7 8
1 Humidity/Temperature Probe: Humidity and
Temperature Sensor inside.
2 LCD Display: 4 ½ digit, LCD, dual display. Displays ⁰F,
⁰C, %RH, dew point, wet bulb, low battery (BAT),
MAX/MIN and HOLD.
3 MAX/MIN Button: Press MAX/MIN button to enter
MAX/MIN recording mode. Select desired range before
entering MAX/MIN mode to ensure that the reading will
not exceed the testing range. Press once to select MIN.
Press again to select MAX. Press and hold the Button
for 3 seconds to turn off MAX, MIN recording.
4 ⁰F, ⁰C, Dew Point,Wbt Select Button: Press button to
toggle between measurement displays: ⁰F, ⁰C, Dew
Point and Wet Bulb measurements.
5 Power ON/OFF Button: Turn the meter power ON/OFF.
6 HOLD /Backlight Button: Press Hold/Backlight button
to enter Hold function. The displayed reading will be
held when Data Hold button is pressed.
Press Hold/Backlight button for 3 seconds to turn on or
off backlight.
7 Battery Compartment: Battery location.
5. HUMIDITY AND AIR TEMPERATURE
MEASUREMENTS
1. Hold the probe in the area to be tested & allow
adequate time for readings to stabilize (< 1 minute).
2. Press ⁰F, ⁰C, dew point, WBT select Button to select ⁰F,
⁰C, dew point Wet Bulb Measurement.
3. Read the Relative Humidity and Air Temperature, dew
point temperature wet bulb temperature.
NOTE:
Do not immerse the probe in liquid; it is intended for use in air
only.
6. BATTERY REPLACEMENT
If the “ ” symbol appears on the LCD display, it indicates
that the battery should be replaced. Open the battery case
and replace the exhausted battery with new battery. (1 x 9V
battery NEDA 1604, 6F22 or equivalent).
Warning:
Do not leave sensor exposed to direct sunlight. Direct,
prolonged exposure to solar radiation can damage sensor