently selected on the first softkey.
4.3.11 To output a Burst waveform
If you want to output a five-cycle wave with a 10ms burst period, continuous or manual
signal trigger.
1. Press 【Burst】key then burst menu is shown in the screen of the current channel.
2. Press 【Burst Mode】and select the ‘Triggered’.
3. Press 〖Burst Period〗softkey then set the value to 10ms using the numeric keypad
or the knob and arrow keys.
4. Press 〖N Cycle〗 softkey then set the value to 5 using the numeric keypad or the
knob. Press the 〖Ok〗 softkey to finish data entry if you are using the numeric
keypad.
At this point, the waveform generator outputs a continuous five-cycle burst at 10 ms
intervals.
14
You also can generate signal burst (still with 5 cycles) by selecting Manual Trig under
pressing 〖Source Int/Ext〗softkey. Then each time you press 〖Manual Trig〗softkey, one
five-cycle burst will be output. Press number depends on the desired burst number.
4.3.12 Frequency Coupling
If you want to couple frequencies between the two channels,
1. Press 【Dual Channel】key. The dual channel menu is shown in the current screen.
2. Press the 〖Freq Cpl〗softkey to turn frequency coupling on. Then press 〖More〗
softkey to configure frequency coupling.
3. Press 【 Continuous】 key to configure CHA frequency, just because channel
frequencies can be linked with a constant ratio or difference between them, CHB
frequency will be changing as CHA changes.
4.3.13 Frequency Counter
If you want to measure a frequency of external signal,
1. Press 【Counter】key to enter into Counter mode and display the counter menu and
count results.
2. Input measured signal through ‘Counter’ connector at rear panel.
3. Press 〖Meas Type〗softkey, the generator begin to measure then display the
measured frequency value.
4. For Square input, press 〖Duty cyc〗softkey to enable the duty cycle measurement
and display the duty cycle.
15
Key
Working Mode
【Continuous】
Continuous output
【Modulate】
Modulate output
【Sweep】
Sweep output
【Burst】
Burst output
【Dual Channel】
Dual channel operation
【Counter】
Frequency counter
No.
Waveform
No.
Waveform
Standard Waveforms 5
00
Sine
03
Pulse
01
Square
04
Noise
02
Ramp
Math Waveforms 36
05
PRBS
23
LogRise
06
PosDC
24
LogFall
07
NegDC
25
PosSquare
08
Cosin
26
NegSquare
5.Features and Functions
This chapter makes it easy to look up details about a particular feature of the waveform
generator. It also covers the front-panel operations. You should read Chpter2
‘Front-panel Menu Operation’ first and master the basic operation so as to get a better
understanding for this chapter.
5.1 Working Mode
There are six working modes for the waveform generator,
While,
CHA and CHB both cover four modes: continuous output, modulate output, sweep output
and burst output.
Notice that the modulate output covers thirteen types: AM, FM, PM, PWM, SUM, FSK,
QFSK, 4FSK, PSK, QPSK, 4PSK, ASK, OSK.
And sweep output covers three types: Frequency Linear Sweep, Frequency Logarithm
Sweep and Frequency List sweep.
5.2 Continuous Output
5.2.1 Waveform Selection
The waveform generator can output 150 waveforms, see below table,
16
09
Tangent
27
PosCube
10
CoTangent
28
NegCube
11
ArgSine
29
SquareRoot
12
ArgCos
30
PosRecipro
13
ArgTan
31
NegRecipro
14
ArgCoTan
32
PNRecipro
15
SineH
33
BiRecipro
16
CosinH
34
PosSemicirc
17
TangentH
35
NegSemicirc
18
CoTangentH
36
Gaussian
19
HaverSine
37
Maxwell
20
Sinc
38
Lorentz
21
ExpRise
39
Laplace
22
ExpFall
40
Besell
Linear Waveforms
41
PosPulse
59
PNTriangl
42
NegPulse
60
HiLoTri
43
PN_Pulse
61
LoHiTri
44
PosBiPulse
62
PosRiseRamp
45
NegBiPulse
63
PosFallRamp
46
PNBiPulse
64
RiFaRamp
47
PMulPulse
65
NegRiseRamp
48
NMulPulse
66
NegFallRamp
49
PNMulPulse
67
FaRiRamp
50
WidePulse
68
Trapezia
51
NarrowPulse
69
RiseStair
52
WiNaPulse
70
FallStair
53
HiLoPulse
71
RiFaStair
54
RisePulse
72
RiStariRamp
55
FallPulse
73
FaStariRamp
56
RiFaPulse
74
Spiry
57
PosTriangl
75
Swallow
58
NegTriangl
76
Chair
Combine Waveforms 40
77
PAllSine
97
SineFSK
78
NAllSine
98
SinePSK
79
PHalfSine
99
SineSum
80
NHalfSine
100
SineSweep
81
SiAmplCut
101
AmplInc
82
BiAmplCut
102
AmplDec
83
SiPhaselCut
103
BurstNoise
84
BiPhaselCut
104
BurstSine
85
SinePulse
105
LowPass
17
86
NoisePulse
106
HighPass
87
BiHarmo
107
BandPass
88
TriHarmo
108
BandPit
89
FourthHarmo
109
PulseOSC
90
FifthHarmo
110
PulseOver
91
SineFM
111
PNCircle
92
SineAM
112
Tripagoda
93
SquareAM
113
Candela
94
NoiseAM
114
ExpSquare
95
PulsePWM
115
ExpSine
96
SineFSK
116
TanSquRoot
Special Waveforms 32
117
TanArcTan
133
Cardiac2
118
ReciInvReci
134
NearQuake
119
HarmInvHarm
135
FarQuake
120
BiReciHarm
136
Blast
121
BiReciCircle
137
Shake
122
CubeGause
138
LandScape
123
TanHarm
139
Cloud
124
HalfBiReci
140
Camel
125
Charge
141
User_arb1
126
Stress
142
User_arb2
127
HeatTreat
143
User_arb3
128
MulHarm
144
User_arb4
129
Syntony
145
User_arb5
130
Stereo
146
User_arb6
131
RainFall
147
User_arb7
132
Cardiac1
148
User_harmo
Edit Waveform 1
149
Edit_wave
(1) 00~04 are standard common used waveforms.
05~140 are built-in waveforms which are used in special occasion.
141~147 are user-defined arbitrary waveforms, allows to be saved by users after
creating with software.
148 is used to save the user-defined harmonic wave.
149 is edit waveform, used to edit and modify arbitrary wave or harmonic wave. If
not save, the data will be lost once exit this function.
(2) Press 【Waveform】 key to see the waveform menu, select Sine, Square, Ramp,
Pulse or Noise from the options. Press〖Arb〗key and then〖Built-in〗option to select
built-in waveforms from 150 waveforms.
18
Built-in waves have five types: Standard, Math, Linear, Combine and Special. Press
the option to enter into the waveform list and then select desired waveform by
left/right key or knob. Finally press〖Confirm〗to output selected waveform from the
output port, the sketch also be shown at the display.
Press 〖Back〗key to return to type menu and you can select other types of built-in
waveform again. Press〖Back〗key again to exit the built-in waveform menu. Or
press 【Waveform】key to return to current function window.
(3) The selected waveform diagram is shown but which is just a sample one with lower
resolution. Observe and test the output waveforms using oscilloscope.
5.2.2 Duty Cycle (Square or Pulse)
The duty cycle represents the fraction of time per cycle that the square wave is at a high
level. Press 【Waveform】key and choose Square or Pulse, press 〖Duty Cycle〗softkey
after selecting 【Continuous】key then set the desired value of duty cycle. Usually, the
duty cycle value remains unchanged if the Square frequency is changing, but duty cycle
is limited by the edge time if output frequency is too high, which should comply with the
below formula:
10ns ≤ ( Duty Cycle × Period ) ≤ ( Period-10ns )
5.2.3 Pulse Width and Edge Time
The pulse width represents the time from the 50% threshold of the rising edge of the
pulse to the 50% threshold of the next falling edge. After selecting the pulse function,
press the 〖Pulse Width〗 softkey. Then use the knob or numeric keypad to enter the
desired pulse width. The pulse width usually remains unchanged if the Square frequency
is changing, but will be limited by the edge time if output frequency is too high, which
should comply with the below formula:
10ns≤Pulse Width≤ Period-10ns
The edge time represents the time from the 10% threshold of rising/falling edge to 90%
of which. If select Pulse waveform in continuous menu, press 〖Edge Time〗 to select the
Edge parameter to set. The edge time setting will be also limited by pulse width, which
should comply with the below formula:
Edge Time≤0.625×Pulse Width
Edge Time≤0.625×Duty Cycle×Period
19
5.2.4 Symmetry (Ramp)
Symmetry represents the fraction of time per cycle that the ramp wave is rising. After
selecting the ramp, press 〖Ramp Symmetry〗then set the desired value of symmetry.
The symmetry will remain unchanged if output frequency is changing. Rising Ramp will
be shown when symmetry is 100%, falling ramp will be shown when symmetry is 0%,
triangle wave display when symmetry is 50%.
5.2.5 Set Frequency
The output frequency range depends on the function currently selected and the upper
limit for sine depends on the model selected. The minimum frequency is 1μHz for all
functions. For detail specification, please see chapter 5. If you change to a function
whose maximum frequency is less than that of the current function, the frequency is
adjusted to the maximum value for the new function. The distortion will go increasing as
frequency rising.
To set the output frequency, press the 【Continuous】key, then the 〖Freq/Period〗
softkey for the selected function. Then use the knob or numeric keypad to enter the
desired frequency. Or press 〖Freq/Period〗 softkey again to set using Period. For
internally applying to frequency synthesis, the display period value is the conversed one
from enter value.
5.2.6 Set Amplitude
You can set amplitude by ‘amplitude’ or ‘level’. If amplitude set is selected, high level and
low level of signal will change at the same time when changing amplitude but DC offset
remain unchangeable. If level set is selected, no matter high level or low level changing,
low or high level remain unchangeable but DC offset will change. The relationships
between Vpp, High, Low and offset are shown below:
Vpp=High-Low High=Offset+Vpp/2 Low=Offset-Vpp/2
In continuous menu, press softkey 〖Ampl/High lev〗to set amplitude or high level. Press
softkey 〖Offset/Low lev〗 to set dc offset or low level.
5.2.7 Amplitude Limitation:
The output amplitude is limited by following factors. Once over the limitation, the
generator will modify the setting as its allowed maximum within the limitation.
(1) Limit Level: press 〖Ouput Menu〗softkey and then〖High Lev Limit〗softkey specify
the limit value of high level. Press〖Low Lev Limit〗softkey specify the limit value of
low level. Even if wrong operation that exceed the limit value, the generator won’t be
20
Waveform
Vpp
Vrms
Sine
2.828Vpp
1Vrms
Square, Pulse
2Vpp
1Vrms
Ramp
3.464Vpp
1Vrms
Vpp
Vrms
dBm
10.0000 Vpp
3.5356 Vrms
23.979 dBm
6.3246 Vpp
2.2361 Vrms
20.000 dBm
2.8284 Vpp
1.0000 Vrms
13.010 dBm
2.0000 Vpp
707.1 mVrms
10.00 0dBm
1.4142 Vpp
500.0 mVrms
6.9897dBm
damage and perform within the limit value. But if specify the high level value to be
+10Vdc and low level to be -10Vdc, the limit function won’t work anymore.
(2) DC offset: except that set DC offset as 0, the amplitude is only limited by limit level,
otherwise, which is further limited by DC offset, as following regulation:
DC offset + Vpp/2≤High Limit
DC offset – Vpp/2≥Low Limit
(3) Frequency: if frequency high enough, allowed maximum amplitude will be limited
(refer to chapter 5, Specifications)
(4) Channel Bandwidth: the output amplitude will decrease if frequency is higher. So
flatness compensation will be needed to ensure the accurate amplitude in
continuous output. But for other functions, once the frequency over 10MHz, the
amplitude will decrease.
(5) For the arbitrary waveform generator, if Vpp don’t reach to full range, the display
value is not agreed with the output value.
5.2.8 Amplitude Units: You can set the output amplitude in Vpp, Vrms, or dBm. Vpp is
available for all functions. For sine, square, ramp and pulse, Vrms can also be used. The
output units for amplitude can also be set to dBm if the external load is currently set to
‘non High Z’.
(1) In continuous mode, press〖Ampl Unit〗select units if current waveforms and load
condition allows. Different unit key will enable the different format displaying.
(2) For different waveforms, the convert relationship between Vrms and Vpp is subject
to, see below table:
(3) The relationship among dBm and Vrms and Vpp is subject to waveform and load,
dBm=10×log10(P/0.001), while, P=(Vrms)2/Load
If waveform is sine, set 50Ω load, the convert among three output units is shown below,
21
632.5 mVpp
223.6 mVrms
0.00 dBm
282.9 mVpp
100.0 mVrms
-6.9879dBm
200.0 mVpp
70.7 mVrms
-10.000 dBm
10.0 mVpp
3.5 mVrms
-36.020 dBm
5.2.9 Set DC offset
Press 〖Offset/Low lev〗softkey then set the desired offset value using the knob or
numeric keypad. Here, enter by knob is strongly recommend because of more
convenience feature. The DC offset setting will limit by amplitude and level, which should
be agreed with below formula:
Low Limit+Vpp/2≤Offset≤High Limit-Vpp/2
If the specified offset out of the range, the waveform generator will automatically adjust it
to the maximum DC voltage allowed with the amplitude specified.
If set amplitude to 0.2mVpp, high level limit to +10Vdc and low level limit to -10Vdc, then
the offset can be set within ±10V range. The waveform generator become the DC power
supply and can output DC voltage signal. But the thing to be noted is that the output
impedance is 50Ω.
5.2.10 Set Phase
Press 〖Phase〗softkey to select ‘Phase’ then set the desired phase using the numeric
keypad or knob.Output phase means the phase difference between output signal and
synchronous signal, and output signal advance to synchronous signal.
If frequency is same for CHA and CHB, press 〖Output Menu〗softkey to select 〖Phase
Sync〗to make the synchronous signal of CHA and CHB are with same phase. So it’s
easy to calculate the phase difference of two channels on basis of the phase setting for
CHA and CHB.
5.2.11 Set Polarity
In continuous mode, press 〖Output Menu〗and then〖Polarity〗softkey to set the polarity
of output signal.
(1) If select Normal, it means output waveform same as sketch. Take Sine for example,
normal polarity means output waveform start from zero phase and voltage goes
rising.
(2) If select Inverse, it means output waveform will be take voltage 0V as axes, and
generates the vertical image opposite to Normal polarity.
(3) Polarity setting has no influence to DC offset volts and Sync signal.
22
Loading...
+ 50 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.