PEAKTECH 1370, 1355, 1356, 1362, 1340 Instructions

...
PeakTech ® 1340 - 1375
User Manual 2 CH & 4 CH
Digital Storage Oscilloscopes
i
Table of Contents
1. General Safety Requirements .......................................................................................... 1
2. Safety Terms and Symbols ............................................................................................... 2
Introduction to the Structure of the Oscilloscope ..................................................................... 5
Front Panel ............................................................................................................................................... 5
Rear Panel ................................................................................................................................................ 6
Control Area ............................................................................................................................................. 7
User Interface Introduction ........................................................................................................ 8
How to Implement the General Inspection ............................................................................. 10
How to Implement the Function Inspection ............................................................................ 10
How to Implement the Probe Compensation .......................................................................... 11
How to Set the Probe Attenuation Coefficient ........................................................................ 12
How to Use the Probe Safely..................................................................................................... 13
How to Implement Self-calibration .......................................................................................... 14
Introduction to the Vertical System ......................................................................................... 14
Introduction to the Horizontal System .................................................................................... 15
Introduction to the Trigger System ......................................................................................... 16
Touchscreen Controls (Touchscreen is optional) ................................................................... 17
Magnifier Function (Only for specific models with touchscreen) ......................................... 21
4. Advanced User Guidebook ............................................................................................ 24
How to Set the Vertical System ................................................................................................ 25
Use Mathematical Manipulation Function ............................................................................. 27
Waveform math .................................................................................................................................... 28
User defined function ............................................................................................................................. 29
Digital Filter ........................................................................................................................................... 29
Using FFT function ................................................................................................................................ 30
Use Vertical Position and Scale Knobs .................................................................................... 32
How to Set the Horizontal System ........................................................................................... 33
Zoom the Waveform .............................................................................................................................. 33
How to Set the Trigger/Decoding System ................................................................................ 34
Single Trigger ......................................................................................................................................... 34
Logic Trigger.......................................................................................................................................... 43
Bus Trigger ............................................................................................................................................. 44
Bus Decoding (Optional)........................................................................................................................ 50
How to Operate the Function Menu ........................................................................................ 55
How to Implement Sampling Setup ....................................................................................................... 55
How to Set the Display System .............................................................................................................. 56
ii
How to Save and Recall a Waveform ..................................................................................................... 59
How to Record/Playback Waveforms .................................................................................................... 66
How to Clone and Recall a waveform .................................................................................................... 70
How to Implement the Auxiliary System Function Setting .................................................................... 74
How to Update your Instrument Firmware............................................................................................. 78
How to Measure Automatically.............................................................................................................. 80
How to Measure with Cursors ................................................................................................................ 84
How to Use Autoscale ............................................................................................................................ 87
How to Use Built-in Help ....................................................................................................................... 88
How to Use Executive Buttons............................................................................................................... 89
How to Print the Screen Image ............................................................................................................... 91
5. Use the Arbitrary Function Generator (Optional to specific models) ............................... 92
Output Connection .................................................................................................................... 92
To Set Channels ......................................................................................................................... 92
To Set Signals ............................................................................................................................. 93
To Output Sine Signals........................................................................................................................... 93
To Set the Frequency ........................................................................................................................................... 93
To Set the Period .................................................................................................................................................. 94
To Set the Start Phase .......................................................................................................................................... 94
To Set the Amplitude ........................................................................................................................................... 94
To Set the Offset .................................................................................................................................................. 95
To Set the High Level .......................................................................................................................................... 95
To Set the Low Level ........................................................................................................................................... 95
To Output Square Signals ....................................................................................................................... 95
To Output Ramp Signals ........................................................................................................................ 95
To Set the Symmetry of Ramp ............................................................................................................................. 95
To Output Pulse Signals ......................................................................................................................... 95
To Set the Pulse Width of Pulse ........................................................................................................................... 95
To Set the Duty Cycle of Pulse ............................................................................................................................ 95
To Output Arbitrary Signals ................................................................................................................... 95
Create a New Waveform ...................................................................................................................................... 96
File Browse .......................................................................................................................................................... 97
Built-in Waveform ............................................................................................................................................... 97
6. Use the Multimeter (Optional) .................................................................................... 100
Input Terminals ....................................................................................................................... 100
DMM Menu .............................................................................................................................. 100
DMM Information Window ................................................................................................... 101
Making Multimeter Measurements ....................................................................................... 102
Measuring AC or DC Current .............................................................................................................. 102
Measuring AC or DC Voltage .............................................................................................................. 102
Measuring Resistance ........................................................................................................................... 103
Testing Diodes...................................................................................................................................... 103
Testing for Continuity .......................................................................................................................... 103
iii
Measuring Capacitance ........................................................................................................................ 103
Multimeter Features ................................................................................................................ 103
Data Hold Mode ................................................................................................................................... 103
Making Relative Measurements ........................................................................................................... 104
Information Display ............................................................................................................................. 104
Auto or Manual Range ......................................................................................................................... 104
Multimeter Recorder............................................................................................................... 104
7. Communication with PC ............................................................................................. 107
Using USB Port ........................................................................................................................ 107
Using LAN Port ....................................................................................................................... 108
Connect directly ................................................................................................................................... 108
Connect through a router ...................................................................................................................... 109
Using Wi-Fi to Connect with PC (Optional) ......................................................................... 111
Connect with PC as Wi-Fi Access Point .............................................................................................. 111
Connect with PC as Wi-Fi Station........................................................................................................ 112
8. Communication wit h Andr oid Device via Wi-Fi (Optional) ...................................................... 115
How to Connect ....................................................................................................................... 115
Connect with APP as Wi-Fi Access Point ............................................................................................ 115
Connect with APP as Wi-Fi Station ..................................................................................................... 117
User Interface ....................................................................................................................................... 119
Gestures Control ................................................................................................................................... 121
9. Demonstration ............................................................................................................. 123
Example 1: Measurement a Simple Signal ............................................................................ 123
Example 2: Gain of a Amplifier in a Metering Circuit ........................................................ 124
Example 3: Capturing a Single Signal ................................................................................... 125
Example 4: Analyze the Details of a Signal ........................................................................... 126
Example 5: Application of X-Y Function .............................................................................. 128
Example 6: Video Signal Trigger ........................................................................................... 129
10. Tr oubleshooting ......................................................................................................... 131
1 1. Technical Specifications ............................................................................................ 132
Oscilloscope .............................................................................................................................. 132
Trigger .................................................................................................................................................. 137
Waveform Generator .............................................................................................................. 139
Multimeter ............................................................................................................................... 139
General Technical Specifications ........................................................................................... 140
12. Appendix .................................................................................................................... 141
Appendix A: Enclosure ........................................................................................................... 141
Appendix B: General Care and Cleaning ............................................................................. 141
iv
Appendix C: Battery Using Guide ......................................................................................... 142
NOTE :
This series of oscilloscopes of
fers a variety of models with many different features and specifications. This manual also describes these functions for models that may not match the features of your model. For detailed information on the built­in features of each model, see the technical specifications.

1.General Safety Requirements

1
. Satefy precautions
1
This product complies with the requirements of the following European Community
Directives: 2014/30/EU (Electromagnetic Compatibility) and 2014/35/EU (Low Voltage) as amended by 2004/22/EG (CE-Marking). Overvoltage category II; pollution degree 2.
To ensure safe operation of the equipment and eliminate the danger of serious injury due
to short-circuits (arcing), the following safety precautions must be observed. Damages resulting from failure to observe these safety precautions are exempt from any legal claims whatever.
Do not use this instrument for high-energy industrial installation measurement. Do not place the equipment on damp or wet surfaces. Do not exceed the maximum permissible input ratings (danger of serious injury and/or
destruction of the equipment).
The meter is designed to withstand the stated max voltages. If it is not possible to exclude
without that impulses, transients, disturbance or for other reasons, these voltages are exceeded a suitable presale (10:1) must be used.
Disconnect test leads or probe from the measuring circuit before switching modes or
functions.
Check test leads and probes for faulty insulation or bare wires before connection to the
equipment.
To avoid electric shock, do not operate this product in wet or damp conditions. Conduct measuring works only in dry clothing and rubber shoes, i. e. on isolating mats. Never touch the tips of the test leads or probe. Comply with the warning labels and other info on the equipment. The measurement instrument is not to be to operated unattended. Do not subject the equipment to direct sunlight or extreme temperatures, humidity or
dampness.
Do not subject the equipment to shocks or strong vibrations. Do not operate the equipment near strong magnetic fields (motors, transformers etc.). Keep hot soldering irons or guns away from the equipment. Allow the equipment to stabilize at room temperature before taking up measurement
(important for exact measurements).
Do not input values over the maximum range of each measurement to avoid damages of
the instrument.
Periodically wipe the cabinet with a damp cloth and mid detergent. Do not use abrasives
solvents.
The meter is suitable for indoor use only Warning: To avoid fire or electrical shock, when the oscilloscope input signal connected is more
than 42V peak (30Vrms) or on circuits of more than 4800VA, please take note of below items:
-Only use accessory insulated voltage probes and test lead.­Check the accessories such as probe before use and replace it if there are any damages.­Remove probes, test leads and other accessories immediately after use.
-Remove USB cable which connects oscilloscope and computer.
-Do not apply input voltages above the rating of the instrument because the probe tip
voltage will directly transmit to the oscilloscope. Use with caution when the probe is set as 1:1.
Do not use exposed metal BNC or banana plug connectors. Do not insert metal objects into connectors.

2.Safety Terms and Symbols

2
2. Safety Terms and Symbols
Safety Terms
Terms in this manual. The following terms may appear in this manual:
Warning: Warning indicates the conditions or practices that could result in
injury or loss of life.
Caution: Caution indicates the conditions or practices that could result in
damage to this product or other property.
Terms on the product. The following terms may appear on this product:
Danger: It indicates an injury or hazard may immediately happen.
Warning: It indicates an injury or hazard may be accessible potentially.
Caution: It indicates a potential damage to the instrument or other property might occur.
Safety Symbols
Symbols on the product. The following symbol may appear on the product:
Hazardous Voltage Refer to Manual
Protective Earth Terminal Chassis Ground
Test Ground
To avoid body damage and prevent product and connected equipment damage, carefully read the following safety information before using the test tool. This product can only be used in the specified applications.
Warning:
The four channels of the oscilloscope are not electrically isolated. The channels should adopt a common ground during measuring. To prevent short circuits, the 2 probe grounds must not be connected to 2 different non-isolated DC levels.
2.Safety Terms and Symbols
3
Ground Clip
Signal Input
Oscilloscope
Electrical Outlet
Probe
Power Cord
Ground Clip
Signal Input
Oscilloscope
(Battery-power)
PC Electrical OutletProbe
USB/VGA/COM/
LAN Cable
Warning:
To avoid fire or electrical shock connected
42V peak (30Vrms) or on circuits of more than
4800VA
Do not apply input voltages above the rating of the instrument
tage will directly transmit to the
The diagram of the oscilloscope ground wire connection:
The diagram of the ground wire connection when the battery-powered oscilloscope is connected to the AC-powered PC through the ports:
It is not allowed to measure AC power when the oscilloscope is AC powered, or when the battery-powered oscilloscope is connected to the AC-powered PC through the ports.
is more than
, please take note of below items:
Only use accessory insulated voltage probes and test lead.
Check the accessories such as probe before use and replace it if
there are any damages.
Remove probes, test leads and other accessories immediately after
use.
Remove USB cable which connects oscilloscope and computer.
because the probe tip vol oscilloscope. Use with caution when the probe is set as 1:1.
Do not use exposed metal BNC or banana plug connectors.
Do not insert metal objects into connectors.
, when the oscilloscope input signal
3.Quick User Guidebook
4

3. Quick User Guidebook

This chapter deals with the following topics mainly:
Introduction to the structure of the oscilloscope
Introduction to the user interface
How to implement the general inspection
How to implement the function inspection
How to make a probe compensation
How to set the probe attenuation coefficient
How to use the probe safely
How to implement an self-calibration
Introduction to the vertical system
Introduction to the horizontal system
Introduction to the trigger system
Touchscreen Controls (Touchscreen is optional)
Magnifier Function (Only for specific models with touchscreen)
3.Quick User Guidebook
5
10
3
5
4
9
8
1
6
2
7

Introduction to the Structure of the Oscilloscope

This chapter makes a simple description of the operation and function of the front panel of the oscilloscope, enabling you to be familiar with the use of the oscilloscope in the shortest time.

Front Panel

The front panel has knobs and function buttons. The 5 buttons in the column on the right side of the display screen or in the row under the display screen are menu selection buttons, through which, you can set the different options for the current menu. The other buttons are function buttons, through which, you can enter different function menus or obtain a specific function application directly.
1. Display area
2. Select the right menu item
3. Control (button and knob) area
4. Probe Compensation: Measurement signal (3.3V/1kHz) output.
5. Input connectors of four channels
6. Remove the left and right menu
7. Select the bottom menu item
8. Copy button: You can save the waveform by just pressing this button in any user
Figure 3-1 Front panel
3.Quick User Guidebook
6
3
10
9
8
7
12
11
1 2
6
4 5
interface.
9. USB Host port: It is used to transfer data when external USB equipment connects to the oscilloscope regarded as "host device". For example: Saving the waveform to USB flash disk needs to use this port.
10. Power on/off

Rear Panel

Figure 3-2 Rear Panel
1. Handle
2. Air vents
3. Input terminals of multimeter (optional)
4. AC power input jack
5. Fuse
6. Foot stool: Adjust the tilt angle of the oscilloscope.
7. VGA port: To connect the oscilloscope with a monitor or a projector as VGA output
(optional).
8. LAN port: the network port which can be used to connect with PC.
9. USB Device port: It is used to transfer data when external USB equipment connects to
the oscilloscope regarded as "slave device". For example: to use this port when connect PC to the oscilloscope by USB.
10. Lock Hole: You can lock the oscilloscope to a fixed location using the security lock
(please buy it yourself) to secure the oscilloscope.
11. Trig Out(P/F) port: Trigger signal output & Pass/Fail output, also can be used as the
port of CH2 Output of dual-channel waveform generator (optional). The output type can be set on the menu (Utility menuOutputOutput).
12. Out 1 port: Output (single-channel) or CH1 Output (dual-channel) of waveform
generator (optional).
Control Area 2 CH
3.Quick User Guidebook
Figure 3-4 Control Area Overview
1. Function button area: Total 11 buttons
2. Waveform generator controls (depending on model)
or
DAQ: Data Acquire, P/F: Pass/Fail, W.REC: Waveform Record
3. Trigger control area with 2 buttons and 1 knob.
The Trigger Level knob is to adjust trigger voltage. Other 2 buttons refer to trigger system setting.
4. Horizontal control area with 1 button and 2 knobs.
"HOR" button refer to horizontal system setting menu, "Horizontal Position" knob control trigger position, " Horizontal Scale" control time base.
5. Vertical control area with 3 buttons and 4 knobs.
"CH1" and "CH2 " correspond to setting menu in CH1 and CH2, "Math" button refer to math menu, the math menu consists of six kinds of operations, including CH1-CH2, CH2-CH1, CH1+CH2, CH1*CH2, CH1/CH2 and FFT. Two "Vertical Position" knob control the vertical position of CH1/CH2, and two "Scale" knob control voltage scale of CH1, CH2.
6. Default: Call out the factory settings.
7. Print
8. Counter: Turn on/off the Counter.
9. DMM (optional) or Snapshot
10. Direction key: Move the cursor of the focused parameter.
11.M knob(Multipurpose knob): when a symbol appears in the menu, it indicates
you can turn the M knob to select the menu or set the value. You can push it to close
the menu on the left and right.
7
7

Control Area 4 CH

3.Quick User Guidebook
Figure 3-3 Control Area Overview
1. Function button area: Total 11 buttons
2. Waveform generator controls (optional)
or
DAQ: Multimeter Recorder (see "Multimeter Recorder" on P104) P/F: Pass/Fail (see "Pass/Fail" on P76) W.R EC : Waveform Record (see "How to Record/Playback Waveforms" on P66)
3. Trigger control area with 2 buttons and 1 knob.
The Trigger Level knob is to adjust trigger voltage. Other 2 buttons refer to trigger system setting.
4. Horizontal control area with 1 button and 2 knobs.
"HOR" button refer to horizontal system setting menu, "Horizontal Position" knob control trigger position, "Horizontal Scale" control time base.
5. Vertical control area with 5 buttons and 2 knobs.
CH1 - CH4 buttons correspond to setting menu in CH1 - CH4. "Math" button provides access to math waveform functions (+, -, ×, /, FFT, user function, digital filter). The "Vertical Position" knob control the vertical position of current channel, and the "Scale" knob control voltage scale of current channel.
6. Default: Call out the factory settings.
Print: Print an image of what appears on the oscilloscope screen.
Decode(optional): Turn on/off Decode function.
DMM (Multimeter, optional) or Snap (Shortcut button for measurement snapshot)
7. Direction key: Move the cursor of the focused parameter.
8. M knob (Multipurpose knob): when a symbol appears on the menu, it indicates
you can turn the M knob to select the menu or set the value. You can push it to close
3. Quick User Guidebook
8
the menu on the left and right.

User Interface Introduction

Figure 3-4 I
1. Waveform Display Area.
2. Run/Stop (touchable on touchscreen) (see "How to Use Executive Buttons" on
P89)
3. The state of trigger, including:
Auto: Automatic mode and acquire waveform without triggering.
Trig: Trigger detected and acquire waveform.
Ready: Pre-triggered data captured and ready for a trigger.
Scan: Capture and display the waveform continuously.
Stop: Data acquisition stopped.
4. Click to auto set (only for touchscreen).
5. Turn on/off the magnifier function (Only for specific models with touchscreen).
6. The two blue dotted lines indicates the vertical position of cursor measurement.
7. The pointer indicates the trigger position in the record length.
8. The T pointer indicates the horizontal position for the trigger.
9. It shows present triggering value and displays the site of present window in
internal memory.
llustrative Drawing of Display Interfaces
3.Quick User Guidebook
9
10. Touchable icon is to enable ( ) or disable ( ) the touchscreen controls
(only for touchscreen).
11. It shows setting time (see "Config" on P74).
12. WI-FI is activated.
13. It indicates that there is a USB disk connecting with the oscilloscope.
14. Indicating battery power status (see "Display" on P75).
15. Multimeter window.
16. The waveform of CH1.
17. The pointer shows the trigger level position of the source in trigger menu.
18. The two blue dotted lines indicate the horizontal position of cursor measurement.
19. The waveform of CH2.
20. The waveform of CH3.
21. The waveform of CH4.
22. The frequency of the trigger signal.
23. The icon shows the selected trigger type, e.g. represents triggering on the
rising edge for an Edge trigger. The reading shows the trigger level value of the corresponding channel.
24. Click to show/hide the touchable shortcut menu (only for touchscreen).
25. It indicates the measured type and value of the corresponding channel. "T" means
period, "F" means frequency, "V" means the average value, "Vp" the peak-peak value, "Vr" the root-mean-square value, "Ma" the maximum amplitude value, "Mi" the minimum amplitude value, "Vt" the Voltage value of the waveform's flat top value, "Vb" the Voltage value of the waveform's flat base, "Va" the amplitude value, "Os" the overshoot value, "Ps" the Preshoot value, "RT" the rise time value, "FT" the fall time value, "PW" the +width value, "NW" the -Width value, "+D" the +Duty value, "-D" the -Duty value, "PD" the Delay A->B value, "ND" the Delay A->B value, "TR" the Cycle RMS, "CR" the Cursor RMS, "WP" the Screen Duty, "RP" the Phase A->B , "FP" the Phase A->B , "+PC" the +Pulse count, "-PC" the - Pulse count, "+E" the Rise edge count, "-E" the Fall edge count, "AR" the Area, "CA" the Cycle area.
26. The readings show the record length.
27. The readings show current sample rate.
28. The readings indicate the corresponding Voltage Division of the channels.
"BW" indicates bandwidth limit.
The icon shows the coupling mode of the channel.
"—" indicates direct current coupling
"" indicates AC coupling
" " indicates GND coupling
29. The reading shows the setting of main time base.
30. The green pointer indicates the grounding datum point (zero point position) of the
waveform of the CH1 channel.
31. The orange pointer indicates the grounding datum point (zero point position) of
3.Quick User Guidebook
10
the waveform of the CH1 channel.
32. It is cursor measure window, showing the absolute values and the readings of the
cursors.
33. The blue pointer indicates the grounding datum point (zero point position) of the
waveform of the CH1 channel.
34. The yellow pointer indicates the grounding datum point (zero point position) of
the waveform of the CH1 channel.

How to Implement the General Inspection

After you get a new oscilloscope, it is recommended that you should make a check on the instrument according to the following steps:
1. Check whether there is any damage caused by transportation.
If it is found that the packaging carton or the foamed plastic protection cushion has
suffered serious damage, do not throw it away first till the complete device and its
accessories succeed in the electrical and mechanical property tests.
2. Check the Accessories
The supplied accessories have been already described in the "Appendix A: Enclosure" of
this Manual. You can check whether there is any loss of accessories with reference to
this description. If it is found that there is any accessory lost or damaged, please get in
touch with the distributor of PeakTech responsible for this service or the PeakTech
local offices.
3. Check the Complete Instrument
If it is found that there is damage to the appearance of the instrument, or the instrument can not work normally, or fails in the performance test, please get in touch with the PeakTech distributor responsible for this business or the PeakTech local offices. If there is damage to the instrument caused by the transportation, please keep the package. With the transportation department or the PeakTech distributor responsible for this business informed about it, a repairing or replacement of the instrument will be arranged by the PeakTech.

How to Implement the Function Inspection

Make a fast function check to verify the normal operation of the instrument, according to the following steps:
1. Connect the power cord to a power source. Long press the
bottom left of the instrument.
The instrument carries out all self-check items and shows the Boot Logo. Push the
Utility button, select Function in the bottom menu. Select Adjust in the left menu,
button on the
3.Quick User Guidebook
11
select Default in the bottom menu. The default attenuation coefficient set value of the probe on the menu is 10X.
2. Set the Switch in the Oscilloscope Probe as 10X and Connect the Oscilloscope
with CH1 Channel.
Align the slot in the probe with the plug in the CH1 connector BNC, and then tighten the probe with rotating it to the right side.
Connect the probe tip and the ground clamp to the connector of the probe compensator.
3. Push the Autoset Button on the front panel.
The square wave of 1 KHz frequency and 3.3V peak-peak value will be displayed in
several seconds (see Figure 3-5).
Figure 3-5 Auto set
Check CH2, CH3 and CH4 by repeating Step 2 and Step 3.

How to Implement the Probe Compensation

When connect the probe with any input channel for the first time, make this adjustment to match the probe with the input channel. The probe which is not compensated or presents a compensation deviation will result in the measuring error or mistake. For adjusting the probe compensation, please carry out the following steps:
1. Set the attenuation coefficient of the probe on the menu as 10X and that of the
switch in the probe as 10X (see "How to Set the Probe Attenuation Coefficient"
on P12), and connect the probe with the CH1 channel. If a probe hook tip is used, ensure that it keeps in close touch with the probe. Connect the probe tip with the signal connector of the probe compensator and connect the reference wire clamp with the ground wire connector of the probe connector, and then push the
3.Quick User Guidebook
12
Caution:
the probe attenuation coefficient in the
Autoset button on the front panel.
2. Check the displayed waveforms and regulate the probe till a correct
compensation is achieved (see Figure 3-6 and Figure 3-7).
Overcompensated Compensated correctly Under compensated
Figure 3-6 Displayed Waveforms of the Probe Compensation
3. Repeat the steps mentioned if needed.
Figure 3-7 Adjust Probe

How to Set the Probe Attenuation Coefficient

The probe has several attenuation coefficients, which will influence the vertical scale factor of the oscilloscope.
To change or check the probe attenuation coefficient on the menu of oscilloscope:
(1) Push the function menu button of the used channels (CH1 - CH4 button).
(2) Select Probe in the bottom menu; select Attenu in the right menu, turn the M
knob to select the proper value corresponding to the probe.
This setting will be valid all the time before it is changed again.
The default attenuation coefficient of the probe on the instrument is preset to 10X. Make sure that the set value of the attenuation switch in the probe is the same as the menu selection of oscilloscope.
The set values of the probe switch are 1X and 10X (see Figure 3-8).
3.Quick User Guidebook
13
Caution:
the attenuation switch is set to 1X, the probe will limit the bandwidth
connect the probe to the
Figure 3-8 Attenuation Switch
When of the oscilloscope in 5MHz. To use the full bandwidth of the oscilloscope, the switch must be set to 10X.
Identify the Probe Attenuation Coefficient Automatically
The oscilloscope can identify the probe attenuation coefficient of the 100:1 (impedance 5K±20%) or 10:1 (impedance 10K±20%) probe with the identifying pin. When you attach the probe, the oscilloscope set the attenuation automatically on the oscilloscope vertical menu for the channel to match the probe.
For example, if you attach a 10:1 probe with the identifying pin, the screen will prompt "The probe attenuation factor is X10", and set the attenuation to 10X automatically on the oscilloscope vertical menu for the channel.

How to Use the Probe Safely

The safety guard ring around the probe body protects your finger against any electric
shock, shown as Figure 3-9.
Figure 3-9 Finger Guard
Warning:
To avoid electric shock, always keep your finger behind the safety guard ring of the probe during the operation.
To protect you from suffering from the electric shock, do not touch any metal part of the probe tip when it is connected to the power supply.
Before making any measurements, always instrument and connect the ground terminal to the earth.
3.Quick User Guidebook
14

How to Implement Self-calibration

The self-calibration application can make the oscilloscope reach the optimum condition rapidly to obtain the most accurate measurement value. You can carry out this application program at any time. This program must be executed whenever the change of ambient temperature is 5 or over.
Before performing a self-calibration, disconnect all probes or wires from the input connector. Push the Utility button, select Function in the bottom menu, select Adjust. in the left menu, select Self Cal in the bottom menu; run the program after everything is ready.

Introduction to the Vertical System

As shown in Figure 3-10, there are a few of buttons and knobs in Vertical Controls.
The 4 channels are marked by different colors which are also used to mark both the corresponding waveforms on the screen and the channel input connectors. Press one of the channel buttons to open the corresponding channel menu, and press again to turn off the channel. Press the Math button to display the math menu in the bottom. The pink M waveform appears on the screen. Press again to turn off the math waveform.
The 4 channels use the same Vertical Position and Vertical Scale knobs. If you want to set the vertical scale and vertical position of a channel, please press CH1, CH2, CH3 or CH4 at first to select the desired channel. Then turn the Vertical Position and Vertical Scale knobs to set the values.
Figure 3-10 Vertical Control Zone
The following practices will gradually direct you to be familiar with the using of the vertical setting.
1. Press CH1, CH2, CH3 or CH4 to select the desired channel.
2. Use the Vertical Position knob to show the selected channel waveform in the center of the waveform window. The Vertical Position knob functions the regulating of the
3.Quick User Guidebook
15
vertical display position of the selected channel waveform. Thus, when the Vertical Position knob is rotated, the pointer of the earth datum point of the selected channel is directed to move up and down following the waveform, and the position message at the center of the screen would change accordingly.
Measuring Skill
If the channel is under the DC coupling mode, you can rapidly measure the DC component of the signal through the observation of the difference between the wave form and the signal ground.
If the channel is under the AC mode, the DC component would be filtered out. This mode helps you display the AC component of the signal with a higher sensitivity.
Vertical offset back to 0 shortcut key
Turn the Vertical Position knob to change the vertical display position of the selected channel, and push the position knob to set the vertical display position back to 0 as a shortcut key, this is especially helpful when the trace position is far out of the screen and want it to get back to the screen center immediately.
3. Change the Vertical Setting and Observe the Consequent State Information Change.
With the information displayed in the status bar at the bottom of the waveform window, you can determine any changes in the channel vertical scale factor.
Turn the Vertical Scale knob and change the "Vertical Scale Factor (Voltage Division)" of the selected channel, it can be found that the scale factor of the selected channel in the status bar has been changed accordingly.

Introduction to the Horizontal System

Shown as Figure 3-11, there are a button and two knobs in the Horizontal Controls.
The following practices will gradually direct you to be familiar with the setting of horizontal time base.
Figure 3-11 Horizontal Control Zone
3.Quick User Guidebook
16
1. Turn the Horizontal Scale knob to change the horizontal time base setting and
observe the consequent status information change. Turn the Horizontal Scale knob to change the horizontal time base, and it can be found that the Horizontal Time Base displayed in the status bar changes accordingly.
2. Use the Horizontal Position knob to adjust the horizontal position of the signal
in the waveform window. The Horizontal Position knob is used to control the triggering displacement of the signal or for other special applications. If it is applied to triggering the displacement, it can be observed that the waveform moves horizontally with the knob when you rotate the Horizontal Position knob.
Triggering displacement back to 0 shortcut key
Turn the Horizontal Position knob to change the horizontal position of channel and push the Horizontal Position knob to set the triggering displacement back to 0 as a shortcut key.
3. Push the Horizontal HOR button to switch between the normal mode and the
wave zoom mode.

Introduction to the Trigger System

As shown in Figure 3-12, there are one knob and three buttons make up Trigger
Controls. The following practices will direct you to be familiar with the setting of the
trigger system gradually.
Figure 3-12 Trigger Control Zone
1. Push the Trigger Menu button and call out the trigger menu. With the operations
of the menu selection buttons, the trigger setting can be changed.
2. Use the Trigger Level knob to change the trigger level setting.
By turning the Trigger Level knob, the trigger indicator in the screen will move up and down. With the movement of the trigger indicator, it can be observed that the trigger level value displayed in the screen changes accordingly.
Note: Turning the Trigger Level knob can change trigger level value and it is also the hotkey to set trigger level as the vertical mid point values of the amplitude of the trigger signal.
3. Push the Force button to force a trigger signal, which is mainly applied to the
3. Quick User Guidebook
17
Press repeatedly to
switch the options
Introduction to the Touchscreen Controls (Depending on
Model) If the LCD is touchscreen, you can control the oscilloscope by different gestures.
The touchable icon at the top right of the screen is used to enable ( (
) the touchscreen controls. The instruction of touchscreen controls is as below. You can also use the buttons / knobs enclosed in brackets to do the same thing.
Operate the Menu through Touchscreen
Select a menu item: Touch the menu items in the bottom menu, or in the right menu,
or in the left menu.
Switch menu items: If there are options that can be switched in the menu, you can
repeatedly touch the area of the menu item to switch, or push the corresponding button to switch. See figure below:
) or disable
Scroll the list: If there is a scroll bar in the left menu or in the file system window,
you can swipe up and down to scroll the list.
Touchable menu pane: Click the
pane will be shown. Clicking the menu item is same as pushing the corresponding button.
icon on the left top of the display area, a menu
Gestures in Normal Mode
Select a channel (CH1 or CH2 button): Touch the pointer on the left side of
3.Quick User Guidebook
18
Touch to select the channel
Selected
Unselected
Control the vertical position
of the selected channel
Control the trigger level of
the source in trigger menu
corresponding channel to make it in selected state.
Set the vertical position of the selected channel (Vertical Position knob): Swipe up
or down on the screen.
Set the trigger level of the source in trigger menu (Trigger Level knob): Swipe up or
Set the horizontal position (Horizontal Position knob): Swipe left or right on the
display area.
down on the area nearby the trigger pointer on the right.
3.Quick User Guidebook
19
Control the horizontal position
Double Zoom and Single Zoom
In touchable menu pane, if DoubleZoom is selected, in the display area, pinch and spread horizontally to change the time base; pinch and spread vertically to change the voltage division of current channel.
3.Quick User Guidebook
20
Pinch and spread vertically
to change the voltage
division of current channel
Pinch and spread
horizontally to change
the time base
Click to increase
the voltage
division of CH1
Click to decrease
the voltage
division of CH1
Click to
decrease
the time base
Click to
increase
the time base
Click to increase
the voltage
division of CH2
Click to decrease
the voltage
division of CH2
In touchable menu pane, if SingleZoom is selected, click anywhere on the Display area to show the touching control panel.
Set the voltage division (Vertical Scale knob): Click on the left upper area to increase
3.Quick User Guidebook
21
Control the vertical position
of the selected channel
Main Window
Zoom Window
Control the trigger level of the source in trigger menu
the voltage division of CH1; click on the left lower area to decrease the voltage division of CH1. Click on the right upper area to increase the voltage division of CH2; click on the right lower area to decrease the voltage division of CH2.
Set the horizontal time base (Horizontal Scale knob): Click on the left area to
increase the voltage division; click on the right area to decrease the voltage division.
Gestures in Wave Zoom Mode
Push the Horizontal HOR button to enter into zoom mode. The top half of the display shows the Main window and the bottom half displays the Zoom window. The Zoom window is a magnified portion of the Main window.
3.Quick User Guidebook
22
Control the horizontal position
of Zoom Window
Pinch and spread vertically
to change the voltage
division of current channel
Pinch and spread
horizontally to change
the time base
of Zoom Window
(Double Zoom)
3.Quick User Guidebook
23
Click to increase
the voltage division of CH1
Click to decrease
the voltage division of CH
1
Click to decrease
the time base of
Zoom Window
Click to increase
the time base of
Zoom Window
Click to increase
the voltage division of CH2
Click to decrease
the voltage division of CH2
Press nearby any cursor line until both lines are selected, drag them up and down simultaneously.
Switch horizontal or vertical lines
If vertical lines are selected, drag left and right.
Press nearby a cursor line to make it selected, drag it up and down.
Selected
Unselected
Other Operations Using Touchscreen
(Single Zoom)
Measure with Cursors:
Run/Stop: Double tap on the display area, or click the or on the left top of
the display area to run or stop the waveform sampling.
Touch keyboard: Click to input.
Adjust value in the menu item: .
3.Quick User Guidebook
24
Click to increase the
value of cursor position
Move the cursor
Click to decrease the
value of cursor position
Loading...
+ 109 hidden pages